Nellcor SRC-MAX Technical Manual

SRC-MAX
Pulse Oximetry Functional Tester Technical Manual
+
--
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 Canada. To obtain information about a warranty, if any, contact Nellcor’s Technical
Services Department, or your local representative. Purchase of this instrument confers no express or implied license under any
Nellcor Puritan Bennett patent to use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign equivalents: 4,621,643; 4,653,498; 4,700,708; 4,770,179; Re.35,122 ; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,368,026; 5,485,847; 5,533,507; 5,662,106; and 5,853,364.
Contents
Contents .................................................................................i
Figures ................................................................................. iii
Tables .................................................................................. iv
Safety Information ........................................ ..... ..... ............. 1
Cautions ......................................................................... 1
Introduction .......................................................................... 3
Intended Use .................................................................. 3
Controls, Indicators, and Symbols ....................................... 5
Front Panel ..................................................................... 5
Back Panel ..................................................................... 6
Side View ...................................................................... 6
SRC-MAX Buttons and Symbols .................................. 7
Select Options ................................................................ 8
SRC-MAX Components and Accessories ...........................9
List of Components ....................................................... 9
Optional Accessories: .................................................... 9
SRC-MAX Operation with a Pulse Oximeter or Monitor . 11
ON / OFF Operation .................................................... 11
Power-Saving Mode .................................................... 11
Introduction ................................................................. 11
Routine Maintenance .............................................. ........... 17
Cleaning ....................................................................... 17
Battery Replacement .................................................... 18
Performance Verification ................................................... 21
Introduction ................................................................. 21
Equipment Required .................................................... 21
LED Current to Photodetector Current Test ................ 22
Testing the RED LED .................................................. 23
Testing the IR LED ...................................................... 25
BPM (Beats per Minute) Test ...................................... 27
Three-Lead ECG Cable ........................................... 9
C-Lock Cable ........................................................ 10
Test Setup .............................................................. 12
Test 1: BPM (Beats per minute) ........................... 14
Test 2: Saturation (Sp0
2) ....................................... 15
Test 3: Modulation ................................................ 15
Test 4: Light .......................................................... 16
Equipment Required ............................................. 27
SRC-MAX i
Contents
Setup ......................................................................27
BPM Test (Alternate Method) ......................................29
Equipment Required: .............................................29
Setup ......................................................................29
Troubleshooting ..................................................................31
Introduction ..................................................................31
Obtaining Technical Assistance ...................................31
Troubleshooting ...........................................................31
Electromagnetic Interference .......................................32
Packing for Shipment ..........................................................35
Introduction ..................................................................35
Returning the SRC-MAX .............................................35
Specifications ......................................................................37
Controls ........................................................................37
Rate ........................................................................37
Light ....................................................................... 37
Modulation ............................................................. 37
Saturation ............................................................... 37
Performance .................................................................38
Saturation (SpO2) ..................................................38
Pulse Rate ..............................................................38
Light Levels ...........................................................38
Modulation (%MOD) ............................................38
ECG .......................................................................38
C-Lock ...................................................................38
Timing ....................................................................39
Simulated Waveform .............................................39
Power ............................................................................ 39
Physical Characteristics .........................................39
Environmental Conditions ............................................40
Standards ......................................................................40
Electrical Safety Requirements ..............................40
EMC/EMI ..............................................................41
International and European Union Compliance ....41
Index ...................................................................................43
ii
Figures
Figure 1: Front Panel..............................................5
Figure 2: Back Panel..............................................6
Figure 3: SRC-MAX Tester Side View ................... 6
Figure 4: Three-Lead ECG Cable .......................... 9
Figure 5: ECG Cable Warning Label...................... 9
Figure 6: C-Lock Cable ........................................ 10
Figure 7: DOC-10 Sensor Cable..........................13
Figure 8: SpO2 sensor/cable connector............... 13
Figure 9: Battery Compartment Access ............... 18
Figure 10: Pinout Reference SRC-MAX................. 21
Figure 11: Block Diagram....................................... 22
Contents
Figure 12: RED LED Low Light Setting..................24
Figure 13: RED LED High Light Setting................. 25
Figure 14: IR LED Low Light Setting...................... 26
Figure 15: IR LED High Light Setting..................... 27
Figure 16: 3 Volts Peak-to-Peak at 3 Hz................28
Figure 17: 200 BPM ............................................... 29
SRC-MAX iii
Contents
Tables
Table 1: Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . 32
SRC-MAX iv
Safety Information
Note: In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Cautions
Cautions are identified by the CAUTION symbol shown above.
XIMAX technology-based
Cautions alert the user to exercise care necessary for the safe and effective use of the SRC-MAX.
Caution: The case of the SRC-MAX should be opened only by qualified service personnel. There are no user serviceable parts inside. Opening the battery compartment does not constitute opening the tester case.
Caution: Use your institution’s appropriate bench test procedures when using the SRC-MAX functional tester.
Caution: Install batteries into the SRC-MAX before using. See Routine Maintenance, page 17.
Caution: Do not open the case of the unit, as the device could be damaged.
Caution: Do not connect ground directly to any pins on the DB 9, as the device could be damaged. Refer to Block Diagram, Figure 17 on page 22.
SRC-MAX 1
Safety Information
Caution: Prior to connecting the oscilloscope’s negative terminal to the +5 volt supply (+) terminal, ensure that the oscilloscope negative is isolated from the +5 volt supply ground.
Caution: This device has been designed to provide resistance to electromagnetic interference. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in the health-care environment (for example, cellular phones, mobile two-way radios, electrical appliances), it is possible that high levels of such interference due to close proximity or strength of a source, may result in disruption of performance of this device. Disruption may be evidenced by erratic readings, cessation of operation, or other incorrect functioning. If this occurs, the site of use should be surveyed to determine the source of disruption, and actions taken to eliminate the source. If assistance is required, contact Nellcor Puritan Bennett’s Technical Services Department or your local Nellcor Puritan Bennett representative.
Caution: Do not spray, pour, or spill any liquid on the SRC-MAX, its accessories, connectors, switches, or openings.
Caution: Do not use the SRC-MAX to test non-O pulse oximeters or non-O
XIMAX OEM monitors.
XIMAX
Caution: Dispose of batteries in accordance with local requirements and regulations.
2
Introduction
Note: In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Intended Use
The SRC-MAX is intended for qualified technicians to functionally test Nellcor O oximeters and OEM O The SRC-MAX tester will perform the following tests:
Red and infrared light-emitting diode (LED) drive circuitry
XIMAX technology-based
XIMAX technology-based pulse
XIMAX technology-based monitors.
Heart rate (beats per minute, BPM)
Oxygen saturation (Sp0
2)
Light level
Percent modulation
The SRC-MAX functional tester is designed to simulate an operational Nellcor O
XIMAX compatible sensor and check
performance of the pulse oximeter or monitor.
SRC-MAX 3
(Blank Page)
Controls, Indicators, and Symbols
Note: In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O monitors refer to OEM O monitors.
Front Panel
XIMAX technology-based
1. Pulse oximeter/monitor cable connector 6. % Modulation select button
2. RED LED drive indicator 7. % Sp0
3. C-LOCK output connector port 8. Light level output button
4. ECG output connector port 9. Pulse rate beats per minute (BPM) button
5. Battery low indicator 10. INFRARED LED drive indicator
SRC-MAX 5
2 select button
Figure 1: Front Panel
Controls, Indicators, and Symbols
Back Panel
Side View
1. C-LOCK connector port 2. ECG Port
Figure 2: Back Panel
Figure 3: SRC-MAX Tester Side View
6
Controls, Indicators, and Symbols
SRC-MAX Buttons and Symbols
Front panel symbology for the SRC-MAX are as follows: Pulse rate/beats per minute (BPM) selector Button
Light output selector button
Percentage of Sp0
2 selector button
Modulation selector button
Feature not available at this time
Battery low indicator
Red LED indicator
C-Lock port symbol
ECG Port symbol
IR (infrared) LED indicator
SRC-MAX 7
Controls, Indicators, and Symbols
Select Options
There are four button switches on the SRC-MAX Front Panel with two associated LEDs.
Pulse/Beats per Minute
Light Level
% Sp0
2
Percent Modulation
The pulse oximeter or monitor will display readings for both Saturation (Sp0
2) and Pulse Rate. The numeric displays will
correlate with the SRC-MAX control panel.
8
SRC-MAX Components and Accessories
List of Components
(1) SRC-MAX Functional Tester (2) AA Alkaline Batteries (1) Technical Manual
Optional Accessories:
Three-Lead ECG Cable
See Figure 4. The three-lead ECG cable is shipped with a warning label, Figure 5. Attach the warning label to the three­lead ECG cable before using the cable.
Figure 4: Three-Lead ECG Cable
SRC-MAX 9
Figure 5: ECG Cable Warning Label
SRC-MAX Components and Accessories
C-Lock Cable
See Figure 6.
Figure 6: C-Lock Cable
10
SRC-MAX Operation with a Pulse Oximeter or Monitor
Note: In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O monitors refer to OEM O monitors.
ON / OFF Operation
The SRC-MAX will automatically turn ON when connected to the sensor/patient cable and a pulse oximeter or monitor that is turned on.
The SRC-MAX will automatically turn OFF when detached from the pulse oximeter or monitor, or if the pulse oximeter or monitor is turned off.
XIMAX technology-based
Install two AA batteries in the battery compartment of the SRC-MAX before connecting to the sensor/patient cable and the pulse oximeter or monitor. Refer to Battery Replacement on page 18.
Power-Saving Mode
When the SRC-MAX is connected to a pulse oximeter or monitor and no buttons are pressed for 2 minutes the SRC­MAX goes into the power-saving mode. All of the LEDs turn off, with the exception of the IR and RED LEDs. The SRC­MAX continues to function normally . Press any button on the SRC-MAX and the LEDs will illuminate.
Introduction
The SRC-MAX functional tester allows qualified technicians to functionally test Nellcor O oximeters and OEM O
SRC-MAX 11
XIMAX technology-based pulse
XIMAX technology-based monitors.
Loading...
+ 37 hidden pages