Oldham OLCT 10N User Manual

Page 1
OLCT 10N
User Manual
Digital Gas
Detector
The Fixed Gas Detection Experts
Page 2
2
OLCT 10N
User Manual
Copyright 2019 by OLDHAM S.A.S March 2019 All rights reserved. No reproduction of all or part of this document, in any
form, is permitted without the written consent of OLDHAM S.A.S. All information provided in this document is accurate to the best of our
knowledge. As a result of continuous research and development, the specifications of
this product may be changed without prior notice. OLDHAM S.A.S
Rue Orfila Z.I. Est – CS 20417 F – 62027 ARRAS Cedex Tel: +33 (0)3 21 60 80 80 Fax: +33 (0)3 21 60 80 00
Page 3
3
Contents
Chapter 1 | Overview .............................................................. 7
Purpose ...................................................................................................... 7
Composition of the Detector Assembly ...................................................... 7
Type of detector available .......................................................................... 8
Chapter 2 | Installation ........................................................... 9
Regulations and conditions of use ............................................................. 9
Location of the detector.............................................................................. 9
Detector positioning ................................................................................... 9
Connection ............................................................................................... 10
Configuration of communication .............................................................. 12
Scope of use ............................................................................................ 14
Chapter 3 | Maintenance ...................................................... 15
Maintenance Frequency........................................................................... 15
Semi-automatic calibration ....................................................................... 16
Sensor replacement ................................................................................. 18
Manual calibration .................................................................................... 20
Chapter 5 | Spare Parts ........................................................ 25
Chapter 6 | UE Declaration ................................................... 27
Chapter 7 | Specifications .................................................... 29
Detector .................................................................................................... 29
Sensor specifications ............................................................................... 30
Line length (use with MX 32) ................................................................... 32
Line length (use with MX 43) ................................................................... 35
Chapter 8 | Specific instructions for use in explosive atmospheres in accordance with European ATEX Directive 94/9/EC 39
General comments ................................................................................... 39
Marking for LEL versions ......................................................................... 39
Marking for toxic gas or oxygen detector versions (except CO2 versions) 40
Page 4
4
OLCT 10N
User Manual
Thank you for choosing this OLDHAM instrument. All necessary actions have been taken to ensure your complete satisfaction with
this equipment. It is important that you read this entire manual carefully and thoroughly.
Limitation of Liability
OLDHAM shall not be held responsible for any damage to the equipment or for
any physical injury or death resulting in whole or in part from the inappropriate use or installation of the equipment, non-compliance with any and all instructions, warnings, standards and/or regulations in force.
No business, person or legal entity may assume responsibility on behalf of
OLDHAM, even though they may be involved in the sale of OLDHAM products.
OLDHAM shall not be responsible for any direct or indirect damage, or any
direct or indirect consequence, resulting from the sale and use of any of its products UNLESS SUCH PRODUCTS HAVE BEEN SELECTED BY OLDHAM ACCORDING TO THE APPLICATION.
Ownership clauses
The drawings, specifications, and information herein contain confidential
information that is the property of OLDHAM.
This information shall not, either in whole or in part, by physical, electronic, or
any other means whatsoever, be reproduced, copied, divulged, translated, or used as the basis for the manufacture or sale of OLDHAM equipment, or for any other reason without the prior written consent of OLDHAM.
Page 5
5
Warnings
This is not a contractual document. OLDHAM reserves the right to alter the
technical features of its equipment at any time and for any reason without prior notice.
READ THESE INSTRUCTIONS CAREFULLY BEFORE USING FOR THE
FIRST TIME: these instructions should be read by all persons who have or will have responsibility for the use, maintenance, or repair of the instrument.
This instrument shall only be deemed to be in conformance with the published
performance if used, maintained, and repaired in accordance with the instructions of OLDHAM, by OLDHAM personnel, or by personnel authorized by OLDHAM.
Warranty
Under normal conditions of use and on return to the factory, parts and
workmanship are guaranteed for 2 years, excluding consumables such as sensors, filters, etc.
Destruction of equipment
European Union only. This symbol indicates that, in conformity with directive DEEE (2002/96/CE) and in accordance with local regulations, this product must not be discarded with household waste.
It must be disposed of in a collection area that is designated for this purpose, for example at a site that is officially designated for recycling of electrical and electronic equipment (EEE) or a point of exchange for authorized products in the event of the acquisition of a new product of the same type.
Page 6
6
OLCT 10N
User Manual
Page 7
1- Overview
7
Chapter 1 | Overview
Purpose
The OLCT 10N is a digital detector designed to detect a particular gas, depending on the type of sensor used. This range of digital detectors is compatible only with the Oldham MX 43 or MX32v2 controller.
Composition of the Detector Assembly
An OLCT 10N detector contains the following parts:
Rep.
Description
1.
Enclosure
2.
Cable gland
3.
PCB
4.
Sensor
5.
Terminal
6.
Gas input
7.
LED
8.
Cover
Page 8
8
OLCT 10N
User Manual
Figure 1: Component parts of an OLCT 10N
Type of detector available
Gas
Range
Methane (CH4)
0-100% LEL (5% vol.) 0-100% LEL (4.4% vol.)
Hydrogen (H2)
0-100% LEL (4% vol.)
Butane (C4H10)
0-100% LEL (1,5% vol.)
Propane (C3H8)
0-100% LEL (2% vol.)
Oxygen (O2)
0-30% vol.
Carbone monoxide (CO)
0-300 ppm 0-1000 ppm
Carbon dioxide (CO2)
0-5000ppm 0-5%vol. 0-100% vol.
Hydrogen sulfide (H2S)
0-30 ppm 0-100 ppm
Nitrogen monoxide (NO)
0-100 ppm 0-300 ppm
Nitrogen dioxide (NO2)
0-10 ppm 0-30 ppm
Ammonia (NH3)
0-100 ppm 0-1000 ppm
Page 9
2- Installation
9
Chapter 2 | Installation
Please read the guides on the installation, use and maintenance of flammable gas and oxygen detectors (standard EN/IEC 60079-29-2) and toxic gas detectors (standard EN 45544-4).
Regulations and conditions of use
Where applicable, the installation should meet all the regulations currently in
force for installations in explosive atmospheres, in particular standards IEC/EN 60079-14 and IEC/EN 60079-17 (current editions).
Generally speaking, the ambient temperatures, supply voltages and power
outputs that are mentioned in this document relate to explosion safety. They
do not relate to the detector’s operating temperatures.
The equipment is authorized for use in ATEX zones 2 and 22 at ambient
temperatures between -20°C and +55°C.
The sensor in the transmitter should always be in contact with ambient air.
Therefore:
- Do not cover the detector
- Do not paint the detector
- Avoid dust
The OLCT 10N may be used only with the OLDHAM MX 43 or MX 32v2
measurement and alarm controller.
CO2 versions are not certified to use in ATEX area
Location of the detector
Depending on the density of the gas to be detected or the application, the detector shall be positioned at ground level, or on the ceiling at the same height as the airflow, or near air extraction ducts. Heavy gases may be detected at the ground level, while light gases will be found at ceiling height.
Detector positioning
The detectors shall preferably be positioned in an accessible place to facilitate inspection and maintenance as well as ensure the safety of the operators. The detectors shall be protected against shock hazards and must not be obstructed by
Page 10
10
OLCT 10N
User Manual
anything that will prevent them from measuring the ambient environment for which they are intended.
The detector shall be installed with the cable gland pointing downwards. If installed at an angle of more than 45° from the vertical or in a horizontal position (ceiling-type installation), the detector will not measure correctly and will need to be re-calibrated.
Figure 2: OLCT 10N dimensions (in millimeters)
Connection
As a reminder, the OLCT 10N can only be connected to the OLDHAM MX 43 or MX 32v2 controller.
RS485 transmission
The OLCT 10N has 2 cable glands that are used for the input cable and the output cable to the next module.
The detectors must be connected using MPI-22A cable type, minimum core size is
0.22 mm² (RS485 cable with 2 shielded twisted pairs, nominal impedance of 100 Ohm). This cable carries the RS485 signal on one pair and the power supply for the modules on the other pair. The +24VDC, 0V, A and B terminals are respectively connected to the +24VDC, 0V, A and B terminals of the other
Page 11
2- Installation
11
modules in the line and then to the corresponding line connector of the MX 43 or MX 32v2 controller.
The cable shielding must be connected to a ground terminal. The 120 end-of-line resistor must be placed at the end of the line. This simply
involves switch #8 being moved to the ON position on the last module (cf. Fig. 5).
Before connecting the detector to the control unit, please disable any installation alarms to avoid unexpected triggering during operation.
*
Connection is made according to the following schematic:
Figure 3: Schematic for connecting modules on a MX 43 or MX 32v2 line
No portion of the bare end of the terminal wires should be visible. For protection against any electromagnetic disturbances, the data as well as screen wires (or braids) must be cut as short as possible.
The connection of the cable between the detector and controller should be made with the power off. The site should be at equal potential.
Controller line terminal
Ground terminal
Detector terminal
Page 12
12
OLCT 10N
User Manual
Configuration of communication
Module address
All the digital modules on a line must be identified by a unique address.
Switches 1 to 5 of the configuration block of each module make it possible to establish an address number (1 to 32) in binary mode.
In the illustration to the right, the address 9 (10010) has been defined.
The Addressing Table below lists the possible combinations.
Figure 4: Switches for address configuration
Address
module
Switches
(ON = 1 ; OFF = 0)
Address
module
Switches
(ON = 1 ; OFF = 0)
1 2 3 4 5 1 2 3 4 5
1 1 0 0 0 0
17 1 0 0 0 1 2 0 1 0 0 0
18 0 1 0 0
1
3 1 1 0 0 0
19 1 1 0 0
1
4 0 0 1 0 0
20 0 0 1 0 1 5 1 0 1 0 0
21 1 0 1 0
1
6 0 1 1 0 0
22 0 1 1 0
1
7 1 1 1 0 0
23 1 1 1 0 1 8 0 0 0 1 0
24 0 0 0 1
1
9 1 0 0 1 0
25 1 0 0 1
1
10 0 1 0 1 0
26 0 1 0 1 1 11 1 1 0 1 0
27 1 1 0 1
1
12 0 0 1 1 0
28 0 0 1 1
1
13 1 0 1 1 0
29 1 0 1 1 1 14 0 1 1 1 0
30 0 1 1 1
1
15 1 1 1 1 0
31 1 1 1 1
1
16 0 0 0 0 1
32 0 0 0 0
0
Table 1: Addressing table (address depends on switch positions)
Page 13
2- Installation
13
Remarks:
During module replacement, all the configuration switches of a new module must be positioned in the same configuration as those of the previous module.
Switch #6 (FRAME FILLING) must be set to OFF (unused option).
Switch #7 (DELAY) must be set to ON.
End of line resistance
For the last detector, set switch #8 (EOL RESISTOR) to ON.
Figure 5: End of line resistor switch in position “ON”
Page 14
14
OLCT 10N
User Manual
Scope of use
The OLCT 10N may be used only with the OLDHAM MX 43 or MX 32v2 measurement and alarm controller.
The sensors have certain limitations; it is essential to fully recognize these limitations.
Presence of specific components
Vapor from components containing silicone or sulfur can affect the catalytic sensors and thereby distort measurements. If sensors have been exposed to these types of compound, inspection or calibration will be necessary.
High concentrations of organic solvents (e.g. alcohols, aromatic solvents, etc.) or exposure to quantities of gas greater than the specified range of measurement can damage the electrochemical sensors. Inspection or calibration is then recommended.
In the presence of high concentrations of carbon dioxide (CO2 > 1% vol.), the oxygen-measuring electrochemical sensors can slightly overestimate the concentration of oxygen (0.1 to 0.5% vol. O2 overestimate).
Operation under low oxygen levels
If an electrochemical sensor is used in an atmosphere containing less than 1% vol. oxygen for over one hour, the measurement may be an underestimate.
If a catalytic sensor is used in an atmosphere containing less than 10% vol. oxygen, the measurement may be an underestimate.
Page 15
3- Maintenance
15
Chapter 3 | Maintenance
Maintenance primarily involves sensor calibration and changing any sensors that no longer meet their initial performance characteristics.
Since they are liable to affect detection reliability, the tasks described in this chapter are reserved for authorized trained personnel only.
Inspection and maintenance shall be carried out in accordance with standard EN/IEC 60079-17. The instructions contained in the ATEX marking shall be observed (see Chapter 8).
When first powered up or after a sensor replacement, wait for 1 hour before performing a calibration.
It is impossible to calibrate an OLCT 10N within 10 minutes after a power cycle.
Maintenance Frequency
Gas detectors are safety devices. OLDHAM recommends the regular testing of fixed gas detection installations. This type of test consists of injecting the calibration gas into the detector at a sufficient concentration to activate the pre-set alarms. It is to be understood that this test is in no way a replacement for a detector calibration.
The frequency of gas tests depends on the industrial application where the detector is in use. Frequent inspections should be made in the months following the commissioning of the installation, and should then become more widely spaced provided that no significant deviation is observed. If a detector should fail to react in contact with the gas, calibration is essential. The frequency of calibrations shall be appropriate according to the results of the tests (humidity, temperature, dust, etc.); however, it must not exceed one year.
The general manager should put safety procedures in place on-site. OLDHAM cannot be held responsible for their enforcement.
Page 16
16
OLCT 10N
User Manual
The OLCT 10N can be calibrated either:
- Manually
- Automatically
The manual calibration must be done only after each sensor or detector replacement.
A manual calibration must be followed by an automatic calibration.
Please find below the different calibration gas concentrations.
Gas
Range
Calibration gas concentration
Methane (CH4)
0-100% LEL (5% vol.) 0-100% LEL (4.4% vol.)
2.5% CH4/Air (50% LEL)
2.2% CH4/Air (50% LEL)
Hydrogen (H2)
0-100% LEL
2.0% H2/Air (50% LEL)
Butane (C4H10)
0-100% LEL
0.9% C4H10/Air (60% LEL)
Propane (C3H8)
0-100% LEL
1.1% C3H8/Air (55% LEL)
Oxygen (O2)
0-30% vol.
20.9% O2
Carbone monoxide (CO)
0-300 ppm 0-1000 ppm
100 ppm CO 300 ppm CO
Carbone dioxide (CO2)
0-5000 ppm 0-5% vol. 0-100% vol.
3000 ppm CO2
2.5% CO2 50% CO2
Hydrogen sulfide (H2S)
0-30 ppm 0-100 ppm
25 ppm H2S 25 ppm H2S
Nitrogen monoxide (NO)
0-100 ppm 0-300 ppm
50 ppm NO 50 ppm NO
Nitrogen dioxide (NO2)
0-10 ppm 0-30 ppm
10 ppm NO2 10 ppm NO2
Ammonia (NH3)
0-100 ppm 0-1000 ppm
100 ppm NH3 1000 ppm NH3
Semi-automatic calibration
While the OLCT 10N is calibrating, the MX 43 or the MX 32v2 inhibits alarms triggered by the detector and displays a maintenance code. Up to 32 sensors can be calibrated at the same time.
Page 17
3- Maintenance
17
Cal ON Attachment of gas injection device
Cal OFF Removal of gas injection device
Cal OFF
LED is OFF
GREEN blinking
GREEN steady
RED blinking
RED steady
LED sensor status
START
END
Normal Operating
Mode
Inject zero air
(Nitrogen in the case of
an OLCT 10N O2)
and wait for stabilization
Zero is stable
Internal processing
(20 to 60 s)
LED turns off as it
passes or blinks as it
fails
Cal ON
Calibration is
complete and
validated
LED turns off as it
passes or blinks as it
fails
Zeroing is complete
and validated
Remove the cap
and apply it again
within 30 seconds
Inject span gas and wait for stabilization
Signal is stable
Internal processing
(20 to 60 s)
Cal OFF
Cal OFF
Cal ON
Page 18
18
OLCT 10N
User Manual
Semi-automatic calibration device (P/N 6331163)
Page 19
3- Maintenance
19
Sensor replacement
Sensors should be replaced as a part of regular preventative maintenance or following a failed calibration test.
After a sensor replacement, both manual and automatic calibrations must to be conducted.
To change the sensor:
1. Remove the detector cover
2. Push and hold down the sensor replacement button (1) for 5 seconds until the solid green
LED (2) is on.
3. Release the button
4. Replace the sensor and conduct both manual and automatic calibrations (This is a mandatory step)
Figure 6: Sensor
replacement
In order to comply with the ATEX rules, the sensor of LEL detectors is soldered to the card and cannot be changed separately. The entire OLCT 10N detector will need to be replaced. As the detector is factory calibrated, there is no point attempting manual calibration. However, a sensor change must be declared using the maintenance menu of the MX 43 or MX 32v2 controller (see MX 43 or MX 32v2 manual).
Page 20
20
OLCT 10N
User Manual
Manual calibration
A voltmeter and the female connector provided by Oldham (P/N 6116291) are required.
1. Remove the detector cover.
2. Connect the connector to the circuit’s male connector
3. Connect the voltmeter between S+ and S- (value in mV)
Zero the sensor
1. Ensure that the sensor is in clean air (in absence of O2 in the case of OLCT 10N O2). If not, inject zero air (Nitrogen in the case of an OLCT 10N O2) into the sensor at a flow rate of 1 l/min, then wait for voltmeter signal to stabilize (use the calibration gas kit: zero air cylinder, calibration cap, tube).
2. Adjust the zero with the potentiometer’s ZERO until the voltmeter reads 0 mV.
Sensitivity
1. Now inject the calibration gas (1 l/min) into the sensor. Wait for the voltmeter signal to stabilize (use the calibration gas kit: calibration gas cylinder, calibration cap, tube).
2. Adjust the sensitivity if necessary with the potentiometer “SENS” until the signal value (in mV) corresponds to the concentration in the reference gas used. Use the following formula to calculate the correct value for the signal.
3. Stop injecting gas (remove the calibration pipe from the sensor).
4. Wait for the voltmeter to “return to zero”.
U
(mV)
=
1600
x
reference gas concentration
measurement scale
Page 21
3- Maintenance
21
MAINTENANCE WIRES: S- (grey) / S+ (yellow) = signal from 0 mV to 1600 mV for zero and sensitivity settings GND (black) = circuit board ground +VCC (red) = + power supply Note: Blue and green wires are not used.
OLCT 10N connector - P/N 6116291
S+
VCC
V
OLCT 10N
S-
Page 22
22
OLCT 10N
User Manual
Page 23
4- Part numbers and accesories
23
Chapter 4 | Part Numbers and
Accessories
Part Number
Description
OLCT 10N-001
OLCT 10N Digital detector, range 0-100% LIE CH4 (5% vol.)
OLCT 10N-002
OLCT 10N Digital detector, range 0-100% LIE CH4 (4.4% vol.)
OLCT 10N-003
OLCT 10N Digital detector, range 0-100% LIE H2 (4% vol.)
OLCT 10N-004
OLCT 10N Digital detector, range 0-100% LIE C4H
10
(1,5% vol.)
OLCT 10N-005
OLCT 10N Digital detector, range 0-100% LIE C3H8 (2% vol.)
OLCT 10N-200
OLCT 10N Digital detector, range 0-30% vol. O2 (life expectancy 2 years)
OLCT 10N-272
OLCT 10N Digital detector, range 0-30% vol. O2 (life expectancy 5 years)
OLCT 10N-204
OLCT 10N Digital detector, range 0-300 ppm CO
OLCT 10N-205
OLCT 10N Digital detector, range 0-1000 ppm CO
OLCT 10N-252
OLCT 10N Digital detector, range 0-5000 ppm CO2
OLCT 10N-239
OLCT 10N Digital detector, range 0-5% vol. CO2
OLCT 10N-241
OLCT 10N Digital detector, range 0-100% vol. CO2
OLCT 10N-213
OLCT 10N Digital detector, range 0-30 ppm H2S
OLCT 10N-214
OLCT 10N Digital detector, range 0-100 ppm H2S
OLCT 10N-216
OLCT 10N Digital detector, range 0-100 ppm NO
OLCT 10N-217
OLCT 10N Digital detector, range 0-300 ppm NO
OLCT 10N-219
OLCT 10N Digital detector, range 0-10 ppm NO2
OLCT 10N-220
OLCT 10N Digital detector, range 0-30 ppm NO2
OLCT 10N-231
OLCT 10N Digital detector, range 0-100 ppm NH3
OLCT 10N-232
OLCT 10N Digital detector, range 0-1000 ppm NH3
Accessory
Description
Illustration
Part Number
Calibration kit
Kit containing pipe, automatic calibration cap and gas injection tube
Cf. Page 18
6331163 24
Page 24
OLCT 10N
User Manual
Page 25
5- Spare Parts
25
Chapter 5 | Spare Parts
Spare parts list for the different detectors.
Part Number
Description
6798301
CO sensor
6314196
CO2 sensor - 5000ppm range
6314195
CO2 sensor - 5% vol range
6314197
CO2 sensor - 100% vol range
6313990
H2S sensor
6314118
NH3 sensor - 100 ppm range
6314119
NH3 sensor - 1000 ppm range
6113331
NO sensor
6113332
NO2 sensor
6799345
O2 sensor (life expectancy 2 years)
6113720
O2 sensor (life expectancy 5 years)
Page 26
26
OLCT 10N
User Manual
Page 27
6- UE Declaration
27
Chapter 6 | UE Declaration
Page 28
28
OLCT 10N
User Manual
Page 29
7- Specifications
29
Chapter 7 | Specifications
Detector
Dimensions
118x110x63 mm
Ingress protection
IP65
Enclosure
UV-resistant ABS
Cable entries
2 cable glands, M16, outer diameter 4 to 8 mm
Type of cable
Type MPI-22A
Power supply
12-30V
Consumption (under normal operating conditions @ 24 V)
Electrochemical sensor: 2.5 mA Catalytic sensor: 50 mA Infrared sensor: 20 mA
Status indication during calibration
LED red/green
Calibration
Non-intrusive and automatic (with the special magnetic calibration cap) or manual calibration (by potentiometer)
Standard gas value set by the detector
ATEX certification temperature
Certified for use in ATEX zones between -20°C and +55°C
Humidity
15 to 90% RH for the electrochemical sensor 0 to 100% RH uncondensed for the catalytic sensor
Storage conditions
Electrochemical sensor: 4 to 20°C 20 to 60% RH
1 bar ± 10% 6 months maximum
Infrared sensor: 4 to 20°C 10 to 60% RH
1 bar ± 10% 6 months maximum
Catalytic sensor: -50 to +70°C 20 – 60% RH 1 bar ± 10% 6 months maximum
Page 30
30
OLCT 10N
User Manual
Certification
EMC according to EN 50270:2006 (Type 1) Explosive Atmospheres :
II 3 GD Ex nA nC IIC T4 Gc for LEL versions
Ex nA IIC T4 Gc for Toxic (except CO2 versions) and O2
versions
Ex tc IIIC T135°C Dc
-20°C<Ta<+55°C
Sensor specifications
Gas Type
Measuring
range (ppm)
Temp. Range
(°C)
Accuracy
Average
Life
Expectancy
(month)
T50 / T90
(sec)
CH4 Methane
0-100% LEL
-20 to +55
± 1% LEL
(0-70% LEL)
48
6/15
H2 Hydrogen
0-100% LEL
-20 to +55
± 1% LEL
(0-70% LEL)
48
C3H8 Propane
0-100% LEL
-20 to +55
± 1% LEL
(0-70% LEL)
48
C4H10 Butane
0-100% LEL
-20 to +55
± 1% LEL
(0-70% LEL)
48
CO Carbon monoxide
300
1000
-20 to +50
± 3%
36
15/40
CO2 Carbon dioxide
5000
0-5% vol.
0-100% vol.
-40 to +50
± 3%
60
18/32 19/33 13/22
H2S Hydrogen sulfide
30
100
-20 to +50
± 3%
36
15/30
Page 31
7- Specifications
31
NH3 Ammonia
100
1000
-20 to +40
± 5 ppm
± 20 ppm
24
25/70 20/60
NO Nitric oxide
100 300
-20 to +50
± 3%
36
10/30
NO2 Nitrogen dioxide
10 30
-20 to +50
± 3%
24
30/60
O2 (2 years) Oxygen
0-30 Vol.%
-20 to +50
0,4 Vol.%
(15-22 %)
28
6/15
O2 (5 years) Oxygen
0-30 Vol.%
-40 to +50
± 1.5%
60
15/25
Page 32
32
OLCT 10N
User Manual
Line length (use with MX 32)
The following table summarizes the maximum line lengths to be used according to the cross-section and number of sensors connected to a MX 32v2 controller.
TOX and O2
MX 32v2 with internal 100-240Vac power supply or powered externally 24Vdc
Distance in meters
Line cross-section
Number of OLCT 10N detectors used to detect toxic gases or oxygen (except CO2 versions)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm²
(AWG 24)
1
1000
1000
1000 2 1000
1000
1000
3
1000
1000
1000 4 1000
1000
1000
5
1000
1000
1000 6 1000
1000
1000
7
1000
1000
1000 8 1000
1000
1000
Page 33
7- Specifications
33
CO2
MX 32v2 with internal 100-240Vac power supply or powered externally 24Vdc
Distance in meters
Line cross-section
Number of OLCT 10N detectors used to detect CO2
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm²
(AWG 24)
1
1000
1000
1000
2
1000
1000
500
3
1000
1000
450 4 1000
800
350 5 1000
600
250
6
900
500
230
7
800
450
200 8 700
400
170
Page 34
34
OLCT 10N
User Manual
Catalytic
MX 32v2 with internal 100-240Vac power supply
Distance in meters
Line cross-section
Number of OLCT 10N detectors equipped with a catalytic sensor (LEL versions)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm² (AWG 24)
1
1000
1000
500
2
1000
600
250
3
600
400
180 4 450
300
150 5 350
250
100
6
300
200
100
Above 6
Maximum load for an MX 32V2 line (>610mA). An external supply must be used
MX 32v2 powered externally 24Vdc
Distance in meters
Line cross-section
Number of OLCT 10N detectors equipped with a catalytic sensor (LEL versions)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm² (AWG 24)
1
1000
1000
500
2
1000
600
250
3
600
400
180 4 450
300
150 5 350
250
100
6
300
200
100
7
250
175
75 8 225
150
50
Page 35
7- Specifications
35
Line length (use with MX 43)
The following table summarizes the maximum line lengths to be used according to the cross-section and number of sensors connected to a MX 43 controller.
TOX and O2
With any MX 43 mother board version
Distance in meters
Line cross-section
Number of OLCT 10N detectors used to detect toxic gases or oxygen (except CO2 versions)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm²
(AWG 24)
10
1000
1000
1000
20
1000
1000
900
25
1000
1000
500
32
1000
800
300
Page 36
36
OLCT 10N
User Manual
CO2
MX 43 with Rev. A or Rev. B mother board
Distance in meters
Line cross-section
Number of OLCT 10N detectors used to detect CO2
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm²
(AWG 24)
1
1000
1000
1000
2
1000
1000
500
5
1000
600
250
10
550
300
125
12
450
250
100
Above 12
Maximum load for an MX 43 line (>500mA). An external supply must be used
MX 43 with Rev. C mother board
Distance in meters
Line cross-section
Number of OLCT 10N detectors used to detect CO2)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm²
(AWG 24)
1
1000
1000
1000 2 1000
1000
500
5
1000
600
250
10
550
300
125
20
250
150
60
30
180
100
40
Above 30
Maximum load for an MX 43 line (>1200mA). An external supply must be used
Page 37
7- Specifications
37
Catalytic
MX 43 with Rev. A or Rev. B mother board
Distance in meters
Line cross-section
Number of OLCT 10N detectors equipped with a catalytic sensor (LEL versions)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm² (AWG 24)
1
1000
1000
500
2
1000
600
250
3
750
400
180 4 600
330
150 5 450
250
100
Above 5
Maximum load for an MX 43 line (>500mA). An external supply must be used
MX 43 with Rev. C mother board
Distance in meters
Line cross-section
Number of OLCT 10N detectors equipped with a catalytic sensor (LEL versions)
0.75 mm²
(AWG 18)
0.5 mm²
(AWG 20)
0.22 mm² (AWG 24)
1
1000
1000
500 2 1000
600
250
3
600
400
180
4
450
300
150 5 350
250
100 6 300
200
100
7
250
175
75
8
225
150
50 9 200
125
50
10
175
100
50
13
125
75
30
Above 13
Maximum load for an MX 43 line (>1200mA). An external supply must be used
Page 38
38
OLCT 10N
User Manual
Page 39
8- Specific instructions
39
Chapter 8 | Specific instructions
for use in explosive atmospheres in accordance with European ATEX Directive 94/9/EC
General comments
OLCT 10N detectors comply with the requirements of European ATEX Directive 94/9/EC on Gas and Dust explosive atmospheres. OLCT 10N transmitter detectors are used to measure certain explosive or toxic gases or oxygen. They can be installed in ATEX zones 2 or 22 (except for CO2 versions which should be installed in safe area only).
The equipment passed a mechanical shock test for low risk conditions and therefore must be protected against shock hazards in ATEX zones.
The information given in the following sections should be taken into account and observed by the manager of the site where the equipment is installed. Refer to European ATEX Directive 1999/92/EC on minimum requirements for improving the safety and health protection of workers potentially at risk from explosive atmospheres.
OLCT 10N detectors also comply with the requirements of the IECEx international certification scheme relating to Gas and Dust explosive atmospheres.
Marking for LEL versions
The marking will be as follows: OLDHAM Arras
OLCT 10N EXPLO CE
II 3 GD Ex nA nC IIC T4 Gc
Ex tc IIIC T135°C Dc IP 65 Amb. Temp.: -20°C to +55°C INERIS 11ATEX3023X IECEx INE 11.0029X
Warning: Do not open when an explosive atmosphere may be present. Potential electrostatic charge hazard. Polish or wipe only with damp cloth.
Page 40
40
OLCT 10N
User Manual
Avertissement: Ne pas ouvrir si une atmosphère explosive peut être présente Danger potentiel de charges électrostatiques. Ne frotter ou n'essuyer qu'avec un chiffon humide.
Marking for toxic gas or oxygen detector versions (except CO2 versions)
The marking will be as follows: OLDHAM Arras
OLCT 10N TOX CE
II 3 GD Ex nA IIC T4 Gc
Ex tc IIIC T135°C Dc IP 65 Amb. Temp.: -20°C to +55°C INERIS 11ATEX3023X IECEx INE 11.0029X
Warning: Do not open when an explosive atmosphere may be present. Potential electrostatic charge hazard. Polish or wipe only with damp cloth. Avertissement: Ne pas ouvrir si une atmosphère explosive peut être présente Danger potentiel de charges électrostatiques. Ne frotter ou n'essuyer qu'avec un chiffon humide.
CO2 versions are not certified to use in ATEX area
Page 41
8- Specific instructions
41
Page 42
42
OLCT 10N
User Manual
Page 43
Page 44
The Fixed Gas Detection Experts
EUROPEAN PLANT AND OFFICES
Z.I. Est – rue Orfila CS 20417 – 62027 Arras Cedex FRANCE
Tél: +33 (0)3 21 60 80 80 – Fax: +33 (0)3 21 60 80 00
Website: https://gasdetection.3m.com
AMERICAS
Tel: +1-713-559-9280
Fax: +1-281-292-2860
ASIA PACIFIC
Tel: +86-21-3127-6373
Fax: +86-21-3127-365
EUROPE
Tel: +33-321-608-080
Fax: +33-321-608-000
Contact: gasandflamedetection@mmm.com
Loading...