RKI Instruments RI-215A User manual

Part Number: 71-0045RK Revision 0
RI-215A
Operator’s Manual
Released: 10/3/05
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Absorbent cartridges Batteries Pump diaphragms and valves Filter elements Fuses
W arranty is voided by abuse including mechanical damage, alteration, r ough handling, or repairs procedures not in accor dance with the instruction manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES
AND REPRESENTATIONS OR LIABILITIES ON THE PART OF LIMITED TO THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE FOR INDIRECT CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY
, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS
RKI INSTRUMENTS, INC., INCLUDING BUT NOT
. IN NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE
.
This warranty covers instruments and parts sold to users only by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnication for any accident or damage caused by the operation of this gas monitor and our warranty is limited to replacement of parts or our complete goods.
2 • Warranty RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
About the RI-215A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measuring Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Maintenance Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Alarm Adjust Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
About this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
External Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Cable Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Internal Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Printed Circuit Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
IR Gas Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Fuse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 3: Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Mounting the RI-215A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Wiring the RI-215A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Notes on Wiring the RI-215A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Connecting the Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connecting the RI-215A to a Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connecting the RI-215A to a Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Connecting the RI-215A to an External Alarm Device. . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Introducing Power to the RI-215A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Warm Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Preparing to Set the Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Chapter 4: Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Alarm Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Responding to Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Alarm Point and Ventilation Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Alarm Adjust Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
RI-215A Operator’s Manual Table of Contents • 3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Verifying the Alarm Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Adjusting the Alarm Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Chapter 5: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Daily. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Quarterly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Replacing Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Replacing the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Replacing the Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Preparing for Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Setting the Response Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Returning to Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Appendix A: Changing the Detection Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4 • Table of Contents RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Chapter 1: Introduction
Overview
This chapter briey describes the RI-215A. This chapter also describes the RI-215A
Operator’s Manual (this document). Table 1 at the end of this chapter lists the specications
for the RI-215A.
About the RI-215A
The RI-215A Air Quality Monitor detects the presence of CO2 (carbon dioxide) using an
infrared (IR) gas sensor. The RI-215A shows increasing levels of CO
display (LCD). It also displays a visual alarm should levels of CO
scale (e.g., 2,000 ppm). Alarm contacts on the RI-215A are also provided, which enable
you to connect an external alarm to the unit.
Three ranges of gas detection are available for the RI-215A: 0 - 2,000 ppm, 0 - 5,000 ppm,
and 0 - 9990 ppm. Detection ranges are set at the factory. When you ordered the RI-215A,
you should have specied one of these gas detection ranges for your unit. It becomes
necessary to change the detection range, See “Appendix A: Changing the Detection
Range” on page 24.
on its liquid crystal
2
exceed the unit’s full
2
The RI-215A operates in three modes: Measuring Mode, Maintenance Mode, and Alarm
Adjust Mode.
Measuring Mode
Measuring Mode is the mode in which the RI-215A detects, measures, and displays CO
(normal operation).
2
Maintenance Mode
Maintenance Mode allows you to perform maintenance tasks. These tasks include:
adjusting the zero setting
adjusting the span setting
changing the detection range
CAUTION: The RI-215A is not in operate as a gas monitor when you are in Maintenance Mode.
Alarm Adjust Mode
Alarm Adjust Mode allows you to verify and change the alarm point.
RI-215A Operator’s Manual Overview • 5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Specications
Table 1 lists specications for the RI-215A.
Power Source 24 VAC or 24 VDC
Table 1: Specifications
Target Gas CO Sampling Method Diffusion
Area Classication Indoor, nonhazardous locations Signal Output 4 to 20 mA Accuracy ± 5% of full scale
1
Fuse
1
A 0.5A, 250 V fuse is an acceptable replacement.
About this Manual
The RI-215A Operator’s Manual uses the following conventions for notes, cautions, and
warnings.
NOTE: Describes additional or critical information.
(carbon dioxide)
2
0. 5A, 125 V (5 mm x 20 mm)
CAUTION: Describes potential damage to equipment.
WARNING: Describes potential danger that can result in injury or death.
6 • Specifications RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Chapter 2: Description
Overview
This section describes the external and internal components of the RI-215A.
External Description
Case
The RI-215A’s steel case is radio frequency (RF)-resistant. It is suitable for installation in
indoor areas where general purpose equipment is in use.
The case has a front housing and a base plate, which snap together. The front housing has
a LCD window, which is shown in the front view of Figure 1.
LCD Display Window
Cable Connector Entry Slot
Front View
Cal Gas Inlet Rubber Cover
Bottom View
Figure 1: Component Location of the RI-215A
RI-215A Operator’s Manual Overview • 7
Mode Switch
Up/Alarm Switch Down Switch
Set/Calibration Switch
Back View
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
On the bottom of the front housing is a series of slots and a calibration port. The slots
allow the target gas sample to enter the IR gas sensor in the RI-215A. The calibration port
1
is covered with a rubber cap. The calibration port accepts
exible tubing. A series of switches are accessible to the right of the calibration port, as
shown in Figure 1 on page 7. These switches include:
Mode (M)
/4-inch ID (inside diameter)
Up/Alarm (
Down (¢)
Set/Calibration (S)
The base plate has a wire entry slot for the cable assembly. The cable assembly connects
directly to the printed circuit board inside the case through the cable entry slot. The cable
assembly is used to connect to an external power source. You also use it to make
connections to an external alarm and to a monitoring or recording device. The base plate
has two mounting holes for convenient mounting on vertical surfaces at the monitoring
site.
Cable Assembly
The cable assembly connects to a six-pin receptacle on the printed circuit board through a
slot on the rear of the case. It allows you to make external wiring connections to the RI-
215A.
Internal Description
Printed Circuit Board
On the printed circuit board is located the IR gas sensor, LCD, and fuse. Connected to the
printed circuit board is the cable assembly.
6)
The printed circuit board converts the output of the IR gas sensor to a 4-to-20 mA signal
(that is proportional to the detection range of the RI-215A). The printed circuit board also
displays a digital reading on the LCD.
The printed circuit board has four switches: Mode (M), Up/Alarm (
Set/Calibration (S). These switches are located at the bottom of the printed circuit board
beneath the gas sensor. Table 2 lists each switch and its function.
Table 2: Model RI-215A Program Switches
Switch Function
Mode (M) Enters Maintenance Mode
Up/Alarm (
Down (¢) Decreases the setting on the LCD
Set/Calibration (S) Prepares a setting to be changed
6)
Increases the setting on the LCD
Displays the alarm point
Enters a new setting
6, Down (¢), and
8 • Internal Description RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
LCD
The LCD is mounted to the printed circuit board. The LCD shows the current gas reading
of the RI-215A’s target gas. The LCD also shows operational messages when the RI-215A
exceeds its full scale.
IR Gas Detector
The IR detector uses infrared light and an infrared sensor to detect CO2. A small infrared
light source is located in the detector. As CO
in a characteristic way as the concentration increases. The infrared sensor detects this
change in infrared light and the unit converts this to a gas concentration.
enters the detector, it absorbs infrared light
2
Fuse
One extractor-type fuse protects the AC line circuitry from short circuit or overload. The
fuse is rated at 0. 5A, 125 V. See “Parts List” on page 23.
RI-215A Operator’s Manual Internal Description • 9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Chapter 3: Installation
Overview
This chapter describes how to install the RI-215A at the monitoring site. This section also
includes procedures to make power and other wiring connections to the RI-215A.
Mounting the RI-215A
CAUTION: The RI-215A is suitable for installation in indoor areas where general purpose
equipment is in use.
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site:
Select a site near the sampling area in which the RI-215A is not likely to be
bumped or disturbed.
Choose a location not subject to signicant air movement, for example, near an air
conditioning or heating vent.
Mount the RI-215A in an area not subject to large temperature and humidity
changes.
Mount the RI-215A on the wall at least six feet from the oor to minimize the
possibility of someone breathing on the unit. The exhaled CO alarm.
Make sure there is sufcient room to make wiring connections through the back
of the unit, and to remove the unit from the base plate to perform maintenance, such as replacing the fuse.
2. For wall-mounting, remove the base plate of the RI-215A.
Using a small screw driver, press down on the small tab on top of the
Model RI-215A.
While pushing down on the tab, gently move the base plate away from the front
housing of the RI-215A.
CAUTION: Be very careful not to pull on the cable assembly connected to the RI-215A ’s internal
printed circuit board when removing the base plate. Pulling on the cable assembly may damage the printed circuit board.
may cause an
2
Completely separate the base plate from the front housing.
10 • Overview RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3. Use two screws to secur e the base plate to the wall. Figur e 2 below shows the location of the screw holes in the base plate.
3.08
Front Vie w
1.23
3.08
Right Side View
2.37
.25
.17 Diameter
Mounting Hole Location
Back Side
Figure 2: RI-215A Outline and Mounting Dimensions
4. Reassemble the base plate and the front housing of the RI-215A in the reverse order as described in step 2 above. Line up the bottom of the base plate with the front housing, then carefully snap the base plate and front housing together.
RI-215A Operator’s Manual Mounting the RI-215A • 11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Wiring the RI-215A
This section describes procedures to connect power to the RI-215A. This section also describes optional procedures to connect a controller or recorder and external alarm device to the RI-215A.
WARNING: To reduce the risk of electric shock, always verify that power to the RI-
Notes on Wiring the RI-215A
Use a stable 24 VAC or 24 VDC power source. Do not use a noisy power source. If the power source is noisy, use a line lter to eliminate the noise.
Use shielded cable for the output signal line in cases where noise is a potential problem.
The alarm relay contacts in the RI-215A are rated at 0.5A, 125 VAC. Do not use these contacts to control an electrical load larger than this rating. For large load contr ol, use the RI-215A alarm contacts to control a slave relay with an appropriate contact rating.
Figure 3 below shows the general wiring diagram of the RI-215A, and is not specic to any device to which the unit could be connected. The red wires of the cable assembly, as shown in Figure 3 below, connect to an external 24 VAC or 24 VDC power supply. The white wires serve as a relay contact and can be used to control an external alarm. The yellow and blue wires carry a 4 - 20 mA output signal. These wires can be connected to a gas monitoring controller or a recording device.
215A has been turned off before making any wiring connections to the RI­215A. Make sure to operate the RI-215A from the specied power voltage of 24VAC/VDC, and follow the local electric code when making all electrical connections.
RI-215A
1 (red)
2 (red) 3 (white)
4 (white)
5 (yellow)
4 - 20 mA Standard, 0 - 10V Optional
6 (blue)
24 VAC or 24 VDC Power
Normally Open Relay Contact
+
Recording Device
Figure 3: General Wiring Diagram of the RI-215A
12 • Wiring the RI-215A RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Figure 4 below shows how to wire the RI-215A to a controller that accepts 3-wire, 4-20 mA transmitters.
RI-215A
Controller Terminals
1 (red)
2 (red)
3 (white)
4 (white) 5 (yellow)
6 (blue)
Figure 4: Wiring Diagram of the RI-215A as a 3-Wire 4-20 mA Transmitter
+ 24 VDC
- (DC Ground)
4 - 20 mA In (S or FB)
Normally Open Relay Contact
Not Used
Connecting Power
1. Verify that power to the RI-215A has been turned OFF.
2. Connect the two red AC line cables to the 24 VAC or 24 VDC power source. Refer to the wiring diagram in Figure 3 on page 12.
3. If you do not plan to make any more wiring connections, then terminate the remaining cables appropriately to avoid an electrical short.
Connecting the RI-215A to a Recorder
Perform the following procedure to connect the RI-215A to a recorder. The output of the RI-215A is a 4-to-20 mA signal, which is proportional to the detection range.
NOTE: The standard output of the RI-215A is 4 - 20 mA. An optional output of 0 -10 V is
available and must be specied at the time of your order.
1. Verify that power to the RI-215A has been turned OFF.
2. Connect the yellow (+) and blue (-) wires to the contr oller or recor ding device. Refer to the wiring diagram in Figure 3.
3. If you do not plan to make any more wiring connections, then terminate the remaining wires.
RI-215A Operator’s Manual Wiring the RI-215A • 13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Connecting the RI-215A to a Controller
Perform the following procedure to connect the RI-215A as a 3-wire 4 - 20 mA transmitter to a controller.
1. Verify that power to the RI-215A has been turned OFF.
2. Connect one of the red wir es from the RI-215A to the +24V transmitter terminal at the controller.
3. Connect the other red wire to the - (DC ground) transmitter terminal at the controller.
4. Connect the yellow wire to the S or FB transmitter terminal at the controller.
Connecting the RI-215A to an External Alarm Device
CAUTION: The maximum current rating of the normally-open, dry contacts is 0.5 amp. If the
alarm device exceeds this rating, use the alarm contacts to control a relay with an appropriate current rating.
1. Verify that power to the RI-215A has been turned OFF.
2. Connect the white wires to the external alarm. Refer to the wiring diagram in Figure 3 on page 12.
3. If you do not connect the white wires to an alarm device, terminate these unused wires.
Start Up
This section describes procedures to start up the RI-215A and place the RI-215A into normal operation.
Introducing Power to the RI-215A
Complete all installation procedures described earlier in this chapter.
Verify that the wiring supplying power to the RI-215A has been connected correctly.
Turn on or plug in the incoming power at the power source.
Verify that the RI-215A shows a readout on its LCD as described in the next subsection.
Warm Up
After you have turned on the power to the RI-215A, the unit takes about one minute to begin operating. During this initial warm-up period, “CO2” is shown on the unit’s LCD.
After the initial warm up is completed, the gas sensor inside the RI-215A detects the presence of CO
the RI-215A also outputs a 4-to-20 mA signal (0 to 10 V optional) which corresponds to the gas concentration.
CAUTION: Allow the RI-215A to warm up for an additional 30 minutes before continuing.
, and displays the concentrations of this gas on its LCD. Simultaneously,
2
14 • Start Up RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Preparing to Set the Zero
Connect the sample tubing from your calibration kit to the hose barb on the xed ow regulator. The xed ow regulator should be a 0.5 LPM regulator.
NOTE: Do not connect the regulator to the calibration cylinder at this time.
Setting the Zero Reading
1. Locate the function switches on the bottom of the unit.
2. Press and hold the MODE switch (M) for three seconds to enter Maintenance Mode. The display will show “1” to indicate Zero Adjustment Mode.
3. Press the SET switch (S) to enter Zero Adjustment Mode. The display will show a reading with a ashing decimal point before the rst and third digits, for example “. 1.5“.
4. Remove the rubber cap covering the calibration gas inlet port at the bottom of the unit.
5. Insert the calibration tubing into the gas inlet opening at the bottom of the RI-215A and push the tubing over the calibration tting.
6. Screw a 100% nitrogen calibration cylinder (a CO
7. Allow the gas to ow for one minute or until the gas reading stabilizes.
8. Press the SET switch (S) and the unit will adjust the reading to zero. Be careful not to breath onto the RI-215A since the CO
9. Press the SET switch (S) and the display will again show a “1”.
10. Unscrew the regulator from the nitrogen cylinder.
11. To return to normal operation, press the MODE switch for three seconds to exit Maintenance Mode.
in your breath may affect the reading.
2
-free source) into the regulator.
2
RI-215A Operator’s Manual Start Up • 1 5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Chapter 4: Operation
Overview
This chapter describes Measuring Mode and Alarm Adjust Mode. Under Measuring Mode, alarm indications are discussed along with responding to alarms, analog output, and alarm point and ventilation control. Under Alarm Adjust Mode, verifying the alarm point and adjusting the alarm point are described.
Measuring Mode
When the RI-215A is in Measuring Mode (normal operation), it displays the current concentration of CO2on its LCD and outputs a 4 - 20 mA signal that corresponds to the
CO
concentration. The typical background CO2level is 200 - 400 ppm.
2
Alarm Indications
If the CO2concentration rises above the alarm point, the relay contact will actuate. If the gas concentration exceeds the full scale, the LCD shows:
Responding to Alarms
1. Follow your established procedure for handling high levels of CO2 gas.
2. When the display reading falls below the alarm point, the alarm circuit automatically resets itself.
16 • Overview RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Analog Output
The RI-215A’s analog output, available via the yellow and blue wires of the cable assembly (refer to Figure 3 on page 12), is proportional to the concentration of CO2 for all
gas ranges (e.g., 0 - 2,000 ppm, 0 - 5,000 ppm, and 0 - 9990 ppm), as shown in Figure 4.
Maximum Output
Should the concentration of CO scale (i.e., it exceeds the upper gas range of your unit), the highest output will be 22 mA,
or 11 V, depending on the output, current, or voltage, selected at the factory. The LCD of the RI-215A will show a series of brackets (upside down “Us”) between 20 mA (10 V) and 22 mA (11V). The standard output setting for the RI-215A is 4- 20 mA.
22 mA (11V)
20 mA (10V)
Output
signal
4 mA (0V)
0 ppm Gas
Figure 5: Model RI-215A’s CO
present in the monitoring environment be above full
2
Concentration
Level to Output Signal
2
2000 ppm 5000 ppm
9990 ppm
Alarm Point and Ventilation Control
The RI-215A’s alarm relay has a built-in hysteresis of 50 ppm (refer to Figure 6 on page 18). After the alarm point has been reached (e.g., 1000 ppm), the alarm relay will activate. As the CO2concentration begins to fall, the alarm relay will deactivate 50 ppm
below the activation point (e.g., 950 ppm). The 50 ppm hysteresis keeps the alarm contacts from “chattering” should the CO
RI-215A Operator’s Manual Measuring Mode • 17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
concentration hover around the alarm point.
2
So if the alarm relay contacts are wired to turn on a ventilation system when the concentration of CO2reaches the alarm point (e.g., 1000 ppm), the ventilation system will
remain on until the concentration of CO
falls to 50 ppm below the alarm point (e.g., 950).
2
Alarm Adjust Mode
Use the Up (6) switch to enter Alarm Adjust Mode.
Verifying the Alarm Point
Three ranges of gas detection are available for the RI-215A: 0 - 2,000 ppm, 0 - 5,000 ppm, and 0 - 9990 ppm. Detection ranges are set at the factory. When you ordered the RI-215A, you should have specied one of these gas detection ranges for your unit.
1000 ppm
950 ppm
CO
2
Closed
Relay contact
Figure 6: Model RI-215A’s Built-in Hysteresis of 50 ppm
Open
Open
Hysteresis
50 ppm
The alarm point of the RI-215A is also set at the factory (e.g., 1000 ppm). To display the current alarm point from the normal LCD, do the following:
1. Press and hold the Up ( (alarm verication screen). In this example, 10:00 indicates an alarm point of 1000 ppm. Note that the colon (:) will be ashing.
2. Release the Up (6)switch to return to the normal LCD. Do not hold the (6) switch for more than three seconds or the unit will display the Alarm Adjust screen.
Further note that the RI-215A’s alarm relay has a built-in hysteresis (lag) of 50 ppm. After the alarm point has been reached (e.g., 1000 ppm), the alarm relay will activate. As the
concentration begins to fall, the alarm relay will deactivate 50 ppm below the
CO
2
activation point (e.g., 950 ppm). The 50 ppm hysteresis keeps the alarm contacts from “chattering” should the CO
18 • Alarm Adjust Mode RI-215A Operator’s Manual
6) switch. The alarm point appears (e.g., 10:00) on the LCD
concentration hover around the alarm point.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Adjusting the Alarm Point
To adjust the alarm point, do the following:
1. Press and hold the Up ( screen. The LCD will ash the alarm point setting and the colon (:) alternately.
2. Use the Up ( want (e.g., 11:00 = 1100 ppm, 12:00 = 1200ppm, 12:30 = 1230ppm, 13:00 = 1300 ppm, etc.)
3. Press the SET switch to save the new alarm point. The unit returns to normal operation and the LCD displays the current gas reading.
6) and Down (¢) switches to adjust the alarm point to the setting you
6) switch for three seconds to display the Alarm Adjust
RI-215A Operator’s Manual Alarm Adjust Mode • 19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Chapter 5: Maintenance
Overview
This chapter describes maintenance and troubleshooting procedures for the RI-215A. It includes instructions for the replacement of the RI-215A’s gas sensor and fuse and calibration procedures. A spare parts list is located at the end of this chapter.
Preventive Maintenance
This section describes a preventive maintenance schedule.
Daily
Verify a display reading for CO2 in ppm. Investigate signicant changes in the reading. Typical background levels of CO
Quarterly
are 200 to 400 ppm, depending on location.
2
Calibrate the unit. See See “Calibration” on page 22.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable causes, and recommended action for problems you may encounter with the RI-215A.
The LCD does not show a gas reading or is blank.
Table 3: Troubleshooting the RI-215A
Symptoms Probable Causes Recommended Action
Power may have been terminated to the unit.
The unit may have been incorrectly connected to a power source.
The wrong power source is being used to run the unit.
The fuse may have blown.
The unit may be in Maintenance Mode.
1. Verify that the power source has been turned on and is working.
2. If a circuit breaker has been used in the installation of the RI-215A, make sure that the circuit breaker has not been activated.
3. Verify that the wiring to the power source has been connected correctly.
4. Make sure you have used the correct power source: 24 VAC or 24 VDC.
5. Open the unit and check the fuse. Replace the fuse if required. The RI-215A takes the following replacement fuse: 0. 5A, 125 VAC (5.2 mm x 20 mm).
6. The unit may be in Maintenance Mode. Press the MODE switch for three seconds, or until a gas reading appears.
7. If difficulties continue, contact RKI.
20 • Overview RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table 3: Troubleshooting the RI-215A
Symptoms Probable Causes Recommended Action
The LCD shows abnormally high or low CO2 readings but other
gas detection instru­ments show a normal level of CO
Unable to zero or span reading during calibration.
2.
Replacing Components
The section includes procedures to replace the sensor and fuse.
Replacing the Detector
The CO2 gas detector inside the RI-215A is a factory replacement part. Should you determine that the RI-215A’s gas detector is no longer working or is working improperly,
you will need to ship the RI-215A to RKI Instruments, Inc. to have the detector replaced.
Replacing the Fuse
1. Turn off the power to the RI-215A.
The RI-215A may need to be recalibrated.
The calibration cylinder is low, outdated, or defective.
1. Recalibrate the unit.
1. Verify that the calibration cylinder contains an adequate supply of sample.
2. If difficulties continue, contact RKI.
2. Remove the base plate of the RI-215A.
Using a small screw driver, press down on the small tab on top of the
Model RI-215A.
While pushing down on the tab, gently move the base plate away from the front housing of the RI-215A.
CAUTION: Be very careful not to pull on the cable assembly connected to the RI-215A ’s internal
printed circuit board when removing the base plate. Pulling on the cable assembly may damage the printed circuit board.
Completely separate the base plate from the front housing.
3. Find the fuse on the left side of the printed circuit board. Carefully remove the fuse from the bracketing holding it on the printed circuit board.
4. Insert a new fuse in the brackets. The replacement fuse must be 0. 5A, 125 V or 250 V (5 mm x 20 mm).
5. Reassemble the base plate and the front housing of the RI-215A in the reverse order as described in step 2 above. Line up the bottom of the base plate with the front housing, then carefully snap the base plate and front housing together.
6. Turn the power back on to the RI-215A.
RI-215A Operator’s Manual Replacing Components • 21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Calibration
This section describes how to calibrate the RI-215A. It includes procedures to prepare for calibration, set the zero reading, set the response (span) reading, and return to normal operation.
WARNING: The RI-215A is not an active gas monitoring device during the calibration
procedure.
Preparing for Calibration
Turn on the power to the RI-215A, and let the unit warm up for at least 30 minutes. Connect the sample tubing from your calibration kit to the hose barb on the xed ow regulator. The xed ow regulator should be a 0.5 LPM (liters per minute) regulator.
NOTE: Do not screw the regulator onto the calibration cylinder at this time.
Setting the Zero Reading
1. Locate the function switches on the bottom of the unit.
2. Press and hold the MODE switch (M) for three seconds to enter Maintenance Mode. The display will show “1” to indicate Zero Adjustment Mode.
3. Press the SET switch (S) to enter Zero Adjustment Mode. The display will show a reading with a ashing decimal point before the rst and third digits, for example “. 1.5“.
4. Remove the rubber cap covering the calibration gas inlet port at the bottom of the unit.
5. Insert the calibration tubing into the gas inlet opening at the bottom of the RI-215A and push the tubing over the calibration tting.
6. Screw a 100% nitrogen calibration cylinder (a CO
7. Allow the gas to ow for one minute or until the gas reading stabilizes.
8. Press the SET switch (S) and the unit will adjust the reading to zero. Be careful not to breath onto the RI-215A since the CO
9. Press the SET switch (S) and the display will again show a “1”.
10. Unscrew the regulator from the nitrogen cylinder.
11. To proceed with setting the span, press the Up ( indicates Span Adjustment Mode and continue with the next section, Setting the Response Reading.
12. To return to normal operation, skip to Returning to Normal Operation below.
in your breath may affect the reading.
2
-free source) into the regulator.
2
6) switch to display “2” which
Setting the Response Reading
1. With “2” showing in the display, press the SET switch (S) to enter Span Adjustment Mode. The display will show the current gas reading with a ashing decimal point before the second and third digits, for example “.6.50“.
2. Press the SET switch (S) again and the reading will start to ash.
22 • Calibration RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3. If not already in, insert the calibration tubing into the gas inlet opening at the bottom of the RI-215A and push the tubing over the calibration tting.
4. Screw the calibration cylinder into the regulator.
5. Allow the calibration gas to ow for one minute or until the gas reading stabilizes.
Parts List
6. If the gas reading does not match the gas concentration, use the Up ( Down (¢) switches to adjust the reading to the gas concentration.
7. Press the SET switch (S) to set the reading and exit Span Adjustment Mode. The display will show a “2”.
8. Unscrew the regulator from the calibration cylinder.
9. Remove the calibration tubing from the calibration tting on the RI-215A.
10. Allow time for the gas reading on the RI-215A display to return to a normal level.
6) and
Returning to Normal Operation
1. Press the MODE switch for thr ee seconds to exit the Maintenance Mode and return to normal operation.
NOTE: If you do not allow the gas reading to return to normal, then unwanted alarms
may occur.
2. Store the components of the calibration kit in a safe and convenient place.
Table 4 lists replacement parts and accessories for the RI-215A carbon dioxide transmitter.
Table 4: Parts List
Part Number Description
06-1252RK Sample tubing (1/4 x 1/8 in.; specify length when ordering) 43-4139RK Replacement fuse, 0. 5A, 250 V, 5 mm x 20 mm 81-0070RK-01 Calibration cylinder (2,000 ppm carbon dioxide; 34 liter) 81-0071RK-01 Calibration cylinder (5,000 ppm carbon dioxide; 34 liter) 81-0078RK-01 100% nitrogen calibration cylinder (34 liter) 81-1003RK Regulator, 0.5 liter/minute; continuous ow
(for 17 and 34 liter calibration cylinders)
RI-215A Operator’s Manual Parts List • 23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Appendix A: Changing the Detection Range
The RI-215A can be setup for one of thr ee detection ranges: 0 - 2,000 ppm; 0 - 5,000 ppm; 0
- 9990- ppm. When the RI-215A is ordered, the detection range must be specied. If it is
necessary to change the detection range in the eld, this can be done in Maintenance Mode.
WARNING: The RI-215A is not an active gas monitoring device during the full scale
adjustment procedure.
Perform the following to change the detection range.
1. With the RI-215A on and in Measuring Mode, locate the function switches on the bottom of the unit.
2. Press and hold the MODE switch (M) for 3 seconds to enter Maintenance Mode. The display will show “1” to indicate Zero Adjustment Mode.
3. Press the Up ( Mode.
4. Press the SET switch (S) to enter Full Scale Adjustment Mode. The display will show a full scale value with a ashing decimal point after the rst digit, for example “2.000“.
5. Use the Up ( until the desired full scale value is displayed.
6. Press the SET switch (S) to accept the displayed full scale value. The display will again show a “3”.
7. Press the MODE switch for 3 seconds to exit Maintenance Mode and return to Measuring Mode.
8. Calibrate the RI-215A as described in "Calibration" on page 22.
WARNING: The RI-215 must be calibrated after the detection range is changed. The
6) switch until the display shows “3” to indicate Full Scale Adjustment
6) and Down (¢) switches to scroll through the full scale value choices
RI-215A readings will not be accurate unless a calibration is performed.
24 • Parts List RI-215A Operator’s Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...