SYSTEM SERVICE TEST RECORD -- FORM 250
Red LED flashes and Alarm sounds three times
Green LED on front panel and power switch are lit
All fields blue except Air Trap
Roller Pump Door changing from Blue to Red when opened.
Check Prime Switch changing from Blue to Red when pressed
no Air Trap Level changing from Red to Blue when filled
Set time and date are to local time
TP400 set at 37°C. Display shows value from 36.9 – 37.1°C
With T1 and T2 connected, system immediately goes into RUN
no Audible alarm and message when T2 disconnected
no Yellow screen appears after pressing Standby/Run button
no System goes into RUN mode after selecting “Yes”.
no Pump is turning counter clockwise
no Coolant is circulating within Coldwell
Target temperature is set to 36.5 °C
no The rate is set to MAX
“Pump Door Open” message on screen
no Audible alarm and system goes into STANDBY
no Mute button silenced alarm
When Air Trap is removed, message display is
“AIR TRAP FAULT – Check Saline Level”
Verify the Target temperature is set to 38.0°C
When the Rate Deg/Hr button is pushed in Standby Mode,
following modes displayed: Max Power/Controlled Rate/ Fever
Select the Controlled Rate Mode and verify that the rate
selection changes from 0.10 to 0.65 ºC/Hr.
System time when display changes from MAX Red to MED Red
System time when display changes from MED Blue to Max Blue
Subtract the first time from the second time. The difference
between these two times must be less than 30 minutes.
Number of cycles in one minute. Should be between 35 and 43
SYSTEM SERIAL #:
HOSPITAL:
For use with TM-071
SYSTEM MODEL:
DATE:
TEST DESCRIPTION REQUIREMENT RESULT OK
3.1 Power up/Self Test
3.2 System Set-up
3.4
3.5
3.6
3.7 Air Trap test
3.9 Rate Set-up
3.10 Verify Slew Rate
3.11
Run Mode
Verification
Pump Door and
Mute Button Test
yes no
yes no
All tests and readings are within the specified limits: Pass________ Fail ________
Service Engineer (signature): _________________________________ Date: ____________
FORM-250 REVISION 3
ZOLL PAGE 1 0F 2