1.1 INTRODUCTION
The Keithley Model 181 is a highly sensitive nanovoltmeter
with a large, easy to read 5 K or 6 ‘/ digit display. The Model
181 is unique in that it combines microprocessor technology
with a new concept in low-noise, high-gain front ends,
resulting in a programmable instrument with sensitivity
down to 10nV. The Model 181 provides highly accurate,
stable, low-noise readings on seven ranges for DC voltage
measurements between 1OnV and 1OOOV. The mV ranges
use a special low-thermal input connector, while connections for the higher voltage ranges are made through two
S-way binding posts. Additional versatility is afforded by the
inclusion of an IEEE-488 interface which allows the unit to
communicate with other instrumentation.
1.2 MODEL 181 FEATURES
The Model 181 includes the following features:
l
High Sensitivity. The resolution of the Model 181 on the
2mV range is IO ~* volts (10nV).
l
5% or 6% Digit Resolution. Normal 5% digit display
resolution may be increased to 6 % digits at the touch of a
button.
l
IEEE-488 Interface. A built in IEEE-488 interface allows
the instrument to communicate with other devices such
as a central controller or printer.
l
Analog Output. An analog output, which accurately
reflects the displayed readings, is available from the rear
pallel.
l
3-p& Digital Filter. The internal 3-p& filter minimizes the
effects of noise in voltage readings and may be controlled
from the front panel or IEEE bus.
l
Separate Inputs. A special input connector is used for the
mV ranges to minimize thermal emf generation.
l
Isolated Low Terminals. The low signal connections for
both inputs are isolated from power line ground and
from IEEE low to minimize ground loop problems.
l
Color Coded Front Panel. Inputs, range switches, and
other front panel controls are marked to form color-coded
groups for easier operation.
1.3 OPTIONAL ACCESSORIES
1
Model 1483 Low-Thermal Connection Kit. The Model
1483 kit contains a crimp tool, pure copper lugs, Lowe
thermal cadmium solder, copper alligator clips, and
assorted hardware. It may be used for constructing experimental circuits with low-thermal connections to
minimize thermal emf effects.
2
Model 1484 Refill Kit. The Model 1484 kit contains
replacement parts for the Model 1483.
3
Model 1485 Low-Thermal Female Connector. The
Model 1485 connector is used for the mV INPUT on the
front panel of the Model 181.
4
Model 1486 Low-Thermal Male Connector. The Model
1486 connector mates with the Model 1485 female connector. It can be used to construct custom cables of
various lengths. This connector is used with the Model
1506 and 1507.
5
Model 1488 Low-Thermal Shorting Plug. The Model
1488 provides a means of shorting the mV INPUT to
check instrument offset and drift.
6
Model 1503 LawThermal Solder. The Model 1503 kit
contains low-thermal cadmium solder to make solder
connections for low voltage measurements.
7
Model 1506 Low-Thermal Input Cable. The Model 1506
cable is supplied with the unit It is a specially designed,
four foot triaxial cable that provides excellent shielding
for sensitive measurements. The Model 1506 has two
color coded alligator clips on one end, and a Model
1486 low-thermal male connector at the other end.
8
Model 1507 Low-Thermal Cable. The Model 1507 cable
is similar to the Model 1506. except that the alligator
clips are replaced by spade lugs. The Model 1507 is cons
strutted of a four foot triaxial cable and has a Model
1486 low-thermal male connector on one end.
9
Model 1815 Maintenance Kit. The Model 1815 kit cons
tains a calibration cover and extender cables that are
helpful when making service adjustments to the Model
181. The calibration cover replaces the top cover while
making these adjustments. The extender cables allow
individual PC cards to be partially removed from the unit
during maintenance.
10
Model 1019 Rack Mounting Kit. The Model 1019 kit
allows the Model 181 to be conveniently mounted in a
standard 19 inch rack.
A summary of the many optional Model 181 accessories is
listed in the following paragraphs. These accessories are
designed to enhance the capabilities of the instrument and
are described in more detail in the Model 181 Service
Manual, Document Number 30816. Contact the nearest
Keithley representative or the factory to obtain accessories.
1.4 WARRANTY INFORMATION
Warranty information may be found inside the front cover of
this manual. If warranty service is required, contact the
Keithley representative in your area or the factory to detw
mine the correct course of action. Keithlev maintains service
l-l
facilities in the United States, West Germany, Great Britain,
France, the Netherlands, Switzerland and Austria. lnforma-
tion concerning the application, operation or service of your
instrument may be directed to the applications engineer at
any of the previously mentioned locations. Check inside
front cover of this manual for addresses.
1.5 MANUAL ADDENDA
Because of a policy of constant improvement, it may
become necessary to make changes to the unit. Any modifi-
cations will be listed in an addendum attached to the inside
back cover of this manual. Be sure to note these changes
before attempting to operate the instrument.
1.6 SAFETY SYMBOLS AND TERMS
Safety symbols used in this manual are as follows:
The WARNING used in this manual explains dangers that
could result in personal injury or death.
The CAUTION used in this manual explains hazards that
could damage the instrument.
1.7 SCOPE OF OPERATOR’S MANUAL
This manual is intended to familiarize the operator with the
operating controls and features of the Model 181 nanovolto-
meter. Some of the items covered in this manual include:
basic and nanovolt measurement techniques, possible problems that could result when making measurements, addi-
tional Model 181 uses, operation of the Model 181 on the
IEEE-488 bus, and programming examples. For technical information including performance verification, theory of
operation, and maintenance procedures, refer to the Model
181 Service Manual.
The symbol
A on the instrument denotes that the
user should refer to the operating instructions.
The symbol 1/2)
on the ,nstrument denotes that 1OOOV
or more may be present on the terminal(s).
1.8 SPECIFICATIONS
For Model 181 detailed specifications, refer to the specifications that precede this section.
1-2
Figure l-l. Model 181 Front Panel View
SECTION 2
OPERATION
2.1 INTRODUCTION
This section contains information needed for basic Model
181 operation. Be sure to read this entire section before
attempting to operate the unit.
2.2 UNPACKING AND INSPECTION
The Model 181 was carefully inspected before shipment.
Upon receiving the unit, unpack all the items from the shipping carton and check for any damage that might have
occurred during shipment. Report any damage to the shipping agent at once. Save the original packing material for
possible future reshipment. Contact your nearest Keithley
representative or the factory if the unit fails to function
properly.
The following items are included with every Model 181
shipment:
1. Model 181 Nanovoltmeter
2. Model 181 Operator’s Manual
3. Model 181 Service Manual
4. Model 1506 Low-Thermal Input Cable
5. List of computer programs.
6. Additional accessories as ordered.
2.3 PREPARING FOR OPERATION
transformer must be installed. Contact your
Keithley representative or the factory for
information.
To remove the top cover. remove the two screws securing
the cover to the rear panel. Then lift off the cover from the
back until the tabs at the front of the cover clear the front
panel. Then remove the cover entirely.
Refer to Figure 2-1 for the location of the voltage switch.
Set the switch to the appropriate voltage. Also make sure
the proper fuse is installed; refer to Table 2-l for the proper
type.
Replace the top cover in the reverse order. Make sure the
tabs at the front of the cover mate with the slots in the front
panel. Finally, install the two screws that secure the top
cover to the rear panel.
Table 2-1. Fuse Selection
3AG. SLO BLO
3AG. SLO BLO
Before operating the Model 181, the appropriate line voltage
must be selected and the unit must be plugged into a proper
power source. This section covers each of these steps; be
sure to observe any precautions that are given.
1. Line Voltage Selection. The operating voltage of the
Model 181 was set at the factory as indicated on the rear
panel. Do not attempt to operate the unit with power line
voltages outside the indicated range. If it is necessary to
change the operating voltage, the top cover of the instrument must be removed to allow access to the line
voltage selection switch.
WARNING
These instructions are intended for use
only by qualified service personnel. Do not
remme the top cover unless qualified to
do so because of the possibility of electric
shock.
NOTE
The Model 181 is designed to operate with
105.125V or 210.250V as selected by the inter-
nal switch. For operation on 90.IlOV and
180.22OV power sources, a special power
POWER
TRANSFORMER
REAR PANEL
/
/
FRONT
/-
WV=
VOLTAGE VOLTAGE
SWITCH SWITCH
Figure 2-l. Line Voltage Switch Location
Y
2-l
2.
Power Line Connection. The Model 181 power cord is
supplied with a 3.prong plug that is designed to be used
with grounded outlets. Connect the female end of this
cord to the power receptacle on the rear panel of the
unit. Connect the other end to an appropriate power
SO”,lX.
CAUTION
Make sure the proper line voltage is
selected as described in the last section.
Failure to do so may result in damage to
the instrument,
possibly voiding the
warranty.
3.
Power-up Procedure. Once the power connections have
been made, the unit may be turned on by depressing the
front panel power switch. The Model 181 display should
show the line frequency and software revision level (e.g.
F60 b7) for approximately one second. After that, the
unit will revert to the normal display mode. In addition,
the 1OOOV range indicator light should be on. This is one
of the power-on default conditions that are explained
more fully in paragraph 4.5.
2.4 OPERATING CONTROLS AND CONNECTIONS
Front Panel Controls. The front panel controls are shown
in Figure 2-2. In addition to the power switch previously
described, the Model 181 has a number of other front
panel switches. The 2mV. 20mV. and 200mV switches
are used to select one of the mV measurement ranges.
The 2V. 2OV. 2OOV. and 1OOOV switches are used to
select one of the normal voltage ranges. The light above
the selected range will turn on when the appropriate
switch is depressed. Note that these switches may be
superseded by IEEE commands as outlined in Section 4.
In addition to the range switches, the Model 181 has
several other front panel controls. These include: the HI
RES switch to select 5% or 6% resolution, the ZERO
switch to enable baseline suppression, and the FILTER
and DAMPING switches, which alter the response of the
internal 3-pole filter. These features will be described in
more detail in later sections.
Front Panel Connections. The front panel has two input
connectors. The two 5.way binding posts are used for
measurements on the 2V through 1OOOV ranges, while
the low-thermal mV INPUT connector is used for
measurements on the 2mV through 200mV ranges.
When using the mV INPUT, be sure to use the supplied
low-thermal cable to minimize errors caused by thermal
emfs.
Display. The 6% digit display is used to make Model 181
voltage readings. The display may be switched to either
5% or 6% digits at the touch of a button. A leading
minus sign appears when negative voltages are
measured, and the decimal point is automatically placed.
Overrange is indicated by an “OFLO” message.
4. IEEE Status Lights. The TALK, LISTEN, and REMOTE indicator lights show the present IEEE status of the Model
181. For complete IEEE information, refer to Section 4.
5. Rear Panel Controls and Connections. The rear panel
Controls and connections are shown in Figure 2-3. An
analog output is available through the two 5-way binding
posts. The switches and connector shown in the lower
left corner are for use with the IEEE-488 bus. The functions and operation of these connectors and switches will
be covered in more detail in later paragraphs.
6. Tilt Sail. The tilt bail is useful for elevating the front panel
of the instrument to a convenient height. To extend the
tilt bail, rotate it 90’ away from the bottom cover; then
push the bail upward until it locks into place. To retract
the bail, first pull the bail down away from the front cover
to release the locking mechanism; then rotate the bail until it is flush with the bottom cover.
2.5 BASIC VOLTAGE MEASUREMENT
Normal voltage measurements are made on the 2V through
IOOOV ranges. To use one of these ranges, the source to be
measured must be connected to the V INPUT. The follow-
ing paragraphs describe the basic procedure for making
these voltage measurements.
Turn on the Model 181 by depressing the front panel
power switch. As previously described, the unit should
momentarily display the line frequency and software revision level. Allow a one hour warm-up period to obtain
rated accuracy. Four hours are required for minimum
drift.
Select the desired voltage range by depressing the
appropriate range button. Select a range that can easily
handle the maximum voltage to be measured.
Select other front panel operating modes, such as HI
RES, ZERO, DAMPING,and FILTER, as required. Refer
to paragraph 2.8 for further information on these
controls.
Connect the source to be measured to the V INPUT terminals es shown in Figure 2-4. Note that circuit ground is
normally connected to the LO terminal, while the HI terminal should be connected to the point to be measured.
CAUTION
Do not exceed IOOOV between the HI and
LO V INPUT terminals or the instrument
might be damaged. Note that the LO
INPUT terminal floats and is not connected to power line ground. Therefore, it
is important that the potential between the
LO input terminal and power line ground
not exceed 14OOV. or the instrument might
be damaged.
WARNING
Observe normal safety precautions when
connecting the Model 181 to potentially
lethal voltage sources. Failure to observe
these precautions may result in serious
personal injury because of electric shock.
2-2
mV RANGES
DISPLAY
IEEE STATUS LIGHTS
- mV INPUT
ANALOG
OUTPUT
POWER
ON/OFF
0
II I
RESOLUTION
Figure 2-2. Front Panel Controls and Connections
^ ^ ^ .
U,SLUNNC
DISPLAY
* ,\
A?
ZERO
FILTER CONTROLS
V HANGES
ENABLE
AC RECEPTACLE
I ,Nt ClArlNG
.,,,,,. ,,, :r,. ,...
1
,. t.;,
,.,. ,,I_, . “,.L
ANALOG OUTPUT
RANGE SWITCH
IEEE CONNECTOR
Figure 2.3. Rear Panel Controls and Connections
2-3
5. Observe the display; if an “OFLO” is shown, switch to
the next higher range. Use the lowest range possible to
make the measurement. This procedure will achieve the
best resolution.
6. Make the voltage reading. The display shows the reading
directly in DC volts with a leading minus sign for negative
voltages. No conversion is necessary as the decimal
point is automatically placed on all ranges.
7. The Model 181 input impedance is greater than 10% on
the 2V range and equal to lOMl7 on the 20V through
IOOOV ranges. Thus, loading should not be a problem
except with very high source resistance values. Refer to
paragraph 2.10 for precautions to be taken under those
conditions.
VOLTAGE SOURCE
,100” MAX,
7
-
CIRCUIT GROUND
,WHERE APPLICABLE1
-L
-
i
-I-
4. Connect the low-thermal cable to the mV input. Connect
the alligator clips of the cable to the voltage source to be
measured as shown in Figure 2-5.
CAUTION
Do not exceed 120V momentary. 35V con-
tinous, between the mV INPUT terminals.
or 1400V between the mV low terminal and
ground. Failure to observe these precau-
tions may result in damage to the unit.
5. Observe the display reading; if the unit is in overflow,
select the next higher range. If an overflow condition
exists on the 200mV range, use the V INPUT and appropriate range as outlined in the preceding paragraph.
6. Take the voltage reading. The reading may be made
directly, in millivolts,
automatically placed. A leading minus sign will be
displayed for negative voltages.
7. Because of the very low signal levels involved, unwanted
no,se, as CleSCrlDea I” paragrapn z.11, may upset the ac-
curacy of the measurement.
since the decimal point is
^
Figure 2-4. Basic Voltage Measurements
NANOVOLT AND MICROVOLT MEASUREMENTS
2.6
The Model 161 may be used to make very low voltage
readings down to a resolution of 10nV. These readings are
made on one of the mV ranges by using the mV INPUT on
the front panel.
The following paragraphs describe the basic procedure for
making these measurements.
1. Turn on the Model 181 with the front panel POWER
switch. Allow the unit to warm-up for at least an hour for
rated accuracy. To guarantee low drift, allow at least four
hours.
2. Select the desired mV range with the appropriate front
panel switch. Use a range appropriate for the voltage to
be measured.
3. Select other parameters such as HI RES, DAMPING,
FILTER, and ZERO as needed. Refer to paragraph 2.8 for
more details on these controls.
2-4
I
Figure 2-5. mV and nV Measurements
2.7 SPECIAL MEASURING SITUATIONS
Some situations may call for a wide range of voltage
measurements that neither the V input nor mV input can
handle alone. In those cases, it may be convenient to use a
common ground for both the V and mV inputs. Since the
LO terminals of the mV and V inputs are internally connected together, it is only necessary to connect the mV Lo
terminal (black lead of the Model 1506 low-thermal cable) to
common of the circuit under test, as shown in Figure 2-6.
Using this method, either the V HI or mV HI terminal can be
used as the test probe, depending on the voltage to be
measured.
CAUTION
Do not exceed the maximum input limit for
the Model 181, especially when the mV HI
terminal is connected. or damage to the in-
strument may occur. Never parallel the mV
and V leads to prevent accidental overload
to the mV input or inadvertent loading of
the circuit under test.
Figure 2-6. Common Ground Connection for V and mV
The zero function is especially useful for nulling out offset
voltages, including internal offsets of the Model 181. To use
the zero in this manner, short the test leads together with
the instrument on the desired range and depress the ZERO
switch; the ZERO indicator light should turn on. This stores
the residual voltage level as the baseline. All voltage reading
taken with zero enabled will then be the actual voltage level
since the unwanted voltage will be subtracted from the
reading.
Note that baseline suppression for the V and mV ranges
operates separately. Switching the unit between a mV and
V range, for example, will cancel the ZERO, also causing the
front panel ZERO indicator light to turn off.
Controlling the Filter. The Model 181 has an internal 3-p&
digital filter that can be controlled by the front panel FILTER
and DAMPING controls. Normally, the filter is switched on
and off as a function of the rate of change in input signal.
Depressing the FILTER button increases the RC time conk
stant of the filter. At the same time, the front panel FILTER
light will turn on. The digital filter cannot be totally disabled
by the front panel controls. However, it may be disabled by
commands given over the IEEE bus. Operating with the filter
disabled allows the user to customize Model 161 response
by using external filtering. For further information on IEEE
commands that control the filter, consult Section 4 of this
manual.
2.8 ADDITIONAL FRONT PANEL CONTROLS
The Model 181 has additional front panel controls that can
be used to enhance the capabilities of the unit. These
switches which include HI RES, ZERO, FILTER, and DAMP-
ING, are shown in Figure 2-2. The following paragraphs will
describe the operation of these controls in more detail.
HI RES. The display resolution of the Model 181 upon
power-up is 5% digits. The display resolution may be increased to 6 % digits by depressing the HI RES switch. Once
the unit is in the 6% digit mode, the display may be returned
to the 5% digit mode by depressing the HI RES switch a
second time. Readings made in the 5% digit mode have the
least significant digit rounded off. HI RES switch affects
only the data on the display; data transmitted over the IEEE
bus always contains 6% digit information. For further infor-
mation on IEEE operation, refer to Section 4.
Zero. The Zero mode serves as a means for baseline sup-
pression. The front panel ZERO indicator light will turn on
when the zero mode is enabled. All readings taken with the
zero enabled will be the difference between the stored
baseline and the actual voltage level.
The baseline is obtained by connecting the instrument to
the voltage to be zeroed. For example, if the baseline
voltage is IOmV, all subsequent readings will have 1OmV
subtracted from the actual voltage level.
The DAMPING button controls whether or not the filter is
continuously enabled. When the DAMPING is off, the
microprocessor automatically disables the filter when the
input voltage changes to permit rapid display update. Once
the reading is within 25 digits of the final value on the 2mV
range, and within 6 digits on the remaining ranges. the
microprocessor then enables the filter to minimize noise in
the final reading. When the DAMPING is on. the digital filter
is permanently enabled. The unit would normally be
operated in this mode only for signals that vary slowlv, or
with extremely noisy ambient signals.
Through careful use of the FILTER and DAMPING controls,
the user can optimize the Model 161 to the required perfw
mance, keeping in mind the resulting speed/ noise compromises. Figure 2-7 shows four curves resulting from
operating the unit with various combinations of the DAMPS
ING and FILTER controls. Curve A shows the fastest
response time because the filter RC time constant is at a
minimum. Also, with DAMPING off, the microprocessor
initially disables the filter as previously described.
Depressing the FILTER switch as with curve 6, has little
effect on the response time since the filter is initially off.
Curves C and D, on the other hand, show that enabling the
DAMPING slows the response down considerably. This can
be seen in more detail in Table 2-2, which lists the settling
times of the various control combinations.
2-5
Table 2-2 Settling Times
~~~~~-~~~~“~~~~~
(The readings all settle to within 0.002% of the Full Range in the specified time.)
2.9 USING THE ANALOG OUTPUT
The analog output of the Model 181 is useful for monitoring
the input signal with an external device such as a chart
recorder. The analog signal is reconstructed from digital
data (supplied by the internal microprocessor) by a 12 bit
D/A converter. Because of this configuration, the analog
output will accurately reflect the display until an overflow
condition is reached. The analog output is optically isolated
from the front panel LO terminal to avoid potential ground
loop problems. The following paragraphs describe the basic
procedure for using the analog output.
1. Connect the measuring device to the two analog output
terminals on the rear panel as shown in Figure 2-8.
CAUTION
The potential between the analog output
LO terminal and ground must not exceed
30V. Make sure the external device does
not exceed this voltage on its common or
ground connections. Failure to observe
this precaution may damage the Model
181. possibly voiding the warranty. IEEE
common is connected to analog output
IOW.
Select the Xl or Xl000 range by using the analog output
gain switch on the rear panel. This switch is combined
with those used to set the IEEE mode in the lower left cor-
ner of the rear panel and is clearly marked. (See Figure
2.3.) In the Xl position, the most significant +2000
counts of the display reading can be covered, while the
Xl000 position will change the range to cover the least
significant f2000 counts. In this manner, the entire 6%
digits of the display may be represented.
If necessary, the analog output may be zeroed with the
front panel ZERO control. To do so, depress the ZERO
button.
The Model 181 will display an “OFLO” message when the
capability of a specific range is exceeded. When this
message is displayed. the analog output value will be + 2V if
the polarity of the input voltage is positive, and -2V if the
input voltage polarity is negative.
An analog output range overflow can occur when the Model
181 analog range switch is in the Xl000 position. An example of the analog ouput voltage under these conditions is
shown in Figure 2-9.” The conditions shown are for the 2mV
range. but the output will react similarly on the other voltage
ranges if the proper scaling factor is applied. For each ten-
fold increase in voltage range, the scale of the horizontal
axis must also be multiplied by a factor of ten.
2-6
Figure 2-8. Analog Output Connections
The horizontal axis of Figure 2-9 has an input voltage range
between -10&V and +lOpV. The vertical axis shows an
analog output voltage between -2V and +ZV. which is the
maximum range of the analog output. Beginning at the OV
point on the graph, the analog output follows the input
voltage linearly until the input voltage reaches +2pV. The
analog output will then suddenly switch to the maximum
negative output value of -2V. Thus, for each 4uV increment
* Units with B-7 software.
in input voltage, the output pattern repeats. ihe action of
the analog output for negative input voltages is the same,
except that the slope of the graph is negative for these
negative-going input voltages.
Figure 2-9. Xl000 Analog Output
By counting the number of repeating waveforms on a chart
recorder, the user can easily determine the actual voltage at
the input, even though the range of the analog output was
exceeded. If, for example, the +lV point on the second
peak with a positive-going slope is noted, it can be clearly
determined that the input voltage was +5@V at that particular time.
A summary of analog output information is shown in Table
2.3. Each range of input values corresponds to the incre-
ment necessary to cause the output to go through its entire
0 to 2V range. Note that the sensitivity is increased by a fac-
tor of a thousand on the Xl000 range. For example, when
the Model 181 is in the 200mV range, and the analog switch
is in the Xl position, the output voltage will swing from 0 to
2V in a smooth manner as the input voltage increases
gradually from 0 to 200mV. When the analog output switch
is changed to the Xl000 position, the input need only swing
between 0 and 2OOpV to obtain the same voltage swing at
the analog output. Beyond those input limitations, the out-
put voltage will repeat as shown in Figure 2-9.
The output resistance of the analog ouput is Ikll for all
voltage ranges regardless of the position of the analog range
switch. Thus, loading problems caused by external devices
are minimized. To keep loading errors below I%, the input
resistance of any device connected to the analog output
should be greater than lOOk0.
Table 2-3. Analog Output Parameters
2.10 SOURCE RESISTANCE CONSIDERATION
The Model 181 has an input resistance greater that IGIl
flO% on the 2mV. 20mV. 200mV. and 2V ranges. The in-
strument will meet this input resistance specification on the
mV ranges even when in overflow with voltages up to 1V.
The input resistance on the remaining voltages ranges is
lOML2. Thus, the Model 181 input resistance is sufficiently
high to minimize loading errors in most measuring situations. For voltage sources with very high source resistance,
two precautions should be observed when using the Model
181.
Shielding becomes more critical when the source resistance
is very high. Otherwise, interference signals may be picked
up by the test leads. Noise picked up in this manner can affect the mV ranges more severely. but shielding might be
necessary for connections to the V INPUT in extreme
situations.
Loading of the voltage source by the Model 181 can become
important with high source resistance values. As the source
resistance increases, the error due to loading increases.
Figure Z-10 shows the method used to determine the Peru
cent error due to loading. The voltage source has an internal
resistance R,, while the internal resistance of the Model 181
is represented by R,. The source voltage is E, while the
voltage actually measured by the meter is E,.
The voltage actually seen by the meter is attenuated by the
voltage-divider action of R and R, and can be found by
using the relationship: E, = &R,IIRL + I?,).
We can modify this relationship to obtain a formula for per-
cent errors as follows: Percent Error = lOOR,/(R, * R,i~
From the above, it is obvious that the input resistance of the
Model 181 must be at least 99 times greater that the source
resistance if the loading error is to be kept to 1%. This maximum 1% error limitation will be achieved on the 2mV
through 2V ranges with sources resistances up to lO.lMI1,
while the source resistance should be no greater than IOlklI
if the same 1% error limitation is to be maintained on the
2OV through IOOOV ranges. If lower errors are required, the
source resistance must be correspondingly less.
Rs
20mV
200mV
2v
20 v
200 v
1 kV
*IV Full Range Maximum
INPUT FOR
IG OUTPUT
Figure 2-10. Source Resistance Considerations
2-7
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.