Installation of the Ink Ribbon ........................... 16
Installation of the Print Sheets ......................... 18
3.3 Setup of Connected Equipment .................. 21
3.4 Inspection of the LCD Panel Display ......... 22
3.5 Setup of This Instrument .............................23
3.6 Inspection of the Monitor Display .............. 24
3.7 Check the Printed Picture ........................... 25
Table of Contents
3
Page 4
Labels and Symbols
Safety-related labels and symbols are attached to this
instrument (OEP-5 color video printer) at the locations
shown below.
Color video printer (OEP-5)
Electrical rating
The product name, rated voltage and
frequency are shown.
If labels or symbols are missing or illegible, contact
Olympus.
Equipotential terminal
Instruction manual reference mark
UDI label
The UDI label is required by some countries’ regulations regarding the identification of a medical device also
known as Unique Device Identification (UDI).
For US Customers only
For a Symbols Glossary, visit us:
http://www.olympus-global.com/en/common/pdf/symbolsglossary.pdf
C-UL mark
FCC mark
4
Labels and Symbols
Page 5
Important Information –
Please Read Before Use
Intended Use
The OEP-5 color video printer has been designed to be
used with Olympus endoscopes for recording and
printing endoscopic documentation.
Do not use this instrument for any purpose other than its
intended use.
SDTV
Standard Definition Television. It is the format used by
ordinary television systems.
HDTV
High Definition Television. It is a format featuring
higher image definition than the SDTV format.
SDI
Serial Digital Interface. It is a digital interface standard
used in video distribution.
Color printing pack
Set of ink ribbon and print sheets used to print pictures
on this instrument.
Instruction Manual
This instruction manual contains essential information
on using this instrument safely and effectively. Before
use, thoroughly review this manual and the manuals of
all equipment that will be used during the procedure and
use the equipment as instructed.
In addition, the “INSTRUCTIONS” for this instrument
are incorporated by “BASIC” and “ADVANCED.”
This “INSTRUCTIONS” is “BASIC,” and provides the
most basic information about this instrument.
More detailed information is described in
“ADVANCED.”
Keep this and all related instruction manuals in a safe,
accessible location. If you have any questions or
comments about any information in this manual, please
contact Olympus.
Terms used in this manual
Wall mains outlet
An electrical outlet that has a terminal used exclusively
for grounding.
Isolation transformer
The isolation transformer is a safety device that is used
to isolate non insulated equipment with potentially high
leakage currents to decrease the possibility of electric
shock.
Freeze
The freeze function creates a stationary view of the
moving image.
Video system center/Video system
These are the devices used to convert the endoscopic
image captured using a videoscope or fiberscope into the
signal to be output to the monitor device.
Ultrasound center
The ultrasound center converts the ultrasonic signals
from an ultrasonic endoscope or probe into monitor
images.
Source image
Live image being input from the video signal input
source.
Memory image
Image stored in the memory of this instrument.
Capture
Stores the image, sent from connected equipment, in the
memory of this instrument for printing.
Wireless remote control
Remote control that operates this instrument by radio
communications.
Wired remote control
Remote control that operates this instrument by cable
connecting.
Instrument Compatibility
Refer to the “System Chart” on page 31 in the
“Appendix” to confirm that this instrument is
compatible with the ancillary equipment being used.
Using incompatible equipment can result in patient or
operator injury and/or equipment damage. It may also
impair the functionality of the instrument.
This instrument complies the EMC standard for medical
electrical equipment edition 2 (IEC 60601-1-2 : 2001)
and edition 3 (IEC 60601-1-2 : 2007). However when
connecting to an instrument that complies with the EMC
standard for medical electrical equipment edition 1 (IEC
60601-1-2 : 1993), the whole system then complies with
edition 1. (See, “EMC Information” on page 36 for
EMC compliance level.)
For the customers in Canada
This unit has been certified according to Standard CSA
C22.2 No.601.1.
Important Information – Please Read Before Use
5
Page 6
Repair and Modification
Dangers, Warnings and Cautions
This instrument does not contain any user-serviceable
parts. Do not disassemble, modify or attempt to repair it;
patient or operator injury and/or equipment damage can
result.
Some problems that appear to be malfunctions may be
correctable by referring to “Chapter 7 Troubleshooting”
in “INSTRUCTIONS (ADVANCED).” If the problem
cannot be resolved using the information in Chapter 7 in
“INSTRUCTIONS (ADVANCED),” contact Olympus.
Signal Words
The following signal words are used throughout this
manual:
DANGER
Indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WAR NING
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Follow the dangers, warnings and cautions given below
when handling this instrument. This information is to be
supplemented by the dangers, warnings and cautions
given in each chapter.
Especially, about a power supply, refer to “2.4
Connection to the AC Mains Power Supply” on page 13.
DANGER
• Strictly observe the following precautions. Failure to
do so may place the patient and medical personnel in
danger of an electric shock:
– Keep fluids away from all electrical equipment. If
fluids are spilled on or into this instrument,
immediately stop operating it and contact Olympus.
– Do not prepare, inspect or use this instrument with
wet hands.
• Never install and operate this instrument in locations
where:
– The concentration of oxygen is high.
– Oxidizing agents (such as nitrous oxide (N
2O)) are
present in the atmosphere.
– Flammable anesthetics are present in the
atmosphere.
Otherwise, explosion or fire may result because this
color video printer is not explosion-proof.
Indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury. It may
also be used to alert against unsafe practices or potential
equipment damage.
NOTE
Indicates additional helpful information.
WARNING
• Be sure to turn this instrument OFF and unplug the
power cord before proceeding to care and
maintenance of this instrument. If the power cord is
left plugged in, an electric shock may result.
• Never insert metal and/or thing that are easy to burn,
or do not drop it into this instrument inside from the
ventilation grills. Electric shock and/or fire may result.
Moreover, if something go into the inside of this
instrument, turn OFF the power immediately and stop
use, then disconnect the power cord and connecting
cables. Contact Olympus.
• When there are abnormally sound, smell and/or heat
from this instrument, stop use immediately and
contact Olympus.
CAUTION
• Do not use a pointed or hard object to press the
switches on the front panel and/or remote control unit
(MAJ-898) (optional). This may damage the switches.
• Do not use this instrument roughly. Otherwise,
equipment damage, patient and operator injury can
occur.
• Strictly observe the following precautions. Otherwise,
damage and/or malfunction can occur.
– Do not touch the electrical contacts inside this
instrument’s terminals.
6
Important Information – Please Read Before Use
Page 7
– Do not apply excessive force to the terminals of this
instrument.
• When using spray-type medical agents such as
lubricate, anesthetic, or alcohol, use them away from
this instrument so that the medical agents will not fall
on this instrument. The medical agents possibly may
come into this instrument from its ventilation grills
and cause the failure.
• Do not use a humidifier by this instrument as dew
condensation possibly might occur and it may cause
the failure.
• Be sure that this instrument is not used adjacent to or
stacked with other equipment (other than the
components of this instrument or system) to avoid
electromagnetic interference.
• Electromagnetic interference may occur to this
instrument near equipment marked with the following
symbol or other portable and mobile radio
telecommunications equipment such as cellular
phones. If radio interference occurs, mitigation
measures may be necessary, such as reorienting or
relocating this instrument or shielding the location.
For the State of California, USA only
Perchlorate Material - special handling may apply, See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Perchlorate Material : Lithium battery contains
perchlorate.
For the customers in Taiwan only
Important Information – Please Read Before Use
7
Page 8
1
Chapter 1 Package Contents
1.1Checking the
Package Contents
Match all items in the package with the components
shown below. Inspect each item for damage. If this
instrument is damaged, a component is missing, or you
Package Contents
have any questions, do not use this instrument;
immediately contact Olympus.
Color video printer (OEP-5)
Color video printer
(OEP-5)
Paper trayPower cordINSTRUCTIONS
Remote control unit (MAJ-898) (optional)
Remote control unit (MAJ-898)Remote cable
Trial pack (100 sheets)Ink ribbon holder
(BASIC)
“AA”-size batteries are not
supplied. Prepare batteries
before use.
INSTRUCTIONS
(ADVANCED)
8
1.1 Checking the Package Contents
Page 9
Chapter 2 Installation and Connection
Prepare this instrument and compatible equipment
(shown in the “System Chart” on page 31 in
“Appendix”) before each use, and refer to the instruction
manuals of each system component. Install and connect
the equipment as follows:
In addition, this “INSTRUCTIONS (BASIC)” explains
the EVIS EXERA II Video System Center (CV-180) and
the VISERA ELITE Video System Center (OTV-S190).
For other equipment, refer to “Chapter 2 Installation and
Connection” in “INSTRUCTIONS (ADVANCED).”
WARNING
• Review this chapter thoroughly before each use. If the
equipment is not properly prepared before each use,
equipment damage, patient and operator injury and/or
fire can occur.
• When non medical electrical ancillary equipment is
used, connect its power cord via an isolation
transformer prior to connecting it to this instrument.
Failure to do so can cause electric shock, burns and/or
fire.
• Use this instrument, the remote control unit (MAJ-
898) and color printing pack (UPC-55) under the
conditions described in “Environment” on page 33
and “Specifications” on page 34 in “Appendix”.
Otherwise, improper performance, compromised
safety and/or equipment damage may result.
• Connect the power cord to a wall mains outlet, and the
connecting cables to each equipment firmly. If
connections are not made correctly/firmly, fire or
electric shock may result.
2.1Installation of the
Color Video Printer
(OEP-5)
CAUTION
• Do not place any object on the top of this instrument.
Otherwise, equipment damage may result.
• Keep the ventilation grilles of this instrument clear.
Blockage may cause overheating and equipment
damage.
• Place this instrument on a stable and level surface.
Otherwise, this instrument topple down or drop.
• If a trolley other than the mobile workstation (WMNP1, WM-WP1) is used, confirm that the trolley can
withstand the weight of the equipment installed on it.
• Do not use this instrument in an inclined position.
Unstable operation, damage and/or abnormal printing
may result.
• Do not install this instrument near a source of strong
magnetic radiation (microwave treatment device,
short wave treatment device, MRI, radio equipment,
etc.). Otherwise, this instrument may cause
malfunction.
2
Installation and Connection
CAUTION
• Turn OFF all equipment before connecting them.
Otherwise, equipment damage or malfunction may
result.
• Use appropriate cables. Otherwise, equipment
damage or malfunction may result.
• The cables should not be sharply bent, pulled, twisted
or crushed.
• When you connect another piece of video equipment
between this instrument and the video system center,
be sure not to turn off the power of the video
equipment while this instrument is being used. If you
turn off the power of the video equipment, the video
signal may not be input to this instrument properly,
and this instrument may not work correctly, depending
on the video equipment connected.
2.1 Installation of the Color Video Printer (OEP-5)
9
Page 10
2
Installation and Connection
Installation on the mobile workstation
(WM-NP1, WM-WP1)
1
Place the mobile workstation on a level surface.
Lock the caster brakes by pushing them down (see
Figure 2.1).
Caster
m
Engaging the caster
brakes
Caster brake
Figure 2.1
2
Install the mobile shelf of the mobile workstation
according to the configuration of the equipment
installed on it, as described in the mobile
workstation’s instruction manual.
3
Place the color video printer (OEP-5) on the mobile
shelf of the mobile workstation.
4
Be sure the portion of the color video printer (OEP-
5) does not contact with the mobile workstation,
etc., when you attach or remove the paper tray, or
open the front door.
If the color video printer (OEP-5) comes into
physical contact with the mobile workstation, etc.,
move the color video printer (OEP-5) to a location
where it will not contact with the mobile
workstation.
Releasing the
caster brakes
10
2.1 Installation of the Color Video Printer (OEP-5)
Page 11
2.2Connection of the EVIS EXERA II Video System
Center (CV-180)
Attach the two black
screws, provided with the
remote cable (MH-995).
Color video printer (OEP-5)
BNC cable (MB-677)
PRINTER OUT
terminal
RS-232C terminal
2
Installation and Connection
INPUT
terminals
VIDEO OUT
terminal
SBGR
RGB cable (MH-984)
Remote Cable (MH-995)
PRINTER IN
terminal
NOTE
PRINTER REMOTE
terminal
For connecting the EVIS EXERA II Video System
Center (CV-180) to a moniotr, refer to the instruction
manual supplied with the EVIS EXERA II Video
System Center (CV-180).
EVIS EXERA II Video System Center (CV-180)
Figure 2.2
2.2 Connection of the EVIS EXERA II Video System Center (CV-180)
11
Page 12
2.3Connection of the VISERA ELITE Video System
Center (OTV-S190)
2
Installation and Connection
SDI 1 INPUT terminal
RS-232C terminal
SDI cable (MAJ-1951)
VIDEO INPUT terminal
High Definition LCD Monitor (OEV261H)
REMOTE2 terminal
BNC cable (MB-677)
VIDEO OUT terminal
Color video printer (OEP-5)
Multi-monitor remote cable (MAJ-1957)
MONITOR REMOTE terminal
VISERA ELITE Video System Center (OTV-S190)
12
2.3 Connection of the VISERA ELITE Video System Center (OTV-S190)
SDI INPUT terminal
SDI cable (MAJ-1951)
HD/SD SDI OUT2 terminal
HD/SD SDI OUT1 terminal
Figure 2.3
Page 13
2.4Connection to the
AC Mains Power
Supply
DANGER
Always power this instrument from a wall mains outlet
that has an exclusive terminal for grounding this
instrument. Otherwise, fire or electric shock may result.
WARNING
• Strictly observe the following precautions for the
power cord. Damaging the power cord may result in
fire or an electric shock:
Moreover, if the power cord is damaged, turn OFF the
power and stop using this instrument immediately and
contact Olympus.
– Take care to place the cord such that it is not crushed
between the instrument and a wall, mobile
workstation or a shelf, etc.
– Do not modify or damage the power cord.
– Do not use damaged power cord.
– Do not stretch or place a heavy object on the power
cord.
– Do not place the power cord near heating equipment
or expose it to excessive heat.
– Always grasp and pull the plug when unplugging the
power cord. Do not pull on the cord itself.
• Always use this instrument with the voltage specified
in “Specifications” on page 34 in “Appendix”. With an
incorrect supply voltage, fire and/or electric shock
may result.
• If there is metal and/or dust on the power cord plug or
near the plug of the power cord, disconnect the power
cord and remove it with a dry cloth. Otherwise, fire
and/or electric shock may result.
• Always use the power cord and connection cables that
were shipped with this instrument or mobile
workstation. Using other power cords or connection
cables may result in electric shock and/or malfunction.
• Do not allow the power plug to become wet. Also do
not hold the power cord with a wet hand. A wet power
plug may cause electric shock.
• Confirm that the wall mains outlet has adequate
electrical capacities. Failure to do so may cause fire or
a power failure to all of the equipment connected to
the same power circuit.
• Do not bend, pull or twist the power cord. Equipment
damage, including separation of the power plug or
disconnection of the cord wire, as well as fire or
electric shock, can result.
• Do not use the adapter that the power cord plug
changed into 3-pin to 2-pin. Electric shock may result.
CAUTION
• Use a power supply that suits the specifications in
“Specifications” on page 34 in “Appendix.”
Otherwise, malfunction may result.
• Be sure to connect the power plug securely.
Otherwise, the equipment will not function.
• Do not use of a power supply with insufficient
electrical capacities. Otherwise, malfunction may
result.
• Be especially careful not to allow the power cord to
become disconnected. If the power cord disconnects
during use, any picture being printing will be lost, or
may not be printed correctly.
• Do not connect or disconnect the power cord while
power is ON. Otherwise, malfunction may result.
To connect to the AC mains power supply
1
Confirm that the color video printer (OEP-5) is
turned OFF.
Power switch
Figure 2.4
2
Connect the power cord provided with the mobile
workstation to the AC mains power inlet of the
color video printer (OEP-5) (see Figure 2.5).
AC mains power inlet
Power cord
Figure 2.5
3
Connect the power cords of the ancillary equipment
to the mobile workstation by referring to the
respective instruction manuals.
2
Installation and Connection
2.4 Connection to the AC Mains Power Supply
13
Page 14
2
Installation and Connection
4
Connect the power cord of the mobile workstation
to a wall mains outlet.
WARNING on power connection for medical
use
Please use the following power supply cord.
With connectors (plug or female) and cord types other
than those indicated in this table, use the power supply
cord that is approved for use in your area.
United States and Canada
Plug TypeHOSPITAL GRADE*
Co r d Ty p eMin. Ty p e S J T
Minimum Rating
for Plug and
Appliance
Couplers
Safety ApprovalUL Listed and CSA
*Note: Grounding reliability can only be achieved when the
equipment is connected to an equivalent receptacle marked ‘Hospital
Only’ or ‘Hospital Grade’.
Min. 18 AWG
10A/125V
14
2.4 Connection to the AC Mains Power Supply
Page 15
Chapter 3 Inspection Before Use
WARNING
Before each use, inspect this instrument as instructed
below. Inspect other equipment to be used with this
instrument as instructed in their respective instruction
manuals. Should the slightest irregularity be observed,
do not use this instrument and see “Chapter 7
Troubleshooting” in “INSTRUCTIONS
(ADVANCED).” If the irregularity is still observed,
contact Olympus.
Damage or irregularity may compromise the patient or
user safety, and may result in more severe equipment
damage.
Prepare this instrument and other equipment (shown in
“System Chart” on page 31 in the “Appendix”) to be
used with this instrument for each particular case. Refer
to the respective instruction manuals for each piece of
ancillary equipment.
To disconnect the main power, unplug the AC IN
connector.
3.1Inspection of the
Power Supply
1
Press the power switch to turn ON the color video
printer (OEP-5) (see Figure 3.1).
2
Confirm that the “PRINT” lamp and “ALARM”
lamp lights and then turns off (see Figure 3.1).
3
Press the power switch again.
4
Confirm that the LCD panel turns off.
PRINT lamp
ALARM lamp
LCD panel
3
Inspection Before Use
Power switch
Figure 3.1
If the power fails to come on
If the power fails to come on, inspect according to
“Chapter 7 Troubleshooting” in “INSTRUCTIONS
(ADVANCED).”
If power still fails to turn ON, contact Olympus.
3.1 Inspection of the Power Supply
15
Page 16
3
2
Take out the ink ribbon holder (see Figure 3.3).
3.2Installation of the
Color Printing Pack
(UPC-55)
Install the separately available color printing pack
(UPC-55) on this instrument. Never use printing packs
other than UPC-55. When using this instrument for the
first time, the supplied trial pack can be used for an
operation check.
CAUTION
The color printing pack (UPC-55) is a set of ink ribbon
and print sheets. Be sure to use the ink ribbon and print
sheets provided in the same set. Otherwise, the color
tones in printed pictures may be abnormal.
Always replace the ink ribbon and print sheets
Inspection Before Use
simultaneously, even if either of them is remaining.
Otherwise, the color tones in printed pictures may be
abnormal.
Installation of the Ink Ribbon
Install the ink ribbon on the provided ink ribbon holder.
1
Open the front door by flipping it downward from
the top edge (see Figure 3.2).
Front door
CAUTION
• Do not apply excessive force to the ink ribbon
holder. Otherwise, damage may result.
• Remove the ink ribbon holder squarely from this
instrument. If you take out the ink ribbon holder
at a slant, this instrument and/or the ink ribbon
holder may be damaged.
3
Remove the used ink ribbon from the ink ribbon
holder.
While pushing the take-up side gear (gray ink
ribbon gear) in the direction of the arrow, remove
the ink ribbon from the take-up side gear shaft. Also
remove it from the feed side gear shaft by pushing
the feed side gear (white ink ribbon gear) in the
direction of the arrow (see Figure 3.4).
Ink ribbon holder
Figure 3.3
Gear (Gray)
Figure 3.2
CAUTION
• Be careful not to catch your hand in the front
door. Otherwise, injury may result.
• Do not apply excessive force to the front door.
Otherwise, damage may result.
Gear (White)
Caution label
Figure 3.4
CAUTION
Do not attempt to reuse a used ink ribbon. Also do
not attempt to rewind and reuse an ink ribbon.
Otherwise, the printing will not only be incorrect
but damage to this instrument may also occur.
NOTE
Also refer to the caution label on the ink ribbon
holder.
16
3.2 Installation of the Color Printing Pack (UPC-55)
Page 17
4
Set a new ink ribbon on the ink ribbon holder.
Hold the ink ribbon right and remove the gear cover
(see Figure 3.5).
Gear cover
Figure 3.5
CAUTION
• Do not place an ink ribbon in a location exposed
to direct strong electromagnetic radiation.
Electromagnetic radiation may damage the IC
chip built into the ink ribbon, and the remaining
ink level cannot be correctly displayed.
• Do not touch the surface of the ink ribbon for
longer than is necessary. Do not leave it in a dusty
environment. If smudges or dust attach to the ink
ribbon, printing will not only be incorrect but also
damage to this instrument may occur.
5
Firstly, firmly secure the white gear in the ink
ribbon, and then the gray gear (see Figure 3.6).
Gear (White)
Caution label
m
Gear (Gray)
Caution label
3
Inspection Before Use
Figure 3.6
CAUTION
Install the ink ribbon securely on the ink ribbon
holder, and be sure that the ink ribbon film is
stretched tightly (not loose), otherwise printing will
not only be incorrect but may result in damage to
this instrument.
NOTE
Also refer to the caution label on the ink ribbon
holder.
6
Turn the gears in the arrow direction until the
magenta section of the ink ribbon appears all over
to remove slack from the ink ribbon (see Figure
3.7).
3.2 Installation of the Color Printing Pack (UPC-55)
17
Page 18
Magenta section
Gears
Caution label
Figure 3.7
8
Close the front door of the color video printer
(OEP-5) (see Figure 3.9).
Front door
Figure 3.9
3
NOTE
Also refer to the caution label on the ink ribbon
holder.
Inspection Before Use
7
Push the ink ribbon holder all the way into the color
video printer (OEP-5) (see Figure 3.8).
CAUTION
• Be careful not to catch your hand in the front
door. Otherwise, injury may result.
• Do not apply excessive force to the front door.
Otherwise, damage may result.
If the ink ribbon is torn in the middle
1
Connect the torn sections with transparent plastic
tape (see Figure 3.10).
2
Turn the gray gear on the take-up side in the
direction of the arrow to take up the ink ribbon until
the plastic tape is invisible (see Figure 3.10). Be
sure to remove slack from the ink ribbon.
Figure 3.8
WARNING
Do not insert or put anything except the ink ribbon
holder into the slot. Otherwise, damage, fire and/or
electric shock may result.
Gear (Gray)
CAUTION
Push the ink ribbon holder squarely into this
instrument. If you push the ink ribbon holder in at a
slant, the ink ribbon may be torn, or this instrument
and/or the ink ribbon holder may be damaged.
18
3.2 Installation of the Color Printing Pack (UPC-55)
Plastic tape
Figure 3.10
Installation of the Print Sheets
NOTE
When removing the print sheets from this instrument
(for storage), put the print sheets in its original package
bag or a similar bag according to the instruction manuals
supplied with the color printing pack (UPC-55).
Page 19
1
Press the “PUSH EJECT” area of the paper tray and
take out the paper tray from the color video printer
(OEP-5) (see Figure 3.11).
Figure 3.11
CAUTION
• When attaching or detaching the paper tray, be
careful not to move this instrument together.
Otherwise, this instrument may drop and cause
injury or malfunction.
• Do not apply excessive force to the paper tray.
Otherwise, damage may result.
• Remove the paper tray squarely from this
instrument. If you take out the paper tray at a
slant, it may result in damage to the paper tray or
to this instrument.
• Do not remove the paper tray while printing, as
damage to this instrument may result.
warped print sheets. Otherwise, paper jam may
result.
3
Open the cover of the paper tray.
If print sheets remain in the paper tray, remove
them.
Align the direction of the arrow of the positioning
mark on the protective sheet with that of the arrow
of the positioning mark on the paper tray, and place
the print sheets in the paper tray together with the
protective sheet, so that the print surfaces face
upward (see Figure 3.13).
Positioning mark
Protective sheet
Positioning mark
Figure 3.13
3
Inspection Before Use
2
Flip through the print sheets together with the
protective sheet (see Figure 3.12).
Figure 3.12
CAUTION
• Do not touch the print surface of print sheets with
a hand. Always handle the print sheets without
removing the print surface protective sheet. If
smudges or dust attach to the print surface,
printing will not only be incorrect but also
damage to this instrument may occur.
• If you use print sheets that have stuck to each
other, these stuck print sheets will be double-fed
into this instrument and may cause frequent paper
jamming.
• Do not bend and/or fold print sheets when you
flip print sheets. Also, do not use extremely
CAUTION
• Set print sheets so that they accurately fit the
paper tray. Otherwise, paper jam may result.
• Do not add additional print sheets to the paper
tray while paper sheets remain. If you add
additional print sheets, this may cause paper
jamming.
• Do not reuse the print sheets left in the paper tray.
NOTE
Also refer to the indications on the paper tray and
the protective sheet.
3.2 Installation of the Color Printing Pack (UPC-55)
19
Page 20
4
Remove the protective sheet (see Figure 3.14).
6
Pull out the stopper of the paper tray until it clicks
(see Figure 3.16).
3
Figure 3.14
5
Insert the paper tray into the color video printer
(OEP-5) until it clicks (see Figure 3.15).
Inspection Before Use
Figure 3.15
CAUTION
• When attaching or detaching the paper tray, be
careful not to move this instrument together.
Otherwise, this instrument may drop and cause
injury or malfunction.
• If the paper tray cannot be inserted all the way,
check that the print sheet is remaining in the
paper tray slot. If any is remaining, remove it and
do not attempt to reuse it.
• Do not place your hand in the paper exit.
Otherwise, injury may result from catching in the
internal roller.
• Insert the paper tray into this instrument squarely.
If you insert the paper tray at a slant, it may result
in damage to the paper tray or to this instrument.
Also, abnormal noise or paper jamming may
occur during printing.
• Insert the paper tray into this instrument after
turning the power on. Otherwise, damage to this
instrument may result.
Figure 3.16
Stopper
20
3.2 Installation of the Color Printing Pack (UPC-55)
Page 21
3.3Setup of Connected Equipment
This document, “INSTRUCTIONS (BASIC),” explains
the EVIS EXERA II video system center (CV-180) and
VISERA ELITE video system center (OTV-S190).
Refer to the instruction manuals of the connected
equipment, which must be set up as required for use with
this instrument.
Refer to “INSTRUCTIONS (ADVANCED)” when you
use equipment other than the equipment introduced in
“INSTRUCTIONS (BASIC).”
Ta b l e 3 . 1
Connected Device Instruction Manual
References
CV-180“9.2 System setup”
“Printer”
“9.3 User preset”
“Remote switch and foot
switch (EXERA and
VISERA)”
“9.3 User preset”
“Release function”
OTV-S190“4.4 System setup
(connection setup)”
Setup ItemSetupNote
“Type”“OEP4”Select the printer type. Be sure to
set “Type” to “OEP4” when using
this instrument.
“Qty. “N””“4” – “ 9”Select the print quantity when the
“Caption”Max. 32 characters.Input the characters to be printed
“Release Time”“0.5sec” “1sec”
“1.5sec” “2sec”
“OEP4 mode”“Mode1” “Mode2”Select the output signal format
“Scope switch”
“Switch1”
“Switch2”
“Switch3”
“Switch4”
“Foot switch”
“Switch1”
“Switch2”
“Release setting”
“Release1”
“Release2”
“Printer”
“Type”
“Release1”
“Release2”
“CVP”Select the instrument to be
“Remote”Select the model of the connected
“Print quantity indicator” on the
keyboard is set to “N.” Select any
setting, as required.
in the margin of print sheet.
Set the time while still images
from the endoscope are captured
before the images are read by the
printer.
for the endoscope image. Select
“Mode1” when HDTV signals
are input.
Select the function to be
controlled from the scope switch
and foot switch on the camera
head. Select any setting, as
required.
controlled from “Release1” and
“Release2” that are assigned to
the scope switch and foot switch
on the camera head. Be sure to set
this item to “CVP” when using
this instrument. The RELEASE
key of the keyboard provides the
same function as “Release1.”
printer. Set “Remote” when this
instrument is connected.
3
Inspection Before Use
3.3 Setup of Connected Equipment
21
Page 22
3.4Inspection of the
LCD Panel Display
The LCD panel displays the information and messages
according to the conditions of this instrument.
CAUTION
If an error or warning message is displayed, refer to
“Countermeasures Against Messages” in “7.1
Troubleshooting Guide” in “INSTRUCTIONS
(ADVANCED)”, and take the appropriate measures.
1
Turn the color video printer (OEP-5) ON as
described in “3.1 Inspection of the Power Supply”
on page 15.
Ta b l e 3 . 2
DisplayInput signal format
VIDEOVBS composite video
SVIDEOY/C signals
SDRGB signals (SDTV)
HDRGB signals (HDTV)
SDAUTOAuto input detection (for SDTV input)
HDAUTOAuto input detection (for HDTV input)
SD-SDISDI signal (SDTV)
HD-SDISDI signal (HDTV)
SSDAUTAuto SDI signal input detection (for SDTV
input)
HSDAUTAuto SDI signal input detection (for HDTV
input)
3
2
Confirm that the LCD panel displays the startup
Inspection Before Use
message and that the display then changes as shown
in Figure 3.17.
For details of each item, refer to “INSTRUCTIONS
(ADVANCED).”
Startup message
Caption display
This field shows “C” when the printed picture is
set to contain a comment (caption).
Input signal display
This field shows the
selected input signal
format (see Table 3.2).
Print quantity display
This field shows the
number of prints.
Memory page display
This field shows the
selected memory page.
Print layout (ZOOM MODE
Settings) display
This field shows the setting of
the print layout (see Table 3.3).
Remaining ink ribbon
display
This field shows the
remaining ink ribbon.
Source/Memory display
This field shows “S”
when the source image
is displayed on the
monitor, and shows “M”
when the memory image
is displayed on the
monitor.
Ta b l e 3 . 3
DisplayZOOM MODE Settings
OFFOFF (Print normally.)
PTPART (Only the “4:3” and “5:4” aspect
ratios are available.)
ALLALL (Only the “4:3” and “5:4” aspect
ratios are available.)
RMREMOTE (All aspect ratios are available.
ZOOM MODE activates depending on the
signal being input.)
Operation mode display
This field shows the operation mode of this
instrument.The number on the left is the number of
images used in multi-picture printing. Also, in the case of
a printed image with white frame, “F” is displayed on the
right side of the number of images. The number on the
right indicates the current position of the pointer. It is the
location where the image is stored next.
22
3.4 Inspection of the LCD Panel Display
Figure 3.17
Page 23
Ta b l e 3 . 4
3.5Setup of This
Instrument
The procedures in this section describe basic operation
when used with the EVIS EXERA II video system
center (CV-180) or VISERA ELITE video system center
(OTV-S190).
For other setups of this instrument, set them as desired
by “Chapter 5 System Setup” in “INSTRUCTIONS
(ADVANCED).”
Cursor keys
“v” “V” “b” “B”
MENU button
LCD panelENTER button
ConnectingProcedure
OTV-S1 901 Press “v” or “V” until “SIGNAL
SETUP” appears on the LCD panel,
and press the “ENTER” button.
2 When “INPUT SETUP” appears,
press “v” or “V” to display
“INPUT SEL,” and press the
“ENTER” button.
3 Press “v” or “V” until “SDI”
appears below “INPUT SEL,” and
press the “ENTER” button.
4 Press the “MENU” button. Press
“v” or “V” until “SIGNAL
SETUP” appears on the LCD panel,
and press the “ENTER” button.
5 When “INPUT SETUP” appears,
press “v” or “V” to display “SD/
HD SEL,” and press the “ENTER”
button.
6 Press “v” or “V” until “AUTO”
appears below “SD/HD SEL,” and
press the “ENTER” button.
3
Inspection Before Use
Figure 3.18
1
Press the “MENU” button.
2
Depending on the connected equipment, perform
setup of the color video printer (OEP-5) by
following the procedures below.
Ta b l e 3 . 4
ConnectingProcedure
CV-1801 Press “v” or “V” until the LCD
panel shows “SIGNAL SETUP,”
and press the “ENTER” button.
2 When “INPUT SETUP” appears,
press “v” or “V” to display
“INPUT SEL,” and press the
“ENTER” button.
3 Press “v” or “V” until
“COMPONENT” appears below
“INPUT SEL,” and press the
“ENTER” button.
4 Press the “MENU” button. Press
“v” or “V” until “SIGNAL
SETUP” appears on the LCD panel,
and press the “ENTER” button.
5 When “INPUT SETUP” appears,
press “v” or “V” to display “SD/
HD SEL,” and press the “ENTER”
button.
6 Press “v” or “V” until “AUTO”
appears below “SD/HD SEL,” and
press the “ENTER” button.
3.5 Setup of This Instrument
23
Page 24
3.6Inspection of the Monitor Display
3
The monitor displays information and messages
according to the conditions of this instrument.
3
When the color video printer (OEP-5) is connected
to a video system center, display the image from the
color video printer (OEP-5) on the monitor by
Inspection of the output image
referring to the instruction manuals for the video
system center and monitor.
1
Turn the color video printer (OEP-5) ON as
described in “3.1 Inspection of the Power Supply”
on page 15.
2
Display the output image from the color video
4
Confirm that the monitor first displays the startup
message, and then shows the display as shown in
Figure 3.19. For details of each item, refer to
“INSTRUCTIONS (ADVANCED).”
printer (OEP-5) on the monitor.
Message display 1
This field shows less urgent
messages that may be
displayed during normal use.
Print layout display
Inspection Before Use
This field shows “ZOOM” except
when “ZOOM PART” is set to
“OFF” in “ZOOM MODE.”
Caption display
This field shows “CAPTION”
when the printed picture is set to
contain a comment (caption).
White frame display
It is displayed in the case of a
printed picture with a white
frame.
Message display 2
This field shows more urgent
message.
Remaining ink ribbon display
This field shows the remaining
ink ribbon quantity.
Pointers
These indicate the locations
where the images are stored.
They can be moved using the
cursor keys (“v” “V” “b” “B”).
The green blinking pointer
indicates the location where the
image is stored next. The
pointers are not displayed on the
monitor when the number of
multi-picture images is set to “1.”
Print quantity display
This field shows the number of
prints. It blinks during printing.
Memory page display
This field shows the selected
memory page. The memory
page number blinks during
printing or when it is reserved for
printing.
CAUTION
Operation mode display
This field shows the operation mode of this
instrument. The symbol on the left shows how the
screen is split for displaying the multi-picture images
to be printed. The number on the right shows the
location where the pointer is now means that this is
the image to be printed next.
DisplayNumber of multi-picture images
If messages are displayed in message display 1 or 2,
refer to “Countermeasures Against Messages” in “7.1
Troubleshooting Guide” in “INSTRUCTIONS
(ADVANCED).”
Source/Memory display
This field shows “S” when the
source image is displayed on the
monitor, and shows “M” when
the memory image is displayed
on the monitor.
1
2
3
4
Figure 3.19
24
3.6 Inspection of the Monitor Display
Page 25
Inspection of the CVP counter
When this instrument is connected to the following
equipment, perform inspection of the CVP counter.
– EVIS EXERA II video system center (CV-180, CV-
160)
– VISERA ELITE video system center (OTV-S190)
– VISERA Pro video system center (OTV-S7Pro)
1
Display the output image from the video system
center on the monitor by referring to the instruction
manuals for the video system center and the
monitor.
2
Confirm that the monitor screen displays the “CVP
counter” (see Figure 3.20).
CVP counter
3.7Check the Printed
Picture
Check whether the printed picture is desired.
1
Observe, for example, the palm of your hand using
the endoscope, while referring to the instruction
manual of any connected equipment.
2
Capture a picture, while referring to the instruction
manual of any connected equipment.
Confirm that the number displayed on the “CVP
counter” is increased by one. If the number does not
increase, refer to “Irregularity: The number
displayed on the CVP counter does not increase
even if an image has been captured” in “7.1
Troubleshooting Guide” in “INSTRUCTIONS
(ADVANCED).”
3
Print.
4
Check that the printed picture is of a suitable
quality.
If it is not of the desired picture quality, adjust the
picture quality, referring to “Chapter 5 System
Setup” in “INSTRUCTIONS (ADVANCED).”
After adjustment, print and check again.
3
Inspection Before Use
Figure 3.20
3
If the “CVP counter” is not displayed on the
monitor screen, refer to “Irregularity: The CVP
counter is not displayed” in “7.1 Troubleshooting
Guide” in “INSTRUCTIONS (ADVANCED).”
3.7 Check the Printed Picture
25
Page 26
Chapter 4 Operation
WAR NING
• Do not use this instrument in a location exposed to
direct strong electromagnetic radiation (for example,
microwave treatment device, short wave treatment
device, MRI or radio equipment). Electromagnetic
radiation can interfere with the monitor display.
• It is recommended to use only Olympus highfrequency electrosurgical equipment with this
instrument. Non Olympus equipment can cause
interference on the monitor display, or a loss of
endoscopic image.
NOTE
Prepare the spare color printing pack (UPC-55), in case
the current one is nearly finished.
4.1Turning the Power
ON
Press the power switch (see Figure 4.1).
Power switch
4
Operation
Figure 4.1
CAUTION
If an error or warning message is displayed, take proper
measures, referring to “Countermeasures Against
Messages” in “7.1 Troubleshooting Guide” in
“INSTRUCTIONS (ADVANCED).”
NOTE
For information and messages displayed on the monitor
and the LCD panel, see “3.4 Inspection of the LCD
Panel Display” on page 22 and “3.6 Inspection of the
Monitor Display” on page 24.
26
4.1 Turning the Power ON
Page 27
4.2Control from the
Connected
Equipment
This document, “INSTRUCTIONS (BASIC),” explains
the use of the EVIS EXERA II video system center (CV-
180) and VISERA ELITE video system center (OTVS190) with this instrument.
Refer to “INSTRUCTIONS (ADVANCED)” when you
use equipment other than the equipment introduced in
“INSTRUCTIONS (BASIC)” or when you attempt to
control this instrument from the panel or the remote
control unit (MAJ-898) (optional).
CAUTION
• When this instrument is controlled from the EVIS
EXERA II video system center (CV-180), do not
perform the following tasks from this instrument or
remote control unit (MAJ-898) (optional).
– Setting up the print quantity
– Setting up the number of multi-pictures
– Changing memory page
– Clearing a captured image
– Moving the pointer
Because of the difference between settings of the
EVIS EXERA II video system center (CV-180) and
settings of this instrument, desired operation may not
be carried out, or image data may disappear.
• The number of multi-picture images cannot be set to
“8” or “16.”
• Do not set the number of multi-picture images to “8”
or “16” from EVIS EXERA II video system center
(CV-180). If it is set as “8” or “16,” it will be set as “1.”
• Do not turn OFF or take out the paper tray while
printing is in progress. Otherwise, paper jam and/or
malfunction may result.
• Do not open the front door or take out the paper tray
while printing is in progress. Otherwise, paper jam
and/or malfunction may result.
• Do not leave 10 or more print sheets on the paper exit.
Otherwise, paper jam may result. Also, even when the
number of print sheets is less than 10, message
“REMOVE PRINTS” may be displayed on the
monitor and LCD panel and printing is interrupted
depending on the print sheet conditions. If this occurs,
remove the print sheets from the paper exit. Printing
will resume automatically.
• If you turn this instrument OFF while the PRINT lamp
is lit, the ink ribbon quantity displayed on the
remaining ink ribbon display on the LCD panel and/or
the monitor may not match the actual ink ribbon
quantity remaining.
Be sure to turn this instrument OFF after the printing
is completed and the PRINT lamp turns off.
NOTES
• Refer to the instruction manuals supplied with the
connected equipment.
• If print sheets jam during printing, take the
measures recommended in to “4.3 In the Case of
a Paper Jam” on page 28.
• An image cannot be stored in a memory page that
is currently printing, or is reserved for printing.
Wait until the printing completes.
• If you open the front door during printing,
printing is interrupted. If you close the front door,
the print sheet being printed is ejected. To print
again, press the “PRINT” button.
• If the remaining print sheets in the paper tray
decreases, a rattling sound may be heard during
printing. This is not a malfunction.
• Depending on the environment in which this
instrument is used, or the condition of the print
sheets, printed sheets output from the paper exit
may go over the stopper and drop from the paper
exit. Printed sheets may slide between previously
ejected print sheets, and may not stack up in
printed order.
1
Set up the connected equipment as shown in Table
3.1 in “3.3 Setup of Connected Equipment” on page
21.
2
Set up the color video printer (OEP-5) according to
the connected equipment as described in “3.5 Setup
of This Instrument” on page 23.
3
Control as described in the instruction manuals for
the connected equipment (see Table 4.1).
Ta b l e 4 . 1
connected
equipment
CV-180“5.5 Printing images”
OTV-S190“7.9 Recording the frozen image
The part indicated to the
instructions manual
(releasing)”
4
Operation
4.2 Control from the Connected Equipment
27
Page 28
4
Operation
4.3In the Case of a
Paper Jam
If a paper jam occurs during printing, correct it as
described below.
The flash memory is built into this instrument to store all
image data stored in the memory pages and all current
settings of this instrument, in case a print sheet jams
unexpectedly. When the power of this instrument is
turned OFF, image data stored in the memory pages is
cleared. However, data stored in the flash memory will
not be cleared even if the power is turned OFF.
Removing the Jammed Paper Sheet
When paper jam occurs, message “SAVING IMAGES..”
is displayed together with the error message, and the
images captured in this instrument are stored together
with the menu setups in the flash memory. When these
message have disappeared, remove the jammed paper in
the following steps.
WAR NING
Do not insert your hand or put anything into the ink
ribbon holder and paper exit while the power is turned
on. Otherwise, damage on the instrument, fire and/or
electric shock may result.
CAUTION
Do not turn OFF this instrument during the message is
displayed. If the power is turned OFF, a image data is not
only saved in the flash memory, but damage may result.
2
Push the “PUSH EJECT” area on the paper tray and
take out the paper tray (see Figure 4.3).
Figure 4.3
3
Remove the jammed paper sheet (see Figure 4.4).
Figure 4.4
CAUTION
Do not reuse the removed print sheet. Paper jam
may occur.
4
Confirm that the print sheets are placed correctly on
the paper tray, and correct if otherwise.
5
Install the paper tray.
1
Press the power switch to turn OFF (see Figure 4.2).
Power switch
28
4.3 In the Case of a Paper Jam
Figure 4.2
If jammed paper is not removed in the
above-mentioned procedure
1
Be sure to turn OFF the power, and disconnect the
power cord and other connecting cables.
2
Remove the ink ribbon holder and the paper tray
from the color video printer (OEP-5), and then turn
the color video printer (OEP-5) upside down.
Page 29
3
Remove the single screw by which the back cover
plate is secured, using a coin, etc. (see Figure 4.5).
7
Secure the back cover plate by the screw removed,
using a coin, etc. (see Figure 4.9).
Figure 4.5
4
Pull the back cover plate up and remove it from the
main body (see Figure 4.6).
Figure 4.6
5
Remove the jammed paper slowly (see Figure 4.7).
Confirm that the screw holes of rear of the
main body and the back cover plate align
before securing.
Figure 4.9
8
Reset the paper tray and ink ribbon holder into the
color video printer (OEP-5).
Reconnect the power cord and other connecting
cables as they were.
Confirm that any connecting cables and the power
cord are securely connected.
NOTE
If jammed paper cannot be removed, or an error message
is displayed even after taking the above procedure, stop
using this instrument and contact Olympus.
Restart Printing
Even if a paper jam occurs during printing, the image of
the print is held in the flash memory of this instrument
so it can be reprinted.
4
Operation
Figure 4.7
6
Replace the back cover plate (see Figure 4.8).
Figure 4.8
CAUTION
• The image of a print stored in the flash memory is
overwritten each time a paper jam occurs. If image
data is overwritten, you will not be able to print the
image data stored before the paper jam occurred.
• Use the buttons on the front panel of this instrument
when restarting printing. Restart cannot be executed
by the keyboard of the connected equipment or remote
control unit (MAJ-898).
1
Press the power switch to turn ON.
2
When the message “RESTORE IMAGES? YES/
NO” is displayed on the monitor and the LCD
panel, press the “b” or “B” button to select “YES”
and press the ENTER button.
4.3 In the Case of a Paper Jam
29
Page 30
4
Operation
NOTES
• If “NO” is selected in step 2, refer to “Setting
“STORE IMAGE”” in “5.7 Setup of the
“FUNCTION SETUP” Menu” in
“INSTRUCTIONS (ADVANCED)” to restore
and print the image manually.
• The message in step 2 will only appear when
“OEP-4 MODE” is selected from “Setting
“STORE MODE”” in “5.7 Setup of the
“FUNCTION SETUP” Menu” in
“INSTRUCTIONS (ADVANCED).”
3
The message “READING IMAGES” is displayed
on the monitor and the LCD panel. And then the
display returns to the normal display.
4
When image readout completes, message
“RESTORE FINISHED” is displayed on the
monitor and the LCD panel.
5
Select the memory page where images that you
want to print are stored.
6
Press the PRINT button to start printing.
4.4Exiting from
Examination
CAUTION
When this instrument is controlled from connected
equipment, the captured images will not be printed
automatically until their number reaches the set number
of multi-picture images. Confirm that there are no
unprinted images in the memory pages before turning
this instrument OFF. Once this instrument is turned
OFF, all images captured in the memory pages are
cleared and cannot be printed.
1
Display the memory image on the monitor, and
confirm that there are no images left unprinted in
the memory page (see “3.6 Inspection of the
Monitor Display” on page 24).
If there are any, print them (see “3.7 Check the
Printed Picture” on page 25).
2
After printing is completed, turn the color video
printer (OEP-5) OFF.
NOTE
When you control the printer from the EVIS EXERA II
video system center (CV-180), EVIS EXERA video
system center (CV-160) or VISERA Pro video system
center (OTV-S7Pro), change the memory page or print
images using the keyboard of the connected equipment.
For details, refer to the instruction manuals of the
connected equipment.
In any case other than the above, see “Selecting the
Memory Page” and “Printing” in “4.2 Control From the
Front Panel” in “INSTRUCTIONS (ADVANCED).”
30
4.4 Exiting from Examination
Page 31
Appendix
System Chart
The recommended combinations of equipment that can
be used with this instrument are listed below. New
products released after the introduction of this
instrument may also be compatible for use in
combination with it. For further details, contact
Olympus.
WARNING
If combinations of equipment other than those shown in
the following chart are used, the full responsibility is
assumed by the medical treatment facility.
Appendix
System Chart
31
Page 32
EVIS EXERA II Video System Center (CV-180)
EVIS EXERA Video System Center (CV-160)
VISERA ELITE Video System Center (OTV-S190)
VISERA Pro Video System Center (OTV-S7Pro)
VISERA Video System Center (OTV-S7V)
Mobile Workstation
(WM-NP1, WM-WP1)
Video System (OTV-SI)
Color Video Printer (OEP-5)
Appendix
EUS EXERA Compact Endoscopic
Ultrasound Center (EU-C60)
32
Remote control unit (MAJ-898)
High Definition LCD Monitor (OEV261H,
OEV191H)
LCD Monitor (OEV191)
Medical Control Unit For Endosurgery (UCES-3)
Control Unit For Endosurgery (UCES-2)
System Chart
Endoscopic Ultrasound Center (EU-M30S)
Universal Endoscopic Ultrasound Center
(EU-ME1)
EUS EXERA Endoscopic Ultrasound Center
(EU-M60)
Image Management Hub (IMH-20, IMH-10)
Page 33
Environment
Color video printer (OEP-5)
Operating
environment
Standard
storage
environment
(e.g. within a
hospital)
Transport and
storage
environment
Ambient temperature 10 – 35°C (50 to
Relative humidity30 – 80% (without
Atmospheric pressure 700 – 1060 hPa
Ambient temperature 5 to 40°C (41 to
Relative humidity20 – 90% (without
Atmospheric pressure 700 – 1060 hPa
Ambient temperature –20 to +60°C (–4 to
Relative humidity20 – 90% (without
Atmospheric pressure 700 – 1060 hPa
95°F)
condensation)
(0.7 – 1.1 kgf/cm2)
(10.2 – 15.4 psia)
104°F)
condensation)
140°F)
condensation)
Remote control unit (MAJ-898)
Operating
environment
Standard
storage
environment
(e.g. within a
hospital)
Transport and
storage
environment
Ambient temperature 10 – 35°C (50 to
95°F)
Relative humidity30 – 80% (without
condensation)
Atmospheric pressure 700 – 1060 hPa
(0.7 – 1.1 kgf/cm2)
(10.2 – 15.4 psia)
Ambient temperature 5 to 40°C (41 to
104°F)
Relative humidity20 – 90% (without
condensation)
Atmospheric pressure 700 – 1060 hPa
Ambient temperature –20 to +60°C (–4 to
140°F)
Relative humidity20 – 90% (without
condensation)
Atmospheric pressure 700 – 1060 hPa
NOTE: Except the “trial pack.”
Color printing pack (UPC-55)
Operating
environment
Transport and
storage
environment
Ambient temperature 10 – 35°C (50 to
95°F)
Relative humidity30 – 80% (without
condensation)
Atmospheric pressure 700 – 1060 hPa
Ambient temperature –20 to +30°C (–4 to
86°F)
Relative humidity30 – 80% (without
condensation)
Atmospheric pressure 700 – 1060 hPa
Appendix
Environment
33
Page 34
ItemSpecifications
Specifications
Color video printer (OEP-5)
ItemSpecifications
Power supplyRated voltage100 – 120 V AC
Voltage fluctuation ±10%
Rated frequency50/60 Hz
Frequency
fluctuation
Input current2.8 A
Apparent power280 VA
Dimensions and
weight
Classification
(Electromedical
equipment)
Storage and
printing
Appendix
Storage and
printing
Dimensions280(W) × 125(H) ×
Weight9 kg
Type of protection
against electric
shock
Degree of
protection against
explosion
Printer packUPC-55 only.
Printing methodDye sublimation
Paper size178 × 152 mm (Special
Print area2528H × 1920V dots
Resolution379 dpi
Print speedNormal: Approx. 29
Multi-picture
splitting
Print quantity1 to 10 copies
Image capturing
instruction
±1 Hz
398(D) mm
Class 1
Use prohibited in
flammable
environment.
“A5” size)
(169.4 × 128.6 mm)
sec.
High : Approx. 21 sec.
1, 2, 3 or 4
Image storage in
synchronism with the
capture signal from the
connected equipment.
Image storage in
synchronism with the
CAPTURE key on the
front panel.
Image storage in
synchronism with the
CAPTURE key on the
remote control unit
(MAJ-898).
Memory save
after setup
Input signalHDTV signal (1080/
59.94i):
RGB/YPbPr video
signal.
Y/C video signals
Composite video signal
SDI video signal
Tolerance (75 Ω
termination)
RGB (HDTV) video
signals:
0 to 0.7 Vp-p (Video)
±0.3 Vp-p (Sync).
RGB video signals:
0 to 0.785 Vp-p.
Composite sync:
1.0 to 4.0 Vp-p.
Y/C video signals:
0 to 0.714 Vp-p.
Composite video:
0 to 0.714 Vp-p
SDI (HDTV) video
signal
SMPTE292M standard
1080/59.94i
SDI (SDTV) video
signal
SMPTE259M standard
480/59.94i
Output signalHDTV signal (1080/
59.94i):
RGB/YPbPr video
signal.
Y/C video signals
Composite video signal
SDI video signal
Tolerance (75 Ω
termination)
RGB (HDTV) video
signals:
0 to 0.7 Vp-p (Video)
±0.3 Vp-p (Sync).
RGB video signals:
0 to 0.785 Vp-p.
Composite sync:
0.6 ± 0.2 Vp-p.
Y/C video signals:
0 to 0.714 Vp-p.
Composite video:
0 to 0.714 Vp-p
SDI (HDTV) video
signal
SMPTE292M standard
1080/59.94i
SDI (SDTV) video
signal
SMPTE259M standard
480/59.94i
Setups are held in memory even after the
power switch is set to OFF.
34
Specifications
Page 35
ItemSpecifications
Ye a r o f
production
EMCApplicable
UDI labelThe UDI label is required by some countries’
7101234
standard:
IEC60601-1-2:
2001
IEC60601-1-2:
2007
regulations regarding the identification of a
medical device also known as Unique Device
Identification (UDI).
The following information is being coded in
the 2-dimensional barcode (GS1 Data
Matrix):
- (01) 14-digit GS1 Global Trade Item
Number;
- (11) 6-digit date of manufacture;
- (21) 7-digit serial number.
Last digit of the year of
production.
This instrument
complies with the
standard listed in the
left column.
CISPR 11 of emission:
Group 1, Class B
This instrument
complies with the EMC
standard for medical
electrical equipment;
edition 2 (IEC 60601-12: 2001) and edition 3
(IEC 60601-1-2: 2007).
However, when
connected with an
instrument that
complies with the EMC
standard for medical
electrical equipment;
edition 1 (IEC 60601-12: 1993), the whole
system complies with
edition 1.
Appendix
Specifications
35
Page 36
EMC Information
This model is intended for use in the electromagnetic
environments specified below. The user and the medical
staff should ensure that it is used only in these
environments.
Magnetic emission compliance information and recommended electromagnetic
environments
Emission standardComplianceGuidance
RF emissions
CISPR 11
Radiated emissions
CISPR 11
Main terminal conducted emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Group 1This instrument uses RF energy only for its internal function.
Therefore, its RF emissions are low and are not likely to cause any
interference in nearby electronic equipment.
Class BThis instrument’s RF emissions are low and are not likely to cause any
interference in nearby electronic equipment.
Class AThis instrument’s harmonic emissions are low and are not likely to cause any
problem in the typical commercial power supply connected to this
instrument.
CompliesThis instrument stabilizes its own radio variability and has no affect such as
flicker in lighting apparatus.
Appendix
Electromagnetic immunity compliance information and recommended electromagnetic
environments
Immunity testIEC 60601-1-2
test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
Contact: ±2, ±4, ±6 kV
Air: ±2, ±4, ±8 kV
±2 kV for power supply lines
±1 kV for input/output lines
3 A/mSame as leftIt is recommended to use this instrument by maintaining
(> 95% dip in UT)
T
(60% dip in UT) for
T
(30% dip in UT) for
T
(> 95% dip in UT)
T
Compliance levelGuidance
Same as leftFloors should by be made of wood, concrete, or ceramic
tile that hardly produces static. If floors are covered with
synthetic material that tends to produce static, the relative
humidity should be at least 30%.
Same as leftMains power quality should be that of a typical
commercial (original condition feeding the facilities) or
hospital environment.
Same as leftMains power quality should be that of a typical
commercial or hospital environment.
Same as leftMains power quality should be that of a typical
commercial or hospital environment. If the user of this
instrument required continued operation during power
mains interruptions, it is recommended that this
instrument be powered from an uninterruptible power
supply or a battery.
enough distance from any equipment that operates with
high current.
36
EMC Information
Page 37
Cautions and recommended electromagnetic environment regarding portable and
3.5
3.5
mobile RF communications equipment such as a cellular phones
Immunity testIEC 60601-1-2
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTES
test level
3 Vrms
(150 kHz to 80 MHz)
3 V/m
(80 MHz to 2.5 GHz)
• Where “P” is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and “d” is the recommended separation
Compliance levelGuidance
3 V (V
)Formula for recommended separation distance
1
3 V/m (E
)
1
(V1=E1=3 according to the compliance level)
d=[]
V
1
d=[] 80 MHz – 800 MHz
E
1
7
d=[]800 MHz – 2.5 GHz
E
1
• Electromagnetic interference may occur to this
instrument near a high-frequency electrosurgical
equipment and/or other equipment marked with the
following symbol:
distance in meters (m).
• This instrument complies with the requirements of
IEC 60601-1-2: 2001 and IEC60601-1-2: 2007.
However, under electromagnetic environment that
exceeds its noise level, electromagnetic interference
may occur to this instrument.
Recommended separation distance between portable and mobile RF communications
equipment and this instrument
Appendix
Rated maximum output
power of transmitter
P (W)
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
NOTE
Separation distance according to frequency of transmitter (m)
(calculated as V
150 kHz to 80 MHz
d=1.2
=3 and E1=3)
1
80 MHz to 800 MHz
d=1.2
800 MHz to 2.5 GHz
d=2.3
The guidance may not apply in some situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Portable and mobile RF communications equipment
such as cellular phones should be used no closer to any
part of this instrument, including cables than the
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
EMC Information
37
Page 38
FCC
This equipment has been tested and found to comply
with the limits for a class B digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful
interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio
communications. However, there is no guarantee that
interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio
or television reception, which can be determined by
turning the equipment off and on, the user is encouraged
to try to correct the interference by one or more of the
following measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the equipment and
receiver.
– Connect the equipment into an outlet on a circuit
different from that to which the receiver is connected.
– Consult the dealer or an experienced radio/TV
technician for help.
Appendix
FCC WARNING
Change or modifications not expressly approved by the
party responsible for compliance could void the user’s
authority to operate the equipment.
All interface cables used to connect peripherals must be
shielded in order to comply with the limits for a digital
device pursuant to Subpart B of part 15 of FCC Rules.
Declaration of Conformity
Trade Name : OLYMPUS MEDICAL SYSTEMS
Model : OEP-5
Responsible Party : OLYMPUS AMERICA INC.
Address :Two Corporate Center Drive, Melville,
N.Y. 11747-3157, U.S.A.
Telephone Number : (631)844-5000
This device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference,
and (2) this device must accept any interference
received, including interference that may cause
undesired operation.
38
FCC
Page 39
Index
C
Capture 5
Check of the printed picture 25
Color printing pack 5, 16
Color video printer 4, 8
Connection
AC mains power supply 13
EVIS EXERA II Video System
Center 11
VISERA ELITE Video System
Center 12
CVP counter 25
E
EMC information 36
EVIS EXERA II video system center