Rohde&Schwarz TS8977 System Manual

TS8977 System Manual
Test and Measurement Division
Author: R&S 1SP1 - W.Tiwald Revision: 1.1 Date: 2008-10-27 File: TS8977-System-Manual.doc
Printed in the Federal Republic of Germany
© Rohde & Schwarz GmbH & Co. KG 2007
Page: 1
TS8977 System Manual Introduction
Contents
Page
1 Introduction .......................................................................................................... 3
2 TS8977 Hardware ................................................................................................. 4
2.1 Overview .................................................................................................................................... 4
2.2 Description of Instruments ...................................................................................................... 7
2.2.1 System Control Instruments................................................................................................. 7
2.2.2 Signal Generating Instruments ............................................................................................ 7
2.2.3 Measurement Instruments ................................................................................................... 8
2.2.4 Switch Units ......................................................................................................................... 9
2.2.5 Other Instruments .............................................................................................................. 10
2.3 Interfaces and Connectors..................................................................................................... 11
2.3.1 Interfaces ........................................................................................................................... 11
2.3.2 Connectors......................................................................................................................... 11
2.4 Operating Conditions ............................................................................................................. 13
2.5 Calibration Requirements ...................................................................................................... 13
3 Starting Operation.............................................................................................. 14
3.1 Preparing for Use .................................................................................................................... 14
3.1.1 Cabling............................................................................................................................... 14
3.1.2 Devices .............................................................................................................................. 14
3.1.3 System Configuration......................................................................................................... 14
3.2 System Switch On/Off and Start/Shutdown ......................................................................... 15
3.2.1 Instrument Switch On/Off................................................................................................... 15
3.2.2 System Start and Shut Down............................................................................................. 15
3.2.3 Start and Shut Down of the Signalling Unit........................................................................ 15
4 WiMAX Applications........................................................................................... 17
4.1 Test Case List.......................................................................................................................... 17
4.1.1 Receiver Tests ................................................................................................................... 17
4.1.2 Transmitter Tests............................................................................................................... 19
4.2 Transmit/Receiver Tests (TRX).............................................................................................. 20
4.2.1 Test Description ................................................................................................................. 20
5 Glossary .............................................................................................................. 22
Figures
.....................................................................................................................................................................
Fig. 2-1 Block diagram of the SSCU.......................................................................................................... 5
Fig. 2-2 Rack layout of the TS8977 test system ........................................................................................ 6
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 2 Revision: 1.0
TS8977 System Manual Introduction
1 Introduction
This document contains all system specific information that is necessary to operate and maintain the TS8977 Test System.
The test system supports numerous RF test cases, based on test methods defined in the EN 302-544 Part 2 and EN 302-623: “Harmonized EN for TDD User Equipment covering essential requirements of article 3.2 of the R&TTE Directive”. These test methods can be executed with defined set of parameters, being compliant to the relevant sections of the EN 302-544 respective EN 302 623 test specification. However, the free selection of parameters in order to define customer specific R&D test scenarios is supported as well.
The test system is designed to perform RF tests for WiMAX subscriber (mobile) stations. The system support following WiMAX Profiles: MP05 (2,5 - 2,69 GHz, 5MHz and 10MHz bandwidth), MP09 (3,4 – 3,6 GHz, 5MHz bandwidth) MP10 (3,4 – 3,6 GHz, 7MHz bandwidth) MP12 (3,4 – 3,6 GHz, 10MHz bandwidth)
To simulate interfering signals, R&S SMU vector signal generators (VSG) are integrated in the TS8977 Test System. The RF output of the DUT is analysed by means of the built-in vector signal analyser (VSA).
The in-band signal switching and conditioning unit SSCU is required to connect the DUT to the RF devices within the test system. It contains all RF components which are not WiMAX specific.A separate Advanced Signal Conditioning Unit (ASCU) is used to filter, combine, split, attenuate or amplify band specific signals which are WiMAX specific.
The system software to operate and maintain the system, namely the RS-PASS Software, is described in the RS-PASS Operation Manual.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 3 Revision: 1.0
TS8977 System Manual TS8977 Hardware
2 TS8977 Hardware
2.1 Overview
The TS8977 hardware consists of the following components:
No Type Device or Option Manufacturer
1 PSL3
2 FSU26
FSP-B10 External generator control for ESCI, ESPO
3 SMU 200A
SMU-B106 RF path A 100kHz to 6 GHz SMU-B11 Baseband generator (16 Msample) SMU-B13 Baseband main module SMU-B17 Analog baseband input SMU-B81 Rear panel connectors for 1st RF path SMU-K49 Digital Standard IEEE 802.16 SMU-K62 Additional white gaussian noise
4 SMR27
SMR-B11 Frequency Extension 0.1 – 1 GHz SMR-B15 RF attenuator 20 GHz
5 NRP-Z21
NRP-Z4 USB Adapter (passive) 6 NRP-Z55 7 SSCU-77 8 ASCU-77
--------- Available from August 2008 onwards --------
9 CMW270
CMW-B200A Signalling Unit universal
CMW-B270A WiMAX signalling extension module
CMW-B612A IEEE Interface Module
CMW-P752 CMW270 Mainframe 01
CMW-KS700 CMW license – signalling (base station emulation)
System controller 17’’ flat screen monitor Mouse Keyboard Spectrum analyser
RF generator inband (1)
RF generator out of band (2)
Power Sensor 10MHz to 18GHz
Thermal Power Sensor 0-40GHz Inband Signal Switching and Conditioning Unit Advanced Signal Switching and Conditioning Unit
WiMAX Communication Tester
b b b b b  b b b b b b b b b b b b b b b b
b b b b b b
R&S
R&S
R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S R&S
R&S R&S R&S R&S R&S R&S
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 4 Revision: 1.0
TS8977 System Manual TS8977 Hardware
Fig. 2-1 Block diagram of the SSCU
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 5 Revision: 1.0
TS8977 System Manual TS8977 Hardware
TS 8977 Conformance Test System
Fig. 2-2 Rack layout of the TS8977 test system
In signalling mode a connection between the protocol and signalling unit and the DUT has to be established.
For transceiver and transmitter tests the signal of the DUT is directed not only to the protocol and signalling unit, but also to the spectrum analyser FSU. This transmission path is the so-called measurement or analyser path.
For receiver tests, interferer signals are generated by the SMU and SMR signal generators. The interferer signals are combined with the wanted signals of the protocol and signalling unit. This transmission path is the so-called stimulus or generator path.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 6 Revision: 1.0
TS8977 System Manual TS8977 Hardware
2.2 Description of Instruments
2.2.1 System Control Instruments
2.2.1.1 System Controller
The system controller controls all devices within the TS8977. Most of the instruments are commanded via GPIB connection. The exceptions include the switching units SSCU and ASCU, which are connected via parallel interface.
The system controller hosts the RS-PASS software, including the system software with all required device drivers, the applications performing the tests as defined, the graphical user interface and several tools. Please refer to the RS-PASS operation manual for more details.
The system is being delivered with the Rohde & Schwarz PSL3, configured with 512 MB RAM, 2 hard disks and two Ethernet interfaces. The operating system is WINDOWS XP.
Before the system controller can be powered up the monitor, the mouse and the keyboard have to be connected to the TS8977 system panel. To switch “ON” the system, the rocker switch on the front panel has to be pushed behind the flap.
2.2.1.2 Ethernet HUB
The first Ethernet adapter of the system controller is connected to the Ethernet HUB. All TCP/IP devices installed on the system have to be connected to that HUB. However, the second adapter card of the PSL3 shall not be connected to the HUB and cannot be used for a connection to the customer's company network.
Be aware that for connections to the HUB cross link cables can not be used! Only for direct connections between instruments a cross link network cable is necessary.
The SMC Ethernet HUB is not equipped with a power switch. In order to apply power to the device, the power cord needs to be plugged in the AC source. A "power" LED will illuminate when main power is applied. Several activity and state LEDs indicate the status of the connections.
2.2.2 Signal Generating Instruments
2.2.2.1 Signalling Unit CMW270
In most test cases it is necessary to establish a running connection between the test system and the DUT before a RF performance test can be performed. As a consequence the test system has to simulate a base station. In the case of the simulation of a base station, the test system has to page the DUT, enabling it to access one of the simulated WiMAX cells and simulating the call establishment process. To achieve this, the corresponding protocol
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 7 Revision: 1.0
TS8977 System Manual TS8977 Hardware
functionalities, defined in the WiMAX specifications, have to be implemented by the use of a signalling device.
In advance of the execution of a test case, a call set-up has to be performed. The system controller instructs the signalling unit to perform the required call set-up via ethernet. After a successful call set-up (initial state), the measurement will be carried out.
In the case of receiver tests, the CMW270 is capable of measuring the PER (Packet Error Rate).
When doing transmitter tests, the CMW270 delivers sophisticated trigger signals. This is especially important for the analysis of signal waveforms performed by the spectrum analyser and triggered by the CMW270.
Wanted Signals
Usually the wanted signal is a standard WiMAX signal.
2.2.2.2 RF Signal Generators SMU and SMR
The RF generator SMU200 is employed to produce interfering signals with frequencies of up to 6 GHz. They are also used to stimulate several RFC and SFT measurements. To generate different kinds of signals, the SMU200s are also equipped with a modulation coder and a data generator.
There is an AC voltage selection feature implemented in the RF signal generator SMU200 that automatically sets itself according to the applied AC voltage. The SMU200 is switched on by first pressing the rear panel AC mains switch followed by the front panel power switch "ON/STBY". Switching off the system at the rear is only necessary, when the instrument is to be disconnected completely from the mains.
Interfering signals
The SMR27 generators can create interfering cw signals up to 27 GHz. The SMR27 is used as generator for the blocking test and the RF path calibration for the spurious emission tests.
2.2.3 Measurement Instruments
2.2.3.1 Spectrum Analyser FSU26
The spectrum analyser FSU26 measures the properties of the DUT´s signal during numerous tests, such as output power as a function of frequency or time, or spurious emissions and spurious response rejection.
Both devices are equipped with an AC voltage selection feature, which automatically sets itself according to the applied power. It has to be checked that the ”ON/STBY” switch, on the front at the bottom left hand corner, is in the correct position. Moreover, an AC mains supply ”ON/OFF” switch is situated on the rear of the device at the top right hand corner. Pressing this rocker switch will apply the AC mains to the instrument.
Measurement Modes of the FSU
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 8 Revision: 1.0
TS8977 System Manual TS8977 Hardware
Being dependent on the test purpose, the FSQ is operated in different measurement modes:
Peak hold: The maximum value occurring for a certain property during the measurement time is
taken.
Average: The measurement value is averaged over the measurement time.
Sample: To analyse the dependence of a certain property against time, each individual
measurement value is taken.
RMS: The RMS deviation from the average measurement value is taken.
Min/max: An upper and lower envelope is derived from the individual measurement values.
More than one measurement mode can be applied in a single test run.
2.2.3.2 Power Meter
The power meters NRP-Z21 and NRP-Z55 are used for the RFC and for SFT measurements. It is delivered with an USB adapter NRP-Z4. The NRP-Z21 and NRP-Z55 are the most accurate devices of the TS8977 system with the typical measurement uncertainty being less than 0.2 dB. The NRP-Z21 is supplied with power via USB.
2.2.4 Switch Units
2.2.4.1 Main Switching Unit (SSCU-77)
This SSCU-77 is used to interconnect all RF instruments within the system and connects the DUT to the system. This system is indispensable for running test cases. It includes components to split, combine, filter, attenuate or amplify RF signals which are not frequency band specific. The SSCU-77 is designed for a frequency range from 400 MHz to 6 GHz.
The SSCU-77 is powered by DC +5 V, +15 V and +28 V. These voltages are supplied by the switched mode power supply (CS-PSSU), located behind the cover plate at the top of the rack. In order to apply the AC mains to the SSCU, the rocker switch, situated on the right hand of the instrument, has to be depressed. All three yellow LED's, located at the SSCU front panel, should be illuminated.
2.2.4.2 Advanced Signal Switching and conditioning Unit (ASCU-77)
The ASCU-77 is an extension for the SSCU and contains all RF components which are frequency specific for WiMAX frequency bands like filters and amplifiers. The ASCU-77 is designed for a frequency range from 400 MHz to 26 GHz.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 9 Revision: 1.0
TS8977 System Manual TS8977 Hardware
2.2.5 Other Instruments
2.2.5.1 AC Power Distributor
The Line distributor (LDE) is used to provide the AC mains voltage to the devices in the test system. It is equipped with an EMI filter that prevents the system from high frequencies on the power line. The LDE has a switch enabling the user switching the system on and off (refer to section 3.2).
The LDE is mounted into the system on the rear side at about the same height as the system controller PSL 3.
2.2.5.2 DC Power Supply
The power supply NGPQ32 is used to provide the DUT with an adjustable amount of energy. It is set into operation with the power switch on the front panel. After being switched on, the digital display shows the last settings of the voltage and current of the outputs. When the system is switched on for the first time, the display will show 0 voltage and 0 current until it has been programmed. The outputs are "off" at this stage.
The NGPQ2 supports four different AC mains supply voltage levels 100 V, 120 V, 220 V and 240 V with system frequencies of 47 Hz to 63 Hz. The AC mains voltage range must be set to:
100 V for e.g. Japan
120 V for e.g. USA
230 V for e.g. central Europe (exceptions: 220 V e.g. Turkey, 240 V e.g. United Kingdom)
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 10 Revision: 1.0
TS8977 System Manual TS8977 Hardware
2.3 Interfaces and Connectors
2.3.1 Interfaces
In the table below are the hardware remote interface settings of the system instruments:
Device Resource Name / Remote Address Comments
Signalling Unit CMW270 tbd
Signal generator SMU GPIB::18 Signal generator SMR ASRL1
Spectrum analyser FSU GPIB::01 Power meter NRP USB::0x0aad::0x000c::_ _ _ _ _ _
SSCU LPT::1
DC Power supply NGMO GPIB::07
NRVS GPIB::20
The resource settings are checked and updated, when running the System Explorer.
The interfaces and resources are indicated by the parameter resource of the physical devices, listed in the system configuration file …RS PASS COMMON/ini/testsystem.ini. The name of the system configuration file can be chosen by the user.
2.3.2 Connectors
These are the HF connector settings of the system instruments:
Device SSCU-77 input SSCU-77 output Device input Device output
Signalling Unit CMW270 SIG RX SIG TX Signal generator SMU SG2 RF-A Signal generator SMR SG1 RF-50 Ohm Spectrum analyser FSU SA RF INP Power meter NRP-Z11 PM Power Probe SSCU LPT ASCU LPT
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 11 Revision: 1.0
TS8977 System Manual TS8977 Hardware
The connectors are defined in the system and must not be changed.
he settings are indicated by the parameters SSCU output connector, SSCU input connector,
T device output connector and device input connector of the physical devices listed in the system configuration file …RS PASS COMMON/ini/testsystem.ini.
Not all connectors offered with the system instruments are indicated in the system configuration files, as not all are relevant. The connectors of the SSCU are the following:
Front side SSCU-77
Connector Remark
SIGNAL ANALYSER RF INPUT Connection to FSU 'RF_INP' REFERENCE OUT RF output used for RFC MS/BS DUT TX TX1/RX1 TX2/RX2 SIG TX Connection to signalling unit CMW270 (receiver part) SIG RX Connection to signalling unit CMW270 (transmit part) SG1 Connection to SMR RF 50 Ohm
Connection to mobile station. Mobile Transmitter Combined connector Mobile TX/RX Combined connector MobileTX/RX
Rear side SSCU-77
Connector Remark
PM Connection to Power Meter NRP-Z21 PARALLEL PORT IN Connection to system controller ‘LPT 1’ for ISSCU control PARALLEL PORT OUT Not connected SG2 Connection to SMU RF-A ANALYSER BOX Input Output GENERATOR BOX Input Output
Connection from ASCU-77 Coonection to ASCU-77
Not connected Not connected
Front side ASCU-77
Connector Remark
none
Rear side ASCU-77
Connector Remark
ANALYSER BOX Input output GENRATOR BOX Input
Connection from SSCU-77 Connection to SSCU-77
Not connected
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 12 Revision: 1.0
TS8977 System Manual TS8977 Hardware
Connector Remark
Output Not connected
The connectors of the signal generating and measurement devices are described thoroughly in the corresponding manuals.
2.4 Operating Conditions
The optimum environmental conditions for the TS8970 system are the following.
Temperatures below the specified values are not critical.
Temperatures above the specified values should be avoided in order to prevent damages!
Operating temperature for the TS8977 : 23°C ± 3°C
Temperature range after RF calibration : ± 2°C
Temperature of the rack mounted components : 20°C to 40°C
Temperature of the SSCU-77 : 20°C to 40°C
Temperature of the ASCU-77 : 20°C to 40°C
Temperature of the components inside SSCU : 20°C to 45°C
Temperature of the components inside ASCU : 20°C to 45°C
Humidity : 20 to 80 %
2.5 Calibration Requirements
The system device calibration is recommended to be done within an interval of 1 year. It is independent from the instrument´s internal calibration or the RFC, described in the RS-PASS Operation Manual.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 13 Revision: 1.0
TS8977 System Manual Starting Operation
3 Starting Operation
3.1 Preparing for Use
3.1.1 Cabling
It is recommended to check the cabling of the system. Every original cable, used for the cabling of the system, is labelled with the name of the device to be connected to it.
It has to be taken care, that all cables are connected correctly!
The DUT cable has to be connected to the DUT Tx/Rx connector. It has to be made sure that the specific DUT cable is indicated as active in the system configuration *.ini file. The DUT cable configuration is done in the system editor of the TSCC and is described in detail in the RS-PASS Operation Manual.
Before running a WiMAX or another RFC application, the RFC calibration of the DUT cable has to be accomplished.
(The calibration of the DUT cable is described in the RS-PASS Operation Manual.
A rough overview over the system´s internal connections is given in section 2.3.2.
3.1.2 Devices
The device´s internal calibration can be updated by running the self test calibrate_all. The internal
calibration is independent to the RFC and is described in detail in the RS-PASS Operation Manual.
3.1.3 System Configuration
After the start of the system the configuration, especially the device resource (bus systems used to control the devices and device addresses) and connector settings, has to be checked. This is done by
running the System Explorer, thoroughly described in the RS-PASS Operation Manual.
When the system is started for the first time, i.e. after installation of the RS-PASS software, the System Explorer has to be executed before other applications can be carried out.
The device interfaces and connectors are described in section 2.3.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 14 Revision: 1.0
TS8977 System Manual Starting Operation
3.2 System Switch On/Off and Start/Shutdown
3.2.1 Instrument Switch On/Off
SYSTEM POWER ON/OFF
In order to start the system, the instruments have to be switched ON. Main power is applied by pushing the switch at the rear bottom of the system. With this switch all instruments can be switched ON and OFF .
Additionally, most instruments can be switched ON/OFF, or put into standby mode individually by pressing the relevant power button at the front panel:
STBY / ON to set the device into standby mode OFF / ON to switch the device off/on
Most of the devices have a main power switch at the rear panel to connect or disconnect it from the mains. For more details please refer to the instrument descriptions in section 2.2 .
If the main power connection at the rear of an instrument is switched off, then this device cannot be switched on with the SYSTEM POWER ON/OFF switch or with the ON/STBY and ON/OFF button at the front panel of the instrument.
3.2.2 System Start and Shut Down
After switching on the test system, the system controller automatically starts booting. After booting is finished, a login window is displayed and the username and password has to be entered. The default logins are:
System Manager: User name: ‘administrator’
Password : delivered without password
Account TS8977: User name: ‘ts8977’
Password : ‘ts8977’
Both accounts hold system manager rights, required e.g. for a system software update. To ensure, that users cannot damage the operating system, please make sure to create individual user accounts for daily work!
3.2.3 Start and Shut Down of the Signalling Unit
When the main switch "SYSTEM POWER ON/OFF" at the rear system panel is used to start the TS8977 test system, the signalling unit is automatically started as well. When switching off the system it also powers down the signalling unit.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 15 Revision: 1.0
TS8977 System Manual Starting Operation
he CMW270 has to be shut down regularly - like the WINDOWS main system - to
T avoid damages to the software!
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 16 Revision: 1.0
TS8977 System Manual WiMAX Applications
4 WiMAX Applications
4.1 Test Case List
This section lists the conformance test cases delivered by R&S and gives an overview which test methods are provided. All descriptions and tables in this document are composed according to the EN 302-504 part 2 and EN 302 623 test specifications. In both specification similar test cases are described, but unfortunately the test case numbering is different. For example, the spectrum emission mask test has in EN 302-544-2 the test case number 4.2.2 whereas in the EN 302 623 the number is
4.2.3. In order to use the same test method for the similar test cases in the RS PASS SW there is following mapping to achieve same number for same test cases:
RS PASS TC Number
TX Tests
1.1 Spectrum Emission Mask 4.2.2 4.2.3
1.2 TX Adjacent Channel Leakage
1.3 TX Spurious Emissions 4.2.4
1.4 Accuracy of max. Output Power 4.2.5 4.2.2
1.5 TX Power Control
1.6 Control & Monitoring Functions 4.2.7 4.2.9
RX Tests
2.1 RX Intermodulation Rejection 4.2.11 n.a.
2.2 RX Adjacent and Alternate
2.3 RX Spurious Emissions
2.4 RX Blocking
2.5 RX Spurious Response 4.2.12 n.a.
Test Case TC Number in
EN 302 544-2 EN 302 623
4.2.3 4.2.4
Power Ration
4.2.5
4.2.6 4.2.6
4.2.9 4.2.8
Channel Rejection
4.2.8 4.2.7
4.2.10 n.a
For details and up-to-date information on test cases, please refer to the latest version of the mentioned specifications and release notes of the RS-PASS Software.
As the test system devices are hardlock protected, the availability of the test cases may depend on individual registered options.
4.1.1 Receiver Tests
The following test cases are currently available:
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 17 Revision: 1.0
TS8977 System Manual WiMAX Applications
4.1.1.1 EN 302-544 part 2
Section of EN 302-504 part 2 Specification
4.2.8
4.2.9
Test Cases Testmethod
Receiver Spurious Emissions Receiver Adjacent Channel Selectivity
4.1.1.1.1 EN 302-544 part 2
Section of EN 302-544 part 2 Specification
4.2.8
4.2.9
Parameter files Remark
MS_2.03_norm_3A5_Rx_SpuriousEmissions.par MS_2.03_norm_3A10_Rx_SpuriousEmissions.par MS_2.02_norm_3A5_AdjacentChannelRejection.par MS_2.02_norm_3A10_AdjacentChannelRejection.par
WiMaxHarmonizedMsTrx.exe WiMaxHarmonizedMsReceiver. exe
4.1.1.2 EN 302- 623
Section of EN 302-623 Specification
4.2.7
4.2.8
4.1.1.2.1 EN 302-623
Section of EN 302-623 Specification
4.2.7
4.2.8
Test Cases Testmethod
Receiver Spurious Emissions Receiver Adjacent and Alternate Channel Rejection
Parameter files Remark
MS_2.03_norm_5AL_Rx_SpuriousEmissions.par MS_2.03_norm_5BL_Rx_SpuriousEmissions.par MS_2.03_norm_5CL_Rx_SpuriousEmissions.par MS_2.02_norm_5AL_AdjacentChannelRejection.par MS_2.02_norm_5BL_AdjacentChannelRejection.par MS_2.02_norm_5CL_AdjacentChannelRejection.par
WiMaxHarmonizedMsTrx.exe WiMaxHarmonizedMsReceiver. exe
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 18 Revision: 1.0
TS8977 System Manual WiMAX Applications
4.1.2 Transmitter Tests
The following test cases are currently available:
4.1.2.1 EN 302 544 part 2
Section of EN 302-544 part 2 Specification
4.2.2
4.2.3
4.2.4
4.2.5
4.2.6
4.2.7
Parameter files Testmethod
Spectrum Emission Mask TX Adjacent Channel Leakage Power Ratio
Transmit Spurious Emissions Accuracy of max. Output Power
Transmitter Power Control
Control and Monitoring Functions
WiMaxHarmonizedMsTrx.exe WiMaxHarmonizedMsModulatio nQuality.ex WiMaxHarmonizedMsTrx.exe WiMaxHarmonizedMsStaticPow er.exe WiMaxHarmonizedMsStaticPow er.exe WiMaxHarmonizedMsStaticPow er.exe
4.1.2.1.1
Parameter files Remark
4.2.2
4.2.3
4.2.4
4.2.5
4.2.6
4.2.7
MS_1.01_norm_3A5_SpectrumEmissionMask.par MS_1.01_norm_3A10_SpectrumEmissionMask.par MS_1.02_norm_3A5_ACLR.par MS_1.02_norm_3A10_ACLR.par MS_1.03_norm_3A5_Tx_SpuriousEmissions.par MS_1.03_norm_3A10_Tx_SpuriousEmissions.par MS_1.04_norm_3A5_MaxOutputPower.par MS_1.04_norm_3A10_MaxOutputPower.par MS_1.05_norm_3A5_TransmitterPowerControl.par MS_1.05_norm_3A10_TransmitterPowerControl.par MS_1.06_norm_3A5_ControlAndMonitoringFunctions.par MS_1.06_norm_3A10_ControlAndMonitoringFunctions.par
4.1.2.2 EN 302-623
Section of EN 302-544 part 2 Specification
4.2.2
4.2.3
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Parameter files Remark
Accuracy of max. Output Power
Spectrum Emission Mask
Page: 19 Revision: 1.0
WiMaxHarmonizedMsStaticPo wer.exe WiMaxHarmonizedMsTrx.exe
TS8977 System Manual WiMAX Applications
4.2.4
4.2.5
4.2.6
4.2.9
TX Adjacent Channel Leakage Power Ratio
Transmit Spurious Emissions Transmitter Power Control
Control and Monitoring Functions
WiMaxHarmonizedMsModulati onQuality.ex WiMaxHarmonizedMsTrx.exe WiMaxHarmonizedMsStaticPo wer.exe WiMaxHarmonizedMsStaticPo wer.exe
4.1.2.2.1
Parameter files Remark
4.2.2
4.2.3
4.2.4
4.2.5
4.2.6
4.2.7
MS_1.04_norm_5AL_MaxOutputPower.par MS_1.04_norm_5BL_MaxOutputPower.par MS_1.04_norm_5CL_MaxOutputPower.par MS_1.01_norm_5AL_SpectrumEmissionMask.par MS_1.01_norm_5BL_SpectrumEmissionMask.par MS_1.01_norm_5CL_SpectrumEmissionMask.par MS_1.02_norm_5AL_ACLR.par MS_1.02_norm_5BL_ACLR.par MS_1.02_norm_5CL_ACLR.par MS_1.03_norm_5AL_Tx_SpuriousEmissions.par MS_1.03_norm_5BL_Tx_SpuriousEmissions.par MS_1.03_norm_5CL_Tx_SpuriousEmissions.par MS_1.05_norm_5AL_TransmitterPowerControl.par MS_1.05_norm_5BL_TransmitterPowerControl.par MS_1.05_norm_5CL_TransmitterPowerControl.par MS_1.06_norm_5AL_ControlAndMonitoringFunctions.par MS_1.06_norm_5BL_ControlAndMonitoringFunctions.par MS_1.06_norm_5CL_ControlAndMonitoringFunctions.par
4.2 Transmit/Receiver Tests (TRX)
4.2.1 Test Description
Transmitter Tests (TC numbers according to EN 302-544 part 2)
4.2.2 Spectrum Emission Mask
Spectrum emission mask defines an out of band emission requirement for the transmitter. These out of band emissions are unwanted emissions outside the channel bandwidth resulting from the modulation process and non-linearity in the transmitter but excluding spurious emissions.
The spectrum emission mask of the UE applies to frequency offsets between 2,5 MHz and 12,5 MHz on both sides of the UE centre carrier frequency. The out-of-channel emission is specified as power level measured over the specified measurement bandwidth but relative to Pnom of the UE carrier measured in Eval_BW1 centred in the 5MHz channel.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 20 Revision: 1.0
TS8977 System Manual WiMAX Applications
The spectrum emission mask of the UE applies to frequency offsets between 5 MHz and 25 MHz on both sides of the UE centre carrier frequency. The out-of-channel emission is specified as a power level relative to Pnom of the UE carrier measured in Eval_BW1 centred in the 10 MHz channel.
4.2.3 Transmitter Adjacent Channel Power Ratio (ACLR)
Adjacent channel leakage power ratio (ACLR) is the ration measured through a filter pass band centred on the assigned channel frequency to the mean power measured through a filter pass band centred on a first or second adjacent channel.
4.2.4 Transmitter Spurious Emissions
Transmitter spurious emissions are emissions which are caused by unwanted transmitter effects such as harmonics emission, parasitic emission, intermodulation products and frequency conversion products, but exclude out of band emissions. This is measured at the RF output port.
4.2.5 Nominal maximum output power and tolerance
The nominal maximum output power (Pnom) is the nominal mean power level measured over total allocated channel bandwidth available at the antenna connector. The tolerance of the output power is the difference between the declared Pnom and the actual maximum output power with the corresponding tolerance limits.
4.2.6 Transmitter power control
Transmit Power Control (TPC) is a mechanism that shall be used by the equipment to ensure a mitigation factor on the aggregate power from a large number of devices to improve the spectrum sharing conditions.
4.2.7 Control and Monitoring Functions
This requirement verifies that the control and monitoring functions of the UE prevent it from transmitting in the absence of a real network.
Receiver Tests
(TC numbers according to EN 302-544 part 2)
The receiver test method covers various test cases for the subscriber station. Most of these test cases are based on the criterion that a connection between the DUT and the emulator is accomplished with a PER (packet error rate) below a certain limit.
4.2.8 Receiver Spurious Emissions
. The spurious emissions power is the power of emissions generated or amplified in a receiver that appear at the UE antenna connector
4.2.9 Receiver adjacent channel selectivity (ACS)
The receiver and alternate channel selectivity (ACS) is a measure of the receivers ability to receive a wanted signal at its assigned channel frequency in the presence of an adjacent channel signal at a given frequency offset from the center frequency of the assigned channel. ACS is the interferer power level (in dB) relative to the thermal noise.
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 21 Revision: 1.0
TS8977 System Manual Glossary
5 Glossary
AC BER BPSK BS BSSE BW CINR DC DL DUT FDD FCH GPIB HF H-FDD IEEE IF IP IQ ISSCU ITU LAN LED LPT MCS MS MRC MRCT OFDM PASS PER PSD PSK QPSK QAM RF RFC R&S RSSI RX SBS SFT SIG SS SSRTG SSTTG SW TCP TDD
Alternating Current Bit Error Ratio
inary PSK
B Base Station Base Station and Subscriber Emulator Band Width Carrier-to-Interference-and-Noise Ratio Direct Current Down Link Device Under Test Frequency Division Duplex Frame Control Header General Purpose Interface Bus High Frequency FDD Half Duplex Institute of Electrical and Electronic Engineers Intermediate Frequency Internet Protocol Inphase Quadrature Inband Signal Switching and Conditioning Unit International Telecommunication Union Local Area Network Light Emitting Diode Local Parallel Port Modulation and Coding Scheme Mobile Station Maximum Ratio Combining Mobile Radio Conformance Test Orthogonal Frequency Division Multiplex Parametrical Application Software for Systems Packet Error Rate Power Spectral Density Phase Shift Keying Quadrature PSK Quadrature Amplitude Modulation Radio Frequency Radio Frequency Calibration Rohde & Schwarz Receive Signal Strength Indicator receive Serving Base Station Self Test Signal Subscriber Station SS receive to transmit turnaround time SS transmit to receive turnaround time Software Transmission Control Protocol Time Division Duplex
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 22 Revision: 1.0
TS8977 System Manual Glossary
TSCC TX UL USB VSA VSG WiMAX
Test System Control Center transmit Up Link Universal Serial Bus Vector Signal Analyser Vector Signal Generator Worldwide Interoperability for Microwave Access
© Rohde & Schwarz GmbH & Co. KG 2007 File: TS8970-System-Manual
Page: 23 Revision: 1.0
Loading...