KEYENCE FS-V1P User Manual

96M11226
Digital Calibration Fiberoptic Sensors
Warning
• This product is just intended to detect the object(s). Do not use this product for the purpose to protect a human body or a part of human body.
• This product is not intended for use as explosion-proof product. Do not use this product in a hazardous location and/or potentially explosive atmosphere.
FS-V1(P)
Instruction Manual
Read this manual before using the product in order to achieve maximum performance. Keep this manual in a safe place after reading it so that it can be used at any time.
Warning
NOTICE

PART NAMES AND FUNCTIONS

Output timer selector switch UP/DOWN key
Note: Outputs of both A and B are set to the same mode.
Calibration indicator
Output A indicator
Output B indicator
Output selector switch
*Use the UP/DOWN key together with the MODE selector button.
It indicates a hazardous situation which, if not avoided, could result in death or serious injury.
It indicates a situation which, if not avoided, could result in product damage as well as property damage.
OFF-delay 40 ms
ON-delay 40 ms
Timer OFF
AB
DARK-ON for both outputs A and B
LIGHT-ON for output A and DARK-ON for output B
LIGHT-ON for both outputs A and B
• Changes sensitivity setting value.
• Selects channel (expansion unit).
MODE selector button
• Selects displayed data.
• Excess gain
• Light intensity
• Setting value
• Selects channel (expansion unit). *
• Enables/disables key protect. *
MODE
CH
2
#
SET
SET button
Sets sensitivity.
LCD display monitor
Turns display red when the selected channel is ON. Turns display green when the selected channel is OFF.
UL Certificate
This product is an UL/C-UL Listed product.
• UL File No. E301717
• Category NRKH,NRKH7
• Enclosure Type 1 (Based on UL50)
Be sure to consider the following specifications when using this product as an UL/C-UL Listed Product.
• Use the power supply with Class 2 output defined in NFPA70 (NEC: National Electrical Code).
• Power supply/ Control input/ Control output circuits shall be con­nected to a single Class 2 source only.
• Use with the over current protection device which is rated 24 V or more and not more than 2 A.
Accessories
• Instruction manual (x 1)
• Mounting bracket (x 1)

SPECIFICATIONS

Model NPN output
Light source Red LED Response time 250 μs Operation mode LIGHT-ON/DARK-ON (switch-selectable) Indicators
Timer function ON-delay: 40 ms, OFF-delay: 40 ms,
External calibration input signal
Control output (2 outputs)
Protection circuit Reverse polarity protection, Over-current protection,
Power supply voltage 12 to 24 VDC ±10%, Ripple (P-P) 10% max., Class 2 Current consumption 50 mA max. Ambient illumination Candescent lamp: 10,000 lx max., Sunlight: 20,000 lx max. Ambient temperature -10 to +55°C (14 to 131°F), No freezing Relative humidity 35 to 85%, No condensation Vibration 10 to 55 Hz, 1.5 mm double amplitude in X, Y and
Shock immunity 500 m/s2 in X, Y and Z directions,
Housing material Body/Cover: Polycarbonate Weight Approx. 80 g (including 2 m cable)
1 A "P" following the model number indicates PNP-output type. 2 When several units are connected, the allowable ambient temperature changes depending
on the following conditions. To connect several units, be sure to mount them to a DIN rail (metal DIN rail). Make sure that the output current is 20 mA. max.
• When 3 to 10 units are connected: -10 to +50 °C
• When 11 to 16 units are connected: -10 to +45 °C
PNP output
Output indicators: 2 Red LED’s,
Received light intensity display: LCD (Red/Green LED back-lit)
Calibration indicator: Yellow LED,
Timer OFF (switch selectable)
NPN or PNP open-collector 24 V 100 mA max., Residual voltage : 1 V max.
FS-V1
FS-V1P
contact, solid state
Surge protection
Z directions for two hours
three times each

SELECTING DISPLAYED DATA

Excess gain (%)
CH
#
2
MODE
Press this button once.
Channel display: Indicates the channel of the expansion unit currently being monitored.
* The setting value flashes on and off.
(Factory-setting is Excess gain.)
MODE
Hold down this button for longer than 3 seconds.
Setting value (4 digits)
CH
#

Received light intensity (4 digits)
CH
#

MODE
Press this button once.
Displaying excess gain
Received light intensity is converted by defining the setting value as 100 P (%).
• Stable LIGHT status is obtained with 110 P (%) or more.
• Stable DARK status is obtained with 90 P (%) or less.
You can adjust the sensitivity in increments of 1%.
CH
# 2
¨ Refer to P.2.
Displaying received light intensity
Received light intensity is displayed approxi­mately 4000 is the maximum setting.
Note: The MAX and MIN values vary depending on the fiber unit con­nected.
CH
# 
Displaying a setting value
The current setting value is displayed. You can adjust the setting value in increments of 1 .
¨ Refer to P.2.
1

SETTING SENSITIVITY

Select the sensitivity setting procedure according to the target condi­tion.

CHANGE THE SETTING VALUE

You can change the setting value in “excess gain display” mode or “set­ting value display” mode.
For sensitivity adjustment using a moving target
z Fully-automatic Calibration
1. Pass a target through the optical axis while press­ing the SET button.
At least 3 seconds
2. Confirm that the calibra­tion indicator (yellow LED) flashes.
3. Release the SET button.
AB
CH
# 2
MODE
SET
The calibration indicator (yellow LED) goes off.
To detect a minute difference
z Two-point Calibration
1. With a target in place, press the SET button and release it. The calibration indicator (yel­low LED) lights.
2. With the target removed, press the SET button and release it. The calibration indicator (yellow LED) goes off.
For target positioning
z Positioning Calibration
1. With no target, press the SET button and release it. The calibration indicator (yellow LED) lights.
2. Place a target in the posi­tion where it is to be
At least 3 seconds
stopped.
3. Press the SET button until the calibration indicator flashes.
AB
CH
# 2
MODE
SET
4. Release the SET button.
For stable detection unaffected by dust
z Maximum Sensitivity Setting
1. Under the following con-
At least 3 seconds
ditions press the SET button until the calibration indicator (yellow LED) flashes.
2. Release the SET button. The calibration indicator (yellow LED) goes off.
AB
Reflective type Thrubeam type
With no target With target
CH
# 2
CUE: When the sensitivity difference is insufficient
If sensitivity has no allowance after calibration, “------” flashes for 3 seconds on the LCD display monitor.
Note: Sensitivity is set and registered even when the sensitivity differ­ence is insufficient. Be sure to confirm that detection is properly per­formed.
CUE: Locking the operation button (Key protect)
Hold down [MODE] and [UP] simultaneously for 3 seconds to lock the operation button. When the operation button is locked, “Loc” is dis­played in flashing letters.
To unlock the operation button, repeat the procedure above. “vnL” (unlocked) is displayed in flashing letters. Even when the operation button is locked, display data and channel can be changed.
MODE
SET
Increase sensitivity (=Decrease the setting value)
z In excess gain display mode The sensitivity increases by 1% each time you press [UP].
CH
# 2
MODE
SET
CH
# 2
MODE
SET
z In setting value display mode
The setting value decreases by 1 each time you press [DOWN].
CH
# 
MODE
SET
CH
# 
MODE
SET
Decrease sensitivity (=Increase the setting value)
The method for decreasing the sensitivity is the same as for increasing the sensitivity, except you must press [DOWN] instead.
CH
# 2
MODE
SET
CUE: Holding down [UP] or [DOWN] changes the setting value rapidly.
For sensitivity adjustment using an external signal
z External Calibration Function
1. Hold down [MODE] and [UP] simultaneously for 3 seconds to display “Loc”. (If “vnL” appears, repeat the same procedure.)
2. Connect the pink cable to a switch or PLC.
NPN
Brown
Blue
Pink
Contact or solid-state input
The minimum input time is 20 ms.
3. The external calibration procedure is the same as using the SET but­ton. Turn the input device on instead of pressing the SET button. Also, turn the input device off instead of releasing the SET button. (You can perform all the setting procedures described on the left.)
CUE: Calibration input during detection
You can re-adjust sensitivity during detection without stopping the operation. The sensor continues operation with the previous sensitivity setting until sensitivity readjustment is completed.
CUE: External calibration with several units
When several units are connected, the sensitivity of all the FS-T2(P) expansion units can be simultaneously set using the external calibra­tion input through the FS-V1(P). You can select a unit to be externally calibrated by setting its key pro­tection switch to the LOCK position. External calibration is effective for all the units which are locked. Set the key protection switch to LOCK on the FS-T2(P) expansion unit to be externally calibrated.
PNP
+V
0V
Brown
Blue
Pink
Contact or solid state input
+V
0V
2
Loading...
+ 2 hidden pages