Keysight N5224/5/7B, N5244/5/7B User Manual

Keysight 4-Port N5224/5/7B and N5244/5/7B Models PNA Direct Digital Synthesizer (DDS) For Version 7 Dual-Source Synthesizer Upgrade Kit
This manual provides documentation for the following models:
To Upgrade 4-Port B models to 4-Port B models with DDS
N5224BU-4S7 (for N5224B), N5225BU-4S7 (for N5225B), N5227BU-4S7
(for N5227B), N5244BU-4S7 (For N5244B), N5245BU-4S7 (For N5245B),
and N5247BU-4S7 (For N5247B)
Upgrade Kit Number: N5225-60143
Installation Note
Notices
© Keysight Technologies, Inc. 2007-2021
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5225-90143
Edition
Edition 1, January 14, 2021
Supersedes: January 2021
Printed in USA/Malaysia
Published by: Keysight Technologies
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or
commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Keysight Add Direct Digital Synthesizer (DDS) 4-Port Upgrade Kit Upgrade Kit Numbers: N5225-60143
Installation Note

Description of this Upgrade

This upgrade adds direct digital synthesis capability to your microwave network analyzer. This is done by removing the existing frequency reference,
13.5 GHz synthesizer boards (x3), and installing a new direct digital synthesizer assembly.
Refer to “Overview of the Installation Procedure” on page 10.
This upgrade adds the following items to your network analyzer:
— direct digital synthesizer (DDS) assembly
— new left outer and inner panels
— new cables
This repair must be done at a service center or a self-maintainer service center! Refer to “Getting Assistance from Keysight” on page 4.
N5224/5/7B or N5244/5/7B
3
-
Description of this Upgrade Getting Assistance from Keysight

Getting Assistance from Keysight

By internet or phone, get assistance with all your test and measurement needs.
Installing this upgrade kit requires special skills and experience. If you think you may not be qualified to do the work, or need advice, contact Keysight.

Contacting Keysight

Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at:
http://www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field engineer.
In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number. With this information, the Keysight representative can determine whether your product is still within its warranty period.
If You Have Problems With the Upgrade Kit Contents
Keysight stands behind the quality of the upgrade kit contents. If you have problems with any item in the kit, refer to www.keysight.com and the Contact
Keysight () link.
4 Installation Note N5225-90143
Description of this Upgrade Getting Prepared

Getting Prepared

The PNA contains extremely sensitive components that can be ruined if mishandled. Follow instructions carefully when making cable connections, especially wire harness connections.
The person performing the work accepts responsibility for the full cost of the repair or replacement of damaged components.
To successfully install this upgrade kit, you will need the following:
— A PDF copy or a paper copy of the PNA Service Guide - refer to
— An ESD-safe work area - refer to “Protecting Your Workspace from
— Correct tools - refer to “Tools Required for the Installation” on page 7.
— Enough time - refer to “About Installing the Upgrade” on page 7.
“Downloading the Online PNA Service Guide” on page 5 below.
Electrostatic Discharge” below.
— Test equipment for the post-upgrade adjustments and full instrument
calibration. To view the equipment list, click the Chapter 3 bookmark “Tests and Adjustments” in the PDF Service Guide

Downloading the Online PNA Service Guide

To view the Service Guide for your PNA model number, use the following steps:
1. Go to www.keysight.com.
2. In the Search box, enter the model number of the analyzer (Ex: N5247B) and click Search.
3. Click Services & Support > Document Library > Manuals or under Refine Your Results:
4. Click Manuals > Service Manual.
5. Click the service guide title to download the PDF file.
6. When the PDF of the Service Guide is displayed, scroll through the
Contents section bookmarks to locate the information needed.
1
.
1. See “Downloading the Online PNA Service Guide” on page 5.
Installation Note N5225-90143 5
-
Description of this Upgrade Getting Prepared

Protecting Your Workspace from Electrostatic Discharge

For information, click on the Chapter 1 bookmark, "Electrostatic Discharge Protection" in the PDF of your instrument’s Service Guide on
1
www.keysight.com
ESD Equipment Required for the Installation
Description Keysight Part Number
ESD grounding wrist strap 9300-1367
5-ft grounding cord for wrist strap 9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire 9300-0797
ESD heel strap (for use with conductive floors) 9300-1308
.
1. See “Downloading the Online PNA Service Guide” on page 5.
6 Installation Note N5225-90143
Description of this Upgrade Tools Required for the Installation

Tools Required for the Installation

Description Qty
T-10 TORX driver - set to 9 in-lbs (1.02 N.m) 1
T-20 TORX driver - set to 21 in-lbs (2.38 N.m) 1
5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m) 1
5/16-in (8 mm) nutsetter or open end torque wrench - set to 21 in-lbs (2.38 N.m) 1
T-6 TORX driver - set to 6 in-lbs (0.68 N.m) 1
5/8-in (16 mm) nutsetter or open end torque wrench - set to 21 in-lbs (2.38 N.m) 1
Needle nose pliers 1
Flat head screwdriver 1
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections except the front and rear panel bulkhead connectors. On these, use a 9 mm nutsetter or open end torque wrench set to 21 in-lb.

About Installing the Upgrade

Products affected All N5224/5/7B, N5244/5/7B 4-port microwave network
Installation to be performed by Keysight service center, personnel qualified by Keysight, or
Estimated installation time 3 hours
Estimated adjustment time 3 hours
Estimated full instrument calibration time 4.5 hours
analyzers
customer
Installation Note N5225-90143 7
-
Description of this Upgrade Items Included in the Upgrade Kit

Items Included in the Upgrade Kit

Check the contents of your kit against the following list. If any part is missing or damaged, contact Keysight Technologies. Refer to “Description of this
Upgrade” on page 3.
This section applies to multiple models and upgrade kits. Please refer to the model and upgrade kit for your instrument’s upgrade.
Extra quantities of items such as protective plastic caps, screws, cable ties, and cable clamps may be included in this upgrade kit. It is normal for some of these items to remain unused after the upgrade is completed.
Table 1 Contents of Upgrade Kit N5225-60143 Option 4S7 (DDS)
Ref Desg.
- Installation Note (this document) 1 N5225-90143
- China RoHS Addendum 1 9320-6722
A15 Direct digital synthesizer (DDS) assembly (Assembly contains four synthesizers) 1 N5240-60102
A55 Solid State Drive Assembly for 2.2 GHz CPU and Win10 in N52xxB 1 N5242-60135
Screw-Machine W/Crest-Cup-Con-Washer Pan-HD Torx-T10 M3X0.510mm-LG
-
SST-300 Passivated – (x2 bracket-center, stabilizer (all options), x2 short and rear panel bracket)
Screw-Machine W/Crest-Cup-Con-Washer Pan-HD Torx-T10 M3X0.5 8mm-LG
­SST-300 Passivated – (x2 secure DDS to chassis)
Description
Qty
4 0515-0374
4 0515-0380
Part Number
- Microwave Plug SMA Male Straight 50-Ohm 1 1250-4261
W1 Cable assy-RF, A15 13.5 GHz (SRC 1) direct digital synthesizer (J4)-SRC 1 RF IN 1 N5240-20124
W51/ W61
W2 Cable assy -RF, A15 13.5 GHz (SRC 2) direct digital synthesizer (J6)-SRC 2 RF IN 1 N5240-20126
W201 Cable assy –RF, A15 DD synthesizer (J7) to rear panel (J13) – (Option 422/423 only) 1 N5240-20127
W200 Cable assy, RF CA, SYNTH J11-SPAM J3 1 N5240-60115
W208 A70/A75 LFE board to A15 DD Synth Source 1 J12 to A70 LFE J20 1 N5240-60112
W209 A70/A75 LFE board to A15 DD Synth Source 2 J14 to A70 LFE J21 1 N5240-60114
W210 A70/A75 LFE board to A15 DD Synth LO J13 to A70 LFE J18 1 N5240-60113
8 Installation Note N5225-90143
Cable assy -RF, A15 13.5 GHz (LO) direct digital synthesizer (J5)-HMA RF IN
- Bracket, rear plate (Option 422/423 only) 1 N5240-00015
- Bracket-center, stabilizer, long 1 N5240-20136
1
N5240-20125
Description of this Upgrade Installation Procedure for the Upgrade
Table 1 Contents of Upgrade Kit N5225-60143 Option 4S7 (DDS)
Ref Desg.
- Label, SRC3 J13 – Rear panel – (Option 422/423 Only) 1 N5240-80042
Description
Qty
Part Number

Installation Procedure for the Upgrade

The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation.
This installation requires the removal of the analyzer’s protective outer covers. The analyzer must be powered down and disconnected from the mains supply before performing this procedure.
Installation Note N5225-90143 9
-
Description of this Upgrade Overview of the Installation Procedure

Overview of the Installation Procedure

The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation.
This installation requires the removal of the analyzer’s protective outer covers. The analyzer must be powered down and disconnected from the mains supply before performing this procedure.

All instruments:

“Step 1. Verify PNA and Erase Flash” on page 11
“Step 2. Remove the Inner and Outer covers” on page 13
“Step 3. Remove Some Cables (All Models) and Some Bottom Assemblies (Options 405, 420, and 425 Only)” on page 13
“Step 4. Remove Some Top Cables” on page 17
“Step 5. Remove the Left and Right Inner Bracket’s #4 Card Guides” on page 19
“Step 6. Install the Center Bracket” on page 20
“Step 7. Reinstall Bottom Cables” on page 21
“Step 8. Reinstall Top Assemblies Reinstall the A5 through A16 (Except A15 Direct Digital Synthesis (DDS) Assembly) and Cables” on page 22
“Step 9. Install *NEW* A15 Direct Digital Synthesizer (DDS) Assembly” on page 23
“Step 10. Install *NEW* and Reused Direct Digital Synthesis (DDS) Assembly Cables” on page 25
“Step 11. Route Gray Cables” on page 30
“Step 12. Add Back Plate and Cable to Rear Panel (Option 422/423 With and Without 029 Only)” on page 32
“Step 13. Install "SRC3 J13" Label on Rear Panel (Option 422/423 With and Without Option 029 Only)” on page 34
“Step 14. Remove and Replace the A55 Solid State Drive (SSD)” on page 34
“Step 15. Boot the Instrument” on page 35
“Step 16. Reinstall the Inner and Outer Covers” on page 35
“Step 17. Verify the PNA Analyzer Program is Running with the Correct Options” on page 35
“Step 18. Perform Post-Upgrade Adjustments and Calibration” on page 36
10 Installation Note N5225-90143
Description of this Upgrade Step 1. Verify PNA and Erase Flash
“Step 19. Prepare the PNA for the User” on page 39

Step 1. Verify PNA and Erase Flash

This section verifies the PNA boots normally and if OK, deletes the flash in preparation for installing and calibrating the new synthesizer assembly. Refer to Figure 1 on page 12.
1. Turn on your PNA.
2. Verify it boots up without errors.
If there are errors, refer to your instrument’s Chapter 4 "Troubleshooting" in the PDF Service Guide
3. If your PNA boots up normally.
4. Open the SCPI Parser Console: System > System Setup… > Remote Interface… > then click Show SCPI Parser Console
1
.
1. See “Downloading the Online PNA Service Guide” on page 5.
Installation Note N5225-90143 11
-
Description of this Upgrade Step 1. Verify PNA and Erase Flash
Figure 1 Open SCPI Parser Console
12 Installation Note N5225-90143
Description of this Upgrade Step 2. Remove the Inner and Outer covers
5. In the SCPI parser window: Enter the following command to erase the flash ROM:
DIAGnostic:FLASH:ERASe
For assistance with sending SCPI commands, refer to the PNA Help file on
www.keysight.com.
6. Enter the following query to verify the flash was erased:
DIAGnostic:FLASH:CATalog?
7. If the does not return any data, then power down the PNA and continue to the next step.
8. If the DIAG:FLASH:CAT? returns any data, repeat steps 5 and 6.
9. If you repeat steps 5 and 6. and still have data being returned, then refer to
“Contacting Keysight” on page 4.

Step 2. Remove the Inner and Outer covers

For instructions, click the Chapter 7 bookmark “Removing the Covers” in the PDF Service Guide
1
.

Step 3. Remove Some Cables (All Models) and Some Bottom Assemblies (Options 405, 420, and 425 Only)

Carefully, turn the instrument so it is bottom side up on a flat surface.
This section contains two sub-steps. Choose the ones that are applicable for your model’s options:
“Step 2a. Remove Some Cables (All Models)” on page 14
“Step 2b. Remove/Reinstall Some Bottom Assemblies and Old/New
Cables (LFE Models, With Options 405, 420, and 425)” on page 14
Be careful not to damage the center pins of the semi-rigid cables. Some flexing of the cables may be necessary but do not over-bend them.
Leave the gray flexible cables, the wire harnesses, and the ribbon cables connected where possible. Any that are removed should be labeled for reconnection later.
1. See “Downloading the Online PNA Service Guide” on page 5.
Installation Note N5225-90143 13