laerdal LSU Directions For Use Manual

Laerdal Suction Unit
Cat. No. 78 00 00 - Reusable
Cat. No. 78 00 30 - Serres
Directions for Use
LSU
English
3
1 Cautions and Warnings 4
2 Intended Use 4
3 User Interface
3.1 Operating Knob 5
3.2 Indicators 5
3.3 TEST-Button 6
3.4 Power Sources 6
3.5 Automatic Power-save Function 6
4 LSU Reusable Canister System
4.1 Overview 7
4.2 Parts 7
4.3 Use 8
4.4 Quick Reference 11
4.5 After Use 11
4.6 Decontamination of Laerdal Reusable
Canister 11
5 Serres Suction Bag System
5.1 Overview 12
5.2 Parts 12
5.3 Use 13
5.4 Quick Reference 15
5.5 After Use 15
6 Cleaning and Maintenance
6.1 General 16
6.2 Clean the Cabinet and parts 16
6.3 Disinfection of reusable parts 16
6.4 Charge the Battery 16
6.5 External Battery Charger 17
6.6 Check the Battery Quality 17
6.7 Replace the Battery 17
6.8 Fastening the brackets 18
6.9 Release arm for Canister holder 18
7 Device Test 18
7.1 Setup 18
7.2 Run the Test 19
7.3 Evaluation of Device Test Results 20
8 Servicing 21
9 Troubleshooting 21
10 Accessories and Parts 22
11 Specifications 24
11.1 Classification 24
11.2 General Tolerance 24
11.3 Physical Characteristics 24
11.4 Operation 24
11.5 Power Requirements 24
11.6 Enviromental Conditions 25
11.7 Material Chart 25
11.8 Symbols 26
– 11.9 EMC Specifications 27
12 Limited Warranty 28
13 Addresses 28
English
4
These Directions for Use covers two configurations of the Laerdal Suction Unit (LSU): Reusable, Serres Suction Bag System. Unless otherwise specified the information in this Directions for Use applies to both configurations.
Read these Directions for Use carefully, and become thoroughly familiar with the operation and maintenance of the LSU before using it.
1 Cautions and Warnings
The LSU should only be used by persons trained in the use of medical suction equipment.
The LSU is not suitable for use in the presence of flammable liquids or gases; there can be a danger of explosion or fire.
Do not use the LSU under environmental conditions that are outside the ranges specified under section 11.6. This can endanger safety and adversely affect operation of the device.
Do not block the Exhaust Outlet during use. This will lead to reduced flow and can also cause damage to the LSU.
Always use the Reusable version with the filter and the Float Ball in place.
Overflow of suctioned material can damage the device. If overflow of liquid from the Canister into the pump is suspected, the LSU must be returned for service (see section 8).
Disconnect the LSU from external power prior to cleaning. Use a minimum of liquid to prevent any electrical shock hazard.
Do not immerse the LSU or allow it to stand in water or other liquids. This might damage the device, and cause electrical hazard.
Do not pump any cleaning solution or other liquids through the vacuum pump, i.e. through the Vacuum Connector. This can damage the LSU.
Use only accessories supplied by Laerdal Medical or one of our authorised distributors to ensure that the LSU operates satisfactorily.
Battery
• Only use batteries approved by Laerdal Medical. Other batteries will have problems related to the battery status indicator of the LSU, the batter y operation time, and safety.
• To maintain satisfactory operation of the battery, it is recommended to place the LSU on continuous charge when not in use.
• The LSU must be placed on charge for a minimum of 24 hours to reach full battery capacity. The fast charging gives approximately 80% battery capacity after 3 hours (for a new battery). Please note that repetitive 3 hour charging is not recommended.
• If it is not possible to place the LSU on continuous charge when not in use, make sure the battery is charged for a minimum of 24 hours at least once a month.
• Do not store the battery when it is discharged; always fully charge the battery before storage.
• We recommend charging a spare battery every 6 months when stored in room temperature at 25°C (77°F ).
2 Intended Use
The LSU is a portable, electrically powered, medical suction equipment intended for field and transport use. It is intended for intermittent operation to remove secretions, blood or vomit from a patient´s airway to allow ventilation.
Higher vacuum levels are generally selected for oropharyngeal suctioning, and lower vacuum levels are usually selected for tracheal suctioning and the suctioning of children and infants.
English
5
3 User Interface (all configurations)
3.1 Operating Knob
The Operating Knob is a combined ON/OFF switch and vacuum selector, located at the bottom of the User Interface.
The LSU is switched ON by turning the Operating Knob to the required setting. Each setting indicates the maximum achievable vacuum level for the selected position (80, 120, 200, 350, 500+ mmHg). The free air flow value changes along with the increasing vacuum settings (see section 11.4 for details). The LSU is switched OFF by resetting the Operating Knob to “0”.
Even if the Operating Knob is in “0” position, voltage is present on some of the internal circuitry, when the LSU is connected to mains. To fully remove power, pull out the Power Cord.
3.2 Indicators
3.2.1 Power ON Indicator
This green LED has 3 modes:
• It is lit continuously while the LSU is switched ON.
• It flashes rapidly (approx. twice per second) during the Device Test.
• It flashes slowly (approx. once per second) while the Automatic Power-save Function is activated, if the Device Test is interrupted or when the battery is discharged.
3.2.2. External Power Indicator
This green LED is continuously lit while external AC or DC power is connected.
3.2.3 Failure Mode Indicator
The red LED is lit when a possible malfunction of the LSU has been detected. If lit, turn the LSU OFF, and then ON again to check if the indication disappears. Provided that the indication does not occur again the LSU can be operated. If the Indicator
continues to lit after three OFF/ON cycles and after replacing the battery with a fully charged battery, discontinue use and return the LSU for service. (See section 8).
3.2.4 Vacuum Indicator
This green LED bargraph displays the actual vacuum level during operation of the LSU. Each fully lit segment represents 50 mmHg. If a segment has a weak light, this represents 25 mmHg (e.g 125 mmHg means 2 fully lit and 1 weakly lit segments).
mmHg 80 120 200 350 500
kPa 11 16 27 47 67
mBar 107 160 267 467 667
Pressure conversion chart
3.2.5 Battery Status Indicator
This green LED bargraph has 3 functions:
• Indicates remaining battery capacity during operation from internal battery
• Indicates approx. achieved battery capacity during charging.
• Indicates which test is currently in progress during device test
During operation from internal battery: Indicates approx. remaining battery capacity. During operation from internal battery and during charging the displayed values must only be used as indications. Several parameters can influence them e.g. device settings, state of battery, temperature etc.
Note: Immediately after turning the LSU ON and immediately after switching from external power to internal battery operation, all 4 LED’s will flash for 5 seconds before the remaining battery capacity is displayed.
English
6
Typical remaining Battery capacity
Accuracy is dependent upon the lifetime and condition of the battery, as well as, the condition of the unit. Outside variables such as temperature may also affect accuracy.
Battery Capacity Output
> 50 % 4 LEDs lit.
30 - 50% 3 LEDs lit.
20 - 30% 2 LEDs lit.
< 20% 1 LED lit.
Battery Low; (min. 1 minute of operation left based on 500+ mmHg/free flow). 1 LED flashes.
During charging: indicates approx. achieved battery capacity.
Capacity Output*
< 75% The LED’s will be lit sequentially
75 - 80% 3rd LED’s lit and 4th flashing
> 80% 4 LED’s lit.
*Note: If no battery is installed, the battery status indicator will be fully lit for approx. 5 sec. until it is turned off.
During the Device Test
Indicates which step of the test is currently in progress or which corresponding test result is being displayed: LED 1 (lower LED) lit = STEP 1, LED 2 lit = STEP 2 etc.
3.3 TEST-Button
This button allows you to run a 4-step user initiated Device Test program to identify whether the LSU is operating satisfactorily or if it needs service (see section 7).
3.4 Power Sources
The LSU is delivered with an AC Power Cord for connection to AC mains and a DC Power Cord for connection to DC mains. The device can be operated from the internal battery, and can be operated or charged from one of the following external Power Sources:
• AC mains when used with the AC Power Cord: 100-240 VAC (50 / 60 Hz).
• DC mains when used with the DC Power Cord: 12-28 VDC.
A Wall Bracket to hold the LSU during operation and (optional) charging is available separately (see section 10).
AC Connector Socket
DC Connector Socket
External power connection will cause the External Power Indicator to be lit. Always check that it is lit when connecting external power or inserting the LSU into the Wall Bracket.
3.5 Automatic Power-save Function
The LSU has an Automatic Power­save Function that switches the pump motor off.
While in this mode, the Power ON Indicator will flash slowly (approx. once per second).
This function will be activated when the Operating Knob is set to 200, 350 or 500+ mmHg and the actual vacuum level has been continuously higher than 120 mmHg for more than 2 minutes. To exit the Power-save Mode and revert to normal operation, set the Operating Knob to any other position and then go back to required setting.
4 LSU Reusable Canister System
4.1 Overview
Aerosol Filter
Vacuum inlet
User Interface
Canister Holder
PATIENT Inlet
Patient Suction Tubing
Suction Catheter Adapter
Holder for Water Bottle/Patient Suction Tubing.
Water Bottle
Canister Lid with Float Ball Battery Door
Release Arm
Canister
4.2 Parts
Gasket
Vacuum seal
PATIENT Inlet
Canister
Aerosol Filter
Angled Connector
Patient Suction Tubing
Canister Lid w/Float Ball
Suction Catheter Adapter
English
7
The LSU Reusable Canister System consists of:
• A 1000 ml transparent plastic Canister.
• A Canister Lid w/Float Ball
• An Aerosol Filter with tubing
The Canister Lid contains a Float Ball that shuts off the vacuum if the canister is full or the LSU tips over. To restore the vacuum, remove the Angled
Connector from the VACUUM Inlet. The Float Ball will be released and the Angled Connector can be reattached,
Note: The LSU Reusable Canister System must not be used without the Aerosol Filter or the Float Ball.
Decontamination of reusable parts must be performed in accordance with section 4.6.
English
8
The Aerosol Filter protects the LSU by preventing aerosols from entering the Pump Unit. It is not intended for microbiological or for particle filtration. The Aerosol Filter is not designed for decontamination.
It is recommended that the Aerosol Filter be replaced after each use or at least once every shift. If the LSU is used on patients in areas where cross contamination is not an issue, the Aerosol Filter should be replaced at least once a month.
It is recommended always to have extra Aerosol Filters with the LSU in case one has to be discarded. If the Aerosol Filter becomes wet, it should be replaced immediately or as soon as possible after use.
4.3 Use
Note: Read all the Cautions and Warnings listed
in section 1 thoroughly before you use the LSU. The parts are color coded to help re-assembly after cleaning.
4.3.1 Check list
• Check that no parts are missing and that all parts are clean.
• If you decide to operate the LSU from external power, connect to either external AC or DC power as described in section 3.4. If you decide to operate the LSU from internal battery, check that the battery is installed.
• Check that the Patient Suction Tubing is securely connected to the PATIENT Inlet on the Canister Lid; and the Aerosol filter is securely fastened in the LSU and the lid.
4.3.2 Using the LSU
Note: The LSU must be operated and
transported in the upright position to prevent overflow of suctioned material.
1 Unwind the Patient Suction Tubing
2 Set the Operating Knob to the required
vacuum level. The LSU will automatically be switched ON and start to operate. The green Power ON Indicator is lit continuously while the LSU is switched ON.
3 When suction is complete, set the Operating
Knob to “0”
4.3.3 Emptying the Canister
Note: To prevent damage and keep the LSU in
good working order; empty the canister when 3/4 full. Check the filter after each use.
If the filter is broken so liquid penetrates the membrane, the pump will be contaminated, and the LSU must be returned for service (see section 8).
When liquid reaches the top of the Canister, the LSU will stop suctioning. To continue suctioning, empty canister and replace the filter.
Overflow of suctioned material can damage the device.
English
9
To remove the Canister, proceed as follows:
1 Disconnect the Filter Tubing from the blue
Vacuum Inlet (see section 4.2 regarding handling instructions for the Aerosol Filter).
2 To release the Canister Holder, press down the
Release Arm while sliding the Holder towards you. Remove the Canister from the Holder.
3 Release the Patient Suction Tubing (if
applicable).
4 Safely dispose of the contents of the Canister
according to local protocols.
5 Clean the LSU according to section 6.2 and
its reusable parts according to section 4.6.
4.3.4 Assembly of the Reusable Canister System
1 Place the yellow Gasket in the Lid, make sure
that the gasket is placed flat in the Lid and not twisted.
2 If the Float Ball has been removed, hold the
lid upside down and snap the ball into place as shown.
Note: Never use the LSU without the Float Ball.
3 Secure the Lid to the Canister by turning it
clockwise until it is securely fastened.
English
10
4 Connect the Suction Catheter Adapter to
the Patient Tube.
5 Connect the Angled Connector to the Patient
Tube.
6 Mount the yellow Vacuum Seal in the centre
hole of the Lid and push it into place.
7 Connect the Aerosol Filter (blue ring) to the
blue Vacuum Inlet on the LSU. (Color code: Blue on Blue)
8 Insert the Canister in the canister holder and
slide it into the LSU.
9 Connect the yellow Angled Connector to the
yellow Vacuum Seal
10 Connect the Angled Connector on the Patient
tubing to the PATIENT Inlet on the canister Lid.
11 Perform the Device Test according to section 7.
12 Wind the Patient Tubing on the Patient
Tubing Holder.
13 Place the LSU on charge (see section 6.4)
Loading...
+ 22 hidden pages