RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in
materials and workmanship, and performance for a period of one year from date of
shipment from RKI Instruments, Inc. Any parts found defective within that period will be
repaired or replaced, at our option, free of charge. This warranty does not apply to those
items which by their nature are subject to deterioration or consumption in normal service,
and which must be cleaned, repaired, or replaced on a routine basis. Examples of such
items are:
Absorbent cartridgesBatteries
Pump diaphragms and valvesFilter elements
Fuses
W arranty is voided by abuse including mechanical damage, alteration, r ough handling, or
repairs procedures not in accor dance with the instruction manual. This warranty indicates
the full extent of our liability, and we are not responsible for removal or replacement costs,
local repair costs, transportation costs, or contingent expenses incurred without our prior
approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES
AND REPRESENTATIONS
OR LIABILITIES ON THE PART OF
LIMITED TO THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE
FOR INDIRECT
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO
FUNCTION OR OPERATE PROPERLY
, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS
RKI INSTRUMENTS, INC., INCLUDING BUT NOT
. IN NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE
.
This warranty covers instruments and parts sold to users only by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of
this gas monitor and our warranty is limited to replacement of parts or our complete
goods.
3.Use two screws to secur e the base plate to the wall. Figur e 2 below shows the location
of the screw holes in the base plate.
3.08
Front Vie w
1.23
3.08
Right Side View
2.37
.25
.17 Diameter
Mounting Hole Location
Back Side
Figure 2: RI-215A Outline and Mounting Dimensions
4.Reassemble the base plate and the front housing of the RI-215A in the reverse order as
described in step 2 above. Line up the bottom of the base plate with the front housing,
then carefully snap the base plate and front housing together.
RI-215A Operator’s ManualMounting the RI-215A • 11
This section describes procedures to connect power to the RI-215A. This section also
describes optional procedures to connect a controller or recorder and external alarm
device to the RI-215A.
WARNING: To reduce the risk of electric shock, always verify that power to the RI-
Notes on Wiring the RI-215A
•Use a stable 24 VAC or 24 VDC power source. Do not use a noisy power source. If the
power source is noisy, use a line filter to eliminate the noise.
•Use shielded cable for the output signal line in cases where noise is a potential
problem.
•The alarm relay contacts in the RI-215A are rated at 0.5A, 125 VAC. Do not use these
contacts to control an electrical load larger than this rating. For large load contr ol, use
the RI-215A alarm contacts to control a slave relay with an appropriate contact rating.
Figure 3 below shows the general wiring diagram of the RI-215A, and is not specific to
any device to which the unit could be connected. The red wires of the cable assembly, as
shown in Figure 3 below, connect to an external 24 VAC or 24 VDC power supply. The
white wires serve as a relay contact and can be used to control an external alarm. The
yellow and blue wires carry a 4 - 20 mA output signal. These wires can be connected to a
gas monitoring controller or a recording device.
215A has been turned off before making any wiring connections to the RI215A. Make sure to operate the RI-215A from the specified power voltage of
24VAC/VDC, and follow the local electric code when making all electrical
connections.
Figure 4 below shows how to wire the RI-215A to a controller that accepts 3-wire, 4-20 mA
transmitters.
RI-215A
Controller Terminals
1 (red)
2 (red)
3 (white)
4 (white)
5 (yellow)
6 (blue)
Figure 4: Wiring Diagram of the RI-215A as a 3-Wire 4-20 mA Transmitter
+ 24 VDC
- (DC Ground)
4 - 20 mA In (S or FB)
Normally Open
Relay Contact
Not Used
Connecting Power
1.Verify that power to the RI-215A has been turned OFF.
2.Connect the two red AC line cables to the 24 VAC or 24 VDC power source. Refer to
the wiring diagram in Figure 3 on page 12.
3.If you do not plan to make any more wiring connections, then terminate the
remaining cables appropriately to avoid an electrical short.
Connecting the RI-215A to a Recorder
Perform the following procedure to connect the RI-215A to a recorder. The output of the
RI-215A is a 4-to-20 mA signal, which is proportional to the detection range.
NOTE: The standard output of the RI-215A is 4 - 20 mA. An optional output of 0 -10 V is
available and must be specified at the time of your order.
1.Verify that power to the RI-215A has been turned OFF.
2.Connect the yellow (+) and blue (-) wires to the contr oller or recor ding device. Refer to
the wiring diagram in Figure 3.
3.If you do not plan to make any more wiring connections, then terminate the
remaining wires.
Perform the following procedure to connect the RI-215A as a 3-wire 4 - 20 mA transmitter
to a controller.
1.Verify that power to the RI-215A has been turned OFF.
2.Connect one of the red wir es from the RI-215A to the +24V transmitter terminal at the
controller.
3.Connect the other red wire to the - (DC ground) transmitter terminal at the controller.
4.Connect the yellow wire to the S or FB transmitter terminal at the controller.
Connecting the RI-215A to an External Alarm Device
CAUTION: The maximum current rating of the normally-open, dry contacts is 0.5 amp. If the
alarm device exceeds this rating, use the alarm contacts to control a relay with an
appropriate current rating.
1.Verify that power to the RI-215A has been turned OFF.
2.Connect the white wires to the external alarm. Refer to the wiring diagram in
Figure 3 on page 12.
3.If you do not connect the white wires to an alarm device, terminate these unused
wires.
Start Up
This section describes procedures to start up the RI-215A and place the RI-215A into
normal operation.
Introducing Power to the RI-215A
•Complete all installation procedures described earlier in this chapter.
•Verify that the wiring supplying power to the RI-215A has been connected correctly.
•Turn on or plug in the incoming power at the power source.
•Verify that the RI-215A shows a readout on its LCD as described in the next
subsection.
Warm Up
After you have turned on the power to the RI-215A, the unit takes about one minute to
begin operating. During this initial warm-up period, “CO2” is shown on the unit’s LCD.
After the initial warm up is completed, the gas sensor inside the RI-215A detects the
presence of CO
the RI-215A also outputs a 4-to-20 mA signal (0 to 10 V optional) which corresponds to the
gas concentration.
CAUTION: Allow the RI-215A to warm up for an additional 30 minutes before continuing.
, and displays the concentrations of this gas on its LCD. Simultaneously,
Connect the sample tubing from your calibration kit to the hose barb on the fixed flow
regulator. The fixed flow regulator should be a 0.5 LPM regulator.
NOTE: Do not connect the regulator to the calibration cylinder at this time.
Setting the Zero Reading
1.Locate the function switches on the bottom of the unit.
2.Press and hold the MODE switch (M) for three seconds to enter Maintenance Mode.
The display will show “1” to indicate Zero Adjustment Mode.
3.Press the SET switch (S) to enter Zero Adjustment Mode. The display will show a
reading with a flashing decimal point before the first and third digits, for example
“. 1.5“.
4.Remove the rubber cap covering the calibration gas inlet port at the bottom of the
unit.
5.Insert the calibration tubing into the gas inlet opening at the bottom of the RI-215A
and push the tubing over the calibration fitting.
6.Screw a 100% nitrogen calibration cylinder (a CO
7.Allow the gas to flow for one minute or until the gas reading stabilizes.
8.Press the SET switch (S) and the unit will adjust the reading to zero. Be careful not to
breath onto the RI-215A since the CO
9.Press the SET switch (S) and the display will again show a “1”.
10. Unscrew the regulator from the nitrogen cylinder.
11. To return to normal operation, press the MODE switch for three seconds to exit
Maintenance Mode.
This chapter describes Measuring Mode and Alarm Adjust Mode. Under Measuring
Mode, alarm indications are discussed along with responding to alarms, analog output,
and alarm point and ventilation control. Under Alarm Adjust Mode, verifying the alarm
point and adjusting the alarm point are described.
Measuring Mode
When the RI-215A is in Measuring Mode (normal operation), it displays the current
concentration of CO2on its LCD and outputs a 4 - 20 mA signal that corresponds to the
CO
concentration. The typical background CO2level is 200 - 400 ppm.
2
Alarm Indications
If the CO2concentration rises above the alarm point, the relay contact will actuate. If the
gas concentration exceeds the full scale, the LCD shows:
Responding to Alarms
1.Follow your established procedure for handling high levels of CO2 gas.
2.When the display reading falls below the alarm point, the alarm circuit automatically
resets itself.
The RI-215A’s analog output, available via the yellow and blue wires of the cable
assembly (refer to Figure 3 on page 12), is proportional to the concentration of CO2 for all
gas ranges (e.g., 0 - 2,000 ppm, 0 - 5,000 ppm, and 0 - 9990 ppm), as shown in Figure 4.
Maximum
Output
Should the concentration of CO
scale (i.e., it exceeds the upper gas range of your unit), the highest output will be 22 mA,
or 11 V, depending on the output, current, or voltage, selected at the factory. The LCD of
the RI-215A will show a series of brackets (upside down “Us”) between 20 mA
(10 V) and 22 mA (11V). The standard output setting for the RI-215A is 4- 20 mA.
22 mA (11V)
20 mA (10V)
Output
signal
4 mA (0V)
0 ppmGas
Figure 5: Model RI-215A’s CO
present in the monitoring environment be above full
2
Concentration
Level to Output Signal
2
2000 ppm
5000 ppm
9990 ppm
Alarm Point and Ventilation Control
The RI-215A’s alarm relay has a built-in hysteresis of 50 ppm (refer to Figure 6 on
page 18). After the alarm point has been reached (e.g., 1000 ppm), the alarm relay will
activate. As the CO2concentration begins to fall, the alarm relay will deactivate 50 ppm
below the activation point (e.g., 950 ppm). The 50 ppm hysteresis keeps the alarm contacts
from “chattering” should the CO
So if the alarm relay contacts are wired to turn on a ventilation system when the
concentration of CO2reaches the alarm point (e.g., 1000 ppm), the ventilation system will
remain on until the concentration of CO
falls to 50 ppm below the alarm point (e.g., 950).
2
Alarm Adjust Mode
Use the Up (6) switch to enter Alarm Adjust Mode.
Verifying the Alarm Point
Three ranges of gas detection are available for the RI-215A: 0 - 2,000 ppm, 0 - 5,000 ppm,
and 0 - 9990 ppm. Detection ranges are set at the factory. When you ordered the RI-215A,
you should have specified one of these gas detection ranges for your unit.
1000 ppm
950 ppm
CO
2
Closed
Relay contact
Figure 6: Model RI-215A’s Built-in Hysteresis of 50 ppm
Open
Open
Hysteresis
50 ppm
The alarm point of the RI-215A is also set at the factory (e.g., 1000 ppm). To display the
current alarm point from the normal LCD, do the following:
1.Press and hold the Up (
(alarm verification screen). In this example, 10:00 indicates an alarm point of 1000
ppm. Note that the colon (:) will be flashing.
2.Release the Up (6)switch to return to the normal LCD. Do not hold the (6) switch for
more than three seconds or the unit will display the Alarm Adjust screen.
Further note that the RI-215A’s alarm relay has a built-in hysteresis (lag) of 50 ppm. After
the alarm point has been reached (e.g., 1000 ppm), the alarm relay will activate. As the
concentration begins to fall, the alarm relay will deactivate 50 ppm below the
CO
2
activation point (e.g., 950 ppm). The 50 ppm hysteresis keeps the alarm contacts from
“chattering” should the CO
18 • Alarm Adjust ModeRI-215A Operator’s Manual
6) switch. The alarm point appears (e.g., 10:00) on the LCD
This chapter describes maintenance and troubleshooting procedures for the RI-215A. It
includes instructions for the replacement of the RI-215A’s gas sensor and fuse and
calibration procedures. A spare parts list is located at the end of this chapter.
Preventive Maintenance
This section describes a preventive maintenance schedule.
Daily
Verify a display reading for CO2 in ppm. Investigate significant changes in the reading.
Typical background levels of CO
Quarterly
are 200 to 400 ppm, depending on location.
2
Calibrate the unit. See See “Calibration” on page 22.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable causes, and
recommended action for problems you may encounter with the RI-215A.
•The LCD does not show
a gas reading or is blank.
Table 3: Troubleshooting the RI-215A
SymptomsProbable CausesRecommended Action
•Power may have been
terminated to the unit.
•The unit may have
been incorrectly
connected to a power
source.
•The wrong power
source is being used to
run the unit.
•The fuse may have
blown.
•The unit may be in
Maintenance Mode.
1. Verify that the power source has been
turned on and is working.
2. If a circuit breaker has been used in the
installation of the RI-215A, make sure
that the circuit breaker has not been
activated.
3. Verify that the wiring to the power source
has been connected correctly.
4. Make sure you have used the correct
power source: 24 VAC or 24 VDC.
5. Open the unit and check the fuse.
Replace the fuse if required. The RI-215A
takes the following replacement fuse: 0.
5A, 125 VAC (5.2 mm x 20 mm).
6. The unit may be in Maintenance Mode.
Press the MODE switch for three
seconds, or until a gas reading appears.
•The LCD shows
abnormally high or low
CO2 readings but other
gas detection instruments show a normal
level of CO
•Unable to zero or span
reading during
calibration.
2.
Replacing Components
The section includes procedures to replace the sensor and fuse.
Replacing the Detector
The CO2 gas detector inside the RI-215A is a factory replacement part. Should you
determine that the RI-215A’s gas detector is no longer working or is working improperly,
you will need to ship the RI-215A to RKI Instruments, Inc. to have the detector replaced.
Replacing the Fuse
1.Turn off the power to the RI-215A.
•The RI-215A may
need to be
recalibrated.
•The calibration
cylinder is low,
outdated, or defective.
1. Recalibrate the unit.
1. Verify that the calibration cylinder
contains an adequate supply of sample.
2. If difficulties continue, contact RKI.
2.Remove the base plate of the RI-215A.
•Using a small screw driver, press down on the small tab on top of the
•Model RI-215A.
•While pushing down on the tab, gently move the base plate away from the front
housing of the RI-215A.
CAUTION: Be very careful not to pull on the cable assembly connected to the RI-215A ’s internal
printed circuit board when removing the base plate. Pulling on the cable assembly
may damage the printed circuit board.
•Completely separate the base plate from the front housing.
3.Find the fuse on the left side of the printed circuit board. Carefully remove the fuse
from the bracketing holding it on the printed circuit board.
4.Insert a new fuse in the brackets. The replacement fuse must be 0. 5A, 125 V or 250 V
(5 mm x 20 mm).
5.Reassemble the base plate and the front housing of the RI-215A in the reverse order as
described in step 2 above. Line up the bottom of the base plate with the front housing,
then carefully snap the base plate and front housing together.
This section describes how to calibrate the RI-215A. It includes procedures to prepare for
calibration, set the zero reading, set the response (span) reading, and return to normal
operation.
WARNING: The RI-215A is not an active gas monitoring device during the calibration
procedure.
Preparing for Calibration
Turn on the power to the RI-215A, and let the unit warm up for at least 30 minutes.
Connect the sample tubing from your calibration kit to the hose barb on the fixed flow
regulator. The fixed flow regulator should be a 0.5 LPM (liters per minute) regulator.
NOTE: Do not screw the regulator onto the calibration cylinder at this time.
Setting the Zero Reading
1.Locate the function switches on the bottom of the unit.
2.Press and hold the MODE switch (M) for three seconds to enter Maintenance Mode.
The display will show “1” to indicate Zero Adjustment Mode.
3.Press the SET switch (S) to enter Zero Adjustment Mode. The display will show a
reading with a flashing decimal point before the first and third digits, for example
“. 1.5“.
4.Remove the rubber cap covering the calibration gas inlet port at the bottom of the
unit.
5.Insert the calibration tubing into the gas inlet opening at the bottom of the RI-215A
and push the tubing over the calibration fitting.
6.Screw a 100% nitrogen calibration cylinder (a CO
7.Allow the gas to flow for one minute or until the gas reading stabilizes.
8.Press the SET switch (S) and the unit will adjust the reading to zero. Be careful not to
breath onto the RI-215A since the CO
9.Press the SET switch (S) and the display will again show a “1”.
10. Unscrew the regulator from the nitrogen cylinder.
11. To proceed with setting the span, press the Up (
indicates Span Adjustment Mode and continue with the next section, Setting the
Response Reading.
12. To return to normal operation, skip to Returning to Normal Operation below.
in your breath may affect the reading.
2
-free source) into the regulator.
2
6) switch to display “2” which
Setting the Response Reading
1.With “2” showing in the display, press the SET switch (S) to enter Span Adjustment
Mode. The display will show the current gas reading with a flashing decimal point
before the second and third digits, for example “.6.50“.
2.Press the SET switch (S) again and the reading will start to flash.
3.If not already in, insert the calibration tubing into the gas inlet opening at the bottom
of the RI-215A and push the tubing over the calibration fitting.
4.Screw the calibration cylinder into the regulator.
5.Allow the calibration gas to flow for one minute or until the gas reading stabilizes.
Parts List
6.If the gas reading does not match the gas concentration, use the Up (
Down (¢) switches to adjust the reading to the gas concentration.
7.Press the SET switch (S) to set the reading and exit Span Adjustment Mode. The
display will show a “2”.
8.Unscrew the regulator from the calibration cylinder.
9.Remove the calibration tubing from the calibration fitting on the RI-215A.
10. Allow time for the gas reading on the RI-215A display to return to a normal level.
6) and
Returning to Normal Operation
1.Press the MODE switch for thr ee seconds to exit the Maintenance Mode and return to
normal operation.
NOTE: If you do not allow the gas reading to return to normal, then unwanted alarms
may occur.
2.Store the components of the calibration kit in a safe and convenient place.
Table 4 lists replacement parts and accessories for the RI-215A carbon dioxide transmitter.
Table 4: Parts List
Part NumberDescription
06-1252RKSample tubing (1/4 x 1/8 in.; specify length when ordering)
43-4139RKReplacement fuse, 0. 5A, 250 V, 5 mm x 20 mm
81-0070RK-01Calibration cylinder (2,000 ppm carbon dioxide; 34 liter)
81-0071RK-01Calibration cylinder (5,000 ppm carbon dioxide; 34 liter)
81-0078RK-01100% nitrogen calibration cylinder (34 liter)
81-1003RKRegulator, 0.5 liter/minute; continuous flow
The RI-215A can be setup for one of thr ee detection ranges: 0 - 2,000 ppm; 0 - 5,000 ppm; 0
- 9990- ppm. When the RI-215A is ordered, the detection range must be specified. If it is
necessary to change the detection range in the field, this can be done in Maintenance
Mode.
WARNING: The RI-215A is not an active gas monitoring device during the full scale
adjustment procedure.
Perform the following to change the detection range.
1.With the RI-215A on and in Measuring Mode, locate the function switches on the
bottom of the unit.
2.Press and hold the MODE switch (M) for 3 seconds to enter Maintenance Mode. The
display will show “1” to indicate Zero Adjustment Mode.
3.Press the Up (
Mode.
4.Press the SET switch (S) to enter Full Scale Adjustment Mode. The display will show a
full scale value with a flashing decimal point after the first digit, for example
“2.000“.
5.Use the Up (
until the desired full scale value is displayed.
6.Press the SET switch (S) to accept the displayed full scale value. The display will again
show a “3”.
7.Press the MODE switch for 3 seconds to exit Maintenance Mode and return to
Measuring Mode.
8.Calibrate the RI-215A as described in "Calibration" on page 22.
WARNING: The RI-215 must be calibrated after the detection range is changed. The
6) switch until the display shows “3” to indicate Full Scale Adjustment
6) and Down (¢) switches to scroll through the full scale value choices
RI-215A readings will not be accurate unless a calibration is performed.