Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defecti ve.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility . Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechar geable
batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT , INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF
THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS
BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION,
LOSSES SUSTAINED AS THE RESULT OF INJURY T O ANY PERSON, OR DAMAGE T O PROPER TY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Le vel letter increases alphabetically as the manual under goes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially . When a new Re vision is created, all Addenda associated with the previous Re vision
of the manual are incorporated into the new Revision of the manual. Each ne w Revision includes
a revised copy of this print history page.
Revision A (Document Number 2420-902-01)................................................................April 1997
Addendum A (Document Number 2420-902-02).............................................................April 1997
Revision B (Document Number 2420-902-01)..................................................................July 1998
Revision C (Document Number 2420-902-01)............................................................ January 1999
Revision D (Document Number 2420-902-01).................................................................June 2000
Revision E (Document Number 2420-902-01) ............................................................October 2001
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
S
afety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body
ment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel
voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state
if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel
trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Installation Category I and Installation Category II,
as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data
I/O signals are Installation Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Installation Cate gory II connections require protection for high transient over -voltages often associated with
local A C mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures.
The American National Standards Institute (ANSI) states that a shock hazard exists when v oltage le vels greater than 30V RMS,
42.4V peak, or 60VDC are present.
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock.
If the circuit is capable of operating at or above 1000 volts,
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equip-
use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
perform routine procedures on the product to keep it operating properly, for example, setting the line
are trained to work on live circuits, and perform safe installations and repairs of products. Only properly
A good safety practice is to expect that hazardous voltage is present in any unknown
no conductive part of the circuit may be exposed.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry , insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of
normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
information very carefully before performing the indicated procedure.
CAUTION
The
ranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
T o maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer ,
test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from
other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only
through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability
of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for
proper cleaning/servicing.
heading in a manual explains dangers that might result in personal injury or death. Al ways read the associated
heading in a manual explains hazards that could damage the instrument. Such damage may inv alidate the war -
Use the procedures in this section to verify that Model 2420 accuracy is within the limits stated
in the instrument’s one-year accurac y specifications. You can perform these verification procedures:
•When you first receive the instrument to make sure that it was not damaged during shipment.
•To verify that the unit meets factory specifications.
•To determine if calibration is required.
•Following calibration to make sure it was performed properly.
WARNING
NOTE
The information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, which
could cause personal injury or death if contacted. Use standard safety precautions when working with hazardous voltages.
If the instrument is still under warranty and its performance is outside specified limits, contact your Keithle y r epr esentative or the factory to determine the correct course of action.
V erification test requirements
Be sure that you perform the verification tests:
•Under the proper environmental conditions.
•After the specified warm-up period.
•Using the correct line voltage.
•Using the proper test equipment.
•Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•An ambient temperature of 18-28°C (65-82°F).
•A relative humidity of less than 70% unless otherwise noted.
W arm-up period
Allow the Model 2420 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. T ypically , allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specifi ed temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manuf acturer .
Line power
The Model 2420 requires a line voltage of 90 to 250V and a line frequency of 50 to 60Hz.
Verification tests must be performed within this range.
Recommended test equipment
T able 1-1 summarizes recommended verification equipment. You can use alternate equipment
as long as that equipment has specifications at least as good as those listed in Table 1-1. Keep in
mind, howev er, that test equipment uncertainty will add to the uncertainty of each measurement.
Generally, test equipment uncertainty should be at least four times better than corresponding Model
2420 specifications. Table 1-1 lists the specifications of the recommended test equipment, including maximum allowable uncertainty for alternate test equipment, which is shown in parentheses.
Table 1-1
Recommended verification equipment
DescriptionManufacturer/ModelSpecifications
Digital MultimeterHewlett Packard
HP3458A
DC Voltage*1V:
Performance Verification 1-3
±5.6ppm
10V:
100V:
±4.3ppm
±6.3ppm
µ
DC Current*10
Resistance CalibratorFluke 5450AResistance**1.9
Precision Resistor*** Isotec RUG-Z-1R00-0.11
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of
alternate test equipment shown in parentheses.
*** Required for verification of 3A current range. Characterize resistor to ±300ppm or better using recommended DMM
before verifying 3A current measurement range.
A:
µ
100
1mA:
10mA:
100mA:
1A:
Ω
Ω
19
Ω
190
1.9k
Ω
19k
190k
1.9M
19M
Ω
, ±0.1%, 100W
±25ppm
A:
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
:
±65ppm (±460ppm)
:
±23ppm (±280ppm)
:
±10.5ppm (±230ppm)
Ω
:
±8ppm (±200ppm)
:
±7.5ppm (±195ppm)
Ω
:
±8.5ppm (±200ppm)
Ω
:
±11.5ppm (±180ppm)
Ω
:
±30ppm (±635ppm)
1-4Performance Verification
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before verifying
the 3A current measurement range. (You need not characterize the resistor if you are checking
only the 3A current source range.) Use the 4-wire ohms function of the DMM recommended in
T able 1-1 to measure the resistance value, and then use that measured value to calculate the cur rent during the 3A current measurement range test procedure.
V erification limits
The verification limits stated in this section have been calculated using only the Model 2420
one-year accuracy specifications, and they do not include test equipment uncertainty. If a particular measurement falls outside the allowable range, recalculate new limits based on Model
2420 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 20V DC
output range using a 20V output value. Using the Model 2420 20V range one-year accuracy
specification of ±(0.02% of output + 2.4mV offset), the calculated output limits are:
Output limits = 20V ± [(20V
Output limits = 20V ± (0.004 + 0.0024)
Output limits = 20V ± 0.0064V
Output limits = 19.9936V to 20.0064V
Resistance limits calculation
When verifying the resistance measurement accuracy, it will probably be necessary to recalculate resistance limits based on the actual calibrator resistance values. You can calculate resistance reading limits in the same manner described above, b ut be sure to use the actual calibrator
resistance values and the Model 2420 normal accuracy specifications for your calculations.
As an example, assume you are testing the 20k
inal 19k
accuracy specifications of ±(0.063% of reading + 3
Ω
calibrator resistor is 19.01kΩ. Using the Model 2420 20kΩ range one-year normal
Reading limits = 19.01k
Reading limits = 19.01k
Reading limits = 18.9950kΩ to 19.0250k
Ω
Ω
×
0.02%) + 2.4mV]
Ω
range, and that the actual value of the nom-
Ω
), the recalculated reading limits are:
± [(19.01kΩ × 0.063%) + 3Ω]
±15
Ω
Ω
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory front panel
(bench) defaults as follows:
1.Press the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL
2.Select SAVESETUP, and then press ENTER. The unit then displays:
SAVESETUP MENU
GLOBAL SOURCE-MEMORY
3.Select GLOBAL, and then press ENTER. The unit then displays:
GLOBAL SETUP MENU
SAVE RESTORE POWERON RESET
4.Select RESET, and then press ENTER. The unit displays:
RESET ORIGINAL DFLTS
BENCH GPIB
5.Select BENCH, and then press ENTER. The unit then displays:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
6.Press ENTER to restore bench defaults, and note the unit displays the following:
RESET COMPLETE
BENCH defaults are now restored
Press ENTER to continue
7.Press ENTER then EXIT as necessary to return to normal display.
Performance Verification 1-5
Performing the verification test procedures
T est summary
•DC voltage output accuracy
•DC voltage measurement accuracy
•DC current output accuracy
•DC current measurement accuracy
•Resistance measurement accuracy
If the Model 2420 is not within specifications and not under warranty , see the calibration pro-
cedures in Section 2 for information on calibrating the unit.
1-6Performance Verification
T est considerations
When performing the verification procedures:
•Be sure to restore factory front panel defaults as previously outlined.
•Make sure that the test equipment is properly warmed up and connected to the Model
2420 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are selected with
the TERMINALS key.
•Make sure the Model 2420 is set to the correct source range (see below).
•Be sure that the Model 2420 output is turned on before making measurements.
•Be sure the test equipment is set up for the proper function and range.
•Allow the Model 2420 output signal to settle before making a measurement.
•Do not connect test equipment to the Model 2420 through a scanner, multiplexer, or
other switching equipment.
WARNING
CAUTION
The maximum common-mode voltage (voltage between LO and chassis
ground) is 250V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
The maximum voltage between INPUT/OUTPUT HI and LO or 4-WIRE
SENSE HI and LO is 75V peak. The maximum voltage between INPUT/
OUTPUT HI and 4-WIRE SENSE HI or between INPUT/OUTPUT LO
and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result
in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output v al-
ue as outlined below.
1.Press either the SOURCE V or SOURCE I ke y to select the appropriate source function.
2.Press the EDIT key as required to select the source display field. Note that the cursor will
flash in the source field while its value is being edited.
3.With the cursor in the source display field flashing, set the source range to the lo west possible range for the value to be sourced using the up or do wn RANGE ke y. For example,
you should use the 20V source range to output a 20V source value. With a 20V source
value and the 20V range selected, the source field display will appear as follows:
Vsrc:+20.0000 V
4.With the source field cursor flashing, set the source output to the required value using
either:
• The SOURCE adjustment and left and right arrow keys.
• The numeric keys.
5.Note that the source output value will be updated immediately; you need not press
ENTER when setting the source value.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range is
coupled to the source range, and you cannot independently control the measure range. Thus, it
is not necessary for you to set the measure range when testing voltage or current measurement
accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 3.15A.
Conversely, when sourcing current, you can set the SourceMeter to limit voltage from 0.2mV to
63V. The SourceMeter output will not exceed the programmed compliance limit.
T ypes of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on which
value is lower , the output will clamp at either the displayed compliance setting (“real”) or at the
maximum measurement range reading (“range”).
Performance Verification 1-7
The “real” compliance condition can occur when the compliance setting is less than the highest possible reading of the measurement range. When in compliance, the source output clamps
at the displayed compliance value. For example, if the compliance voltage is set to 1V and the
measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible reading of the selected measurement range. When in compliance, the source output clamps at the
maximum measurement range reading (not the compliance value). For example, if the compliance voltage is set to 1V and the measurement range is 200mV, the output voltage will clamp
(limit) at 210mV.
1-8Performance Verification
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in Table 1-2.
Table 1-2
Maximum compliance values
Measurement
range
200mV
2V
20V
60V
10
µ
A
100
µ
A
1mA
10mA
100mA
1A
3A
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e., “mA”)
for the compliance display will flash.
Maximum
compliance value
210mV
2.1V
21V
63V
10.5
µ
A
105
µ
A
1.05mA
10.5mA
105mA
1.05A
3.15A
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows.
They assume that the measurement function is the same as the compliance function.
•Compliance Setting < Measurement Range = Real Compliance
•Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compliance
setting to the present measurement range. If the compliance setting is lower than the maximum
possible reading on the present measurement range, the compliance setting is the compliance
limit. If the compliance setting is higher than the measurement range, the maximum reading on
that measurement range is the compliance limit
T aking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance. For
purposes of the verification tests, the SourceMeter can be taken out of compliance by going into
the edit mode and increasing the compliance limit.
NOTE
Do not take the unit out of compliance by decreasing the sour ce value or changing the
range. Always use the recommended range and source settings when performing the
verification tests.
Output voltage accuracy
Follow the steps belo w to verify that Model 2420 output v oltage accurac y is within specified
limits. This test involves setting the output voltage to each full-range value and measuring the
voltages with a precision digital multimeter.
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks, as shown in Figure 1-1.
Figure 1-1
Connections for voltage
verification tests
Performance Verification 1-9
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
MEAS
EDIT
V
I
Ω
DISPLAY
1
LOCAL
67
DIGITS SPEED
230
REL
FILTER
89
STORE
TOGGLE
POWER
2420 3A SourceMeter
SOURCE
FCTN
I
V
4
EDIT
5
TRIG
SWEEP
LIMIT
+/-
EXIT ENTER
RECALL
CONFIG MENU
Model 2420
LO
250V
PEAK
RANGE
AUTO
TERMINALS
ON/OFF
FRONT/
RANGE
REAR
OUTPUT
Input HI
Input LO
Digital Multimeter
2.Select the multimeter DC volts measuring function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V
to access the CONFIGURE V-SOURCE menu. Select PROTECTION, and set the voltage source protection limit to NONE.
4.Press the Model 2420 SOURCE V key to source voltage, and make sure the source output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-3. For each test
point:
• Select the correct source range.
• Set the Model 2420 output voltage to the indicated value.
• Verify that the multimeter reading is within the limits given in the table.
6.Repeat the procedure for negative output voltages with the same magnitudes as those
listed in Table 1-3.
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
1-10Performance Verification
Table 1-3
Output voltage accuracy limits
Model 2420
source range
200mV
2V
20V
60V
Model 2420 output
voltage setting
200.000mV
2.00000V
20.0000V
60.0000V
V oltage measurement accuracy
Follow the steps below to verify that Model 2420 voltage measurement accuracy is within
specified limits. The test inv olves setting the source v oltage to full-range values, as measured by
a precision digital multimeter, and then v erifying that the Model 2420 voltage readings are within required limits.
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks , as shown in Figure 1-1.
2.Select the multimeter DC volts function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V
to access the CONFIGURE V-SOURCE menu. Select PROTECTION, and set the voltage source protection limit to NONE.
4.Set the Model 2420 to both source and measure voltage by pressing the SOURCE V and
MEAS V keys, and make sure the source output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-4. For each test
point:
• Select the correct source range.
• Set the Model 2420 output voltage to the indicated value as measured by the digital
multimeter.
• Verify that the Model 2420 voltage reading is within the limits given in the table.
Output voltage limits
(1 year, 18°C-28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
19.9936 to 20.0064V
59.9808 to 60.0192V
NOTE
It may not be possible to set the voltage sour ce to the specified value . Use the closest
possible setting, and modify reading limits accordingly.
6.Repeat the procedure for negative source voltages with the same magnitudes as those
listed in Table 1-4.
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Performance Verification 1-11
*Measure range coupled to source range when simultaneously sourcing and measuring voltage.
**As measured by precision digital multimeter. Use closest possible value, and modify reading limits
Output current accuracy
limits. The test involves setting the output current to each full-range value and measuring the
currents with a precision digital multimeter.
10µA to 1A range accuracy
Figure 1-2
Connections for 10
to 1A range current
verification tests
Model 2420 source
and measure range*
200mV
2V
20V
60V
Source voltage*
200.000mV
2.00000V
20.0000V
60.0000V
Model 2420 voltage reading
limits (1 year, 18°C-28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
19.9960 to 20.0040V
59.9880 to 60.0120V
accordingly if necessary.
Follow the steps belo w to verify that Model 2420 output current accuracy is within specified
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks, as shown in Figure 1-2.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
µ
A
MEAS
EDIT
V
I
Ω
DISPLAY
1
LOCAL
67
DIGITS SPEED
230
REL
FILTER
89
STORE
TOGGLE
POWER
2420 3A SourceMeter
SOURCE
FCTN
I
V
4
5
EDIT
TRIG
SWEEP
LIMIT
+/-
EXIT ENTER
RECALL
CONFIG MENU
PEAK
LO
250V
PEAK
RANGE
AUTO
TERMINALS
ON/OFF
FRONT/
RANGE
REAR
OUTPUT
Model 2420
Input LO
Amps
Digital Multimeter
2.Select the multimeter DC current measuring function.
3.Press the Model 2420 SOURCE I key to source current, and make sure the source output
is turned on.
1-12Performance Verification
4.Verify output current accuracy for the 10µA-1A range currents listed in Table 1-5. For
each test point:
• Select the correct source range.
• Set the Model 2420 output current to the correct value.
• Verify that the multimeter reading is within the limits given in the table.
5.Repeat the procedure for negative output currents with the same magnitudes as those
listed in Table 1-5.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
3A range accuracy
1.With the power of f, connect the digital multimeter and the 1Ω resistor to the Model 2420
INPUT/OUTPUT jacks, as shown in Figure 1-3.
Figure 1-3
Connections for
3A range current
verification tests
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
V
4
TRIG
+/-
CONFIG MENU
2420 3A SourceMeter
SOURCE
I
5
EDIT
SWEEP
EXIT ENTER
MEAS
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
FCTN
I
Ω
1
230
REL
LIMIT
FILTER
89
RECALL
STORE
LO
250V
RANGE
AUTO
RANGE
PEAK
TERMINALS
ON/OFF
FRONT/
REAR
OUTPUT
1Ω Resistor
Model 2420
Input HI
Input LO
Digital Multimeter
2.Select the multimeter DC volts measuring function.
3.Press the Model 2420 SOURCE I key to source current, and make sure the source output
is turned on.
4.Verify output current accuracy for the 3A range. Be sure to:
• Select the 3A source range.
• Set the Model 2420 output current to the correct 3A output value.
• Verify that the multimeter reading is within the 3A range limits given in Table 1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal value,
the DMM voltage reading is the same as the sourced current.)
5.Repeat the procedure for a negative 3A current output value.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Performance Verification 1-13
Model 2420
source range
10µA
100µA
1mA
10mA
100mA
1A
3A*
* See separate procedure for 3A range. DMM voltage reading is same as sourced current.
1
Specifications valid for continuous output currents below 105mA. For operating abov e 105mA on the 1A range
for >1 minute, derate accuracy 10%/100mA above 105mA. For operating abov e 105mA on the 3A range for >1
minute, derate accuracy 10%/300mA above 105mA.
Model 2420
output current setting
10.0000µA
100.000µA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
Current measurement accuracy
Follow the steps below to verify that Model 2420 current measurement accuracy is within
specified limits. The procedure inv olves applying accurate currents from the Model 2420 current
source and then verifying that Model 2420 current measurements are within required limits.
10µA to 1A range accuracy
1.With the power of f, connect the digital multimeter to the Model 2420 INPUT/OUTPUT
jacks, as shown in Figure 1-2.
2.Select the multimeter DC current function.
3.Set the Model 2420 to both source and measure current by pressing the SOURCE I and
MEAS I keys, and make sure the source output is turned on.
4.V erify measure current accurac y for the 10µA-1A range currents listed in Table 1-6. For
each measurement:
• Select the correct source range.
• Set the Model 2420 source output to the correct value as measured by the digital multimeter .
• Verify that the Model 2420 current reading is within the limits given in the table.
Output current limits
(1 year, 18°C-28°C)
9.9947 to 10.0053µA
99.949 to 100.051µA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
2.99553 to 3.00447A
1
1
NOTEIt may not be possible to set the current source to the specified value. Use the closest
possible setting, and modify reading limits accordingly.
5.Repeat the procedure for negative calibrator currents with the same magnitudes as those
listed in Table 1-6.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
NOTETest currents above 105mA cannot be maintained longer than 1 minute without
affecting accuracy. See derating information in Note 1 under Table 1-5.
1-14Performance Verification
3A range accuracy
NOTEThe 1Ω resistor should be characterized to within ±300ppm before verifying the 3A
current measur ement range. Use the 4-wir e ohms function of the DMM to measur e the
resistance value, and then use that measur ed value to calculate the curr ent during the
measurement procedure.
1.With the power off, connect the 1Ω resistor and digital multimeter to the Model 2420
INPUT/OUTPUT jacks, as shown in Figure 1-3.
2.Select the multimeter DC volts function.
3.Set the Model 2420 to both source and measure current by pressing the SOURCE I and
MEAS I keys, and make sure the source output is turned on.
4.Verify measurement current accuracy for the 3A range as follows:
• Select the 3A source range.
• Set the Model 2420 source output to the correct 3A value as measured by the digital
multimeter.
• Note the DMM voltage reading, and then calculate the current from the voltage reading
and characterized 1Ω resistance value as I = V/R, where V is the DMM voltage reading
and R is the characterized resistance value.
• V erify that the Model 2420 current reading is within the 3A limits gi ven in the Table 1-6.
NOTEIt may not be possible to set the current sour ce to the specified 3A value . Use the clos-
est possible setting, and modify reading limits accordingly.
5.Repeat the procedure for a negative 3A current.
6.Repeat the procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the
rear panel jacks with the front panel TERMINALS key.
NOTETest currents above 105mA cannot be maintained longer than 1 minute without
affecting accuracy. See derating information in Note 1 under Table 1-5.
Table 1-6
Current measurement accuracy limits
Model 2420 source
and measure range*
10µA
100µA
1mA
10mA
100mA
1A
3A
*Measure range coupled to source range when simultaneously sourcing and measuring current.
**As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
***Current calculated as follows: I = V/R, where V is the DMM voltage reading, and R is the character-
ized value of the 1Ω resistor.
Source current**
10.00000µA
100.000µA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
Model 2420 current reading limits (1
year, 18°C-28°C)
9.9966 to 10.0034µA
99.969 to 100.031µA
0.99967 to 1.00033mA
9.9959 to 10.0041mA
99.939 to 100.061mA
0.99883 to 1.00117A
2.99673 to 3.00327A***
Resistance measurement accuracy
Use the following steps to verify that Model 2420 resistance measurement accuracy is
within specified limits. This procedure involves applying accurate resistances from a resistance calibrator and then verifying that Model 2420 resistance measurements are within required limits.
CAUTIONBefore testing the 2Ω and 20Ω ranges make sure your resistance calibrator
can safely handle the default test currents for those ranges (see Model 2420
and calibrator specifications). If not, use the CONFIG OHMS menu to select
the MANUAL source mode, then set the source current to an appropriate safe
value. When using the manual source mode, total resistance reading uncertainty includes both Source I and Measure V uncertainty (see specifications),
and calculated reading limits should take the additional uncertainty into account.
If using the Fluke 5450A resistance calibrator, you cannot use the Auto Ohms
mode of the Model 2420 to verify the 2Ω range. The 1A test current for the 2Ω
range of the Model 2420 will damage the calibrator. On the Model 2420, use
the CONFIG OHMS menu to select the MANUAL source mode, and then set
the source (test) current to 100mA.
Performance Verification 1-15
1.With the power off, connect the resistance calibrator to the Model 2420 INPUT/OUT-
Figure 1-4
Connections for resistance
accuracy verification
PUT and 4-WIRE SENSE jacks, as shown in Figure 1-4. Be sure to use the 4-wire connections as shown.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
MEAS
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
FCTN
I
Ω
V
1
230
4
TRIG
REL
LIMIT
FILTER
+/-
89
RECALL
STORE
CONFIG MENU
Model 2420
Resistance Calibrator
2420 3A SourceMeter
SOURCE
I
EDIT
5
SWEEP
EXIT ENTER
LO
RANGE
AUTO
ON/OFF
RANGE
OUTPUT
Output HI
Output LO
250V
PEAK
TERMINALS
FRONT/
REAR
Sense HI
Sense LO
1-16Performance Verification
2.Select the resistance calibrator external sense mode.
3.Configure the Model 2420 ohms function for the 4-wire sense mode as follows:
• Press CONFIG then MEAS Ω. The instrument will display the following:
CONFIG OHMS
SOURCE SENSE-MODE GUARD
• Select SENSE-MODE, and then press ENTER. The following will be displayed:
SENSE-MODE
2-WIRE 4-WIRE
• Select 4-WIRE, and then press ENTER.
• Press EXIT to return to normal display.
4.Press MEAS Ω to select the ohms measurement function, and make sure the source out-
put is turned on.
5.Verify ohms measurement accuracy for each of the resistance v alues listed in Table 1-7.
For each measurement:
• Set the resistance calibrator output to the nominal resistance or closest av ailable value.
NOTEIt may not be possible to set the resistance calibrator to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
• Select the appropriate ohms measurement range with the RANGE keys.
• Verify that the Model 2420 resistance reading is within the limits given in the table.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE SENSE
jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-7
Ohms measurement accuracy limits
Model 2420 rangeCalibrator resistance*
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
*Nominal resistance value.
**Reading limits based on Model 2420 normal accuracy specifications and nominal resistance values. If actual
resistance values differ from nominal values sho wn, recalculate reading limits using actual calibrator resistance
values and Model 2420 normal accuracy specifications. See Verification limits earlier in this section for details.
1.9Ω
19Ω
190Ω
1.9kΩ
19kΩ
190kΩ
1.9MΩ
19MΩ
Model 2420 resistance reading limits**
(1 year, 18°C-28°C)
1.89649 to 1.90351Ω
18.9784 to 19.0216Ω
189.824 to 190.176Ω
1.89845 to 1.90155kΩ
18.9850 to 19.0150kΩ
189.847 to 190.153kΩ
1.89861 to 1.90139MΩ
18.9517 to 19.0483MΩ
2
Calibration
2-2Calibration
Introduction
Use the procedures in this section to calibrate the Model 2420. These procedures require
accurate test equipment to measure precise DC voltages and currents. Calibration can be performed
either from the front panel or by sending SCPI calibration commands over the
IEEE-488 bus or RS-232 port with the aid of a computer.
WARNING
This information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages.
Environmental conditions
T emperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18-28°C (65-82°F) with relative humidity of less than 70% unless otherwise noted.
W arm-up period
Allow the Model 2420 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. Typically,
allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature
range.
Also, allow the test equipment to warm up for the minimum time specified by the manuf acturer .
Line power
The Model 2420 requires a line voltage of 90 to 250V at line frequency of 50 to 60Hz. The
instrument must be calibrated while operating from a line voltage within this range.
Calibration considerations
When performing the calibration procedures:
•Make sure that the test equipment is properly warmed up and connected to the Model
2420 front panel INPUT/OUTPUT jacks. Also be certain that the front panel jacks are
selected with the TERMINALS switch.
•Always allow the source signal to settle before calibrating each point.
•Do not connect test equipment to the Model 2420 through a scanner or other switching
equipment.
•If an error occurs during calibration, the Model 2420 will generate an appropriate error
message. See Appendix B for more information.
Calibration 2-3
WARNING
CAUTION
The maximum common-mode voltage (voltage between LO and chassis
ground) is 250V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
The maximum voltage between INPUT/OUTPUT HI and LO or 4-WIRE
SENSE HI and LO is 75V peak. The maximum voltage between INPUT/
OUTPUT HI and 4-WIRE SENSE HI or between INPUT/OUTPUT LO
and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result
in instrument damage.
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its specifications.
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use alternate equipment as long that equipment has specifications at least as good as those listed in the
table. For optimum calibration accuracy, test equipment specifications should be at least four
times better than corresponding Model 2420 specifications.
2-4Calibration
Table 2-1
Recommended calibration equipment
DescriptionManufacturer/ModelSpecifications
Digital MultimeterHewlett Packard
HP3458A
Precision Resistor**Isotec RUG-Z-1R00-0.11
*90-day, full-range accuracy specifications of ranges required for various measurement points.
** Necessary for calibration of 3A current range. Resistor must be characterized to ±300ppm or better using recom-
mended DMM before calibrating 3A range.
DC Voltage*
DC Current*
1V:
10V:
100V
µ
A:
10
µ
A:
100
1mA:
10mA:
100mA:
1A:
Ω
, ±0.1%, 100W
±5.6ppm
±4.3ppm
±6.3ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
1Ω resistor characterization
The 1Ω resistor must be characterized to ±300ppm or better before calibrating the 3A current
range. Use the 4-wire ohms function of the DMM recommended in Table 1-1 to measure the resistance value, and then use that measured value to calculate the current during the 3A current
range calibration procedure.
Unlocking calibration
Before performing calibration, you must first unlock calibration by entering or sending the
calibration password as explained in the following paragraphs.
Unlocking calibration from the front panel
1.Press the MENU key , and then choose CAL, and press ENTER. The instrument will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
2.Select UNLOCK, and then press ENTER. The instrument will display the following:
PASSWORD:
Use ,,,, ENTER or EXIT.
3.Use the up and down RANGE keys to select the letter or number, and use the left and
right arrow keys to choose the position. (Press do wn RANGE for letters; up RANGE for
numbers.) Enter the present password on the display. (Front panel default: 002420.)
Calibration 2-5
4.Once the correct password is displayed, press the ENTER key. If the password was correctly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5.Press EXIT to return to normal display. Calibration will be unlocked and assume the
states summarized in T able 2-2. Attempts to change any of the settings listed below with
calibration unlocked will result in an error +510, "Not permitted with cal un-locked."
NOTE
With calibr ation unlock ed, the sense function and rang e trac k the sour ce function and
range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC setting will be
'VOLT:DC'. When :SOUR:FUNC is set to CURR, the :SENS:FUNC setting will be
'CURR:DC'. A similar command coupling exists for :SOUR:V OL T :RANG/
:SENS:VOLT:RANG and SOUR:CURR:RANG:SENS:CURR:RANG.
Table 2-2
Calibration unlocked states
ModeStateEquivalent remote command
Concurrent Functions
Sense Function
Sense Volts NPLC
Sense Volts Range
Sense Current NPLC
Sense Current Range
Filter Count
Filter Control
Filter A veraging
Source V Mode
Volts Autorange
Source I Mode
Current Autorange
Autozero
Trigger Arm Count
Trigger Arm Source
Trigger Count
Trigger Source
OFF
Source
1.0
Source V
1.0
Source I
10
REPEAT
ON
FIXED
OFF
FIXED
OFF
ON
1
Immediate
1
Immediate
:SENS:FUNC:CONC OFF
:SENS:FUNC <source_ function>
:SENS:VOLT:NPLC 1.0
:SENS:VOLT:RANG <source_V_range>
:SENS:CURR:NPLC 1.0
:SENS:CURR:RANG <source_ I_ range>
:SENS:AVER:COUN 10
:SENS:AVER:TCON REPeat
:SENS:AVER:STAT ON
:SOUR:VOLT:MODE FIXED
:SOUR:VOLT:RANG:AUTO OFF
:SOUR:CURR:MODE FIXED
:SOUR:CURR:RANG:AUTO OFF
:SYST:AZERO ON
:ARM:COUNT 1
:ARM:SOUR IMMediate
:TRIG:COUNT 1
:TRIG:SOUR IMMediate
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE '<password>'
For example, the following command uses the default password:
:CAL:PROT:CODE 'KI002420'
2-6Calibration
Changing the password
The default password may be changed from the front panel or via remote as discussed belo w.
Changing the password from the front panel
Follow the steps below to change the password from the front panel:
1.Press the MENU key, then choose CAL and press ENTER. The instrument will display
the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
2.Select UNLOCK, then enter the password. (Default: 002420.)
3.Select CHANGE-P ASSWORD, and then press ENTER. The instrument will display the
following:
New Pwd: 002420
Use , , , , ENTER or EXIT.
4.Using the range keys, and the left and right arrow keys, enter the new password on the
display.
5.Once the desired password is displayed, press the ENTER key to store the ne w password.
Changing the password by remote
T o change the calibration password by remote, first send the present passw ord, and then send
the new password. For example, the following command sequence changes the password from
the 'KI002420' remote default to 'KI_CAL':
:CAL:PROT:CODE 'KI002420'
:CAL:PROT:CODE 'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTE
If you change the first two characters of the password to something other than
"KI", you will not be able to unlock calibration from the front panel.
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the CAL
pads, which are located on the display board. Doing so will also reset the password to the factory
default (KI002420).
See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the display board component layout drawing at the end of Section 6 for the location of the CAL pads.
Calibration 2-7
V iewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be acces-
sible in the calibration menu. To view calibration dates and calibration count at any time:
1.From normal display, press MENU, select CAL, and then press ENTER. The unit will
display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
2.Select VIEW -D ATES, and then press ENTER. The Model 2420 will display the next and
last calibration dates and the calibration count as in the following example:
NEXT CAL: 02/15/98
Last calibration: 02/15/97 Count: 0001
Calibration errors
The Model 2420 checks for errors after each calibration step, minimizing the possibility that
improper calibration may occur due to operator error.
Front panel error reporting
If an error is detected during comprehensive calibration, the instrument will display an appropriate error message (see Appendix B). The unit will then prompt you to repeat the calibration
step that caused the error.
Remote error reporting
You can detect errors while in remote by testing the state of EAV (Error A vailable) bit (bit 2)
in the status byte. (Use the *STB? query to request the status byte.) Query the instrument for the
type of error by using the appropriate :SYST:ERR? query. The Model 2420 will respond with the
error number and a text message describing the nature of the error. See Appendix B for details.
Front panel calibration
The front panel calibration procedure described below calibrates all ranges of both the current
and voltage source and measure functions. Note that each range is separately calibrated by repeating the entire procedure for each range.
Step 1: Prepare the Model 2420 for calibration
1.Turn on the Model 2420 and the digital multimeter, and allow them to warm up for at
least one hour before performing calibration.
2-8Calibration
2.Press the MENU key, then choose CAL and press ENTER. Select UNLOCK, and then
press ENTER. The instrument will display the following:
PASSWORD:
Use , , , , ENTER or EXIT.
3.Use the up and down range keys to select the letter or number, and use the left and right
arrow keys to choose the position. Enter the present password on the display. (Front
panel default: 002420.) Press ENTER to complete the process.
4.Press EXIT to return to normal display. Instrument operating states will be set as summarized in Table 2-2.
Step 2: Voltage calibration
Perform the steps below for each voltage range, using Table 2-3 as a guide.
1.Connect the Model 2420 to the digital multimeter, as shown in Figure 2-1. Select the
multimeter DC volts measurement function.
NOTE
Figure 2-1
Voltage calibration
test connections
The 2-wire connections shown assume that remote sensing is not used. Remote
sensing may be used, if desired, b ut it is not essential when using the r ecommended digital multimeter.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
2420 3A SourceMeter
LO
MEAS
EDIT
V
DISPLAY
TOGGLE
LOCAL
POWER
67
DIGITS SPEED
SOURCE
FCTN
I
I
Ω
V
1
230
4
5
EDIT
TRIG
REL
SWEEP
LIMIT
FILTER
+/-
89
EXIT ENTER
RECALL
STORE
CONFIG MENU
250V
PEAK
RANGE
AUTO
TERMINALS
ON/OFF
FRONT/
RANGE
REAR
OUTPUT
Model 2420
Input HI
Input LO
Digital Multimeter
2.From normal display, press the SOURCE V key.
3.Press the EDIT key to select the source field (cursor flashing in source display field), and
then use the down RANGE key to select the 200mV source range.
4.From normal display, press MENU.
5.Select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
6.Select EXECUTE, and then press ENTER. The instrument will display the following message:
V-CAL
Press ENTER to Output +200.00mV
Calibration 2-9
7.Press ENTER. The Model 2420 will source +200mV and simultaneously display the following:
DMM RDG: +200.0000mV
Use , , , ,
ENTER, or EXIT.
8.Note and record the DMM reading, and then adjust the Model 2420 display to agree
exactly with the actual DMM reading. Use the up and down arro w keys to select the digit
value, and use the left and right arrow k eys to choose the digit position (or use the number keys, 0-9, +/-). Note that the display adjustment range is within ±10% of the present
range.
9.After adjusting the display to agree with the DMM reading, press ENTER. The instrument will then display the following:
V-CAL
Press ENTER to Output +000.00mV
10.Press ENTER. The Model 2420 will source 0mV and at the same time display the following:
DMM RDG: +000.0000mV
Use , , , ,ENTER, or EXIT.
11.Note and record the DMM reading, and then adjust the Model 2420 display to agree with
the actual DMM reading. Note that the display value adjustment limits are within ±1%
of the present range.
12.After adjusting the display value to agree with the DMM reading, press ENTER. The
unit will then display the following:
V-CAL
Press ENTER to Output -200.00mV
13.Press ENTER. The Model 2420 will source -200mV and display the following:
DMM RDG: -200.0000mV
Use , , , , ENTER, or EXIT.
14.Note and record the DMM reading, and then adjust the Model 2420 display to agree with the
DMM reading. Again, the maximum display adjustment is within ± 10% of the present
range.
15.After adjusting the display value to agree with the DMM reading, press ENTER and note
that the instrument displays:
V-CAL
Press ENTER to Output -000.00mV
16.Press ENTER The Model 2420 will source -0mV and simultaneously display the following:
DMM RDG: +000.0000mV
Use , , , ,ENTER or EXIT.
17.Note and record the DMM reading, and then adjust the display to agree with the DMM
reading. Once again, the maximum adjustment is within ±1% of the present range.
18.After adjusting the display to agree with the DMM reading, press ENTER to complete
calibration of the present range.
19.Press EXIT to return to normal display, and then select the 2V source range. Repeat steps
2 through 18 for the 2V range.
20.After calibrating the 2V range, repeat the entire procedure for the 20V and 60V ranges
using Table 2-3 as a guide. Be sure to select the appropriate source range with the EDIT
and RANGE keys before calibrating each range.
21.Press EXIT as necessary to return to normal display.
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
___________ V
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
Step 3: Current calibration
Perform the steps below for each current range, using Table 2-4 as a guide.
1.Connect the Model 2420 to the digital multimeter, as shown in Figure 2-2. Select the
multimeter DC current measurement function.
Figure 2-2
µ
A to 1A range current
10
calibration test connections
MEAS
EDIT
V
I
Ω
DISPLAY
1
230
TOGGLE
LOCAL
REL
FILTER
POWER
67
89
DIGITS SPEED
STORE
Model 2420
Digital Multimeter
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
2420 3A SourceMeter
LO
SOURCE
FCTN
I
V
4
5
EDIT
TRIG
SWEEP
LIMIT
+/-
EXIT ENTER
RECALL
CONFIG MENU
250V
PEAK
RANGE
AUTO
TERMINALS
ON/OFF
FRONT/
RANGE
REAR
OUTPUT
Input LO
Amps
Calibration 2-11
2.From normal display, press the SOURCE I key.
3.Press the EDIT key to select the source display field, and then use the down RANGE k ey
to select the 10µA source range.
4.From normal display, press MENU.
5.Select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES SAVE LOCK CHANGE-PASSWORD
6.Select EXECUTE, and then press ENTER. The instrument will display the following message:
I-CAL
Press ENTER to Output +10.000µA
7.Press ENTER. The Model 2420 will source +10µA and simultaneously display the fol-
lowing:
DMM RDG: +10.00000µA
Use
, , , , ENTER, or EXIT.
8.Note and record the DMM reading, and then adjust the Model 2420 display to agree
exactly with the actual DMM reading. Use the up and down arro w k eys to select the digit
value, and use the left and right arrow k eys to choose the digit position (or use the number
keys, 0-9, +/-). Note that the display adjustment range is within ±10% of the present range.
9.After adjusting the display to agree with the DMM reading, press ENTER. The instrument will then display the following:
I-CAL
Press ENTER to Output +00.000µA
10.Press ENTER. The Model 2420 will source 0mA and at the same time display the following:
DMM RDG: +00.00000µA
Use , , , , ENTER, or EXIT.
11.Note and record the DMM reading, and then adjust the Model 2420 display to agree with
the actual DMM reading. Note that the display value adjustment limits are within ±1%
of the present range.
12.After adjusting the display value to agree with the DMM reading, press ENTER. The
unit will then display the following:
I-CAL
Press ENTER to Output -10.000µA
13.Press ENTER. The Model 2420 will source -10µA and display the following:
DMM RDG: -10.00000µA
Use , , , , ENTER, or EXIT.
14.Note and record the DMM reading, and then adjust the Model 2420 display to agree with
the DMM reading. Again, the maximum display adjustment is within ± 10% of the
present range.
15.After adjusting the display value to agree with the DMM reading, press ENTER and note
that the instrument displays:
I-CAL
Press ENTER to Output -00.000µA
2-12Calibration
16.Press ENTER The Model 2420 will source -0µA and simultaneously display the follo wing:
DMM RDG: +00.00000µA
Use , , , , ENTER, or EXIT.
17.Note and record the DMM reading, and then adjust the display to agree with the DMM
reading. Once again, the maximum adjustment is within ±1% of the present range.
18.After adjusting the display to agree with the DMM reading, press ENTER to complete
calibration of the present range.
19.Press EXIT to return to normal display , and then select the 100µA source range use the
EDIT and up RANGE keys. Repeat steps 2 through 18 for the 100µA range.
20.After calibrating the 100µA range, repeat the entire procedure for the 1mA through 1A
ranges using T able 2-4 as a guide. Be sure to select the appropriate source range with the
EDIT and up RANGE keys before calibrating each range.
21.After calibrating the 1A range, connect the 1Ω characterized resistor and DMM to the
Model 2420 INPUT/OUTPUT jacks, as shown in Figure 2-3.
22.Select the DMM DC volts function.
23.Repeat steps 2 through 18 for the 3A range using Table 2-4 as a guide. When entering
the DMM reading, use the calculated current as follows: I = V/R, where V is the DMM
voltage reading, and R is the characterized value of the 1Ω resistor.
Figure 2-3
3A range current
calibration test
connections
MEAS
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
FCTN
I
Ω
1
230
REL
LIMIT
FILTER
89
RECALL
STORE
Model 2420
Digital Multimeter
V
4
TRIG
+/-
CONFIG MENU
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
75V
5V
PEAK
PEAK
PEAK
2420 3A SourceMeter
SOURCE
I
5
EDIT
SWEEP
EXIT ENTER
LO
250V
RANGE
AUTO
RANGE
PEAK
TERMINALS
ON/OFF
FRONT/
REAR
OUTPUT
1Ω Resistor
Input HI
Input LO
Table 2-4
Front panel current calibration
Source range*Source current
10µA+10.000µA
+00.000µA
-10.000µA
-00.000µA
100µA+100.00µA
+000.00µA
-100.00µA
-000.00µA
1mA+1.0000mA
+0.0000mA
-1.0000mA
-0.0000mA
10mA+10.000mA
+00.000mA
-10.000mA
-00.000mA
100mA+100.00mA
+000.00mA
-100.00mA
-000.00mA
1A+1.0000A
+0.0000A
-1.0000A
-0.0000A
3A***+3.0000A
+0.0000A
-3.0000A
-0.0000A
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
***Current calculated as follows: I = V/R, where V is the DMM reading, and R is the
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ A
___________ A
___________ A
___________ A
___________ A
___________ A
___________ A
___________ A
Calibration 2-13
Step 4: Enter calibration dates and save calibration
NOTEFor temporary calibration without saving new calibration constants, proceed to
Step 5: Lock out calibration.
1.From normal display, press MENU.
2.Select CAL, and then press ENTER. The Model 2420 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES SAVE LOCK CHANGE-PASSWORD
2-14Calibration
3.Select SAVE, and then press ENTER. The instrument will display the following message:
SAVE CAL
Press ENTER to continue; EXIT to abort calibration sequence.
4.Press ENTER. The unit will prompt you for the calibration date:
CAL DATE: 2/15/97
Use , , , , ENTER, or EXIT.
5.Change the displayed date to today's date, and then press the ENTER key . Press ENTER
again to confirm the date.
6.The unit will then prompt for the calibration due date:
NEXT CAL: 2/15/98
Use , , , , ENTER, or EXIT.
7.Set the calibration due date to the desired value, and then press ENTER. Press ENTER
again to confirm the date.
8.Once the calibration dates are entered, calibration is complete, and the following message will be displayed:
CALIBRATION COMPLETE
Press ENTER to confirm; EXIT to abort
9.Press ENTER to save the calibration data (or press EXIT to abort without saving calibration data.) The following message will be displayed:
CALIBRATION SUCCESS
Press ENTER or EXIT to continue.
10.Press ENTER or EXIT to complete process.
Step 5: Lock out calibration
1.From normal display, press MENU.
2.Select CAL, and then press ENTER. The Model 2420 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES SAVE LOCK CHANGE-PASSWORD
3.Select LOCK, and then press ENTER. The instrument will display the following message:
CALIBRATION LOCKED.
Press ENTER or EXIT to continue.
4.Press ENTER or EXIT to return to normal display.
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands o ver
the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are separately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed. ( For example, omitting a negative full-scale step).
Ranges that calibrated successfully will be saved if calibration is unlocked. Ranges that failed will not be saved.
Calibration subsystem.
Cal commands protected by password.
Unlock cal; changes password if cal is already unlocked.
(Default password: KI002420.)
Query number of times 2420 has been calibrated.
Save calibration data to EEPROM.*
Lock calibration, inhibit SAVE command operation.
Request cal lock status.
Program calibration year, month, day.
Query calibration year, month, day.
Program calibration due year, month, day.
Query calibration due year, month, day.
Calibrate active measure range. (See Table 2-6 parameters.)
Query measurement cal constants for active range.
Calibrate active source range. (See Table 2-7 parameters.)
Query source cal constants for active range.
Recommended calibration parameters
The maximum calibration command parameter ranges are: 75% to 150% of full scale for positive and negative full scale calibration points; ± zero calibration steps have ±50% of full scale
for valid entry ranges. However, for optimum calibration, it is recommended that you use calibration points within the ranges listed in Table 2-6 and Table 2-7. Note that each sense range
requires three parameters: zero, negativ e full scale, and positive full scale. Similarly , each source
range requires four parameters: two zero parameters, a positive full-scale parameter, and a negative full-scale parameter.
0.2V-0.002 to +0.002-0.18 to -0.22+0.18 to +0.22
2V-0.02 to +0.02-1.8 to -2.2+1.8 to +2.2
20V-0.2 to +0.2-18 to -22+18 to +22
60V-0.5 to +0.5-50 to -65+50 to +65
10µA-1E-7 to +1E-7-9E-6 to -11E-6+9E-6 to +11E-6
100µA-1E-6 to +1E-6-90E-6 to -110E-6+90E-6 to +110E-6
1mA-1E-5 to +1E-5-0.9E-3 to -1.1E-3+0.9E-3 to +1.1E-3
10mA-1E-4 to +1E-4-9E-3 to -11E-3+9E-3 to +11E-3
100mA-1E-3 to +1E-3-90E-3 to -110E-3+90E-3 to +110E-3
1A-1E-2 to +1E-2-0.9 to -1.1+0.9 to +1.1
3A-3E-2 to +3E-2-2.7 to -3.15+2.7 to +3.15
Note: Parameter steps for each range may be performed in any order, but all three parameter steps for each
range must be completed. For optimum calibration, use parameters within recommended limits.
0.2V-0.18 to -0.22-0.002 to +0.002+0.18 to +0.22-0.002 to +0.002
2V-1.8 to -2.2-0.02 to +0.02+1.8 to +2.2-0.02 to +0.02
20V-18 to -22-0.2 to +0.2+18 to +22-0.2 to +0.2
60V-50 to -65-0.5 to +0.5+50 to +65-0.5 to +0.5
10µA-9E-6 to -11E-6-1E-7 to +1E-7+9E-6 to +11E-6-1E-7 to +1E-7
100µA-90E-6 to -110E-6-1E-6 to +1E-6+90E-6 to +110E-6-1E-6 to +1E-6
1mA-0.9E-3 to -1.1E-3-1E-5 to +1E-5+0.9E-3 to +1.1E-3-1E-5 to +1E-5
10mA-9E-3 to -1E-3-1E-4 to +1E-4+9E-3 to +11E-3-1E-4 to +1E-4
100mA-90E-3 to -110E-3-1E-3 to +1E-3+90E-3 to +110E-3-1E-3 to +1E-3
1A-0.9 to -1.1-1E-2 to +1E-2+0.9 to +1.1-1E-2 to +1E-2
3A-2.7 to -3.15-3E-2 to +3E-2+2.7 to +3.15-3E-2 to +3E-2
Note: Parameter steps for each range may be performed in any order, b ut all four parameter steps for each range
must be completed. For optimum calibration, use parameters within recommended limits.
Remote calibration procedure
Step 1: Prepare the Model 2420 for calibration
1.Connect the Model 2420 to the controller IEEE-488 interface or RS-232 port using a
shielded interface cable.
2.Turn on the Model 2420 and the test equipment, and allo w them to w arm up for at least
one hour before performing calibration.
3.If you are using the IEEE-488 interface, make sure the primary address of the Model 2420
is the same as the address specified in the program you will be using to send commands.
(Use the MENU key and the COMMUNICATION menu to access the IEEE-488 address.)
Step 2: Voltage calibration
1.Connect the Model 2420 to the digital multimeter (see Figure 2-1), and select the multimeter DC volts function.
2.Send the commands summarized in Table 2-8 in the order listed to initialize voltage calibration. (When the :CAL:PROT:CODE command is sent, the instrument will assume
the operating states listed in Table 2-2.)
3.Perform the range calibration steps listed in Table 2-9 for all ranges. For each range:
• Send the :SOUR:VOLT:RANG command to select the source and sense range being
calibrated.. For example, for the 2V range, the following command would be sent
:SOUR:VOLT:RANG 2
• Program the source to output the negative full-range value using the :SOUR:VOLT
command.
For example:
:SOUR:VOLT -2
• Note and record the multimeter reading.
• Use the multimeter reading as the parameter for the :CAL:PRO T:SOUR and
• :CAL:PROT :SENS commands. F or example, a typical value for the 2V range w ould be:
:CAL:PROT:SOUR -1.998
:CAL:PROT:SENS -1.998
• Program the voltage source for 0V output using the :SOUR:VOLT 0.0 command.
• Note the multimeter reading.
• Send the source and sense calibration commands using the multimeter reading for the
parameter. For example:
:CAL:PROT:SOUR 1E-3
:CAL:PROT:SENS 1E-3
• Set the source to the positive full-range value using the :SOUR:VOLT command. For
example:
:SOUR:VOLT 2
Calibration 2-17
2-18Calibration
• Note and record the multimeter reading.
• Send the source and sense commands using the multimeter reading as the parameter .
For example:
:CAL:PROT:SOUR 1.997
:CAL:PROT:SENS 1.997
• Send the SOUR:VOLT 0.0 command to set the source voltage to 0V.
• Note and record the multimeter reading.
• Send the :CAL:PROT :SOUR command using the multimeter reading as the command
parameter. For example:
:CAL:PROT:SOUR -1.02E-3
Table 2-8
Voltage calibration initialization commands
CommandDescription
*RSTRestore GPIB defaults.
:SOUR:FUNC VOLTActivate voltage source.
:SENS:CURR:PROT 0.1Current limit when voltage source is active.
:SENS:CURR:RANG 0.1Make sure 1A range is not active.
:SOUR:VOLT:PROT MAXMaximum allowable source voltage.
:SYST:RSEN OFFDisable remote sensing. *
:CAL:PROT:CODE 'KI002420' Unlock cal.
:OUTP:STAT ONTurn source on.
* Remote sensing may be used if desired, but is not essential when using the recommended
digital multimeter. To use remote sensing, send :SYST:RSEN ON.
Calibration 2-19
Table 2-9
Voltage range calibration commands
Step Command/procedure*Description
1:SOUR:VOLT:RANGE <Range>Select source range.
2:SOUR:VOLT -<Source_value>Establish negative full-range polarity.
3Take DMM reading.Read actual output value.
4:CAL:PROT:SOUR <DMM_Reading>Calibrate sense function negative full scale.
5 Check 2420 for errors.
6:CAL:PROT:SENS <DMM_Reading>Calibrate source function negative full scale.
7Check 2420 for errors.
8:SOUR:VOLT 0.0Set output to 0V.
9Take DMM reading.Read actual output value.
10:CAL:PROT:SOUR <DMM_Reading>Calibrate sense function negative zero.
11Check 2420 for errors.
12:CAL:PROT:SENS <DMM_Reading>Calibration source function negative zero.
13 Check 2420 for errors.
14:SOUR:VOLT +<Source_value>Establish positive full-range polarity.
15Take DMM reading.Read actual output value
16:CAL:PROT:SOUR <DMM_Reading>Calibrate sense function positive full scale.
17 Check 2420 for errors.
18 :CAL:PROT:SENS <DMM_Reading>Calibrate source function positive full scale.
19 Check 2420 for errors.
20:SOUR:VOLT 0.0Set output to 0V.
21Take DMM reading.Read actual output value.
22:CAL:PROT:SOUR <DMM_Reading>Calibrate source positive zero.
*1. Perform complete procedure for each range, where <Range> = 0.2, 2, 20, and 60, and
<Source_value> = 0.2, 2, 20, and 60.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
Step 3: Current calibration
1.Connect the Model 2420 to the digital multimeter (see Figure 2-2), and select the multimeter DC current function.
2.Send the commands summarized in Table 2-10 in the order listed to initialize current calibration.
3.Calibrate the 10µA to 1A current ranges using the procedure summarized in T able 2-11.
For each range:
•Send the :SOUR:CURR:RANG command to select the source and sense range
being calibrated. For example, for the 1mA range, the command is:
:SOUR:CURR:RANG 1E-3
2-20Calibration
•Program the source to output the negative full-range value using the
:SOUR:CURR command. For example:
:SOUR:CURR -1E-3
• Note and record the multimeter reading.
•Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 1mA range
would be:
:CAL:PROT:SOUR -1.025E-3
:CAL:PROT:SENS -1.025E-3
•Program the current source for 0A output using the :SOUR:CURR 0.0 command.
•Note the multimeter reading.
•Send the source and sense calibration commands using the multimeter reading for
the parameter. For example:
:CAL:PROT:SOUR 1E-6
:CAL:PROT:SENS 1E-6
•Set the source to the positive full-range v alue using the :SOUR:CURR command.
For example, for the 1mA range:
:SOUR:CURR 1E-3
•Note and record the multimeter reading.
•Send the source and sense commands using the multimeter reading as the param-
eter. For example:
:CAL:PROT:SOUR 1.03E-3
:CAL:PROT:SENS 1.03E-3
•Send the :SOUR:CURR 0.0 command to set the source current to 0A.
•Note and record the multimeter reading.
•Send the :CAL:PROT:SOUR command using the multimeter reading as the com-
mand parameter. For example:
:CAL:PROT:SOUR 1E-6
4.Connect the 1Ω resistor and DMM to the Model 2420 INPUT/OUPUT jacks, as shown
in Figure 2-3. Select the DMM DC volts function.
5.Repeat step 3 for the 3A range using the calculated current as follows: I = V/R, where V
is the DMM voltage reading, and R is the characterized value of the 1Ω resistor.
Table 2-10
Current calibration initialization commands
CommandDescripton
:SOUR:FUNC CURRSelect source current mode.
:SENS:VOLT:PROT 20Voltage limit when current source is active.
:SENS:VOLT:RANG 20Make sure 60V range is not active.
:OUTP:STAT ONTurn source on.
Calibration 2-21
Table 2-11
Current range calibration commands
Step Command/procedure*Description
1
:SOUR:CURR:RANGE <Range>
2
:SOUR:CURR -<Source_value>
3
Take DMM reading.
4
:CAL:PROT:SOUR <DMM_Reading>
5
Check 2420 for errors.
6
:CAL:PROT:SENS <DMM_Reading>
7
Check 2420 for errors.
8
:SOUR:CURR 0.0
9
Take DMM reading.
10
:CAL:PROT:SOUR <DMM_Reading>
11
Check 2420 for errors.
12
:CAL:PROT:SENS <DMM_Reading>
13
Check 2420 for errors.
14
:SOUR:CURR +<Source_value>
15
Take DMM reading.
16
:CAL:PROT:SOUR <DMM_Reading>
17
Check 2420 for errors.
18
:CAL:PROT:SENS <DMM_Reading>
19
Check 2420 for errors.
20
:SOUR:CURR 0.0
21
Take DMM reading.
22
:CAL:PROT:SOUR <DMM_Reading>
*1. Perform complete procedure for each range, where <Range> and <Source_value> = 10E-6, 100E-6, 1E-3,
10E-3, 100E-3, 1, or 3.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A.
Read actual output value.
Calibrate source positive zero.
Step 4: Program calibration dates
Use following commands to set the calibration date and calibration due date:
Note that the year, month, and date must be separated by commas. The allowable range for
the year is from 1997 to 2096, the month is from 1 to 12, and the date is from 1 to 31.
Step 5: Save calibration constants
Calibration is now complete, so you can store the calibration constants in EER OM by sending
the following command:
:CAL:PROT:SAVE
2-22Calibration
NOTECalibration will be temporary unless you send the SAVE command. Also, calibra-
tion data will not be saved if (1) calibration is locked, (2) invalid data exists, or
(3) all steps were not completed.
Step 6: Lock out calibration
T o lock out further calibration, send the follo wing command after completing the calibration
procedure:
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the entire
instrument is properly calibrated. In some instances, however, you may want to calibrate only
certain ranges. To do so, complete the entire procedure only for the range(s) to be calibrated.
Keep in mind, however, that you must complete all parameter steps for each source or sense
range. Also be sure to set calibration dates and save calibration after calibrating the desired
range(s).
3
Routine Maintenance
3-2Routine Maintenance
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
Introduction
The information in this section deals with routine type maintenance that can be performed by
the operator.
Line fuse replacement
Figure 3-1
Rear panel
WARNING
Disconnect the line cord at the rear panel, and remove all test leads connected to the instrument (front and rear) before replacing the line fuse.
The power line fuse is accessible from the rear panel, just abov e the AC po wer receptacle (see
Figure 3-1).
5V
HI
75V
PEAK
4-WIRE
SENSE
WITH FRONT PANEL MENU)
5V
PEAK
LO
INPUT/
OUTPUT
IEEE-488
(ENTER IEEE ADDRESS
75V
PEAK
PK
250V
PEAK
PEAK
5V
V, Ω,
GUARD
GUARD
SENSE
RS232
MADE IN
U.S.A.
TRIGGER
LINK
LINE FUSE
SLOWBLOW
3.15A, 250V
LINE RATING
90-250VAC
50, 60HZ
220VA MAX
INTERLOCK-
DIGITAL I/O
Perform the following steps to replace the line fuse:
1.Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to the
fuse holder.
2.Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTION
To prevent instrument damage, use only the fuse type specified in Table 3-1.
3.Re-install the fuse carrier.
Routine Maintenace 3-3
NOTE
If the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional information.
Table 3-1
Power line fuse
Line voltageRatingKeithley part no.
90-250V250V, 3.15A, Slow
×
Blow 5
20mm
FU-106-3.15
3-4Routine Maintenance
4
Troubleshooting
4-2Troubleshooting
Introduction
This section of the manual will assist you in troubleshooting and repairing the Model 2420.
Included are self-tests, test procedures, troubleshooting tables, and circuit descriptions. Note
that disassembly instructions are located in Section 5, and component layout drawings are at the
end of Section 6.
WARNING
The information in this section is intended for qualified service personnel
only. Do not perform these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages that could
cause personal injury or death. Use caution when working with hazardous
voltages.
Repair considerations
Before making any repairs to the Model 2420, be sure to read the following considerations.
CAUTION
•Repairs will require various degrees of disassembly. However, it is recommended that
the Front Panel Tests be performed prior to any disassembly. The disassembly instructions for the Model 2420 are contained in Section 5 of this manual.
•Do not make repairs to surface mount PC-boards unless equipped and qualified to do so
(see previous CAUTION).
•When working inside the unit and replacing parts, be sure to adhere to the handling precautions and cleaning procedures explained in Section 5.
•Many CMOS devices are installed in the Model 2420. These static-sensitive devices
require special handling as explained in Section 5.
•Whenever a circuit board is remov ed or a component is replaced, the Model 2420 must
be recalibrated. See Section 2 for details on calibrating the unit.
The PC-boards are built using surface mount techniques and require specialized equipment and skills for repair. If you are not equipped and/or qualified,
it is strongly recommended that you send the unit back to the factory for repairs or limit repairs to the PC-board replacement level. Without proper
equipment and training, you could damage a PC-board beyond repair.
Power-on self-test
During the power-on sequence, the Model 2420 will perform a checksum test on its EPROM
and test its RAM. If one of these tests fails, the instrument will lock up.
Front panel tests
There are three front panel tests: one to test the functionality of the front panel keys and tw o
to test the display . In the e vent of a test f ailure, refer to Display board checks for details on troubleshooting the display board.
KEYS test
The KEYS test lets you check the functionality of each front panel key . Perform the follo wing
steps to run the KEYS test.
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select KEYS, and press ENTER to start the test. When a key is pressed, the label name
for that key will be displayed to indicate that it is functioning properly. When the key is
released, the message “No keys pressed” is displayed.
5.Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT
aborts the test and returns the instrument to the SELF-TEST MENU. Continue pressing
EXIT to back out of the menu structure.
Troubleshooting 4-3
DISPLA Y P A TTERNS test
The display test lets you verify that each pixel and annunciator in the v acuum fluorescent dis-
play is working properly. Perform the following steps to run the display test:
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select DISPLAY-PATTERNS, and press ENTER to start the display test. There are five
parts to the display test. Each time a front panel key (except EXIT) is pressed, the next
part of the test sequence is selected. The five parts of the test sequence are as follows:
• Checkerboard pattern (alternate pixels on) and all annunciators.
• Checkerboard pattern and the annunciators that are on during normal operation.
• Horizontal lines (pixels) of the first digit are sequenced.
• Vertical lines (pixels) of the first digit are sequenced.
• Each digit (and adjacent annunciator) is sequenced. All the pixels of the selected digit
are on.
5.When finished, abort the display test by pressing EXIT . The instrument returns to the
FRONT P ANEL TESTS MENU. Continue pressing EXIT to back out of the menu structure.
4-4Troubleshooting
CHAR SET test
The character set test lets you display all characters. Perform the following steps to run the
character set test:
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select CHAR-SET, and press ENTER to start the character set test. Press an y ke y except
EXIT to cycle through all displayable characters.
5.When finished, abort the character set test by pressing EXIT. The instrument returns to
the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu
structure.
Principles of operation
The following information is provided to support the troubleshooting tests and procedures
covered in this section of the manual. Refer to the following drawings:
Figure 4-1 shows an overall block diagram of the Model 2420. Circuitry may be divided into
three general areas:
•Analog circuits — includes sourcing circuits such as the DACs, clamps, output stage,
and feedback circuits, as well as measurement circuits such as the A/D converter.
•Digital circuits — includes the microcomputer that controls the analog section, front
panel, and GPIB and RS-232 ports, as well as associated interfacing circuits.
•Power supplies — converts the AC line voltage into DC voltages that supply the power
for the digital and analog circuits, and the output stage.
Analog circuits
Figure 4-2 shows the analog circuitry block.
Troubleshooting 4-5
Figure 4-1
Overall block
diagram
Display,
Keyboard
Digital
I/O
Trigger
Link
Front
Panel
Controller
Trigger,
Digital
I/O
DACs
Analog Section
Clamps
Feedback
A/D
Converter
Microcomputer
Digital Section
Output
Stage
RS-232
GPIB
Interface
Guard
Buffer
Output
Guard
RS-232 I/O
GPIB I/O
To Analog
CircuitsToOutput Stage
±15V
+5V±42V
Analog
Power
Supply
Output
Stage
Power
Supply
Power Supply
To
Digital Circuits
±85V +5V +12V
Digital
Power
Supply
Line In
Figure 4-2
Analog circuitry block diagram
V DAC
I DAC
Control
FB
VFB
A/D
IFB
V Clamp
I Clamp
+7
MUX
Output
Stage
+42
+85
Sense
Resistors
O
Output
HI
S+
Output
LO
S-
VFB
Error
Amp
-42
-85
O
Remote
IFB
Protection
O
S
+
Protection
Guard
Out
Guard
Sense
4-6Troubleshooting
D/A converters control the programmed v oltage and current, or voltage compliance and current
compliance. Each DA C has two ranges, a 10V full-scale output or a 1V full-scale output. The D A C
outputs are fed to the summing node, FB. Either the V DA C or the I D A C has the ability to control
the main loop. If the unit is set for SV (source voltage), it will source voltage until the compliance
current is reached (as determined by the I DA C setting), and the current loop will o verride the voltage loop. If, howev er , the unit is set for SI (source current), it will source current until the compliance voltage is reached (as determined by the V DAC setting), and the voltage loop will ov erride
the current loop. A priority bit in the Vclamp/I clamp circuit controls these functions.
The error amplifier adds open-loop gain and slew-rate control to the system to assure accuracy and provide a controllable signal for the output stage, which provides the necessary voltage
and current gain to drive the output. Sense resistors in the HI output lead pro vide output current
sensing, and a separate sense resistor is used for each current range. The 1A and 3A ranges use
0.2V full-scale for a full-range output, while all other ranges use 2V output for full-scale current.
Voltage feedback is routed either internally or externally.
There are four voltage ranges: 0.2V, 2V, 20V, and 60V. The feedback gain changes only for
the 20V and 60V ranges, resulting in three unique feedback gain values. A multiplexer directs
the voltage feedback, current feedback, reference, or ground signal to the A/D converter. An
opto-isolated interface provides control signals for both DACs, analog circuit control, and A/D
converter communication to the digital section.
Power supply
Figure 4-3 shows a block diagram of the Model 2420 power delivery system.
Figure 4-3
Power supply block diagram
Line
Neutral
Analog Board
+15V+5V
FFF
Constant Frequency
Low Noise Floating
Switching Supply
D
PFC
DC/DC
Converter
F
+24Vdc
+24dc
D
-15V
Output Stage
-85V -42V+42V +85V
+12Vdc
DC
Regulators
+5Vdc
D
O
Digital
Circuits
The offline switching power supply provides all power for the instrument while providing
universal inputs and power factor correction for the 110/120V line. The digital board runs off
of 5V and 12V supplies derived from the switcher . The +12VD supply is set to program the flash
ROM. (See Digital circuitry below.)
A constant-frequency switching supply runs off the +24VD supplies and generates all the fl oating and output supply voltages for the analog board: +5V and ±15V, ±42V and ±85V.
Output stage
Figure 4-4 shows a simplified schematic of the output stage.
Troubleshooting 4-7
Figure 4-4
Output stage simplified
schematic
Maindrive
O
+15Vf
300K
-15Vf
F
HI Drive
+
-
+85V
Q500
+42V
Q504
Q514
Q516
Q518
F
Q521
F
Q523
HI Drive
Q525
Q507
-42V
Q503
-85V
4-8Troubleshooting
The Model 2420 output stage serves two purposes: (1) it converts signals from floating common to output common, and (2) it provides both voltage and current amplification. The output
stage drive transistors are biased in class B configuration to prevent the possibility of thermal
runaway with high-current output values.
Output transistors Q518 and Q521 are cascoded with output MOSFETs Q516 and Q523. All
other MOSFETs and transistors are slaves, and the v oltages across these devices are determined
by the resistor-capacitor ladder circuits shown. High-current drive capability is provided by
Q500-Q511. Coarse current limits are built into the output stage.
A/D converter
The SourceMeter unit uses a multi-slope charge balance A/D converter with a single-slope
run-down. The converter is controlled by gate array U610. Commands are issued by the MPU
on the digital board through communications opto-isolators to U610, and U610 sends A/D reading data back through opto-isolators to the digital board for calibration and processing.
Active guard
The Model 2420 has an active guard or "six-wire ohms" circuit used to measure complex devices. This circuitry provides a low-current (50mA) equivalent of the voltage on output HI. If
the unit is in the SV mode, the low-current equivalent of the source voltage will appear on the
guard terminal. If the unit is in the SI mode, the voltage on output HI is equal to the source cur rent multiplied by the external resistance value. An equivalent voltage will be generated by the
guard circuit, and a guard sense terminal is provided to sense around the voltage drop in the
guard leads since significant current can flow (50mA).
Digital circuitry
Refer to Figure 4-5 for the following discussion on digital circuitry.
The core digital circuitry uses a Motorola 68332 microcontroller running at 16.78MHz. The
memory configuration includes two 256K X 8-bit flash EEPR OMs and two 128K X 8-bit RAMs
used in parallel to utilize the 16-bit data bus of the MPU. The RAM is battery backed-up, providing continued storage of data buffer information during power-down c ycles, and flash ROM
support allows internal firmware upgrades using either the serial or GPIB port for downloading
new firmware. All calibration constants and the save 0 setup are stored in a separate serial EEPROM. Setups 1 through 4 are stored in battery backed-up RAM.
External communication is provided via GPIB and serial interfaces. A 9914 GPIB IEEE-488
standard interface IC is used for the GPIB, and a 68332 Queued Serial Module (QSM) provides
the serial UART. For internal communications, the Time Processing Unit (TPU) is used for serial
communications with the front panel display module, and both the TPU and QSM handle digitalto-analog interfacing.
Troubleshooting 4-9
Figure 4-5
Digital circuitry
block diagram
A/D
Control/Data
Reset
2
E PROM
U17
A/D Interface
U9, U25
Voltage Source
Control
ROM
U15, U16
Microprocessor
U3
RAM
U12, U14
16.78MHz
Serial
Interface
U4
GPIB
U6, U13
U20
Digital I/O
Trigger
U23
U7
RS-232 Interface
IEEE-488 Interface
To Display
Board Controller
Trigger
Digital
I/O
Display board circuit theory
Display board components are shown in the digital circuitry block diagram in Figure 4-5.
U902 is the display microcontroller that controls the VFD (vacuum fluorescent display) and
interprets key data. The microcontroller has four peripheral I/O ports that are used for the v arious
control and read functions.
Display data is serially transmitted to the microcontroller from the digital board via the TXB
line to the microcontroller PD0 terminal. In a similar manner, key data is serially sent back to
the digital board through the RXB line via PD1. The 4MHz clock for the microcontroller is generated on the digital board.
DS901 is the VFD (vacuum fluorescent display) module, which can display up to 49 characters. Each character is organized as a 5
bar segment to act as a cursor.
The display uses a common multiplexing scheme with each character refreshed in sequence.
U903 and U904 are the grid drivers, and U901 and U905 are the dot dri vers. Note that dot dri ver
and grid driver data is serially transmitted from the microcontroller (PD3 and PC1).
×
7 matrix of dots or pixels and includes a long under-
4-10Troubleshooting
The front panel keys (S901-S931) are organized into a row-column matrix to minimize the
number of microcontroller peripheral lines required to read the keyboard. A k ey is read by strobing the columns and reading all rows for each strobed column. K e y down data is interpreted by
the display microcontroller and sent back to the main microprocessor using proprietary encoding schemes.
T roubleshooting
Troubleshooting information for the various circuits is summarized below.
Display board checks
If the front panel display tests indicate that there is a problem on the display board, use Table
4-1. See Principles of operation for display circuit theory.
Table 4-1
Display board checks
Step Item/componentRequired conditionRemarks
1
2
3
4
5
6
Front panel test
J1033
U902, pin 1
U902, pin 43
U902, pin32
U902, pine 33
Verify that all segments operate.
+5V, ±5%
Goes low briefly on power up, and
then goes high.
4MHz square wave.
Pulse train every 1 ms.
Brief pulse train when front panel key
is pressed.
Use front panel display test.
Digital +5V supply.
Microcontroller RESET.
Controller 4MHz clock.
Control from main processor.
Key down data sent to main
processor.
Power supply checks
Power supply problems can be checked out using Table 4-2. See Principles of operation for
circuit theory on the power supply.
Table 4-2
Power supply checks
StepItem.componentRequired conditionRemarks
1
2
3
4
5
6
7
8
9
Line fuse
Line power
TP502
TP503
TP504
TP505
TP507
TP508
TP510
Digital circuitry checks
Digital circuit problems can be checked out using Table 4-3. See Principles of operation for
a digital circuit description.
Check continuity.
Plugged into live receptacle,
power on.
+80V, ±10%
-80V, ±10%
+38V, ±10%
-38V, ±10%
+15V, ±5%
-15V, ±5%
+5V, ±5%
Troubleshooting 4-11
Remove to check.
Check for correct power-up
sequence.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
+15VF, referenced to TP500.
-15VF, referenced to TP500.
+5VF, referenced to TP500.
Table 4-3
Digital circuitry checks
StepItem/componentRequired conditionRemarks
Verify that RAM and ROM are
functional.
All signals referenced to digital
common.
Digital logic supply.
MPU RESET line.
MPU address bus.
MPU data bus.
MPU clock.
RS-232 RX line.
RS-232 TX line.
IEEE-488 data bus.
IEEE-488 command lines.
IEEE-488 REN line.
IEEE-488 IFC line.
D_ADDATA
D_DATA
D_CLK
D_STB
>60V voltage protection
SOURCE +10V
SOURCE +10V (SVMI)
SOURCE +10V
SOURCE +10V
OUTPUT COM
OUTPUT COM
SVMI, OUTPUT ON, 20V
Bench defaults
-13V ±1V
-5V ±0.5V
-10V ±1V
-10.5V ±1V
0V ±0.1V
7V ±0.7V
7V ±0.7V
20V ±0.5V
6.4V ±0.6V
WARNING
The volatile memories of the Model 2420 are protected by a replaceable battery when po wer is
off. Typical battery life is approximately 10 years. The battery should be suspected if the instrument no longer retains buffer data or user -defined operating parameters such as instrument setups,
source memory , and math expressions. If the battery is absent or totally e xhausted, the display will
show the “Reading buf fer data lost” message shortly after the Model 2420 is switched on.
The battery is a 3V wafer-type lithium cell, Panasonic type BR-2330 or equivalent (Keithley
part number BA-46), which is located on the digital board. Replacement of the battery requires
removal of the case cover and analog board assembly. (See Section 5.)
Disconnect the instrument from the power line and all other equipment
before changing the battery.
Battery replacement precautions
WARNING
WARNING
There is a danger of explosion if battery is incorrectly replaced. Replace
only with the same or equivalent type recommended by the manufacturer.
Dispose of used batteries according to federal, state, and local laws.
The precautions below must be followed to avoid personal injury.
• Wear safety glasses or goggles when working with lithium batteries.
• Do not short the battery terminals together.
• Keep lithium batteries away from all liquids.
• Do not attempt to recharge lithium batteries.
• Observe proper polarity when inserting the battery in its holder.
• Do not incinerate or otherwise expose the battery to excessive heat (>60°C).
• Bulk quantities of lithium batteries should be disposed of as a hazardous
waste.
Battery replacement procedure
1.Remove the case cover and analog board assembly as covered in Section 5.
2.Locate the battery on the digital board.
3.Carefully unsolder and remove the old battery.
4.Install and solder the new battery in place.
5.Re-assemble the instrument, and turn it on. The “Reading buf fer data lost” error message
will be displayed.
6.Send the :SYST:MEM:INIT command via remote to perform the following:
• Clear the reading buffer.
• Initialize instrument setups 1 through 4 to present instrument settings.
• Initialize all 100 source memory locations to present instrument settings.
• Delete user math expressions.
No comm link error
A “No Comm Link” error indicates that the front panel processor has stopped communicating
with the main processor, which is located on the digital board. This error indicates that one of the
main processor ROMs may require re-seating in its sock et. ROMs may be reseated as follows:
1.Turn off the po wer, and disconnect the line cord and all other test leads and cables from
the instrument.
2.Remove the case cover as outlined in Section 5.
3.Remove the analog board assembly as outlined in Section 5.
4.Locate the two firmware ROMS, U15 and U16, located on the digital board. These are
the only ICs installed in sockets. (Refer to the component layout drawing at the end of
Section 6 for exact locations.)
5.Carefully push down on each ROM IC to make sure it is properly seated in its socket.
Troubleshooting 4-13
CAUTION
6.Connect the line cord, and turn on the power . If the problem persists, additional troubleshooting will be required.
Be careful not to push down excessively, or you might crack the digital
board.
4-14Troubleshooting
5
Disassembly
5-2Disassembly
Introduction
This section explains how to handle, clean, and disassemble the Model 2420. Disassembly
drawings are located at the end of this section.
Handling and cleaning
T o av oid contaminating PC board traces with body oil or other foreign matter , av oid touching
the PC board traces while you are repairing the instrument. Motherboard areas covered by the
shield have high-impedance de vices or sensitive circuitry where contamination could cause degraded performance.
Handling PC boards
Observe the following precautions when handling PC boards:
•Wear cotton gloves.
•Only handle PC boards by the edges and shields.
•Do not touch any board traces or components not associated with repair.
•Do not touch areas adjacent to electrical contacts.
•Use dry nitrogen gas to clean dust off PC boards.
Solder repairs
Observe the following precautions when you must solder a circuit board:
•Use an OA-based (organic activated) flux, and take care not to spread the flux to other
areas of the circuit board.
•Remove the flux from the work area when you have finished the repair by using pure
water with clean, foam-tipped swabs or a clean, soft brush.
•Once you have removed the flux, swab only the repair area with methanol, then blowdry the board with dry nitrogen gas.
•After cleaning, allow the board to dry in a 50°C, low-humidity environment for several
hours.
Static sensitive devices
CMOS devices operate at very high impedance le vels. Therefore, an y static that b uilds up on
you or your clothing may be sufficient to destroy these de vices if the y are not handled properly.
Use the following precautions to avoid damaging them:
Disassembly 5-3
CAUTION
•Transport and handle ICs only in containers specially designed to prevent static buildup. Typically, you will receive these parts in anti-static containers made of plastic or
foam. Keep these devices in their original containers until ready for installation.
•Remove the devices from their protective containers only at a properly grounded work
station. Also, ground yourself with a suitable wrist strap.
•Handle the devices only by the body; do not touch the pins.
•Ground any printed circuit board into which a semiconductor device is to be inserted to
the bench or table.
•Use only anti-static type desoldering tools.
•Use only grounded-tip solder irons.
•Once the device is installed in the PC board, it is normally adequately protected, and you
can handle the boards normally.
Many CMOS devices are installed in the Model 2420. Handle all semiconductor devices as being static sensitive.
Assembly drawings
Use the assembly drawings located at the end of this section to assist you as you disassemble
and re-assemble the Model 2420. Also, refer to these drawings for information about the Keithley part numbers of most mechanical parts in the unit. Assembly drawings include:
Follow the steps below to remove the case cover to gain access to internal parts.
WARNING
1.
Remove handle
gently pulling it away from the sides of the instrument case and swinging it up or do wn.
To remove the handle, swing the handle below the bottom surface of the case and back
until the orientation arrows on the handles line up with the orientation arrows on the
mounting ears. W ith the arro ws lined up, pull the ends of the handle a way from the case.
Before removing the case cover, disconnect the line cord and any test leads
from the instrument.
— The handle serves as an adjustable tilt-bail. Adjust its position by
5-4Disassembly
2.
Remove mounting ears
and out on each mounting ear.
— Remove the screw that secures each mounting ear . Pull do wn
NOTE
NOTE
When re-installing the mounting ears, make sure to mount the right ear to the
right side of the chassis, and the left ear to the left side of the chassis. Eac h ear is
marked “RIGHT” or “LEFT” on its inside surface.
3.
Remove rear bezel
rear bezel to the chassis, then pull the bezel away from the case.
4.
Remove grounding screws
to the chassis. They are located on the bottom of the case at the back.
5.
Remove chassis
fully slide the chassis forward. Slide the chassis out of the metal case.
To gain access to the components under the analog board shield, remove the
shield, which is secured to the analog board by a single screw.
— T o remov e the case, grasp the front bezel of the instrument, and care-
Analog board removal
Perform the following steps to remove the analog board . This procedure assumes that the
case cover is already removed.
1.Disconnect the front and rear input terminals.
You must disconnect these input terminal connections for both the front and rear inputs:
• INPUT/OUTPUT HI and LO
• 4-WIRE SENSE HI and LO
•V,
Ω
, GUARD and GUARD SENSE (rear panel only)
Remove all the connections by pulling the wires off the pin connectors, then remove the ferrite noise filters from the chassis. During re-assembly, use the following table to identify input
terminals:
— To remove the rear bezel, loosen the two screws that secure the
— Remove the two grounding screws that secure the case
Input terminalsFront wire colorRear wire color
INPUT/OUTPUT HI
INPUT/OUTPUT LO
4-WIRE SENSE HI
4-WIRE SENSE LO
Ω
, GUARD
V,
GUARD SENSE
2. Unplug cables.
• Carefully unplug the ribbon cables at J1027, J1028, and J1029.
• Unplug the ON/OFF cable at J1034.
Red
Black
Yellow
Gray
-
-
White/Red
White/Black
White/Yellow
White/Gray
White
Blue/White
Disassembly 5-5
3.Remove screws.
• Remove the two fastening screws that secure the analog board assembly to the
chassis. These screws are located on the side of the board opposite from the heat sink.
• Remove the two screws that secure the heat sink to the chassis.
4. Remove analog board assembly.
After all screws have been removed, carefully lift the analog board assembly free of the
main chassis.
5. Disassemble analog board assembly.
• Remove the screws that secure the analog board and heat sink to the analog board
subchassis.
• Carefully remove the heat sink by sliding the clips off the power transistors.
CAUTION
NOTE
Be careful not to damage the heat sink insulation layer.
• Remove the analog board from the subchassis.
• Remove the four screws that secure the bottom co ver , then remov e the cover
from the bottom of the PC board.
When re-installing the heat sink, make sure that all clips are properly installed
and centered on each pair of output transistors.
Digital board removal
Perform the following steps to remov e the digital board. This procedure assumes that the an-
alog board assembly is already removed.
NOTE
In order to remove the digital board, the display board must first be removed.
1.Remove the IEEE-488, Digital I/O, and RS-232 fasteners.
The IEEE-488, Digital I/O, and RS-232 connectors each have two nuts that secure the
connectors to the rear panel. Remove these nuts.
2.Remove the POWER switch rod.
At the switch, place the edge of a flat-blade screw driver in the notch on the pushrod.
Gently twist the screw driver while pulling the rod from the shaft.
3.Unplug cables:
• Unplug the display board ribbon cable.
• Unplug the cables going to the power supply.
• Unplug the rear panel power module cable.
• The fan may need to be removed.
4.Remove digital board.
Slide the digital board forward until it is free of the guide pins, then remove the board.
During re-assembly , replace the board, and start the IEEE-488, Digital I/O, and RS-232
connector nuts and the mounting screw. Tighten all the fasteners once they are all in
place and the board is correctly aligned.
5-6Disassembly
Front panel disassembly
Use the following procedures to remov e the display board and/or the pushb utton switch pad.
1.Unplug the display board ribbon cables.
2.Remove the front panel assembly.
This assembly has four retaining clips that snap onto the chassis over four pem nut studs.
Two retaining clips are located on each side of the front panel. Pull the retaining clips
outward and, at the same time, pull the front panel assembly forward until it separates
from the chassis.
3.Using a thin-bladed screw driver, pry the plastic PC board stop (located at the bottom of
the display board) until the bar separates from the casing. Pull the display board from the
front panel.
4.Remove the switch pad by pulling it from the front panel.
Removing power components
The following procedures to remove the power supply and/or power module require that the
case cover and motherboard be removed, as previously explained.
Power module removal
Perform the following steps to remove the rear panel power module:
1.Remove the analog board.
2.Unplug the cable connecting the power module to the digital board.
3.Disconnect the power module's ground wire. This green and yellow wire connects to a
threaded stud on the chassis with a kep nut.
4.Squeeze the latches on either side of the power module while pushing the module from
the access hole.
WARNING
To avoid electrical shock, which could result in injury or death, the ground
wire of the power module must be connected to chassis ground. When installing the power module, be sure to re-connect the green and yellow
ground wire to the threaded stud on the chassis.
Instrument re-assembly
Re-assemble the instrument by reversing the previous disassembly procedures. Make sure
that all parts are properly seated and secured, and that all connections are properly made. To ensure proper operation, replace the analog signal wire ferrite noise filters, and securely fasten the
shield.
Disassembly 5-7
WARNING
To ensure continued protection against electrical shock, verify that power
line ground (green and yellow wire attached to the power module) is connected to the chassis.
Also make sure the two bottom case screws are properly installed to secure
and ground the case cover to the chassis.
5-8Disassembly
6
Replaceable Parts
6-2Replaceable Parts
Introduction
This section contains replacement parts information and component layout drawings for the
Model 2420.
Parts lists
The electrical parts lists for the Model 2420 are shown tables at the end of this section. For
part numbers to the various mechanical parts and assemblies, use the Miscellaneous parts list
and the assembly drawings provided at the end of Section 5.
Ordering information
To place an order, or to obtain information concerning replacement parts, contact your Keithley representative or the f actory (see inside front cover for addresses). When ordering parts, be
sure to include the following information:
•Instrument model number (Model 2420)
•Instrument serial number
•Part description
•Component designation (if applicable)
•Keithley part number
Factory service
If the instrument is to be returned to Keithley Instruments for repair, perform the following:
•Call the Repair Department at 1-800-552-1115 for a Return Material Authorization
(RMA) number.
•Complete the service form at the back of this manual, and include it with the instrument.
•Carefully pack the instrument in the original packing carton.
•Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping
label.
Component layouts
The component layouts for the various circuit boards are provided on the following pages.