OMRON products are manufactured for use according to proper procedures by a qualified operator
and only for the purposes described in this manual.
The following conventions are used to indicate and classify precautions in this manual. Always heed
the information provided with them. Failure to heed precautions can result in injury to people or damage to the product.
DANGERIndicates information that, if not heeded, is likely to result in loss of life or serious injury.
!
WARNINGIndicates information that, if not heeded, could possibly result in loss of life or serious injury.
!
CautionIndicates information that, if not heeded, could result in relatively serious or minor injury,dam-
!
age to the product, or faulty operation.
OMRON Product References
All OMRON products are capitalized in this manual. The word “Unit” is also capitalized when it refers
to an OMRON product, regardless of whether or not it appears in the proper name of the product.
The abbreviation “Ch,” which appears in some displays and on some OMRON products, often means
“word” and is abbreviated “Wd” in documentation in this sense.
The abbreviation “PC” means Programmable Controller and is not used as an abbreviation for anything else.
Visual Aids
The following headings appear in the left column of the manual to help you locate different types of
information.
Ó OMRON, 1998
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any
form, or by any means, mechanical, electronic, photocopying, recording, or otherwise, without the prior written permission of OMRON.
No patent liability is assumed with respect to the use of the information contained herein. Moreover, because OMRON is
constantly striving to improve its high-quality products, the information contained in this manual is subject to change
without notice. Every precaution has been taken in the preparation of this manual. Nevertheless, OMRON assumes no
responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of the information contained in this publication.
Note Indicates information of particular interest for efficient and convenient operation
of the product.
1, 2, 3...1. Indicates lists of one sort or another, such as procedures, checklists, etc.
v
Page 4
vi
Page 5
TABLE OF CONTENTS
PRECAUTIONSxi.................................
1 General Precautionsxii..........................................................
This manual describes the installation and operation of the K3NP Period Meter and includes the sections
described below.
Please read this manual carefully and be sure you understand the information provided before attempting
to install and operate the K3NP.
Section 1 describes the functions of the K3NP. The main components are also described.
Section 2 provides instructions required for mounting and wiring the K3NP.
Section 3 provides instructions for setting the parameters of the K3NP.
Section 4 provides instructions for operating the K3NP in RUN mode.
Section 5 provides information on the teaching function, output test, and maintenance mode.
Section 6 provides information on the use of the K3NP with the BCD Output Board.
Section 7 provides information for troubleshooting the K3NP
The Appendices provide specifications, a list of settings, a list of standard models, and a list of available
menu items.
!
WARNING Failure to read and understand the information provided in this manual may result in
personal injury or death, damage to the product, or product failure. Please read each
section in its entirety and be sure you understand the information provided in the section
and related sections before attempting any of the procedures or operations given.
ix
Page 8
PRECAUTIONS
This section provides precautions for using the K3NP Period Meter and related devices.
The information contained in this section is important for the safe and reliable application of the K3NP .You must read
this section and understand the i nformation contained before attempting to set up or operate the K3NP.
1 General Precautionsxii..........................................................
The user must operate the product according to the performance specifications
described in the operation manuals.
Before using the product under conditions which are not described in the manual
or applying the product to nuclear control systems, railroad systems, aviation
systems, vehicles, combustion systems, medical equipment, amusement machines, safety equipment, and other systems, machines, and equipment that
may have a serious influence on lives and property if used improperly, consult
your OMRON representative.
Be sure to read this manual before attempting to use the product and keep this
manual close at hand for reference during operation.
2Safety Precautions
WARNING Never attempt to disassemble any Units while power is being supplied. Doing so
!
may result in serious electrical shock or electrocution.
WARNING Never touch any of the terminals while power is being supplied. Doing so may
!
result in serious electrical shock or electrocution.
3SectionApplication Precautions
3Application Precautions
Observe the following precautions when using the product.
· Always use the power supply voltage specified in the specifications.
· Do not use the product in locations subject to flammable gases or combustible
objects.
· Be sure to confirm terminal names when wiring.
· Be sure to tighten the screws on the terminal blocks.
Observe the following precautions when mounting the product.
· Mount the product on level surfaces.
· Mount the product on a panel which has a thickness of 1 to 3.2 mm.
Do not mount the product in the following places.
· Locations subject to strong shock or vibration.
· Locations subject to temperature or humidity exceeding the rated levels or
where icing is liable to occur.
· Locations subject to dust.
· Locations subject to corrosive gases (particularly sulfuric gases or ammonium
gases).
· Locations subject to direct sunlight or outdoor conditions.
· Locations near devices (high-frequency welders or high-frequency sewing
machines) that produce high-frequency noise.
xii
Page 10
4Noise Prevention
Provide the following countermeasures when using the product in an environment where the product is exposed to noise.
· Countermeasures for protecting the product against high-frequency noise or
abnormal voltages.
4SectionNoise Prevention
Line filter
Period Meter
Signal
input
+++
Period Meter
--
Power
input
Power
input
· Countermeasures for protecting the product against inductive noise produced
from the input line.
+
Period Meter
--
2-conductor shield wire
xiii
Page 11
SECTION 1
Introduction
This section describes the functions ofthe K3NP. The main components arealso described. Referto the remaining sections of
this manual for the operation of the K3NP and its menus in detail.
The K3NP Period Meter displays pulse input after performing the numeric conversion of the input.
The K3NP has the following functions.
MeasurementThe internal system clock counts the period between pulses or pulse-ON time
and displays it in a variety of modes.
In the following example, the internal system clock counts the time (T) between
pulse INA and pulse INB. If there are 100,000 counts during the time (T), a passing speed (F1) of 1,200 (m/min) will be displayed according to the following calculation.
T = 1 system clock count (0.5 ms) x 100,000 = 0.05 s
F1 = 1/T x 60 (m/min) according to the internal calculation method.
Therefore, 1/0.05 s x 60 = 1,200 (m/min)
Example: Passing Speed F1
Internal system
clock
PrescalingInput pulses are converted into a value within a specified range.
Comparative Output
Selection
Linear Output
BCD OutputA digital data output format where every four binary bits is numerically equivalent
Communications Output
HOLD
RESET
Output TestThis function is convenient for checking a system to which the K3NP is con-
Comparison output patterns can be selected from the standard, level, or zone
output depending on the application.
Refer to Comparative Output Patterns, page 61.
Refer to Linear Output Range, page 64.
to one decimal digit.
Refer to Section 7 BCD Output.
Refer to the Communications Manual.
HOLD is an external input which is used to stop the A/D process and freeze the
display. The comparative, linear, and BCD outputs are also retained.
Refer to 5-3 External Input Signals for details.
RESET is an external input to reset to the initial state and the present maximum
and minimum values. Resetting can be performed using the front panel keys.
Refer to pages 85 and 86.
nected, especially when some inputs cannot be operated. The K3NP simulates
an input to check the output conditions.
Refer to 6-2 Test Mode for details.
Remote/Local SelectionThe K3NP can be operated remotely through a host computer or locally with key
inputs.
Remote Mode: For programming remotely by downloading setup parameters
from a host computer via RS-232C, RS-485, or RS-422.
Local Mode:Programming is performed with the front panel key input.
Refer to Remote/Local Programming, page 67.
2
Page 13
Front of the Meter
1-2Front of the Meter
Section 1-2
SV display status
indicators
Escape Key
Comparative output
status indicators
Mode Key
Up Key
PV display
Status indicators
Teaching indicator
Unit of measure
SV display
RESET/TEACH Key
Shift Key
Five-digit (0 to 99999), seven-segment, 14.2-mm-high LED display with a programmable decimal point.
The displays show the process value, maximum value, minimum value, operations/parameters when setting, and error messages.
PV DisplayK3NP-jjjA Basic Model
RUN Mode:Displays the process, maximum, and minimum values. Also
displays setting values while the SV indicator is lit. When
changing a value, all digits other than those that can be set become dimmer.
Setting Mode: Displays the menu, parameter, or setting value. When chang-
ing a value, all digits other than those that can be set become
dimmer.
K3NP-jjjC Set Value LED Display Model
RUN Mode:Displays the process, maximum, and minimum values.
Setting Mode: Displays the menu and parameters.
SV Display (Setting value
LED Display Models Only)
RUN Mode:Displays comparative set values setting values. When chang-
ing a value, all digits other than those that can be set become
dimmer.
Setting Mode: Displays setting values. When changing a value, all digits other
than those that can be set become dimmer.
Comparative Output Status
Indicators
Indicates the status of the comparative output.
3
Page 14
Front of the Meter
Section 1-2
Status IndicatorsHOLD Indicator
Lit when the HOLD input signal is ON.
MAX Indicator
Lit when the value displayed on the PV display is the maximum value.
MIN Indicator
Lit when the value displayed on the PV display is the minimum value.
PROG Indicator
Lit when the setting mode menu is displayed. The indicator flashes while parameters are displayed.
Teaching IndicatorLit when displayed parameters can be set in teaching operation. The indicator
flashes when the process value is indicated as a setting value.
SV Display Status Indicators Indicates which set value is on the PV or SV display.
Unit of MeasureAttach the appropriate label showing the unit of measure (enclosed).
Escape Key
Mode Key
Process value
Maximum value
Minimum value
Used to select the process, maximum, or minimum value to be displayed on the
PV display in RUN mode.
Process valueMaximum valueMinimum value
ESCESC
Used to return from the setting, protect, or maintenance mode to the RUN mode.
This key is also used to return to the previous operation during the setting, protect, or maintenance mode.
Displays a setting value (out of HH, H, L, and LL setting values in this order) on
the PV display in RUN mode when this key is pressed. Unless another operation
key is pressed within five seconds after this key has been pressed, the display
automatically changes to the one for process values.
HH is lit.
HH setting
value
No key input
for 5 seconds.
H is lit.L is lit.LL is lit.
H setting
value
No key input
for 5 seconds.
L setting
value
No key input
for 5 seconds.
LL setting
value
No key input
for 5 seconds.
In the RUN mode, this button terminates the measurement process and allows
you to enter the setting mode, advancing through the menus and parameters.
Menu 1Menu 2
Parameter 1Parameter 2Parameter n
Menu n
In the setting mode, this button will store changes in the non-volatile memory
while at the same time advancing the display to the next menu item.
4
Page 15
Front of the Meter
Section 1-2
Up Key
Shift Key
Used to select a parameter to be displayed for setting value change.
Used to increment the current digit in the setting value by one.
The value increases in the following order:
0, 1, 2, 3, 4, 5, 6, 7, 8, 9
Used to change the parameter displayed in setting mode.
Used to scroll the digit to the right of the presently displayed digit.
RESET/TEACH Key
Used to reset the K3NP to its initial state or to reset the max./min. value in RUN
mode.
Used to select the teaching function. Refer to 6-1 Teaching Function for details.
5
Page 16
Rear of the Meter
1-3Rear of the Meter
Terminal arrangement varies depending on the selected Output Board.
For wiring, refer to Section 2 Setup.
K3NP with Relay Output Board, K31-C1, -C2, -C5
K3NP with Transistor Output Board, K31-T1, -T2
K3NP with Linear Output Board, K31-L1, -L2, -L3, -L4, -L5, -L6, -L7, -L8, -L9, -L10
K3NP with RS-485 Output Board, K31-FLK2, -FLK5
Section 1-3
Output Board
Input Board
K3NP with BCD Output Board, K31-B2, -B4
K3NP with RS-232C Output Board, K31-FLK1
Output Board
Input Board
Communications
Output Board
Input Board
6
Page 17
Modes
K3NP with RS-422 Output Board, K31-FLK3
Communications
Output Board
Terminator
K3NP with RS232C + Transistor Output Board, K31-FLK4
K3NP with RS-422 + Transistor Output Board, K31-FLK6
Communications
Output Board
Terminator
Section 1-4
Input Board
Transistor
Output Board
1-4Modes
Input Board
The following four modes are available.
· RUN mode for normal operations (see Section 5 Operations in RUN Mode)
· Setting mode for initializing parameter input (see Section4 ParameterSetting)
· Protect mode for lock-out configuration (see 4-1 Protect Mode)
· Maintenance mode for initialization (see 6-3-2 Initialization)
Refer to the following for the relationship among these modes and selection of
the modes.
Press the Mode and Shift Keys
while turning on the K3NP.
Maintenance mode
1 second
Power on
RUN mode
RUNmode
Press the Escape and up
Keys for 1 second.
Protect mode
Setting mode
RUN ModeK3NP is in RUN when the K3NP is turned ON.
The K3NP in this mode provides an output signal as a result of the comparison of
the measured and setting values.
7
Page 18
Communications Function
The basic model in this mode usually displays the process value. The maximum
and minimum values are displayed by pressing the Escape Key. The parameters and setting values are displayed by pressing the Mode Key.
Refer to Section 5 Operations in RUN Mode for RUN mode in detail.
Setting ModeValues are set in the K3NP in this mode by key input or using the teaching func-
tion.
Refer to Section 4 Parameter Setting for value setting by key input and 6-1Teaching Function for the teaching function in detail.
Protect ModeUse this mode to prohibit some operations in order to lock out the setting values.
Refer to 4-1 Protect Mode for details.
Maintenance ModeThe setting values are reset to factory-set values in this mode. Refer to 6-3-2
Initialization for details.
Section 1-5
1-5Communications Function
The communications function of the K3NP makes it possible for the host computer to perform the following operations.
· Confirmation and change of setting values. Communications conditions cannot be changed.
· Reading and resetting the maximum and minimum values.
· Forced-zero setting and resetting.
· Confirmation of model data.
Use a model with a Communications Board if the communications function is
required.
Refer to the Communications Manual for the communications function in detail.
RS-232CUse the K31-FLK1 or K31-FLK4 Output Board to use the RS-232C interface.
RS-422Use the K31-FLK3 or K31-FLK6 Output Board to use the RS-422 interface.
RS-485Use the K31-FLK2 or K31-FLK5 Output Board to use the RS-485 interface.
8
Page 19
This section provides instructions required for mount ing and wiring the K3NP.
Do not mount more than one Unit closely in the horizontal or vertical direction. Be
sure to keep the distance between adjacent Units.
Page 21
Input Block
Mounting Method
Section 2-2
Panel
1, 2, 3...1. Insert the K3NP into the mounting hole on the panel.
2. Hook the fixture claws onto the side holes.
3. Mount a fixing metal to the right and left sides as shown above and while
keeping them in balance, alternately tighten each screw until the ratchet becomes idle.
2-2Input Block
2-2-1Terminal Arrangement
Sensor power supply
(80mAmax.at12VDC)
100 to 240 to VAC
12 to 24 VDC
Note Voltage pulse input is available for the K3NP-NBjj-jj.
2-2-2Wiring Precautions
· Do not make any mistake in polarity when supply DC power to the K3NP.
· Do not wire power lines alongside the signal lines of the K3NP in order to pre-
vent the K3NP from noise interference.
Open collector
input
Voltage pulse
input
11
Page 22
Input Block
· Wire the terminal block with crimp terminals.
· Tighten each screw to a torque of 0.78 N S m(8kgfS cm).
2-2-3Wiring
Power SupplyApply 100 to 240 VAC or 12 to 24 VDC to terminals 8 and 9.
Open Collector InputConnect the pulse output from sensor A to terminal 1.
Connect the pulse output from sensor B to terminal 2.
Terminals 10 and 11 are exclusively used for a power supply with an output of
80 mA at 12 VDC to sensors A or B. If power is supplied to the sensors A or B
from a different power source, do not use terminal 10. Do not connect a sensor
with open collector output to terminal 10.
Refer to the following for sensor connections.
Section 2-2
Sensor (with NPN output)
+12 V
Sensor (with PNP output)
· Residual voltage with sensor turned on: 3 V max.
· Current leakage with sensor turned off: 1.5 mA max.
· Switching load current: 20 mA or greater. Must be able to dependable switch
a load current of 5 mA max.
Photoelectric sensors, proximity sensors, rotary encoders, and relays can be
connected as sensors to the K3NP.
Voltage Pulse InputConnect the pulse output of sensor A to terminal 3.
Connect the pulse output of sensor B to terminal 4.
Terminals 10 and 11 are exclusively used for a power supply with an output of
80 mA at 12 VDC to sensors A or B. If power is supplied to the sensors A or B
from a different power source, do not use terminal 10. Do not connect a sensor
with voltage pulse output to terminal 10.
Refer to the following for sensor connections.
+12 V
Sensor
+12 V
H level (sensor output ON): 4.5 to 30 VDC
L level (sensor output OFF): -- 30 to 2 VDC
Auxiliary Power SupplyTerminals 10 and 11 are exclusively used for power supply to sensors with an
output of 80 mA at 12 VDC ± 10%.
12
Page 23
Input Block
External Signal InputHOLD Input
RESET Input
BANK Input
Connect external signal inputs to terminals 5 through 7 and 13 through 15. Terminals 7 and 13 are connected to each other internally.
Connect HOLD input to terminal 5.
Connect RESET input to terminal 6.
Connect BANK inputs to terminals 14 and 15 for BANK1 and BANK2.
If open collector input is used as external signal input, the transistor must satisfy
the following conditions.
· Residual voltage with transistor turned on: 3 V max.
· Current leakage with transistor turned off: 1.5 mA max.
· Switching load current: 20 mA or greater.
Approximately 5 V is imposed between COM and terminals 5 to 7 with a current
flow of approximately 18 mA (a nominal value) at the time of external input shortcircuiting.
Section 2-2
Short-circuited together internally.
13
Page 24
Output Board
2-3Output Board
2-3-1Terminal Arrangement
K3NP with Relay Output Board, K31-C1, -C2, -C5
K3NP with Transistor Output Board, K31-T1, -T2
K3NP with Linear Output Board, K31-L1, -L2, -L3, -L4, -L5, -L6, -L7, -L8, -L9, -L10
K3NP with RS-485 Output Board, K31-FLK2, -FLK5
Section 2-3
K3NP with BCD Output Board, K31-B2, -B4
K3NP with RS232C + Transistor Output Board, K31-FLK4
14
Page 25
Output Board
K3NP with RS-422 + Transistor Output Board, K31-FLK6
K3NP with RS-232C Output Board, K31-FLK1
Section 2-3
K3NP with RS-422 Output Board, K31-FLK3
15
Page 26
Output Board
K3NP with RS-422 Output Board, K31-FLK3
2-3-2Relay Output Board
The following figures show the connections for relay output.
Section 2-3
K3NP with 3 Relay
Output Boards,
K31-C1
H
comparative
output
PASS
output
L
comparative
output
K3NP with 5 Relay
Output Boards,
K31-C5
K3NP with 5 Relay
Output Boards,
K31-C2
HH comparative
output
H comparative
output
PASS
output
L comparative
output
LL comparative
output
HH comparative
output
H comparative
output
PASS
output
L comparative
output
LL comparative
output
The following contact output conditions are required.
5 A (resistive load) at 250 VAC
1.5 A (inductive load) at 250 VAC
5 A (resistive load) at 30 VDC
1.5 A (inductive load) at 30 VDC
16
Page 27
Output Board
2-3-3Transistor and Combination Output Board
Section 2-3
K3NP with Transistor Output
Board, K31-T1 or K31-T2
K3NP with Linear Output Board,
K31-L4, -L5, -L6, -L9, -L10
K3NP with RS-485 + 5 Relay
Output Boards, K31-FLK5
HH comparative
21
output
H comparative
22
output
PASS output
23
L comparative
24
output
LL comparative
25
output
26
COM
The following transistor output conditions are required.
Maximum rated voltage: 24 VDC
Load current: 50 mA
Current leakage with transistor turned off: 100 mA.
K3NP with BCD Output
Board, K31-B2 or K31-B4
HH comparative
32
output
H comparative
33
output
PASS output
34
L comparative
35
output
LL comparative
36
output
COM
37
K3NP with RS232C + 5
Transistor Output Boards,
K31-FLK4
K3NP with RS-422 + 5
Transistor Output Boards,
K31-FLK6
HH comparative
1
output
H comparative
2
output
PASS output
4
L comparative
5
output
LL comparative
6
output
3
COM
2-3-4Linear Output Board
The following figures show connections for linear output.
4to20mA1to5V
The following linear output conditions are required.
Linear outputPermissible load
4to20mA600 W max.4096±0.5% FS
1to5V500 W min.4096±0.5% FS
1 mV/10 digit1kW min.4096±1.5% FS
2-3-5BCD Output Board
Refer to Section 7 BCD Output for the terminal arrangement and interface.
K3NP with Linear
Output Board,
K31-L1 or K31-L4
+
--
K3NP with Linear
Output Board,
K31-L2 or K31-L5
resistance
K3NP with Linear
Output Board,
K31-L3 or K31-L6
+
1 mV/10 digit
--
ResolutionOutput error
+
--
17
Page 28
Operating Modes
This section provides information on the ba sic operation of the K3NP in operating mode.
Measures the time taken by an object
to travel the distance between A and B.
Basic Operation
Displays the speed of an object passing between A and B. The speed is obtained
by multiplying the reciprocal of T by 60. T (time) is the time interval between INA
turning ON and INB turning ON. If the distance between the INA sensor and INB
sensor is 1 m, the displayed value is m/min. If the distance is other than 1 m, or if
another unit of measurement is required, input an appropriate prescale value.
Obtain display value D as follows:
D (m/min) = 1/T x 60 x a
T: Time interval between INA turning ON and INB turning ON (sec)
a: Prescale value
Hold Measured Value
When the HOLD input turns ON while the K3NP is measuring, the measuring
operation is canceled and the K3NP holds the last measured value (displayed
value). While the HOLD input is ON, the K3NP holds comparative output and
BCD output. Measurement will not begin while the K3NP is in HOLD status. This
is illustrated in the following diagram.
When the comparative output terminals of the output board and the HOLD
input terminals are connected, the value measured immediately after the occurrence of an error can be obtained.
20
Display
Value previously
measured
TR: Recovery time
Page 30
Passing Speed:
SETTING
REFERENCE
f1
Section 3-1
Refer to the following table to set the most suitable prescaling value.
ModeUnit of displayPrescale value
Passing speedmm/sec1000L/60
m/secL/60
m/minL
cm/sec100L/60
cm/min100L
km/h0.06L
Where,
L = Distance between sensors (m)
Refer to 4-2 Setting Mode.
Performance Characteristics
Accuracy of measurement
Measurement range10 ms to 3200 s
Response time of HOLD input20 ms max.
Recovery time (TR)20 ms min.
ON/OFF pulse width
Response timeOutput configuration
Relay outputTransistor
Comparative output 25 ms max.20 ms max.35 ms max.35 ms max.20 ms max .
Sensor with transistor output: 9 msmin.
Sensor with relay output: 15 ms min.
output
BCD and
transistor output
Linear and
transistor output
Communication
and transistor
output
The response time of the HOLD input is the time required for the K3NP to
accept HOLD input after the HOLD input turns ON.
Recovery Time (T
)Recovery time is the period required for the K3NP to become ready for the
R
next measuring operation after a measuring operation is completed. For
Transistor Output Models, comparative output remains ON during recovery
time. For other models, comparative output requires more time to turn ON.
When the comparative output terminals of the output board and the HOLD
input terminals are connected, set a sufficient recovery period, considering
the response time of the HOLD input. This is illustrated in the following diagram.
21
Page 31
(
)
Passing Speed:
f1
Operating Mode F1 (Transit Speed)
Section 3-1
Measurement
starts
T
Comparative output
Delay in comparative
outputs
HOLD input
Acceptance of
HOLD input
1
Measurement
completed
Recovery time
(20 ms min.)
T
2
Response time of HOLD input
(20 ms max.)
Available functions in this mode are indicated as “Yes” in the following table.
MenuFunctionDisplayed
Character
---Max./Min. value display and reset---Yes82
sUset (note 1)
Set value bank no. of set valuess.bank
HH set valuesU*.hh
HsetvaluesU*.h
LsetvaluesU*.l
LL set valuesU*.ll
pscl
Set value bank no. of prescale valuesp.bankYes
Prescaling value of input A
X (mantissa)
Y (exponent)
ps*.ax
ps*.ay
Decimal point positiondecp.*Ye s
setup
Operating modefuncYes50
Input A sensor typeinaYe s
Input B sensor typeinbYe s
Time unittimeNo54
Communications unit no. (note 1)u- noYe s
Baud rate (note 1)bpsYe s
Word length (note 1)lenYe s
Stop bits (note 1)sbitYes
Parity bits (note 1)prtyYes
opt
Output pattern (note 1)c-outYes61
H linear output range (note 1)lset.hYes
L Linear output range (note 1)lset.lYes
Remote/Local programming (note 1)r- lYes67
AvailabilityReference
Yes42
46
Yes
52
56
58
64
page
Note1. The availability of the parameters depends on the type of selected Output
Board.
2. The selected bank number will be displayed where the asterisk (
*) appears.
22
Page 32
Cycle:
f2
3-2Cycle: f2
Section 3-2
FUNCTION
Application example
Measures the cycle with
which items pass a sensor.
Basic Operation
Displays in seconds the interval between successive occurrences of INA turning
ON. Input an appropriate prescale value to display units other than seconds. Obtain display value D as follows:
D (sec) = T x a
T: Interval between successive occurrences of INA turning ON
a: Prescale value
By using
time in the time unit display setting parameter, the passing time can
be displayed in units of hours, minutes, and seconds.
Display
SETTING
T
1
Value previously
measured
To set a prescale value, refer to 4-2-4 Prescaling Menu.
T
2
T
1
T
3
T
T
4
2
T
5
T
5
Refer to the following table to set the most suitable prescaling value.
ModeUnit of display
value
Cyclesec1
min1/60
Prescale value
Where,
INB input is disregarded for operating mode 2.
Note The K3NP can display time, minute, and second. Refer to page 54 for details.
Refer to 4-2 Setting Mode.
REFERENCE
23
Page 33
(
)
Cycle:
f2
Performance Characteristics
Section 3-2
Accuracy of measurement
Measurement range20 ms to 3200 s
Response time of HOLD input20 ms max.
Recovery time (TR)20 ms min.
ON/OFF pulse width
Response timeOutput configuration
Relay outputTransistor
Comparative output 25 ms max.20 ms max.35 ms max.35 ms max.20 ms max .
Sensor with transistor output: 9 msmin.
Sensor with relay output: 15 ms min.
output
BCD and
transistor output
Linear and
transistor output
Communication
and transistor
output
The response time of the HOLD input is the time required for the K3NP to
accept HOLD input after the HOLD input turns ON. This is illustrated in the
following diagram.
Available functions in this mode are indicated as “Yes” in the following table.
MenuFunctionDisplayed
Character
---Max./Min. value display and reset---Yes82
sUset (note 1)
pscl
setup
opt
Set value bank no. of set valuess.bank
HH set valuesU*.hh
HsetvaluesU*.h
LsetvaluesU*.l
LL set valuesU*.ll
Set value bank no. of prescale valuesp.bankYes
Prescaling value of input A
X (mantissa)
Y (exponent)
Decimal point positiondecp.*Ye s
Operating modefuncYes50
Input A sensor typeinaYe s
Input B sensor typeinbNo
Time unittimeYes54
Communications unit no. (note 1)u- noYe s
Baud rate (note 1)bpsYe s
Word length (note 1)lenYe s
Stop bits (note 1)sbitYes
Parity bits (note 1)prtyYes
Output pattern (note 1)c-outYes61
H linear output range (note 1)lset.hYes
L Linear output range (note 1)lset.lYes
Remote/Local programming (note 1)r- lYes67
ps*.ax
ps*.ay
AvailabilityReference
Yes42
46
Yes
52
56
58
64
page
Note1. The availability of the parameters depends on the type of selected Output
Board.
2. The selected bank number will be displayed where the asterisk (
*) appears.
24
Page 34
Time Difference:
f3
3-3Time Difference: f3
Section 3-3
FUNCTION
Application example
Measures the time taken by an
object to pass between A and B.
Basic Operation
Displays the time (T) in seconds between INA turning ON and INB turning ON. If
another unit of measurement is required, input an appropriate prescale value.
Obtain the display value D as follows:
D (m/min) = T x a
T: The time (sec) between INA turning ON and INB turning ON
a: Prescale value
To set a prescale value, refer to 4-2-4 Prescaling Menu.
Display
SETTING
REFERENCE
Value previously
measured
T
R
T
: Recovery time
R
Refer to the following table to set the most suitable prescaling value.
ModeUnit of display
value
Time differencesec1
min1/60
Prescale value
Note The K3NP can display time, minute, and second. Refer to page 54 for details.
Refer to 4-2 Setting Mode.
25
Page 35
Time Difference:
f3
Section 3-3
Performance Characteristics
Accuracy of measurement
Measurement range10 ms to 3200 s
Response time of HOLD input20 ms max.
Recovery time (TR)20 ms min.
ON/OFF pulse width
Response timeOutput configuration
Relay outputTransistor
Comparative output 25 ms max.20 ms max.35 ms max.35 ms max.20 ms max .
Sensor with transistor output: 9 msmin.
Sensor with relay output: 15 ms min.
output
BCD and
transistor output
Linear and
transistor output
Communication
and transistor
output
Response Time of HOLD
Input
The response time of the HOLD input is the time required for the K3NP to
accept HOLD input after the HOLD input turns ON. This is illustrated in the
following diagram.
Recovery Time (TR)Recovery time is the period required for the K3NP to become ready for the
next measuring operation after a measuring operation is completed. For transistor output Models, comparative output remains ON during recovery time.
For other models, comparative output requires more time to turn ON. When
the comparative output terminals of the output board and the HOLD input
terminals are connected, set a sufficient recovery period, considering the response time of the HOLD input. This is illustrated in the following diagram.
Comparative output
HOLD input
Acceptance of
HOLD input
Measurement
starts
T
Delay in comparative
outputs
1
Measurement
completed
Recovery time
(20 ms min.)
T
2
Response time of HOLD input
(20 ms max.)
26
Page 36
(
)
Time Difference:
f3
Section 3-3
Available functions in this mode are indicated as “Yes” in the following table.
MenuFunctionDisplayed
Character
---Max./Min. value display and reset---Yes82
sUset (note 1)
pscl
setup
opt
Set value bank no. of set valuess.bank
HH set valuesU*.hh
HsetvaluesU*.h
LsetvaluesU*.l
LL set valuesU*.ll
Set value bank no. of prescale valuesp.bankYes
Prescaling value of input A
X (mantissa)
Y (exponent)
Decimal point positiondecp.*Ye s
Operating modefuncYes50
Input A sensor typeinaYe s
Input B sensor typeinbYe s
Time unittimeYes54
Communications unit no. (note 1)u- noYe s
Baud rate (note 1)bpsYe s
Word length (note 1)lenYe s
Stop bits (note 1)sbitYes
Parity bits (note 1)prtyYes
Output pattern (note 1)c-outYes61
H linear output range (note 1)lset.hYes
L Linear output range (note 1)lset.lYes
Remote/Local programming (note 1)r- lYes67
ps*.ax
ps*.ay
AvailabilityReference
Yes42
46
Yes
52
56
58
64
page
Note1. The availability of the parameters depends on the type of selected Output
Board.
2. The selected bank number will be displayed where the asterisk (
*) appears.
27
Page 37
Elapsed Time:
f4
3-4Elapsed Time: f4
Section 3-4
FUNCTION
Application example
Measures the pressing time of a printing machine.
Basic Operation
Displays in seconds the time period (T) that INA is ON. If another unit of measurement is required, input an appropriate prescale value. Obtain display value
D as follows:
D (s) = T x a
T: Time period (sec) INA is ON
a: prescale value
HOLD
Display
SETTING
INA
Value previously
measured
To set a prescale value, refer to 4-2-4 Prescaling Menu.
T
1
T
R
T
2
T
3
T
4
T4X aT2X aT1X a
: Recovery time
T
R
Refer to the following table to set the most suitable prescaling value.
ModeUnit of display
value
Elapsed timesec1
min1/60
Prescale value
Where,
with operating mode 4, INB input is disregarded.
Note The K3NP can display time, minute, and second. Refer to page 54 for details.
Refer to 4-2 Setting Mode.
REFERENCE
28
Page 38
Elapsed Time:
f4
Performance Characteristics
Section 3-4
Accuracy of measurement
Measurement range10 ms to 3200 s
Response time of HOLD input20 ms max.
Recovery time (TR)20 ms min.
ON/OFF pulse width
Response timeOutput configuration
Relay outputTransistor
Comparative output 25 ms max.20 ms max.35 ms max.35 ms max.20 ms max .
Sensor with transistor output: 9 msmin.
Sensor with relay output: 15 ms min.
output
BCD and
transistor output
Linear and
transistor output
Communication
and transistor
output
The response time of the HOLD input is the time required for the K3NP to
accept HOLD input after the HOLD input turns ON. This is illustrated in the
following diagram.
Recovery Time (T
)Recovery time is the period required for the K3NP to become ready for the
R
next measuring operation after a measuring operation is completed. For transistor output Models, comparative output remains ON during recovery time.
For other models, comparative output requires more time to turn ON. When
the comparative output terminals of the output board and the HOLD input
terminals are connected, set a sufficient recovery period, considering the response time of the HOLD input. This is illustrated in the following diagram.
Measurement startsMeasurement completed
INA
Comparative output
Delay in comparative outputs
HOLD input
Acceptance of HOLD input
T
1
Response time of HOLD input
(20 ms max.)
Recovery time (20 ms min.)
29
Page 39
(
)
Elapsed Time:
f4
Section 3-4
Available functions in this mode are indicated as “Yes” in the following table.
MenuFunctionDisplayed
Character
---Max./Min. value display and reset---Yes82
sUset (note 1)
pscl
setup
opt
Set value bank no. of set valuess.bank
HH set valuesU*.hh
HsetvaluesU*.h
LsetvaluesU*.l
LL set valuesU*.ll
Set value bank no. of prescale valuesp.bankYes
Prescaling value of input A
X (mantissa)
Y (exponent)
Decimal point positiondecp.*Ye s
Operating modefuncYes50
Input A sensor typeinaYe s
Input B sensor typeinbNo
Time unittimeYes54
Communications unit no. (note 1)u- noYe s
Baud rate (note 1)bpsYe s
Word length (note 1)lenYe s
Stop bits (note 1)sbitYes
Parity bits (note 1)prtyYes
Output pattern (note 1)c-outYes61
H linear output range (note 1)lset.hYes
L Linear output range (note 1)lset.lYes
Remote/Local programming (note 1)r- lYes67
ps*.ax
ps*.ay
AvailabilityReference
Yes42
46
Yes
52
56
58
64
page
Note1. The availability of the parameters depends on the type of selected Output
Board.
2. The selected bank number will be displayed where the asterisk (
*) appears.
30
Page 40
Length Measurement:
f5
3-5Length Measurement: f5
FUNCTION
Application example
Basic Operation
Counts the number of pulses of INA while INB is ON and displays the result. Obtain display value D as follows:
Section 3-5
Measures the length of an object.
INA
INB
HOLD
Display
SETTING
C
1
Value previously
measured
D (pulse count) = C x a
C: Pulse count of INA while INB input is ON
a: Prescale value
To set a prescale value, refer to 4-2-4 Prescaling Menu.
C
T
R
C1X aC2X aC4X a
2
C
3
C
4
TR: Recovery time
Refer to the following table to set the most suitable prescaling value.
ModeUnit of display
value
Length measurementmm
cm
m
Prescale value
1000 pDa/Na
100 pDa/Na
pDa/Na
REFERENCE
Na: Number of pulses per revolution at A input
pDa: Circumference (m) per revolution from A input
Note The K3NP can display time, minute, and second. Refer to page 54 for details.
Refer to 4-2 Setting Mode.
31
Page 41
Length Measurement:
f5
Section 3-5
Performance Characteristics
Maximum input frequency of INASensor with transistor output: 50 kHz
Sensor with relay output: 30 Hz
Measurement range (C)0 to 4 G (with 32-bit c ounter)
Response time of HOLD input20 ms max.
Recovery time (TR)20 ms min.
ON/OFF pulse width
Response timeOutput configuration
Relay outputTransistor
Comparative output 25 ms max.20 ms max.35 ms max.35 ms max.20 ms max .
BCD output------Refer to page 93.------
Linear output---------40 ms max.---
Sensor with transistor output: 9 msmin.
Sensor with relay output: 15 ms min.
BCD and
output
transistor output
Linear and
transistor output
Communication
and transistor
output
Maximum Input FrequencyThe maximum input frequency is the maximum number of INA input pulses
that the K3NP can count accurately.
Response Time of HOLD
Input
The response time of the HOLD input is the time required for the K3NP to
accept HOLD input after the HOLD input turns ON. This is illustrated in the
following diagram.
Recovery Time (T
)Recovery time is the period required for the K3NP to become ready for the
R
next measuring operation after a measuring operation is completed. For transistor output Models, comparative output remains ON during recovery time.
For other models, comparative output requires more time to turn ON. When
the comparative output terminals of the output board and the HOLD input
terminals are connected, set a sufficient recovery period, considering the response time of the HOLD input. This is illustrated in the following diagram.
Measurement starts
INA
Measurement completed
INB
Comparative output
Delay in comparative outputs
HOLD input
Acceptance of HOLD input
32
C
C
1
Response time of HOLD input
(20 ms max.)
Recovery time (20 ms min.)
2
TR
Page 42
(
)
Length Measurement:
f5
Section 3-5
Available functions in this mode are indicated as “Yes” in the following table.
MenuFunctionDisplayed
Character
---Max./Min. value display and reset---Yes82
sUset (note 1)
pscl
setup
opt
Set value bank no. of set valuess.bank
HH set valuesU*.hh
HsetvaluesU*.h
LsetvaluesU*.l
LL set valuesU*.ll
Set value bank no. of prescale valuesp.bankYes
Prescaling value of input A
X (mantissa)
Y (exponent)
Decimal point positiondecp.*Ye s
Operating modefuncYes50
Input A sensor typeinaYe s
Input B sensor typeinbYe s
Time unittimeNo54
Communications unit no. (note 1)u- noYe s
Baud rate (note 1)bpsYe s
Word length (note 1)lenYe s
Stop bits (note 1)sbitYes
Parity bits (note 1)prtyYes
Output pattern (note 1)c-outYes61
H linear output range (note 1)lset.hYes
L Linear output range (note 1)lset.lYes
Remote/Local programming (note 1)r- lYes67
ps*.ax
ps*.ay
AvailabilityReference
Yes42
46
Yes
52
56
58
64
page
Note1. The availability of the parameters depends on the type of selected Output
Board.
2. The selected bank number will be displayed where the asterisk (
*) appears.
33
Page 43
Interval:
f6
3-6Interval: f6
Section 3-6
FUNCTION
Application example
Measures the distance between the slits.
Basic Operation
Counts the number of pulses of INA between successive instances of INB turning ON and displays the result. Obtain display value D as follows:
D (pulse count) = C x a
C: Pulse count of INA while INB input is ON
a: Prescale value
To set a prescale value, refer to 4-2-4 Prescaling Menu.
INA
INB
HOLD
Display
SETTING
C
1
Value previously
measured
C
T
R
C1X aC2X aC4X a
2
C
3
C
4
TR: Recovery time
Refer to the following table to set the most suitable prescaling value.
ModeUnit of display
value
Intervalmm
cm
m
Prescale value
1000 pDa/Na
100 pDa/Na
pDa/Na
Na: Number of pulses per revolution at A input
pDa: Circumference (m) per revolution from A input
Note The K3NP can display time, minute, and second. Refer to page 54 for details.
Refer to 4-2 Setting Mode.
REFERENCE
34
Page 44
Interval:
f6
Section 3-6
Performance Characteristics
Maximum input frequency of INASensor with transistor output: 50 kHz
Sensor with relay output: 30 Hz
Measurement range (C)0 to 4 G (with 32-bit c ounter)
Response time of HOLD input20 ms max.
Recovery time (TR)20 ms min.
ON/OFF pulse width
Response timeOutput configuration
Relay outputTransistor
Comparative output 25 ms max.20 ms max.35 ms max.35 ms max.20 ms max .
BCD output------Refer to page 93.------
Linear output---------40 ms max.---
Sensor with transistor output: 9 msmin.
Sensor with relay output: 15 ms min.
BCD and
output
transistor output
Linear and
transistor output
Communication
and transistor
output
Maximum Input FrequencyThe maximum input frequency is the maximum number of INA input pulses
that the K3NP can count accurately.
Response Time of HOLD
Input
The response time of the HOLD input is the time required for the K3NP to
accept HOLD input after the HOLD input turns ON. This is illustrated in the
following diagram.
Recovery Time (T
)Recovery time is the period required for the K3NP to become ready for the
R
next measuring operation after a measuring operation is completed. For transistor output Models, comparative output remains ON during recovery time.
For other models, comparative output requires more time to turn ON. When
the comparative output terminals of the output board and the HOLD input
terminals are connected, set a sufficient recovery period, considering the response time of the HOLD input. This is illustrated in the following diagram.
Measurement starts
INA
Measurement completed
INB
Comparative output
Delay in comparative outputs
HOLD input
Acceptance of HOLD input
C
1
Recovery time (20 ms min.)
C
2
Response time of HOLD input
(20 ms max.)
TR
35
Page 45
(
)
Interval:
f6
Section 3-6
Available functions in this mode are indicated as “Yes” in the following table.
MenuFunctionDisplayed
Character
---Max./Min. value display and reset---Yes82
sUset (note 1)
pscl
setup
opt
Set value bank no. of set valuess.bank
HH set valuesU*.hh
HsetvaluesU*.h
LsetvaluesU*.l
LL set valuesU*.ll
Set value bank no. of prescale valuesp.bankYes
Prescaling value of input A
X (mantissa)
Y (exponent)
Decimal point positiondecp.*Ye s
Operating modefuncYes50
Input A sensor typeinaYe s
Input B sensor typeinbNo
Time unittimeYes54
Communications unit no. (note 1)u- noYe s
Baud rate (note 1)bpsYe s
Word length (note 1)lenYe s
Stop bits (note 1)sbitYes
Parity bits (note 1)prtyYes
Output pattern (note 1)c-outYes61
H linear output range (note 1)lset.hYes
L Linear output range (note 1)lset.lYes
Remote/Local programming (note 1)r- lYes67
ps*.ax
ps*.ay
AvailabilityReference
Yes42
46
Yes
52
56
58
64
page
Note1. The availability of the parameters depends on the type of selected Output
Board.
2. The selected bank number will be displayed where the asterisk (
*) appears.
36
Page 46
SECTION 4
Parameter Setting
This sectionprovides instructions for setting the parameters of theK3NP.Be sure to read this section before using the K3NP
Period Meter for the first time.
The following symbols are used for headings in this section.
This symbol precedes an explanation of the parameter’s meaning and function.
FUNCTION
This symbol precedes a description of the settings, setting range, and default
value.
SETTING
This symbol precedes an explanation of procedures for parameters that specify
operations.
PROCEDURE
This symbol precedes a listing of references and related parameters.
REFERENCE
This symbol precedes a listing of the models in which this parameter can be
used.
MODELS
4-1-2Setting Procedures
· The K3NP has four modes: RUN mode for normal operations, Setting mode for
initial parameter input, Protect mode for lock-out configuration, and Maintenance mode for initializing set values. The parameters that are accessible on
any individual K3NP will vary depending on the Output Board installed. Refer
to Appendix D Available Parameters.
· The K3NP is in RUN mode when the K3NP is turned on. Parameter settings in
protect or setting mode are described below on the basis that the parameters
are set for the first time.
For the operation in RUN mode, refer to Section 5 Operations in RUN Mode.
· The setting examples are provided on condition that the factory-set values of
the K3NP have not been changed.
38
Page 48
Setting Mode
4-2Setting Mode
4-2-1Selecting Setting Mode
· The K3NP in RUN mode will go into setting mode if the Mode Key is pressed for
1 s minimum.
· The K3NP in setting mode will go into RUN mode if the Escape Key is pressed.
Press the Mode and Shift Keys
while turning the K3NP on.
Section 4-2
Power On
Maintenance mode
1 second
RUN mode
RUNmode
Setting mode
Press the Escape and Up
Keys for 1 second.
Protect mode
· The menu in each mode changes whenever the Mode Key is pressed.
· If the Mode Key is pressed for more than one second while a menu is dis-
played, a parameter will be displayed.
· The parameter changes whenever the Mode Key is pressed.
· If the Shift Key is pressed while a parameter is displayed, the parameter will be
ready to change.
· Press the Up Key to change parameters.
· The digit of a set value is selected with the Shift Key and changed with the Up
Key.
· The PROG indicator is lit while a menu or parameter is displayed.
· The PROG indicator flashes during a set value change.
Note If the operating mode is changed, all the other parameters will be set to default
values except for communication settings and set values in protect mode.
Therefore, set the operating mode and sensor type first.
39
Page 49
Setting Mode
4-2-2Menu Overview
Power On
RUN Mode
Refer to Section 5 Operations in RUN Mode.
Press the Mode and Shift Keys
while turning the K3NP on.
Press the Escape and
Up Keys for 1 second.
Section 4-2
Maintenance Mode
Refer to Section 6 Useful Functions.
Protect Mode
Setting Mode
Output Test
Press the
Mode Key
for 1 second.
Setting value menu
Press the Mode Key for 1 second.
Bank no. of set
value
HH setting value
H setting value
L setting value
LL setting value
1to4
Refer to 4-3 Protect Mode.
Prescaling menu
Bank no. of prescale
Presc aling value of input A
X ( mantiss a))
Presc aling value of input A
Y ( exponent)
Decimal point
position
Press the Mode Key for 1 second.
Refer to Section 6Useful Functions.
Note1. When making new settings or changing settings of each parameter, press the
Shift Key to the setting state.
2. The K3NP stops measurement in setting mode.
Some menus cannot be set according to the Output Board selected.
If the operating mode is changed, all the other parameters will be set to default
values except communication setting and set values in protect mode. Therefore, set the operating mode and sensor type.
40
Page 50
Setting Mode
Section 4-2
Setup menu
Operating mode
Input A sensor
type
Input B sensor
type
Time unit
Communications
unit no.
Baud rate
Word length
Press the Mode Key for 1 second.
Option menu
Comparative output pattern
H linear output
range
L linear output
range
Remote/Local
programming
Press the Mode Key for 1 second.
Stop bits
Parity bits
41
Page 51
Setting Mode
4-2-3Setting Value Menu (sUset)
Section 4-2
s.bank
sU*.hh
sU*.h
sU*.l
sU*.ll
FUNCTION
SETTING
Bank No. of Set Value
HH Set V
HSetV
LSetValue
LL Set Value
alue
alue
· There are two basic methods for setting HH, H, L, and LL set values: by entering during RUN mode via the front-panel buttons, setting in Setting mode, or by
the teaching function.
SettingSetting rangeDefault
Set value bank1to41
HH set value0 to 9999999,999
Hsetvalue0 to 9999999,999
Lsetvalue0 to 999990
LL set value0 to 999990
MODELS
Set the decimal point position in the prescaling menu.
· The menu is only available for the K3NP, with Comparative Output Board.
42
Page 52
Setting Mode
Section 4-2
SETTING
EXAMPLE
Follow the steps described below to input the following.
Setting value bank = 2
HH setting value = “8000”
H setting value = “6000”
L setting value = “4000”
LL setting value = “2000”
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the sUset setting value
menu is displayed. The
Set Value LED Display ModelBasic Model
s.bank setting value bank setting will be displayed.
2. Press the Shift Key to display the set value 1 for changing. The PROG indicator will flash.
Set Value LED Display ModelBasic Model
3. Press the Up Key to set the value to 2. The input will be validated automatically if no change is made for five seconds. The
s.bank setting value bank
setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The
sU2.hh
HH setting value setting of bank 2 will be displayed for setting the next
parameter.
Set Value LED Display ModelBasic Model
4. Press the Mode Key to display the sU2.hh HH setting value of bank 2 setting.
Set Value LED Display ModelBasic Model
5. Press the Shift Key to display the set value 99999 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
43
Page 53
Setting Mode
Section 4-2
6. Press the Up and Shift Keys to set the value to 8000. The input will be validated automatically if no change is made for five seconds. The
setting value of bank 2 setting will be displayed again.
sU2.hh HH
Note Press the Mode Key to enter the set value immediately. The
sU2.h H
setting value of bank 2 setting will be displayed for setting the next
parameter.
Set Value LED Display ModelBasic Model
7. Press the Mode Key to display the sU2.h H setting value of bank 2 setting.
Set Value LED Display ModelBasic Model
8. Press the Shift Key to display the set value 99999 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
9. Press the Up and Shift Keys to set the value to 6000. The input will be validated automatically if no change is made for five seconds. The
sU2. h Hset-
ting value setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The
sU2. l L
setting value of bank 2 setting will be displayed for setting the next
parameter.
Set Value LED Display ModelBasic Model
10. Press the Mode Key to display the sU2.l L setting value of bank 2 setting.
Set Value LED Display ModelBasic Model
1 1. Press the Shift Key to display the set value 00000 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
44
Page 54
Setting Mode
Section 4-2
12. Press the Up and Shift Keys to set the value to 4000. The input will be validated automatically if no change is made for five seconds. The
ting value of bank 2 setting will be displayed again.
sU2. l Lset-
Note Press the Mode Key to enter the input immediately. The
sU2.ll LL
setting value setting will be displayed for setting the next parameter.
Set Value LED Display ModelBasic Model
13. Press the Mode Key to display the sU2.ll LL setting of bank 2 value setting.
Set Value LED Display ModelBasic Model
14. Press the Mode Key to display the set value 00000 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
15. Press the Up and Shift Keys to set the value to 2000. The input will be validated automatically if no change is made for five seconds. The
sU2.ll LL
setting value setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The
s.bank
setting bank setting will be displayed for setting the next parameter.
Set Value LED Display ModelBasic Model
16. Press the Escape Key to display the sUset setting value menu.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
45
Page 55
Setting Mode
4-2-4Prescaling Menu (pscl)
Section 4-2
p.bank
ps*.ax
ps*.ay
decp.*
FUNCTION
SETTING
Bank No. of Prescale
Prescaling Value X (Mantissa) of Input
Prescaling Value Y (Exponent) of Input
Decimal Point Position
The measured data of passing speed or cycle is displayed after being multiplied
by a preset value (i.e., prescaling value).
Display value = Measured data x Prescaling value
Input typeSettingDefault
p.bank: Bank no. of prescaleoff/1 to 4off
ps*.ax: Prescaling value X (mantissa)
of input A
ps*.ay: Prescaling value Y (exponent)
of input A
decp.*: Decimal point positionOne of the 1st to 4th digits
0.0001 to 9.99991.0000
--9to90
from the right
A
A
No decimal
point
position
setting
REFERENCE
SETTING
EXAMPLE
Refer to Section 3 Operating Modes.
Follow the steps described below to input the following.
Operating mode = F1
Prescaling bank = OFF
Prescaling value X (mantissa) of input A = 0.5000
Prescaling value Y (exponent) of input A = --1
Decimal point = jjjj.j (First digit from the right)
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the pscl prescaling
menu is displayed. The
Set Value LED Display ModelBasic Model
p.bank prescaling bank setting will be displayed.
46
Page 56
Setting Mode
Section 4-2
2. Press the Shift Key to display off for changing. The PROG indicator will
flash.
Set Value LED Display ModelBasic Model
3. Press the Up Key to change the prescaling bank setting. The input will be
validated automatically if no change is made for five seconds. The
prescaling bank setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The
prescaling value X (mantissa) of input A setting will be displayed for
setting the next parameter.
Set Value LED Display ModelBasic Model
4. Press the Mode Key to display the ps.ax prescaling value X (mantissa) of
input A setting.
Set Value LED Display ModelBasic Model
p.bank
ps.ax
5. Press the Shift Key to display the set value 1.0000 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
6. Press the Up and Shift Keys to set the value to 0.5000. The input will be validated automatically if no change is made for five seconds. The
ps.ax pres-
caling value X (mantissa) of input A setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The
ps.ax
prescaling value X (mantissa) of input A setting will be displayed for
setting the next parameter.
Set Value LED Display ModelBasic Model
7. Press the Mode Key to display the ps.ay prescaling value Y of input A setting.
Set Value LED Display ModelBasic Model
47
Page 57
Setting Mode
Section 4-2
8. Press the Shift Key to display the set value 10 00 for changing.
Set Value LED Display ModelBasic Model
9. Press the Up and Shift Keys to set the value to 10 -1. The input will be validated automatically if no change is made for five seconds. The
caling value Y (exponent) of input A setting will be displayed again.
ps.ay pres-
Note Press the Mode Key to enter the set value immediately. The
decimal point position setting will be displayed for setting the next parameter.
Set Value LED Display ModelBasic Model
10. Press the Mode Key to display the decp decimal point position setting.
Set Value LED Display ModelBasic Model
1 1. Press the Shift Key to display %%%%% for changing.
Set Value LED Display ModelBasic Model
decp
48
12. Press the Shift Key to set %%%%.%. The input will be validated automatically if
no change is made for five seconds. The
decp decimal point position setting
will be displayed again.
Note Press the Mode Key to enter the set value immediately. The
p.bank
prescaling bank setting will be displayed for setting the next parameter.
Set Value LED Display ModelBasic Model
When no operation is executed for five seconds
Set Value LED Display ModelBasic Model
Page 58
Setting Mode
Section 4-2
13. Press the Escape Key to display the pscl prescaling menu.
Set Value LED Display ModelBasic Model
49
Page 59
Setting Mode
4-2-5Setup Menu (setup)
Section 4-2
func
FUNCTION
SETTING
Operating Mode
All parameters will be set to default values if any change in made in this parameter, except to those for the communications and protect settings.
f1: Displays the reciprocal of the time (T) between the moment input A is
turned ON and the moment input B is turned ON after being multiplied by
60.
f2: Measures and displays the cycle (T) of input A as is.
f3: Displays the time between the moment input A is turned ON and the mo-
ment input B is turned ON as is.
f4: Displays the elapsed time (T) of input A.
f5: Displays the number of pulses of input A counted while input B is ON.
f6: Displays the number of pulses of input A counted between the moment in-
put B is turned ON and the moment input B is turned ON again.
f1: Passing speed
f2: Cycle
f3: Time difference
f4: Elapsed time
f5: Length measurement
f6: Interval
SettingDefault
f1
REFERENCE
SETTING
EXAMPLE
Refer to Section 3 Operating Modes.
Follow the steps described below to select the f4 elapsed time setting.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the setup setup menu
is displayed. The
2. Press the Shift Key to display f1 for changing. The PROG indicator will flash.
func operating mode setting will appear.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
50
Page 60
Setting Mode
Section 4-2
3. Repeatedly press the Up Key until f4 is displayed. The displayed setting will
be validated automatically if no change is made for five seconds. The
operating mode setting will be displayed again.
Note Press the Mode Key to enter the displayed setting immediately. The
next parameter will be displayed for setting.
Set Value LED Display ModelBasic Model
4. Repeatedly press the Escape Key until the setup setup menu is displayed.
Set Value LED Display ModelBasic Model
func
51
Page 61
Setting Mode
Section 4-2
ina
inb
FUNCTION
SETTING
Sensor T
· Specifies the type of sensors for input A and input B.
· The sensor type of input B cannot be selected if the operating mode of the
K3NP is set to
· Open Collector Input
Transistor input0001
Relay input1011
Normally Open Model: The sensor output is OFF (open) when the sensor is
Normally Closed Model: The sensor output is ON (closed) when the sensor is
· Voltage Pulse Input
Transistor input000100
ype
f2 (cycle) or f4 (elapsed time).
Sensor typeNormally openNormally closedDefault
00
not sensing an object.
not sensing an object.
Sensor typeActive high (H)Active low (L)Default
SETTING
EXAMPLE
Follow the steps described below to set input A to 11 in operating mode f1 (passing
speed).
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the setup setup menu
is displayed. The
2. Press the Mode Key to display ina input A sensor type setting.
3. Press the Shift Key to display 00 for changing. The PROG indicator will flash.
func operating mode setting will appear.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
52
Page 62
Setting Mode
Section 4-2
4. Press the Up and Shift Keys to display 11. The displayed setting will be validated automatically if no change is made for five seconds. The
sensor type setting will be displayed again.
Note Press the Mode Key to enter the displayed setting immediately. The
next parameter will be displayed.
Set Value LED Display ModelBasic Model
5. Press the Escape Key to display the setup setup menu.
Set Value LED Display ModelBasic Model
ina input A
53
Page 63
Setting Mode
Section 4-2
time
FUNCTION
SETTING
Time
Unit
The time unit can be selected to display the calculation results of F2 (cycle), F3
(time difference), and F4 (elapsed time).
UnitDisplay rangeDefault
scal0 to 99,999
secDisplayed in seconds within a range
between 0 and 99,999 s.
minDisplayed minutes within a range
between 0 and 99,999 min.
mm.ss.dDisplayed in minutes and seconds
(1/100 s) within a range between 00
min, 00 s 0 and 59 min, 59 s, 9.
scal
SETTING
EXAMPLE
Follow the steps described below to set sec second unit setting.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the setup setup menu
is displayed. The
func operating mode setting will appear.
Set Value LED Display ModelBasic Model
2. Repeatedly press the Escape Key until time time unit setting is displayed.
Set Value LED Display ModelBasic Model
54
3. Press the Shift Key to display scal for changing. The PROG indicator will
flash.
Set Value LED Display ModelBasic Model
Page 64
Setting Mode
Section 4-2
4. Press the Up Key to display sec. The input will be validated automatically if
no change is made for five seconds. The
played again.
Set Value LED Display ModelBasic Model
When no operation is executed for five seconds
Set Value LED Display ModelBasic Model
5. Press the Escape Key to display the setup setup menu.
Set Value LED Display ModelBasic Model
time time unit setting will be dis-
55
Page 65
Setting Mode
Section 4-2
u-no
bps
FUNCTION
SETTING
Communications Unit
Baud Rate
· Set a communications unit number as an identification number by which the
host computer is connected to the K3NP.
· If more than one K3NP is connected in parallel, make sure that each communications unit number is unique.
· The baud rate should be set to the baud rate of the host computer.
This setting is available for the K3NP with the Communications Output Board.
Follow the steps described below to set the communications unit number to 15 and the
baud rate to 19,200 bps.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the setup setup menu
is displayed. The
2. Repeatedly press the Mode Key until the u- no communications unit number
setting is displayed.
func operating mode setting will appear.
Set Value LED Display ModelBasic Model
56
Set Value LED Display ModelBasic Model
Page 66
Setting Mode
Section 4-2
3. Press the Shift Key to display the prior set value 00 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
4. Press the Up and Shift Keys to set the value to 15. The input value will be
validated automatically if no change is made for five seconds. The
communications unit number setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
5. Press the Mode Key to display the bps baud rate setting.
u- no
Set Value LED Display ModelBasic Model
6. Press the Shift Key to display the prior set value 9600 for changing. The
PROG indicator will flash.
Set Value LED Display ModelBasic Model
7. Press the Up Key to set the value to 19200. The input will be validated automatically if no change is made for five seconds. The
bps baud rate setting
will be displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed again for setting.
Set Value LED Display ModelBasic Model
8. Press the Up Key to enter the set value for setting the next parameter. The
input value will be validated automatically if no change is made for five seconds. The
bps baud rate setting will be displayed again.
Set Value LED Display ModelBasic Model
57
Page 67
Setting Mode
Section 4-2
len
sbit
prty
FUNCTION
SETTING
Word
Length
Stop Bits
Parity Bits
· The communications format used for communicating with the host computer is
set in the setup menu.
· Refer to the Communications Manual for the communications format in detail.
· Word Length
7/8bit7
· Stop Bits
1/2bit2
· Parity Bit
none: None
eUen:Even
odd:Odd
SettingUnitDefault
SettingUnitDefault
SettingDefault
eUen
MODELS
SETTING
EXAMPLE
This setting is available for the K3NP with the Communications Output Board.
Follow the steps described below to set the following.
Word length: 8 bits
Number of stop bits: 1
Parity bits:
1, 2, 3...1. Press the Mode Key for more than one second while the setup setup menu
is displayed. The
2. Repeatedly press the Mode Key until the len word length setting is displayed.
none
Set Value LED Display ModelBasic Model
func operating mode setting will appear.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
58
Page 68
Setting Mode
Section 4-2
3. Press the Shift Key to display the prior set value 7 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
4. Press the Up Key to set the value to 8. The input value will be validated automatically if no change is made for five seconds. The
will be displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
5. Press the Mode Key to display the sbit stop bit setting.
Set Value LED Display ModelBasic Model
len word length setting
6. Press the Shift Key to display the set value 2 for changing.
Set Value LED Display ModelBasic Model
7. Press the Up Key to set the value to 1. The input will be validated automatically if no change is made for five seconds. The
sbit stop bit setting will be
displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
8. Press the Mode Key to display the prty parity bit setting.
Set Value LED Display ModelBasic Model
9. Press the Shift Key to display eUen for changing.
Set Value LED Display ModelBasic Model
59
Page 69
Setting Mode
Section 4-2
10. Press the Up Key to display none. The setting will be validated automatically
if no change is made for five seconds. The
played again.
Note Press the Mode Key to enter the setting immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation is executed for five seconds
Set Value LED Display ModelBasic Model
prty parity bit setting will be dis-
60
Page 70
Setting Mode
4-2-6Option Menu (opt)
Section 4-2
c-out
FUNCTION
SETTING
REFERENCE
Comparative Output Pattern
· The pattern of HH, H, L, LL, and PASS comparative outputs is set in the option
menu.
SettingDefault
nomal: Standard output
=one: Zone output
leUel: Level output
Standard Output
H or HH Comparative Output:
Turns ON when the measured value is larger than the H or HH setting value.
nomal
PASS Output:
Turns ON when LL, L, H, and HH comparative outputs are all OFF.
L or LL Comparative Output:
Turns ON when the measured value is smaller than the L or LL setting value.
HH setting value
H setting value
L setting value
LL setting value
HH comparative output
H comparative output
PASS
L comparative output
LL comparative output
Measured value
61
Page 71
Setting Mode
Section 4-2
Level Output
LL, L, H, or HH Comparative Output:
Turns ON when the measured value exceeds the LL, L, H, or HH setting value.
PASS Output:
Turns ON when the LL, L, H, and HH comparative outputs are all OFF .
HH setting value
H setting value
L setting value
LL setting value
HH comparative output
H comparative output
PASS
L comparative output
LL comparative output
Measured value
Zone Output
HH Comparative Output:
Turns ON when the measured value exceeds the HH setting value.
H Comparative Output:
Turns ON when the measured value is between the H and HH setting values.
PASS Output:
Turns ON when the measured value is between the L and H setting values.
L Comparative Output:
Turns ON when the measured value is between the LL and L setting values.
LL Comparative Output:
Turns ON when the measured value falls below the LL setting value.
Be sure to set the setting values so they satisfy the following formula:
LL<L<H<HH
62
HH setting value
H setting value
L setting value
LL setting value
HH comparative output
H comparative output
PASS
L comparative output
LL comparative output
Measured value
Page 72
Setting Mode
MODELS
Section 4-2
This setting is only available for the K3NP with the Comparative Output Unit.-
SETTING
EXAMPLE
Follow the steps described below to set the comparative output pattern to level output.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the opt option menu is
displayed. The
aUg process time for averaging measured value setting will
appear.
Set Value LED Display ModelBasic Model
2. Repeatedly press the Mode Key until the c-out comparative output pattern
setting is displayed.
Set Value LED Display ModelBasic Model
3. Press the Shift Key to display the prior setting nomal for changing. The
PROG indicator will flash.
Set Value LED Display ModelBasic Model
4. Press the Up Key twice to display leUel. The setting will be validated automatically if no change is made for five seconds. The
c-out comparative out-
put pattern setting will be displayed again.
Note Press the Mode Key to enter the setting immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation is executed for five seconds
Set Value LED Display ModelBasic Model
63
Page 73
Setting Mode
Section 4-2
lset.h
lset.l
FUNCTION
Upper Limit (H) of Lin ear Output
Range
Lower Limit (L) of Linear Output Range
Linear output setting is made in the option menu to enable the K3NP to have
voltage or current output in proportion to the change in display value.
· The maximum and minimum values of linear output are set in this parameter.
Linear output
20 mA or 5 V
4mAor1V
· L can be greater or less than H.
· L cannot be the same as H, otherwise H will be automatically set to a value
obtained by adding 1 to L.
· The teaching function can be used for setting linear output ranges.
Display value
SETTING
REFERENCE
MODELS
Setting rangeDefault
0 to 99999
Refer to 6-1 Teaching Function.
This setting is available for the K3NP with the Linear Output Board.
H linear output range99999
L linear output range0
64
Page 74
Setting Mode
Section 4-2
SETTING
EXAMPLE
Follow the steps described below to set the following.
H: 100.00
L: 1.00
(Assume that the decimal point is set between the 2nd and 3rd digit from the right in the
prescale menu.)
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the opt option menu is
displayed. The
aUg process time for averaging measured value setting will
appear.
Set Value LED Display ModelBasic Model
2. Repeatedly press the Mode Key until the lset.h H linear output range setting is displayed.
Set Value LED Display ModelBasic Model
3. Press the Shift Key to display the prior set value 999.99 for changing. The
PROG indicator will flash.
Set Value LED Display ModelBasic Model
4. Press the Up and Shift Keys to set the value to 100.00. The setting will be
validated automatically if no change is made for five seconds. The
lset.h H
linear output range setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
+
5. Press the Mode Key to display the lset.l L linear output range setting.
Set Value LED Display ModelBasic Model
65
Page 75
Setting Mode
Section 4-2
6. Press the Shift Key to display the prior set value 000.00 for changing. The
PROG indicator will flash.
Set Value LED Display ModelBasic Model
7. Press the Up and Shift Keys to set the value to 001.00. The setting will be
validated automatically if no change is made for five seconds. The
linear output range setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation is executed for five seconds
lset.l L
Set Value LED Display ModelBasic Model
66
Page 76
Setting Mode
Section 4-2
r-l
FUNCTION
SETTING
MODELS
SETTING
EXAMPLE
Remote/Local
· The K3NP can be set to remote or local mode in the option menu. The K3NP in
remote mode is operated through the host computer and the K3NP in local
mode is operated through the front panel key input.
Remote: rmt
Local: lcl
This setting is available for the K3NP with the Communications Output Board.
Follow the steps described below to set the K3NP to remote programming.
Programming
SettingDefault
lcl
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the opt option menu is
displayed. The
2. Repeatedly press the Mode Key until the r- l remote/local setting is displayed.
3. Press the Shift Key to display the prior setting lcl for changing. The PROG
indicator will flash.
4. Press the Up Key to display rmt.
aUg setting will appear.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
67
Page 77
Setting Mode
Section 4-2
5. The setting will be validated automatically if no change is made for five seconds. The
c-out comparative output pattern again.
Note Press the Mode Key to enter the setting immediately. The
mote/local setting will be displayed again.
Set Value LED Display ModelBasic Model
r- l re-
68
Page 78
Protect Mode
4-3Protect Mode
4-3-1Selecting Protect Mode
· The K3NP in RUN mode will go into protect mode if the Escape and Up Keys
are pressed for more than 1 second.
· The K3NP in protect mode will go into RUN mode if the Escape Key is pressed.
Press the Mode and Shift Keys
while turning the K3NP on.
Section 4-3
Power On
Maintenance mode
RUN mode
1 second
Setting mode
Press the Escape and Up
Keys for 1 second.
Protect mode
· If the Mode Key is pressed for more than one second while a menu is displayed, a parameter will be displayed.
· The parameter changes whenever the Mode Key is pressed.
· If the Shift Key is pressed while a parameter is displayed, the parameter will be
ready to change.
· Press the Up Key to change parameters.
69
Page 79
Protect Mode
4-3-2Menu Overview
Section 4-3
Press the Mode and Shift Keys
Power On
while turning the K3NP on.
RUN Mode
Refer to Section 5 Operations in RUN Mode.
Press the Escape and
Up Keys for 1 second.
Protect Mode
Protect menu
All key protect
No key input
for 2 seconds.
Setting value
change prohibit
No key input
for 2 seconds.
Press the Mode Key
for 1 second.
Key protect ON
Key protect OFF
Key protect ON
Key protect OFF
Maintenance Mode
Refer to Section 6 Useful Functions.
Setting Mode
Refer to 4-2 Setting Mode.
Reset prohibit
Security
Key protect ON
No key input
for 2 seconds.
Key protect OFF
All menus prohibited: 2
Setup menu prohibited: 1
No key input
for 2 seconds.
All menu allowed: 0
Note1. The K3NP stops measurement in setting mode.
2. Some menus cannot be set due to the display type or output type selected.
70
Page 80
Protect Mode
4-3-3Protect Menu (prot)
Section 4-3
all
FUNCTION
SETTING
SETTING
EXAMPLE
All Key Protect
· The operation of all keys can be prohibited in the protect menu.
SettingDefault
kpon: Key protect ON
kpoff: Key protect OFF
Follow the steps described below to set the key protect to ON.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the prot protect menu
is displayed. The
all all key protect setting will appear.
Set Value LED Display ModelBasic Model
kpoff
2. Press the Shift Key to display the prior setting kpoff for changing. The
PROG indicator will flash.
Set Value LED Display ModelBasic Model
3. Press the Up Key to display kpon. The setting will be validated automatically
if no change is made for five seconds. The
all all key protect setting will be
displayed again.
Note Press the Mode Key to enter the setting immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation
is executed for five
seconds
71
Page 81
Protect Mode
Section 4-3
sUset
FUNCTION
SETTING
REFERENCE
MODELS
Setting Value Change
· The setting value change of the K3NP in RUN mode with the front-panel input
can be prohibited in the protect menu.
SettingDefault
kpon: Key protect ON
kpoff: Key protect OFF
Refer to 5-1 Checking and Changing Setting Values.
This setting is only available for the K3NP with the Comparative Output Unit.
Prohibit
kpoff
SETTING
EXAMPLE
Follow the steps described below to set the setting value change prohibit to ON.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the prot protect menu
is displayed. The
2. Press the Mode Key to display the sUset setting value prohibit setting.
3. Press the Shift Key to display the prior setting kpoff for changing. The
PROG indicator will flash.
all all key protect setting will appear.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
72
Set Value LED Display ModelBasic Model
Page 82
Protect Mode
Section 4-3
4. Press the Up Key to display kpon. The setting will be validated automatically
if no change is made for five seconds. The
ting will be displayed again.
Note Press the Mode Key to enter the setting immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation
is executed for five
seconds
sUset setting value prohibit set-
73
Page 83
Protect Mode
Section 4-3
reset
FUNCTION
SETTING
MODELS
Reset
Prohibit
· Prohibits the resetting of the K3NP to the initial state and the resetting of max./
min. value with key input.
· This function does not prohibit the resetting of the K3NP with external signal
input.
SettingDefault
kpon: Key protect ON
kpoff: Key protect OFF
Refer to 5-3 External Input Signals.
kpoff
SETTING
EXAMPLE
Follow the steps described below to set key protect ON.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the prot protect menu
is displayed. The
2. Press the Mode Key to display reset Reset prohibit setting.
3. Press the Shift Key to display kpoff for changing. The PROG indicator will
flash.
all all key protect setting will appear.
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
74
Set Value LED Display ModelBasic Model
Page 84
Protect Mode
Section 4-3
4. Press the Up Key to display kpon. The displayed setting will be validated automatically if no change is made for five seconds. The
setting will be displayed again.
Note Press the Mode Key to enter the setting immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation is executed for five seconds
Set Value LED Display ModelBasic Model
reset reset prohibit
75
Page 85
Protect Mode
Section 4-3
secr
FUNCTION
SETTING
Security
· Settings in setting mode can be prohibited in the protect menu.
· The following table shows what set values for menus can be prohibited. The
Setting valueProhibited
ScalingProhibited
SetupProhibitedProhibited
OptionProhibited
· The value changes in the following order with the Up Key: 0, 1, 2, and 0
default is 0.
Menu
Set value
012
SETTING
EXAMPLE
Follow the steps described below to set the security setting to 1.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the Mode Key for more than one second while the prot protect menu
is displayed. The
all all key protect setting will appear.
Set Value LED Display ModelBasic Model
2. Repeatedly press the Mode Key until the secr security setting is displayed.
Set Value LED Display ModelBasic Model
76
3. Press the Shift Key to display the prior set value 0 for changing. The PROG
indicator will flash.
Set Value LED Display ModelBasic Model
Page 86
Protect Mode
Section 4-3
4. Press the Up Key to display 1. The setting will be validated automatically if
no change is made for five seconds. The
played again.
Note Press the Mode Key to enter the setting immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display ModelBasic Model
When no operation
is executed for five
seconds
secr security setting will be dis-
77
Page 87
Operations in RUN Mode
This section provides instructions for operating the K3NP in RUN mode.
5-1Displaying and Changing Setting Values80......................................
· When the Mode Key is pressed in RUN mode, the K3NP displays a setting value on the PV display (in the order of HH, H, L, and LL).
· While the setting value is displayed, the corresponding SV display status indicator is lit.
· Unless another operation key is pressed within five seconds after the setting
value is displayed, the process, maximum, or minimum value is displayed
again.
Section 5-1
HH is lit.
Process value
Maximum value
Minimum value
HH setting
value
No key input
for 5 seconds.
Set Value LED Display Model
· The setting value appears on the SV display and the corresponding SV display
status indicator is lit while the process, maximum, or minimum value is displayed on the PV display
· When the Mode Key is pressed, the K3NP displays a setting value (in the order
of HH, H, L, and LL) on the SV display.
5-1-2Changing Setting Values
· Setting values can be prohibited against change when key protect or setting
value change prohibit is ON in protect mode.
· Select the setting value to be changed with the Mode Key.
· All digits will be displayed and ready for changing if the Shift Key is pressed.
· Press the Up and Shift Keys to change the displayed setting value. The input
will be entered if nothing else is input within five seconds. The input is entered
immediately by pressing the Mode Key.
· The next setting value will be displayed and ready for changing if the Mode Key
is pressed. If nothing else is input within five seconds, the setting value will be
entered and the previous setting value will be displayed.
· The setting value can be changed through communications if the K3NP is a
model with a communications function.
H is lit.L is lit.LL is lit.
H setting
value
No key input
for 5 seconds.
L setting
value
No key input
for 5 seconds.
LL setting
value
No key input
for 5 seconds.
80
Page 89
Displaying and Changing Setting Values
SETTING
EXAMPLE
Basic ModelFollow the steps below to change the H setting value from 600 to 700 while the
process value is displayed in RUN mode.
1, 2, 3...1. Press the Mode Key to display the HH setting value and to light the HH indi-
cator of the SV display status indicators.
2. Press the Mode Key again to display the H setting value H and light the H
indicator.
Section 5-1
3. Press the Shift Key to display all the digits of the prior setting value 600 for
changing.
4. Press the Up and Shift Keys to set the value to 00700.
5. The input will be entered if nothing else is input within five seconds.
or Press the Mode Key to display the L setting value for changing. The H set-
ting value will appear again if nothing else is input within five seconds.
6. To return to the process value display, perform one of the following.
· Repeatedly press the Mode Key until the process value appears.
· Leave the K3NP with no key input for five seconds.
81
Page 90
Displaying and Resetting of Maximum and Minimum Values
SETTING
EXAMPLE
Section 5-2
Set Value LED Display
Model
1, 2, 3...1. Press the Mode Key to display the H setting value and to light the H indicator
Follow the steps below to change the H setting value from 600 to 700 while the
process value is displayed in RUN mode, provided that the HH setting value is
already displayed on the SV display.
of the SV display status indicators.
2. Press the Shift Key to display all the digits of the setting value 600 for changing.
3. Press the Up and Shift Keys to set the value to 00700.
4. The input will be entered if nothing else is input within five seconds.
or Press the Mode Key to display the L setting value for changing. The H set-
ting value will appear again if nothing else is input within five seconds.
5-2Displaying and Resetting of Maximum and Minimum
Values
The maximum and minimum values are refreshed automatically while the K3NP
is in measurement operation.
5-2-1Displaying Maximum and Minimum Values
· The Escape Key is used to select the process, maximum, or minimum value to
be displayed on the PV display in RUN mode.
82
Page 91
Displaying and Resetting of Maximum and Minimum Values
· The corresponding SV display indicator (i.e., the MAX or MIN indicator) is lit
while the maximum or minimum value is displayed.
MAX is lit.MIN is lit.
Process valueMaximum valueMinimum value
5-2-2Resetting Maximum and Minimum Values
· The maximum and minimum values are reset when the K3NP is turned on, set
to RUN mode, or reset.
· The K3NP will be reset when one of the following is performed.
The RESET/TEACH Key is pressed for more than one second while the
maximum or minimum value is displayed.
The external RESET input signal is turned ON.
The K3NP receives the reset command through communications.
· When the K3NP is reset, the maximum and minimum values are set to the process value.
Section 5-2
Note The K3NP cannot be reset with the RESET/TEACH Key if the maximum/mini-
mum value clear prohibit is ON in protect mode.
83
Page 92
p
g
External Input Signals
SETTING
EXAMPLE
1, 2, 3...1. The process value 401 is displayed.
Section 5-3
Set Value LED Display ModelBasic Model
2. Press the Escape Key to display the maximum value 1024 and light the MAX
indicator.
Set Value LED Display ModelBasic Model
3. Press the RESET/TEACH Key to set the maximum and minimum values to
the process value 401.
4. Press the Escape Key to check that the minimum value is set to 401.
5-3External Input Signals
5-3-1Bank Selection
· The K3NP uses BANK1 and BANK2 signals to select the setting value, prescaling value, and decimal point position of the bank 1, 2, 3, or 4.
Refer to the following for the relationship between bank input signals and bank
numbers.
Bank no.
1OFFOFFsU1.**ps1.**
2ONOFFsU2.**ps2.**
3OFFONsU3.**ps3.**
4ONONsU4.**ps4.**
Set Value LED Display ModelBasic Model
Set Value LED Display ModelBasic Model
Control input
BANK 1BANK 2
Comparative set valuePrescaling value
Note If the prescaling bank is set to OFF, the pscl prescaling value for each bank will
be fixed.
· When a bank is selected, the corresponding BANK indicator 1, 2, 3, or 4 will be
lit.
84
Page 93
External Input Signals
· It takes a maximum of 100 ms for a bank switching after BANK1 and BANK2
signals are input.
The bank number can be displayed on the PV or SV display by pressing the Shift
Key for more than one second while the K3NP is in measurement operation. If
there is no key input for three seconds, the K3NP will be in measurement operation again.
5-3-2 Max./Min. Value RESET
· Refer to 5-2 Displaying and Resetting of Maximum and Minimum Values (Op-
erating Modes F1 to F6 for details.
· The following shows the operation timing of maximum and minimum values.
Section 5-3
BANK1
BANK2
Bank 1Bank 2
100 ms max.
5-3-3 RESET
RESET/TEACH Key input
RESET input
Display
Previous measured
value
First time (max. value)
Second time
(min. value)
Max.
value
Third time (PV)
Min. value
= Prescaling value of input A
Note The K3NP holds the maximum and minimum values that have been recorded
since the K3NP was turned ON or while the RESET input is ON.
Whenever the RESET/TEACH Key is pressed, the maximum, minimum, or
present value will appear on the PV display in this order.
If RESET input is ON or the RESET/TEACH Key is pressed, the maximum and
minimum values will be cleared and the K3NP will start measurement.
· The K3NP will be reset to the initial state when the power is turned ON or when
the RESET input is turned ON.
· All the displayed values will be reset to “00000”.
· The comparative outputs (HH, H, P, L, and LL) will be turned OFF.
· The Maximum and Minimum values will be reset and the reset function of the
RESET/TEACH Key will become invalid.
· The reset state will continue while the RESET is ON and no measurement will
be started.
85
Page 94
External Input Signals
5-3-4HOLD
Section 5-3
It is possible to reset the process value by pressing the RESET/TEACH Key.
· The K3NP will stop the measurement if the HOLD input is ON.
· When the HOLD input is ON, the K3NP will retain the process value, output,
and BCD data effective immediately before the HOLD input.
· The HOLD indicator is lit while HOLD input is ON.
· If the HOLD input is turned ON while the measurement is being performed, the
measurement will be cancelled and the previous measurement value (displayed value) will be retained.
· No measurement will be started while the HOLD input is ON.
Process value
86
Page 95
SECTION 6
Useful Functions
This section provides information on the output test and maintenance mode functions of the K3NP.
· The K3NP is provided with a teaching function that can set an actual measured
value as a set value without any front panel key input.
This function is useful for setting parameters while checking the operating status of the K3NP.
· The teaching function can be used to set the setting, scaling values, and linear
output range. The TEACH indicator will be lit if a parameter that can use the
teaching function is displayed.
6-1-1Set Value
· A setting value can be set with the actual input instead of key input in the setting
mode.
· Follow the steps below to teach the setting value.
1, 2, 3...1. Press the RESET/TEACH Key for more than one secondwhile the parame-
ter is displayed. The process value will be displayed and the teaching indicator will flash.
2. Press the RESET/TEACH Key again to retrieve the process value immediately before the key was pressed as a set value. The teaching indicator will
be lit and the parameter will be displayed. Press the Escape Key to interrupt
teaching.
SETTING
EXAMPLE
Follow the steps described below to set the HH set value of bank 1 by using the teaching
function.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the RESET/TEACH Key for more than one secondwhile the parame-
ter is displayed. The process value will be displayed and the teaching indicator will flash.
Set Value LED Display ModelBasic Model
2. Press the RESET/TEACH Key again to retrieve the process value immediately before the key was pressed as a set value for changing. The teaching
indicator will be lit and the parameter is displayed.
Set Value LED Display ModelBasic Model
88
Page 97
Teaching FunctionSection 6-1
6-1-2Linear Output Range
The teaching function can be also used to set the linear output range of the
K3NP with the Linear Output Board.
· The H and L linear output ranges can be set using the actual measured value
instead of key input in the option menu.
· Follow the steps below to teach the setting value.
1, 2, 3...1. Press the RESET/TEACH Key for more than one secondwhile the parame-
ter is displayed. The process value will be displayed and the teaching indicator will flash.
2. Press the RESET/TEACH Key again to retrieve the process value immediately before the key was pressed as a set value. The teaching indicator will
be lit and the parameter will be displayed. Press the Escape Key to interrupt
teaching.
SETTING
EXAMPLE
Follow the steps described below to set the H linear output range by using the teaching
function.
Set Value LED Display ModelBasic Model
1, 2, 3...1. Press the RESET/TEACH Key for more than one secondwhile the parame-
ter is displayed. The process value will be displayed and the teaching indicator will flash.
Set Value LED Display ModelBasic Model
2. Press the RESET/TEACH Key again to retrieve the process value immediately before the key was pressed as a set value for changing. The teaching
indicator will be lit and the parameter will be displayed. Press the Escape
Key to interrupt teaching.
Set Value LED Display ModelBasic Model
89
Page 98
Output Test
6-2Output Test
Note The K3NP has output according to the simulated input in this menu. If there is
1, 2, 3...1. While the K3NP is in RUN mode, press the Escape Key for more than one
Section 6-2
This function is convenient for checking a system to which the K3NP is connected, especially when some inputs cannot be operated. The K3NP simulates
an input signal to check the output conditions.
any device connected to the output of the K3NP, be sure that the output will not
have a negative influence on the device before testing the system.
· Follow the steps described below to perform the test.
second to set the K3NP to the setting mode.
2. Repeatedly press the Mode Key until the
played.
RUN mode
test output test setting is dis-
Setting Mode
1 second
Setting values menu
sUset
Output TestOption menuSetup menu
testopt
Prescaling menu
pscl
3. Press the Mode Key for more than one second to display 0, which is a simulated input value.
4. The simulated input value increases when the Up Key is pressed. Comparative outputs are output according to the output pattern that has been preset.
5. The simulated input value decreases when the Shift Key is pressed. Comparative outputs are output according to the output pattern that has been
preset.
6. Press the Escape Key after testing. The
test output test setting will be dis-
played again.
7. Press the Escape Key to return to the RUN mode.
· To change the simulated input value continuously, keep pressing the Up or
Shift Key.
setup
90
Page 99
Maintenance Mode
6-3Maintenance Mode
The set values of the K3NP can be initialized in maintenance mode. The operations of the K3NP in this mode are described below.
6-3-1Maintenance Mode
· The K3NP will be in maintenance mode if the Mode and Shift Keys are pressed
simultaneously while the K3NP is turned on.
· The K3NP in maintenance mode will go into RUN mode if the Escape Key is
pressed.
Press the Mode and Shift Keys
while turning the K3NP on.
Section 6-3
Power On
6-3-2Initialization
1, 2, 3...1. Press the Mode and Shift Keys simultaneously while turning the K3NP on.
Maintenance mode
1 second
RUN mode
RUNmode
Setting mode
Press the Escape and Up
Keys for 1 second.
Protect mode
Follow the steps described below to reset the set values of the K3NP to factoryset values.
mant maintenance mode setting will be displayed.
The
Set Value LED Display ModelBasic Model
2. Press the Mode Key for more than one second. The re-in initialization setting will be displayed.
Set Value LED Display ModelBasic Model
3. Press the Up Key to display no.
Set Value LED Display ModelBasic Model
91
Page 100
Maintenance Mode
Section 6-3
4. Press the Up Key to display yes. Press the Mode Key to initialize all set values. The K3NP will go into RUN mode.
Set Value LED Display ModelBasic Model
RUN mode
92
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.