Rice Lake Discovery Series Semi User Manual

Ohaus Discovery Balances Instruction Manual
Ohaus Corporation, 19A Chapin Road, P.O. Box 2033 Pine Brook, New Jersey, 07058, USA
Declaration of Conformity We, Ohaus Corporation, declare under our sole responsibility that the balance models listed below marked with “CE”
- are in conformity with the directives and standards mentioned.
EC Marking: EC Directive Applicable Standards
Low Voltage
EN61010-1: 2001
89/336/EEC
Electromagnetic compatibility
For non-automatic weighing instruments used in an Article 1, 2.(a) application, additional metrological marking according to Annex IV of Council directive 90/384/EEC must be attached to the instrument
EN45501:1992 *
Ted Xia
President
Ohaus Corporation
Pine Brook, NJ USA
Urs Müller
General Manager
Ohaus Europe
Greifensee, Switzerland
* Applies only to certified non-automatic weighing instruments. Date: April 1, 2006
Instrument Type: Discovery Series
EN61326:1997+A1:A2:01+A3:03 (class B) EN61326:1997+A1:A2:01+A3:03 (minimal requirements)
90/384/EEC
Non-Automatic Weighing Instruments
Compliance to the following standards is indicated by the corresponding marking on the product.
Mark Standard
AS/NZS CISPR 22, AS/NZS 61000.4.3
CAN/CSA C22.2 No. 61010-1-04 UL Std No 61010A-1
FCC Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environ­ment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-003. Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May 15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
N13123
MC173467
Disposal
In conformance with the European Directive 2002/96/ EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
Weighing Instruments verified at the place of manufacture bear one of the preceding mark on the packing label and the green ‘M’ (metrology) sticker on the descriptive plate. They may be put into service immediately.
Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive plate and bear one of the preceding identification mark on the packing label. The second stage of the initial verification must be carried out by the approved service organization of the authorized representative within the EC or by the national weight & measures (W+M) authorities.
The first stage of the initial verification has been carried out at the manufacturers work. It comprises all tests according to the adopted European standard EN 45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the respective W+M authorities.
Important notice for verified weighing instruments
EN-1Ohaus Discovery
1. INTRODUCTION ............................................................................................................................................................... EN-3
1.1 Description ..................................................................................................................................................................... EN-3
1.2 Features ......................................................................................................................................................................... EN-3
1.3 Safety Precautions ........................................................................................................................................................... EN-3.
2. INSTALLATION ................................................................................................................................................................. EN-4
2.1 Unpacking ...................................................................................................................................................................... EN-4
2.2 Installing Components ..................................................................................................................................................... EN-4
2.2.1 Assembly ......................................................................................................................................................... EN-4
2.3 Selecting the location ....................................................................................................................................................... EN-4
2.4 Leveling the Balance ....................................................................................................................................................... EN-4
2.5 Connecting Power .......................................................................................................................................................... EN-5
2.5.1 AC Adapter ......................................................................................................................................................... EN-5
2.5.2 Turning Power On and Off .................................................................................................................................... EN-5
2.6 Initial Calibration ............................................................................................................................................................. EN-5
3. OPERATION ....................................................................................................................................................................... EN-7
3.1 Overview of Controls & Display Functions .......................................................................................................................... EN-7
3.2 Button Control Functions .................................................................................................................................................. EN-9
3.3 Using the Button Control Functions .................................................................................................................................. EN-10
3.3.1 Setting the Balance to Zero .............................................................................................................................. EN-10
3.3.2 Taring ............................................................................................................................................................. EN-10
3.3.3 Changing Units of Measure ............................................................................................................................... EN-10
3.3.4 Changing Application Modes ............................................................................................................................. EN-10
3.3.5 Printing Data ................................................................................................................................................... EN-10
3.3.6 Calibration ...................................................................................................................................................... EN-10
3.4 Menu ..................................................................................................................................................................... EN-11
3.4.1 Menu Structure ................................................................................................................................................. EN-11
3.4.2 Menu Navigation ............................................................................................................................................. EN-12
3.4.3 Changing Settings ........................................................................................................................................... EN-12
3.5 Application Modes ......................................................................................................................................................... EN-13
3.5.1 Weighing ........................................................................................................................................................ EN-13
3.5.2 Parts Counting ................................................................................................................................................ EN-13
3.5.3 Percent Weighing ............................................................................................................................................ EN-14
3.5.4 Check Weighing .............................................................................................................................................. EN-15
3.5.5 Dynamic/Animal Weighing ............................................................................................................................... EN-15
3.5.6 Gross, Net, Tare Weighing ................................................................................................................................ EN-16
3.5.7 Totalize .......................................................................................................................................................... EN-17
3.5.8 High Point ...................................................................................................................................................... EN-17
3.5.9 Density ........................................................................................................................................................... EN-18
3.5.10 Statistics ......................................................................................................................................................... EN-25
3.3.11 Pipette Calibration ........................................................................................................................................... EN-28
TABLE OF CONTENTS
EN-2
Ohaus Discovery
3.6 Additional Features ........................................................................................................................................................ EN-30
3.6.1 Weigh Below .................................................................................................................................................... EN-30
3.7 Balance Settings ........................................................................................................................................................... EN-30
3.7.1 Menu Setup .................................................................................................................................................... EN-30
3.7.2 Calibration ...................................................................................................................................................... EN-31
3.7.3 Setup ............................................................................................................................................................. EN-35
3.7.4 Readout ......................................................................................................................................................... EN-35
3.7.5 Mode ............................................................................................................................................................. EN-35
3.7.6 Unit ................................................................................................................................................................ EN-36
3.7.7 Print ............................................................................................................................................................... EN-37
3.7.8 RS232 ........................................................................................................................................................... EN-37
3.7.9 GLP Data ........................................................................................................................................................ EN-37
3.7.10 GLP Print ........................................................................................................................................................ EN-38
3.7.11 Reset ............................................................................................................................................................. EN-38
3.7.12 Lockout .......................................................................................................................................................... EN-39
3.7.13 End ................................................................................................................................................................ EN-39
3.8 Sealing Access to the Balance Settings ............................................................................................................................ EN-40
3.9 Printing Data ................................................................................................................................................................ EN-40
4. MAINTENANCE ................................................................................................................................................................. EN-41
4.1 Calibration ................................................................................................................................................................... EN-41
4.2 Cleaning ..................................................................................................................................................................... EN-41
4.3 Troubleshooting ............................................................................................................................................................ EN-41
4.4 Service Information ........................................................................................................................................................ EN-42
4.5 Parts ..................................................................................................................................................................... EN-42
4.6 Accessories .................................................................................................................................................................. EN-42
5. TECHNICAL DATA ............................................................................................................................................................... EN-43
5.1 Drawings ..................................................................................................................................................................... EN-43
5.2 Specifications ............................................................................................................................................................... EN-44
5.3 Capacity x Readability ................................................................................................................................................... EN-45
5.4 Communication ............................................................................................................................................................ EN-45
5.4.1 Commands ..................................................................................................................................................... EN-45
5.4.2 Connections .................................................................................................................................................... EN-46
Limited Warranty ..................................................................................................................................................................... EN-47
TABLE OF CONTENTS (Cont.)
EN-3Ohaus Discovery
1 INTRODUCTION
This manual contains installation, operation and maintenance instructions for the Ohaus Discovery balances. Please read the manual com­pletely before using the balance.
1.1 Description
Ohaus Discovery balances are precision weighing instruments that will provide you with years of service if properly cared for. The Ohaus Discovery balances are available in capacities from 110 grams to 310 grams.
1.1.1 Model Name Designations
Example: DV214CD DV = Product family abbreviation Discovery 214 = 210g model with 0.0001g readability C = With internal calibration (AutoCal) D = Dual range model
1.2 Features
The Ohaus Discovery balances include many application modes.
• Weighing, Parts Counting
• Animal Weighing
• Check Weighing
• Percent Weighing
• Totalization
• Density, G/N/T
• Pipette Calibration
• High Point, Statistics Other Standard features include:
• Two line backlit LCD
• In-use cover
• RS232 interface
• Weigh below hook
• Automatic Internal Calibration
• Overload and Underload indicators
• Lock-out switch safeguards pre-set parameters
• Legal for Trade models
• AC Adapter operation (AC Adapter included)
• Adjustable feet and level bubble
• Anti static wipes
1.3 Safety Precautions
Please follow these safety precautions:
• Verify that the input voltage printed on the AC Adapter matches the local AC power supply
• Use the balance only in dry locations
• Do not operate the balance in hostile environments
• Do not drop loads on the pan
• Service should be performed only by authorized personnel
EN-4
Ohaus Discovery
2 INSTALLATION
2.1 Unpacking
Carefully remove your Discovery balance and each of its components from the package. Save the packaging to ensure safe storage and trans­port.
2.2 Installing Components
Refer to the illustration and instructions below to identify and assemble your Discovery balance with its components. All components must be installed before using the balance.
2.2.1 Assembly
1. Open the side doors of the balance, then Install the Shield, Wind Ring and Pan as shown in the illustration.
2. Place the In-Use cover on the front panel of the balance.
2.3 Selecting the Location
Locate the balance on a firm, steady surface. Avoid locations with excessive air currents, vibrations, heat sources, or rapid temperature changes.
2.4 Leveling the Balance
Before the balance is used, the feet should be adjusted so that the scale is level with the surface. This will enable accurate weighing. You will see a level bubble indicator in a small round window inside the weighing compartment. Level the balance by adjusting the leveling feet so the bubble is centered in the circle, as shown below.
Balance level
Installation of Balance Components
In-Use Cover
Pan
Wind Ring
Shield
EN-5Ohaus Discovery
2.5 Connecting Power
2.5.1 AC Adapter
Connect the AC Adapter to the wall outlet. Connect the plug into the receptacle on the rear of the balance.
For use with CSA Certified (or equivalent approved) power source, which must have a limited circuit output.
2.5.2 Turning Power On and Off
Press the On/Zero button to turn power on.
The balance performs a segment check. The balance then displays the last selected application mode.
To turn power off, press and hold the Off button until OFF is displayed, then release the button.
2.6 Initial Calibration
When the balance is first installed, and when it is moved to another location, it must be calibrated to ensure accurate weighing results. The Discovery balance has built in Advanced AutoCal which can calibrate the balance automatically and does not require calibration masses. If preferred, the balance can be manually calibrated with external masses. Have the appropriate calibration masses available before beginning calibration. Refer to Section 3.7.2 for masses and calibration procedure.
IMPORTANT: Balance should acclimate to its new surroundings for at least 8 hours. Balance electronics should be warmed up for 2 hours prior to use.
EN-6
Ohaus Discovery
EN-7Ohaus Discovery
3 OPERATION
3.1 Overview of Controls & Display Functions
TOP VIEW BOTTOM VIEW REAR VIEW
TABLE 3-1. DISCOVERY CONTROL FUNCTIONS.
1 Level Bubble
2 Pan
3 Wind Ring
4 Tare Buttons
5 On / Zero Off Button
6 Print Unit Button
7 Function Mode Button
8 Calibrate Menu Button
9 Weigh Below Cover and Screw
10 Leveling Feet
11 Lockout Reset Switch
12 COM 1 Connector
13 Power Connector
1
2
3
4 5 6 7 8
9
10
11 12 13
4
EN-8
Ohaus Discovery
3.1 Overview of Controls & Display Functions (Cont.)
Digital Display Window
1
2
3
4
56789
10
11
12
13
14
1 Primary (7 Segment) Display
2 Brackets
3 Units of Measurement
4 Memory Indicator
5 Pieces Indicator
6 Secondary (14 Segment) Display
7 Gross Indicator
8 Brutto Indicator
9 Tare Indicator
10 Preset Tare Indicator
11 Net Indicator
12 Stability Indicator
13 Negative Indicator
14 Center of Zero Indicator
TABLE 3-2. DISCOVERY DISPLAY FUNCTIONS.
EN-9Ohaus Discovery
3.2 Button Functions
Four multifunction buttons and two tare buttons are used to to operate the balance and navigate the menus. The function of each button is
shown below.
Press and Release Press and Hold
Menu Function
Primary Function Tare
• Performs tare operation.
On/Zero
• If balance is off, turns balance on.
• Zeros balance.
Print
• Sends the current display value to the serial interface.
Function
• Operation is depen­dent on the applica­tion mode.
Cal
• Initiates calibration function.
Secondary Function Off
• Turns balance off.
Unit
• Change weighing unit. Press and hold will scroll through units.
Mode
• Change application mode.
Menu
• Press and hold allows entering the menu.
Menu Function Yes
• Accepts the current (blinking) setting on the display.
• When editing in the alpha-numeric field, increments to the next digit.
No
• Rejects the current setting (blinking) on the display.
• Increments a value being entered.
Back
• Reverts back to the previous menu item.
• Decrements a value being entered.
Exit
• Immediately exits menu mode.
• Aborts calibration in progress.
EN-10
Ohaus Discovery
3.3 Using the Button Control Functions
3.3.1 Setting the Balance to Zero
Remove the load from the pan and press the Zero button to set the display to zero. When the weighing pan is empty, the Center of Zero indicator turns on when the measurement is within + 1/4 division (d) of the zero setting.
3.3.2 Taring
Taring refers to the action of allowing for the weight of a container so that only the weight of objects held in the container (net weight) is dis­played.
To Tare Place the empty container on the pan and press the Tare button. Add material to the container. The net weight of the material is displayed. To clear the Tare value, remove the container from the pan and press the Tare button.
Auto Tare Auto Tare automatically tares the first item placed on the pan. Auto Tare must be set ON in the Setup sub-menu (See Section 3.7.3 Setup Sub-menu). The secondary display shows CONTAINER (blinking). When the container is placed on the pan, it is automatically tared and Net weight is displayed. The Tare value is automatically cleared when the container is removed from the pan.
Preset Tare A preset tare may be entered by using a computer connected to the COM 1 Connector. To enter a Preset Tare, send the xT command to the bal­ance where x is the value in grams. Enter a value of 0 to clear the preset tare.
3.3.3 Changing Units of Measure
The Discovery balance can be configured to measure in a variety of units, including custom units. The Unit Sub-menu is used to enable or dis­able a specific unit. To select a unit of measure: Press and hold the Unit button, then release it when the desired unit is displayed. Note: If the desired unit is not displayed, it must be turned on in the Unit menu (See Section 3.7.6).
3.3.4 Changing Application Modes
The Discovery balance can be configured to operate in various application modes. The Mode Sub-menu is used to enable or disable a specific application mode. Press and hold the Mode button and release it when the desired mode is shown on the secondary display. Note: If the desired mode is not displayed, it must be turned on in the Mode menu. (See Section 3.7.5).
3.3.5 Printing Data
Press the Print button to send the displayed value to the COM port (See Section 3.10).
3.3.6 Calibration
Press the Cal button to enter the calibration mode. (See Section 3.7.2).
EN-11Ohaus Discovery
3.4 MENU
3.4.1 Menu Structure
The Discovery balance menu structure is illustrated below.
CALIBRATE (3.7.2)
AutoCal Span Calibration Test (Cal Test) Calibration Adjust (Cal Adj) End Cal
SETUP (3.7.3)
Legal Trade Auto Tare Auto Off End Setup
READOUT (3.7.4)
Auto Zero Filter Gross Ind. End Read
PRINT (3.7.7)
Output When Stable GLP Tare Auto Print Content Num Only Header, Gross, Net, Tare, Ref, Result, GLP Layout Line Format 4LF Form Feed List End Print
UNIT (3.7.6)
mg g ct oz ozt GN dwt m t HK t SG t TW Custom End Unit
MODE (3.7.5)
Weighing Parts Counting Percent Weighing Check Weighing Animal Weighing GNT Totalize High Point Density Statistics Pipette End Mode
RS232 (3.7.8) Baud Parity Handshake End RS232
GLP DATA (3.7.9)
User ID Proj ID Time Date End GLP Data
GLP PRINT (3.7.10)
Time Balance ID User ID Project ID Difference Name End GLP Prt
RESET (3.7.11)
Setup GLP Data Readout GLP Print Mode Lockout Unit Global Print End Reset RS232
LOCKOUT (3.7.12)
Cal RS232 Setup GLP Data Readout GLP Print Mode Reset Unit Lock Set Print End Lockout
END (3.7.13)
EN-12 Ohaus Discovery
3.4.2 Menu Navigation
The Discovery menu structure consists of three levels. The top level is the main menu which consists of a number of sub-menus.
CALIBRATE SETUP READOUT MODE UNIT PRINT
END LOCKOUT RESET GLP PRINT GLP DATA RS232
TOP LEVEL
SECOND LEVEL
The second level of the menu structure is the sub-menu level. Each sub-menu is a loop that consists of a number of menu items.
BOTTOM LEVEL
The bottom level of the menu structure is the menu item level. Each menu item level consists of two or more settings
3.4.3 Changing Settings
To change a menu setting, perform the following steps:
ENTER THE MENU
Press and hold the Menu button until MENU appears on the primary (7-Segment) display. Release the button and the first sub-menu appears on the secondary display (14-Segment) Calibration (blinking).
Note: When the secondary display is blinking, press Yes to accept. Press No or Back or view another selection.
SELECT THE SUB-MENU
Press the No button to select the next sub-menu or the Back button to select the previous sub-menu in the loop. Press the Yes button to select the sub-menu that appears on the secondary display (blinking). The sub-menu name now appears on the primary display and the first menu item in the sub-menu now appears on the secondary display (blinking).
SELECT THE MENU ITEM
Press the No button to select the next setting or the Back button to select the previous setting in the loop. Press the Yes button to select and save the setting on the secondary display (blinking) as the new menu item setting.
EXIT THE MENU
Press the Exit button at any time to quickly exit the menu or select END sub-menu and press the Yes button.
Enter
END RS232 BAUD
HANDSHAKE
PARITY
Typical Sub-Menu
Typical Menu Item
Loading...
+ 36 hidden pages