Meade LightBridge Laser User Manual

Instruction Sheet
Laser Collimator
For use with Meade's LightBridge Telescopes
Assembly
Please read the entire instruction sheet before using your Laser Collimator.
Meade's Laser Collimator simplifies the collimation process. Collimation is a method to align your telescope's optics. Your telescope is aligned at the factory, but shipping and handling can sometimes misalign collimation. Misaligned collimation can mean dimmer and blurrier images in your telescope eyepiece. Your instruction manual describes how to collimate your telescope. Meade's Laser Collimator provides a simpler, second method of collimation.
Be aware of the following as you use your Laser Collimator:
Only turn on your laser when you are going to use it. When you turn on the laser, only turn the "on screw" until the laser turns on. Do not overtighten this screw! Overtightening may damage the collimator.
Note the small "doughnut" attached to your telescope's primary mirror. The doughnut is attached to help you perform alignment with the Laser Collimator. When you insert the Laser Collimator into your eyepiece and turn it on, the pinpoint of laser light will normally be close to or on top of this doughnut. However, if it is any distance away from the doughnut, perform the collimation procedure described in your LightBridge instruction manual before you use the method described below.
DANGER: Invisible laser radiation. Avoid eye or skin exposure to direct or scattered radiation.
All lasers use up battery power very quickly and have short battery life.
STEP 1:
Remove the eyepiece or eyepiece cap from the telescope.
STEP 2:
Point the laser downward. Turn the screw on the side of the laser until the laser turns on. Do not overtighten this screw.
window
doughnut
STEP 3:
Insert the laser collimator into the eyepie holder. Orient the window on the side of the collimator so that it is pointing towards the bottom of the telescope. Tighten the
.
collimator in place using the e
y
w
STEP 4:
Standing at the side of the telescope pinpoint. If the laser pinpoint is not near or on the "doughnut" attached to the mirror, remove the laser collimator and use the collimation method described in your
idge telescope man
LightBr If the pinpoint is v
ery near or on the doughnut, proceed to step 5.
ual.
, check out the primary mirror and look for the laser
tilt screws
(800) 626-3233
www.meade.com
STEP 5:
Using a Philips head screwdriver, adjust the one, two or all three tilt screws on the secondary mirror holder until the the laser pinpoint is in the center of the doughnut. Notice that when the pinpoint is on top of the doughn
ably. When it is in the center, it dims again.
sider
ut that the pinpoint brightens con-
STEP 6:
Go to the bottom side of the primary mirror tube and look up at the Laser Collimator. Adjust the one, two or three of the collimation screws on the bottom, until the pinpoint in the window (see Step #3) of the laser collimator is in the center of the window. Note that you will have to loosen one or all of the locking screws to tighten or loosen the collimation screws.
locking screw
pinpoint misaligned
collimation screw
pinpoint centered
STEP 7:
Return to the secondary mirror holder and readjust the tilt screws, if necessary, to make sure the pinpoint is centered in the doughnut. Your telescope should now be aligned.
14-7002-00 1/06