Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. OXIMAX is a
trademark of Nellcor Puritan Bennett Inc.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
To obtain information about a warranty, if any, contact Nellcor’s Technical
Services Department, or your local representative.
Purchase of this instrument confers no express or implied license under any
Nellcor Puritan Bennett patent to use the instrument with any sensor that is
not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign
equivalents: 4,621,643; 4,653,498; 4,700,708; 4,770,179; Re.35,122 ;
4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,368,026; 5,485,847; 5,533,507;
5,662,106; and 5,853,364.
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Cautions
Cautions are identified by the CAUTION symbol shown
above.
XIMAX technology-based
Cautions alert the user to exercise care necessary for the safe
and effective use of the SRC-MAX.
Caution: The case of the SRC-MAX should be opened
only by qualified service personnel. There are no user
serviceable parts inside. Opening the battery
compartment does not constitute opening the tester case.
Caution: Use your institution’s appropriate bench test
procedures when using the SRC-MAX functional tester.
Caution: Install batteries into the SRC-MAX before
using. See Routine Maintenance, page 17.
Caution: Do not open the case of the unit, as the device
could be damaged.
Caution: Do not connect ground directly to any pins on
the DB 9, as the device could be damaged. Refer to Block
Diagram, Figure 17 on page 22.
SRC-MAX1
Page 8
Safety Information
Caution: Prior to connecting the oscilloscope’s negative
terminal to the +5 volt supply (+) terminal, ensure that
the oscilloscope negative is isolated from the +5 volt
supply ground.
Caution: This device has been designed to provide
resistance to electromagnetic interference. However,
because of the proliferation of radio-frequency
transmitting equipment and other sources of electrical
noise in the health-care environment (for example,
cellular phones, mobile two-way radios, electrical
appliances), it is possible that high levels of such
interference due to close proximity or strength of a
source, may result in disruption of performance of this
device. Disruption may be evidenced by erratic readings,
cessation of operation, or other incorrect functioning. If
this occurs, the site of use should be surveyed to
determine the source of disruption, and actions taken to
eliminate the source. If assistance is required, contact
Nellcor Puritan Bennett’s Technical Services Department
or your local Nellcor Puritan Bennett representative.
Caution: Do not spray, pour, or spill any liquid on the
SRC-MAX, its accessories, connectors, switches, or
openings.
Caution: Do not use the SRC-MAX to test non-O
pulse oximeters or non-O
XIMAX OEM monitors.
XIMAX
Caution: Dispose of batteries in accordance with local
requirements and regulations.
2
Page 9
Introduction
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Intended Use
The SRC-MAX is intended for qualified technicians to
functionally test Nellcor O
oximeters and OEM O
The SRC-MAX tester will perform the following tests:
•Red and infrared light-emitting diode (LED) drive
circuitry
XIMAX technology-based
XIMAX technology-based pulse
XIMAX technology-based monitors.
•Heart rate (beats per minute, BPM)
•Oxygen saturation (Sp0
2)
•Light level
•Percent modulation
The SRC-MAX functional tester is designed to simulate an
operational Nellcor O
XIMAX compatible sensor and check
performance of the pulse oximeter or monitor.
SRC-MAX3
Page 10
(Blank Page)
Page 11
Controls, Indicators, and
Symbols
Note:In this manual, pulse oximeters refer to Nellcor
See Figure 4. The three-lead ECG cable is shipped with a
warning label, Figure 5. Attach the warning label to the threelead ECG cable before using the cable.
Figure 4: Three-Lead ECG Cable
SRC-MAX9
Figure 5: ECG Cable Warning Label
Page 16
SRC-MAX Components and Accessories
C-Lock Cable
See Figure 6.
Figure 6: C-Lock Cable
10
Page 17
SRC-MAX Operation with a Pulse
Oximeter or Monitor
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
ON / OFF Operation
The SRC-MAX will automatically turn ON when connected
to the sensor/patient cable and a pulse oximeter or monitor
that is turned on.
The SRC-MAX will automatically turn OFF when detached
from the pulse oximeter or monitor, or if the pulse oximeter
or monitor is turned off.
XIMAX technology-based
Install two AA batteries in the battery compartment of the
SRC-MAX before connecting to the sensor/patient cable and
the pulse oximeter or monitor. Refer to Battery Replacement
on page 18.
Power-Saving Mode
When the SRC-MAX is connected to a pulse oximeter or
monitor and no buttons are pressed for 2 minutes the SRCMAX goes into the power-saving mode. All of the LEDs turn
off, with the exception of the IR and RED LEDs. The SRCMAX continues to function normally . Press any button on the
SRC-MAX and the LEDs will illuminate.
Introduction
The SRC-MAX functional tester allows qualified technicians
to functionally test Nellcor O
oximeters and OEM O
SRC-MAX11
XIMAX technology-based pulse
XIMAX technology-based monitors.
Page 18
SRC-MAX Operation with a Pulse Oximeter or Monitor
The technician must perform the test setup procedure before
performing tests 1 through 4. The following is a brief
description of each test:
•Test Setup — This procedure establishes the
baseline for all the other tests. The Test Setup
procedure must be performed before performing any
or all of the SRC-MAX tests.
•Test 1: BPM — This procedure simulates an
O
XIMAX sensor attached to a patient indicating 60
BPM and 200 BPM. The test setup procedure sets up
Test 1 for 60 BPM.
Test Setup
•Test 2: SpO
XIMAX sensor attached to a patient ,indicating 75
O
2 — This procedure simulates an
percent blood oxygen saturation and 90 percent
blood oxygen saturation. The test setup procedure
sets up Test 2 for 75 percent blood oxygen saturation.
•Test 3: Modulation — This procedure simulates an
O
XIMAX sensor attached to a patient indicating low
and high pulse strength. The test setup procedure sets
up Test 3 for low pulse strength.
•Test 4: Light — This procedure simulates an
O
XIMAX sensor attached to a patient indicating low
and high light level passing throug h the patien t at the
sensor site. The test setup procedure sets up Test 4
for low light level.
Note:Refer to the applicable pulse oximeter or monitor
1.Locate the Sp0
pulse oximeter or monitor.
2.Connect a DOC-10 pulse oximetry cable to the pulse
oximeter’s Sp0
to the monitor’s Sp0
12
operator’s manual for the pulse oximeter or monitor
operation and test procedures.
2 Sensor Port on the front panel of the
2 sensor port or the ap plicable OEM cable
2 sensor port. See Figure 7.
Page 19
SRC-MAX Operation with a Pulse Oximeter or Monitor
Figure 7: DOC-10 Sensor Cable
3.Connect the SRC-MAX Sp0
2 sensor/cable connector to
other end of the DOC-10 cable or applicable OEM cable.
See Figure 8.
Figure 8: SpO
2 sensor/cable connector
4.Turn on the pulse oximeter or monitor. Refer to the
applicable pulse oximeter or monitor operator’s manual.
5.The IR and RED LEDs will alternately flash, turn off,
turn on and remain lit once the SRC-MAX is powered up.
6.The Green and Amber LED () and the Red Battery
Low LED will illuminate for 2 seconds and turn off,
indicating the LEDs are functioning properly.
7.The four selectable parameters (BPM, light level,
%SpO
condition (LED will be lit closest to the selector button).
SRC-MAX13
2, and percent modulation) will initiate to a default
Page 20
SRC-MAX Operation with a Pulse Oximeter or Monitor
Pulse/Beats per Minute
Light Level
% Sp0
2
Percent Modulation
8.The SRC-MAX is ready to test the pulse oximeter or
monitor.
Test 1: BPM (Beats per minute)
The BPM test checks the pulse oximeter or monitor beats per
minute operation at 60 and 200 BPM.
Prerequisite Procedure — Test Setup on page 12.
1.The pulse oximeter or monitor should be displaying a
BPM of 60 (57 to 63 BPM).
2.Press and release the SRC-MAX Heart Rate/BPM button.
3.The BPM LED indicator on the SRC-MAX will switch to
illuminate the 200 numeric reading.
4.Observe the display of the pulse oximeter or monitor.
The numeric reading for BPM will gradually rise to the
200 (197 to 203 BPM) level. Allow the pulse oximeter or
monitor 10-30 seconds to achieve the test result.
14
Page 21
SRC-MAX Operation with a Pulse Oximeter or Monitor
Test 2: Saturation (Sp02)
The saturation test checks the operation of the pulse oximeter
or monitor percent of blood oxygen saturation at 75% and
90%.
Prerequisite Procedure — Test Setup on page 12.
1.The pulse oximeter or monitor should be displaying a
%SpO
2 of 75 (72 to 78).
2.Press and release the SRC-MAX %Sp0
3.The %Sp0
from a value of 75 and illuminate the 90 value.
4.Observe the display of the pulse oximeter or monitor. The
numeric reading for %Sp0
(87 to 93). Allow the oximeter several seconds to achieve
the test result.
Test 3: Modulation
The modulation test checks to ensure that the pulse oximeter
or monitor is functioning properly under low and high
modulation conditions. The pulse oximeter or monitor shall
display the correct %SpO
in tests 1 and 2, under low and high modulation conditions.
2 button.
2 LED indicator on the SRC-MAX will switch
2 will flash and change to 90
2 and pulse rate reading, as defined
Prerequisite Procedure — Test Setup on page 12.
1.Perform Test 1: BPM on page 14.
2.Perform Test 2: %SpO
3.Press and release the SRC-MAX MOD button.
SRC-MAX15
2 on page 15.
Page 22
SRC-MAX Operation with a Pulse Oximeter or Monitor
4.The Modulation LED indicator on the SRC-MAX will
switch from the low (arrow down) value and illuminate
the high value (arrow up).
5.The Plethysmographic waveform or blip bar on the pulse
oximeter or monitor display will increase in amplitude
within a few seconds.
6.Perform Test 1: BPM on page 14.
7.Perform Test 2: %SpO
Test 4: Light
The light test checks to ensure that the pulse oximeter or
monitor is functioning properly under low and high light
conditions. The pulse oximeter or monitor shall display the
correct %SpO
and 2, under low and high light conditions.
Prerequisite Procedure — Test Setup on page 12.
1.Perform Test 1: BPM on page 14.
2.Perform Test 2: %SpO
3.Press and release the Light button on the SRC-MAX.
2 on page 15.
2 and pulse rate reading, as defined in tests 1
2 on page 15.
4.The HIGH (arrow up) LED will illuminate.
5.Pleth View — The Plethysmographic waveform on the
pulse oximeter display may briefly disappear and then
reappear slightly altered from the previously displayed
waveform.
Blip View — The blip display will saturate momentarily
and then return to normal indications.
6.Perform Test 1: BPM on page 14.
7.Perform Test 2: %SpO
16
2 on page 15.
Page 23
Routine Maintenance
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Cleaning
Caution: Do not spray, pour or spill any liquid on the
SRC-MAX, its accessories, connectors, or openings in the
chassis.
The SRC-MAX may be surface-cleaned by using a soft cloth
dampened with either a commercial, non-abrasive cleaner or
a solution of 70% alcohol in water, and lightly wiping the
surface of the unit.
XIMAX technology-based
SRC-MAX17
Page 24
Routine Maintenance
Battery Replacement
Figure 9: Battery Compartment Access
Press and slide the battery compartment door down to expose
the batteries. See Figure 9, above.
The SRC-MAX requires two AA alkaline batteries. The unit
will run continuously for a minimum of 45 hours with a new
set of batteries.
Caution: Install the batteries in the unit according to
polarity markings within the battery compartment.
When the SRC-MAX is connected via cable to a pulse
oximeter or monitor, the SRC-MAX will automatically
power on. The unit will turn off within two seconds of being
disconnected from the pulse oximeter or monitor.
18
Page 25
Routine Maintenance
The Low Battery Indicator LED will illuminate and blink
when the voltage drops below 1.75 volts. There is
approximately 15 minutes of use after first indication of low
battery LED. It is recommended to cease operation as soon as
possible and replace the batteries in the SRC-MAX.
SRC-MAX19
Page 26
(Blank Page)
Page 27
Performance Verification
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Introduction
This section describes the procedures for verifying specified
functionality of the SRC-MAX.
All measurements/tests shall be performed through the 9-pin
sensor/cable connector. See Figure 10.
Caution: Do not open the case of the SRC-MAX. The
device could be damaged.
XIMAX technology-based
Figure 10: Pinout Reference SRC-MAX
Caution: Do not connect ground to any pins on the SRCMAX 9 pin connector or the 9-pin D-sub female connector
used in the test set up. The device could be damaged.
Refer to Block Diagram, Figure 11 on page 22.
Equipment Required
•Oscilloscope; 20 MHz bandwidth, 2 channels, able to
resolve 500 millivolts peak-to-peak to 3 volts peakto-peak at 1,000 Hz, 1 and 3 Hz
•10 ohm ±1% resistor; 1/10 Watt Metal Film
•1 meg-ohm ±1% resistor; 1/10 Watt Metal Film
•Power Supply, able to deliver +5 volts 100 mA
SRC-MAX21
Page 28
Performance Verification
•Signal/Function Generator, able to deliver 0 to ±5
volts peak-to-peak at 1,000 Hz, 100 mA
•Pulse oximeter or monitor
•ECG monitor (RA, LA, LL)
LED Current to Photodetector Current Test
Note:The initial setting on the Signal Generator is 1000
Hz, Square wave, 0 to 5 volts peak-to-peak. The
RED and IR LEDs will be tested separately.
1.Connect a 10 ohm resistor (R1 in Figure 11) to pin 3 of a
9-pin D-sub female connector. The other side of the 10
ohm resistor should be connected to the Signal
Generator’s ground. See Figure 11.
2.Connect the positive side of the Signal generator to pin 2
of the 9 pin D-sub female connector.
3.Connect the oscilloscope channel 1 (+) to the junction of
pin 3 and the 10 ohm resistor.
4.Connect the oscilloscope channel 1 (-) to the junction of
the Signal Generator’s ground and the 10 ohm resistor.
22
Figure 11: Block Diagram
Page 29
Testing the RED LED
1.Turn on the signal generator.
2.Adjust the output until there is 475 to 525 millivolts peak
across the 10 ohm resistor.
3.Connect one end of a 1 meg ohm resistor (R2 in Figu re
11) to pin 9 of the 9 pin D-sub female connector.
4.Connect the other end to the positive terminal of the
power supply set for +5 volts.
5.Connect the negative terminal of the +5 volt supply to
pin 5 of the 9 pin D-sub female connector.
Performance Verification
Caution: See Figure 11. Prior to connecting the
oscilloscope’s negative terminal to the +5 volt supply (+)
terminal, ensure that the oscilloscope negative is isolated
from the +5 volt supply ground.
6.Connect the oscilloscope channel 2 (+) lead to the
junction of pin 9 and the 1 meg ohm resistor.
7.Connect the oscilloscope channel 2 (-) lead to the
junction of the positive terminal of the power supply (set
for +5 volts) and the other end of the 1 meg ohms.
8.With Low-Light setting selected on the SRC-MAX,
measure the peak-to-peak voltage across the 1 meg ohm
resistor on channel 2 of the oscilloscope. the peak-topeak voltage should be 600 to 900 millivolts peak-topeak. See Figure 12. The peak-to-peak voltage tolerance
includes a possible probe impedance error.
SRC-MAX23
Page 30
Performance Verification
Figure 12: RED LED Low Light Setting
Note:This figure and the following waveform figures
represent typical measurements. The actual
measured waveforms may vary in amplitude and
scale.
9.Select the High Light setting on the SRC-MAX. Measure
the peak-to-peak voltage across the 1 meg ohm resistor
on Channel 2 of the oscilloscope. The peak-to-peak
voltage should be 2.4 to 3.6 volts peak-to-peak. See
Figure 13.
24
Page 31
Performance Verification
Figure 13: RED LED High Light Setting
Note:The peak-to-peak voltage tolerance includes a
possible probe impedance error.
Testing the IR LED
1.Turn on the signal generator.
2.Adjust the signal generator output until there is -475 to 525 millivolts peak across the 10 ohm resistor (R1 in
Figure 11).
3.Select the Low Light setting on the SRC-MAX.
4.Measure the peak-to-peak voltage across the 1 meg ohm
resistor (R2 in Figure 11) on Channel 2 of the
oscilloscope. The peak-to-peak voltage should be 600 to
900 volts peak-to-peak . See Figure 14.
SRC-MAX25
Page 32
Performance Verification
Figure 14: IR LED Low Light Setting
Note: The peak-to-peak voltage tolerance includes a
possible probe impedance error.
5.Select the High Light Setting on the SRC-MAX.
6.Measure the peak-to-peak voltage across the 1 meg ohm
resistor on Channel two of the Oscilloscope. The peak-topeak voltage should be 2.4 to 3.6 volts peak-to-peak. See
Figure 15.
26
Page 33
Performance Verification
Figure 15: IR LED High Light Setting
Note:The peak-to-peak voltage tolerance includes a
1.Connect the SRC-MAX to the powered up pulse
oximeter or monitor. The SRC-MAX should power up
normally.
2. Select the BPM for 60 on the SRC-MAX.
3.Connect a Sub-mini 3/32” phone plug to the C- Lock
output of the SRC-MAX.
SRC-MAX27
Page 34
Performance Verification
4.Connect the oscilloscope positive lead to the phone
plug’s tip and the ground to the sleeve.
5.Adjust the oscilloscope for 3 volts peak-to-peak at 1 Hz.
See Figure 16.
.
Figure 16: 3 Volts Peak-to-Peak at 3 Hz
Note:The period of the signal should be 0.96 to 1.04
seconds.
6.The pulse width of the signal should be 50 to 100
milliseconds.
7.Select 200 BPM on the SRC-MAX.
8.Adjust the Oscilloscope for 3 volts peak-to-peak at 3 Hz.
See Figure 17.
28
Page 35
Performance Verification
Figure 17: 200 BPM
9.The period of the signal should be 0.27 to 0.33 seconds.
10. The pulse width of the signal should be 50 to 100
milliseconds.
BPM Test (Alternate Method)
Equipment Required:
•Pulse oximeter or monitor
•ECG monitor
•Three lead ECG cable
Setup
T olerances for this procedure are in the Specifications section,
page 37.
1.Connect the SRC-MAX to the powered-up pulse
oximeter or monitor. The SRC-MAX should power up
normally.
SRC-MAX29
Page 36
Performance Verification
2.Select 60 BPM on the SRC-MAX.
3.Connect the 3-Lead ECG cable to the ECG monitor and
the ECG port of the SRC-MAX.
4.Select Lead 1 on the ECG monitor.
5.Verify the 60 BPM reading on the ECG monitor.
6.Repeat step 5 for Lead selection 2 and 3.
7.Select 200 BPM on the SRC-MAX.
8.Select Lead 1 on the ECG monitor.
9.Verify the 200 BPM reading on the ECG monitor.
10. Repeat step 9 for Lead selection 2 and 3.
30
Page 37
Troubleshooting
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Introduction
Caution: The case of the SRC-MAC should be opened
only by qualified service personnel. There are no userserviceable parts inside.
Obtaining Technical Assistance
XIMAX technology-based
Contact Nellcor Technical Services Department,
1.800.635.5267 or your local Nellcor representative.
Troubleshooting
Troubleshooting symptoms, causes, and fixes are provided in
Table 1.
SRC-MAX31
Page 38
Troubleshooting
Table 1: Troubleshooting
SymptomCause/Fix
SRC-MAX does not powerup when attached to a pulse
oximeter or monitor.
SRC-MAX powers on and
turns off after about 3
seconds.
All the LEDs turn off, with
the exception of the IR and
RED LEDs; functionality
continues.
1.Batteries not installed,
inserted incorrectly, or
discharged.
2.Pulse oximeter or monitor
is not powered-on.
3.Sensor/patient cable not
connected.
4.Non-compatible pulse
oximeter or monitor.
5.Disconnect and reconnect
the SRC-MAX.
1.Disconnect and reconnect
SRC-MAX.
2.Pulse oximeter or monitor
has shutdown or reset.
Normal operation, the SRCMAX goes into power-saving
mode after about 2 minutes.
Press any button and the LEDs
will illuminate.
All of the LEDs turn on and
then turn off; this action
repeats.
Electromagnetic Interference
This device has been tested and found to comply with the
limits for medical devices to the IEC 60101-1-2: 1993,
EN60601-1-2:1994, Medical Device Directive 93/42/EEC.
These limits are designed to provide rea sonable protection
against harmful interference in a typical medical installation.
However, because of the proliferation of radio-frequency
transmitting equipment and other sources of electrical noise
32
Batteries are too low for normal
operation. Replace batteries.
Page 39
Troubleshooting
in the health-care environment (for example, cellular phones,
mobile two-way radios, electrical appliances), it is possible
that high levels of such interference due to close proximity or
strength of a source, may result in disruption of performance
of this device.
This equipment generates, uses and can radiate radio
frequency energy and if not used in accordance with the
instructions, may cause harmful interference with other
devices in the vicinity. Disruption or interference may be
evidenced by erratic readings, cessation of operation, or other
incorrect functioning. If this occurs, the site of use should be
surveyed to determine the source of this disruption, and
actions taken to eliminate the source.
The user is encouraged to try to correct the interference by
one or more of the following measures:
•Turn equipment in the vicinity off and on to isolate
the offending equipment
•Reorient or relocate the other receiving device
•Increase the separation between the interfering
equipment and this equipment
•If assistance is required, contact Nellcor Puritan
Bennett’s Technical Services Department or your
local Nellcor Puritan Bennett representative.
SRC-MAX33
Page 40
(Blank Page)
Page 41
Packing for Shipment
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Introduction
To return the tester for any reason, follow the instructions in
this section.
Returning the SRC-MAX
Contact Nellcor Technical Services Department,
1.800.635.5267 or your local Nellcor representative for
shipping instructions including a Returned Goods
Authorization (RGA) number. Unless Otherwise instructed
by Nellcor T echnical Services Department, it is not necessary
to return any other accessory item with the tester.
XIMAX technology-based
If the original carton is not available, use a suitable carton
with appropriate packing material to protect the unit during
shipping.
Mark the shipping carton and any shipping documents with
the Returned Goods Authorization (RGA) number . Return the
SRC-MAX by any method that provides proof of delivery.
SRC-MAX35
Page 42
(Blank Page)
Page 43
Specifications
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Controls
Rate
Two positions, Simulated Heart Rate:
60 BPM
200 BPM
XIMAX technology-based
Light
Two Positions, Simulated light attenuation level :
Low light level
High light level
Modulation
Two positions, Simulated Pulse Strength:
Low % Modulation
High % Modulation
Saturation
Two Positions:
75 % Sp02
90 % Sp02
SRC-MAX37
Page 44
Specifications
Performance
Saturation (SpO2)
75% ±1 Sat Point
90% ±1 Sat Point
Pulse Rate
60 bpm ±1 bpm
200 bpm ±2 bpm
Light Levels
Low Level = 750 nAi ±10%
High Level = 3,000 nAi ±10%
(nAi = nano-Amperes photodetector current)
Modulation (%MOD)
Low Modulation = 0.4% to 0.6%
High Modulation = 4.5% to 5.5%
ECG
Analog Outputs (R-Wave)
Output 1: = RA (White)
Output 2: = LA (Black)
Output 3: = LL (Red)
This is a 3 lead ECG compatible output.
C-Lock
Output 1: SRC-MAX System Ground
38
Page 45
Output 2: CMOS Compatible Output
Timing
C-Lock timing is synchronous with each optical pulse.
C-Lock synchronization signals start to 45-55 msec prior to
the optical pulse.
C-Lock synchronization signal is 50 msec to 100 msec in
width.
Simulated Waveform
Rise Time = 95 msec to 105 msec for all heart rates.
Fall Time = 190 msec to 210 msec for 200 BPM.
Specifications
Fall Time = 855 msec to 945 msec for 60 BPM.
Power
2 AA Alkaline Batteries
Minimum of 45 hours of continuous operation with new
batteries.
Typical power consumption: 80 milliwatts
Physical Characteristics
Overall Dimensions: 7 inches x 3.52 inches x 1.45 inches
Weight: .45 pounds
SRC-MAX39
Page 46
Specifications
Environmental Conditions
Temperature0 ºC to 40 ºC (32 ºF to 104 ºF)
Relative Humidity15% to 95% non-condensing
Altitude2,000 m (6,561.6 feet)
Transport and Storage (not in shipping container)
Temperature-20 ºC to 70 ºC
Operation
(-4 ºF to 158 ºF)
Relative Humidity15% to 95% non-condensing
Transport and Storage (in shipping container)
Temperature-20 ºC to 70 ºC
Relative Humidity15% to 95% non-condensing
Standards
Electrical Safety Requirements
UL 3101-1/UL 3111-1: Safety Requirements for Electrical
Equipment for measurement, control and laboratory use.
(-4 ºF to 158 ºF)
IEC 61010-1 (EN61010): Safety Requirements for Electrical
Equipment for measurement, control and laboratory use.
CSA 22.2 No. 1010.1: Safety Requirements for Electrical
Equipment for measurement, control and laboratory use.
40
Page 47
EMC/EMI
Specifications
IEC 61326-1, Amendment 1 (EN 61326-1): Electrical
Equipment for measurement, control and laboratory use EMC Requirements.
Radiated emissions per Federal Communication Commission
(FCC) part 15.
This ISM device complies with Canadian ICES-001, (CISPR
11, Class b, Group 1).
CISPR 11: Industrial, Scientific and Medical (ISM) RadioFrequency Equipment - Electromagnetic Disturbance
Characteristics - Limits and Methods of Measurement.