Olympus Fluoview User Manual

Users Manual
Fluoview
Scanning Laser Biological Microscope
Ver. 1.1
WARNING
A
Caution:
Before using your microscope, the items in this manual Identified by the mark shown to the left should be read carefully until completely understood to ensure safe operation.
Thank you for purchasing an Olympus Fluoview Microscope. Before using your microscope, read this manual thoroughly to make sure you obtain full performance from all of the functions provided by this system. This manual is divided into the following five sections:
For Safe Usage Introduction to Fluoview Operation Hardware Troubleshooting Q & A
OIVMPUS'
August 1996
raj
?3
n
IF=°
-About This Manual
This manual explains potential risks and provide important safety information designed to ensure the safe use bf the FLUOVIEW System. Thoroughly study this manual before operating the system.
1 Safety Precautions
2 Warning Labels
A
CONTENTS
1-1
2-1
2-1 Warning Labels Concerning Laser Safety 2-1
2-1-1 Warning Labels 2-1 2-1-2 Aperture Label 2-3 2-1-3 Protective Housing Label 2-5
3 Precautions for Use of the System 3J^
1.
Safety Precautions
1 Safety Precautions
A
Depending on the laser configuration, this laser product is classified as follows:
Ar laser configuration — CLASS n
OAyilONl
LASER RADIATION DO NOT
S[fPt
IN1D
Sm
ARGON LASER InW Mf\X
CW
450-515nn
(LASS n LASKPRGDUCT
Kr and Ar laser configuration
CLASS in a
ANGE
LASER LIGHT
/M)IDD!fB]TBEB«RRE
KR. AR LASER SniW
MAX CW
450-570na
CLASS la
LASER
PRdXJCT
Precautions for Use
Always pay careful attention to the following points.
Avoid engaging mirror cubes into the light path in such a manner that the laser t>eam is directed outward.
This is extremely dangerous as unprotected eyes may be exposed to the laser beam.
If the cube turret of the reflected light fluorescence module, the DIC slider (optional) or the dummy slider is removed during laser output (during laser scanning), laser beam radiation will occur inside the microscope. This is extremely dangerous. Never remove the cube turret, the DIC slider, or the
dummy slider. I The cube turret, the DIC slider (optional), or the dummy slider is to be removed only by trained L personnel, and only when power has been removed from the laser system.
']
1-1
1.
Safety Precautions
If the objective revolving nosepiece or the objective lens is removed during laser output (during laser scanning), the laser beam will be emitted uncontrollably. This is extremely dangerous. Never remove the objective revolving nosepiece or the objective lens.
TThe nosepiece or the objective lens is to be removed only by trained personnel, and only when [_ power has been removed from the laser system.
]
If the transmitted light lamp housing or the lamp socket is removed during laser output (during laser scanning), the laser beam light will be emitted uncontrollably. This is extremely dangerous. Never remove the reflected light lamp housing or the lamp socket.
The transmitted light lamp housing or the lamp socket is to be removed only by trained personnel,
and only when power has been removed from the laser system and from the
microscope.
• If the scan unit's (2) detection mode selector slider is detached during laser output (during laser scanning), laser beam radiation will occur inside the microscope (beam reflected from the specimen). This is very dangerous. Never remove the scan unit's detection mode selector slider.
The scan unit's (2) detection mode selector slider is to be detached only by trained personnel, |
_and only when power has been removed from the laser system. J
• Using tools to remove the protective housing, or otherwise dismantling/remodeling units, is extremely dangerous as this may lead to uncontrollable laser beam radiation inside the unit Never attempt to dismantle/remodel the system.
• Never excessively bend, strain, or step on the optical fibers. Damaged optical fibers may result in laser beam leakage. This is extremely dangerous. Should the optical fibers be damaged, immediately turn OFF the laser power supply unit and contact your Olympus representative.
• Hot air is emitted from the laser cooling fan exhaust outlet. Keep flammable materials and materials adversely affected by heat at a safe distance of the outlet.
• The power consumption of both the power unit and the laser unit is very large. Ensure that each unit obtains power from separate electrical systems.
• To prevent shock hazards, always ground the equipment.
• Always unplug the power cord before replacing fuses.
• Since the mixture (ether (70%) and alcohol (30%), etc.) used to clean the optical components of
the system is highly flammable, be careful to keep these chemicals away from open fire and potential sources of electrical sparks, such as when the main switches are moved to
"I"
(ON) or
"O"
(OFF).
1-2
1.
Safety Precautions
A
Please note that the laser beam is emitted when the display's function
pane!
([Acquire]
panel) looks as shown below.
"Laser beam emitted" is shown.
w}iFjmt^axv/),^^.\jiu»:p^
Disposal of Laser Tubes
Strict adherence to applicable industrial guidelines and laws is required when disposing of laser tubes. Before disposing of laser
tubes,
contact your Olympus representative for further information.
Safety Symbols
The following symbols are found on the equipment Study the meaning of
the
symbols,
and always use
the equipment
in
the safest possible manner.
Symbol
A
O
Explanation Before use, carefully read the instmction manual. Improper handling could result in injury to the user and/or damage to the equipment Indicates that the main switch is ON.
Indicates that the main switch is OFF.
1-3
2.
Warning Labels
2 Warning Labels
2-1 Warning Labels Concerning Laser Safety
2-1-1 Warning Labels
(1)
Ar
laser configuration
©AUTIOff^
LASER RADIATION
DONdTSIWE INTDB^ ARGON LASER
Inn IMX
cn
45Ch515nn
CLASS n LASBIPRCOUCT
(2)
Kr
and
Ar
laser configuration
ANGE
LASER LIGKT MIDDIIECrBEenHIE
Ka
AR
LASER
SaWMAX CW45O-570niD
CLASS
la
LASe)
PRODUCT
Attached position
BX50 configuration (Similar
in
case
of
BXWI configuration)
(Dor (2)
*^^MiiSr'
r
-
2-1
2.
Warning Labels
IX configuration
(Dor
(2)
2-2
2.
Warning Labels
2-1-2 Aperture Label
(Indicates where the laser beam is emitted.)
(3)
AVOID EXPOSURE
Laser radiation is emitted from this aperture
BX50 Configuration
(3) The laser beam is emitted from the
objective and the sockets on the revolving nosepiece.
(3) The laser beam Is emitted from the
condenser lens.
BXWI Configuration
(3) The laser beam Is emitted from the
objective and the sockets on the revolving nosepiece.
The laser beam is emitted from the condenser lens. 13)
2-3
2.
Warning Labels
IX Configuration
The laser beam is emitted (3) from the objective.
(3) The laser beam is emitted from
the condenser lens.
2-4
2.
Warning Labels
2-1-3 Protective Housing Label
(4)
CAUTION - Laser radiation when open.
DO NOT STARE INTO BEAM
t/-^f3tS®^ajit«L»c
t
(5)
DANGER - Laser light
when
open.
AVaO DIRECT
EYE
EXPOSURE.
CAUTION - Laser
^
radiation when open.
jDO NOT STARE INTO BEAM
BX50,
BXWI configuration
Ar
: (4)
Kr •»
Ar : (5)
Do
not
remove the cube turret.
Do
not
remove the revolving nosepiece, the objective or the DIC slider or the dummy slider.
When
the
transmitted light detector
is not
mounted (left) and when mounted (right)
(30
not
remove
the
transmitted light lamp
housing or the lamp socket.
2-5
2.
Warning Labels
IX configuration
Ar : (4)
Kr + Ar : (5)
Do not remove the cube tuaet.
When the transmitted detection unit is not mounted (left) and when mounted (right)
Do not remove the transmitted light lamp housing or the lamp socket.
2-6
2.
Warning Labels
Common to all configurations
When the slider cover is removed
(6) Do not remove the detection mode
selector slider.
In addition to the locations shown in this manual, a protective housing label is also attached.
For replacement of spiled brp warriihg labblSi^ pibaise cohtact
your Olympus representative. r ^^^'..'
•.:'.'
:.'r ••• '••••:;:'
•. •
vr
2-7
3. Precautions for Use of the System
3 Precautions for Use of the System
Installation
• The FLUOVIEW system will be assembled and setup by qualified technicians. Avoid moving the system as this may adversely affect the adjustment of the optical system. In case it becomes necessary to move the system, please consult your Olympus representative. Olympus shall not be liable for improper adjustments, damages, or other problems occum'ng as a result of the system being moved by the customer or other unauthorized personnel.
• Pulling the optical fiber in case of the KrAr laser configuration will change the laser output Never modify by pulling the optical fiber.
• To prevent overheating, leave at least 30 cm of free space t>etween the exhaust opening of the
laser cooling fan and walls or other objects. Laser tube overheating will result in unstable laser oscillation or the laser oscillation may stop. Excessively high temperature inside the unit may also
result in malfunction and/or damage.
• Openings for a cooling fan for ventilation are provided on the rear panel of the control unit
computer, and the magneto-optical disk unit. To prevent the risk of overiieating and other damage, the openings should never be blocked. Leave at least 10 cm of free space between the openings and walls or other objects.
Avoid installation in places exposed to direct sunlight high humidity, high temperatures, and dust Operating environment: Ambient room temperature 10~35°C; relative humidity 30~80%.
• Avoid installation in places exposed to vibrations. Consult your Olympus representative if the location is subjected to vibrations.
Handling
• The FLUOVIEW system is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact
• Never excessively bend, strain, or subject the laser fiber to crushing force as this will result in reduced performance.
• The piezoelectric stage and the piezoelectric nosepiece (option) employ a piezoelectric element which is fragile. Never subject to impact and ensure that undue force is never applied to these units.
Precautions for Use
• When the built-in power supply for transmitted light is tumed on for a long time, accumulated heat may cause metal to stretch and the focus to drift. To achieve highly precise image data, it is recommended to turn OFF the main switch on the microscope frame.
3-1
3. Precautions for Use of
the
System
• Standard objectives for biological use are employed for both the upright microscope system configuration (BX) and the inverted microscope system configuration (IX). Accordingly, use cover glass with a thickness of 0.17 mm or Petri dishes with a bottom thickness of 0.17 mm. It is recommended to use special cover glass for LSM water immersion objectives.
• When using the FLUOVIEW system, it is recommended to dim the light in the room to reduce the adverse effects of extraneous light.
• To ensure a stable laser beam output allow at least 10 minutes for warm-up after the laser power supply is tumed ON.
• After tuming the reflected light power supply OFF, wait at least 10 minutes before tuming ON again.
Tuming ON sooner will reduce the service life of the mercury burner.
About the Computer
and
Software
• Microsoft Windows are pre-installed on the computer for use with the FLUOVIEW system. Microsoft Windows are pre-installed on the computer for use with the FLUOVIEW system. The customer should make a back-up of the pre-installed program and store the back-up floppy disks carefully. (Olympus offers no support for back-up, missing back-up, etc.) For details on Microsoft Windows, refer to the User's Manual for Microsoft Windows.
• When the empty capacity of the hard disk becomes low, data processing will become very slow. Periodically erase unnecessary data files from the hard disk. For details on erasing data files, refer to the User's Manual for Microsoft Windows.
• When closing the software, always click the <Exit> button at the bottom of the [File I/O] panel. If you display Windows' [Task List] dialog box and click the <End Task> button to close the software, the file conditions, etc. of the last session will not be saved and subsequently not recalled next time the software is started.
• When using the Z-motor, bring the specimen into focus by operating the coarse (fine) adjustment knob on the microscope frame after clearing the check from the [Keep locked] check box on the [Z Stage] panel of the [Acquire] panel (refer to Section 4-2-3-1 in the Operation Manual). Operating the focus adjustment knobs on the microscope frame while the [Keep locked] check box remains checked may result in damage to the Z-motor.
3-2
Concerning This Section
This section provides an overview of the Fluoview System
and should be read before the system is used. After reading and understanding this section, please read the Operation section.
Contents
CAUTIONS 2 REGISTERED TRADEMARKS 2 CARE AND MAINTENANCE 3 MOVING THE FLUOVIEW 3
SECTION 1 INTRODUCTION H
1-1 Fluoview Manual Configuration 1-1
1-1-1 Users Manual Configuration 1-1
1-2 Conventions Used in This Manual 1-2
SECTION 2 SYSTEM OVERVIEW 2J.
2-1 Principle of Operation 2-1 2-2 Fluoview Features 2-2 2-3 Optical Path Diagram 2-3 2-4 System Configuration 2-4
2-4-1 System Units and Their Roles 2-4
2-5 Software Function Configuration 2-7
2-5-1 Software Panel Configuration 2-7 2-5-2 Function Panel and Display Panel 2-8 2-5-3 Drag and Drop Function Execution Icon 2-9
2-6 System Setup 2-10
2-6-1 Power Consumption 2-10 2-6-2 Tuming on the Power Supply 2-11
2-7 System Operational Procedure Overview 2-14
2-7-1 Fluorescent Light Observation Procedure 2-15 2-7-2 Transmitted Light Observation Procedure 2-16
2-8 Identifying Images From Different Methods of Observation 2-17
Cautions
(1) This software and manual may not be reproduced in part or in their entirety without the express written
permission of Olympus.
(2) The contents of this manual are subject to change without notice.
Registered Trademarks
Microsoft, Microsoft Windows and Excel for Windows are registered trademarks of Microsoft Corporation of the U.S. All other company names and product names are the trademarks or the registered trademarks of the respective companies.
Care and Maintenance
(1) When not in use, always use the accessory cover to protect your microscope from dust. (2) This microscope is a precision instrument, so do not disassemble any of the components. (3) Keep dirt, fingerprints off the lenses and filters. Remove any dirt by wiping lightly with soft gauze.
Stubborn dirt can be removed by wetting the gauze with a mixture of alcohol and ether (3:7 ratio), or benzene.
,::Ethei^
is highly <x>mbusUble^^^ cautipri when turiiihg'
* \ the main switch oh arid off.
(4) Clean the various parts by wiping with a soft cloth moistened with diluted detergent. Do not use organic
solvents because these can cause deterioration of the paint and plastic parts.
Moving the Fluoview
Avoid moving this microscope since this can adversely affect adjustment of the optical system.
Please consult with your Olympus sales representative before moving this system.
Olympus is not responsible for any problems resulting from moving this system.
Section 1 Introduction
Section 1 Introduction
1-1 Fluoview Manual Configuration
There are two Fluoview manuals; the Users Manual and the On-screen Manual (on-line help). The Users Manual consists of five sections. The contents of these sections are described below.
1-1-1 Users Manual Configuration
• For Safe Usage This section explains the requests, cautions and wamings related to usage of the Fluoview System.
• Introduction to Fluoview This section provides an overview of the Fluoview System.
• Operation This section explains the method of
operation,
including the input of images and different
types of image processing.
• Hardware This section provides detailed explanations of the Fluoview System hardware functions and specifications.
• Troubleshooting 0 & A This section explains the various countermeasures that can be taken in case a problem should occur.
1-1
Section 1 Introduction
1-2 Conventions Used in This Manual
The following is an explanation of
the
various conventions used in this manual.
<•
<•
Caution,
remari< and
Symbol
*
one-point advisory symbols.
Explanation
Caution items are indicated by (*). Remarks to be observed and one-point advisories are
by (@).
Menu,
command button and dialogue box conventions.
Usage
Explanation
indicated
[Config] panel
<0K> button <Open File> button
The names of panels, dialogue boxes, list boxes, check boxes,
etc., are enclosed in square brackets. buttons are enclosed in triangular parentheses.
•» Mouse Operation
Convention
Explanation
Click This means to press and immediately release the mouse
button.
Double click This means to rapidly press and release the mouse button
twice.
Drag This means to hold down the mouse button, move the pointer to
the desired location and release the mouse button.
Note:
Unless othen/vise
specified,
the temns click, double click and drag are used in this manual in
relation to the left mouse button.
••• Key Operation Convention
Explanation
( Enter )
( Alt ) + fTT"
Direction keys
The names of keys are enclosed in rounded boxes(_ The plus (+) sign indicates combined key operations.
For
example,
( Alt ] + ( Fl ] means to hold down the( Alt
key and press the( Fl ] key.
( - ). CED. ( T
land,
nrikeys.
1-2
^
Section 1 Introduction
> Terms Unique to This System
Convention Explanation
XY obsen/ation This means using XY scan to obsen/e. (Other observation) (The same is for
XZ,
Xt, XVt and XYZt observation.)
1-3
Section 2 System Overview
Section 2 System Overview
The Olympus Fluoview is a confocal scanning type laser fluorescent microscope that utilizes a common focal point optical system to realize high resolution and high contrast as well as a spectacular improvement in resolution in the optical axis. This microscope provides researchers with the ability to perform automatic sectioning, three-dimensional structuring and time fluctuation observation as well as various types of image processing and analysis.
2-1 Principle of Operation
Light detector
The scanning laser microscope focuses the laser beam into a tiny spot through the objective lens and scans specimens in the X and Y directions. The scan unit detects fluorescent and transmitted light from specimens and outputs specimen
images to the image monitor.
As shown in the diagram, the confocal optical system
has a pinhole located at a position (confocal plane) which is optically conjugate with the in-focus position and which eliminates all light from outside the in-focus
position.
Consequently, these lights appears dark in
images so thick systematized specimens can be sliced optically. With ordinary optical microscopes, light from outside the in-focus position is superimposed over the image fomiing rays at the in-focus position, which results in images that are blurred. The laser beam that passes through a specimen is detected by the transmitted light detector and a transmitted light image is formed, but this is not a confocal image.
However, extremely important data can be obtained by synthesizing the transmitted light image and the confocal fluorescent light image.
Light detector
c)^
Laser
Objective lens
Specimen
2-1
Section 2 System Overview
2-2 Fluoview Features
1.
The detector has a resolution of 12 bits, so images are extremely clear,
2.
1 A high-resolution 1024 x 768 pixel system is used. Non-interlaced output signals are used
for clear, flickeriess images.
3. Transmitted light is detected by a photomultiplier detector, so transmitted light images are sharp and clear.
4.
An auto-gain function eliminates any need for bothersome sensitivity adjustment
5. Two image modes, 2-channel fluorescent images or fluorescent image + transmitted light image,
can be selected with one touch.
6. The scan unit is corrected up to the infrared range to assure compatibility with a variety of lasers. Use is also possible with both erect image and inverted image microscopes.
7. The confocal aperture is a 5-position turret, so the correct confocal aperture for each objective lens can be selected with one touch.
2-2
Section 2 System Overview
2-3 Optical Path Diagram
CHI CH2
FL TR
Laser
CH2
CH2
CHI CH2
TR FL
CH2
PM2
• —
2-3
Section 2 System Overview
2-4 System Configuration
2-4-1 System Units and Their Roles
Scan unit This unit scans the laser beam in the X and Y directions and detects the light that retums.
Microscope A BX50 erect image microscope for fluorescent observation.
type
light
Vertical stand This stand supports the scan unit to prevent excessive load being applied to the microscope.
Transmitted light:
detector (option) This unit is used to obtain transmitted light images.
Vibration-free stand Special rubber legs eliminate vibration.
Microscope desk
Hard copy device This device copies images onto 35mm or Polaroid
film.
Control Unit Used to control the scan unit and convert detected signals into images.
Electromagnetic disk A large-capacity storage device for storing images as files.
Image monitor Used to display laser scanned images, operation panel, etc.
Power controller A special table-top controller for power supply connections other than the laser and Illumination bulb power supplies.
Computer Used for LSD control and for storing Images as files.
Computer desk
Laser power supply The power supply for the laser oscillator.
Illumination power supply
_ The power supply for the
mercury vapor illumination.
Photo
2-1
BX50 System
L^ser
unit Consists of a laser oscillator and fiber coupling (argon laser shown to the left).
2-4
Section 2 System Overview
Microscope A BXWI fixed stage erect image microscope for fluorescent light observation.
VIbration-free stand A special pneumatic type stand for eliminating vibration.
^fis^.f •••"
Photo 2-2 BXWI System
Compressor Supplies air to the vibration-free stand.
2-5
Section 2 System Overview
Microscope AN 1X70 Inverted Image microscope
for
fluorescent light observation.
Vibration-free stand A special pneumatic type stand for eliminating vibration.
Photo 2-3
1X70
System
Pt^^ii^
^W
^r^F pjfr-:
T*
fc^^
RK?
^m
mmM:n^
ij^feg
PH . ^-.«"-^'
'•"•••
i. .f^^S:
^^^ •^'•nn
'
'MF^'^Cz^m^D
"
iri'''**?**SPyt1rtit' • :K--<.'*S<
-':H^9BS^BHli ''^'''^''
t-*.—
'|~
"•'Vk'^'"^''*
%•
!)fj|;5l|;
^•^^^'.-'f-J.%'\
..'"' (,..ii^-—.^".
"^M'J^'.^'^
• •
••:V..<>rT"
""" •
iL
^
T^^W^HWll
• 1
^8H
Transmitted light detector (option) This unit
Is
used
to
obtain transmitted light
images.
Compressor Supplies air to the vibration-free stand.
2-6
Section 2 System Overview
2-5 Software Function Configuration
This software uses panel type windows. With conventional software, it is necessary to select a menu and then select the command to be executed.
With panels, however, software functions can easily be executed by merely selecting
the panel index for the function to be executed, like a system notebook or file folder.
2-5-1 Software Panel Configuration
The panel indexes for all of the functions cannot be displayed at one time but must be scrolled.
The panel list is shown below for use as reference when scrolling.
— Acquire
— Scan — Z Stage — Time Series — Config
— File I/O — Tile
— Process
— Math — Filters — Histogram — Logical
— Analyze
— Single
'— Series
— Visualize
— Orientation
'— Other Options
2-7
Section 2 System Overview
2-5-2 Function Panels and Display Panels
The Fluoview software utilizes two types of panels: the function panel and the display panel. The functions panel includes the [Acquire], [File I/O], [Tile], [Process], [Analyze], and [Visualize] panels. The [Live] panel and [(File Name)] panel which are read in from a file are displayed in the display panel.
Function panel ^ Display panel
•k With this software, a panel is simply called the '0000 panel' instead of the '0000 panel of the function panel'. For example, the [File I/O] panel of the function panel is simply called the [File I/O] panel.
2-8
Section 2 System Overview
2-5-3 Drag and Drop Function Execution Icon
This software uses
the
drag
and
drop method
of
selecting image files
and
observation
methods (dye color name
or
transmitted light). With this intuitive method, selection requires
only picking
up
[Item
to
Select (image file
or
obsen/ation method)] with
the
mouse
and
dragging
it
to the setting location, where
it is
dropped.
*^ "•.•:.^-\ •::••••'.•,:'..
.Oi1:FITC!:."=.---:'':-; S-'-.--.
,;CH2:Rhodamin... ;,•.
;v.'
bVutV
-.-
-
tl
lniMit2'
1.
-1 ...,„
.:••• 1 dsar Bafcwl
6
::''\
^':;i^.TVl»,-;is;,j:-:<--y
:'0i
• •
f
TRITC
f
CY3
1
Texat
Re/
.
^^rn'.
Lucifei Yellow
_J,
Da
±1
.xyz-2.Tr
D
;XYCCE(iill2)
&![•
Scalaf Qpi/jrr
SciibrVakw
r''•'AJd
to'im»B»T
I I' Sufahact hop Imagtf] JMulliplyliMgBiiah
I
..|-. DivijB Image hy"
|
<r'.-.
MiitB Image Opt ~
Add 2 Ireaget.-
' |
Sufattact 2 Image!
MuBiiijf 2 Imaget
Experiments
in Men ory
CMMAGEV«YZ-2.TIF..'
Done
nmm^^md^
2-9
Section 2 System Overview
2-6 System Setup
This system must be set up by a special technician. To maintain the perfonnance and for safety, never disassemble or re-adjust this system.
2-6-1 Power Consumption
As shown below, some of the units consume considerable
power;
therefore, check the
capacity of the power outlet before connecting this system.
Microscope
Control unit
Electromagnetic
disk
IHard copy unit
Monitor
3A
Max.
3A
Max.
j 0,3A Max. IA Max,
Power controller
(up to 15A max, can be connected)
Personal computer
I.SAMax,
3A Max.
AC100V12.1AMax.
O
Fluorescent Illumination
power supply
(BH2-RFL-T3)
AC100V2.8AMax,
O
Argon laser power
supply
ACIOOVIOAMax.
O
Krypton/argon laser
power supply
200VAC, single-phase, 20Amax.
<
2-10
Section 2 System Overview
2-6-2 Turning on the Power Supply
Tum on the power supply of each unit.
When only reading data, the laser power supply and illumination power supply are not
necessary and need not be tumed on.
To obtain a stable laser output, tum on the laser power supply and allow the system to warm
up for 10 min. or more before using. Once the illumination power supply (mercury vapor lamp power supply) has been tumed off, do not turn it on again for at least 10 min. to avoid shortening the service life of the mercury
vapor lamp.
• Tuming on the laser power supply Argon laser:
(1) Turn on the power supply switch. (2) Turn the key to the ON position.
Krypton/argon laser:
Please refer to the laser instruction manual.
Please refer to the laser instruction manual.
(1) Tum on the power supply switch.
(D
2-11
Section 2 System Overview
• Tuming on the power controller power supply (1) Tum on the main switch.
For details, refer to the Power Controller
SSI-9501 instruction manual.
(2) Tuming on the switches
If
all
of the switches are tumed on, the power for the entire system can be tumed on and off
with
the main switch as described in item
(1)-
The units are connected as shown below.
Computer
Monitor
Hard copy device
Electromagnetic disk
Microscope
2-12
Section 2 System Overview
Tuming on the control unit power supply
(1) Turn on the power supply switch.
This unit is connected to the power controller, so the power switch can be left turned on.
• Tuming on the electromagnetic disk power supply (Please read the electromagnetic disk instruction manual.) This unit is connected to the power controller, so the power switch can be left turned on.
• Tuming on the microscope power supply (Please read the instruction manual of the microscope used.) This unit is connected to the power controller, so the power switch can be left turned on.
2-13
Section 2 System Overview
2-7 System Operational Procedure Overview
A flow chart is used here to explain the operational procedures so that operation of the entire system can be more easily understood. For reference, the relevant manual sections (f j) and numbers (( )) are indicated at the right side of the flow chart for each procedure. Please refer to 2-7-1 Fluorescent Light Observation Procedure and 2-7-2 Transmitted Light Observation Procedure.
2-14
Section 2 System Overview
2-7-1 Fluorescent Light Observation Procedure
Start the system.
Prepare the system
(tum on the power supply).
Introduction to Fluoview (2-6)
Start the Fty<M/iew software.
Op9fati«>n(
2-1-1)
Look through the eiej^Hnti J^ns and focus on the speclmea
ISelect the li^pf^for
100%
for binocular
tube and
feiCttSidn
the specimen.
Har<iW»e(S-l-1)
Set to thft LSM light
path.
Set to the tSM light
path.
Hardware (3-1-2)
Set ^ scan unit.
[
Operate the det^t^tion mode setting
knob and^et the channel.
Har<IW8re(
3-1-4)
Insert the absorptK«t filter into the light
path!
Har<tware(
3-1-5)
Operate the pinhole turret and
se{.dd a pinhole.
Har(JWB(re(
3-1-3)
Operate th6 ND filter turret and
seled;*
ND filter
(100%.
50%, 20%, 6%).
.H^tJware
(3*1-6)
'••"• • --••• •)*-«••«•»'+ fy^yy^^ifr^i'^-^*
i
Replace with a high transmt^lifQn
rate ND filter (100%, 50%, 20%»6%).
Hardware -,
(3-1-6) . . ,
Read in an image.
Operation (3-1)
Set the ot)$en/ation
conditions.
\i
Scan once
<H:
repeatedly.
If the image is not displayed,
adjust In
accordance with the
;
The
imagels \
dsplayediofe^ ; !software[Uv8] > !
panel.
!
If^ima^jefs. not dis{)layad.
( Adjust the PMt voltage, ^\
V lllll r [ f
tltspl«^ed.
iC
stop scanning.
I *M -It 4i * I
Shut down the system.
Shut
doWCi
the Ruoview software.
,
<;^>eration(
2-1-2)
•*:*•«
•«*(••
Shgl down the system.
(Tum off the power supply.)
2-15
Section 2 System Overview
2-7-2 Transmitted Light Observation Procedure
start the system.
Prepare the system
(tum on the power supply).
Introduction to Fluoview (2-6)
start
the
fluoview softvyrare.
Operatton (2-1-1)
.
Look thraugti
the eyd(jto0^l0ns and
focus on
ttie specimen.
" Select
the
^; path for
100%
for ^
binocular
tube
and
focus
on the specimen.
:
H*dware
•Set to the LSM light
path.
Set to 1t)e LSM light
path.
HardWWe (3-2-3)
I f « Y.|»« I
Set Ihe scan unit.
Remove the microscope caps and
filters from the light
path.
[
Operate the dialection mode setting
knob and $et the channel.
Hardware (3-2-4)
Insert
the
absoi|7lJqc^ filter into the light
path.
Hardware (3-2-5)
Operate the ND filter turret and
sef0<:^
^ ND filter
(100%,
50%.
20%, 6%).
H^dw^re
(3-2-6)
^Replace with a high transmis^i;^ rate^
ND filter (100%, 50%, 20%,^5%).
Hardware '
(3-2-6)
Read in an image.
Operation (3-1)
-w-^
*.j*
.•
<
Set the Ot)idervation
con<itlons.
Scan once dftepeatedly.
If the image is not displayed,
adjust
In accordance with
the relevant
procedure.
%
m
The
image
is \ displayed l(t|he^ software [LjveJ » panel.
*
if;dr)im«S«)9
Adjust the PMT voltage* J
If an iinei^e i$ tSsplayed,'
Stop scann^g.
Shut down the system.
Shut down the Fluoview softv\^are.
Operation (2-1-2)
Shtltdown the system.
(Turn off the power supply.)
2-16
Section 2 System Overview
2-8 Identifying Images From Different Methods of Observation
Fluoview displays various image icons
which
can be used to identify the method of
obsen/ation
used to read in images. (See the table to the right) When the [File I/O], [Tile], (Process], [Analyze] and [Visualize] panels are selected, the icon of the image of the selected [Display] panel is displayed inside the box at the top of the respective panel. Also, when the icon area of
the
[Files] list box of
the
[File I/O] panel or image file is dragged, the icon of the image is used in various aspects. This sen/es in the identification of the
method of observation.
Image icon
0 0
^
^
D
[h
^^
IS
BHI
Icon Meaning
XZ observation XZ obsen/ation, 2-channel
mode
XT obsen/ation XT observation, 2-channel
mode
XZT observation
XZT obsen/ation, 2-channel mode
XY observation XY observation, 2-channel
mode XYt obsen/ation
XYt obsen/ation, 2-channel mode
XYZ observation
XYZ obsen/ation, 2-channel mode
XYZt observation
XYZt
observation,
2-channel
mode
Animated image
Binocular erect image/erect image viewed through colored glasses
2-17
This instmction manual describes functions, specifications, assembly (connections) and adjustment of the hardware. Before reading this instmction manual, please thoroughly review the preceding manual "INTRODUCTION TO FLUOVIEW" for an outline of the system.
LASER SAFETY PRECAUTION
Units indicated as "OPERATOR SERVICE" should be removed only by a person who has had laser safety training,
and only after tuming off the laser unit. This should not be done by any untrained persons.
Removing any of the following units is considered to be OPERATOR SERVICE :
Cube turret
Objective revolving nosepiece
Objective lens
DIC slider or dummy slider
Transmitted light lamphousing and / or lamp socket
Detection mode selector slider ( Scan unit)
See
2-1,
2-7, 2-10.
1 STANDARD CONFIGURATIONS
1 STANDARD CONFIGURATIONS
1-1 BX Upright Microscope System Configuration
bfesi&;SftM!3escnpbonMH?#g^
Scan Unit Excitation Cube Filter
Dual Wavelength Excitation Cube
Banier Filter Barrier Filter Kr and Ar Laser Line Filter
Polarizing Filter for Upright Microscope
Control Unit Pupil Lens Laser Tube Extension Unit Stand for BX
Rubber Feet Anti-Vibration Table for BX
Ar Laser Unit 2 Ar Power Supply Unit 100 Laser Combiner Z-Motor
Unit Desk for Microscope Frame Computer Desk Microscope Computer Monitor 17" Power Controller SCSI Kit Polaroid Recorder
FVX-SU FVX-DM488 FVX-BA565IF
FVX-DM488/568 FVX-BA550RIF
FVX-BA585IF FVX-LLF-KR
FVX-PO-U
FVX-CU FVX-PL-IBX50 FVX-LT FVX-EXTU FVX-ST
FVX-UVT-BX FVX-LU-AR2
FVX-PS-AR100
FVX-COM-KRAR
FVX-ZM FVX-DK8070 FVX-DK6570 BX50-FLA-FVX PC-AT-P95 M0NIT0R17 SSI-9501 SCSI-1510-JPN FVX-POLA
ilBX5oy^ti§
1 1 1
S^^^ii
1
1-1
CONTENTS
1 STANDARD CONFIGURATIONS t£t
1-1 BX Upright Microscope System Configuration 1-1 1-2 IX Inverted Microscope System Configuration 1-2 1-3 Optional Accessories 1-3
2 MAIN UNITS AND DESCRIPTION OF CONTROLS 2£L
2-1 Scan Unit 2-1 2-2 Microscope Frame 2-7 2-3 Transmitted Light Detector (Optional) 2-13
3 PREPARATIONS FOR OBSERVATION Z=l
3-1 Fluorescence Observation 3-1
3-1-1 Bringing the Specimen into Focus 3-1 3-1
-2
Selecting the LSM Light Path 3-3 3-1-3 Selecting the Pinhole 3-5 3-1-4 Selecting the Detection Mode 3-5 3-1-5 Engaging the Barrier Filter 3-5 3-1-6 Selecting the ND Filter 3-6 3-1-7 Selecting the Laser Line Filter (Kr and Ar Laser Combination) 3-6
3-2 Transmitted Observation 3-7
3-2-1 Selecting the Transmitted Light Detector Light Path (Visual Setting) 3-7 3-2-2 Bringing the Specimen into Focus 3-7 3-2-3 Selecting the LSM Light Path 3-8 3-2-4 Selecting the Detection Mode 3-10 3-2-5 Disengaging the Banier Filter 3-11 3-2-6 Selecting the ND Filter 3-11
4 SPECTRAL CHARACTERISTICS OF FILTERS 4=1
5 SPECIFICATIONS 5=1
1 STANDARD CONFIGURATIONS
1-2 IX Inverted Microscope System Configuration
^nimsii^
Scan Unit Excitation Cube Filter
Dual Wavelength Excitation Cube
Banier Filter Banier Filter Kr and Ar Laser Line Filter
Polarizing Filter for Inverted Microscope
Control Unit Pupil Lens Air Anti-Vibration Table for IX Compressor Ar Laser Unit 2 Ar Power Supply Unit 100 Laser Combiner Z-Motor
Unit Computer Desk Microscope Computer Monitor 17" Power Controller SCSI Kit Polaroid Recorder
«Siiii^«»
FVX-SU FVX-DM458 FVX-BA565IF FVX-DM488/568
FVX-BA550RIF FVX-BA585IF
FVX-LLF-KR
FVX-PO-U FVX-CU
FVX-PL-IBX50 FVX-ST FVX-UVT-BX FVX-LU-AR2 FVX-PS-AR100 FVX-COM-KRAR FVX-ZM FVX-DK6570 IX70-FLA-FVX
PC-AT-P95 M0NIT0R17 SSI-9501 SCSI-1510-JPN
FVX-POLA
mmm^.
1 1 1
iHlli^
1
1-2
1 STANDARD CONFIGURATIONS
1-3 Optional Accessories
Green HeNe Laser Unit
Dual Wavelength Excitation
~
Cube Banier Filter
Transmitted Light Detector for BX
Transmitted Light Detector for IX
UCD Lens Unit
LWUCD Lens Unit
Piezoelectric Z-Stage
IX Frame (With Piezo)
Anti-Vibration Platform for IX
Barrier Filter
Banier Filter Barrier Filter Optional Banier Filter Slider
Optional Emission dichromatic min-or Cube
FVX-LU-HEG
FVX-DM488/543
FVX-BA530RIF
FVX-TD-BX
FVX-TD-IX
LSM-THI-UCD
LSM-THI-WUCD
FVX-PZT-BX1
IX70-FVXPZ-F
FVX-UVT-TA FVX-BA590
FVX-610IF FVX-BA510-540 FVX-BA-OPT
FVX-SDM-OPT
Used In combination with the Ar laser. Suitable for TRITC and PI obsen/ation. 543 nm excitation wavelength. Can not be combined witii the Kr laser.
Dual wavelength excitation cube for Ar laser and HeNe green laser.
Short-pass filter for blocking ttie 543 nm wavelength beam emitted by the HeNe green laser.
Used for 488/543 nm dual wavelengUi excitation. Transmitted Ught Detector for BX50/BXWI for
providing transmitted light to the scan unit via an optical fiber connection.
Transmitted Light Detector for IX for providing transmitted light to Uie scan unit via an optical fiber connection.
Lens unit for use with the UCD condenser, mounted on the FVX-TD-IX.
Lens unit for use with the LWUCD condenser, mounted on the FVX-TD-IX.
When used with extemal sensor for fijil-cross conti-ol.
this stage offers Z-movement with extremely high position reproducibility. Minimum step of 0.1 ^m;
stroke length 100 \im.
When used with extemal sensor for fijil-cross control,
this IX frame offers Z-movement with extremely high position reproducibility. Minimum step of
0.1
^m; stroke lengtii 100 ^m.
Simple anti-vibration Platform for IX.
To minimize the fluorescence cross-over when observing double-stained specimens, such as FITC+TRITC, using ttie Ar laser (488 nm).
Band pass filter for use
writh
FITC.
Optional banier filter slider for customized filter placement.
Compatible with a filter diameter of 13
.Q'3
mm and a filter thickness of 2-3 mm. Optional Emission DM cube for mounting a custom
dichromatic mirror to separate the fluorescence emission for detection by the Channel 1 and Channel 2 PMT. Compatible writti 18mm x 11 mm, 1 mm thickness dichroic minors.
1-3
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
Also referto Section 2-4, SYSTEM LAYOUT in the "INTRODUCTION TO FLUOVIEW manual.
2-1 Scan Unit
(5) Filter slider
cover screws
USBt IIIEKITr
0
(4) ND filter tun^t
(6) Laser line filter turret (only for
Kr and Ar laser combination)
"OPERATOR SERVICE" (2) Detection mode _ (D Pinhole tun^t
selector slider
o(E)«
:iECTION WOE
Ofl OC
H. IR H. H. TR R.
OLYMPUS
FLUOVIEW
(D
&
(3) Banier filter slider
(1) Pinhole turret
The pinhole sizes are as follows
1:60nm
The pinhole turret contains 5 pinholes. Allowing selection of an optimum pinhole for all objectives. Select the pinhole number displayed on the control panel ([Acquire] panel). If the acquired image is sufficiently bright and noise is not aparent even when the photomultiplier HV is raised,
an even better image may be obtained by operating the pinhole turret and selecting a smaller pinhole.
2:
lOOum 3: 150nm 4:
200nm
5: 300nm
}
Select either 1, 2 or 3 in accordance with the pinhole size indicated for a given objective (see below).
Use to obtain a brighter image, Confocality is compromised.
2-1
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
In
general,
selecting a smaller pinhole than the pinhole size recommended for a given
objective
will
not increase the axial resolution. However, in some cases the lateral
resolution may
increase.
The effect differs with the refractive index of
the
specimen and
the dispersion of
the
light
The following table shows-the recommended pinhole number for a given objective.
o
\i^
Nl
•^i
, -
N.
N
Objective PLAPO 40X PLAPO 60XO PLAPO 100XO PLAPO 40XWLSM PLAPO 60XWLSM PLAPO 60XOLSM UPLAPO 10X UPLAPO 20X UPLAPO 20XO UPLAPO 40X UPLAPO 40XO UPLAPO 60X UPLAPO 60XWPSF UPLAPO 100XO UPLFL10X UPLFL20X UPLFL40X UPLFL60XO UPLFL100XO UMPLFL10XW UMPLFL20XW LUMPLFL40XW LUMPLFL60XW
NA
0.95
1.40
1.40
0.90
1.00
1.10
0.40
0.70
0.80
0.85
1.00
0.90
1.20
1.35
0.30
0.50
0.75
1.25
1.30
0.30
0.50
0.80
0.90
Pinhole No.
2 2 3 2 3 3
1 1 1
2 2 3 2 3
1
2 3 2 3
1
2 2 3
2-2
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
(2) Detection mode selector slider
Slider for selecting either two fluorescence signals or one fluorescence signal + transmitted light 3 settings. This slider is placed at the A-section shown in Fig, (I).
• Slide'r at the pushed-in position Select this mode when observing a transmitted light and / or a fluorescence with emission wavelength shorter than 570nm, such as FITC, GFP, DiO, etc. As shown in Fig. (I) on the following page, a double-sided mirror is engaged into the light path. Consequently, fluorescence light is reflected to the CHI (PMI). When the transmitted light detector (option) is
attached,
the transmitted light is reflected via the optical filter
connection to the CH2 (PM2).
• Slider at the middle position Select this mode when observing a double-stained specimen, such as FITC+PI, etc. As shown in Fig. (II) on the following page, a beam splitting dichroic mirror is engaged into the light
path.
The characteristics of this dichroic minor make it reflect light with wavelengths below 570 nm and pass the longer wavelengths. Consequently, the fluorescence light can be split into light for the 1CH (PMI) and the 2CH (PM2). With the slider at this position, the transmitted light from the optical fiber connection is blocked.
Slider at the pulled-out position Select this mode when observing a transmitted light image and / or fluorescence specimen with emission wavelengths longer than 570 nm, such as PI, TRITC, etc. As shown in Fig. (Ill) on the following page, nothing is engaged into the light
path. Consequently, fluorescence light is directed to the CH2 (PM2). When the transmitted light detector (option) is attached, the transmitted light is directed via the optical fiber connection to the
CH1
(PMI).
2-3
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
Laser
Slider at pushed-in position
CHI CH2
FL TR
Fig. (I)
Middle position
Fig.(n)
Slider at pulled-out position
Fig. (IE)
2-4
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
1
(3) Bamer filter slider
Up to two filters can be placed for both CHI and CH2. At the pushed-in position, the filter is engaged into the light
path.
At the pulled-out position, it is disengaged.
* If erroneously placed at the middle position, no image will appear. Always stop at the
correct position.
(4) ND filter turret
There are five positions. Transmission ratio: 0, 6, 20, 50, and 100%.
(5) Filter slider cover screws
Operated when the bamer filter slider or the detection mode selector slider is replaced. When this screw is loosened, the covers can be pulled off. Then replace the sliders, retum the cover to its original position, and tighten the screw again.
r Detection mode selector slider should be removed only, by a person who has had laser L safety training, and only after tuming off the laser unit
(6) KrAr laser line filter tun-et
(Only attachable with the Kr and Ar laser combination)
Unit
writh
built-in 5-filter turret for selection in accordance with the 488 nm and 568 nm
excitation wavelength of the Kr
and
Ar laser.
(1)488:
The specimen is excited at 488 nm. Used for observation of single-stained
specimens, such as FITC.
(2) 568: The specimen is excited at 568 nm. Used for observation of single-stained
specimens, such as TRITC, PI.
(3) 568: The specimen is excited at 488 nm and 568 nm. However, the 488 nm output
488
°"'y'®
attenuated to 6% of the normal excitation light. Used for preventing the Pl
AT6 from eclipsing the FITC when observing double-stained specimens such as
FITC+PI.
(4) 568: The specimen is excited at 488 nm and 568 nm. However, the 488 nm output
488 only is attenuated to 25% of the normal excitation light Used for preventing the PI
AT25 from eclipsing the FITC when observing double-stained specimens such as
FITC+PI.
(5) 568: The specimen is excited at 488 nm and 568 nm. Used for observation of double-
+
^og Stained specimens, such as FITC+PI.
2-5
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
• FITC emission tails longer than 570 nm. Accordingly, FITC fluorescence may be detected on the CH2 where it overlaps the Pl fluorescence. (See the figure below.) The problem may be remedied by cutting the FITC by adjusting the OFFSET on the operation panel ([Acquire] panel). If not, It will be necessary to balance the excitation light by reducing the intensity of the 488 nm excitation light used for exciting the FfFC.
When observing a double-labeled specimen, such as FITC+PI, attempt balancing the
intensities ofthe emissions by engaging the (3), (4), (5) filters in the described order.
CH,
A
°°N
AT25 A A \
(FITC overlapping the j \ 1 I 1 CH2 can be reduced by / Vl 1 \ attenuating the excitation 1 \ \
Y
\\
CH2
/~
/
^ /
/\ /
PI
1
400 500 600
wavelength / nm
700
2-6
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
2-2 Microscope Frame
The illustration below shows the main controls of the microscope frame. The stage, revolving
nosepiece, etc., may differ from those illustrated. For details on operation of the microscope frame, refer to the instruction manual for tiie microscope frame.
BX50 Confiouration
(1) Light path selector
"OPERATOR SERVICE"
(2) Cube turret
"OPERATOR SERVICE"
nosepiece
"OPERATOR SERVICE"
objective lens
(6) Universal
(5) Filters
LBD ND6
ND25
(3) Analyzer
U-AN
(optional)
"OPERATOR SERVICE" (4) DIC prism
U-DICT(dummy slider) (optional)
"OPERATOR SERVICE"
• transmitted light lamphousing
• lamp socket
2-7
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
(1) Light path selector
• When the knob is pushed-in, visual observation is possible.
• When the knob is pulled-out laser microscopy is possible.
(2) Cube turret
• Engage the designated cube for visual fluorescence observation.
• When used as a laser microscope and for visual transrnitted light observation, operate the turret to place the index at the [ { Q )] position.
(Set the cube turret so tiiat no cube is engaged.)
(3) Analyzer U-AN
(Optional)
• Engage the analyzer into the light path for visual transmitted Nomarski obsen/ation and transmitted polarized light observation. The analyzer is engaged at the pushed-in position.
• Always remove the analyzer from the light path when used as a laser microscope. The analyzer is disengaged at the pulled-out position. If the analyzer is engaged, a good image cannot be achieved.
(4) DIC prism
U-DICT
(Optional)
• Engaged into the light path for laser Nomarski observation and visual transmitted Nomarski observation. An improved image is achieved if the
U-DICT
is removed from the light path in case of laser fluorescence observation. For fluorescence observation alone, disengage the U­DICT from the light
path.
[Leaving the
U-DICT
engaged during laser fluorescence observation will
degrade the image quality somewhat]
(1) Filter
• Always disengage filters from the light path for laser transmitted observation. A good image will not be obtainable if a fllter is left engaged.
2-8
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
(2) Universal condenser
For Nomarski observation, engage the Nomarski prism (optional) suitable for the objective in use.
(Similar for both visual and laser Nomarski observation.) Note that tiie polarizer should also be engaged in the case of laser Nomarski obsen/ation. (Note that the analyzer U-AN should be removed from Uie light path in tiie case of laser Nomarski observation.)
[Note that laserphase contrast observation js hot poisfsil)le due to construction
'
"i''
',
limKatioris^"
.-!.••'^^'.'•;.?"''.,•.••'!•;,
\, '''•'"••'••'.';•
;j;i..''v--^''':'v^^;''v.-J'-''".'
"•"
2-9
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
IX Confiquration
(6) Filter
"OPERATOR SERVICE"
objective lens
(1) Light path selector
"OPERATOR SERVICE"
(2) Cube tunet
(7) Magnification selector knob
(3)
Analyzer
IX-AN
"OPERATOR SERVICE"
• transmitted light lamphousing
• lamp socket
"OPERATOR SERVICE" (4) DIC prism
U-DICT(dummy
slider)
(optional)
2-10
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
(1) Light path selector
• When tiie knob is set at the ^[ position, visual observation is possible.
• Set the knob at the ( SP ] position for laser microscope use.
(2) Cube turret
• Engage the designated cube for visual fluorescence observation.
• When used as a laser microscope and for visual transmitted light observation, operate tiie turret to place the index at the [ [ O )] position.
(Set the cube tun'et so that no cube is engaged.)
(3) Analyzer IX-AN
(Optional)
• Engage the analyzer into the light path for visual transmitted Nomarski observation and transmitted polarized light observation. The analyzer is engaged at the pushed-in position.
• Always remove the analyzer from the light path when used as a laser microscope. The analyzer is disengaged at the pulled-out position. If the analyzer is engaged, a good image cannot be achieved.
(4) DIC prism
U-DICT
(Optional)
• Engaged into the light path for laser Nomarski observation and visual transmitted Nomarski observation.
• A better image is achieved if the
U-DICT
is removed from the light path in case of laser
fluorescence observation. For fluorescence observation alone, disengage the
U-DICT
from the light
path.
[Leaving the
U-DICT
engaged during laser fluorescence observation will
degrade the image quality somewhat.]
(1) Filter
• Always disengage filters from the light path for laser transmitted observation. A good image will not be obtainable if a filter is left engaged.
2-11
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
(2) Condenser
For Nomarski obsen/ation, engage the Nomarski prism (optional) conesponding to the objective in use. (This applies to both visual and laser Nomarski obsen/ation.) Note that the polarizer should also be engaged for Nomarski obsen/ation. (Note that the analyzer IX-AN should be disengaged for laser Nomarski observation.)
(3) Magnification selector knob
For LSM use, always select IX (knob pushed-in).
The 1.5X setting cannot be used.
[Note that laser phase coritrast obseivatloh is'.nbt possible due to cohstructibn
•^irhitations.]'-.-••.''-•;;:''•'••':•;•:'' • •:
.:";;""';\'\-'"-^-''•:••"*>•'>;
•.•• '•'•'• '"'• '',_ •'••'.:
2-12
2 MAIN UNITS AND DESCRIPTION OF CONTROLS
2-3 Transmitted Light Detector (Optional)
FVX-TD-BX (for BX50, BXWI)
(1) Light paUi selector
• When the knob is pushed-in, laser transmitted light observation is possible.
• When the knob is pulled-out visual transmitted light observation is possible.
(D Light path selector
FVX-TD-IX (for IX)
• When the knob is pushed-in, laser transmitted light observation is possible.
• When the knob is pulled-out, visual transmitted light observation is possible.
(1) Light path selector
2-13
3 PREPARATIONS FOR OBSERVATION
3 PREPARATIONS FOR OBSERVATION
This section explains the order of procedures for specimen obsen/ation.
3-1 Fluorescence Observation
3-1-1 Bringing the Specimen into Focus
3-1-1-1
BX50, BXWI Configuration
1.
Push in the light path selector knob (1) on tiie
binocular observation tube to select Uie 100% at
binocular eyepieces setting.
2.
Operate the cube turret to engage the cube
corresponding to the specimen fluorochrome.
3. While looking through tiie eyepieces, bring the specimen into focus. Make sure to adjust the eyepiece diopter correctly. (Refer to tiie BX
instmction manual.)
stage clips
* When using the Z-motor, clear the check from
the [Engage motor] check box (see Section 4­2-2-1 of the OPERATION MANUAL) on the [Z Stage] panel on the [Acquire] panel. Then operate the microscope frame's focusing adjustment knob (fine) to bring the specimen into focus. Damage to the Z-motor may occur if the microscope frame's focusing adjustment knob is operated while the [Engage motor] check box Is checked.
* The specimen may float during oil immersion
observation. While refening to tiie figure on the left, attach ttie stage clips provided with the microscope frame.
3-1
3 PREPARATIONS FOR OBSERVATION
3-1-1-2
IX Configuration
1.
Turn the light paUi selector dial (1) on ttie right
side of Uie microscope to the ^ position. While looking through the eyepieces, bring the specimen into focus. Make sure to adjust the diopter adjustment ring of Uie eyepiece. (Refer to the 1X50/70 instiuction manual.)
2.
Operate the cube tun-et to engage the cube
corresponding to the specimen fluorochrome.
3. While looking through the eyepieces, bring the specimen into focus. Make sure to adjust Uie eyepiece diopter correctiy. (Refer to the IX instmction manual.)
* When using the Z-motor, clear the check from
the [Engage motor] check box (see Section 4­2-2-1 ofthe OPERATION MANUAL) on the \Z Stage] panel on the [Acquire] panel. Then operate the microscope frame's focusing adjustment knob (fine) to bring the specimen into focus. Damage to the Z-motor may occur if the microscope frame's focusing adjustment knob is operated while the [Engage motor] check box is checked.
* The specimen may float during oil immersion
observation. While refemng to the figure on ttie left, attach Uie stage clips provided witii the microscope frame.
3-2
3 PREPARATIONS FOR OBSERVATION
3-1-2 Selecting the LSM Light Path
3-1-2-1 BX50, BXWI Configuration
1.
Pull out the light path selector (1) on the trinocular obsen/ation tube
to
the stop position.
2.
Operate the cube turret (2) on the vertical illuminator to align
the
index with
the [( Q ) ]
position.
3.
If
the analyzer U-AN
(3) is
mounted, disengage
it
by pulling
it
out to the pulled-out
clickstop. (Leaving
the
U-DICT
engaged during laser fluorescence observation will
degrade the image quality somewhat)
(D Light path selector
"OPERATOR SERVICE"
(2) Cube tun-et
"OPERATOR SERVICE"
nosepiece
"OPERATOR SERVICE"
objective lens
(6) Universal condenser
(5) Filters
LBD ND6 ND25
(3) Analyzer
U-AN (Optional)
"OPERATOR SERVICE"
(4) DIC prism
U-DICT(dummy slider) (optional)
"OPERATOR SERVICE"
• transmitted light lamphousing
• lamp socket
3-3
3 PREPARATIONS FOR OBSERVATION
3-1-2-2
IX Configuration
1.
2.
3.
4.
Tum the light path selector (1) to the [[ SP J ] position. Set the magnification selector knob (7) to lx. Rotate the cube turret to select [ ( Q ]]. If the analyzer IX-AN (3) is mounted, disengage it by pulling it out to the far clickstop. (Leaving the
U-DICT
engaged during laser fluorescence observation will degrade tiie
image quality somewhat.)
(6) Filter
OPERATOR SERVICE"
transmitted light lamphousing lamp socket
(5) Condenser
"OPERATOR SERVICE"
objective lens
"OPERATOR SERVICE" (4) DIC prism
U-DICT(dummy slider) (optional)
(1) Light path selector
"OPERATOR SERVICE"
(2) Cube tun-et
(7) Magnification selector knob
(3) Analyzer
IX-AN
(optional)
3-4
3 PREPARATIONS FOR OBSERVATION
3-1-3 Selecting the Pinhole
Operate the pinhole tun-et (1) to select ttie pinhole corresponding to tiie pinhole number indicated for
each
objective on tiie control panel.
(Refer
to
Section
2-1,
Scan Unit)
3-1-4 Selecting the Detection Mode
Set ttie detection mode selector slider to the designated position in accordance
with
Uie
fluochrome ofthe specimen to be
observed.
If
in
doubt about
Uie
setting,
refer to Section 4­1-2, Adjusting the Scan Unit, in the Operation Manual and follow the prompts of the [Microscope Configuration] window.
Example for reference
Fluochrome
FITC
Lucifer Yellow etc.
FITC + TRITC FITC + PI etc.
PI
TRITC
Detection Channel
CHI
CH1/CH2
CH2
Detection Mode Selector Slider
Position
Pushed in
Middle position
* Pulled out
(Referto Section
2-1,
Scan Unit)
* In case of single-stained specimens, a brighter
image can be observed if ttie slider is set at the pushed-in or out positions. Avoid the niiddle position.
3-1-5 Engaging the Barrier Filter
Depending on the
fluochrome
of
the
specimen to
be
observed,
engage barrier
Alters
as required. The banier
filter
slider is engaged into the light
path
at ttie pushed-in position, and
disengaged at
Uie
pulled-out position.
* No image will appear if the slider is stopped between the positions. Always set at
correct position.
If in doubt about the
setting,
refer to Section
4-1-2,
Adjusting the Scan Unit, in the Operation Manual and follow the prompts of the [Microscope Configuration] window.
Example for reference
Fluorochrome
FITC
FITC + PI
TRITC
CHI BA510IF BA510IF
-
(BA500-540)
CH2
BA565IF
BA565IF
(BA590)
3-5
3 PREPARATIONS FOR OBSERVATION
(5) FIKer slider cover
screws
(2) [}etection mode
selector slider
UBBt IMTEI6ITY
0
(4) ND filter turret
(7) Laser line filter turret
(only
for
Kr and
Ar
laser
combination)
(1) Pinhole turret
~B|
9(S)9
OLYMPUS
FLUOVIEW
0
Jt
(3) Barrier filter slider
3-1-6 Selecting
the
ND Filter
Operate tiie
ND
filter
turret (4) to select the appropriate ND
filter.
Select the ND
filter
in
accordance with the specimen brightness and level
of
photobleaching.
The
ND20%
filter
may be appropriate to start
with.
3-1-7 Selecting
the
Laser Line Filter
(Kr
and
Ar
Laser Combination)
Operate the laser
filter
turret (7) to select the laser line
filter.
Examples for reference
Fluochrome
FITC
Lucifer Yellow etc.
FITC + TRITC FITC + PI etc.
PI
TRITC etc.
Laser line filter
488
568 ^^ ^^
4'***
AT25 AT6
568
(Refer
to
"(6) Laser line
filter
tun-et" in Section
2-1,
Scan Unit)
3-6
3 PREPARATIONS FOR OBSERVATION
3-2 Transmitted Observation
The procedures for
transmitted
observation are similar to those for
fluorescence
observation. Also refer
to Section
3-1,
Fluorescence obsen/ation.
3-2-1 Selecting the Transmitted Light Detector Light Path (Visual Setting)
3-2-1-1
BX50, BXWI Configuration
Pull out the light
path
selector (1).
(D Light path selector
3-2-1-2
IX Configuration
Pull out the light path selector (1).
(D Light path selector
3-2-2 Bringing the Specimen into Focus
(Refer to Section
3-1-1,
Bringing Uie Specimen into Focus.) When using Nomarski DIC attachment, refer to the instruction manuals for the respective microscope frames.
3-7
3 PREPARATIONS FOR OBSERVATION
3-2-3 Selecting the LSM Light Path
3-2-3-1
BX50, BXWI Configuration
1.
Pull out
Uie
light path selector (1) on Uie trinocular observation tube to the pulled-out
clickstop.
2.
Operate the cube turret (2) on the vertical illuminator to align the index with Uie
[Col]
3. If
the
analyzer U-AN (3) is
mounted,
disengage it by pulling it out to the pulled-out
clickstop. (It makes no difference if
the
DIC prism
U-DICT
(4) remains engaged.)
4.
Disengage all the
filters
(5) from Uie light
path.
5. Select the LSM light path on the transmitted light detector. (Push in the light path selector (1) on the
ttansmitted
light detector.)
(1) Light path selector
"OPERATOR SERVICE"
(2) Cube tunet
"OPERATOR SERVICE"
nosepiece
"OPERATOR SERVICE"
objective lens
(6) Universal condenser
(5)Filters
LBD ND6 ND25
(3) Analyzer
U-Mi (Optional)
"OPERATOR SERVICE" (4) DIC prism
U-DICT(dummy
slider)
(optional)
"OPERATOR SERVICE"
• transmitted light lamphousing
• lamp socket
3-8
3 PREPARATIONS FOR OBSERVATION
3-2-3-2
IX Configuration
1.
Tum the light path selector (1) to select the [[ sP W setting.
2.
Set the magnification selector knob (7) to IX.
3. Operate the cube turret4o select the
[ (
Q J position.
4.
If the analyzer U-AN (3) is mounted, disengage it by pulling it out to the far clickstop. (It
makes no difference if the DIC prism
U-DICT
(4) remains engaged into the light paUi.)
5. Remove all the filters from the light
path.
6. Select the LSM light path for the transmitted light detector. (The Light path selector of the transmitted light detector should be set at the pushed-in position.)
(6) Filter
(5) Condenser
"OPERATOR SERVICE"
objective lens
"OPERATOR SERVICE"
• transmitted light lamphousing
• lamp socket
(1) Light path selector
"OPERATOR SERVICE" (4) DIC prism
U-DICT(dummy sikler) (optional)
"OPERATOR SERVICE"
(2) Cube tun-et
(7) Magnification selector knob
(3) Analyzer
IX-AN (Optional)
3-9
3 PREPARATIONS FOR OBSERVATION
3-2-4 Selecting the Detection Mode
When the detection mode selector slider (2) is:
• Pushed in, transmitted light is detected by CH2.
• Pulled out, transmitted light is 'detected by CHI. Consequently, when only transmitted observation is performed, the detection mode selector slider (2) may be set at either position. When performing fluorescence observation simultaneously with transmitted observation, set the detection mode selector slider at the position specified for the specimen fluochrome. If in doubt about the setting, refer to Section 4-1-2, Adjusting the Scan Unit in the Operation Manual and follow the prompts of the [Microscope Configuration] window. (Refer to Section
2-1,
Scan Unit in this manual.)
Examples for reference
Observation modes
FITC + TR(Transmitted)
TRITC + TRfTransmitted)
CHI
FITC
TR(Transmitted)
*
CH2
TRfTransmitted)
TRITC
Detection mode
selector slider
Pushed in Pulled out
[
The middle position of the detection mode selector slider is used for observation double-stained specimens like FITC+PI.
ion of ^
(5) Filter sliders cover screws
"OPERATOR SERVICE" (2) Detection mode (D Pinhole turret
selector slider
USm INIEKSITY
0
GOIFOCAL.
BMRIBtR
OLYMPUS
FLUOVIEW
CD
Jf
(4) ND fllter turret
(6) Laser line filter turret
(3) Banier filter slider
3-10
3 PREPARATIONS FOR OBSERVATION
3-2-5 Disengaging the Barrier Filter
Remove all the barrier
filters
from the channel to detect transmitted light. (The banier filter
slider
(3)
should
be
at ttie pulled-out position.)
If
in
doubt about tiie setting, refer to Section
4-1-2,
Adjusting the Scan Unit in the Operation
Manual and follow ttie prompts of ttie [Microscope Configuration] window.
3-2-6 Selecting the ND Filter
Operate the ND
filter
turret (4) to engage a suitable ND
filter.
Select tiie ND
filter
in accordance witii the specimen brightness
and
level of photobleaching.
The
ND20%
filter
may be appropriate to start
with.
3-11
4 SPECTRAL CHARACTERISTICS OF FILTERS
4 SPECTRAL CHARACTERISTICS OF FILTERS
Barrier filters
BA510IF
BA585IF
Representative filter combinations for Ar laser use
BA565IF (
DM488-
460
Ar laser Excitation wavelength 488 nm
700
EDM570 is a dichroic mirror for splitting fluorescence light into CH1/CH2. DM488 is dichroic mirror for excitation.
4-1
4 SPECTRAL CHARACTERISTICS OF FILTERS
• Representative filter combination for the Kr and Ar laser combination
DM488/568 \ , BA510IF
100
50 -
450
Excitation wavelength 488nm
EDM570 is a dichroic min-or for splitting fluorescence light into CH1/CH2. DM488/568 is dichroic mirror for double excitation for use with Uie KrAr
laser.
700
wavelength 568nm
4-2
5 SPECIFICATIONS
5 SPECIFICATIONS
-:^^M^:r^M^^M^'^:^ir^'$^
Laser unit
Scan unit
Control unit
Ar laser
Kr laser
Scan mode Laser input Shutter ND filter Scan area
Scan center position
Scan mode Detection mode
setting Banier filter
Confocal pinhole Light detector
Applicable wavelengths
Dimensions (mm) Weight Image input
Real-time processing Image memory
Input memory Image display Image processing Image analysis image output
Z-movement
Dimensions (mm) Weight Power requirement
'^^i}:k;^:/i:l-^r!-f''!'^^U%^
5 mW output, random or linear polarization. Air cooled Argon ion laser (488 nm) Scan unit connection: Single mode optical fiber (3 m) Power requirements: 100V 10A (MAX)
15 mW output, linear polarization. Air cooled Kr ion laser (568 nm) Scan unit connection: Single mode optical fiber (3 m) Power requirements: 200V single phase, 20A or more
Polarized light produced by dual scanning galvanometer min-ors. Via fiber connection Blocks the laser when not scanning 5-aperture switchable tun-et type: 100, 50, 20,6, 0% transmission 16X12 rectangular area inscribed in the F.N. 20 boundary circle.
ZOOM:
1 to 10.
ZOOM 1.5 to 10: The center position may be moved within the X-Y direction within the ZOOM 1 scanning area.
XY, XZ, XT, XYZ, XYT, XYZ
1.
volume projection
3-step
switchable (CH1/CH2: FL/TR, FtTFL, TR/FL)
2 filters per channel. Sharp-cut filter (filters for laser use) Band pass filter (user-replaceable)
5-pinhole tun-et (60,100,150, 200, 300 nm) Photomultiplier 350 - 650 nm
390(VV)x 280(D) X 197(H) 11kg 2-channel input
OFFSET: 0-100% GAIN:
1-1
ox
/VD conversion. 4096 gradations (12 bit)
Kalman filter processing. Peak detection integration.
1024 pixel X 768 pixel X 4096 gradations LUT alteration, tiling, zooming, 3D display. Filter processing, image processing. Line profile, intensity map, histogram, length, area measurement Gradations: R, G, B, 256 gradations each
Output signal: R, G, B Sync non-interiace. standard:
Z-motor (resolution 0.1 nm)
Optional:
Piezoelectric stage (for BX), piezoelectric nosepiece (for
IX).
stroke: 100 nm, resolution: 0.1 nm
124 (W)x 342 (D)x 382(H)
9kg
100/120/220/240V AC
1.6/1.6/0.8/O.SA
50/60Hz
5-1
5 SPECIFICATIONS
5 SPECIFICATIONS
;:«§p^p5if^6^^ipjKfi|ps^?i^^
Laser unit
Scan unit
Control unit
Ar laser
Kr laser
Scan mode Laser input Shutter ND filter Scan area
Scan center position
Scan mode Detection mode
setting Barrier filter
Confocal pinhole Light detector
Applicable wavelengths
Image input
Real-time processing Image memory
Input memory Image display Image processing Image analysis Image output
Z-movement
Dimensions (mm) Power requirement
•li^^^li^fiiPi^a«iii«
5 mW output, random or linear polarizaition. Air cooled Argon ion laser (488 nm) Scan unit connection: Single mode optical fiber (3 m) Power requirements: 100V 10A (M/VX)
15 mW output, linear polarization. Air cooled Kr ion laser (568 nm) Scan unit connection: Single mode optical fiber (3 m) Power requirements: 200V single phase, 20A or more
Polarized light produced by dual scanning galvanometer min-ors. Via fiber connection Blocks the laser when not scanning 5-aperture switchable turret type: 100, 50, 20, 6, 0% transmission 16X12 rectangular area inscribed in the F.N. 20 boundary circle.
ZOOM:
1 to 10.
ZOOM 1.5 to 10: The center position may be moved within the X-Y direction within the ZOOM 1 scanning area.
XY, XZ, XT, XYZ, XYT, XYZl, volume projection 3-step
switchable (CH1/CH2: FL/TR, FUFL, TR/FL)
2 filters per channel. Sharp-cut filter (filters for laser use) Band pass filter (user-replaceable)
5-pinhole tun-et (60, 100,150, 200, 300 nm) Photomultiplier 350 - 650 nm
2-channel input OFFSET: 0-100% GAIN:
1-1
OX
A/D conversion. 4096 gradations (12 bit) Kalman filter processing. Peak detection integration.
1024 pixel X 768 pixel X 4096 gradations LUT alteration, tiling, zooming, 3D display. Filter prcxessing, image processing. Line profile, intensity map, histogram, length, area measurement Gradations: R, G, B, 256 gradations each
Output signal: R, G, B Sync non-interiace. Standard:
Z-motor (resolution 0.1 nm)
Optional:
Piezoelectric stage (for BX), piezoelectric nosepiece (for
IX).
Sti-oke: 100 nm, resolution: 0.1 nm 124 (VV)x 342 (D)x 382(H) 100V3A
5-1
5 SPECIFICATIONS
Microscope fi-ame [Upright] BX50-FLA-FVX
BXWI-FLA-FVX [Inclined] IX70-FLA-FVX Stand
for
BX
Computer system
Power controller
Z-motor
Computer desk Stand
for
BX
Anti-vibration table forBX
Anti-vibration table for IX, BXWI
Options
Microscope ft-ame Trinocular
observation tube light path selection
Objective Microscope ft-ame Objective
Computer
Monitor Magneto-optical disk
No.
of
outlets
Capacity
Dimensions (mm) Dimensions (mm)
Anti-vibration system
Dimensions
(mm)
Anti-vibration system
Dimensions (mm)
Piezoelectric
Z-stage
Transmitted light detector
Objective micrometer
Green HeNe laser unit
Rubber feet
anti-
vibration table
for IX
?^-ri^^.h;M;^i^^i*^«iispedfi(a^
BX50,
BXWI
2 settings (100%
for
binocular eyepieces; 100%
for
LSM)
UIS series IX70-FVX-F, IX70-FVXPZ-F UIS series Scan unit,
for
support
Specially designed
for BX
microscopes
IBM PC-AT compatible, Windows 3.11 (English)
17"
230 MB 6 / JP
(not
usable
for
the laser power supply unit, reflected light
module power supply)
100V 15A (MAX)
Microscope fine adjustment knob axis. One full tum: 100
nm
Minimum resolution: 0,1
nm*
650 (W) X 700 (D) x 670
(H)
800 (W) x 700 (D) X 670
(H)
Rubber feet
790 (W) X 667 (D) x 73
(H)
Air
1000
(W)x
700
(D)x
750(H)
Cross-control
by
extemal sensor
Stroke: 100
nm
Minimum resolution: 0.1
nm
Position reproducibility:
± 0.2 nm
2 light path settings; transmitted detection/transmitted illumination Connected
to
the scan unit via optical fiber connection.
1
mm
100 degrees,
for
reflected light
Random polarization
1.5 mW
(543
nm)
Direct connection,
ND
filter tunet provided
1000 (W) X 680 (D) x 73 (H) (mm)
5-2
5 SPECIFICATIONS
Operating environment
Indoor use. Altitude: Max. 2000m Ambient temperature: 5°C to 40t
(41*
F to 109* F)
10°C to35t (Performance guarantees)
Maximum relative humidity 80% for temperatures up to 31°C
(88'
F) decreasing lineariy tiirough 70% at 34t (93' F), 60% at
3rt(99*
F). to 50% relative humidity at 40°C (109" F)
Main supply voltage fluctuations not to exceed ±10% of the nominal voltage.
Pollution Degree 2 (in accordance writti lEC 664) Installation / Over voltage Category n (in accordance
virith
lEC
664)
5-3
II mil
inie^QQinij Q ^'
Concerning
This section explains countermeasures to be taken in the event that a problem occurs.
Please read this section and try the countermeasures before contacting Olympus. If the problem persists, then contact Olympus.
Contents
Section 1 Problem Countermeasures 1-1
Section 1 Problem Countermeasures
Section 1 Problem Countermeasures
Due to Uie usage, the functions of the system sometimes are not fully effective, although there has been no failure; tiierefore, when a problem occurs, refer to the following chart and apply the appropriate countermeasures. If the same phenomenon continues to occur after the countermeasures have been taken, contact your local Olympus dealer.
fea^AJi^,SyniptOmff;;^ia^
^:j;^.^.CountermeasureF^^'<"»
;Manual;Referencel
1.
No fluorescent light
image.
The laser is not oscillating.
Make sure ttie laser unit power supply is tumed on and ttie emission key is on.
Inttoduction to Fluoview, 2-6-2.
The light path selector lever is not set to the LSM light path.
Set the light path selector lever to the LSM light
path.
Hardware,
2-2,
3-1-2.
The visual observation dichroic mirror and analyzer are on the microscope side.
Remove the visual observation dichroic mirror and analyzer from the light path.
The detector photomultiplier voltage is low.
Use the PMT scale of the [Acquire] panel to increase the photomultiplier voltage.
Operation,
4-1-3-10.
The excitation light is weak.
Change to a high transmission ratio ND fllter.
Hardware, 2-1,
3-1-6.
The fluorescent light dye
color and excitation
wavelength are not matched
(krypton/argon laser combination).
Select a laser line filter that matches the fluorescent light dye color. Refer to Operation, 4-1-2 Scan Unit Setting. Follow the instructions in the [Microscope Conflguration] window.
Operation,
4-1-2.
Hardware, 3-1-7.
The fluorescent light dye color and absorption filter
are not matched.
Select an absorption filter that matches the fluorescent light dye color. Refer to Operation, 4-1-2 Scan Unit Setting, r Follow the insti^ctions in the [Microscope Configuration] window.
Operation,
4-1-2.
Hardware, 3-1-5
and 4.
The confocal aperture is too small.
Increase the aperture.
Hardware, 2-1.
The offset value is too large.
Lower the offset to an appropriate value.
Operation,
4-1-3-10.
1-1
Section 1 Problem Countermeasures
^^K^^^^S
1.
No fluorescent light
image.
2.
No transmitted light
image.
3. Images are disrupted.
4.
Visibility is
poor.
5. Bluning and uneven clarity.
6. Flare is visible.
:^Kl^aus^i^Si^
The photomultiplier of the channel set with the detection mode setting knob is not set.
The transmitted light detector is not correctiy set.
The
U-ansmitted
light
filter
at ttie microscope base is in the light
patti.
The absorption filter is in the light
path.
Too much vibration at the setting location.
Fluorescent or other extemal light is being detected.
The analyzer is in the light path.
Scanning speed is fast
The fluorescent light dye color and absorption filter are not matched.
The excitation light is too weak.
Too much vibration at Uie setting location.
The specimen is mounted inclined.
Non-fluorescent glass is not being used.
g^^^'JCountemrieasureliJ^sg
Check the [Chi] and [Ch2] check boxes of the [Acquire]
panel.
Con-ectly set ttie transmitted
light path selector knob of Uie transmitted light detection unit.
Correctly set the detection
mode setting slider of the scan unit
Remove the fllter from the
light
patti.
Remove the absorption filter
in the ti-ansmitted light detection channel of the scan unit from the light
path. Consult your Olympus
dealer.
Darken the room and read the image again.
Remove ttie analyzer from the light
path.
Lower the scanning speed to an appropriate speed or
integrate the images. Select an absorption filter
that matches the fluorescent
light dye
color.
Refer to Operation, 4-1-2 Scan Unit Setting. Follow the
instmctions in the [Microscope Configuration] window.
Change to a high transmission ratio ND
filter.
Consult your Olympus dealer.
Set the specimen con-ectty.
Use non-fluorescent glass.
?Manual!Reference^ Hardware, 2-1.
Operation,
4-1-3-3.
Hardware, 3-2-3.
Hardware, 2-1, 3-2-4
Hardware, 2-2, 3-2-3.
Hardware, 3-2-5.
Hardware, 3-1-2, 3-2-3.
Operation,
4-1-3-11,
4-1-5.
Operation,
4-1-2. Hardware, 3-1-5 and 4.
Hardware, 2-1, 3-1-6.
Refer to the microscope instmction manual.
1-2
Section 1 Problem Countermeasures
^^^«^
6. Flare is visible.
7. Circular flares appear in the center of the image.
8. Blun-ing is visible.
9. Images are dari< with excessive noise.
10.
Poor reproduction of Z
motor positions.
11.
The Fluoview software
will not start.
12.
The [Acquire] panel is
not displayed.
•^m^amm
The cover glass ttiickness is
not suitable.
The specimen is
over-
stained. The excitation wavelength
and absorption fllter type are not matched.
An absorption filter has not been inserted.
The focus is not correctly adjusted.
The con-ect pinhole has not
been selected.
The dye color is too weak.
The pinhole is too small.
The photomultiplier HV exceeds-800.
Scanning speed is too fast
The excitation light is too weak.
The Z position slips because the tension of ttie microscope coarse movement handle is too light
The Fluoview software has already been started.
Anotiier application software is operating.
The control unit is not recognized.
ftmmQounleTmeaswe"MMi Use a cover glass witti a
thickness of 0.17mm. Use an appropriate dye color
or increase the offset value. Select an absorption fllter
that matches the excitation wavelength.
Refer to Operation, 4-1-2. Follow ttie insU^ctions in tiie [Microscope Configuration] window.
Re-focus using visual observation.
Select the correct pinhole.
Set an appropriate dye
color.
Adjust to an appropriate position.
Use an HV of less than -800. If
dari<,
adjust the
gain.
Lower the scanning speed to an appropriate speed or integrate the images.
Change to a high transmission ratio ND
filter.
Increase ttie tension by tuming the adjustment ring and adjusting until the coarse movement handle tension increases suitably.
Press the keyboard (Alt) and fTab) keys to switch to Fluoview.
Shut down the other application software and start the Fluoview software.
Check Uie connection between the computer and contt-ol unit Tum on tiie control unit power supply. Turn on the AUX2 switch of the power controller.
;iManuai,iReference^'
— •
Operation,
4-1-3-10.
Operation,
4-1-2. Hardware, 3-1-5 and 4.
Refer to ttie microscope instiojction manual.
Hardware, 2-1.
Hardware, 2-1.
Operation,
4-1-3-10.
Operation,
4-1-3-11.
Hardware, 2-1, 3-1-6.
Refer to the microscope instmction manual.
"
Operation,
2-1-1.
Inti-oduction to Fluoview, 2-6-2.
1-3
Section 1 Problem Countermeasures
teg5l3i?^tSymptomMs4;!Si;#Si
13.
The scale of tiie [Z
Stage] in the [Acquire] panel is not woridng.
14.
The fine movement
handle of the microscope will not move or moves with difficulty.
15.
Images cannot be saved
to disk.
16.
Images are not output to
the printer.
17.
The mouse pointer does
not move when the mouse is moved.
The Z stage (Z revolver) motor is not being excited.
The Z stage (Z revolver) motor is being excited.
The disk has not been formatted (in the case of a floppy disk or photoelectric disk).
The photomagnetic disk is not recognized.
Insufficient empty space on
Uie disk.
The printer is not being recognized.
The software is misoperating.
^!/i^seountemiMisdre-^<|?s^i Check the [Engage motor]
check box of ttie [Z Stage] panel.
Clear the [Engage motor] check box of ttie [Z motor] panel in the [Acquire] panel.
Format the disk.
Make sure the photoelectric disk power supply is on.
Check ttie cable connections.
Increase the available space by deleting unnecessary data or use a new disk (formatting is necessary).
Make sure the printer power supply is on.
Check the cable connections.
Press the keyboard +(ctri)
( Alt 1 -^ (Dellete ] keys and observe the messages that appear. Shut down Fluoview. Shut down Windows and restart the computer.
IManuaHReferenbe^ Operation,
4-1-3-8.
Operation,
4-1-3-8.
Operation,
Appendix
A.
Refer to Uie photoelectric disk instruction manual.
Operation,
Appendix
A.
Refer to the printer instruction manual.
Operation,
2-1.
1-4
Loading...