Model HEM-629
TABLE OF CONTENTS
INTRODUCTION
Introduction ……………………………………………………………3 Know Your Unit…………………………………………………………4 Battery Installation/Replacement………………………………………5 Suggestions Before Measuring Blood Pressure ………………………6 How To Apply The Wrist Cuff …………………………………………7 How To Take A Reading ……………………………………………8-9 Memory Feature Instruction …………………………………………10 Troubleshooting ………………………………………………………11 Caution ………………………………………………………………12 Care and Maintenance ………………………………………………13 Warranty………………………………………………………………14 Specifications …………………………………………………………15
Thank you for purchasing the Omron HEM-629 Portable Wrist Blood Pressure Monitor. This remarkable instrument is ideal for people who frequently monitor their own blood pressure because it’s so compact and easy to use. There is no tricky arm cuff, just a small, pre-formed wrist cuff that is easy to apply. With the push of a button the Omron Wrist Monitor measures your blood pressure and pulse rate and displays the reading on a clear digital panel. Perfect for quick, easy readings at home, at work, and during travel.
The Omron Wrist Monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through the artery in your wrist and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use and measurements are easy to obtain.
Clinical research has proven a direct relationship between blood pressure in the wrist and blood pressure in the arm. Changes in wrist bloodpressure reflect changes in arm blood pressure because the arteries in the wrist and the arm are close to each other. Frequently measuring the blood pressure in your wrist will provide you and your doctor with an accurate indication of changes in your blood pressure.
Consult your physcician before measuring blood pressure at the wrist if you have any of the following conditions: severe hypertension, severe diabetes, severe arterialsclerosis, severe kidney disease, or vasculor disease that may compromise circulation.
Please read this instruction manual thoroughly before using the Omron Portable Wrist Blood Pressure Monitor. For specific information on your own blood pressure, CONTACT YOUR PHYSICIAN.
2
3
SUGGESTIONS BEFORE
MEASURING BLOOD PRESSURE
KNOW YOUR UNIT
1. Eating, smoking, and exercising within 30 minutes
before taking a measurement may affect your blood pressure. To take baseline blood pressure, rest for at least 5 minutes before taking a reading.
2. Mental stress raises blood pressure.
3. Throughout the measurement, the
unit should be held at your heart level. The carring case is large enough to raise the arm to heart level.
4. Remain still and do not talk
during the measurement.
Digital Display
START Button
Systolic Blood Pressure
Diastolic Blood Pressure/ Pulse Rate
MEMORY
button “M”
Wrist Cuff
ON/OFF Button
Battery Cover
4
5
BATTERY INSTALLATION/REPLACEMENT
HOW TO APPLY THE WRIST CUFF
1. Slide the battery cover off
in the direction of the arrow.
2. Insert or replace 2 "AAA" size
batteries so that the + (positive) and - (negative) markings match those in the battery compartment.
3. Replace the battery cover.
4. If the Low Battery Indicator
appears on the display, replace both batteries. Long-life alkaline batteries are recommended.
5. Remove the batteries if the
monitor will not be used for an extended period of time.
Low Battery Indicator
1. Hold your left arm in front of you
with your palm facing up. Apply the cuff so that the monitor is on the inside of your wrist with the display in easy view.
2. The cuff’s edge should be
approximately 1/4 to 1/2 inch from below your palm.
3. Wrap the cuff snugly around your
wrist. The top and bottom edges of the cuff should be tightened evenly around your wrist.
4. When the cuff is positioned
correctly, secure it in place by pressing the pile material firmly against the hook material.
1/4" - 1/2"
g
H m m
C
I L O
T
S Y S
g H m
m
IC
L
TO
n i
S
m /
IA
E S
D
L U P
F F
/O
N O
T R
A T S
g
H m m
C
I L O
T
S Y S
g
H m m
C I
in
L
/m
O
E
T
S
L
S
U
A
P
I D
F
F O /
N
O
T
R A T
S
6
7
HOW TO TAKE A READING
SYSTOLIC mm Hg
DIASTOLIC mm Hg
PULSE/min
HOW TO TAKE A READING
Read and follow directions in "Suggestions Before Measuring Blood Pressure" and "How To Apply The Wrist Cuff" before taking a reading.
1. Press the ON/OFF Button.
a) All the display symbols appear
for approximately one second. This is the initial LCD test.
SYSTOLIC mm Hg
b) The display symbols disappear
and the Deflation Indication Symbol ( ) starts to flash.
c) When the monitor completes
the necessary preparations before measurement, the
Heart Symbol () appears
with a zero.
2. Hold the unit at your heart level and press the START button.
The cuff automatically inflates to approximately 180 mm Hg.
a) Keep the unit at the heart level until the measurement is completed.
DIASTOLIC mm Hg
PULSE/min
SYSTOLIC mm Hg
DIASTOLIC mm Hg
PULSE/min
b) If during your measurement (deflation) the
monitor determines that a higher pressure value is required, it will momentarily stop deflation and automatically increase the pressure value to 230 mmHg.
c) To inflate higher than 230 mm Hg, during
deflation press and hold the START button until the necessary pressure value is reached. DO NOT INFLATE HIGHER THAN 280 mmHg.
3. When the cuff reaches your correct inflation
SYSTOLIC mm Hg
level, it begins to deflate automatically. As the cuff deflates, decreasing numbers appear on the display. The Heart Symbol
(
♥) flashes at every heart.
DIASTOLIC mm Hg
PULSE/min
SYSTOLIC mm Hg
4. The monitor exhausts the residual air when the
measurement is complete. The Heart Symbol
) stops flashing and your blood pressure
(
and pulse rate display alternately.
DIASTOLIC mm Hg
PULSE/min
NOTE: Wait 2-3 minutes before taking another blood pressure measurement. Waiting allows the engorged blood vessels to return to normal. You may require more rest time between readings depending on your individual physiological characteristics.
Remember To Keep A Record Of Your Blood Pressure And Pulse Measurements.
SYSTOLIC mm Hg
DIASTOLIC mm Hg
PULSE/min
8
9
MEMORY FEATURE INSTRUCTIONS
TROUBLESHOOTING
1. How to store the measured blood pressure values...
The measured and displayed blood pressure and pulse rate readings are automatically stored in the memory.
The memory can store the readings of up to fourteen measurements.
The memory stores the most recent 14 readings. When more than 14 readings have been taken the earliest readings are erased and replaced by the most recent readings. The memory is not erased when the batteries are replaced.
2. How to recall previously measured readings from memory...
While the () symbol is displayed, you can recall and display readings
from the memory starting with the previous measured readings.
When you push the MEMORY Button “M”, if there is any measured reading stored in the memory, M1 indicating the latest memory, the systolic and the diastolic blood pressures, and the pulse rate are displayed alternatively.
In order to recall the readings measured previously, push the MEMORY Button “M” again. The values will be displayed as M2, M3,... etc.
If there is no measured reading in the memory, a blank screen with dashed lines will be displayed, when the MEMORY Button “M” is pushed.
3. How to erase the memory...
All the measured readings stored in the memory can be erased simultaneously by operating the buttons as follows:
While the ( Buttons simultaneously. Hold both buttons down for at least two seconds, a blank screen with dashed lines will be displayed and the measured readings stored in the memory will be erased.
) symbol is displayed, push the MEMORY “M” and the START
Error Indicator Cause Correction
Cuff under-inflated
Movement during measurement.
No pulse or blood pressure reading due to very weak pulse.
Cuff over-inflated (more than 300 mmHg).
Battery voltage is
excessively low
Retake measurement by setting cuff inflation level 40 mmHg higher than previous level, by keep pressing the START BUTTON until the cuff pressure reaches the necessary level.
Be sure not to move and talk during the measurement.
Wrap the cuff snugly around your wrist, carefully read and repeat steps listed under “How To Apply The Wrist Cuff” (page 6).
Release the START BUTTON before the pressure indication reaches 280 mmHg or higher.
Replace both “AAA” batteries (See page 5).
NOTES:
- Wait 2-3 minutes before taking another measurement.
- If you continue to have difficulty obtaining a reading, call
OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350. Please tell the representative
you have an Omron HEM-629 Portable Wrist Blood Pressure Monitor.
10
11
CAUTION
CARE AND MAINTENANCE
CAUTION:
Changes or modifications not expressly approved by Omron Healthcare, Inc. will void the user’s warranty.
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules.These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Connect the product into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only) This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference-causing equipment standard entitled "Digital Apparatus", ICES-003 of the Canadian Department of Communications.
Cet appareil numerique respecte les limites de bruits radioelectriques applicables aux appareils numeriques de Classe B prescrites dans la norme sur le materiel brouilleur: "Appareils Numeriques," NMB-003 edictee par le ministre des communications.
LICENSED UNDER U.S. PAT. 4829419, 4942516, 4896260, 4531182, 4825364, 4686622, 4121284 4396976, 4371923, 4471385, 4120583, 4435732, 4672457, 4739396
To protect your monitor from damage and loss of warranty, please AVOID the following:
Subjecting your monitor to extreme temperatures, humidity, and direct sunlight.
Washing or moistening the cuff.
Disassembling the monitor.
Subjecting the monitor to strong shocks
(for example, dropping the monitor on the floor).
Cleaning the monitor with volatile liquids. THE MONITOR
SHOULD BE CLEANED WITH A SOFT, DRY CLOTH.
12
13
LIMITED FIVE YEAR WARRANTY
SPECIFICATIONS
Your Omron HEM-629 Portable Wrist Blood Pressure Monitor is warranted to be free from manufacturing defects for a period of five years under normal use. The five year warranty excludes the monitor cuff. The cuff is warranted for a one year period. This warranty extends only to the original retail purchaser.
Should repair be needed within the warranty period, ship the unit prepaid to Omron Healthcare, Inc., Attn: Repair Dept., 300 Lakeview Parkway, Vernon Hills, IL 60061, include $5.00 for the return shipping and insurance. Be sure to include the model number of your unit and your phone number on any correspondence.
We will either repair or replace (at our option) free of charge any parts necessary to correct defects in the materials or workmanship.
The above warranty is complete and exclusive. The warrantor expressly disclaims liability for incidental, special, or consequential damages of any nature. (Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above warranty may not apply to you.)
Any implied warranties arising by the operation of law shall be limited in duration to the term of this warranty. (Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you.)
This warranty gives you specific legal rights and you may have other rights which vary from state to state. As a condition to operation of your warranty, the enclosed registration card must be completed and sent to us within 10 days from the date of purchase.
FOR CUSTOMER SERVICE CALL TOLL FREE:
1-800-634-4350
Model: HEM-629 Display: Liquid Crystal Digital Display Measurement Range: Pressure: 0 to 280mmHg
Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading
Inflation: Automatic by electric pump Deflation: Automatic deflation system Rapid Pressure Release: Automatic pressure release valve Pressure Detection: Capacitive pressure sensor Measurement Method: Oscillometric method Pulse Wave Detection: Capacitive pressure sensor Power Source: 2 "AAA" batteries Battery Life: Approximately 1 year when used
Operating Temperature/Humidity: 50°F to 104°F (10°C to 40°C)
Storage Temperature/Humidity: 14°F to 140°F (-10°C to 60°C)
Console Weight: Approximately 5 oz. (140g) with batteries Outer Dimensions: Approximately 3" (l) x 3" (w) x 1 1/2" (h)
Cuff Dimensions: Approximately 3" x 12" (76mm x 304mm) Wrist Circumference Range: Approximately 5 1/4" x 7 3/4" (135mm x 195mm) Accessories: Carrying case, instruction manual, 2 "AAA" batteries*
These specifications are subject to change without notice for purposes of improvement. * The attached batteries are only to confirm the unit operation. They may run out before
indicated battery life.
Pulse: 40 to 200/minute
Pulse: ±5% of reading
once a day for two minutes*
30 to 85% RH maximum
10 to 95% RH maximum
(76mm x 76mm x 38mm)