The DataVU 6 paperless recorder features a resistive touchscreen and an intuitive, icon-based
operation and visualization concept that makes it very easy to use.
There are different versions of the DataVU 6 available for process data recording. These range
from the device version without measuring input in which up to 24 process val-ues are read
(master) or received (slave) from external systems via Modbus, through to a device version with
six measuring inputs (universal analog inputs), two analog outputs, 12 digital inputs, and 12
individually switchable digital inputs/outputs. In the version with FDA-compliant data re-cording
all the requirements are met in accordance with 21 CFR Part 11.
The DataVU 6 can display data using the default visualizations, such as curve diagram (vertical
or horizontal), bar graph, text image (numerical), or digital diagram. For batch-related processes
a special batch recording is available which allows the storage of additional information. In
addition, users can create up to six individual process screens with up to 100 ob-jects per
process screen to fit their requirements using the setup program.
In addition to the setup program further powerful PC programs are available, e.g. for the
evaluation of archived data and for the administration of access control.
Type 706520/ ...
Block diagram
Inputs/outputs
Option 1, option 2:
3 analog inputs,
6 digital inputs,
1 analog output
each
Option 3:
12 digital inputs/outputs,
incl. 1 high-speed counting input
(up to 12.5 kHz)
Inputs via interface
As standard:
24 external analog inputs and
24 external digital inputs and
14 external text (10 batch texts,
4 event texts)
Relay output
As standard:
1 relay (changeover contact)
Display/operation
Display
14,5 cm (5,7”) TFT color screen
640 x 480 pixel,
65536 colors
Operation
Touchscreen (resistive)
Voltage supply
AC 110 to 240 V +10/-15 %,
48 to 63 Hz
AC/DC 20 to 30 V, 48 to 63 Hz
Interfaces
As standard :
1x Ethernet 10/100 MBit/s
1x USB host (flash drive)
1x USB device (setup)
1x RS232/RS485
(Modbus master/slave or
barcode scanner)
Measurem.-data memory
Internal memory:
1 GByte
(data transfer via interface
or USB flash drive)
Fe-CuNi"L"DIN 43710ITPS-68-200 to +900 °C≤ 0.25 %
Fe-CuNi"J"IEC 60584-1ITS-90-210 to +1200 °C≤ 0.25 % from -100 °C
Cu-CuNi"U"DIN 43710ITPS-68-200 to +600 °C≤ 0.25 % from -100 °C
Cu-CuNi DIN"T"IEC 60584-1ITS-90-270 to +400 °C≤ 0.25 % from -150 °C
NiCr-Ni DIN"K"IEC 60584-1ITS-90-270 to +1372 °C≤ 0.25 % from -80 °C
NiCr-CuNi"E"IEC 60584-1ITS-90-270 to +1000 °C≤ 0.25 % from -80 °C
NiCrSi-NiSi"N"IEC 60584-1ITS-90-270 to +1300 °C≤ 0.25 % from -80 °C
Pt10Rh-Pt "S"IEC 60584-1ITS-90-50 to 1768 °C≤ 0.25 % from 20 °C
Pt13Rh-Pt "R"IEC 60584-1ITS-90-50 to 1768 °C≤ 0.25 % from 50 °C
Pt30Rh-Pt6Rh"B"IEC 60584-1ITS-900 to 1820 °C≤ 0.25 % from 400 °C
W5Re/W26Re"C"ASTM E230M-11ITS-900 to 2315 °C≤ 0.25 % from 500 °C
W3Re/W25Re"D"ASTM E1751M-09ITS-900 to 2315 °C≤ 0.25 % from 500 °C
W5Re/W20Re"A1"GOST R 8.585-2001ITS-900 to 2500 °C≤ 0.25 % from 500 °C
Chromel-Copel"L"GOST R 8.585-2001ITS-90-200 to +800 °C≤ 0.25 % from -80 °C
Chromel-AlumelGOST R 8.585-2001ITS-90-270 to1372 °C≤ 0.25 % from -80 °C
Ambient temperature influence≤ 100 ppm/K
Smallest measuring spanType L (Fe-CuNi), J, U, T, K, E, N, Chromel-Alumel: 100 K
Type S, R, B, C, D, A1, Chromel-Copel: 500 K
Measuring range start/endFreely programmable within the limits in steps of 0.1 K
Cold junctionInternal (Pt100) or external (constant)
Reference point accuracy (inter-
± 1K
nal)
Reference point temperature
-30 to +85 °C (adjustable)
(external)
Sampling rate3 or 6 channels: 125 ms
Input filterDigital filter, 2nd order; filter constant can be set from 0 to 100.0 s
Galvanic isolationSee "Galvanic isolation"
Base measuring range20 to 70 mV
a
The accuracy value refers to the maximum measuring range. Small measuring ranges lead to reduced linearization accuracy.
a
RTD temperature probe
DescriptionStandardITSConnection type Measuring rangeAccuracy
Pt50IEC 751: 2008ITS-902-/3-/4-wire-200 to +850 °C≤ 0.1 %500 μA
Pt100IEC 751: 2008ITS-902-/3-/4-wire-200 to +850 °C≤ 0.1 %500 μA
Pt500IEC 751: 2008ITS-902-/3-/4-wire-200 to +850 °C≤ 0.1 %100 μA
Pt1000IEC 751: 2008ITS-902-/3-/4-wire-200 to +850 °C≤ 0.1 %100 μA
Pt100JIS 16042-/3-/4-wire-200 to +650 °C≤ 0.1 %500 μA
Pt50GOST 6651-2009 A.2ITS-902-/3-/4-wire-200 to +850 °C≤ 0.1 %500 μA
Pt100GOST 6651-2009 A.2ITS-902-/3-/4-wire-200 to +850 °C≤ 0.1 %500 μA
Cu50GOST 6651-2009 A.3ITS-902-/3-/4-wire-180 to +200 °C≤ 0.4 %500 μA
Cu100GOST 6651-2009 A.3ITS-902-/3-/4-wire-180 to +200 °C≤ 0.4 %500 μA
Ni100DIN 43760ITPS-682-/3-/4-wire-60 to +250 °C≤ 0.2 %500 μA
Ni100GOST 6651-2009 A.5ITPS-682-/3-/4-wire-60 to +180 °C≤ 0.2 %500 μA
Ambient temperature influence≤ 50 ppm/K
Smallest measuring span15 K
Sensor lead wire resistanceMax. 10 Ω per lead for two-wire circuit
Max. 30 Ω per lead for three/four-wire circuit
Measuring range start/endFreely programmable within the limits in steps of 0.1 K
Sampling rate3 or 6 channels: 125 ms
Input filterDigital filter, 2nd order; filter constant can be set from 0 to 100.0 s
Galvanic isolationSee "Galvanic isolation"
a
The accuracy value refers to the maximum measuring range. Small measuring ranges lead to reduced linearization accuracy.
a
Measuring
current
DS-DV6-D1-UK-1701
DS-DV6-1-US-1804
Resistance transmitter and resistor/potentiometer
DescriptionMeasuring rangeAccuracy
a
Measuring current
Resistance transmitter0 to 4000 Ω≤0.1 %100 μA
Resistance/potentiometer0 to 400 Ω≤0.1 %500 μA
0to4000Ω≤0.1 %100 μA
Ambient temperature influence≤ 100 ppm/K
Connection type
Resistance transmitterThree-wire circuit
Resistance/potentiometerTwo/three/four-wire circuit
Smallest measuring span60 Ω
Sensor lead wire resistanceMax. 10 Ω per cable for two-wire and three-wire circuits
Resistance valuesFreely programmable within the limits in steps of 0.1 Ω
Sampling rate3 or 6 channels: 125 ms
Input filterDigital filter, 2nd order; filter constant can be set from 0 to 100.0 s
Galvanic isolationSee "Galvanic isolation"
a
The linearization accuracy value refers to the maximum measuring range. Small measuring ranges lead to reduced linearization accuracy.
Voltage, current (standard signals)
DescriptionMeasuring rangeAccuracy
Voltage0to70mV≤ 0.1 %> 500 kΩ
0to10V≤ 0.05 %> 500 kΩ
-10 to +10 V≤ 0.05 %> 500 kΩ
-1 to +1 V≤ 0.08 %> 500 kΩ
0to1V≤ 0.08 %> 500 kΩ
Current4to20mA≤ 0.1 %< 2V
0to20mA≤ 0.1 %< 2V
a
Input resistance or burden
voltage
Ambient temperature influence≤ 100 ppm/K
Smallest measuring span
5 mVVoltage
Current0.5 mA
Measuring range start/end
Voltage
Current
Deviation below/above the measuring
Freely programmable within the limits in steps of 0.01 mV
Freely programmable within the limits in steps of 0.01 mA
According to NAMUR recommendation NE 43 (only current input 4 to 20 mA)
range
Sampling rate3 or 6 channels: 125 ms
Input filterDigital filter, 2nd order; filter constant can be set from 0 to 100.0 s
Galvanic isolationSee "Galvanic isolation"
a
The accuracy value refers to the maximum measuring range. Small measuring ranges lead to reduced linearization accuracy.
Measuring circuit monitoring
The device response in the event of a fault is configurable.
Measuring probeProbe breakShort-circuitPolarity
Thermocoupleis detectedis not detectedis detected in certain condi-
RTD temperature probeis detectedis detectedis not detected
Resistance transmitteris detectedis not detectedis not detected
Resistance/potentiometeris detectedis not detectedis not detected
Voltage 0to 70 mVis detectedis not detectedis detected
Voltage 0 to 10 Vis not detectedis not detected is detected
Voltage -10 to +10 Vis not detectedis not detectedis not detected
Voltage 0to 1 Vis detectedis not detectedis detected
Voltage -1 to +1 Vis detectedis not detectedis not detected
Current 0 to 20 mAis not detectedis not detectedis not detected
Current 4 to 20 mAis detectedis detectedis detected
a
dependent on the set characteristic line
tions
a
DS-DV6-1-US-1804
Digital inputs (options 1 and 2)
Quantity0, 6, or 12
Connector number (back of device)6 and 10
Input