Olympus G3 Instruction Manual

Page 1
Instructions Guide for G3 WorkStation Procedure Setup
G3 WorkStation
ENT
Page 2
2 G3 WORKSTATION
G3 WORKSTATION
1 2 4
3 6 7
5
1
Return Electrode Connection
2
TCRF Handpiece Cable Connection
3
TCRF Set Target Temperature/
PlasmaCision, PlasmaBlend Settings
4
Standby/Power Button
5
Mode/ Menu
6
TCRF Set Target
7
PlasmaCision Handpiece Connection
Page 3
PROCEDURE SETUP GUIDE
Using Temperature Controlled Radio Frequency (TCRF)
Initial Setup:
A
Connect the power cord and footswitch on the rear panel of the WorkStation.
B
Turn the machine on at the back of the unit.
C
Press the Stand-By button on the center of the front panel (the “apple” switch).
D
Place the grounding pad on the patient and connect the cord to the front of the
controller.
E
CAREFULLY connect the handpiece (Green or Gray) cable to the front of the controller (Use the green cable for turbinate, tonsil, and dual-channel tongue-base and the gray cable for palate and single-channel tongue base).
F
Set the temperature using arrows on the left side of the controller.
G
Set the Joules using arrows on the right side of the controller.
H
See the charts below for temperature and Joules settings for each procedure.
Rapid Lesion Mode:
A
Press the temperature arrow until 85ºC is reached.
B
Release and press once more to reach the RAPID LESION menu.
C
The temperature is now set for rapid lesion.
D
Maximum Joule output for single channel instruments using rapid lesion is 300.
E
Maximum Joule output for dual channel instruments using rapid lesion is 600.
F
Do not use rapid lesion for palate reduction.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
Page 4
4 G3 WORKSTATION
Begin Treatment:
A
Press and release the BLUE foot pedal OR press the MODE button (located in the center of the controller with a round arrow) ONCE (This will set the system to RF SENSING OFF).
B
Press and release the BLUE foot pedal OR press the Mode button a second time.
(This will set the system to RF SENSING ON, Impedance values are recognized here).
C
Press and release the BLUE foot pedal OR press the Mode button a third time.
(This initiates the energy delivery sequence).
Pause: Press the BLUE foot pedal• Restart: Press BLUE foot pedal once more
Reset and Additional treatment:
A
After each completed lesion, press the BLACK mode foot pedal once to reset the
system to RF SENSING OFF.
B
Press and release the BLUE foot pedal will set the system to RF SENSING ON
(Impedance values are recognized).
C
Press the BLUE foot pedal once more to initiate energy delivery to the additional
site.
D
Audible tone is recognized every 50 Joules.
Settings using TCRF for turbinate:
Temperature: 85ºC• Joules: 350-500
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode• Joules: Single channel instruments: 300 peak• Impedance: 110 - 160 OHMS
Page 5
Settings using TCRF for palate reduction:
Temperature: 85ºC• Joules: 600 to 750 for midline lesion• Joules: 300 to 350 for lateral lesions• Impedance: 110 - 160 OHMS
IMPORTANT NOTE: Rapid Lesion is not recommended for soft palate.
Settings using TCRF for base of tongue:
Temperature: 85ºC• Joules: Single channel instrument: 600 - 750 per site• Joules: Dual channel instrument: 1200 - 1500 per site
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode• Joules: Single channel instruments: 300 peak• Impedance: 110 - 160 OHMS
Settings using TCRF for base of tongue:
Temperature: 85ºC• Joules: 800 - 1000
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode• Joules: Dual channel instrument: 600 peak• Impedance: SETTINGS USING TCRF FOR BASE OF TONGUE: 120 - 170 OHMS
Page 6
6 G3 WORKSTATION
Using Plasmacision
Initial Setup:
A
Connect the power cord and footswitch on the rear panel of the WorkStation.
B
Turn the machine on at the back of the unit.
C
Press the Stand-By button on the center of the front panel (the “apple” switch).
D
Connect the PlasmaKnife cable to the purple “PC” plug on at the far right side of
the G3 WorkStation.
E
The WorkStation will automatically identify the PlasmaKnife and set the appropriate
default PlasmaBlend setting.
F
To change the PlasmaBlend CUT/COAG ratio, press the BLACK mode foot pedal
for 1 second until a beep is heard.
G
Press the YELLOW pedal to increase the cut component in 5% increments, or the
BLUE pedal to increase the COAG component in 5% increments.
H
Press the BLACK mode foot pedal to select the new PlasmaBlend ratio.
I
Alternatively, use the “up/down” PlasmaBlend buttons on the left side of G3
WorkStation to adjust the CUT/COAG to desired settings.
J
Press YELLOW pedal for PlasmaBlend CUT output.
K
Press BLUE pedal for COAG output.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
Page 7
Cleaning the cutting surface:
Keep the PlasmaKnife clean for optimum performance.• Wipe the instrument tip frequently with a wet 4 x 4 gauze.• Periodic dipping in a container of saline may help to keep the suction channel
patent.
If the WorkStation gives a “clean tip” message; clean the instrument tip and reset
the WorkStation by pressing the BLACK mode foot pedal.
EG70353001 G3 WorkStation
EG70353002 Footswitch for G3 WorkStation
EG720100901 Powercord USA
EG70353003 Tonsil PlasmaKnife (4-pack)
EG70353005 Dissector PlasmaKnife (4-pack)
EG70353006 jPlasmaKnife (5-pack)
EG1120411005 TCRF Turbinate Handpiece (5-pack)
EG1010211005 TCRF Palate Handpiece (5-pack)
EG2220221005 TCRF Dual Channel Tongue Base Handpiece (5-pack)
EG1200211005 TCRF Single Channel Tongue Base Handpiece (5-pack)
EG2420132805 TCRF Tonsil Handpiece (5-pack)
EG7201013 RC2 Cable for 1010 and 1200 (20 reuses)
EG7201017 RC20 Cable for 1120, 2220 and 2420 (20 reuses)
EG7201010 Bending Tool
EG7201001 Return Electrode (5-pack)
Ordering Information
Cat. No. Description
Page 8
8 G3 WORKSTATION
Specifications, design and accessories are subject to change without any notice or obligation on the part of the manufacturer.
Postbox 10 49 08 20034 Hamburg, Germany www.olympus-europa.com
E0499049 · 3.000 · 03/10 · e7
Loading...