Omega Products PX655-0.5DI User Manual

HIGHLY ACCURATE, LOW-PRESSURE LABORATORY TRANSMITTER
0-0.1 to 0-50 inH
O
2
0-25 Pa to 0-1.25 kPa
PX655 Series
Standard
Ideal Applications:
U Clean Rooms U HVAC U Fume Hoods U Laboratory Monitoring
PX655-0.5BDI, with DP3002-E meter, shown smaller than actual size.
SPECIFICATIONS
Excitation: 12 to 36 Vdc Output: 4 to 20 mA (2 wire) Linearity: 0.3% FS (BFSL) Hysteresis: 0.02% FS Repeatability: 0.05% FS Operating Temp: -29 to 72°C
(-20 to 160°F) Compensated Temp: 2 to 57°C
(35 to 135°F) Thermal Effects:
Zero: 0.015% FS/°F Span: 0.015% rdg/°F
Proof Pressure: 15 psi Burst Pressure: 20 psi Static Pressure: 25 psi Gage Type: Capacitance Max Loop Resistance:
(Supply voltage – 12) x 50 Ω Calibration Report: NIST cal at 25,
50, 75 and 100% FS; upscale and downscale provided
Response Time: 250 ms Wetted Parts: Dry, clean, non-corrosive
gases only
Enclosure: NEMA 2 (IP62) Dimensions: 107 H x 117 W x 53 mm D
(4.2 x 4.6 x 2.1")
Pressure Port:
(tubing TY-316-100)
Electrical Conn.: Screw terminal Weight: 368 g (13 oz)
1
4 barbed fittings
To Order
inH
O Pa/kPa MODEL NO. COMPATIBLE METERS
2
0 to 0.1 0 to 25 PX655-0.1DI DP25B-E, DP41-E, DP460-E 0 to 0.25 0 to 62 PX655-0.25DI DP25B-E, DP41-E, DP460-E 0 to 0.50 0 to 125 PX655-0.5DI DP25B-E, DP41-E, DP460-E 0 to 0.75 0 to 187 PX655-0.75DI DP25B-E, DP41-E, DP460-E 0 to 1 0 to 249 PX655-01DI 0 to 1.5 0 to 374 PX655-01.5DI 0 to 2 0 to 498 PX655-02DI DP25B-E, DP41-E, DP460-E 0 to 3 0 to 748 PX655-03DI DP25B-E, DP41-E, DP460-E 0 to 5 0 to 1.25 PX655-05DI DP25B-E, DP41-E, DP460-E 0 to 10 0 to 2.49 PX655-10DI 0 to 25 0 to 6.23 PX655-25DI DP25B-E, DP41-E, DP460-E 0 to 50 0 to 12.5 PX655-50DI DP25B-E, DP41-E, DP460-E
BI-DIRECTIONAL RANGES
±0.1 25 PX655-0.1BDI DP25B-E, DP41-E, DP460-E ±0.25 62 PX655-0.25BDI DP25B-E, DP41-E, DP460-E ±0.50 125 PX655-0.5BDI DP25B-E, DP41-E, DP460-E ±1 249 PX655-01BDI DP25B-E, DP41-E, DP460-E ± 2.5 623 PX655-2.5BDI DP25B-E, DP41-E, DP460-E ±5 1.25 PX655-05BDI DP25B-E, DP41-E, DP460-E ±10 2.49 PX655-10BDI DP25B-E, DP41-E, DP460-E ± 25 6.23 PX655-25BDI DP25B-E, DP41-E, DP460-E ± 50 12.5 PX655-50BDI DP25B-E, DP41-E, DP460-E
Comes complete with NIST traceable calibration and operator’s manual. Ordering Examples: PX655-01DI, 0 to 1 inH
output pressure transmitter. PX655-01BDI, ±1 psi differential pressure range with 4 to 20 mA output (0 set at 12 mA).
Visit omega.com/px655 for Pricing and Details
RANGE
DP25B-E, DP41-E, DP3002-E DP25B-E, DP41-E, DP3002-E
DP25B-E, DP41-E, DP3002-E
O pressure transducer range with 4 to 20 mA
2
B-223