Roche Elecsys1010 Service manual

Elecsys 1010 Service Manual
Service ManualService Manual
Service Manual
Service ManualService Manual
Elecsys 1010Elecsys 1010
Elecsys 1010
Elecsys 1010Elecsys 1010
FINAL 2.1 - February 2000
Elecsys 1010 Service Manual
This manual is published by Roche Diagnostics GmbH,
Technical Product Management and Service Department.
© Roche Diagnostics GmbH All rights reserved.
No part of this publication may be reproduced without the expressed written permission of Roche Diagnostics GmbH.
Disclaimer
Roche Diagnostics GmbH makes no rep­resentation or warranties with respect to the contents of this documentation and specifically disclaims any implied warranties, including the implied warranties of merchantability and fitness for a particular purpose.
In no case shall Roche Diagnostics GmbH be liable for incidental or consequential damages.
FINAL 2.1 - February 2000
Elecsys 1010 Service Manual
How to Use this ManualHow to Use this Manual

How to Use this Manual

How to Use this ManualHow to Use this Manual
It may occur that identical components occur several times in Elecsys 1010 and perform different functions. Every component has an ID number in the figures.
Photo Coupler Assy, preadjusted (416)Photo Coupler Assy, preadjusted (416)
Photo Coupler Assy, preadjusted (416)
e.g.
Photo Coupler Assy, preadjusted (416)Photo Coupler Assy, preadjusted (416) Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
416
418
D-10
In Chapter 5 Electronics the components are referred to as functional units.
Photo Coupler Magnet Home (418)Photo Coupler Magnet Home (418)
Photo Coupler Magnet Home (418)
e.g .
Photo Coupler Magnet Home (418)Photo Coupler Magnet Home (418) Photo Coupler Magnet Cell (416)Photo Coupler Magnet Cell (416)
Photo Coupler Magnet Cell (416)
Photo Coupler Magnet Cell (416)Photo Coupler Magnet Cell (416)
FINAL 2.1 - February 2000
Elecsys 1010 Service Manual
In Chapter 4 Mechanics the part name is used. In addition, the information is contained whether the part can be ordered as a spare part ,or if it consists of several components.
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)
e.g.
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
Replaceable Components:
- Peltier Assy Preheater (302)
- Preheater Sipper (306)
- PCB Preamp High Voltage (310)
- Holder Photo Coupler MD1 (326) consisting of:
- Photo Coupler Holder (419)
- Photo Coupler Holder (327)
--
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
-
Photo Coupler Assy, preadjusted (418)
--
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
In Chapter 8.2 Part Identification shows where this part is used in Chapter 4 or 5.
Photo Coupler Assy, preadjusted (416)Photo Coupler Assy, preadjusted (416)
Photo Coupler Assy, preadjusted (416)
e.g.
Photo Coupler Assy, preadjusted (416)Photo Coupler Assy, preadjusted (416) Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
NoNo
Part Name EnglishPart Name English
No
Part Name English
NoNo
Part Name EnglishPart Name English
412 TUBING SIPPER (2) 0,8 Detection Module
416416
Photo Coupler Assy, preadjustedPhoto Coupler Assy, preadjusted
416
Photo Coupler Assy, preadjusted
416416
Photo Coupler Assy, preadjustedPhoto Coupler Assy, preadjusted
417 Eccentric Ring, 2 pcs Detection Module
418418
Photo Coupler Assy, preadjustedPhoto Coupler Assy, preadjusted
418
Photo Coupler Assy, preadjusted
418418
Photo Coupler Assy, preadjustedPhoto Coupler Assy, preadjusted
419 Photo Coupler Holder Detection Module 527 Cable M10 Detection Module
In Chapter 8.3 Spare Parts, components that cannot be ordered individually do not have an ID. The part name is followed by a cross reference to the spare part, in which the component is contained.
Photo Coupler Assy, preadjusted (416)Photo Coupler Assy, preadjusted (416)
Photo Coupler Assy, preadjusted (416)
e.g.
Photo Coupler Assy, preadjusted (416)Photo Coupler Assy, preadjusted (416) Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)
Photo Coupler Assy, preadjusted (418)Photo Coupler Assy, preadjusted (418)
NoNo
No
NoNo
IDID
ID
IDID
Part Name EnglishPart Name English
Part Name English
Part Name EnglishPart Name English
ElectronicsElectronics
Electronics
ElectronicsElectronics
BP 1BP 1
BP 1
BP 1BP 1 BP 1BP 1
BP 1
BP 1BP 1
MechanicsMechanics
Mechanics
MechanicsMechanics
Detection ModuleDetection Module
Detection Module
Detection ModuleDetection Module Detection ModuleDetection Module
Detection Module
Detection ModuleDetection Module
Spare PartSpare Part
Spare Part
Spare PartSpare Part
BTOBTO
BTO
BTOBTO
RR
AA
R
RR
ABCABC
A
ABC
AA
ABCABC
326 1234567890 HOLDER PHOTO COUPLER MD1 X A
416416
416
416416 418418
418
418418
419 Photo Coupler Holder, Part of 326 327 Photo Coupler Holder, Part of 326
FINAL 2.1 - February 2000
Photo Coupler Assy, preadjusted, Part of 326Photo Coupler Assy, preadjusted, Part of 326
Photo Coupler Assy, preadjusted, Part of 326
Photo Coupler Assy, preadjusted, Part of 326Photo Coupler Assy, preadjusted, Part of 326 Photo Coupler Assy, preadjusted, Part of 326Photo Coupler Assy, preadjusted, Part of 326
Photo Coupler Assy, preadjusted, Part of 326
Photo Coupler Assy, preadjusted, Part of 326Photo Coupler Assy, preadjusted, Part of 326
Elecsys 1010 Service Manual
11
Application / IntroductionApplication / Introduction
1
Application / Introduction
11
Application / IntroductionApplication / Introduction
22
InstallationInstallation
2
Installation
22
InstallationInstallation
33
FluidicsFluidics
3
Fluidics
33
FluidicsFluidics
44
MechanicsMechanics
4
Mechanics
44
MechanicsMechanics
55
ElectronicsElectronics
5
Electronics
55
ElectronicsElectronics
66
SoftwareSoftware
6
Software
66
SoftwareSoftware
77
Trouble ShootingTrouble Shooting
7
Trouble Shooting
77
Trouble ShootingTrouble Shooting
88
Spare PartsSpare Parts
8
Spare Parts
88
Spare PartsSpare Parts
1010
10
1010
99
Host InterfaceHost Interface
9
Host Interface
99
Host InterfaceHost Interface MaintenanceMaintenance
Maintenance
MaintenanceMaintenance AppendixAppendix
Appendix
AppendixAppendix
FINAL 2.1 - February 2000
Elecsys 1010 Service Manual
FINAL 2.1 - February 2000
Elecsys 1010 Service Manual
11
1
11
Application / IntroductionApplication / Introduction
Application / Introduction
Application / IntroductionApplication / Introduction
1.11.1
1.1
1.11.1
1.21.2
1.2
1.21.2
11
1
11
System DescriptionSystem Description
System Description
System DescriptionSystem Description
1.1.11.1.1
1.1.1
1.1.11.1.1
1.1.21.1.2
1.1.2
1.1.21.1.2
1.1.31.1.3
1.1.3
1.1.31.1.3
System SpecificationSystem Specification
System Specification
System SpecificationSystem Specification
.3.3
Operator Precautions andOperator Precautions and
.3
Operator Precautions and
.3.3
Operator Precautions andOperator Precautions and HazardsHazards
Hazards
HazardsHazards
1.3.11.3.1
1.3.1
1.3.11.3.1
Instrument IntroductionInstrument Introduction
Instrument Introduction
Instrument IntroductionInstrument Introduction Principles of the AnalysisPrinciples of the Analysis
Principles of the Analysis
Principles of the AnalysisPrinciples of the Analysis Pipetting SchemePipetting Scheme
Pipetting Scheme
Pipetting SchemePipetting Scheme
1.1.3.11.1.3.1
1.1.3.1
1.1.3.11.1.3.1
Contact PersonsContact Persons
Contact Persons
Contact PersonsContact Persons
Timing DiagramTiming Diagram
Timing Diagram
Timing DiagramTiming Diagram
__________________________________________________________________________________________
_____________________________________________
__________________________________________________________________________________________
________________________________________________________________________________
________________________________________
________________________________________________________________________________
______________________________________________________________________________________________
_______________________________________________
______________________________________________________________________________________________
__________________________________________________________________________
_____________________________________
__________________________________________________________________________
____________________________________________________________________________________________
______________________________________________
____________________________________________________________________________________________
____________________________________________________________________________________
__________________________________________
____________________________________________________________________________________
________________________________________________________________________________
________________________________________
________________________________________________________________________________
__________________________________________________________________________________
_________________________________________
__________________________________________________________________________________
1515
15
1515
1818
18
1818
2020
20
2020
2020
20
2020
11
1
11
11
1
11 55
5
55 66
6
66
1.41.4
1.4
1.41.4
1.3.21.3.2
1.3.2
1.3.21.3.2
1.3.31.3.3
1.3.3
1.3.31.3.3
Service ConceptService Concept
Service Concept
Service ConceptService Concept
1.4.11.4.1
1.4.1
1.4.11.4.1
1.4.21.4.2
1.4.2
1.4.21.4.2
Proved SecurityProved Security
Proved Security
Proved SecurityProved Security Precautions and HazardsPrecautions and Hazards
Precautions and Hazards
Precautions and HazardsPrecautions and Hazards
Service LevelService Level
Service Level
Service LevelService Level RA-Procedure (Replacement of defective parts)RA-Procedure (Replacement of defective parts)
RA-Procedure (Replacement of defective parts)
RA-Procedure (Replacement of defective parts)RA-Procedure (Replacement of defective parts)
______________________________________________________________________________________________
_______________________________________________
______________________________________________________________________________________________
______________________________________________________________________________
_______________________________________
______________________________________________________________________________
__________________________________________________________________________________
_________________________________________
__________________________________________________________________________________
__________________________________________________________________________________________________
_________________________________________________
__________________________________________________________________________________________________
________________________________________
____________________
________________________________________
2020
20
2020 2020
20
2020
2222
22
2222
2222
22
2222 2222
22
2222
FINAL 2.1 - February 2000 Chapter 1
Elecsys 1010 Service Manual
FINAL 2.1 - February 2000 Chapter 1
Elecsys 1010 Service Manual
1.11.1
1.1
1.11.1
1.1.11.1.1
1.1.1
1.1.11.1.1
System DescriptionSystem Description

System Description

System DescriptionSystem Description
Instrument IntroductionInstrument Introduction

Instrument Introduction

Instrument IntroductionInstrument Introduction
F-9F-9
F-9
F-9F-9
The sequence of operation of Elecsys 1010 is as follows:
Pick-up of sample and reagents
Incubation
Pick-up of incubate and addition of wash and buffer solutions
Measurement of luminescence, data evaluation and output
S/R rotor MFA s/r S/R dilutor Clot detection
Incubation rotor Temperature (incubator)
Bottle temperature MFA assy (sipper)
Sipper dilutor
Measuring cell Potentiostat PMT
Temperature
FINAL 2.1 - February 2000
Chapter 1Page 1
Elecsys 1010 Service Manual
IntroductionIntroduction
Introduction
IntroductionIntroduction
When the instrument cover top is closed, loading of two This System Description explains functions and their interrelationship.
Short Description of the InstrumentShort Description of the Instrument
Short Description of the Instrument
Short Description of the InstrumentShort Description of the Instrument The Elecsys 1010 is a Multi-Batch-Analyzer which analyzes
immunological tests according to the principle of electro­chemical luminescence (ECL). This measuring system is also called ORIGEN® technology. The emitted photons are measured with a photomultiplier (photon counts).
STAT (short-turn-around-time) samples is possible outside
of the direct range of action of the MFA arms and rotors. Pull
out the control panel like a drawer, behind which the holder
for the above mentioned STAT samples is located. By
pushing the control panel back in, the STAT samples are
within the direct range of the pipette needle under the closed
instrument cover without mechanically or logically
interrupting an ongoing process.
Elecsys 1010 has two driving blocks for the sample/
reagent rotor and the incubation rotor. Elecsys 1010 has all the modules and components
necessary for executing the entire analysis including the measurements and evaluation. When operating the instrument (e.g. for test selection and sample placement), the user is guided by menus on the keyboard. Softkeys and function keys can be used for interacting with the instrument. Samples (both in primary tubes and in secondary cups), reagents and calibrators are to be positioned on the sample/ reagent (s/r) rotor. A special reagent bottle combination (RackPack) contains the reagents needed for a test. Each RackPack has a 2-dimensional bar code with all data specific of the test. Positive user identification (PSID) is possible with samples in primary tubes. With samples in secondary cups, patients are identified manually by using the keyboard. After reading the bar code, the RUN is started and performed automatically.
CalibrationCalibration
Calibration
CalibrationCalibration
At the beginning of a RUN a calibration can be performed. The system imposes a full calibration
- if a new reagent (with a new batch number) is put in.
- at regular time intervals (specific of the reagent) This is done by calibrators specific of the test (normally 2 levels).
Quality ControlQuality Control
Quality Control
Quality ControlQuality Control
The instrument software contains all programs necessary for the quality control. Data input, management and security is done by using the keyboard.
Instrument AssemblyInstrument Assembly
Instrument Assembly
Instrument AssemblyInstrument Assembly ELECSYS 1010 is a desk top model and is suitable for a
standard laboratory bench. The instrument can be loaded from the front. The instrument cover top must be closed during the RUN.
The sample/reagent rotor is divided into 6 segments. The
two outer trays can hold samples in so-called primary tubes
of various diameters and heights. Control and calibrator
liquids are normally also positioned in these trays. A
maximum of 6 RackPacks with 2 reagents and one bead
container each are arranged radially between the sample
segments. The two inner trays are provided for secondary
cups. The labels on the primary tubes on the two outer trays
and the RackPacks can be read by a bar-code reader
(BCR).
The rotor assy is equipped with toothed disks and LEDs for
controlling and checking the rotor functions. The sample/
reagent rotor is removable.
The incubation rotor is an aluminum block heated to 37°C.
It can hold 4 identical segments with 32 incubation vessels
each. These incubation vessels are small conical plastic
tubes that hold the incubation material. When preparing the
RUN, the user loads the incubation rotor with the 4 segments.
The incubation rotor is not removable.
Detection ModuleDetection Module
Detection Module
Detection ModuleDetection Module
The ECL measuring cell is located in the detection module.
The photomultiplier (PMT), which is shielded from
electromagnetic radiation by a special housing, as well as
the holder for the pre-amplifier and the high-voltage module
are located above the measuring cell. The magnet with the
drive for the bead capturing is under the measuring cell. The
magnet is moved vertically from below to the bottom of the
measuring cell by a motor/spindle unit. This internal
assembly is incorporated in a housing, which serves as an
optical, electrical and thermal shield. The optimum
temperature for the process of ORIGEN
®
technology is 28°C. Therefore, the temperature of the entire detection module is set at 28°C.
MFA AssysMFA Assys
MFA Assys
MFA AssysMFA Assys
The instrument is designed to be serviced exclusively from the front, i.e. all components can be taken out and exchanged from the front.
The functional elements (dilutors, rotors, wash stations, bottle temperature assys) have covers that can be easily removed for service. The keyboard is located outside of the closed instrument cover top.
FINAL 2.1 - February 2000
The MFA assys transport the a) sample/reagent from the sample/reagent rotor to
the incubation vessels and
b) incubation material as well as ProCell and CleanCell
to the measuring cell.
The ProCell is a liquid required for generating the ECL signal. CleanCell is an auxiliary liquid used for cleaning the measuring cell and preparing it for the next measurement.
Page 2
Chapter 1
Elecsys 1010 Service Manual
All components involved in the sample/reagent processes are cleaned with system water.
Arms are mounted on the MFAs. The sample arm holds the sample needle and the bead mixer. The sipper arm holds the sipper needle. The LLD electronic system and the contact sensory system (crash detection) are also located in the arms. The assembly of both MFAs is identical except for the arm adapters.
The contact sensory system consists of mechanical and electronic hardware components. It records the contact of the needle and the bead mixer with a solid surface, e.g. the bottom of the container or a closed RackPack.
Rotation and lifting are controlled by toothed disks / vanes and photo couplers. Horizontal and vertical movements are normally performed in succession.
Bottle TemperatureBottle Temperature
Bottle Temperature
Bottle TemperatureBottle Temperature Elecsys 1010 has 2 bottle temperature systems, one for
ProCell bottle and one for CleanCell bottle. Bottles are refrigerated to approx. 8°C and will reach the desired temperature of 28°C after approx. 1hr. The actual temperature is controlled by NTC resistors in the housing walls.
by two membrane pumps. Another membrane pump is located on the back of the s/r dilutor. This pump has the following functions:
1. Fill the s/r system with system water.
2. Rinse s/r tubing and needle after pipetting.
ValvesValves
Valves
ValvesValves
The flow of system water to the wash stations for bead mixer, s/r needle and sipper needle is controlled by four valves (2-way). The water discharge from the wash stations is controlled by four valves (3-way). Various liquid levels in the wash stations are possible by setting the 3-way valves at different positions. Another valve controls the flow of the measuring material to the measuring cell.
Wash StationsWash Stations
Wash Stations
Wash StationsWash Stations
Elecsys 1010 has four wash stations. There are two wash stations for the s/r needle and one wash station each for the sipper needle and the bead mixer. The wash stations use system water.
LLDLLD
LLD
LLDLLD
DilutorsDilutors
Dilutors
DilutorsDilutors Elecsys 1010 has 2 dilutors. These are glass cylinders in
which pistons are precisely moved at a defined number of increments. When going down, liquids are aspirated. When going up, an exact amount of liquid is discharged by displacement. The direction of the liquid flow can be controlled by reversing valves at the entrance of the dilutors. The dilutors are moved by stepping motors.
The two dilutors have different functions: The needle of the s/r dilutor aspirates the sample, reagents
and microparticles in a defined sequence. The media are separated from the system liquid (H The aspirated material is discharged into the incubation
O) by an air bubble.
2
vessels through the sample needle. After the sample/ reagent process, the needle is internally cleaned with system water by means of the rinsing pump.
The sipper dilutor is located behind the measuring cell and pulls the incubate, ProCell and CleanCell in a defined sequence through the measuring cell. Then the measurement is performed. After that the needle is internally rinsed with CleanCell. The needle is externally cleaned with system water.
PumpsPumps
Pumps
PumpsPumps
The s/r and the sipper needle have one liquid level detection unit (LLD) each. The s/r-LLD detects samples, reagents, incubate and wash water. The sipper LLD detects incubate, ProCell, CleanCell and wash water.
Clot DetectionClot Detection
Clot Detection
Clot DetectionClot Detection On the sample/reagent side, there is a risk of (blood) clot
formation in the samples. This can be caused by pollution and/or belated clotting of the samples. Changes in the vacuum profile when picking up the samples are identified as clots by means of a differential pressure converter. The appropriate cleaning steps will be taken.
PMTPMT
PMT
PMTPMT The photomultiplier tube (PMT) receives and amplifies the
luminescence signal and converts it to an equivalent voltage. The PMT window is located at a defined distance vertically above the cell work electrode. High-voltage power supply is required for amplification.
High-VoltageHigh-Voltage
High-Voltage
High-VoltageHigh-Voltage High-voltage is used for amplifying the luminescence signal
generated in the measuring cell. The control range spans from 500 V to1100 V.
The instrument has 4 built-in pumps. The waste pump, a peristaltic pump, pumps the measuring material and system liquids into the waste container. The wash stations are fed
FINAL 2.1 - February 2000
PotentiostatPotentiostat
Potentiostat
PotentiostatPotentiostat The potentiostat controls the electrochemical redox process
in the cell.
Chapter 1Page 3
Elecsys 1010 Service Manual
Measuring CellMeasuring Cell
Measuring Cell
Measuring CellMeasuring Cell
Power Supply AssyPower Supply Assy
Power Supply Assy
Power Supply AssyPower Supply Assy
The cell has three electrodes: the work electrode (on which the microparticles are deposited), the counter electrode and the reference electrode. For the measurement, a constant voltage is applied between work electrode and counter electrode.
It is equipped with an input for input voltages of approx. 80
- 260 volts and 50-60 Hz. The fuses are not accessible to the customer and can only be exchanged by service personnel.
Rotor / Dilutor / MFA DrivesRotor / Dilutor / MFA Drives
Rotor / Dilutor / MFA Drives
Rotor / Dilutor / MFA DrivesRotor / Dilutor / MFA Drives The rotors, dilutor pistons and the MFAs are moved by
stepping motors. In order to avoid the loss of steps, the motions are controlled by photo couplers and toothed disks. Moreover, forward/backward motions can be recognized by two photo couplers placed at an angle of 90°.
TemperaturesTemperatures
Temperatures
TemperaturesTemperatures Elecsys 1010 has 4 systems for setting the temperatures
of liquids. The luminescence liquid is measured at a temperature of
28°C. The temperatures of the incubate picked up by the incubator and the liquids used for the measurements (CleanCell and ProCell) are preset inside the detection module. The immunological reaction occurs at 37°C (incubator temperature).
PCPC
PC
PCPC The PC is a standard PC assembly (CPU 486 DX 2,
66MHz, 8MB RAM) with Roche-owned functions. An 1.44 MB floppy disk drive serves as drive a. The system can be started and reloaded from disk drive a:. A flash-E
2
PROM serves as a hard disk replacement. The general program and changeable parameters are filed here. The LC display is controlled by the VGA controller.
Barcode Reader / BCRBarcode Reader / BCR
Barcode Reader / BCR
Barcode Reader / BCRBarcode Reader / BCR A bar-code reader (BCR) is attached to the sample/reagent
rotor. It reads and decodes primary sample tubes (on the 1st and 2nd tray on the rotor) and reagent RackPacks whenever the rotor passes the reader and transfers the data to the PC by a built-in RS 232 interface. The reader decodes 2-dimensional (PDF 417) and all standard bar codes.
Software Distribution /ProcessorsSoftware Distribution /Processors
Software Distribution /Processors
Software Distribution /ProcessorsSoftware Distribution /Processors
The entire Elecsys 1010 software is distributed over a total of 8 processors. The master board has a 486 processor. The measuring, sipper, temperature, rotor and s/r modules have an 80196 processor. The incubation module and the clot detection are controlled by an 8051derivative. The keyboard assy and the printer have one CPU each.
Keyboard AssyKeyboard Assy
Keyboard Assy
Keyboard AssyKeyboard Assy The keyboard assy consists of the keyboard, the keyboard
controller, an acoustic signal transmitter, the LC display and some LEDs. The keys are short-stroke touch contact keys. An additional keyboard can be connected to the bottom side of the keyboard assy by using a 5-pin plug (standard keyboard plug). One non-standard function is a signal transmitter (piezo) that can be activated on several frequencies. The display is a 640 x 480 dot LC display. In order to prevent reflections when reading the display in the keyboard assy drawer, it is possible to incline it.
PrinterPrinter
Printer
PrinterPrinter Elecsys 1010 has a thermoprinter with a printing width of
640 pixels. This corresponds to the display resolution. An immediate hard copy output is possible.
FINAL 2.1 - February 2000
Page 4
Chapter 1
Elecsys 1010 Service Manual
1.1.21.1.2
1.1.2
1.1.21.1.2
Principles of the AnalysisPrinciples of the Analysis

Principles of the Analysis

Principles of the AnalysisPrinciples of the Analysis
See Chapter 5 of the Operator's Manual for details of the test principles used on Elecsys
1010.
FINAL 2.1 - February 2000
Chapter 1Page 5
Elecsys 1010 Service Manual
1.1.31.1.3
1.1.3
1.1.31.1.3
Run This chapter describes processes going on in different
tests, e g. the placement of STAT samples during the run, the automatic rerun and the system preparation after start, as well as system evaluation at the end of a run.
Assay Protocols There are 24 independent assay protocols used for
determining the pipetting scheme of tests. Further protocols (no. 24 to no. 26) are used for checking the instrument.
The assay protocols are stored in the system (max. 64) and can only be modified and/or extended by a software update.
In each test a minimum of four liquids, the sample itself, reagent 1, reagent 2 and the microparticles, are processed. The following abbreviations will be used for the liquids:
MP: Microparticles B: Microparticles (test RackPack, assay RackPack) R0: Diluent ("all-purpose diluent", special RackPack) RM: Pretreatment solution for IgM or IgG tests
PS: Prereaction solution (pretreatment solution in RI: Reagent 1 (assay RackPack )
R2: Reagent 2 (assay RackPack ) S: Undiluted sample/ calibrator/ check DL: Diluted sample/ calibrator/ check D: Aspiration into the measuring cell and optical
i: Incubation time
Pipetting SchemePipetting Scheme

Pipetting Scheme

Pipetting SchemePipetting Scheme
(in special RackPack) special RackPack)
detection
5. Aspirate R2 --> discharge R2 into the third incubation vessel.
6. Wait until incubation time i2 is over.
7. Measurement.
One step is taken after the other, allowing for the incubation time. The last step is always the measurement.
The processing of these liquids is determined by the four basic protocols 0 through 3. They define which liquids are processed in which pipetting step. On the basis of the basic protocols, the processing with pretreatment liquids (RO, RM or PS) is defined in the protocols 4 through 23. Read the sequence of pipetting steps within one protocol from left to right. The example shows the following sequence:
1. Aspirate R0 and sample S --> discharge both liquids
2. Aspirate R0 and the diluted sample DL from the first
3. Aspirate microparticles, R1 and the diluted sample
4. Wait until the incubation time i1 is over.
FINAL 2.1 - February 2000
Protocol 9 (dilution with 2 different assays)Protocol 9 (dilution with 2 different assays)
Protocol 9 (dilution with 2 different assays)
Protocol 9 (dilution with 2 different assays)Protocol 9 (dilution with 2 different assays)
into the first incubation container.
incubation vessel --> discharge both liquids into the second incubation vessel.
DL from the second incubation vessel --> discharge all three liquids into the third incubation vessel.
Page 6
Chapter 1
Elecsys 1010 Service Manual
Assay ProtocolsAssay Protocols
Assay Protocols
Assay ProtocolsAssay Protocols TableTable
Table
TableTable
No. Pipetting Step 0 Inc 0
Pipetting Step
1
Inc 1
Pipetting Step
2
Inc 2 Det.
0 B R1 R2 S i1 D X
1 B R1 S i1 R2 i2 D X
2 R1 R2 S i1 B i2 D X
3 R1 S i1 B R2 i2 D X
4 R0 S B R1 R2 DL i1 D
5 R0 S B R1 DL i1 R2 i2 D
6 R0 S R1 R2 DL i1 B i2 D
7 R0 S R1 DL i1 B R2 i2 D
8 R0-> DL1 R0 B R1 R2 DL i1 D
9 R0 S-> DL1 R0 B R1 DL i1 R2 i2 D
10 R0 S-> DL1 R0 R1 R2 DL i1 B i2 D
11 R0 S-> DL1 R0 R1 DL i1 B R2 i2 D
12 PS S i0 B R1 R2 i1 D
13 PS S i0 B R1 i1 R2 i2 D
14 PS S i0 R1 R2 i1 B i2 D
Main assay Remarks
15 PS S i0 R1 i1 B R2 i2 D
16 RM S i0 B R1 R2 DL i1 D
17 RM S i0 B R1 DL i1 R2 i2 D
18 RM S i0 R1 R2 DL i1 B i2 D
19 RM S i0 R1 DL i1 B R2 i2 D
20 RM S-> DL1 RM i0 B R1 R2 DL i1 D
21 RM S-> DL1 RM i0 B R1 DL i1 R2 i2 D
22 RM S-> DL1 RM i0 R1 R2 DL i1 B i2 D
23 RM S-> DL1 RM i0 R1 DL i1 B R2 i2 D
24 R1 R1 D X
25 R1 R2 D X
26 R2 R2 D X
27 R1 B i1 D X
28
29...
... Reserve
only needed
for instrument
checks,
not part of
the RUN
...63
E-Tab1: Assay Protocols
FINAL 2.1 - February 2000
Chapter 1Page 7
Elecsys 1010 Service Manual
Pipetting StepsPipetting Steps
Pipetting Steps
Pipetting StepsPipetting Steps The processing of the assay protocols is divided into four
pipetting steps: pipetting step 0, the so-called pretreatment step, pipetting steps1 and 2, and the measuring step. The pipetting steps are divided into partial steps (e.g. pipette R1). The sequence of the partial steps is (regarding the liquid pick­up by the s/r needle): R1, R2, B, S.
The pipetting steps / partial steps according to the assay protocols are described below. The pick-up sequence during pretreatment is always: reagent before sample (in both steps).
For each pipetting step and/or measuring step, a generally applicable step can be described. Figures 1 to 3 give an overview of these pipetting steps.
Pipetting Step 0Pipetting Step 0
Pipetting Step 0
Pipetting Step 0Pipetting Step 0 In pipetting step 0 the sample is pretreated. It has one or two
increments. In the figure, these increments are called "Block 1 and 2".
- Block 1 is passed through for protocols 4 through 23.
- Block 2 is passed through for protocols 8 through 11 and 20 through 23.
Each block consists of
- aspiration of the separation air bubble
- aspiration of the pretreatment liquid from the sample rotor
- washing in wash station 1
- aspiration of the sample
- washing and rinsing in wash station 1 or rinsing in wash station 2, rehomogenizing of the microparticles and washing
Rinsing and washing in wash station 2 with simultaneous rehomogenizing of the microparticles can only be performed one time, always at the end of a pipetting step.
The difference between block 1 and block 2 is the aspiration of the diluted sample from the incubation rotor.
- In block 1, the sample is aspirated from the sample rotor, and the samples are discharged into the first incubation vessel.
- In block 2, the sample is aspirated from the first
incubation vessel, and the liquids are discharged into the second incubation vessel.
FINAL 2.1 - February 2000
Page 8
Chapter 1
Elecsys 1010 Service Manual
Pipetting Step 0Pipetting Step 0
Pipetting Step 0
Pipetting Step 0Pipetting Step 0
Sample rotor Incubation rotor
Block 1Block 1
Block 1
Block 1Block 1
Aspirate
Aspirate
Block 2Block 2
Block 2
Block 2Block 2
Aspirate
Wash and rinse in
wash station 1
Aspirate separation air bubble
R0 or PS or RM
Aspirate sample by using
LLD
Wash in wash station 1
S
Aspirate sample by using
LLD
Discharge
Aspirate separation air bubble
R0 or RM
Aspirate pretreatment
liquid by using LLD
Discharge
Wash and rinse in
wash station 2 and
rehomogenize microparticles
Wash and rinse
in wash station 1
Figure 1: Pipetting step 0
FINAL 2.1 - February 2000
Wash in wash station 1
DL
Aspirate pretreated sample
by using LLD
Discharge
Aspirate
Discharge
Wash and rinse in
wash station 2 and
rehomogenize microparticles
Chapter 1Page 9
Elecsys 1010 Service Manual
Pipetting Step 1Pipetting Step 1
Pipetting Step 1
Pipetting Step 1Pipetting Step 1 Pipetting step 1 is also divided into blocks and always
starts with block 1.
- Aspirate the separation air bubble.
In blocks 2, 3 and 4, reagents are aspirated. All three blocks are passed through in the following sequence:
- Aspiration of R1, R2 or microparticles.
- Discharging of fluids after the last reagent has been picked up and the sample pretreated with PS. The sample pretreated with PS will be further processed in block 6.
- Washing in wash station 1.
The assay protocols determine which and how many of the blocks will be passed through in pipetting step 1.
In block 5, the sample is aspirated. The pretreated sample DL is
- aspirated from the first incubation vessel and is discharged into a second incubation vessel in block 6, if it has been pretreated with R0 or RM in one increment (see pipetting step 0).
- aspirated from the second incubation vessel and is discharged into a third incubation vessel in block 6, if it has been pretreated with R0 or RM in two increments (see pipetting step 0).
If sample S has not been pretreated, it will be aspirated from the sample rotor and discharged into the first incubation vessel in block 6.
In block 6, the liquids are discharged into the incubation rotor, and the s/r needle is washed and rinsed. If the sample has been pretreated with PS, the aspirated liquids will be added to the first incubation vessel (see pipetting step 0). The rehomogenizing of the microparticles and the was­hing /rinsing in wash station 2 will be performed if in the next cycle microparticles are pipetted in one pipetting step 1.
FINAL 2.1 - February 2000
Page 10
Chapter 1
Elecsys 1010 Service Manual
Pipetting Step 1Pipetting Step 1
Pipetting Step 1
Pipetting Step 1Pipetting Step 1
Sample rotor Incubation rotor
Block 1Block 1
Block 1
Aspirate
Aspirate
Block 1Block 1
Block 2Block 2
Block 2
Block 2Block 2
Block 3Block 3
Block 3
Block 3Block 3
Aspirate separation air bubble
R1
Aspirate reagent 1
by using LLD
Wash in wash station 1
R2
Aspirate reagent 2
by using LLD
Aspirate
Block 5Block 5
Block 5
Block 5Block 5
Aspirate
Block 6Block 6
Block 6
Block 6Block 6
Block 4Block 4
Block 4
Block 4Block 4
Aspirate sample by
using LLD (clot and
foam detection)
Wash and rinse in wash station 1
Wash in wash station 1
Microparticles
Aspirate microparticles
by using LLD
Wash in wash station 1
Discharge
Wash and rinse in wash
station 2 and rehomogenize
DLS
Aspirate pretreated sample by using LLD
microparticles
Aspirate
Discharge sample pretreated with PS
Figure 2: Pipetting step 1
FINAL 2.1 - February 2000
Chapter 1Page 11
Elecsys 1010 Service Manual
Pipetting Step 2Pipetting Step 2
Pipetting Step 2
Pipetting Step 2Pipetting Step 2 Pipetting step 2 is divided into 4 blocks. Like pipetting
step 1, it always starts with block 1:
- Aspirate the separation air bubble.
In blocks 2 and 3, reagents are aspirated. Both blocks are passed through in the following sequence:
- Aspirate R2 or microparticles.
- Discharge liquid into the incubation rotor after the last reagent has been picked up.
- Wash in wash station 1.
The assay protocols determine which and how many of the blocks will be passed through in pipetting step 2.
In block 4, the liquids are discharged into the incubation rotor, and the s/r needle is washed and rinsed. The liquids will be
- added to the first incubation vessel, if the sample has not been pretreated or pretreated with PS.
- added to the second incubation vessel, if the sample has been pretreated with R0 or RM in one increment.
- added to the third incubation vessel, if the sample has been pretreated with R0 or RM in two increments.
The rehomogenizing of the microparticles and the was­hing /rinsing in wash station 2 will be performed if in the next cycle microparticles are pipetted in one pipetting step 1.
FINAL 2.1 - February 2000
Page 12
Chapter 1
Elecsys 1010 Service Manual
Pipetting Step 2Pipetting Step 2
Pipetting Step 2
Pipetting Step 2Pipetting Step 2
Sample rotor Incubation rotor
Block 1Block 1
Block 1
Aspirate
Block 1Block 1
Block 2Block 2
Block 2
Block 2Block 2
Aspirate separation air bubble
R2
Aspirate reagent 2
by using LLD
Wash in wash station 1
Aspirate
Block 4Block 4
Block 4
Block 4Block 4
Block 3Block 3
Block 3
Block 3Block 3
Wash and rinsei n wash station 1
Microparticles
Aspirate microparticles by
using LLD
Wash in wash station 1
Discharge
wachen und spülen in
Wash and rinse in
wash station 2 and
rehomogenize microparticles
Figure 3: Pipetting step 2
FINAL 2.1 - February 2000
Chapter 1Page 13
Elecsys 1010 Service Manual
Measuring StepMeasuring Step
Measuring Step
Measuring StepMeasuring Step Measuring step actions take place around the measuring
cell, e.g. in the incubator, sipper needle, measuring cell or in the sipper system dilutor. The measuring step has the following increments:
- Pick-up of the incubate ( - sampling, pick -up ) (pre­conditioning with ProCell included).
- Transport of the incubate with ProCell (pre­conditioning with ProCell included).
- Capturing with ProCell
- Washing with ProCell
- Measuring -> optical measurement
- Pick-up of CleanCell solution for cleaning 1
- Pick-up of air and CleanCell alternately
- Solution for cleaning 2
- Pick-up of ProCell for cleaning 3
- Pick-up of ProCell for post -conditioning
FINAL 2.1 - February 2000
Page 14
Chapter 1
Elecsys 1010 Service Manual
1.1.3.11.1.3.1
1.1.3.1
1.1.3.11.1.3.1
Timing DiagramTiming Diagram
Timing Diagram
Timing DiagramTiming Diagram
FINAL 2.1 - February 2000
diag1.tif
Chapter 1Page 15
Elecsys 1010 Service Manual
FINAL 2.1 - February 2000
Page 16
diag2.tif
Chapter 1
Elecsys 1010 Service Manual
FINAL 2.1 - February 2000
diag3.tif
Chapter 1Page 17
Elecsys 1010 Service Manual
1.21.2
1.2
1.21.2
Analyzer dimensions Height: 620 (mm)
Electrical Supply voltage 100 - 240 (V) AC at 50/60 (Hz)
Environmental conditions Temperature 18 (°C) to 32 (°C)
System SpecificationSystem Specification

System Specification

System SpecificationSystem Specification
Depth: 780 (mm)
Width: 960 (mm)
Weight approx. 110 (kg)
Power consumption max 0.61 (kVA)
Fuse circuit 16 (A) time-lag at 230 (V) AC
Heat generation
Ambient temperature note change during one batch run
approx. 2000 (kJ/h) resp. 481 (kcal/h) resp. 1912 (BTU/h)
max. ± 3 (°C)
rel. humidity 20 (%) to 85 (%) without condensation
Noise level according to DIN 43635
Water supply Water tank approx. 4,2 (l)
Waste water connection
Throughput ECL measurements approx. 50 (tests/h)
Samples Sample volume per test 10 (µl) to 50 (µl)
Continuous operation max. 60 (dBA)
Peak level max. 65 (dBA)
Water quality (conductivity) sub-micron filtered <10 µS/cm
Water consumption approx. 3 (l) for 128 test
Waste container for measuring material, system water, ProCell, CleanCell
Sample detection
Sample rotor for routine samples, calibrators and controls
Usable sample bar codes
approx. 5 (l)
by Liquid Level Sensor and Clot Detector of the sample / reagent needle
42 positions for primary tubes, 18 positions for secondary cups
2 positions for STAT samples
NW7 (Codabar), Code 39, Code 128, Interleave 2 of 5
Reagent Reagent capacity 6 reagent channels
by Liquid Level Sensor of the sample/ reagent needle
E-Tab2
FINAL 2.1 - February 2000
Reagent detection
Page 18
Chapter 1
Elecsys 1010 Service Manual
Bottle volume for CleanCell and ProCell 380 (ml)
Reagent ID by 2-dim. bar code (PDF 417)
Incubator Incubator capacity 128 plastic assay cups (4 x 32)
Volume of assay cups 200 (µl)
ambient temperature 18°C to 20°C
Incubation temperature
Incubation time 9 / 18 (min)
37°C + 0.3°C - 0.8°C ambient temperature 20°C to 32°C 37°C + 0.3°C - 0.3°C
Measuring system Measuring method
Calibration method 2 point calibration
PC Disk drive Floppy disk 3,5" / 1.44 MB
Interfaces
LCD
Thermoprinter Paper width 110 (mm)
E-Tab3
Integral measurement of luminescence signal
1 Centronics for printer, 1 RS 232 for laboratory EDP
Black-and-white VGA LCD with 640 x 480 pixels
FINAL 2.1 - February 2000
Chapter 1Page 19
Elecsys 1010 Service Manual
1.3.3 Precautions and Hazards1.3.3 Precautions and Hazards

1.3.3 Precautions and Hazards

1.31.3
1.3
1.31.3
The data and information provided in this manual correspond to the state of knowledge existing at the time of introducing the Elecsys 1010 onto the market. Any important changes will be taken into account in the next edition of this manual.
The respective packaging leaflet should be regarded as authoritative.
Operator Precautions andOperator Precautions and
Operator Precautions and

Operator Precautions andOperator Precautions and HazardsHazards

Hazards
HazardsHazards
1.3.3 Precautions and Hazards1.3.3 Precautions and Hazards
All electrical equipment is potentially hazardous. Never remove covers without first ensuring that it is isolated from the AC supply, unless specific maintenance instructions or repairs are being carried out by authorized Roche Diagnostics personnel.
The hard- and software is subject to a program of continuous evaluation and improvement and, therefore, may be changed in the future. This also concerns service requirements.
This service manual was created for the telephone and technical service staff.
1.3.1 Contact Persons1.3.1 Contact Persons

1.3.1 Contact Persons

1.3.1 Contact Persons1.3.1 Contact Persons
Technical Product Management and Support Department: LI-TT
Phone: 0621 / 759 / 3227
8802
Fax No: 0621/ 759 / 4591
Logistics Hotline (RA):
Phone: 0621 / 759 / 4389 Fax No.: 0621 / 759 / 4613
8093
When calling from abroad, drop the "0" and add the country code (+ 49).
1.3.2 Proved Security1.3.2 Proved Security

1.3.2 Proved Security

1.3.2 Proved Security1.3.2 Proved Security
All samples and reagents should be treated with caution accorded to those known to contain pathogenic organisms. Similarly, the cleaning of component part of Elecsys 1010 should be done with respect to human health.
Warning:Warning:
Warning:
Warning:Warning: All components used on Elecsys 1010 must be regarded
as potentially dangerously contaminated when doing repair work. Use rubber gloves or double gloves whenever cleaning or sterilizing components. the most frequently used components (tubings, cuvette, measuring chamber flap, sip. lever and waste container) should be cleaned and treated with a suitable disinfecting solution (75% alcohol) prior to doing any service work.
Disinfect and wash hands after work is completed. This instrument has been constructed and checked in
accordance with Standards IEC 1010. When the instrument leaves our factory it is in perfect order from the point of view of work safety. To keep it that way and to ensure safe operation the user must follow the instructions and warnings given in the operation instruction manual.
The electrical protection of the apparatus to Class I (it has a protective earth).
This instrument has been constructed according to the regulations of DIN IEC 601, part: "Security of Electromedical Instruments: General Guidelines" and left our factory in perfect condition regarding work safety.
The instrument obtained the GS mark for proved security from the TÜV.
In order to ensure safe operation the user must follow the instructions and warnings given in the instruction manual.
This instruments meets protection 1 (earth conductor wire).
FINAL 2.1 - February 2000
The power plug must only be inserted into a socket that has a protective contact. The protection must not be abolished by using an extension lead that does not have a protective earth wire.
Warning:Warning:
Warning:
Warning:Warning: Any break in the earth wire inside or outside the apparatus
and any loose connection of the earth wire can make the operation of the apparatus dangerous. A deliberate break or interruption is not allowed.
When the housing is opened or components are being removed (except when this can be done by hand), live parts may be exposed. Connection may also be live.
Therefore, if the carrying out of an adjustment, service or repair on the open apparatus in the live state is unavoidable, this must be done by an expert who is familiar with the dangers involved.
Page 20
Chapter 1
Elecsys 1010 Service Manual
Make sure you only use fuses of the specified type and rated amperage when replacing. Repaired fuses must not be used and the fuse holder must not be bypassed.
If you have any reason to believe that the instrument can no longer operate safely, take it out of use and make sure no one can use it accidentally.
It must be assumed that safe operation is no longer possible when the instrument:
- shows visible signs of damage
- fails to operate
- has been stored under unfavourable conditions for a fairly long time
- has been transported under rough conditions
The Elecsys 1010 should be used by appropriately qualified persons only.
Warning:Warning:
Warning:
Warning:Warning: In order to avoid electrostatic charging when working on the
electronics, use a grounding strap and a grounding mat. For more information, please refer to operator manual.
FINAL 2.1 - February 2000
Chapter 1Page 21
Elecsys 1010 Service Manual
1.4 Service Concept1.4 Service Concept

1.4 Service Concept

1.4 Service Concept1.4 Service Concept
1.4.11.4.1
1.4.1
1.4.11.4.1
From the early stage of development, Elecsys 1010 was designed for simple error detection and easy exchangeability of modules. This gives the service workshops the possibility of a fast and easy repair of the instrument on service level A (module level). No big stock or expensive equipment is necessary and service technicians are easier to train. Also, a permanent technical improvement in layout and components took and takes place for better productivity and economic production.
On repairable modules the quality and function is always provided by the manufacturer according to the latest technology. This keeps Elecsys 1010 always on the highest quality level.
The exchange price for modules will be kept on a low level to guarantee repairs, on an economical basis.
1.4.21.4.2
1.4.2
1.4.21.4.2
Service LevelService Level

Service Level

Service LevelService Level
RA-Procedure (Replacement ofRA-Procedure (Replacement of
RA-Procedure (Replacement of
RA-Procedure (Replacement ofRA-Procedure (Replacement of defective parts)defective parts)
defective parts)
defective parts)defective parts)
- All defective parts ( non-"R" and "A" parts ) should be kept on stock for a period of 7 months. In case the manufacturer needs the part for investigation it will be requested from Mannheim.
- All parts returned to Mannheim and not requested by Mannheim will be send back at the expense of the countries.
- The RA claim for warranty has to be in Mannheim no later than 8 weeks after the problem date.
Exclusion of warrantyExclusion of warranty
Exclusion of warranty
Exclusion of warrantyExclusion of warranty The aforementioned warranties do not apply in case of
improper use, handling, transportation or storage, faulty installation, repair or maintenance, chemical influence or contamination as well as damages resulting from that, failure to follow operating instructions, alterations or modifications of instruments or parts thereof not authorized or recommended by RD GmbH and resulting damages, normal wear and tear and in case of other circumstances beyond the control of RD GmbH.
Handling of repairsHandling of repairs
Handling of repairs
Handling of repairsHandling of repairs As a general rule, all instrument repairs should be carried
out by authorized and trained personnel only.
Warranty period for instruments and spare partsWarranty period for instruments and spare parts
Warranty period for instruments and spare parts
Warranty period for instruments and spare partsWarranty period for instruments and spare parts The warranty period for instruments is 16 months starting
with the date of shipment ex works Mannheim/Federal Republic of Germany or 12 months starting with the date of the first installation, whichever period is shorter. The warranty period for spare parts is 6 months from installation date of the part, however not longer than 24 months after having delivered ex works Mannheim/Federal Republic of Germany. Note: In case the instrument has a remaining warranty period of more than 6 months, the parts remain under warranty until the warranty period of the instrument expires.
Handling of warranty claimsHandling of warranty claims
Handling of warranty claims
Handling of warranty claimsHandling of warranty claims The warranty claim has to be handled via Return
Authorization procedure or accepted equivalent. Please answer all the questions on the RA form with the greatest care.Especially a detailed fault description is needed or the warranty claim will not be accepted by the manufacturer. Complete instruments are not accepted unless this has been agreed with the service department of the relevant product group responsible at RD GmbH.
Repair of parts marked with "R"Repair of parts marked with "R"
Repair of parts marked with "R"
Repair of parts marked with "R"Repair of parts marked with "R" Parts which are economically worth repairing are marked
with "R" in the spare parts price list. New and repaired parts could be recognized by different material numbers (language version).
(e.g. new part: 1234567-001, repaired part: 1234567-984) Repaired parts should be ordered together with new parts
via the order processing department in Mannheim (OU­VDG). Parallel to the ordering process of a repaired part, the defective part should be returned together with the filled RA form (giving full details of the defect and marked choice box with repair) to Logistic Instruments (Goods Receipt) in MA-Wohlgelegen (LI-GS). Repair of instrumentsComplete instruments are not accepted for replacement or repair unless this has been agreed with the product group responsible at RD GmbH. Before replacement or repair can take place, the validity of the request must be examined and the question of costs must be settled in a written agreement with RD GmbH.
Terms of deliveryTerms of delivery
Terms of delivery
Terms of deliveryTerms of delivery
Important informationImportant information
Important information
Important informationImportant information
- Only parts marked with "A" in the price list are generally accepted under warranty.
- Only return those parts marked with "R" in the spare parts price list.
- Warranty claim for „R“ parts will be accepted, if the part was returned to Mannheim.
FINAL 2.1 - February 2000
Shipments to the countries with the routine truck are c.i.f./ shipments outside this procedure are ex works Mannheim. Emergency shipments require additional costs to be charged.
Page 22
Chapter 1
Loading...
+ 455 hidden pages