Omron MC-521-E Product Information Sheet

Page 1
X
New Product Information Sheet
Specifications subject to change
© OMRON HEALTHCARE EUROPE B.V.
Sales Name: Gentle Temp 521
Item Number: MC-521-E
Table of Contents
Page
General Information
Sales name and Item number of product 1 Model/Category of the product 1 Legal Manufacturer, Production Facilities 1 Photo of Product 1 Packaging Content (including accessories) 2
List of Harmonized EN standards` 2
Intended Use 3
Particularity’s
How it works 4 Efficiency of the medial product 5 Description of the design and test procedure (Application) 6
Warnings 7-8
Specifications 8
Materials Used 9
Title page of Instruction Manual 10
Packages Design 11
Operation and Care / Storage / Servicing / Repair 12
Disposal / Warranty / Lifetime 13
Comparison with Previous Product / Comparative Table 14-15
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33 2132 LR Hoofddorp The Netherlands
P.O. Box 2050 2130 GL Hoofddorp The Netherlands
Phone: +31 (0)23 55 44 700 fax +31 (0)23 55 44 701
BANKERS BANK OF TOKYO-MITUSBISHI (HOLLAND ) N.V.
CHAMBER OF COMMERCE NR. 342.10306
www.omron.healthcare.com
SWIFT BOTKNL 2
BTW/VAT NR. NL8132.86.281.B 01
Page 2
Infrared Ear Thermometer
Omron Gentle Temp 521 (MC-521-E)
LEGAL MANUFACTURER
Address: OMRON HEALTHCARE Co., Ltd 53, Kunotsubo, Terado-cho, Muko, Kyoto 617-0002 Japan Phone/Fax: +31 23 5544700 / +31 23 5544 701 e-mail: info@omronhealthcare@eu.omron.com
PRODUCION FACILITIES
Address: Radiant Innovation Inc. No. 20, TaiHongRoad, WuSong Jiang Development Zone YuShan Town, KunShan City, Jiang Su, China Phone/Fax: +31 23 5544700 / +31 23 5544 701 e-mail: info@omronhealthcare@eu.omron.com
Ear Thermometer Omron Gentle Temp 521 is a medical device meant for
measurement of temperature. The thermometer detects the infrared heat given off by the eardrum and surrounding tissues, and it converts this heat into an equivalent ear temperature. The Gentle Temp 521 also measures the surface temperature of an item for example baby’s milk. The device also shows the room temperature when turned on.
2
PM-1157-01-07/2013
Page 3
The manufacturer produces the ear thermometer Omron Gentle Temp 521 with the applicable accessories included, necessary for the application of its intended purpose.
1. Main unit
2. Instruction manual
3. Test battery
4. Probe cap
5. 21 Probe covers
6. Connection Ring
7. Warranty card
The complete set of the medical device is specified in the Instruction Manual in the Technical Data section
The above mentioned medical device belongs to a IIa (MDD Annex IX Rule 10) potential risk class.
Applicable Directives: Medical Device Directive (MDD) 93/42/EEC
Standards: EN980:2008 EN1041:2008 EN60601-1:2006 EN60601-1-2:2007 EN60601-1-6:2007 EN ISO14971:2012 EN ISO 10993-1:2009 EN ISO 10993-5:2009 EN ISO 10993-10:2010 EN62304:2006 EN62366:2008 EN12470-5:2003
3
PM-1157-01-07/2013
Page 4
Intended Use
Medical product’s intended use as determined by the manufacturer.
The device is a digital ear thermometer to measure the infrared heat given off by the eardrum and surrounding tissues, and converts this heat into an equivalent ear temperature.
Application field:
This device is designed for use in a physician’s office, hospitals, clinics and other medical facilities as well as home use.
User:
This device may be used by a medical professional or adult on themselves or children. The device is intended for use on adults and children as well as babies.
Intended Use of the Probe Covers
Medical product’s intended use as determined by the manufacturer.
The probe cover is designed to be used on the Omron Gentle Temp 520 digital ear thermometer. It protects the sensor from dirt and ear wax to ensure accurate measurements.
Application field:
This device is designed for use in a physician’s office, hospitals, clinics and other medical facilities as well as home use.
User:
This device may be used by a medical professional or adult on themselves or children. The device is intended for use on adults and children as well as babies.
4
PM-1157-01-07/2013
Page 5
Particularities
OMRON is introducing the Gentle Temp 521 ear thermometer. It uses a highly accurate technology and includes some of its advanced features, including one­second measurement, selectable centigrade/Fahrenheit readings and a memory for up to 25 readings. It also comes with 21 probe covers. OMRON recommends changing the covers each time to ensure good hygiene and the most accurate measurements. To assist with readings in the dark (on a ward or in a child’s room at night) the device also has backlight function. The 3 in 1 function enables various types of measurement, ear, surface and room temperature.
The Gentle Temp 521 is a supplement to our current EFT portfolio.
Key features:
3 in 1 measurement (ear, surface, room temp. measurement)
Fast 1 second measurement
25 memories Backlight function 21 probe covers Selectable Celsius/Fahrenheit New ergonomic design Large display Automatic shut-off
The digital ear thermometer Omron Gentle Temp 521 measures the infrared heat given off by the eardrum and surrounding tissues, and converts this heat into an equivalent ear temperature. For surface and room temperature the device measures the infrared heat given off by the air or surface.
Efficient measurement of the body temperature is one of the most important element of disease control. The device allows the care giver to easily and accurately measure the patient’s body temperature. The infrared method is used for detecting the temperature.
How it works
Efficiency of the medical produc t
5
PM-1157-01-07/2013
Page 6
Description of the design and test procedure
Infrared Sensor Probe Probe cap Probe cover detector Display ON/MEM button Battery cover pick hole Battery compartment Connection ring Probe cover
Application:
Review the Instruction Manual for the full details on the application of this product.
1. Chose which temperature indication (°C and °F)
2. Set the buzzer (to turn the buzzer on or off)
3. Attach the probe cover
4. Press ON/MEM button
5. Insert probe comfortably into the ear in the direction of the eardrum
6. Press START button
7. At the long beep, the thermometer can be removed
8. Read the temperature measurement
9. Press and hold ON/MEM button to shut off
6
PM-1157-01-07/2013
Page 7
Warnings
• Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
• Conducting self-diagnosis based on the measurement results and/or treatment can be dangerous. Please follow the instructions of your doctor. Self-diagnosis may worsen the symptoms.
• A high or prolonged fever requires medical attention, especially for small children. Please contact your doctor.
• Please keep still during measurement.
• Do not forcibly insert the probe in the ear.
• If you feel discomfort such as a pain during the measurement, stop using the unit immediately. It may injure the external auditory canal.
• Do not use this unit if suffering from ear disease such as otitis externa or otitis media. It may worsen the condition.
• Do not use this unit when the external auditory canal is wet such as after swimming or taking a bath. It may injure the external auditory canal.
• Do not use this unit without attaching a probe cover.
• Please ensure that the ear canal is clean and free of earwax.
• If the probe cover becomes dirty with earwax or other substances, replace it with a new one.
• Do not use a probe cover after someone else has used it. This can lead to cross infections such as otitis externa.
• Correct measurement result may not be obtained if dirty probe covers are used.
• Proper installation of the probe cover ensures accurate measurements.
• When the infrared sensor becomes dirty, wipe it lightly with a soft dry cloth or a cotton swab. Do not wipe the infrared sensor with tissue paper or a paper towel.
• Do not use more than one probe cover at a time.
• If there is any temperature difference between the places where the unit is stored and where you are going to measure, leave the unit in the room where you are going to use it for more than 30 minutes to allow it to reach room temperature first, then measure.
• If the ear is cold, wait until the ear is warmed up before taking a temperature measurement. The measured result may indicate low when you use an ice bag or an ice pack or immediately after coming in from the outside in winter.
• Do not touch the infrared sensor with a finger or breathe on it.
• Do not attempt measurements when the unit is wet as inaccurate readings may result.
• Check the symbol on the display before and after the measurement so that the measurement is taken in the appropriate mode.
• When taking temperature of the subject with low emissivity such as gold or aluminum inaccurate readings may result.
• Keep the unit out of children’s reach.
• Avoid children trying to measure themselves or others as they may damage the ear.
• In an emergency case, if a child swallows a battery or a probe cover, immediately consult with a doctor.
• Do not throw batteries into a fire. The battery may explode.
• Remove the battery when the unit will not be used for 3 months or more. Failure to do so may lead to fluid leakage, heat generation or bursting, resulting in damage to the unit.
• Do not use the unit in places where strong static electricity or electromagnetic fields are present. Doing so may lead to inaccurate readings and may contribute to instrument failure.
• Do not use the unit to measure any item with a temperature higher than 80.0°C (176.0°F).
7
PM-1157-01-07/2013
Page 8
(
)
General Precautions
• Do not apply a strong shock to, drop, step on, or vibrate the main unit.
• Do not use a mobile phone near the unit.
• The main unit is not waterproof. Be careful when handling this unit so that no liquid (alcohol, water, or hot water) will get into the main unit. When the unit is wet with vapor, wait until it dries or wipe it lightly with a soft dry cloth.
• Do not disassemble, repair, or modify the unit.
• When you inform your doctor of your temperature, make sure you state that you measured
the temperature in the ear.
Specifications of OMRON Gentle Temp 521 (MC-521-E)
Product Description: Digital Ear Thermometer Model: OMRON Gentle Temp 521 (MC-521-E) Sensing Unit: Thermopile Temperature Display:
Measurement Accuracy: Ear Measurement Mode:
Measurement Range: Power Supply: 3.0V DC, 1 CR2032 Lithium Button Battery Power Consumption 0.015 W Battery Life With a new battery approx. 2,500 measurements or
Operating environment Temp and Humidity:
Storage environment Temp and Humidity:
Protection against electric shock: Internally powered ME equipment Applied Part:
4-digit, F display in 0.1 degree increments 3-digit, C display in 0.1 degree increments
± 0.2°C (± 0.4°F) within 35.5°C to 42.0°C (95.9°F to 107.6°F), ± 0.3°C (± 0.5°F) for other range Surface Mode: ±0.3°C (± 0.5°F) within 22.0°C to 42.2°C (71.6°F to 108.0°F), for other range it is ±2°C (± 3.6°F) or 4% whichever is greater.
34.0C (93.2F) to 42.2C (108.0F)
more
Ambient environment 25 ± 15C, 50±40% RH
10 °C (50°F) to 40°C (104°F), 0 RH 85%
-20 °C (-4°F) to 50°C (122°F), 0 RH 85%
Weight: Approx. 85 g (including battery installed) Outer Dimensions: 36 mm (w) x 161 mm (h) x 56 mm (d) Package Contents: Test Battery (Lithium Button Battery CR2032), probe
Options: Probe cover exclusively for Gentle Temp 521, MC-EP2
= Type BF
cap, 21 probe covers, connection ring, instruction manual, warranty card
8
PM-1157-01-07/2013
Page 9
Materials Used
Component Material
CAP Polycarbonates Plastic
Battery Cover Acrylonitrile - Butadiene - Styrene Plastic
Cosmetic cover Acrylonitrile - Butadiene - Styrene Plastic
Rear Cover Acrylonitrile - Butadiene - Styrene Plastic
Action Key Polypropylene Plastic
Front Cover Acrylonitrile - Butadiene - Styrene Plastic
Power Button Silicone Rubber
Probe cover ZIP Lock Bags_ LLDPE Film + HDPE Ring
Probe cover Detect Polyoxymethylene, polyacetal
9
PM-1157-01-07/2013
Page 10
Title page of Instruction Manual
(9 languages – English / French / German / Italian / Spanish / Dutch /
Russian / Turkish / Arabic
10
PM-1157-01-07/2013
Page 11
OMRON Gentle Temp 521 (MC-521-E) Package Design
Packaging specifications :
Main Unit
Package
Master Carton
Dimensions WxDxH Qty Weight
36x56x161mm - 85g
115x59.5x197mm 1 343g
497x420x340 40 15KG
11
PM-1157-01-07/2013
Page 12
Operation and Care
Please check the device if damaged after it is dropped. If unsure, please contact the OMRON retail outlet or distributor for having the device checked.
Surface cleaning
Wipe the external surface with a soft dry cloth. The probe is the most delicate part of the unit. Use care when cleaning the Infrared sensor to avoid damage.
Storage
• Do not store the unit in the following types of places. Doing so may damage the unit.
- Wet locations.
- Locations with high heat and humidity or those that are exposed to direct sunlight. Areas close to heating equipment, dusty locations, or environments where there are high salt concentrations in the air.
- Locations where the unit will be subjected to leaning over, falling, shock or vibration.
- Pharmaceutical storage areas or locations where corrosive gases are present.
Servicing
The device requires no routine service other than cleaning, and visually checking for damage.
Repair
In case of malfunction of the device the user must turn to the OMRON retail outlet or distributor to test the device and repair the failure. The user shall not repair the device by themselves.
Warranty card text as included in the product
This product is guaranteed by OMRON for a period of 3 years after the date of purchase. The proper construction, workmanship and materials of this product is guaranteed by OMRON. During this period of guarantee OMRON will, without charge for labour or parts, repair or replace the defect product or any defective parts.
The guarantee covers only products purchased in Europe, Russia and other CIS countries, Middle East and Africa.
The guarantee does not cover any of the following: a. Transport costs and risks of transport. b. Costs for repairs and / or defects resulting from repairs done by unauthorized persons. c. Periodic check-ups and maintenance. d. Failure or wear of accessories or other attachments other than the main device itself, unless explicitly guaranteed above.
12
PM-1157-01-07/2013
Page 13
e. Costs arising due to non-acceptance of a claim (those will be charged for). f. Damages of any kind including personal caused accidentally or from misuse.
Disposal
Description
As there is a risk of environmental pollution, follow your applicable national and local legal regulations regarding disposal or recycling of this equipment and batteries. The main constituents of each part are listed in the table “Materials List”’ on page 9.
Warranty
Warranty term for the ear thermometer- 3 years.
Warranty term for probe cover – None as this is a single use item.
Lifetime
Lifetime of the ear thermometer. 5 years
Life of the probe cover: no lifetime as these are considered disposable items
Warranty terms are specified in the warranty card inside the packaging of the Omron Gentle Temp 521.
13
PM-1157-01-07/2013
Page 14
(
Comparison with Previous Product
The Gentle Temp 521 (MC-521-E) with the legal manufacturer of OMRON HEALTHCARE Co., Ltd at the production location of Radiant No. 20, TaiHongRoad, WuSong Jiang Development Zone, YuShan Town, KunShan City, Jiang Su, China, is an substantially equivalent to a similar device MC-510-E2.
The device under registration is equivalent to the earlier registered equivalent device in the context of its intended use, field of application, technology, quality, reliability, main specifications, lifetime, operating and storage conditions, safety, functional peculiarities, efficiency and other parameters.
Comparative table of the product with the Previous Model
Medical device under registration
Pictures of the devices
Earlier registered analogue
Product Description:
Model: OMRON Gentle Temp 521
Sensing Unit: Thermopile Thermopile
Temperature Display:
Measurement Accuracy:
Digital Ear Thermometer Digital Ear Thermometer
Omron Gentle Temp 510 (MC-510-E2)
(MC-521-E)
4-digit, F display in 0.1 degree increments 3-digit, C display in 0.1 degree increments
± 0.2C (± 0.4F) within 35.5C to
42.0C (95.9F to 107.6F) ± 0.3C
± 0.5F) forother range
In 3 digits of +C, every 0.1 F or C
± 0.2C (± 0.4F) within 34.0C to
42.2C (93.2F to 107.9F)
PM-1157-01-07/2013
14
Page 15
y
y
y
Measurement Range: Power Supply: 3.0V DC, 1 CR2032 Lithium Button
Power Consumption Battery Life With a new battery approx. 2,500
Operating environment Temp and Humidit Storage environment Temp and Humidity: Protection against electric shock:
Applied Part:
Weight: Approx. 85 g (including battery
Outer Dimensions: Package Contents:
Options: Probe cover exclusively for Gentle
:
34.0C (93.2F) to 42.2C (108.0F) 34.0C (93.2F) to 42.2C (108.0F)
3.0V DC, 1 CR2032 Lithium Button
Batter
0.015 W 0.05W
measurements or more (Ambient environment 25 ± 15C, 50±40% RH) 10 °C (50°F) to 40°C (104°F) 0 RH 85%
-20 °C (-4°F) to 50°C (122°F), 0 RH 85%
Internally powered ME equipment Internally powered ME equipment
installed) 36 mm (w) x 161 mm (h) x 56 mm (d) 46 mm (w) x 93 mm (h) x 57 mm (d)
Test Battery (Lithium Button Battery CR2032), probe cap, 21 probe covers, connection ring, instruction manual, warranty card
Temp 521, MC-EP2
= Type BF = Type BF
Batter
With a new battery approx. 5000 measurements
10 °C (50°F) to 40°C (104°F) 30 - 85% RH
-20 °C (-4°F) to 60°C (140°F) 10 - 95% RH
Approx. 50 g (including battery installed)
Main unit, Storage case, 10 probe covers, instruction manual, warranty card, information sheet (2 pieces)
Probe cover exclusively for Gentle Temp 510, MC-EP1
On the basis of protocol of tests for conformity to standards requirements, technical specifications of medical devices and the comparative table above we can make a conclusion that the functional peculiarities, safety, efficiency and quality of the new device are equivalent to the previous device.
15
PM-1157-01-07/2013