1 Outline of the Product ...................................................................................................................................... 3
2 Important No tice on Safety .............................................................................................................................. 4
2-3. Precauti ons .............................................................................................................................................. 5
3-1. Main unit a nd st an d ard accessories ........................................................................................................ 6
3-2. Names and fu nc t i o ns for each part ......................................................................................................... 8
4 How to Install ................................................................................................................................................. 11
4-1. Precauti ons for installati o n po i nts ......................................................................................................... 11
4-2. Precautions for system designing ......................................................................................................... 12
4-3. Install at io n of m ai n un i t .......................................................................................................................... 14
4-4. Precauti ons for wiring ............................................................................................................................ 17
5 How to Use .................................................................................................................................................... 21
5-1. Before using the monitor ....................................................................................................................... 21
5-2. Preparation for start-up ......................................................................................................................... 21
5-4. User mode ............................................................................................................................................. 23
5-5. Air calibration ......................................................................................................................................... 25
5-6. Setting v alue check ............................................................................................................................... 26
6 Alarm Activation an d Fu n c ti ons ..................................................................................................................... 29
6-1. Gas alarm activation .............................................................................................................................. 29
6-3. External out p ut operation ...................................................................................................................... 31
6-4. Suppression function ............................................................................................................................. 32
6-5. Air Follower Function ............................................................................................................................. 32
6-6. Auto Air Cal Function ............................................................................................................................. 33
7-3. Preparation for zero and span adjus tment ............................................................................................ 41
7-4. Zero adjustment ..................................................................................................................................... 42
7-7. Return to us er m ode .............................................................................................................................. 51
7-8. Componen t R e p lac ement ...................................................................................................................... 51
7-9. Replacing a Sensor Cable ..................................................................................................................... 53
7-10. Adding a Sensor Cable ........................................................................................................................ 57
8 Storage and D is posal .................................................................................................................................... 62
8-1. Procedures to store the monitor or leave it for a long time ................................................................... 62
8-2. Procedures to relocate the monitor or use it again ............................................................................... 62
8-3. Disposal of products .............................................................................................................................. 63
10-1. List of specifications ............................................................................................................................ 66
10-2. List of accessories ............................................................................................................................... 67
11 Spare Parts List ........................................................................................................................................... 68
12-1. Detect ion pr i nciple of galvan ic c e l l type .............................................................................................. 69
Important precautions ............................................................................................................................... 69
12-2. Definition of terms ................................................................................................................................ 70
• Before turning on the monitor, always check that the voltage is compliant with the specifications.
Operating on an u ns t ab l e power supply may cause malfunctions.
• When air calibration is performed, check the atmosphere for freshness before beginning it. If
interference gases exist, th e a ir calibration ca nn o t b e per f or med properly, thus leading t o da n gers
when an oxygen defi ciency occurs.
• Do not oper ate th is m onitor in a p lace where c om bustibl e/ex plos ive gas es or vap ors are pres ent.
Operating the monitor in such an environment will lead to extreme dangers.
• Issuance of an oxygen d eficiency alarm indicates tha t there are ex treme dangers . Take proper
actions base d o n your ju d g ment.
• Perform spa n adj us tment at fixed in ter v a ls .
• Make sure that the monitor is installed in a place where it is in contact with the atmosphere. If the
monitor is ins ta l l ed in a p l a c e wh ere it is in less con tact with the atmosphere, it m a y not d et ec t the
correct ox yge n c onc en t r at i on , thus leading to an oxygen deficiency accident.
• Do not run the power cable and remote sensor cable of the monitor in parallel with cables of
high-frequency or high-voltage and other device's power cables.
• If a cable of high-frequency or high-voltage and the power cable need to int e r sect wit h ea ch othe r,
it should be orthogonally connected.
• When wiring, be careful not to apply any stresses on the cables by pulling, tightening or twisting.
• Do not disassemble/modify the monitor. It may invalidate the warranty of the performance.
Changing the s et t in gs wi th ou t un de r s t an d in g th em may cause alarm m alf unctions. Pleas e use the
monitor prop er ly in accordance wit h t he operating man ua l .
2-3. Precaut io ns
• Do not use a de vice, such as a transcei ver, which tr ansmits a radio wav e near the mon itor or its
cables. It ma y affec t t he m easur em ent. If a tr ansc eive r or other radio wave transmitting device is
used, it must be used in a place away from the monit or wh ere it does not disturb the monitor.
• Restarting the monitor within five seconds after turning it off may cause errors.
• This is not a control unit. It is not allowed to use the ex ternal output of t he monitor to co nt r o l ot h er
units.
• This is a saf ety unit. N ever fail t o perform a regular maintenanc e to ens ure safet y. Contin uing to
use the monitor without performing maintenanc e will compromise the sensitivity of the sensor,
thus resulting in inaccurate gas detection.
• Do not pick the sensor or buzzer opening with a sharp-pointed it em. T he u nit m ay m alfunc tion or
get damaged, possibly resulting in incorrect measurements.
• Do not let the m onit or dr aw in wa ter. D o n ot ins tal l the mon itor i n a place wher e the moni tor m a y
get wet. Ignoring this may cause malfunction because the monitor is not water- and drip-proof.
• This is a precis i on de v ice . Do not apply strong shock or vibration to the monitor.
• When the c ase is open ed for wiri ng or other op eration , do not touch inner parts . W hen wiring, be
sure that excessive pressur e is n ot a pplied to the powe r ca b le and remote sensor cab l e.
After opening the shipping box, check the monitor and accessories.
If there is anything missing, contact RKI.
Main unit
For names and functions of individual parts of the monitor and LCD display, see "3-2. Names and
functions for each part" on page 8.
NOTE
• The OX-600’s standard configuration is with the sensor installed in the main unit. Specify when
ordering if you w a nt a n ex t en d er c ab l e wi th a r emote sensor.
• The sensor int egrated type does n ot i nclude the remote sensor.
• The cable lengths available for the remote sensor type are: 3, 5, 10 and 20 m.
• 3.2 m AC power cable (1 pc.) *Supplied only with AC specification
Main unit
Remote sensor
Power switch
Cable outlet
Cable outlet
Cable outlet
Cable outlet
Display
Sensor part
Mounting hole f or
3-M3
Use one of the two
for lower holes
To main unit
To main unit
To main unit
Standard accessories
machine screw (2 pcs.)
• Cross-recessed r ou n d he a d wo od
• Operating manual (1 pc.)
CAUTION
• The main unit and sensor unit (remote type) are precision devices. Be careful not to drop the
monitor when installing or uninstalling the main unit or sensor unit. Dropping the monitor may
compromise its original performance or cause malfunctions.
Optional accessories
• Mounting pl at e (1 pc.) RKI par t number: 21-1927
• Gas calibration cup (1 pc.) RKI part number: 81-1153
Displays the operating status. Blinks at a normal state.
(2)
First alarm display
Lights up in orange at a first alarm state.
(3)
Second alarm display
Lights up in red at a second alarm state.
(4)
Fault alarm display
Lights up in red at a fault alarm state.
Concentrat io n v al u e
Maintenance indicator
Displays the gas concentration.
Maintenance items and other s ar e dis played during se tu p.
(6)
Unit display
Displays the unit (%) according to the specification.
Gas name disp lay
Maintenance display
Displays the gas name (O2: Oxygen).
Maintenance items and other s ar e dis played during setup.
(8)
AC/DC power display
Lights up when the monitor is operating on AC or DC power.
Displays the battery level when the monitor is operating on dry
batteries.
(10)
This is not used for th e m o ni t or.
CAUTION
characteristic of rechargeable batteries.
Display
(5)
display
(7)
(9) Battery level icon
Installation of batteries (for dry battery type)
When the moni tor is us e d f or the first time, or wh en the battery level is lo w, install or re p lac e w ith the new
AA alkaline dry batteries according to the following procedure.
• Turn off the power of the monitor b ef ore replacing the ba t t er ies.
• Replace the b att eries in a safe plac e wh ere explosive gases are not prese nt.
• Never fail to use alkaline batteries. If a rechargeable (secondary battery) nickel-cadmium batter y
or nickel metal hydride battery is used, the specifications cannot be met, such as continuous
operating time.
• Replace bot h ba tt eries at the same time.
• Pay attention to the polarities of the batteries whe n i ns t a lli n g th em.
• Completely lock the battery cover after installing the batteries. If the battery cover is not
completel y locked, the dry batt er ies may fall out.
• Do not use rechargeable batteries that may interrupt a measurement due to the discharge
www.GlobalTestSupply.com
1
Check that the power of the
2
Remove the battery case from
the monitor.
Unlock the tab of the ba ttery case.
4
Put the battery case back in t he
New battery
main unit is turned off.
If the power is turned on, slide the
power switch down to turn it off.
3
Put new batteries (tw o AA
alkaline dry batteries) in the
battery case.
Confirm that the polarities of the
batteries are correct.
monitor.
Push the circled area an d
confirm that the tab is locked.
<Do not install the monitor in a place with vibrations or shocks.>
<Do not install the monitor in a place exposed to water, oil or
<Do not install the monitor in a place where the temperature
CAUTION
4
4 How to Install
4-1. Precaut io ns for installation points
When installing the monitor, never fail to observe the following precautions.
Ignoring the pr ecautions ma y dam ag e th e m on i to r, resultin g i n inac curate gas detec t io n .
• This is a pr ecision device. Because the monitor may not provide the specified performance in
some places (environments), check the environment in the installation site, and then take
appropriate actions if necessary. Because the monitor plays an important role for safety and
disaster prevention, it must be installed in appropriate points.
• Do not instal l this product in any of th e fo l l o wi ng loc ations.
• Place exposed to direct sunl i g ht or o utside
• Place exposed to water
• Place exposed to ventilation from an air conditioner, etc.
• Place exposed to soot, smoke or steam
• Place wher e the temper ature drops below -10ºC or ris es over 40ºC or where the tem peratur e
changes suddenly
• Place with h ig h h umidity like a bath ro om
• Place with bad ventilation such as behind a curtain or under an overhang.
The monitor consists of sensitive electronic parts. The monitor must be installed in
a stable place wit h ou t v i brations or shocks , etc . an d it c an no t drop.
chemicals, etc.>
When select in g i nstallation poi nts , a voi d a p lace where the m onitor is exposed to
liquids such as water, oil, or chemicals.
drops below -10ºC or rises over 40ºC.>
The operatin g temperature of the m oni t or is -10 to +40ºC. Install the monitor in a
stable place not exceeding the operating temperature range.
<Do not install the monitor in a place exposed to direct sunlight
or the monitor cannot adjust to sudden changes in the temperature.
<Keep the monitor (and its cables) away from noise source
high-frequency/high-vo lt a g e de v ices exist.
<Do not install the monitor in a place where maintenance of the monitor cannot
in a place where m ai nt en a nc e in v olv es d an g er s, for example, n ear a hi gh -voltage cable.
Power supply
voltage
Up to 10 milliseconds
system), etc. outside the monitor.
Do not use it with a power supply of large power load or high-frequency noise.
Use a line filter, etc. to avo i d th e no ise source if necessary.
CAUTION
or sudden changes in temperature.>
Avoid a place where the monitor is exposed to direct sunlight or radiant heat
(infrared rays emitted from a high-temperature object), and where the unit
temperature changes suddenly. Condensation may be formed inside the monitor,
devices.>
When select in g i nstallation poi nts , a voi d a p lace where
be performed or where handling the monitor involves dangers.>
Regular maintenance of the monitor must be performed.
Do not install the monitor in a place where the machinery must be stopped when mainten a nc e is
performed, w h er e parts of the machin ery must be removed to per f or m maintenance, or where the
monitor cannot be removed because racks or other things prevent access to it. Do not install the monitor
4-2. Precautions for system designing
Note the following precautions for system designing of the monitor.
• An unstable power supply and noise may cause malfunctions or false alarms.
Using a stable power supply
The external output and alarm contact of the monitor may be activated when the power is turned on, when
momentary blackout occurs, or when the system is being stabilized. In such cases, use a UPS
(uninterruptible power system), or take appropriate actions on the receiving side.
The monitor m ust b e pro v i d ed wit h t he f o llo w ing power supply.
100 - 120 VAC ±10% (50/60 Hz), 24 VD C ± 10 % or AA alkaline dry batter y (2 pcs.)
Allowed time of
momentary
blackout
Others
(To recover from the momentary blackout for 10 milliseconds or more, restart the
monitor.)
Example of actions
To ens ure continuous op er at i on a nd ac t i va ti o n, i ns t a l l a UPS (uninterruptible power
The alarm contact of the monitor is used to transmit signals to activate an external buzzer, alarm lamp or
rotating lamp. Do not use it for controlling purpose (e.g., controlling the shutdown valve). The contacts can
be configured as Form A or Form B but are factory set as Form A contacts and they are normally
de-energized.
CAUTION
• The b contact (break contact) under de-energized state may be open ed momen t ari ly by a phy sical
The specifications for the external output gas alarm contact of the monito r are based on th e re sistanc e load
conditions. If inductive load is used at the alarm contact, the following errors will occur easily because
counter electromotive force is generated at the contact.
• Deposition, defective ins u la t ion or defective contact at the relay contact
• Damage of any electric parts due to high-voltage generated inside the monitor
• Abnormal o perations by an out-of-control CPU
If load is to be activated, appropriate measures must be taken to stabilize the operation of the monitor and
protect the alarm contact referring to the following information.
• Relay it with an ex ternal relay at a lower vo lt a g e of 120 VAC or below (contact amplification). At the same
• In addition, the surge absorbing part SK2 must be attached to the loaded side of the external rela y if
• It may be recommended that the surge absorbing part should be attached to the contact for certain load
shock, such as external force.
When the b contact is se lected for the alarm contac t, take appropr iate actions to prepare for a
momentar y activat ion s uch as ad ding a s igna l dela y oper ation (appr oxim atel y one sec ond) to the
receiving side of the b contact.
time, the surge absorbing part SK1 suitable for the specifications must be attached to the external relay.
necessary.
conditions. It must be attached to an appropriate position by checking how the load is activated.
(Lower voltage relay)
CAUTION
• In princip le, do not activa te an inductive load at the alarm contact of the monitor. In part icular,
never use the i nd uctive load to activate a fluorescent lamp or motor.
• If inductive load is activated, relay it with an external relay (contact amplification). However,
because the coil of an external relay also involves inductive load, select a relay at a lower voltage
(120 VAC or below), and then protect the contact of the monitor with an appropriate surge
absorbing par t , suc h as a CR c ircuit.
www.GlobalTestSupply.com
Mounting hole (for M4
screw)
*Supporting the switch box
CAUTION
4-3. Install ation of main unit
Install the main unit on the wall 50 to 180 cm up from the floor.
If wall screws are available, remove the screw at the lower part of the main unit to open the cover and
install the unit using the mounting holes on the back of the unit.
• Install the sensor in a place not directly exposed to ventilation from an air conditioner.
Sudden changes in the temperature may cause the readings to disappear.
NOTE
• To install the main unit to the wall with screws, use the mounting holes on the back of the unit
according to the following procedure.
(1) Loosen the screw at the lower part of the main unit and open the surface cover of the case.
(2) Fix the main unit with two screws (M4) through the mounting holes.
(3) Put the surface cover back on the case and tighten the screw at the lower part of the main unit.
If the monitor operating on dry batteries needs to be removed with the power on, use the mounting plate
(option). When the m ounting plate is used, install the plate before installing the main unit.
<Installat io n of Remote Sensor>
Tighten the screws through one upper mounting hole and one of the two lower mounting holes.
NOTE
• To install the r emote sensor to the wall using screws, use the mounting holes on the back of the
sensor accor d i ng to th e following procedure.
(1) Open the surface cover of the remote sensor.
(2) Attach the sensor with two screws (M4 ) th ro ug h the mo un tin g hol e s (one up pe r moun ti ng hole and
<Maintenance Space>
For the dry battery type, be sure to leave space below the unit for battery replacement. If the mounting
plate (optional) is used, be sure to leave space above the unit so that it can be installe d by sliding.
If the monitor operates on AC or DC power of if an inductive load is use d a t th e a l ar m contact, wirin g wor k
is required.
The following cables are recomm end ed for wir i n g th e monitor with the p o wer supply, signal cab l e an d
contact.
<Recommended Cables>
2
2
• Be careful not to damage the internal electronic circuit when wiring. In addition, be careful not to
apply stresses on the monitor when (overweight) cables are installed.
• The power and signal cables must be wired separately from the motor power cables.
• When stranded wires are used, prevent wires from contacting each other.
<Cable Connection Conditions>
Connectable cable, bare wire length and connection tools are as follows:
• Cable: 0.2 - 1.5 mm
• Bare w ir e le n gt h: 10 - 11 mm
• Connecting tools: Dedicated screwdrivers manufactured by WAGO and equivalent (edge width 3.0 to 4.5
mm x 0.5 mm)
The specified bare wire length must be observed when the wire insulation is peeled off.
• Improper clamping of the wire due to a shorter bare wire length may cause defective electric
conduction or heating.
• Catching the wire insulation due to a shorter bare wire length may cause defective electric
conduction or heating.
• Exposing the wire due to a longer bare wire length may cause defective insulation or a short
circuit.
• Be careful not to break up the wire. If the wire is broken up when inserted to the terminal, this may
cause defect i ve ins u la t io n or he at i n g.
CVVS: 0.2 - 1.5 mm
2
2
<Compatible Bar Terminal>
For a bar terminal, the following items are available. See the spare parts list for RKI part numbers.
• Bar terminal (ferrule): Model 216 Series (manufactured by WAGO)
• Crimping tool: Model V arioCrimp 4 (206-204) (manufactured by WAGO)
• A bar terminal of the specified model must be used. Using other bar terminals invalidates the
warranty of th e per f ormance.
www.GlobalTestSupply.com
1
Push the push button strai ght
equivalent to open the spri ng.
2
Insert a wire with a specified
3
Release the screwdri ver.
The wire is conn ec t e d.
CAUTION
How to connect to terminal plate
When cables (wires) are connected to the terminal plate inside the main unit, use the dedicated
screwdriver or a compatible flathead screwdriver.
When connecting a stranded wire, be sure to press the push button and open the spring while connecting
the wire.
downward using the
compatible screwdriver or
bare wire length (10 mm) until
the end of it reaches the
deepest point.
• Never fail to use the correct tool.
• Do not insert more than one wire into one wiring hole. If the total size (mm
within the maximum wire connection range of the terminal plate, it may cause reduced spring
clamping for ce, defective i n sul at i o n d ue to clogged wire sheath, d ef ec t i ve c o nt ac t or coming off of
wires.
Compatibl e sc r ew d riv e r man ufactured by WAGO
Screwdriver (M) straight type
210-120J
Screwdriver (M) straight type (short shaft & grip)
210-350/01
210-657
Screwdriver (M) straight type (insulated shaft type)
210-720
NOTE
<Compatibl e S cr ewdriver>
• When opening the spring, use the compatible screwdriver manufactured by WAGO or equivalent (a
screwdriver with an edge width of 3.0 to 4.5 mm x 0.5 mm which can fully open the spring: See the
table below). In doing this wo rk, be ca reful no t to apply excessive force. Ignor ing thi s may damage the
housing/push buttons or cause dropping off of the push buttons.
Follow the operating precautions.
Ignoring the pr ecautions ma y dam ag e th e m on i to r, resultin g i n inac curate gas detec t io n .
CAUTION
• After you r ece i v e t he monitor, star t us i n g t he monitor with in the specified operation sta rt l imit of its
5-2. Preparation for start-up
Check the following points before starting oxygen concentration monitoring.
- Before turn in g o n th e power, check that the monitor is inst alled properly and that extern a l wiring is done
properly.
- Check that th e p o w er s up ply voltage is com pl ia nt wi t h th e s pecifications.
- Because the ex t ernal contact may be activated during the adjustm en t, take measures to prevent an
activated contact from having influences on external circuits.
state. The LCD backlight for AC and
24 VDC versio ns wi l l be on under
non-alarm conditions. The LCD
backlight f or t he bat tery powered
version will b e off under non-alarm
conditions in order to conserve
• Do not turn off the monitor during the wa rm up. W hen tur ning it on aga in, a bnor m al op erat ion ma y
occur.
• After initia l warmup, perfor m air calibration.
• Air calibration must be performed within one hour of power-on.
NOTE
• External out pu t d ur i n g ini t ia l c l ear is as follows:
• 4 - 20 mA: About 17.4 mA fixed
• 0 - 1 V: About 0. 84 V fixed
• To turn off the power, slide the power switc h down.
Press the SET button.
* If air calibration fails, "FAI L"
6
After checking the setting
state.
CAUTION
5-5. Air ca l ibration
An air calibration sets the fresh air reading to 20.9%. By default, the OX-600 automatically does an air
calibration once a month. See 6-6. Auto Air Cal Function on page 33 for more discussion.
button once.
2
Press the SET button to d is pl ay the
current concentration.
The concentration display
blinks.
5
adjust the concentratio n val ue
to "20.9%".
Air calibration is performed.
After air calibration i s
completed, "PASS" is
displayed and the monitor
returns to the display in st ep 2.
• When turning on the power, check the atmosphere for freshness and never fail t o perform air
calibration.
www.GlobalTestSupply.com
1
In the user mode, press the
2
Press the SET button.
<First Alarm Setpoint
3
Press the ˅ or ˄ but t on to
check the setting value.
<Second Alarm Setpoint
<Alarm Delay Time Display
<Zero Suppression Value
5-6. Setting value check
Typical setting values can be checked by the following procedure.
In the user mode, setting values cannot be changed. To change setting values, you must enter the
maintenance mode as described on P.36.
˄ butt on twice.
Press the SET button to display
the first alar m setpoint.
Slide the power switch down to power off.
After turning off the power switch of the monitor, turn off the power supply of the monitor.
CAUTION
• When the monitor is turned off, an alarm may be triggered on the upper (central) system.
Before turning off the monitor, the inhibit (point skip) on the upper (central) system must be
activated.
Decide whether the power can be turned off by checking the operation of the devices connected to
the external o ut pu t or ex t ernal contact out p ut t er minal of the monit or .
A gas alarm is activated when detected gas concentration reaches the preset alarm setpoint.
NOTE
• The alarm setpoint (fi r st alar m an d secon d ala r m) is fact o ry-set. The setti ng v alu es can be c ha nge d in
the maintenance mode (P.36).
• Although t he alarm delay time ( standard: 2 seconds) is set to preven t a false acti vation, it can be
changed in the maintenance mode (P.36).
Display operation
The LCD backlight for AC and 24 VDC vers ions will be on under non-alarm conditions. The LCD backlight
for the battery powered version will be off under no n-alarm conditions in order to conserve ba ttery.
<Gas Alarm Display (In case of oxygen deficiency state)>
The alarm disp l ay is a two-step alarm consisting of the first alarm display (ALM1) and the second alarm
display (ALM2).
It lights up i n o range w hen the concentration drops below the first alarm state (ALM1: 19.5%) or reaches its
alarm setpoint and in red when concentration drops below the second alarm state (ALM2: 23.5%) or
reaches its alarm setpoint.
<Alarm Display at Over Scale>
If a gas alarm is activated when the concentration display is over scale (oxygen concentration exceeds
The contact is ac t i v at ed wh en the gas concentration drops below the alarm setpoint value. The contact
activation is reset automatically when the gas concentration exceeds the alarm setpoint value (except for
the dry battery type).
• If the MODE or SET button is presse d to en ter another mode w h i le a n al ar m is triggered, the
Response to gas alarm (In case of responding to leaked gas)
When a gas alarm is triggered, take actions in acc ordance with your management rul es of gas alarm.
Normally, take the following actions.
• Check the reading of the monitor.
• If the gas alarm display continues to be displayed, close the main valve of the gas, and then
check that the oxygen concentration reading is back to normal.
NOTE
• If an oxygen deficiency is momentary, the reading may already have returned to normal when
checking it . In a ddi tio n, wh en th e al arm is tr igger ed by n oise or other inc ide nta l c ond itions ot her than
a gas, the readi ng m a y ha v e a lre ady returned to nor m al .
6-2. Fault alarm activation
A fault alarm is triggered when the monitor detects abnormalities. When a fault alarm is triggered, "FAULT"
in the LCD display lights up in red.
An error message is displayed in the LCD. Determine the cause and take appropriate action.
After the monitor is successfully returned from the fault, it enters the 25 second warm up per i od .
If the monitor has problems and is repeatedly malfunctioning, contact RKI immediately.
NOTE
• For informati o n on m alf u nc t ions (error messages), see "Troubleshooting" (P.64).
6-3. External output operation
<4 - 20 mA Output Table>
Status
Normal
Scale over
4.0 - 20.0mA
20.1mA -
Remarks
17.4mA
Air calibration
Alarm setpoint check
Date/time ch ec k an d
setting
Alarm test
*External out pu t 4 - 20 mA can be adjusted in the maintenance mode "2-48 MA20".
The monitor is i nf l ue nced by environmental changes (temperature, humidit y and o th er characteristics ),
which affects the reading.
Therefore, the reading might fluctuate because of the atmosphere, ev e n i n a nor mal environment .
The suppression function is used to hide (suppress) the fluctuation of the reading under the setting value,
indicating a l e v el that can obscure i nf l ue nc es by environment al c ha ng es in t h e atmosphere.
This function can be set in the maintenance mode (P.36).
fluctuation near the atm ospher e lev el
included.
The atmosphere level can be
displayed within the standard setting
(±0.5%).
6-5. Air Follower Function
OX-600s manufactured after mid-2019 have a DIP switch that turns the air follower function on (default) or
off.
If the air follower function is turned on (default) the OX-600 automatica lly adjusts the rea di n g to 20.9%
every hour if the indicated value is within ±0.5% of 20.9%.
The DIP switch is locat ed on the sen sor board. To adjust the switch, either open the main unit or the sensor
unit (if using a remote-mounted sensor). The air follower switch is the leftmost of the two switches. When
the switch is up , th e a ir fo l lo w er is o n. When the switch is d o wn, the air follower is off .
OX-600s manufac tured af ter mid-2019 have a DIP switch that turns the auto air cal function on (default) or
off.
If the auto air cal func t ion is t urn ed o n ( def aul t), the OX-60 0 autom atica lly perf orm s an air ca libra tion onc e a
month. If the OX-600 is not in a fresh air environment, the air calibration will be inaccurate.
The DIP switch is loc ated on the s ens or boar d. To adjust the switch, either ope n the m ain u nit or t he se nsor
unit (if us ing a rem ote-mount ed sensor) . The auto air cal switch is the rig htmost of the two s witches. W hen
the switch is up, the auto air cal function is on. When the switch is down, the auto air cal function is off.
Inspect the a larm circuit by usin g t h e
alarm test function.
Span
Adjustment
Perform spa n adj us tment using a
calibration gas.
Gas alarm
check
Check the gas al ar m using a
calibration gas.
7
7 Maintenance
The monitor is an important instrument for the purpose of safety.
To m aint a in the per f or mance of the monit or an d improve the reliability of safety, perform regular
maintenance.
7-1. Maintenance intervals and items
This is a safet y un it . Ne v er fail to perform dai ly and regular main te n ance before use.
• Dai ly maintenance: P erf orm maintenance before commencing each work .
• Regular maintenance: Perform maintenance once or more for every six months to maintain the
performanc e as a safety unit.
displa y ch ec k
WARNING
• If an abnorm ality is found in the mon it or, contact RKI im mediately.
value is "20. 9% ".
When the reading is incorrect,
perform the air calibration after
ensuring that no other gases exist
NOTE
• The built-in sensor of the monitor has a validity period and must be replaced regularly.
Checks the power supply voltage.
(Verifies that relevant points can be identified on the system.)
Concentration
display check
Verifies that the concentration display value is "20.9%" at the atmosphere level.
Performs the air calibration if the reading is incorrect.
Inspects the alarm circuit by using the alarm test function.
Checks the alarm activation such as a buzzer.
Span adjustm e nt
Performs span adjustment usin g a c a li br ation gas.
Checks the gas alarm by using the calibration gas.
Checks the alarm activation such as a buzzer.
Checks dust or damage on the surface, cover or internal parts of the unit, and
(Visual diagnosis) Replaces parts which are cracked or damaged.
Unit operation
check
Replacement of
parts
About maintenance services
We provide services on regular maintenance including span adjustment, other adjustments and
maintenance.
Our qualified service engineers have expertise and knowledge on the dedicated tools used for services,
along with oth er products.
To maintain the safety operation of the monitor, please use our maintenance service.
Typical maintenance services are listed as follows. Please contact RKI for more information.
<Typical Maintenance Services>
Power supply
check
Alarm test
Gas alarm check
Cleaning and
repair of the unit
consumable
Checks that the LCD display lights up.
Checks the alarm display (Checks the activation each for ALM1 and ALM2.)
Checks the alarm. (Checks triggering of alarm when the alarm setpoint is reached.)
Checks the del ay time. (Check s time t o d e lay until the alarm is tr ig g er e d.)
Checks the alarm display (Checks the activation each for ALM1 and ALM2.)
cleans or repairs such parts as needed.
Operates the buttons to check the operation of functions and parameters, etc.
Perform the span adjustment.
Refer to "Span adjustment" (P.43) for details.
Not used.
Specify the en v ir onmental settin gs.
2-40: Not used.
2-41: Not used.
2-42: Alarm Value Setting (AL-P)
2-43: Alarm Delay Time Setting (AL-D)
2-44: Alarm Pattern Setting (AL-T)
2-45: Zero Suppression Type Setting (SP-T)
2-46: Zero Suppression Value Setting (SAPP)
2-47: Not used.
2-48: External Output Adjustment (MA20)
2-49: Sensor Operation Start Setting (SSET)
2-4A: Date/Tim e S ett ing (DATE)
2-4b: Air Pressure Correction Setting ON/OFF (P-A)
After making the necessary adjustment s, hol d down the
(2-7 UMODE)>
The monitor returns to the user mode.
Press the SET button twice to return to the user mode.
Refer to "Return to user mode" (P.51) for details.
MODE button.
The monitor returns to the measuring state.
WARNING
• After air calibr ation, hold down the MODE button to return to the measuring state.
The monitor automatically returns to the measuring state in 10 hours.
Press the SET button
to perform a gas test.
Similar to the detection
condition, the reading
changes and an alarm
is displayed after the
gas is introduced, but
the contact is n ot
Press the SET button
to perform an alarm
test.
Use the ˅ button to
decrease the display
value to the alarm
setpoint to trigger an
alarm.
The externa l out p u t
signal and contact
activation are enabled.
To disable them,
specify the ac t i vat io n
setting for each in 2-4E
Press the SET button
to perform a fault
alarm test.
Use the ˅ or ˄ button
to enable the function
to trigger an al arm.
For details ab ou t op eration,
For details ab ou t op eration,
(P.43).
7-3. Preparation for zero and span adjustment
Perform zero and span adjustment of sensors using a calibration gas more than once every six months.
The zero and span adjustment requires dedicated equipment and a calibration gas. Please contact RKI.
<Required Equipment/Material>
Zero
adjustment
Span
adjustment
1. Screw the fixed flow regul ator into the calibration cylinder.
2. Use the calibration tubing to connect the regulator to the calibration cup’s inlet.
Perform the preparation (P.41) before performing a span adjustment. If a fresh air environment cannot be
verified, a cylinder of zero air will have to be used.
Navigate to the span
screen.
to display the current
concentration.
Turn the regulator’s
knob counterclockwise
to open the regula tor.
Allow zero air to flow
concentration SET
button again. While the
concentration display
is blinking, use the ˅
or ˄ butto n to adjust
the reading to "20. 9%" .
when adjustment is
successful, and "FAIL"
is displayed when it
fails.
sensor. Ignoring this may bend and damage the sensor pins.
• After the sensor is installed, always perform a calibration (zero adjustment and span adjustment).
• Do not dis assemble t he sensor bec ause it c ontains elec trolyte. If c ontact occ urs, rinse the area
immediate l y wit h a la r g e qu an t ity of water.
• The sensor mus t be stor ed under norm al t emper ature /hum idit y in a cle an p lace awa y from direc t
sunlight.
www.GlobalTestSupply.com
1
Remove the sensor cover of the
2
Remove the old sensor.
3
Install the new sensor.
4
Put the sensor cover back on.
2
Remove the sensor cover of the old
OX-600.
<Replacing the Sensor (Remote Sensor Type)>
sensor unit.
Push the tabs on the both sides with
your fingers to release the sensor
cover latch.
Remove the sensor cover to check
the sensor.
Be sure the sensor’s pins are lined
up with the boar d ’s sockets.
7-9. Repl acing a Sensor Cable
NOTE: The wire colors in the extender cable may vary depending on the type of cable used. The
Type A cable has red/green/white/black wires and the Type B cable has
white/yellow/white/blue wires. The pictures in this section show the Type A cable.
Turn off or disconnect power to the
1
cable’s remote sensor housing.
Push the tabs on bot h sides of the sens or
cover with your fingers to release the sensor
cover latch.
Connect the cable’s wires to the
sensor terminal strip. See the table
Note that the Type B cable
the cable. Tighten the zip tie to
Terminal 1
Terminal 2
Terminal 3
Terminal 4
T ype A
Red
Green
White
Black
Type B
White wire of
white/yello w pair
Yellow wire of
white/yello w pair
White wire of
white/blue pair
Blue wire of
white/blue pair
the hole in the housing to the right of
the sensor terminal str ip and install a
zip tie in the case and
cable as show n in this picture but do
not tighten it yet.
9
below for the wire colors. Type A wire
colors are shown in the pictures
below.
has two pairs of wires extending from
NOTE: The wire colors in the extender cable may vary depending on the type of cable used. The
Type A cable has red/green/white/black wires and the Type B cable has
white/yellow/white/blue wires. The pictures in this section show the Type A cab le.
Connect the cable’s wires t o the
sensor terminal strip. See the
table below for the wire colors.
. Note that
Terminal 1
Terminal 2
Terminal 3
Terminal 4
T ype A
Red
Green
White
Black
Type B
White wire of
white/yello w pair
Yellow wire of
white/yello w pair
White wire of
white/blue pair
Blue wire of
white/blue pair
Secure the top half of
to the bottom half.
Type A wire colors are shown in
the pictures below
the Type B cable has two pairs of
wires extending from the cable.
Tighten the zip tie to secure th e
cable in place.
8-1. Procedures to store the monito r or leave it for a
long time
The monitor must be stored under the following environmental conditions.
• In a dark plac e wi t h normal temperature and humidity and away from direct sunlight
• In a place wher e g as es, solvents, vapors, etc. are not present
Store the monitor in a shipping carton, in whic h th e product was deli v ere d, if possible.
Store the monitor away from dust, etc. if the shipping carton is not available.
• If the monito r operating on dr y batteries is not used f or a long time, stor e it after removin g the
batteries. B att ery leaks may resul t i n fire, injury, etc.
8-2. Procedures to relocate the monitor or use it again
When the moni tor is rel ocated, refer to "H o w to Install" for reloca ti o n s it es a nd wir i ng w ork.
The unpowere d t ime must be minimized when the monitor is relocated.
When the mon i tor is used again after long-term storage, perform a calibration.
• Contact RKI for information on readjustment including calibration.
When the monitor is dis p osed of, it must be tre ated properl y as an industrial waste i n ac cordance with th e
local regulations, etc.
If liquid is leaked from the sensor, do not touch the liquid. The sensor must be put into a plastic bag to
prevent the liquid from leaking to the outside. If any liquid leaked from the sensor is found in the monitor
unit, turn off the power and contact RK I immediately.
• Do not disassemble the galvanic cell type sensor because it contains electrolyte.
Electrolyte may cause severe skin burns if it contacts skin, and ma y cause b lindness if it c ontac ts
eyes. If electrolyte gets on clothing, th e material will di s col or or det eriorate.
If contact occ ur s, r i nse the area imm edia te ly with a large quan ti ty of water.
• Dispose of batteries in accord a nc e wit h a procedure spec if i e d by the local author ity.
Check wheth er t h e m ai n u n it is pr o per ly attached to
the wall-mounted unit.
Cable abnormalities
connected/short circuit)
The batteries are dead.
type
Turn off and restart the monitor .
appropriat e m easures to elim inat e the noise.
The sensor is n ot
connected.
Check that the sensor is connected and the
fastened.
Errors in communication
with the unit
Make a setting of date/time. If a symptom like this is
contact RKI.
The rated voltage is not
supplied to the monitor.
Check the power supply, and supply the rated
voltage.
Abnormalities of ROM,
the monitor
Drifting of sensor output
Perform the air calibration.
Disturbanc es by interference gases, such as
please contact RKI.
Perform the air cali bration. In particular, the
galvanic cell type is affected by the air pressur e.
9
9 Troubleshooting
This Troubleshooting section does not explain the causes of all the malfunctions which could occ ur on the
monitor. This simply helps to find the causes of malfunctions which frequently occur.
If the monitor shows a symptom which is not explained in this manual or still has malfunctions even though
remedial actions are taken, please contact RKI.
Turn on the power switch.
The power cannot
be turned on.
Abnormal
operations
Sensor
abnormalities
E-1
Clock
abnormalities
E-9
Abnormalities/momentary
blackout of power supply
system
(open circuit/not
*Only for the dry battery
Disturbances by sudden
surge noise, etc.
connected or improperly
Abnormalities of the clock
inside the monitor
Take measures such as checking or adding the
UPS, power supply line filter and insulation
Check the wiring of the monitor and related devices
around it.
Replace the t wo dry batteries.
If a symptom like this is observed frequently, take
connectors of the se nsor unit board ar e sec urely
Replace the sensor unit board with a new one.
observed repeatedly, the built-in clock is seemingly
malfunctioning. Thus, it must be replaced. Please
solvents, cannot be eliminated completely. For
information on actions, such as removal filter,
www.GlobalTestSupply.com
Symptom/Display
Causes
Actions
Disturbanc es by interference gases, such as
please contact RKI.
Turn off and restart the monitor .
appropriat e m easures to elim inat e the noise.
When the environment (temperature, etc.) changes
them.
Deteriorated sensor
sensitivity
Improper calibration gas
concentration
Deteriorated sensor
sensitivity
An oxygen
deficiency alarm is
triggered even
though there are no
abnormaliti e s at
the detection point.
Span adjustment
impossible
Presence of interference
gas
Disturbanc e by noise
Sudden change in the
environment
solvents, cannot be eliminated completely. For
information on actions, such as removal filter,
If a symptom like this is observed frequently, take
suddenly, the monitor cannot adjust to it and is
affected by it.
In some cases, the monitor triggers an indication
alarm.
Because the monitor cannot be used under sudden
and frequent environmental changes, the user
should take some preventive actions to eliminate
By immersing p reci ou s meta l an d le ad i n el e ct rolyte and connecting t he m wi th a lea d w i re, a b at te ry can be
made (galvanic cell).
When oxygen passes through the barrier, deoxidizing reaction occurs at the precious metal electrode while
oxidizing reaction occurs at the lead electrode.
The electric current generate d by this reaction goes through load r esistance (ther mister), where it is
converted in t o v oltage so that it can b e re ad. When the oxyge n c onc en t r at io n is decreased, deoxidizing
reaction at the pr ecious metal elec tr o d e is also decreased , th us l o wering the voltage at bo t h en ds of th e
thermister.
Important precautions
1. The readings of the monitor fluctuate slightly in response to changes in the air pressure. In particular,
be careful of alarm activation when a low air pressure is brought in by typhoon, etc.
In addition, it may be fluctuated by environmental (temperature, humidity, etc.) changes in the
installation site.
2. The alarm must be set within a range where the performance of the monitor can be ensured. The
standard alarm setpoint values are 18 vol% for the first and 19 vol% for the second.
3. This is a safety unit. This is not a control unit.
The alarm contact output of the monitor must be used for an external alarm lamp/buzzer, while the
analog signa l o ut pu t m us t be us e d for an indicator or ex t ernal recorder. If these outputs are used to
control other units, we shall not be responsible for any malfunctions.
4. For maintenance of the monitor, it must go through a regular maintenance, including replacement and
adjustment of the regular replacement parts as specified in the operating manual . I n ad d it ion, because
this is a safety unit, i t is recommended that a regular maintena nce a nd s pa n adjustment are perf ormed
every six months.
Gas concentration indicated in the unit of one-hundredth of the vo lume
Calibration
Find relationship of the readings, display values or setpoints with the actual values
by using the calibration gas.
Maintenance
When mainte na nce is perform ed on th e m on i to r, the alarm cont act is disconnect e d,
monitor.
Initial clear
The reading is uns t ab l e for seconds after t he power is turned on. To prevent
indicate the i ni t ia l cl ea r st at us is s en t o ut t o th e ex t er n al ou tp ut .
Zero
suppression
A function to cut off the influe nc es of environmental changes, interference gas es,
etc.
Alarm delay time
A function which temporarily suspends activation to prevent a false alarm caused by
noise from its outside.
12-2. Definition of terms
mode
and a signal to indicate the maintenance mode status is sent out to the external
output signal. As a result, maintenance can be performed on a single unit of the
malfunctions for that period, the alarm contact is deactivated. In addition, a signal to