Manual Supplement
OX-500
This supplement includes calibration instructions and a spare parts list for the OX-500.
Calibrating the OX-5 00 in Ma intenance Mode
Prepari ng for Ca libration
NOTE: The following procedure assumes the use of a calibration kit which includes a
100% nitrogen calibration gas cylinder, a zero air cylinder, a 2.0 LPM fixed flow
regulator with an on/off knob, a calibration cup, and a short piece of sample
tubing to connect the regulator to the calibration cup.
NOTE: The following procedure assumes that the OX-500 is not in a fresh air
environment. If a fresh air envir onment can be verified, applying zer o air to the
detector is not necessary when setting the span adjus tment.
1. Connect the calibration kit sample tubing to the calibration cup hose barb.
2. Connect the fixed flow regulator to the calibration cup using the sample tubing. Be
sure that you are using a 2.0 LPM fixed flow regula tor.
WARNING: Failure to use the recommended gas flow rate will result in an inaccurate
reading.
Adjusting the Zero Reading
1. While in Measuring Mode, press and hold the MAINTE Switch and the MODE Switch
simultaneously for 3 seconds to enter Maintenance Mode. Release the switches when
the following screen appears.
NOTE: The MAINTE Switch is a small recessed push button switch loca ted to the left of
the MODE Switch. It is labelled on page 5 of the attached OX-500 manual. A tool
such as a small tipped screwdriver or a pencil may be used to press the MAINTE
Switch.
2. Press and release the MODE Switch to proceed to item 2-2, Zero Adjustment Mode.
3. Press and release the RESET switch to select item 2-2. The display will indicate the
current gas reading.
4. Screw the 100% nitrogen cylinder into the fixed flow regulator.
5. Turn the regulator knob counterclockwise to open the regulator.
6. Hold the calibration cup against the case over the diffusion grill that covers the sensor
face in the lower left corner of the front.
7. Allow 100% nitrogen to flow for two minutes. The reading should decrease and
stabilize after two minutes.
P. 1 of 3
Released: 1/25/13
www.rkiinstruments.com

Manual Supplement
OX-500
8. Press and release the MODE switch to perform the zero adjustment. The reading will
momentarily flash and the OX-500 will perform a zero adjustment.
9. If the zero adjustment is successful, the buzzer will sound a double tone and the
display reading will be set to 0.0 %.
If the zero adjustment is not successful, the LCD will indicate the error message Err.
10. Remove the calibration cup from the front face of the OX-500.
11. Turn the regulator knob clockwise to close the regulator.
12. Unscrew the 100% nitrogen cylinder from the fixed flow regulator.
13. Press and release the RESET Switch to exit Zero Adjustment Mode and return to the
screen that indicates item 2-2.
Adjusting the Span Setting
1. Press and release the MODE Switch to proceed to item 2-3, Span Adjustment Mode.
2. Press and release the RESET Switch to select item 2-3. The display will ind icate the
current gas reading.
NOTE: If you can confirm a fresh air environment, you do not need to apply zero air to
set the span.
3. Screw the zero air calibration gas cylinder into the fixed flow regulator.
4. Turn the regulator knob counterclockwise to open the regulator.
5. Hold the calibration cup against the case over the diffusion grill that covers the sensor
face in the lower left corner of the front.
6. Allow zero air calibration gas to flow for two minutes.
7. Adjust the gas reading so that it matches the ca libration gas concentration using the
MAINTE Switch. The MAINTE Switch will only increase the gas reading. When the
reading reaches the maximum, it will “wrap around” and begin increasing from zero.
8. Press and release the MODE Switch to accept the span setting. The gas reading will
momentarily flash and the span setting will be saved.
9. If the span adjustment is successful, the buzzer will sound a double tone and the gas
reading will be displayed.
If the span adjustment is not successful, the display will indicate Err.
10. Remove the calibration cup from the front face of the OX-500.
11. Turn the regulator knob clockwise to close the regulator.
12. Unscrew the zero air calibration gas cylinder from the fixed flow regulator.
13. Press and release the RESET Switch to exit Span Adjustment Mode and return to the
screen that indicates item 2-3.
14. Press and hold the MAINTE Switch for 3 seconds to exit Maintenance Mode and
return to Measuring Mode.
15. Store the components of the calibration kit in a safe place.