Leica RM2235 Instruction Manual

Page 1
Instruction Manual
Leica RM2235 V1.4; Rev B, English – 02/2010 Always keep this manual with the instrument. Read carefully before working with the instrument.
/HLFD50
Rotary Microtome
Page 2
Page 3
NOTE
The information, numerical data, notes and val­ue judgments contained in this manual repre­sent the current state of scientific knowledge and state-of-the-art technology as we under­stand it following thorough investigation in this field. We are under no obligation to update the present manual periodically and on an ongoing basis according to the latest technical develop­ments, nor to provide our customers with addi­tional copies, updates etc. of this manual. For erroneous statements, drawings, technical illustrations etc. contained in this manual we exclude liability as far as permissible according to the national legal system applicable in each individual case. In particular, no liability whatsoever is accepted for any financial loss or consequential damage caused by or related to compliance with state­ments or other information in this manual. Statements, drawings, illustrations and other information as regards contents or technical details of the present manual are not to be con­sidered as warranted characteristics of our products.
These are determined only by the contract provisions agreed between ourselves and our customers. Leica reserves the right to change technical specifications as well as manufacturing pro­cesses without prior notice. Only in this way is it possible continuously to improve the technolo­gy and manufacturing techniques used in our products. This document is protected under copyright laws. Any copyrights of this document are re­tained by Leica Biosystems Nussloch GmbH. Any reproduction of text and illustrations (or of any parts thereof) by means of print, photocopy, microfiche, web cam or other methods – includ­ing any electronic systems and media – requires express prior permission in writing by Leica Biosystems Nussloch GmbH. For the instrument serial number and year of manufacture, please refer to the name plate at the back of the instrument.
© Leica Biosystems Nussloch GmbH
Published by: Leica Biosystems Nussloch GmbH
Heidelberger Str. 17 - 19 D-69226 Nussloch Germany Phone: +49 (0) 6
224 143-0
Fax: +49 (0) 62 24 143-268 Internet: http://www.leica-microsystems.com
3
Page 4
Table of contents
1. Important notes ........................................................................................................................... 6
2. Safety ............................................................................................................................................ 7
2.1 Safety notes ................................................................................................................................ 7
2.2 Warnings ...................................................................................................................................... 7
2.3 Integrated safety devices ........................................................................................................ 10
3. Instrument components and specifications ....................................................................... 12
3.1 Overview – instrument components ...................................................................................... 12
3.2 Instrument specifications ........................................................................................................ 13
3.3 Technical data ........................................................................................................................... 14
4. Startup ........................................................................................................................................ 15
4.1 Standard delivery ...................................................................................................................... 15
4.2 Installation site requirements.................................................................................................. 15
4.3 Unpacking and installation ...................................................................................................... 16
4.4 Assembling the handwheel .....................................................................................................18
4.5 Inserting the universal cassette clamp ................................................................................. 19
4.6 Inserting the knife holder ......................................................................................................... 20
5. Operation ................................................................................................................................... 21
5.1 Operating elements and their functions ................................................................................ 21
5.1.1 Section thickness setting ......................................................................................................... 21
5.1.2 Coarse driving wheel ................................................................................................................ 21
5.1.4 Mechanical trimming function ................................................................................................ 22
5.1.3 Specimen retraction ................................................................................................................. 22
5.1.5 Specimen holder with precision orientation ......................................................................... 23
5.2 Adjusting the clearance angle ................................................................................................24
5.3 Fine adjustment of the force balance .................................................................................... 25
5.4 Clamping the specimen ............................................................................................................ 26
5.5 Clamping the knife / disposable blade ................................................................................... 26
5.6 Sectioning ................................................................................................................................... 27
5.7 Changing the specimen or interrupting sectioning ............................................................. 28
5.8 Finishing daily routine ............................................................................................................... 28
6. Optional accessories .............................................................................................................. 29
6.1 Assembly for fixture for specimen clamps ............................................................................ 29
6.1.1 Rigid fixture for specimen clamps ..........................................................................................29
6.1.2 Directional fixture for specimen clamps ............................................................................... 29
6.1.3 Fine-directional fixture for specimen clamps ....................................................................... 30
4
Instruction manual V 1.4 – 02/2010
Page 5
Table of contents
6.1.4 Quick clamping system ............................................................................................................. 31
6.2 Specimen clamps and holders ................................................................................................ 32
6.2.1 Standard specimen clamp ....................................................................................................... 32
6.2.2 Vee insert .................................................................................................................................... 32
6.2.3 Foil clamp type 1 ........................................................................................................................ 33
6.2.4 Foil clamp type 2 ........................................................................................................................ 34
6.2.5 Universal cassette clamp......................................................................................................... 35
Universal cassette clamp, ice-cooled ................................................................................... 35
6.2.6 Super mega-cassette clamp ................................................................................................... 36
6.2.7 Holder for round specimens .................................................................................................... 37
6.3 Knife holder base and knife holder ......................................................................................... 38
6.3.1 Knife holder base, without lateral movement feature ......................................................... 38
6.3.2 Knife holder E/E-TC ................................................................................................................... 39
6.3.3 Knife holder N/NZ ...................................................................................................................... 42
6.4 Section waste tray .................................................................................................................... 44
6.5 Backlighting................................................................................................................................ 44
6.6 Tray .............................................................................................................................................. 45
6.7 Freezer pack ............................................................................................................................... 45
6.8 Universal microscope carrier .................................................................................................. 46
6.9 Magnifying lens ......................................................................................................................... 48
6.10 Ordering information ................................................................................................................. 49
7. Customized solutions .............................................................................................................. 51
7.1 Possible faults ............................................................................................................................ 51
7.2 Instrument malfunctions .......................................................................................................... 51
8. Cleaning and maintenance .................................................................................................... 52
8.1 Cleaning the instrument ........................................................................................................... 52
8.2 Maintenance instructions ........................................................................................................ 54
8.3 Lubricating the instrument ....................................................................................................... 55
9. Warranty and service .............................................................................................................. 56
5
Page 6
1. Important notes
Symbols in the text and their meanings
Dangers, warnings and cautions ap­pear in a gray box and are marked by
a warning triangle .
Notes, i.e. important user information appear in a gray box and are marked
by an information symbol .
(5)
Numbers in parentheses refer to item numbers in illustrations.
Qualification of personnel
The Leica RM2235 may be operated by trained laboratory personnel only.
All laboratory personnel designated to oper­ate the Leica instrument must read this in­struction manual carefully and must be fa­miliar with all technical features of the instrument before attempting to operate it.
Intended use of instrument
The Leica RM2235 is a manually operated rota­tion microtome for creating thin sections of specimens of varying hardness for use in rou­tine and research laboratories in the fields of bi­ology, medicine and industry.
It is designed for sectioning soft paraffin speci­mens as well as harder specimens, as long as they are suitable for being cut manually.
Any other use of the instrument is considered
improper!
Instrument type
All information provided in this manual applies only to the instrument type indicated on the cover page.
An identification label indicating the instrument serial number is attached at the left side of the instrument.
Fig. 1
6
Instruction manual V 1.4 – 02/2010
Page 7
Be sure to comply with the safety instructions and warnings provided in this chapter. Be sure to read these instructions, even if you are already familiar with the operation and use of other Leica products.
2.1 Safety notes
2. Safety
This instruction manual contains important instructions and information regarding the op­erational safety and maintenance of the instru­ment. The instruction manual is an important part of the product, which must be read carefully prior to startup and use and must always be kept near the instrument.
If additional requirements on accident prevention and environmental protec­tion exist in the country of operation, this instruction manual must be sup­plemented by appropriate instructions to ensure compliance with such requirements.
The protective devices on both instrument and accessories may neither be removed nor modi­fied. Only service personnel qualified by Leica may repair the instrument and access the in­strument’s internal components.
2.2 Warnings
This instrument is built and inspected according to the Safety requirements for laboratory instru­ments.
For current information about applicable stan­dards, please refer to the CE declaration of con­formity on our Internet site:
www.leica-microsystems.com
In order to maintain this condition and ensure safe operation, the operator must observe all the instructions and warnings contained in this instruction manual.
The safety devices installed in this instrument by the manufacturer only constitute the basis for accident prevention. Primarily responsible for accident-free operation is above all the own­er of the instrument and, in addition, the designated personnel who operates, services or cleans the instrument. To ensure trouble-free operation of the instrument, make sure to comply with the following instructions and warnings.
7
Page 8
2. Safety
Warnings – Safety instructions / warning labels attached to the instrument
Safety instruction labels on the instrument marked with a warning triangle indicate that the correct operating instructions (as described in this manual) must be followed when operating or replacing the instrument component bearing the label.
Failure to adhere to these instructions may result in an accident, personal injury, damage to the instrument or accessory equipment.
Warnings – Transport and installation
Once removed from the crate, the instrument may only be transported in an upright position.
Do not transport the instrument by holding it by the handwheel grips, coarse driving wheel or the knob for setting the section thickness.
The protective devices on both instrument and accessories must neither be removed nor mod­ified.
Warnings – Working at the instrument
Take care when handling microtome knives and disposable blades. The cutting edge is extremely sharp and can cause severe injury!
Always remove the knife / blade before detaching the knife holder from the instrument. Always put the knives back into the knife case when not in use!
Never place a knife anywhere with the cutting edge facing upwards and never try to catch a falling knife!
Always clamp the specimen block BEFORE clamping the knife.
8
Instruction manual V 1.4 – 02/2010
Page 9
Warnings – Working at the instrument
Prior to manipulating the knife and specimen, or changing the specimen or knife, and during
breaks, always lock the handwheel and cover the cutting edge with the knife guard!
ALWAYS turn the handwheel clockwise; otherwise, the brake will not work properly.
Always wear protective glasses when sectioning brittle specimens. Specimen may splinter!
Specimen blocks must not be oriented during the retraction phase! If a block is oriented during
retraction, the block will advance by the retraction value PLUS the selected section thickness before the next section. This may cause damage to both specimen and knife!
Warnings – Cleaning and maintenance
Always lock the handwheel before cleaning!
Do not use any solvents containing acetone or xylene for cleaning!
Ensure that no liquids enter the interior of the instrument when cleaning!
When using detergents please comply with the safety precautions of the manufacturer.
2. Safety
9
Page 10
2. Safety
2.3 Integrated safety devices
5
Locking the handwheel
There are two ways of locking the handwheel (12):
Using the lever (3) on the right side of the micro­tome base plate, the handwheel can be braked in almost any position.
To brake, rotate the lever in a counterclock­wise direction to position z.
12
Caution! The braking lever (3) must be exactly in position z, so that the handwheel brake is applied correctly. If the lever is moved beyond this point, it is possible that the handwheel is no longer braked.
To unlock the handwheel brake turn the lever (3) back to its original position. Position
3
To lock the handwheel, press the lever (5)
c.
outwards and continue to turn the hand­wheel slowly until it locks exactly in the 12 o'clock position.
When using both brake systems at the same time, always move the lever (3) to position c first. Otherwise, it may be not be possible to release the lever (5).
10
Fig. 2
Instruction manual V 1.4 – 02/2010
Page 11
2. Safety
7
Knife holder N
8
7
Knife guard on the knife holder
Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi­ble to cover completely the cutting edge in every knife or blade position.
Knife holder N/NZ
The knife guard (8) of the knife holder N/NZ can be easily positioned via the two handles (7)
Fig. 3).
( To cover the knife edge, push both cover strips of the knife guard to the center.
Fig. 3
Knife holder E
11
9
10
Knife holder E/E-TC
The knife guard on knife holder E/E-TC consists of a red foldaway handle (9). To cover the cutting edge, fold the knife guard handle (9)
Knife holder E-TC
upwards as illustrated in
The clamping levers on the knife hold­er E are not interchangeable. The two clamping levers (10, 11) must remain in the position shown at all times, as otherwise isolated malfunc­tions of the knife holder can occur. Clamping lever for the blade (10) at the right, clamping lever for the later­al displacement (11) at the left.
Fig. 4.
9
Fig. 4
11
Page 12
3. Instrument components and specifications
3.1 Overview – instrument components
Universal cassette clamp
Knife holder base, without lateral movement feature
Removable section waste tray
Directional
fixture for specimen
clamps
Coarse driving wheel
Tray
Handwheel
locking
mechanism
Smooth-
turning
handwheel
Lever for
activating the
handwheel brake
Clamping lever
of the knife
holder base
Window for
displaying the
section
thickness
Adjusting knob
for setting the
section
thickness
Lever for
activating the
mechanical
trimming function
Clamping lever for lateral movement function of knife holder
12
Knife holder E
Fig. 5
Instruction manual V 1.4 – 02/2010
Page 13
3.2 Instrument specifications
Basic instrument with mechanical trim function, lateral coarse drive
The Leica RM2235 rotary microtome is equipped with a low-maintenance,­slack-free micrometer drive, with vertical and horizontal specimen feed realized via low-maintenance cross roller bearings.
The instrument is equipped with two independent handwheel locking systems for even greater operating safety.
Leica's patented, user-adjustable force balance system compensates centrifugal forces arising while cutting via a pretensioned spring for extremely light handwheel action.
Benefit: a heavy counterweight in the handwheel is no longer needed. The spring tension is individually adjustable, corresponding to the weight of the respective attached specimen clamp or specimen.
The patented specimen retraction system can be switched on and off by the user. The instrument thus offers all of the advantages of specimen retraction while supporting work in "rocking mode", i.e. without a full handwheel rotation.
We recommend disabling specimen retraction when working in "rocking mode".
The coarse drive wheel is ergonomically positioned. (For more information on the rotating direction, see
3. Instrument components and specifications
chapter 5.1.2)
13
Page 14
3. Instrument components and specifications
3.3 Technical data
General
Approvals: The instrument-specific marks are located on the
rear panel of the instrument next to the name plate. Operating temperature range: +10 °C to +35 °C Temperature range during storage: +5 °C to +55 °C
Relative humidity: max. 80 % non-condensing Humidity during storage: < 80 % Section thickness range: 1.0 - 60.0 µm
Section thickness settings: from 1.0 - 10.0 µm in 1.0 µm increments
from 10.0 - 20.0 µm in 2.0 µm increments
from 20.0 - 60.0 µm in 5.0 µm increments
Specimen feed: approx. 24 mm, ± 2 mm Vertical stroke: 70 mm Max. sectioning range without retraction: 69 mm (without specimen orientation at 1 µm) Max. sectioning range with retraction: 62 mm The specimen retraction can be turned off manually: 200 µm
Dimensions and weight
Width (including handwheel): 413 mm Depth (including waste tray): 618 mm Height (total): 305 mm (with tray on the hood)
Working height (knife blade): 168 mm (measured from the table)
Weight (without accessories): approx. 37 kg
Optional equipment and optional accessories
Object orientation (option) horizontal: 8° vertical: 8 ° Rotating range: ± 90°
Trimming increments: 10 µm, 50 µm
Repositioning of knife holder base north-south: ± 25 mm
14
Instruction manual V 1.4 – 02/2010
Page 15
4.1 Standard delivery
The Leica RM2235 standard delivery includes:
1 Leica RM2235 basic instrument
1 handwheel, complete ................................................................. 14 0500 38181
1 section waste tray ....................................................................... 14 0502 37931
1 tool set – consisting of: .............................................................. 14 0500 38600
1 Allen key with handle, size 5 ................................................ 14 0194 04760
1 Allen key with handle, size 4 ................................................ 14 0194 04782
1 Allen key, size 3 ...................................................................... 14 0222 04138
1 screwdriver 3x50, 186 long ................................................... 14 0170 11568
1 bottle (50 ml) of oil for drives, type 405 ............................... 14 0336 06086
1 brush w/magnet ..................................................................... 14 0183 40426
1 dust protective cover.................................................................. 14 0212 30350
1 operating manual D/E + language CD ...................................... 14 0500 80001
4. Startup
The accessories ordered are included in a separate box. Carefully check the delivery against the packing list and the delivery note. Should there be any discrepancy, please contact the Leica selling unit handling your order.
4.2 Installation site requirements
Stable, vibration-free laboratory bench with horizontal and even stage plate; practically vibration-free floor.
No other instruments nearby which might cause vibrations.
Room temperature permanently between +10°C and +35 °C.
Obstruction-free access to the handwheel.
Never operate the instrument in rooms with explosion hazard.
15
Page 16
4. Startup
4.3 Unpacking and installation
Fig. 6
22
2
22
When the instrument is delivered, check the tilt indicators on the packaging. If the arrowhead is blue, the shipment was transported laying flat, was tilted at too great an angle or fell over during transport. Note this on the shipping documents and check the shipment for possible damage.
11
1
11
Loosen and unscrew the six upper
screws (2).
Remove the cover (1).
Take the accessory carton (optional accessories) (3) and the cartons (4) from the standard scope of delivery.
16
The transport crate and includ­ed retaining elements should be kept in case a return ship­ment is necessary later.
33
3
33
44
4
44
55
5
55
44
4
44
44
4
44
44
4
44
Fig. 7
Instruction manual V 1.4 – 02/2010
Page 17
4.3 Unpacking and installation (cont.)
55
5
55
77
7
77
4. Startup
Take out the fixing module (5). To do so, hold it by the top edge of the module and in the recessed grip (6) and pull it out by pull­ing upwards.
Lift the instrument (7) by holding it by the
66
6
66
base plate and under the instrument on the back and lift it out of the formed cushion (8).
Never hold the instrument for transport by the handwheel or the rotary knob for section thickness adjustment.
Fig. 8
88
8
88
Place the instrument on a stable laboratory table.
The two sliding elements (9) located on the rear of the base plate make it easier to move the instrument on the table.
To move the instrument, hold it by the front of the base plate, lift it up gently and slide it on its slides.
9
Observe the correct resting angle to the table to avoid pinching your fingers.
17
Page 18
4. Startup
4.4 Assembling the handwheel
Fig. 9
The handwheel has to be assembled before attempting to use the instrument. The necessary parts and tools can be found in the tool­kit supplied in the delivery.
4
3
1
2
2a
5
The feather key (4) is loosely placed in the handwheel shaft (1) and fixed in place with a tie-rap during transport.
Remove the cable tie (3). Caution! Make sure not to lose the feather key!
Place the handwheel (2) on the handwheel shaft (1) as shown.
Tighten the screw (2a) located in the center hole of the hand­wheel with an Allen key size 4 (5).
Remove the cover foil from the self-adhesive cover disk and fix the cover disk on the hand­wheel.
18
Instruction manual V 1.4 – 02/2010
Page 19
4.5 Inserting the universal cassette clamp
4. Startup
60
There are two versions of the specimen holder, one with and one without specimen orientation, which are interchangeable. The object orientation allows for simple position correction of the specimen surface when the specimen is clamped into place.
You can use the quick clamping system (64) to hold all available accessory specimen clamps (for more information, see Chapter 6 "Optional
12
accessories").
To do so, proceed as follows:
Move the object head (60) to the upper end position by turning the handwheel (12) and engage the handwheel lock.
62
60
To release the clamping system, turn the screw (61) of the quick clamping system (64) counterclockwise using an Allen key size 4 (71).
64
63
61
71
Push the guide (63) of the universal cassette clamp (62) from the left into the quick clamp­ing system (64) as far as it will go.
To clamp the cassette clamp turn the screw (61) clockwise as far as it will go.
Since all stage clamps available as accessories are equipped with the same kind of guide on the back, they are inserted in the same way described here using the example of the cassette clamp.
Fig. 10
19
Page 20
4. Startup
4.6 Inserting the knife holder
55
53
52
Enlarged detail: Scale for better repositioning of the knife holder for varying specimen heights.
51
54
50
Setting up the knife holder base
Release the clamping lever (50) by rotating it counterclockwise.
Insert the knife holder base (51) with the notch (52) on the bottom into the T-piece (55) of the microtome base plate (53).
To secure the knife holder base, turn the clamping lever (50) clockwise.
The knife holder base (51) can be moved back and forth on the microtome base plate. This allows bringing the knife holder to optimal sec­tioning position in relation to the specimen.
There is a scale (54) on the right side of the mi­crotome base plate. This enables faster and better positioning of the knife holder at the specimen if various combinations of standard specimens and specimen holders are used. The rear edge of the knife holder base (51) functions as the scale reference.
20
57
56
Fig. 11
71
58
Fig. 12
Inserting the knife holder
Loosen the screw (58) using an Allen key size 4 (71) until the knife holder (57) can be moved.
Place the knife holder (57) with the under­side groove onto the T-piece (56) of the knife holder base (51).
To clamp, retighten the screw (58).
Instruction manual V 1.4 – 02/2010
Page 21
5. Operation
5.1 Operating elements and their functions
38
33
34
Fig. 13
5.1.1 Section thickness setting
The section thickness is set by turning the ad­justing knob (33) at the front of the microtome on the right. The adjusting knob has a notch for each value that can be set.
Setting range: 1 - 60 µm from 1 - 10 µm in 1 µm increments from10 - 20 µm in 2 µm increments from20 - 60 µm in 5 µm increments.
The section thickness set in each case is displayed in the window (34). The selected section thickness (on the scale) must agree with the red pointer (38)
5.1.2 Coarse driving wheel
The instrument can be ordered with clockwise or counterclockwise rotation. The given direc­tion of rotation means "forwards" and relates to the feed movement of the specimen towards the knife. The coarse motion serves for a fast horizontal forwards movement of the object - towards the knife - and backwards - away from the knife. When reaching the rear/front end positions, the coarse driving wheel can only be turned with difficulty. In the front end position, no more feed motion takes place.
The coarse driving wheel also turns during sectioning. Therefore it must not block whilst the handwheel is be­ing turned during sectioning; other­wise, no feed motion can take place and thus also no sectioning.
Fig. 14
21
Page 22
5. Operation
Magnet for holding the wrench size 4
5.1.3 Specimen retraction
The specimen retraction serves for protecting the knife and the specimen. When the retrac­tion is switched on, the object is drawn back 200 µm into the upper end position after the sectioning stroke during the return movement. Before the feed motion of the new section thickness, the feed motion for the retraction
For turning the retraction Off, turn the slot into the horizontal position (off).
For turning the retraction On, turn the slot into the vertical position (on).
value takes place. The specimen retraction can be switched off manually at the back of the instrument (Fig. 15), if required, using the slotted-screwdriver sup­plied with the delivery. Before switching the specimen retraction On and Off, run the object head to the upper end
Fig. 15
position by turning the handwheel.
5.1.4 Mechanical trimming function
The RM2235 is fitted with a mechanical trimming function. The trimming lever has 3 notching positions (0, 10 µm, 30 µm).
The points (36) mark the two trimming stages:
= 10 µm  = 30 µm
22
36
Fig. 16
For activating the trimming function, press the lever downwards into one of the two notching positions and keep depressed. After each rotation of the handwheel, a feed motion of 10 µm or 30 µm takes place.
After letting go of the lever, it automatically springs back to its original position (zero position). The trimming function is thereby deactivated.
The section thickness that has been set is not added to the selected trimming value. If the section thickness that has been set is greater than the selected trimming value, the section thickness is fed.
Instruction manual V 1.4 – 02/2010
Page 23
5.1.5 Specimen holder with precision orientation
In the quick clamping device of the specimen holder fixture with precision orientation, all object clamps available as optional accessories can be used (implemented).
30
32
29
32
31
5. Operation
The object orientation allows for simple position correction of the specimen surface when the specimen is clamped into place.
Orienting the specimen
Specimen blocks must not be oriented during the retraction phase! If a block is oriented during retrac­tion, the block will advance by the re­traction value PLUS the selected sec­tion thickness before the next section. This may cause damage to both speci­men and knife!
Fig. 17
Display of the zero position
For better display of the zero position, the orien­tation has two red indicators (32).
When both indicators are visible and both set­screws are in zero position at the same time (notch point, white marking on "R"), the speci­men is in zero position.
When the large standard specimen clamp (50 x 55 mm) is used, the speci­men orientation of 8° in north-south direction is no longer possible. The usable angle is only about 4° in this case.
Raise the object head to the upper end posi­tion and activate the handwheel lock.
To release the clamp, turn the eccentric lever (29) forwards.
Turn setscrew (30) to orient the specimen in north-south direction. Turn setscrew (31) to orient the specimen in east-west direction.
Each complete turn of the screw inclines the specimen by 2°. A total of 4 complete turns = 8° are possible in every direction. The accu­racy is approximately ± 0.5°.
For ease of estimation, there is a white marking on the handle and a notch point that is noticeable during turning.
To lock the current orientation, turn the eccentric lever (29) backwards.
23
Page 24
5. Operation
5.2 Adjusting the clearance angle
57
51
Enlarged detail: Index marks for adjusting the clearance angle.
58
The index marks (0°, 5° and 10°) for adjust-
ment of the clearance angle (59.1) are locat­ed on the right side of the knife holder (57).
There is also an index mark (59.2) on the right side of the knife holder basis (51) which serves as a reference point when adjusting the clearance angle.
Loosen the screw (58) using an Allen key size 4 (71) until the knife holder (57) can be moved.
Move the knife holder until the index mark of the desired clearance angle coincides with the reference line on the knife holder base.
Example: Enlarged detail showing a clearance angle setting of 5°.
24
59.1
59.2
71
The recommended clearance angle setting for knife holder E is approx. 5°.
Hold down the knife holder in this position and retighten the screw (58) for clamping.
Fig. 18
Instruction manual V 1.4 – 02/2010
Page 25
5. Operation
5.3 Fine adjustment of the force balance
If another accessory of a different weight is mounted on the object head (33), you must check whether it is necessary to readjust the force balance.
Checking the correct setting:
33
Fig. 19
Attach the new accessory and clamp the
specimen.
Set the object head to half the height of the
vertical travel range by turning the hand­wheel (
Fig. 19).
If the object head remains in this exact position, the setting is correct. If the object head moves, i.e. it is raised or low­ered, fine adjustment is necessary.
Important! Never turn the screw more than 1/2 turn at
a time.
The force balance is adjusted using the screw (34), which can be accessed by removing the
Failure to adjust the force balance may result in injury while working.
section waste tray on the bottom of the base plate of the microtome. Use the Allen key provided, size 5 (with handle!) for the adjustment.
If the object head moves downwards, turn the
screw approx.
1
/2 turn clockwise.
If the object head moves upwards, turn the
screw (34) approx.
1
/2 turn counterclock-
wise.
Continue this procedure until the object
head no longer moves once released.
34
Fig. 20
25
Page 26
5. Operation
5.4 Clamping the specimen
Rotate the handwheel until the specimen clamp is in the uppermost po­sition.
Always clamp the specimen block BEFORE clamping the knife. Lock the handwheel and cover the knife edge with the knife guard prior to any manipulation of knife or specimen, prior to changing the specimen block and during all work breaks!
Block the handwheel (allow lever (5) Fig. brake.
Insert a specimen block into the specimen clamp.
A detailed description for inserting the specimen into various specimen clamps and specimen holders is provided in Chapter 6 "Optional accessories".
5.5 Clamping the knife / disposable blade
Be very careful when handling microtome knives or blades. The cutting edge is extremely sharp and can cause severe injury!
Carefully insert knife or disposable blade into the knife holder and clamp.
Make sure that the blade is clamped parallel to the upper edge of the pressure plate. (For more information see
2 to notch) and activate the
Chapter 6.3.2, Fig. 33)
26
A detailed description for inserting the blade or the knife into the individual knife holders is provided in Chapter 6, "Optional accessories".
Instruction manual V 1.4 – 02/2010
Page 27
5.6 Sectioning
Cutting into the specimen (trimming)
Run the specimen to the rear end position by turning the coarse driving
wheel.
Push the knife holder on the knife-holder base almost until it is just
before the object.
Orientate the position of the specimen surface (only in the case of
specimen holders that can be orientated).
Release the handwheel lock, or handwheel brake, respectively.
Using the trimming lever select the required trimming stage.
Begin the cutting process by turning the handwheel.
Stop the cutting process when the required specimen level has been
reached.
Let go of the trimming lever.
Remove the sections
Set the required section thickness, or check the value setting on the
display, respectively.
Always use a different area of the cutting edge for trimming and sectioning.
To do so, laterally displace the blade or knife in the knife holder.
When using the knife holder E with lateral movement, it is sufficient to move the knife holder sideways.
For sectioning, turn the handwheel evenly in a clockwise direction.
5. Operation
Always turn the handwheel evenly in clockwise direction; otherwise, the brake will not work properly. The rotation speed of the handwheel must be adapted to suit the hardness of the spec­imen. For harder specimens, use a slower speed.
Take care not to block the coarse driving wheel when turning the handwheel! Otherwise there will be no feed motion of the section thickness and thus no sectioning will occur.
Pick up the sections and mount them on microscope slides.
27
Page 28
5. Operation
5.7 Changing the specimen or interrupting sectioning
Lock the handwheel and cover the knife edge with the knife guard prior to any manipulation of knife or object head, as well as prior to changing the specimen block and during all work breaks!
Raise the specimen to the upper end position and activate the mechanical handwheel lock.
Cover the sectioning edge with the knife guard.
Remove the specimen from the specimen clamp and mount a new sample to continue.
Run the object clamps with the coarse driving wheel back far enough until the new specimen can start being cut.
5.8 Finishing daily routine
Raise the specimen to the upper end position by turning the handwheel and engage the handwheel lock.
28
Always remove the knife / blade before detaching the knife holder from the instrument. Always put the knives back into the knife case when not in use! Never place a knife anywhere with the cutting edge facing up­wards and never try to catch a falling knife!
Remove the blade from the knife holder and insert it in the receptacle at the bottom of the dispenser, or remove the knife from the knife hold­er and put it back in the knife case.
Remove the specimen from the specimen clamp.
Push all section debris into the section waste tray and empty the tray.
Clean the instrument (see Chapter 8.1).
Instruction manual V 1.4 – 02/2010
Page 29
6.1 Assembly for fixture for specimen clamps
Depending upon the purchase order, the basic instrument is delivered with the directional or rigid fixture for specimen clamps which must be assembled first. All specimen clamps avail­able as accessories can be used in both fixtures for specimen clamps. Before assembling the fixture for specimen clamps, activate the mechanical handwheel lock!
6. Optional accessories
2
4
3
5
Fig. 211
35b5a5 6
7 + 8
2
1
4
9a
Fig. 22
7 + 8
9b 9b
9a
Finally, place the dovetail holder (2) and fasten by screwing in the 4 screws (1) using an Allen key size 3.
6.1.1 Rigid fixture for specimen clamps
Screw the rigid fixture for specimen clamps (4) onto the object head (3): Remove the screw (1), place the fixture for specimen clamps (4) onto the object head (3) from the front and tighten the screws (2) with an Allen key size 3. Next, insert the screw (1) from the side and briefly tighten it with an Allen key size 4.
Remove the rubber ring only after attaching the object head!
6.1.2 Directional fixture for specimen
clamps
Loosen the eccentric bolt (6) by turning it
counterclockwise.
Completely unscrew the thrust piece (5) with a flat-tip screwdriver and pull it out with spring (5a) and pin (5b).
Completely unscrew the setscrews (3) and (4).
Attach the directional fixture for specimen clamps as shown.
Insert the screws (7+8) in the bore (2 screws (8) are accessible through the bore (9a)) and evenly screw them in using a size 3 Allen key.
Insert the spring (5a) and pin (5b) with the flatter side into the thrust piece (5). Completely screw in the thrust piece with a flat-tip screwdriver.
Completely screw in the setscrews (3+4).
29
Page 30
6. Optional accessories
10
11
6.1.3 Fine-directional fixture for specimen clamps
Before the fine-directional fixture for speci-
men clamps can be mounted, loosen 4 screws (10) (Allen key size 3) and carefully remove the fixture for specimen clamps from the baseplate (9).
9
Using the 4 supplied screws (11) and the
12
Allen key size 3, fasten the baseplate to the specimen head (12).
30
11
Now, screw the fine-directional fixture for
specimen clamps with the 4 screws (10) and the Allen key size 3 onto the object head.
If the fine-directional fixture for spec­imen clamps is not used, retain the baseplate and 4 screws (11) together with the fine-directional fixture for specimen clamps!
Fig. 23
Instruction manual V 1.4 – 02/2010
Page 31
6. Optional accessories
6.1.4 Quick clamping system
It is used as specimen holder for use with the fine-directional fixture for specimen clamps with zero point indicators or the directional fixture for specimen clamps.
Screw the 4 screws (13) into bore A with an Allen key size 2.5 and tighten them.
A
13
A
Fig. 22
31
Page 32
6. Optional accessories
6.2 Specimen clamps and holders
All specimen clamps available as accessories can be integrated into either the directional or non-directional specimen holder fixture.
66
67
68
6.2.1 Standard specimen clamp
The standard specimen clamps are available in two sizes: 40 x 40 mm and 50 x 55 mm. They are designed for direct clamping of rect­angular blocks. In addition, they accommodate the foil clamps.
Turn the knurled screw (66) counterclock­wise to move the movable jaw (68) down­ward.
Mount the sample (67) as required.
Turn the knurled screw (66) clockwise to move the movable jaw upward against the fixed jaw to securely clamp the sample.
Fig. 25
32
70
70.1
6.2.2 Vee insert
66
The vee insert (70) is mounted in the hole pro­vided in the lower movable jaw of the standard specimen clamp. This makes it possible to clamp round speci­mens in the standard specimen clamp.
Turn the knurled screw (66) counterclock­wise to move the movable jaw (68) down-
69
68
Fig. 26
ward.
Insert the pin (70.1) of the vee insert (70) in the hole (69) of the lower jaw (68).
Mount the sample as required.
Turn the knurled screw (66) clockwise to move the movable jaw with the vee insert upward against the fixed jaw to securely clamp the sample.
Instruction manual V 1.4 – 02/2010
Page 33
6. Optional accessories
6.2.3 Foil clamp type 1
The foil clamp type 1 is appropriate both for clamping very small foil pieces and flat, angular samples. It is mounted in the standard speci­men clamp.
66
73
72
76
74
71
75
Fig. 27
7473
77
Clamping of foil pieces
Move the movable jaw (74) to the right as re-
quired by turning the set screw with an Allen key size 4 (71).
Place the foil (72) between the movable jaw
(74) and the fixed jaw (73).
To clamp the foil, screw the movable jaw
(74) against the fixed jaw (73) by using the Allen key.
Insert the foil clamp (75) in the standard
specimen clamp as shown.
Turn the knurled screw (66) clockwise to
clamp the foil clamp in the standard speci­men clamp.
Clamping of flat, angular samples
To clamp angular samples, replace the long set screw (76) with the short set screw (77) provid­ed with the foil clamp.
Unscrew the long set screw (76) to the left
with an Allen key size 4 (71).
Screw the short set screw (77) in the hole.
Place the sample (67) between the movable
jaw (74) and the fixed jaw (73).
To clamp the sample, screw the movable
jaw (74) by turning the set screw (77) against the fixed jaw (73).
Insert the foil clamp in the standard speci-
men clamp as shown.
Turn the knurled screw (66) clockwise to
clamp the foil clamp in the standard speci­men clamp.
67
Fig. 28
33
Page 34
6. Optional accessories
6.2.4 Foil clamp type 2
The foil clamp type 2 is appropriate for large foil ribbons. It is mounted in the standard specimen clamp.
72 7978
66
80
To open the jaws (78) and (79) lightly loosen
the 3 screws (
71).
(
81) with an Allen key size 4
Insert the foil ribbon (72) from behind to position it between the movable jaw (79) and
82
the fixed jaw (78).
To clamp the foil, first tighten the screw in the middle and then the other two screws (81) with the Allen key (71).
Place the foil clamp (82) in the standard
81
specimen clamp so that the beveled surface (80) on the back of the foil clamp points to
71
the right or left.
Turn the knurled screw (66) clockwise to securely clamp the foil clamp in the standard specimen clamp.
Fig. 29
34
Instruction manual V 1.4 – 02/2010
Page 35
6.2.5 Universal cassette clamp
6. Optional accessories
60
65
Fig. 30
Universal cassette clamp, ice-cooled
a
The universal cassette clamp (UCC) is designed to for horizontal or vertical clamping of all kinds of commercial cassettes. Laboratory personnel MUST check correct, firm seating be­fore cutting.
Push the lever (60) forwards.
Mount the cassette (65) horizontally or verti-
cally as required.
To clamp the cassette, release the lever (60).
Use the ice-cooled UCC with the non­directional specimen holder fixture only!
Ice cubes can be made using the included
Paraflex mold (a).
Quick clamping system (b), see page 31.
b
Fig. 31
35
Page 36
6. Optional accessories
6.2.6 Super mega-cassette clamp
2
4
5
3
Assembly of the super mega-cassette clamp
The super mega-cassette clamp should preferably be used with the rigid fixture for specimen clamps. To do so, proceed as follows:
Remove the rubber ring (5) only after having attached the cassette clamp on the object head!
Screw the rigid fixture for specimen clamps (4) onto the object head (3): Remove the screw (1), place the fixture for specimen clamps (4) onto the object head (3) from the front and tighten the screws (2) with an Allen key size 3.
Next, insert the screw (1) from the side and briefly tighten it with an Allen key size 4.
Insert the super mega-cassette clamp from the side on the left into the dovetail guide of the rigid fixture for specimen clamps and tighten screw (1).
If the directional fixture for specimen clamps is used with the rigid knife holder base, the orientation must be in position "0" and the cover for backlighting illumination must be detached. (Danger of collision if not observed!)
1
Fig. 32
36
NEVER use the super mega-cassette clamp with backlighting illumination!
Instruction manual V 1.4 – 02/2010
Page 37
6.2.7 Holder for round specimens
The holder for round specimens is designed to accommodate cylindrical samples. Inserts for specimens of 6, 15 and 25 mm diameter are available.
6. Optional accessories
67
90
89.1
89.2
89.3
92
91
To mount the required insert (89.1-3) turn the clamping ring (90) counterclockwise and remove it.
Place the required insert into the tension ring (90) and screw the tension ring onto the thread (91) by turning it clockwise.
Mount the sample (67) and fix by turning the clamping ring (90) clockwise.
To orient the inserted sample, insert the pin (92) into the bore (93) and rotate it counter­clockwise to release the clamp. You can now rotate the specimen so that the side you want faces upwards.
93
To lock it in the position you have chosen, tighten the pin (92) by turning it clockwise.
Fig. 33
37
Page 38
6. Optional accessories
6.3 Knife holder base and knife holder
94
6.3.1 Knife holder base, without lateral movement feature
The plastic handles of all clamping levers on the instrument and knife holders can be turned to the position that is most convenient for each user.
Pull the grip (94) out of the lever, hold it in this position, and rotate it to the desired position. It will then lock automatically when released.
Fig. 34
Repositioning the knife holder base
The one-piece knife holder base (rigid) (51) can be moved forwards and backwards on the microtome base plate.
This vertical displacement allows bringing the knife holder into the optimal cutting position in relation to the specimen.
38
51
To release, rotate the clamping lever (50) on the right side of the microtome base plate counterclockwise.
Reposition the knife holder together with the knife holder base forward or backward as appropriate.
Secure the clamping mechanism by rotating the lever (50) clockwise.
50
Fig. 35
Instruction manual V 1.4 – 02/2010
Page 39
6.3.2 Knife holder E/E-TC
6. Optional accessories
The knife holder E-TC is designed for the Leica TC-65 tungsten carbide blades.
Prior to inserting the blade, both knife holder and knife holder base must have been installed on the instrument!
Inserting the blades, knife holder E and E-TC
Knife holder E-TC
Fold knife guard (9) downward.
To insert the blade, flap the right clamping
lever (10) forward and down.
Carefully insert the blade from the side.
Make sure that the blade is clamped parallel to the upper edge of the pressure plate.
To clamp the blade, rotate clamping lever
(10) back upwards.
9
The knife holder E is designed for conventional disposable blades from all current manufac­turers. It is available in two models: one for narrow-band blades and one for broad-band blades. The knife holder has a lateral movement, so that the entire width of the blade can be used.
10
Fig. 36
11
Knife holder E
9
10
The clamping levers on the knife hold­er are not interchangeable. The two clamping levers (10, 11) must remain in the position shown at all times, as otherwise isolated malfunctions of the knife holder can occur. Clamping lever for the blade (10) at the right, clamping lever for the later­al displacement (11) at the left.
Fig. 37
39
Page 40
6. Optional accessories
Knife holder E
13
11
A
Note on the knife holder E:
Lateral displacement (only for knife holder E)
10
The lateral movement feature of the knife hold­er base enables the use of the entire length of the blade or knife, eliminating the need for read­justing the knife holder. The knife holder E con­sists of a segment arch A (with lever (11)) and the clamp mount B (with lever (10) and ejector (13)). The extreme left and right positions as well as the middle position are each marked with a
B
notch point.
To release the clamp, rotate the lever (11) on
12
Fig. 38
the left side of the knife holder forwards.
Move the knife holder sideways.
To clamp, rotate the lever (11) back.
The knife holder E is an important precision component, the quality and precise adjustment of which have a lasting effect on the entire function of the microtome. In case of malfunctions or damages to the clamp mount, it is always required to replace the clamp mount, including the pertinent clamping lever. Leica Biosystems offers special prices for new clamp mounts in case of damages to the clamp mount after the warranty has expired. In this way, perfect function of the device can be ensured over the course of many years.
Setting of the clamping system of the clamp mount on the segment arch
To ensure a proper sectioning result, clamp mount B must be securely clamped onto seg­ment arch A. The clamping is carried out using an eccentric at the lever (11). The clamping force is adjusted with the setscrew (12) on the underside of the segment arch. The setting of the clamping is carried out so that the clamping lever can be
Adjust the clamping with a 2.5 mm Allen key at the setscrew (12) so that the lever initially "bars" when the lever is activated. Continue turning the setscrew (12) a little bit (approx. 1/4 turn to the left or right), then check that the lever no longer "bars", but also does not jam "heavily". (The language CD features a short video film for
this purpose.) rotated to the stop with constantly increasing resistance.
40
Instruction manual V 1.4 – 02/2010
Page 41
6. Optional accessories
Knife holder E with water bath for narrow-band and broad-band blades
Fig. 39
Use
9
11
10
Knife holder E with water bath is available for both narrow-band and broad-band blades.
The knife guard on knife holder E consists of a red foldaway handle (9). To cover the cutting edge, fold the knife guard handle (9) upwards as illustrated in Figure.
The clamping levers on the knife holder are not interchangeable. The two clamping levers (10 and 11) must remain in the position shown at all times, as otherwise isolated malfunctions of the knife holder can occur. Clamping lever for the blade (10) at the right, clamping lever for the lateral displacement (11) at the left.
Floating thin paraffin sections (for example, for subsequent immunostain­ing procedures) on the surface of the water. Flat paraffin sections can be removed from the surface of the water using glass slides.
The vessel is filled with water up to the blade. After trimming, remove the section waste from the tray and create the sections to be prepared. The sections floating on the surface of the water can be removed using the glass slide.
Fig. 40
41
Page 42
6. Optional accessories
6.3.3 Knife holder N/NZ
The knife holders N and NZ are appropriate for standard steel and tungsten carbide knives, profile c and d, up to 16 cm long. The integrated height adjustment feature allows you to also use knives that have been resharpened numerous times.
Mounting the knife support bar
Push knife guard (8) to the center.
Set the knife support bar (46) onto the height adjustment screws (not vis­ible) in the position shown. The flat ends of the height adjustment screws must be located in the slots at each end of the knife support bar.
Enlarged detail: Knife inserted and height-adjusted
Knife holder N For holding convention-
al knives up to 16 cm in length.
Prior to inserting the knife, both knife holder and knife holder base must have been installed on the instrument!
Knife holder NZ
49
39
For holding conventional and hard metal knives up to 16 cm in length. Knife pressure plate (56) for extreme stability and full utilization of the knife blade.
42
49
8
47
56
48
46
Fig. 41
Instruction manual V 1.4 – 02/2010
Page 43
Inserting the knife
Knife height adjustment
6. Optional accessories
Rotate the knurled nuts (48) on the right and left of the knife holder
forward in opposite directions, lowering the knife support bar to the lowest possible position, thus ensuring that the knife edge will not be damaged when inserting the knife.
Unscrew the clamping screws (49) as far out as possible (rotate coun-
terclockwise).
Hold the knife (47) at the knife back and carefully insert it in the holder
from the side as shown with the cutting edge facing upward.
When adjusting the clearance angle, the knife edge should be positioned as exactly as possible in the actual center of rotation of the knife holder. The lay-on edge (39) of the rear clamping chucks serves as a reference position for correct knife height adjustment. The knife edge should be par­allel with the locating edges.
Rotate the knurled nuts (48) uniformly and backwards until the knife
blade is parallel to the lay-on edge (39) (see detailed illustration) of the rear clamping chucks.
To clamp the knife (47) evenly screw the two knife clamping screws
(49) inward (rotate clockwise).
Lateral repositioning of the knife
Push knife guard (8) to the center.
Loosen the clamping screws (49) by turning them counterclockwise.
Push the knife (47) to the left or right as required.
To clamp the knife (47), always tighten the clamping screw (49) first
which is located on the side to which the knife has been repositioned by turning it clockwise.
43
Page 44
6. Optional accessories
6.4 Section waste tray
53
39
18
6.5 Backlighting
Push the section waste tray (18) from the front to the microtome base plate (53) until it is held in place by the two magnets (39) (on the front of the microtome base plate). To remove the section waste tray, lift it slightly and pull it off towards the back.
Fig. 42
44
1
2
3
1
5
Remove the two screws (1) using a slotted
The backlighting is inserted at the front into the one-piece knife holder base.
screwdriver and then remove the cover plate (2).
Insert the backlighting (3) in the recess at
4
the rear of the knife holder base.
Insert the plug (4) for the backlighting into the socket (5) of the microtome and connect the plug of the power adapter to an AC pow­er socket.
The backlighting illuminates once the micro­tome is turned on with the mains switch.
Fig. 43
NEVER use the backlighting illumina­tion with the super mega-cassette clamp!
Instruction manual V 1.4 – 02/2010
Page 45
6.6 Tray
6. Optional accessories
The tray is mounted on the hood of the micro­tome so that the small feet on the underside fit into the cutouts on the hood. It is for storage of the utensils used during sectioning as well as the sectioned specimens.
Fig. 44
6.7 Freezer pack
98
98
99
Fig. 45
The freezer pack consists of the freezer plate (98) and the insulation jacket (99). It is for cool­ing (or keeping cold) paraffin specimens. Up to 35 standard cassettes can be cooled at the same time. The specimens require approx. 20 min. for being cooled from room temperature (approx. 20 °C) to "sectioning temperature". The cooling performance depends on the ambi­ent temperature and the volume of the speci­mens being cooled. The insulation jacket can be placed on the hood of the microtome in the same manner as the tray, or it can be placed next to the instrument.
When using the freezer pack on the instrument the tray (Fig. 44) must be removed.
Remove the freezer plate from the insulation
jacket and place it in a deep freezer, ideally overnight (but for at least six hours) at approx. -23°C.
Reassemble the freezer plate and insulation
jacket and place them on the bench or the instrument.
45
Page 46
6. Optional accessories
6.8 Universal microscope carrier
a
1
1 - Baseplate with drilling (a) 2 - Vertical column with size 8 cap screw (b) and
Fig. 46
d
e
8
2
3
c
b
7
6
4
5
Fig. 47
lock washer (c)
3 - Horizontal arm with cross-member (d) and
support ring (e)
4 - Support plate, large (for RM2235, RM2245 and
RM2255) 5 - Support plate, small (for RM2265) 6 - Allen key size 3 7 - 4 countersunk screws to install support plate 8 - Allen key size 8
Unpack all accessories in the package and check for completeness.
46
Assembling the universal microscope carrier
1. Attach baseplate
Select the large (5) or small (4) support plate,
depending on the microtome to be used.
Attach the support plate to the baseplate with
the included 4 countersunk screws (7) using the Allen key size 3 (6).
Fig. 48
Instruction manual V 1.4 – 02/2010
Page 47
Fig. 50
f
6. Optional accessories
2. Attach vertical column
b
c
Fig. 49
2
A
2
h
d
e
Insert the cap screw (b) into the hole of the baseplate from below. Place the lock washer (c) on the cap screw from above.
Thread the silver vertical column (2) onto the baseplate from above and tighten with the Allen key size 8.
The lock washer must be positioned between the baseplate and vertical column to prevent unintentional turn­ing of the column.
Extreme danger of tipping! After install­ing the vertical column, immediately place the microtome onto the baseplate so that the front feet of the microtome are positioned in the shallow recesses (A).
3. Attach horizontal arm
Slide the support ring (e) onto the vertical column and position it so that the lock nut (f) is facing to the back. Tighten the lock nut.
Slide the cross-member (d) onto the column.
g
Ensure that the lock screw (g) is facing to the right of the baseplate. The horizontal arm must be centered over the microscope.
Slide the horizontal arm (3), flat side facing the lock screw (g), into the cross-member (d) and tighten.
For more information on connecting and using the microscope, magnifying lens or cold-light source, please see the appropriate manual.
Fig. 51
47
Page 48
6. Optional accessories
6.9 Magnifying lens
1
4
2
Open the screw (3) on the horizontal arm of
The magnifying lens provides a 2x magnification and can be used with all Leica 2200 series rotary micro­tomes.
the microscope carrier in a counterclock­wise direction.
Insert the silver connection piece (1) as far as it will go. Tighten the screw (3).
Fig. 52
The adapter (2) allows a fiber-optic light guide to be installed.
3
Fig. 53
Adjust the position of the magnifying lens to the specimen being processed.
The magnifying lens can be swiveled completely to the side if necessary.
48
Always protect the magnifier glass from exposure to direct sunlight! Exposure to sunlight may cause a burning glass effect. Danger of fire!
Use the provided protective cap (4) to cover the magnifying lens.
Fig. 54
Instruction manual V 1.4 – 02/2010
Page 49
6. Optional accessories
6.10 Ordering information
Fixture for specimen clamps, rigid, silver ........................................................................ 14 0502 38160
Fixture for specimen clamps, directional, silver ............................................................ 14 0502 38949
Fixture for specimen clamps, directional, with 2 zero point indicators, silver .......... 14 0502 37717
Quick clamping system, complete .................................................................................... 14 0502 37718
Knife holder-base rigid, silver ........................................................................................... 14 0502 37962
Knife holder N RM22xx, silver ........................................................................................... 14 0502 37993
Knife holder NZ RM22xx, silver ......................................................................................... 14 0502 37994
Knife holder E narrow band RM22xx, silver, with ejector ............................................. 14 0502 40508
Knife holder E broad band RM22xx, silver, with ejector ............................................... 14 0502 40509
Knife holder E narrow band Run-off tray RM22xx.......................................................... 14 0502 38961
Knife holder E broad band Run-off tray RM22 ................................................................ 14 0502 39114
Knife holder E-TC RM22xx silver ....................................................................................... 14 0502 37997
Standard specimen clamp with adapter 50x60, silver .................................................. 14 0502 38005
Standard specimen clamp with adapter 40x40, silver .................................................. 14 0502 37998
V-insert for Standard specimen clamp, silver................................................................ 14 0502 38000
Universal cassette clamp, with adapter RM2125 (silver) ............................................. 14 0502 37999
Universal cassette clamp - ice-cooled ........................................................................... 14 0502 37793
Megaclamp assembly RM22xx, silver ............................................................................. 14 0502 38967
Foil clamp - type 1, black ................................................................................................... 14 0402 09307
Foil clamp - type 2, black ................................................................................................... 14 0402 26922
Round specimen holder, silver ......................................................................................... 14 0502 38001
Round specimen holder with adapter and 3 clamping rings, silver............................ 14 0502 38002
Clamping ring for round specimen holder, d = 6 mm .................................................... 14 0356 08322
Clamping ring for round specimen holder, d = 15 mm .................................................. 14 0356 09200
Clamping ring for round specimen holder, d = 25 mm .................................................. 14 0356 08320
49
Page 50
6. Optional accessories
Universal microscope carrier ........................................................................................... 14 0502 40580
Magnifying lens assembly ................................................................................................. 14 0502 42790
Cold light sources
Leica CLS 100X, 100-120V/50-60Hz ........................................................................... 14 0502 30214
Leica CLS 100X, 230V/50-60Hz.......................................................................... 14 0502 30215
Leica CLS 100X, 240V/50-60Hz........................................................................... 14 0502 30216
Fiber-optic light guide ................................................................................................... 14 0502 30028
Backlighting assembly ....................................................................................................... 14 0502 38719
External power supply, complete ..................................................................................... 14 0500 31244
Handrest for waste tray ..................................................................................................... 14 0502 38770
Cooling device assembly ................................................................................................... 14 0502 38325
Section waste tray, large .................................................................................................. 14 0502 37931
Section waste tray, small .................................................................................................. 14 0503 39060
Demo Box RM22xx series .................................................................................................. 14 0502 38930
Tray ....................................................................................................................................... 14 0502 37932
‘Leica’ brush w/magnet ..................................................................................................... 14 0183 40426
CoolClamp ............................................................................................................................ 14 0502 46573
50
Instruction manual V 1.4 – 02/2010
Page 51
In the following table there is a list of the most common problems which can arise while working with the instrument, along with possible causes and troubleshooting procedures.
Problem Corrective actionPossible cause
7.1 Possible faults
1. Thick/thin sections
The sections alternate between being thick and thin. In extreme cases, there are no sections what­soever.
2. Section compression
The sections are very compressed, show folds or are squeezed togeth­er.
3. "Stripes" in sections
For knife holder E
4. Noises during sectioning
The knife "sings" when sectioning hard specimens. The sections show scratches or chatter marks.
• The blade is not clamped properly.
• The blade is dull.
• The pressure plate is damaged or incorrectly adjusted.
• Clearance angle of the knife/blade too small.
• The blade is dull.
• The specimen is too warm.
• The sectioning speed is too fast.
• There is an accumulation of paraf­fin on the rear pressure plate of the knife holder.
• The sectioning speed is too fast.
• The clearance angle is too wide.
• Insufficient clamping of object and/ or knife holder.
7. Customized solutions
• Reclamp the blade.
• Laterally displace the knife holder or insert a new blade.
• Insert a new pressure plate or use a new knife holder.
• Readjust the pressure plate.
• Methodically experiment with larger clearance angle settings until you have found the optimum angle.
• Use another area of the blade or a new blade.
• Cool the specimen before section­ing.
• Reduce the sectioning speed.
• Remove paraffin from this area on a regular basis.
• Turn the handwheel at a slower speed.
• Methodically reduce the clearance angle setting until you have found the optimum angle.
• Check all screw and clamp connec­tions on the object holder system and the knife holder. If necessary, tighten the levers and screws.
7.2 Instrument malfunctions 1 There is no further feed motion and
thus no sectioning.
2. High blade consumption
• The front end position has been reached.
• Too great of a sectioning force was applied.
• Run the specimen back by turning the coarse driving wheel.
• Adjust the sectioning speed and/or section thickness when trimming. Select a smaller section thickness, rotate the handwheel more slowly.
51
Page 52
8. Cleaning and maintenance
8.1 Cleaning the instrument
Always remove the knife / blade before detaching the knife holder from the instrument. Always put the knives back into the knife case when not in use! Never place a knife anywhere with the cutting edge facing upwards and never try to catch a falling knife! When using cleaners, comply with the safety instructions of the manufacturer and the labor-safety regulations of your laboratory! When cleaning the outer surfaces, do not use xylene, scouring powders or solvents containing acetone or xylene. Xylene or acetone will damage the finished surfaces! Ensure that liquids do not enter the interior of the instrument during cleaning!
Before each cleaning carry out the following preparatory steps:
Raise the specimen clamp to the upper end position and activate the handwheel lock.
Switch the unit off and unplug it.
Remove the blade from the knife holder and insert it in the receptacle at the bottom of the dispenser, or remove the knife from the knife hold­er and put it back in the knife case.
Remove knife holder base and knife holder for cleaning.
Remove the specimen from the specimen clamp.
Remove section waste with a dry brush.
Remove specimen clamp and clean separately.
Instrument and outside surfaces
If necessary, the varnished outside surfaces of the control panels can be cleaned with a mild commercial household cleaner or soap water and then be wiped with a moist cloth. To remove paraffin residue, xylene substitutes, paraffin oil, or paraffin remov­ers such as "Para Gard" (Polysciences) can be used. The instrument must be completely dry before it can be used again.
Cleaning the knife
Always wipe the knife from the back of the knife to the cutting edge. NEVER wipe in the opposite direction - risk of injury!
Clean using an alcohol-based solution or acetone.
52
Instruction manual V 1.4 – 02/2010
Page 53
Knife holder E
11
8. Cleaning and maintenance
9
10
Take the knife holder apart for cleaning. To do so, proceed as follows:
Fold knife guard (9) downward.
83
Rotate the clamping lever (11) of the lateral
displacement forwards and pull it out side­ways.
86
Push the knife holder base plate (86) with the pressure plate (83) until it can be
Fig. 55
11
removed from the segment arch (87).
Rotate the blade clamping lever (10) down­ward and pull it out sideways.
Remove pressure plate (83).
Clean all parts of the knife holder.
If several knife holders are cleaned at
87
86
the same time, the parts must not be mixed up!
For cleaning and removal of paraffin, do not use xylene or cleaning fluids containing alcohol (e.g. glass cleaner).
Dry knife holder and reassemble.
83
10
Fig. 56
After cleaning the moving parts, apply a thin
• coat of drive part oil to them (see also Chapter 8.3).
When installing the pressure plate (83) make sure that the upper edge is parallel and level with the rear edge of the knife holder base plate (86).
53
Page 54
8. Cleaning and maintenance
Universal cassette clamp
Detach cassette clamp (13) for a thorough cleaning, removing all
60
13
Fig. 57
8.2 Maintenance instructions
Only authorized and qualified service personnel may access the internal components of the in­strument for service and repair!
The instrument is basically maintenance-free. To ensure trouble-free operation of the instrument over a long period of time, the following is recommended by Leica:
paraffin residues.
For cleaning, do not use xylene. Use xylene substitutes or paraffin removers such as "Para Gard."
The cassette clamp (13) can also be placed in an oven heated to a maximum of 65 °C, until the liquid wax escapes.
Remove paraffin residues with a dry cloth.
After such a cleaning procedure in an oven, always be sure to lubricate the axle and spring of the clamping lever (60) (see also
chapter 8.3).
1. Thoroughly clean the instrument on a daily basis.
2. Enter into a service contract at the latest at the end of the warranty period. For more in­formation, please contact your local Leica
3. Have the instrument inspected on a regular basis by a qualified service technician authorized by Leica.
The intervals depend on how heavily the instrument is used.
technical service center.
We recommend the following inspection intervals depending on the general workload of the instru­ment as defined in the two categories listed in the table:
Number of sections per day:
Workload (hours per day):
Sectioning speed:
Specimen material:
Maintenance:
54
Category I
> 8.000 sections per day
> 5 hours per day
Predominantly high sectioning speed
Working with soft and hard specimens
Every 12 months
Category II
< 8.000 sections per day
< 5 hours per day
Slow to medium sectioning speed
Predominantly soft specimen material
Every 24 months
Instruction manual V 1.4 – 02/2010
Page 55
8. Cleaning and maintenance
95
8.3 Lubricating the instrument
Once a month, lubricate the following parts with the included drive parts oil No. 405 (1 - 2 drops are enough).
96
Instrument and specimen holder:
The clamping key (95) of the quick clamping
system.
The T-piece (55) on the microtome base
55
plate.
Guide rails (96) for the knife holder base on
56
the microtome base plate.
The T-piece on the knife holder base (56).
Knife holder E:
11
88
10
Clamping lever (11) for the lateral displace-
ment.
T-piece (88) and guide (97) of the knife hold-
97
er for the lateral displacement.
Clamping lever (10) for the blade.
90
Knife holders N and NZ:
91
8
Sliding surfaces of the finger guard (8).
Knurled nuts (73) for measuring height
adjustment.
Universal cassette clamp:
Shaft (90) and flat coil spring (91) of the
clamping lever of the cassette clamp. After lubrication, move the lever back and forth
73
Fig. 58
several times.
55
Page 56
9. Warranty and service
Warranty
Leica Biosystems Nussloch GmbH guarantees that the contractual prod­uct delivered has been subjected to a comprehensive quality control pro­cedure based on the Leica in-house testing standards, and that the prod­uct is faultless and complies with all technical specifications and/or agreed characteristics warranted.
The scope of the warranty is based on the content of the concluded agreement. The warranty terms of your Leica sales organization or the or­ganization from which you have purchased the contractual product shall apply exclusively.
Service information
If you are in need of technical customer service or spare parts, please contact your Leica representative or the Leica dealer where you pur­chased the instrument.
Please provide the following information:
Model name and serial number of the instrument.
Location of the instrument and name of the person to contact.
Reason for the service call.
Delivery date.
Decommissioning and disposal
The instrument or parts of the instrument must be disposed of according to existing local applicable regulations.
56
Instruction manual V 1.4 – 02/2010
Loading...