Keithley 487, 486 User Manual

Model 486
PICOAMMETER
Model 487
PlCOAMMETER/SOURCE
QUICK REFERENCE GUIDE
INTRODUCTION
~21990, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Document Number: 486-903-01 Rev. A
1
Asyst ts a registered trademark of Asyst Software technologies.
Hewtett-Packard Is a registered trademark of Hewlett-Packard
Company.
IBM and AT are registered trademarks and PC/XT is a trademark of International Business Machines, I~C. PC&488 is a registered trademark of Capital Equipment Cow ration.
2
TABLE OF CONTENTS
SAFETY WARNINGS
ERROR MESSAGES ......................... 7
CURRENT MEASUREMENTS
VOLTAGE SOURCE (Model 487) .............. 14
V/l OHMS MEASUREMENTS (Model 487)
MENU .................................... 21
FRONT PANEL TRIGGERING ................. 22
EXTERNAL TRIGGERING .................... 24
DATA STORE OPERATION .................. 26
ANALOG OUTPUT .......................... 27
TALK-ONLY
IEEE-488 PROGRAMMING .................. .29
Device-dependent Commands ................. 29
Display Intensity ........................ 29
Reading Source ........................ 29
Zero Check and Correct .................. 29
Display V/lOhms Data Format
Hit Control ............................ 31
Self-Test EOI and Bus Hold-off
Default Conditions or Calibration ...........
SRQ
Data Store ............................ 32
Operate
Filters ................................ 33
Interval ............................... 33
Range................................3 3
Integration
...............................
.............................
...............................
.........................
..................
...........................
...........................
..............................
...................
..............................
............................
.......
19
28
..3 0 ..3 0
39 31
.31
32
..3 2
33
34
5 8
3
Trigger ...............................
Status.. ..............................
Voltage Source (487)
Delay ................................
Execute ............................
Terminator ..........................
Relative ..............................
Data Formats SRQ Mask and Serial Poll Byte StatusWords
..............................
............................
....................
................
i: l
35
..3 5
35 a
..3 5
36 36 40
..4 1
a
a
4
SAFETY WARNINGS
The following precautions should be observed before us-
ing Model 488/487. Refer to main manual for detailed
safety information and complete operating instructions.
The Model 48W487 is intended for use by qualified per­sonnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible in­jury. Read over the instruction manual carefully before
using the instrument.
Before operating the instrument, make sure the line cord
is connected to a properly grounded power receptacle.
Exercise extreme caution when a shock hazard is pre-
sent. Lethal voltages may be present on the test fixture or
the Model 487 output jacks. The American National Stan-
dards Institute (ANSI) states that a shock hazard exists
when voltage levels greater than 30V RMS or 42.4V peak
are present. A good safety practice is to expect that
hazardous voltage is present in any unknown circuit
before measuring.
Inspect the connecting cables, test leads, and jumpers
for possible wear, cracks, or breaks before each use.
5
For maximum safety, do not touch the Model 487 con­nections, test fixture, test cables or connections to any other instruments while power is applied to the circuit un­der test. Turn off all power and discharge all capacitors before connecting or disconnecting cables or jumpers. Also, keep the test fixture lid closed while power is ap­plied to the device under test. Safe operation requires the use of the lid interlock.
Do not touch any object which could provide a current
path to the common side of the circuit under test or power
line (earth) ground.
Do not exceed the maximum signal levels of the instru­ment, as shown on the rear panel and as defined in the
specifications and operation section of the instruction
manual.
Connect the chassis of the test fixture to safety earth ground using #I 8 AWG or larger wire.
Instrumentation and accessories should not be con-
nected to humans.
Maintenance should only be performed by qualified serv­ice personnel. Before performing any maintenance, dis­connect the line cord and all test cables from the instru­ment.
6
ERROR MESSAGES
Table 1. Error Messages
Message
NO DEFLTS UNCAL ROM ERROR
RAM ERROR TRG OVERUN
CAL VALUE CAL ERROR
CAL LOCK
CAL ZCHK INTERLOCK
Description
Checksum error in default condi­tion during power-up. Checksum error in calibration during power-up Failed ROM self-test. Failed RAM self-test. Trigger overrun; instrument trig­gered while still processing a pre­vious trigger. Calibration value conflicts with the currently selected range. Calibration value not within allow­able limits. Sent calibration command with calibration switch (CAL LOCK) disabled.
Cannot calibrate picoammeter with zero check enabled. Tried to put V-Source of Model
487 in operate while an enabled
safety interlock switch is open.
CURRENT MEASUREMENTS
Current Ranges
SHIFT
q
The available current ranges of the Model 4861487 are listed in Table 2.
Table 2. Current Ranges
v RANGE A
El 0
AUTO
RANGE
.
2nA
20nA
200nA
2c1A
20/.iA
2oopA
2mA
‘Higher voltages require current to be limited to 3mA.
2.19999nA
21.9999nA
219.999nA
2.19999pA
21.9999pA
219.999uA
2.19999inA
35OVdc 35OVdc 35OVdc 35OVdc
5OVdc 50Vdc 50Vdc
Autorange is enabled by pressing SHIFTand then AUTO
RANGE in that order.
To avoid possible damage to the instru­ment, do not apply more than 5OVdc to the input when autorange is enabled (unless an external series resistor to
limit current to 3mA is used).
CAUTION
Zero Check and Correct
ZERO
SHIFT
CHECK
q El
CORRECT
When zero check is enabled, the input amplifier is inter­nally disconnected from the input connector of the instru­ment. A 1 OOkG resistor shunts the input connector, and the instrument displays the offset of the selected range.
When zero correction is performed, the offset is meas­ured and algebraically subtracted from every subsequent reading (including zero checked readings).
Zero check is enabled by pressing ZERO CHECK. Pressing ZERO CHECK a second time disables the fea­ture.
9
When zero correction is performed, only the present
range is zero corrected. Before making measurements, it is recommended that each current range be zero cor­rected.
To perform zero correction:
1. Enable zero check.
2. To perform zero correction, press SHIFT and then CORRECT in that order.
Filters
SHIFT
cl
Filtering is used to stabilize noisy measurements. The Model 486/487 has two available filters; a digital filter and an analog filter. The digital filter bases the reading on the weighted average of a number of measurement conver­sions. The analog filter is a simple RC filter whose time constant varies with the selected range.
When the filter is enabled, the selected filter(s) (digital fil­ter, analog filter or both) will be used.
10
To check or change the filter selection:
Press SHIFT and then FILTER SELECT in that or­der. One of the following messages will be dis­played to indicate the currently selected filter(s):
FILTER DIGITAL FILTER ANALOG FILTER DIG+AN
To select a different filter, use the rotary knob or the
4 and b keys to display the desired filter selection.
Enter the displayed filter selection by pressing SHIFT and then FILTER SELECT.
(Digital filter selected) (Analog filter selected) (Both digital and analog
filters selected)
Rel
REL
0
q
The rel (relative) feature serves as a means of baseline suppression by allowing a stored offset value to be sub­tracted from subsequent readings. When rel is enabled, the instrument takes the currently displayed reading as a baseline value. All subsequent readings represent the difference between the applied signal level and the stored baseline.
11
A baseline can be established for both current and VII ohm measurements and is “remembered” by both func­tions.
Rel is enabled by pressing the REL key. Pressing REL a
second time disables the feature.
Making Current Measurements
To make current measurements:
1. Select a current range that is appropriate for the ex­pected measurement or enable autorange (press SHIFT and then AUTO RANGE).
2. Enable zero check and, if the display is not zeroed, perform zero correction by pressing SHIFT and then CORRECT.
3. Connect the current to be measured to the input of the Model 486/487 (see Figures 1 and 2).
4. Disable zero check and read the measured current on the display of the Model 486/487.
12
Input Connector
Max Overload
2nA - 2pA Range; 35OVdc
2OpA - 2mA Range;
50 Vdc
Chassis Gmund
Fiaure 1. Triax InDut Connector (3-Lua)
Model 486
Figure 2. Picoammeter Connections
VOLTAGE SOURCE
(Model 487)
The voltage source may float. However,
CAUTION
neither V-Source terminal may ever be more than 500V above chassis.
V-SOURCE OUTPUT
500V MAX TO ,& OR.+
I HI
LO 1
5oov MAX
Figure 3. Model 487 V-Source Output Terminals
Table 3. Model 487 V-Source
14
Loading...
+ 33 hidden pages