Leica LED1, LED3, LED8, LED5 Instructions For Use Manual

From Eye to Insight
LEICA LED1 LEICA LED3 LEICA LED5 LEICA LED8
External light sources for Leica microscopes
Instructions for use
Leica Microsystems CMS GmbH. Instructions for use, 11934173, Revision 1.0, 2019.05.01 Document Number 57-10022 Revision A
1
Copyright
Copyright
All rights to this documentation are held by Leica Microsystems CMS GmbH. Reproduction of text or illustrations (in whole or in part) by print, photocopy, microfilm or other method (including electronic systems) is not allowed without express written permission from Leica Microsystems CMS GmbH.
The instructions contained in the following docu­mentation reflect state-of-the-art technology and knowledge standards. We have compiled the texts and illustrations as accurately as possible. Nevertheless, no liability of any kind shall be ac­cepted for the accuracy of this manual's con­tents. Still, we are always grateful for comments and suggestions regarding potential errors within this documentation.
The information included in this manual may be changed without prior notice.
As the official manufacturer of Leica branded light engine products, Lumencor is the respon­sible owner of this instructions for use 57-10022, Revision A.
Leica Microsystems CMS GmbH is the distributor of the Lumencor illumination products detailed herein, 11934173, Revision 1.0 2019.05.01 is for Leica internal tracking purposes only.
Leica Microsystems CMS GmbH Ernst-Leitz-Strasse 17–37 D-35578 Wetzlar (Germany) http://www.leica-microsystems.com
Responsible for contents: Marketing CMS
2
Contents
Contents
1. Important Notes about this Manual .............4
2. Function of the Compact Light Source ........6
3. Safety Notes .....................................................7
3.1 Regulatory safety notes .................................7
3.1.1 Introduction LED1 ............................................8
3.1.2 Introduction LED3 ............................................8
3.1.3 Introduction LED5/LED8 ..................................8
3.1.4 Precautions and Warnings ...........................9
3.2 Electrical safety .............................................12
3.3 Disposal ..........................................................15
4. Transport, Storage, Unpacking ..................16
5. Notes on Warranty ........................................17
6. Overview of the Instrument .........................18
6.1 Intended use ..................................................18
6.2 Identification of the instrument ...................18
6.3 Special features of the instrument .............18
6.4 Overview Light Sources ...............................19
6.5 Connecting LED1 ...........................................23
6.6 Connectable light guides (LED3-8) .............23
7. Instrument Setup ...........................................24
7.1 Installation ......................................................24
7.2 Connecting the LED1 .....................................24
7.3 LED3/5/8 – connecting the light guide .......24
7.4 Connecting the Leica POD (optional) .........26
7.5 Connecting the power cable .......................26
8. Operation ........................................................27
8.1 Operation LED1 ..............................................27
8.2 Operation LED3 ..............................................27
8.3 Operation LED5/8 ...........................................29
8.3.1 Start Up ...........................................................30
9. Service ............................................................31
9.1 Maintenance and cleaning ..........................31
9.2 Repairs ............................................................31
10. Troubleshooting ............................................32
11. EC Declaration of Conformity .....................33
3

1. Important Notes about this Manual

1. Important Notes about this Manual
Attention!
This User Manual is an essential component of the instrument and must be read carefully before assembly, commissioning or use.
Text symbols, pictograms and their meanings:
This User Manual contains important instruc­tions and information for the operational safety and maintenance of the instrument and acces­sories. Therefore, it must be kept in a safe place.
(1.2)
page 20
WARNING
CAUTION
!
Numbers in parentheses refer to illustrations; for example, (1.2) refers to Fig. 1, item 2.
Numbers with a pointer arrow (for example page 20), point to a certain page of this manual.
WARNING indicates a potentially hazardous situation with a medium degree of risk that, if not avoided, could result in death or serious injury.
CAUTION indicates a potentially hazardous situa­tion with a low degree of risk that, if not avoided, may result in minor or moderate injury.
Attention! In this manual, additional safety notes are in­dicated with the triangle symbol shown here, and have a gray background.
Attention! The instrument and its accessories can be damaged when operated incorrectly.
4
1. Important Notes about this Manual
Warning of hazardous electrical voltage! Danger of electric shock!
Warning of hot surface.
Warning: Possibly hazardous optical radiation emitted from this product. Do not look at operat­ing lamp. Eye injury may result.
UV
Warning: UV emitted from this product. Avoid eye and skin exposure to unshielded product.
Warning: IR emitted from this product. Do not
IR
look at operating lamp.
Warning of electromagnetic field
Instructions for disposing of the instrument, its accessories and consumables.
Earth (Ground) terminal
Explanatory note.
*
Item not contained in all configurations.
MM/YYYY
Manufacturer's address with manufacturing date
China RoHS 50 years EFUP (Environmentally friendly use period)
5

2. Function of the Compact Light Source

2. Function of the Compact Light Source
The Leica LED1, LED3, LED5 and LED8 compact light sources, with which this instructions for use are associated, are intended for biological rou­tine and research work as well as examinations in materials science or mineralogy.
Reasonably foreseeable misuse
The following are prohibited:
• Using the light source for any purpose not in accordance with the Declaration of Conformity (e.g. use as a medical product in accordance with EU Directive 93/42/EEC).
• Cleaning the light source in a way other than specified in the manual.
• Modifying the built-in safety circuits in the in­strument.
• Allowing unauthorized personnel to open the instrument.
• Using cables that Leica has not provided or permitted.
• Using combinations with non-Leica compo­nents that go beyond the scope of the User Manual.
• Operating the light source while it is not con­nected to the stand.
The manufacturer assumes no liability for damage caused by, or any risks arising from, using the light source for other purposes than those for which they are intended or not us­ing them within the specifications of Leica Microsystems CMS GmbH. In such cases, the Declaration of Conformity shall be invalid.
This instrument is not intended for use in a patient environment as defined by DIN VDE 0100-710. It is also not intended to be com­bined with medical devices as defined by EN 60601-1. If a microscope is electrically connected to a medical instrument in accor­dance with EN 60601-1, the requirements de­fined in EN 60601-1-1 shall apply.
Attention!
Attention!
6

3. Safety Notes

3. Safety Notes

3.1 Regulatory safety notes

Leica Microsystems utilizes regulatory model names for all certified and CE marked products. The regulatory model names are traceable to all regulatory documentation, third party reports and certifications.
“Regulatory Model: LED1/LED3/LED5/LED8 is used as a representative model for all certified and CE marked LED1/LED3/LED5/LED8 Products.
Emissions
This equipment has been tested and found to comply with the limits of EMC directive 2014/30/ EU. These limits are designed to provide rea­sonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interfer­ence to radio communications.
It also meets the requirements of EN 61326-1,
Electrical equipment for measurement, control and laboratory use – EMC requirements.
Photobiological Safety
Safety Certifications
TÜV SÜD America, CB Certification (IEC 61010-1:2010) TÜV SÜD America, NRTLus Certification (UL 61010-1:2012) TÜV SÜD America, cNRTL Certification (CAN/CSA-C22.2 No. 61010-1:2012) TÜV SÜD America, EN Certification (EN 61010-1:2010)
CE Marking
Low Voltage Directive (2014/35/EU) EMC Directive (2014/30/EU) RoHS Directive (2011/65/EU) plus (2015/863/EU) REACH Regulation (EC) No. (1907/2006/EC)
In accordance with the REGULATION (EU) No. 1194/2012 OF THE COMMISSION for the imple­mentation of the Directive 2009/125/EC of the European Parliament and Council with regard to the requirements for the environmentally friendly design of lamps with highly focused light, LED lamps and associated instruments, the products described here, Leica LED1, LED3, LED5 and LED8, are special products because they are not suitable for general lighting purposes.
The compact light sources have also been tested in accordance with EN 62471 / IEC 62471 Photo­biological safety of lamps and lamp systems and is classified in risk group 3 (high risk).
7
3. Safety Notes

3.1.1 Introduction LED1

The LED1 light source is a compact solid-state light source for fluorescence microscopy and other bioanalytical applications. The light source incorporates efficient thermal management, re­quiring no fan and providing quiet and vibration­free operation. This enables direct coupling to the fluorescence epi-illumination port of a micro­scope.
The light source provides four distinct color bands of light output that are controlled using either manual control knobs on the side panel or the light source control POD accessory. The COLOR knob located on the side panel is used for on/off control, selection of color channels and for enabling external tablet control. The INTENSITY knob provides four different levels of output at­tenuation between 100% and 10% of maximum.

3.1.2 Introduction LED3

The LED3 light sources are designed for labo­ratory use by bioanalytical researchers and/ or developers of life science instrumentation. The light sources generate white light output by combining the outputs of multiple solid state light sources. The light output on/off status is manu­ally controlled via a rocker switch located on the front panel or a foot pedal toggle switch acces­sory that plugs into the 3.5 mm connector on the rear panel.

3.1.3 Introduction LED5/LED8

The LED light source consists of 4 (LED5) and 8 (LED8) individually addressable solid-state light sources with integrated electronic control sys­tems. The constituent light sources may be LEDs or luminescent light pipes. The outputs of the constituent light sources are refined by bandpass filters and merged into a common optical train directed to the light output port on the front panel. The light output port has a built-in adapter for connection to a liquid light guide (LLG). An onboard photodiode continuously monitors the light output and generates refer­ence signals that can be applied to the constitu­ent sources in a feedback loop to maintain con­stant light output over time. The LED5/8 light sources are controlled by on­board micro-processors that connect to LAS X imaging software via a RS-232 (PC) serial in­terface. The user can enable or disable each source independently by serial commands, as well as change the intensity of each source in­dependently. Alternatively, the light sources may be turned on and off by TTL inputs from a trigger device such as a camera or a real-time control­ler. The only manual control is the master power switch on the front panel. Optimal internal op­erating temperature is maintained by negative pressure air cooling with the air intake at the front of the light engine and the exhaust fan at the the rear.
8
3. Safety Notes

3.1.4 Precautions and Warnings

A few simple practices will ensure trouble-free operation for the life of the light engine.
Safety Instructions:
Please read and follow all safety instructions provided BEFORE using the light source. Failure to comply with the safety instructions may result in fire, electrical shock, or personal injury and may damage or impair protection provided by equipment. Please save all safety instructions.
In order to maintain this condition and to ensure safe operation, the user must follow the instruc­tions and warnings contained in this User Man­ual.
Safety Definitions
Warning: Statements identify conditions or prac­tices that could result in personal injury
Warning: DO NOT look into the output of the light engine. The brightness of this light source is higher than most commercial lighting fixtures and is intended to couple directly into a micro­scope or other bioanalytjcal instrument.
Warning: DO NOT turn on the light without the output end of the light guide safely directed into an enclosed optical path. DO NOT point the light output directly onto any flammable or burn­susceptible material. This includes all animal or vegetable tissues, plastics, fabrics, paper and liquids
RISK GROUP 3
Applicable for LED1/3/5/8
RISK GROUP 3
WARNING Possibly hazardous optical radiation emitted from this product. Do not look at operating lamp. Eye injury may result.
Groupe de risque 3
AVERTISSEMENT Ce produit émet un rayonnement optique potentiellement dangereux. Ne regardez pas la lampe lorsqu‘elle est allumée. Des lésions oculaire peuvent en résulter.
Caution: Statements identify conditions or prac­tices that could result in damage to your equip­ment.
Safety Items
Warning: ONLY use the power supply provided by Leica. It is imperative that the DC power sup­ply has output over-current protection, as the power input of the light sources are not fused. The DC power supply must have the AC power cord connected to a receptacle with a protective safety (earth) ground terminal
Applicable for LED3/5/8
WARNING IR emitted from this product. Do not look at operating lamp. Avoid eye exposure. Use appropriate shielding or eye protection.
IR
AVERTISSEMENT Ce produit émet un rayonnement infrarouge. Ne regardez pas la lampe lorsqu‘elle est allumée. Évitez tout exposition aux yeux. Utilisez un écran ou dispositif de protection oculaire approprié.
RISK GROUP 3
Groupe de risque 3
9
3. Safety Notes
Applicable for LED1/3/5/8
RISK GROUP 3
WARNING UV emitted from this product. Avoid eye and skin exposure to unshielded product
UV
Groupe de risque 3
AVERTISSEMENT Ce produit émet un rayonnement ultraviolet. Évitez d‘exposer les yeux et la peau à un produit sans protection.
The Leica LED1, LED3, LED5 and LED8 com­pact light sources generate high-energy light with invisible UV components and has been classified in risk group 3 (high risk) in accor­dance with EN62471-1 while the light guide is not connected to the microscope/system. There is a general risk of being dazzled or blinded by light! Never look directly into the light guide output of the instrument or into the output of the light guide connected to the instrument. Always ensure that the output of the connected light guide is securely con­nected to the system to be illuminated before switching on the compact light source. Fur­thermore, before switching on the compact light source, dim the light output to prevent damage to the connected system. During op­eration, the light guide must not be removed from the microscope/system. While connect­ed to the microscope/system, the risk group 1 (low risk) has been classified. Please see the instructions on this topic in the microscope instructions for use.
Warning: Possibly hazardous optical radiation emitted from this product. Do not look at operat­ing lamp. Eye injury may result.
Warning: Infrared (IR) emitted from this product. Do not look at operating lamp.
Warning: UV emitted from this product. Avoid eye and skin exposure to unshielded product.
Caution: DO NOT open the unit. There are no serviceable parts inside and opening the light engine enclosure will void the manufacturer’s warranty.
Caution: DO NOT set liquids on the light engine. Spilled liquids may damage your light engine.
Caution: DO NOT drop the light engine. It con­tains glass optical components that could be damaged or misaligned by the shock produced by a drop onto a hard surface.
DISCLAIMER: Leica shall not be liable for injury to the user or damage to the product resulting from the light source being used in a way for which it was not intended and in complete disregard for all posted safety precautions and warnings.
10
Attention!
The instruments and accessories described in this Instructions for use have been tested for safety and potential hazards. The responsible Leica representative or the main plant in Wetzlar must be consulted whenever the instrument is altered, modified or used in conjunction with non-Leica com­ponents that are outside of the scope of this manual!
Unauthorized alterations to the instrument or noncompliant use shall void all rights to any warranty claims!
Attention!
If any safety defects or malfunctions of the compact light sources are identified, the instrument must be immediately taken out of operation, disconnected from the power system at all poles and secured against fur­ther use. For repair, the compact light source must be sent to the vendor or an authorized representative of the vendor.
3. Safety Notes
11