Read and understand this instruction manual before
operating instrument. Improper use of the gas monitor
could result in bodily harm or death.
Periodic calibration and maintenance of the gas
monitor is essential for proper operation and correct
readings. Please calibrate and maintain this instrument
regularly! Frequency of calibration depends upon the
type of use you have and the sensor types. Typical
calibration frequencies for most applications are
between 3 and 6 months, but can be required more
often or less often based on your usage.
M2A Transmitter Operator’ s Manual
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment s old by us t o be free from def ects in material s,
workmanship, and performance for a period of one year from date of shipment from RKI
Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our
option, free of charge. This warrant y does not apply to those items which by their nature are subject
to deterioration or consumption in normal service, and which must b e cleaned, repaired, o r replaced
on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair
procedures not in accordance with the operator’s manual. This warranty indicates the full extent of
our liability, and we are not responsible for removal or replacement costs, local repair costs,
transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED T O, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and
representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas
monitor, and our warranty is limited to the replacement of parts or our complete goods.
(Appendix C available from RKI Instruments, Inc. Not included in manual as normally provided
with M2A detector head.)
M2A Transmitter Operator’s Manual
Chapter 1: Introduction
Overview
This chapter briefly describes the M2A Transmitter. This chapter also describes the M2A
Transmitter Operator’s Manual (this document). Table 1 at the end of this chapter lists the
specifications for the M2A.
About the M2A Transmi tter
The M2A transmitter is a fixed mount, continuous-monitoring detector head. All user adjustable
parameters may be accessed using pu sh bu tto n switc hes . In addition, calibration may be performed
non-intrusively by use of a magnetic wand accessory which activates magnetic switches through a
window at the front face of the detector head.
The M2A displays the current gas reading on an OLED display which is visible through the
window in the cover (front face) and provides a 4 - 20 mA signal which indicates the target gas
reading for use by a gas monitoring controller, recording device, or programmable controller. The
M2A also provides an RS-485 Modbus output. Three sets of relay contacts, two controlled by the
gas alarms and one by the f ail alarm, rated at 115 VAC 5 amps are available for con trolling devices
such as lights or horns or for controlling higher rated relays.
Four operating modes allow you to display and change setup and calibration settings and change
the gas type. They are Calibration Mode, Configuration Mode, Gas Type Mode, and Modbus
Mode.
About this Manual
The M2A Transmitter Operator’s Manual uses the following conventions for notes, cautions, and
warnings.
NOTE:Describes additional or critical information.
CAUTION:Describes potential damage to equipment.
WARNING:Describes potential danger that can result in injury or death.
1M2A Transmitter Operator’ s Manual
Specifications
Table 1 lists specif ications for the M2A.
Table 1: M2A Specifications
T ar get Gas/Detection RangeCombustible Gas/Carbon Dioxide (CO
•Catalytic LEL: 0 - 10 0% L EL, 1% LEL i ncremen ts (CH4 calibration unless otherwise
specified. H
Construction (ho usi n g)Explosion-pro of Ju n cti on Bo x, NE MA 4X
Area ClassificationExplosion-proof for Class I, Groups B, C, and D (Combustible, CO
H2S, and CSA Type oxygen)
Sampling MethodDiffusion
, CSA Type CO and
2
M2A Transmitter Operator’s Manual2
Table 1: M2A Specifications
Input Power 10 - 30 VDC
Controls•Three push butto n sw it ch e s
•Three magnetic switches for non-intrusive calibration
Weight4.5 lbs.
Signal Output•4 to 20 mA with impedance maximums listed below
-12 VDC input: 500 ohms impedance max
-24 VDC input: 1000 ohms impedance max
•RS-485 Modbus
Explosion ProofNon-Explosion Proof
LEL-40°C to 75°CN/A
Oxygen-20°C to 45°C-20°C to 50° C
H
S-40°C to 40°C-40°C to 50° C
2
Operating Temperature
CO-5°C to 40°C-20°C to 50°C
Toxic (ESM-01)N/A-10°C to 40°C
Infrared LEL-20°C to 50°C-40°C to 50° C
Infrared CO
2
AccuracyCatalytic Combustible Gas
-20°C to 50°C-40°C to 50° C
:
± 5% of reading or ± 2% LEL (whichever is greater)
IR Combustible Gas/IR CO2:
± 5% of reading or ± 2% of full scale (whichever is greater)
Oxygen
:
± 0.5% O
2
Hydrogen Sulfide:
± 5% of reading or ± 2 ppm H
S (whichever is greater)
2
Carbon Monoxid e :
± 5% of reading or ± 5 ppm CO (wh ichever is greate r)
ESM-01 Toxic Sensors
:
± 10% of reading or ± 5% of ful l scale (whichever is greater)
*
These are the RKI factory settings. You can change the alarm settings in Configuration Mode. See
“Viewing & Changing M2A Paramete rs” on page 31.
WARNING:When using the M2A, you must follow the instructions and warnings in this
manual to assure proper and safe operation of t he M2A and to minimize the risk
of personal injury. Be sure to maintain and periodically calibrate the M2A as
described in this manual.
3M2A Transmitter Operator’ s Manual
Chapter 2: Description
Overview
This chapter describes external and internal components of the M2A Transmitter.
External Description
This section describes the junction box and all external components of the M2A transmitter.
3/4 NPTConduit Hub, P lugged
Junction Box Cover
Mounting Slot (2x)
Magnetic
Wand
Window
3/4 NPT
Conduit Hub
Detector
(Cat alytic LEL Detector Shown)
Figure 1: M2A Ex ternal Components
Junction Box
The M2A’s cast aluminum junction box is dust and weather resistant. The junction box also
protects the M2A and all connections made to it. Use the three 3/4 in. conduit hubs to mount the
detector to the junction box (factory installed in the bottom hub) and connect wiring from an
external device (left h ub). T he top hub i s shipp ed with a factory i nstalle d and sealed co nduit plug to
avoid leaks into the junction box. If necessary, the conduit plug can be removed and the top hub can
be used for wiring.
WARNING:If the top conduit hu b is used for wiring, be sure to seal the threads to ensure
water does not enter the junction box. See “Wiring the M2A Transmitter” on
page 18 for complete wiring instructions.
M2A Transmitter Operator’s Manual4
Use the junction box’s two mounting holes to mount the M2A to a vertical surface at the
monitoring site. The window in the cover on the front of the junction box allows you to view the
OLED display and use the magnetic wand to actuate the magnetic control switches so you can
perform non-intrusive calibration. Removing the cover allows you to access the interior of the
junction box.
Magnetic Wand
The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate the
magnetic control switches on the control PCB while the junction box cover is still installed so that
non-intrusive calibration can be performed.
Gas Detector
The gas detector senses the target gas and is mounted in a 3/4” cond uit hu b on the r ight bo ttom si de
of the M2A. A variety of detectors may be used with the M2A. Table 2 below lists the detectors that
can be used with the M2A.
Table 2: M2A Detectors
Target Gas, RangeDetector(s) Used
Combustible Gas, 0 - 100% LEL•61-0140RK (Catalytic UL version)
•61-0140RK-05 (Catalytic CSA version)
Hydrogen, 0 - 100% LE L•NC-6205-01 (Catalytic UL version, hyd rogen specific)
•NC-6205-05 (Catalytic CSA version, hyd r ogen specific)
CH
, 0 - 100% LEL•61-0190RK-CH 4 ( IR UL)
4
HC, 0 - 100% LEL (cal. to propane)•61-0190RK-HC (IR UL)
CH
, 0 - 100% volume•61-0192RK- C H4 ( IR UL)
4
CO
, 0 - 5,000 ppm •61-0191RK-02 (IR UL)
2
CO2, 0 - 5%•61-0191RK -03 (IR U L)
CO
, 0 - 50%•61-0191RK-05 (IR UL)
2
, 0 - 100%•61-0191RK -10 (IR UL)
CO
2
Oxygen, 0 - 25%•65-2494RK (capil lary)
CO, 0 - 300 ppm•65-2433RK-05 (CSA)
•61-0198RK-02 (IR)
•61-0198RK-03 (IR)
•61-0198RK-05 (IR)
•61-0198RK-10 (IR)
•65-2510RK (partial pressure)
•65-2511 RK (capilla ry )
•65-2514RK (capillary, CSA)
•65-2496RK
H
S, 0 - 100 ppm•65-2423RK-05 (C SA)
2
NH3, 0 - 75 ppm•65-2300RK- NH3
AsH
, 0 - 1.5 ppm•65-2300RK-ASH3
3
5M2A Transmitter Operator’ s Manual
•65-2495RK
Table 2: M2A Detectors
Target Gas, RangeDetector(s) Used
Cl2, 0 - 3 ppm•65-2300RK-CL2
Cl
, 0 - 10 ppm•65-2300RK-CL- 10
2
PH
, 0 - 1 ppm•65-2300RK-PH3
3
SO
, 0 - 6 ppm•65-2300RK-SO2
2
Catalytic LEL Detectors
The catalytic LEL detectors have a 1/2” NPT thread and require a 3/4”NPT x 1/2”NPT reducer to
install in the detector hub. The rest of the detectors have a 3/4” NPT mounting thread and are
installed directly in the detector hub.
IR Detectors
The IR combustible detectors are generally used instead of the catalytic combustible detectors in
applications where there may be catalyst poisons such as silicone present or where oxygen is not
present in the monitoring environment.
Oxygen Detectors
There are four types of oxygen detectors available, a non-explosion proof partial pressure type, a
non-explosion proof capillary type, a non-explosion proof capillary type with a replaceable sensor,
and an explosion proof (CSA) capillary type with a replaceable sensor. The capillary type’s output
is less susceptible to variations in atmospheric pressure than the partial p r essure type. The CSA
type oxygen detector (65-2514RK) and one of the non-explosion proof capillary type detectors
(65-2494RK) have a replaceable plug-in sensor ins ide the detector ho using while the entire oxygen
detector must be removed if replacement is required for the other two.
Detector Housing Body
Cap Gasket
Oxygen Sensor
Detector Housing Cap
Flame Arrestor
Guard
Figure 2: 65-2514RK Oxygen D etector, CSA
M2A Transmitter Operator’s Manual6
Detector Housing Body
e
r
Cap Gasket
Plug-In Oxygen Sensor
Detector Housing Cap
Hydr ophobic Membran
Figure 3: 65-2494RK Oxygen Detector, Non-Explosion Proof, Capillary Type
CO and H
Both types of CO and H
S Detectors
2
S detectors have replaceable plug-in sensors inside the detector housing.
2
The CO detector includes a charcoal filter disk which is held onto the sensor with a rubber retaining
boot. The charcoal filter disk scrubs out interfering gases to which the CO sensor responds such as
H
S and certain hydrocarbons. The 65- 2423RK- 05 explo sion-p roof H2S detector includes a rubber
2
boot and a spacer which help ensur e that the sensor remains plugged into th e detector housing body.
The 65-2495RK non-explosion-proof H
S detector includes an adhesive backed gasket inside the
2
detector housing cap which helps ensure that the sensor remains plugged into the detector housing
body.
Detector Housing Body
Cap Gasket
Plug-In CO Sensor
Rubber Boot W ith
Charcoal F ilter
Detector Housing Cap
Flame Arresto
Guard
Figure 4: 65-2433RK-05 CO Detector, CSA
7M2A Transmitter Operator’ s Manual
Detector HousingBody
Cap Gasket
CO Plug-in Sensor
Charcoal Filter
w/Rubber Boot
Detector
Housing Cap
Hydrophobic
Membrane
Figure 5: 65-2496RK CO Detector Non-Explosion Proof
The constr uction of the various ESM-01 detectors is similar. Each of the ESM-01 detectors has a
different replaceable plug-in sensor that protrudes through the detector housing cap. Although
ESM-01 detectors look the same except for the plug-in detector, the detector housing which
includes the detector socket and conditioning circuitry is specific to a sensor type. For example, a
detector housing from a Cl
detector cannot be used with an NH3 detector. One of the two detector
2
wires is color coded for each specific detector type. The ESM-01 detectors include a removable
splashguard that screws onto the detector housing cap. When it is necessary to apply gas to an
ESM-01 detector, remove the splashguard to allow access to the sensor that protrudes through the
detector housing cap. You can then push the appropriate ESM-01 calibration cup onto the sensor.
Detector Housing Body
(different for each detector type)
Cap Gasket
Sensor
(different for each detector type)
Detector Housing
Cap
Splash Guard
(removable)
Figure 8: ESM-01 Type Detector
9M2A Transmitter Operator’ s Manual
Internal Description
This section describes the internal components of the M2A. The internal components of the M2A
include the terminal PCB which provides for all the wiring connections to the M2A and the control
PCB which displays the gas reading and has the control buttons.
3/4" NPT ConduitOpening, Plugged
Power/Signal
TerminalStrip
Detector Terminal Strip
(Comb/CO2 Version Shown)
Detector Current
Adjust Pot
(Comb/CO2 Type Only,
FactoryAdjust)
3/4" NPT
Conduit
Opening
fo r Wi r e
Entry
Ground Jumper
Programming
Connector
(Factory Use)
ModbusTerminal Strip
TerminationJumper
RibbonCableConnector
Fail LED
Alarm 1 LED
Alarm2 LED
RX LE D
TX LED
Relay
Terminal
Strips
Junctio n Box
TerminalPCB
Control PCB
Magnetic
Switches
ENTER
DOWN/NO
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
UP/YES
NO
RKI INSTRUMENTS
M2AT RANSMITTER
Push Button
Switches
Cont rol
UP/YES
DOWN/NO
ENTER
OLED Display
View WithCover and
Control PCB Removed
View With Cover Removed
Figure 9: M2A Internal Components
Terminal PCB
The terminal PCB is encapsulated in epoxy for protection against moistur e and physi cal damage. It
is mounted into the rear of the junction box with three standoffs and rests on a thin layer of foam. A
banana jack is screwed into each of the standoffs and used for mounting the control PCB. The
terminal PCB converts the electrical output from the detector to a signal wh i ch can be displayed by
the OLED display, a 4 - 20 mA signal (that is proportional to the detection range), and an RS-485
Modbus output signal. The 4 - 20 mA signal may be used by a recording device, gas monitor
controller, or programmable controller. The Modbus output may be used to connect the M2A to a
Modbus network. The terminal PCB also controls three relays, one fail and two gas alarm relays.
Two columns of plug-in style terminal strips ar e used to make all wiring connections to the M2A.
The column on the left consists of the power/signal, detector, and Modbus terminal strips. The
column on the right consists of the relay terminal strips. A 20 position connector at the bottom of
the terminal PCB is used to connect the terminal PCB to the control PCB with a ribbo n cable. A 5
position connector on the left side of the terminal PCB is used by factory or field service pers on nel
M2A Transmitter Operator’s Manual10
to program the M2A. On the combustible and CO2 versions of the M2A, a factory adjust pot just
above the programming connector is used to set the detector current.
Power/Sign al Terminal Strip
The power/signal terminal strip is a three position plug-in style terminal strip located at the top of
the left terminal column. It is used to connect 24 VDC power to the M2A and to connect the 4 - 20
mA output signal to a device.
The signal output, the S terminal, does not have to be connected for the M2A to function. The S
terminal is used if you want to connect the 4 - 20 mA output signal to another device such as a gas
monitoring controller, chart recorder, or programmable controller (PLC).
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip and is the middle ter minal
strip in the left terminal column. For a combustible or CO
M2A, all four terminals are used to
2
connect the detector to the M2A. For a toxic o r oxygen M2A, only two of the terminals are used for
connecting a detector: two terminals are dedicated for connecting a toxic detector and two are
dedicated for connecting an oxygen detector (only one detector can be installed).
Comb/CO2
Detector
Terminal
Strip
Toxic/Oxy
Detector
Terminal
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
C NC NO
C NC NO
C NC NO
FAIL
ALARM 1
ALARM 2
Strip
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
C NC NO
C NC NO
C NC NO
FAIL
ALARM 1
ALARM 2
Figure 10: Detector Terminal Strip
NOTE:The detector is factory-wired to the M2A. See “Wiring the M2A Transmitter ” on
page 18 for all wiring procedures related to M2A.
Modbus Terminal Str ip
The Modbus terminal strip is a three position plug-in style terminal strip and is the bottom terminal
strip in the left terminal column. It allows connection of the M2A into a Modbus network.
11M2A Trans mitte r Op erato r’ s Manu al
Relay Terminal Strips
The right column of terminal strips consists of, from top to bottom, the fail, alarm 1, and alarm 2
relay terminal strips. They are three-position plug-in style terminal strips. The relay terminal strips
are used to connect devices such as lights and horns that are controlled by the relay contacts. The
relay contacts are rated at 115 VAC, 5 amps. The relay contacts may also be used to control higher
rated relays.
Termination Jumper
A two pin header with a termination jumper installed is located below the Modbus terminal strip.
Leave this jumper installed unless directed to remove it for a Modbus installation. See “Chapter 8:
RS-485 Modbus Output” on page 51 for a description of using the M2A in a Modbus system.
Ground Jumper
A two pin header with a jumper installed is located along the left side of the terminal PCB. Leave
this jumper installed unless directed to remove it for a Modbus installation. See “Chapter 8: RS-485
Modbus Output” on page 51 for a description of using the M2A in a Modbus system.
Control PCB
The OLED display and control switches are located o n the control PCB. It is installed on top of the
terminal PCB by lining up its three spacing standoffs with the banana jacks in the terminal PCB
mounting standoffs and pushing it onto the banana jacks. The jacks retain the control PCB.
OLED Display
The OLED display is located at the top of the control PCB. It indicates the current gas reading and
displays messages and parameters in the M2A’s programs.
Control Buttons
The M2A includes three push button switches that allow you to enter the M2A’s operating modes,
navigate through the modes, update settings, and save changes to the settings. The push button
switches are located along the bottom edge of the control PCB (see Figure 9). The UP/YES button
is on the left, the DOWN/NO button is in the middle, and the ENTER button is on the right.
Tabl e 3: M2A Control Button Functions
SwitchFunction
UP (YES)•Saves settings
•Changes the displayed setting
•Enters the Calibration Program
•Enters Gas Type Mode (press with DOWN/NO button)
•Enters Configuration Mode (press with ENTER button)
DOWN (NO)•Cancels setting changes
•Changes the displayed setting
•Enters Gas Type Mode (press with UP/YES button)
•Enters Modbus Mode (press wi th ENTER button)
•Displays the Information Screen
ENTER•Initiates operations
•Enters Configuration Mode (press with UP /YES button)
•Enters Modbus Mode (press with DOWN/NO button)
•Functions as an alarm reset switch
Just above each push button switch is a magnetic switch with the same function as the push button
M2A Transmitter Operator’s Manual12
switch below it. The magnetic switches are for use in non-intrusive calibration. They are actuated
by bringing the magnetic wand close enough to them to actuate them. Although the magnetic
switches have the same functions as the push button switches, it is not practical to use them for
operations other than calibration because it is not possible to actuate two magnetic switches at once
with only one magnetic wa nd. Si nce disp layi ng the I nformat ion Scr een onl y requir es th e use o f one
switch, the wand may be used to show the Information Screen (see “Information Screen” on
page 26).
Status LEDs
The M2A includes five status LEDs that are located above the display (see Figure 9).
•Fail LED
The fail LED turns on when the M2A is experiencing a fail condition. A fail condition can be
caused by a detector failure or low detector signal.
•Alarm 1 LED
The alarm 1 LED is on when the M2A is experiencing an alarm 1 condition.
•Alarm 2 LED
The alarm 2 LED is on when the M2A is experiencing an alarm 2 condition.
•RX & TX LED’s
These LED’s indicate data being received (RX) and transmitted (TX) when the M2A’ s Modbus
output is operating.
13M2A Trans mitte r Op erato r’ s Manu al
Chapter 3: Installation & Startup
Overview
This chapter describes procedures to mount the M2A Transmitter in the monitoring environment
and wire it to input power and devices.
Mounting the M2A Transmitter
1.Select a mounting site that is representative of the monitoring environment. Consider the
following when you select the mounting site.
•Select a site where the M2A is not likely to be bumped or disturbed. Make sure there is
sufficient room to perform start-up, maintenance, and calibration procedures.
•Select a site where the target gas is likely to be found fi rst. For lighter gases, mount the
detector near the ceiling; for heavier gases, mount the detector near the floor.
2.At the monitoring site, use #10 screws throu gh the junction box’ s two mou nting holes to secure
the junction box to a vertical surface. Figure 11 - Figure 16 show the outline and mounting
dimensions for each version of the M2A. Mount the M2A with the detector facing down as
shown in the following figures.
WARNING:Always verify that the power source is OFF before making any wiring
connections.
1.Remove th e junction bo x cover.
2.Grasp the control PCB by its edges.
3.Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull
too hard and damage the cable which connects the control and terminal PCBs.
4.Let the control PCB hang by the cable. The terminal strips are now visible on the terminal
PCB. The control PCB may be left hanging while wiring is done. If desired, the control PCB
may be disconnected from the cable and set aside while wiring.
5.The detector leads are factory wired. Verify that the detector leads are wired to the detector
terminal strip as shown in the applicable figure below:
•C ata lytic LEL/Co mbustible IR/CO
Detectors
2
The wiring for these types of detectors is the same. Red wire to terminal labeled LEL R,
white wire to terminal labeled LEL W , g reen wire to terminal labeled LEL G, black wire to
terminal labeled LEL B.
Black
White
Red
Green
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
Figure 17: Catalytic LEL Detector Wiring
Black
Green
White
Red
- S +
FAIL
C NC NO
ALA R M 1
C NC NO
ALARM 2
C NC NO
Figure 18: Combusti bl e IR/ CO
LEL
R W G B
RS 485PWR/SIG
A B C
Detector Wiring
2
M2A Transmitter Operator’s Manual18
•Oxygen Detector, Partial Pressure, Capillary Type, and CSA Capillary Type
The wiring for all the oxygen detectors is the same. White wire to terminal labeled OXY
+, green wire to terminal labeled OXY -.
Figure 21: CSA Capillary Type Oxygen Detector Wiring
19M2A Trans mitte r Op erato r’ s Manu al
Green
White
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
Figure 22: Capillary Type Replaceable Sensor Oxygen Detector Wiring
•CSA type CO or H
S gas detector.
2
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
- S +
PWR/SIG
Black
Red
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
C NC NO
C NC NO
FAIL
ALARM 1
ALARM 2
FAIL
C NC NO
ALA R M 1
C NC NO
ALARM 2
C NC NO
Figure 23: H
S/CO Detector Wiring, CSA
2
•Non Explosion Proof CO or H2S gas detector
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
- S +
PWR/SIG
Black
Red
Figure 24: H
S/CO Detector Wiring
2
TOXIC OXY
+ - + -
RS 485
A B C
M2A Transmitter Operator’s Manual20
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
•ESM-01 Toxic Detectors
The ESM-01 toxic detectors have one black wire and one color coded wire. Connect the color
coded wire (see chart below) to the terminal labeled TOXIC + and the black wire to the terminal
labeled TOXIC -.
Black
See Chart
Wire ColorGasType
BrownAsH3
YellowCl2
NH3Red
OrangeNO
GreenPH3
BlueSO2
Figure 25: ESM-01 Toxi c Detector Wiring
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
6.To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the
wall between two terminal positions with needle nose pliers and pulling. Be careful not to exert
too much force on the wall to avoid damage to the terminal strip. The detector terminal strip
and the alarm 1 relay terminal strip may be removed by grasp in g them with yo ur fi nger s if th e
adjacent terminal strips have been removed.
W ARNING:If the M2A is installed in a hazardous location, use appropriately rated conduit,
conduit fittings, and appropriate construction technique that complies with the
local electrical code.
In addition, use appropriately rated conduit, conduit fittings, and appropriate
construction technique to maintain the NEMA 4X environmental ratin g of the
junction box and avoid water entering the junction box at the conduit hubs.
7.Guide multi conductor shielded cable or cables or wires in conduit through the conduit hub on
the left side of the junction box.
WARNING:If the top conduit hu b is used for wiring, be sure to seal the threads to ensure
water does not enter the junction box. A conduit seal, which is normally
required to maintain the junction box’ s hazardous location rating, will also help
to prevent water from entering through the top port.
21M2A Trans mitte r Op erato r’ s Manu al
The number of cables or wires needed will depend on whether the M2A is wired to a gas
monitoring controller or just to power, whether any relays are used, and whether the Modbus
output is used.
NOTE:If the M2A is being wired to a PLC or DCS device, see “Appendix B: PLC and DCS
Device Wiring” on page 66, then continue with step 8.
Use the following recommendations to determine how to wire the M2A:
•If Modbus connections will not be used and only the PWR/SIG connections will be used,
use a two or three conductor shielded cable or two or three wires in conduit for
connections to the power/signal terminal strip depending on whether or not the signal (S)
terminal is used. The S terminal has a 4 - 20 mA output, but if you do not need to monitor
this signal and do not connect to the S terminal to access this signal, the M2A will still
function completely.
Table 4: Wire Size for PWR/SIG Connections
Max Distance to Controller
w/18 Gauge Wire
2,500 ft.5,000 ft.8,000 ft.
Max Distance to Controller w/
16 Gauge Wire
Max Distance to Controller
•If the PWR/S IG connection s and one or more relays are used, route the connections to the
M2A in conduit. Use shielded cable in the conduit for the PWR/SIG connections and
unshielded cable or individual wires for the relay connections. Make sure any wire or
cable used for relay wiring is appropriately rated for the power that it will carry.
NOTE:If shielded cable is used for the PWR/SIG connections, leave the cable shield’s drain
wire insulated and disconnected at the M2A. You will connect the opposite end of the
cable’s drain wire at the controller or device.
•If the M2A will be wired into a Modbus network, see “Chapter 8: RS-485 Modbus
Output” on page 51.
w/14 Gauge Wire
M2A Transmitter Operator’s Manual22
RKI Controller
Terminals
See Figure 26 below for field wiring connections to the M2A.
Ala rm De vice
Power
(24 VDC) +
4 - 20 mA In (S)
(24 VDC) -
Fail Alarm
Device
FAIL
C NC NO
Alarm 1
Ala rm De vice
ALARM 1
C NC NO
Alarm 2
Ala rm De vice
ALARM 2
C NC NO
Typical Alarm
Wiring Shown
See
Detector
Wiring
See
Modbus
Wiring
- S +
LEL
R W G B
RS 485PWR/SI G
A B C
Figure 26: Wiring the M2A to a Controller and Alarm Devices
8.Re-install the control PCB (and ribbon cable if necessary). Be sure the ribbon cable is routed
down below the control PCB so it will not be damaged by the cover when it is screwed back
on.
9.Secure the junction box cover to the junction box.
Start Up
10. Make controller, device, and relay connections as appropriate. If shielded cable is used for the
PWR/SIG connections, connect the cable shield’s drain wire to an available chassis ground at
the gas monitoring controller, recording device, or programmable controller .
Introducing Incoming Power
1.Complete the installation procedures described earlier in this m anual.
2.Verify that all wiring connections are correct and secure.
3.Turn on the incoming power.
4.If necessary, turn on the controller or other monito ri ng device th at is connected to the M2A.
5.The OLED display will indicate the firmware version when the M2A is first powered up and
will then count down a one minute warm-up period before normal operation begins. During
normal operation, the display will indicate the target gas and current gas reading. Verify that
the display is indicating the target gas and current gas reading after the warm-up period is
complete and normal operation begins.
METHANE
0 %LEL
23M2A Trans mitte r Op erato r’ s Manu al
NOTE:Allow the M2A’ s detector to warm up for 15 minutes bef ore you cont inue with t he next
section, “Adjusting the Fresh Air Reading.”
Adjusting the Fresh Air Reading
When the M2A is shipped from RKI Instruments, Inc., it is factory calibrated. If a full calibration is
desired at startup, see “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 42 or
“Calibration, Oxygen Version” on page 45.
Verify that the M2A is in a fresh air environment (environment known to be free of the target gas
and combustible or toxic gas vapors and of normal oxygen content, 20.9%).
CAUTION:If you suspect the presence of combustible gas, CO2, toxic gas, or that the oxygen
content is not normal in the monitoring environment, use the calibration kit and the
zero air calibration cylinder to introduce “fresh air” to the detector and verify an
accurate fresh air setting. See the “Calibratio n, Combus tible Gas , C O2, and Toxic
Versions” on page 42 or “Calibration, Oxygen Version” on page 45 for
instructions on how to use a zero air cylinder when performing a fresh air
adjustment.
Non-Intrusive Fresh Air Signal Adjustment
If the M2A is installed in a classified area and non-intrusive zero adjustment is requi red, fol low the
instructions below, but do not remove the junction box cover. Use the magnetic wand accessory
to actuate the magnetic switches instead of pushing the control switch buttons. To actuate a
magnetic switch and perform the same operation as pressing a control switch button, touch the
magnet in the end of the magnetic wand to the M2A’s glass cover directly over the magnetic switch
you wish to actuate. Touching the glass and removing the wand is the same as pressing and
releasing a button. Touching the glass and keeping the wand in place is the same as pressing and
holding a button.
W ARNING:The M2A is not an active gas monitoring device during the fresh air adjustment
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA for a toxic,
combustible gas, or CO
M2A, or at 17.4 mA for an oxygen M2A, and all r el ays
2
will remain in their non-alarm state while the M2A is in Calibration Mode. The
output signal will not indicate current rea dings and the relays will not resume
operating normally until the M2A is in norma l operation again.
NOTE:While in the calibration program, if there is no switch activity for the calibration time-
out period the unit will return to normal operation. The factory set time-out is 15
minutes. If you want a different time-out period , see “Viewing & Changing M2A
Parameters” on page 31 for instructions to change the calibration time-out.
Combustible Gas, CO
, and Toxic Versio ns, Adjus ting the F resh Air Readin g
2
1.While in normal operation, press and hold the UP/YES button for 5 seconds to enter
Calibration Mode. Release the button when the following screen appears.
Calib?
YES/NO
M2A Transmitter Operator’s Manual24
2.Press and release the UP/YES button to continue. The display will indicate the target gas and
CAL Mode for a few seconds before showing FreshAir Adjust?.
3.Press and release the UP/YES button. ENTER will alternate with FreshAir on the top display
line and the current gas reading will be on the bottom display line.
4.Press and release the ENTER button. The M2A will perform a zero operation and the display
will indicate SPAN w/Cal Gas?.
5.Press and release the DOWN/NO button. The display will indicate Leaving CAL Mode and
the M2A will return to normal operation.
Oxygen Version, Adjusting the Fresh Air Reading
1.While in normal operation, press and hold the UP/YES button for 5 seconds to enter
Calibration Mode. Release the button when the following screen appears.
Calib?
YES/NO
2.Press and release the UP/YES button to continue. The display will indicate OXYGEN CAL
Mode for a few seconds before showing FreshAir Adjust?.
3.Press and release the UP/YES button. ENTER will alternate with FreshAir on the top display
line and the current oxygen reading will be on the bottom display line.
4.Press and release the ENTER button. The M2A will perform a span operation and then the
display will indicate ZERO w/Cal Gas?.
5.Press and release the DOWN/NO button. The display will indicate Leaving CAL Mode and
the M2A will return to normal operation.
25M2A Trans mitte r Op erato r’ s Manu al
Chapter 4: Operation
Overview
This chapter describes the M2A in normal operation. This chapter also describes the M2A in alarm
1, alarm 2, and fail conditions and suggests responses to these conditions.
Normal Operation
Normal operation is defined as follows:
•The start-up procedure is complete.
•The M2A is not indicating an alarm 1, alarm 2, or fail condition.
•The M2A is not in Calibration, Configuration, or Gas Type modes.
During normal operation, the M2A simultaneously displays the current gas reading, unit of
measure, and target gas. The example below illustra tes a typical combustible gas M2A.
Information Screen
The Information Screen displays the M2A’s operating voltage and firmware and hardware
information. To display the Information Screen, do the following:
1.Press and hold the DOWN/NO button for 3 seconds. The Information Screen will appear.
2.Continue holding the DOWN/NO button to keep the Information Screen on the display. The
top line indicates the operating voltage that is connected to the M2A. The second line indicates
the firmware that is running, version 6.0 in this example, and the M2A’s hardware version,
version 2A in this example. This information may be useful if you need to contact the factory
with questions about the M2A.
3.Release the DOWN/NO button when you are done viewing the Information Screen. The
display will return to the normal operation screen within a couple of seconds.
METHANE
0 %LEL
24.2V
v6.0L 2A
M2A Transmitter Operator’s Manual26
4 - 20 mA Signal Output Operation
The output at the S terminal of the power/signal terminal strip is a 4 - 20 mA signal that
corresponds to the detection range of the M2A. During normal operation, this signal tracks the gas
concentration on the OLED display.
There are several circumstances where the signal output will not track the di splay reading but will
behave as follows:
•When the M2A is in its warm-up period, the signal output will be fixed at 3.5 mA (zero) for all
gas types except oxygen. For oxygen types, the output will be fixed at 17.4 mA (20.9%
oxygen) while the M2A is in warm-up.
•When the M2A’s gas type is changed, the M2A will enter Configuration Mode for you to
verify the parameter settings. When you exit Configuration Mode, the display will indicate
NEEDS CALIBRATION and will continue to indicate this until Calibration Mode is entered
and a calibration is performed. In this situation, the signal output will be fixed at 3.5 mA for
LEL and toxic versions and 17.4 mA for oxygen versions from the time Gas Type Mode is
entered until the M2A is calibrated and returns to normal operation.
•If you enter Calibration Mode, Configuration Mode, Gas Type Mode, or Modbus Mode, the
signal output will be fixed at 3.5 mA for combustible gas, CO
mA for oxygen versions until the M2A returns to normal operation.
, and toxic versions and 17.4
2
•If the M2A’s input power decreases below 9.5 volts so that the M2A is in a low power alarm,
the signal output is fixed below 2.4 mA until the low power alarm is cleared.
•If the M2 goes into a fail condition, after a 30 second delay, the signal output is fixed below 2.4
mA until the fail alarm is cleared. During the 30 second delay, the signal output follows the
detector output. In the case of a downscale reading, the display and the signal output continue
to track the reading down to -99% of full scale (1.15 mA).
27M2A Trans mitte r Op erato r’ s Manu al
Alarm Indications
NOTE:The M2A includes alarm on and alarm off delay settings for alarm 1 and alarm 2. The
ConditionCauseVisual Indication(s)
alarm indications described in this section operate according to the factory set alarm
settings. See Table 6 on page 32 for all the factory settings.
Table 5: Visual and Audible Alarm Indications
1
Alarm 1
1
Alarm 2
Fail•Disconnect ed or misconnected detector
Low PowerDC power source less than 9.5 volts.•F LED is on
1
*
If the M2A is in both an alarm 1 and an alarm 2 conditio n, bo th ala rm LEDs are on an d the
display alternates between the gas reading and the ALMS 1&2 message.
Increasing (decreasing for O2) gas reading at
or above the alarm 1 setpoint
Increasing gas reading at or above the alarm 2
setpoint
wiring
•Display reading at -10% of full scale or
lower
•Defective components
•A1 LED is on
•Gas reading alternates with
ALARM-1 message
•A2 LED is on
•Gas reading alternates with
ALARM-2 message
•F LED is on
•FAIL message replaces gas
reading
NOTE: There is a 30 second delay
on the fail condition.
•LowPower message and actual
voltage of incoming DC power
NOTE:You can select normally energized (N. EN) or normally de-energized (N. DE-EN)
alarm 1 and alarm 2 relay settings in Configuration Mode. The following sections
describe the standard factory setting for these relays which is N. DE-EN.
The fail relay is factory set as N. EN and is not user-adjustable.
Alarm 1 Condition
Alarm 1 Condition Indications
When the gas reading reaches the alarm 1 setpoint, the M2A senses an alarm 1 condition. The M2A
alerts you to an alarm 1 condition as follows:
•The A1 LED turns on.
•The gas reading alternates with the ALARM-1 message .
•The alarm 1 relay energizes.
Responding to an Alarm 1 Condition
1.Follow your established procedure for a low level combustible or toxic gas condition or a
decreasing oxygen concentration condition.
2.After the gas reading falls below (above for oxygen) the alarm 1 setpoint, press the ENTER
button to reset the alarm 1 circuit. Resetting the alarm 1 circuit turns off the A1 LED, resets the
M2A Transmitter Operator’s Manual28
OLED display, and de-energizes the alarm 1 relay.
NOTE:If the ENTER button is pressed while the M2A is in an alarm 1 condition, the A1 LED
will flash but all other indications will remain unchanged.
You cannot de-energize the alarm 1 relay until the gas reading falls below (above for
oxygen) the alarm 1 setpoint.
Alarm 2 Condition
Alarm 2 Condition Indications
When the gas reading reaches the alarm 2 setpoint, the M2A senses an alarm 2 condition. The M2A
alerts you to an alarm 2 condition as follows:
•The A2 LED turns on.
•The gas reading alternates with the ALARM-2 message .
•The alarm 2 relay energizes.
NOTE:If the M2A is in both an alarm 1 and alarm 2 condition, both the A1 and A2 LEDs will
be on, the gas reading will alternate with the ALMS 1&2 message, and both alarm
relays will energize.
Responding to an Alarm 2 Condition
1.Follow your established procedure for a high level combustible or toxic gas condition or an
increasing oxygen concentration condition.
2.After the gas reading falls below the alarm 2 setpoint, press the ENTER button to reset the
alarm circuit. Resetting the alarm circuit turns off the A2 light, resets the OLED display, and
de-energizes the alarm 2 relay.
NOTE:If the ENTER button is pressed while the M2A is in an alarm 2 condition, the A2 LED
will flash but all other indications will remain unchanged.
You cannot de-energize the alarm 2 relay until the gas reading falls below the alarm 2
setpoint.
Fail Condition
Fail Condition Indications
The M2A indicates a fail condition for any of the following:
•The detector wiring is disconnected or incorrectly connected.
•The display reading is -10% of full scale or lower.
When the M2A senses a fail condition, it alerts you as follows:
•The F LED turns on.
•The gas reading is replaced by the FAIL message.
•The fail relay de-energizes.
NOTE:The fail alarm has a 30 second delay.
29M2A Trans mitte r Op erato r’ s Manu al
Responding to a Fail Condition
1.Verify that the detector wiring is correctly and securely connected.
2.If the M2A has a replaceable plug-in sensor, verify that the replaceable plug-in sensor in the
detector housing is properly installed.
Low Power Alarm
Low Power Alarm Indications
The M2A senses a low power condition when the DC power source is 9.5 volts or less.
WARNING :While in a low power condition, the M2A is not an active gas monitor.
When the M2A senses a low power condition, it alerts you as follows:
•The F LED turns on.
•Th e message LowPower is indicated on the top line of the OLED display and the input voltage
is displayed on the bottom line of the OLED display.
•The fail relay de-energizes.
NOTE:The low power alarm cannot be cleared using the ENTER button.
When the voltage increases to 10.0 volts, the low power alarm is cleared and the M2A will begin its
warm-up sequence.
Responding to a Low Power Condition
1.Determine and correct the cause of the low power condition.
2.When the input power increases above 10.0 volts, the M2A will begin its warm-up sequence.
3.Verify that the M2A enters normal operation after its warm-up sequence. If necessary , perform
a fresh air adjustment. See “Adjusting the Fresh Air Reading” on page 24.
M2A Transmitter Operator’s Manual30
Chapter 5: Configuration Mode
Overview
This chapter describes how to view and change M2A parameters using Configuration Mode. It is
accessed using the program buttons.
Configuration Mode incl udes a 5-mi nute time-o ut featur e. If you do not press a cont rol but ton for 5
minutes, the M2A automatically returns to normal operation.
NOTE:If the M2A returns to normal operation because of a time-out, it enters a warm-up
period just as it does when it is first turned on.
If you are installing a new M2A, it has been setup at the factory. Use Configuration
Mode only if you want to change the M2A’s setup. If you want to change the detector
type see “Chapter 6: Gas Type Mode” on page 34.
Viewing & Changing M2A Parameters
1.While in normal op eration, simultaneou sly pr ess and h old the UP/ YES and EN TER button s for
5 seconds to enter Configuration Mode. Release the buttons when the following screen
appears.
Enter
Config?
2.If you want to exit Configuration Mode, press and release the DOWN/NO button and the M2A
will return to normal operation.
If you want to continue in Configuration Mode, press and release the UP/YES button. The
target gas and the full scale will be displayed for a few seconds before the first adjustable
parameter, the alarm 1 setpoint, is displayed.
3.If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO
buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter is OK, press the ENTER button to proceed to the next
parameter.
Table 6 lists the M2A parameters you can set. Table 6 also lists the factory set value for each
parameter.
31M2A Trans mitte r Op erato r’ s Manu al
Tabl e 6: Configuration Parame ters
Parameter
(Factory Set Value)
ALARM-1 (level)
(See “M2A Specifications” on page 2)
ALARM-1 (activation)
(Decrease for oxygen M2A, Increase for
all other types)
ALARM-1 (relay action)
(N. DE-EN)
ALARM-1 (relay reset)
(LATCH)
A1 OnDy (alarm 1 on delay)
(1 secs)
A1 OffDy (alarm 1 off delay)
(0 sec.)
Description
The gas reading at which the M2A initiates an alarm 1 condition.
Indicates if the alarm 1 circuit is activate d by gas readin gs increas ing (Increase)
or decreasing (Decrease) to the ALARM-1 Level.
If set as N. DE-EN, the alarm 1 relay is de-energized in normal operation and
energizes when an alarm 1 condition is initiated.
If set as N. EN, the alarm 1 rel a y is energized in normal operation and de-energizes when an alarm 1 condition is initiated.
If set as LATCH, you must press the ENTER button to reset the alarm 1 circuit
after the alarm 1 condition passes.
If set as SELF-RST, the M2A automatically resets the alarm 1 circuit after the
alarm 1 condition passes.
The amount of time the M2A delays activatio n of the ala rm 1 circuit on ce an
alarm 1 condition is initiated. It can be set in 1 second increments from 0 - 60
seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 - 60 minutes.
The amount of time the M2A delays turning off the alarm 1 circuit once an alarm
1 condition passes. Th is pa ramete r appe ars o nl y if th e ala rm 1 re la y re set se ttin g
is set to SELF-RST. It can be set in 1 second increments from 0 - 60 seconds, in
1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 60 minutes.
ALARM-2 (level)
(See “M2A Specifications” on page 2)
ALARM-2 (activation)
(Increase for all types)
ALARM-2 (relay action)
(N. DE-EN)
ALARM-2 (relay reset)
(LATCH)
A2 OnDy (alarm 2 on delay)
(1 secs)
A2 OffDy (alarm 2 off delay)
(0 sec.)
The gas reading at which the M2A initiates an alarm 2 condition.
Indicates if the alarm 2 circuit is activate d by gas readin gs increas ing (Increase)
or decreasing (Decrease) to the ALARM-1 Level.
If set as N. DE-EN, the alarm 2 relay is de-energized in normal operation and
energizes when an alarm 2 condition is initiated.
If set as N. EN, the alarm 2 rel a y is energized in normal operation and de-energizes when an alarm 2 condition is initiated.
If set as LATCH, you must press the ENTER button to reset the alarm 2 circuit
after the alarm 2 condition passes.
If set as SELF-RST, the M2A automatically resets the alarm 2 circuit after the
alarm 2 condition passes.
The amount of time the M2A delays activatio n of the ala rm 2 circuit on ce an
alarm 2 condition is initiated. It can be set in 1 second increments from 0 - 60
seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 - 60 minutes.
The amount of time the M2A delays turning off the alarm 2 circuit once an alarm
2 condition passes. Th is pa ramete r appe ars o nl y if th e ala rm 2 re la y re set se ttin g
is set to SELF-RST. It can be set in 1 second increments from 0 - 60 seconds, in
1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 60 minutes.
M2A Transmitter Operator’s Manual32
Table 6: Configuration Parameters (Continued)
Parameter
(Factory Set Value)
ZeroSupp
(0.0% of the detection range in terms of
the detection units for carbon dioxide
channel types, 0.5% oxygen for oxygen
types, 2% of the detection range in terms
of the detection units for all other types)
FILTER
(5 secs)
CAL Time
(15 mins)
Auto Zero
(Disabled for carbon dioxide channels;
Enabled for all other channel types)
4.When you have scrolled through all the adjustable parameters, SA VE IT? YES/NO appears on
the display.
Description
The zero suppression feature helps prevent “jumpy” readings near the fresh air
reading.
For example, if the zero suppression setting for a combustible LEL detector is
2.0% LEL, the M2A will display a reading of 0% LEL for gas readings from 2% LEL to 2% LEL. It is settable from 0 to 6% of the detection range.
The filter feature helps “smooth out” jumpy or noisy signals from the detector.
You can set the filter from 0 seconds to 60 seconds in 5 seconds increments.
The displayed gas reading is the average reading over the previous time period
defined by the filter setting. So if the filter is set to 5 seconds, the displayed gas
reading is the average over the past 5 seconds.
The calibration time-out sets the amount of time after the last button push while
in Calibration Mode that the M2A will wait before returning to normal operation. If the calib ration t ime-out is set to 15 mins, then M2A will return to normal
operation automatically 15 minu tes af ter the last butto n pu sh.
If set to Enabled, the M2A’s firmware will monitor the zero (fresh air) reading
and make automatic zero adjustme nts under precisely specified conditions if it
determines that a reading change is due to normal sensor drift.
If set to Disabled, the Auto Zero function is turned off.
5.If you do not wish to save the adjustments and want to exit Configuration Mode, press and
release the DOWN/NO button. The DO OVER? YES/NO message will display. Press and
release the DOWN/NO button. The ABORT? YES/NO message will display. Press the UP/
YES button to return to normal operation.
If you wish to change some of the adjustments made, pr ess and release the DOWN/NO button.
The DO OVER? YES/NO message will display. Press and release the UP/YES button. The
Re-do Config message will display and the M2A will return to the first adjustab le parameter.
Go back to step 3 and continue.
If you wish to save t he adjus tments made, pres s and release the UP /YES button . Config Saved
is indicated on the display for a few seconds and the M2A returns to normal operation.
33M2A Trans mitte r Op erato r’ s Manu al
Chapter 6: Gas Type Mode
Overview
This chapter describes how to use Gas Type Mode to select the M2A’s gas type. The gas type
determines the target gas and detection range. The combustible gas/CO
have a different terminal PCB and run on different firmware. Your M2A will only have gas type
choices available that it can support.
Gas Type Mode includes a 5-minute time-out feature. If you do not press a button for 5 minutes, the
M2A automatically returns to normal operation.
NOTE:If the M2A enters normal operation because of a program time-out, it enters a warm-up
period just as it does when it is first turned on.
Selecting the Gas Type
and toxic/oxygen M2As
2
1.While in normal operation or during the warm-up sequence, press and hold the UP/YES and
DOWN/NO buttons for five seconds. Release them when the following screen appears.
Select
GasType?
2.If you want to exit Gas Type Mode, press and release the DOWN/NO button. The display will
indicate NO CHANGE and the M2A will return to normal operation.
If you want to continue in Gas Type Mode, press and release the UP/YES button. The top
display line will indicate GasType? and the bottom display lin e will indicate the target gas
choice.
3.Use the UP/YES or DOWN/NO button to scroll through the choices of target gas. Table 7 and
Table 8 below list the choices of target gas for the M2A.
NOTE:Consult factory for availability of detectors for toxic M2A types not listed in “M2A
Specifications” on page 2.
4.When the desired target gas is on the display, press and release the ENTER button. The display
will ask SAVE IT? YES/NO.
5.To discard the gas type change, press and release the DOWN/NO button. The display will ask
DO OVER? YES/NO. Press the DOWN/NO button. The display will ask ABORT? YES/
NO. Press the UP/YES button. The M2A will return to normal operation without changing the
gas type.
T o save the gas type, p ress and release the UP/YES butto n. The display will indicate Gas Type
Updated, the display will indicate the target gas and the detection range for a few seconds, and
then the M2A will enter Configuration Mode automatically so that the parameter settings in
Configuration Mode can be verified or changed.
6.While in Configuration Mode press and release the ENTER button to accept the displayed
parameter setting and move to the next one. Use the UP/YES and DOWN/NO buttons to adjust
a setting. See “Chapter 5: Configuration Mode” on page 31 for a complete description of
Configuration Mode.
7.When you have reviewed and accepted the last item in Configuration Mode by pressing the
ENTER button, display will ask SA VE IT? YES/NO.
8.To save the configuration settings, press and release the UP/YES button.The display will
indicate Config Saved and the M2A will begin its warm-up sequence.
To discard the settings and review them again, press the DOWN/NO button. The display will
ask DO OVER? YES/NO. Press the UP/YES button and the display will indicate Re-do Config, then display the target gas and the detection range for a few seconds before returning
to Configuration Mode. Repeat steps 6 - 8 until the settings are as desir e d.
9.After the M2A completes its warm-up sequence, the display will indicate CAL NEEDED.
Since the gas type has been changed, a successful calibration must be performed before the
M2A can enter normal operation.
10. Press and release the UP/YES button to enter Calibration Mode. Normally, you must hold the
UP/YES button for 5 seconds to enter Calibration Mode, but since a calibration is required ,
only a momentary push is needed to enter Calibration Mode after changing the gas type.
11. See “Calibration, Combustible Gas, CO2, and Toxic Versions” on page 42 or “Calibration,
Oxygen Version” on page 45 for calibration in structions.
NOTE:When calibrating an M2A after changing the gas type, the M2A will not ask if you want
to calibrate, or whether you want to perform a fresh air adjustment, span adjustment or
zero adjustment. Since a complete calibration is required, the calibration sequence will
proceed without asking if you want to perform certain operations.
M2A Transmitter Operator’s Manual36
Chapter 7: Maintenance
Overview
This chapter describes procedures for performing preventive maintenance, troubleshooting,
calibrating the M2A, and replacing field replaceable parts. It includes separate procedures for
calibrating the combustible gas/CO
Preventive Maintenance
This section describes a recommended preventive maintenance schedule to ensure the optimum
performance of the M2A. It includes daily, monthly, quarterly, and biannual procedures.
Daily
Verify a display reading of zero (20.9% for an oxygen M2A). Inves ti gate s ign if icant ch anges in th e
reading.
/toxic and oxygen versions of the M2A.
2
Monthly
This procedure describes a test to verify that the M2A responds properly to the target gas.
NOTE:Performing a response tes t on the M2A may cause alarms. If you do not want external
alarms to actuate, disable any external alarms during the response test to avoid
unwanted external alarm indications.
NOTE:The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a fixed flow regulator with an on /off knob, a calibration cup for
the detector, and a short piece of sample tubing to connect the regulator to the
calibration cup.
1.Disable all external alarms if external alarm indications are not desired during the response
test.
2.Connect the calibration kit sample tubing to the calibration cup hose barb.
3.Screw the calibration cup onto the detector.
NOTE:For an ESM-01 detector, you must unscrew the removable splashguard from the
detector housing cap so you can push the appropriate calibration cup onto the ESM-01
sensor . Make sure to use the correct calibra tion cup for the sensor being used. There is a
Cl
type calibration cup for a Cl2 sensor, an NH3 type calibration cup for an NH3
2
sensor, and a general type (usually labeled SO
sensors.
) calibration cup for all other ESM-01
2
4.Connect the fixed flow regulator to the calibration cup using the sample tubing.
37M2A Trans mitte r Op erato r’ s Manu al
NOTE:When applying gas to an NH3 ESM-01, the 0.25 LPM flowrat e reg ul at or mus t be u sed .
All other M2As require a 0.5 LPM flowrate regulator.
5.Screw the calibration gas cylinder into (onto for an H
6.Turn the regulator knob counterclockwise to open the regulator.
7.Allow calibration gas to flow for one minute.
8.Verify that the display reading is within ± 20% of the gas concentration.
NOTE: If the reading is not within ± 20% of the gas concentration, calibrate the M2A as
described in “Calibration, Combustib le Ga s, CO2, and Toxic Versions” on page 42 or
“Calibration, Oxygen Version” on page 45.
9.Turn the regulator knob clockwise to close the regulator.
10. Unscrew the calibration gas cylinder from the fixed flo w regulator and unscr ew the calib ration
cup from the detector.
For an ESM-01 detector, pull the calibration cup off the ESM-01 sensor and screw the
splashguard back onto the detector housing cap.
11. If external alarms have been disabled, enable them when the display returns to a normal fresh
air reading.
12. Store the components of the calibration kit in a safe place.
S cylinder) the fixed flow regulator.
2
Quarterly
Calibrate M2As that do not have an IR detector as described in “Calibration, Combustible Gas,
CO2, and Toxic Versions” on page 42 or “Calibration, Oxygen Version” on page 45. See
“Calibration Frequency” on page 41 for a discussion of calibration frequency guidelines.
Biannually
Calibrate M2As that have an IR detector as described in “Calibration, Combustible Gas, CO2, and
T oxic Versions ” on page 42. See “Calibration Frequency” on p age 41 for a discussion of calibratio n
frequency guidelines.
M2A Transmitter Operator’s Manual38
Troubleshoo ting
The troubleshooting guide describes symptoms, probable causes, and recommended action for
problems you may encounter with the M2A.
NOTE: This troubleshooting guide describes M2A problems only. If the M2A is connected to a
controller, see the controller o perator’s manual for problems you may encounter with the
controller.
Table 9:Troubleshooting the Combustible Gas Detector
The M2A alerts you to
frequent or suspect alarms
while the fresh air
readings remain on zero
(20.9 for oxygen).
The display reading
flickers often.
•The power wiring is
disconnected or
misconnected.
•The display cable is
disconnected or
misconnected.
•The M2A is
experiencing false
readings due to RFI or
EMI.
•The detector wiring is
disconnected,
misconnected, or
intermittent.
•The M2A is
experiencing false
readings due to RFI or
EMI.
•The noise filter setting is
too low.
•The zero suppression
setting is too low .
•The display screen is
malfunctioning.
1. Verify that the wiring to the power sou rce
or controller is correct and secure.
2. Verify that the ribbon cable connecting
the terminal PCB to the control PCB is
securely installed. This ribbon cable
plugs into a connector on the top edge of
the control PCB and on the bottom of the
terminal PCB.
3. If the power dif fic ulti es con tin ue, c ont act
RKI for further instruction.
1. Verify that the M2A wiring is properly
shielded. See “Wiring the M2A
Transmitter” on page 18.
2. Verify that the detector wiring is correct
and secure.
3. Increase the alarm on delay setting in
Configuration Mode.
4. If the frequent or suspect alarm
difficulties continue, contact RKI for
further instruction.
1. Verify that the M2A wiring is properly
shielded. See “Wiring the M2A
Transmitter” on page 18.
2. Verify that the detector wiring is correct
and secure.
3. Increase the filt er setting in Conf iguration
Mode.
4. Increase the zero suppression setting in
Configuration Mode.
5. If the display difficulties continue,
contact RKI for further instruction.
39M2A Trans mitte r Op erato r’ s Manu al
Table 9:Troubleshooting the Combustible Gas Detector (Continued)
Slow or No
Response/
Difficult or
Unable to
Calibrate
condition.
•Controller indicates a
fail condition
•M2A responds slowly
or does not respond to
response test.
•Unable to successfully
set the fresh air or
span (zero for oxygen)
reading during
calibration.
•M2A requires
frequent calibration.
Note: See “Calibration
Frequency” on page 41
for calibration frequency
guidelines.
•The detector wiring to
the terminal PCB is
disconnected or
misconnected.
•The wiring from the
M2A to the controller is
disconnected or
misconnected.
•The detector fresh air
signal is low enough to
cause a fail condition.
•The detector is
malfunctioning.
•The terminal or display
PCB is malfunctioning.
•The calibration cyl inder
is low, out-dated, or
defective.
•The detector is
malfunctioning.
•The terminal PCB is
malfunctioning.
1. Verify that the detector wiring is correct
and secure.
2. Verify that the wiring between the M2A
and the controller is correct and secure.
3. Calibrate the M2A.
4. If the fail condition continues, replace the
detector.
5. If the fail condition continues, contact
RKI for further instruction.
1. Verify that the calibration cylinder
contains an adequate supply of a fresh test
sample.
2. If the calibration/response difficulties
continue, replace the dete ctor.
3. If the calibration/response difficulties
continue, contact RKI for further
instruction.
M2A Transmitter Operator’s Manual40
Calibration Frequency
Some M2As have IR (infrared) type detectors , such as an M2A for CO2 or an LEL M2A with an IR
detector. Most of the M2As do not have an IR detector. For example, an H
electrochemical detector and some of the combustible M2As use catalytic detectors. The M2As that
use IR detectors typically need less frequent calibration as is discussed below. See Table 1, “M2A
Specifications,” on page 2 to see which M2As are IR versions.
Although there is no particular calibration frequency that is correct for all applications, a calibration
frequency of every 3 to 6 months for M2As that do not have an IR detector and every 6 to 12
months for M2As that have an IR detector is adequate for most M2A applications. Unless
experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a
calibration frequency of every 3 months for M2As that do not have an IR detector and every 6
months for M2As that have an IR detector.
If an application is not very demanding, for example detection in a clean, temperature controlled
environment where the toxic or combustible target gas is not normally present, and calibration
adjustments are minimal at calibration, then a calibration frequency of every 6 mont hs f or a non-IR
M2A or 12 months for an IR M2A is adequate.
If an application is very demanding, for example if a combustible or toxic gas is present often and
in significant concentrations or the environment is not well controlled, then more frequent
calibration than every 3 months for a non-IR M2A or every 6 month s for an IR M2A may be
necessary. For combustible gas detection with an M2A that uses a catalytic detector, if potential
catalyst poisons are known or likely to be present, more frequent calibration than every 3 months
will be necessary.
S M2A has an
2
41M2A Trans mitte r Op erato r’ s Manu al
Calibration, Combustible Gas, CO2, and Toxic Versions
This section describes how to calibrate the combustible gas, CO2, and toxic gas versions of the
M2A. It includes procedures to prepare for calibration, enter Calibration Mode, adjust the fresh air
(zero) setting, adjust the span setting, and return to normal operation. See the next section,
“Calibration, Oxygen Version” for instructions to calibrate an oxygen M2A.
WARNING :The M2A is not an active gas monitoring device during th e calibration
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA and all relays will
remain in their non-alarm state while the M2A is in Calibration Mode. The
output signal will not indicate current rea dings and the relays will not resume
operating normally until the M2A is in norma l operation again.
Prepari ng for Ca libration
NOTE:The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a zero air cylinder, a fixed flow regulator with an on/off knob,
a calibration cup for the appropriate sensor, and a short piece of sample tubing to
connect the regulator to the calibration cup.
NOTE:While in the calibration program, if there is no switch activity for the calibration time-
out period the unit will return to normal operation. See “Viewing & Changing M2A
Parameters” on page 31 for instructions to set the calibration time-out.
NOTE:The following procedure assumes that the target gas is present in a high enough
concentration to affect the fresh air (zero) reading. If a fresh air environment can be
verified, applying zero air to the detector is not necessary when setting the zero (span
for oxygen) reading.
1.Connect the calibration kit sample tubing to the calibration cup hose barb.
2.Screw the calibration cup onto the detector.
NOTE:For an ESM-01 detector, you must unscrew the removable splashguard from the
detector housing cap so you can push the appropriate calibration cup onto the ESM-01
sensor . Make sure to use the correct calibra tion cup for the sensor being used. There is a
Cl
type calibration cup for a Cl2 sensor, an NH3 type calibration cup for an NH3
2
sensor, and a general type (usually labeled SO
sensors.
3.Connect the fixed flow regulator to the calibration cup using the sample tubing.
NOTE:When calibrating an NH3 ESM-01, the 0.25 LPM flowrate regulator must be used. All
other M2As require a 0.5 LPM flowrate regulator.
) calibration cup for all other ESM-01
2
M2A Transmitter Operator’s Manual42
Non-Intrusive Calibration
If the M2A is installed in a classified area and non-intrusive calibration is required, follow the
instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to
actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic
switch and perform the same operation as pressing a control switch button, touch the magnet in the
end of the magnetic wand to the M2A’s glass cover directly above the magnetic switch you wish to
actuate. Touching the glass and removing the wand is the same as pressing and releasing a button.
Touching the glass and keeping the wand in place is the same as pressing and holding a button.
Calibration Gas Response Memory Feature
The M2A has the capability to “remember” the detector’s response to the calibration gas after the
gas is removed from the detector during the fresh air and span adjustment procedure. This feature
enables one person to perform a calibration if the detector is mounted remotely from the M2A.
When zero air is applied to the M2A during a fresh air adjustment, the M2A will freeze the display
reading at the lowest response and the M2A will continue to display this reading and retain it in its
memory until the fresh air adjustment procedure is completed. When calibration gas is applied to
the M2A during a span adjustment, the M2A will freeze the display reading at the highest response
to the calibration gas. The calibration gas can then be removed and the M2A will continue to
display this reading and retain it in its memory until the span adjustm e nt procedure is completed.
Adjusting the Fresh Air Reading
1.While in normal operation, press and hold the UP/YES button for 5 seconds to enter
Calibration Mode. Release the button when the following screen appears.
Calib?
YES/NO
2.If you want to continue with calibration, press and release the UP/YES button. The display will
indicate the target gas and CAL Mode for a few seconds before showing FreshAir Adjust?.
If you want to exit Calibration Mode, press and release the DOWN/NO button. The M2A will
indicate Leaving CAL Mode and the M2A will return to normal operation.
3.T o continue with adjusting the fresh air reading, press and release the UP/YES button. ENTER
will alternate with FreshAir on the top display line and the current gas reading will be on the
bottom display line.
To skip adjusting the fresh air reading, press and release the DOWN/NO button. The display
will indicate SPAN w/Cal Gas?. Skip to the next section, “Adjusting the Span Setting”.
4.Screw the zero air cylinder into the fixed flow regulator.
5.Turn the regulator knob counterclockwise to open the regulator.
6.Allow zero air to flow for two minutes. If there was gas present causing a reading on the M2A,
the reading should decrease and stabilize after two minutes.
7.Turn the regulator knob clockwise to close the regulator.
8.Unscrew the zero air cylinder from the fixed flow regulator leaving the calibration cup on the
detector. The M2A wi ll continue t o display th e minimum gas response on the displ ay and
retain the response level in its memory.
9.Press and release the ENTER button. The M2A will perform a fresh air adjustment and the
display will indicate SPAN w/Cal Gas?.
NOTE:If the fresh air adjustment fails, see “Troubleshooting” on page 39 for recommended
actions.
43M2A Trans mitte r Op erato r’ s Manu al
Adjusting the Span Setting
1.If you want to continue with adjusting the span setting, press and release the UP/YES butto n.
APPLY will alternate with SPAN Gas on the top display line and the current gas reading will
be on the bottom display line.
If you want to skip adjusting the span reading, press and release the DOWN/NO button. The
display will indicate Leaving Cal Mode and the M2A will return to normal operation.
2.Screw the calibration gas cylinder into (onto for an H
S cylinder) the fixed flow regulator.
2
3.Turn the regulator knob counterclockwise to open the regulator.
4.Allow calibration gas to flow for two minutes. The gas reading should be stable after two
minutes.
5.Turn the regulator knob clockwise to close it.
6.Unscrew the calibration gas cylinder from the fixed flow regulator and unscre w the calibrat ion
cup from the detector. For an ESM-01 detector, pull the calibration cup off of the ESM-01
sensor and screw the splashguard back onto the detector housing cap. The M2A will continue
to display the maximum gas response on the display and retain the response level in its
memory.
7.Adjust the gas reading up or down to match the calibration gas cylinder concentration by using
the UP/YES and DOWN/NO buttons, then press and release the ENTER button.
8.The M2A will perform a span operation. The display will indicate SPAN Gas PASS for a few
seconds, then indicate SPAN Gas SAVED before indicating Leaving CAL Mode for a few
seconds.
NOTE:If the span adjustment fails, see “Troubleshooting” on page 39 for recommended
actions.
9.The display will now alternate between the normal operation screen and the message
REMOVE CAL GAS for 1 minute. If the calibration gas has not been removed from the
detector, remove it now to avoid unwanted alarms.
During this 1 minute period, the signal output will remain fixed at 3.5 mA and the relays will
remain in their non-alarm state to avoid unwanted alarms wh ile the calibration gas clears fr om
the detector. At the end of the 1 minute period, the REMOVE CAL GAS message will stop
appearing and the M2A will return to normal operation.
10. Store the components of the calibration kit in a safe place.
M2A Transmitter Operator’s Manual44
Calibration, Oxygen Version
This section describes how to calibrate the oxygen version of the M2A. It includes procedures to
prepare for calibration, enter Calibration Mode, set the fresh air (span) reading, set the zero reading,
and return to normal operation.
WARNING :The M2A is not an active gas monitoring device during th e calibration
procedure. The 4-20 mA output signal will “freeze” at 17.4 mA and all relays
will remain in their non-alarm state while the M2A is in Calibration Mode. The
output signal will not indicate current rea dings and the relays will not resume
operating normally until the M2A is in norma l operation again.
Prepari ng for Ca libration
NOTE:The following procedure assumes the use of a calibration kit which includes a zero air
(20.9% oxygen) calibration cylinder to set the fresh air reading, a 100% nitrogen (0%
oxygen) calibration gas cylinder to set the zero reading, a fixed flow regulator with an
on/off knob, a calibration cup, and a short piece of sample tubing to connect the
regulator to the calibration cup. A cylinder with up to an 18% oxygen concentration
may be used instead of a 100% nitrogen cylinder to set the zero reading.
NOTE:While in calibration mode, if there is no switch activity for the calibration time-out
period the unit will return to normal operation. See “Viewing & Changing M2A
Parameters” on page 31 for instructions to set the calibration time-out.
NOTE:The following procedure assumes that the oxygen concentration in the calibration area
is not normal, 20.9%, but is oxygen deficient. If a fresh air environment can be verified,
applying zero air to the detector is not necessary when adjusting the fresh air (span)
setting.
1.Connect the calibration kit sample tubing to the calibration cup hose barb.
2.Screw the calibration cup onto the detector.
3.Connect the fixed flow regulator to the calibration cup using the sample tubing.
Non-Intrusive Calibration
If the M2A is installed in a classified area and non-intrusive calibration is required, follow the
instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to
actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic
switch and perform the same operation as pressing a control switch button, touch the magnet in the
end of the magnetic wand to the M2A’s glass cover directly above the magnetic switch you wish to
actuate. Touching the glass and removing the wand is the same as pressing and releasing a button.
Touching the glass and keeping the wand in place is the same as pressing and holding a button.
Calibration Gas Response Memory Feature
The M2A has the capability to “remember” the detector’s response to the calibration gas after the
gas is removed from the detector during the fresh air or zero adjustment procedure. This feature
enables one person to perform a calibration if the detector is mounted remotely from the M2A.
When zero air is applied to the M2A during a fresh air adjustment, the M2A will freeze the display
45M2A Trans mitte r Op erato r’ s Manu al
reading at the highest response and the M2A will continue to display this reading and retain it in its
memory until the fresh air adjustment procedure is completed. When calibration gas is applied to
the M2A during a zero adjustment, the M2A will freeze the display reading at the lowest response
to the calibration gas. The calibration gas can then be removed and the M2A will continue to
display this reading and retain it in its memory until the zero adjustment p roced ure is completed.
Adjusting the Fresh Air Setting
1.While in normal operation, press and hold the UP/YES button for 5 seconds to enter
Calibration Mode. Release the button when the following screen appears.
Calib?
YES/NO
2.If you want to continue with calibration, press and release the UP/YES button to continue. The
display will indicate OXYGEN CAL Mode for a few seconds before showing FreshAir Adjust?.
If you want to exit Calibration Mode, press and release the DOWN/NO button. The M2A will
indicate Leaving CAL Mode and the M2A will return to normal operation.
3.To continue with ad jus ting the fresh air (span) reading, press and release the UP/YES button.
ENTER will alternate with FreshAir on the top display line and the current oxygen reading
will be on the bottom display line.
To skip adjusting the fresh air reading, press and release the DOWN/NO button. The display
will indicate ZERO w/Cal Gas?. Skip to the next section, “Adjusting the Zero Setting”.
4.Screw the zero air cylinder into the fixed flow regulator.
5.Turn the regulator knob counterclockwise to open the regulator.
6.Allow zero air to flow for 2 minutes. If the oxygen concentration in the area is less than 20.9,
the reading should stabilize after 2 minutes.
7.Turn the regulator knob clockwise to close the regulator.
8.Unscrew the zero air cylinder from the fixed flow regulator leaving the calibration cup on the
detector. The M2A will continue t o display th e maximum ga s response on the display and
retain the response level in its memory.
9.Press and release the ENTER button. The M2A will perform a fresh air adjustment and the
display will indicate ZERO w/Cal Gas?.
NOTE:If the fresh air adjustment fails, see “Troubleshooting” on page 39 for recommended
actions.
Adjusting the Zero Setting
1.If you want to continue with adjusting the zero setting, press and release the UP/YES button.
APPLY will alternate with ZERO Gas on the top display line and the current oxygen reading
will be on the bottom display line.
If you want to skip adjusting the zero reading, press and release the DOWN/NO button. The
display will indicate Leaving Cal Mode and the M2A will return to normal operation.
2.Screw the 100% nitrogen gas cylinder into the fixed flow regulator.
3.Turn the regulator knob counterclockwise to open the regulator.
4.Allow gas to flow for 2 minutes. The gas reading should be stable after 2 minutes.
M2A Transmitter Operator’s Manual46
5.Turn the regulator knob clockwise to close the regulator.
6.Unscrew the nitrogen cylinder from the fixed flow regulator and unscrew the calibration cup
from the detector. The M2A will continue to display the minimum gas response on the display
and retain the response level in its memory.
7.Adjust the oxygen reading up or down to match the 100% nitrogen cylinder oxygen
concentration of 0 by using the UP/YES and DOWN/NO buttons.
NOTE:If a cylinder other than a 100% nitrogen (0% oxygen) cylinder is used to set the zero
reading, such as a cylinder with 12% oxygen, set the zero reading to match the
cylinder’s oxygen concentration.
8.Press and release the ENTER button. The M2A will perform a zero operation. The display will
indicate ZERO Gas PASS for a few seconds, then indicate ZERO Gas SAVED before
indicating Leaving CAL Mode for a few seconds.
NOTE:If the zero adjustment fails, see “Troubleshooting” on page 39 for recommended
actions.
9.The display will now alternate between the normal operation screen and the message
REMOVE CAL GAS for 1 minute. If the calibration gas has not been removed from the
detector remove it now to avoid unwanted alarms.
During this 1 minute period, the signal output will remain fixed at 17.4 mA and the relays will
remain in their non-alarm state to avoid unwanted alarms wh ile the calibration gas clears fr om
the detector. At the end of the 1 minute period, the REMOVE CAL GAS message will stop
appearing and the M2A will return to normal operation.
10. Store the components of the calibration kit in a safe place.
Replacing Components of the M2A
Replacing the Gas Detector
NOTE:For M2A versions that do not have a replaceable plug-in sensor, or if you determine
that it is necessary to replace the entire detector, use the procedure in this section to
change the entire detector. For M2A versions that include a replaceable plug-in sensor,
see the appropriate instructions in the sections that follow this one to replace the plug-in
sensor.
1.Turn off or disconnect power to the M2A.
2.Remove th e junction bo x cover.
3.Grasp the control PCB by its edges.
4.Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull
too hard and damage the cable which connects the control and terminal PCBs.
5.Let the control PCB hang by the cable. The terminal strips are now visible on the terminal
PCB. The control PCB may be left hanging while wiring is done. If desired, the control PCB
may be disconnected from the cable and set aside while wiring.
6.Remove th e detector terminal stri p from its socket.
7.Disconnect the detector leads from the detector terminal strip. Note the position of the color-
47M2A Trans mitte r Op erato r’ s Manu al
coded leads as you remove them.
8.Unscrew the detector from the junction box.
9.Guide the replacement detector leads through the bottom con du it hub o f the junction box, then
screw the mounting threads of the detector into the conduit hub.
10. Connect the detector leads to the detector terminal strip as follows:
•Catalytic Combustible Gas Detector, IR Combustible Gas Detector, and IR CO
Detector
2
Red wire to terminal labeled LEL RED, white to terminal labeled LEL WHT, green wire
to terminal labeled LEL GRN, black wire to terminal labeled LEL BLK.
•CO or H2S gas detector.
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
•ESM-01 Gas Detector
Color coded wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
See Figure 25, “ESM-01 Toxic Detector Wiring” on page 21 for the wire color coding.
•Oxygen gas detector.
Green wire to terminal labeled OXY +, white wire to terminal labeled OXY -.
11. Re-install the detector terminal strip into its socket.
12. Re-install the control PCB (and ribbon cable if necessary). Be sure the ribbon cable is routed
down below the control PCB so it will not be damaged by the cover when it is screwed back
on.
13. Secure the junction box cover to the junction box.
14. Turn on or reconnect power to the M2A.
NOTE:Allow the replacement detector to warm up for 15 minutes before you continue with the
next step.
15. Calibrate the replacement detector as described in “Calibration, Combustible Gas, CO2, and
Toxic Versions” on page 42 or “Calibration, Oxygen Version” on page 45.
Replacing the CO or H2S Plug-in Sensor
NOTE:This procedure is applicable for both the CSA type, explosion proof CO and H2S
detectors and the non-explosion proof CO and H2S detectors.
1.Turn off or disconnect power to the M2A.
2.Unscrew the detector cap from the housing body. Make sure not to lose the cap gasket. See
Figure 4 - Figure 7 for illustr a tions of the CO or H
3.Unplug and remove the sensor.
4.If you are replacing a CO sensor, remove the rubber boot and charcoal filter from the old
sensor.
NOTE:RKI instruments recommends replacing the charcoal disk filter whenever a CO sensor
is replaced.
If you are replacing the 65-2423RK-05 explosion-proof H
S detectors.
2
S detector’s sensor, remove the
2
M2A Transmitter Operator’s Manual48
rubber boot and spacer from the old sensor.
5.If you are replacing a CO sensor, install the rubber boot and charcoal filter onto the new
sensor’s face.
If you are replacing the 65-2423RK-05 explosion-proof H
S detector’s sensor, install the
2
rubber boot and spacer on the new sensor’s face.
6.Carefully match the replacement sens or’ s male pi ns with the 4- socket patter n in the top section
and plug it in.
7.Make sure the cap gasket is in place and screw the detector cap back onto the housing body.
8.Turn on or recon nect power t o the M2A.
CAUTION:Allow the replacement sensor to warm up for 15 minutes before you continue with
the next step.
9.Calibrate the detector as described in “Calibration, Com bustible Gas, CO2, and Toxic
Versions” on page 42.
Replacing the CO Sensor Charcoal Filter
The charcoal filter is used to scrub out interfering gases from the environment being monitored,
such as H
CO M2A, the charcoal filter may be saturated and no longer scrubbing out interfering gases and it
may be necessary to change the charcoal filter.
1.Turn off or disconnect power to the M2A.
2.Unscrew the detector housing cap from the housing body. Make sure not to lose the cap gasket.
3.Remove the rubber retaining boot with the charcoal filter disk from the sensor.
4.Replace the charcoal filter disk in the rubber retaining boot and install the boot onto the sensor.
S and certain hydrocarbons. If you are experiencing unexplained upscale readings on a
2
See Figure 4 for an illustration of the CO detector.
5.Make sure that the cap gasket is in place and screw the detector cap back onto the housing
body.
6.Turn on or recon nect power t o the M2A.
Replacing an Oxygen Plug-in Sensor
NOTE:This procedure is applicable for both the CSA type, explosion proof oxygen detector
(65-2514RK) and non-explosion proof, replaceable sensor oxygen detector (652494RK).
1.Turn off or disconnect power to the M2A.
2.Unscrew the detector cap from the detector hou sing b ody. Take care not to lose the cap gasket.
See Figure 2 or Figure 3 for an illustration of the oxygen detect or.
3.Unplug and remove the sensor.
4.Carefully match the replacement sens or’ s male pi ns with the 2- socket patter n in the top section
and plug it in.
5.Make sure the cap gasket is in place and screw the detector cap back on to the detector
housing.
6.Turn on or recon nect power t o the M2A.
49M2A Trans mitte r Op erato r’ s Manu al
NOTE:Allow the replacement sensor to warm up for 15 minutes before you continue with the
next step.
7.Calibrate the detector as described in “Calibration, Com bustible Gas, CO2, and Toxic
Versions” on page 42.
Replacing an ESM-01 Plug-in Sensor
1.Turn off or disconnect power to the M2A.
2.Unscrew the detector cap from the detector hou sing b ody. Take care not to lose the cap gasket.
If the splashguard begins to unscrew from the detector cap, make sure to remove the detector
cap also. See Figure 8 for an illustration of the ESM-01 detector.
3.Unplug and remove the sensor.
4.Carefully match the replacement sensor’s plug with the socket in the housing body and plug it
in.
W ARNING:The ESM-01 detectors cannot be changed from one type to another. Replace the
plug-in sens or only with th e same type of E S M-01 sensor.
5.Make sure the cap gasket is in place and screw the detector cap back onto the housing body. If
the splash guard was unscrewed from the detector cap, screw it back onto the detector cap.
6.Turn on or recon nect power t o the M2A.
NOTE:Allow the replacement sensor to warm up for 15 minutes before you continue with the
next step.
7.Calibrate the detector as described in “Calibration, Com bustible Gas, CO2, and Toxic
Versions” on page 42
M2A Transmitter Operator’s Manual50
Chapter 8: RS-485 Modbus Output
Overview
This chapter describes the M2A’s RS-485 Modbus output and how to configure the M2A to make
use of it. It also discusses how to wire the M2A into a Modbus system.
The M2A provides an RS-485 serial communications interface. It is a Modbus Slave Device,
supporting 2-wire RS-485 Modbus RTU serial communications.
Wiring the M2A in a Modbus System
The M2A is a 2-wire Modbus RTU device. When wiring the M2A into a Modbus system, adhere to
standard Modbus wiring practices per the Modbus Over Serial Line Specification and
Implementation Guide V1.0 or later. This document can be found online at www.modbus.org/
specs.php.
The Modbus protocol supports a maxi mum of 247 uni qu e slave add ress es (1-2 47). The M2A ’s line
driver provides for up to 128 M2As to be connected together without the need for a repeater.
Figure 29 & Figure 30 below illustrate typical M2A wiring configurations. If more than 128 M2As
need to be connected together, RS-485 repeater(s) should be used such that no more than 128
M2As reside on any given ne twork segment.
CAUTION:The network segment can only support 128 units if RKI M2As are being used. The
use of any other instruments will lower the number of units that may reside on any
given network segment.
The MODBUS terminal strip is located on the terminal PCB (see Figure 9). The following signals
are available at the Modbus terminal strip:
Table 10 : Modbus Termi nal Strip Signals
Modbus Terminal Label (RS-485
Name)
AD0
BD1
CCommon
Modbus Signal Name
Ground Jumper
The M2A includes a 2-pin ground header (see Figure 9) that is used when the M2A is used in a
Modbus system. Every M2A is supplied with a ground jumper (a jumper block) installed onto this
header. If the M2A isn’t wired into a Modbus system or if it is wired into one as shown in
Figure 28, the ground jumper should be installed. If the M2A is wired in a Modbus system as
shown in Figure 27, the ground jumper should be removed.
51M2A Trans mitte r Op erato r’ s Manu al
Recommended Modbus Wiring
l
l
The recommended Modbus wiring for the M2A is illustrated in Figure 27 below. In this
configuration, 5 wires are used for wiring the M2A into a Modbus system. Figure 27 also illustrates
typical alarm device wiring. The ground jumper must be removed in this wiring configuration.
Power Su pply
10-30 VDC
Remove Ground Jumper
Input
Terminals
Modbus Controller
+
-
Comm on
D1
D0
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
See
Detector
Wiring
- S +
LEL
R W G B
RS 485PW R/SIG
A B C
Figure 27: Recommended Modbus Wiring
See
Detector
Wiring
- S +
LEL
R W G B
RS 485PW R/SIG
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
To
Additiona
M2As
Power Supply
10-30 VDC
Input
Terminals
Adding an M2A to an Existing M2 Modbus Installation with Recommended Wiring
An M2A can be wired into an existing M2 Modbus system. The shape of the terminal PCB will be
different but the wiring connections are identical and the ground jumper on the M2A must still be
removed.
Alternate Modbus Wiring For Existing Installations
Although the wiring shown in Figure 27 is recommended, it is possible to wire the M2A into a
Modbus system with only 4 wires in situations where a pre-existing system is being replaced and
wiring is already in place. This should only be done if wiring for a system that is being replaced is
already installed and it is not practical to run another wire. See Figure 28 below for this wiring
configuration. In this configuration, the wire between the “C” terminal on the M2A and the
Common terminal on the Modbus controller is omitted. Instead, the Common terminal on the
Modbus controller is connected to the “-” of the 24 VDC power supply. The ground jumper mus t be
installed in this wiring configuration.
Comm on
D1
D0
+
-
See
Detector
Wiring
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
FAIL
C NC NO
See
ALARM 1
C NC NO
ALARM 2
C NC NO
Detector
Wiring
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
To
Additiona
M2As
Modbus Controller
Figure 28: Alternate Modbus Wiring
M2A Transmitter Operator’s Manual52
5
ID = 128
M2A
Termination
Jumper
Installed
Adding an M2A to an Existing M2 Modbus Installation with Alternate Wiring
An M2A can be wired into an existing M2 Modbus system. The shape of the terminal PCB will be
different but the wiring connections are identical and the ground jumper on the M2A must still
remain installed.
Termination Jumper
The M2A includes a 2-pin termination header (see Figure 9) that is used when the M2A is used in a
Modbus system. Every M2A is supplied with a termination jumper (a jumper block) installed onto
this header. If the M2A is not used in a Modbus system, this jumper has no function. When the
M2A is installed in a Modbus system, this jumper must be installed in an M2A that is at the end of
a Modbus line. Any M2A in a Modbus system that is not at the end of a line must have the
termination jumper removed (see Figure 29 & Figure 30 below).
Modbus
Master
RS-485RS-48
M2A
Termination
Jumper
Not Installed
Up to 12 8 M2As canbe connected without a repeater
M2A
Termination
Jumper
Not Installed
ID = 126ID = 127
Figure 29: Multiple M2As in a Daisy Chain Configuration
M2A
Termination
Jumper
Not Installed
ID = 125
M2A
Termination
Jumper
Not Installed
ID = 1
53M2A Trans mitte r Op erato r’ s Manu al
ID = 1
M2A
Termination
Jumper
Installed
RS-485
M2A
Termination
Jumper
Not Install ed
Termination
Not Installed
ID = 2ID = 3
Up to 128 M2As can be connected without a repeater
Figure 30: Multiple M2As in a Two Branch Configurati on
M2A
Jumper
Modbus
Master
M2A
Termination
Jumper
Not Installed
ID = 4
ID = 128
M2A
Termination
Jumper
Installed
RS-485
M2A
Termination
Jumper
Not Install ed
ID = 127
Using the M2A in a 4-wire Modbus System
Although the M2A is a 2-wire Modbus R TU device, it can be used with a 4-wire Modbus controller
if the system wiring is modified as follows:
•Connect the controller’s TxD0 and RxD0 wires together and use this connection as the 2-wire
Modbus D0 signal.
•Connect the controller’s TxD1 and RxD1 wires together and use this connection as the 2-wire
Modbus D1 signal.
•Connect these D0 and D1 signals and the common wire from the controller to the 2-wire
Modbus Network of M2As.
Modbus Mode
Modbus Mode allows you to configure various parameters relating to the Modbus setup of the
M2A. Modbus Mode includes a 5-minute time-out feature. If you do not press a control button for
5 minutes, the M2A automatically returns to normal operation.
NOTE:If the M2A returns to normal operation because of a time-out, it enters a warm-up
period just as it does when it is first turned on.
1.While in normal operation, simultaneously press and hold the DOWN/NO and ENTER buttons
M2A Transmitter Operator’s Manual54
for 5 seconds to enter Modbus Mode. Release the buttons when the following screen appears.
Set Up
Modbus?
2.To exit Modbus Mode, press and release the DOWN/NO button. The M2A will return to
normal operation.
To continue in Modbus Mode, press and release the UP/YES button. The current ENABLED/
DISABLED setting will appear.
3. If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO
buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter setting is OK, press the ENTER button to proceed to the
next parameter.
T able 11 lists the M2A Modb us Mode par ameters you can set. Table 11 also lists the fact ory set
value for each parameter.
Enabled/DisabledCan be set to ENABLED or DISABLED (factory setting).
Enables or disables the Modbus output .
Slave IDThe Slave ID can be set to values from 1 (factory setting) to 247.
The M2A will only receive messages from t he M aster which are addressed to
this Slave ID (except for broadcast messages whic h are received by all slaves).
Baud RateCan be set to 1200, 2400, 4800, 9600 (factor y setting), 14440, & 19200.
ParityEVEN (factory setting), ODD, NONE
Response DelayThe response delay can be set from 0 (factory setting) to 20 mS.
This is an optional additional delay inserted by the M2A prior to returning a
response message to the Master. It may be helpful in some installations where
the Master’s preparedness to receive responses might be delayed.
4.When you have scrolled through all the adjustable parameters, SA VE IT? YES/NO appears on
the display.
5.If you do not wish to save the adjustments, press and release the DOWN/NO button. The DO
OVER? YES/NO message will display. Press and release the DOWN/NO button. The
ABORT? YES/NO message will display. Press the UP/Y ES button to return to normal
operation.
If you wish to change some of the adjustments made, pr ess and release the DOWN/NO button.
The DO OVER? YES/NO message will display. Press and release the UP/YES button. The
Re-do MB Setup message will display and the M2A will return to the first adjustable
parameter. Go back to step 3 and continue.
If you want to save the adjustments made, press and release the UP/YES button. Modbus
Saved is indicated on the display for a few seconds and the M2A returns to normal operation.
NOTE:All M2As on a Modbus network must be configured w ith the same baud r ate and parity,
and each M2A must have a unique slave ID. For an M2A to participate on a network,
the ENABLED/DISABLED setting must be set to ENABLE.
M2A Transmitter Operator’s Manual56
Supported Modbus Functions
The M2A supports Function Code 03: Read Holding Registers and Function Code 16: Write
Registers.
Function Code 03: Read Holding Registers
There are a total of 40 registers available to be read via this Function Code.
Registers 1 - 2
The format (bit & field assignments) for registers 1 - 2 is shown below in Table 12 - Table 13.
Table 12: Register 1, Numerical Reading with Sign & Decimal Point Bit & Field Assignments
Bit(s)Value & Field
[15]Sign of Reading
0 = Positive
1 = Negative
[14:13]Format of Reading
0 = XXX (no decimal point)
1 = XX.X
2 = X.XX
3 = .XXX
[12:0]Numerical value of display reading (8000 max)
Table 13 : Register 2, Operating State, Alarms & Relay s B it & Field Assignments
Bit(s)Value
[15:12]Operating State
0 = Undefined
1 = Initialization
2 = Normal
3 = Warm-up
4 = Fail
5 = Calibration
6 = Post-Calibration
7 = Low Power
8 = Setup
[11]Gas Type Change
0=Not Changed
1=Changed
[10]Configuration Change
0=Not Changed
1=Changed
[9]Calibration Activity Flag
0=No Calibration Activity
1=Calibra tion Activ ity Has Occu rred
[8]Fail Status
0 = Fail Not Asserted
1 = Fail Asserted
57M2A Trans mitte r Op erato r’ s Manu al
Table 1 3: Register 2, Operating State, Alarms & Relays Bit & Field Assignments ( Continued)
Bit(s)Value
[7:6]Alarm 2 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Unused Code
[5:4]Alarm 1 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Unused Code
[3]Overscale Status
0 = Not Overscale
1 = Overscale
[2]Fail Relay Status
0 = Not Energized
1 = Energized
[1]Alarm 2 Relay Status
0 = Not Energized
1 = Energized
[0]Alarm 1 Relay Status
0 = Not Energized
1 = Energized
Register 3
Register 3 is the supply voltage (0.1 volt per count).
Registers 4 - 11
The 16 bytes in these registers contain the same ASCII text information that is displayed on the
M2A’s OLED (not NULL Terminated).
Register 12
Register 12 is the range (full scale readout).
Register 13
Register 13 is the alarm 1 set point.
The decimal point location is the same as specified in Register 1.
Register 14
Register 14 is the alarm 2 set point.
The decimal point location is the same as specified in Register 1.
Registers 15 - 19
Registers 15 - 19 are the Gas Name ASCII String (NULL Terminated).
Registers 20 - 22
Registers 20 - 22 are the Gas Units String (NULL Terminated).
M2A Transmitter Operator’s Manual58
Register 23
Register 23 is the alarm 1 trigger.
0=Decreasing
1=Increasing
Register 24
Register 24 is the alarm 1 normal relay state.
0=Norma lly De-Ener g i zed
1=Normally Energized
Register 25
Register 25 is the alarm 1 relay reset.
0=Latching
1=Self-Resetting
Register 26
Register 26 is the alarm 1 ON delay.
Integer values are in seconds.
Register 27
Register 27 is the alarm 1 OFF delay.
Integer values are in seconds.
Register 28
Register 28 is the alarm 2 trigger.
0=Decreasing
1=Increasing
Register 29
Register 29 is the alarm 2 normal relay state.
0=Norma lly De-Ener g i zed
1=Normally Energized
Register 30
Register 30 is the alarm 2 relay reset.
0=Latching
1=Self-Resetting
Register 31
Register 31 is the alarm 2 ON delay.
Integer values are in seconds.
Register 32
Register 32 is the alarm 2 OFF delay.
Integer values are in seconds.
Register 33
Register 33 is zero suppression.
Toxics and LEL: 0%-6% of full scale
Oxygen: 0%-0.7% Oxygen
The decimal point location is the same as specified in Register 1.
59M2A Trans mitte r Op erato r’ s Manu al
Register 34
Register 34 is noise filter.
Integer values in seconds.
0-60 in 5 second increments.
Register 35
Register 35 is calibration timeout in minutes (5, 10, 15, 20, 25, or 30).
Register 36
Register 36 is calibration month (upper byte) and day (lower byte).
This is zero based (e.g. January=0 and the 1st=0).
Register 37
Register 37 is the calibration year.
Register 38
Register 38 is the calibration outcome.
Upper=Gas Operation
Lower=Air Operation
Calibration Outcome Codes:
1=Sensor Missing
2=Sensor Too Weak
3=Signal Too Strong
4=Not Enough Span
5=Negative Reading
6=Span Gas Value Is Not OK
7=Calculated Zero Is Too Low
8=Calculated Zero Is Too High
9=Zero Failure
10=Span Failure
11=Calibration Is Overscale
12=LEL Sensor Offset Is Too High
13=LEL Null Operation Failed
14=Operation Not Performed
Register 39
Register 39 is the auto-zero disable.
0=AutoZero Enabled
1=AutoZero Disabled
Register 40
Register 40 is the remote configuration register access level
0=None
1=Alarm Rese t
M2A Transmitter Operator’s Manual60
Function Code 16: Write Registers
There are 21 registers in Function Code 16 that can be used to write to the M2A. This manual only
describes Register 16 because it can be used to reset an alarm condition. For a complete description
of Function Code 16, request Appendix C from RKI Instruments, Inc.
Register 16
Register 16 performs the same function as the ENTER button on the M2A. It is useful for remotely
silencing alarms. With the M2A in an alarm condition, bit [0] can be used to silence it.
Table 14: Register 16, Button Function
NOTE:Register 40 in Function Code 03, the remote configuration register, must be set to 1 in
order for Register 16 to work.
BitValue & Field
[0]ENTER (used for “Alarm Reset”)
0=Not Pressed
1=Pressed
Parts List
Table 15 lists replacement parts and accessories for the M2A Transmitter.
Table 15: Parts List
Part NumberDescription
06-1248RKCalibration kit sample tubing (3/16 in. x 5/16 i n.; specify length when ordering)
07-0033RKCap gasket, for CSA type CO and H
07-0125RKCap gasket, for ESM-01 detector
07-0203RKRubber retaining boot for charcoal filter and for 65-2423RK-05 H2S detector
14-2101RKSpacer for 65-2423RK-05 H
18-0431RK-01Junction box
33-7101RKCharcoal filter disk
45-6132RKTermination and group jumper
61-0140RKLEL detector, catalytic type, UL version
61-0140RK-05LEL detector, catalytic type, CSA version
61-0190RK-CH4CH
LEL detector, IR type
4
2
2
S detector
S detector
61-0190RK-HCHC LEL detector, IR type, propane calibration
61-0191RK-02CO
61-0191RK-03CO
61-0191RK-05CO
61M2A Trans mitte r Op erato r’ s Manu al
detector, IR type, 0 - 5,00 0 ppm
2
detector, IR type, 0 - 5% volume
2
detector, IR type, 0 - 50.0% volume
2
Table 15: Parts List
Part NumberDescription
61-0191RK-10CO
61-0192RK-CH4CH
61-0198RK-02CO
61-0198RK-03CO
61-0198RK-05CO
61-0198RK-10CO
detector, IR type, 0 - 100% volume
2
detector, IR type, 0 - 100% volume
4
detector, IR type, replaceable sensor, 0 - 5000 ppm
2
detector, IR type, replaceable sensor, 0 - 5% volume
2
detector, IR type, replaceable sensor, 0 - 50.0% volume
2
detector, IR type, replaceable sensor, 0 - 100% volume
65-2433RK-05CO detector assembly, CSA version
65-2494RKOxygen detector assembly, non-explosion proof, r e placeable sensor
65-2495RKH2S detector assembly, 0 - 100 ppm, non-explosion proof, replaceable sensor
65-2496RKCO detector assembly, 0 - 300 ppm, non-explosion proof, replaceable sensor
65-2510RKOxygen detector, partial pressure type
65-2511RKOxygen detector, capillary type
65-2514RKOxygen detector , CSA version , capilla ry type
71-0305RKM2A Transmitter Operator’s Manual (this document)
81-0002RK-01Calibration cylinder, 50% LEL hydrogen in air, 34 liter steel
81-0002RK-03Calibration cylinder, 50% LEL hydrogen in air, 103 liter steel
81-0004RK-01Ca libration cylinder, 50% LEL propane in air, 34 liter steel
81-0004RK-03Calibration cyli nder, 50% LEL propane in air, 103 liter steel
81-0007RK-01Calibration cylinder, 15% LEL hexane in air, 34 liter steel
81-0012RK-01Ca libration cylinder, 50% LEL methane in air, 34 liter steel
81-0012RK-03Ca libration cylinder, 50% LEL methane in air, 103 liter steel
81-0013RK-01Calibration cylinder, 50% volume methane in nitrogen, 34 liter steel
M2A Transmitter Operator’s Manual62
Table 15: Parts List
Part NumberDescription
81-0013RK-05Calibration cylinder, 50% volume methane in nitrogen, 58 liter steel
81-0064RK-01Calibration cylinder, CO, 50 ppm in air, 34 liter steel
81-0064RK-03Calibration cylinder, CO, 50 ppm in air, 103 liter steel
81-0070RK-01Calibration cylinder , CO
81-0070RK-03Calibration cylinder , CO
81-0072RK-01Calibration cylinder , CO
81-0072RK-03Calibration cylinder , CO
81-0073RK-01Calibration cylinder , CO
81-0073RK-03Calibration cylinder , CO
, 2,000 ppm in nitrogen, 34 liter steel
2
, 2,000 ppm in nitrogen, 103 liter steel
2
, 2.5% in nitrogen, 34 liter st ee l
2
, 2.5% in nitrogen, 103 liter steel
2
, 15% in nitrogen, 34 liter steel
2
, 15% in nitrogen, 103 liter steel
2
81-0076RK-01Zero air calibration cylinder, 34 liter steel
81-0076RK-03Zero air calibration cylinder, 103 liter steel
81-0078RK-01Calibration cylinder, 100% nitrogen, 34 liter steel (used for setting oxygen zero)
81-0078RK-03Calibration cyl inder, 100% nitrogen, 103 liter steel (used for setti ng oxygen zero)
81-0151RK-02Calibration cylinder , H
81-0151RK-04Calibration cylinder , H
81-0170RK-02Calibration cylinder , 5 ppm SO
81-0170RK-04Calibration cylinder , 5 ppm SO
81-0175RK-02Calibration cylinder, 10 ppm NH
S, 25 ppm in nitrogen, 58 liter aluminum
2
S, 25 ppm in nitrogen, 34 liter aluminum
2
in nitrogen, 58 liter
2
in nitrogen, 34 liter
2
in nitrogen, 58 liter
3
81-0175RK-04Calibration cylinder, 10 ppm NH
in nitrogen, 34 liter
3
81-0185RK-02Calibration cylinder , 0.5 ppm PH3 in nitrogen, 58 liter
81-0185RK-04Calibration cylinder , 0.5 ppm PH3 in nitrogen, 34 liter
81-0190RK-02Calibration cylinder, 5 ppm Cl2 in nitrogen, 58 liter
81-0190RK-04Calibration cylinder, 5 ppm Cl
81-0192RK-02Calibration cylinder, 2 ppm Cl
81-0192RK-04Calibration cylinder, 2 ppm Cl
in nitrogen, 34 liter
2
in nitrogen, 58 liter
2
in nitrogen, 34 liter
2
81-1050RKRegulator, with gauge and knob, 0.5 liter/minute continuous flow, for combustible
gas, nitrogen, zero air, and CO cylinders.
81-1051RKRegulator, with gau ge an d k n ob , 0.5 lite r/m in ut e con tinu o us flo w, for H2S cylinder
81-1051RK-25Regulator, fixed flow, w/gauge & knob, 0.25 LPM, for 34 liter aluminum/58 liter/
103 liter cylinder (Used to calibrate NH3 ESM-01 M2A only)
81-1103RKCalibration c up, for 3/4” NPT IR detectors, combus tible gas and CO
2
63M2A Trans mitte r Op erato r’ s Manu al
Table 15: Parts List
Part NumberDescription
81-1117RKCalibration cup, for catalytic LEL, CO, H2S, and oxygen detectors
81-1138RKCalibration cup, general ( SO
81-1138RK-CL2Calibrati on cup, for ESM-01 Cl2, use 0.5 LPM flow
81-1138RK-NH3Calibration cup, for ESM-01 NH3, use 0.25 LPM flow
81-1190RKSplashguard, for ESM-01 detector cap
ES-1531-COCO sensor, plug -i n
ES-1537-H2SH2S sensor, plug-in
ESM-015-NOESM-01 plug-in sensor, 0 - 100 ppm nitric oxide
ESM-01DH-ASH3ESM-01 plug-in sensor, 0 - 1.50 ppm arsine
ESM-01DH-D-SO2ESM-01 plug-in sensor, 0 - 6.00 ppm sulphur dioxide, diffusion type only
ESM-0DH-PH3ESM-01 plug- in sensor, 0 - 1.00 ppm phosphine
ESM-01R-D-NH3ESM-01 plug-in sensor, 0 - 75.0 ppm ammonia, diffusion type only
ESM-K01-D-CL2ESM-01 plug-in sensor, 0 - 3.00 ppm chlorine, diffus ion type only
ESM-K01D-CL2-10ESM-01 plug-in sensor, 0 - 10.0 ppm chlorine, diffusion type only
NC-6205-01Hydrogen specific LEL detector, catalytic type, UL version
NC-6205-05Hydrogen specific LEL detector, catalytic type, CSA version
), for ESM-01, use 0.5 LPM flow
2
M2A Transmitter Operator’s Manual64
Appendix A: Control Button Quick Reference Guide
The M2A’s control buttons allow access to operational modes, resetting of alarms, and display of
the Information Screen. Table 16 shows which button combinations perform these functions and
which parameters are available for adjustment while in the operational modes. While in these
modes, display prompts showing a “?” require you to respond by pressing either the UP/YES (for
yes) or DOWN/NO (for no) button. To change parameter settings, u se the UP/YES an d DOWN/NO
buttons to get the desired setting, then press and release the ENTER button to accept the setting.
Table 16: Control Button Quick Reference Chart
Mode Entered or
Function Performed
Alarm ResetP ress and Release
Show Information ScreenPress and Hold
Calibration ModePress and Hold UP/
Configuration ModePress and Hold UP/
Gas Type ModePress and Hold UP/
Button
Combination
ENTER
DOWN/NO
YES for 5 seconds
YES & ENTER for 5
seconds
YES and DOWN/
NO for 5 seconds
Adjustable or Viewable Parameters
n/a
•Input Voltage
•M2A Version/Revision Information
•Fresh Air Adjust
•Span (or Zero for Oxygen) Adjust with Calibration
Gas
•Alarm Levels
•Increasing or Decreasing Alarms
•Normally Energized or Normally De-energized
Relays
•Latching or Self-resetting Alarms
•Alarm On Delay
•Alarm Off Delay (if self-resetting)
•Zero Suppression
•Noise Filter
•Calibration Time-out Period
Gas Type
Modbus ModePress and Hold
DOWN/NO and
ENTER for 5
seconds
65M2A Trans mitte r Op erato r’ s Manu al
•Enable or Disable Modbus
•Slave ID
•Baud Rate
•Parity
•Response Delay
Appendix B: PLC and DCS Device Wiring
The M2A can be wired to a PLC or DCS device if desired.
1.Guide multi conductor shielded cable or cables or wires in conduit through the top conduit hub
of the junction box. The number of cables or wires needed will depend on whether any relays
are used and whether the Modbus output is used. Use the following recommendations to
determine how to wire the M2A:
•If Modbus connections will not be used and only the PWR/SIG terminal stri p conn ect ions
will be used, use four conductor shielded cable or four wires in conduit for connections to
the power/signal terminal strip.
Table 17: Wire Size for PWR/SIG Connections
Max Distance to Controller
w/18 Gauge Wire
2,500 ft.5,000 ft.8,000 ft.
•If the PWR/SIG terminal strip connections and one or more relays are used, route the
connections to the M2A in conduit. Use shielded cable in the conduit for the PWR/SIG
connections and unshielded cable or individual wi res for the relay conn ections . Make su re
any wire or cable used for relay wiring is appropriately rated for the power that it will
carry.
NOTE:If shielded cable is used for the PWR/SIG connections, leave the cable shield’s drain
wire insulated and disconnected at the M2A. You will connect the opposite end of the
cable’s drain wire at the controller or device.
•If the M2A will be wired into a Modbus network, see “Chapter 8: RS-485 Modbus
Output” on page 51.
Max Distance to Controller w/
16 Gauge Wire
Max Distance to Controller
w/14 Gauge Wire
M2A Transmitter Operator’s Manual66
e
24 VDC
e
Power Supply
24 V DC +
24 VDC -
See Figure 31 below for field wiring connections to the M2A.
Alarm Device
Power
4-20mA In (S)
(24 VD C) -
Controller
(PLC, DCS)
- S +
FAIL
C NC NO
See
Detector
Wiring
See
Modbus
Wiring
LEL
R W G B
RS 485PWR/SIG
A B C
C NC NO
C NC NO
ALARM 1
ALARM 2
Figure 31: PLC and DCS Device Wiring
Fail Alarm
Device
Alarm 1
Alarm Devic
Alarm 2
Alarm Devic
Typical Alarm
WiringShown
67M2A Trans mitte r Op erato r’ s Manu al
Appendix C: Function Code 16 Registers
The M2A suppo rts Function Code 16 that allows writing to the M2A. There are 21 registers in this
Function Code.
Register 1
Register 1 is the alarm 1 set point.
Value must be an integer. Any decimals are omitted.
Register 2
Register 2 is the alarm 1 trigger.
0=Decreasing
1=Increasing
Register 3
Register 3 is the alarm 1 relay state.
0=Norma lly De-Ener g i zed
1=Normally Energized
Register 4
Register 4 is the alarm 1 relay reset.
0=Latching
1=Self-Resetting
Register 5
Register 5 is the alarm 1 ON delay.
Integer value is in seconds.
Register 6
Register 6 is the alarm 1 OFF delay.
Integer value is in seconds.
Register 7
Register 7 is the alarm 2 set point.
Value must be an integer. Any decimals are omitted.
The decimal point location is the same as specified in Register 1 of Function Code 03.
Register 8
Register 8 is the alarm 2 trigger.
0=Decreasing
1=Increasing
Register 9
Register 9 is the alarm 2 relay state.
0=Norma lly De-Ener g i zed
1=Normally Energized
M2A Transmitter Operator’s Manual68
Register 10
Register 10 is the alarm 2 relay reset.
0=Latching
1=Self-Resetting
Register 11
Register 11 is the alarm 2 ON delay.
Integer value is in seconds.
Register 12
Register 12 is the alarm 2 OFF delay.
Integer value is in seconds.
Register 13
Register 13 is the zero suppression.
Toxics and LEL: 0%-6% Full Scale
Oxygen: 0%-0.7% Oxygen
Value must be an integer. Any decimals are omitted.
Register 14
Register 14 is the noise filter.
Integer value is in seconds (0-60 in 5 second increments).
Register 15
Register 15 is the calibration timeout.
Value is in minutes (5, 10, 15, 20, 25, or 30).
Register 16
Register 16 is explained in “Supported Modbus Functions” in the M2A manual.
Register 17
Register 17 is the calibration month (MSB) and day (LSB).
This is zero-based (e.g. January=0 and the 1st=0).
Register 18
Register 18 is the calibration year.
Register 19
Register 19 is for clearing change flags. When parameters are changed at the M2A, a flag is raised
at the controller. This register can be used to clear those flags.
Table 18: Register 19, Clear Change Flags
Bit(s)Value & Field
[15:2]Unused
[1]Write “1” to this bit to clear “Gas Type Changed” flag
[0]Write “1” to this bit to clear “Configuration Changed” flag
69M2A Trans mitte r Op erato r’ s Manu al
Register 20
Register 20 is the AutoZero Disable.
0=AutoZero Enabled
1=AutoZero Disabled
Register 21
Register 21 is the Remote Configuration Register.
0=No Remote Configuration
1=Reset Alarm
M2A Transmitter Operator’s Manual70
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.