Thank you for purchasing the Omron HEM-412C Manual Inflation 
Blood Pressure Monitor. This easy-to-use monitor quickly detects 
your blood pressure and pulse measurements and displays them on 
a large, clear digital panel.
Your new HEM-412C uses the oscillometric method of blood pressure 
measurement. This means the monitor detects your blood's movement 
through your brachial artery and converts the movements into a digital 
reading. An oscillometric monitor does not need a stethoscope so the 
monitor is simple to use. The monitor features an easy-to-squeeze inflation 
bulb that allows you to quickly fill the cuff with air. Stop pumping and the 
cuff deflates automatically. Error indicators alert you to incorrect procedures 
and an automatic power-off function conserves batteries.
Please read this instruction manual thoroughly before using the 
Omron HEM-412C Manual Inflation Blood Pressure Monitor. 
For specific information on your own blood pressure, 
CONTACT YOUR PHYSICIAN.
INTRODUCTION

4
Diastolic Blood Pressure 
/Pulse Display
Systolic 
Blood Pressure Display
Air 
Jack
Air Tube
Air Inflation Bulb
Air 
Release 
Button
Cuff Tab
Arm Cuff
Air Plug 
(Connects 
to the 
air jack)
On/Off 
Button
KNOW YOUR UNIT

5
1. Avoid eating, smoking, and exercising for at least 30 minutes 
before taking a measurement.
2. Put your left arm through the cuff loop. The bottom of the cuff 
should be about 1/2'' above the elbow. The cuff tab should lie 
over the brachial artery on the inside of the arm.
3. Pull the cuff so that it is tightened evenly around your arm. 
Press the sewn hook material firmly against the pile side of 
the cuff. Make sure that your skin will not be pinched in the 
D-ring once you inflate the cuff. Immediately deflate the cuff 
and readjust it if skin becomes pinched.
4. Sit in a chair with your feet flat on the floor and place your left 
arm on a table so that the cuff is at the same level as your heart.
5. After the Ready To Measure Symbol (  ) appears next to a zero 
on the digital panel, quickly squeeze the inflation bulb to inflate the 
cuff. Stop pumping when the numbers on the display are approximately 
40 mmHg higher than your normal systolic blood pressure.
7. The cuff will deflate automatically. Remain still throughout 
the entire reading.
8. When measurement is complete, the monitor alternately 
displays your blood pressure and pulse on the digital panel.
9. Wait 5-10 minutes before taking another blood pressure 
measurement. You may need more rest time between 
readings depending on your individual physical characteristics.
QUICK REFERENCE GUIDE

6
1. Slide the battery cover off in the 
direction of the arrow.
2. Install 4 “AA” size batteries so 
+ (positive) and – (negative) 
polarities match the polarities 
of the battery compartment as indicated.
3. Replace the battery cover.
4. If the Low Battery Indicator appears 
on the display, replace all four batteries. 
Long-life alkaline batteries 
are recommended.
5. Remove the batteries if the monitor will 
not be used for an extended period of time.
Low Battery Indication
BATTERY INSTALLATION/REPLACEMENT

7
HOW TO APPLY THE ARM CUFF
1. Locate your brachial artery by 
pressing two fingers approximately 
one inch above the elbow crease on the 
inside of the left arm. Determine where 
your pulse is strongest.
2. The cuff should be assembled 
correctly when it is removed from 
the box. If it is not, pass the end 
of the cuff furthest from the tubing 
through the metal D-ring to form 
a loop. The smooth cloth should 
be on the inside of the cuff loop.
3. If the cuff is assembled correctly, the 
sewn hook material will be on the 
outside of the cuff loop and the metal 
D-ring will not touch your skin.
4. Put your left arm through the cuff 
loop. The bottom of the cuff should 
be approximately 1/2 inch above the 
elbow. The cuff tab should lie over 
the brachial artery on the inside 
of the arm.

8
HOW TO APPLY THE ARM CUFF
5. Pull the cuff so that the top and 
bottom edges are tightened evenly 
around your arm.
6. When the cuff is positioned correctly, 
press the sewn hook material FIRMLY 
against the pile side of the cuff.
7. Make certain the cuff fits snugly 
around your arm. The cuff should 
make good contact with your skin.
8. Sit in a chair with your feet flat on the
floor and place your arm on a table so 
that the cuff is at the same level 
as your heart.
9. Relax your arm and turn your 
palm upward.
10. Be sure there are no kinks in 
the air tubing.

9
HOW TO TAKE A READING
1. Press the On/Off button. 
a) All display symbols appear 
for about one second. This is the 
initial LCD Test.
b) The display symbols 
disappear and the Deflation 
Indication Symbol (  ) 
starts to flash.
c) When the monitor completes 
the necessary preparations 
before measurement, the 
Ready To Measure Symbol 
(  ) appears next to a zero.
NOTE: Wait for the Ready To Measure 
Symbol (  ) before taking a measurement.
2. Begin inflating the cuff by rapidly squeezing the inflation bulb 
until you have reached approximately 40 mmHg higher than 
your normal systolic reading. If your normal systolic blood 
pressure is unknown, inflate the cuff to 180 mmHg.
Please read “How To Apply The Arm Cuff” before taking a reading.
➧
Deflation Indication
Ready To Measure
Initial LCD Test
SYSTOLIC mm Hg
DIASTOLIC mm Hg
PULSE/MIN

6. When the Deflation Indication Symbol (  ) appears on the 
display, push the air release button to release any air left in the cuff.
NOTE: Wait 5-10 minutes before taking another blood pressure measurement. 
Waiting allows the engorged blood vessels to return to normal. You may require 
more rest time between readings depending on your individual 
physical characteristics.
REMEMBER TO KEEP A RECORD OF YOUR BLOOD PRESSURE 
AND PULSE MEASUREMENTS.
➧
10
HOW TO TAKE A READING
3. Stop pumping the inflation bulb when 
you have reached your correct inflation 
level and remain still. The cuff will 
begin to deflate automatically.
4. As the cuff deflates, decreasing 
numbers appear on the display. 
The Ready To Measure Symbol 
(   ) flashes at the onset of 
oscillation signals.
5. When measurement is complete, 
the Ready To Measure Symbol 
(  ) stops flashing and your blood 
pressure and pulse display alternately.
Indicated Value 
Drops
➞
Measuring
Blood Pressure 
Reading Display 
(for approx. 
3.5 sec.)
Pulse Reading Display 
(for approx. 2 sec.)
➞
➞

11
To protect your monitor from damage, please do the following:
• Do not subject your monitor to extreme temperatures, 
humidity, and direct sunlight.
• Do not fold the cuff and tubing tightly.
• Do not inflate the monitor over 280 mmHg.
• Do not disassemble the monitor.
• Do not subject the monitor to strong shocks 
(for example, dropping the monitor on the floor).
• Do not clean the monitor with volatile liquids. 
THE MONITOR SHOULD BE CLEANED WITH 
A SOFT, DRY CLOTH.
CARE AND MAINTENANCE

12
ERROR INDICATORS
Error Symbol Cause Correction
If you continue to have difficulty obtaining a reading, call OMRON HEALTHCARE 
CUSTOMER SERVICE TOLL FREE 1-800-634-4350. Please tell the representative
you have a Omron Model HEM-412C Manual Inflation Blood Pressure Monitor.
Cuff under-inflated.
Wait 10 minutes before 
taking another measurement. 
Repeat steps listed under 
“How To Tke a Reading”, 
but manually inflate the 
cuff 40 mmHg higher 
than the previous 
inflation value.
Replace all 4 four 
“AAA” batteries.
Batteries have run down.
Cuff over-inflated 
(more than 300 mmHg.)
Residual air pressure 
in cuff.
Movement 
during measurement.
No pulse or blood 
pressure reading due 
to very weak artery pulse.
Carefully read and repeat 
steps linsted under “How 
to Take a Reading.”

13
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) 
This product has been tested and found to comply with the limits for a Class B digital device, 
pursuant to part 15 of the FCC rules. 
These limits are designed to provide reasonable protection against harmful interference in 
a residential installation. The product generates, uses, and can radiate radio frequency energy 
and, if not installed and used in accordance with the instructions, may cause harmful interference 
to radio communications. However, there is no guarantee that interference will not occur 
in a particular installation. If the product does cause harmful interference to radio or television 
reception, which can be determined by turning the product on and off, the user is encouraged 
to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Connect the product into an outlet on a circuit different from that to which 
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only) 
This digital apparatus does not exceed the Class B limits for radio noise emissions from digital 
apparatus as set out in the interference-causing equipment standard entitled "Digital Apparatus", 
ICES-003 of the Canadian Department of Communications.
LICENSED UNDER U.S, PAT. 
4829419, 4942516, 4896260, 4531182, 4825364, 4686622, 4121284, 
4396976, 4371923, 4471385, 4120583, 4435732, 4672457, 4739396
CAUTION INDICATORS

14
Model: HEM-412C 
Display: LCD digital display 
Measurement Range: Pressure: 0-280 mmHg
Pulse: 40-200/min.
Accuracy/Calibration: Pressure: ±3 mmHg or 2% of reading
Pulse: ±5% of reading 
Inflation: Manual by inflation bulb 
Deflation: Automatic pressure release valve 
Rapid Air Release: Push button valve 
Pressure Detection: Capacitive pressure sensor 
Measurement Method: Oscillometric method 
Pulse Wave Detection: Capacitive pressure sensor 
Power Source: 4 "AA" batteries (not included) 
Battery Life: Approximately 1 year when used once a day
for 2 minutes 
Operating Temperature/ 50°F to 104°F (10°C to 40°C)
Humidity: 30 to 85% RH maximum 
Storage Temperature/ -4°F to 140°F (-20°C to 60°C)
Humidity: 10 to 95% RH maximum 
Console Weight: Approximately 8 oz (225g) with batteries 
Outer Dimensions: Approximately 4 7/8" (w) x 2 1/5'' (l) 
x 3 1/6" (h) (124mm x 56mm x 80mm) 
Cuff Dimensions: Approximately 5 1/2" (w) x 18 7/8" (l)
(140mm x 480mm) 
Arm Circumference Range 9'' to 13'' 
UPC Code: 0 73796 04122 9
NOTE: These specifications are subject to change without notice.
SPECIFICATIONS

15
LIMITED ONE YEAR WARRANTY
Your HEM-412C Manual Inflation Blood Pressure Monitor is warranted to be free from 
manufacturing defects for a period of one year under normal use. This warranty extends 
only to the original retail purchaser.
Should repair be needed within the warranty period, ship the unit prepaid to 
Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061, 
Attn: Service Dept., together with $5.00 for return shipping and insurance. 
Be sure to include the model number of your unit and your phone number 
on any correspondence.
We will either repair or replace (at our option) free of charge any parts necessary 
to correct defects in the materials or workmanship.
The above warranty is complete and exclusive. The warrantor expressly disclaims 
liability for incidental, special, or consequential damages of any nature. (Some states 
do not allow the exclusion or limitation of incidental or consequential damages, 
so the above warranty may not apply to you.)
Any implied warranties arising by the operation of law shall be limited in duration 
to the term of this warranty. (Some states do not allow limitations on how long an 
implied warranty lasts, so the above limitation may not apply to you.)
This warranty gives you specific legal rights and you may have other rights which 
vary from state to state. As a condition to operation of your warranty, the enclosed 
registration card must be completed and sent to us within 10 days 
from the date of purchase.
FOR CUSTOMER SERVICE CALL TOLL FREE:
1-800-634-4350

Manufactured for:
OMRON HEALTHCARE, INC.
300 Lakeview Parkway 
Vernon Hills, Illinois 60061 
www.omronhealthcare.com
Copyright © 2002 OMRON HEALTHCARE, INC. 
HEM-412CINST Rev A
For Customer Service