Roche URISYS 2400 Operator's Manual

URISYS 2400
Operator’s Manual
Version 1.0

Revision History

URISYS 2400
Version Revision date
1.0 April 2001
Language Order No.
English
Edition notice URISYS 2400 Operator’s Manual
manual is correct at the time of printing. However, Roche Diagnostics GmbH reserves the right to make any changes necessary without notice as part of ongoing product development.
Any customer modification to the instrument will render the warranty or service agreement null and void.
Intended use The URISYS 2400 is a fully automated urinalysis system intended for in vitro
qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and erythrocytes.
Copyrights © 2001, Roche Diagnostics GmbH. All rights reserved.
Trademarks The following trademarks are acknowledged:
, URISYS 2400 are registered trademarks of the Roche group. All other trademarks that are mentioned in this manual are the property of the respective trademark holders: KOVA® and Monovette®.
03184315.018
Instrument approvals The URISYS 2400 analyzer meets the requirements stated in
Directive 9879EC of the European Parliament and the Council of the European Union (EU) on in vitro diagnostic medical devices. Furthermore, the URISYS 2400 analyzer is manufactured and tested according to International Standard IEC 1010-1, "Safety requirements for electrical equipment for measurement, control, and laboratory use, Part 1: General requirements". This International Standard is equivalent to the standards UL 3101-1 for the USA, CSA C 22.2 No. 1010.1 for Canada, and DIN EN 61010-1 for Germany.
The URISYS 2400 fulfils the EMC immunity requirements for laboratory use equipment in industrial areas, according to the EMC standards EN 50082-2 and EN 61326-A1. According to EN 55011, the URISYS 2400 analyzer is a device of limit class A.
Roche Diagnostics ii Operator’s Manual · Version 1.0
URISYS 2400
Contact addresses
Compliance is demonstrated by the following marks:
Complies with the IVD directive 9879EC.
Issued by Underwriters Laboratories, Inc. (UL) for
CUS
®
Canada and the US.
Manufacturer:
Hitachi Science Systems Ltd. Ibaraki Japan
Warranty
Authorized Representative
EC REP
USA Roche Diagnostics
Roche Diagnostics GmbH Sandhofer Strasse 116 D-68305 Mannheim Germany
Corporation 9115 Hague Road PO Box 50457 Indianapolis, IN 46250 USA
Refer to the URISYS 2400 analyzer purchase agreement for warranty conditions. Contact your local Roche Technical Support for further information.
Roche Diagnostics Operators Manual · Version 1.0 iii
URISYS 2400
Roche Diagnostics iv Operators Manual · Version 1.0
URISYS 2400

Table of contents

Revision History ii Table of contents v Preface vii How to use this manual vii Conventions used in this manual viii Safety classifications x Safety information xi
Overview Part A
1 Introduction to the system
Description of the URISYS 2400 analyzer A-2
System description Part B
2Hardware
Hardware description B-2 Installation B-5
3Software
Installation B-12
4 Cassette
First-time installation B-30
Maintenance Part D
7 General maintenance actions
Maintenance procedures D-2 Maintenance menu D-4 Exchange user-replaceable parts D-8
Troubleshooting Part E
8 Data alarms (flags)
Flags on result data printout E-2 Handle alarms E-3 Reset the analyzer E-9
9 Instrument alarms (messages)
Instrument alarms (messages) E-12
Appendix Part F
10 Technical specifications
Technical specifications F-2 Barcode specifications F-3 Concentration ranges F-5
Operation Part C
5 Daily operation
Initial preparations C-3 System start C-4 Sample preparation C-8 Termination of analysis C-11 Remove processed samples C-12 Process additional samples C-13 Process STAT (emergency) samples C-14 Process control samples C-17 Replace an empty URISYS 2400 cassette C-18 Replace an expired URISYS 2400 cassette C-19 Re-insert a used URISYS 2400 cassette C-20 Analysis and documentation of results C-21 Switch off the analyzer C-34
6 Special operations
Calibration C-36 Checks C-41
Glossary and index Part G
Glossary G-1 Index G-5
Roche Diagnostics Operators Manual · Version 1.0 v
URISYS 2400
Roche Diagnostics vi Operators Manual · Version 1.0
URISYS 2400

Preface

The URISYS 2400 analyzer is a fully automated, benchtop urine analyzer. It is designed for use with URISYS 2400 cassettes.
The URISYS 2400 is optimized for workloads of between 100 and 1000 urine samples per day. Only trained personnel working in a professional laboratory environment must operate the URISYS 2400. The analyzer has been developed and designed to assay urine samples only.
It is important that the operator reads this manual thoroughly before using the analyzer.
Any disregard of the instructions in the Operators Manual may result in a safety risk.

How to use this manual

This manual is designed to help you perform all tasks required as part of your work with the URISYS 2400.
Content The manual is divided into the following parts:
o Part A – Overview: The overview provides general information about the
URISYS 2400 analyzer and specific details about the safety precautions.
o Part B – System description: The system description outlines the hardware
and software components of the analyzer, the installation procedure, and the use of the touch screen.
o Part C – Operation: This section defines the daily and special operations of
the URISYS 2400 analyzer in a step-by-step guide.
o Part D – Maintenance: This section provides details on all of the general
maintenance actions.
o Part E – Troubleshooting: This section contains information on how to
respond to data and instrument alarms, and provides troubleshooting guidelines for alarms that cannot be solved by the operator. Alarms that require intervention by trained Roche Technical Support personnel are indicated.
o Part F – Appendix: This section contains additional information,
including technical specifications, glossary, and index.
Roche Diagnostics Operators Manual · Version 1.0 vii

Conventions used in this manual

Visual cues are used to help locate and interpret information in this manual quickly. This section explains formatting conventions used in this manual.
Symbols The following symbols are used:
Symbol Used for
a Procedural step
o List item
e Cross-reference
i Note and tip
[Parameter] Name of screen (enclosed in brackets)
<Start> Name of button (enclosed in less and greater)
URISYS 2400
Abbreviations The following abbreviations are used:
Abbreviation Definition
A
ASCII American Standard Code for Information Interchange
B
BIL Bilirubin
C
CCD Charge-coupled device
C/D Check digit
CLA Clarity
COL Color
COM Compensation pad
COMM.Parameter Communication parameters for host-communication
available on the [System Parameters] screen
E
EMC Electromagnetic Compatibility
EN European Standard
ERY Erythrocytes
G
GLU Glucose
Roche Diagnostics viii Operators Manual · Version 1.0
URISYS 2400
Abbreviation Definition
I
ID Identification number
IEC International Electrotechnical Commission
K
KET Ketones
L
LED Light Emitting Diode
LEU Leukocytes
L.TRB Lightly turbid
N
NEG Negative
NIT Nitrite
nm Nanometer
NORM Normal
O
ORANG Orange
P
P.YEL Pale Yellow
POS Positive
PRO Protein
S
SG Specific Gravity
SI Standard International
T
TS Test strip
TURBD Turbid
U
UBG Urobilinogen
Y
YELLO Yellow
Roche Diagnostics Operators Manual · Version 1.0 ix

Safety classifications

Before operating with the URISYS 2400 it is essential that the warnings, cautions, and safety requirements contained in this manual are read and understood by the user. This section explains how precautionary information is formatted in the manual.
The safety precautions and important user notes are classified according to ANSI Z535 standards. Familiarize yourself with the following meanings and icons:
Warning
Indicates a possibly hazardous situation which, if not avoided, may result in death or serious injury. Examples of a “serious injury” include loss of eyesight, burn (high temperature, low temperature), electric shock, bone fracture, or poisoning. These injuries require medical assistance.
URISYS 2400
Biohazard
Samples containing material of human origin must be treated as potentially infectious. The relevant laboratory guidelines on safe use must be observed.
Caution
Indicates a possibly hazardous situation which, if not avoided, may result in slight or minor injuries, and/or damage to equipment. “Minor injury” refers to injuries that may require medical assistance. “Equipment” refers to extended damage to buildings, furniture, and so on.
Spillage
Avoid all liquid spills on the analyzer.
Roche Diagnostics x Operators Manual · Version 1.0
URISYS 2400

Safety information

Electrical safety
Connect the analyzer to grounded power outlets only (protection class 1). All peripheral devices that are connected to the URISYS 2400 must comply with safety standard IEC 950 (UL 1950) for information technology equipment, or with IEC 1010 (UL 3101) for laboratory use instruments.
Instrument in use
Keep hands or other objects away from the sampling probe area to prevent personal injury while the instrument is in use. Do not touch parts of the analyzer other than those specified.
User qualification
Only appropriately trained operators are qualified to operate the analyzer.
Correct use
Any disregard of the instructions in the Operators Manual may result in a safety risk. Use the URISYS 2400 analyzer to analyze urine samples only. It is not intended for any other application.
Transport
At least two persons must carry the analyzer, by holding the baseplates on the left side and right side of the analyzer. Be careful not to hurt your hands or fingers when putting the analyzer in place.
Environmental conditions
The URISYS 2400 is approved for indoor use only.
Biological safety
Liquid waste and strip waste are potentially biologically hazardous. Always wear gloves if handling those materials. Do not touch parts of the analyzer other than those specified. Consult your laboratory protocol for handling biohazardous materials.
Spilling and cleaning
If a sample is spilled on the analyzer, wipe up immediately and apply disinfectant. When handling the washing solution, ensure that you wear gloves and safety goggles to avoid direct contact with the solution. If the solution contacts the body, wash immediately with a large volume of water. Consult your laboratory protocol for handling biohazardous materials.
Roche Diagnostics Operators Manual · Version 1.0 xi
URISYS 2400
Radio interference
URISYS 2400 is a class A device. In residential areas it may cause radio interference. The user must take precautions as required.
Installation
Follow the specified installation conditions carefully. Otherwise, inaccurate results, or damage to the analyzer may occur.
Waste
Handle liquid waste and used test strips properly, according to legislation on water pollution, and on the treatment of drainage and waste matter.
Electromagnetic waves
Devices that emit electromagnetic waves may affect measured data, or cause the analyzer to malfunction. Do not operate the following devices in the same room where the analyzer is installed: mobile phone, transceiver, cordless phone, other electrical devices that generate electromagnetic waves.
o
Operating and maintenance
Carefully follow the procedures specified in the Operators Manual for the operation and maintenance of the analyzer. Leave maintenance of other areas to trained Roche Technical Support personnel.
o
Calibration
Calibrate the analyzer every four weeks to ensure the analyzer operates optimally. Failure to do so may lead to inaccurate results.
o
Test strips
Handle and store test strips according to the instructions provided by Roche Diagnostics. Refer to the instructions in the package insert of the URISYS 2400 cassette.
o
Particular care must be taken when performing the following procedures: Preparing and processing samples (routine, STAT and control) Using the washing solution Draining the liquid waste tank Cleaning the TS transfer base Disposing of used test strips Cleaning the waste box
Roche Diagnostics xii Operators Manual · Version 1.0
URISYS 2400
D
C B
A
E
Figure 0-1 URISYS 2400 analyzer with labels
A
Housing
B STAT sample position
D Strip waste container door E Power supply
C Housing before sampling area
Symbol Validity Symbol Validity
o STAT position o Rack transfer
o STAT position
o Rack transfer o Waste box o TS transfer base o Power supply
o Sampling probe
o STAT sample position o Rack transfer o Rinsing station o Waste box o TS transfer base
Roche Diagnostics Operators Manual · Version 1.0 xiii
URISYS 2400
Roche Diagnostics xiv Operators Manual · Version 1.0

Overview

Part A is an overview of the URISYS 2400.
A
URISYS 2400 1 Introduction to the system
Table of contents

Introduction to the system

This chapter contains an introduction to the URISYS 2400 analyzer.
In this chapter A
Description of the URISYS 2400 analyzer .........................................................2
Measuring principle ......................................................................................2
Chapter
1
Roche Diagnostics Operators Manual · Version 1.0 A-1
1 Introduction to the system URISYS 2400
Description of the URISYS 2400 analyzer
Description of the URISYS 2400 analyzer
The URISYS 2400 analyzer is a fully automated urinalysis system intended for in vitro qualitative or semi-quantitative determination of urine analytes, including pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and erythrocytes. In addition, the analyzer determines specific gravity, color, and clarity. The URISYS 2400 is intended for professional use only.
The primary functions of the URISYS 2400 analyzer include:
o Sample identification o Automatic dispensing of test strips o Robotic pipetting of samples o Controlled incubation period o Photometric measurements o Result memory o Optional formats for data output
Measuring principle
The URISYS 2400 analyzer uses the URISYS 2400 cassette containing 400 test strips. Each test strip has ten individual test pads that are used to test for different substances or characteristics. The test strips are analyzed as they move automatically through the analyzer. One strip is used per sample. When a strip is dispensed for use, an aliquot of the urine sample is pipetted onto each of the test pads. The test results are based on the measurement of reflected light intensity. The specific gravity, color, and clarity of each sample are also measured.
The URISYS 2400 is a fully automated reflectance photometer for in vitro semi-quantitative measurements of urine test strips. The light sources (LED’s [light-emitting diodes]) and reading times are optimized for the reaction chemistry and color development that occur on the test pads.
The measuring head of the URISYS 2400 contains LED’s of three different wavelengths. The test strip remains stationary at the measuring position as the measuring head moves over each test pad. Measurement starts at the reference plate, which is used to test the optical system.
The URISYS 2400 verifies that the test strip is properly positioned under the measuring head, by measuring the reflected light. If a test strip is not properly positioned under the measuring head, the URISYS 2400 displays a message.
Roche Diagnostics A-2 Operators Manual · Version 1.0
URISYS 2400 1 Introduction to the system
Description of the URISYS 2400 analyzer
Reflected light is measured electro-optically, and the process is illustrated in the following figure.
C
A
B
Optical transmission Electronic transmission
Figure A-1 Process of measuring
LED
A B Te st pad C Photodiode detector
An LED (
A) transmits light of a defined wavelength, at an optimum angle,
onto the surface of a test pad (
B). The light hits the surface of the test pad and
D
D Analog-to-digital converter E Microprocessor F Concentration result
E
F
is reflected with an intensity that is dependent on the color of the test pad.
A photodiode detector (
C), positioned directly above the test pad, receives the
reflected light. The photodiode detector transmits an analog electrical signal to the analog-to-digital converter (
D), which changes the analog signal to a
digital value.
A microprocessor (
E) adjusts the digital value, based on a value from an
internal reference plate. It converts the digital value to a relative value by using a calibration standard scale. Then, the microprocessor calculates a reflectance value.
The semi-quantitative concentration result (
F) is determined by comparing
the reflectance value with the so-called range limits (parameter-specific reflectance values stored in the analyzer). The full list of concentration range values for all test parameters is shown in:
Table F-1 on page F-5
The results are stored in the memory and can be printed, saved to a diskette, or sent to another computer.
Each of the three LEDs transmits light of a different wavelength. The use of different wavelengths improves the results that are obtained from measuring each of the test strip parameters. The wavelengths used to measure the different parameters are listed in the table below.
Roche Diagnostics Operators Manual · Version 1.0 A-3
1 Introduction to the system URISYS 2400
Description of the URISYS 2400 analyzer
Test parameter Measuring wavelength (nm)
pH 555, 620
Leukocytes 555
Nitrite 555
Protein 555
Glucose 555
Ketones 555
Urobilinogen 555
Bilirubin 555
Erythrocytes 555, 620
Color 470, 555, 620
Table A-1 Wavelengths used to measure the reflectance values
The photometric reflectance measurement for all of the test parameters is performed after an incubation time of 60 seconds. As with previous urine analyzers from Roche Diagnostics, compensation for intrinsic urinary color, which is a recognized interfering factor, is made through the measurement of a blank compensation pad on the test strip. The compensation pad assists in the prevention of false positives when a urine sample is strongly colored.
The URISYS 2400 determines the urine color by evaluating the reflectance values of all three measuring wavelengths (470 nm, 555 nm, and 620 nm) on the compensation pad. Color results are reported as PALE YELLOW, YELLOW, AMBER, BROWN, ORANGE, RED, GREEN, or OTHERS.
The measurement of specific gravity (SG) is performed in a flow cell that is filled with a urine sample during aspiration of the sample. An LED (650 nm) transmits light to the flow cell. A charge-coupled device (CCD) determines the angle of total reflection. The refractive index of the sample is calculated from the angle of total reflection using a calibration curve. Transformation into the SG value is performed by a nomogram stored in the analyzer.
The flow cell that is used to measure specific gravity also determines the clarity of the sample. However, light of a different LED is used to measure clarity. This LED (660 nm) transmits light to the flow cell, and the amount of light that passes through the flow cell is measured.
The system is blanked (zeroed) during the washing process, between sample measurements, by setting the water transmission to 100%. The transmission values of samples are transformed to clarity results (CLEAR, LIGHT TURBID or TURBID), by using the clarity range table that is programmed in the software of the URISYS 2400.
Roche Diagnostics A-4 Operators Manual · Version 1.0
URISYS 2400 1 Introduction to the system
Description of the URISYS 2400 analyzer
When a sample has been pipetted onto the test pads of a test strip, and the specific gravity and clarity have been measured, the remaining urine is ejected into the rinse station. The needle and the flow cell are washed with water before the next sample is taken.
Roche Diagnostics Operators Manual · Version 1.0 A-5
1 Introduction to the system URISYS 2400
Description of the URISYS 2400 analyzer
Roche Diagnostics A-6 Operators Manual · Version 1.0

System description

Part B is the system description of the URISYS 2400.
B
URISYS 2400 2 Hardware
Table of contents

Hardware

This chapter contains the description of the system hardware.
In this chapter B
Hardware description ........................................................................................2
Standard accessories .....................................................................................4
Additional items ............................................................................................5
Installation .........................................................................................................5
Basic requirements .......................................................................................5
Surface and space measurements ............................................................5
Power supply ............................................................................................6
Water supply .............................................................................................6
Assemble the parts .........................................................................................7
Set up the water supply tank ..................................................................7
Liquid waste tank ....................................................................................8
TS transfer base ........................................................................................8
External printer ............................................................................................9
External host connection ............................................................................9
Environmental conditions .........................................................................10
Chapter
2
Roche Diagnostics Operators Manual · Version 1.0 B-1
2 Hardware URISYS 2400

Hardware description

Hardware description
The URISYS 2400 analyzer consists of several major components:
o Rack transport system o Liquid handling system o Test strip cassette compartment o Automated test strip processing area o 3-wavelength reflectance photometer o Combined, flow-through refracto-/turbidimeter o User interface screen o Inbuilt computer o Battery-buffered memory o Diskette drive (for MS-DOS compatible diskettes)
Roche Diagnostics B-2 Operators Manual · Version 1.0
URISYS 2400 2 Hardware
Hardware description
C
B A
Figure B-1 The URISYS 2400 analyzer
Diskette drive
A B URISYS 2400 test strip cassette
compartment
C Touch screen D Pipette mechanism E Wash and sample syringes
D E
FGHIJ
F Rack output line G STAT sampling position H Rack cross-feed line I Barcode reader J Rack input line
Roche Diagnostics Operators Manual · Version 1.0 B-3
2 Hardware URISYS 2400
Hardware description
Standard accessories
The standard accessories that accompany the URISYS 2400 analyzer are:
o Water supply tank with a liquid level detector o Water supply tube with filter o Two rack trays o Liquid waste tube 5 m (16.4 ft) length o Strip waste container o Standard maintenance kit (O-rings, seals) o Fuses o Power cord o Three software installation diskettes o Operators Manual (English version)
Water supply tank with liquid
level detector
Rack trays Two identical trays are supplied with the analyzer. Each tray holds up to 15
Strip waste container Used test strips are deposited in a strip waste container that has a minimum
The water supply tank that supplies deionized water to the analyzer is located externally. It has a maximum capacity of 5 L (5.3 qts). This volume is sufficient to process approximately 1000 samples.
racks at one time. One tray is used to load the racks onto the rack input line. The second tray is used as a stationary rack output tray during sample
processing and serves to remove racks from the rack output line.
capacity of 420 test strips. (The capacity is calculated on the basis of the contents of one URISYS 2400 cassette plus a safety margin.)
The strip waste container is located on the right side of the analyzer, behind the door panel. The container is designed to minimize the contact an operator has with potentially infectious and toxic materials.
When the strip waste container is full, a message is displayed on the touch screen.
Roche Diagnostics B-4 Operators Manual · Version 1.0
URISYS 2400 2 Hardware

Installation

Additional items
Items that are necessary or optional for the installation and operation of the URISYS 2400 analyzer, but are not supplied with the analyzer, are:
o Liquid waste tank (with or without an optional overflow sensor) o Racks o Sample tubes o Wash in g solu tion o URISYS 2400 cassette o URISYS 2400 calibration strip o External printer and standard printer cable (parallel/Centronics interface) o Waste container cartons o Serial interface (host) cable o 1.44 MB diskette for storage of data
Installation
Basic requirements
Surface and space measurements
Read this manual carefully before installation, to ensure correct operation of the analyzer.
If you have any questions after initial installation, contact Technical Support.
Only trained Roche Technical Support personnel, or similarly qualified personnel supervised by authorized service agents of Roche Diagnostics, are qualified to install the URISYS 2400 analyzer. At least two persons must carry the analyzer, by holding the baseplates on the left side and right side of the analyzer. Be careful not to hurt your hands or fingers when putting the analyzer in place. Follow the specified installation conditions carefully. Otherwise, inaccurate results, or damage to the analyzer may occur.
The following surface and space requirements must be met for the use of the URISYS 2400 analyzer, and the proper ventilation of the analyzer when it is in use:
o 5 cm (2.0 inches) or more (rear) o 20 cm (7.9 inches) or more (left side) o 30 cm (11.8 inches) or more (right side)
Use a table that is able to support the mass of the analyzer, which is approximately 85 kg (187 lbs). The table must be level. Any inclination must be less than 1.5 degrees in every direction.
Roche Diagnostics Operators Manual · Version 1.0 B-5
2 Hardware URISYS 2400
Installation
Power supply A standard 230 V/50 Hz or 115 V/60 Hz power line is required to operate the
analyzer. It must be connected to a grounded power supply socket (protection class 1).
Although the maximum power consumption of the instrument is 200 VA, the external power supply should fulfill the following minimum requirements to cover for peaks at switching on.
Electric power supply capacity:
o Minimum 500 VA o Minimum 5 Amp, 3-pin power supply socket
The power switch and power supply socket are located on the right side of the analyzer.
AB
Figure B-2 Power switch and power supply socket
Power switch B Power supply socket
A
Water supply An external water supply tank for deionized water is supplied with the
URISYS 2400. The deionized water must be germ free and have an electric conductance value that is less than or equal to 1 µS/cm.
Do not connect the analyzer directly to a pressurized water supply.
Roche Diagnostics B-6 Operators Manual · Version 1.0
Loading...
+ 136 hidden pages