Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. OXIMAX is a
trademark of Nellcor Puritan Bennett Inc.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
To obtain information about a warranty, if any, contact Nellcor’s Technical
Services Department, or your local representative.
Purchase of this instrument confers no express or implied license under any
Nellcor Puritan Bennett patent to use the instrument with any sensor that is
not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign
equivalents: 4,621,643; 4,653,498; 4,700,708; 4,770,179; Re.35,122 ;
4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,368,026; 5,485,847; 5,533,507;
5,662,106; and 5,853,364.
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Cautions
Cautions are identified by the CAUTION symbol shown
above.
XIMAX technology-based
Cautions alert the user to exercise care necessary for the safe
and effective use of the SRC-MAX.
Caution: The case of the SRC-MAX should be opened
only by qualified service personnel. There are no user
serviceable parts inside. Opening the battery
compartment does not constitute opening the tester case.
Caution: Use your institution’s appropriate bench test
procedures when using the SRC-MAX functional tester.
Caution: Install batteries into the SRC-MAX before
using. See Routine Maintenance, page 17.
Caution: Do not open the case of the unit, as the device
could be damaged.
Caution: Do not connect ground directly to any pins on
the DB 9, as the device could be damaged. Refer to Block
Diagram, Figure 17 on page 22.
SRC-MAX1
Safety Information
Caution: Prior to connecting the oscilloscope’s negative
terminal to the +5 volt supply (+) terminal, ensure that
the oscilloscope negative is isolated from the +5 volt
supply ground.
Caution: This device has been designed to provide
resistance to electromagnetic interference. However,
because of the proliferation of radio-frequency
transmitting equipment and other sources of electrical
noise in the health-care environment (for example,
cellular phones, mobile two-way radios, electrical
appliances), it is possible that high levels of such
interference due to close proximity or strength of a
source, may result in disruption of performance of this
device. Disruption may be evidenced by erratic readings,
cessation of operation, or other incorrect functioning. If
this occurs, the site of use should be surveyed to
determine the source of disruption, and actions taken to
eliminate the source. If assistance is required, contact
Nellcor Puritan Bennett’s Technical Services Department
or your local Nellcor Puritan Bennett representative.
Caution: Do not spray, pour, or spill any liquid on the
SRC-MAX, its accessories, connectors, switches, or
openings.
Caution: Do not use the SRC-MAX to test non-O
pulse oximeters or non-O
XIMAX OEM monitors.
XIMAX
Caution: Dispose of batteries in accordance with local
requirements and regulations.
2
Introduction
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
Intended Use
The SRC-MAX is intended for qualified technicians to
functionally test Nellcor O
oximeters and OEM O
The SRC-MAX tester will perform the following tests:
•Red and infrared light-emitting diode (LED) drive
circuitry
XIMAX technology-based
XIMAX technology-based pulse
XIMAX technology-based monitors.
•Heart rate (beats per minute, BPM)
•Oxygen saturation (Sp0
2)
•Light level
•Percent modulation
The SRC-MAX functional tester is designed to simulate an
operational Nellcor O
XIMAX compatible sensor and check
performance of the pulse oximeter or monitor.
SRC-MAX3
(Blank Page)
Controls, Indicators, and
Symbols
Note:In this manual, pulse oximeters refer to Nellcor
See Figure 4. The three-lead ECG cable is shipped with a
warning label, Figure 5. Attach the warning label to the threelead ECG cable before using the cable.
Figure 4: Three-Lead ECG Cable
SRC-MAX9
Figure 5: ECG Cable Warning Label
SRC-MAX Components and Accessories
C-Lock Cable
See Figure 6.
Figure 6: C-Lock Cable
10
SRC-MAX Operation with a Pulse
Oximeter or Monitor
Note:In this manual, pulse oximeters refer to Nellcor
XIMAX technology-based pulse oximeters and
O
monitors refer to OEM O
monitors.
ON / OFF Operation
The SRC-MAX will automatically turn ON when connected
to the sensor/patient cable and a pulse oximeter or monitor
that is turned on.
The SRC-MAX will automatically turn OFF when detached
from the pulse oximeter or monitor, or if the pulse oximeter
or monitor is turned off.
XIMAX technology-based
Install two AA batteries in the battery compartment of the
SRC-MAX before connecting to the sensor/patient cable and
the pulse oximeter or monitor. Refer to Battery Replacement
on page 18.
Power-Saving Mode
When the SRC-MAX is connected to a pulse oximeter or
monitor and no buttons are pressed for 2 minutes the SRCMAX goes into the power-saving mode. All of the LEDs turn
off, with the exception of the IR and RED LEDs. The SRCMAX continues to function normally . Press any button on the
SRC-MAX and the LEDs will illuminate.
Introduction
The SRC-MAX functional tester allows qualified technicians
to functionally test Nellcor O
oximeters and OEM O
SRC-MAX11
XIMAX technology-based pulse
XIMAX technology-based monitors.
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.