This document contains all system specific information that is necessary to operate and maintain the
TS8977 Test System.
The test system supports numerous RF test cases, based on test methods defined in the EN 302-544
Part 2 and EN 302-623: “Harmonized EN for TDD User Equipment covering essential requirements of
article 3.2 of the R&TTE Directive”. These test methods can be executed with defined set of
parameters, being compliant to the relevant sections of the EN 302-544 respective EN 302 623 test
specification. However, the free selection of parameters in order to define customer specific R&D test
scenarios is supported as well.
The test system is designed to perform RF tests for WiMAX subscriber (mobile) stations. The system
support following WiMAX Profiles:
MP05 (2,5 - 2,69 GHz, 5MHz and 10MHz bandwidth),
MP09 (3,4 – 3,6 GHz, 5MHz bandwidth)
MP10 (3,4 – 3,6 GHz, 7MHz bandwidth)
MP12 (3,4 – 3,6 GHz, 10MHz bandwidth)
To simulate interfering signals, R&S SMU vector signal generators (VSG) are integrated in the TS8977
Test System. The RF output of the DUT is analysed by means of the built-in vector signal analyser
(VSA).
The in-band signal switching and conditioning unit SSCU is required to connect the DUT to the RF
devices within the test system. It contains all RF components which are not WiMAX specific.A separate
Advanced Signal Conditioning Unit (ASCU) is used to filter, combine, split, attenuate or amplify band
specific signals which are WiMAX specific.
The system software to operate and maintain the system, namely the RS-PASS Software, is described
in the RS-PASS Operation Manual.
The TS8977 hardware consists of the following components:
No Type Device or Option Manufacturer
1PSL3
2FSU26
FSP-B10 External generator control for ESCI, ESPO
3SMU 200A
SMU-B106 RF path A 100kHz to 6 GHz
SMU-B11 Baseband generator (16 Msample)
SMU-B13 Baseband main module
SMU-B17 Analog baseband input
SMU-B81 Rear panel connectors for 1st RF path
SMU-K49 Digital Standard IEEE 802.16
SMU-K62 Additional white gaussian noise
In signalling mode a connection between the protocol and signalling unit and the DUT has to be
established.
For transceiver and transmitter tests the signal of the DUT is directed not only to the protocol and
signalling unit, but also to the spectrum analyser FSU. This transmission path is the so-called measurement or analyser path.
For receiver tests, interferer signals are generated by the SMU and SMR signal generators. The
interferer signals are combined with the wanted signals of the protocol and signalling unit. This
transmission path is the so-called stimulus or generator path.
The system controller controls all devices within the TS8977. Most of the instruments are commanded
via GPIB connection. The exceptions include the switching units SSCU and ASCU, which are connected
via parallel interface.
The system controller hosts the RS-PASS software, including the system software with all required
device drivers, the applications performing the tests as defined, the graphical user interface and several
tools. Please refer to the RS-PASS operation manual for more details.
The system is being delivered with the Rohde & Schwarz PSL3, configured with 512 MB RAM, 2 hard
disks and two Ethernet interfaces. The operating system is WINDOWS XP.
Before the system controller can be powered up the monitor, the mouse and the keyboard have to be
connected to the TS8977 system panel. To switch “ON” the system, the rocker switch on the front panel
has to be pushed behind the flap.
2.2.1.2 Ethernet HUB
The first Ethernet adapter of the system controller is connected to the Ethernet HUB. All TCP/IP devices
installed on the system have to be connected to that HUB. However, the second adapter card of the
PSL3 shall not be connected to the HUB and cannot be used for a connection to the customer's
company network.
Be aware that for connections to the HUB cross link cables can not be used! Only for
direct connections between instruments a cross link network cable is necessary.
The SMC Ethernet HUB is not equipped with a power switch. In order to apply power to the device, the
power cord needs to be plugged in the AC source. A "power" LED will illuminate when main power is
applied. Several activity and state LEDs indicate the status of the connections.
2.2.2 Signal Generating Instruments
2.2.2.1 Signalling Unit CMW270
In most test cases it is necessary to establish a running connection between the test system and the
DUT before a RF performance test can be performed.
As a consequence the test system has to simulate a base station. In the case of the simulation of a
base station, the test system has to page the DUT, enabling it to access one of the simulated WiMAX
cells and simulating the call establishment process. To achieve this, the corresponding protocol
functionalities, defined in the WiMAX specifications, have to be implemented by the use of a signalling
device.
In advance of the execution of a test case, a call set-up has to be performed. The system controller
instructs the signalling unit to perform the required call set-up via ethernet. After a successful call set-up
(initial state), the measurement will be carried out.
In the case of receiver tests, the CMW270 is capable of measuring the PER (Packet Error Rate).
When doing transmitter tests, the CMW270 delivers sophisticated trigger signals. This is especially
important for the analysis of signal waveforms performed by the spectrum analyser and triggered by the
CMW270.
Wanted Signals
Usually the wanted signal is a standard WiMAX signal.
2.2.2.2 RF Signal Generators SMU and SMR
The RF generator SMU200 is employed to produce interfering signals with frequencies of up to 6 GHz.
They are also used to stimulate several RFC and SFT measurements. To generate different kinds of
signals, the SMU200s are also equipped with a modulation coder and a data generator.
There is an AC voltage selection feature implemented in the RF signal generator SMU200 that
automatically sets itself according to the applied AC voltage. The SMU200 is switched on by first
pressing the rear panel AC mains switch followed by the front panel power switch "ON/STBY". Switching
off the system at the rear is only necessary, when the instrument is to be disconnected completely from
the mains.
Interfering signals
The SMR27 generators can create interfering cw signals up to 27 GHz. The SMR27 is used as
generator for the blocking test and the RF path calibration for the spurious emission tests.
2.2.3 Measurement Instruments
2.2.3.1 Spectrum Analyser FSU26
The spectrum analyser FSU26 measures the properties of the DUT´s signal during numerous tests,
such as output power as a function of frequency or time, or spurious emissions and spurious response
rejection.
Both devices are equipped with an AC voltage selection feature, which automatically sets itself
according to the applied power. It has to be checked that the ”ON/STBY” switch, on the front at the
bottom left hand corner, is in the correct position. Moreover, an AC mains supply ”ON/OFF” switch is
situated on the rear of the device at the top right hand corner. Pressing this rocker switch will apply the
AC mains to the instrument.
Being dependent on the test purpose, the FSQ is operated in different measurement modes:
Peak hold: The maximum value occurring for a certain property during the measurement time is
•
taken.
•Average: The measurement value is averaged over the measurement time.
•Sample: To analyse the dependence of a certain property against time, each individual
measurement value is taken.
•RMS: The RMS deviation from the average measurement value is taken.
•Min/max: An upper and lower envelope is derived from the individual measurement values.
More than one measurement mode can be applied in a single test run.
2.2.3.2 Power Meter
The power meters NRP-Z21 and NRP-Z55 are used for the RFC and for SFT measurements. It is
delivered with an USB adapter NRP-Z4.
The NRP-Z21 and NRP-Z55 are the most accurate devices of the TS8977 system with the typical
measurement uncertainty being less than 0.2 dB.
The NRP-Z21 is supplied with power via USB.
2.2.4 Switch Units
2.2.4.1 Main Switching Unit (SSCU-77)
This SSCU-77 is used to interconnect all RF instruments within the system and connects the DUT to the
system. This system is indispensable for running test cases. It includes components to split, combine,
filter, attenuate or amplify RF signals which are not frequency band specific. The SSCU-77 is designed
for a frequency range from 400 MHz to 6 GHz.
The SSCU-77 is powered by DC +5 V, +15 V and +28 V. These voltages are supplied by the switched
mode power supply (CS-PSSU), located behind the cover plate at the top of the rack. In order to apply
the AC mains to the SSCU, the rocker switch, situated on the right hand of the instrument, has to be
depressed. All three yellow LED's, located at the SSCU front panel, should be illuminated.
2.2.4.2 Advanced Signal Switching and conditioning Unit (ASCU-77)
The ASCU-77 is an extension for the SSCU and contains all RF components which are frequency
specific for WiMAX frequency bands like filters and amplifiers. The ASCU-77 is designed for a
frequency range from 400 MHz to 26 GHz.
The Line distributor (LDE) is used to provide the AC mains voltage to the devices in the test system. It is
equipped with an EMI filter that prevents the system from high frequencies on the power line. The LDE
has a switch enabling the user switching the system on and off (refer to section 3.2).
The LDE is mounted into the system on the rear side at about the same height as the system controller
PSL 3.
2.2.5.2 DC Power Supply
The power supply NGPQ32 is used to provide the DUT with an adjustable amount of energy. It is set
into operation with the power switch on the front panel. After being switched on, the digital display shows
the last settings of the voltage and current of the outputs. When the system is switched on for the first
time, the display will show 0 voltage and 0 current until it has been programmed. The outputs are "off" at
this stage.
The NGPQ2 supports four different AC mains supply voltage levels 100 V, 120 V, 220 V and 240 V with
system frequencies of 47 Hz to 63 Hz.
The AC mains voltage range must be set to:
•100 V for e.g. Japan
•120 V for e.g. USA
•230 V for e.g. central Europe (exceptions: 220 V e.g. Turkey, 240 V e.g. United Kingdom)
In the table below are the hardware remote interface settings of the system instruments:
DeviceResource Name / Remote AddressComments
Signalling Unit CMW270tbd
Signal generator SMUGPIB::18
Signal generator SMRASRL1
Spectrum analyser FSUGPIB::01
Power meter NRPUSB::0x0aad::0x000c::_ _ _ _ _ _
SSCULPT::1
DC Power supply NGMOGPIB::07
NRVSGPIB::20
The resource settings are checked and updated, when running the System Explorer.
The interfaces and resources are indicated by the parameter resource of the physical devices, listed in
the system configuration file …RS PASS COMMON/ini/testsystem.ini. The name of the system
configuration file can be chosen by the user.
2.3.2 Connectors
These are the HF connector settings of the system instruments:
Signalling Unit CMW270SIG RXSIG TX
Signal generator SMUSG2RF-A
Signal generator SMRSG1RF-50 Ohm
Spectrum analyser FSUSARF INP
Power meter NRP-Z11PMPower Probe
SSCULPT
ASCULPT
The connectors are defined in the system and must not be changed.
he settings are indicated by the parameters SSCU output connector, SSCU input connector,
T
device output connector and device input connector of the physical devices listed in the system
configuration file …RS PASS COMMON/ini/testsystem.ini.
Not all connectors offered with the system instruments are indicated in the system configuration files, as
not all are relevant. The connectors of the SSCU are the following:
Front side SSCU-77
Connector Remark
SIGNAL ANALYSER RF INPUT Connection to FSU 'RF_INP'
REFERENCE OUT RF output used for RFC
MS/BS DUT
TX
TX1/RX1
TX2/RX2
SIG TX Connection to signalling unit CMW270 (receiver part)
SIG RX Connection to signalling unit CMW270 (transmit part)
SG1 Connection to SMR RF 50 Ohm
Connection to mobile station.
Mobile Transmitter
Combined connector Mobile TX/RX
Combined connector MobileTX/RX
Rear side SSCU-77
Connector Remark
PM Connection to Power Meter NRP-Z21
PARALLEL PORT IN Connection to system controller ‘LPT 1’ for ISSCU control
PARALLEL PORT OUT Not connected
SG2 Connection to SMU RF-A
ANALYSER BOX
Input
Output
GENERATOR BOX
Input
Output
The connectors of the signal generating and measurement devices are described thoroughly in the
corresponding manuals.
2.4 Operating Conditions
The optimum environmental conditions for the TS8970 system are the following.
Temperatures below the specified values are not critical.
Temperatures above the specified values should be avoided in order to prevent
damages!
•Operating temperature for the TS8977 : 23°C ± 3°C
•Temperature range after RF calibration :± 2°C
•Temperature of the rack mounted components : 20°C to 40°C
•Temperature of the SSCU-77: 20°C to 40°C
•Temperature of the ASCU-77: 20°C to 40°C
•Temperature of the components inside SSCU : 20°C to 45°C
•Temperature of the components inside ASCU : 20°C to 45°C
•Humidity : 20 to 80 %
2.5 Calibration Requirements
The system device calibration is recommended to be done within an interval of 1 year. It is independent
from the instrument´s internal calibration or the RFC, described in the RS-PASS Operation Manual.
It is recommended to check the cabling of the system. Every original cable, used for the cabling of the
system, is labelled with the name of the device to be connected to it.
It has to be taken care, that all cables are connected correctly!
The DUT cable has to be connected to the DUT Tx/Rx connector. It has to be made sure that the
specific DUT cable is indicated as active in the system configuration *.ini file. The DUT cable
configuration is done in the system editor of the TSCC and is described in detail in the RS-PASS
Operation Manual.
Before running a WiMAX or another RFC application, the RFC calibration of the DUT
cable has to be accomplished.
(The calibration of the DUT cable is described in the RS-PASS Operation Manual.
A rough overview over the system´s internal connections is given in section 2.3.2.
3.1.2 Devices
The device´s internal calibration can be updated by running the self test calibrate_all. The internal
calibration is independent to the RFC and is described in detail in the RS-PASS Operation Manual.
3.1.3 System Configuration
After the start of the system the configuration, especially the device resource (bus systems used to
control the devices and device addresses) and connector settings, has to be checked. This is done by
running the System Explorer, thoroughly described in the RS-PASS Operation Manual.
When the system is started for the first time, i.e. after installation of the RS-PASS
software, the System Explorer has to be executed before other applications can be
carried out.
The device interfaces and connectors are described in section 2.3.
In order to start the system, the instruments have to be switched ON. Main power is applied by pushing
the switch at the rear bottom of the system. With this switch all instruments can be switched ON and
OFF .
Additionally, most instruments can be switched ON/OFF, or put into standby mode individually by
pressing the relevant power button at the front panel:
STBY / ON to set the device into standby mode
OFF / ON to switch the device off/on
Most of the devices have a main power switch at the rear panel to connect or disconnect it from the
mains. For more details please refer to the instrument descriptions in section 2.2 .
If the main power connection at the rear of an instrument is switched off, then this
device cannot be switched on with the SYSTEM POWER ON/OFF switch or with the
ON/STBY and ON/OFF button at the front panel of the instrument.
3.2.2 System Start and Shut Down
After switching on the test system, the system controller automatically starts booting. After booting is
finished, a login window is displayed and the username and password has to be entered. The default
logins are:
•System Manager: User name: ‘administrator’
Password : delivered without password
•Account TS8977: User name: ‘ts8977’
Password : ‘ts8977’
Both accounts hold system manager rights, required e.g. for a system software
update. To ensure, that users cannot damage the operating system, please make sure
to create individual user accounts for daily work!
3.2.3 Start and Shut Down of the Signalling Unit
When the main switch "SYSTEM POWER ON/OFF" at the rear system panel is used to start the
TS8977 test system, the signalling unit is automatically started as well. When switching off the system it
also powers down the signalling unit.
This section lists the conformance test cases delivered by R&S and gives an overview which test
methods are provided. All descriptions and tables in this document are composed according to the EN
302-504 part 2 and EN 302 623 test specifications. In both specification similar test cases are
described, but unfortunately the test case numbering is different. For example, the spectrum emission
mask test has in EN 302-544-2 the test case number 4.2.2 whereas in the EN 302 623 the number is
4.2.3. In order to use the same test method for the similar test cases in the RS PASS SW there is
following mapping to achieve same number for same test cases:
RS PASS
TC Number
TX Tests
1.1 Spectrum Emission Mask 4.2.2 4.2.3
1.2 TX Adjacent Channel Leakage
1.3 TX Spurious Emissions 4.2.4
1.4 Accuracy of max. Output Power 4.2.5 4.2.2
1.5 TX Power Control
1.6 Control & Monitoring Functions 4.2.7 4.2.9
RX Tests
2.1 RX Intermodulation Rejection 4.2.11 n.a.
2.2 RX Adjacent and Alternate
2.3 RX Spurious Emissions
2.4 RX Blocking
2.5 RX Spurious Response 4.2.12 n.a.
Test Case TC Number in
EN 302 544-2 EN 302 623
4.2.3 4.2.4
Power Ration
4.2.5
4.2.6 4.2.6
4.2.9 4.2.8
Channel Rejection
4.2.8 4.2.7
4.2.10 n.a
For details and up-to-date information on test cases, please refer to the latest version of the mentioned
specifications and release notes of the RS-PASS Software.
As the test system devices are hardlock protected, the availability of the test cases
may depend on individual registered options.
Transmitter Tests (TC numbers according to EN 302-544 part 2)
4.2.2 Spectrum Emission Mask
Spectrum emission mask defines an out of band emission requirement for the transmitter. These out of
band emissions are unwanted emissions outside the channel bandwidth resulting from the modulation
process and non-linearity in the transmitter but excluding spurious emissions.
The spectrum emission mask of the UE applies to frequency offsets between 2,5 MHz and 12,5 MHz on
both sides of the UE centre carrier frequency.
The out-of-channel emission is specified as power level measured over the specified measurement
bandwidth but relative to Pnom of the UE carrier measured in Eval_BW1 centred in the 5MHz channel.
The spectrum emission mask of the UE applies to frequency offsets between 5 MHz and 25 MHz on
both sides of the UE centre carrier frequency.
The out-of-channel emission is specified as a power level relative to Pnom of the UE carrier measured
in Eval_BW1 centred in the 10 MHz channel.
4.2.3 Transmitter Adjacent Channel Power Ratio (ACLR)
Adjacent channel leakage power ratio (ACLR) is the ration measured through a filter pass band centred
on the assigned channel frequency to the mean power measured through a filter pass band centred on
a first or second adjacent channel.
4.2.4 Transmitter Spurious Emissions
Transmitter spurious emissions are emissions which are caused by unwanted transmitter effects such
as harmonics emission, parasitic emission, intermodulation products and frequency conversion
products, but exclude out of band emissions. This is measured at the RF output port.
4.2.5 Nominal maximum output power and tolerance
The nominal maximum output power (Pnom) is the nominal mean power level measured over total
allocated channel bandwidth available at the antenna connector. The tolerance of the output power is
the difference between the declared Pnom and the actual maximum output power with the
corresponding tolerance limits.
4.2.6 Transmitter power control
Transmit Power Control (TPC) is a mechanism that shall be used by the equipment to ensure a
mitigation factor on the aggregate power from a large number of devices to improve the spectrum
sharing conditions.
4.2.7 Control and Monitoring Functions
This requirement verifies that the control and monitoring functions of the UE prevent it from transmitting
in the absence of a real network.
Receiver Tests
(TC numbers according to EN 302-544 part 2)
The receiver test method covers various test cases for the subscriber station. Most of these test cases
are based on the criterion that a connection between the DUT and the emulator is accomplished with a
PER (packet error rate) below a certain limit.
4.2.8 Receiver Spurious Emissions
.
The spurious emissions power is the power of emissions generated or amplified in a receiver that
appear at the UE antenna connector
4.2.9 Receiver adjacent channel selectivity (ACS)
The receiver and alternate channel selectivity (ACS) is a measure of the receivers ability to receive a
wanted signal at its assigned channel frequency in the presence of an adjacent channel signal at a given
frequency offset from the center frequency of the assigned channel. ACS is the interferer power level (in
dB) relative to the thermal noise.
AC
BER
BPSK
BS
BSSE
BW
CINR
DC
DL
DUT
FDD
FCH
GPIB
HF
H-FDD
IEEE
IF
IP
IQ
ISSCU
ITU
LAN
LED
LPT
MCS
MS
MRC
MRCT
OFDM
PASS
PER
PSD
PSK
QPSK
QAM
RF
RFC
R&S
RSSI
RX
SBS
SFT
SIG
SS
SSRTG
SSTTG
SW
TCP
TDD
Alternating Current
Bit Error Ratio
inary PSK
B
Base Station
Base Station and Subscriber Emulator
Band Width
Carrier-to-Interference-and-Noise Ratio
Direct Current
Down Link
Device Under Test
Frequency Division Duplex
Frame Control Header
General Purpose Interface Bus
High Frequency FDD
Half Duplex
Institute of Electrical and Electronic Engineers
Intermediate Frequency
Internet Protocol
Inphase Quadrature
Inband Signal Switching and Conditioning Unit
International Telecommunication Union
Local Area Network
Light Emitting Diode
Local Parallel Port
Modulation and Coding Scheme
Mobile Station
Maximum Ratio Combining
Mobile Radio Conformance Test
Orthogonal Frequency Division Multiplex
Parametrical Application Software for Systems
Packet Error Rate
Power Spectral Density
Phase Shift Keying
Quadrature PSK
Quadrature Amplitude Modulation
Radio Frequency
Radio Frequency Calibration
Rohde & Schwarz
Receive Signal Strength Indicator
receive
Serving Base Station
Self Test
Signal
Subscriber Station
SS receive to transmit turnaround time
SS transmit to receive turnaround time
Software
Transmission Control Protocol
Time Division Duplex
Test System Control Center
transmit
Up Link
Universal Serial Bus
Vector Signal Analyser
Vector Signal Generator
Worldwide Interoperability for Microwave Access