Keithley 1920 Instruction Manual

INSTKUCTIW MANUAL
IlUUEL 1YZU
TKWS UPTIUII
Copyright 1981 Keithley Instruments, Inc.
First Printing, July, 1981, Cleveland, Ohio, U.S.A.
Document No. 32000, Kev. A
Table of Contents
1.
2.
3.
4.
5.
ClENEKAL INFORMATION
.............................
l-l.
Introduction.
.............................
l-3. Warranty Information.
.........................
1-5.
Manual Addendums.
...........................
l-7. Safety Symbols and Terms.
.......................
l-9. Specifications.
............................
OPERATING INSTRUCTIONS.
...........................
2-l.
Operation in Model 191.
........................
2-4. Operation in Model 192.
........................
MAINTENANCE .................................
3-2. Installation in Model 191
.......................
3-3. Installation in Model 192
.......................
3-4.
Performance Verification.
.......................
3-6. Kecommended Test Equipment.
......................
3-7.
Environmental Conditions.
.......................
3-9.
Performance Verification Procedure.
..................
3-11.
Initial Conditions.
..........................
3-13. Calibration
..............................
3-15. Kecommended Test Equipment.
......................
3-17. Environmental Conditions.
.......................
X-19.
Calibration Procedure
.........................
3-22. Troubleshooting
............................
3-24. Troubleshootiny Procedure
.......................
TtiEURY OF UPEKATIUN
.............................
REPLACEABLE PAKTS
..............................
5-l.
General
................................
5-3. Ordering Information.
.........................
5-5.
Factory Service
............................
5-7.
Schematic and Component Layout.
....................
Page
l-l l-l l-l l-l l-l l-l
2-1 2-l 2-1
3-l 3-l 3-2
3-3 3-3 3-3 3-3 3-4 3-6
3-6
3-G
3-6
3-9
3-9
4-l
5-l 5-l
5-l 5-i 5-l
Model 1920
General Information
SECTION 1. GENERAL INFORMATION
l-l. INTRODUCTION
l-2.
The Keithley Model
1920
is a True Root Mean Square (TKMS) AC plug-in option for the
Models
192
and
191.
The 1920 enables the 191 or 192 to measure the TKMS value of an AC
siynal.
When the 1920 is installed in the Model
192
an ACtUC function is available. The
Model
1920
has four ranges on which an AC siynal can be measured.
It
is field
installable.
1-3.
Warranty Information
l-4.
The warranty is yiven on the inside front cover of this manual. If there is a need
to exercise the Warranty,
contact the Keithley Representative in your area to determine
the proper action to be taken.
Keithley maintains complete repair and calibration
facilities in the United States, West tiermany,
tireat Britain, France, the Netherlands,
Switzerland and Austria.
Information concerning the application, operation or service of
your instrument may be directed to the applications engineer at any of the above
locations.
Check the inside front cover of this manual for addresses.
l-5. Manual Addendums
1-G.
Improvements or changes to the instrument that occur after printing of the [manual
will be explained on an addendum which will be attached to the inside back cover.
1-7.
Safety Symbols and Terms
l-8. The symbol
A
on the instrument denotes that the user should refer to the operating
instructions.
t
The symbol
on the instrument denotes that up to 5UllV may be present on the terminals.
The WARNING used in this manual explains dangers that could result in personal injury
or death.
The CAUTION used in this manual explains hazards that could damage the instrument.
l-9.
Specifications
l-10.
Uetailed specifications for the Model 1920 are given in Table
l-l.
l-l
General Information
Model 1920
TABLE 1-l
1920 Specifications
5-l/2 Digit Accuracy, +(% Rdg + Counts),
1
yr 18-28'C
5-l/2 Digit
20-50Hz"
DC, 50Hz-lOkHz* lOK-20kHz"
20kHz-lOOkHz**
Range
RESOLUTION
2v 1OPV 1% t
100
0.25% t 100 0.35% + 300
1% t 500
2ov lOOld
1% t 100
0.25% t 100 0.35% + 300
1% t 500
2oov 1mV 1% t
100
0.25% + 100 0.35% + 300 1% t 500
7oov 1OmV
1% t 100
0.354: t 100 0.5% + 300
1% t 500
RESPONSE: CREST FACTOR AC + DC MODE: riAxIriuri INPUT:
SETTLING TIME: INPUT IMPEDANCE:
TEMP. COEFFICIENT:
3dB BANTMIDTH:
CFIRR:
TRUE RFlS
Add 60 counts to specified accuracy (not wailable in We1
191)
1OOOV Peak AC + DC, 2 x lo7 V* Hz
0.5 set to within 0.1% of change in reading 211 ohm shunted by less than 5OpF.
(O"-18OC and 28"-5O'C): Less than
0.1
x applicable accuracy
specification per "C below 50kHz; 0.2 x for 50kHz to 1OOkHz 500kHz Typical Greater than 60dB at 50 and 60Hz
(lk
ohm unbalanced)
* Above 2000 counts
** Above 20000 counts; 3% + 500 typical below 20000
l-2
Model 1920
SECTION 2. OPERATING INSTRUCTIONS
Operating Instructions
Z-l. Operating Instructions for the rlcdel 1920 installed in the ilodel 191 are as follows:
2-2.
Ilith the ilodel 1920 option, the ilcdel 191
reads AC voltages froln 10 r;licrovolts/digit
to 1000v.
The instrument displays the True Koot llean Square (TKilS) value of an AC
voltage.
It has a frequency response of 20Hz to 100ktlz.
2-3.
The rnaximm reading is 199YYY. Werrange is irdicatcd by (-) l-----, except on
the 1000 volt range.
The 700 volt range is selected idith the 1000 volt button. On the
700 volt range, the display can read beyond the rnaxiinuln allobiable input voltage. ilaximun
allobcable input: 1OOOV Peak AC + DC; 2 x lo7 V'Hz. Use the ilodel 191 to lneasure AC
voltage as follobis:
CAUTION
Do not exceed maximum allohcable input voltage. Instrument damage may occur.
A. Turn the pokier on IJith the UN/OFF pushbutton,
ati depress the ACV pushbutton.
8.
Select the desired range from the 4 ranges available.
The decimal point is positioned
by the range pushbutton.
C.
Ensure that the I0JLL pushbutton is out (light off) unless measurements are to be made
as deviations from a preset value.
NOTE
Do not use NULL to zero the range.
A small residual zero readirq is normal. If NULL is used to zero this offset, readings in specified accuracy range will be low by the offset amount.
This is because rms measurement signals add in a root mean square fashion arrl not linearly.
The NULL function is a linear offset axl not usable in TKllS
measurements.
0.
Connect the signal to be measured between the INPUT HI ati LO binjiny posts. The birding posts accept wires, spade lugs or banana plugs for ease of connectiq the circuit to be measured.
Observe the displayed digits ard decimal point.
2-1
Operating Instructions
Model 1920
E.
For specified accuracy (pulse widths ,lOns, peak voltage (1.5 x full scale) Figure 2-1 shows the allowable input signal vs. crest factor.
The figure illustrates that for a
crest factor of three the displayed reading must be less than 100,000 counts (1V on
the 2V range).
So long as the maximum input is not exceeded no damage will result in crest factors exceeding that shown in Figure2-I, but accuracy will slowly degrade (For CF >3 but<10 typical accuracy is degraded by (CF-3) x 0.36% arti peak signal must be less than 5 x full scale for that range).
CF
I
I
0
100000
200000
Displayed Counts
Figure 2-l.
Crest Factor VS. Displayed Counts
2-4. Operating Instructions for the Model lY20 installed in the Model 192 are as
follows:
2-5.
With the Model 1920 option, the Model
192
reads AC voltages from 1 microvolt/digit
to 1ooov.
The instrument displays the True Koot Mean Square (TKMS) value of an AC voltage
signal.
2-G.
The maximum reading is 199999.
Overrange is indicated by OFLO. Maximum allowable
input is 1OOOV Peak AC f DC; 2 x 107V.Hz. Use the Model
192
to measure AC or AC + DC
voltages as follows:
CAUTION
Do not exceed maximum allowable input voltage. Instrument darnage may occur.
A. TURN POWER UN with UN/OFF pushbutton.
B.
Press the ACV button once to select the ACV (TKMS) function. Press the ACV button ayain and the 192 will measure the T'KMS value of an AC + UC signal (AC signal superimposed with a DC level). The ACV and DCV LEDs will light simultaneously when the AC + UC function is enabled.
Pressing the ACV button again will return the 192 to
the ACV function. Pressiny the DC button while in AC t DC returns the
192
the DCV
function.
2-2
UPEKATINti INSTKUCTIUNS
NOTE
IWDEL 1YZU
The 14odel 192 will display "NO AC" if ACV is selected without the 1920 or 1910 ACV option.
The AC + UC function is available only with C-4
and dbove software and Kev U and above Analoy Board.
C. SELECT KANtiE from the four ranyes available. The decimal point is positioned by the
ranye pushbutton. The 1UOUV ranye is selected with the 2000 button. If the ZUIQ button is inadvertently pressed when in ACV function,
the 192 will set the ranye to
1uuo.
Il. ZEKU OFF unless [measurements are to be made as deviations frown a preset value.
NOTE
Uo not use ZEKU to zero the ranye. A small residual zero readiny is normal. If LEKU is used, the residual voltaye readiny in specified accuracy ranye will be low by the amplitude of the zeroed residual voltaye.
This is because rms measurement signals add in a root mean square fashion and NUT linearly. THe LEKU function is a linear offset and not usable in TKMS measurelnents.
E.
CONNECT IllPUT to be measured between the HI HCV and LO bindiny posts.
F.
TAKE KEAUING.
ti. For specified accuracy (pulse widths LlUus, peak voltage (1.5 x full scale) Figure Z-l
shows the allowable input siynal vs. crest factor.
The figure illustrates that for d crest factor of three the displayed readiny (must be less than lUU,UUU counts (1V on the ZV ranye).
So long as the maximum input is not exceeded no damage will result in
crest factors exceediny that shown in Fiyure 2-1,
but accuracy will slowly degrade
(For CF >3 but 510 typical accuracy is degraded by (CF-3) x U.3GX and peak signal must
be less than 5 x full scale for that range).
Model 1920
SECTION 3. MAINTENANCE
Maintenance
3-l. The Model 1920 is a plug-in option that can be installed in your Model 191 or 192
Instdllation of the 1920 into your instrument is as follows.
WARNING
All service information is intended for qualified electronic maintenance
personnel only.
3-2. INSTALLATION: (For use with Model
191)
A. Disconnect the line power cord of the Model
191
and remove all test leads from its
input terminals for safety.
B. Turn the DMM bottom side up and loosen the four screws in the bottom cover.
C.
Hold the top and bottom cover together to prevent their separation and turn the DMM over to the normal position. Remove the top cover.
D. Remove the RF shield by grasping it and FIRMLY lifting it away from the motherboard.
E. Carefully insert the connector of the
1920
into its mating receptacle on the 191
motherboard. Take note of the orientation of the
1920
into the 191 motherboard as
shown in the figure below. Connect the brown wire of the
1920
to Pin 7 of the DCV
switch, and connect the blue wire to Pin 9 of the ACV switch.
F. Replace the RF shield being careful not to pinch any wires.
G. Calibrate the
1920
per the procedure located on Table 3-5.
H.
Replace the top cover.
3-l
Maintenance
Model 1920
NOTE
Pin locations for the switches are shown on the decal on top of the RF
shield.
WARNING
All service information is intended for qualified electronic maintenance personnel only.
3-3.
INSTALLATION: (For use with Model 192)
A. Uisconnect the line power cord of the Model 192 and remove all test leads from its
input terminals for safety.
B.
Remove two screws that hold the top cover to the rear panel and remove the cover.
C.
Remove the top shield from the Analog Board by grasping it and FIRMLY lifting it off, with a prying motion of the four retaining clips.
u. Carefully insert the connector of the 1920 into its mating receptacle on the Analog
Uoard.
Take note of the orientation of the 1920 onto the 192 Analog Hoard as shown in
the figure below.
3-2
Model 1920
Maintenance
E.
Connect the brown wire of the 1920 to P1014 on the Analog Board as shown in the figure
below.
Connect the blue wire to PlOlG which is located on the ACV HI input terminal.
This is also shown in the figure below.
F.
Replace the shield to the Analoy Board beiny careful not to pinch any wires when engaying the shield into the retaining clips.
ti.
Calibrate the lY20 per the procedure located on Table 3-4.
3-3
Maintenance
Model
1920
3-4. PERFORMANCE VERIFICATION
3-5.
Performance verification may be performed upon receipt of the instrument to ensure
that no damaye or misadjustment has occurred duriny transit.
NOTE
For instruments that are still under warranty (less than 12 months since date of shipment), if the instrument's performance falls outside specifications at any point, contact your Keithley representative or the factory immediately.
3-6. RECOMMENDED TEST EQUIPMENT
3-7. Kecommended test equipment for performance verification is listed in Table 3-l. Alternate test equipment may be used.
However, if the accuracy of the alternate test equipment is not at least 4 times better than the instrument specifications, additional allowance must be made in the readinys obtained.
ITEM
TABLE 3-l
Recommended Test Equipment For Performance Verification
UESCKIPTIUN
SPECIFICATION
MFK,
I
O.lV, lV,
lOV, 1oov
H-P
+O.WzX
1OOOV
H-P
+0.04%
AC Calibrator
High Voltage Amplifier
(Used with Model 745A)
UC Calibrator
I
1v +0.002x
Flukl
746A
343A
3-7.
ENVIRONMENTAL CONDITIONS
3-8.
All measurements should be made at an ambient temperature within the range of 18" to
28'C (GS" to 82'F), and a relative humidity of less than 8o'k.
3-9. PERFORMANCE VERIFICATION PROCEDURE
3-10.
Use the following procedure to verify the basic accuracy for AC voltage in the
Model
191
or the Model 192.
3-4
Model 1920
Maintenance
WARNING
Performance verification should be performed by qualified personnel
using accurate and reliable test equipment.
3-11. INITIAL CONDITIONS
3-12. Before beginning the verification procedure,
the instrument must meet the following
conditions:
A. If the instrument has been subject to extremes of temperature, allow sufficient time
for internal temperatures to reach environmental conditions specified (1X0-28'C). Typically, it takes one hour to stabilize a unit that is 10°C (1U'F) out of the specified temperature range.
B. Turn on the instrument and allow it to warm up for two hours before using it with the
Model 192 and one hour before using it with the Model 1Yl.
WARNING
Some procedures require the use of high voltage. Take care to prevent contact with live circuits which could cause electrical shock resulting in injury or death.
3-5
Maintenance
Model 1920
Performance Verification
2V 2UV zoov
7oov
2V ZUV
zoov
2V NV 2UUV
7oov
ZUV
zov
ZV
2V
APPLIED INPUT
AT
1kHz
l.UUUOUV 1o.ouoov lUO.UUUV
7oo.oov
AT ZUkHz
l.UUUUOV lU.UUUUV lUU.UUUV
AT 1UUkHz
l.UUUUUV lU.UUUUV lUU.UOUV
AT 3UkHz
700. oov
AT ZOHz
1o.uuuuv
AT 50Hz
10. ououv
AC + UC Mode
+1.uuouov UC
-1.uuuuuv UC
T
ALLUWAfJLf
For 191
.9965UV to 1.0035UV
9.965OV to 10.035UV
99.65UV to 100.35UV
696.55V to 703.45V
.9935UV to l.UU65OV
9.Y35UV to lU.UG5UV
99.35UV to 100.65UV
.985UUV to 1.01500v
9.x5uuv to 10.1500v Y8.5UOV to 101.5uuv
688.UOV to 712.oov
9.89UUV to 10.1100v
9.89UUV to 10.1100v
L
:EAUIIVGS
For
192
.Y9G5UV to 1.0035OV
9.965OV to 10.035UV
99.GWV to lOU.35OV
696.55V to 703.45V
.9935OV to l.UU65UV
9.935011 to lU.UG5OV
99.35UV to lUU.65OV
.985UUV to 1.015uuv
9.85OUV to lU.15OOV 9x.5uuv to lU1.5OOV
688.OOV to 712.OOV
Y.89OUV to 10.1100v
9.89UOV to 10.1100v
.995YUV to 1.ou41uv
-.9959uv to -1.uu410
3-6
Model 1920
Maintenance
3-13.
CALIBRATION
3-14.
This section contains information necessary to calibrate the Model 1920 in your
instrument (191 or lY2). Calibration should be performed when any of the followiny
conditions occur:
A. Annually B. Installation of the 1920
C. Performance Verification indicates 1920 as out of specification
3-15. RECOMMENDED TEST EQUIPMENT
3-16. Recommended test equipment for calibration is listed in the following table.
Alternate equipment may be used. However, the accuracy of the alternate equipment must be
at least 4 times better than the Model 1920 specifications,
or equal to the specifications
listed in the Table below.
TABLE 3-3
Recommended Test Equipment For Calibration
ITEM
UESCRIPTION
SPECIFICATION MFK. MOUEL
-
A DC Calibrator
+1v
Fluke
343A
r.uw'/
B
AC Calibrator
U.lV,
lV, lOV, 1uuv
H-P 745A
Hl.U22%
C High Voltage Amplifier
IUOUV H-P 746A
(Used with Model 74SA)
+0.04x
II
Calibration Cover
--
Keithley 1913 (191)
3-17. ENVIRONMENTAL CONDITIONS
3-18. Calibration should be performed under laboratory conditions having an ambient
temperature of 23
_+l”C,
and a relative humidity of less than 70%. If the instrument has been subjected to temperatures outside of this range, or the higher humidity, allow two hours minimum for the instrument to stabilize at the specified environmental conditions before beginning the calibration procedure.
3-19. CALIBRATION PROCEDURE
3-20.
Remove the top cover of your instrument and replace it with the appropriate
calibration cover.
With the calibration cover in place allow the internal temperature of
the instrument to stabilize for two hours before performing the calibration.
3-7
Maintenance
WARNING
Model 1920
3-8
STEP
1
2
*AUJU
3
4
5 6
Some procedures require the use of high voltage. Take care to prevent contact with live circuits which could cause electrical shock
resulting in injury or death.
R42G
R406 C412
Model 1920 Calibration Adjustments
TABLE 3-4
Model 1920 Installed in the 192
Calibration Procedure
KANtiE
FWCTIUN
APPLIEU INPUT
AUJIJSTMENT
KEAUINti
2
ACV + KY
1.lllllloov UC
*Calibrator
I.UOOOO ?lO counts
2
ACV + UCV
-1.UOUUUV UC
*K411
l.UOOUU QO counts
I
I
KNT: Adjust the UC calibrlator until the display reads 1.00000 +I0 digits.
leads and take note of the reading.
Keverse th
Adjust K411 to l/Z the difference (Example:
Keadiny is l.U04lJU, adjust K411 to read 1.00200).
?peat Steps and 2 until there is < 10 digits
change in reading from + to -.
ACV
l.UlxlUUV AC
at SUUtir
ACV
.llJlJUlJV AC
INepeat Steps 4 and 5 until they
K407
1.00000 ?lO counts
K4Z6
.lOOOlJ i-10 collnts
are within tolerance.
Wodel
1920
Maintenance
c
I;iP
7
20
c, mu
Y I 2lJlJu
( 700V AC Max)
10 2UU
11
2
12
20
ACV ACV
5UU.UUV AC IUU.UUUV AC at
1Wktiz
/\cv
ACV
l.UUUUUV AC
at 1UUkHr
lU.UUUUV AC
I
I
I
at 1UUktir
13
Kepeat Steps lU-12 until no chanye.
14
Kepeat Steps l-l:! to verify the readings.
NOTE
T
TABLE 3-4
Model 1920 Installed in the 192
Calibration Procedure (cont.)
FUNCTION APPLIED INPUl AUJUSTMENl
ACV
ACV
1o.oLwLlv HC
at SUUtir
lUU.UUUV AC
at 5UUtlr
K4U0
K405
K404
C413 Caution: Use an insulated aliynlnent tool
c4u2
C4U7
KEAUINti
5LlO.UU r10 counts
lOO.llU +25u counts
l.OUOU *50 counts
lU.ULlUU ~100 counts
tliyh frequency (100kHr) measurements are sensitive to component location. Uo nut move or bend the components in the input area (C401, iKWJ, etc.) Kecalibration is necessary if these components are moved.
If the 1uuv
1UUkHr cannot be brouytit into specification,
refer to the troubleshootiny
section.
WARNING
Some procedures require the use of high voltage. Use an insulated alignment tool.
Take care to prevent contact with live circuits which could cause
electrical shock resulting in injury or death.
3-9
Maintenance
Model 1920
TABLE 3-5
Model 1920 Installed in the 191
Calibration Procedure
STEP
KANtiE
FWCTIUI~
I\PPLIEU INPUT
AUJUSTMENT
KEAUING
1
2 ACV
1UUmV AC
K411 Minimum Reading
at 50011r
2 2 ACV
1UmV AC
K411
Minimum Reading
at SUUHr
3 2 ACV
l.UUUUUV AC
K4U7 l.UUOUU *lo counts
at SUUHZ
4 2 ACV .lUUUUV AC 11426
.lUUUU *lo counts
at 5UUHr
:, IRepeat Steps 3 dnd 4 until they are within specification.
ti
20
ACV
11).uouuv AC
K4UG
10.0000 *lo counts
at 5UUHz
7 2uu ACV
IUU.UUUV AC
I<405
100.000 ?lO counts
at SUUHZ
8
zuuu
ACV
suu.ouv AC
11404 5u0.00 +10 counts
( 7UOb AC Max)
9,
LOU
ACV
lUU.UUUV AC
c413
100.00 i25ll counts
at 1lJUkHr
Caution: Use an insulated align­ment tool.
10
ii
ACV
l.UUUUUV AC
c4uil 1.0000 +50 counts
at 1UUkHz
11 20 ACV lU.UUUUV AC c4u7
10.0000 *loo counts
at 1OUkHr
12 Hepeat Steps 9 thru 11 until no change.
13 Kepeat Steps 1 thru 11 to verify the readings.
NOTE
Hiyh frequency (100kHz) measurements are sensitive to component location. Do not move or bend the components in the input area (C401, K403, etc.) Recalibration is necessary if these components are inoved.
If the 1OUV
1UUkHz cannot be bt-ouyht into specification,
refer to the troubleshooting
section.
3-22.
Troubleshooting
3-23. The troubleshootiny instructions contained in this section are intended for qualified personnel having a basic understanding of analog and digital electronic principles and components used in precision electronic test equipment. Instructions have been written to assist in isolating the defective circuit or subcircuit.
Isolation of the
specific defective component has been left to the technician.
3-10
Model 1920
Maintenance
NOTE
For 1920's that are still under warranty (less than 12 months since date of shipment), if the 1920's performance is outside of specifications at any point,
contact your Keithley representative
or the factory before attempting troubleshootiny or repair.
3-24. Troubleshooting Procedure
3-25. Table lists step by step checks of the lnajor circuit blocks of the 1920. Follow Table 3-G to locate the trouble.
NOTE
Performance Verification of the Model 1920 is necessary if any of the following occurs.
1)
Kemoval /rep1 acement of 1920
2) Removal/replacement of shields on the 1920
WARNING
Some procedures require the use of High Voltage. Take care to prevent contact with live circuits which could cause electrical shock resulting in
injury or death.
TABLE 3-6
Model 1920 Troubleshooting (cont.)
STEP
1
7 8
9
ITEWCOf1PONENT
REQUIKEO CONDITIOll KEHAKKS
J1006,
Pin 1 U401, Pin 7 U401, Pin 4
U403, Pin 8
(Brown wire 51008)
U401, Pin G
Q408, Pins 2
anJ. 5
Q408, Pin 3
Turn on Power. Select the AC function ard 2V range.
t5V <lOOnV AC noise t15V *lV DC <lOOnV AC noise
-15V +lV DC <lOOnV AC noise Apply 1 volt at 1ktlZ
1 volt DC
+5v supply t15v supply
-15v supply
1920 output
1 Volt AC at IkHZ, no DC Offset
u401 output
+3V to t8V DC, within
1OmV
Differential Output
of each other
of Q408
UV t 25mV
Input summifq junction
Maintenance
Model 1920
-
TABLE 3-6
Model 1920 Troubleshooting (cont.)
STEP
10
11
12
13
14
ITEM/COMPONENT
REQUIRED CONDITION
Select 2ov range a,j apply 1OV AC at 1kHZ
Select 2oov ranye ard apply 100 V AC at 1kHZ
Select 700V range ard apply 350V AC at 1kHZ
Select AC + DC function ard the 2V rarqe. Apply 1v DC Select AC apply IV DC
KEIlAKKS
(virtual ground)
If display overranges check (1406 an3 Q407 for gate drive ard signal
If
display met-ranges check Q404 ard Q405 for gate drive anl signal
If display wet-ranges check Q402 arrl Q403 for gate drive drd signal Check display for 1V
if not, check (1401 an3
K401 Oisplay approx. OV
if not, check Q401
K401, c401
NOTE
If the 1OOV 1OUkHz adjustment cannot be brought into specification,
check the spaciny between C4Ul and K403. If these components have
been bent toward or away from each other, C413 may not have adequate
calibration ranye.
Moviny C401 and K403 closer toyether increases the
1OOkHz response on all ranges.
Check and adjust, if necessary.
3-12
Model 1920
Theory of Operation
SECTION 4. MODEL 1920 THEORY OF OPERATION
4-l. The Model 1920 is a plug in AC(+UC) to RMS converter with variable yain. The gain factor conditions the AC input voltaye for application to the A/U converter.
4-2. The input siynal is applied through GIN (0.05uF) and KIN (2Mrl) to 4408. For
AC+UC operation GIN is shorted by Relay K4Ul.
The two stage amplifier (op amp) is a variable yain inverting type with gain from-U.001 to-l (see table below). The gain is selected accordingly by the double FET switches Q402-Q407. Two FETs are used for each range.
They are configured in a T-type attenuator with the 1Uk resistors K413, K414 and
K41G to effectively attenuate high frequency (1UUkHz) signals.
4-3.
The op amp is a two stage amplifier. The first stage is confiyured around dual FET
Q408. 4408 allows the reduction of input bias current.
This reduction is necessary because an auto zero cycle is not possible on the AC or AC+UC functions. Any input bias current will show up as an input offset voltage.
The second stage is configured around u401. The output of U4Ul is applied to the KMS converter U403. C409, C41U and K42U make up the two pole filter that is located at the output of U403. The output of U403 is a UC signal which is applied to the A/U converter.
A small UC offset may be applied to the
auto zero A/D input by adjusting R426.
This allows compensation for the KMS converter's
output offset.
TABLE 4-l
Ranging Information
Ranye tiain
UP Amp Feedback Resistance Engergized FETs
2v 1
None
2ov
l/10 (217k f"K4U6) 11 2M
Q406, q4u7
2ouv
l/100 (19.6k + K405)1) 2M
Q404, Q405
7UOV
l/lOUU
(2.12k + K404)I)
Q402, Q403,
(19.GK + R405)1/ 2M
Q404, Q4U5
.I
4-1
Model 1920 Biock Diagram
FET
Orive
-
-
Range Select
Model 1920
Repldieabl
e Parts
SECTION 5.
REPLACEABLE PARTS
5-l.
General
5-2. This section contains information for ordering replacement parts. The replaceable
parts list is arranged in alphabetical order of the circuit designations of the components.
A crass reference list of manufacturers containing their addresses is
given in Table 5-1.
5-3.
Ordering Information
5-4.
To place an order or to obtain information concerning replacement parts contact your Keithley representative or the factory.
See the inside front cover for addresses.
When ordering,
include the following information:
a. Instrument Model Number b.
Instrument Serial Number c. Part Uescription d.
Circuit Designation (if applicable)
e. Keithley Part Number
5-5. Factory Service
5-6. If the instrument is to be returned to the factory for service, please complete the
Service Form which follows this section and return it with the instrument.
5-7. Schematic and Component Layout
5-B. The Schematic and Component Layout follow this section.
TABLE 5-l
Cross Reference of Manufacturers
MFti Code
A-U
NAME AND AUURESS
Analog Uevices, Inc. Norwood, MA 02026
FEUEKAL SUPPLY COUE
01121
BKN
Bourns, Inc.
Kiverside, CA 92507
80294
5-l
Replaceable Parts
Model 1920
MFti Code
c-u
CLU
CLK
ULE
EC1
EFJ
EKI
INT
K-I
NAT
N I c
NYT
MEP
TABLE 5-l
Cross Reference of Manufacturers (cont.)
INAME ANU AUUKESS
Cornell-Uubilier Newark, NJ 07101
Centralab Uivision
Milwaukee, WI 532021
Clarostat Manufacturing Co. Uover, NH 03820
Uale Electronics Columbus, NE 68601
Electra-Cube, Inc. San Gabriel, CA
91776
E. F. Johnson Co.
Waseca, MN 56093
Erie Technological Products Erie, PA
16512
Intersil,
Inc.
Cupertino, CA
95014
Keithley Instruments, Inc. Cleveland, Ohio 44139
National Semi Corp. Santa Clara, CA 95051
Nichicon Corp. Chicago, IL 60645
Nytronics Components Group Darlington, SC 29532
Mepco, Inc. Morristown,
NJ 07960
FEUERAL SUPPLY CODE
-
1
14655
71590
12697
91637
14752
74979
72982
32294
80164
27014
83125
80031
5-2
Model 1920 Replaceable Parts
TABLE 5-1
Cross Reference of Manufacturers (cont.)
MFG Code
PKP
NAME ANU AUUKESS
Precision Kesistive Products Mediapolis, IA 53237
FEUERAL SUPPLY COUE
SIL
Siliconix, Inc. Santa Clara, CA 95054
17856
STD
Standard Condenser Chicago, IL
Texas Instruments,
Inc.
Uallas, TX 75231
97419
T-I
01295
5-3
Replaceable Parts
Model 1920
TABLE 5-2
Replaceable Parts List (cont.)
Circuit Schematic PC-Board Mfr. Mfr.
Keithley
Lksig. Description
Location Item No./Location Code Lksig.
Part No.
c401 .05,#, lUUUV, Metal Poly A4 5/EZ STU CAPACPLYEEOX
C-285-.05
c402 .25pF to 1.5pF, ZOOUV Trimmer 23 6/E2
EFJ 273-0001-002 31863A
c403
IpF, XV, Alum Elect tI3 7/u
NIC SOVKBID c-325-1.0
c404 luF, 16V, Alum Elect 04 8/E2
NIC SIJVKBIU c-325-1.0
c4u5 33OpF, 5OUV, Poly
Cl Y/D2 CLtJ CPK330J C-138-33UpF
C406 3llpF. 5OUV, Mica
c2 IO/D2 C-O
UClOEUUOJ3 C-236-30pF
c407 .25pF to 1.5pF, 2UUUV,
c3
11/w
K-l 273-0001-002 3186311
Trimmer
c4w luF, 5UV, Metal Poly F4 12/c3
EC1 625U
c-335-1.0
C4OY
l,,F, SUV, Metal Poly FS 13/c3
EC1 6258 c-335-1.0
c41u
l"F, 5UV, Metal Poly F5
14/C3 EC1 62513
c-335-1.0
c411 33pF, 5UlJV, Mica 04
15/w C-D
UClUEU3OUJ3 C236-33pF
c412 G.BpF, SOV, Tube Cer
lUU/W ERI
301.oooco c-282-6.8pF HO15
c413 .25pF tu l.SpF, XJOOV, u4
101/u K-I 273.OOUl-002
3186311
Trimmer
102/E2
CKL KlCC2U
C-77-0.75pF
IN914
RF-28
IN914
RF-28
IN914
RF-28
KL-69
5-4
Model 1920
Replaceable Parts
Circuit lksig.
L401 L4U2
q401 q402 q403 4404 q405 4406 4407
441)s
R401 R4O2 R403 R404 R405 R406 R407 R408 R409 K410 R411
Description
lUU,,H Choke
lOU$l Choke
Transistor, PNP N-Channel JFET (selected) N-Channel JFET N-Channel JFET (selected) N-Channel JFET N-Channel JFET (selected) N-Channel JFET Uual N-Channel JFET
lk, 1%. 1/2W, rntf lk, I%, i/ZW, lntf 2M,
.5x,
lW, mtf Pot, lUUl1, lor, 3/4w Pot, lk,
lo’b,
J/4W
Pot, 1Uk. lOY, 3/4W
Pot, lk, lo%, 3/4W
2.12k,
.Yb,
1/8W, intf
19.6k, .a, l/IIW, mtf
217k,
.5x,
1/8W,
mtf
Pot, lOk, lo'b, 3/4W
TABLE 5-2
laceable Parts List (cont.)
thematic ocati on
A3
A3
nz lx E2 u2
E2
II3
E3
c4
84 04 B4 CZ C2 C3
F5 c2 CZ C3 05
17/w
l&l/U3
34lE3
35/U2 36102 37/w 38/U2 39/w 4U/U2 41/E2
57/F2 58!FZ 5Y/E2 (IU/UZ 61/02 62/W 63/1)2 64102 65lU2 a/u2 67lE2
I
PC-Board
,tem No./Location
Mfr. Code
NYT NYT
A-U
K-I
INT
K-l
INT
K-l
INT
SIL
ULE
ULE
ULE BRN BRN BRN BHN ULE ULE OLE URN
Mfr.
Desig.
swu-100 swu-1uu
AU-820
lTE43Y2
lTE4392
ITE4392
E411
MFF-l/2 MFF-l/2 MFF-l/2-31 36UU, IiPlUU
3600, GPlOUU
36OO, GPlOUUU
3600,
tiPlOO
MFF
l/U
MFF-l/8 MFF-l/8
3600, GPlOUOO
T
Keithley Part No.
CH-14
CM-14
Tti-84 TG-128 Tti-77
Tti-128
Tti-77
FIG-12X
TG-77 Tti-118
K-94-lk K-Y4-lk K-303-21.1 IRP-XY-100 KP-XY-lk RP-&l9-1Ok KP-89-lk R-246-2.12k R-246-19.6k
K-246-217k
RP-89-1Ok
5-5
Reolaceable Parts
Model 19'20
TABLE 5-2
Replaceable Parts List (cant
Circuit Desig.
R412 R413 R414
11415 K416 K417 K418 11419
R420 K421
R422
11423 K424 R425 K426 K427 K428 R42Y lK430 K431 H432 K433 K434
Iescription LM, 5X, 1/4W, Carb LUk, 5'h, 1/4W, Carb IOk, 5%, 1/4W, Carb LM, 5%. 1/4W, Carb lOk, Y/S, 1/4W, Garb lM, 5X, 1/4W Carb ZM, .b%, 1/8W, i&f j.65k, lY, l/ilW, mtf 24k, 5X, 1/4W, Carb IOk, I%, 1/8W, mtf
2.20, 5X, 1/4W, Carb Z.Zk, 5%. 1/4W, Carb J.3k, 5X, 1/4W, Carb
2.2k,
5y,
1/4W, Cat-b
Pot, lOk, lo%, 1/2W.
2.2k, 5$, 1/4W. Garb 49912, la, l/&l, mtf
SG.Zk, .5x,
1/8W,
mtf
56.2k, .5%, l/SW, mtf
44.2k, I%, 1/8W, mtf lOk, 5X, 1/4W, Carb
16.9n, 1X, l/&J, mtf 33k, 5X, 1/4W, Carb
ichematic .ocation
U2 UZ II3 U3 u3 U3 c4 04
F5 F2 GZ F4 G4
A2
F2 F5 E5 c4 C4 c5 c5 c5 u4
t-
PC-Board
tern No./Location
68/C2 69/U 7o/c3 7l/C3 72ic3 73/a 74/E2 75/U2 76JC3 77/u 78/C?!
7Y/CZ UUICZ Xl/E2 82/C2 83/U3 84/w 85/E2 86/U2 87/E2 @S/E2 89/U YO/OZ
Mfr. Code
MEP MEP MEP MEP MEP MEP ULE PRP MEP
PRP MEP MEP
MCP MEP
URN
MEP
PRP
ULE
ULE
PKP
MEP
PKP
MEP
Yfr. Jesig.
X25* CK25* CK25* CK25* CK25* CR25* YFF-l/II
**
CR25* **
CK25* CR25* CR25* CRZS*
3386H-1-103
UK25* **
MFF-l/8 MFF-l/8
**
cn25*
**
CR25*
Keithley Part No.
R-76.1M IK-76-1Ok IK-76-10k K-76.1M H-76-1UK K-76-1M IR-246-2M K-88-3.65k IR-76.24k IK-88.1Ok R-76-2.2
R-76.2.2k IK-76-3.3k
R-76-2.2k
KP-111.1Ok K-76-2.2k R-88-499 K-246-56.2k K-246-56.2k R-88-44.2k R-76-10k K-88-16.9 R-76-33k
CK25* Manufacturers Uesignation includes parts description e.g. CK25 33k, 5'6, 1/4W, Carb for 11434 **
Manufacturers &sign&ion is GP1/4, 1X, 7100, Kesistance value
5-6
Model 1920
Replaceable Parts
I
Circuit
Uesiq.
u401 u4w u4u3
2P
RI
VOI tage Quad comparator
1
t
TABLE 5-2
laceable Parts List (cont.)
khematic aation
u4 I73,4
E5
41/u
1:
4Y/U3
49/u3
Mfr. Code
NAT
Mfr. Desig.
LF356H LM339 AUS36AK
5-7
KEITHLEY INSTRUMENTS, INC.
28775 AURORA ROAD
CLE'.'ELAND, OHIO 44139
SERViCE FORWl
MODEL NO. SERIAL ?lO,~
P.O. NO. DATE R-
_. ,--- _..~___
1
NAME
PHONE COMPANY ADDRESS
CITY
STATE
ZIP
I3 :
Describe problem and symptoms using quantitative data whenever possible (enclose readings, chart recordings, etc.)
(Attach additional sheets as necessary).
El
Show a block diagram of your measurement system including all instruments connected
(whether power is turned on or not). Also describe signal source.
q
the instrument.
List the positions of alJ controls and switches on both front and rear panels of
El
Describe input signal source levels, frequencies, etc.
q .
List and describe all cables used in the experiment (length, shielding, etc.).
q
for each.
List and describe all other equipment used in the experiment. Give control settings
El
Environment:
Where is the measurement being performed?
(Factory, controlled laboratory, out-of-doors, etc.) What power line volt=is used?
Variation? Frequency?
Ambient temperature?
"F. Variation?
"F. Rel. Humidity?
Other
lxl
Additional Information. (If special modifications have been made by the user, please describe below.)
REV 0774
Loading...