Keysight N5166B Start Guide

Page 1
Getting Started Guide
Keysight N5166B CXG N5171B/72B/73B EXG N5181B/82B/83B MXG X-Series Signal Generators
Page 2

Notices

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N5180-90054
Edition
Edition 1, January 2021
Supersedes: September 2019
Printed in USA/Malaysia
Published by: Keysight Technologies
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software
documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Page 3

Where to Find the Latest Information

Documentation is updated periodically. For the latest information about these products, including instrument software upgrades, application information, and product information, browse to one of the following URLs, according to the name of your product:
http://www.keysight.com/find/mxg
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing instrument damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Keysight Technical Support website at:
http://www.keysight.com/find/techsupport
3
Page 4
4
Page 5

Table of Contents

1. Safety Information
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
General Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Instrument Markings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2. Getting Started
Checking the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Verifying Pre-Installed Software License . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Signal Generator Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Meeting Environmental and Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Line Setting Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Connecting the AC Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Configuring the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Configuring for Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
LAN Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
GPIB Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Ordering Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Proper Use and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Cleaning Suggestions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Returning a Signal Generator to Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Contacting Keysight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3. Operation Verification
Running Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Self Test Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Frequency Range and Accuracy Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Frequency Counter Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Troubleshooting Problems with the Frequency Accuracy Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Checking the Output Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
N5171B/81B Test Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
N5171B/81B Alternative Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
N5166B/72B/82B Test Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
N5166B/72B/82B Alternative Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
N5173B/83B Test Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
N5173B/83B Alternative Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Troubleshooting Problems with the Output Power Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4. Regulatory Information
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5
Page 6
Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Statement of Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Compliance with Canadian EMC Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
South Korean Class A EMC Declaration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Compliance with European Machinery Directive Acoustic Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5. Open Source Licenses
Apple mDNSresponder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Sun ONC/RPC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
WCELIBCEX - Windows CE C Library Extensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6
Page 7

Documentation Overview

Getting Started Guide
User’s Guide
— Safety Information — Receiving the Instrument — Environmental & Electrical Requirements —Basic Setup — Accessories — Operation Verification — Regulatory Information
— Signal Generator Overview — Preferences & Enabling Options — Basic Operation —Optimize Performance — Avionics VOR/ILS (Option N5180302B) — Analog Modulation (Option UNT) — Pulse Modulation (
Options UNW or N5180320B)
— Basic Digital Operation—No BBG Option — Basic Digital Operation (
Options 653/655/656/657)
— Digital Signal Interface Module (Option 003 /004) — Baseband Operating Mode—Primary, BERT, or N5102A
— BERT (Option N5180UN7B) — Real–Time Noise—AWGN (Option N5180403B)
— Real–Time Phase Noise Impairments (Option N5180432B)
— Real–Time Fading (Option 660) — Custom Digital Modulation (
Option N5180431B)
— Multitone and Two-Tone Waveforms (Option N5180430B) — Troubleshooting — Working in a Secure Environment
Programming Guide
SCPI Reference
— Getting Started with Remote Operation — Using IO Interfaces — Programming Examples — Programming the Status Register System — Creating and Downloading Files — Creating and Downloading User–Data Files
— SCPI Basics — Basic Function Commands — System Commands — Analog Modulation Commands — Arb Commands — Avionics VOR/ILS Commands — Bit Error Rate Test (BERT) Commands — Digital Signal Interface Module Commands — Real-Time Commands
7
Page 8
Programming Compatibility Guide
— Provides a listing of SCPI commands and programming codes for
signal generator models that are supported by the Keysight CXG, EXG, and MXG X- Series signal generators.
Service Guide
— Troubleshooting — Replaceable Parts — Assembly Replacement — Post–Repair Procedures — Safety and Regulatory Information — Instrument History
Error Messages Guide
— Error Messages — Error Message Format — Error Message Type — List of Error Messages
Key Help
a
— Key function description — Related SCPI commands
a. Press the Help key, and then the key for which you wish help.
8
Page 9
Keysight Technologies X-Series Signal Generators
Getting Started Guide

1 Safety Information

— Warnings, Cautions, and Notes on page 9
— General Safety Considerations on page 10
— Instrument Markings on page 11

Warnings, Cautions, and Notes

The documentation for this product uses the following safety notations. Familiarize yourself with each notation and its meaning before operating the signal generator.
Warning denotes a hazard. It calls attention to a condition or situation that could result in personal injury or loss of life. Do not proceed beyond a warning until you fully understand the indicated conditions or situations.
Caution calls attention to a condition or situation that could result in damage to or destruction of the signal generator, or in the loss of a user’s settings or data. Do not proceed beyond a caution until you fully understand the indicated conditions.
Note calls the user’s attention to an important point or special information in the text.
9
Page 10
Safety Information General Safety Considerations

General Safety Considerations

If the signal generator is not used as specified, the protection provided by the equipment could be impaired. The signal generator must be used in a normal condition only, in which all means for protection are intact.
Personal injury may result if the signal generator covers are removed. There are no operator serviceable parts inside. To avoid electrical shock, refer servicing to qualified personnel.
Safety of any system incorporating the equipment is the responsibility of the assembler of the system.
10 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 11
Safety Information Instrument Markings

Instrument Markings

The signal generator may have the following markings. Familiarize yourself with each marking and its meaning before operating the signal generator.
This symbol marks the standby position of the power line switch.
This symbol indicates that the input power required is AC.
This symbol indicates Frame or chassis Terminal.
The instruction documentation symbol. The product is marked with this symbol when it is necessary for you to refer to instructions in the documentation.
This symbol identifies the Protective Conductor Terminal.
The CE mark is a registered trademark of the European Community. If this symbol is accompanied by a year, it is the year when the design was proven. It indicates that the product complies with all relevant directives.
The Keysight email address is required by EU directives applicable to our product.
The CSA mark is a registered trademark of the CSA International.
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
This is a combined marking to indicate product compliance with industry Canadian Interference-Causing Equipment Standard (ICES-001). This is also a symbol of an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 5).
ICES/NMB-001 This is a marking to indicate product compliance with the Industry Canadian
Interference-Causing Equipment Standard (ICES-001), Cet appareil ISM est conforme a la norme NMB du Canada.
This is the symbol of an Industrial Scientific and Medical Group 1 Class A Product. (CISPER 11, Clause 5)
This symbol indicates conformance to the standard specifications.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 11
Page 12
Safety Information Instrument Markings
China Restricted Substance Product Label. The EPUP (environmental protection use period) number in the center Indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
This WEEE symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE Directive, 2002/96/EC).
Universal recycling symbol. This symbol indicates compliance with the China standard GB 18455-2001 as required by the China RoHS regulations for paper/fiberboard packaging.
South Korean Certification (KC) mark. It includes the marking’s identifier code which follows this format: R-R-Kst-ZZZZZZZZZ
To view the EMC Declaration, see “South Korean Class A EMC
Declaration” on page 47
.
12 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 13
Keysight Technologies X-Series Signal Generators
Getting Started Guide

2 Getting Started

To avoid damaging or degrading the performance of the signal generator, do not exceed 27 dBm (0.5W) for N5173B/83B models, or 33 dBm (2W) for N5166B/N5171B/72B/81B/82B models, or maximum of reverse power levels at the RF input. See also Tips for Preventing Signal Generator Damage on
www.keysight.com.
Software Updates
Is your product software up-to-date? Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Keysight Technical Support website at
www.keysight.com/find/ss_firmware.
Documentation Updates
Is your product documentation up-to-date? To search for documentation updates for your product, go to the Keysight X-Series Signal Generators website at
www.keysight.com/find/X-Series_SG.
— Checking the Shipment on page 14
— Verifying Pre-Installed Software License on page 15
— Signal Generator Physical Characteristics on page 16
— Meeting Environmental and Electrical Requirements on page 17
— Configuring the Display on page 20
— Configuring for Remote Control on page 21
— Ordering Accessories on page 23
— Proper Use and Cleaning on page 25
— Returning a Signal Generator to Keysight Technologies on page 25
— Contacting Keysight on page 26
13
Page 14
Getting Started Checking the Shipment

Checking the Shipment

1. Inspect the shipping container for damage.
Signs of damage can include a dented or torn shipping container or cushioning material that indicates signs of unusual stress or compacting.
2. Carefully remove the contents from the shipping container and verify that your order is complete.
The following items are included with each signal generator:
— Getting Started Guide
— three-prong AC power cord specific to geographic location
3. Verify that the options/accessories you ordered are included with the
shipment by checking the serial number label on the rear of the signal generator and the packing literature included with the shipment.
See also, “Ordering Accessories” on page 23.
4. If a pre-installed software license is part of the instrument order, verify its installation using the directions outlined in “Verifying Pre-Installed
Software License” on page 15.
14 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 15
Getting Started Verifying Pre-Installed Software License

Veri f ying Pre-I nstalled Software License

If you purchased a signal generator where the order included the pre-installation of a fixed perpetual Signal Studio license (N76xx), or a Test Management Environment (TME) software license (N78xxA), or both, verify the license installation using the following key path:
Utility > Instrument Info > Options Info
This opens a menu with the choices: Instrument Options, Auxiliary Software Options, and Waveform Licenses. To view the license, press the appropriate menu key:
— For a Signal Studio license, press Instrument Options.
— For the TME license, press Auxiliary Software Options.
If a license is pre-installed, the software for that license still needs to be downloaded onto your PC. Use the following links to access the software for download and installation:
— Signal Studio software: www.keysight.com/find/signalstudio
—TME software: www.keysight.com/find/calibrationsoftware
If the software license(s) was ordered with the instrument for pre-installation and it does not show in the instrument, please contact Keysight Technologies:
www.keysight.com/find/contactus
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 15
Page 16
Getting Started Signal Generator Physical Characteristics

Signal Generator Physical Characteristics

Dimensions

88 mm H x 426 mm W x 489 mm L (length includes rear panel feet)
(3.5 in H x 16.8 in W x 19.2 in L)
Max length (L) including RF connector tip to end of rear panel feet is 508 mm (20 in)

We i g h t

N5171B/81B: 13.6 kg (30 lb)
N5166B/72B/82B: 15.9 kg (35 lb)
N5173B/83B: 14.5–15 kg (32–33 lb)
16 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 17
Getting Started Meeting Environmental and Electrical Requirements

Meeting Environmental and Electrical Requirements

To avoid the loss of data, GPIB settings, or current user instrument states that have not been permanently saved to non-volatile memory, the signal generator should always be powered down either via the signal generator's front panel power button or the appropriate SCPI command. Signal generator's installed in rack systems and powered down with the system rack power switch rather than the signal generator's front panel switch display a Error -310 due to the signal generator not being powered down correctly.
The input terminals for this product are classified as Measurement Category None.

Environment

— For indoor use only. —altitudes
Only with Option 660:
—0 to 55°C temperature, unless otherwise specified —Maximum Relative Humidity (non-condensing): 95%RH up to 40°C,
decreases linearly to 45%RH at 55°C
— Storage: –40 to 70°C
≤ 15,092 feet (4,600 meters)
≤ 9,843 feet (3,000 meters)
1
This product is designed for use in INSTALLATION CATEGORY II and POLLUTION DEGREE 2, per IEC 61010-1 Third Edition.
1. From 40°C to 55°C, the maximum % Relative Humidity follows the line of constant dew point.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 17
Page 18

Ventilation

Getting Started Meeting Environmental and Electrical Requirements
The instruments can operate with mains supply voltage fluctuations up to ± 10% of the nominal voltage.
Ventilation holes are located on the rear panel and all four sides of the signal generator cover. To ensure proper air flow through the signal generator, do not allow these holes to be obstructed.
Rack-mounted units
When installing the instrument(s) into a cabinet, consideration shall be given to the convection flow into and out of the cabinet. Consideration shall also be given to the individual instruments to avoid having the heated discharge of one instrument, now becoming the cooling intake air for another instrument.
Another area of concern is verification that the maximum ambient operating temperature of the instrument(s) is not exceeded by cabinet installation. Keysight recommends forced air convection whenever an instrument(s) are installed in a cabinet and further recommends that the maximum operating temperature of the cabinet be reduced 10°C from the lowest, of the maximum operating temperature of a single instrument.
If there are any concerns or special requirements a Keysight Field Engineer should be consulted to assure instrument(s) temperature compliance and performance.

Line Setting Requirements

The signal generator has autoranging line voltage input; ensure that the supply voltage is within the specified range.
Voltage: 100/120 volts nominal
Frequency: for 100/120 volts: 50/60/400 Hz nominal
Power: 160 W maximum (N5171B & N5181B)
220/240 volts nominal
for 220/240 volts: 50/60 Hz nominal
300 W maximum (N5166B, N5172B & N5182B)
280 W maximum (N5173B & N5183B)
18 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 19
Getting Started Meeting Environmental and Electrical Requirements

Connecting the AC Power Cord

This is a Safety Class 1 Product provided with a protective earth ground incorporated into the power cord. The front panel switch is only a standby switch; it is not a line switch. The AC power cord is the disconnecting device that disconnects the signal generator mains circuits from the mains supply. Alternatively, an external switch or circuit breaker, readily identifiable and easily reached by the operator, may be used as a disconnecting device.
The instrument shall be set flat in its normal use position so that the detachable power cord is readily identifiable and is easily reached by the operator. It shall not be placed on the four feet of the rear panel in normal use.
The mains plug shall be inserted only in a socket outlet provided with a protective earth contact. Always use the three-prong AC power cord supplied with the signal generator. Personal injury can occur if there is any interruption of the protective conductor inside or outside of the signal generator. Intentional interruption is prohibited.
The mains wiring and connectors shall be compatible with the connector used in the premise electrical system. Inadequate earth grounding can damage the signal generator. Always use the three-prong AC power cord supplied with the signal generator.
To avoid possible instrument damage, use the Keysight supplied power cord or one with same or better electrical rating.
Connect the AC power cord as follows:
1. Ensure that the power cord is not damaged.
2. Install the signal generator so that one of the following items is readily
identifiable and easily reached by the operator: AC power cord, alternative switch, or circuit breaker.
3. Insert the mains plug into a socket outlet provided with a protective earth grounding.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 19
Page 20
Getting Started
For details on a key, press Help and then the desired key.
Range: 0—100
Range: 1—12 hours, in 1 hour increments
Light Only turns the display light off, leaving the text visible at a low intensity.
Light & Text turns the display light and the text off.
If the display remains unchanged for long periods of time, use this mode to prevent the text from burning the display.
When on, commands executed through the remote control bus update the signal generator display accordingly.
The display blanks, except for the message Secure Display Activated, and the front panel keys are disabled. Cycle the instrument power to exit secure display mode.
For information on using the secure display, refer to the User’s Guide.
Dark text on a light background.
Instrument frequency information is hidden and front panel frequency control is disabled. Preset the instrument or cycle the power to re-enable the front panel.
For information on using the restricted display, refer to the User’s Guide.

Configuring the Display

Configuring the Display
Screen saver settings are persistent states; they are unaffected by preset or a power cycle.
Use the arrow keys, numeric keypad, or front panel knob to adjust numeric values.
X-Series signal generators are shipped from the factory with default display settings. When the signal generator display is not required to be active for long periods of time, consider using the Screen Saver Mode to extend the life of the display.
Figure 2-1 Display Softkeys
20 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
With the brightness set to minimum, the display may be too dark to see the softkeys. If this happens, use Figure 2-1 to locate the brightness softkey and adjust the value so that you can see the display.
Page 21
Getting Started
NOTES
Use a 100Base-T LAN cable to connect the signal generator to the LAN.
For details on using the instrument remotely, see the Programming Guide.
For details on a key, press Help and then the desired key.

Configuring for Remote Control

Configuring for Remote Control

LAN Configuration

Configuring the LAN Interface
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 21
Page 22
Getting Started
Use a browser to control the signal generator
Use a browser to view signal generator files.
For details on a key, press Help and then the desired key.
Select the desired GPIB language. See also the Programming Guide and the SCPI Command Reference.
NOTES USB is also available. It is not shown in the menu because it requires no configuration. For details on using the instrument remotely, see the Programming Guide.
Configuring for Remote Control
Enabling LAN Services: Browser, Sockets, and VXI-11

GPIB Configuration

22 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 23
Getting Started Ordering Accessories

Ordering Accessories

Only Keysight approved accessories should be used.
You can purchase accessories and find the latest documentation at:
http://www.keysight.com/find/mxg
If you do not have access to the Internet, please contact your Keysight field engineer.
See also, “Contacting Keysight” on page 26.
Description
Rack Mount Kit 1CM110A
Rack Mount Kit with Handles 1CP104A
Rack Slide Kit 1CR112A
Front Handle Kit 1CN106A
Transit Case (See
page 24
CXG Data Sheet (N5166B)
EXG Data Sheet (N5171B/72B/73B)
MXG Data Sheet (N5181B/82B/83B)
Getting Started Guide
User’s Guide
SCPI Reference
Programming Guide
a
Figure 2-2 on
)
b
b
b
b
b
b
b
Part Number
N5180-80053
5992-3959EN
5991-0039EN
5991-0038EN
N5180-90054
N5180-90056
N5180-90057
N5180-90074
Programming Compatibility Guide
Service Guide
Error Messages Guide
b
b
b
N5180-90069
N5180-90059
N5180-90075
a. For a description of the contents of each guide and
reference, see page 7.
b. Always refer to www.keysight.com for the
most up-to-date documentation.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 23
Page 24
Getting Started
Lid Insert
Base Insert
Standard PRV
Rotomold Case - Black
Standard Plastic Handl
Ordering Accessories
Figure 2-2 Transit Case
Retractable Handle
Casters
24 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 25
Getting Started Proper Use and Cleaning

Proper Use and Cleaning

The signal generator cover protects against physical contact with internal assemblies that contain hazardous voltages, but does not protect internal assemblies against contact with liquids. To avoid damage and personal injury, ensure that liquids are positioned away from the signal generator.
Personal injury may result if the signal generator is not used as specified. Unspecified use impairs the protection provided by the equipment. The signal generator must be used with all means for protection intact.

Cleaning Suggestions

To prevent electrical shock, disconnect the instrument from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Cleaning connectors with alcohol shall only be done with the instruments power cord removed, and in a well-ventilated room. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument.
To ensure good connections, regularly clean the instrument’s front and rear panel connectors with alcohol.
To prevent dust build-up that could potentially obstruct ventilation, periodically clean the instrument’s cover. Use a dry cloth or a cloth slightly dampened with water to clean the external case parts.

Returning a Signal Generator to Keysight Technologies

1. Gather as much information as possible about the signal generator’s problem.
2. Contact Keysight using the phone number that is specific to your geographic location. These phone numbers are listed on the Internet at
http://www.keysight.com/find/assist. If you do not have access to the
Internet, contact your Keysight field engineer.
After you provide information about the signal generator and its condition, you will receive information about where to ship your signal generator for repair.
3. Ship the signal generator in the original factory packaging materials, if available, or use similar packaging to properly protect the signal generator.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 25
Page 26
Getting Started Contacting Keysight

Contacting Keysight

Assistance with test and measurements needs, information on finding a local Keysight office, and information on purchasing accessories and documentation are available on the Internet at: http://www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field engineer.
In correspondence or telephone conversation, refer to the signal generator by its model number and full serial number. With this information, the Keysight representative can determine if your unit is still within its warranty period.
26 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 27
Keysight Technologies X-Series Signal Generators
Getting Started Guide

3 Operation Verification

To avoid damaging or degrading the performance of the signal generator, do not exceed 27 dBm (0.5W) for N5173B/83B models, or 33 dBm (2W) for N5166B/71B/72B/81B/82B models, or maximum of reverse power levels at the RF input. See also Tips for Preventing Signal Generator Damage on
www.keysight.com.
For the instrument to meet performance specifications allow a warm up period of 45 minutes within an operational temperature range of 0 to 55 °C. For more information, refer to the CXG, EXG, and MXG signal generator Data Sheets.
Operation verification is a series of tests used to confirm that the signal generator is operating properly, or to diagnose problems. Operation verification does not verify performance to instrument specifications.
Perform operation verification when you initially set up the signal generator, after a minor repair (refer to the Service Guide for details), or when the integrity of the signal generator is in question.
— Running Self Test on page 28
— Frequency Range and Accuracy Check on page 30
— Checking the Output Power on page 32
27
Page 28

Running Self Test

It takes about 5 minutes to run the self test.
Opens a table in which you can select specific tests and view details of those test.
Operation Verification Running Self Test
Self Test is a series of internal tests of signal generator functions. If this test fails, refer to “Self Test Failure” on page 29 for further instructions.
Use the following procedure to run self test:
1. Disconnect all external cables, including GPIB, LAN, and USB cables.
2. Preset the signal generator: Press Preset > Utility > Instrument Info > Self Test.
The following message appears:
— The current status of the self-test is: In complete.
Not all tests have been run.
3. Press Run Complete Self Test.
An activity bar displays on the screen indicating the test progress.
If you press Abort while self-test is running, the following message displays:
— The current status of the self-test is: Incom plete.
Not all tests have been run.
When self-test completes, one of the following messages displays:
— The current status of the self-test is : Passed — The current status of the self-test is : Failure. One
or more tests have failed. System diag nostics indicate this test as the root failure : xxx
If the signal generator fails only one test, the title of the failed test displays. If the signal generator fails more than one test, the test number of the most significant failure (root failure) displays.
28 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 29
Operation Verification
If Self Test fails, the summary indicates the most significant failure (root failure).
Utility > Instrument Info > Self Test
This information is supplemental. Please disregard unless Keysight specifically requests this information.
Running Self Test
NOTE The root failure is the error to report to Keysight Support.

Self Test Failure

1. Confirm that all external cables, including GPIB, LAN, and USB cables, are disconnected from the signal generator and repeat the self-test.
2. If the self-test continues to fail, the signal generator requires service. If you are unable to service the signal generator, send it to an Keysight service center for repair. Include a detailed description of the most significant failure (root failure) and any displayed error messages. Also include the model, serial number, installed options, and firmware version.
See “Returning a Signal Generator to Keysight Technologies” on page 25 for return instructions.
Refer to “Contacting Keysight” on page 26.
Viewing Test Results
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 29
Page 30
Operation Verification Frequency Range and Accuracy Check

Frequency Range and Accuracy Check

The frequency range is tested by determining the frequency accuracy relative to the timebase at the frequency limits of the signal generator. This test can be performed with a frequency counter that meets the frequency accuracy limits in Table 3-1.

Frequency Counter Procedure

Test Setup 1. Connect the equipment as shown.
2. Preset the signal generator: Press Preset.
3. Turn modulation off: Press the Mod On/Off so that the MOD On/Off LED
turns off.
4. Set the amplitude: Press Amplitude and enter 0 dBm.
5. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
6. Verify that the frequency counter is locked to the 10 MHz external
reference frequency (±1 Hz).
7. For maximum accuracy, set the gate time on the frequency counter to >5 seconds. (For example, press Gate & ExtArm twice and use the arrow keys to set the value.)
8. Set the frequency: Press Frequency and set the signal generator to the first frequency listed in Table 3-1 .
9. Confirm that the measured frequency is within the limits listed.
10.Repeat step 8 and step 9 for all of the frequencies in the table that are
within the frequency range of your signal generator.
For frequencies <200 MHz, use Channel 3 on the frequency counter (press Freq Ratio until
CH3: displays).
30 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 31
Operation Verification Frequency Range and Accuracy Check
Table 3-1 Frequency Accuracy Limits
Frequency (MHz) Limit (Hz)
0.01 MHz ±2
0.1 MHz ±2
200 MHz ±2
300 MHz ±2
500 MHz ±2
1000 MHz ±2
2000 MHz ±2
3100 MHz ±2
6000 MHz ±2
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 31
Page 32
Operation Verification Checking the Output Power

Troubleshooting Problems with the Frequency Accuracy Check

— Verify the cables are connected correctly.
— If you are using a frequency counter, verify that you are using the correct
channel for the frequencies you are measuring.

Checking the Output Power

This test verifies that the CW output power from the signal generator is within defined limits. The following table lists the preferred equipment for this test.
Test Equipment Recommended Model
Power Meter Keysight E4418B or E4419A/B E-Series or equivalent
Power Sensor, Input: Type-N (m)
Keysight E9304A
— N5171B/81B Test Procedure on page 33
— N5171B/81B Alternative Test Procedure on page 34
— N5166B/72B/82B Test Procedure on page 36
— N5166B/72B/82B Alternative Test Procedure on page 39
— N5173B/83B Test Procedure on page 40
— N5173B/83B Alternative Test Procedure on page 41
If this test fails, refer to “Troubleshooting Problems with the Output Power
Check” on page 43 for further instructions.
32 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 33
Operation Verification Checking the Output Power

N5171B/81B Test Procedure

Test Setup 1. Connect the equipment as shown:
2. Zero and calibrate the power sensor.
USB U2000A Series Power Sensors do not require the sensor to be zeroed or calibrated.
3. Connect the power sensor to the RF output of the signal generator. Once the LED on the power sensor turns off, the sensor is ready to measure.
4. Preset the signal generator: Press Preset
5. Turn the internal power meter channel on: Press Aux Fctn > Power Meter Measurements > Channel A On (or Channel B On).
6. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
7. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
off.
8. Set the frequency: Press Frequency and enter the first frequency value listed in Tab le 3-2.
9. Set the amplitude: Press Amplitude and enter the amplitude value for that frequency.
10.Measure the output power level.
11.Repeat steps 8 through 10 to measure power at each of the frequencies
listed in Tab le 3-2.
12.Confirm that the measured power levels are within the limits listed in the table.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 33
Page 34
Operation Verification Checking the Output Power

N5171B/81B Alternative Test Procedure

If a USB power sensor is not available, use a power meter to measure the output power of the signal generator.
Test Setup 1. Zero and calibrate the power sensor to the power meter:
2. Connect the equipment as shown:
3. Preset the signal generator: Press Preset.
4. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
5. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
off.
6. Set the frequency: Press Frequency and enter the first frequency value listed in Tab le 3-2.
7. Set the amplitude: Press Amplitude and enter the amplitude value for that frequency.
8. Configure the power meter for the measurement.
a. Press the Frequency Cal Fac button on the power meter.
b. Select a power meter channel (if applicable).
c. Use the arrow keys to enter the frequency at which to measure the
power.
9. Measure the output power level.
10.Repeat steps 6 through 9 to measure power at each of the 15 frequencies
listed in Tab le 3-2.
11.Confirm that the measured power levels are within the limits listed in the table.
Limit values are due to power meter uncertainty.
34 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 35
Operation Verification Checking the Output Power
Table 3-2 Leveled Output Power Limits
N5171B/81B Output Power
Frequency Amplitude (dBm) Limits
125 MHz 7 ±2
275 MHz 7 ±2
338 MHz 7 ±2
425 MHz 7 ±2
538 MHz 7 ±2
675 MHz 7 ±2
850 MHz 7 ±2
1075 MHz 7 ±2
1350 MHz 7 ±2
(dB)
1700 MHz 7 ±2
2150 MHz 7 ±2
2700 MHz 7 ±2
3400 MHz 7 ±2
4300 MHz 7 ±2
5400 MHz 7 ±2
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 35
Page 36
Operation Verification Checking the Output Power

N5166B/72B/82B Test Procedure

Test Setup 1. Connect the equipment as shown:
2. Zero and calibrate the power sensor.
USB U2000A Series Power Sensors do not require the sensor to be zeroed or calibrated.
3. Connect the power sensor to the RF output of the signal generator. Once the LED on the power sensor turns off, the sensor is ready to measure.
Without Modulation
4. Preset the signal generator: Press Preset
5. Turn the internal power meter channel on: Press Aux Fctn > Power Meter Measurements > Channel A On (or Channel B On).
6. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
7. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
off.
8. Set the frequency: Press Frequency and enter the first frequency value listed in Tab le 3-3.
9. Set the amplitude: Press Amplitude and enter the amplitude value for that frequency.
10.Measure the output power level.
11.Repeat steps 8 through 10 to measure power at each of the frequencies
listed in Tab le 3-3.
12.Confirm that the measured power levels are within the limits listed in the table.
a. Set the signal generator frequency to the first value listed in Table
3-3:
Press Frequency > 250 > MHz.
b. Set the amplitude to 7 dBm:
Press Amplitude > 7 > dBm.
c. If using a power meter, configure the power meter as follows:
36 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 37
Operation Verification Checking the Output Power
1. On the power meter, press the Frequency Cal Fac button.
2. If applicable, select a power meter channel.
3. Use the arrow keys to enter the frequency at which to
measure the power.
d. Measure the output power level.
e. Repeat steps a through d for the remaining frequencies in the table,
and confirm that the power level at each point is within limits.
Table 3-3 Output Power without Modulation
N5166B/72B/82B Output Power without
Modulation
Frequency
(MHz)
250 7 ±2
338 7 ±2
425 7 ±2
538 7 ±2
675 7 ±2
850 7 ±2
1075 7 ±2
1350 7 ±2
1700 7 ±2
2150 7 ±2
2700 7 ±2
3400 7 ±2
4300 7 ±2
Amplitude
(dBm)
Limits
(dB)
5400 7 ±2
With Modulation 13.Preset the signal generator: Press Preset.
14.Select the factory-supplied waveform SINE_TEST_WFM:
a. Press Mode > Dual ARB > Select Waveform. b. Highlight the SINE_TEST_WFM waveform.
c. Press Select Waveform.
15.Turn the arbitrary waveform player on: Press the ARB softkey to highlight On.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 37
Page 38
Operation Verification Checking the Output Power
16.Set the frequency to the first value listed in Table 3-4 :
Press Frequency > 250 > MHz.
17.Set the amplitude to 7 dBm:
Press Amplitude > 7 > dBm.
18.If using a power meter, configure the power meter as follows:
a. On the power meter, press the Frequency Cal Fac button.
b. Select a power meter channel (if applicable).
c. Use the arrow keys to enter the frequency at which to measure the
power.
19.Measure the output power.
20.Repeat steps 16 through 19 for the remaining frequencies listed in
Table 3-4, and confirm that the power level at each point is within limits.
Table 3-4 Output Power with Modulation
N5166B/72B/82B Output Power with
Modulation
Frequency
(MHz)
250 7 ±2
323 7 ±2
512 7 ±2
814 7 ±2
1275 7 ±2
2025 7 ±2
2750 7 ±2
3750 7 ±2
5250 7 ±2
Amplitude
(dBm)
Limits
(dB)
38 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 39
Operation Verification Checking the Output Power

N5166B/72B/82B Alternative Test Procedure

If a USB power sensor is not available, use a power meter to measure the output power of the signal generator.
Test Setup 1. Zero and calibrate the power sensor to the power meter:
2. Connect the equipment as shown below:
3. Preset the signal generator: Press Preset.
4. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
5. Turn modulation off:
Press Mod On/Off so that the Mod On/Off LED turns off.
6. Continue the procedure from “Without Modulation” on page 36.
7. Continue the procedure from “With Modulation” on page 37 to complete
the output power measurements of the N5166B/72B/82B.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 39
Page 40
Operation Verification Checking the Output Power

N5173B/83B Test Procedure

Test Setup 1. Connect the equipment as shown:
2. Zero and calibrate the power sensor.
USB U2000A Series Power Sensors do not require the sensor to be zeroed or calibrated.
3. Connect the power sensor to the RF output of the signal generator. Once the LED on the power sensor turns off, the sensor is ready to measure.
4. Preset the signal generator: Press Preset
5. Turn the internal power meter channel on: Press Aux Fctn > Power Meter Measurements > Channel A On (or Channel B On).
6. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
7. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
off.
8. Set the frequency: Press Frequency and enter the first frequency value listed in Tab le 3-6.
9. Set the amplitude: Press Amplitude and enter the amplitude value for that frequency.
10.Measure the output power level.
11.Repeat steps 8 through 10 to measure power at each of the frequencies
listed in Tab le 3-6.
12.Confirm that the measured power levels are within the limits listed in the table.
40 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 41
Operation Verification
*Refer to
*Refer to
Checking the Output Power

N5173B/83B Alternative Test Procedure

If a USB power sensor is not available, use a power meter to measure the output power of the signal generator.
Test Setup 1. Zero and calibrate the power sensor to the power meter:
2. Connect the equipment as shown:
3. Preset the signal generator: Press Preset.
4. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
5. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
off.
6. Set the frequency: Press Frequency and enter the first frequency value listed in Tab le 3-6.
7. Set the amplitude: Press Amplitude and enter the amplitude value for that frequency.
8. Configure the power meter for the measurement.
a. Press the Frequency Cal Fac button on the power meter.
b. Select a power meter channel (if applicable).
c. Use the arrow keys to enter the frequency at which to measure the
power.
9. Measure the output power level.
10.Repeat steps 6 through 9 to measure power at each of the 15 frequencies
listed in Tab le 3-6.
11.Confirm that the measured power levels are within the limits listed in the table.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 41
Page 42
Operation Verification Checking the Output Power
Table 3-5 Power Sensors by Frequency and Options
N5173B/83B Frequency Power Sensor
Option 513/520/540 < 5GHz E9304A
Option 520 > 5GHz 8485A
Option 532/540 > 5GHz 8487A
Table 3-6 Leveled Output Power Limits
N5173B/83B Output Power
Frequency Amplitude (dBm) Standard
Power
513/520 532/540 513/520 532/540
200 MHz 11 7 15 14 ±2
300 MHz 11 7 15 14 ±2
500 MHz 11 7 15 14 ±2
800 MHz 11 7 15 14 ±2
1.0 GHz 11 7 15 14 ±2
2.0 GHz 11 7 15 14 ±2
3.1 GHz 11 7 15 14 ±2
5.0 GHz 11 7 18 14 ±2
10 GHz 11 7 18 14 ±2
20 GHz (Option
520)
31.8 GHz (Option
532)
-- 7 -- 13 ±2
-- 7 -- 13 ±2
Amplitude (dBm) Option 1EA
Limits
ab
(dB)
40 GHz (Option
540)
-- 7 -- 12 ±2
a. Limit values are due to power meter uncertainty. b. For questions around measurement uncertainty, refer to the Data Sheet.
42 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 43
Operation Verification Checking the Output Power

Troubleshooting Problems with the Output Power Check

— Verify that you are using the appropriate power sensor.
— Normally, power sensor calibration factors are automatically downloaded to
the power meter when the power meter turns on. If this does not occur, manually enter the correct calibration factors for the power sensor you are using.
— Verify that the power sensor is properly calibrated to the power meter.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 43
Page 44
Operation Verification Checking the Output Power
44 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 45
Keysight Technologies X-Series Signal Generators
Getting Started Guide

4 Regulatory Information

— General on page 46
— Certification on page 46
— Assistance on page 46
— Statement of Compliance on page 46
— Compliance with Canadian EMC Requirements on page 46
— South Korean Class A EMC Declaration on page 47
— Compliance with European Machinery Directive Acoustic Requirement on
page 47
45
Page 46

General

Certification

Regulatory Information General
This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.

Assistance

Product maintenance agreements and other customer assistance agreements are available for Keysight Technologies products. For assistance, contact Keysight Technologies (see page 26).

Statement of Compliance

This product(s) complies with applicable Safety and EMC regulations and directives. A copy of the manufacturer’s Declaration of Conformity for this product(s) can be obtained by contacting your local Keysight Technologies sales representative.

Compliance with Canadian EMC Requirements

This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB du Canada.
46 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 47
Regulatory Information South Korean Class A EMC Declaration

South Korean Class A EMC Declaration

This equipment has been conformity assessed for use in business environments. In a residential environment this equipment may cause radio interference.

Compliance with European Machinery Directive Acoustic Requirement

Table 4-1
Acoustic statement (European Machinery Directive 2002/42/EC, 1.7.4.2u)
Acoustic noise emission
LpA < 70 dB
Operator position
Normal operation mode
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 47
Page 48
Regulatory Information Compliance with European Machinery Directive Acoustic Requirement
48 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 49
Keysight Technologies X-Series Signal Generators
Getting Started Guide

5 Open Source Licenses

This chapter contains the licenses for open source software used in the Keysight CXG, MXG and EXG X-Series signal generators.
49
Page 50
Open Source Licenses Apple mDNSresponder

Apple mDNSresponder

Apache License
Version 2.0, January 2004
http://www.apache.org/licenses/
TERMS AND CONDITIONS FOR USE, REPRODUCTION, AND DISTRIBUTION
1. Definitions.
"License" shall mean the terms and conditions for use, reproduction,
and distribution as defined by Sections 1 through 9 of this document.
"Licensor" shall mean the copyright owner or entity authorized by
the copyright owner that is granting the License.
"Legal Entity" shall mean the union of the acting entity and all
other entities that control, are controlled by, or are under common
control with that entity. For the purposes of this definition,
"control" means (i) the power, direct or indirect, to cause the
direction or management of such entity, whether by contract or
otherwise, or (ii) ownership of fifty percent (50%) or more of the
outstanding shares, or (iii) beneficial ownership of such entity.
"You" (or "Your") shall mean an individual or Legal Entity
exercising permissions granted by this License.
"Source" form shall mean the preferred form for making modifications,
including but not limited to software source code, documentation
source, and configuration files.
"Object" form shall mean any form resulting from mechanical
transformation or translation of a Source form, including but
not limited to compiled object code, generated documentation,
and conversions to other media types.
"Work" shall mean the work of authorship, whether in Source or
Object form, made available under the License, as indicated by a
50 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 51
Open Source Licenses Apple mDNSresponder
copyright notice that is included in or attached to the work
(an example is provided in the Appendix below).
"Derivative Works" shall mean any work, whether in Source or Object
form, that is based on (or derived from) the Work and for which the
editorial revisions, annotations, elaborations, or other modifications
represent, as a whole, an original work of authorship. For the purposes
of this License, Derivative Works shall not include works that remain
separable from, or merely link (or bind by name) to the interfaces of,
the Work and Derivative Works thereof.
"Contribution" shall mean any work of authorship, including
the original version of the Work and any modifications or additions
to that Work or Derivative Works thereof, that is intentionally
submitted to Licensor for inclusion in the Work by the copyright owner
or by an individual or Legal Entity authorized to submit on behalf of
the copyright owner. For the purposes of this definition, "submitted"
means any form of electronic, verbal, or written communication sent
to the Licensor or its representatives, including but not limited to
communication on electronic mailing lists, source code control systems,
and issue tracking systems that are managed by, or on behalf of, the
Licensor for the purpose of discussing and improving the Work, but
excluding communication that is conspicuously marked or otherwise
designated in writing by the copyright owner as "Not a Contribution."
"Contributor" shall mean Licensor and any individual or Legal Entity
on behalf of whom a Contribution has been received by Licensor and
subsequently incorporated within the Work.
2. Grant of Copyright License. Subject to the terms and conditions of
this License, each Contributor hereby grants to You a perpetual,
worldwide, non-exclusive, no-charge, royalty-free, irrevocable
copyright license to reproduce, prepare Derivative Works of,
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 51
Page 52
Open Source Licenses Apple mDNSresponder
publicly display, publicly perform, sublicense, and distribute the
Work and such Derivative Works in Source or Object form.
3. Grant of Patent License. Subject to the terms and conditions of
this License, each Contributor hereby grants to You a perpetual,
worldwide, non-exclusive, no-charge, royalty-free, irrevocable
(except as stated in this section) patent license to make, have made,
use, offer to sell, sell, import, and otherwise transfer the Work,
where such license applies only to those patent claims licensable
by such Contributor that are necessarily infringed by their
Contribution(s) alone or by combination of their Contribution(s)
with the Work to which such Contribution(s) was submitted. If You
institute patent litigation against any entity (including a
cross-claim or counterclaim in a lawsuit) alleging that the Work
or a Contribution incorporated within the Work constitutes direct
or contributory patent infringement, then any patent licenses
granted to You under this License for that Work shall terminate
as of the date such litigation is filed.
4. Redistribution. You may reproduce and distribute copies of the
Work or Derivative Works thereof in any medium, with or without
modifications, and in Source or Object form, provided that You
meet the following conditions:
(a) You must give any other recipients of the Work or
Derivative Works a copy of this License; and
(b) You must cause any modified files to carry prominent notices
stating that You changed the files; and
(c) You must retain, in the Source form of any Derivative Works
that You distribute, all copyright, patent, trademark, and
attribution notices from the Source form of the Work,
52 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 53
Open Source Licenses Apple mDNSresponder
excluding those notices that do not pertain to any part of
the Derivative Works; and
(d) If the Work includes a "NOTICE" text file as part of its
distribution, then any Derivative Works that You distribute must
include a readable copy of the attribution notices contained
within such NOTICE file, excluding those notices that do not
pertain to any part of the Derivative Works, in at least one
of the following places: within a NOTICE text file distributed
as part of the Derivative Works; within the Source form or
documentation, if provided along with the Derivative Works; or,
within a display generated by the Derivative Works, if and
wherever such third-party notices normally appear. The contents
of the NOTICE file are for informational purposes only and
do not modify the License. You may add Your own attribution
notices within Derivative Works that You distribute, alongside
or as an addendum to the NOTICE text from the Work, provided
that such additional attribution notices cannot be construed
as modifying the License.
You may add Your own copyright statement to Your modifications and
may provide additional or different license terms and conditions
for use, reproduction, or distribution of Your modifications, or
for any such Derivative Works as a whole, provided Your use,
reproduction, and distribution of the Work otherwise complies with
the conditions stated in this License.
5. Submission of Contributions. Unless You explicitly state otherwise,
any Contribution intentionally submitted for inclusion in the Work
by You to the Licensor shall be under the terms and conditions of
this License, without any additional terms or conditions.
Notwithstanding the above, nothing herein shall supersede or modify
the terms of any separate license agreement you may have executed
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 53
Page 54
Open Source Licenses Apple mDNSresponder
with Licensor regarding such Contributions.
6. Trademarks. This License does not grant permission to use the trade
names, trademarks, service marks, or product names of the Licensor,
except as required for reasonable and customary use in describing the
origin of the Work and reproducing the content of the NOTICE file.
7. Disclaimer of Warranty. Unless required by applicable law or
agreed to in writing, Licensor provides the Work (and each
Contributor provides its Contributions) on an "AS IS" BASIS,
WITHOUT WARRANTIES OR CONDITIONS OF ANY KIND, either express or
implied, including, without limitation, any warranties or conditions
of TITLE, NON-INFRINGEMENT, MERCHANTABILITY, or FITNESS FOR A
PARTICULAR PURPOSE. You are solely responsible for determining the
appropriateness of using or redistributing the Work and assume any
risks associated with Your exercise of permissions under this License.
8. Limitation of Liability. In no event and under no legal theory,
whether in tort (including negligence), contract, or otherwise,
unless required by applicable law (such as deliberate and grossly
negligent acts) or agreed to in writing, shall any Contributor be
liable to You for damages, including any direct, indirect, special,
incidental, or consequential damages of any character arising as a
result of this License or out of the use or inability to use the
Work (including but not limited to damages for loss of goodwill,
work stoppage, computer failure or malfunction, or any and all
other commercial damages or losses), even if such Contributor
has been advised of the possibility of such damages.
9. Accepting Warranty or Additional Liability. While redistributing
the Work or Derivative Works thereof, You may choose to offer,
and charge a fee for, acceptance of support, warranty, indemnity,
or other liability obligations and/or rights consistent with this
54 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 55
Open Source Licenses Apple mDNSresponder
License. However, in accepting such obligations, You may act only
on Your own behalf and on Your sole responsibility, not on behalf
of any other Contributor, and only if You agree to indemnify,
defend, and hold each Contributor harmless for any liability
incurred by, or claims asserted against, such Contributor by reason
of your accepting any such warranty or additional liability.
END OF TERMS AND CONDITIONS
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 55
Page 56

Sun ONC/RPC

Open Source Licenses Sun ONC/RPC
Portions of this software are copyright Sun Microsystems, Inc. and are licensed under the following terms:
=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*
Sun's ONC/RPC Copyright
=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*
* Sun RPC is a product of Sun Microsystems, Inc. and is provided for * unrestricted use provided that this legend is included on all tape * media and as a part of the software program in whole or part. Users * may copy or modify Sun RPC without charge, but are not authorized to * license or distribute it to anyone else except as part of a product * or program developed by the user. * * SUN RPC IS PROVIDED AS IS WITH NO WARRANTIES OF ANY KIND INCLUDING * THE WARRANTIES OF DESIGN, MERCHANTIBILITY AND FITNESS FOR A * PARTICULAR PURPOSE, OR ARISING FROM A COURSE OF DEALING, USAGE OR * TRADE PRACTICE. * * Sun RPC is provided with no support and without any obligation on * the part of Sun Microsystems, Inc. to assist in its use, correction, * modification or enhancement. * * SUN MICROSYSTEMS, INC. SHALL HAVE NO LIABILITY WITH RESPECT TO THE * INFRINGEMENT OF COPYRIGHTS, TRADE SECRETS OR ANY PATENTS BY SUN RPC * OR ANY PART THEREOF. * * In no event will Sun Microsystems, Inc. be liable for any lost * revenue or profits or other special, indirect and consequential * damages, even if Sun has been advised of the possibility of such * damages. * * Sun Microsystems, Inc. * 2550 Garcia Avenue * Mountain View, California 94043
56 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 57
Open Source Licenses WCELIBCEX - Windows CE C Library Extensions

WCELIBCEX - Windows CE C Library Extensions

The source code of the WCELIBCEX library is licensed under MIT License:
http://opensource.org/licenses/mit-license.php
Permission is hereby granted, free of charge, to any person obtaining a copy of this software and associated documentation files (the "Software"), to deal in the Software without restriction, including without limitation the rights to use, copy, modify, merge, publish, distribute, sublicense, and/or sell copies of the Software, and to permit persons to whom the Software is furnished to do so, subject to the following conditions:
The above copyright notice and this permission notice shall be included in all copies or substantial portions of the Software.
THE SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT. IN NO EVENT SHALL THE AUTHORS OR COPYRIGHT HOLDERS BE LIABLE FOR ANY CLAIM, DAMAGES OR OTHER LIABILITY, WHETHER IN AN ACTION OF CONTRACT, TORT OR OTHERWISE, ARISING FROM, OUT OF OR IN CONNECTION WITH THE SOFTWARE OR THE USE OR OTHER DEALINGS IN THE SOFTWARE.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide 57
Page 58
Open Source Licenses WCELIBCEX - Windows CE C Library Extensions
58 Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide
Page 59
Index
Index
A
AC
power cord, connecting
symbol accessories address, GPIB
11
23
22
alcohol, cleaning with altitude requirements
19
25
17
Australian Communications Authority (C-tick)
11
mark auto-IP
21
B
brightness adjustment
20
C
Canadian
EMC requirements
Standards Association (CSA) mark cautions certification
9
46, 50
checking the shipment cleaning suggestions contrast adjustment
46
11
14
25
20
D
DHCP
21
dimensions, signal generator display adjustment documentation
20
7
documentation, ordering
16
23
I
ICES symbol ICES-001 IEC Publication 61010 inspection, shipping container installation guide content instrument warm up interface, GPIB interface, LAN IP address, setting ISM1-A symbol
11
46
46, 57
14
7
27
22
21
21
11
K
Keysight, contacting
26
L
LAN configuration line setting requirements
21
18
LXI standard specifications symbol
M
manuals, content of manuals, ordering
7
23
N
noise requirements
47
O
operation verification operation, remote output power, checking
27
21
32
11
E
European Community (CE) trademark
F
frequency accuracy check
30
frequency range and accuracy check FTP server
22
G
German noise requirements GPIB address, setting guides, content of
22
7
47
H
hostname, setting humidity requirements
21
17
11
30
P
power
checking connecting disconnecting line cord requirements symbol
programming
32
19
19
19
18
11
21
programming guide content
R
references, content of regulatory information remote operation requirements
17, 18, 47
7
45, 49
21
7
59
Page 60
returning a signal generator 25
S
safety information
9
SCPI
reference content SCPI, enabling self test
28
server, enabling
7
22
22
service
guide content
7
service, Keysight sales and service offices shipping container, inspection shipping requirements sockets, enabling
25
22
South Korean Emc Requirements standby symbol
11
14
47
T
temperature requirements troubleshooting output power
17
32
26
U
user
documentation content
V
ventilation requirements verification, operation voltage requirements VXI-11, enabling
27
18
22
W
warnings web server
9
22
weight, signal generator
7
18
16
60
Page 61
This information is subject to change without notice. © Keysight Technologies 2012-2021
Edition 1, January 2021
N5180-90054
www.keysight.com
Loading...