RKI Instruments SM-2009U User Manual

Operator’s Manual
SM-2009U Single Module
Calibration Station
Part Number: 71-0164RK
Revision: A
www.rkiinstruments.com
Warranty
RKI Instruments, In c. warrant s gas alarm equi pment sold by us to be free from defects in materials and workmanship, and performance f or a period of one year from date of shipment from RKI Instruments, Inc. Any parts found de fective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to tho se items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Absorbent cartridges Batteries Pump diaphragms and
Filter elements
valves Fuses
Warrant y is voi ded by abuse including mechanical damage, alteration, rough handling, or repairs procedures not in accordance with the instruction manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
HIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
T
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
NSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO THE WARRANTY
I
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users only by authorized distributor s, dealers, and represen tatives as appoin ted by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor and our warranty is limited to replacement of parts or our complete goods.
Warranty
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About the SM-2009U. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
About This Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions & Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Air Filter & Sample Tubing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Instrument Cradle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Back Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Control Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
USB Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Chapter 3: Preparing to Use the SM-2009U . . . . . . . . . . . . . . . . . . . . . 12
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Hardware Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting the Bump Test and Calibration Parameters. . . . . . . . . . . . . 13
Connecting Calibration Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Installing the Single Module Data Viewer Software. . . . . . . . . . . . . 24
Table of Contents
Chapter 4: Using the SM-2009U. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Bump Testing Instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Calibrating Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Charging an Instrument in a Calibration Station . . . . . . . . . . . . . . . 35
Calibration and Bump Test Records . . . . . . . . . . . . . . . . . . . . . . . . 37
Chapter 5: Single Module Data Viewer Software . . . . . . . . . . . . . . . . . 41
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Launching the Single Module Data Viewer Software. . . . . . . . . . . . 41
Using the Single Module Data Viewer Software . . . . . . . . . . . . . . . 42
Spare Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
CAUTION: Read and understand this manual before using the
SM-2009U. Also read and understand the GX-2009 operator’s manual.
Table of Contents
Chapter 1: Introduction
Overview
This chapter briefly describes the SM-2009U Single Module Calibration Station and the Single Module Data View Software. This chapter also describes the SM-2009U Single Module Calibrat ion Station Operator’s Manual (this document). Table 1 at the end of this chapter lists the SM-2009U’s specifications.
About the SM-2009U
The SM-2009U Single Module Calibration Station is an advanced, reliable system that provides charging, calibration, bump testing, and calibration and bump test records for the GX-2009 gas detect or. It is designed to save the records to a USB flash drive. The Single Module Data Viewer Software can then be used with a Windows-based personal computer to retrieve calibration and bump test data files from the USB flash drive or from the computer’s hard drive if the files have been transferred to the hard drive from the flash drive.
The purpose of this manual is to explain how to set up and use the SM­2009U and the Single Module Data Viewer Software. You will learn how to:
install and launch the software
prepare the SM-2009U for use
perform a bump test
perform a calibration
save calibration and bump test records to a USB flash drive
view, print, and export calibration and bump test reco rds
use the SM-2009U to charge a GX-2009
Overview • 1
CAUTION: The GX-2009 detects oxygen deficiency and elevated levels
of oxygen, combustible gases, carbon monoxide, and hydrogen sulfide, all of which can be dangerous or life threatening. When using the GX-2009, you must follo w th e instructions and warnings in the GX-2009 Operator’s Manual to assure proper and safe operation of the unit and t o minimize the risk of personal injury.
CAUTION: The operator of this instrument is advised that if the
equipment is used in a manner not specified in this manual, the protection provided by the equipment may be impaired.
System Requirements
To use the Single Module Data Viewer Sof tware, your personal comput er must meet the following requirements:
Operating Systems: Windows® 2000, Windows® XP, Windows
®
Vista.
Processor: IBM
®
compatible PC running Pentium® 2 processor or
equivalent minimum
Memory: 32 MB RAM minimum
Hard Disk Space: 32 MB minimum
CD-ROM Drive
Available USB port
2 • System Requirements
Specifications
Table 1: SM-2009U Specifications
Input Power 12 VDC
NOTE: AC Adapter with 100 - 240 VAC, 50/ 60 Hz, 0.6A input and 12 VDC, 1.2A output provided as standard.
Environmental Conditions • For Indoor Use Only
• -10° C to 40° C, below 80% Relative Humidity, Non-Condensing
Applicable Instrument GX-2009 Memory Capacity 64 KB Maximum Record Size 256 bytes Maximum Number of
200
Records Saved Number of Calibration
Gas Cylinders
One calibration gas cylinder per bump test or calibration
Standard Accessories AC Adapter
USB Flash Drive
Single Module Data Viewer Software
Inlet Air Filter
Instruction Manual
Specifications • 3
About this Manual
The SM-2009U Single Module Calibration Stations Operator’s Manual uses the following conventions for notes, cautions, and warnings.
NOTE: Describes additional or critical information.
CAUTION: Describes potential damage to equipment.
WARNING: Describes potential danger that can result in injury or
death.
Cautions & Safety Information
Use only polyurethane sample tubing with the SM-2009U. Consult RKI Instruments, Inc. for other materials.
Do not subject the SM-2009U to infrared or intense light. This may cause communication errors.
Do not expose the SM-2009U to water.
Do not subject the SM-2009U to any hard impact.
4 • About this Manual
Chapter 2: Description
Overview
This section describes the SM-2009U single module calibration station. It is designed to be used on a table top and consist s of the AC adaptor, air filter and sample tubing, instrument cradle, back panel, control panel, status LEDs, and USB port.
AC Adapter
The AC adapter is a wall plug style adapter with a 5 foot cable. The end of the cable has a plug that connects to the power jack on the SM­2009U’s back pan el. The AC adapter is rated 100 - 240 VAC input, 12 VDC 1.2 A output.
To Power Jack
on SM-2009U
Back Pane l
Figure 1: AC Adapter
Air Filter & Sample Tubing
A cylindrical particle filter with a short length of tubing is supplied wi th the SM-2009U for installation to the AIR fit ting on the back panel. The filter keeps particulate contamination out of the calibration station.
Two sample tubes are included with the calibration station. A 3 foot length of 3/16” ID polyurethane tubing is provided to connect the regulator on a calibration cylinder to the CAL. GAS fitting on the back panel. A 10 foot length of 5/16” ID polyurethane tubing is provided for connection to the exhaust fitting on the b ack panel to allow routing of the exhaust to a location such as open window where the exhaust can disperse.
Overview • 5
Calibration G as S ample Tubing, 3 feet
Exhaust Tubing,10 feet
Particle Filter for Air Inlet
Figure 2: Air Filter & Sample Tubing
WARNING: Do not use an exhaust tube that is longer than 10 feet.
Using an exhaust tube longer than 10 feet will cause inaccurate gas response by the GX-2009 during a bump test or calibration.
6 • Air Filter & Sample Tubing
Instrument Cradle
The instrument cradle is a recessed area on the top of the SM-2009 that is designed to accept the GX-2009. Insert the GX-2009 in the inst rumen t cradle when you perform a bump test or calibration. The cradle has charging contacts on the bottom that mate with the contacts on the back of the GX-2009. There are sample ports in the front of the cradle that match up with the GX-2009’ s sensors and route air and calibration gas to the sensors during a bump test or calibration. An infrared (IR) por t at the back of the cradle lines up with the GX-2009’s IR port when it is inserted in the cradle and is used to communicate with the GX-2009. A retaining tab that holds the GX-2009 in the cradle is located to the right of the IR port.
BUMP
CAL.
SM-2009U
EDIT
ENTER
DATA CLEAR
GX- 2 009 O F F
COPY
CHARGE
PO WER
1SEC ON 3SEC OFF
IR Port Instrument Retaini ng
Tab Instrument Cradle
Charging Contacts
SM-2009U With GX-2009SM-2009U WithoutGX-2009
Figure 3: Instrument Cradle
Instrument Cradle • 7
Back Panel
The back panel includes the power jack and sample fittings.
CAL. GAS
Fitting
AIR Fitting
Power J ack
Figure 4: Back panel
Exhaust Fitting
Power Jack
The power jack is located in the lower left corner of the back panel. The plug on the end of the AC adapter cable mates to it.
Sample Fittings
Three sample fittings are located on the back of the SM-2009 U. The AIR fitting is in the upper left corner and draws air into the SM-2009U. The CAL. GAS fitting is next to the AIR fitting and is used to connect the SM­2009U to a calibration gas cylinder. Both the AIR and CAL. GAS fittings accept 3/16” ID tubing.
An exhaust fitting is located in t he up per right corne r of t he ba ck p anel. It allows routing of the exhausted calibration gas to a convenient location. This fitting accepts 5/16” ID tubing. Even though the exhaust gas can be routed to an area to be safely dispersed, the docking station should still be installed in a well ventilated area.
8 • Back Panel
Control Panel
The control panel is used to setup and operate the calibr ation station. It is located at the front of the calibration station. It includes the control buttons, the control button LEDs, and the CHARGE status LED.
CHARGE LED
COPY LED
CAL. LED
BUMP LED
CHARGE
CAL.BUMP
EDIT
ENTER
DATA CLEAR
GX-2009 OFF
COPY
POWER
1SEC ON 3SEC OFF
Control Buttons
Figure 5: Control Panel
Five control buttons are located on the control panel. From left to right they are BUMP ▼, CAL. , EDIT ENTER, COPY, and POWER. The BUMP ▼, CAL. , and COPY control buttons each have an L ED above them that indicates the status of t he function controlled by that button. The CHARGE LED is located above the POWER butt on and functions as a pilot LED, a system failure LED, and a charge indication LED.
Control Panel • 9
Table 2: Control Button Functions
Control
Control Button Function(s)
Button
BUMP
• Initiates a bump test
• Cancels a bump test
• Moves down a list of parameters
• Decreases an adjustable parameter
CAL. • Initiates a calibration
• Cancels a calibration
• Clears data from station memory (when used with COPY button)
• Moves up a list of parameters
• Increases an adjustable parameter
Control Button LED
Function(s)
Indicates status of a bump test in progress
Indicates status of a calibration in progress
EDIT ENTER
• Puts station into various edit modes
• Makes a displayed parameter editable
• Escapes or cancels an operation
• Turns off connected GX-2009 (when used with POWER button)
COPY • Copies data to USB flash
drive.
• Clears data from station memory (when used with CAL. button)
n/a
• Indicates amount of calibration station memory used
• Indicates status of copying function
• Indicates the result of a copy operation
10 • Control Panel
Table 2: Control Button Functions
Control
Control Button Function(s)
Button
POWER • Turns on the station
• Turns off the station
• Turns off connected GX-2009 (when used with EDIT ENTER button)
Control Button LED
Function(s)
n/a
USB Port
A USB port is located on the right front of the calibratio n statio n below the COPY and POWER buttons. The port can be used to save calibration and bump test data to a USB flash drive.
USB Port
Figure 6: Control Panel
NOTE: The SM-2009U does not support connection of a computer
to the USB port, only a USB flash drive.
USB Port • 1 1
Chapter 3: Preparing to Use the SM-2009U
Overview
There are four tasks t hat must be completed befo re you ca n begin to use the SM-2009U: hardware assembly, setting the bump test and calibration parameters, connecting calibration gas, and installing the Single Module Data View Software on your computer. This chapter describes how to assemble the parts that are shipped with the SM-2009U and set the bump test and calibration parameters. It also describes how to install the Single Module Data Viewer Software on a Windows based personal computer.
Hardware Assembly
The hardware assembly consists of connecting the AC adapter, installing the air filter, connecting the sample tubing, and setting the SM-20090U’s bump test and calibration parameter s. Perform the following to complete the hardware assembly
1. Place the SM-2009U on a convenient table top near an AC wall socket in a well ventilated area. A location near a window that can be opened is best so that the exhaust can be routed to the window.
2. Insert the round plug on the end of the AC adapter’s cable into the power jack on the back of the SM-2009U.
3. Connect the AC adapter’s wall plug into a wall AC socket.
4. Install the air filter so that the arrow on the filter that indicates direction of flow is pointing towards the AIR fitt i ng. Push the open end of the flexible tube that is on one end the filter onto the AIR fitting on the back of the SM-2009U.
5. Install the 10 foot long 5/16” ID flexible tube that is included with the SM-2009U on the exhaust fitting. Route the tube to an area where the exhaust can be safely dispersed, such as an open window.
12 • Overview
CAUTION: The maximum recommended length for the exhaust tube is
10 feet. Do not use more than 10 f eet of tubing or tubing with an ID of less tha n 5/16” for the exhaust tube or t he bump test and calibration accuracy will be adversely affected. The tube that is shipped with the SM-2009U has an ID of 5/16” and is 10 feet long.
6. Install the 3 foot long 3/16” ID tube that is included with the SM-2009U on the CAL. GAS fitting.
Setting the Bump Test and Calibration Parameters
The bump test parameters def ine how long fresh air and calibration gas are applied to an instrument during a bump test. They also define the tolerance used in determining whether a unit fails or passes a bump test and whether or not a calibration automatically takes place if a bump test fails. The calibration parameters defi ne how long fresh air and calibration gas are applied to an instrument during a calibration.
The bump test and calibration parameters are saved in the SM-2009U’s memory. If a paramete r is changed with one particular GX-2009 installed in the SM-2009U, the change will be in effect for the bump test or calibration of any subsequent GX-2009 until the parameter is changed again.
Bump Test & Calibration Parameters
There are four bump test parameters and two calibration parameters. The two calibration parameters, air sample time and calibration gas sample time, are also bump test parameters. The parameters are described below. Table 3 below shows the factory settings for the bump test and calibration parameters. If you wish t o use the factory settings, then you do not n eed to make any parameter adjustments. If you wish to confirm or change the parameter settings, follow the instructions below beginning with “Turning On the SM-2009U with a GX-2009” on page 15.
Setting the Bump Test and Calibration Parameters • 1 3
Loading...
+ 38 hidden pages