RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from
defects in materia ls and workmans hip, and perform ance for a perio d of one year from
date of shipment from RKI Instruments, Inc. Any parts found defective within that
period will be repair e d or repl ace d, at our optio n, fre e of char ge . This w arra nty do es
not apply to those items which by their nature are subject to deterioration or
consumption in normal service, and which must be cle aned, repaired, or replac ed o n
a routine basis. Examples of such items are:
Absorbent cartridgesBatteries
Pump diaphragms and valvesFilter elements
Fuses
Warranty is voided by abuse including mechanical damage, alteration, rough
handling, or repairs procedures not in accordance with the instruction manual. This
warranty indicates the full extent of our liability, and we are not responsible for
removal or replacement costs, local repair costs, transportation costs, or contingent
expenses incurred without our prior approval.
THISWARRANTYISEXPRESSLYINLIEUOFANYANDALLOTHER
WARRANTIESANDREPRESENTATIONS, EXPRESSEDORIMPLIED, AND
ALLOTHEROBLIGATIONSORLIABILITIESONTHEPARTOF RKI
NSTRUMENTS, INC. INCLUDINGBUTNOTLIMITEDTOTHEWARRANTY
I
OFMERCHANTABILITYORFITNESSFORAPARTICULARPURPOSE. IN
NOEVENTSHALL RKI INSTRUMENTS, INC. BELIABLEFORINDIRECT,
INCIDENTAL, ORCONSEQUENTIALLOSSORDAMAGEOFANYKIND
CONNECTEDWITHTHEUSEOFITSPRODUCTSORFAILUREOFITS
PRODUCTSTOFUNCTIONOROPERATEPROPERLY.
This warranty covers instruments and parts sold to users only by authorized
distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by th e
operation of th is gas mo nitor an d our war ranty is li mited to r eplacement of par ts or our
complete goods.
CAUTION:Read and understand this manual before using the SDM-2009. Also
read and understand the GX-2009 Operator’s Manual.
Table of Contents
Introduction
This section briefl y describes the S DM-2009 D ocking S t ation, the Single Modul e Data
Viewer Program, and the SDM-GX Docking Station PC Controller Program. This
section also describes the SDM-2009 Docking Station PC Controlled Configuration Operator’s Manual (this document). Table 1 at the end of this section lists the SDM2009’s specifications.
The SDM-2009 Docking Station is an advanced, reliable system that provides
charging, calibration, bump testing, and calibration and bump test records for the
GX-2009 portable gas monitor. It is designed to save the calibration and bump test
records to a USB flash drive (standal one functional ity) or to be conn ected directly to a
computer (PC contr olled f unctiona lity). If calibra tion an d bump test recor ds are stor ed
to a USB flash drive while operating in the standalone configuration, the Single
Module Data Viewer Program can then be used with a Windows-based personal
computer to retrieve calibration and bump test data files from the USB flash drive or
from the comput er’s hard drive if the fi les have b een transferre d to the hard driv e from
the flash drive. If you are using the PC Controller Program while operating in the PC
controlled configuration, you may retrieve instrument data, bump test, and calibrate
up to 10 units at once. Instrument info rmation and data for each instrument can be
viewed directly using the PC Controller Program and can be printed from the PC
Controller Program. For instructions to use the SDM-2009 in the Standalone
configuration, see the SDM-2009 Docking Station Standalone Configuration Operator’s Manual.
The purpose of this manual is to explain how to set up and use the SDM-2009 in PC
Controlled config uratio n. It also explai ns how to use the SD M-GX Doc king S t ation PC
Controller Program. You will learn how to:
•install and launch the SDM-GX Docking Station PC Controller Program
•setup the SDM-GX Docking Station PC Controller Program for use with the
SDM-2009
•prepare the SDM-2009 for use
•bump test and calibrate up to 10 units using the PC controlled configuration
•use the SDM-2009 to charge a GX-2009
•view, print, and export calibr ation and bump test records
•view instrumen t inf or m ati on and data using the PC Controller Program
•update instrument parameters using the PC Controller Program
1 • Introduction
CAUTION:The GX-2009 detect s oxygen defi ciency and elevated levels of oxygen ,
combustible gases, carbon monoxide, and hydrogen sulfide, all of
which can be dange r ous or life threatening. Wh en using the GX-2009,
you must follow the instructions and warnings in the GX-2009
Operator’s Manual to assure proper and safe operation of the unit and
to minimize the risk of personal injury.
CAUTION:The operator of this instr umen t is advised that if the equi pment is u sed
in a manner not specifie d in this manual, the protection pr ovided by the
equipment may be impaired.
System Requirements
To use the SDM-GX Do cking S t ation PC Con troller Sof tware, your personal comp uter
must meet the following requirements:
•Operating System s: Windows® XP, Windows® Vista, Windows® 7.
®
•Processor: IBM
compatible PC running Pentium® 2 processor or equivalent
minimum
•Memory: 3 2 MB RAM minimum
•Hard Disk Space: 32 MB minimum
•CD-RO M Drive
•One (for 4 or less SDM-2009s) or two (for 5 or more SD M-2009s) available
USB port(s), one or two USB hubs may also be needed depending on the
number of SDM-2009s in your system. See “USB Hub Requirements for
Multiple Station Systems” on page 11.
System Requirements • 2
Specifications
Table 1: SDM-2009 Specifications
Input Power12 VDC
NOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz,
0.6A input and 1 2 VDC, 1.2A output provided as
standard.
Environmental Conditions•For Indoor Use Only
•-10° C to 40° C, below 80% Relative Humidity, NonCondensing
Applicable InstrumentGX-2009
Memory Capaci ty64 KB (S tandalone configuration only)
Maximum Record Size256 bytes (Standalone configuration only)
Maximum Number of
Records Saved
Number of Calibration Gas
Cylinders
Standard Accessories•AC Adapter
200 (Standalone configuration only)
One calibration gas cylinder per bump test or
calibration
•USB Flash Drive
•Single Module Data Viewer Program
•SDM-GX Docking Station PC Controller
Program
•Inlet Air Filter
•Instruction Manual
•10 Foot Long 5/16 Inch Exhaust Tube
•One 3 Foot Long 3/16 Inch Tube for GAS
Fitting
•10 Foot Long 3/16 Inch Tube for GAS Fitting
Manifolding
3 • Specifications
•2 T-Fittings
•Check Valve
•USB Cable, Type A to Type B
NOTE:The bump test function of the GX-2009 is included in instruments with
firmware version 03955 or later installed. GX-2009 instruments
shipped after May 2011 include the bump test function. Instruments
that have a serial nu mb er starting with 04516 or later shi pp ed fr om th e
factory with the bump test function available. Instruments that have a
serial number starting with 04515 did not include the bump test
function when shipped from the factory. The factory setting for the
bump test function is off. Any GX-2009 can be bump tested using the
SDM-2009 and the PC Controlle r Program regardless of its firmware
version.
About this Manual
The SD M-2009 D ocking S ta tion PC Controll ed Config uration O perator’s Man ual uses
the following conventions for notes, cautions, and warnings.
NOTE:Describes additional or critical information.
CAUTION:Describes potentia l damage to equipm en t.
WARNING:Describes potential danger that can result in injury or death.
Cautions & Safety Information
•Use only polyurethane sample tubing with the SDM-2009. Consult RKI
Instruments, Inc. for other materials.
•Do not subject the SDM-2009 to infrared or intense light. This may cause
communication errors.
•Do not expose the SD M-2009 to water.
•Do not subject the SDM-2009 to any hard impact.
About this Manual • 4
Description
To Pow er Jack
on SDM-2009
Back Panel
Figure 1: Single-Port AC Adapter
To Power Jack
on SDM-2009
Back Panels
Figure 2: 4-Port AC Adapter
This section describes the SDM-2009 docking station. It is designed to be used on a
table top and cons ists of the AC ad aptor , T ype A to T y pe B USB cable, air filte r , check
valve, 3 plasti c T-fittings, sample tubing, instrument panel, back panel, control panel,
status LEDs, and 2 USB ports.
AC Adapter
Single-Port AC Adapter
The single-port AC ada pt er is a w all plug style ada pter w ith a 5 f oot cable. The en d of
the cable has a plug that connects to the power jack on the SDM-2009’s back panel.
The AC adapter is rated 100 - 240 VAC input, 12 VDC 1.2 A output.
4-Port AC Adapter
The 4-port AC adapt er is a wa l l p lug style adapter with four 5-foot cables . The end of
each cable has a plug that connects to t he power jack on the SDM-2009’s back
panel. The AC adapter is rated 100 - 240 VAC input, 12 VDC 2.0 A output.
5 • Description
USB Cable
Type A, for
connection to
PC USB port
Type B, for connection
to USB port on the
SDM-2009's back panel
Exhaust Tubing, 5/16 Inch
ID,10 feet
Calibration Gas Sample Tubing, 3/16
Inch ID, 3 feet
Particle Filter for Air Inlet
Manifold Tubing, 3/16 Inch
ID,10 feet
Figure 4: Air Filter & Sample Tubing
A Type A to Type B USB cable is provided with the docking station. It is used to
connect the USB port on the back of the SDM-2009 to a computer.
Figure 3: USB Cable
Air Filter, Sample Tubing, and Chec k Valve
A cylindrical particle filter with a short length of tubing is supplied with the SDM-2009
for installation to the AIR fitting on the back panel. The filter keeps particulate
contamination out of the docking station.
Three types of sample tubes are included with the docking station. One 3 foot length
of 3/16 inch ID polyurethane tubing is provided to connect the regulator on a
calibration cylinder to the GAS fitting on the back panel. One 10 foot length of 3/16
inch ID polyurethane tubing is provided for GAS fitting manifold construction (if
desired). In addition, a 10 foot length of 5/16 inch ID polyurethane tubing is provided
for connection to the exhaust fitting on the back panel to allow routing of the exhaust
to a location such as an open window where the exhaust can disperse.
WARNING:Do not use an exhaust tube that is longer than 10 feet. The increased flow
restriction caused by a longer tube may affect gas response and cause
inaccurate calibration and bump test results.
Description • 6
A check valve is included with the SDM-2009. It is intended for use on the exhaust
Figure 5: Check Valve
For Exhaust
Tubing
For GAS
Tubing
Figure 6: T- Fittings
fitting when manifolding multiple docking stations together. See “Assembling a
Manifold for Multiple SDM-200 9 Units” on page 13 for manifolding instructions.
Two T-fittings are included with the SDM-2009. The larger fitting is for the exhaust
tubing manifold construction. The smaller fitting is for the GAS tubing manifold
construction.
7 • Description
Instrument Cradle
EDIT
ENTER
CAL.
POWER
CHARGE
SDM-2009
1SEC ON
3SEC OFF
COPY
GX-200 9 OFF
BUMP
SDM-2009 Without GX-2009
H2SCO%LELO
2
RKIGX-2009
POWER
MODE
AIR
BUMP
GX-200 9 OFF
COPY
1SEC ON
3SEC OFF
SDM-2009
CHARGE
POWER
CAL.
EDIT
ENTER
Charging Contacts
SDM-2009 With GX-2009
IR Port
Inst rum ent Cradle
Instrument Retaining
Tab
Figure 7: Instrument Cradle
The instrument cradle is a recessed area on the top of the SM-2009 that is designed
to accept the GX-2009. Insert the GX-2009 in the instrument cradle when you
perform a bump test or calibration. The cradle has charging contacts on the bottom
that mate with the contacts on the back of the GX-2009. There are sample ports in
the front of the cradle that match up with the GX-2009’s sensors and route air and
calibration gas to the sensors during a bump test or calibration. An infrared (IR) port
at the back of the cra dle l i ne s up wi th th e GX -2009’s IR port when it is inserted i n th e
cradle and is used to communicate with the GX-2009. A retaining tab that holds the
GX-2009 in the cradle is located to the right of the IR port.
Description • 8
Back Panel
Gas Fitti ng
Exhaust Fitting
Back Panel USB
Port (Type B),
For Computer
Connection
Air Fitting
Power Jack
Figure 8: Back Panel
The back panel includes the power jack, sample fittings, and a USB PC connector.
Power Jack
The power jack is located in the bottom left corner of the back panel. The plug on the
end of the AC adapter cable mates to it.
Sample Fittings
Three sample fittings are located on the back of the SDM-2009. The AIR fitting is in
the upper left corner and draws air into the SDM-2009. The GAS fitting is to the right
of the AIR fitting and is used to connect the SDM-2009 to a calibration gas cylinder.
Both fittings accept 3/16 inch ID tubing.
An exhaust fitting is located in the upper right corner. It allows routing of the
exhausted calibration gas to a convenient location. This fitting accepts 5/16 inch ID
tubing. Even though the exhau st gas can be r outed to an area to b e safely disp ersed,
the docking station should still be installed in a well ventilated a rea.
PC Connection
A type B USB connection is located beneath the exhaust fitting on the SDM-2009’s
back panel. It is used to connect the SDM-2009 to a PC.
9 • Description
Control Panel
CHARGE
1SEC ON
3SEC OFF
COPY
GX-2009 OFF
POWER
SDM-2009
BUMPCAL.
Control Buttons
EDIT
ENTER
BUMP LED
CHARGE LED
COPY LED
CAL LED
Figure 9: Control Panel
The control panel is us ed to setup and operate the docking station in the Standalone
configuration. It is located at the front of the docking station. It includes the control
buttons, the control button LEDs, and the CHARGE status LED.
Five control butto ns are located on the control p anel. From lef t to right the y are BUMP
T , CAL S , EDIT ENTER, COPY, and POWER. The BUMP T , CAL S , EDIT
ENTER, and COPY control buttons are not used in the PC Controlled configuration of
the SDM-2009. The BUMP T LED and CAL S LEDs indicate the results of bump
tests and calibrations, respectively. The COPY LED does not indicate anything in the
PC Controlled configuration but will be on if the SDM-2009 was used in the
Standalone configuration and calibration and bump test records are still stored in the
SDM-2009’s memory. The CHARGE LED is located above the POWER button and
functions as a pilot LED, a system failure LED, and a charge indication LED.
The POWER button turns the SDM-2009 on and off.
Description • 1 0
Front Panel
Figure 10: Front Panel
A type A USB port is located on the front of the docking station. This port can be used
to save calibration and bump test data to a USB flash drive. This USB port is for use
only in the Standalone configuration of the SDM-2009 and is not used in the PC
Controlled configuration.
NOTE:The SDM-2009 does not support connection of a computer to the front
USB port, only a USB flash drive.
USB Hub Requirements for Multiple
Station Systems
If you have more docking stations than you have available USB ports, you will need a
USB hub. RKI provides both a 4-port hub and a 7-port hub. If you have more than 7
docking stations, you will need to purchase one of each hub. Do not connect one hub
to the other. Make sure that each USB hub has its own USB port on your computer.
You may also purchase your own USB hub of any port number from your local
electronics store but it must meet the following requirements:
•USB 2.0
®
•compatible with your Windows
An AC adapter comes with the USB hub(s) sold by RKI but does not need to be
plugged in for SDM-2009 operation. If the hub will be used to charge other devices,
the AC adapter needs to be plugged in.
operating system
11 • USB Hub Requirements for Multiple Station Systems
Hardware Setup
Figure 11: SDM-2009 Arrangement
This section describes how to assemble the hardware that came with the SDM-2009,
assemble a manifold, and connect calibration gas.
Hardware Assembly
The hardware assembly consists of connecting the AC adapter(s), installing the air
filter(s), installing the check valve(s), connecting the sample tubing and connecting
the USB cable(s). The SDM-GX Docking Station PC Controller software can support
1-10 SDM-2009 docking stations connected at the same time. Perform the following
steps to complete the hardware assembly for each SDM-2009:
1.Place the SDM-2009(s) on a convenient table top near an AC wall socket or
power strip in a well ventilated area. A location near a window that can be
opened is best so that the exhaust can be routed to the window.
2.If you have more than one SDM-2009, arrange them side by side as shown in
Figure 11 below. Each docking station has three connection tabs on the left
side near the bottom of the station and three connection recesses on the right
side near th e bottom of the station. The tabs from one docking statio n can be
pushed into the recesses on another station to mechanically connect the
docking stations.
3.Insert the round plug on the end of each AC adapter’s cable into the power
jack on the back of each SDM-2009.
NOTE:If you have multiple SDM-2009s and are using a 4-port AC adapter,
plug each of the round plugs on the end of the AC adapter into the
power jack on th e back of 4 separate SDM-2009s.
4.Connect the AC adapter’s wall plug into a wall AC socket or power strip.
5.Install th e air filter to the AI R fitting of ea ch SD M-2009 so that th e ar row on the
filter that in dicates dir ection of fl ow is poin ting toward s t he AI R fitt ing. Push the
open end of the flexible tube that is on one end of the filter onto the AIR fitting
on the back of the SDM-2009.
6.Install the 10 foot long 5/16 inch ID flexible tube that is included with each
Hardware Setup • 1 2
SDM-2009 on the exh aust fitting of e ach SDM-2009 . Route the t ube to an area
where the exhaust can b e safe ly d isperse d, such as an ope n w indow. Exhaust
tubing from multiple units can be daisy chained together in a manifold for more
convenient operation. In this case, the check valves that are provided with th e
docking stations need to be used. See "Assembling a Manifold for Multiple
SDM-2009 Units" below for instructions.
CAUTION:The maximum recomme nded le ngth for th e exha ust tube is 10 feet. Do
not use more than 10 feet of tubing or tubing with an ID of less than
5/16 inch for the exhaust tube or the bump test and calibration
accuracy will be adversely affected. The exhaust tube that is shipped
with the SDM-2009 has an ID of 5/16 inch and is 10 feet long.
7.Install the 3 foot long 3/16 inch ID tube that is included with the SDM-2009 on
the GAS fitting. GAS tubing from multiple units can be daisy chained together
in a manifold for more convenient operation. See "Assembling a Manifold for
Multiple SDM-2009 Units" below for instructions.
8.Connect the provided USB cord from the type B USB port on the bac k of the
SDM-2009 to an available USB port on your computer or to a USB hub that is
connected to your computer. See “USB Hub Require ment s for M ultipl e Station
Systems” on page 11 for more information about USB hubs.
Assembling a Manifold for Multiple SDM-2009 Units
Up to 10 SDM-2009s can b e manifo lde d toget her. The GAS fitting can be manifolde d
to allow the stations to draw from one calib rati on cyli nd er. The exhaust line s ca n al so
be manifolded together and be routed away from the docking stations with only one
10 foot long 5/16 inch piece of tubing.
Each SDM-2009 is sh ipped with exha ust tubing, GA S tubing, T-fittings in two differ ent
sizes, and a check valve.
Exhaust Tubing
Reference Figure 12 for the in structions below.
1.Cut a 1- 2” pi ece o f 5/ 16 in ch tubi ng fo r each SDM- 2009 a nd con ne ct it to each
unit’s exhaust fi tting.
2.Connect the provided check valves to the short pieces of tubing already
installed at the exhaust fittings. Be sure that the arrows that appear on the
check valve are pointing away fr om the exhaust fitting.
3.Cut a 2-3” piece of 5/16 inch tubing for every SDM-2009 except the first one
and connect it to the other end of the check valve.
4.Insert the larger T-fittings into the tubing so that the remaining two ports on the
T-fitting are perpendicular to the exhaust tube and check valve.
5.For the first SDM-2009, cut a 9-10” piece of 5/16 inch tubing and connect it
from the check valve on the first SDM-2009 to the closest port on the second
SDM-2009’s exhaust T-fitting.
13 • Hardware Setup
6.Use one of the provided 10 foot lengths of tubing and connect it to the T-fitting
7 inches
T-Fitting
10 Foot Max
Exhaust Tube
To Op en Are a
9 inc hes
AIR
GAS1
GAS1
Unit 2Unit 10
Check Valve
Unit 9
GAS1
AIR
AIR
GAS1
AIR
Unit 1
Figure 12: Exhaust Tubing Connections
on the last SDM-2009 . The maxim um recomm ended e xhaust tub e lengt h is 10
feet.
7.For the r emaini ng SDM- 2009s, cut 6-7” pi eces of 5/16 inch tubing and con nect
the remaining T-fittings.
Hardware Setup • 1 4
GAS Tubing
Reference Figure 13 for the instructions b elow.
1.Cut a 3-4” piece of 3/16 inch tubing for each GAS fitting on every SDM-2009
except the last one.
2.Connect the tubing to the GAS fitting on every SDM-2009 except the last one.
3.Insert the smaller T-fittings into the tubing so that the remaining two ports on
the T-fitting are perpendicular to the GAS fitting.
4.Cut 6-7” pieces of 3/16 inch tubing and connect the T-fittings.
5.Cut o ne m or e 6- 7” piece of 3/16 inch tubing and conne ct th e GAS fitting of the
last SDM-2009 to the T-fitting from the second to last SDM-2009.
6.Cut 1 piece of 3/16 inch tubing long enough for easy access to the calibration
cylinder and connect it to the T-fitting on the first SDM-2009.
15 • Hardware Setup
Unit 1
Unit 10
Unit 2
Unit 9
GAS1
Unit 6
Unit 5
AIR
GAS1
GAS1
GAS1
AIR
GAS1
AIRAIR
GAS1
GAS1
AIRAIR
AIR
GAS1
GAS1
AIR
AIR
GAS1
AIR
Unit 7
Unit 4
Unit 8
Unit 3
Figure 13: Gas Manifold Connections
Hardware Setup • 1 6
Connecting Calibration Gas
The GAS fitting on the back of the docking station is designed to be used with a
calibration gas cylinder that is fitted with a demand flow regulator. The AIR fitting may
be used with a demand flow regulator and a cylinder of zero emissions air, but this is
not normally necessary since the docking station will generally be in a fresh air area.
The type of calibratio n g as cyli nd er u s ed de pen ds on the gas sensors installed in the
instrument bein g used with th e calibration station . A 4-gas mix, LEL/Oxygen/CO/H2S,
is used if the instrument being used with the calibration station is a 4-gas unit. If the
instrument does not have an H2S channel, then a 3-gas mix, LEL/Oxygen/CO, is
used. Although a 4-ga s cylinder will work for an instrume nt of any gas combina tion, if
you have multiple instruments with various gas combinations, you may want to keep
a 4-gas cylinder a nd a 3-gas cylinder to help preserve the charcoal filter on the CO
sensor in instruments without an H2S channel. Use Table 2 below as a guide in
determining which calibration gas cylinder is appropriate for your system. See “S pa re
Parts List” on page 108 for a list of available cylinder part numbers.
Table 2: Recommended Gas Cylinders
Typical Instrument Types
Recommended Calibration Gas
Cylinder(s)
LEL/Oxy/H2S/CO 4-gas mix with CH4/Oxy/H2S/CO
LEL/Oxy/CO 3-gas mix with CH4/Oxy/CO
LEL/Oxy/H2S 4-gas mix with CH4/Oxy/H2S/CO
LEL/Oxy3-gas mix with LEL/Oxy/CO
Oxy/CO3-gas mix with LEL/Oxy/CO
Oxy/H2S4-gas mix with LEL/Oxy/H2S/CO
To connect calibration gas to the SDM-2009, do the following:
1.If the area around the docking station is not considered a fresh air area (an
area free of combustible and toxic gases and of normal oxygen content,
20.9%) install a tube not longer than 10 feet on the filter attached to the AIR
fitting on the back of the docking station and route it to a fresh air area or
connect a cylinder of zero air with a demand flow regulator to the AIR fitting.
2.Install the demand flow regulator on the calibration gas cylinder.
3.Connect the demand flow regulator to the GAS inlet fitting using the 3 foot
length of 3/16 inch ID sample tubing provided with the docking station.
17 • Hardware Setup
EAGLE 2 and GX Type Instrument Connection
There are 2 different versions of the Docking Station PC Controller program. The
SDM-E2 Docking Station PC Controller program is intended for use with SDM-E2
docking stations and EAGLE 2 instruments. The SDM-GX Docking Station PC
Controller program is intended for use with GX t ype docking stations and GX type
instruments.
Bump tests and calibrations can be performed on multiple EAGLE 2s and GX type
instruments at the same time. The SDM-E2 Docking Station PC Controller program
controls the EAGLE 2 functions while the SDM-GX Docking Station PC Controller
program controls the GX type instrument functions. Both programs can be used
simultaneousl y on your PC. Fo llow all i nstructions i n the SDM- E2 Docking Station PC
Controlled Conf iguration Op erator’s Manual as well as th e instructions in thi s manual.
Installing the SDM-GX Docking Station
PC Controller Software
1.Launch Windows®.
2.Exit from all applications and open windows.
3.There are two ways to install the SDM-GX Docking Station PC Controller
Software: by using the SDM-2009 product CD or by using the SDM-GX
Docking Station PC Controller Software Installation CD.
•If you are using the SD M-2009 Produc t CD, insert the Product CD int o your
computer’s CD-ROM drive. The CD will automatically open revealing
several folders. Open the S DM-GX Docking S tation PC Controller Software
folder, double click on setup.exe and continue with step 4.
•If you are using the SDM-GX Docking Station PC Controller Software
Installation CD, it will automatically begin the installation process. Insert it
in your computer’s CD-ROM drive, then continue with step 4.
4.The SDM-GX Docking Station PC Controller InstallShield Wizard comes up to
guide you through installation. Click Next to proceed to the License Agreement
window.
5.Read the license agreement and click the agreement acceptance selection
box, then click Next to proceed to the Customer Information window.
6.Enter a user name and organization and select if you want to install the
program for all users on the computer or just for your user account, then click
Next to proceed to the Destination Folder window.
7.The default installation folder (C:\Program Files\SDM-GX\) is displayed. If you
want to install the software in the default folder continue with step 8. If you
want to install the software in a different location, click Change and choose a
new installation folder and then continue with step 8.
8.Click Next to proceed t o the Ready to Install the Program window.
EAGLE 2 and GX Type Instrument Connection • 1 8
9.Review the installation settings. If they are OK, click Install and the installation
process will begin. If you want to change installation settings, click Back and
change them to the desired settings.
10. During software installation, the installation program may find newer versions
of Windows files on your computer than those in the Installation CD. If this
happens, the installation software will ask you if you want to keep these newer
files. Click Yes to do so.
11. A Device Driver Window will appear prompting you to install necessary drivers.
You cannot continue the installation without installing the drivers.
Figure 14: Device Driver Installation
12. Click Next to install the drivers. The Wizard will find the appropriate drivers.
19 • Installing the SDM-GX Docking Station PC Controller Software
13. If this is the first time you are installing a PC Controller Program, a window will
appear saying that the file did not pass Windows logo testing. Click Continue
Anyway.
14. Click Finish once the drivers are successfully installed.
Figure 15: Finish Device Driver Installation
The installation will continue.
15. Follow the on-screen instructions to complete software installation.
16. To complete the driver installation, ensure that all of your docking stations are
connected to your computer.
Installing the SDM-GX Docking Station PC Controller Software • 20
17. Turn on a docking station by pressing and holding the POWER button for at
least 1 second.
18. The first time an SDM-2009 is turned on after being connected to the
computer, a Found New Hardware window will appear.
NOTE:The following instructions do not apply to computers running Windows
Vista or Windows 7. The hardware driver will automatically in stall and
you will be notified that the USB device is ready for use.
Figure 16: Found New Hardware Wizard
19. Select the “Install the software automatically” option and click Next.
21 • Installing the SDM-GX Docking Station PC Controller Software
20. The wizard will search for the driver files.
Figure 17: Searching
21. Once the files have been found, a window will appear saying that the file did
not pass Windows logo testing. Click Continue Anyway.
Figure 18: Windows Logo Error
22. The installation will continue. Click Finish when the installation has comple ted.
23. Repeat steps 17 through 22 for every docking station in your system.
Installing the SDM-GX Docking Station PC Controller Software • 22
Launching the SDM-GX Docking Station
PC Controller Program
1.Click Start on the Windows Icon Tray, then select Programs/SDM-GX
Docking Station PC Controller. You may also double click the shortcut
created on yo ur desktop. The SDM-GX Docking Station PC Controller
Program is launched and the main program window appears.
Figure 19: Main Program Window
2.If you are starting the software for the first time, a message window appears
informing you that a database has been created. Click OK in that window.
When starting the software for the first time, or if you wish to change the
software setup, proceed to the next section, "Overview of the SDM-GX
Docking Station PC Controller Program".
If you have al ready performed the software setup, proceed to “Connecting
Instruments to the PC Controller Program” on page 36.
23 • Launching the SDM-GX Docking Station PC Controller Program
Overview of the SDM-GX Docking Station PC
Controller Program
This section pr ovides a brief overview of the program and a description of the main
program window. Instructions for using the program are given in other parts of this
manual. Use this section to become familiar with the main program window, but
before attempting to use the program, make sure to perform the program setup as
described in “Setting Up the PC Controller Program” on page 26.
The PC Controller Program is designed to provide manual or automated data
downloading, bump testing, and calibration of the GX-2009. It also provides the
capability to update instrument parameters such as alarm points. The program
compiles a database of the instruments that have been connected to the program
which consists of calibration and bump test history, logged data (if a unit’s data is
downloaded), an d memos that record user entered information about an ins trument
at a particular time.
The GX-2009 is a gas monitoring device for combustible gas, O2, H2S, and CO. It
stores logged data and calibration information in its internal memory. See the GX-
2009 Operator’s Manual for a complete description of the GX-2009.
PC Controller Program Functions
The PC Controller Program has the following basic functions accessed or controlled
by buttons in the main program screen.
•Gas cylinder setup accessed by the Cylinders control button
•Program and system configuration accessed by the Config control button (this
function is password protected)
•Data viewing accessed by the Logs control bu tton
•Instrument bump testing controlled by the Bump test control button
•Instrument calibration controlled by th e Calibration control button
Overview of the SDM-G X Docking Station PC Controller Program • 24
Main Program Window
Message Area
SDM-2009/Instrument
Display Area
The main program window is where the current status of the program operation is
shown. Any SDM-2009s and instruments that are currently connected to it and any
instruments that have previously been but are not currently connected to it are also
shown. The program’s various functions are initiated or accessed from this screen.
The current date and time are always displayed in the lower right corner of the main
program window.
Control Buttons
The Cylinders, Config, Logs, and Exit control buttons are located along the top
right of the window. The Bump test and Calibration control buttons are located in
the lower right of the window along with the Force Calibration selecti o n box. The use
of these control buttons and the Force Calibration selection box is described in other
parts of this manual.
SDM-2009/Instrument Displa y Are a
The large middle portion of the main program window displays SDM-2009s and
instruments that are currently connected to the PC Controller Program and
25 • Overview of the SDM-GX Docking Station PC Controller Program
Figure 20: Parts of the Main Program Window
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.