RKI Instruments M2A Operating Manual

M2A Transmitter
Operator’s Manual
Part Number: 71-0305RK
Revision: H
Released: 12/3/18
RKI Instruments, Inc.
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating instrument. Improper use of the gas monitor could result in bodily harm or death.
Periodic calibration and maintenance of the gas monitor is essential for proper operation and correct readings. Please calibrate and maintain this instrument regularly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage.

Product Warranty

RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS W ARRAN T Y IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries
b) Pump diaphragms and valves e) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
M2A Transmitter Operator’s Manual

Table of Contents

Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
About the M2A Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
External Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Internal Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Chapter 3: Installation & Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Mounting the M2A Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Wiring the M2A Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Chapter 4: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Information Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4 - 20 mA Signal Output Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Chapter 5: Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Viewing & Changing M2A Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Chapter 6: Gas Type Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Selecting the Gas Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Chapter 7: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Calibration Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Calibration, Combustible Gas, CO2, and Toxic Versions . . . . . . . . . . . . . . . . . . . . . . . . 51
Calibration, Oxygen Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Replacing Components of the M2A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Chapter 8: RS-485 Modbus Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Wiring the M2A in a Modbus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Using the M2A in a 4-wire Modbus System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Modbus Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Supported Modbus Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Appendix A: Control Button Quick Reference Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Appendix B: PLC and DCS Device Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Appendix C: Function Code 16 Registers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
(Appendix C available from RKI Instruments, Inc. Not included in manual as normally provided with M2A detector head.)
M2A Transmitter Operator’s Manual

Chapter 1: Introduction

Overview

This chapter briefly describes the M2A Transmitter. This chapter also describes the M2A Transmitter Operator’s Manual (this document). Table 1 at the end of this chapter lists the
specifications for the M2A.

About the M2A Transmitter

The M2A transmitter is a fixed mount, continuous-monitoring detector head. All user adjustable parameters may be accessed using push button switches. In addition, calibration may be performed non-intrusively by use of a magnetic wand accessory which activates magnetic switches through a window at the front face of the detector head.
The M2A displays the current gas reading on an OLED display which is visible through the window in the cover (front face) and provides a 4 - 20 mA signal which indicates the target gas reading for use by a gas monitoring controller, recording device, or programmable controller. The M2A also provides an RS-485 Modbus output. Three sets of relay contacts, two controlled by the gas alarms and one by the fail alarm, rated at 115 VAC 5 amps are available for controlling devices such as lights or horns or for controlling higher rated relays.
Four operating modes allow you to display and change setup and calibration settings and change the gas type. They are Calibration Mode, Configuration Mode, Gas Type Mode, and Modbus Mode.

About this Manual

The M2A Transmitter Operator’s Manual uses the following conventions for notes, cautions, and
warnings.
NOTE: Describes additional or critical information.
CAUTION: Describes potential damage to equipment.
WARNING: Describes potential danger that can result in injury or death.
6 M2A Transmitter Operator’s Manual

Specifications

Table 1 lists specifications for the M2A.
Table 1: M2A Specifications
Target Gas/Detection Range Combustible Gas/Carbon Dioxide (CO
Catalytic LEL: 0 - 100% LEL, 1% LEL increments (CH4 calibration unless otherwise specified. H
specific detector available)
2
Catalytic CH4 ppm: 0 - 9000 ppm, 20 ppm increments
IR (infrared) CH
LEL: 0 - 100% LEL, 1% LEL increments
4
IR HC LEL: 0 - 100% LEL, 1% LEL increments (propane calibration)
IR CH
Volume: 0 - 5.00% volume, 0.01% increments
4
IR CH4 Volume: 0 - 100 % volume, 1% increments
IR CO2 ppm: 0 - 5,000 ppm, 20 ppm increments
IR CO
Volume: 0 - 5.00 % volume, 0.01% increments
2
IR CO2 Volume: 0 - 50.0 % volume, 0.1% increments
IR CO
Volume: 0 - 100 % volume, 1% increments
2
Oxygen
Oxygen (O2): 0 - 25.0 % volume, 0.1% increments
CO/H
S
2
Carbon Monoxide (CO): 0 - 300 ppm, 1 ppm increments
Carbon Monoxide (CO): 0 - 200 ppm, 1 ppm increments
Carbon Monoxide (CO): 0 - 100 ppm, 1 ppm increments
Hydrogen Sulfide (H
S): 0 - 100 ppm, 1 ppm increments
2
To xi c
Ammonia (NH3): 0 - 75.0 ppm, 0.1 ppm increments
Arsine (AsH
Chlorine (Cl
): 0 - 1.50 ppm, 0.01 ppm increments
3
): 0 - 3.00 ppm, 0.01 ppm increments
2
Chlorine (Cl2): 0 - 10.0 ppm, 0.1 ppm increments
Chlorine Dioxide (ClO2): 0 - 1.00 ppm, 0.01 ppm increments
Hydrogen Cyanide (HCN): 0 - 15.0 ppm, 0.1 ppm increments
Phosphine (PH
Sulphur Dioxide (SO
): 0 - 1.00 ppm, 0.01 ppm increments
3
): 0 - 6.00 ppm, 0.01 ppm increments
2
)
2
Alarm Settings (Alarm 1/Alarm 2)
Ammonia: 12 ppm/25 ppm Arsine: 0.20 ppm/0.50ppm Carbon Dioxide 0 - 5,000 ppm: 2,500 ppm/5,000 ppm Carbon Dioxide 0 - 5.00%: 0.50 %/3.00 % Carbon Dioxide 0 - 50.0%: 50.0%/50.0% Carbon Dioxide 0 - 100%: 100%/100% Carbon Monoxide (all ranges): 25 ppm/50 ppm Chlorine (0 - 3.00 ppm): 0.50 ppm/1.50 ppm Chlorine (0 - 10.0 ppm): 1.0 ppm/3.0 ppm Chlorine Dioxide (0 - 1.00 ppm): 0.10 ppm/0.20 ppm Combustible Gas (%LEL): 10 %LEL/50 %LEL Combustible Gas (0-100%Volume): 100 %/100 % Hydrogen Cyanide: 5.0 ppm/10.0 ppm Hydrogen Sulfide: 10 ppm/50 ppm Methane (ppm): 5000 ppm/8000 ppm Methane (0-5.00%Volume): 0.5%/2.5 % Oxygen: 19.5%(decreasing)/23.5%(increasing) Phosphine: 0.10 ppm/0.30 ppm Sulphur Dioxide: 1.00 ppm/2.00 ppm
M2A Transmitter Operator’s Manual 7
Table 1: M2A Specifications
Construction (housing) Explosion-proof Junction Box, NEMA 4X
Area Classification Explosion-proof for Class I, Groups B, C, and D (Combustible, CO
and CSA Type oxygen)
Sampling Method Diffusion
Input Power 10 - 30 VDC
Controls Three push button switches
Three magnetic switches for non-intrusive calibration
Weight 4.5 lbs.
Signal Output 4 to 20 mA with impedance maximums listed below
-12 VDC input: 500 ohms impedance max
-24 VDC input: 1000 ohms impedance max
RS-485 Modbus
Explosion Proof Non-Explosion Proof
Catalytic LEL/ppm -40°C to 75°C N/A
Oxygen -20°C to 45°C -20°C to 50°C
S -40°C to 40°C -40°C to 50°C
H
2
Operating Temperature
CO -5°C to 40°C -20°C to 50°C
Toxic (ESM-01) N/A -10°C to 40°C
Toxic (CT-7 Series) N/A -20°C to 40°C
, CSA Type CO and H2S,
2
Infrared Combustible -20°C to 50°C -40°C to 50°C
Infrared CO
2
Accuracy Catalytic Combustible Gas (LEL range)
-20°C to 50°C -40°C to 50°C
: ± 5% of reading or ± 2% LEL (whichever is greater) Catalytic CH4 (ppm range):
± 5% of reading or ± 50 ppm (whichever is greater)** IR Combustible Gas/IR CO
:
2
± 5% of reading or ± 2% of full scale (whichever is greater)
:
Oxygen ± 0.5% O
2
Hydrogen Sulfide: ± 5% of reading or ± 2 ppm H
S (whichever is greater)
2
Carbon Monoxide: ± 5% of reading or ± 5 ppm CO (whichever is greater) Toxic Sensors (ESM-01 Type and CT-7 Series Type)
:
± 10% of reading or ± 5% of full scale (whichever is greater)
*
These are RKI factory settings. See “Viewing & Changing M2A Parameters” on page 41 to change the alarm settings.
**
Independent of zero drift.
WARNING: When using the M2A, you must follow the instructions and warnings in this
manual to assur e proper a nd safe op eration of the M2A and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the M2A as described in this manual.
8 M2A Transmitter Operator’s Manual

Chapter 2: Description

Magne tic
Wand
Junction Box Cover
3/4 NPT Conduit Port, Plugged
3/4 NPT Conduit Port
Window
Mounting Slot (2x)
Detector
(Catalytic LEL Detector Shown)

Overview

This chapter describes external and internal components of the M2A Transmitter.

External Description

This section describes the junction box and all external components of the M2A transmitter.
M2A Transmitter Operator’s Manual 9
Figure 1: M2A External Components

Junction Box

The M2A’s cast aluminum junction box is dust and weather resistant. The junction box also protects the M2A and all connections made to it. Use the three 3/4 NPT conduit ports to mount the detector to the junction box (factory installed in the bottom port) and connect wiring from an external device (left port). The top port is shipped with a factory installed and sealed conduit plug to avoid leaks into the junction box. If necessary, the conduit plug can be removed and the top port can be used for wiring.
WARNING: If the top conduit port is used for wiring, be sure to seal the threads to ensure
water does not enter the junction box. See “Wiring the M2A Transmitter” on page 27 for complete wiring instructions.
Use the junction box’s two mounting holes to mount the M2A to a vertical surface at the
Detector PartNumbers
61-0140RK
61-0140RKA
NC-6205-01
Detector PartNumbers
61-0140RK-05
61-0140RKA-05
NC-6205-05
monitoring site. The window in the cover on the front of the junction box allows you to view the OLED display and use the magnetic wand to actuate the magnetic control switches so you can perform non-intrusive calibration. Removing the cover allows you to access the interior of the junction box.

Magnetic Wand

The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate the magnetic control switches on the control PCB while the junction box cover is still installed so that non-intrusive calibration can be performed.

Gas Detector

The gas detector senses the target gas and is mounted in a 3/4 NPT conduit port on the right bottom side of the M2A. A variety of detectors may be used with the M2A. See the sections below for descriptions of each type of detector.
Catalytic Detectors
The catalytic detectors have a 1/2 NPT thread and require a 3/4 NPT x 1/2 NPT reducer to install in the detector port.
The table below outlines the M2A part numbers and replacement detector part numbers for the catalytic detectors.
Table 2: Catalytic Detectors
Target Gas, Range M2A Part Number Detector Used
Combustible Gas, 0 - 100% LEL 65-2640RK 61-0140RK (Catalytic UL version)
65-2640RK-05 61-0140RK-05 (Catalytic CSA version)
Hydrogen, 0 - 100% LEL 65-2641RK NC-6205-01 (Catalytic UL version, hydrogen specific)
65-2641RK-05 NC-6205-05 (Catalytic CSA version, hydrogen specific)
Methane, 0 - 9000 ppm 65-2647RK 61-0140RKA (Catalytic UL version)
65-2647RK-05 61-0140RKA-05 (Catalytic CSA version)
10 M2A Transmitter Operator’s Manual
Figure 2: Catalytic Sensors
Infrared (IR) Detectors
The IR combustible detectors are generally used instead of the catalytic combustible detectors in applications where there may be catalyst poisons such as silicone present or where oxygen is not present in the monitoring environment.
The table below outlines the M2A part numbers and replacement detector part numbers for the infrared (IR) detectors.
Table 3: Infrared (IR) Detectors
Target Gas, Range M2A Part Number Detector Used
CH4, 0 - 100% LEL 65-2649RK-CH4 61-0190RK-CH4 (UL)
65-2664RK-CH4 61-0197RK-CH4
, 0 - 5.00% volume 65-2649RK-CH4-5 61-0190RK-CH4 (UL)
CH
4
CH
, 0 - 100% volume 65-2658RK-CH4 61-0192RK-CH4 (UL)
4
65-2665RK-CH4 61-0199RK-CH4
, 0 - 5,000 ppm 65-2660RK-02 61-0191RK-02 (UL)
CO
2
65-2661RK-02 61-0198RK-02
, 0 - 5% volume 65-2660RK-03 61-0191RK-03 (UL)
CO
2
65-2661RK-03 61-0198RK-03
, 0 - 50% volume 65-2660RK-05 61-0191RK-05 (UL)
CO
2
65-2661RK-05 61-0198RK-05
, 0 - 100% volume 65-2660RK-10 61-0191RK-10 (UL)
CO
2
65-2661RK-10 61-0198RK-10
HC, 0 - 100% LEL (cal. to propane) 65-2649RK-HC 61-0190RK-HC (UL)
65-2664RK-HC 61-0197RK-HC
M2A Transmitter Operator’s Manual 11
Non-XP, Repla ceable-SensorType
Cap Gasket
Detector Housing Body
Hydrophobic Membrane
Detector Part Numbers
61-0197RK-CH4
61-0197RK-HC
61-0198RK-02 61-0198RK-03 61-0198RK-05 61-0198RK-10
61-0199RK-CH4
Detector Housing Cap
Plug-In IR LEL Sensor
Detector Part Numbers
61-0190RK-CH4
61-0190RK-HC
61-0191RK-02 61-0191RK-03 61-0191RK-05 61-0191RK-10
61-0192RK-CH4
61-0192-HC
XP,
Non-Replaceable-
Sensor Type
Figure 3: Infrared (IR) Sensors
Oxygen Detectors
There are three types of oxygen detectors available: a non-explosion proof partial pressure type, a non-explosion proof capillary type with a replaceable sensor, and an explosion proof (CSA) capillary type with a replaceable sensor. The capillary type’s output is less susceptible to variations in atmospheric pressure than the partial pressure type. The CSA type oxygen detector (65-2514RK) and the non-explosion proof capillary type detector (65-2494RK) has a replaceable plug-in sensor inside the detector housing. The entire oxygen detector must be removed if replacement is required for the 65-2510RK.
The table below outlines the M2A part numbers and replacement detector part numbers for the oxygen detectors.
Table 4: Oxygen Detectors
Target Gas, Range M2A Part Number Detector Used
Oxygen, 0 - 25% 65-2643RK-05 65-2514RK (capillary, CSA)
65-2644RK 65-2510RK (partial pressure)
65-2666RK 65-2494RK (capillary)
12 M2A Transmitter Operator’s Manual
Figure 4: Oxygen Detectors
Cap Gasket
Detector HousingBody
Detector Housing Cap
Hydrophobic Membrane
Detector Housing Cap
Plug-In O xygen Sensor
Cap Gasket
Non-XP,
Replaceable-SensorC apillary Type
Non-XP, Partial
Pressure Type
Oxygen Sen sor
Detector Housing Body
Flame Arrestor Guard
Detector
PartNumber
65-2494RK
Detector Part
Number
65-2510RK
Detector Part Number
65-2514RK
XP,
Replaceable-Sensor Capillary Type
CO and H
S Detectors
2
Both types of CO and H2S detectors (explosion-proof and non-explosion-proof) have replaceable plug-in sensors inside the detector housing.
The CO detectors include a charcoal filter disk which is held onto the sensor with a rubber retaining boot. The charcoal filter disk scrubs out interfering gases to which the CO sensor responds such as H
S and certain hydrocarbons.
2
The 65-2428 and 65-2428-01 explosion-proof H
S detectors both include a rubber boot and a
2
spacer which help ensure that the sensor remains plugged into the detector housing body. The only difference between the 65-2428 and the 65-2428-01 is that the sensor used in the 65-2428-01 is resistant to degradation caused by frequent exposure to low levels of H
high levels of H
The 65-2495RK non-explosion-proof H
S.
2
S detector includes an adhesive-backed gasket inside the
2
detector housing cap which helps ensure that the sensor remains plugged into the detector housing body.
The table below outlines the M2A part numbers and replacement detector part numbers for the CO and H
S detectors.
2
Tab le 5 : CO and H2S Detectors
Target Gas, Range M2A Part Number Detector Used
S or occasional exposure to
2
CO, 0 - 300 ppm 65-2646RK-05 65-2438 (CSA)
65-2663RK 65-2496RK
CO, 0 - 200 ppm 65-2646RK-05-02 65-2438-01
CO, 0 - 100 ppm 65-2646RK-05-01 65-2438-01
S, 0 - 100 ppm 65-2645RK-05 65-2428 (CSA)
H
2
65-2645-05-07 65-2428-01 (CSA)
65-2662RK 65-2495RK
M2A Transmitter Operator’s Manual 13
Figure 5: H2S Detectors
Cap Gasket
Flame Arrestor Guard
H2S Plug-in Sensor
Spacer
RubberBoot
Detector Housing Cap
Non-XP Type
Detector Housing Cap
Hydrophobic Memb rane
Cap Gasket
H2S Plug-in Sensor
Detector Housing B ody
Detector PartNumber
65-2495RK
Detector Part Numbers
65-2428
65-2428-01
Detector Housing Body
XP Type
Detector Housing C ap
CharcoalFilter w/Rubber Boot
CO Plug-inS ensor
Cap Gasket
Detector Housing Body
Cap Gasket
Hydrophobic Membrane
CO Plug-inS ensor
Flame Arrestor Guard
Detector Housing C ap
Non-XP Type
Detector Part Number
65-2496RK
Detector Housing B ody
RubberBoot w / CharcoalFilter
XP Type
Detector PartNumber
65-2438
65-2438-01
Figure 6: CO Detectors
ESM-01 Detectors
The construction of the various ESM-01 detectors is similar. Each of the ESM-01 detectors has a different replaceable plug-in sensor that protrudes through the detector housing cap. Although ESM-01 detectors look the same except for the plug-in detector, the detector housing which includes the detector socket and conditioning circuitry is specific to a sensor type. For example, a detector housing from a Cl
detector cannot be used with an NH3 detector. One of the two detector
2
wires is color coded for each specific detector type. The ESM-01 detectors include a removable splashguard that screws onto the detector housing cap. When it is necessary to apply gas to an ESM-01 detector, remove the splashguard to allow access to the sensor that protrudes through the detector housing cap. You can then push the appropriate ESM-01 calibration cup onto the sensor.
The table below outlines the M2A part numbers and replacement detector part numbers for the
Target Gas, Range M2A Part Number Detector Used
AsH
, 0 - 1.5 ppm 65-2648RK-ASH3 65-2300RK-ASH3
3
ESM-01 detectors.
Table 6: ESM-01 Detectors
14 M2A Transmitter Operator’s Manual
Target Gas, Range M2A Part Number Detector Used
Splash Guard (removable)
Cap Gasket
Det ec t or Ho usi ng Bo dy (di fferent for each d etector type )
Sensor (differe nt for each detector type)
Detector Housing Cap
Cl
, 0 - 3 ppm 65-2648RK-CL2 65-2300RK-CL2
2
Cl
, 0 - 10 ppm 65-2648RK-CL-10 65-2300RK-CL-10
2
HCN, 0 - 15 ppm 65-2648RK-HCN 65-2300RK-HCN
NH3, 0 - 75 ppm 65-2648RK-NH3 65-2300RK-NH3
PH
, 0 - 1 ppm 65-2648RK-PH3 65-2300RK-PH3
3
SO
, 0 - 6 ppm 65-2648RK-SO2 65-2300RK-SO2
2
Figure 7: ESM-01 Type Detector
CT-7 Series Detectors
The CT-7 Series detectors for Cl
and ClO2 have a replaceable plug-in sensor inside the detector
2
housing. An adhesive backed gasket inside the detector housing cap helps ensure that the sensor remains plugged into the detector housing body.
The table below outlines the M2A part numbers and replacement detector part numbers for the CT­7 Series detectors.
Target Gas, Range M2A Part Number Detector Used
Cl
, 0 - 3 ppm 65-2670RK-CL2 65-2302-CL2
2
, 0 - 10 ppm 65-2670RK-CL210 65-2302-CL2-10
Cl
2
ClO
, 0 - 1 ppm 65-2670RK-CLO2 65-2302-CLO2
2
M2A Transmitter Operator’s Manual 15
Table 7: CT-7 Series Detectors
Figure 8: CT-7 Series Detectors
Hydrophobic Membrane
Plug-in Sensor
Detector Housing Cap
Cap Gasket
Detector Housing Body (different for each detector type)
16 M2A Transmitter Operator’s Manual

Internal Description

NO
Control PCB
M2A TRA N SMITTER
RKI INSTRUMENTS
Junction Box

Terminal PCB

3/4" NPT Conduit
Open i ng f or W ire Entry
3/4" NPT Conduit Opening, Plugged
This section describes the internal components of the M2A. The internal components of the M2A include the terminal PCB which provides for all the wiring connections to the M2A and the control PCB which displays the gas reading and has the control buttons.
Terminal PCB
The terminal PCB is encapsulated in epoxy for protection against moisture and physical damage. It is mounted into the rear of the junction box with three standoffs and rests on a thin layer of foam. A banana jack is screwed into each of the standoffs and is used for mounting the control PCB. The terminal PCB converts the electrical output from the detector to a signal which can be displayed by the OLED display, a 4 - 20 mA signal (that is proportional to the detection range), and an RS-485 Modbus output signal. The 4 - 20 mA signal may be used by a recording device, gas monitor controller, or programmable controller. The Modbus output may be used to connect the M2A to a Modbus network. The terminal PCB also controls three relays: one fail and two gas alarm.
Two columns of plug-in style terminal strips are used to make all wiring connections to the M2A. The column on the left consists of the power/signal, detector, and Modbus terminal strips. The column on the right consists of the relay terminal strips. A 20 position connector at the bottom of the terminal PCB is used to connect the terminal PCB to the control PCB with a ribbon cable. A 5 position connector on the left side of the terminal PCB is used by factory or field service personnel to program the M2A. On the combustible and CO
above the programming connector is used to set the detector current.
Figure 9: M2A Internal Components
versions of the M2A, a factory adjust pot just
2
M2A Transmitter Operator’s Manual 17
Figure 10: Terminal PCB Component Location
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Modbus Terminal Strip
Ground Jumper
Programming Connector
(Factory Use)
Power/Signal Terminal Strip
Termination Jumper
Ribbon Cable Connector
Detec tor Curren t Adjust Pot
(Comb/CO 2 T y p e O n l y, Factory Adj us t )
Detector Terminal Strip
(Comb/CO2 Version Shown)
Relay Terminal Strips
Toxic/Oxy
Detector
Terminal
Strip
Comb/CO2
Detector
Terminal
Strip
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALA RM 2
C NC NO
FAIL
- S + A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALA RM 2
C NC NO
FAIL
Power/Signal Terminal Strip
The power/signal terminal strip is a three position plug-in style terminal strip located at the top of the left terminal column. It is used to connect 24 VDC power to the M2A and to connect the 4 - 20 mA output signal to a device.
The signal output, the S terminal, does not have to be connected for the M2A to function. The S terminal is used if you want to connect the 4 - 20 mA output signal to another device such as a gas monitoring controller, chart recorder, or programmable controller (PLC).
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip and is the middle terminal strip in the left terminal column. For a combustible or CO
connect the detector to the M2A. For a toxic or oxygen M2A, only two of the terminals are used for connecting a detector; two terminals are dedicated for connecting a toxic detector and two are dedicated for connecting an oxygen detector (only one detector can be installed).
18 M2A Transmitter Operator’s Manual
Figure 11: Detector Terminal Strip
M2A, all four terminals are used to
2
NOTE: The detector is factory-wired to the M2A. See “Wiring the M2A Transmitter” on
ENTER
DOWN/NO
UP/YES
Push Button
Control
Switches
Magnetic Switches
ENTER
UP/YES
DOWN/NO
Alar m 1 LED
Alar m 2 LED
Fail LED
RX LED
TX LED
M2A TRANSMITTE R
RKI INSTRUMENTS
OLEDDisplay
page 27 for all wiring procedures related to M2A.
Modbus Terminal Str ip
The Modbus terminal strip is a three position plug-in style terminal strip and is the bottom terminal strip in the left terminal column. It allows connection of the M2A into a Modbus network.
Relay Terminal Strips
The right column of terminal strips consists of, from top to bottom, the fail, alarm 1, and alarm 2 relay terminal strips. They are three-position plug-in style terminal strips. The relay terminal strips are used to connect devices such as lights and horns that are controlled by the relay contacts. The relay contacts are rated at 115 VAC, 5 amps. The relay contacts may also be used to control higher rated relays.
Termination Jumper
A two pin header with a termination jumper installed is located below the Modbus terminal strip. Leave this jumper installed unless directed to remove it for a Modbus installation. See “Chapter 8: RS-485 Modbus Output” on page 61 for a description of using the M2A in a Modbus system.
Ground Jumper
A two pin header with a jumper installed is located along the left side of the terminal PCB. Leave this jumper installed unless directed to remove it for a Modbus installation. See “Chapter 8: RS-485 Modbus Output” on page 61 for a description of using the M2A in a Modbus system.

Control PCB

The OLED display and control switches are located on the control PCB. It is installed on top of the terminal PCB by lining up its three spacing standoffs with the banana jacks in the terminal PCB mounting standoffs and pushing it onto the banana jacks. The jacks retain the control PCB.
Figure 12: Control PCB Component Location
M2A Transmitter Operator’s Manual 19
OLED Display
The OLED display is located at the top of the control PCB. It indicates the current gas reading and displays messages and parameters in the M2A’s programs.
Control Buttons
The M2A includes three push button switches that allow you to enter the M2A’s operating modes, navigate through the modes, update settings, and save changes to the settings. The push button switches are located along the bottom edge of the control PCB (see Figure 12). The UP/YES button is on the left, the DOWN/NO button is in the middle, and the ENTER button is on the right.
Table 8: M2A Control Button Functions
Switch Function
UP (YES) Saves settings
Changes the displayed setting
Enters the Calibration Program
Enters Gas Type Mode (press with DOWN/NO button)
Enters Configuration Mode (press with ENTER button)
DOWN (NO) Cancels setting changes
Changes the displayed setting
Enters Gas Type Mode (press with UP/YES button)
Enters Modbus Mode (press with ENTER button)
Displays the Information Screen
ENTER Initiates operations
Enters Configuration Mode (press with UP/YES button)
Enters Modbus Mode (press with DOWN/NO button)
Functions as an alarm reset switch
Just above each push button switch is a magnetic switch with the same function as the push button switch below it. The magnetic switches are for use in non-intrusive calibration. They are actuated by bringing the magnetic wand close enough to them to actuate them. Although the magnetic switches have the same functions as the push button switches, it is not practical to use them for operations other than calibration because it is not possible to actuate two magnetic switches at once with only one magnetic wand. Since displaying the Information Screen only requires the use of one switch, the wand may be used to show the Information Screen (see “Information Screen” on page 36).
Status LEDs
The M2A includes five status LEDs that are located above the display (see Figure 12).
Fail LED
The fail LED turns on when the M2A is experiencing a fail condition. A fail condition can be caused by a detector failure or low detector signal.
Alarm 1 LED
The alarm 1 LED is on when the M2A is experiencing an alarm 1 condition.
Alarm 2 LED
The alarm 2 LED is on when the M2A is experiencing an alarm 2 condition.
RX & TX LED’s
These LED’s indicate data being received (RX) and transmitted (TX) when the M2A’s Modbus output is operating.
20 M2A Transmitter Operator’s Manual

Chapter 3: Installation & Startup

3/4 NPT Conduit Port, Plugged
7.7 max
2.3 max
5.23
5.00
2.41
.30 (2X)
.47 (2X)
3/4 NPT Conduit Port
1 1/2-20 For Calibration Cup
4.59
5.86
1.02

Overview

This chapter describes procedures to mount the M2A Transmitter in the monitoring environment and wire it to input power and devices.

Mounting the M2A Transmitter

1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the M2A is not likely to be bumped or disturbed. Make sure there is
sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site where the target gas is likely to be found first. For lighter gases, mount the
detector near the ceiling; for heavier gases, mount the detector near the floor.
2. At the monitoring site, use #10 screws through the junction box’s two mounting holes to secure
the junction box to a vertical surface. Figure 13 - Figure 20 show the outline and mounting dimensions for each version of the M2A. Mount the M2A with the detector facing down as shown in the following figures.
Figure 13: Outline & Mounting Dimensions, Catalytic LEL and ppm, UL Version
M2A Transmitter Operator’s Manual 21
.47 ( 2X)
.30 ( 2X)
3/4" NPT Conduit Port, Plugged
2.41
5.00
2.3 max
1.02
3/4" NPT Conduit Port
4.59
5.23
5.86
7.7 max
1.9 max
3/4 NPT Condui t Por t, Plugged
7.4 max
1.02
5.86
4.59
5.23
.30 (2 X)
.4 7 (2 X)
2.41
5.00
3/4 NPT Condui t Por t
Figure 14: Outline & Mounting Dimensions, Catalytic LEL and ppm, CSA Version
Figure 15: Outline & Mounting Dimensions, IR Combustible & CO
22 M2A Transmitter Operator’s Manual
2
3/4 NPT Conduit Port, Plugged
7.9 m a x
2.6 m a x
5.23
.47 (2X)
2.41
5.00
.30 (2 X)
3/4 NPT Cond u it P ort
1 1/2-20 F or Calibration Cup
1.02
5.86
4.59
3/4 NPT Conduit Port, Plugged
1.5 max
1 1/2-20 For Calibrat ion Cup
3/4 NPT Conduit Port
6.9 max
5.23
.47 (2X)
.30 (2X)
2.41
5.00
5.86
1.02
4.59
Figure 16: Outline & Mounting Dimensions, Oxygen, Partial Pressure Type
Figure 17: Outline & Mounting Dimensions, Replaceable Sensor, H
S, CO, and Oxygen,
2
Non Explosion Proof
M2A Transmitter Operator’s Manual 23
3.2 max
3/4 NPT Conduit Port, Plugged
8.4 max
4.59
1.02
5.86
.30 (2X)
5.00
2.41
.47 (2X)
5.23
3/4 NPT Conduit Port
Figure 18: Outline & Mounting Dimensions, H2S, CO, and Oxygen, CSA, Explosion-Proof
24 M2A Transmitter Operator’s Manual
3/4 NPT Conduit Por t, Plugged
4.82 max
10.4 max
5.86
1.02
4.59
5.00
5.23
.47 (2X)
.30 (2X)
2.41
3/4 NPT Conduit Port
Figure 19: Outline & Mounting Dimensions, ESM-01 Toxic
M2A Transmitter Operator’s Manual 25
3/4" NPT Conduit Opening Plugged
1.40 max
1.02
4.59
CT-7 Detector
5.00
2.41
5.23
.47 (2X)
.30 (2X)
5.86
6.80 max
3/4" NPT Conduit Opening
Figure 20: Outline & Mounting Dimensions, CT-7 Series
26 M2A Transmitter Operator’s Manual

Wiring the M2A Transmitter

Green
Black
White
Red
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Green
Black
Red
White
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
WARNING: Always verify that the power source is OFF before making any wiring
connections.
1. Remove the junction box cover.
2. Grasp the control PCB by its edges.
3. Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull too hard and damage the cable which connects the control and terminal PCBs.
4. Let the control PCB hang by the cable. The terminal strips are now visible on the terminal PCB. The control PCB may be left hanging while wiring is done. If desired, the control PCB may be disconnected from the cable and set aside while wiring.
5. The detector leads are factory wired. Verify that the detector leads are wired to the detector terminal strip as shown in the applicable figure below:
Catalytic LEL/Catalytic ppm/IR Combustible/IR CO2 Detectors
The wiring for these types of detectors is the same. Red wire to terminal labeled LEL R, white wire to terminal labeled LEL W, green wire to terminal labeled LEL G, black wire to terminal labeled LEL B.
Figure 21: Catalytic LEL and ppm Detector Wiring
Figure 22: IR Combustible/IR CO
M2A Transmitter Operator’s Manual 27
Detector Wiring
2
Oxygen Detectors: Partial Pressure, Capillary Type, and CSA Capillary Type
Green
White
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
Green
White
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
Green
White
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
The wiring for all the oxygen detectors is the same. White wire to terminal labeled OXY +, green wire to terminal labeled OXY -.
Figure 23: Partial Pressure Oxygen Detector Wiring
Figure 24: CSA Capillary Type Oxygen Detector Wiring
Figure 25: Capillary Type Replaceable Sensor Oxygen Detector Wiring
28 M2A Transmitter Operator’s Manual
Black
Red
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
Black
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Red
CSA Type CO or H
S Gas Detectors
2
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
Figure 26: H
Non Explosion Proof CO or H2S Gas Detectors
S/CO Detector Wiring, CSA
2
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
Figure 27: H
M2A Transmitter Operator’s Manual 29
S/CO Detector Wiring
2
ESM-01 Toxic Detectors
Black
See Chart
Gas TypeWire Color
Red NH3
Cl2Yellow
AsH3Brown
Green PH3 Blue SO2
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
The ESM-01 toxic detectors have one black wire and one color coded wire. Connect the color coded wire (see chart below) to the terminal labeled TOXIC + and the black wire to the terminal labeled TOXIC -.
Figure 28: ESM-01 Toxic Detector Wiring
30 M2A Transmitter Operator’s Manual
CT-7 Series Toxic Detectors
FAIL
Not
Used
C NC NO
S
PWR / SIG
A B C
ALAR M 2
C NC NO
RS 485
TOXIC OXY
ALARM 1
Gas Type Table
Black
C NC NO
CL2 yellow CL2-10 yellow CLO2 violet
0 - 10 ppm Cl2
0 - 3 ppm Cl2
See Chart
Gas Type + Wire
The CT-7 Series toxic detectors have one black wire and one color coded wire. Connect the color coded wire (see chart below) to the terminal labeled TOXIC + and the black wire to the terminal labeled TOXIC -.
Figure 29: CT-7 Series Toxic Detector Wiring
6. To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the wall between two terminal positions with needle nose pliers and pulling. Be careful not to exert too much force on the wall to avoid damage to the terminal strip. The detector terminal strip and the alarm 1 relay terminal strip may be removed by grasping them with your fingers if the adjacent terminal strips have been removed.
W ARNING: If the M2A is installed in a hazardous location, use appropriately rated conduit,
conduit fittings, and appropriate construction techni que that complies with the local electrical code. In addition, use appropriately rated conduit, conduit fittings, and appropriate construction technique to maintain the NEMA 4X environmental rating of the junction box and avoid water entering the junction box at the conduit ports.
7. Guide multi conductor shielded cable or cables or wires in conduit through the conduit port on the left side of the junction box.
WARNING: If the top conduit port is used for wiring, be sure to seal the threads to ensure
water does not enter the junction box. A conduit seal, which is normally required to maintain the junction box’s hazardous location rating, will also help to prevent water from entering through the top port.
The number of cables or wires needed will depend on whether the M2A is wired to a gas monitoring controller or just to power, whether any relays are used, and whether the Modbus output is used.
M2A Transmitter Operator’s Manual 31
NOTE: If the M2A is being wired to a PLC or DCS device, see “Appendix B: PLC and DCS
Ala rm Devi ce
Power
Fail Alarm Device
Alarm 1 Ala rm De vice
Typical Alarm Wiring Shown
See
Modbus
Wiring
RKI Controller
Terminals
(24 VDC) +
4 - 20 mA In (S)
(24 V DC) -
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
See
Detector
Wiring
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Alarm 2 Ala rm De vice
Device Wiring” on page 76, then continue with step 8.
Use the following recommendations to determine how to wire the M2A:
If Modbus connections will not be used and only the PWR/SIG connections will be used,
use a two or three conductor shielded cable or two or three wires in conduit for connections to the power/signal terminal strip depending on whether or not the signal (S) terminal is used. The S terminal has a 4 - 20 mA output, but if you do not need to monitor this signal and do not connect to the S terminal to access this signal, the M2A will still function completely.
Table 9: Wire Size for PWR/SIG Connections
Max Distance to Controller
w/18 Gauge Wire
Max Distance to Controller w/
16 Gauge Wire
Max Distance to Controller
2,500 ft. 5,000 ft. 8,000 ft.
If the PWR/SIG connections and one or more relays are used, route the connections to the
M2A in conduit. Use shielded cable in the conduit for the PWR/SIG connections and unshielded cable or individual wires for the relay connections. Make sure any wire or cable used for relay wiring is appropriately rated for the power that it will carry.
NOTE: If shielded cable is used for the PWR/SIG connections, leave the cable shield’s drain
wire insulated and disconnected at the M2A. You will connect the opposite end of the cable’s drain wire at the controller or device.
If the M2A will be wired into a Modbus network, see “Chapter 8: RS-485 Modbus
Output” on page 61.
See Figure 30 below for field wiring connections to the M2A.
w/14 Gauge Wire
Figure 30: Wiring the M2A to a Controller and Alarm Devices
32 M2A Transmitter Operator’s Manual

Start Up

METHANE
0 %LEL
8. Re-install the control PCB (and ribbon cable if necessary). Be sure the ribbon cable is routed down below the control PCB so it will not be damaged by the cover when it is screwed back on.
9. Secure the junction box cover to the junction box.
10. Make controller, device, and relay connections as appropriate. If shielded cable is used for the PWR/SIG connections, connect the cable shield’s drain wire to an available chassis ground at the gas monitoring controller, recording device, or programmable controller.

Introducing Incoming Power

1. Complete the installation procedures described earlier in this manual.
2. Verify that all wiring connections are correct and secure.
3. Turn on the incoming power.
4. If necessary, turn on the controller or other monitoring device that is connected to the M2A.
5. The OLED display will indicate the firmware version when the M2A is first powered up and will then count down a one minute warm-up period before normal operation begins. During normal operation, the display will indicate the target gas and current gas reading. Verify that the display is indicating the target gas and current gas reading after the warm-up period is complete and normal operation begins.
NOTE: Allow the M2A’s detector to warm up for 15 minutes before you continue with the next
section, “Adjusting the Fresh Air Reading.”

Adjusting the Fresh Air Reading

When the M2A is shipped from RKI Instruments, Inc., it is factory calibrated. If a full calibration is desired at startup, see “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51 or “Calibration, Oxygen Version” on page 54.
Verify that the M2A is in a fresh air environment (environment known to be free of the target gas and combustible or toxic gas vapors and of normal oxygen content, 20.9%).
CAUTION: If you suspect the presence of combustible gas, CO2, toxic gas, or that the oxygen
content is not normal in the monitoring environment, use the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate fresh air setting. See the “Ca libration, Combustibl e Gas, C O2, and Toxic Versions” on page 51 or “Calibration, Oxygen Version” on page 54 for instructions on how to use a zero air cylinder when performing a fresh air adjustment.
M2A Transmitter Operator’s Manual 33
Non-Intrusive Fresh Air Signal Adjustment
Calib?
YES/NO
Calib? YES/NO
If the M2A is installed in a classified area and non-intrusive zero adjustment is required, follow the instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic switch and perform the same operation as pressing a control switch button, touch the magnet in the end of the magnetic wand to the M2A’s glass cover directly over the magnetic switch you wish to actuate. Touching the glass and removing the wand is the same as pressing and releasing a button. Touching the glass and keeping the wand in place is the same as pressing and holding a button.
W ARNING: The M2A is not an active gas monitoring device during the fresh air adjustment
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA for a toxic, combustible gas, or CO
M2A, or at 17.4 mA for an oxygen M2A, and all relays
2
will remain in their non-alarm state while the M2A is in Calibration Mode. The output signal will not indicate current rea dings and the relays will not resume operating normally until the M2A is in normal operation again.
NOTE: While in the calibration program, if there is no switch activity for the calibration time-
out period the unit will return to normal operation. The factory set time-out is 15 minutes. If you want a different time-out period, see “Viewing & Changing M2A Parameters” on page 41 for instructions to change the calibration time-out.
Combustible Gas, CO
, and Toxic Versions, Adjusting the Fresh Air Reading
2
1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter Calibration Mode. Release the button when the following screen appears.
2. Press and release the UP/YES button to continue. The display will indicate the target gas and CAL Mode for a few seconds before showing FreshAir Adjust?.
3. Press and release the UP/YES button. ENTER will alternate with FreshAir on the top display line and the current gas reading will be on the bottom display line.
4. Press and release the ENTER button. The M2A will perform a zero operation and the display will indicate SPAN w/Cal Gas?.
5. Press and release the DOWN/NO button. The display will indicate Leaving CAL Mode and the M2A will return to normal operation.
Oxygen Version, Adjusting the Fresh Air Reading
1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter Calibration Mode. Release the button when the following screen appears.
2. Press and release the UP/YES button to continue. The display will indicate OXYGEN CAL Mode for a few seconds before showing FreshAir Adjust?.
3. Press and release the UP/YES button. ENTER will alternate with FreshAir on the top display
34 M2A Transmitter Operator’s Manual
line and the current oxygen reading will be on the bottom display line.
4. Press and release the ENTER button. The M2A will perform a span operation and then the display will indicate ZERO w/Cal Gas?.
5. Press and release the DOWN/NO button. The display will indicate Leaving CAL Mode and the M2A will return to normal operation.
M2A Transmitter Operator’s Manual 35

Chapter 4: Operation

METHANE
0 %LEL
24.2V v6.0L 2A

Overview

This chapter describes the M2A in normal operation. This chapter also describes the M2A in alarm 1, alarm 2, and fail conditions and suggests responses to these conditions.

Normal Operation

Normal operation is defined as follows:
The start-up procedure is complete.
The M2A is not indicating an alarm 1, alarm 2, or fail condition.
The M2A is not in Calibration, Configuration, or Gas Type modes.
During normal operation, the M2A simultaneously displays the current gas reading, unit of measure, and target gas. The example below illustrates a typical combustible gas M2A.

Information Screen

The Information Screen displays the M2A’s operating voltage and firmware and hardware information. To display the Information Screen, do the following:
1. Press and hold the DOWN/NO button for 3 seconds. The Information Screen will appear.
2. Continue holding the DOWN/NO button to keep the Information Screen on the display. The top line indicates the operating voltage that is connected to the M2A. The second line indicates the firmware that is running, version 6.0 in this example, and the M2A’s hardware version, version 2A in this example. This information may be useful if you need to contact the factory with questions about the M2A.
3. Release the DOWN/NO button when you are done viewing the Information Screen. The display will return to the normal operation screen within a couple of seconds.
36 M2A Transmitter Operator’s Manual

4 - 20 mA Signal Output Operation

The output at the S terminal of the power/signal terminal strip is a 4 - 20 mA signal that corresponds to the detection range of the M2A. During normal operation, this signal tracks the gas concentration on the OLED display.
There are several circumstances where the signal output will not track the display reading but will behave as follows:
When the M2A is in its warm-up period, the signal output will be fixed at 1.2 mA.
When the M2A’s gas type is changed, the M2A will enter Configuration Mode for you to verify the parameter settings. When you exit Configuration Mode, the display will indicate NEEDS CALIBRATION and will continue to indicate this until Calibration Mode is entered and a calibration is performed. In this situation, the signal output will be fixed at 3.5 mA for LEL and toxic versions and 17.4 mA for oxygen versions from the time Gas Type Mode is entered until the M2A is calibrated and returns to normal operation.
If you enter Calibration Mode, Configuration Mode, Gas Type Mode, or Modbus Mode, the signal output will be fixed at 3.5 mA for combustible gas, CO
mA for oxygen versions until the M2A returns to normal operation.
If the M2A’s input power decreases below 9.5 volts so that the M2A is in a low power alarm, the signal output is fixed below 2.4 mA until the low power alarm is cleared.
, and toxic versions and 17.4
2
If the M2 goes into a fail condition, after a 30 second delay, the signal output is fixed below 2.4 mA until the fail alarm is cleared. During the 30 second delay, the signal output follows the detector output. In the case of a downscale reading, the display and the signal output continue to track the reading down to -99% of full scale (1.15 mA).
M2A Transmitter Operator’s Manual 37

Alarm Indications

NOTE: The M2A includes alarm on and alarm off delay settings for alarm 1 and alarm 2. The
Condition Cause Visual Indication(s)
alarm indications described in this section operate according to the factory set alarm settings. See Table 11 on page 42 for all the factory settings.
Table 10: Visual and Audible Alarm Indications
1
Alarm 1
1
Alarm 2
Fail Disconnected or misconnected detector
Low Power DC power source less than 9.5 volts. F LED is on
1
*
If the M2A is in both an alar m 1 and an ala r m 2 condition, both alarm LEDs are on and the
display alternates between the gas reading and the ALMS 1&2 message.
Increasing (decreasing for O2) gas reading at
or above the alarm 1 setpoint
Increasing gas reading at or above the alarm 2 setpoint
wiring
Display reading at -10% of full scale or lower
Defective components
A1 LED is on
Gas reading alternates with ALARM-1 message
A2 LED is on
Gas reading alternates with ALARM-2 message
F LED is on
FAI L message replaces gas reading
NOTE: There is a 30 second delay on the fail condition.
LowPower message and actual voltage of incoming DC power
NOTE: You can select normally energized (N. EN) or normally de-energized (N. DE-EN)
alarm 1 and alarm 2 relay settings in Configuration Mode. The following sections describe the standard factory setting for these relays which is N. DE-EN.
The fail relay is factory set as N. EN and is not user-adjustable.

Alarm 1 Condition

Alarm 1 Condition Indications
When the gas reading reaches the alarm 1 setpoint, the M2A senses an alarm 1 condition. The M2A alerts you to an alarm 1 condition as follows:
The A1 LED turns on.
The gas reading alternates with the ALARM-1 message.
The alarm 1 relay energizes.
Responding to an Alarm 1 Condition
1. Follow your established procedure for a low level combustible or toxic gas condition or a decreasing oxygen concentration condition.
2. After the gas reading falls below (above for oxygen) the alarm 1 setpoint, press the ENTER button to reset the alarm 1 circuit. Resetting the alarm 1 circuit turns off the A1 LED, resets the
38 M2A Transmitter Operator’s Manual
OLED display, and de-energizes the alarm 1 relay.
NOTE: If the ENTER button is pressed while the M2A is in an alarm 1 condition, the A1 LED
will flash but all other indications will remain unchanged.
You cannot de-energize the alarm 1 relay until the gas reading falls below (above for oxygen) the alarm 1 setpoint.

Alarm 2 Condition

Alarm 2 Condition Indications
When the gas reading reaches the alarm 2 setpoint, the M2A senses an alarm 2 condition. The M2A alerts you to an alarm 2 condition as follows:
The A2 LED turns on.
The gas reading alternates with the ALARM-2 message.
The alarm 2 relay energizes.
NOTE: If the M2A is in both an alarm 1 and alarm 2 condition, both the A1 and A2 LEDs will
be on, the gas reading will alternate with the ALMS 1&2 message, and both alarm relays will energize.
Responding to an Alarm 2 Condition
1. Follow your established procedure for a high level combustible or toxic gas condition or an increasing oxygen concentration condition.
2. After the gas reading falls below the alarm 2 setpoint, press the ENTER button to reset the alarm circuit. Resetting the alarm circuit turns off the A2 light, resets the OLED display, and de-energizes the alarm 2 relay.
NOTE: If the ENTER button is pressed while the M2A is in an alarm 2 condition, the A2 LED
will flash but all other indications will remain unchanged.
You cannot de-energize the alarm 2 relay until the gas reading falls below the alarm 2 setpoint.

Fail Condition

Fail Condition Indications
The M2A indicates a fail condition for any of the following:
The detector wiring is disconnected or incorrectly connected.
The display reading is -10% of full scale or lower.
When the M2A senses a fail condition, it alerts you as follows:
The F LED turns on.
The gas reading is replaced by the FAI L message.
The fail relay de-energizes.
NOTE: The fail alarm has a 30 second delay.
M2A Transmitter Operator’s Manual 39
Responding to a Fail Condition
1. Verify that the detector wiring is correctly and securely connected.
2. If the M2A has a replaceable plug-in sensor, verify that the replaceable plug-in sensor in the detector housing is properly installed.

Low Power Alarm

Low Power Alarm Indications
The M2A senses a low power condition when the DC power source is 9.5 volts or less.
WARNING: While in a low power condition, the M2A is not an active gas monitor.
When the M2A senses a low power condition, it alerts you as follows:
The F LED turns on.
The message LowPower is indicated on the top line of the OLED display and the input voltage is displayed on the bottom line of the OLED display.
The fail relay de-energizes.
NOTE: The low power alarm cannot be cleared using the ENTER button.
When the voltage increases to 10.0 volts, the low power alarm is cleared and the M2A will begin its warm-up sequence.
Responding to a Low Power Condition
1. Determine and correct the cause of the low power condition.
2. When the input power increases above 10.0 volts, the M2A will begin its warm-up sequence.
3. Verify that the M2A enters normal operation after its warm-up sequence. If necessary, perform a fresh air adjustment. See “Adjusting the Fresh Air Reading” on page 33.
40 M2A Transmitter Operator’s Manual

Chapter 5: Configuration Mode

Enter
Config?

Overview

This chapter describes how to view and change M2A parameters using Configuration Mode. It is accessed using the program buttons.
Configuration Mode includes a 5-minute time-out feature. If you do not press a control button for 5 minutes, the M2A automatically returns to normal operation.
NOTE: If the M2A returns to normal operation because of a time-out, it enters a warm-up
period just as it does when it is first turned on.
If you are installing a new M2A, it has been setup at the factory. Use Configuration Mode only if you want to change the M2A’s setup. If you want to change the detector type see “Chapter 6: Gas Type Mode” on page 44.

Viewing & Changing M2A Parameters

1. While in normal operation, simultaneously press and hold the UP/YES and ENTER buttons for 5 seconds to enter Configuration Mode. Release the buttons when the following screen appears.
2. If you want to exit Configuration Mode, press and release the DOWN/NO button and the M2A will return to normal operation.
If you want to continue in Configuration Mode, press and release the UP/YES button. The target gas and the full scale will be displayed for a few seconds before the first adjustable parameter, the alarm 1 setpoint, is displayed.
3. If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter is OK, press the ENTER button to proceed to the next parameter.
Table 11 lists the M2A parameters you can set. Table 11 also lists the factory set value for each parameter.
M2A Transmitter Operator’s Manual 41
Table 11: Configuration Parameters
Parameter
(Factory Set Value)
ALARM-1 (level) (See “M2A Specifications” on page 7)
ALARM-1 (activation) (Decrease for oxygen M2A, Increase for all other types)
ALARM-1 (relay action) (N. DE-EN)
ALARM-1 (relay reset) (LATCH)
A1 OnDy (alarm 1 on delay) (1 secs)
A1 OffDy (alarm 1 off delay) (0 sec.)
Description
The gas reading at which the M2A initiates an alarm 1 condition.
Indicates if the alarm 1 circuit is activated by gas readings increasing (Increase) or decreasing (Decrease) to the ALARM-1 Level.
If set as N. DE-EN, the alarm 1 relay is de-energized in normal operation and energizes when an alarm 1 condition is initiated. If set as N. EN, the alarm 1 relay is energized in normal operation and de-ener­gizes when an alarm 1 condition is initiated.
If set as LATCH, you must press the ENTER button to reset the alarm 1 circuit after the alarm 1 condition passes. If set as SELF-RST, the M2A automatically resets the alarm 1 circuit after the alarm 1 condition passes.
The amount of time the M2A delays activation of the alarm 1 circuit once an alarm 1 condition is initiated. It can be set in 1 second increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute incre­ments from 15 - 60 minutes.
The amount of time the M2A delays turning off the alarm 1 circuit once an alarm 1 condition passes. This parameter appears only if the alarm 1 relay reset setting is set to SELF-RST. It can be set in 1 second increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 ­60 minutes.
ALARM-2 (level) (See “M2A Specifications” on page 7)
ALARM-2 (activation) (Increase for all types)
ALARM-2 (relay action) (N. DE-EN)
ALARM-2 (relay reset) (LATCH)
A2 OnDy (alarm 2 on delay) (1 secs)
A2 OffDy (alarm 2 off delay) (0 sec.)
The gas reading at which the M2A initiates an alarm 2 condition.
Indicates if the alarm 2 circuit is activated by gas readings increasing (Increase) or decreasing (Decrease) to the ALARM-1 Level.
If set as N. DE-EN, the alarm 2 relay is de-energized in normal operation and energizes when an alarm 2 condition is initiated. If set as N. EN, the alarm 2 relay is energized in normal operation and de-ener­gizes when an alarm 2 condition is initiated.
If set as LATCH, you must press the ENTER button to reset the alarm 2 circuit after the alarm 2 condition passes. If set as SELF-RST, the M2A automatically resets the alarm 2 circuit after the alarm 2 condition passes.
The amount of time the M2A delays activation of the alarm 2 circuit once an alarm 2 condition is initiated. It can be set in 1 second increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute incre­ments from 15 - 60 minutes.
The amount of time the M2A delays turning off the alarm 2 circuit once an alarm 2 condition passes. This parameter appears only if the alarm 2 relay reset setting is set to SELF-RST. It can be set in 1 second increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 ­60 minutes.
42 M2A Transmitter Operator’s Manual
Table 11: Configuration Parameters (Continued)
Parameter
(Factory Set Value)
ZeroSupp (0.0% of the detection range in terms of the detection units for carbon dioxide channel types, 0.5% oxygen for oxygen types, 2% of the detection range in terms of the detection units for all other types)
FILTER (5 secs)
CAL Time (15 mins)
Auto Zero (Disabled for carbon dioxide channels; Enabled for all other channel types)
4. When you have scrolled through all the adjustable parameters, SAVE IT? YES/NO appears on the display.
Description
The zero suppression feature helps prevent “jumpy” readings near the fresh air reading. For example, if the zero suppression setting for a combustible LEL detector is
2.0% LEL, the M2A will display a reading of 0% LEL for gas readings from ­2% LEL to 2% LEL. It is settable from 0 to 6% of the detection range.
The filter feature helps “smooth out” jumpy or noisy signals from the detector. You can set the filter from 0 seconds to 60 seconds in 5 seconds increments. The displayed gas reading is the average reading over the previous time period defined by the filter setting. So if the filter is set to 5 seconds, the displayed gas reading is the average over the past 5 seconds.
The calibration time-out sets the amount of time after the last button push while in Calibration Mode that the M2A will wait before returning to normal opera­tion. If the calibration time-out is set to 15 mins, then M2A will return to normal operation automatically 15 minutes after the last button push.
If set to Enabled, the M2A’s firmware will monitor the zero (fresh air) reading and make automatic zero adjustments under precisely specified conditions if it determines that a reading change is due to normal sensor drift. If set to Disabled, the Auto Zero function is turned off.
5. If you do not wish to save the adjustments and want to exit Configuration Mode, press and release the DOWN/NO button. The DO OVER? YES/NO message will display. Press and release the DOWN/NO button. The ABORT? YES/NO message will display. Press the UP/ YES button to return to normal operation.
If you wish to change some of the adjustments made, press and release the DOWN/NO button. The DO OVER? YES/NO message will display. Press and release the UP/YES button. The Re-do Config message will display and the M2A will return to the first adjustable parameter. Go back to step 3 and continue.
If you wish to save the adjustments made, press and release the UP/YES button. Config Saved is indicated on the display for a few seconds and the M2A returns to normal operation.
M2A Transmitter Operator’s Manual 43

Chapter 6: Gas Type Mode

Select GasType?

Overview

This chapter describes how to use Gas Type Mode to select the M2A’s gas type. The gas type determines the target gas and detection range. The combustible gas/CO
have a different terminal PCB and run on different firmware. Your M2A will only have gas type choices available that it can support.
Gas Type Mode includes a 5-minute time-out feature. If you do not press a button for 5 minutes, the M2A automatically returns to normal operation.
NOTE: If the M2A enters normal operation because of a program time-out, it enters a warm-up
period just as it does when it is first turned on.

Selecting the Gas Type

and toxic/oxygen M2As
2
1. While in normal operation or during the warm-up sequence, press and hold the UP/YES and DOWN/NO buttons for five seconds. Release them when the following screen appears.
2. If you want to exit Gas Type Mode, press and release the DOWN/NO button. The display will indicate NO CHANGE and the M2A will return to normal operation.
If you want to continue in Gas Type Mode, press and release the UP/YES button. The top display line will indicate GasType? and the bottom display line will indicate the target gas choice.
3. Use the UP/YES or DOWN/NO button to scroll through the choices of target gas. Table 12 and Table 13 below list the choices of target gas for the M2A.
Table 12: Combustible Gas/CO2 Gas Types
Gas Type
Choices
HC 0 - 100% LEL
iBUTANE 0 - 100% LEL
PROPANE 0 - 100% LEL
HEXANE 0 - 100% LEL
Detection Range
HYDROGEN 0 - 100% LEL
METHANE 0 - 100% Volume
METHANE 0 - 100% LEL
CO2 0 - 5,000 ppm
44 M2A Transmitter Operator’s Manual
Table 12: Combustible Gas/CO
Gas Type
Choices
CO2 0 - 2,000 ppm
CO2 0 - 100% volume
CO2 0 - 50.0% volume
CO2 0 - 5.00% volume
METHANE 0 - 5.00% volume
METHANE 0 - 8980 ppm*
*Use this gas type for 0-9000 ppm CH
is the highest displayed value. Anything higher is displayed as OVER.
Table 13: Toxic/Oxygen Gas Types
Gas Type
Choices
H2S 0 - 100 ppm
Detection Range
Detection Range
Gas Types
2
. 8980 ppm
4
CO 0 - 300 ppm
CO 0 - 200 ppm
CO 0 - 100 ppm
CL2 0 - 3.00 ppm
CL2 0 - 10.0 ppm
NH3 0 - 75.0 ppm
SO2 0 - 6.00 ppm
ASH3 0 - 1.50 ppm
B2H6 0 - 5.00 ppm
CLO2 0 - 1.00 ppm
F2 0 - 3.00 ppm
GEH4 0 - 1.50 ppm
HCL 0 - 15.0 ppm
HCN 0 - 15.0 ppm
HF 0 - 9.00 ppm
NO 0 - 100 ppm
O3 0 - 1.00 ppm
PH3 0 - 1.00 ppm
M2A Transmitter Operator’s Manual 45
Table 13: Toxic/Oxygen Gas Types
Gas Type
Choices
SIH4 0 - 15.0 ppm
OXYGEN 0 - 25.0% volume
Detection Range
NOTE: Consult factory for availability of detectors for toxic M2A types not listed in “M2A
Specifications” on page 7.
4. When the desired target gas is on the display, press and release the ENTER button. The display will ask SAVE IT? YES/NO.
5. To discard the gas type change, press and release the DOWN/NO button. The display will ask
DO OVER? YES/NO. Press the DOWN/NO button. The display will ask ABORT? YES/ NO. Press the UP/YES button. The M2A will return to normal operation without changing the
gas type.
To save the gas type, press and release the UP/YES button. The display will indicate Gas Type Updated, the display will indicate the target gas and the detection range for a few seconds, and then the M2A will enter Configuration Mode automatically so that the parameter settings in Configuration Mode can be verified or changed.
6. While in Configuration Mode press and release the ENTER button to accept the displayed parameter setting and move to the next one. Use the UP/YES and DOWN/NO buttons to adjust a setting. See “Chapter 5: Configuration Mode” on page 41 for a complete description of Configuration Mode.
7. When you have reviewed and accepted the last item in Configuration Mode by pressing the ENTER button, display will ask SAVE IT? YES/NO.
8. To save the configuration settings, press and release the UP/YES button.The display will indicate Config Saved and the M2A will begin its warm-up sequence.
To discard the settings and review them again, press the DOWN/NO button. The display will ask DO OVER? YES/NO. Press the UP/YES button and the display will indicate Re-do Config, then display the target gas and the detection range for a few seconds before returning to Configuration Mode. Repeat steps 6 - 8 until the settings are as desired.
9. After the M2A completes its warm-up sequence, the display will indicate CAL NEEDED. Since the gas type has been changed, a successful calibration must be performed before the M2A can enter normal operation.
10. Press and release the UP/YES button to enter Calibration Mode. Normally, you must hold the UP/YES button for 5 seconds to enter Calibration Mode, but since a calibration is required, only a momentary push is needed to enter Calibration Mode after changing the gas type.
11. See “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51 or “Calibration, Oxygen Version” on page 54 for calibration instructions.
NOTE: When calibrating an M2A after changing the gas type, the M2A will not ask if you want
to calibrate, or whether you want to perform a fresh air adjustment, span adjustment or zero adjustment. Since a complete calibration is required, the calibration sequence will proceed without asking if you want to perform certain operations.
46 M2A Transmitter Operator’s Manual

Chapter 7: Maintenance

Overview

This chapter describes procedures for performing preventive maintenance, troubleshooting, calibrating the M2A, and replacing field replaceable parts. It includes separate procedures for calibrating the combustible gas/CO

Preventive Maintenance

This section describes a recommended preventive maintenance schedule to ensure the optimum performance of the M2A. It includes daily, quarterly, and biannual procedures.

Daily

Verify a display reading of zero (20.9% for an oxygen M2A). Investigate significant changes in the reading.
/toxic and oxygen versions of the M2A.
2

Quarterly

Calibrate M2As that do not have an IR detector as described in “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51 or “Calibration, Oxygen Version” on page 54. See “Calibration Frequency” on page 50 for a discussion of calibration frequency guidelines.

Biannually

Calibrate M2As that have an IR detector as described in “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51. See “Calibration Frequency” on page 50 for a discussion of calibration frequency guidelines.
M2A Transmitter Operator’s Manual 47

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the M2A.
NOTE: This troubleshooting guide describes M2A problems only. If the M2A is connected to a
controller, see the controller operator’s manual for problems you may encounter with the controller.
Table 14:Troubleshooting the Combustible Gas Detector
Condition Symptom(s) Probable Causes Recommended Action
No Power The display backlight is
off and the display screen is blank.
Frequent or Suspect Alarms
Flickering Display
The M2A alerts you to frequent or suspect alarms while the fresh air readings remain on zero (20.9 for oxygen).
The display reading flickers often.
The power wiring is disconnected or misconnected.
The display cable is disconnected or misconnected.
The M2A is experiencing false readings due to RFI or EMI.
The detector wiring is disconnected, misconnected, or intermittent.
The M2A is experiencing false readings due to RFI or EMI.
The noise filter setting is too low.
The zero suppression setting is too low.
The display screen is malfunctioning.
1. Verify that the wiring to the power source or controller is correct and secure.
2. Verify that the ribbon cable connecting the terminal PCB to the control PCB is securely installed. This ribbon cable plugs into a connector on the top edge of the control PCB and on the bottom of the terminal PCB.
3. If the power difficulties continue, contact RKI for further instruction.
1. Verify that the M2A wiring is properly shielded. See “Wiring the M2A Transmitter” on page 27.
2. Verify that the detector wiring is correct and secure.
3. Increase the alarm on delay setting in Configuration Mode.
4. If the frequent or suspect alarm difficulties continue, contact RKI for further instruction.
1. Verify that the M2A wiring is properly shielded. See “Wiring the M2A Transmitter” on page 27.
2. Verify that the detector wiring is correct and secure.
3. Increase the filter setting in Configuration Mode.
4. Increase the zero suppression setting in Configuration Mode.
5. If the display difficulties continue, contact RKI for further instruction.
48 M2A Transmitter Operator’s Manual
Table 14:Troubleshooting the Combustible Gas Detector (Continued)
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition M2A indicates a fail
condition.
Controller indicates a fail condition
Slow or No Response/ Difficult or Unable to Calibrate
Unable to successfully set the fresh air or sp an ( zer o fo r ox yge n) reading during calibration.
M2A requires frequent calibration.
Note: See “Calibration Frequency” on page 50 for calibration frequency guidelines.
The detector wiring to the terminal PCB is disconnected or misconnected.
The wiring from the M2A to the controller is disconnected or misconnected.
The detector fresh air signal is low enough to cause a fail condition.
The detector is malfunctioning.
The terminal or display PCB is malfunctioning.
The calibration cylinder is low, out-dated, or defective.
For non-XP CO, H
detectors: The
O
2
S, or
2
membrane on the detector housing cap is blocked with dirt of some other particulate contamination.
For CT-7 Series detectors: The hydrophobic membrane on the plug-in sensor face is blocked with dirt or some other particulate contamination.
The detector is malfunctioning.
The terminal PCB is malfunctioning.
1. Verify that the detector wiring is correct and secure.
2. Verify that the wiring between the M2A and the controller is correct and secure.
3. Calibrate the M2A.
4. If the fail condition continues, replace the detector.
5. If the fail condition continues, contact RKI for further instruction.
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. For non-XP CO, H2S, or O2 detectors: Check the face of the detector housing cap and remove any particulate contamination from the hydrophobic membrane, if necessary. If the membrane appears saturated with contamination or is damaged, replace the membrane as described in “Replacing the Hydrophobic Membrane (Non-XP CO, H
S, and O2
2
Detectors)” on page 60.
3. For CT-7 Series detectors: Check the plug-in sensor face and remove any particulate contamination from the hydrophobic membrane, if necessary. If the membrane appears saturated with contamination or is damaged, replace the membrane as described in “Replacing the Hydrophobic Membrane (CT-7 Series Toxic Detector)” on page 60.
4. If the calibration/response difficulties continue, replace the detector.
5. If the calibration/response difficulties continue, contact RKI for further instruction.
M2A Transmitter Operator’s Manual 49

Calibration Frequency

Some M2As have IR (infrared) type detectors, such as an M2A for CO2 or an LEL M2A with an IR
detector. Most of the M2As do not have an IR detector. For example, an H electrochemical detector and some of the combustible M2As use catalytic detectors. The M2As that
use IR detectors typically need less frequent calibration as is discussed below. See Table 3, “Infrared (IR) Detectors,” on page 11 to see which M2As are IR versions.
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 to 6 months for M2As that do not have an IR detector and every 6 to 12 months for M2As that have an IR detector is adequate for most M2A applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months for M2As that do not have an IR detector and every 6 months for M2As that have an IR detector.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where the toxic or combustible target gas is not normally present, and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months for a non-IR M2A or 12 months for an IR M2A is adequate.
If an application is very demanding, for example if a combustible or toxic gas is present often and in significant concentrations or the environment is not well controlled, then more frequent calibration than every 3 months for a non-IR M2A or every 6 months for an IR M2A may be necessary. For combustible gas detection with an M2A that uses a catalytic detector, if potential catalyst poisons are known or likely to be present, more frequent calibration than every 3 months will be necessary.
S M2A has an
2
50 M2A Transmitter Operator’s Manual

Calibration, Combustible Gas, CO2, and Toxic Versions

This section describes how to calibrate the combustible gas, CO2, and toxic gas versions of the M2A. It includes procedures to prepare for calibration, enter Calibration Mode, adjust the fresh air
(zero) setting, adjust the span setting, and return to normal operation. See the next section, “Calibration, Oxygen Version” for instructions to calibrate an oxygen M2A.
WARNING: The M2A is not an active gas monitoring device during the calibration
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA and all relays will remain in their non-alarm state while the M2A is in Calibration Mode. The output signal will not indicate current rea dings and the relays will not resume operating normally until the M2A is in normal operation again.

Special Note About Calibrating a ClO2 Detector

RKI Instruments, Inc. recommends using a ClO2 generator to set the ClO2 detector’s response reading. The flow rate on the generator needs to be set to 0.5 LPM and RKI Instruments, Inc.
recommends that 0.5 ppm ClO
related to the use of a calibration gas cylinder in the instructions below can be disregarded.
be used. If a ClO2 generator is used for calibration, the steps
2

Preparing for Calibration

NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a zero air cylinder, a fixed flow regulator with an on/off knob, a calibration cup for the appropriate sensor, and a short piece of sample tubing to connect the regulator to the calibration cup.
WARNING: RKI Instruments, Inc. recommends that you dedicate a regulator for use
with chlorine (Cl for any other gases, particularly hydrogen sulfide (H2S).
NOTE: While in the calibration program, if there is no switch activity for the calibration time-
out period the unit will return to normal operation. See “Viewing & Changing M2A Parameters” on page 41 for instructions to set the calibration time-out.
NOTE: The following procedure assumes that the target gas is present in a high enough
concentration to affect the fresh air (zero) reading. If a fresh air environment can be verified, applying zero air to the detector is not necessary when setting the zero (span for oxygen) reading.
) gas and that you do not use that dedi cated regulator
2
CAUTION: If calibrating a CT-7 Series M2A, allow the instrument to warm up for 15 minutes
before perfo rming a calibration.
1. Connect the calibration kit sample tubing to the calibration cup hose barb.
2. Screw the calibration cup onto the detector.
M2A Transmitter Operator’s Manual 51
NOTE: For an ESM-01 detector, you must unscrew the removable splashguard from the
Calib?
YES/NO
detector housing cap so you can push the appropriate calibration cup onto the ESM-01 sensor. Make sure to use the correct calibration cup for the sensor being used. There is a Cl
type calibration cup for a Cl2 sensor, an NH3 type calibration cup for an NH3
2
sensor, and a general type (usually labeled SO sensors.
3. Connect the fixed flow regulator to the calibration cup using the sample tubing.
NOTE: When calibrating an NH3 ESM-01, the 0.25 LPM flowrate regulator must be used. All
other M2As require a 0.5 LPM flowrate regulator.
) calibration cup for all other ESM-01
2

Non-Intrusive Calibration

If the M2A is installed in a classified area and non-intrusive calibration is required, follow the instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic switch and perform the same operation as pressing a control switch button, touch the magnet in the end of the magnetic wand to the M2A’s glass cover directly above the magnetic switch you wish to actuate. Touching the glass and removing the wand is the same as pressing and releasing a button. Touching the glass and keeping the wand in place is the same as pressing and holding a button.

Calibration Gas Response Memory Feature

The M2A has the capability to “remember” the detector’s response to the calibration gas after the gas is removed from the detector during the fresh air and span adjustment procedure. This feature enables one person to perform a calibration if the detector is mounted remotely from the M2A. When zero air is applied to the M2A during a fresh air adjustment, the M2A will freeze the display reading at the lowest response and the M2A will continue to display this reading and retain it in its memory until the fresh air adjustment procedure is completed. When calibration gas is applied to the M2A during a span adjustment, the M2A will freeze the display reading at the highest response to the calibration gas. The calibration gas can then be removed and the M2A will continue to display this reading and retain it in its memory until the span adjustment procedure is completed.

Adjusting the Fresh Air Reading

1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter Calibration Mode. Release the button when the following screen appears.
2. If you want to continue with calibration, press and release the UP/YES button. The display will indicate the target gas and CAL Mode for a few seconds before showing FreshAir Adjust?.
If you want to exit Calibration Mode, press and release the DOWN/NO button. The M2A will indicate Leaving CAL Mode and the M2A will return to normal operation.
3. To continue with adjusting the fresh air reading, press and release the UP/YES button. ENTER will alternate with FreshAir on the top display line and the current gas reading will be on the bottom display line.
To skip adjusting the fresh air reading, press and release the DOWN/NO button. The display will indicate SPAN w/Cal Gas?. Skip to the next section, “Adjusting the Span Setting”.
4. Screw the zero air cylinder into the fixed flow regulator.
5. Turn the regulator knob counterclockwise to open the regulator.
52 M2A Transmitter Operator’s Manual
6. Allow zero air to flow for two minutes. If there was gas present causing a reading on the M2A, the reading should decrease and stabilize after two minutes.
7. Turn the regulator knob clockwise to close the regulator.
8. Unscrew the zero air cylinder from the fixed flow regulator leaving the calibration cup on the detector. The M2A will continue to display the minimum gas response on the display and retain the response level in its memory.
9. Press and release the ENTER button. The M2A will perform a fresh air adjustment and the display will indicate SPAN w/Cal Gas?.
NOTE: If the fresh air adjustment fails, see “Troubleshooting” on page 48 for recommended
actions.

Adjusting the Span Setting

1. If you want to continue with adjusting the span setting, press and release the UP/YES button. APPLY will alternate with SPAN Gas on the top display line and the current gas reading will be on the bottom display line.
If you want to skip adjusting the span reading, press and release the DOWN/NO button. The display will indicate Leaving Cal Mode and the M2A will return to normal operation.
2. Screw the calibration gas cylinder into (onto for an H
S cylinder) the fixed flow regulator.
2
3. Turn the regulator knob counterclockwise to open the regulator (or start the flow of gas from a generator).
4. Allow calibration gas to flow for two minutes. The gas reading should be stable after two minutes.
5. Turn the regulator knob clockwise to close it (or stop the flow of gas from a generator).
6. Unscrew the calibration gas cylinder from the fixed flow regulator and unscrew the calibration cup from the detector. For an ESM-01 detector, pull the calibration cup off of the ESM-01 sensor and screw the splashguard back onto the detector housing cap. The M2A will continue to display the maximum gas response on the display and retain the response level in its memory.
7. Adjust the gas reading up or down to match the calibration gas cylinder concentration by using the UP/YES and DOWN/NO buttons, then press and release the ENTER button.
8. The M2A will perform a span operation. The display will indicate SPAN Gas PASS for a few seconds, then indicate SPAN Gas SAVED before indicating Leaving CAL Mode for a few seconds.
NOTE: If the span adjustment fails, see “Troubleshooting” on page 48 for recommended
actions.
9. The display will now alternate between the normal operation screen and the message REMOVE CAL GAS for 1 minute. If the calibration gas has not been removed from the detector, remove it now to avoid unwanted alarms.
During this 1 minute period, the signal output will remain fixed at 3.5 mA and the relays will remain in their non-alarm state to avoid unwanted alarms while the calibration gas clears from the detector. At the end of the 1 minute period, the REMOVE CAL GAS message will stop appearing and the M2A will return to normal operation.
10. Store the components of the calibration kit in a safe place.
M2A Transmitter Operator’s Manual 53

Calibration, Oxygen Version

This section describes how to calibrate the oxygen version of the M2A. It includes procedures to prepare for calibration, enter Calibration Mode, set the fresh air (span) reading, set the zero reading, and return to normal operation.
WARNING: The M2A is not an active gas monitoring device during the calibration
procedure. The 4-20 mA output signal will “freeze” at 17.4 mA and all relays will remain in their non-alarm state while the M2A is in Calibration Mode. The output signal will not indicate current rea dings and the relays will not resume operating normally until the M2A is in normal operation again.

Preparing for Calibration

NOTE: The following procedure assumes the use of a calibration kit which includes a zero air
(20.9% oxygen) calibration cylinder to set the fresh air reading, a 100% nitrogen (0% oxygen) calibration gas cylinder to set the zero reading, a fixed flow regulator with an on/off knob, a calibration cup, and a short piece of sample tubing to connect the regulator to the calibration cup. A cylinder with up to an 18% oxygen concentration may be used instead of a 100% nitrogen cylinder to set the zero reading.
NOTE: While in calibration mode, if there is no switch activity for the calibration time-out
period the unit will return to normal operation. See “Viewing & Changing M2A Parameters” on page 41 for instructions to set the calibration time-out.
NOTE: The following procedure assumes that the oxygen concentration in the calibration area
is not normal, 20.9%, but is oxygen deficient. If a fresh air environment can be verified, applying zero air to the detector is not necessary when adjusting the fresh air (span) setting.
1. Connect the calibration kit sample tubing to the calibration cup hose barb.
2. Screw the calibration cup onto the detector.
3. Connect the fixed flow regulator to the calibration cup using the sample tubing.

Non-Intrusive Calibration

If the M2A is installed in a classified area and non-intrusive calibration is required, follow the instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic switch and perform the same operation as pressing a control switch button, touch the magnet in the end of the magnetic wand to the M2A’s glass cover directly above the magnetic switch you wish to actuate. Touching the glass and removing the wand is the same as pressing and releasing a button. Touching the glass and keeping the wand in place is the same as pressing and holding a button.

Calibration Gas Response Memory Feature

The M2A has the capability to “remember” the detector’s response to the calibration gas after the gas is removed from the detector during the fresh air or zero adjustment procedure. This feature enables one person to perform a calibration if the detector is mounted remotely from the M2A. When zero air is applied to the M2A during a fresh air adjustment, the M2A will freeze the display
54 M2A Transmitter Operator’s Manual
reading at the highest response and the M2A will continue to display this reading and retain it in its
Calib? YES/NO
memory until the fresh air adjustment procedure is completed. When calibration gas is applied to the M2A during a zero adjustment, the M2A will freeze the display reading at the lowest response to the calibration gas. The calibration gas can then be removed and the M2A will continue to display this reading and retain it in its memory until the zero adjustment procedure is completed.

Adjusting the Fresh Air Setting

1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter Calibration Mode. Release the button when the following screen appears.
2. If you want to continue with calibration, press and release the UP/YES button to continue. The display will indicate OXYGEN CAL Mode for a few seconds before showing FreshAir Adjust?.
If you want to exit Calibration Mode, press and release the DOWN/NO button. The M2A will indicate Leaving CAL Mode and the M2A will return to normal operation.
3. To continue with adjusting the fresh air (span) reading, press and release the UP/YES button. ENTER will alternate with FreshAir on the top display line and the current oxygen reading will be on the bottom display line.
To skip adjusting the fresh air reading, press and release the DOWN/NO button. The display will indicate ZERO w/Cal Gas?. Skip to the next section, “Adjusting the Zero Setting”.
4. Screw the zero air cylinder into the fixed flow regulator.
5. Turn the regulator knob counterclockwise to open the regulator.
6. Allow zero air to flow for 2 minutes. If the oxygen concentration in the area is less than 20.9, the reading should stabilize after 2 minutes.
7. Turn the regulator knob clockwise to close the regulator.
8. Unscrew the zero air cylinder from the fixed flow regulator leaving the calibration cup on the detector. The M2A will continue to display the maximum gas response on the display and retain the response level in its memory.
9. Press and release the ENTER button. The M2A will perform a fresh air adjustment and the display will indicate ZERO w/Cal Gas?.
NOTE: If the fresh air adjustment fails, see “Troubleshooting” on page 48 for recommended
actions.

Adjusting the Zero Setting

1. If you want to continue with adjusting the zero setting, press and release the UP/YES button. APPLY will alternate with ZERO Gas on the top display line and the current oxygen reading will be on the bottom display line.
If you want to skip adjusting the zero reading, press and release the DOWN/NO button. The display will indicate Leaving Cal Mode and the M2A will return to normal operation.
2. Screw the 100% nitrogen gas cylinder into the fixed flow regulator.
3. Turn the regulator knob counterclockwise to open the regulator.
4. Allow gas to flow for 2 minutes. The gas reading should be stable after 2 minutes.
M2A Transmitter Operator’s Manual 55
5. Turn the regulator knob clockwise to close the regulator.
6. Unscrew the nitrogen cylinder from the fixed flow regulator and unscrew the calibration cup from the detector. The M2A will continue to display the minimum gas response on the display and retain the response level in its memory.
7. Adjust the oxygen reading up or down to match the 100% nitrogen cylinder oxygen concentration of 0 by using the UP/YES and DOWN/NO buttons.
NOTE: If a cylinder other than a 100% nitrogen (0% oxygen) cylinder is used to set the zero
reading, such as a cylinder with 12% oxygen, set the zero reading to match the cylinder’s oxygen concentration.
8. Press and release the ENTER button. The M2A will perform a zero operation. The display will indicate ZERO Gas PASS for a few seconds, then indicate ZERO Gas SAVED before indicating Leaving CAL Mode for a few seconds.
NOTE: If the zero adjustment fails, see “Troubleshooting” on page 48 for recommended
actions.
9. The display will now alternate between the normal operation screen and the message REMOVE CAL GAS for 1 minute. If the calibration gas has not been removed from the detector remove it now to avoid unwanted alarms.
During this 1 minute period, the signal output will remain fixed at 17.4 mA and the relays will remain in their non-alarm state to avoid unwanted alarms while the calibration gas clears from the detector. At the end of the 1 minute period, the REMOVE CAL GAS message will stop appearing and the M2A will return to normal operation.
10. Store the components of the calibration kit in a safe place.

Replacing Components of the M2A

Replacing the Gas Detector

NOTE: For M2A versions that do not have a replaceable plug-in sensor, or if you determine
that it is necessary to replace the entire detector, use the procedure in this section to change the entire detector. For M2A versions that include a replaceable plug-in sensor, see the appropriate instructions in the sections that follow this one to replace the plug-in sensor.
1. Turn off or disconnect power to the M2A.
2. Remove the junction box cover.
3. Grasp the control PCB by its edges.
4. Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull too hard and damage the cable which connects the control and terminal PCBs.
5. Let the control PCB hang by the cable. The terminal strips are now visible on the terminal PCB. The control PCB may be left hanging while wiring is done. If desired, the control PCB may be disconnected from the cable and set aside while wiring.
6. Remove the detector terminal strip from its socket.
56 M2A Transmitter Operator’s Manual
7. Disconnect the detector leads from the detector terminal strip. Note the position of the color­coded leads as you remove them.
8. Unscrew the detector from the junction box.
9. Guide the replacement detector leads through the bottom conduit port of the junction box, then screw the mounting threads of the detector into the conduit port. If necessary for environmental conditions, apply thread sealant or Teflon tape to the hub and/or detector threads to seal them.
10. Connect the detector leads to the detector terminal strip as follows:
Catalytic Combustible Gas Detector, IR Combustible Gas Detector, and IR CO
Detector
2
Red wire to terminal labeled LEL RED, white to terminal labeled LEL WHT, green wire to terminal labeled LEL GRN, black wire to terminal labeled LEL BLK.
CO or H2S Gas Detector
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
ESM-01 Toxic Gas Detector or CT-7 Series Toxic Gas Detector
Color coded wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -. See Figure 28 on page 30 and Figure 29 on page 31 for the wire color coding.
Oxygen Gas Detector
Green wire to terminal labeled OXY +, white wire to terminal labeled OXY -.
11. Re-install the detector terminal strip into its socket.
12. Re-install the control PCB (and ribbon cable if necessary). Be sure the ribbon cable is routed down below the control PCB so it will not be damaged by the cover when it is screwed back on.
13. Secure the junction box cover to the junction box.
14. Turn on or reconnect power to the M2A.
NOTE: Allow the replacement detector to warm up for 15 minutes before you continue with the
next step.
15. Calibrate the replacement detector as described in “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51 or “Calibration, Oxygen Version” on page 54.

Replacing the CO, H2S, or CT-7 Series Plug-in Sensor

NOTE: This procedure is applicable for: the CSA type, explosion proof CO and H2S detectors,
the non-explosion proof CO and H2S detectors, and the CT-7 Series detectors.
1. Turn off or disconnect power to the M2A.
2. Unscrew the detector cap from the housing body. Make sure not to lose the cap gasket. See Figure 5, Figure 6, and Figure 8 for illustrations of the detectors.
3. Unplug and remove the sensor.
M2A Transmitter Operator’s Manual 57
4. If you are replacing a CO sensor, remove the rubber boot and charcoal filter from the old
Remove jumper
sensor.
NOTE: RKI instruments recommends replacing the charcoal disk filter whenever a CO sensor
is replaced.
If you are replacing the 65-2428 explosion-proof H
S detector’s sensor, remove the rubber
2
boot and spacer from the old sensor.
5. Remove the replacement sensor from its packaging and remove the wire jumper. This wire jumper is installed on the sensor pins for shipment or storage but must be removed for the sensor to operate properly when installed in a detector.
WARNING: The sensor will not operate properly if the wire jumper is not removed.
Figure 31: Plug-In Sensor Jumper Removal
6. If you are replacing a CO sensor, install the rubber boot and charcoal filter onto the new sensor’s face.
If you are replacing the 65-2428 explosion-proof H
S detector’s sensor, install the rubber boot
2
and spacer on the new sensor’s face.
7. Carefully match the replacement sensor’s male pins with the 4-socket pattern in the top section and plug it in.
8. Make sure the cap gasket is in place and screw the detector cap back onto the housing body.
9. Turn on or reconnect power to the M2A.
CAUTION: Allow the replacement sensor to warm up for 15 minutes before you continue with
the next step.
10. Calibrate the detector as described in “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51.

Replacing the CO Sensor Charcoal Filter

The charcoal filter is used to scrub out interfering gases from the environment being monitored, such as H
CO M2A, the charcoal filter may be saturated and no longer scrubbing out interfering gases and it may be necessary to change the charcoal filter.
1. Turn off or disconnect power to the M2A.
2. Unscrew the detector housing cap from the housing body. Make sure not to lose the cap gasket.
3. Remove the rubber retaining boot with the charcoal filter disk from the sensor.
4. Replace the charcoal filter disk in the rubber retaining boot and install the boot onto the sensor.
58 M2A Transmitter Operator’s Manual
S and certain hydrocarbons. If you are experiencing unexplained upscale readings on a
2
See Figure 6 for an illustration of the CO detector.
5. Make sure that the cap gasket is in place and screw the detector cap back onto the housing body.
6. Turn on or reconnect power to the M2A.

Replacing an Oxygen Plug-in Sensor

NOTE: This procedure is applicable for both the CSA type, explosion proof oxygen detector
(65-2514RK) and non-explosion proof, replaceable sensor oxygen detector (65-2494RK).
1. Turn off or disconnect power to the M2A.
2. Unscrew the detector cap from the detector housing body. Take care not to lose the cap gasket. See Figure 4 for an illustration of the oxygen detectors.
3. Unplug and remove the sensor.
4. Carefully match the replacement sensor’s male pins with the 2-socket pattern in the top section and plug it in.
5. Make sure the cap gasket is in place and screw the detector cap back on to the detector housing.
6. Turn on or reconnect power to the M2A.
NOTE: Allow the replacement sensor to warm up for 15 minutes before you continue with the
next step.
7. Calibrate the detector as described in “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51.

Replacing an ESM-01 Plug-in Sensor

1. Turn off or disconnect power to the M2A.
2. Unscrew the detector cap from the detector housing body. Take care not to lose the cap gasket. If the splashguard begins to unscrew from the detector cap, make sure to remove the detector cap also. See Figure 7 for an illustration of the ESM-01 detector.
3. Unplug and remove the sensor.
4. Carefully match the replacement sensor’s plug with the socket in the housing body and plug it in.
W ARNING: The ESM-01 detectors cannot b e change d from one type to another. Replace the
plug-in sens or only with th e same type of E S M-01 sensor.
5. Make sure the cap gasket is in place and screw the detector cap back onto the housing body. If the splash guard was unscrewed from the detector cap, screw it back onto the detector cap.
6. Turn on or reconnect power to the M2A.
NOTE: Allow the replacement sensor to warm up for 15 minutes before you continue with the
next step.
7. Calibrate the detector as described in “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 51.
M2A Transmitter Operator’s Manual 59

Replacing the Hydrophobic Membrane (Non-XP CO, H2S, and O2 Detectors)

1. Turn off or disconnect power to the M2A.
2. Unscrew the detector housing cap from the detector housing body.
3. Gently pry up the edge of the white hydrophobic membrane with a small flat blade screwdriver or a similar tool.
4. Peel off the hydrophobic membrane. It may be necessary to clean off the detector housing cap face to remove any residue left from the adhesive backed membrane.
5. Install the new membrane in the recess on the face of the detector housing cap.
6. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
7. Turn on or reconnect power to the M2A.

Replacing the Hydrophobic Membrane (CT-7 Series Toxic Detector)

1. Turn off or disconnect power to the M2A.
2. Unscrew the detector housing cap from the detector housing body.
3. Unplug the plug-in sensor from the four-socket pattern in the detector housing body.
4. Gently pry up the edge of the white hydrophobic membrane from the sensor face with a small flat blade screwdriver or a similar tool.
5. Peel off the hydrophobic membrane. It may be necessary to clean off the sensor face to remove any residue left from the adhesive backed membrane.
6. Install the new membrane on the face of the plug-in sensor.
7. Plug the sensor back into the four-socket pattern in the detector housing body.
8. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
9. Turn on or reconnect power to the M2A.
60 M2A Transmitter Operator’s Manual

Chapter 8: RS-485 Modbus Output

Overview

This chapter describes the M2A’s RS-485 Modbus output and how to configure the M2A to make use of it. It also discusses how to wire the M2A into a Modbus system.
The M2A provides an RS-485 serial communications interface. It is a Modbus Slave Device, supporting 2-wire RS-485 Modbus RTU serial communications.

Wiring the M2A in a Modbus System

The M2A is a 2-wire Modbus RTU device. When wiring the M2A into a Modbus system, adhere to standard Modbus wiring practices per the Modbus Over Serial Line Specification and Implementation Guide V1.0 or later. This document can be found online at www.modbus.org/ specs.php.
The Modbus protocol supports a maximum of 247 unique slave addresses (1-247). The M2A’s line driver provides for up to 128 M2As to be connected together without the need for a repeater. Figure 34 & Figure 35 below illustrate typical M2A wiring configurations. If more than 128 M2As need to be connected together, RS-485 repeater(s) should be used such that no more than 128 M2As reside on any given network segment.
CAUTION: The network segment can only support 128 units if RKI M2As are being used. The
use of any other instruments will lower the number of units that may reside on any given network segment.
The MODBUS terminal strip is located on the terminal PCB (see Figure 9). The following signals are available at the Modbus terminal strip:
Table 15: Modbus Terminal Strip Signals
Modbus Terminal Label (RS-485
Name)
AD0
BD1
CCommon
Modbus Signal Name

Ground Jumper

The M2A includes a 2-pin ground header (see Figure 10) that is used when the M2A is used in a Modbus system. Every M2A is supplied with a ground jumper (a jumper block) installed onto this header. If the M2A isn’t wired into a Modbus system or if it is wired into one as shown in Figure 33, the ground jumper should be installed. If the M2A is wired in a Modbus system as shown in Figure 32, the ground jumper should be removed.
M2A Transmitter Operator’s Manual 61

Recommended Modbus Wiring

+
-
10-30 VDC
Remove Ground Jumper
Power Supply
Common
D1 D0
Input
Terminals
Modbus Controller
Remove
Ground
Jumper
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
See
Detector
Wiring
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
To
Additional
M2As
See
Detector
Wiring
+
-
10-30 VDC
Power Supply
Input
Terminals
Modbus Controller
D1
Common
D0
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
A B C
RS 485 PW R/SIG
R W G B
LEL
- S +
A B C
RS 485 PW R/SIG
R W G B
LEL
See
Detector
Wiring
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
To
Additional
M2As
See
Detector
Wiring
The recommended Modbus wiring for the M2A is illustrated in Figure 32 below. In this configuration, 5 wires are used for wiring the M2A into a Modbus system. Figure 32 also illustrates typical alarm device wiring. The ground jumper must be removed in this wiring configuration.
Figure 32: Recommended Modbus Wiring
62 M2A Transmitter Operator’s Manual
Adding an M2A to an Existing M2 Modbus Installation with Recommended Wiring
An M2A can be wired into an existing M2 Modbus system. The shape of the terminal PCB will be different but the wiring connections are identical and the ground jumper on the M2A must still be removed.

Alternate Modbus Wiring For Existing Installations

Although the wiring shown in Figure 32 is recommended, it is possible to wire the M2A into a Modbus system with only 4 wires in situations where a pre-existing system is being replaced and wiring is already in place. This should only be done if wiring for a system that is being replaced is already installed and it is not practical to run another wire. See Figure 33 below for this wiring configuration. In this configuration, the wire between the “C” terminal on the M2A and the Common terminal on the Modbus controller is omitted. Instead, the Common terminal on the
Modbus controller is connected to the “-” of the 24 VDC power supply. The ground jumper must be installed in this wiring configuration.
Figure 33: Alternate Modbus Wiring
Adding an M2A to an Existing M2 Modbus Installation with Alternate Wiring
Modbus
Master
Up to 128 M2Ascan be connected withouta repeater
RS-485 RS-485
ID = 126ID = 127
M2A
Termination
Jumper
Not Installed
M2A
Termination
Jumper
Not Installed
ID = 128
M2A
Termination
Jumper
Installed
ID = 125
ID = 1
M2A
Termination
Jumper
Not Installed
M2A
Termination
Jumper
Not Installed
ID = 128
RS-485
M2A
Termination
Jumper
Installed
Modbus
Master
Up to 128 M2As can be connected without a repeater
M2A
Termination
Jumper
Not Installed
M2A
Termination
Jumper
Not Installed
M2A
Termination
Jumper
Not Installed
M2A
Termination
Jumper
Not Installed
ID = 2 ID = 3
RS-485
ID = 1
M2A
Termination
Jumper
Installed
ID = 127
ID = 4
An M2A can be wired into an existing M2 Modbus system. The shape of the terminal PCB will be different but the wiring connections are identical and the ground jumper on the M2A must still remain installed.

Termination Jumper

The M2A includes a 2-pin termination header (see Figure 10) that is used when the M2A is used in a Modbus system. Every M2A is supplied with a termination jumper (a jumper block) installed onto this header. If the M2A is not used in a Modbus system, this jumper has no function. When the M2A is installed in a Modbus system, this jumper must be installed in an M2A that is at the end of a Modbus line. Any M2A in a Modbus system that is not at the end of a line must have the termination jumper removed (see Figure 34 & Figure 35 below).
Figure 34: Multiple M2As in a Daisy Chain Configuration
M2A Transmitter Operator’s Manual 63
Figure 35: Multiple M2As in a Two Branch Configuration

Using the M2A in a 4-wire Modbus System

Set Up Modbus?
Although the M2A is a 2-wire Modbus RTU device, it can be used with a 4-wire Modbus controller if the system wiring is modified as follows:
Connect the controller’s TxD0 and RxD0 wires together and use this connection as the 2-wire Modbus D0 signal.
Connect the controller’s TxD1 and RxD1 wires together and use this connection as the 2-wire Modbus D1 signal.
Connect these D0 and D1 signals and the common wire from the controller to the 2-wire Modbus Network of M2As.

Modbus Mode

Modbus Mode allows you to configure various parameters relating to the Modbus setup of the M2A. Modbus Mode includes a 5-minute time-out feature. If you do not press a control button for 5 minutes, the M2A automatically returns to normal operation.
NOTE: If the M2A returns to normal operation because of a time-out, it enters a warm-up
period just as it does when it is first turned on.
1. While in normal operation, simultaneously press and hold the DOWN/NO and ENTER buttons for 5 seconds to enter Modbus Mode. Release the buttons when the following screen appears.
2. To exit Modbus Mode, press and release the DOWN/NO button. The M2A will return to normal operation.
To continue in Modbus Mode, press and release the UP/YES button. The current ENABLED/ DISABLED setting will appear.
3. If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter setting is OK, press the ENTER button to proceed to the next parameter.
Table 16 lists the M2A Modbus Mode parameters you can set. Table 16 also lists the factory set value for each parameter.
64 M2A Transmitter Operator’s Manual
Table 16: Configuration Parameters
Modbus Mode Parameter Available Settings & Description
Enabled/Disabled Can be set to ENABLED or DISABLED (factory setting).
Enables or disables the Modbus output.
Slave ID The Slave ID can be set to values from 1 (factory setting) to 247.
The M2A will only receive messages from the Master which are addressed to this Slave ID (except for broadcast messages which are received by all slaves).
Baud Rate Can be set to 1200, 2400, 4800, 9600 (factory setting), 14440, & 19200.
Parity EVEN (factory setting), ODD, NONE
Response Delay The response delay can be set from 0 (factory setting) to 20 mS.
This is an optional additional delay inserted by the M2A prior to returning a response message to the Master. It may be helpful in some installations where the Master’s preparedness to receive responses might be delayed.
4. When you have scrolled through all the adjustable parameters, SAVE IT? YES/NO appears on the display.
5. If you do not wish to save the adjustments, press and release the DOWN/NO button. The DO
OVER? YES/NO message will display. Press and release the DOWN/NO button. The ABORT? YES/NO message will display. Press the UP/YES button to return to normal
operation.
If you wish to change some of the adjustments made, press and release the DOWN/NO button. The DO OVER? YES/NO message will display. Press and release the UP/YES button. The Re-do MB Setup message will display and the M2A will return to the first adjustable parameter. Go back to step 3 and continue.
If you want to save the adjustments made, press and release the UP/YES button. Modbus
Saved is indicated on the display for a few seconds and the M2A returns to normal operation.
NOTE: All M2As on a Modbus network must be configured with the same baud rate and parity,
and each M2A must have a unique slave ID. For an M2A to participate on a network, the ENABLED/DISABLED setting must be set to ENABLE.
M2A Transmitter Operator’s Manual 65

Supported Modbus Functions

The M2A supports Function Code 03: Read Holding Registers and Function Code 16: Write Registers.

Function Code 03: Read Holding Registers

There are a total of 40 registers available to be read via this Function Code.
Registers 1 - 2
The format (bit & field assignments) for registers 1 - 2 is shown below in Table 17 - Table 18.
Table 17: Register 1, Numerical Reading with Sign & Decimal Point Bit & Field Assignments
Bit(s) Value & Field
[15] Sign of Reading
0 = Positive 1 = Negative
[14:13] Format of Reading
0 = XXX (no decimal point) 1 = XX.X 2 = X.XX 3 = .XXX
[12:0] Numerical value of display reading (8000 max)
Table 18: Register 2, Operating State, Alarms & Relays Bit & Field Assignments
Bit(s) Value
[15:12] Operating State
0 = Undefined 1 = Initialization 2 = Normal 3 = Warm-up 4 = Fail 5 = Calibration 6 = Post-Calibration 7 = Low Power 8 = Setup
[11] Gas Type Change
0=Not Changed 1=Changed
[10] Configuration Change
0=Not Changed 1=Changed
[9] Calibration Activity Flag
0=No Calibration Activity 1=Calibration Activity Has Occurred
[8] Fail Status
0 = Fail Not Asserted 1 = Fail Asserted
66 M2A Transmitter Operator’s Manual
Table 18: Register 2, Operating State, Alarms & Relays Bit & Field Assignments (Continued)
Bit(s) Value
[7:6] Alarm 2 Status
0 = No Alarm 1 = Unacknowledged Alarm 2 = Acknowledged Alarm 3 = Unused Code
[5:4] Alarm 1 Status
0 = No Alarm 1 = Unacknowledged Alarm 2 = Acknowledged Alarm 3 = Unused Code
[3] Overscale Status
0 = Not Overscale 1 = Overscale
[2] Fail Relay Status
0 = Not Energized 1 = Energized
[1] Alarm 2 Relay Status
0 = Not Energized 1 = Energized
[0] Alarm 1 Relay Status
0 = Not Energized 1 = Energized
Register 3
Register 3 is the supply voltage (0.1 volt per count).
Registers 4 - 11
The 16 bytes in these registers contain the same ASCII text information that is displayed on the M2A’s OLED (not NULL Terminated).
Register 12
Register 12 is the range (full scale readout).
Register 13
Register 13 is the alarm 1 set point. The decimal point location is the same as specified in Register 1.
Register 14
Register 14 is the alarm 2 set point. The decimal point location is the same as specified in Register 1.
Registers 15 - 19
Registers 15 - 19 are the Gas Name ASCII String (NULL Terminated).
Registers 20 - 22
Registers 20 - 22 are the Gas Units String (NULL Terminated).
M2A Transmitter Operator’s Manual 67
Register 23
Register 23 is the alarm 1 trigger. 0=Decreasing 1=Increasing
Register 24
Register 24 is the alarm 1 normal relay state. 0=Normally De-Energized 1=Normally Energized
Register 25
Register 25 is the alarm 1 relay reset. 0=Latching 1=Self-Resetting
Register 26
Register 26 is the alarm 1 ON delay. Integer values are in seconds.
Register 27
Register 27 is the alarm 1 OFF delay. Integer values are in seconds.
Register 28
Register 28 is the alarm 2 trigger. 0=Decreasing 1=Increasing
Register 29
Register 29 is the alarm 2 normal relay state. 0=Normally De-Energized 1=Normally Energized
Register 30
Register 30 is the alarm 2 relay reset. 0=Latching 1=Self-Resetting
Register 31
Register 31 is the alarm 2 ON delay. Integer values are in seconds.
Register 32
Register 32 is the alarm 2 OFF delay. Integer values are in seconds.
Register 33
Register 33 is zero suppression. Toxics and LEL: 0%-6% of full scale Oxygen: 0%-0.7% Oxygen The decimal point location is the same as specified in Register 1.
68 M2A Transmitter Operator’s Manual
Register 34
Register 34 is noise filter. Integer values in seconds. 0-60 in 5 second increments.
Register 35
Register 35 is calibration timeout in minutes (5, 10, 15, 20, 25, or 30).
Register 36
Register 36 is calibration month (upper byte) and day (lower byte). This is zero based (e.g. January=0 and the 1st=0).
Register 37
Register 37 is the calibration year.
Register 38
Register 38 is the calibration outcome. Upper=Gas Operation Lower=Air Operation
Calibration Outcome Codes:
1=Sensor Missing 2=Sensor Too Weak 3=Signal Too Strong 4=Not Enough Span 5=Negative Reading 6=Span Gas Value Is Not OK 7=Calculated Zero Is Too Low 8=Calculated Zero Is Too High 9=Zero Failure 10=Span Failure 11=Calibration Is Overscale 12=LEL Sensor Offset Is Too High 13=LEL Null Operation Failed 14=Operation Not Performed
Register 39
Register 39 is the auto-zero disable. 0=AutoZero Enabled 1=AutoZero Disabled
Register 40
Register 40 is the remote configuration register access level 0=None 1=Alarm Reset
M2A Transmitter Operator’s Manual 69

Function Code 16: Write Registers

There are 21 registers in Function Code 16 that can be used to write to the M2A. This manual only describes Register 16 because it can be used to reset an alarm condition. For a complete description of Function Code 16, request Appendix C from RKI Instruments, Inc.
Register 16
Register 16 performs the same function as the ENTER button on the M2A. It is useful for remotely silencing alarms. With the M2A in an alarm condition, bit [0] can be used to silence it.
Table 19: Register 16, Button Function
NOTE: Register 40 in Function Code 03, the remote configuration register, must be set to 1 in
order for Register 16 to work.
Bit Value & Field
[0] ENTER (used for “Alarm Reset”)
0=Not Pressed 1=Pressed

Parts List

Table 20 lists replacement parts and accessories for the M2A Transmitter.
Table 20: Parts List
Part Number Description
06-1248RK Calibration kit sample tubing (3/16 in. x 5/16 in.; specify length when ordering)
07-0033RK Cap gasket, for CSA type CO and H
07-0039RK Cap gasket, for non-XP CO/H
07-0125RK Cap gasket, for ESM-01 detector
07-0203RK Rubber retaining boot for CO detectors’ charcoal filter and for 65-2428 H
14-2101RK Spacer for 65-2428 H
18-0431-01-01 Junction box with foam pad, label, and conduit plug
33-0157RK Hydrophobic membrane for non-XP CO, H2S, O2, and CT-7 Series detectors
33-7101RK Charcoal filter disk
45-6132RK Termination and group jumper
61-0140RK LEL detector, catalytic type, UL version
S detector
2
S detector
2
S and CT-7 Series detectors
2
S detector’s spacer
2
61-0140RK-05 LEL detector, catalytic type, CSA version
61-0140RKA CH
61-0140RKA-05 CH
70 M2A Transmitter Operator’s Manual
ppm detector, catalytic type, UL version
4
ppm detector, catalytic type, CSA version
4
Table 20: Parts List
Part Number Description
61-0190RK-CH4 CH
0-100% LEL/0-5.00% volume detector, IR type, non-replaceable sensor
4
61-0190RK-HC HC LEL detector, IR type, non-replaceable sensor, propane calibration
61-0191RK-02 CO2 detector, IR type, non-replaceable sensor, 0 - 5,000 ppm
61-0191RK-03 CO
61-0191RK-05 CO
61-0191RK-10 CO
61-0192RK-CH4 CH
61-0197RK-CH4 CH
detector, IR type, non-replaceable sensor, 0 - 5% volume
2
detector, IR type, non-replaceable sensor, 0 - 50.0% volume
2
detector, IR type, non-replaceable sensor, 0 - 100% volume
2
detector, IR type, non-replaceable sensor, 0 - 100% volume
4
detector, IR type, replaceable sensor, 0 - 100% LEL
4
61-0197RK-HC HC detector, IR type, replaceable sensor, 0 - 100% LEL
61-0198RK-02 CO
61-0198RK-03 CO
61-0198RK-05 CO
61-0198RK-10 CO
61-0199RK-CH4 CH
61-5039RK-CH4 Plug-in IR CH
detector, IR type, replaceable sensor, 0 - 5000 ppm
2
detector, IR type, replaceable sensor, 0 - 5% volume
2
detector, IR type, replaceable sensor, 0 - 50.0% volume
2
detector, IR type, replaceable sensor, 0 - 100% volume
2
detector, IR type, replaceable sensor, 0 - 100% volume
4
sensor, LEL range, for 61-0197RK-CH4 detector
4
61-5039RK-HC Plug-in IR HC sensor, LEL range, for 61-0197RK-HC detector
61-5040RK-02 Plug-in IR CO2 sensor, 0 - 5000 ppm range (for 61-0198RK-02 detector)
61-5040RK-03 Plug-in IR CO
61-5040RK-05 Plug-in IR CO
61-5040RK-10 Plug-in IR CO
61-5041RK-CH4 Plug-in IR CH
sensor, 0 - 5% volume range (for 61-0198RK-03 detector)
2
sensor, 0 - 10% volume range (for 61-0198RK-05 detector)
2
sensor, 0 - 100% volume range (for 61-0198RK-10 detector)
2
sensor, %vol range (for 61-0199RK-CH4 detector)
4
65-1025RK Plug-in oxygen sensor (for 65-2514RK and 65-2494RK detectors)
65-2039RK Plug-in H
65-2051-01 CT-7 Series plug-in sensor, 0 - 3.00 ppm and 0 - 10.0 ppm Cl
S sensor (for 65-2428-01 detector)
2
, with membrane (for 65-2302-CL2 and
2
65-2302-CL2-10 detectors)
65-2052-01 CT-7 Series plug-in sensor, 0 - 1.00 ppm ClO2, with membrane (for 65-2302-CLO2 detector)
65-2300RK-ASH3 ESM-01 detector, 0 - 1.50 ppm arsine (includes sensor)
65-2300RK-CL-10 ESM-01 detector, 0 - 10.0 ppm chlorine (includes sensor)
65-2300RK-CL2 ESM-01 detector, 0 - 3.00 ppm chlorine (includes sensor)
65-2300RK-HCN ESM-01 detector, 0 - 15.0 ppm hydrogen cyanide (includes sensor)
M2A Transmitter Operator’s Manual 71
Table 20: Parts List
Part Number Description
65-2300RK-NH3 ESM-01 detector, 0 - 75.0 ppm ammonia (includes sensor)
65-2300RK-PH3 ESM-01 detector 0 - 1.00 ppm phosphine (includes sensor)
65-2300RK-SO2 ESM-01 detector, 0 - 6.00 ppm sulphur dioxide (includes sensor)
65-2302-CL2 CT-7 Series detector, 0 - 3.00 ppm chlorine (includes sensor)
65-2302-CL2-10 CT-7 Series detector, 0 - 10.0 ppm chlorine (includes sensor)
65-2302-CLO2 CT-7 Series detector, 0 - 1.00 ppm chlorine dioxide (includes sensor)
65-2428 H
65-2428-01 H
S detector assembly, CSA version (includes plug-in sensor)
2
S detector assembly, CSA version, for frequent-H2S-exposure applications (includes plug-in sensor)
2
65-2438 CO detector assembly, 0 - 300 ppm, CSA version (includes plug-in sensor)
65-2438-01 CO detector assembly, 0 - 100 ppm and 0 - 200 ppm, CSA version (includes plug-in sensor)
65-2494RK Oxygen detector assembly, non-explosion proof, replaceable sensor type (includes plug-in sensor)
65-2495RK H
S detector assembly, 0 - 100 ppm, non-explosion proof, replaceable sensor type (includes plug-in
2
sensor)
65-2496RK CO detector assembly, 0 - 300 ppm, non-explosion proof, replaceable sensor type (includes plug-in
sensor)
65-2510RK Oxygen detector, partial pressure type
65-2511RK Oxygen detector, capillary type
65-2514RK Oxygen detector, CSA version, capillary type (includes plug-in sensor) 71-0305RK M2A Transmitter Operator’s Manual (this document)
81-0002RK-01 Calibration cylinder, 50% LEL hydrogen in air, 34 liter steel
81-0002RK-03 Calibration cylinder, 50% LEL hydrogen in air, 103 liter
81-0004RK-01 Calibration cylinder, 50% LEL propane in air, 34 liter steel
81-0004RK-03 Calibration cylinder, 50% LEL propane in air, 103 liter
81-0007RK-01 Calibration cylinder, 15% LEL hexane in air, 34 liter steel
81-0010RK-01 Calibration cylinder, 10% LEL (5000 ppm) methane in air, 34 liter steel
81-0010RK-03 Calibration cylinder, 10% LEL (5000 ppm) methane in air, 103 liter
81-0012RK-01 Calibration cylinder, 50% LEL methane in air, 34 liter steel
81-0012RK-03 Calibration cylinder, 50% LEL methane in air, 103 liter
81-0013RK-01 Calibration cylinder, 50% volume methane in nitrogen, 34 liter steel
81-0013RK-05 Calibration cylinder, 50% volume methane in nitrogen, 58 liter
81-0064RK-01 Calibration cylinder, CO, 50 ppm in air, 34 liter steel
81-0064RK-03 Calibration cylinder, CO, 50 ppm in air, 103 liter
72 M2A Transmitter Operator’s Manual
Table 20: Parts List
Part Number Description
81-0070RK-01 Calibration cylinder, CO
81-0070RK-03 Calibration cylinder, CO
81-0072RK-01 Calibration cylinder, CO
81-0072RK-03 Calibration cylinder, CO
81-0073RK-01 Calibration cylinder, CO
81-0073RK-03 Calibration cylinder, CO
, 2,000 ppm in nitrogen, 34 liter steel
2
, 2,000 ppm in nitrogen, 103 liter
2
, 2.5% in nitrogen, 34 liter steel
2
, 2.5% in nitrogen, 103 liter
2
, 15% in nitrogen, 34 liter steel
2
, 15% in nitrogen, 103 liter
2
81-0076RK-01 Zero air calibration cylinder, 34 liter steel
81-0076RK-03 Zero air calibration cylinder, 103 liter
81-0078RK-01 Calibration cylinder, 100% nitrogen, 34 liter steel (used for setting oxygen zero)
81-0078RK-03 Calibration cylinder, 100% nitrogen, 103 liter (used for setting oxygen zero)
81-0151RK-02 Calibration cylinder, H
81-0151RK-04 Calibration cylinder, H
81-0170RK-02 Calibration cylinder, 5 ppm SO
81-0170RK-04 Calibration cylinder, 5 ppm SO
81-0175RK-02 Calibration cylinder, 10 ppm NH
S, 25 ppm in nitrogen, 58 liter
2
S, 25 ppm in nitrogen, 34 liter aluminum
2
in nitrogen, 58 liter
2
in nitrogen, 34 liter aluminum
2
in nitrogen, 58 liter
3
81-0175RK-04 Calibration cylinder, 10 ppm NH
81-0185RK-02 Calibration cylinder, 0.5 ppm PH
81-0185RK-04 Calibration cylinder, 0.5 ppm PH
81-0190RK-02 Calibration cylinder, 5 ppm Cl
81-0190RK-04 Calibration cylinder, 5 ppm Cl
81-0192RK-02 Calibration cylinder, 2 ppm Cl
81-0192RK-04 Calibration cylinder, 2 ppm Cl
in nitrogen, 34 liter aluminum
3
in nitrogen, 58 liter
3
in nitrogen, 34 liter aluminum
3
in nitrogen, 58 liter
2
in nitrogen, 34 liter aluminum
2
in nitrogen, 58 liter
2
in nitrogen, 34 liter aluminum
2
81-0196RK-02 Calibration cylinder, 10 ppm HCN in nitrogen, 58 liter
81-0196RK-04 Calibration cylinder, 10 ppm HCN in nitrogen, 34 liter aluminum
81-1050RK Regulator, with gauge and knob, 0.5 liter/minute continuous flow, for 17 liter and 34 liter steel
calibration cylinders (cylinders with external threads)
81-1051RK Regulator, with gauge and knob, 0.5 liter/minute continuous flow, for 34 liter aluminum/58 liter/103
liter calibration cylinders (cylinders with internal threads)
81-1051RK-25 Regulator, fixed flow, w/gauge & knob, 0.25 LPM, for 34 liter aluminum/58 liter/103 liter cylinders
(Used to calibrate NH
ESM-01 M2A only)
3
81-1103RK Calibration cup, for 3/4 NPT IR detectors
M2A Transmitter Operator’s Manual 73
Table 20: Parts List
Part Number Description
81-1117RK Calibration cup, for catalytic LEL/ppm, CO, H
S, replaceable-sensor-type IR, CT-7 Series, and
2
oxygen detectors
81-1138RK Calibration cup, general (SO2), for all ESM-01 except Cl2 and NH3, use 0.5 LPM flow
81-1138RK-CL2 Calibration cup, for ESM-01 Cl
81-1138RK-NH3 Calibration cup, for ESM-01 NH
, use 0.5 LPM flow
2
, use 0.25 LPM flow
3
81-1190RK Splashguard, for ESM-01 detector cap
ES-1531-CO CO sensor, plug-in (for 65-2438, 65-2438-01, and 65-2496RK detectors)
ES-1537-H2S H
S sensor, plug-in (for 65-2428 and 65-2495RK detectors)
2
ESM-015-NO ESM-01 plug-in sensor, 0 - 100 ppm nitric oxide
ESM-01DH-ASH3 ESM-01 plug-in sensor, 0 - 1.50 ppm arsine
ESM-01DH-D-HCN ESM-01 plug-in sensor, 0 - 15.0 ppm hydrogen cyanide
ESM-01DH-D-SO2 ESM-01 plug-in sensor, 0 - 6.00 ppm sulphur dioxide
ESM-0DH-PH3 ESM-01 plug-in sensor, 0 - 1.00 ppm phosphine
ESM-01R-D-NH3 ESM-01 plug-in sensor, 0 - 75.0 ppm ammonia
ESM-K01-D-CL2 ESM-01 plug-in sensor, 0 - 3.00 ppm chlorine
ESM-K01D-CL2-10 ESM-01 plug-in sensor, 0 - 10.0 ppm chlorine
NC-6205-01 Hydrogen specific LEL detector, catalytic type, UL version
NC-6205-05 Hydrogen specific LEL detector, catalytic type, CSA version
74 M2A Transmitter Operator’s Manual

Appendix A: Control Button Quick Reference Guide

The M2A’s control buttons allow access to operational modes, resetting of alarms, and display of the Information Screen. Table 21 shows which button combinations perform these functions and which parameters are available for adjustment while in the operational modes. While in these modes, display prompts showing a “?” require you to respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button. To change parameter settings, use the UP/YES and DOWN/NO buttons to get the desired setting, then press and release the ENTER button to accept the setting.
Table 21: Control Button Quick Reference Chart
Mode Entered or
Function Performed
Alarm Reset Press and Release
Show Information Screen Press and Hold
Calibration Mode Press and Hold UP/
Configuration Mode Press and Hold UP/
Gas Type Mode Press and Hold UP/
Button
Combination
ENTER
DOWN/NO
YES for 5 seconds
YES & ENTER for 5 seconds
YES and DOWN/ NO for 5 seconds
Adjustable or Viewable Parameters
n/a
Input Voltage
M2A Version/Revision Information
•Fresh Air Adjust
Span (or Zero for Oxygen) Adjust with Calibration Gas
•Alarm Levels
Increasing or Decreasing Alarms
Normally Energized or Normally De-energized Relays
Latching or Self-resetting Alarms
•Alarm On Delay
Alarm Off Delay (if self-resetting)
Zero Suppression
Noise Filter
Calibration Time-out Period
Gas Type
Modbus Mode Press and Hold
DOWN/NO and ENTER for 5 seconds
M2A Transmitter Operator’s Manual 75
Enable or Disable Modbus
•Slave ID
•Baud Rate
•Parity
Response Delay

Appendix B: PLC and DCS Device Wiring

The M2A can be wired to a PLC or DCS device if desired.
1. Guide multi conductor shielded cable or cables or wires in conduit through the top conduit port of the junction box. The number of cables or wires needed will depend on whether any relays are used and whether the Modbus output is used. Use the following recommendations to determine how to wire the M2A:
If Modbus connections will not be used and only the PWR/SIG terminal strip connections
will be used, use four conductor shielded cable or four wires in conduit for connections to the power/signal terminal strip.
Table 22: Wire Size for PWR/SIG Connections
Max Distance to Controller
w/18 Gauge Wire
2,500 ft. 5,000 ft. 8,000 ft.
If the PWR/SIG terminal strip connections and one or more relays are used, route the
connections to the M2A in conduit. Use shielded cable in the conduit for the PWR/SIG connections and unshielded cable or individual wires for the relay connections. Make sure any wire or cable used for relay wiring is appropriately rated for the power that it will carry.
NOTE: If shielded cable is used for the PWR/SIG connections, leave the cable shield’s drain
wire insulated and disconnected at the M2A. You will connect the opposite end of the cable’s drain wire at the controller or device.
If the M2A will be wired into a Modbus network, see “Chapter 8: RS-485 Modbus
Output” on page 61.
Max Distance to Controller w/
16 Gauge Wire
Max Distance to Controller
w/14 Gauge Wire
76 M2A Transmitter Operator’s Manual
See Figure 36 below for field wiring connections to the M2A.
Alarm 1 Alarm Device
Fail Alarm Device
Alarm Device
Power
TypicalAlarm WiringShown
4-20mA In (S)
(24 VD C) -
24 V DC +
24 VDC
Power Supply
24 VDC -
Alarm 2 Alarm Device
See
Modbus
Wiring
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
See
Detector
Wiring
Controller
(PLC, DCS)
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Figure 36: PLC and DCS Device Wiring
M2A Transmitter Operator’s Manual 77

Appendix C: Function Code 16 Registers

The M2A supports Function Code 16 that allows writing to the M2A. There are 21 registers in this Function Code.

Register 1

Register 1 is the alarm 1 set point. Value must be an integer. Any decimals are omitted.

Register 2

Register 2 is the alarm 1 trigger. 0=Decreasing 1=Increasing

Register 3

Register 3 is the alarm 1 relay state. 0=Normally De-Energized 1=Normally Energized

Register 4

Register 4 is the alarm 1 relay reset. 0=Latching 1=Self-Resetting

Register 5

Register 5 is the alarm 1 ON delay. Integer value is in seconds.

Register 6

Register 6 is the alarm 1 OFF delay. Integer value is in seconds.

Register 7

Register 7 is the alarm 2 set point. Value must be an integer. Any decimals are omitted. The decimal point location is the same as specified in Register 1 of Function Code 03.

Register 8

Register 8 is the alarm 2 trigger. 0=Decreasing 1=Increasing

Register 9

Register 9 is the alarm 2 relay state. 0=Normally De-Energized 1=Normally Energized
78 M2A Transmitter Operator’s Manual

Register 10

Register 10 is the alarm 2 relay reset. 0=Latching 1=Self-Resetting

Register 11

Register 11 is the alarm 2 ON delay. Integer value is in seconds.

Register 12

Register 12 is the alarm 2 OFF delay. Integer value is in seconds.

Register 13

Register 13 is the zero suppression. Toxics and LEL: 0%-6% Full Scale Oxygen: 0%-0.7% Oxygen Value must be an integer. Any decimals are omitted.

Register 14

Register 14 is the noise filter. Integer value is in seconds (0-60 in 5 second increments).

Register 15

Register 15 is the calibration timeout. Value is in minutes (5, 10, 15, 20, 25, or 30).

Register 16

Register 16 is explained in “Supported Modbus Functions” in the M2A manual.

Register 17

Register 17 is the calibration month (MSB) and day (LSB). This is zero-based (e.g. January=0 and the 1st=0).

Register 18

Register 18 is the calibration year.

Register 19

Register 19 is for clearing change flags. When parameters are changed at the M2A, a flag is raised at the controller. This register can be used to clear those flags.
Table 23: Register 19, Clear Change Flags
Bit(s) Value & Field
[15:2] Unused
[1] Write “1” to this bit to clear “Gas Type Changed” flag
[0] Write “1” to this bit to clear “Configuration Changed” flag
M2A Transmitter Operator’s Manual 79

Register 20

Register 20 is the AutoZero Disable. 0=AutoZero Enabled 1=AutoZero Disabled

Register 21

Register 21 is the Remote Configuration Register. 0=No Remote Configuration 1=Reset Alarm
80 M2A Transmitter Operator’s Manual