Philips M1013A IntelliVue G1, M1019A IntelliVue G5 Service Manual

M1013A / M1019A IntelliVue G1/G5
Service Guide
IntelliVue G1/G5
M1013A/M1019A
Patient Monitoring
Part Number 4535 643 23271 Issued in Germany 02/2012
*453564323271*
1Table of Contents
1 Introduction 7
Who Should Use This Guide 7 How To Use This Guide 7 Description 7 Responsibility of the Manufacturer 8 Warnings and Cautions 8 Physical Specifications 8 Environmental Specifications 9 MDD Classification 9 Performance Specifications 9
CO2 Measurement 10 AWRR derived from CO2 Waveform 10 N2O Measurement 10 O2 Measurement 10 Anesthetic Agent Measurement 10 Alarm Ranges 11 Alarm Delay 11
awRR Alarm Delay 11
Apnea Alarm 11 INOP Alarms 12
Theory of Operation 12
General Measurement Principles 12 O2 Sensor 12
Measurement Principle 12
Pump 13 Wate rt ra p 13
2 Installation and Patient Safety 15
Site Preparation - Introduction 15 IntelliVue G1/G5 Site Requirements 15 Environment 16 Initial Inspection 16
Mechanical Inspection 16 Electrical Inspection 16
Claims for Damage and Repackaging 17
Claims for Damage 17 Repackaging for Shipment or Storage 17
Making Connections to the IntelliVue G1/G5 17
Connecting the IntelliVue G1/G5 to AC Mains 18
3
Securing the Power Cord 19
Connections to the Sample Gas Exhaust 20
Returning the Gas Sample 20 Setting Up the Gas Return 20 Removing the Gas Sample 21
Installing the Top Mount 21 Mounting Instructions 22 Setup and Configuration Procedures 23
IntelliVue Serial Port Configuration 23 Altitude Configuration 23 Connect Sample Input Tubing 23
Post-Installation Checks 23 Safety Requirements Compliance and Considerations 23
Explanation of Symbols Used 24 Electrical and Safety Requirements (Customer or Philips) 24 Power Supply Requirements 24 Protective Earthing of the System 25 Equipotential Grounding 25 Combining Equipment 25 Connecting Non-Medical Devices 25
3 Software Uploads 27
Checking the Unit for Functionality 28 Uploading the Software 32
4 Testing and Maintenance 41
Introduction 41 Terminology and Definitions 41 Recommended Frequency 42 When to perform Tests 42 Testing Sequence 43 Visual Inspection 43
Before Each Use 43 After Each Service, Maintenance or Repair Event 43 Power On Test 44
Safety Tests 44
Warnings, Cautions, and Safety Precautions 44 Safety Test Procedures 45 Hints for Correct Performance of Safety Tests 47 Guideline for Performance of Safety Tests 47
Electrical Safety Testing 48
S(1): Protective Earth Resistance Test 48 S(2) Equipment Leakage Current Test - Normal Condition 49 S(3) Equipment Leakage Current Test - Single Fault Condition 50 Reference: Allowable Values for IEC 60601-1:1998 and UL 60601-1 Measurements 51 Insulation Resistance 51
4
System Test 51
What is a Medical Electrical System 51 General Requirements for a System 52
Preventive Maintenance Procedures 52
Cleaning 52 Replace PM Parts 53
Replacing the Fan Filter 53 Replacing the Watertrap Manifold Seals 53
Performance Assurance Tests - Checking and Calibrating the Gas Analyzer 54
Access Service Functions of the Gas Analyzer 54 When and how to check the Gas Analyzer 54 Equipment required for checking 54 Annual Checks 56
Connecting the Gas Analyzer to a PC/Laptop 56 Getting started with the VISIA software 56
Zero Calibration 59
Zero Calibration Test 59
Component Status Check 60 Pneumatic Tests 61
Equipment needed: 61
Leak Check 61
Checking for leaks between inlet and pump 61
Flow Rate Check 64 Pressure Sensor Test 64 Flow Rate Adjustment 65 Gas Calibration Test 66
Disposal of Empty Gas Cylinder 69
Mounting Integrity Test 69
Reporting of Test Results 69
Carrying Out and Reporting Tests 70
Tes t R ep or t 7 0 Test and Inspection Matrix - Checks with Patient Monitor 71 Checks with VISIA Tool 73 Evaluation 73
Evaluation of Test Results 74
Other Regular Tests 74 After Installation, Testing or Repair 74
5 Troubleshooting the Gas Analyzer 75
Technical Alarm Messages (INOPs) 76 Troubleshooting 78
6 Repairing the Gas Analyzer 79
Introduction 79 Who Should Perform Repairs 79
5
Tools required 79 Removing the Bottom Quick Release Mount 80
7 Parts List 81
Exchange Parts 81 Replacement Parts 81
6
This book is intended for personnel authorized to install, service or repair an IntelliVue G1 or IntelliVue G5 gas analyzer. A good understanding of the English language is a requirement.
This chapter contains the following information on the M1013A IntelliVue G1 and the M1019A IntelliVue G5:
• A description of the module, including its physical, environmental and performance specifications
• A general explanation of the measurement principles that the module uses to measure gas concentrations
• The theory of operation of the module, its components and how they work.

Who Should Use This Guide

1

1Introduction

This guide is for biomedical engineers or technicians responsible for troubleshooting, repairing, and maintaining Philips’ patient monitoring systems.

How To Use This Guide

This guide is divided into seven sections. Navigate through the table of contents at the left of the screen to select the desired topic. Links to other relevant sections are also provided within the individual topics. In addition, scrolling through the topics with the page up and page down keys is also possible.

Description

The Philips M1013A IntelliVue G1 and the M1019A IntelliVue G5 work together with the IntelliVue patient monitors through an RS232 serial interface. They measure the airway gases of ventilated patients who are under general gas anesthesia, or emerging from it.
The modules produce graphical wave data, and inspired and end-tidal numeric data for the following gases:
•CO
•N
• One volatile anesthetic agent (IntelliVue G1) / Two volatile anesthetic agents (IntelliVue G5)
•O
It also generates numerics for MAC (Minimum Alveolar Concentration) and the patient’s airway respiration rate (awRR).
2
O
2
(optional with IntelliVue G1, standard with IntelliVue G5)
2
7
1 Introduction Responsibility of the Manufacturer

Responsibility of the Manufacturer

Philips only considers itself responsible for any effects on safety, EMC, reliability and performance of the equipment if:
• assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by Philips, and
• the electrical installation of the relevant room complies with national standards, and
• the instrument is used in accordance with the instructions for use.
To ensure safety and EMC, use only those Philips parts and accessories specified for use with the monitor. If non-Philips parts are used, Philips is not liable for any damage that these parts may cause to the equipment.
This document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
Philips Medizin Systeme Böblingen GmbH
Hewlett-Packard Str. 2
71034 Böblingen, Germany
The information contained in this document is subject to change without notice.
Philips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.
Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Warnings and Cautions

In this guide:
•A warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.
•A caution alerts you where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.

Physical Specifications

Size (H x W x D):
93 x 306 x 232 mm (3.66 x 12.05 x 9.13 in).
Weight:
8
less than 4 kg (7.94 lb)
Environmental Specifications 1 Introduction

Environmental Specifications

Operating Temperature: 10 to 40C (50 to 104F)
Storage Temperature: -20 to 65C (-4 to 149F)
Humidity Limit (Operating): 5 to 90% RH max @ 40C (104F).
non-condensing
Humidity Limit (Storage): 5 to 95% RH max @ 65C (149F).
non-condensing
Altitude Range (Operating): -305 to 2900m (-1,000 to 9,515ft)
Altitude Range (Storage): -305 to 5000m (-1,000 to 16,404ft)
Warm-up Time: 1-2 minutes to measure CO
accuracy specifications

MDD Classification

According to the Council Directive 93/42/EEC (Medical Devices Directive) the device classification is 2A, Rule 10.

Performance Specifications

All Performance and accuracy specifications are valid based on gas sample tubing M1658A, including watertrap M1657B, and airway adapter 13902A.
Humidity Correction: For CO
Wet: p [mmHg] = c [Vol%] * (p_abs - p_H
Dry: p [mmHg] = c [Vol%] * p_abs /100
Where p = partial pressure, c = gas concentration, p_abs = pressure in breathing circuit, p_H2O = 21mmHg, partial pressure of water vapor of room temperature gas (23 oC, 100% rh).
For all other gases the readings are always given as dry values.
Sample Flow Rate: 200 ml/min
Sample Delay Time: All measurements and alarms are subject to a delay of 5 seconds.
the humidity correction can be set to “wet” or “dry”.
2
1
20
O)/100
2
, less than 6 minutes for full
2
Total System Response Time = the sum of the delay time and the parameter specific rise time.
1. After warm up or zero the flow rate may be higher than 200 ml/min for about 30 minutes.
9
1 Introduction Performance Specifications

CO2 Measurement

Range: 0 to 76 mmHg
Accuracy: 0.5 vol% or 12% relative, whichever is greater
Resolution: 1 mmHg
Rise-time: 350 msec typical

AWRR derived from CO2 Waveform

Range: 0 to 60 rpm
Accuracy: ± 1 rpm
Resolution: 1 rpm
Detection Criteria: adaptive threshold

N2O Measurement

Range: 0 to 100vol%
Accuracy: 2.0 vol% + 8% relative
Resolution: 1 vol%
Rise-time: 500 msec typical

O2 Measurement

Range: 5 to 100vol%
Accuracy: ± 3 vol%
Resolution: 1 vol%
Rise-time: 500 msec typical

Anesthetic Agent Measurement

Agent Range (vol%) Accuracy Resolution Rise Time
Halothane 0 - 8.5 0.15 vol% + 15.0% relative 0.05 < 500
Enflurane 0 - 10.0 0.15 vol% + 15.0% relative 0.05 < 500
Isoflurane 0 - 8.0 0.15 vol% + 15.0% relative 0.05 < 500
Sevoflurane 0 - 10.0 0.15 vol% + 15.0% relative 0.05 < 500
Desflurane 0 - 20.0 0.15 vol% + 15.0% relative 0.05 < 500
10
Performance Specifications 1 Introduction
Agent ID Response Time 14 s for first agent, 19 s for second agent
First Agent Detection / Identification Threshold
Second Agent Detection / Identification Threshold
All agents max. 0.3 vol%
All agents max. 0.4 vol% of a second agent, except if a
second agent is added to Desflurane, this causes a mixture identification at the latest if the concentration of the second agent exceeds 10 vol% of the current Desflurane concentration.

Alarm Ranges

Agent High Range Low Range
AWRR 10 - 60 rpm (Adult/Pedi)
ETCO
IMCO
inN2O 0 - 82 vol% none
inO2 (optional) 19-100 vol% 18 - 99 vol%
et SEV 0.1 - 9.0 vol% 0.0 - 8.9 vol%
in SEV 0.1 - 9.0 vol% 0.0 - 8.9 vol%
et DES 0.2 - 20.0 vol% 0.0 - 19.8 vol%
in DES 0.2 - 20.0 vol% 0.0 - 19.8 vol%
Halothane, Enflurane, Isoflurane
et 0.1 - 7.5 vol% 0.0 - 7.4 vol%
in 0.1 - 7.5 vol% 0.0 - 7.4 vol%

Alarm Delay

15 seconds if no zero calibration occurs within that time.
awRR Alarm Delay
The alarm delay for the awRR low alarm is 10 sec for awRR > 20rpm and 0 sec for awRR < 20rpm. The alarm delay for the awRR high alarm is 10 sec.
0 - 55 rpm
30 - 60 rpm (Neonatal)
2
2
20 - 76 mmHg 10 - 75 mmHg
2 - 20 mmHg none

Apnea Alarm

Delay Range: 10 - 40 seconds
Criterion No detected breath within the adjusted delay time
Alarm: Within 2 seconds after this criterion is met, if no automatic zero
occurs
11
1 Introduction Theory of Operation

INOP Alarms

INOP alarms are triggered if:
• The gas analyzer is disconnected or switched off.
• The gas analyzer accuracy is in doubt.
• The equipment or any of its components malfunctions.
• Zero calibration has failed.
• The gas sample tube is occluded, or the watertrap is full.
• Any parameter is unable to measure.
• Any parameter is out of range.
• The gas analyzer is in warm-up mode.
• Gas analyzer calibration is running.
• Gas analyzer alarms are suppressed.
• No breath detected.

Theory of Operation

General Measurement Principles

The M1013A IntelliVue G1 and the M1019A IntelliVue G5 use infrared technology to measure the concentration of the gases CO
The gases which can be measured by the gas analyzer absorb infrared (IR) light. Each gas has its own absorption characteristic. The gas is transported into a sample cell, and an optical IR filter selects a specific band of IR light which has passed through the gas. For multiple gas measurement, such as in the IntelliVue G1 and G5, there are multiple IR filters. The higher the concentration of gas in a given volume the more IR light is absorbed. This means that higher concentrations of IR absorbing gas cause a lower transmission of IR light. The amount of IR light transmitted after it has been passed through the gas is measured. From the amount of IR light measured, the concentration of gas present can be calculated. This calculation provides the gas measurement value. Oxygen is measured using a paramagnetic cell.
NOTE The presence of organic cleaning solutions or gases containing freon may impact the accuracy of the
infrared gas measurement.

O2 Sensor

NOTE The O
Sensor is optional with IntelliVue G1 and standard with IntelliVue G5
2
, N2O and the volatile anesthetic agents.
2
Measurement Principle
The O2 sensor uses a fast O2 measurement technique that utilizes the paramagnetic properties of oxygen.
12
Theory of Operation 1 Introduction
Gases with paramagnetic properties are attracted by magnetic fields. In a magnetic field the density and thus the heat conductivity of such gases is increased. The gas analyzer determines the amount of oxygen in the gas sample by measuring its heat conducting properties while switching a magnetic field on and off inside the O measured, and the amount of oxygen in the gas sample can be calculated.
sensor. This way the changes in the oxygen present in the magnetic field can be
2

Pump

The software-controlled pump generates the flow through the system and pulls the gas from the airway adapter through the measurement subsystems to the exhaust outlet. It also delivers the zero calibration gas to the sample cells of the measurement subsystems for the periodic zero procedures and it exhausts the patient’s sample gas, the zero calibration and field calibration gases.
The flow-rate control logic drives the pump as hard as necessary to maintain the selected flow rate. A partial occlusion or an inefficient pump results in the pump being driven harder. A serious occlusion results in the pump being driven at or near its maximum load. If, as a result of this occlusion, the desired flow rate cannot be upheld, an occlusion INOP is triggered.

Watertrap

Figure 1 Watertrap
The watertrap consists of two water separation filters, two water fuses and a water reservoir. The gas sample coming from the patient may contain fluids which are separated from the gas at the first water separation filter. The gas is then split into two paths, the “measurement” path with the main part of the total gas flow (including water vapor) continuing on the “dry” side of the separation filter and the “drainage” path (containing any liquid droplets) with the smaller amount of the total flow continuing on the “wet” side of this filter through the water reservoir. At the pump both gas paths are recombined.
13
1 Introduction Theory of Operation
The watertrap itself includes “water fuses” in both the “measurement” and the “drainage” paths, consisting of a material that swells when getting wet (when the reservoir is full or when fluid penetrates the separation filter and enters the “measurement” path) and blocks the respective path at the inlet of the unit. Once the “water fuses” are blown, any passage of fluid is blocked and the gas flow resistance increases so that an occlusion is detected.
14
2

2Installation and Patient Safety

NOTE The M1013A IntelliVue G1 and the M1019A IntelliVue G5 must be installed by qualified personnel
capable of performing the post-installation checks as outlined in the Test and Inspection Matrix
This chapter describes how to install the Philips M1013A IntelliVue G1 and the M1019A IntelliVue G5. It details the operating environment required by the gas analyzers as well as instructions on how to physically connect them to the monitor and how to fit the gas exhaust return system. Next, the patient safety information is detailed. Finally, this chapter describes the software setup required and any post­installation checks that have to be performed before using the gas analyzer together with a reminder of the preventive maintenance (PM) checks and their frequencies.
CAUTION The gas analyzer must be positioned on a surface with a maximum incline of 15°. To avoid condensed
water collecting in the patient sample tube, it is recommended that the gas analyzer is positioned at or above patient level, wherever possible.

Site Preparation - Introduction

This section describes the procedures you should follow to plan and prepare a site for an IntelliVue G1/ G5 installation.
Refer to the Site Preparation chapter in the respective IntelliVue Patient Monitor Service Guide, for details about:
• Site planning.
• Roles and responsibilities for local and Philips personnel.
• Remote installation planning.
These details are also valid for the IntelliVue G1 / G5.

IntelliVue G1/G5 Site Requirements

For space requirements and environmental requirements refer to chapter 1, Introduction.
15
2 Installation and Patient Safety Environment

Environment

WARNING Possible explosion hazard if used in the presence of flammable anesthetics.
The environment where the gas analyzer is used should be free from vibration, dust, corrosive or explosive gases, and extremes of temperature and humidity.
For a cabinet mounted installation with the monitor, allow sufficient room at the front for operation and sufficient room at the rear for servicing with the cabinet access door open.
The IntelliVue G1 and the IntelliVue G5 operate within specifications at ambient temperatures between 10C and 40C, 6 minutes after switching it on.
Ambient temperatures that exceed these limits could affect the accuracy of this instrument and cause damage to the components and circuits. Allow at least 2 inches (5cm) clearance around the instruments for proper air circulation.
CAUTION If the gas analyzer has been stored at temperatures below freezing, it needs a minimum of 4 hours at
room temperature to warm up before any connections are made to it.
Make sure that the gas analyzer is free of condensation before operation. Condensation can form when equipment is moved from one building to another, thus being exposed to moisture and differences in temperature.

Initial Inspection

Mechanical Inspection

Open the shipping container(s) and examine each part of the INtelliVue G1 / G5 for visible damage, such as broken connectors or controls, or scratches on the equipment surfaces. If the shipping carton/ container is undamaged, check the cushioning material and note any signs of severe stress as an indication of rough handling in transit. This may be necessary to support claims for hidden damage that may only become apparent during subsequent testing.

Electrical Inspection

The IntelliVue G1 / G5 has undergone extensive testing prior to shipment. Safety testing at installation is not required (except in situations where devices are interconnected forming a system). An extensive self check may be performed. This recommendation does not supersede local requirements.
All tests are described in the Testing and Maintenance section of this manual.
16
Claims for Damage and Repackaging 2 Installation and Patient Safety
1
2
3
4
5

Claims for Damage and Repackaging

Claims for Damage

When the equipment is received, if physical damage is evident or if the IntelliVue G1/ G5 does not meet the specified operational requirements of the patient safety checks or the extended self check, notify the carrier and the nearest Philips Sales/Support Office at once. Philips will arrange for immediate repair or replacement of the instrument without waiting for the claim settlement by the carrier.

Repackaging for Shipment or Storage

If the instrument is to be shipped to a Philips Sales/Support Office, securely attach a label showing the name and address of the owner, the instrument model and serial numbers, and the repair required (or symptoms of the fault). If available and reusable, the original Philips packaging should be used to provide adequate protection during transit. If the original Philips packaging is not available or reusable please contact the Philips Sales/Support Office who will provide information about adequate packaging materials and methods.

Making Connections to the IntelliVue G1/G5

All connections to the gas analyzer are made on its rear panel. Refer to Figure 2.
Figure 2 The Rear Panel
1 Local power connector; this is a 3-pin connector, used to connect the gas analyzer to AC Power.
The gas analyzer can be operated from an AC power source of 100 - 240 V ± 10%, 50/60 Hz. The adjustment is made automatically by the power supply inside the module.
2 RS232 Connector (RS232 Interface); this is an RJ45 connector, used to connect the gas analyzer to
the monitor.
The connection to an IntelliVue patient monitor can be made with the following cables:
• For M1013A IntelliVue G1:
– M1013A#K11 1.5 m (M1013-61001)
– M1013A#K12 3 m (M1013-61002)
• For M1019A IntelliVue G5
17
2 Installation and Patient Safety Making Connections to the IntelliVue G1/G5
– M1019A#K11 1.5 m (M1013-61001)
– M1019A#K12 3 m (M1013-61002)
3 Equipotential Grounding Terminal; this is used to connect the gas analyzer to the hospital’s
equipotential grounding system.
4 Gas exhaust. If N
O and/or other inhalation anesthetics are used during anesthesia, pollution of
2
the operating room should be prevented. Once the gas sample has passed through the gas analyzer, it should either be returned to or removed from the anesthesia circuit.
NOTE In some countries where closed loop functionality is not available, gas must not be returned to the
anesthesia circuit.
5 Fan Filter
CAUTION Combinations of medical equipment with non-medical equipment must comply with IEC 60601-1-1.
Never use a multiple portable socket-outlet or extension cord when combining equipment unless the socket outlet is supplied specifically for use with that equipment.

Connecting the IntelliVue G1/G5 to AC Mains

The IntelliVue G1/G5 has a wide-range power supply that allows you to operate the monitor from an AC (alternating current) power source of 100 V to 240 V (± 10%) and 50/60 Hz (± 5%).
WARNING • Always use the supplied power cord with the earthed mains plug to connect the monitor to an
earthed AC mains socket. Never adapt the mains plug from the power supply to fit an unearthed AC mains socket.
• Do not use AC mains extension cords or multiple portable socket-outlets. If a multiple portable socket-outlet without an approved isolation transformer is used, the interruption of its protective earthing may result in enclosure leakage currents equal to the sum of the individual earth leakage currents, so exceeding allowable limits.
18
• Do not connect any devices that are not supported as part of a system.
• Any non-medical device placed and operated in the patient’s vicinity must be powered via an approved isolation transformer that ensures mechanical fixing of the power cords and covering of any unused power outlets.
Making Connections to the IntelliVue G1/G5 2 Installation and Patient Safety

Securing the Power Cord

In order to prevent the power cord from accidentally being unplugged, secure it with the power cord securing bracket.
1 Insert the nose of the power cord securing bracket into the small slit above the power connector.
2 Slide the bracket to the left and secure it with knurled nut.
19
2 Installation and Patient Safety Connections to the Sample Gas Exhaust
1
2
3
4

Connections to the Sample Gas Exhaust

Returning the Gas Sample

You will need the following equipment to return the gas sample to the anesthesia circuit:
Equipment Part Number Comments
Gas Exhaust Return Line M1655B Tubing includes two parts:
Tube A = 300 cm long
Tube B = 30 cm long
Gas Exhaust Return Filter M1656B Single patient use only
Gas Exhaust Tubing M1015-40001 Multi-Patient use

Setting Up the Gas Return

(see diagram Figure 3)
1 Fit the shorter tube tightly to the female side of the filter. Shorten the tube if it is worn or does not
fit tightly onto the filter.
NOTE When using the M1656B Gas Exhaust Return Filter with an old M1655A Gas Exhaust Return Line,
you must cut off the luer lock connection of the shorter tube first before connecting to the filter.
2 Fit the female luer lock connection (2) of the longer tube to the male side of the filter.
3 Fit the open end (5) of the longer tube to the Anesthetic Gas Exhaust outlet.
4 Fit the open end (4) of the shorter tube to the ventilation circuit.
20
Figure 3 Setting Up the M1655B Gas Exhaust Return Line
Installing the Top Mount 2 Installation and Patient Safety
NOTE
Make sure the sample gas is routed through the CO2 absorber before going back to the patient.
1 M1656B Gas Exhaust Return Filter
2 Female luer lock
3 Shorter tube connecting to the ventilation circuit
4 Longer tube connecting to the Anesthetic Gas Exhaust Outlet

Removing the Gas Sample

To remove the gas sample from the anesthesia circuit, a scavenging system needs to be connected to the gas analyzer’s Anesthetic Gas Exhaust. If you intend to use a scavenging system with the gas analyzer, one of the following parts must also be connected to protect it against malfunction:
1 A ventilator reservoir where the suction pressure does not exceed 70 mbar or
2 A scavenging interface, properly set and maintained (see scavenging interface manufacturer’s
instructions).
NOTE If you are not returning the gas sample into the patient’s breathing circuit, install the M1655B Exhaust
Return Tubing without the M1656B Exhaust Return Filter, shorter tube and the luer lock fitting. See the Instructions for Use provided with the tubing and filter for further details.

Installing the Top Mount

1 Remove the three rounded head screws on the top of the gas analyzer.
21
2 Installation and Patient Safety Mounting Instructions
2 Attach the top mount to the gas analyzer using only the three short countersunk screws supplied
with the mount.
NOTE Devices installed on the top mount may not weigh more than 12 kg. Make sure that any device
installed on the top mount snaps in properly and is fixed securely to the mount.

Mounting Instructions

NOTE There are different mounting options available for the IntelliVue G1 / G5. This section covers the
general concepts of safe mount installations and specific steps for the mounting options sold by Philips. Instructions which ship with a mounting solution should always take precedence over the instructions described in this chapter. You MUST follow the instructions that ship with the mounting solution, regardless of manufacturer.
Please mount the IntelliVie G1 / G5 using the Philips Quick Mount solution or another approved mounting solution. The mounting shall be done in a manner that no patient, operator or other person can be harmed by a IntelliVue G1 / G5 removed intentionally or released accidentally from the mount. When using the Quick Mount, be aware of the danger of accidental activation of the Quick Mount release button when lifting or moving items located under the monitor, such as pole mounts, etc.
For instructions on how to mount the monitor using the Quick Mount table mount refer to the Assembly Instructions delivered with the mounting kit M8000-64100 or 453564239731.
WARNING • It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling,
wall, or mounting rail and the construction of the ceiling, wall, or mounting rail evaluated for structural integrity and compliance with all local, state and any other required codes by a registered, professional, structural and/or mechanical engineer.
• Ensure that this commitment has been met before assembling mounts.
22
• Incorrect mounting and use of inappropriate mounting material may lead to injury. It is the customer’s responsibility to ensure that the mounting procedures have been performed correctly and the appropriate mounting devices have been used.
Setup and Configuration Procedures 2 Installation and Patient Safety

Setup and Configuration Procedures

This section describes final setting up and configuration procedures that must be completed after the gas analyzer is connected to the monitor and switched on before the gas analyzer is used for monitoring.

IntelliVue Serial Port Configuration

The MIB port used in the IntelliVue host monitor must be configured to “GM”. To do this, go into service mode and then select Setup followed by Hardware and then MIB.

Altitude Configuration

The altitude setting for the monitor is important as it is used as a reference to check the gas analyzer ambient pressure measurement.
See your monitor service guide for details.

Connect Sample Input Tubing

Connect the sample input tubing to the watertrap at the luer lock connector. For details, refer to the Instructions for Use.

Post-Installation Checks

See Test and Inspection Matrix for details.
WARNING Do not use the instrument for any monitoring procedure on a patient if you identify anything which
indicates impaired functioning of the instrument.

Safety Requirements Compliance and Considerations

The M1013A IntelliVue G1 and the M1019A IntelliVue G5 comply with the following international safety requirements for medical electrical equipment:
IEC 60601-1:1988 + A1:1991 + A2:1995; EN60601-1:1990 + A1:1993 + A2:1995; UL 60601­1:2003; CAN/CSA C22.2#601.1-M90; IEC 60601-1-2:2001; EN 60601-1-2:2001.
Classification (according to IEC 60601-1): Class 1, Type BF, Continuous Operation.
This ISM device complies with Canadian ICES-001. Cet appareil est conforme a la norme NMB-001 du Canada.
The possibility of hazards arising from software errors was minimized in compliance with ISO14971:2000, EN60601-1-4:1996 + A1:1999 and IEC 60601-1-4:1996 + A1:1999.
23
2 Installation and Patient Safety Safety Requirements Compliance and Considerations
2002­06
0366

Explanation of Symbols Used

Symbols
Caution, refer to accompanying documents
Power On/Off*
Alternating current Electrical signal input
Applied part has special protection against electric shocks (Type BF according to IEC 60601-1)and is defibrillator proof
The device complies with the requirements of the Council Directive 93/42/EEC of 14 June 1993 (Medical Device Directive).
Indicates location of catalog number
Setup*
Equipotential grounding
indicator
Gas output indicator Gas input indicator
Indicates location of the date of manufacture and/or name and address of manufacturer
Indicates location of service number
Standby*
Identifies year and month of manufacture
Electrical signal output indicator
Indicates location of serial number
* These symbols are replaced by English text in the U.S.A.
The IntelliVue G1 and the IntelliVue G5 are protected against the effects of defibrillation and electrosurgery.

Electrical and Safety Requirements (Customer or Philips)

Power Supply Requirements

The system and the gas analyzer can both be operated from an AC supply of 100 - 240V ±10%, 50 ­60Hz. The IntelliVue G1/G5 uses <25W typical and <45W peak.
24
Safety Requirements Compliance and Considerations 2 Installation and Patient Safety

Protective Earthing of the System

To protect the patient and hospital personnel, the cabinet of the installed equipment has to be grounded. The equipment is supplied with a detachable 3-wire cable which grounds the instrument to the power line ground (protective earth) when plugged into an appropriate 3-wire receptacle. If a 3­wire receptacle is not available, consult the hospital electrician.
WARNING Do not use a 3-wire to 2-wire adapter.

Equipotential Grounding

Protection class 1 instruments are already included in the protective grounding (protective earth) system of the room by way of grounding contacts in the power plug. For internal examinations on the heart or the brain, Computer Module and Display Module of the System and the gas analyzer must have separate connections to the equipotential grounding system.
One end of the equipotential grounding cable (potential equalization conductor) is connected to the equipotential grounding terminal on the instrument’s rear panel and the other end to one point of the equipotential grounding system. The equipotential grounding system assures that potential differences between conductive parts are limited according to requirements of applicable standards. This safety measure prevents that currents flowing through the heart of a patient caused by potential differences stimulate arrhythmias.
Examinations in or on the heart (or brain) should only be carried out in rooms designed for medical use incorporating an equipotential grounding system.

Combining Equipment

If it is not evident from the instrument specifications whether a particular instrument combination is hazardous or not, for example, due to summation of leakage currents, the user should consult the manufacturers concerned or an expert in the field, to ensure that the necessary safety of all instruments concerned will not be impaired by the proposed combination.

Connecting Non-Medical Devices

Refer to the Site Preparation chapter in the respective IntelliVue Patient Monitor Service Guide for details.
25
2 Installation and Patient Safety Safety Requirements Compliance and Considerations
26
Loading...
+ 58 hidden pages