1.1.11 Analyzer Status Conditions ...........................................................................................................1 - 48
1.2TECHNICAL DATA ....................................................................................................................................1-52
1.2.1 Technical Data for Operation of Instrument................................................................................... 1 - 52
1.3POTENTIAL HAZARD AND SAFETY PRECAUTIONS ....................................................................................1-61
1.4.1 Disk System ..................................................................................................................................... 1 - 71
2.3.5 Mounting a Measuring Cell ............................................................................................................ 2 - 25
2.3.6 Installation of System Software....................................................................................................... 2 - 26
2.3.7 Fill Liquid System ........................................................................................................................... 2 - 26
2.3.8 Adjustments to be Checked During Installation and After Replacement......................................... 2 - 26
2.7TABLE OF CONTENTS ...............................................................................................................................2-92
2.7.1 Method to detach the external covers..............................................................................................2 - 92
2.7.2 Method to attach the external covers .............................................................................................. 2 - 99
2.7.3 Mmethod to detach the PC Unit.................................................................................................... 2 - 107
2.7.4 Method to attach the PC Unit ....................................................................................................... 2 - 109
Contents - 1
Fluidics
Chapter 3 Fluidics
3.1
DESCRIPTION OF FLOW PATH .....................................................................................................................3-1
3.1.2 List of Parts.......................................................................................................................................3 - 3
4.2.2 Rack Sampler System ........................................................................................................................ 4 - 6
4.3LIST OF MOTORS,SENSORS, AND OTHER MECHANISMS ..........................................................................4-10
4.3.1 List of Motors.................................................................................................................................. 4 - 10
4.3.2 Motor Reference List....................................................................................................................... 4 - 10
4.3.3 List of Sensors ................................................................................................................................. 4 - 10
4.3.4 List of Peltier, Heater and Fan Motor ............................................................................................ 4 - 11
4.3.5 List of Motors for Rack Sampler System ......................................................................................... 4 - 12
4.3.6 List of Sensors for Rack Sampler System ........................................................................................4 - 12
4.3.7 List of LEDs and Fan Motor for Rack Sampler System .................................................................. 4 - 12
4.4DETAILED EXPLANATION OF EACH MECHANISM .....................................................................................4-13
4.4.1 Sample Disk Drive Mechanism ....................................................................................................... 4 - 13
4.4.2 Reagent Disk Drive Mechanism......................................................................................................4 - 19
4.4.3 Cap Open/Close Mechanism........................................................................................................... 4 - 28
5.2.1 AC Power ........................................................................................................................................ 5 - 10
5.2.2 DC Power Supply............................................................................................................................ 5 - 11
7.2DATA ALARM LIST ....................................................................................................................................7-2
7.3TROUBLESHOOTING LIST ............................................................................................................................7-4
7.3.1 Reagent Disk ..................................................................................................................................... 7 - 4
7.3.2 Cap Opener....................................................................................................................................... 7 - 5
7.3.13 Distilled Water Float Switch......................................................................................................... 7 - 18
7.3.14 System Reagent Unit ..................................................................................................................... 7 - 19
SPECIAL SERVICE TOOLS ...........................................................................................................................8-1
8.2COMPLETE PARTS LIST ..............................................................................................................................8-2
The Roche Diagnostics cobas e411 Immunoassay System is a fully automated, software-controlled
system for immunoassay analysis. It is designed for both quantitative and qualitative in vitro
determinations using a large variety of tests for analysis.
To assist you in quickly identifying which component is specific to either the disk or rack system, one
of the following graphics appears to the right of the subsection header. If no graphic appears next to
the header, then that component is common to both systems.
Figure 1.1-3 Disk
Figure 1.1-4 Rack
1.1.2.1 The Control Unit
The control unit of the e411 analyzer is a touchscreen, no-keyboard type computer, which is located
on the left-center of the analyzer unit. This monitor unit contains the controlling software and also has
an on-screen keyboard function.
Figure 1.1-5 Control unit
Version 1.0 – May 2006 1 - 2 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
1.1.2.2 The Analyzer Unit
Figure 1.1-6
The analyzer unit on the disk system consists of the:
• sample/reagent area
• consumables area
• measuring area
• operation switch
The only difference on the rack system is in the sample area. The sample disk is replaced by a rack
sampling unit.
Refer to the photo below.
Figure 1.1-7
Version 1.0 – May 2006 1 - 3 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
1.1.2.3 Sample/Reagent Area
The sample/reagent area comprises the left half of the analyzer and consists of a sample disk or rack
sampler (rack system), rack bar code reader (rack system), sample/reagent (S/R) probe, bar code
reader, bar code card reading station, reagent disk, a cap open/close mechanism, a microparticle
mixer, probe/ mixer rinse station and sample/reagent (S/R) pipettor.
The sample disk accommodates up to 30 samples. The A-Line of the rack sampler accommodates 75
samples on a single tray (15 racks at a time; each rack with five positions) and 25 samples in the input
buffer for a total capacity of 100 samples. The reagent disk, temperature controlled at 20 ± 3 °C,
accommodates up to 18 reagent packs.
1.1.2.4 Consumables Area
The consumables area is on the right of the analyzer, consisting of three tip trays, three AssayCup
trays, a gripper unit, cup disposal opening, liquid waste container, solid waste tray and liner and
system water container.
1.1.2.5 Measuring Area
The measuring area includes the incubator, the sipper probe, sipper rinse station, system reagents
(ProCell and CleanCell), an aspiration station, sipper pipettor and the detection unit. The sipper probe
aspirates the incubated reaction mixture into the detection unit for result determination.
1.1.2.6 Operation Switch
The operation ON/OFF switch is located on the front left of the analyzer. In addition, there is a circuit
breaker for the analyzer located on the right side panel and a rack sampler circuit breaker located on
the left side of the rack sampler.
Version 1.0 – May 2006 1 - 4 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
1.1.3 Control Unit Components
The control unit consists of a color touchscreen monitor, host interface and external printer.
1.1.3.1 Touchscreen Monitor
The touchscreen monitor is located on the left-center of the analyzer and displays the software. For
details on the cobas e411 software, refer to the Software Guide.
Figure 1.1-8 Touchscreen monitor
Version 1.0 – May 2006 1 - 5 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
1.1.3.2 On-screen Keyboard
The e411 software has an on-screen keyboard. For details refer to the Software Guide.
Figure 1.1-9 On-screen keyboard
1.1.3.3 External Printer
The instrument uses an 80-column, graphics-capable, dot matrix printer.
The printer is connected to the analyzer via a USB port. The analyzer has two USB ports on its left
side.
Figure 1.1-10 Location of the USB ports and Host port
1.1.3.4 Host Interface
The instrument can be bidirectionally interfaced with a host computer.
Please refer the Host Interface manual in detail.
Version 1.0 – May 2006 1 - 6 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
1.1.4 Sample/Reagent Area Components
The sample/reagent area consists of a sample disk or rack sampler (rack system), rack ID bar code
reader (rack system), sample/reagent (S/R) probe, bar code reader, bar code card reading station,
reagent disk, cap open/close mechanism, microparticle mixer, probe/mixer rinse station and
sample/reagent (S/R) pipettor.
Figure 1.1-11 Sample Disk
The sample disk has 30 positions for samples, calibrators and controls. Patient samples may be
placed in either primary sample tubes or sample cups. Built-in adapters allow intermixing of different
size primary sample tubes.
Sample tubes that may be used are listed in chapter 2.7 Technical Data.
Sample cups [2 mL (Standard) Hitachi cups only] may be placed directly on the sample disk or on top
of 16 mm primary sample tubes.
Figure 1.1-12 Sample disk
CAUTION
Micro cups cannot be used on the
e411 analyzer!
Version 1.0 – May 2006 1 - 7 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Note: How to raise the sample disk protective cover
To take out the sample disk, first you must raise the sample disk protective cover. The cover can be
held at a certain angle. To lay the cover down, release the hold by raising the cover up to its limit
angle. Be sure to lay it down before you start operation.
Figure 1.1-13 Sample disk protective cover
Figure 1.1-14 n Raising the sample disk protective cover
Version 1.0 – May 2006 1 - 8 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Figure 1.1-15 Standing angle
Figure 1.1-16 Limit angle
CAUTION
Be sure to lay down the sample disk
protective cover before you start
operation.
Version 1.0 – May 2006 1 - 9 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Figure 1.1-17
Rack Sampler
The rack sampler consists of an A-Line, B-Line, C-Line and STAT position.
1.1.4.1 A-Line
Specimens are placed in 5-position sample racks and are loaded onto a tray. Once a tray is loaded,
additional racks can be added to the tray one at a time during Operation, provided the tray indication
light is green (ON). If the light is out (OFF), the pusher arm is preparing to move. The pusher arm is
located at the far left of the A-Line and pushes the sample racks forward and onto the B-Line.
The A-Line holds a tray that accommodates 15 racks at one time. Another five racks can be in the
input buffer. Therefore, you can have a total of 100 specimens loaded at any one time. Refer to the
photo and graphic below.
Figure 1.1-18
Version 1.0 – May 2006 1 - 10 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Figure 1.1-19
1.1.4.2 B-Line
The B-Line transports the sample racks, single file, first to the rack bar code reader. Here each
position in the rack is scanned for a sample bar code. After the last position is scanned, the bar code
reader scans the rack ID. After the last specimen is sampled, the rack is transferred via the output
buffer onto the tray on the C-Line. Refer to the photo and graphic below.
Figure 1.1-20
Version 1.0 – May 2006 1 - 11 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Figure 1.1-21 B-Line of the rack sampler
1.1.4.3 Rack Bar Code Reader
The rack bar code reader reads both sample bar code labels and the rack bar code label. The bar
code reader is auto-discriminating, allowing the use of various types of bar codes during operation.
Bar code symbologies read include:
• NW7 (Codabar)
• Code 39
• Code 128
• Interleaved 2 of 5
Figure 1.1-22 Rack bar code reader
Version 1.0 – May 2006 1 - 12 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
1.1.4.4 C-Line
Racks are off-loaded from the B-Line into the output buffer. If there is no tray, up to 5 racks can enter
the output buffer, thereafter the sampling procedure is stopped. When the sixth rack is moved into the
output buffer, a rack is pushed onto the tray on the C-Line. You can remove the tray from the C-Line
any time the tray indication light is green (ON). If the light is out (OFF), the system is preparing to push
a rack onto the C-Line tray. You cannot remove single racks from the C-Line. You must remove an
entire tray at one time.
Figure 1.1-23
Figure 1.1-24 C-Line of the rack sampler
If the tray is removed, the system continues to push racks into the output buffer. If the buffer fills and
there is no tray, the analyzer issues an alarm and stops sampling racks.
Version 1.0 – May 2006 1 - 13 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Figure 1.1-25 Output buffer with racks
1.1.4.5 STAT Position
The STAT position is located at the front of the analyzer and is in line to feed directly onto the B-Line.
Place a rack in the position as directed on the label and press the STAT key. When the rack currently
being sampled is completed, the STAT rack is pushed onto the B-Line and is sent on to the rack bar
code reader and sampling position.
Figure 1.1-26
Version 1.0 – May 2006 1 - 14 Chapter 1.1
RD/Hitachi cobas e411 Service Manual
Figure 1.1-27 STAT position of the rack sampler
1.1.4.6 Sample Rack
Sample cups, primary sample tubes, calibrator or control vials are placed in sample racks shown
below. Each sample rack holds a maximum of five samples. Each tube slot contains adapters that
allow the rack to hold different sizes of primary sample tubes. Each rack has a unique ID found on the
bar code label on the back end of the rack. This rack ID is read by the bar code reader and transferred
to the system. This ID appears on the screens in the software and on the reports.
Figure 1.1-28 Sample rack
1.1.4.7 Sample/Reagent (S/R) Probe
The sample/reagent probe is located on the back left wall of the analyzer and is mounted on an arm
(S/R arm) that moves horizontally between the sample and reagent disk. The probe uses disposable
tips to avoid sample carryover, and has liquid level and clot detection for accurate pipetting. Liquid
level detection is accomplished by capacitance measurement. Clot detection is accomplished by a
pressure transducer.
A new AssayTip is utilized with every new pipetting sequence. For example, TSH = 1 tip for R1, R2
and sample, then one new tip for microparticles. The tip is washed externally at the rinse station
between each aspiration. Additional tips are used for sample dilutions or pretreatment.
Version 1.0 – May 2006 1 - 15 Chapter 1.1
Loading...
+ 571 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.