Ocean Optics NeoFox User Manual

and NeoFox Sport
Installation and Operation Manual
Document Number 013-20000-009-02-201203
Offices: Ocean Optics, Inc. World Headquarters
Phone 727.733.2447 Fax 727.733.3962
8 a.m.– 8 p.m. (Mon-Thu), 8 a.m.– 6 p.m. (Fri) EST
E-mail: Info@OceanOptics.com (General sales inquiries)
Orders@OceanOptics.com (Questions about orders) TechSupport@OceanOptics.com (Technical support)
Additional Offices:
Ocean Optics Asia
137 Xianxia Road, Suite 1802, Changning District, Shanghai, PRC. 200051
Phone 86.21.5206.8686
Fax 86.21.5206.8686 E-Mail Sun.Ling@OceanOptics.com
Ocean Optics EMEA
Geograaf 24, 6921 EW DUIVEN, The Netherlands
Phone 31-(0)26-3190500
Fax 31-(0)26-3190505 E-Mail
Copyright © 2010 Ocean Optics, Inc.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means, electronic, mechanical, photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc. This manual is sold as part of an order and subject to the condition that it shall not, by way of trade or otherwise, be lent, re-sold, hired out or otherwise circulated without the prior consent of Ocean Optics, Inc. in any form of binding or cover other than that in which it is published.
Trademarks
All products and services herein are the trademarks, service marks, registered trademarks or registered service marks of their respective owners.
Limit of Liability
Ocean Optics has made every effort to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The information provided is on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect to any loss or damages arising from the information contained in this manual.
Info@OceanOptics.eu

Important Safety Notices

WARNINGS

1. Intense UV radiation is emitted from the NeoFox LED. Do not look directly at the
LED output with the naked eye once the SMA cap is removed. To verify there is power to the instrument, place a white piece of paper in front of LED 1.
2. Do not open the instrument case. High voltage is present. There are no
serviceable parts inside. Return the instrument to the factory for service.
013-20000-009-02-201203 i
Important Safety Notices
ii 013-20000-009-02-201203

Table of Contents

About This Manual .......................................................................................................... vii
Document Purpose and Intended Audience.............................................................................. vii
What’s New in this Document ................................................................................................... vii
Document Summary.................................................................................................................. vii
Product-Related Documentation ............................................................................................... vii
Safety and Compliance .............................................................................................................vii
Chapter 1: Introduction ......................................................................1
NeoFox System Overview ............................................................................................... 1
NeoFox ......................................................................................................................................1
NeoFox Sport ............................................................................................................................ 2
Features..................................................................................................................................... 2
Oxygen Sensor Probes ................................................................................................... 2
Software .......................................................................................................................... 3
Applications ..................................................................................................................... 3
Shipment Components.................................................................................................... 3
NeoFox ......................................................................................................................................3
NeoFox Sport ............................................................................................................................ 4
Additional Recommended Equipment ............................................................................. 4
Chapter 2: Installation ........................................................................7
Overview ......................................................................................................................... 7
Software Installation ........................................................................................................ 7
Hardware Installation....................................................................................................... 10
Installing NeoFox Hardware Drivers................................................................................ 12
Installing Hardware Drivers Using Windows XP ....................................................................... 12
Installing Hardware Drivers Using Windows Vista ....................................................................14
Chapter 3: Using NeoFox with NeoFox Viewer Software................17
Overview ......................................................................................................................... 17
013-20000-009-02-201203 iii
Table of Contents
Entering a New Event...................................................................................................... 18
Data Logging ................................................................................................................... 18
Advanced Settings .......................................................................................................... 19
Setup .........................................................................................................................................20
Status......................................................................................................................................... 21
Advanced Status ....................................................................................................................... 22
Advanced Settings..................................................................................................................... 22
Sensor Type Selection .................................................................................................... 24
APD Voltage.................................................................................................................... 24
Calibrating the Probe....................................................................................................... 25
Multipoint Calibration ................................................................................................................. 25
Two-Point Calibration ................................................................................................................27
Single Point Reset........................................................................................................... 30
Single Point Reset Control ........................................................................................................ 32
Saved Settings ................................................................................................................ 33
Saving Data..................................................................................................................... 35
NeoFox Communication Interfaces ................................................................................. 36
Saving/Restoring Chart Layouts...................................................................................... 36
Saving Settings.......................................................................................................................... 37
Resetting Saved Settings .......................................................................................................... 37
Chapter 4: Using NeoFox Sport.........................................................39
Overview ......................................................................................................................... 39
NeoFox Sport Features ................................................................................................... 39
NeoFox Sport Menu ........................................................................................................ 40
Setting Up the NeoFox Sport .......................................................................................... 43
Performing a Single Point Reset in the Field................................................................... 44
Data Logging / Saving Data ............................................................................................ 45
Logging Data ............................................................................................................................. 45
Viewing the Data Log File Name............................................................................................... 46
Setting the Data Logging Interval .............................................................................................. 46
Data Log File Format................................................................................................................. 46
NeoFox Sport Saved Settings ......................................................................................... 47
Low Battery Detector....................................................................................................... 47
Appendix A: Specifications................................................................48
iv 013-20000-009-02-201203
Table of Contents
Appendix B: Maintenance..................................................................50
Updating Firmware .......................................................................................................... 50
Updating Software........................................................................................................... 50
Troubleshooting............................................................................................................... 51
Appendix C: Analog Output...............................................................54
Analog Output Connector................................................................................................ 54
Configuring Analog Voltage Output Channel with NeoFox Viewer ........................................... 54
Configuring the Analog Current Output Channel with NeoFox Viewer ..................................... 55
Selecting a Data Source............................................................................................................ 56
Selecting a Data Range............................................................................................................. 57
Saving the Analog Output Settings ........................................................................................... 58
Analog Output Pinouts .................................................................................................... 58
Analog Output Parameters List ....................................................................................... 59
Analog Output Specifications .......................................................................................... 59
Interfacing to the 4-20mA Output .................................................................................... 60
4-20 mA Loop Voltage Supply......................................................................................... 60
4-20 mA Isolation ............................................................................................................ 61
Resistor / Volt Meter........................................................................................................ 62
Appendix D: Calculating Drift Per Sensor ........................................64
Index .....................................................................................................66
013-20000-009-02-201203 v
Table of Contents
vi 013-20000-009-02-201203

About This Manual

Document Purpose and Intended Audience

This document provides the users of the NeoFox and NeoFox Sport with instructions for setting up, calibrating and performing experiments with their equipment.

What’s New in this Document

This version of the NeoFox Installation and Operation Manual updates for the firmware version 4.0 release.

Document Summary

Chapter Description
Chapter 1: Introduction Contains descriptive information about the NeoFox. It also
provides a list of system requirements, interface options, and shipment components.
Chapter 2: Installation Provides installation instructions. Chapter 3: Using NeoFox Contains steps to operate NeoFox Viewer software with the
NeoFox hardware. Also contains steps for restoring screen defaults.
Chapter 4: Using NeoFox Sport Contains the menu structure and operational information for
the NeoFox Sport.
Appendix A: Specifications Lists NeoFox specifications. Appendix B: Maintenance Contains information for updating firmware and software and a
troubleshooting table.
Appendix C: Analog Output Provides information about NeoFox’s analog output connector.

Product-Related Documentation

You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com
document from the available drop-down lists. Or, use the Search by Model Number field at the bottom
of the web page.
013-20000-009-02-201203 vii
. Select Technical Operating Instructions, then choose the appropriate
About This Manual
Jaz Installation and Operation Manual
Detailed instructions for the Oxygen Sensor Probes are located at:
http://www.oceanoptics.com/technical/NEW OOISensors FOXY Manual.pdf
Engineering-level documentation is located on our website at Technical Engineering Docs.
NeoFox Calibration Engineering NoteNeoFox Communication Interfaces Engineering Note
You can also access operati
ng instructions for Ocean Optics products from the Software and Technical
Resources CD that ships with the product.

Safety and Compliance

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which the user will be required to correct the interference at his own expense.
WARNING: The authority to operate this equipment is conditioned by the requirement that no
modifications will be made to the equipment unless the changes or modifications are expressly approved by the manufacturer.
viii 013-20000-009-02-201203
Chapter 1

Introduction

NeoFox System Overview

The NeoFox is a device for measuring fluorescence lifetime, phase and intensity. It uses LED excitation and photodiode detection with filter-based wavelength selection for easy experimental set-up and control. Because the unit is self-contained, it is invariant to fiber bending and stray light, and has a wide dynamic range of optical intensity as well as low optical and electronic crosstalk, and low drift and phase noise. Therefore, the NeoFox is especially useful for oxygen sensing applications where stability and sensitivity to drift is important and where sample set-ups must be left undisturbed for long periods of time.

NeoFox

NeoFox is available as a compact, self-contained, benchtop unit.
013-20000-009-02-201203 1
1: Introduction

NeoFox Sport

NeoFox Sport is available as a handheld oxygen­sensing system with a phase fluorometer, an onboard microprocessor with OLED display and a battery module, or as part of the Jaz community of stackable, modular and autonomous component modules (limit of one sensor module per stack).

Features

NeoFox is a fluorescence-sensing detector for our optical sensors line and provides a viable alternative to traditional chemical sensing devices. Our optical sensors consist of transducer materials, applied to the tips of optical fibers or to substrates such as patches or cuvettes, which change optical properties in response to specific analytes in their immediate environment.
Additional features that are available to NeoFox Sport users include:
Battery-powered operation
Real-time display of temperature and oxygen in various measurement units:
Gaseous oxygen partial pressure as a percentage of 1 ATM.
Gaseous oxygen partial pressure in Torr.
Dissolved oxygen partial pressure in ppm (parts per million)
Dissolved oxygen concentration in um/L (micromoles per liter)
Data logging of temperature and oxygen readings in units listed above.
In-field single point reset (also available on desktop NeoFox)

Oxygen Sensor Probes

The NeoFox is used with Ocean Optics’ NeoFox software and custom probes (priced separately). Ocean Optics oxygen sensor probes are low-power, portable devices that offer high sensitivity, reversibility, and stability. Their small size is useful for remote monitoring. What's more, the thin film used in the probe tips consumes no oxygen, allowing for continuous contact with the sample. Ocean Optics oxygen sensors offer other key advantages: they're ideal for viscous samples and are immune to interference caused by pH change or from changes in ionic strength, salinity, and biofouling. The NeoFox supports the following Ocean Optics sensors:
2 013-20000-009-02-201203
1: Introduction
FOXY – Ocean Optics’ standard oxygen sensors designed for monitoring oxygen partial pressure
in gas and aqueous solutions. FOXY is a fiber optic fluorescence probe with proprietary oxygen­sensing thin-film coating on the tip.
HIOXY – Designed for monitoring oxygen partial pressure in nonaqueous vapors and solutions.
The sensor coating chemistry is compatible with oils, alcohols, and hydrocarbon-based vapors and liquids.
FOSPOR – A new generation of highly sensitive sensor coating for monitoring traces of oxygen
in gas and liquids.
See
http://www.oceanoptics.com/products/probes.asp for more information on probes
available from Ocean Optics.
The NeoFox connects to a PC via USB connection and saves your data in an easy-to-use Microsoft Excel format. NeoFox can be configured with single-channel LED excitation and detection, and modulation frequencies to 100 kHz. The on-board pressure transducer measures atmospheric pressure.

Software

The NeoFox standalone module uses NeoFox Viewer software to display oxygen measurements on a connected Windows XP PC. The NeoFox Sport can also use NeoFox Viewer software when connected to a Windows XP PC, or it can be untethered from a computer and display some Oxygen measurements on its OLED screen. See Chapter 4: Using NeoFox Sport NeoFox Sport.
for more information on the functionality of the

Applications

Typical applications include the following:
Bioprocess monitoring
Environmental applications (such as waste water management)
Life science
Modified Atmosphere Packaging (MAP)

Shipment Components

NeoFox

The following equipment and information ships with the standalone NeoFox:
NeoFox Unit
013-20000-009-02-201203 3
1: Introduction
5V Power Adapter Mini USB Cable Software and Technical Resources CD
Each order ships with the Ocean Optics Software and Resources CD. This disc contains software, operating instructions, and product information for all Ocean Optics software, spectrometers, and spectroscopic accessories. You need Adobe Acrobat Reader version 6.0 or higher to view these files. Ocean Optics includes the Adobe Acrobat Reader on the Software and Technical Resources CD.
NeoFox CD
This CD contains this manual and the NeoFox Viewer software.
Packing List
The packing list is inside a plastic bag attached to the outside of the shipment box (the invoice arrives separately). It lists all items in the order, the shipping and billing addresses, and any items on back order.

NeoFox Sport

The following equipment and information ships with the NeoFox Sport:
Jaz stack with the NeoFox Sport module(s)
See your Jaz manual for more information about what ships with Jaz (refer to Product-Related
Documentation
SD card containing the neofox.cal calibration file Software and Technical Resources CD NeoFox CD Packing List
).

Additional Recommended Equipment

NeoFox Viewer Software Temperature Probe FOXY, HIOXY or FOSPOR Oxygen Sensor Probe RedEye
4 013-20000-009-02-201203
®
Patches (FOXY, HIOXY and FOSPOR)
1: Introduction
FOXY-CAL, HIOXY-CAL, FOSPOR-CAL
This is an optional factory calibration service for environments from 0 to 80ºC.
Bifurcated Fibers 21-02 SMA Splice Bushing
This in-line adapter connects the probe to the fiber assembly.
013-20000-009-02-201203 5
1: Introduction
6 013-20000-009-02-201203
Chapter 2

Installation

Overview

There are two steps to completely installing the NeoFox system onto your computer.
First, you must install the NeoFox Viewer software. This is done by running the NeoFox-Viewer
Setup.msi installer program for XP systems or the NeoFox-Vista.msi installer for Vista systems.
The second step is to properly install the NeoFox hardware drivers. To do this, plug in the device
and wait for the new hardware wizard to run.
You must install the NeoFox Viewer software application prior to connecting the NeoFox unit to the computer to install the drivers required for the hardware. If you do not install the software first, the system will not properly recognize the NeoFox hardware.
CAUTION
Do not connect the NeoFox unit to the PC prior to installing the NeoFox Viewer software.

Software Installation

You can install the NeoFox Viewer software from either the CD that came with your NeoFox unit or from the Ocean Optics website.
Use the following procedure to install NeoFox Viewer software on a Windows XP PC. The process on Windows Vista machines is very similar. The important thing to note is to make sure to install the driver from the specific location within the NeoFox Viewer\USB Driver directory.
Note
You may be prompted to install .NET runtime libraries if you do not already have them installed on your computer. You can select to install them from the Internet, or you can run the vc-redist­setup.exe program located on your NeoFox installation CD. However, this installation program will also need to connect to the Internet to download the necessary files.
013-20000-009-02-201203 7
2: Installation
Procedure
To install the NeoFox software,
1. Close all other applications running on the PC.
2. Start the software installation process.

Installing from CD:

a. Insert the CD containing the NeoFox Viewer software.
b. Select the drive on your computer with the software CD.
c. Double-click on the NeoFox-Viewer Setup.msi installer program for Windows XP or the
NeoFox-Vista.msi installer for Vista computers. The installation wizard appears.

Installing from the Web:

d. Go to
http://www.oceanoptics.com/technical/softwaredownloads.asp.
e. Right-click on NeoFox Viewer Software and select Save Target As… to download the
executable to your machine.
f. Click Save. The installation begins. g. Click Run. The NeoFox Viewer Setup screen appears.
3. Click Next. The Select Installation Folder screen appears. The default location is C:\Program
Files\Ocean Optics\NeoFox-Viewer\.
8 013-20000-009-02-201203
2: Installation
4. Click Next to select the default location, or browse to the desired location and click Next. You
are now ready to install the software.
5. Click Install. The Neofox Viewer installs at the selected location. This process can take several
minutes. When the software has successfully installed, the final Install Wizard screen appears.
013-20000-009-02-201203 9
2: Installation
6. Click Finish. Go to Hardware Installation

Hardware Installation

WARNING
.
Ensure that the plastic caps are covering the two SMA connectors on the front of the NeoFox unit. Intense UV radiation is emitted from the LEDS when the unit is powered-up. Do NOT look directly at the LED output with the naked eye.
10 013-20000-009-02-201203
2: Installation
Procedure
1. Unpack the equipment and verify that you have all the necessary components (see Shipment
Components
and Additional Recommended Equipment).
2. Connect power cord from the power supply that came with your NeoFox unit from back of unit to
an AC outlet. Do NOT look directly at the light being emitted with the naked eye.
Caution
Use only the power supply that came with your NeoFox unit. Using a different power supply could damage your equipment.
3. Connect power cord from the power supply that came with your NeoFox unit from back of unit to
an AC outlet. Do NOT look directly at the light being emitted with the naked eye.
Note
On the NeoFox Sport hardware, there is a difference between the power connector on the DPU module and the power connector on the NeoFox module. The external power connector on the NeoFox module is the auxiliary power connector. It provides power only to the NeoFox module itself. It will not charge the battery or provide power to the DPU. Very few users will ever need to use the NeoFox module’s auxiliary power connector.
4. Connect the NeoFox unit to your computer using the USB cable. See Installing NeoFox
Hardware Drivers
.
Note
On the NeoFox Sport hardware, there is a difference between the USB connector on the DPU module and the USB connector on the NeoFox module. The USB connector on the NeoFox module is the NeoFox Sport’s primary USB connector. When the USB cable is attached to the NeoFox module and the stack power on, you can interact with the NeoFox module through the NeoFox Viewer software. The USB connector on the DPU is the unit’s auxiliary USB connector. You will not be able to interact with the device in NeoFox Viewer if you have plugged the USB cable into the DPU module. You will only want to use this USB port if you need to update the DPU firmware.
5. Connect the temperature probe (if you have one) to the rear panel of the NeoFox unit.
6. Locate the bifurcated fiber that came with the system. This optical fiber assembly has a “Y”
shaped design.
7. Connect one arm (it doesn’t matter which one) of the bifurcated end of the probe fiber to LED
connector and the other arm to the Detector connector on front of unit.
013-20000-009-02-201203 11
2: Installation
8. If you are using an Oxygen probe, locate the 21-02 SMA Splice Bushing that came with the
probe. This item is a 0.75" screw with two female ends. Screw one end of the splice bushing into the SMA 905 connector on the end of the probe. If you are using the RedEye Patch, you don’t need the splice bushing.

Installing NeoFox Hardware Drivers

When you connect NeoFox to a computer for the first time using the USB cable, you must then install the appropriate hardware drivers. This procedure differs, depending on whether you are using Windows XP or Windows Vista as the operating system.

Installing Hardware Drivers Using Windows XP

The following Found New Hardware wizard appears when you first attach NeoFox to computers running the XP operating system.
12 013-20000-009-02-201203
2: Installation
Procedure
1. From the first hardware wizard screen, select "No, not this time". Click Next. The next hardware
wizard screen appears.
2. Select “Install the software Automatically (Recommended).” Click Next. The system installs the
necessary driver software to operate your NeoFox device. The next hardware wizard screen appears.
3. When the hardware driver installation completes successfully, the following screen appears:
013-20000-009-02-201203 13
2: Installation

Installing Hardware Drivers Using Windows Vista

The following Found New Hardware wizard appears when you first attach NeoFox to computers running the Vista operating system.
Procedure
1. From the first Vista hardware wizard screen, select Locate and install driver software (recommended). The next hardware wizard screen appears.
2. Select Don’t search online. The next hardware wizard screen appears.
3. Select I don’t have the disc. Show me other options. The next hardware wizard screen appears.
4. Select Browse my computer for driver software (advanced). The next hardware wizard screen
appears.
5. Highlight the driver_support directory (located beneath C:\Program Files\Ocean Optics\NeoFox­Viewer) and click OK followed by Next. The next hardware wizard screen appears.
6. If the following screen appears, select Install this driver anyway.
14 013-20000-009-02-201203
2: Installation
7. When the hardware driver installation completes successfully, the following screen appears:
013-20000-009-02-201203 15
2: Installation
16 013-20000-009-02-201203
Chapter 3
Using NeoFox with NeoFox
Viewer Software

Overview

When you first display the NeoFox Viewer software, the basic NeoFox Viewer screen appears.
This screen’s controls enable you to perform the following actions:
Insert comments into the exported data file using the Enter New Event textbox. See Entering a
New Event
Set scans to average
Select the sensor type
013-20000-009-02-201203 17
for more information.
3: Using NeoFox with NeoFox Viewer Software
Select Automatic or Manual Gain
Start/Stop Data Logging (see Data Logging
)

Entering a New Event

The Enter New Event text box at the top of the screen enables you to insert your comments into a file
containing exported data. Your comment will appear in the Events column at the far right end of each row of data.
You may enter multiple comments. Each comment is time stamped with the instant you clicked on the
Enter New Event field.
When you create the export data file, your comment(s) will appear on the row of data that most closely corresponds to the time you created your comment.
Procedure
To enter a new comment:
1. Click the Enter New Event field at the top of the screen.
2. Enter your message into the text field.
3. Press the Enter key.

Data Logging

You can save a .csv file (useful for Microsoft Excel) via data logging. When you click the Start/Stop Data Logging button on the main NeoFox Viewer screen, the Data Logging Control Panel screen
appears.
18 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
When you click Start Logging, samples are exported from that point onward and continue until you click Stop Logging and then click Save Settings and Close Window.
Note
User-defined events entered in the Enter New Event field on the basic NeoFox Viewer screen
are not written to the data logging export file.
Field Description
Start Logging
Stop Logging
Specify Output File
Specify Logging Interval
Choose Columns to Export
Close Window Without Saving Changes
Save Settings and Close Window
Initiates data logging process. Actual data logging does not start until you
click Save Settings and Close Window.
Click this button to stop data logging. Data logging does not actually stop
until you click Save Settings and Close Window.
Click Choose to select the file to which the data will be exported. If the file
already exists, you will be asked if you want to overwrite its contents.
Select how frequently you want data records to be written to the log file.
Select to export one of the following data types:
All Columns: Exports all available types of data
Standard: Exports a limited subset of the available data types (O
Tau, temperature, pressure)
Closes this window. None of the changes that you have made will take effect or be saved.
All changes that you have made on this window are saved. If you have
clicked Start Logging, data logging will start the moment you click Save Settings and Close Window. If you have clicked Stop Logging, data logging will stop the moment you click Save Settings and Close Window.
2
%,

Advanced Settings

The NeoFox Viewer software provides a lot of information and features. However, most users will probably only use these functions:
Oxygen reading: Use the pick list above the Oxygen Reading to select to display oxygen as O
P (torr), pO
(torr), DO (ppm or mg/L), or DO (µmol/L).
2
Sensor type selection: FOXY, HIOXY or FOSPOR
Calibration
Salinity Correction
This screen is available from the View | Advanced Settings menu.
013-20000-009-02-201203 19
2
%,
3: Using NeoFox with NeoFox Viewer Software
NeoFox Viewer Advanced Settings Window
Note
If the controls on the right side of the screen seem to be disabled, stuck, or if they say Error even after clicking the Manual Gain button (
software versions are not compatible. You will then need to install the appropriate software version onto your computer to use NeoFox. See Software Installation
Maintenance
The Advanced Settings window contains the following tabs (bottom right corner of screen):
SetupStatusAdvanced StatusAdvanced Settings
for more information.
), the problem is likely that the firmware and
and Appendix B:

Setup

Setup allows you to perform such functions as Sensor Type Selection, adjusting the APD Voltage (for
firmware versions below 4.00), and Calibrating the Probe.
20 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software

Status

Clicking on the Status tab displays the status information shown below.
013-20000-009-02-201203 21
3: Using NeoFox with NeoFox Viewer Software

Advanced Status

The Advanced Status tab displays the values of parameters such as Oxygen (base and converted), Tau,
etc. as shown below.

Advanced Settings

This tab displays the advanced settings shown below for sensor waveform, decay waveform, FFT and salinity correction.
22 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
Setting Salinity Correction
NeoFox Viewer provides you with the capability to change the salinity correction setting from its default setting of 0. Salinity correction applies to dissolved oxygen for PPM and µM/L; it does not apply to percentage.
Procedure
To change the salinity setting:
1. Double-click on the Salinity Calibration value (default is 0).
2. Type the new salinity value.
3. Press Enter.
013-20000-009-02-201203 23
3: Using NeoFox with NeoFox Viewer Software

Sensor Type Selection

Before using the software, you must select the appropriate sensor type from the three buttons in the
Sensor field on the right side of the screen. Choose FOXY, HIOXY, or FOSPOR.

APD Voltage

APD voltage is the voltage applied to the APD detector, which determines the detector’s sensitivity towards converting photons to electrical signal. This can be manually adjusted for firmware versions below 4.0 if the signal level of the sensor appears to be low. For firmware versions 4.00 or above, this parameter is set at the factory and cannot be changed.
Procedure
To manually adjust the signal level:
1. From the Basic View screen, select View | Advanced Settings to display the Advanced View
screen. Make sure that the Setup tab (lower right corner) is selected.
24 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
2. Click the
3. Slide the APD Voltage selector to the desired level.
4. Click the
button in the Gain field.
button in the Gain field.

Calibrating the Probe

When you access the software for the first time, the Oxygen field will be blank. To see an oxygen value, you must first enter calibration information for the probe. This calibration information is unique for each oxygen sensor probe itself, but is independent of the NeoFox hardware unit. Therefore, a new calibration must be loaded each time a new sensor is added or before first use.
You have the option of purchasing a factory calibration for your Oxygen Sensor Probe (FOXY-CAL, HIOXY-CAL, FOSPOR-CAL). If you have a factory calibration, follow the instructions for Multipoint
Calibration
. If you do not have a factory calibration, follow the instructions for a Two-Point Calibration.

Multipoint Calibration

Follow these instructions if you purchased a factory calibration file for your sensor. This file can be found on the NeoFox CD. To retrieve it, select Calibration from the Options menu.
013-20000-009-02-201203 25
3: Using NeoFox with NeoFox Viewer Software
Procedure
1. From the Main screen, select Options | Calibration.
2. Click the Setup tab. The Setup screen appears.
3. Under Select Sensor Calibration Method, choose Multi-Point Calibration.
4. Under Select How Temperature Is Measured, choose Use NeoFox Temperature Sensor.
5. Under Enter Sensor Type And Optional Data, choose the type of sensor and add the serial number and any notes needed.
Note
Normalized decay time = τ0/τ.
6. Click the button, then browse to the location of the factory calibration file on the NeoFox CD.
7. Select the Multi-Point Calibration tab. The uploaded calibration table appears on the left side of
the screen. The curve-fitted graphs appear on the right side of the screen.
26 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
8. Click on
Calibration Constants table fills in on the bottom left corner of the screen.
9. Click the Confirmation screen appears.
10. Click Yes to confirm the download and continue.
11. Proceed to the Single Point Reset
button on the bottom right corner of the screen. The Computed
button to use the coefficients for oxygen calculation. A Download
.

Two-Point Calibration

If you want to do your own calibration instead of using a factory calibration, follow these Two-Point Calibration instructions.
Procedure
1. From the Main screen, select Options | Calibration.
2. Click the Setup tab. The Setup screen appears.
013-20000-009-02-201203 27
3: Using NeoFox with NeoFox Viewer Software
3. Under Select Sensor Calibration Method, choose Two-Point Calibration.
4. Under Select How Temperature Will Be Measured, choose None (Two point calibration only).
5. Under Enter Sensor Type And Optional Data, choose the type of sensor and add the serial
number and any notes needed.
6. Select the Two-Point Calibration tab. You need two known concentrations of oxygen to perform
this calibration, one being the 0% oxygen concentration.
28 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
7. Expose your oxygen sensor to 0% oxygen and let the Tau value stabilize. You can observe the Tau value on the bottom of the screen to determine when it stabilizes.
8. Click on the Use Current Tau for Point #1 box once the tau value is stable.
9. Expose your oxygen sensor to the second known concentration of oxygen. You can expose the probe to ambient conditions in air, which is 20.9% oxygen.
10. Enter the known oxygen value in the O2% box for point #2.
11. Once the Tau value is stable, click on the Use Current Tau for Point #2 box.
12. Now that you have the two points of calibration, click anywhere in the graph chart displayed on this screen. The two points will appear on the graph (circled in red on the figure above), which
will Compute Constants m and b.
13. To use this calibration for oxygen measurement, click the
button. A Download
Confirmation screen appears.
14. Click Yes to confirm the download and continue. The main NeoFox Viewer window will now
display oxygen measurements.
013-20000-009-02-201203 29
3: Using NeoFox with NeoFox Viewer Software
Note
Normalized decay time = τ0/τ.

Single Point Reset

A Single Point Reset must always be done after a Multipoint Calibration table is uploaded.
CAUTION
If any settings in the experimental setup change (APD gain, voltage, probe geometry, etc.), it is strongly recommended that you do a Single Point Calibration to offset these changes. This includes any gain setting changes or switching between NeoFox electronics units.
30 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
Procedure
To perform a Single Point Reset,
1. From the Main screen, select Options | Calibration.
2. Click the Setup tab.
3. Under Select Sensor Calibration Method, choose Single Point Calibration.
4. Under Select How Temperature Will Be Measured, choose Use NeoFox Temperature Sensor.
5. Select the Single Point Calibration tab. The Single Point Calibration screen appears.
6. Single point calibration is always performed within the calibration range. This is usually done in ambient conditions at room temperature. Enter the known oxygen concentration the sensor is
exposed to in the Oxygen field. Enter the known temperature reading from the oxygen sensor in the Temperature field. (The NeoFox temperature reading can be observed on the main NeoFox
Viewer window.)
7. Once the sensor is in stable condition with a stable Tau value, click on Use Current Tau.
8. To use this calibration for oxygen measurement, click on the graph, then click the button. A Download Confirmation screen appears.
9. Click Yes to confirm the download and continue.
013-20000-009-02-201203 31
3: Using NeoFox with NeoFox Viewer Software
Si
ngle Point Reset Control
Newer versions of the NeoFox hardware have a S Single Point Reset without a computer. This con unit. The control button itself is inside the recession so you will need a pointed object to reach it.
Before beginning a Single Point Reset using the hardware control, you must tell the device what percen oxygen to calibrate against once you are in the field. By default, NeoFox will use the last percent o value from a Single Point Reset, so the following procedure simulates a Single Point Reset to save that value.
ingle Point Reset control that allows you to perform a
trol is located in the center of the back of the NeoFox
xygen
Note
You must perform this setup procedure every time you change any of the device’s calibration settings because the Viewer may reset the SPR values to their defaults.
Procedure
In the lab:
1. Connec
2. Select Options | Calibration to open the calibration screen.
t NeoFox to a computer and open the NeoFox Viewer software.
t
3. If you have not already downloaded your multi-point calibrat
Calibration button, then clicking the Download button.
4. Click the Single Point Calibration button. A new tab will appear at the top of the window.
5. Click the new Single Point Calibration tab.
6. In the Oxygen field, enter the percent oxygen
the field. For most users, this will be ambient ATM.
7.
You do not necessarily need to enter tau and temperature to actually perform the single point reset, b
8. Click the Download button.
9. Verify the Single Point Reset then following the field proce value.
In th
1. Place you
ut you should do so if possible.
oxygen value was saved correctly by cycling power to the device,
dure below and verifying that the oxygen changes to the set oxygen
e field:
r temperature and oxygen probes into your single point reset conditions.
that you will be using for your single point reset in
air and the percent oxygen will be 20.9% of 1
ion, do so now by clicking the Open
32 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
2. Wait long enough for the temperature and tau values to settle – this amount of time will vary based on the probe type and the number of averages parameter. You should wait at least one minute.
Push the
3. Single Point Reset button for at least five seconds. The Single Point Reset control button is recessed in the center back of the NeoFox unit.
After five seconds, NeoFox will use the following three values to perform the single point reset:
Tau will be the current value of Tau measured from the probe.
Temperature will be either the current temperature reading from
fixed temperature depending on which temperature source you h If no temperature is selected, the SPR will not occur.
Percent oxygen will be the value that was set during the setup phase in the lab.
Once the Single Point Reset has been performed, NeoFox saves your settings. This m of the parameters listed in Saved Settings
will be saved in their current states.

Saved Settings

The following parameters are save are saved, the saved values will be loaded when the device turns on:
Measurement Parameters:
Duty Cycle Seconds o
Duty Cycle Seconds off
Number of Averages
Oxygen Units
Salinity Correc
Calibration Settings:
tion Factor
d onboard the NeoFox in its non-volatile memory. When the settings
n
the temperature probe or the
ave selected for your device.
eans that all
Temperature Se
Temperature Sourc
Fixed Temperature V
Calibration Method
Sensor Type
Single Point R
Single Point – Tau
Single Point – Oxyg
Single Point – Tempera
013-20000-009-02-201203 33
ttings
e
alue
eset Settings
en
ture
3: Using NeoFox with NeoFox Viewer Software
Two Point Calibration Settings
Two Point – Tau0
Two Point – Tau1
Two Point – Oxygen 0
Two Point –
Oxygen 1
Two Point – Slope
Two Point – Offset
Two Point – Temper
Multipoint Calibration Settings
12 Multipoint Original Calib
T1, T2
12 Singl
e Point Adjusted Calibration Coefficients: A0, A1, A2, B0, B1, B2, C0, C1, C2, T0,
T1, T2
n tings:
alog Output SetA
Analog Voltage Ou
tput So
Analog Voltage Lower Bound
Analog Voltage Upper Bound
Analog Current Output
Analog Current Lower Bound
Analog Current Upper bound
CJaz ommunication:
ature
ration Coefficients : A0, A1, A2, B0, B1, B2, C0, C1, C2, T0,
urce
Source
rfac
Enable Jaz Inte
e
Serial and USB Communication:
RS232 Divisor Latch
RS232 Divisor Add V
alue
RS232 Divisor Mul Value
RS232 Enable
Uart Data Copy
Type
Uart Data copy Mode
Uart Transmission Mod
e
Uart Send Alerts
34 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software

Saving Data

You can save your NeoFox in a .csv file, which is comp it to stop or the application is shut down.
Procedures for saving both types of files are described below. See Data Logging for more information on saving a .csv file.
data in either a .fxy file, which can be opened in a NeoFox Viewer window or
atible with Microsoft Excel. NeoFox saves data to a disk file until you request
Procedure
To save the measure
1. Select File | Save. A window appears displaying a file name with the date and the time the file
st saved.
was la
2. Select the desired location to save the file.
ment data in a .fxy file,
Procedure
To save the measure
1. Click the Start/Stop Data Logging… button from Panel screen appears.
ment data in a .csv file,
the main screen. The Data Logging Control
2. Click the Start Logging button.
3. Select the desired location to save the file.
4. Specify the logging interval.
013-20000-009-02-201203 35
3: Using NeoFox with NeoFox Viewer Software
5. Choose the data columns to export, either Standard or Extended (wide range of data). Standard (recommended setting) includes the following:
TimPressure Temperature Tau % O2 Conv
6. Click Save Setting
e stamp (to tenth of a second)
erted O2
s and Close Window to save the data.

NeoFox Communication Interfaces

You may wish to write programs to interface directly to the NeoFox withou There are a number of ways to do so: Through the NeoFox DLL, through the NeoFox RS232 port (only NeoFox-GT units with an communications), or through custom USB drivers. See the NeoFox Communication Interfaces
Engineering Note for more information.
F or later in the third digit of its serial number support RS232
t using the NeoFox Viewer.

Saving/Restoring Chart Layouts

NeoFox Viewer allows you to save and restore the lay can be saved and then restored:
out of the displa
yed charts. The following settings
X-axis duration
Vertical height se
Y-axis scale settings
Type(s) of data being
ttings
displayed on multiple plotting areas within the chart window

Saving a Chart Layout

Procedure
1. Select View | Save Graph Layout
2. Specify the name of the file where the settings will b
Res r
to ing a Chart Layout
Procedure
1. Select View | Restore Gra
ph Layout from the menu.
from
the menu.
e saved.
36 013-20000-009-02-201203
3: Using NeoFox with NeoFox Viewer Software
2. Enter the name of the file containing the saved setting(s).

Saving Settings

Procedure
To save these values in
1. Navigate to the Advanced Settings sc
2. Set the Flash Write parameter to Enabled. It
reverting to Disabled.
3. rough the NeoFox Sport, select Save Settings from the O2 Setup Menu.
To save these settings th
the NeoFox Viewer,
reen.
will remain displayed as Enabled briefly before
Note
When the device saves a parameter through a rite, it must save ALL of its savable settings at the same time. Therefore, actions which ma r to save specific settings to memory will actually save the entire set of savable parameters. Those actions include downloading calibration data, performing a single point reset, or manually prompting a flash write by setting the Flashed
Write parameter to Enabled in the Advanced Settings Panel.
flash w
y appea

Resetting Saved Settings

You can reset the saved settings to their origin
al state.
Procedure
1. Open the device in the NeoFox Viewer and navi
2. Set the value of the Flash Reset parameter to Enabled. This value may revert to disabled
thereafter. When the device is reset, the saved values revert to their factory settings.
3.
Download your calibrations again after resetting the flash memory.
gate to the Advanc
ed Settings screen.
shortly
Note
For devices with firmware version 4.00 or above, the APD Bias Voltage (displayed as DAC-APD Gain) setting value persists through a flash reset.
013-20000-009-02-201203 37
3: Using NeoFox with NeoFox Viewer Software
38 013-20000-009-02-201203
Chapter 4

Using NeoFox Sport

Overview

The NeoFox Sport is a field-portable extension of the NeoFox oxygen measurement system. Users of the NeoFox Sport have the ability to make real-time measurements in the field without a computer or external power supply. To do this, the NeoFox Sport system uses the same NeoFox measurement hardware as the desktop NeoFox system, but also includes a battery (to supply power) and a data processing unit (to provide a user interface).
Because the NeoFox Sport uses the same measurement hardware as the standalone NeoFox, user experience such as setup and measurement are common between both systems. Both systems can be connected to a PC via USB and controlled through the NeoFox Viewer software. In fact, the user interface for the desktop NeoFox and the NeoFox Sport are exactly the same.
This chapter contains information for using NeoFox Sport without the NeoFox Viewer software.

NeoFox Sport Features

Additional features that are available to NeoFox Sport users include:
Battery-powered operation.
Real-time display of temperature and oxygen in various measurement units:
Gaseous oxygen partial pressure as a percentage of 1 ATM.
Gaseous oxygen partial pressure in Torr.
Dissolved oxygen partial pressure in ppm (parts per million)
Dissolved oxygen concentration in µm/L (micromoles per liter)
Gaseous oxygen concentration as a percentage of total pressure in Mole%
Atmospheric pressure from the on-board pressure sensor (in units kPa)
Data logging of temperature, oxygen and pressure readings in units listed above.
In-field “single point reset” (also available on desktop NeoFox)
013-20000-009-02-201203 39
4: Using NeoFox Sport

NeoFox Sport Menu

40 013-20000-009-02-201203
4: Using NeoFox Sport
013-20000-009-02-201203 41
4: Using NeoFox Sport
Menu Tree Description
Menu Item Description
Sensor Readout
Enables you to view the sensed values for Time, O2, temperature and Tau. The values displayed refresh every time a new sample is uploaded from NeoFox with the appropriate values.
Prs Displays the atmospheric pressure measured by the on-board pressure sensor on the
NeoFox Sport.
O2 Displays the oxygen unit label and the current oxygen value.
Temp (°C) Displays the algorithm temperature in degrees Celsius
Tau (µs) Displays the value of tau (fluorescence lifetime)
Sensor Setup
Single Pnt
Reset
Allows you set the parameters shown below.
Allows you to set the following parameter settings to be used when a single point reset is performed:
Mole %: Default = 20.90%; Range = 000.00 – 100.00%
Temperature: Default = algorithm temperature; Range = 000.00 – 200.00 °C
SPR – Pressure (kPa): Default = On-board pressure sensor
Tau: Default = current value of Tau; Range = 00.000 – 50.0000
A message displays informing you that the single point reset has completed.
NOTE: NeoFox must be configured to Single Point mode through NeoFox Viewer.
Data Logging Allows you to enable datalogging to an SD card, set the datalogging interval and see
the filename used for the stored data. Default = current datalog interval. Range = 1000 – 99,999,999 ms
Oxygen Units Enables you to set the units in which the O2 data is displayed. Choose from %
%, ppm , torr and µmol/L, mole%
PPO
2
Temperature Allows you to set the source of the temperature, either none, the sensor temperature
or a fixed temperature in degrees Celsius (Default = current value of fixed temperature variable; Range = 0.0 – 200.0 °C).
Pressure Allows you to set the source of the pressure: either none, the sensor pressure or a
fixed pressure in kPa (Default = current value of fixed pressure variable; Range = 0 – 200 kPa)
Averaging Set the running average of the measurements. Default = current value of variable.
Range = 1– 300.
Duty Cycle Set the duty cycle time in seconds. Range = 1 – 999. Set Off time to 0 to make it
always run.
42 013-20000-009-02-201203
4: Using NeoFox Sport
Menu Item Description
Flashing On/Off Set the flashing status of the probe LED.
Save Settings Select Accept to save your settings or Cancel to discard them.
Sensor Info
Displays sensor information: module ID, FPGA, controller and PCB revision level.

Setting Up the NeoFox Sport

Before using the NeoFox Sport in the field, you must download the appropriate sensor calibration to the NeoFox device. This must be done with your NeoFox device connected to a computer.
Procedure
1. Connect the power supply to the DPU power jack (not the NeoFox auxiliary power jack).
2. Press the red power button (
3. Connect the USB cable between the USB port on the NeoFox module (not the DPU) and the computer.
4. Start the NeoFox Viewer software. See Chapter 3: Using NeoFox with NeoFox Viewer Software
5. Perform a calibration. See Calibrating the Probe
6. Select the sensor type as appropriate.
7. Verify that the calibration has been downloaded to memory by cycling power to the device and then taking a measurement.
) to turn the device on.
.
.
Note
Once you are in the field, you will not be able to update this calibration. However, you will be able to do a Single Point Reset.
013-20000-009-02-201203 43
4: Using NeoFox Sport
The following additional steps must be taken once after a flash reset has been performed. These steps are required to let the device know that it should communicate with the NeoFox Sport DPU.
Procedure
1. In the NeoFox Viewer software, Select View | Advanced Settings.
2. Click on the Advanced Settings tab in the lower right-hand corner of the screen.
4. Scroll down to the Enable Jaz Interface parameter and set it to Enabled.
5. Set the Flash Write parameter to Enabled. It will remain displayed as Enabled briefly before reverting to Disabled.

Performing a Single Point Reset in the Field

Once the calibration has been downloaded and the device is in the field, you should always perform a Single Point Reset before beginning a measurement.
44 013-20000-009-02-201203
4: Using NeoFox Sport
CAUTION
If any settings in the experimental setup change (APD gain, voltage, probe geometry, etc.), it is strongly recommended that you do a Single Point Calibration to offset these changes. This includes any gain setting changes or switching between electronics units.
Procedure
1. On the NeoFox Sport, navigate to the Oxygen Sensor menu. This is the first screen that displays when you power-on the unit.
Note
If there is an SD card with a boot application in one of the device’s SD card slots, the device will
boot to that application. In this case, simply navigate to the root menu and select Oxygen Sensor
to bring up the correct oxygen sensor home screen.
2. Navigate to Sensor Setup | Single Pnt Reset.
3. Place your probe and temperature sensor in your sample and press the Accept button (
4. Proceed through the Single Point Reset screens. A message appears to tell you that your single point reset has completed.
).

Data Logging / Saving Data

NeoFox Sport provides you with the following functionality for logging/saving data:
Log data to an SD card
View the file name of the data log
Set the data logging interval

Logging Data

The NeoFox Sport allows you to log data to an SD card via the Oxygen Sensor | Sensor Setup | Data
Logging menu. When you select Turn datalog on from this menu, a log file is created on the SD card
and NeoFox Sport begins logging data automatically. The name of this log file is automatically generated when the data log begins in the format nf_xxx.tsv, where xxx is a 3-digit number for a filename that does not already exist in the NeoFox directory.
When you select Oxygen Sensor | Sensor Setup | Data Logging | Turn data log off, then the file will be
closed. A new file is created the next time you select Turn data log on.
013-20000-009-02-201203 45
4: Using NeoFox Sport

Viewing the Data Log File Name

You cannot set the file name, but you can view the name of the current log file by selecting Oxygen Sensor | Sensor Setup | Data Logging | Data log filename.

Setting the Data Logging Interval

You can also set the data logging interval via the Oxygen Sensor | Sensor Setup | Data Logging | Data log interval menu selection. This can be used to reduce the size of the log file. The minimum data log interval is 1 second.

Data Log File Format

The datalog file is formatted as a tab separated value file. This file can be easily imported into Microsoft Excel or comparable spreadsheet programs. A single sample in the data log is formatted as follows.
millisecond count <tab> upload date <tab> upload time <tab> oxygen <tab> oxygen unit label <tab> algorithm temperature <tab> sensor temperature <tab> tau <tab> algorithm pressure <tab> sensor pressure
Where: Millisecond Count = The millisecond counter of the NeoFox device at the time of the actual sample. Because this value is stored at the actual sampling time, this value is a more reliable timing metric than the “upload date” and “upload time” fields.
Upload Date / Upload Time = These fields represent the date and time that the sample was uploaded to the DPU.
Oxygen = The oxygen measurement (in whatever units are currently selected).
Oxygen Unit Label = The oxygen unit label; either “ppm”, “%”, “Torr”, or “um/L”
Algorithm Temperature = The temperature that was used to perform the conversion from Tau to Oxygen.
If the NeoFox’s temperature source parameter is set to:
Sensor temp, then this is the sensor temperature. Fixed temp, then this is the fixed temperature value. None, then this is 200.
Sensor Temperature = The sensor temperature (in degrees C).
Tau = The tau value (in microseconds).
Algorithm Pressure = The temperature that was used to perform the conversion from partial pressure of
Oxygen to concentration or mole%.
If the NeoFox’s pressure source parameter is set to:
Sensor Pressure, then this is the sensor pressure. Fixed Pressure, then this is the fixed pressure value. None, then this is 101 – 325.
Sensor Pressure = The pressure (in kPa).
46 013-20000-009-02-201203
4: Using NeoFox Sport

NeoFox Sport Saved Settings

Most of the parameters that can be set through the NeoFox Sport’s user menus can be saved to the NeoFox Sport module’s flash memory, which allows those settings to be restored the next time that the device is powered on. See Saved Settings
To save the flash write settings, navigate to the Sensor Setup | Save Settings menu. Note that this saves
ALL of the savable parameters in their current state. Also, performing a Single Point Reset through the DPU menu may result in all of the savable parameters being overwritten.
for a list of settings that can be saved.

Low Battery Detector

The NeoFox Sport contains a low battery detector that prevents inaccurate readings in low battery conditions. When the device detects that battery conditions are too low for reliable operation, its tau and oxygen measurements will be set to 0 or -1. NeoFox disables its measurements shortly before the device itself turns off.
013-20000-009-02-201203 47
Appendix A

Specifications

Specification Value
Parameters Measured Gasseous oxygen partial pressure, dissolved oxygen concentration,
temperature, pressure (via onboard pressure transducer)
Control Software Windows 2000/XP/Vista (32 and 64-bit) and Windows 7 (32-bit) control
software with data logging capability; controls include: modulation frequency, data rate, signal averaging, APD gain, analog gain, LED intensity
Sampling Rate Acquires data samples every 0.1 second
NOTE: This is not the response speed of the probe, just the rate at which the probe is digitally sampled
Thermistor Probes Closed-end stainless steel tube with thermistor sensor mounted in tip; liquid
immersible rugged design; 1/8” NPT fitting; temperature range 0 to 75 °C
Pressure Measurement On-board pressure transducer monitors atmospheric pressure
Input Power Approximately 2.5W @ 5V
Communications USB, 0-5V analog output; 4-20mA analog output. RS232 is available on the
NeoFox GT model with an F or greater in the third digit of its serial number.
NeoFox Sport
Upload Interval The NeoFox Sport uploads data from the NeoFox module at least once per
second in the Sport configuration. Alternate configurations may have lower data rates.
Battery Life In the NeoFox Sport configuration, a fully charged, gently used battery will
last for approximately three hours.
013-20000-009-02-201203 48
4: Using NeoFox Sport
013-20000-009-02-201203 49

Updating Firmware

Do not unplug the device while the firmware is updating.
Caution
Appendix B

Maintenance

Do not download a firmware version that predates the version you received with your NeoFox unit.
Procedure
1. Download the new NeoFox firmware image from
www.oceanoptics.com/technical/softwaredownloads.asp
2. Run Neofox Viewer and select Options|Firmware Update.
3. In the firmware update dialog, select Download, then choose the file that you just downloaded.

Updating Software

You can overwrite the previous installation.
Procedure
1. Download the installer for the latest version of the Neofox Viewer from the Ocean Optics Software Downloads page:
2. Unzip the directory and then run the “Setup.msi” installer wizard.
013-20000-009-02-201203 50
www.oceanoptics.com/technical/softwaredownloads.asp
B: Maintenance

Troubleshooting

Problem Possible Cause Suggested Solution
Clicking Download during a Two-Point Calibration does not change the measured values of oxygen.
At startup, the device’s controls appear disabled, no values display, or the device shows other signs of not uploading data.
During hardware installation, the Found New Hardware Wizard” can’t find the driver or .sys file for the NeoFox.
After running a VI that calls the Neofox DLL, LabView crashes when that VI is closed.
The Two-Point Calibration was not performed correctly.
After scanning the tau values a single click on the graph area was not performed
The NeoFox drivers and firmware versions are incompatible.
The NeoFox directory isn’t included in search.
The VI passes a function in the library an uninitialized string.
Follow the Two-Point Calibration
instructions in this manual exactly as written.
Update the NeoFox firmware and
NeoFox Viewer software. See Error!
Reference source not found., as
well as Updating Software
information.
When prompted, choose to install from a specific location, then include the NeoFox installation directory in your search. This directory is located at “C:\Program Files\Ocean Optics\Neofox Viewer” by default. The drivers can be found in the USB Drivers directory located here.
The library functions generally assume that any strings passed are already initialized. Therefore, be sure to pass only initialized strings. Better error handling for this error was added in DLL version 1.89, but you should still use this convention to avoid unnecessary complications.
for more
After setting the display units to anything other than ppm, the display will not return to percent oxygen.
Percent oxygen is above 100 or below 0.
A fatal error or critical error warning message appears at startup and prompts to continue or exit the application. After continuing, the FFT view does not seem to work.
013-20000-009-02-201203 51
Previous software version. This problem was resolved in
version 1.92 of the Neofox Viewer. Contact Ocean Optics if you see this error in a more recent software version. Previously, this problem was solved by reinstalling the NeoFox Viewer software.
Sensor probe has been calibrated incorrectly.
This problem seems to occur when the application starts with the FFT View Window active.
Recalibrate the sensor probe.
Select Continue, then select the waveform or decay form view. Close the application and then restart. The problem should not occur again. If the problem persists, reinstall the NeoFox Viewer software.
B: Maintenance
Problem Possible Cause Suggested Solution
A Single Point Reset was performed, but the %O
reading
2
appears to be incorrect, even when the same Tau value is used in the Single Point Reset.
The HIOXY automatic frequency setting defaults to 45 KHz instead of 100 KHz.
HIOXY calibrations performed prior to 31 AUG 2009 yield incorrect results on NeoFox devices with updated firmware.
The temperature probe reading appears erratic or is drastically incorrect.
When performing the Single Point Reset, the temperature may have been set to 0 even though a valid temperature was entered in the Temperature field. This is a known issue with NeoFox Viewer release 2.10 when used with Firmware Version
2.0401.
In Firmware Releases 2.03 and greater, the HIOXY probe’s automatic frequency setting has been changed from 100 KHz to 45 KHz. Using HIOXY probes calibrated at 100 KHz will produce incorrect results. Using older NeoFox units with new calibrations will also produce incorrect results.
Known issue with units manufactured prior to August
2009.
After entering the desired value in the Temperature field, click in one of the other fields (Oxygen or Tau) before clicking the click the
button.
If this does resolve the issue, retry the Single Point Reset.
If your HIOXY probe was calibrated before 01 SEPT 2009, return it to Ocean Optics for a full recalibration at 45 KHz. Or, you can use firmware version 2.02 available here:
http://www.oceanoptics.com/ technical/neofoxsupport.asp
. Or, you can enter the frequency setting manually, although this is not recommended.
Contact Ocean Optics Technical Support.
The waveform windows or settings / status windows are displayed as popups instead of being docked on the side of the Main window
N/A All utility windows can be docked or
undocked by double-clicking the title bar of the utility window.
52 013-20000-009-02-201203
B: Maintenance
013-20000-009-02-201203 53

Analog Output

Analog Output Connector

The Analog Output connector is located on the rear panel of the NeoFox unit.
Appendix C
There are two user programmable analog outputs available on the NeoFox as follows:
4-20 mA current source
0-5V voltage source
Analog threshold settings can be saved by the user to allow a NeoFox unit to operate in a stand alone mode, outputting analog signals without input from a computer.

Configuring Analog Voltage Output Channel with NeoFox Viewer

Procedure
1. Select the Advanced Settings tab on the Advanced View screen to view the advanced settings.
013-20000-009-02-201203 54
C: Analog Output
2. In the Advanced Settings tab, select an appropriate data source from the dropdown box located to the right of the “Voltage Output Source” setting. For more info on valid sources, please see
Selecting a Data Source
3. Select a valid lower bound for your represented data range by typing a number into the box located to the right of the “Analog Voltage Lower Bound” setting. In some versions of the Viewer software, this setting may be called “Voltage Limit Minimum”. For more information on selecting a lower bound for your voltage, see Selecting a Data Range
4. Select a valid upper bound for your represented data range by typing a number into the box located to the right of the “Analog Voltage Upper Bound” setting. In some versions of the Viewer software, this setting may be called “Voltage Limit Maximum”. For more information on selecting an upper bound for your voltage, see Selecting a Data Range
You can save your settings by enabling the “Flash Write” setting momentarily.
.
.
.

Configuring the Analog Current Output Channel with NeoFox Viewer

Procedure
1. Select the Advanced Settings tab on the Advanced View screen to view the advanced settings.
013-20000-009-02-201203 55
C : Analog Output
2. Select an appropriate data source from the dropdown box located to the right of the Current Output Source setting. For more info on valid sources, see Selecting a Data Source
3. Select a valid lower bound for your represented data range by typing a number into the box located to the right of the “Analog Current Lower Bound” setting. In some versions of the Viewer software, this setting may be called “Current Limit Minimum”. For more information on selecting a lower bound for your voltage, see Selecting a Data Range
4. Select a valid upper bound for your represented data range by typing a number into the box located to the right of the “Analog Current Upper Bound” setting. In some versions of the Viewer software, this setting may be called “Current Limit Maximum”. For more information on selecting a upper bound for your voltage, see Selecting a Data Range
You can save your settings by enabling the Flash Write setting momentarily.
.
.
.

Selecting a Data Source

There are six potential data sources that can be driven by the NeoFox. The pressure mode is not currently supported as an analog data source.
56 013-20000-009-02-201203
C: Analog Output
Mode Meaning
0 Off
This mode does not actually turn the current source off. Instead it sets the current source to 4 mA and the voltage source to 0 V. It does not reduce the unit’s power draw.
1 Oxygen
This is the standard percent oxygen mode. The data source will be percent oxygen, not a converted unit such as ppm or torr. See Mode 5 for more information on these alternate units.
2 Temperature
Temperature (in degrees Celsius)
3 Pressure
Not yet supported
4 Tau
In microseconds
5 Alternate O2 Units
In this mode, the dimensionless value of the alternate O2 value is compared with the lower and upper bound parameters.

Selecting a Data Range

The data ranges of the analog outputs are fully adjustable through the lower bound and upper bound settings in NeoFox Viewer. These settings are simply unitless values that are compared to the unitless value of the data source.
Analog Voltage Boundaries
The analog voltage lower bound represents the value of the data source that corresponds to a 0 volt analog output. The analog voltage upper bound represents the value of the data source that corresponds to a 5 volt analog output. Intermediate values are linearly scaled across this region to produce the appropriate output. Values of the data source that fall outside this region result in output voltages that are either 0 volts or 5 volts depending on whether the value of the data source exceeds the upper bound (resulting in 5 V) or falls short of the lower bound (resulting in 0V).
013-20000-009-02-201203 57
C : Analog Output
Analog Current Boundaries
The analog current lower bound represents the value of the data source that corresponds to a 4 mA analog output. The analog current upper bound represents the value of the data source that corresponds to a 20 mA analog current output. Intermediate values are linearly scaled across this region to produce the appropriate output. Values of the data source that fall outside this region result in output currents that are either 4mA or 20 mA depending on whether the value of the data source exceeds the upper bound (resulting in 20 mA) or falls short of the lower bound (resulting in 4 mA).
Analog Data Range Example
As an example, consider the following. A user who is interested in using the analog voltage output to monitor oxygen in the range from 4.5 ppm to 7 ppm would select the following settings. The result would be that any ppm values below 4.5 ppm would result in an output of 0 volts and any values above 7 ppm would result in 5 volts. In this case, 6.3 ppm would produce an analog voltage output of [(6.3 - 4.5) / (7 -
4.5)] * 5 = 3.6 volts.
Analog Voltage Source: 5 – Alternate O2 Units
Analog Voltage Lower Bound: 4.5
Analog Voltage Upper Bound: 7.00
(Oxygen Units) – “DO – ppm – Water”. (Selection of ppm as alternate O2 units)

Saving the Analog Output Settings

To save the analog output settings, open the “Advanced Settings” control panel and then select “Enabled” from the dropdown box located to the right of the “Flash Write” setting. This box will momentarily continue to display “enabled”, but will switch back to “disabled” when the unit has processed the command.

Analog Output Pinouts

Caution
Be careful when wiring. Incorrect wiring could cause permanent damage to the unit.
View of Connector on Back of NeoFox Unit
1 = RS232 Rx 2 = 0-5V – V+ 3 = Current 4-20 - V- (Iout) 4 = Current 4-20 - V+ 5 = RS232 Tx 6 = 0-5V / RS232 GND
This is the view looking at the connector on the back of the NeoFox, with power plug on the right and the USB plug on the left. The connector is a standard PS-2 connector.
58 013-20000-009-02-201203
C: Analog Output

Analog Output Parameters List

NeoFox Viewer Parameter Potential Values
Voltage Output Source 0 – Disable analog output
1 – Oxygen
2 – Temperature (not supported)
3 – Pressure (not supported)
4 – Tau
5 – Alternate O
Current Output Source 0 – Disable analog output
1 – Oxygen
2 – Temperature (not supported)
3 – Pressure (not supported)
4 – Tau
5 – Alternate O
2
Units
2
Units
Analog Voltage Lower Bound (or Voltage Limit Minimum)
Analog Voltage Upper Bound (or Voltage Limit Maximum)
Analog Current Lower Bound (or Current Limit Minimum)
Analog Current Upper Bound (or Current Limit Maximum)
<floating point value> - This floating point value represents
the “lower bound” of the represented range of the input data. Must be < UBOUND and 0.
<floating point value> - This floating point value represents
the “upper bound” of the represented range of the input data. Must be > LBOUND.
<floating point value> - This floating point value represents
the “lower bound” of the represented range of the input data. Must be < UBOUND and 0.
<floating point value> - This floating point value represents
the “upper bound” of the represented range of the input data. Must be > L

Analog Output Specifications

Name Min Typ Max Conditions
4-20 mA Loop Voltage 11 V -- 24V R
4-20 mA Output Accuracy -- < 160 μA -- 25°C, V
Loop
= 2
Loop
= 12V, R
Loop
= 2
0-5V Output Accuracy -- < 50mV -- 25°C
0-5V Output Resistance -- -- 250 --
013-20000-009-02-201203 59
C : Analog Output
A

Interfacing to the 4-20mA Output

The NeoFox includes a non-isolated 4-20 mA current sink that can be used to transmit oxygen, temperature, or tau data. It is extremely important that you set up the interface for this output correctly, or else irreparable damage to the NeoFox unit can occur.
The figure below shows a complete 4-20 mA loop and its 3 main components: the NeoFox unit, an ammeter, and a voltage supply. In the diagram, the NeoFox unit is shown on the left and the customer’s interface circuitry (voltage supply and ammeter) is on the right. Although they are drawn close together, the NeoFox unit and the customer’s interface circuitry may be physically located far apart, connected by a 2-wire cable.
PIN 4
Loop Voltage Supply
Neofox Unit
Ammeter
PIN 3
Typical 4-20 mA Interface

4-20 mA Loop Voltage Supply

As a simple current sink, NeoFox does not provide its own power for the current loop. The user must provide an isolated power supply between 11 V and 30 V (measured at the input to the NeoFox unit) to
power the loop. This power supply must be the grounds of the Neofox and the supply cannot be connected together
Most customers will use separate AC adapters to power their NeoFox unit and their interface circuitry. Usually this will result in the two circuits floating with respect to each other so that the circuit will work properly. However, it is important to note that NeoFox has a grounded metal case that is connected to its circuit ground and its USB shell. Hence, customers who connect the NeoFox’s case ground to their interface circuit’s case ground will find that the measurements are not correct. Some common situations that could lead to unintended ground connections are the following:
completely separate from the NeoFox and its power supply;
.
60 013-20000-009-02-201203
C: Analog Output
Mounting both devices on a common metal surface (such as a DIN rail)
Connecting both devices to a single computer or USB hub
Using the same measurement device to measure the 0-5V output and the 4-20mA output. (The 0-
5V output is referenced to the NeoFox circuit ground.)
Using the same measurement device to communicate with NeoFox via RS232 and to measure the
output of the 4-20 mA output. (RS232 is only available in NeoFox GT models with an F or greater in the third digit of its serial number, which are available by request).
Users should be aware of any possible connections between NeoFox and their interface circuits when setting up their hardware.

4-20 mA Isolation

Some users (especially those in industrial settings) may need to use an isolated 4-20 mA loop. Unfortunately, the current sink on the NeoFox unit is not isolated and is not suitable for use in environments which may cause the loop voltage to exceed 30V. However, there are a number of third­party products available that can convert the output of the NeoFox into an isolated 4-20 mA loop transmitter.
Some suitable products include:
LPI-1
Manufacturer: A Pi (
Suggested Vendor: Mod-tronic. (
Part Number: LPI-1
Datasheet:
931S-A1A1N-IP1
Manufacturer: Rockwell Automation
Suggested Vendor: Varies by geographic region
Part Number: 931S-A1A1N-IP1
Datasheet:
http://www.ab.com/en/epub/catalogs/12768/229240/229268/10202439/10202445/tab3.ht ml
931S-C1A2D-OP*
Manufacturer: Rockwell Automation
Suggested Vendor: Varies by geographic region
Part Number: 931S-C1A2D-OP
http://www.api-usa.com/pdf/api/api_lpi-1_lpi-2.pdf
http://www.api-usa.com)
http://www.mod-tronic.com)
Datasheet:
http://www.ab.com/en/epub/catalogs/12768/229240/229268/10202439/10202445/tab9.ht ml
*This part can be used to convert the 0-5V output of the NeoFox into a 4-20mA signal.
013-20000-009-02-201203 61
C : Analog Output

Resistor / Volt Meter

If an ammeter is not available, as is the case with many data acquisition systems, a voltmeter can be used instead (see following figure). To use a voltage measurement system, users should select a low­impedance resistor and a high-impedance voltage measurement instrument.
Be aware that there will be a voltage drop across the resistor that “substracts” from the loop voltage supply. The voltage drop can be calculated as resistance times current. For example, a 100 Ohm resistor at 20 mA would have a 2 volt drop across it. If the power supply was 12 V, then the loop voltage would actually measure under 10V and the system would be outside of specification (11V is the minimum specified loop voltage). It is important to select a resistor that will not “subtract” too much from the loop voltage.
62 013-20000-009-02-201203
C: Analog Output
013-20000-009-02-201203 63
Appendix D

Calculating Drift Per Sensor

Drift differs depending on the sensor formulation (FOXY, FOSPOR or HIOXY). See the table and example below.
Stability (Continuous LED):
Lifetime Stability (Tau): Oxygen Stability %: 0.01% HR 0.005% HR 0.007% HR
Example: Calculate drift for a FOXY sensor probe while measuring a sample every 2.5 hours for 100
hours with the LED being lit for 1 minute:
Newdrift=reporteddrift*dutycycleon/dutycycleoff 
Newdrift=0.01%/hr*(1min/(2.5hr*60min/hr))2.5*60convertshoursintominutes, soastocanceloutthedimensions Newdrift=0.01%/hr*(1min/150min) Newdrift=0.01%/hr*(.0066) Newdrift=0.000066%/hr
To run the experiment for 100 hours:
Newdrift=0.000066%/hr*100hrs  Newdrift=0.006%
FOXY Formulation FOSPOR Formulation HIOXY Formulation
0.0006 usec/ HR 0.003 usec/ HR 0.0002 usec/HR
013-20000-009-02-201203 64
C: Analog Output
013-20000-009-02-201203 65
A
Adobe Acrobat Reader, 4 advanced settings, 19 Advanced Status tab, 22 analog current boundaries, 57 analog data range selection, 57 analog output, 53
connector, 53 parameters, 58 pinouts, 57 saving settings, 57 specifications, 58
Analog Output
configuring current output channel, 54
configuring voltage output channel, 53 analog voltage boundaries, 56 APD voltage, 24 applications, 3
B
battery detector, 46
C

Index

data source
selecting, 55
document
audience, vii purpose, vii summary, vii
E
enter new event, 17 Enter New Event field, 18 equipment
included, 3 recommended, 4
event
entering new, 18
F
features, 2
NeoFox Sport, 39 fiber, 5 firmware
updating, 49 FOSPOR, 3 FOXY, 3
calibration, 5
multipoint, 25 probe, 25 two-point, 27
chart layout
restore, 36 save, 36
Communication Interfaces, 36
hardware, 10 hardware drivers
installation, 12
Windows Vista, 14
Windows XP, 12 HIOXY, 3
D
data
saving, 35 data log file format, 45 data logging, 18
NeoFox Sport, 44 data range
selecting, 56
013-20000-009-02-201203 66
Install
from CD, 8 from Web, 8
installation
hardware, 10 hardware drivers, 12 software, 7
H
I
isolation
4-20 mA loop, 60
C: Analog Output
S
log, 18 loop voltage supply, 59
menu
NeoFox Sport, 40 menu description
NeoFox Sport, 41
NeoFox
overview, 1 NeoFox Sport, 39
data logging, 44
features, 39
menu, 40
menu description, 41
saved settings, 46
setup, 42
single point reset, 43 NeoFox Viewer, 3, 19 new event
entering, 18
L
M
N
safety warnings, i salinity correction, 23 save
chart layout, 36
saved settings, 33
NeoFox Sport, 46
sensor
oxygen, 2, 4 select type, 24
setup
NeoFox Sport, 42 Setup tab, 20 shipment components, 3 single point reset
NeoFox Sport, 43 Single Point Reset
control button, 32 single-point reset, 30 software, 3, 7, 19
NeoFox Viewer, 3
updating, 49 Software and Resources Library CD, 4 specifications, 47 Status tab, 21
T
O
oxygen sensor probes, 2, 4
P
probe
calibrating, 25 oxygen sensor, 2, 4 temperature, 4
product-related documentation, vii
R
recommended equipment, 4 RedEye, 4 reset
single point, 30
restore
chart layout, 36
013-20000-009-02-201203 67
tab
Advanced Status, 22
Setup, 20
Status, 21 troubleshooting, 50
U
updating, 49
W
what's new, vii
Loading...