Ocean Optics Smart pH Cuvettes User Manual

Smart pH Cuvettes Instructions
Overview
The Fiber Optic pH Sensor system consists of the following:
Smart pH Cuvettes (1 cm x 1 cm PMMA or quartz) and/or patches Ocean Optics VIS-NIR spectrometer (or Jaz Sensor module) SpectraSuite software for reading values Light source (LS-1 Tungsten Halogen Light Source with a blue filter or a white LED) CUV-UV Cuvette Holder Connecting Fibers
Calibration requires recording spectra in high and low pH samples, as well as in at least one pH standard solution (such as a NIST-traceable buffer).
The Smart pH Cuvettes use a sol gel sensing material coated onto the inner walls of a cuvette. The immobilized indicator dye(s) are encapsulated into the sol gel matrix, allowing for the diffusion of ions while preventing leaching of the dye. The cuvettes provide very accurate measurements in the biological range, from pH 5 to 9. Advantages of these cuvettes over traditional potentiometric devices include faster response time, easy storage and no maintenance, and low cost. These are especially useful for monitoring low conductivity samples such as boiler water, where electrode devices fail.
Ocean Optics’ fully integrated pH systems provide full spectral analysis to help eliminate errors from dye leaching or from changes in turbidity, temperature, and ionic strength. Inherent calibration based on the physical properties of the immobilized indicator dye eliminates the need for frequent calibration. The ratiometric algorithm provides accurate and reproducible measurements at a high resolution.
Smart pH Cuvettes are 1 cm by 1 cm and are made of Poly(methyl methacrylate) (PMMA) or quartz. The PMMA material has a temperature range of -5 °C to 70 °C, while quartz is available for specific applications such as high-temperature measurements. The cuvettes are compatible with aqueous solutions, ethanol/methanol solutions, ammonia, peroxides, sodium hypochlorite solutions, while the quartz cuvettes are also compatible with concentrated acids (nitric, sulfuric, acetic, etc) and acetone.
Although PMMA cuvettes are designed to be disposable, they can be used multiple times. The lifetime depends on the extent of exposure to high pH levels and harsh chemicals. See
Storage/Lifetime for more information.
SC-PH-VIS1M-SAM-01-1109 1
Cuvette
Smart Cuvettes Instructions
Cuvette Storage/Lifetime
Smart Cuvettes can be stored dry at room temperature for any amount of time. As they are used, the cuvettes may slowly leach indicator dye from the sensing material. As a rule,
absorbance at pH 11 falls below 0.1, the cuvette should be discarded and replaced
reference of pH 1). The cuvette’s lifetime depends on frequency of use, harshness of the samples it is exposed to, the temperature of samples, and other environmental factors.
once the maximum
(assumes a
Nature of Samples
The analyte solutions being measured should have a pH within the biological range (pH 5-9) for accurate readings. Data obtained from analyte solutions that register values above or below this range should not be considered valid within the specifications of the cuvette. Concentrated acids and acetone will ruin the plastic cuvettes, so these types of chemicals should be avoided. Aqueous solutions, ethanol/methanol solutions, peroxides, ammonia, and sodium hypochlorite solutions are all compatible with the sensor material and the plastic cuvette. Samples should be optically transparent, having no turbidity or sediment present. It is also ideal to have analyte solutions that are colorless, though colored liquids can be compensated for.
Response time is dependent on the ionic strength of the solution, with higher salinity samples responding notably faster. For example, using the calibration buffers of pH 5 – 8 will show a 90% response in 10 seconds or less, but when pure D.I. water is being measured, more time is needed to equilibrate at a final value. Make sure that the cuvette is filled to a sufficient level with the analyte solution. If the liquid level is at or below the optical path, the data will not be valid. To ensure there is sufficient liquid, it is recommended to fill the cuvette about 70% its height. Likewise, more accurate results will be obtained if the cuvette is rinsed once or twice with the analyte solution after calibration. This removes any residual buffer solution that may contaminate your sample. Note that once the cuvette has been affixed into the cuvette holder, it should not be moved or removed until all measurements have been completed.
When immersed in solution, the film dyes may leach very slowly over time and will have to be replaced. The film response rate is limited by diffusion of ions into the material, therefore increasing stirring speed and ionic strength tend to increase the response rate.
pH Smart Cuvette Set Up
The following procedures describe how to connect and calibrate pH Smart Cuvettes using a VIS-NIR spectrometer, a light source and SpectraSuite software. See your spectrometer and SpectraSuite manual for more detailed installation information.
Installing the pH Sensor System
Procedure
Perform the steps below to install the pH Sensor components:
1. Install SpectraSuite on your computer.
2. Connect the spectrometer to your computer using the supplied USB cable.
3. Install the light source as specified in its instructions.
2 SC-PH-VIS1M-SAM-01-1109
Smart Cuvettes Instructions
4. Attach the fibers between the spectrometer, cuvette holder, and light source.
5. Turn on the light source and allow it to warm up for the period specified in the light source instructions.
Caution
Make sure that the cuvette says fastened in the cuvette holder with the tightening screw and that it does not move until all measurements have been completed. Any movement will change the optical signal, disrupting the quality of the measurement.
Calibrating the pH Sensor System
The Smart Cuvettes include a pre-calibrated pK value determined at the factory. This value was originally obtained at 22°C, and it is recalculated using the temperature compensation algorithm based on the temperature that was entered in SpectraSuite’s Calibration Wizard. Using the Factory Calibration method is ideal for being able to start making pH measurements quickly, though it is less accurate than performing an Independent Calibration. The specifications listed for the cuvettes assume a complete Independent Calibration, as this eliminates the errors seen from temperature and other environmental differences.
Using Factory Calibration
Procedure
1. Open SpectraSuite and select File | New | New Sol Gel pH Measurement.
2. Click the Calibration Wizard button to begin the calibration.
3. Select the spectrometer to use and click
4. Select
Use Factory Calibration and click Next. The Experimental Parameters screen appears.
Next.
SC-PH-VIS1M-SAM-01-1109 3
Loading...
+ 7 hidden pages