NULL AND SPAN ADJUSTMENTS (Voltage and Current Models Only)
The PX700 Pressure Sensors have been factory calibrated to give the rated
output and no adjustment should be necessary. However, if so desired,
small residual offsets can be corrected on the voltage and currrent models
by the following procedure:
1. Connect unit as per terminal plate.
2. With zero pressure applied to the transmitter pressure port, adjust NULL
ADJ trimpot for rated output at zero pressure.
To allow precise calibration the trimpots are multiturn types.
3. Apply a known accurate pressure to the transmitter pressure port.
Ideally, this pressure should be the full range pressure of the transmitter;
however, it is acceptable to use an intermediate value. Adjust the SPAN
ADJ for the rated output at the applied pressure.
4. Because of a small interaction between the SPAN ADJ and the NULL ADJ
trimpots. repeat steps 2 ad 3 as required.
MIINIMUM ADJUSTMENT RANGE
VOLTAGE MODELS CURRENT MODELS
0.5 to 5.5V 1 to 11V 4 to 20 mA
Zero 0.2 to 1.3 V 0.4 to 1.3 V 3.6 to 5.0 mA
Span 4.8 to 5.2 V 9.2 to 10.8 V 15 to 17 mA
PX700 SERIES SPECIFICATIONS
EXCITATION: 0.5 to 5.5 Models: 12 to 32 Vdc
1 to 11 V Models: 14 to 32 Vdc
I Models: 12 to 67 Vdc (with appropriate
loop resistance - see chart)
OUTPUT: V Models: 0.5 to 5.5 or 1 to 11 V;
I Models: 4 to 20 mA
PERFORMANCE: 2X full scale or 13,000 psi
PROOF PRESSURE: Whichever is less
ACCURACY: 0.25% full scale at 75% (includes linearity,
hystersis, and repeatability)
ZERO BALANCE: ±2% of full scale
SPAN ACCURACY: ±1% of full scale 75°F
COMPENSATED 30 to 160°F
TEMPERATURE
RANGE:
TOTAL THERMAL EFFECT: 1% of full scale max.
OPERABLE
TEMPERATURE RANGE: 0 to 160°F