Keithley 181 User Manual

Operator’s Manual
Model 181
Digital Nanovoltmeter
01982. Keithley Instruments, Inc Cleveland, Ohio, U.S.A.
Document Number 32421
l
NANOVOLTMETER
IEEE-488 BUS IMPLEMENTATION MULTlLlNE COMMANDS: DCL, LLCJ, xc, GET. “NlLINE COMMANDS: IFC, REN, EOI, SRQ, ATN. PROGRAMMABLE PARAMETERr3
Front Panel Controls: Range, Filter, Zero, Damping, Hi
Resoluli”“.
Internal Pnrametea: SRQ Rcsponsc, Trigger Modes, Data Ter-
minators.
GENERAL
DISPLAY: Seven 0.5 in. LED digits with appropriate decimal point
and polarity.
NOISE: <30”” p-p on lowest range wit,, Filter on.
lNP”T CAPACITANCE: 5OOOpF on mV ranges.
SETTLING TIME: 0.5 sec. to within 25 COLMS of final reading with
Fikr on, Damping off. FILTER: 3-polcdigital; RC = 0.5.l.or2 wands dcpcndingonrangc. CON”CRSlON WEBI): 4 readings/second. OVERLOAD INDICATION: Display indicates polarity and OFI.0. ANALOG OUTPUT: Accuracy: +(0.15% of displayed reading +
ImW Time Constant: 400mscc. Level: f2V full scale on all rongcs;
xl or xl000 gain.
ISOLATION: Input LO to Output LO or power line ground: ,400”
peak, 5 x IOW*Hz, >I09 paralleled by 150OpF. WARM-UP: 1 hour to r&cd accuracy. ENVIRONMENTAL LIMITS: Operating: 0”35”C, O%-80% *da-
tive humidity. Storage: -25” to +WC. POWER: 105-123’ or 21~25OV (internal switch selected), 50-6OHz,
30Vh maximum.
ADDRESS MODES: TALK ONLY and ADDRESSABLE. TRIGGlX MODES: One Shot: Updatcs output buffer once at first
valid conversion after triggeronTALKand/arG~T. Continuous: Updates output buffer at al, valid ~onwr~ion~ after trigger.
INFUT CONNECTOR% Special low thermal for 2”OmV and lower
ranges. Binding posts for 2V to lOO”V ranges.
DIMENSIONS, WEIGHT: 127mm high x 21hmm wide x 359mm
deep (5 in. x 8.5 in. x 14.125 in.). Net weight 3.85kg (8.5 Ibs.). ACCESSORY SUPPLIED: Model ,506 Low ‘Thermal fnput Cable. ACCESSORfES AVAILABLE:
Model 262: Low Thermal Voltage Divider
Model 1019A-1: 5%-i,,. Single Fixed Rack Mounting Kit
Model 1019A-2: 5’/rin. Dual Fixed Rack Mounting Kit
Model 1019SI: 5’,~-in. Single Slide Rack Mounting Kit
Model 1019S.2: Y/1-in. Dual Slide Rack Mounting Kit
Model 1483: Low Thermal Connechan Kit
Model 1484:
Madcl1485: Fcmalc Low Thermal Input Conneck!r
Model 1486: Male Low Thermal Input Connector
Model 1488: Low Thermal Shorting I’lug
Made, 1506: Low Thermal Input Cable (4 ft., clips)
Model 1507:
Model ,815: Maintcnancc Kit
Model 8003: Low Resistance Test Fixture
&fill Kit for 1483 Kit
Low Thermal Input Cable (4 ft., plugs)
TABLE OF CONTENTS
Paragraph
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
2.1
2.2
2.3
2.4
2.5
2.6
2.7
2.8
2.9
2.10
2.11
3.1
3.2
3.3
3.4
3.5
3.6
3.7
Title
SECTION l-GENERAL INFORMATION
Introduction.. ...............................................................................
Model181 Features.. .........................................................................
Optional Accessories
Warranty Information .........................................................................
ManualAddenda..
SafetySymbolsandTerms .....................................................................
ScopeofOperator’sManual...........................................................~.~~ .....
Specifications ................................................................................
SECTION 2-OPERATION
Introduction.. ................................................................................
Unpackingandlnspection
PreparingforOperation.................................................................~...~
Operating Conrtrolsand Connections
BasicVoltageMeasurement..........................................................~.~
Nanovolt and Microvolt Measurements
Special Measuring Situations
Additional Front Panel Controls. .................................................................
UsingtheAnalogOutput .......................................................................
Source Resistance Considerations ..........................................................
Microvolt and Nanovolt Measurement Consideration SECTION 3-APPLICATIONS
Introduction.. ................................................................................
StandardCellComparisons .....................................................................
Low Resistance “Lindeck” Measurements
TemperatureMeasurements .........................................................
ResistanceThermometry .......................................................................
SemiconductorTesting .........................................................................
JosephsonJunctionStudies ....................................................................
..........................................................................
...........................................................................
............................................................
.............................................................
....................................................... ...
....................................................................
...............................................
..................................................... ..
........
........
..........
....
...
Page
l-l 1~1 1~1 l-l 1~2 1~2 1~2 1-2
2~1 2~1 2~1 2-2 2-2 2~4 2-4 2-5 2~6
2-J
2-8
3-1
3~1 3-2 3-3 3~3 3-3 3-3
4.1
4.2
4.3
4.4
4.6
4.6
4.7
4.8
4.9
SECTION 4-IEEE OPERATION
Introduction to the IEEE-488 Bus .................................................................
Descriptionof BusLines ........................................................................
IEEE-488Set-UpProcedure.. ...................................................................
BusCommands.. .............................................................................
Device-Dependentcommands ..................................................................
Data Format.. ................................................................................
StatusByte Format.. ..........................................................................
StatusWordFormat ...........................................................................
ProgrammingExample .........................................................................
4-1 4-l 4-2 4-3 4~5 4-7 4-8 4-9 4-9
LIST OF ILLUSTRATIONS
Figure
l-l 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 2-11 2-12 2-13 2-14 3-l 3-2 3-3 3-4 4-1 4-2 4-3
4-4
4-5
4-6
4-7
4-8
Model 181 Front Panel View. .............
Line Voltage Switch Location ............
Front Panel Controls and Connections Rear Panel Controls and Connections. Basic Voltage Measurements.
mV and nV Measurements ...............
Common Ground Connections for V and mV
Filter Response Graph. ..................
Analog Output Connections ..............
Xl000 Analog Output ...................
Source Resistance Consideration .........
Thermal emf Generation .................
Power Line Ground Loops ...............
Ground Loop Voltage Generation .........
Eliminating Ground Loops ...............
Standard Cell Comparison ...............
Absolute Cell Measurement Connections. Low Resistance Measurement Connections. Minimizing Josephson Junction RFI Effects
IEEE Bus Configuration ..................
IEEEHandshakeSequence ...............
Primary Address and IEEE Mode Switches.
IEEE Contact Configuration ..............
IEEE Bus Data Format ...................
Status Byte Format .....................
Programming Example ..................
Timing Diagram .......................
............
.....
.....
..
Title Page
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
.....
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
.....
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
.....
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
.....
..........
.......... 2-l
.......... 2-3
.......... 2-3
.......... 2-5
.......... 2-6
.......... 2-6
.......... 2-7
.......... 2-7
.......... 2-8
.......... 2-9
......... 2-10
..........
.......... 3-2
..........
.......... 4-l
.......... 4-2
..........
.......... 4-8
.......... 4-9
......... 4-11
l-2
2-4 2-4
2-9
3-l 3-2
3-3
4-3 4-3
4-13
Table
2-l
2-2
2-3
4-l
4-2
4-3
4-4
4-5
4-6
4-7
4-8
4-S
ii
Fuse Selection .........................
Settling Times .........................
Analog Output Parameters. ..............
IEEE Contact Designations ...............
Bus Command Summary ................
Device-Dependent Command Summary
Range Commands ......................
Default Conditions. .....................
Data String Exponent Values .............
Error and Data Code Summary ...........
Status Word Example ...................
HP-85 BASIC IEEE-488 Statements .......
LIST OF TABLES
...
......
......
......
......
......
......
......
......
......
., ....
Page
2-l 2-6 2-7 4-4 4-4 4-6 4-6 4-7 4-8 4-9 4-9
4-10
SECTION 1
GENERAL INFORMATION
1.1 INTRODUCTION The Keithley Model 181 is a highly sensitive nanovoltmeter
with a large, easy to read 5 K or 6 ‘/ digit display. The Model 181 is unique in that it combines microprocessor technology with a new concept in low-noise, high-gain front ends, resulting in a programmable instrument with sensitivity down to 10nV. The Model 181 provides highly accurate, stable, low-noise readings on seven ranges for DC voltage measurements between 1OnV and 1OOOV. The mV ranges use a special low-thermal input connector, while connec­tions for the higher voltage ranges are made through two S-way binding posts. Additional versatility is afforded by the inclusion of an IEEE-488 interface which allows the unit to communicate with other instrumentation.
1.2 MODEL 181 FEATURES The Model 181 includes the following features:
l
High Sensitivity. The resolution of the Model 181 on the
2mV range is IO ~* volts (10nV).
l
5% or 6% Digit Resolution. Normal 5% digit display
resolution may be increased to 6 % digits at the touch of a button.
l
IEEE-488 Interface. A built in IEEE-488 interface allows the instrument to communicate with other devices such as a central controller or printer.
l
Analog Output. An analog output, which accurately
reflects the displayed readings, is available from the rear
pallel.
l
3-p& Digital Filter. The internal 3-p& filter minimizes the
effects of noise in voltage readings and may be controlled from the front panel or IEEE bus.
l
Separate Inputs. A special input connector is used for the
mV ranges to minimize thermal emf generation.
l
Isolated Low Terminals. The low signal connections for
both inputs are isolated from power line ground and
from IEEE low to minimize ground loop problems.
l
Color Coded Front Panel. Inputs, range switches, and
other front panel controls are marked to form color-coded
groups for easier operation.
1.3 OPTIONAL ACCESSORIES
1
Model 1483 Low-Thermal Connection Kit. The Model
1483 kit contains a crimp tool, pure copper lugs, Lowe thermal cadmium solder, copper alligator clips, and assorted hardware. It may be used for constructing ex­perimental circuits with low-thermal connections to minimize thermal emf effects.
2
Model 1484 Refill Kit. The Model 1484 kit contains replacement parts for the Model 1483.
3
Model 1485 Low-Thermal Female Connector. The Model 1485 connector is used for the mV INPUT on the front panel of the Model 181.
4
Model 1486 Low-Thermal Male Connector. The Model 1486 connector mates with the Model 1485 female con­nector. It can be used to construct custom cables of various lengths. This connector is used with the Model 1506 and 1507.
5
Model 1488 Low-Thermal Shorting Plug. The Model 1488 provides a means of shorting the mV INPUT to check instrument offset and drift.
6
Model 1503 LawThermal Solder. The Model 1503 kit
contains low-thermal cadmium solder to make solder connections for low voltage measurements.
7
Model 1506 Low-Thermal Input Cable. The Model 1506 cable is supplied with the unit It is a specially designed, four foot triaxial cable that provides excellent shielding for sensitive measurements. The Model 1506 has two color coded alligator clips on one end, and a Model 1486 low-thermal male connector at the other end.
8
Model 1507 Low-Thermal Cable. The Model 1507 cable is similar to the Model 1506. except that the alligator clips are replaced by spade lugs. The Model 1507 is cons strutted of a four foot triaxial cable and has a Model 1486 low-thermal male connector on one end.
9
Model 1815 Maintenance Kit. The Model 1815 kit cons tains a calibration cover and extender cables that are helpful when making service adjustments to the Model
181. The calibration cover replaces the top cover while making these adjustments. The extender cables allow individual PC cards to be partially removed from the unit during maintenance.
10
Model 1019 Rack Mounting Kit. The Model 1019 kit allows the Model 181 to be conveniently mounted in a standard 19 inch rack.
A summary of the many optional Model 181 accessories is listed in the following paragraphs. These accessories are designed to enhance the capabilities of the instrument and are described in more detail in the Model 181 Service Manual, Document Number 30816. Contact the nearest Keithley representative or the factory to obtain accessories.
1.4 WARRANTY INFORMATION
Warranty information may be found inside the front cover of this manual. If warranty service is required, contact the
Keithley representative in your area or the factory to detw
mine the correct course of action. Keithlev maintains service
l-l
facilities in the United States, West Germany, Great Britain,
France, the Netherlands, Switzerland and Austria. lnforma-
tion concerning the application, operation or service of your
instrument may be directed to the applications engineer at any of the previously mentioned locations. Check inside front cover of this manual for addresses.
1.5 MANUAL ADDENDA
Because of a policy of constant improvement, it may
become necessary to make changes to the unit. Any modifi-
cations will be listed in an addendum attached to the inside
back cover of this manual. Be sure to note these changes
before attempting to operate the instrument.
1.6 SAFETY SYMBOLS AND TERMS
Safety symbols used in this manual are as follows:
The WARNING used in this manual explains dangers that could result in personal injury or death.
The CAUTION used in this manual explains hazards that could damage the instrument.
1.7 SCOPE OF OPERATOR’S MANUAL
This manual is intended to familiarize the operator with the operating controls and features of the Model 181 nanovolto-
meter. Some of the items covered in this manual include: basic and nanovolt measurement techniques, possible pro­blems that could result when making measurements, addi-
tional Model 181 uses, operation of the Model 181 on the
IEEE-488 bus, and programming examples. For technical in­formation including performance verification, theory of operation, and maintenance procedures, refer to the Model
181 Service Manual.
The symbol
A on the instrument denotes that the
user should refer to the operating instructions.
The symbol 1/2)
on the ,nstrument denotes that 1OOOV
or more may be present on the terminal(s).
1.8 SPECIFICATIONS
For Model 181 detailed specifications, refer to the specifica­tions that precede this section.
1-2
Figure l-l. Model 181 Front Panel View
SECTION 2
OPERATION
2.1 INTRODUCTION This section contains information needed for basic Model
181 operation. Be sure to read this entire section before
attempting to operate the unit.
2.2 UNPACKING AND INSPECTION The Model 181 was carefully inspected before shipment.
Upon receiving the unit, unpack all the items from the ship­ping carton and check for any damage that might have occurred during shipment. Report any damage to the ship­ping agent at once. Save the original packing material for
possible future reshipment. Contact your nearest Keithley
representative or the factory if the unit fails to function
properly. The following items are included with every Model 181
shipment:
1. Model 181 Nanovoltmeter
2. Model 181 Operator’s Manual
3. Model 181 Service Manual
4. Model 1506 Low-Thermal Input Cable
5. List of computer programs.
6. Additional accessories as ordered.
2.3 PREPARING FOR OPERATION
transformer must be installed. Contact your Keithley representative or the factory for information.
To remove the top cover. remove the two screws securing the cover to the rear panel. Then lift off the cover from the back until the tabs at the front of the cover clear the front panel. Then remove the cover entirely.
Refer to Figure 2-1 for the location of the voltage switch. Set the switch to the appropriate voltage. Also make sure
the proper fuse is installed; refer to Table 2-l for the proper
type. Replace the top cover in the reverse order. Make sure the
tabs at the front of the cover mate with the slots in the front panel. Finally, install the two screws that secure the top cover to the rear panel.
Table 2-1. Fuse Selection
3AG. SLO BLO
3AG. SLO BLO
Before operating the Model 181, the appropriate line voltage must be selected and the unit must be plugged into a proper power source. This section covers each of these steps; be sure to observe any precautions that are given.
1. Line Voltage Selection. The operating voltage of the Model 181 was set at the factory as indicated on the rear panel. Do not attempt to operate the unit with power line voltages outside the indicated range. If it is necessary to change the operating voltage, the top cover of the in­strument must be removed to allow access to the line voltage selection switch.
WARNING
These instructions are intended for use only by qualified service personnel. Do not
remme the top cover unless qualified to do so because of the possibility of electric shock.
NOTE
The Model 181 is designed to operate with
105.125V or 210.250V as selected by the inter-
nal switch. For operation on 90.IlOV and
180.22OV power sources, a special power
POWER
TRANSFORMER
REAR PANEL
/
/
FRONT
/-
WV=
VOLTAGE VOLTAGE
SWITCH SWITCH
Figure 2-l. Line Voltage Switch Location
Y
2-l
2.
Power Line Connection. The Model 181 power cord is supplied with a 3.prong plug that is designed to be used with grounded outlets. Connect the female end of this cord to the power receptacle on the rear panel of the unit. Connect the other end to an appropriate power SO”,lX.
CAUTION
Make sure the proper line voltage is selected as described in the last section.
Failure to do so may result in damage to the instrument,
possibly voiding the
warranty.
3.
Power-up Procedure. Once the power connections have been made, the unit may be turned on by depressing the front panel power switch. The Model 181 display should show the line frequency and software revision level (e.g.
F60 b7) for approximately one second. After that, the
unit will revert to the normal display mode. In addition, the 1OOOV range indicator light should be on. This is one of the power-on default conditions that are explained
more fully in paragraph 4.5.
2.4 OPERATING CONTROLS AND CONNECTIONS Front Panel Controls. The front panel controls are shown
in Figure 2-2. In addition to the power switch previously described, the Model 181 has a number of other front panel switches. The 2mV. 20mV. and 200mV switches are used to select one of the mV measurement ranges. The 2V. 2OV. 2OOV. and 1OOOV switches are used to select one of the normal voltage ranges. The light above the selected range will turn on when the appropriate switch is depressed. Note that these switches may be superseded by IEEE commands as outlined in Section 4.
In addition to the range switches, the Model 181 has several other front panel controls. These include: the HI RES switch to select 5% or 6% resolution, the ZERO
switch to enable baseline suppression, and the FILTER and DAMPING switches, which alter the response of the internal 3-pole filter. These features will be described in more detail in later sections.
Front Panel Connections. The front panel has two input connectors. The two 5.way binding posts are used for measurements on the 2V through 1OOOV ranges, while the low-thermal mV INPUT connector is used for measurements on the 2mV through 200mV ranges.
When using the mV INPUT, be sure to use the supplied
low-thermal cable to minimize errors caused by thermal emfs.
Display. The 6% digit display is used to make Model 181 voltage readings. The display may be switched to either
5% or 6% digits at the touch of a button. A leading
minus sign appears when negative voltages are measured, and the decimal point is automatically placed. Overrange is indicated by an “OFLO” message.
4. IEEE Status Lights. The TALK, LISTEN, and REMOTE in­dicator lights show the present IEEE status of the Model
181. For complete IEEE information, refer to Section 4.
5. Rear Panel Controls and Connections. The rear panel Controls and connections are shown in Figure 2-3. An analog output is available through the two 5-way binding posts. The switches and connector shown in the lower left corner are for use with the IEEE-488 bus. The func­tions and operation of these connectors and switches will be covered in more detail in later paragraphs.
6. Tilt Sail. The tilt bail is useful for elevating the front panel of the instrument to a convenient height. To extend the
tilt bail, rotate it 90’ away from the bottom cover; then
push the bail upward until it locks into place. To retract the bail, first pull the bail down away from the front cover
to release the locking mechanism; then rotate the bail un­til it is flush with the bottom cover.
2.5 BASIC VOLTAGE MEASUREMENT Normal voltage measurements are made on the 2V through
IOOOV ranges. To use one of these ranges, the source to be
measured must be connected to the V INPUT. The follow-
ing paragraphs describe the basic procedure for making
these voltage measurements.
Turn on the Model 181 by depressing the front panel power switch. As previously described, the unit should momentarily display the line frequency and software revi­sion level. Allow a one hour warm-up period to obtain rated accuracy. Four hours are required for minimum drift.
Select the desired voltage range by depressing the appropriate range button. Select a range that can easily handle the maximum voltage to be measured.
Select other front panel operating modes, such as HI RES, ZERO, DAMPING,and FILTER, as required. Refer to paragraph 2.8 for further information on these controls.
Connect the source to be measured to the V INPUT ter­minals es shown in Figure 2-4. Note that circuit ground is normally connected to the LO terminal, while the HI ter­minal should be connected to the point to be measured.
CAUTION
Do not exceed IOOOV between the HI and LO V INPUT terminals or the instrument might be damaged. Note that the LO INPUT terminal floats and is not con­nected to power line ground. Therefore, it is important that the potential between the LO input terminal and power line ground not exceed 14OOV. or the instrument might be damaged.
WARNING
Observe normal safety precautions when connecting the Model 181 to potentially
lethal voltage sources. Failure to observe these precautions may result in serious personal injury because of electric shock.
2-2
mV RANGES
DISPLAY
IEEE STATUS LIGHTS
- mV INPUT
ANALOG
OUTPUT
POWER
ON/OFF
0
II I
RESOLUTION
Figure 2-2. Front Panel Controls and Connections
^ ^ ^ .
U,SLUNNC
DISPLAY
* ,\
A?
ZERO
FILTER CONTROLS
V HANGES
ENABLE
AC RECEPTACLE
I ,Nt ClArlNG
.,,,,,. ,,, :r,. ,...
1
,. t.;,
,.,. ,,I_, . “,.L
ANALOG OUTPUT
RANGE SWITCH
IEEE CONNECTOR
Figure 2.3. Rear Panel Controls and Connections
2-3
5. Observe the display; if an “OFLO” is shown, switch to the next higher range. Use the lowest range possible to make the measurement. This procedure will achieve the best resolution.
6. Make the voltage reading. The display shows the reading directly in DC volts with a leading minus sign for negative voltages. No conversion is necessary as the decimal point is automatically placed on all ranges.
7. The Model 181 input impedance is greater than 10% on
the 2V range and equal to lOMl7 on the 20V through
IOOOV ranges. Thus, loading should not be a problem except with very high source resistance values. Refer to paragraph 2.10 for precautions to be taken under those conditions.
VOLTAGE SOURCE
,100” MAX,
7
-
CIRCUIT GROUND
,WHERE APPLICABLE1
-L
-
i
-I-
4. Connect the low-thermal cable to the mV input. Connect the alligator clips of the cable to the voltage source to be measured as shown in Figure 2-5.
CAUTION
Do not exceed 120V momentary. 35V con-
tinous, between the mV INPUT terminals.
or 1400V between the mV low terminal and ground. Failure to observe these precau-
tions may result in damage to the unit.
5. Observe the display reading; if the unit is in overflow, select the next higher range. If an overflow condition exists on the 200mV range, use the V INPUT and appro­priate range as outlined in the preceding paragraph.
6. Take the voltage reading. The reading may be made directly, in millivolts,
automatically placed. A leading minus sign will be displayed for negative voltages.
7. Because of the very low signal levels involved, unwanted no,se, as CleSCrlDea I” paragrapn z.11, may upset the ac-
curacy of the measurement.
since the decimal point is
^
Figure 2-4. Basic Voltage Measurements
NANOVOLT AND MICROVOLT MEASUREMENTS
2.6
The Model 161 may be used to make very low voltage
readings down to a resolution of 10nV. These readings are made on one of the mV ranges by using the mV INPUT on
the front panel.
The following paragraphs describe the basic procedure for making these measurements.
1. Turn on the Model 181 with the front panel POWER switch. Allow the unit to warm-up for at least an hour for rated accuracy. To guarantee low drift, allow at least four hours.
2. Select the desired mV range with the appropriate front panel switch. Use a range appropriate for the voltage to be measured.
3. Select other parameters such as HI RES, DAMPING, FILTER, and ZERO as needed. Refer to paragraph 2.8 for more details on these controls.
2-4
I
Figure 2-5. mV and nV Measurements
2.7 SPECIAL MEASURING SITUATIONS
Some situations may call for a wide range of voltage measurements that neither the V input nor mV input can handle alone. In those cases, it may be convenient to use a
common ground for both the V and mV inputs. Since the
LO terminals of the mV and V inputs are internally con­nected together, it is only necessary to connect the mV Lo
terminal (black lead of the Model 1506 low-thermal cable) to
common of the circuit under test, as shown in Figure 2-6. Using this method, either the V HI or mV HI terminal can be used as the test probe, depending on the voltage to be measured.
CAUTION
Do not exceed the maximum input limit for the Model 181, especially when the mV HI terminal is connected. or damage to the in-
strument may occur. Never parallel the mV
and V leads to prevent accidental overload
to the mV input or inadvertent loading of
the circuit under test.
Figure 2-6. Common Ground Connection for V and mV
The zero function is especially useful for nulling out offset voltages, including internal offsets of the Model 181. To use the zero in this manner, short the test leads together with the instrument on the desired range and depress the ZERO switch; the ZERO indicator light should turn on. This stores the residual voltage level as the baseline. All voltage reading taken with zero enabled will then be the actual voltage level since the unwanted voltage will be subtracted from the reading.
Note that baseline suppression for the V and mV ranges operates separately. Switching the unit between a mV and V range, for example, will cancel the ZERO, also causing the front panel ZERO indicator light to turn off.
Controlling the Filter. The Model 181 has an internal 3-p& digital filter that can be controlled by the front panel FILTER and DAMPING controls. Normally, the filter is switched on and off as a function of the rate of change in input signal. Depressing the FILTER button increases the RC time conk stant of the filter. At the same time, the front panel FILTER light will turn on. The digital filter cannot be totally disabled by the front panel controls. However, it may be disabled by commands given over the IEEE bus. Operating with the filter disabled allows the user to customize Model 161 response by using external filtering. For further information on IEEE commands that control the filter, consult Section 4 of this manual.
2.8 ADDITIONAL FRONT PANEL CONTROLS The Model 181 has additional front panel controls that can
be used to enhance the capabilities of the unit. These
switches which include HI RES, ZERO, FILTER, and DAMP-
ING, are shown in Figure 2-2. The following paragraphs will
describe the operation of these controls in more detail.
HI RES. The display resolution of the Model 181 upon power-up is 5% digits. The display resolution may be in­creased to 6 % digits by depressing the HI RES switch. Once the unit is in the 6% digit mode, the display may be returned to the 5% digit mode by depressing the HI RES switch a second time. Readings made in the 5% digit mode have the
least significant digit rounded off. HI RES switch affects
only the data on the display; data transmitted over the IEEE
bus always contains 6% digit information. For further infor-
mation on IEEE operation, refer to Section 4.
Zero. The Zero mode serves as a means for baseline sup-
pression. The front panel ZERO indicator light will turn on
when the zero mode is enabled. All readings taken with the
zero enabled will be the difference between the stored
baseline and the actual voltage level.
The baseline is obtained by connecting the instrument to the voltage to be zeroed. For example, if the baseline
voltage is IOmV, all subsequent readings will have 1OmV
subtracted from the actual voltage level.
The DAMPING button controls whether or not the filter is continuously enabled. When the DAMPING is off, the microprocessor automatically disables the filter when the input voltage changes to permit rapid display update. Once the reading is within 25 digits of the final value on the 2mV range, and within 6 digits on the remaining ranges. the microprocessor then enables the filter to minimize noise in the final reading. When the DAMPING is on. the digital filter is permanently enabled. The unit would normally be operated in this mode only for signals that vary slowlv, or with extremely noisy ambient signals.
Through careful use of the FILTER and DAMPING controls, the user can optimize the Model 161 to the required perfw mance, keeping in mind the resulting speed/ noise com­promises. Figure 2-7 shows four curves resulting from operating the unit with various combinations of the DAMPS
ING and FILTER controls. Curve A shows the fastest response time because the filter RC time constant is at a minimum. Also, with DAMPING off, the microprocessor initially disables the filter as previously described.
Depressing the FILTER switch as with curve 6, has little effect on the response time since the filter is initially off.
Curves C and D, on the other hand, show that enabling the
DAMPING slows the response down considerably. This can be seen in more detail in Table 2-2, which lists the settling times of the various control combinations.
2-5
Table 2-2 Settling Times
~~~~~-~~~~“~~~~~
(The readings all settle to within 0.002% of the Full Range in the specified time.)
2.9 USING THE ANALOG OUTPUT
The analog output of the Model 181 is useful for monitoring
the input signal with an external device such as a chart recorder. The analog signal is reconstructed from digital data (supplied by the internal microprocessor) by a 12 bit
D/A converter. Because of this configuration, the analog output will accurately reflect the display until an overflow condition is reached. The analog output is optically isolated from the front panel LO terminal to avoid potential ground loop problems. The following paragraphs describe the basic procedure for using the analog output.
1. Connect the measuring device to the two analog output terminals on the rear panel as shown in Figure 2-8.
CAUTION
The potential between the analog output
LO terminal and ground must not exceed
30V. Make sure the external device does
not exceed this voltage on its common or ground connections. Failure to observe this precaution may damage the Model
181. possibly voiding the warranty. IEEE
common is connected to analog output
IOW.
Select the Xl or Xl000 range by using the analog output gain switch on the rear panel. This switch is combined
with those used to set the IEEE mode in the lower left cor-
ner of the rear panel and is clearly marked. (See Figure
2.3.) In the Xl position, the most significant +2000 counts of the display reading can be covered, while the Xl000 position will change the range to cover the least significant f2000 counts. In this manner, the entire 6% digits of the display may be represented.
If necessary, the analog output may be zeroed with the front panel ZERO control. To do so, depress the ZERO button.
The Model 181 will display an “OFLO” message when the
capability of a specific range is exceeded. When this message is displayed. the analog output value will be + 2V if the polarity of the input voltage is positive, and -2V if the input voltage polarity is negative.
An analog output range overflow can occur when the Model
181 analog range switch is in the Xl000 position. An exam­ple of the analog ouput voltage under these conditions is
shown in Figure 2-9.” The conditions shown are for the 2mV
range. but the output will react similarly on the other voltage ranges if the proper scaling factor is applied. For each ten-
fold increase in voltage range, the scale of the horizontal
axis must also be multiplied by a factor of ten.
2-6
Figure 2-8. Analog Output Connections
The horizontal axis of Figure 2-9 has an input voltage range between -10&V and +lOpV. The vertical axis shows an analog output voltage between -2V and +ZV. which is the maximum range of the analog output. Beginning at the OV point on the graph, the analog output follows the input voltage linearly until the input voltage reaches +2pV. The analog output will then suddenly switch to the maximum negative output value of -2V. Thus, for each 4uV increment
* Units with B-7 software.
in input voltage, the output pattern repeats. ihe action of the analog output for negative input voltages is the same, except that the slope of the graph is negative for these negative-going input voltages.
Figure 2-9. Xl000 Analog Output
By counting the number of repeating waveforms on a chart
recorder, the user can easily determine the actual voltage at the input, even though the range of the analog output was exceeded. If, for example, the +lV point on the second
peak with a positive-going slope is noted, it can be clearly determined that the input voltage was +5@V at that par­ticular time.
A summary of analog output information is shown in Table
2.3. Each range of input values corresponds to the incre-
ment necessary to cause the output to go through its entire 0 to 2V range. Note that the sensitivity is increased by a fac-
tor of a thousand on the Xl000 range. For example, when
the Model 181 is in the 200mV range, and the analog switch
is in the Xl position, the output voltage will swing from 0 to 2V in a smooth manner as the input voltage increases
gradually from 0 to 200mV. When the analog output switch
is changed to the Xl000 position, the input need only swing
between 0 and 2OOpV to obtain the same voltage swing at
the analog output. Beyond those input limitations, the out-
put voltage will repeat as shown in Figure 2-9.
The output resistance of the analog ouput is Ikll for all
voltage ranges regardless of the position of the analog range
switch. Thus, loading problems caused by external devices
are minimized. To keep loading errors below I%, the input
resistance of any device connected to the analog output
should be greater than lOOk0.
Table 2-3. Analog Output Parameters
2.10 SOURCE RESISTANCE CONSIDERATION The Model 181 has an input resistance greater that IGIl
flO% on the 2mV. 20mV. 200mV. and 2V ranges. The in-
strument will meet this input resistance specification on the mV ranges even when in overflow with voltages up to 1V. The input resistance on the remaining voltages ranges is
lOML2. Thus, the Model 181 input resistance is sufficiently
high to minimize loading errors in most measuring situa­tions. For voltage sources with very high source resistance, two precautions should be observed when using the Model
181.
Shielding becomes more critical when the source resistance is very high. Otherwise, interference signals may be picked up by the test leads. Noise picked up in this manner can af­fect the mV ranges more severely. but shielding might be necessary for connections to the V INPUT in extreme situations.
Loading of the voltage source by the Model 181 can become
important with high source resistance values. As the source resistance increases, the error due to loading increases.
Figure Z-10 shows the method used to determine the Peru
cent error due to loading. The voltage source has an internal resistance R,, while the internal resistance of the Model 181 is represented by R,. The source voltage is E, while the voltage actually measured by the meter is E,.
The voltage actually seen by the meter is attenuated by the voltage-divider action of R and R, and can be found by using the relationship: E, = &R,IIRL + I?,).
We can modify this relationship to obtain a formula for per-
cent errors as follows: Percent Error = lOOR,/(R, * R,i~
From the above, it is obvious that the input resistance of the Model 181 must be at least 99 times greater that the source resistance if the loading error is to be kept to 1%. This max­imum 1% error limitation will be achieved on the 2mV through 2V ranges with sources resistances up to lO.lMI1, while the source resistance should be no greater than IOlklI if the same 1% error limitation is to be maintained on the 2OV through IOOOV ranges. If lower errors are required, the source resistance must be correspondingly less.
Rs
20mV
200mV
2v
20 v
200 v
1 kV
*IV Full Range Maximum
INPUT FOR
IG OUTPUT
Figure 2-10. Source Resistance Considerations
2-7
Loading...
+ 30 hidden pages