Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defecti ve.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility . Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechar geable
batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT , INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF
THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS
BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION,
LOSSES SUSTAINED AS THE RESULT OF INJURY T O ANY PERSON, OR DAMAGE T O PROPER TY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Le vel letter increases alphabetically as the manual under goes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially . When a new Re vision is created, all Addenda associated with the previous Re vision
of the manual are incorporated into the new Revision of the manual. Each ne w Revision includes
a revised copy of this print history page.
Revision A (Document Number 2420-902-01)................................................................April 1997
Addendum A (Document Number 2420-902-02).............................................................April 1997
Revision B (Document Number 2420-902-01)..................................................................July 1998
Revision C (Document Number 2420-902-01)............................................................ January 1999
Revision D (Document Number 2420-902-01).................................................................June 2000
Revision E (Document Number 2420-902-01) ............................................................October 2001
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
S
afety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body
ment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel
voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state
if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel
trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Installation Category I and Installation Category II,
as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data
I/O signals are Installation Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Installation Cate gory II connections require protection for high transient over -voltages often associated with
local A C mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures.
The American National Standards Institute (ANSI) states that a shock hazard exists when v oltage le vels greater than 30V RMS,
42.4V peak, or 60VDC are present.
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock.
If the circuit is capable of operating at or above 1000 volts,
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equip-
use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
perform routine procedures on the product to keep it operating properly, for example, setting the line
are trained to work on live circuits, and perform safe installations and repairs of products. Only properly
A good safety practice is to expect that hazardous voltage is present in any unknown
no conductive part of the circuit may be exposed.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry , insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of
normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
information very carefully before performing the indicated procedure.
CAUTION
The
ranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
T o maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer ,
test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from
other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only
through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability
of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for
proper cleaning/servicing.
heading in a manual explains dangers that might result in personal injury or death. Al ways read the associated
heading in a manual explains hazards that could damage the instrument. Such damage may inv alidate the war -
Use the procedures in this section to verify that Model 2420 accuracy is within the limits stated
in the instrument’s one-year accurac y specifications. You can perform these verification procedures:
•When you first receive the instrument to make sure that it was not damaged during shipment.
•To verify that the unit meets factory specifications.
•To determine if calibration is required.
•Following calibration to make sure it was performed properly.
WARNING
NOTE
The information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, which
could cause personal injury or death if contacted. Use standard safety precautions when working with hazardous voltages.
If the instrument is still under warranty and its performance is outside specified limits, contact your Keithle y r epr esentative or the factory to determine the correct course of action.
V erification test requirements
Be sure that you perform the verification tests:
•Under the proper environmental conditions.
•After the specified warm-up period.
•Using the correct line voltage.
•Using the proper test equipment.
•Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•An ambient temperature of 18-28°C (65-82°F).
•A relative humidity of less than 70% unless otherwise noted.
W arm-up period
Allow the Model 2420 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. T ypically , allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specifi ed temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manuf acturer .
Line power
The Model 2420 requires a line voltage of 90 to 250V and a line frequency of 50 to 60Hz.
Verification tests must be performed within this range.
Recommended test equipment
T able 1-1 summarizes recommended verification equipment. You can use alternate equipment
as long as that equipment has specifications at least as good as those listed in Table 1-1. Keep in
mind, howev er, that test equipment uncertainty will add to the uncertainty of each measurement.
Generally, test equipment uncertainty should be at least four times better than corresponding Model
2420 specifications. Table 1-1 lists the specifications of the recommended test equipment, including maximum allowable uncertainty for alternate test equipment, which is shown in parentheses.
Table 1-1
Recommended verification equipment
DescriptionManufacturer/ModelSpecifications
Digital MultimeterHewlett Packard
HP3458A
DC Voltage*1V:
Performance Verification 1-3
±5.6ppm
10V:
100V:
±4.3ppm
±6.3ppm
µ
DC Current*10
Resistance CalibratorFluke 5450AResistance**1.9
Precision Resistor*** Isotec RUG-Z-1R00-0.11
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of
alternate test equipment shown in parentheses.
*** Required for verification of 3A current range. Characterize resistor to ±300ppm or better using recommended DMM
before verifying 3A current measurement range.
A:
µ
100
1mA:
10mA:
100mA:
1A:
Ω
Ω
19
Ω
190
1.9k
Ω
19k
190k
1.9M
19M
Ω
, ±0.1%, 100W
±25ppm
A:
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
:
±65ppm (±460ppm)
:
±23ppm (±280ppm)
:
±10.5ppm (±230ppm)
Ω
:
±8ppm (±200ppm)
:
±7.5ppm (±195ppm)
Ω
:
±8.5ppm (±200ppm)
Ω
:
±11.5ppm (±180ppm)
Ω
:
±30ppm (±635ppm)
1-4Performance Verification
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before verifying
the 3A current measurement range. (You need not characterize the resistor if you are checking
only the 3A current source range.) Use the 4-wire ohms function of the DMM recommended in
T able 1-1 to measure the resistance value, and then use that measured value to calculate the cur rent during the 3A current measurement range test procedure.
V erification limits
The verification limits stated in this section have been calculated using only the Model 2420
one-year accuracy specifications, and they do not include test equipment uncertainty. If a particular measurement falls outside the allowable range, recalculate new limits based on Model
2420 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 20V DC
output range using a 20V output value. Using the Model 2420 20V range one-year accuracy
specification of ±(0.02% of output + 2.4mV offset), the calculated output limits are:
Output limits = 20V ± [(20V
Output limits = 20V ± (0.004 + 0.0024)
Output limits = 20V ± 0.0064V
Output limits = 19.9936V to 20.0064V
Resistance limits calculation
When verifying the resistance measurement accuracy, it will probably be necessary to recalculate resistance limits based on the actual calibrator resistance values. You can calculate resistance reading limits in the same manner described above, b ut be sure to use the actual calibrator
resistance values and the Model 2420 normal accuracy specifications for your calculations.
As an example, assume you are testing the 20k
inal 19k
accuracy specifications of ±(0.063% of reading + 3
Ω
calibrator resistor is 19.01kΩ. Using the Model 2420 20kΩ range one-year normal
Reading limits = 19.01k
Reading limits = 19.01k
Reading limits = 18.9950kΩ to 19.0250k
Ω
Ω
×
0.02%) + 2.4mV]
Ω
range, and that the actual value of the nom-
Ω
), the recalculated reading limits are:
± [(19.01kΩ × 0.063%) + 3Ω]
±15
Ω
Ω
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory front panel
(bench) defaults as follows:
1.Press the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL
2.Select SAVESETUP, and then press ENTER. The unit then displays:
SAVESETUP MENU
GLOBAL SOURCE-MEMORY
3.Select GLOBAL, and then press ENTER. The unit then displays:
GLOBAL SETUP MENU
SAVE RESTORE POWERON RESET
4.Select RESET, and then press ENTER. The unit displays:
RESET ORIGINAL DFLTS
BENCH GPIB
5.Select BENCH, and then press ENTER. The unit then displays:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
6.Press ENTER to restore bench defaults, and note the unit displays the following:
RESET COMPLETE
BENCH defaults are now restored
Press ENTER to continue
7.Press ENTER then EXIT as necessary to return to normal display.
Performance Verification 1-5
Performing the verification test procedures
T est summary
•DC voltage output accuracy
•DC voltage measurement accuracy
•DC current output accuracy
•DC current measurement accuracy
•Resistance measurement accuracy
If the Model 2420 is not within specifications and not under warranty , see the calibration pro-
cedures in Section 2 for information on calibrating the unit.
1-6Performance Verification
T est considerations
When performing the verification procedures:
•Be sure to restore factory front panel defaults as previously outlined.
•Make sure that the test equipment is properly warmed up and connected to the Model
2420 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are selected with
the TERMINALS key.
•Make sure the Model 2420 is set to the correct source range (see below).
•Be sure that the Model 2420 output is turned on before making measurements.
•Be sure the test equipment is set up for the proper function and range.
•Allow the Model 2420 output signal to settle before making a measurement.
•Do not connect test equipment to the Model 2420 through a scanner, multiplexer, or
other switching equipment.
WARNING
CAUTION
The maximum common-mode voltage (voltage between LO and chassis
ground) is 250V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
The maximum voltage between INPUT/OUTPUT HI and LO or 4-WIRE
SENSE HI and LO is 75V peak. The maximum voltage between INPUT/
OUTPUT HI and 4-WIRE SENSE HI or between INPUT/OUTPUT LO
and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result
in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output v al-
ue as outlined below.
1.Press either the SOURCE V or SOURCE I ke y to select the appropriate source function.
2.Press the EDIT key as required to select the source display field. Note that the cursor will
flash in the source field while its value is being edited.
3.With the cursor in the source display field flashing, set the source range to the lo west possible range for the value to be sourced using the up or do wn RANGE ke y. For example,
you should use the 20V source range to output a 20V source value. With a 20V source
value and the 20V range selected, the source field display will appear as follows:
Vsrc:+20.0000 V
4.With the source field cursor flashing, set the source output to the required value using
either:
• The SOURCE adjustment and left and right arrow keys.
• The numeric keys.
5.Note that the source output value will be updated immediately; you need not press
ENTER when setting the source value.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range is
coupled to the source range, and you cannot independently control the measure range. Thus, it
is not necessary for you to set the measure range when testing voltage or current measurement
accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 3.15A.
Conversely, when sourcing current, you can set the SourceMeter to limit voltage from 0.2mV to
63V. The SourceMeter output will not exceed the programmed compliance limit.
T ypes of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on which
value is lower , the output will clamp at either the displayed compliance setting (“real”) or at the
maximum measurement range reading (“range”).
Performance Verification 1-7
The “real” compliance condition can occur when the compliance setting is less than the highest possible reading of the measurement range. When in compliance, the source output clamps
at the displayed compliance value. For example, if the compliance voltage is set to 1V and the
measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible reading of the selected measurement range. When in compliance, the source output clamps at the
maximum measurement range reading (not the compliance value). For example, if the compliance voltage is set to 1V and the measurement range is 200mV, the output voltage will clamp
(limit) at 210mV.
1-8Performance Verification
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in Table 1-2.
Table 1-2
Maximum compliance values
Measurement
range
200mV
2V
20V
60V
10
µ
A
100
µ
A
1mA
10mA
100mA
1A
3A
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e., “mA”)
for the compliance display will flash.
Maximum
compliance value
210mV
2.1V
21V
63V
10.5
µ
A
105
µ
A
1.05mA
10.5mA
105mA
1.05A
3.15A
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows.
They assume that the measurement function is the same as the compliance function.
•Compliance Setting < Measurement Range = Real Compliance
•Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compliance
setting to the present measurement range. If the compliance setting is lower than the maximum
possible reading on the present measurement range, the compliance setting is the compliance
limit. If the compliance setting is higher than the measurement range, the maximum reading on
that measurement range is the compliance limit
T aking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance. For
purposes of the verification tests, the SourceMeter can be taken out of compliance by going into
the edit mode and increasing the compliance limit.
NOTE
Do not take the unit out of compliance by decreasing the sour ce value or changing the
range. Always use the recommended range and source settings when performing the
verification tests.
Output voltage accuracy
Follow the steps belo w to verify that Model 2420 output v oltage accurac y is within specified
limits. This test involves setting the output voltage to each full-range value and measuring the
voltages with a precision digital multimeter.
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks, as shown in Figure 1-1.
Figure 1-1
Connections for voltage
verification tests
Performance Verification 1-9
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
MEAS
EDIT
V
I
Ω
DISPLAY
1
LOCAL
67
DIGITS SPEED
230
REL
FILTER
89
STORE
TOGGLE
POWER
2420 3A SourceMeter
SOURCE
FCTN
I
V
4
EDIT
5
TRIG
SWEEP
LIMIT
+/-
EXIT ENTER
RECALL
CONFIG MENU
Model 2420
LO
250V
PEAK
RANGE
AUTO
TERMINALS
ON/OFF
FRONT/
RANGE
REAR
OUTPUT
Input HI
Input LO
Digital Multimeter
2.Select the multimeter DC volts measuring function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V
to access the CONFIGURE V-SOURCE menu. Select PROTECTION, and set the voltage source protection limit to NONE.
4.Press the Model 2420 SOURCE V key to source voltage, and make sure the source output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-3. For each test
point:
• Select the correct source range.
• Set the Model 2420 output voltage to the indicated value.
• Verify that the multimeter reading is within the limits given in the table.
6.Repeat the procedure for negative output voltages with the same magnitudes as those
listed in Table 1-3.
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
1-10Performance Verification
Table 1-3
Output voltage accuracy limits
Model 2420
source range
200mV
2V
20V
60V
Model 2420 output
voltage setting
200.000mV
2.00000V
20.0000V
60.0000V
V oltage measurement accuracy
Follow the steps below to verify that Model 2420 voltage measurement accuracy is within
specified limits. The test inv olves setting the source v oltage to full-range values, as measured by
a precision digital multimeter, and then v erifying that the Model 2420 voltage readings are within required limits.
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks , as shown in Figure 1-1.
2.Select the multimeter DC volts function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V
to access the CONFIGURE V-SOURCE menu. Select PROTECTION, and set the voltage source protection limit to NONE.
4.Set the Model 2420 to both source and measure voltage by pressing the SOURCE V and
MEAS V keys, and make sure the source output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-4. For each test
point:
• Select the correct source range.
• Set the Model 2420 output voltage to the indicated value as measured by the digital
multimeter.
• Verify that the Model 2420 voltage reading is within the limits given in the table.
Output voltage limits
(1 year, 18°C-28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
19.9936 to 20.0064V
59.9808 to 60.0192V
NOTE
It may not be possible to set the voltage sour ce to the specified value . Use the closest
possible setting, and modify reading limits accordingly.
6.Repeat the procedure for negative source voltages with the same magnitudes as those
listed in Table 1-4.
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Performance Verification 1-11
*Measure range coupled to source range when simultaneously sourcing and measuring voltage.
**As measured by precision digital multimeter. Use closest possible value, and modify reading limits
Output current accuracy
limits. The test involves setting the output current to each full-range value and measuring the
currents with a precision digital multimeter.
10µA to 1A range accuracy
Figure 1-2
Connections for 10
to 1A range current
verification tests
Model 2420 source
and measure range*
200mV
2V
20V
60V
Source voltage*
200.000mV
2.00000V
20.0000V
60.0000V
Model 2420 voltage reading
limits (1 year, 18°C-28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
19.9960 to 20.0040V
59.9880 to 60.0120V
accordingly if necessary.
Follow the steps belo w to verify that Model 2420 output current accuracy is within specified
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks, as shown in Figure 1-2.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
µ
A
MEAS
EDIT
V
I
Ω
DISPLAY
1
LOCAL
67
DIGITS SPEED
230
REL
FILTER
89
STORE
TOGGLE
POWER
2420 3A SourceMeter
SOURCE
FCTN
I
V
4
5
EDIT
TRIG
SWEEP
LIMIT
+/-
EXIT ENTER
RECALL
CONFIG MENU
PEAK
LO
250V
PEAK
RANGE
AUTO
TERMINALS
ON/OFF
FRONT/
RANGE
REAR
OUTPUT
Model 2420
Input LO
Amps
Digital Multimeter
2.Select the multimeter DC current measuring function.
3.Press the Model 2420 SOURCE I key to source current, and make sure the source output
is turned on.
1-12Performance Verification
4.Verify output current accuracy for the 10µA-1A range currents listed in Table 1-5. For
each test point:
• Select the correct source range.
• Set the Model 2420 output current to the correct value.
• Verify that the multimeter reading is within the limits given in the table.
5.Repeat the procedure for negative output currents with the same magnitudes as those
listed in Table 1-5.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
3A range accuracy
1.With the power of f, connect the digital multimeter and the 1Ω resistor to the Model 2420
INPUT/OUTPUT jacks, as shown in Figure 1-3.
Figure 1-3
Connections for
3A range current
verification tests
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
V
4
TRIG
+/-
CONFIG MENU
2420 3A SourceMeter
SOURCE
I
5
EDIT
SWEEP
EXIT ENTER
MEAS
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
FCTN
I
Ω
1
230
REL
LIMIT
FILTER
89
RECALL
STORE
LO
250V
RANGE
AUTO
RANGE
PEAK
TERMINALS
ON/OFF
FRONT/
REAR
OUTPUT
1Ω Resistor
Model 2420
Input HI
Input LO
Digital Multimeter
2.Select the multimeter DC volts measuring function.
3.Press the Model 2420 SOURCE I key to source current, and make sure the source output
is turned on.
4.Verify output current accuracy for the 3A range. Be sure to:
• Select the 3A source range.
• Set the Model 2420 output current to the correct 3A output value.
• Verify that the multimeter reading is within the 3A range limits given in Table 1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal value,
the DMM voltage reading is the same as the sourced current.)
5.Repeat the procedure for a negative 3A current output value.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Performance Verification 1-13
Model 2420
source range
10µA
100µA
1mA
10mA
100mA
1A
3A*
* See separate procedure for 3A range. DMM voltage reading is same as sourced current.
1
Specifications valid for continuous output currents below 105mA. For operating abov e 105mA on the 1A range
for >1 minute, derate accuracy 10%/100mA above 105mA. For operating abov e 105mA on the 3A range for >1
minute, derate accuracy 10%/300mA above 105mA.
Model 2420
output current setting
10.0000µA
100.000µA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
Current measurement accuracy
Follow the steps below to verify that Model 2420 current measurement accuracy is within
specified limits. The procedure inv olves applying accurate currents from the Model 2420 current
source and then verifying that Model 2420 current measurements are within required limits.
10µA to 1A range accuracy
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks, as shown in Figure 1-2.
2.Select the multimeter DC current function.
3.Set the Model 2420 to both source and measure current by pressing the SOURCE I and
MEAS I keys, and make sure the source output is turned on.
4.V erify measure current accurac y for the 10µA-1A range currents listed in Table 1-6. For
each measurement:
• Select the correct source range.
• Set the Model 2420 source output to the correct value as measured by the digital multimeter .
• Verify that the Model 2420 current reading is within the limits given in the table.
Output current limits
(1 year, 18°C-28°C)
9.9947 to 10.0053µA
99.949 to 100.051µA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
2.99553 to 3.00447A
1
1
NOTEIt may not be possible to set the current source to the specified value. Use the closest
possible setting, and modify reading limits accordingly.
5.Repeat the procedure for negative calibrator currents with the same magnitudes as those
listed in Table 1-6.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
NOTETest currents above 105mA cannot be maintained longer than 1 minute without
affecting accuracy. See derating information in Note 1 under Table 1-5.
1-14Performance Verification
3A range accuracy
NOTEThe 1Ω resistor should be characterized to within ±300ppm before verifying the 3A
current measur ement range. Use the 4-wir e ohms function of the DMM to measur e the
resistance value, and then use that measur ed value to calculate the curr ent during the
measurement procedure.
1.With the power off, connect the 1Ω resistor and digital multimeter to the Model 2420
INPUT/OUTPUT jacks, as shown in Figure 1-3.
2.Select the multimeter DC volts function.
3.Set the Model 2420 to both source and measure current by pressing the SOURCE I and
MEAS I keys, and make sure the source output is turned on.
4.Verify measurement current accuracy for the 3A range as follows:
• Select the 3A source range.
• Set the Model 2420 source output to the correct 3A value as measured by the digital
multimeter.
• Note the DMM voltage reading, and then calculate the current from the voltage reading
and characterized 1Ω resistance value as I = V/R, where V is the DMM voltage reading
and R is the characterized resistance value.
• V erify that the Model 2420 current reading is within the 3A limits gi ven in the Table 1-6.
NOTEIt may not be possible to set the current sour ce to the specified 3A value . Use the clos-
est possible setting, and modify reading limits accordingly.
5.Repeat the procedure for a negative 3A current.
6.Repeat the procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the
rear panel jacks with the front panel TERMINALS key.
NOTETest currents above 105mA cannot be maintained longer than 1 minute without
affecting accuracy. See derating information in Note 1 under Table 1-5.
Table 1-6
Current measurement accuracy limits
Model 2420 source
and measure range*
10µA
100µA
1mA
10mA
100mA
1A
3A
*Measure range coupled to source range when simultaneously sourcing and measuring current.
**As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
***Current calculated as follows: I = V/R, where V is the DMM voltage reading, and R is the character-
ized value of the 1Ω resistor.
Source current**
10.00000µA
100.000µA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
Model 2420 current reading limits (1
year, 18°C-28°C)
9.9966 to 10.0034µA
99.969 to 100.031µA
0.99967 to 1.00033mA
9.9959 to 10.0041mA
99.939 to 100.061mA
0.99883 to 1.00117A
2.99673 to 3.00327A***
Resistance measurement accuracy
Use the following steps to verify that Model 2420 resistance measurement accuracy is
within specified limits. This procedure involves applying accurate resistances from a resistance calibrator and then verifying that Model 2420 resistance measurements are within required limits.
CAUTIONBefore testing the 2Ω and 20Ω ranges make sure your resistance calibrator
can safely handle the default test currents for those ranges (see Model 2420
and calibrator specifications). If not, use the CONFIG OHMS menu to select
the MANUAL source mode, then set the source current to an appropriate safe
value. When using the manual source mode, total resistance reading uncertainty includes both Source I and Measure V uncertainty (see specifications),
and calculated reading limits should take the additional uncertainty into account.
If using the Fluke 5450A resistance calibrator, you cannot use the Auto Ohms
mode of the Model 2420 to verify the 2Ω range. The 1A test current for the 2Ω
range of the Model 2420 will damage the calibrator. On the Model 2420, use
the CONFIG OHMS menu to select the MANUAL source mode, and then set
the source (test) current to 100mA.
Performance Verification 1-15
1.With the power off, connect the resistance calibrator to the Model 2420 INPUT/OUT-
Figure 1-4
Connections for resistance
accuracy verification
PUT and 4-WIRE SENSE jacks, as shown in Figure 1-4. Be sure to use the 4-wire connections as shown.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
MEAS
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
FCTN
I
Ω
V
1
230
4
TRIG
REL
LIMIT
FILTER
+/-
89
RECALL
STORE
CONFIG MENU
Model 2420
Resistance Calibrator
2420 3A SourceMeter
SOURCE
I
EDIT
5
SWEEP
EXIT ENTER
LO
RANGE
AUTO
ON/OFF
RANGE
OUTPUT
Output HI
Output LO
250V
PEAK
TERMINALS
FRONT/
REAR
Sense HI
Sense LO
1-16Performance Verification
2.Select the resistance calibrator external sense mode.
3.Configure the Model 2420 ohms function for the 4-wire sense mode as follows:
• Press CONFIG then MEAS Ω. The instrument will display the following:
CONFIG OHMS
SOURCE SENSE-MODE GUARD
• Select SENSE-MODE, and then press ENTER. The following will be displayed:
SENSE-MODE
2-WIRE 4-WIRE
• Select 4-WIRE, and then press ENTER.
• Press EXIT to return to normal display.
4.Press MEAS Ω to select the ohms measurement function, and make sure the source out-
put is turned on.
5.Verify ohms measurement accuracy for each of the resistance v alues listed in Table 1-7.
For each measurement:
• Set the resistance calibrator output to the nominal resistance or closest av ailable value.
NOTEIt may not be possible to set the resistance calibrator to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
• Select the appropriate ohms measurement range with the RANGE keys.
• Verify that the Model 2420 resistance reading is within the limits given in the table.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE SENSE
jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-7
Ohms measurement accuracy limits
Model 2420 rangeCalibrator resistance*
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
*Nominal resistance value.
**Reading limits based on Model 2420 normal accuracy specifications and nominal resistance values. If actual
resistance values differ from nominal values sho wn, recalculate reading limits using actual calibrator resistance
values and Model 2420 normal accuracy specifications. See Verification limits earlier in this section for details.
1.9Ω
19Ω
190Ω
1.9kΩ
19kΩ
190kΩ
1.9MΩ
19MΩ
Model 2420 resistance reading limits**
(1 year, 18°C-28°C)
1.89649 to 1.90351Ω
18.9784 to 19.0216Ω
189.824 to 190.176Ω
1.89845 to 1.90155kΩ
18.9850 to 19.0150kΩ
189.847 to 190.153kΩ
1.89861 to 1.90139MΩ
18.9517 to 19.0483MΩ
2
Calibration
Loading...
+ 105 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.