RKI Instruments SC-01 User Manual

SC-01 Data Logger
Management Program
Operator’s Manual
Part Number: 71-0138RK
Revision: P1
Released: 9/12/07
www.rkiinstruments.com
Warranty
RKI Instruments, Inc., warrant s gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance for a period of one year fr om date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or repla ced, at our op tion , free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Absorbent cartridges Batteries Pump diaphragms and
Filter elements
valves Fuses
Warrant y is voi ded by abuse including mechanical damage, alteration, rough handling, or repairs procedures not in accordance with the instruction manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
HIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
T
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
NSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO THE WARRANTY
I
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and part s sold to users o nly by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor and our warranty is limited to replacement of parts or our complete goods.
Table of Contents
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installing the SC-01 Data Logger Management Program . . . . . . . . . . . 7
IrDA Downloading Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Installing an IrDA Adapter Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Windows
®
Wireless Link Operation Note . . . . . . . . . . . . . . . . . . . . . 9
Using a USB or Serial Port With an IrDA Cable Installed . . . . . . . . 10
Launching the Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Control Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Download Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Instrument Information Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Data Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Last Calibration Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Set Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Downloading Data from the SC-01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
SC-01 Data Logging Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Viewing, Printing, & Deleting Data in the Last Calibration Window. . 26
Viewing and Printing Last Calibration Data . . . . . . . . . . . . . . . . . . . 26
Deleting Last Calibration Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Viewing, Printing, Exporting, & Deleting Data in the Data Window . . 33
Data Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Calibration History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Table of Contents • 3
Event Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Sensor Change History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Interval Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Alarm Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Deleting Data in the Data Window. . . . . . . . . . . . . . . . . . . . . . . . . . 60
Changing the Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Changing SC-01 Instrument Pa rameters . . . . . . . . . . . . . . . . . . . . . . . 63
Changing the Appearance of the Program Screens . . . . . . . . . . . . . . 66
Spare Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
CAUTION: Read and understand this manual before using the
SC-01 Data Logger Management Program. Also read and understand the SC-01 Operator’s Manual included with the SC-01 portable gas detectio n unit.
4 • Table of Contents
Introduction
Using an advanced microprocesso r controlled detection system, the SC-01 Personal Single Gas Monit or detects the presence of one of a variety of toxic gasses depending on the sensor installed. The SC-01’ s compact size and easy-to-use design makes it ideally suited for a wide range of applications as described in the SC-01 Operator’s Manual, which you should read first before using the SC-01 Data Logger Management Program.
The SC-01 Data Logger Management Program do wnloads dat a stored in the SC-01 to a Windows-based PC. After the data has been downloaded, you can view, save, or print it using your computer and the SC-01 Data Logger Management Program.
The purpose of this manual is to explain how to use and set up the SC-01 Data Logger Management Program. You will learn how to:
install and launch the program
install the downloading cable (if needed)
download data from the SC-01
view, print, and export, and delete data
change data logging parameters
change the appearance of the program screens
change the color of graphed readings for a particular gas Before you get started, be sure to review the system
requirements in the next section.
WARNING: The SC-01 detects toxic gases which can be
dangerous or life threatening. When using the SC-01, you must follow the instructions and warnings in the SC-01 Operator’s Manual to assure proper and safe operation of the unit and to minimize the risk of personal injury.
Introduction • 5
CAUTION: The operator of this instrument is advised that if t he
equipment is used in a manner not specified in this manual, the protection provided by th e equipment may be impaired.
System Requirements
To use the GX-2003 Data Logger Management Program, your personal computer must meet the following requirements:
Operating Systems: Windows® 98, Windows® 2000, or
®
Windows
XP
Processor: IBM
®
compatible PC running Pentium® 2 or
higher.
Memory: 32 MB RAM minimum
Available Har d Disk Space: 32 MB minimum
CD-ROM Drive
Infrared port or Serial port and a serial/IrDA adapter cable or USB port and a USB/IrDA adapter cable
6 • System Requirements
Installing the SC-01 Data Logger Management Program
1. Launch Windows®.
2. Exit from all applications and open windows.
3. Insert the SC-01 Data Logger Management Program
Installation CD in your computer’s CD-ROM drive.
4. After a few seconds, a screen appears indicating that the
SC-01 InstallShiel d Wiza r d is pr epar in g to in stall th e da ta logging program, then the SC-01 InstallShield Wizard window appears to guide you through installation.
Figure 1: SC-01 InstallShield Wizard
5. Follow the on-screen instructions in the Installation Wizard
window to install the program.
Installing the SC-01 Data Logger Management Program • 7
6. If the InstallShield Wizard finds versions of Windows® files
on your computer newer than those in the installation CD, it will ask you if you want to keep these newer files. Click “Y e s”.
7. When the InstallShield Wizard indicates that installation is
complete, click the “Finish” button.
8. Eject the installation CD from the CD-ROM drive a nd store it
in a safe place.
IrDA Downloading Cable
The SC-01 communicates with a computer via an on-board infrared data port that complies with IrDA protocol standards.
NOTE: If your computer has a built -in infrared port, you do not
need an adapter cable to download data.
If your computer does not have an infrar ed port, you will need to install an IrDA/serial adapter cable or an IrDA/USB adapter cable on your computer to download data to the PC from your SC-01. Both types of cables are availab le from RKI In strument s, Inc. See the Spare Parts List at the end of this manual for the RKI part numbers. These cables can also be found in many electronic supply stores and computer stores.
®
Some versions of Windows device drivers loaded in Windows
already have several infrared
®
and will automatically
recognize a cable during the installation process and guide you in installing the drivers. Other versions of Windows
®
will require you to load device drivers provided by the manufacturer of the cable during the install ation pr ocess. RKI makes n o warranty for the operation or compatibility of the drivers with any particular device.
8 • IrDA Downloading Cable
Installing an IrDA Adapter Cable
After instal ling the Data Logger Management Program, connect an IrDA/serial or IrDA/USB cable to your computer and follow the manufacturer’s instructions for installing the cable on your computer. Make sure the cable is compatible with your
®
Windows If you do not have instructions from the cable manufacturer for
installing your cable, see your Windows® documentation. In general, you must go to the Control Panel and use the Add Hardware Wizard to install the cable drivers.
operating system.
Windows
®
Wireless Link Operation Note
When using an IrDA adapter cable and the SC-01 Data Logging Software on a Windows
®
computer, it is necessary to make a special setting in the Wireless Link Configuration window for proper communication between the SC-01 and the Dat a Logger Management Program. This must be done before attempting to use the program. To make this setting, perform the following steps:
®
1. Click Start in the Windows
Icon Tray.
2. If Control Panel is available to select in the Start menu,
select it. The Control Panel will appear. If Control Panel is not selectable in the Start menu but
Settings is, select Settings, then select Control Panel. The Control Panel will appear.
3. Double click on Wireless Link. The Wireless Link
Configuration window will appear.
4. Click on the Image Transfer tab.
IrDA Downloading Cable • 9
5. Deselect the selection box for “Use Wireless Link to transfer
Deselect this box
images from a digital camera to your computer.”
Figure 2: Image Transfer Tab
6. Click OK.
7. Close the Control Panel window.
Using a USB or Serial Port With an IrDA Cable Installed
When an IrDA cable is inst alled on a USB or serial port, that port is always being used by the cable. If you need to use that port for another device, you must disable the IrDA/serial cable or
IrDA/USB cable using Windows® and then shut down your computer before disconnecting it, starting your computer up again, and then using the port with another device or cable. For example, if you have an IrDA/Serial cable installed in a serial port and want to download a different device using a standard straight through serial cable, you will need to disable the IrDA cable, remove it and inst all the straight throug h serial cable, then restart your computer. When you wish to download the SC-01
10 • IrDA Downloading Cable
again, remove the straight through cable and reconnect the IrDA cable, enable the IrDA cable using Windows your computer. See your IrDA cable and Windows
®
, and then restart
®
documentation for instructions to enable and disable an IrDA device. In general, you can enable or disable a device by going to the Control Panel, double clicking System, selecting the Hardware Tab, clicking Device Manager, double clicking the IrDa cable under Infrared Devices, and making the appropriate selection in the Device usage field. The process may differ
®
slightly in your particular Windows
system.
IrDA Downloading Cable • 11
Launching the Program
1. Click Start on the Windows® Icon Tray, then select
Programs, then select the SC-01 icon. Your operating
system may also have a shortcut installed in the Start menu.
2. The program will launch and the Download window will
appear.
Figure 3: The Download Window
3. For convenience, make a shortcut of the SC-01 Data Logger
Management Program and place it on the Windows desktop. See your Windows
about making shortcuts.
12 • Launching the Program
®
®
documentation for information
Control Buttons
This section describes the control buttons. Instructions for using each control button are given in other parts of this manual.
When the program is launched, it opens in the Download window. Along the right side of the Download window are six control buttons that access other windows in the program. Figure 4 below shows the various windows that you can access when you click these control buttons
Figure 4: Windows Accessed by Control Buttons
Control Buttons • 1 3
Download Button
If you are in another program window, clicking the Download button opens the Download window (see Figure 3 on page 12). The Download window has several download commands that allow you to perform various data retrieval functions with an instrument that is connected to the program. Data can be retrieved from the instrument, data can be cleared from the instrument, and the instrument can be turned of f. The Download State area on the left side of the window displays various messages about operations in progress or completed and the condition of the program’s connection to the instrument. See “Downloading Data from the SC-01” on page 19 for a complete description of downloading data from a SC-01.
Instrument Information Button
Clicking the Instrumen t Information button opens the Instrument Information window.
Figure 5: Instrument Information Window
14 • Control Buttons
The Instrument Information window displays various instrument parameters for an instrument that has been downloaded using the Complete Download or Inst rument Informati on download commands in the Download window and is currently connected to the program. If an instrument is turned off after being connected to the program, then the program will lose the connection with the instrument and the fields in th e Instrument Information window will become blank.
Data Button
Clicking the Data button opens the Data window.
Figure 6: Data Window
In the Data window, you can view, print, export, and delete data that has been downloaded from instruments. The following types of data files are accessible in the Data window:
Control Buttons • 1 5
Calibration History Files
A calibration history file is sa ved for each instr ument that has been downloaded. It records the calibration information for every calibration that was downloaded. The SC-01 can save information for up to the 20 most recent calibrations in its memory.
Alarm Event Files
Alarm event files record gas alarm events that have been downloaded from instruments. The SC-01 can save up to 100 alarm events in it’s memory.
Trouble Event Files
Trouble event files record sensor failure and system failure events that have been downloaded from instruments. The SC-01 can save up to 100 trouble events in it’s memory.
Sensor Change History Files
A sensor change history file is created for every downloaded instrument whose downloaded data indicates that the sensor has been replaced or changed to a different ty pe (gas). The sensor replacements /changes a re logged to pr ovi de a reco rd of the instruments’ sensor histor y.
Interval Trend Data Files
Interval trend data is lo gged at the interval trend ti me saved in the SC-01. Each logged point is an average reading over the previous time interval.
Alarm Trend Data Files
Data in an alarm trend data file is logged every five seconds and is logged around an al arm e vent. Each lo gged read ing is the peak reading during the previous f ive seco nd per iod. The SC-01 can save up to 8 alarm trend data files in its memory.
16 • Control Buttons
Last Calibration Button
Figure 7: Last Calibration Window
Clicking the Last Calibration button opens the Last Calibration window.
The Last Calibration window stores the information for the most recent successful calibration for each SC-01 that has been downloaded. You can display the information three ways by using the Need Calibration, Calibration Date, or Calibration Record selection buttons. You can also print the information if you select the Need Calibration or Calibration Date display options.
Control Buttons • 1 7
Set Button
Clicking the Set button opens the Set window.
In the Set window, you can perform the following function s:
Alter the appearance of the program windows with the font
and color buttons
Change the instrument serial number, user ID, or interval
trend time of a connected instrument by editing the fields in the SC-01 Status frame and then clicking the Update button
Update the date and t ime in a con nected instrument t o be the
same as the computer’s by using the Date/Time Set button
18 • Control Buttons
Figure 8: Set Window
Downloading Data from the SC-01
You have the option of downloading data manually or automatically. If you want to download data using the automatic download feature, click the Aut omatic Download selection box in the Download window before attempting to download data from the SC-01. Remember that if Automatic Download is selected, the Instrument Information window will remain blank and the SC-01 will turn off automatically after the data is downloaded. If you want the data in the SC-01 to be cleared auto matically af t er an automatic download, select the Automatic Data Removal box. The Automatic Data Removal box is only available for selection if the Automatic Download box is selected.
Click here for
automatic
downloads
Figure 9: Automatic Download Selection Box
To download data from a SC-01:
1. Launch the SC-01 Data Logger Management Program (see
“Launching the Program” on page 12). The Download window displays.
2. Place the SC-01 within an inch or two of the infrared port on
your computer aligning the infrared port on the side of the SC-01 with the infrared port on your computer.
If your computer does not have a built in infrared port, place the SC-01 within an inch or two of the infrared port on the IrDA adapter cable a s shown in Figure 10 below , aligning the infrared port on the side of the SC-01 with the infrared port on the cable.
Downloading Data from the SC-01 • 19
r
SC-01
Infrared Port
IrDA/USB or IrDA/Ser ial C able
USB Connector
Cable 'sInfrare d Port
Serial Connector
To Computer USB Port
To Compute Serial Port
Figure 10: Aligning the SC-01 With the Cable Infrared Receiver
3. Press and hold the POWER button on the SC-01 until you
hear a beep, then release it. The SC-01 will begin it’s power up sequence.
4. If the CAL. LIMIT menu item in the SC-01’s Setup Mode is
on, then one of the two following startup sequences takes place. If CAL. LIMIT is turned off, the startup sequence in S tep 5 below takes place. See the SC-01 Operat or’s Manual for a description of the CAL. LIMIT Setup Mode menu item.
If calibration is past due, t he following screen appears wit h “CAL.” flashing and the LEDs and buzzer pulsing several times alerting you that the SC-01 is due for calibration. If this screen appears, wait for the buzzer and LED pulsing to stop, then press and release the POWER button to continue in the startup sequence.
CAL. LIMIT
CAL.
The SC-01 will then show the current date for a few seconds before attempting to connect to the computer.
If the calibration is up to d ate , the SC-01 t hen displays th e calibration due date, how many days until the next
20 • Downloading Data from the SC-01
calibration, and the current date bef ore attempting to connect to the computer.
5. If the CAL. LIMIT menu item in Set up Mod e is turned off, the SC-01 shows “WARMUP” in the upper left of the LCD for a few seconds, then displays the current date for a few seconds before attempting to connect to the computer.
CAUTION: If the instrument gives a low battery warning or
dead battery alarm, change the alkaline batteries before attempting to download data. Se e the SC-01 Operator’s Manual for the battery alarm indications.
6. If a successful connection between the SC-01 and the computer occurs, the SC-01 displays the screen shown on the left in Figure 1 1 below, the Connect light in the program’s Download window turns green after a few seconds and “SC­01 Connection Successful.” displays in the Download State area of the Download window as shown on the right in Figure 11 below.
TRANSMIT
SC-01 Display
Figure 11: SC-01 and Program Connection Message
In addition, the Windows
®
icon tray will indicate that a
wireless connection is in effect.
7. If you selected Automatic Download, the downloading process begins automatically after a successful connectio n
Program Indication
is made. If Automatic Download is not selected, the Complete
Download, Instrument Information, and Power Off download command buttons become active.
Downloading Data from the SC-01 • 21
Loading...
+ 47 hidden pages