JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-5866
1-800-554-JUMO
Fax: 315-697-5867
E-mail: info@jumo.us
Internet: www.jumo.us
JUMO Wtrans B
Transmission unit
Transmission
frequency
868.4 MHz
Head transmitter
Power supply
Lithium battery
3.6 V, 2.2 Ah
Interface
Setup
Analog input
+
-
RTD
temperature probe
Thermocouple
Resistance /
Potentiometer
Voltage
Resistance transmitter
Programmable Head Transmitter
with Radio Transmission
Brief description
The Wtrans B head transmitter with wireless data transmission is used in connection with a
Wtrans receiver unit for stationary or mobile temperature measuring with RTD temperature
probes or thermocouples. Alternatively, resistors with an ohmic-value of up to 10 kΩ, voltages up
to 50 mV, and currents up to 20 mA that use an external shunt are measurable. Customer-specific linearizations are possible.
The measured values are transmitted wirelessly to the receiver of the Wtrans measuring system.
These values are shown on the receiver's LCD display and are digitally available on the RS485interface or as analog output(s). Optionally, different alarm types can be indicated with the two
relay outputs.
This head transmitter is designed for industrial applications – it is made up of the transmitter with
an integrated transmission unit and an antenna-battery case. The transmitter is designed to be
installed into type B terminal heads and works in an ambient temperature range from -30 to
85 °C. It is possible to install the transmitter into customer-specific terminal heads. The antennabattery case is connected to the terminal head through a screw connection (M20 × 1.5).
The transmission frequency of the Wtrans measuring system is 868.4 MHz. This frequency is unaffected for the most part by external interferences and it allows for transmissions even in a rough
industrial environment. When using the antenna wall-holder with the 3 m antenna line for the receiver then the maximum open-air range is 300 m.
A 3.6 V, 2.2 Ah (size AA) lithium battery is used as a power supply for the transmitter. For the
easy configuration and parameterization of the head transmitter and the Wtrans receiver on a PC
an optional setup program is available. Additionally, measured values can be recorded on the PC
with the online-chart function.
Data Sheet 707060
Seite 1/10
Block diagram
2011-09-30/00532898
Special features
• Transmission frequency: 868.4 MHz
• Measuring input for RTD temperature
probe, thermocouple, potentiometer and
voltage
• Configurable probe ID
• Battery status monitoring
• Easy battery change
• Configurable via setup program
• Customer-specific linearization (value pairs
in tables or fourth order number), can be
configured on the receiver
• Online chart of the measured values via setup program and receiver
Type 707060/...
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-5866
1-800-554-JUMO
Fax: 315-697-5867
E-mail: info@jumo.us
Internet: www.jumo.us
JUMO Wtrans
Data Sheet 707060 Seite 2/10
The JUMO Wtrans device series also includes the receivers 902931/... and the probes 902930/... in addition to the head transmitter 707060/....
JUMO Wtrans - receiver/probe
Data sheet
JUMO Wtrans receiver
Type T01.EC1
Type T01.EC3
JUMO Wtrans probe
Type T01.G1
Type T02.G1
JUMO Wtrans probe
Type T01.G1
Type T02.G1
JUMO Wtrans probe
Type T01.G2
Type T02.G2
Receiver (voltage supply AC 110 to 240 V or AC/DC 20 to 30 V)
4x analog outputs
2x analog outputs and 2x relay outputs
Insertion RTD temperature probe
Ambient temperature, handle: -30 to +85 °C
Ambient temperature, handle: -25 to +125 °C
Mineral insulated RTD temperature probe with flexible sheath
Ambient temperature, handle: -30 to +85 °C
Ambient temperature, handle: -25 to +125 °C
With M12 connector for RTD temperature probe
Ambient temperature, handle: -30 to +85 °C
Ambient temperature, handle: -25 to +125 °C
JUMO Wtrans B Programmable head transmitter with radio transmission 707060
902931
902930
902930
902930
2011-09-30/00532898
Data Sheet 707060 Seite 3/10
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-5866
1-800-554-JUMO
Fax: 315-697-5867
E-mail: info@jumo.us
Internet: www.jumo.us
Technical data
Analog input
RTD temperature probe
Designation Standard Measuring range Measuring accuracy
Pt100 DIN EN 60751 -100 to +200 °C ± 0.1 K
(TK value = 3.85 × 10
Pt500 DIN EN 60751 -100 to +200 °C ± 0.1 K
(TK value = 3.85 × 10
Pt1000 DIN EN 60751 -100 to +200 °C ± 0.1 K
(TK value = 3.85 × 10
Ni 100 DIN 43760 -60 to +250 °C ± 0.2 K
(TK value = 6.18 × 10
Ni 500 DIN 43760 -60 to +150 °C ± 0.2 K
(TK value = 6.18 × 10
Ni 1000 DIN 43760 -60 to +150 °C ± 0.2 K
(TK value = 6.18 × 10
Pt100 JIS 1604 -100 to +200 °C ± 0.1 K
(TK value = 3.917 × 10
Pt50 ST RGW 1057 1985 -200 to +600 °C ± 0.2 K
(TK value = 3.91 × 10
Pt100 GOST 6651-94 A.1 -100 to +200 °C ± 0.1 K
(TK value = 3.91 × 10
Cu50 GOST 6651-94 A.4 -50 to +200 °C ± 0.2 K
(TK value = 4.26 × 10
Cu100 GOST 6651-94 A.4 -50 to +200 °C ± 0.2 K
(TK value = 4.26 × 10
Connection type 2-wire or 3-wire circuit
Sensor current < 0.5 mA
Lead compensation Not required for 3-wire circuit (max. admissible 11 Ω per line).
-3
1/K) -200 to +600 °C ± 0.2 K
-3
1/K) -200 to +600 °C ± 0.2 K
-3
1/K) -200 to +600 °C ± 0.2 K
-3
1/K)
-3
1/K)
-3
1/K)
-3
1/K) -200 to +600 °C ± 0.2 K
-3
1/K)
-3
1/K) -200 to +600 °C ± 0.2 K
-3
1/K)
-3
1/K)
Max. configurable lead wire resistance for 2-wire circuit: 22 Ω.
Thermocouples
Designation Standard Measuring range Measuring accuracy
Fe-CuNi „L“ DIN 43710 -200 to +900 °C ± 0.1 %
Fe-CuNi „J“ DIN EN 60584 -210 to +1200 °C ± 0.1 % as of -100 °C
Cu-CuNi „U“ DIN 43710 -200 to +600 °C ± 0.1 % as of -100 °C
Cu-CuNi T DIN EN 60584 -270 to +400 °C ± 0.1 % as of -150 °C
NiCr-Ni K DIN EN 60584 -270 to +1372 °C ± 0.1 % as of -80 °C
NiCr-CuNi „E“ DIN EN 60584 -270 to +1000 °C ± 0.1 % as of -80 °C
NiCrSi-NiSi „N“ DIN EN 60584 -270 to +1300 °C ± 0.1 % as of -80 °C
Pt10Rh-Pt „S“ DIN EN 60584 -50 to +1768 °C ± 0.15 % as of 20 °C
Pt13Rh-Pt „R“ DIN EN 60584 -50 to +1768 °C ± 0.15 % as of 50 °C
Pt30Rh-Pt6Rh „B“ DIN EN 60584 0 to 1820 °C ± 0.15 % as of 400 °C
W5Re-W26Re „C“ 0 to 2320 °C ± 0.15 %
W3Re-W25Re „D“ 0 to 2495 °C ± 0.25 %
W3Re-W26Re 0 to 2400 °C ± 0.15 %
Chromel-Copel -200 to +800 °C ± 0.1 % as of -80 °C
Chromel-Alumel -200 to +1372 °C ± 0.1 % as of -80 °C
PLII (Platinel II) 0 to 1395 °C ± 0.15 %
2011-09-30/00532898
a