Invacare ISA COMPACT, ISA STANDARD, ISA XPLUS, ISA PLUS User Manual

Invacare®ISA™
ISA™COMPACT,ISA™STANDARD,ISA™PLUS,ISA™XPLUS
enStandassistpatientlift
UserManual...................................3
deAufstehlifter
Gebrauchsanweisung...........................33
esGrúadebipedestaciónparapacientes
Manualdelusuario.............................65
Manueld'utilisation............................97
itSollevatoreperpazientiStandAssist
Manualed’uso.................................129
nlActievepatiëntentillift
Gebruiksaanwijzing.............................161
ptElevadordetransferênciaposiçãovertical
Manualdeutilização...........................193
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,thismanualMUSTbereadandsaved forfuturereference.
©2019InvacareCorporation Allrightsreserved.Republication,duplicationormodicationinwholeorinpartisprohibitedwithoutpriorwritten permissionfromInvacare.Trademarksareidentiedby™and®.AlltrademarksareownedbyorlicensedtoInvacare Corporationoritssubsidiariesunlessotherwisenoted. – AlleRechtevorbehalten.DieWeiterveröffentlichung,VervielfältigungoderÄnderungimGanzenoderinT eilenohne vorherigeschriftlicheGenehmigungvonInvacareistuntersagt.Markensinddurch™und®gekennzeichnet.Soweit nichtandersangegeben,sindalleMarkenEigentumderInvacareCorporationbzw.derenTochtergesellschaften oderwerdenvondieseninLizenzgenutzt. – Todoslosderechosreservados.Seprohíbevolverapublicar ,copiaromodicarelpresentedocumento,enparteopor completo,sinelprevioconsentimientoporescritodeInvacare.Lasmarcascomercialesseidenticancon™y®.T odas lasmarcascomercialessonpropiedaddeInvacareCorporationodesusliales,obienInvacareCorporationosus lialescuentanconlascorrespondienteslicencias,salvoqueseindiquelocontrario. – Tousdroitsréservés.Larepublication,laduplicationoulamodicationdetoutoupartieduprésentdocumentest interditesansl'accordécritpréalabled'Invacare.Lesmarquescommercialessontidentiéespar™et®.Toutes lesmarquescommercialessontdétenuesparoucédéessouslicenceàInvacareCorporationousesliales,sauf stipulationcontraire. – Tuttiidirittiriservati.Èproibitalariproduzione,laduplicazioneolamodicaparzialeocompleta,salvoprevia autorizzazionescrittadapartediInvacare.Imarchisonocontrassegnatida™e®.Tuttiimarchisonodiproprietào licenzadiInvacareCorporationodisueafliate,salvoindicazionecontraria. – Allerechtenvoorbehouden.Herpublicatie,duplicatieofgeheleofgedeeltelijkewijzigingisverbodenzonder voorafgaandeschriftelijketoestemmingvanInvacare.Handelsmerkenzijnteherkennenaan™en®.Allehandelsmerken zijneigendomvanofgelicentieerdaanInvacareCorporationofhaardochterondernemingen,tenzijandersaangegeven. – Todososdireitosreservados.Arepublicação,duplicaçãooumodicaçãototalouparcialestáinterditasema autorizaçãopréviaporescritodaInvacare.Asmarcascomerciaissãoidenticadaspelossímbolos™e®.Todasas marcascomerciaissãopropriedadedaouestãolicenciadasàInvacareCorporationouàssuassubsidiárias,exceto quandoapresentadainformaçãoemcontrário.
Contents
ThismanualMUSTbegiventotheuseroftheproduct. BEFOREusingthisproduct,thismanualMUSTbereadandsaved forfuturereference.
1General.........................................4
1.1Introduction..................................4
1.1.1Symbolsinthisdocument.....................4
1.2Servicelife...................................4
1.2.1Additionalinformation........................4
1.3Limitationofliability............................4
1.4Warrantyinformation...........................4
1.5Compliance...................................4
1.5.1Product-specicstandards.....................4
2Safety..........................................5
2.1Generalsafetyinformation.......................5
2.1.1Pinchpoints................................6
2.2Safetyinformationonaccessories..................6
2.3Safetyinformationonelectromagneticinterference.....6
2.3.1Batterychargingmode........................6
2.4Labelsandsymbolsontheproduct.................7
2.4.1Labellocation..............................7
2.4.2Identicationlabel...........................7
2.4.3Othersymbols..............................7
3ProductOverview.................................8
3.1Intendeduse..................................8
3.2Mainpartsofthelift............................8
3.3Accessories...................................9
4Setup...........................................10
4.1Generalsafetyinformation.......................10
4.2Scopeofdelivery..............................10
4.3Installingthemasttothebase....................10
4.4Installingthefootplate..........................11
4.5Installingthelegband...........................11
4.6Installingthelegsupport.........................11
4.7Installingtheleverformanuallegspreader...........11
4.8Resettingtheserviceindicator.....................12
5Usage..........................................13
5.1Generalsafetyinformation.......................13
5.2Locking/Unlockingtherearcastors.................13
5.3Raising/LoweringanElectricLift...................13
5.4Closing/OpeningLegs...........................13
5.4.1Closing/OpeningElectricLegs...................13
5.4.2Closing/Openinglegsmanually..................13
5.5Adjustingtheliftingarmtopatientbodyheight........13
5.6Adjustingtheheightofthelegsupport..............14
5.7Usingthelegband.............................14
5.8Emergencyfunctions............................15
5.8.1Performinganemergencystop..................15
5.8.2Activatinganemergencylowering(CBJHome
controlunit)...............................15
5.8.3Activatinganemergencylowering(CBJCare,CBJ1,
CBJ2controlunit)...........................15
5.8.4Activatinganemergencylifting(CBJCare,CBJ1
controlunit)...............................15
5.8.5Activatingamechanicalemergencylowering.......15
5.9Chargingthebattery............................16
5.9.1CBJHomecontrolunit........................16
5.9.2CBJCare,CBJ1,CBJ2controlunit................16
5.9.3Optionalbatterycharger......................17
6PatientTransfer...................................18
6.1Generalsafetyinformation.......................18
6.2Attachingtheslingtothelift......................19
6.3Transferringapatient...........................20
7TransportationandStorage..........................21
7.1Generalinformation............................21
7.2Disassemblingthelift...........................21
8Maintenance.....................................22
8.1Generalmaintenanceinformation..................22
8.2Dailyinspections...............................22
8.3CleaningandDisinfection........................22
8.3.1Generalsafetyinformation.....................22
8.3.2CleaningIntervals............................22
8.3.3Cleaninginstructions.........................22
8.3.4Disinfectioninstructions.......................23
8.4Serviceinterval................................23
8.4.1LOLERStatement............................23
9AfterUse........................................24
9.1Disposal.....................................24
9.2Reconditioning................................24
10Troubleshooting..................................25
10.1Identifyingfaultsandpossiblesolutions.............25
11T echnicalData...................................27
11.1Maximumsafeworkingload.....................27
11.2Dimensionsandweights........................27
11.3Electricalsystem..............................28
11.4Environmentalconditions........................28
11.5Materials....................................29
11.6Operatingforcesofcontrols......................29
12Electromagneticcompatibility(EMC)..................30
12.1GeneralEMCinformation.......................30
12.2Electromagneticemission........................30
12.3ElectromagneticImmunity.......................30
Invacare®ISA™

1General

1.1Introduction

Thisusermanualcontainsimportantinformationaboutthe handlingoftheproduct.T oensuresafetywhenusingthe product,readtheusermanualcarefullyandfollowthe safetyinstructions.
Notethattheremaybesectionsinthisdocument,whichare notrelevanttoyourproduct,sincethisdocumentapplies toallavailablemodels(onthedateofprinting).Ifnot otherwisestated,eachsectioninthisdocumentreferstoall modelsoftheproduct.
Themodelsandcongurationsavailableinyourcountrycan befoundinthecountry-specicpricelists.
Invacarereservestherighttoalterproductspecications withoutfurthernotice.
Beforereadingthisdocument,makesureyouhavethe latestversion.YoundthelatestversionasaPDFonthe Invacarewebsite.
Ifyoundthatthefontsizeintheprinteddocumentis difculttoread,youcandownloadthePDFversionfromthe website.ThePDFcanthenbescaledonscreentoafont sizethatismorecomfortableforyou.
Formoreinformationabouttheproduct,forexampleproduct safetynoticesandproductrecalls,contactyourInvacare representative.Seeaddressesattheendofthisdocument.
Incaseofaseriousincidentwiththeproduct,youshould informthemanufacturerandthecompetentauthorityin yourcountry.
maintenanceintervalsandcorrectuse,statedinthismanual. Theeffectiveservicelifecanvaryaccordingtofrequencyand intensityofuse.
1.2.1Additionalinformation
Theexpectedservicelifeisbasedonanestimatedaverage of4liftingcyclesperday.

1.3Limitationofliability

Invacareacceptsnoliabilityfordamagearisingfrom:
Non-compliancewiththeusermanual
Incorrectuse
Naturalwearandtear
Incorrectassemblyorset-upbythepurchaserorathird party
Technicalmodications
Unauthorisedmodicationsand/oruseofunsuitable spareparts

1.4Warrantyinformation

Weprovideamanufacturer’swarrantyfortheproduct inaccordancewithourGeneralTermsandConditionsof Businessintherespectivecountries.
Warrantyclaimscanonlybemadethroughtheprovider fromwhomtheproductwasobtained.

1.5Compliance

Qualityisfundamentaltothecompany’soperation,working withinthedisciplinesofISO13485.
1.1.1Symbolsinthisdocument
Symbolsandsignalwordsareusedinthisdocumentand applytohazardsorunsafepracticeswhichcouldresultin personalinjuryorpropertydamage.Seetheinformation belowfordenitionsofthesignalwords.
WARNING Indicatesahazardoussituationthatcouldresult inseriousinjuryordeathifitisnotavoided.
CAUTION Indicatesahazardoussituationthatcouldresult inminororslightinjuryifitisnotavoided.
IMPORTANT Indicatesahazardoussituationthatcouldresult indamagetopropertyifitisnotavoided.
TipsandRecommendations Givesusefultips,recommendationsand informationforefcient,trouble-freeuse.

1.2Servicelife

Theexpectedservicelifeofthisproductiseightyearswhen useddailyandinaccordancewiththesafetyinstructions,
ThisproductfeaturestheCEmark,incompliancewiththe MedicalDeviceRegulation2017/745Class1.Thelaunchdate ofthisproductisstatedintheCEdeclarationofconformity.
Wearecontinuouslyworkingtowardsensuringthatthe company’simpactontheenvironment,locallyandglobally , isreducedtoaminimum.
WeonlyuseREACHcompliantmaterialsandcomponents.
WecomplywiththecurrentenvironmentallegislationsWEEE andRoHS.
1.5.1Product-specicstandards
TheproducthasbeentestedandconformstoISO10535 (Hoistsforthetransferofdisabledpersons)andallrelated standards.
Forfurtherinformationaboutlocalstandardsandregulations, contactyourlocalInvacarerepresentative.Seeaddresses attheendofthisdocument.
4
1654371-A
Safety

2Safety

2.1Generalsafetyinformation

Thissectionofthemanualcontainsgeneralsafety informationaboutyourproduct.Forspecicsafety information,refertotheappropriatesectionofthemanual andprocedureswithinthatsection.
WARNING! Riskofinjuryordamage
–Donotusethisproductoranyavailableoptional
equipmentwithoutrstcompletelyreading andunderstandingtheseinstructionsandany additionalinstructionalmaterialsuchasuser manualsorinstructionsheetssuppliedwith thisproductoroptionalequipment.Ifyouare unabletounderstandthewarnings,cautionsor instructions,contactahealthcareprofessional, Invacareproviderorqualiedtechnicianbefore attemptingtousethisproduct.
–Donotmakeanyunauthorizedalterationsor
modicationstotheproduct.
WARNING! Themaximumsafeworkingloadmustnotbe exceeded
–Donotexceedthemaximumsafeworkingload
ofthisproductorusedaccessorieslikeslings etc.Seedocumentationorthelabelingforthe statedmaximumsafeworkingload.
–Thecomponentwiththelowestloadlimit
determinesthemaximumsafeworkingloadof theentiresystem.
WARNING! Riskofinjuryordamage
Improperuseofthisproductmaycauseinjury ordamage.
–Donotattemptanytransferwithoutapproval
ofthepatient’shealthcareprofessional.
–Readtheinstructionsinthisusermanualand
observetrainedpersonnelperformingtransfer procedures.Thenpracticetransfersunder supervisionandwithacapablepersonacting asapatient.
–Patientswhoaretransferredandpositioned
usingthestandassistliftMUSTbecooperative, coherentandhaveheadandneckcontrol. Otherwise,injurymayoccur .
–Patientswhoaretransferredandpositioned
usingaslingwithbacksupportonly,MUSTbe abletosupportamajorityoftheirownweight. Otherwise,injuryordamagemayoccur .Refer totheslingusermanualformoreinformation.
–Donotusetheliftasatransportdevice.It
isintendedtotransferanindividualfromone seatingsurfacetoanother.
WARNING! Riskofinjuryordamage
Inappropriatehandlingofcables,incorrect connectionsanduseofunauthorizedequipment cancauseelectricalshockandproductfailure.
–Donotkink,shearorotherwisedamagethe
cablesoftheproduct.
–Makesurethatnocablesarejammedor
damaged,whentheproductisused.
–Makesurethewiringiscorrectandconnections
areproper.
–Donotuseunauthorizedequipment.
WARNING! Riskofinjuryordamage
Excessivemoisturewilldamagetheproductand maycauseelectricalshock.
–Thepatientliftcanbeusedinabathorshower
areabutmustNOTbeusedundertheshower. Thepatientmustbetransferredtoashower chairoruseothermeansforshowering.
–Ifthepatientliftisusedinamoisture
environment,ensurethatthepatientliftis wipedcleanofanymoistureafteruse.
–Donotplugorunplugthepowercableina
moistureenvironmentorwithwethands.
–Donotstoretheproductinadampareaor
dampcondition.
–Periodicallyinspectallcomponentsofthe
productforsignsofcorrosionordamage. Replacepartsthatarecorrodedordamaged.
–See11.4Environmentalconditions,page28
WARNING! Riskofinjuryordamage
Ignitionsourcescancauseburnsorre.
–Patienttransfermustbeperformedwitha
safetyspacebetweentheliftandpossible ignitionsources(heater,stove,replace,etc.)
–Thepatientandassistantsmustnotsmoke
duringtransfer.
–Theslingmustnotbeplacedoverheatsources
(heater,stove,replace,etc.)
WARNING! Riskofinjuryordamage
Toavoidinjuryordamagewhenoperatingthe product:
–Closesupervisionisnecessarywhentheproduct
isusednearchildrenorpets.
–Donotletchildrenplaywiththeproduct.
CAUTION! Riskofinjuryordamage
Theproductcangethotwhenexposedtosunlight orotherheatsources.
–Donotexposetheproducttodirectsunlightfor
prolongedperiods.
–Keeptheproductawayfromheatsources.
.
IMPORTANT!
Accumulationoflint,dustandotherdirtcan impairtheproduct.
–Keeptheproductclean.
1654371-A5
Invacare®ISA™
2.1.1Pinchpoints
WARNING! RiskofInjury
Pinchpointsarepresentinseverallocationson theliftandbodypartscouldbepinched.
–Alwayskeephands,ngersandfeetclearof
movingparts.

2.2Safetyinformationonaccessories

CAUTION! Riskofinjury
Non-originalorincorrectaccessoriesmayaffect thefunctionandsafetyofthisproduct.
–Duetoregionaldifferences,refertoyour
localInvacarewebsiteorcatalogforavailable accessoriesorcontactyourInvacareprovider .
–Seethemanualdeliveredwiththeaccessory
forfurtherinformationandinstructions.
–Useonlyoriginalaccessoriesfortheproduct
inuse.Undercertaincircumstances,theuse slingsfromothermanufacturersispossible.See additionalinformationinthissection.
–Ariskassessmentisalwaystobecarriedoutby
aprofessionalpriortoissuingliftingequipment. ItisimportantthattheT ask,Individual,Load, EnvironmentandEquipmentareconsideredin theriskassessment.
–Alwayschoosetheslingdesignandsize
accordingtothepatient’sweight,sizeand physicalabilitywhilstconsideringthetypeof transfertobecarriedout.
CAUTION! Compatibilityofslingswithattachmentsystem
Invacareusesacommonattachmentsystem basedonhooksandloops.Loopsontheslings areattachedtohooksonthelift.Therefore, suitableslingsofothermanufacturerscanbe usedonthisliftaswell.
–Useonlyslingswhichareintendedforstand
assistlifts.
–Useonlyslingswithloopattachmentsthatare
suitableforhooksasattachmentpoints.
–Donotuseslingsdesignedfor"Keyholeor
ClipAttachmentSystems"or"TiltingFrame Systems".

2.3Safetyinformationon electromagneticinterference

WARNING! Riskofmalfunctionduetoelectromagnetic interference
Electromagneticinterferencebetweenthis productandotherelectricalequipmentcanoccur anddisturbtheelectricaladjustmentfunctions ofthisproduct.Toprevent,reduceoreliminate suchelectromagneticinterference:
–Onlyuseoriginalcables,accessoriesandspare
parts,tonotincreaseelectromagneticemission orreduceelectromagneticimmunityofthis product.
–Donotuseportableradiofrequency(RF)
communicationsequipmentcloserthan30cm toanypartofthisproduct(includingcables).
–Donotusethisproductnearactive
high-frequencysurgicalequipmentandthe RFshieldedroomofasystemformagnetic resonanceimaging,wheretheintensityof electromagneticdisturbancesishigh.
–Ifdisturbancesoccur,increasethedistance
betweenthisproductandtheotherequipment orswitchitoff.
–Refertothedetailedinformationandfollow
theguidanceinchapter12Electromagnetic
compatibility(EMC),page30
WARNING! Riskofmalfunction
Electromagneticinterferencemaycauseimproper operation.
–Donotusethisproductadjacenttoorstacked
withotherelectricalequipment.Ifsuchuseis necessarythisproductandtheotherequipment mustbecloselyobservedtoverifythatthey areoperatingnormally.
2.3.1Batterychargingmode
IMPORTANT!
Duringbatterychargingthecontrolunitis moresusceptiblefordamagebyanElectrostatic Discharge(ESD).TheoccurrenceofanESDduring batterychargingmodemayleadtocontrolunit malfunctionorlossoffunction.
–AvoidanypossiblesourcesofESDduring
batterycharging.
.
61654371-A

2.4Labelsandsymbolsontheproduct

Product Name
YYYY-MM
= xxx kg
Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio, Portugal
XXXXXX XXXXX_XX
XXXXXXXXXXXXX
Xxxxxxx xxxxx xxx Xxxxxxx xxxxx xxx Xxxxxxx xxxxx xxx Xxxxxxx xxxxx xxx Xxxxxxx xxxxx xxx
ISO 10535
1
9
2.4.1Labellocation
Max.SafeWorkingLoad
CLASSIIequipment
TypeBAppliedPart
WEEEconform
Europeanconformity
Abbreviationsfortechnicaldata:
Iin=IncomingCurrent
Uin=IncomingVoltage
Int.=Intermittence
AC=AlternatingCurrent
Max=maximum
min=minute
Formoreinformationabouttechnicaldata,see11T echnical Data,page27.
2.4.3Othersymbols
Safety
Identicationlabel
A
Readusermanuallabel
B
Nameandsafeworkingloadoflift–depending
C
onmodel
Environmentalconditionsandtotalmaximum
D
weightlabel
Safeworkingloadoflift–dependingonmodel
E
Warrantyvoidlabel–notpresentonallmodels
F
Warninglabel
G
Liftingarmadjustmentlabel
H
2.4.2Identicationlabel
Theidenticationlabelcontainsthemainproduct information,includingtechnicaldata.
RefertoUserManual
Totalweightoftheproductwiththe maximumsafeworkingloadapplied
Donotusetheactuatorasapushbar
Donotremove
Warrantyvoidifremovedorbroken
Temperaturelimit
Humiditylimitation
Atmosphericpressurelimitation
Transportandstorageconditions
Symbols
1654371-A
SerialNumber
ReferenceNumber
ManufacturerAddress
ManufacturingDate
Operationconditions
Liftingarmadjustmentincorrelationto thebodyheightofthepatient
7
Invacare®ISA™

3ProductOverview

3.1Intendeduse

Thestandassistpatientliftisabattery-poweredtransfer deviceandisintendedtotransferandpositionanindividual fromoneseatingsurfacetoanother .Forexample:
Betweenthebedandawheelchair
Toandfromthetoilet
Rehabilitation,assistingapatientfromaseatedposition toastandingposition
Themaximumsafeworkingloadisstatedin11Technical
Data,page27
Thestandassistpatientliftisdesignedtobeusedindoorson alevelsurface,inhospitals,nursingfacilitiesanddomestic areas.Thestandassistpatientliftcanbeturned(rotated)in placefortransferswithlimitedoorspace.
Ahealthcareprofessionalorprivatepersonwhohasreceived propertrainingistheintendedoperatorofthisproduct.
Indications
Thestandassistliftisintendedtotransferandposition partiallyimmobileweight-bearingpatients.

3.2Mainpartsofthelift

Contraindications
Thestandassistliftiscontraindicatedforpatientswho:
arenotcooperative
arenotabletounderstandandfollowinstructions
donothaveheadandneckcontrol
donothavereasonabletrunkstability
Someslingsdesignedforthestandassistliftare contraindicatedforpatientswhocannotsupportthemajority oftheirweight.Refertotheslingusermanualformore information.
Boom
A
Hookforsling
B
Handgrip
C
Liftingarm
D
Pushbar
E
Liftingactuator
F
Mast
G
Legsupport(xedorswiveling)
H
Basewithfootpedalformanuallegspreader
I
Footplate(high,loworinclined)
J
Frontcastors
K
Leg
L
Basewithactuatorsforelectricallegspreader—
M
withorwithoutactuatorcovers
Rearcastorswithbrake
N
Handcontrol
O
Emergencystop
P
CBJHomecontrolunitwithintegratedbattery
Q
81654371-A
ProductOverview
CBJCare,CBJ1orCBJ2controlunitwithdetachable
R
battery
Mechanicalemergencylowering
S
Leverformanuallegspreader
T

3.3Accessories

Duetoregionaldifferences,refertoyourlocal Invacarewebsiteorcatalogfordetailsaboutavailable accessoriesorcontactyourInvacareprovider.
Leverformanuallegspreader
Protectivecoversforlegs
Legband
Wallchargerforadetachablebattery
Extrabattery
Extrafootplate(high,loworinclined)
Slingmodelswithloopattachments,suitableforhooksas attachmentpoints:
Invacare®StandAssistsling(Europe)/Invacare®Standup sling(APAC):2–pointattachmentwithbacksupportonly.
Invacare®TransferStandAssistsling:4–pointattachment withbackandlegsupport.
1654371-A9
Invacare®ISA™

4Setup

4.1Generalsafetyinformation

WARNING!
–Checkallpartsforshippingdamagebeforeuse. –Incaseofdamage,donotusetheequipment.
ContactyourInvacareproviderforfurther instructions.
WARNING! RiskofInjury
Improperassemblymaycauseinjuryordamage.
–UseonlyInvacarepartsintheassemblyofthis
patientlift.
–Aftereachassembly ,checkthatallttings
areproperlytightenedandallpartshavethe correctfunction
–Donotovertightenthemountinghardware.
Thismaydamagethemountingbrackets.
Therearenotoolsrequiredtoassemblethepatient lift.
Ifthereareanyissuesorquestionsduringassembly, contactyourInvacareprovider.
UserManual(1piece)
Battery(1or2pieces)*–CBJCare,CBJ1and CBJ2only
Leverformanuallegspreader(1piece)*
Legband(1piece)*
Sling(1piece)*
*Dependsonmodeland/orconguration
Iftheliftisdeliveredwithasling,refertotheuser manualoftheslingforuse,application,maintenance andwashinginstructions.

4.3Installingthemasttothebase

WARNING! RiskofinjuryorDamage
Themaximumsafeworkingloadstatedonthe boomandonthebasemustbethesame.
–Alwayscomparethemaximumsafeworking
loadvaluesstatedonthebaseandonthe boom.
1.

4.2Scopeofdelivery

Theitemsincludedinyourpackagedependonthemodels andcongurationsavailableinyourcountry.See1.1
Introduction,page4
LockbothrearcastorsBandrotatethehandscrews anti-clockwiseAtoremovethemfromthebase
2.
A
B
C
D
E
101654371-A
Mast/boomassembly(1piece)—incl. controlunitandactuator
Basewithorwithoutactuators(1piece)
Legsupport
Footplate
Handcontrol(1piece)
Mainscable(1piece)
Withtheboompointinginforwarddirection,lowerthe mastCintothesupportofbaseDandreinstallthe handscrewsAtoxthemastCtothebase.
Setup

4.4Installingthefootplate

1.LowerthefootplateontothebaseuntilthebracketsA arefullyseatedontubeB.

4.5Installingthelegband

(optional)
Thetwopartsofthelegbandareattachedontheback sideofthelegsupport.
1.

4.6Installingthelegsupport

1.PushandholdupthereleasehandleAontheleg supportB.
2.SlidetheguidebaronthebackofthelegsupportB ontoitsmountingplateConthemast.
3.Whenthedesiredheightisreached,releasethehandle AandslightlymovethelegsupportBupordownuntil anaudibleclickisheardandthelegsupportBislocked intooneofthe6availableheightpositions.
ThreadthetwohooksAintolugsBonthebackside ofthelegsupport.
2.
4.7Installingtheleverformanualleg
spreader
2x13mmwrench
1.InserttheboltAfrombelowintothelowerholeofthe bracketBandxitwiththenutC.
2.InsertthethreadedendofleverDintotheupperhole ofbracketBontothexedboltA.
3.TurntheleverDclockwisetoscrewitontothebolt.
ThreadtheloopsCofthetwolegbandpartsintothe hooksA.
1654371-A
11
Invacare®ISA™

4.8Resettingtheserviceindicator

(CBJCarecontrolunitonly)
IMPORTANT!
Iftheservicelightisashingyellowduringdaily usageorafterareassembly,theliftrequires service.
–ContactyourInvacareproviderforservice.
Aftertheinitialassemblyofthelift,theserviceindicator needstoberesetbeforeusingthelift.
Toperformaninitialresetoftheserviceindicator:
1.Locatethehandcontrol.
2.PressandholdtheUPbuttonandtheDOWNbuttonat thesametimeforveseconds.
3.Youwillhearasoundwhentheserviceindicatorhas beenreset.
12
1654371-A
Usage

5Usage

5.1Generalsafetyinformation

WARNING! Riskofinjuryordamage
Beforeusingtheliftwithapatient,refertothe followingsafetyinformationandinstructions:
2Safety,page56PatientTransfer,page18

5.2Locking/Unlockingtherearcastors

WARNING! RiskofInjuryorDamage
Theliftcouldtipandendangerthepatientand assistants.
–Invacarerecommendsthattherearcastors
beleftunlockedduringliftingproceduresto allowthepatientlifttostabilizeitselfwhenthe patientisinitiallyliftedfromachair,bedorany stationaryobject.
–Invacarerecommendslockingtherearcastors
oftheliftonlywhenpositioningorremoving theslingfromaroundthepatient.

5.4Closing/OpeningLegs

WARNING! RiskofInjury
Theliftcouldtipandendangerthepatientand assistants.
–Thelegsoftheliftmustbeinthemaximum
openpositionforoptimumstabilityandsafety. Ifitisnecessarytoclosethelegsoftheliftto maneuvertheliftunderabed,closethelegsof theliftonlyaslongasittakestopositionthe liftoverthepatientandliftthepatientoffthe surfaceofthebed.Whenthelegsofthelift arenolongerunderthebed,returnthelegsof thelifttothemaximumopenposition.
5.4.1Closing/OpeningElectricLegs
Thehandcontrolisusedtoopenorclosethelegsofthe base.
1.Toclosethelegs,press andholdthelegsclosed buttonA.
2.Toopenthelegs,pressand holdthelegsopenbutton B.
Tolockthecastor,pushdownpedalAwiththefoot.
Tounlockthecastor,pushuppedalAwiththefoot.

5.3Raising/LoweringanElectricLift

Thehandcontrolisusedtoraiseorlowerthelift.
1.Toraisethelift—Press andholdtheUPAbutton toraisetheboomandthe patient.
2.Tolowerthelift—Press andholdtheDOWNB buttontolowertheboom andthepatient.
Releasethebuttontostopraisingorloweringthelift.
Thelegswillstopmovingwhenthebuttonis released.
5.4.2Closing/Openinglegsmanually
Themanuallegspreaderisoperatedbytwopedals(Aand B)orbytheleverC.
1.Toopenthelegs,presstherightpedalBwithafoot.
2.Toclosethelegs,presstheleftpedalAwithafoot. Withthelever:
1.Toopenthelegs,pulltheleverCtotheright.
2.Toclosethelegs,pushtheleverCtotheleft.

5.5Adjustingtheliftingarmtopatient bodyheight

CAUTION! Riskofinjuryordamage
Awrongboompositionmaycausetheliftingarm tosuddenlyslideinoroutwhenreleased.
–Alwaysadjusttheliftingarmwhenunloaded
andwiththeboominhorizontalposition.
Themainadjustmentcriteriaisthepatientsbodyheight,but thecorrectsettingalsodependsonotherfactorssuchas lengthoftheslingloops,patientconditionandpreference.
1654371-A13
Invacare®ISA™
B
C
A
123456
1
9
1
9
Thefollowingguidelinesarebasedonaveragevaluesand deviationscananddooccur:
Position1(fullyshortened):patientheightfromabout 140cmandtaller.
Position5:patientheightfromabout170cmandtaller.
Position9(fullyextended):patientheightuptoabout 200cm.
IMPORTANT
Anindividualtrialttingisstronglyrecommended todetermineacorrectandsafesettingforeach patient.
1.PushandholdthereleasehandleAonthelegsupport Bupwards.
2.SlidethelegsupportBupordown.
3.Whenthedesiredheightisreached,releasethehandle AandslightlymovethelegsupportBupordownuntil anaudibleclickisheardandthelegsupportBislocked intooneofthe6availableheightpositions.

5.7Usingthelegband

1.
1.PushandholdtheredleverAontheboomupwards.
2.AdjustthelengthoftheliftingarmBtooneofthenine availablepositionsindicatedonthescaleC,byplacing thedesiredpositioneldbeforetheedgeoftheboom.
3.ReleasetheleverAandslightlymovetheliftingarmB backandforthuntilanaudibleclickisheardanditis lockedintoposition.
5.6Adjustingtheheightoftheleg
support
Toapplythelegband,wrapthelongpaddedpart aroundthelegsupportandthepatient´slowerlegsand pushthecatchAintothebuckleclaspBontheback sideofthelegsupport.
2.
Toopen,pressthereleasemechanismCandpullthe catchAoutofthebuckleclaspB.
3.
Toadjustthelengthofthelegband,eitherpullonthe lugAtoshortenorpushthelugAtowardsthebuckle tolengthen.
14
1654371-A
Usage
Whenthereisnopatientinthelift,wraptheleg bandaroundthelegsupport,closeandshortenit totighten.

5.8Emergencyfunctions

5.8.1Performinganemergencystop
1.PresstheredemergencybuttonAonthecontrolunit tostoptheboomandpatientfromraisingorlowering.
2.Toreset,rotatetheemergencybuttonclockwise.
5.8.2Activatinganemergencylowering(CBJ
Homecontrolunit)
Ifthehandcontrolfails,theboomcanbeloweredbyusing thecircularswitchforemergencyrelease.
5.8.4Activatinganemergencylifting(CBJCare, CBJ1controlunit)
Ifthehandcontrolfails,theboomcanberaisedbyusing thecircularswitchforemergencylifting.Useasharpobject, suchasapenciltopressthebutton.
1.RaisetheboombypressingandholdingbuttonBatthe
frontofthecontrolunit.
2.Stopraisingtheboombyreleasingthebutton.
5.8.5Activatingamechanicalemergency lowering
Iftheemergencyloweringonthecontrolunitisnotworking, amechanicalemergencyloweringisavailableasaback-up. Thismayhappenincaseofpartialortotalpowerfailure,or ifthebatteryrunsdownduringuse.
1.LowertheboombypressingandholdingbuttonAat thefrontofthecontrolunit.
2.Stoploweringtheboombyreleasingthebutton.
5.8.3Activatinganemergencylowering(CBJ
Care,CBJ1,CBJ2controlunit)
Ifthehandcontrolfails,theboomcanbeloweredbyusing thecircularswitchforemergencyrelease.Useasharp object,suchasapenciltopressthebutton.
1.LowertheboombypressingandholdingbuttonAat thefrontofthecontrolunit.
2.Stoploweringtheboombyreleasingthebutton.
WARNING! RiskofInjuryorDamage
Repeatedmechanicalemergencyloweringleads toahighincreaseoftheloweringspeed.
–Onlyusethemechanicalemergencylowering
asaback-upwhentheemergencyloweringon thecontrolunitisnotworking.
–Donotusethemechanicalemergencylowering
repeatedlyinclosesuccession.
–Alwayslowerthepatientontoasupporting
surfacelikeabedorachair .
–Afteramechanicalemergencyloweringwas
required,theliftmustbecheckedtoresolve themalfunctionofthecontrolunit.
1.Locatethered emergencygripA atthebottomofthe actuatorpiston.
2.Slowlypullupthered emergencygripAand holditinpositionwhen asafeloweringspeed isreached.
3.Ifnoloweringoccurs withafullypulled emergencygripA, pushdownonthe boomatthesame time.
Theeffectiveloweringspeedisdependingonthepatient weight.Ifthespeedistoolowortoohigh,itcanbeadapted tothepatientweight. Adjustingtheloweringspeed:
1.Locatethescrewinthebottomholeofthered emergencygripA.
2.Turnthescrewcounter-clockwisetoincreasethespeed.
3.Turnthescrewclockwisetodecreasethespeed.
1654371-A15
Invacare®ISA™

5.9Chargingthebattery

IMPORTANT!
–Makesuretheemergencystopisnotactivated
whilechargingthebattery.
–Makesurethatchargingtakesplaceinaroom
withgoodairventilation.
–Electricalfunctionsarenotworkingwhenthe
liftisconnectedtopowersupply.
–Donotattempttousetheliftifthebattery
housingisdamaged.
–Replaceadamagedbatteryhousingbefore
furtheruse.
–Donotmovetheliftwithoutunpluggingfrom
thepoweroutlet.
Itisrecommendedtochargethebatterydailytoensure optimaluseoftheliftandprolongthelifeofthebattery. Furthermore,itisrecommendedtochargethebattery beforerstuse.
5.9.1CBJHomecontrolunit
Thecontrolunitisequippedwithasoundsignal.Abeep indicatesthatthebatteryhaslowcapacity,butloweringthe patientisstillpossible.Itisrecommendedtochargethe batteriesassoonasthesoundsignalisheard.
patientisstillpossible.Itisrecommendedtochargethe batteriesassoonasthesoundsignalisheard.
1.PlugthepowercordAintoapoweroutlet. Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhenthe batteriesarefullycharged. TherightyellowdiodeBwilllightcontinuously duringcharging,andswitchoffwhenfully charged. TheleftgreendiodeCwilllightcontinuously whilethecontrolunitisconnectedtothemains.
2.DisconnectthepowercordAfromthepoweroutlet
afterthebatteryhasbeenfullycharged.
BatteryIndicator
Thecontrolunitmaybeequippedwithabatteryindicator D,indicatingtheremainingbatterycapacity.
CBJCare
Battery
Indicator
Type
Battery State
Full Charge
Description
ThebatteryisOK—noneed forcharging(100–50%).The topLEDisGREEN.
1.PlugthepowercordCintoapoweroutlet. Thebatterywillchargeinapproximately4
hours.Thechargerstopsautomaticallywhenthe batteriesarefullycharged. TheupperyellowdiodeAwillashduring charging,andswitchtocontinuouslightwhen fullycharged. ThelowergreendiodeBwilllightcontinuously whilethecontrolunitisconnectedtothe mains,andlightsupwhenanybuttononthe handcontrolispressedorwhentheelectric emergencyloweringisactivated.
2.Disconnectthepowercordfromthepoweroutletafter
thebatteryhasbeenfullycharged.
5.9.2CBJCare,CBJ1,CBJ2controlunit
CBJ1andCBJ2withLCD
Battery
Indicator
Type
Partial Charge
Low Charge
Low Charge (LED ashing)
Battery State
Full Charge (100%)
Thebatteryneedstobe charged(50–25%).Themiddle LEDisYELLOW.
Thebatteryneedstobe charged(Lessthan25%). Acousticsignalwhenabutton ispressed.ThebottomLEDis YELLOW .
Thebatteryneedstobe charged.
Someofthefunctionalityof theliftislostanditisonly possibletolowertheboom.
Description
ThebatteryisOK—noneed forcharging(100%).
Partial Charge (75%)
Partial Charge (50%)
Thecontrolunitisequippedwithasoundsignal.Abeep indicatesthatthebatteryhaslowcapacity,butloweringthe
161654371-A
ThebatteryisOK—noneed forcharging(75%).
Thebatteryneedstobe charged(50%).
Usage
A
B
B
B
Battery
Indicator
Battery State
Description
Type
Low Charge (25%)
Low Charge (0%)
Thebatteryneedstobe charged(25%).Acousticsignal whenabuttonispressed.
Thebatteryneedstobe charged.
Someofthefunctionalityof theliftislostanditisonly possibletolowertheboom.
Optionalhandcontrol
ForCBJCareaoptionalhandcontrolwithbatteryindicator isavailable.
Battery
Indicator
Battery State
Description
Type
Full Charge
ThebatteryisOK—noneed forcharging(100–50%).The rightLEDisGREEN.
Partial Charge
Thebatteryneedstobe charged(50–25%).Themiddle LEDisYELLOW.
Low Charge
Thebatteryneedstobe charged(Lessthan25%). Acousticsignalwhenabutton ispressed.TheleftLEDis YELLOW .
Low Charge (LED ashing)
Thebatteryneedstobe charged.
Someofthefunctionalityof theliftislostanditisonly possibletolowertheboom.
5.9.3Optionalbatterycharger
(forcontrolunitswithdetachablebatteryonly)
Removingthebattery
1.2.
1.LiftuponthehandleAonthebackofthebatteryB.
2.Liftthebatteryupandoutawayfromthecontrolunit orcharger.
Installingthebattery
CAUTION!
Installingthebatteryimproperlymaycauseinjury ordamage.
–Makesurethereisanaudibleclickwhen
installingthebatteryonthecontrolunitor chargertoconrmproperinstallation.
1.PlacethebatteryB onthecontrolunitor chargerasshownand makesurethereisan audibleclick.
Whenthebatteryisplacedonthecharger,thecharge LEDwillilluminate.Whenchargingiscomplete, chargeLEDwillswitchoff .
Abatteryneedingtobefullyrechargedwilltake approximatelyfourhours.
Theprocedureofremovingorinstallingthebattery isthesameforthecontrolunitandthecharger.
1654371-A
17
Invacare®ISA™

6PatientTransfer

6.1Generalsafetyinformation

WARNING! RiskofInjuryorDamage
Improperuseofthisproductmaycauseinjury ordamage.
–Beforetransferringtoastationaryobject
(wheelchair,bed,commodeorothersurface) checkthatitsweightcapacitycanwithstand thepatient’sweight.
–Ifapplicablethebrakesofthestationaryobject
(wheelchair,bedetc.)mustbeappliedbefore loweringthepatientontoorliftingthepatient off.
–Whenusingtheliftinconjunctionwithbedsor
wheelchairs,beawareofthepositionofthelift inrelationshiptothoseotherdevicessothat theliftdoesnotbecomeentangled.
–Beforepositioningthelegsofthepatientlift
underabed,makesurethattheareaisclear ofanyobstructions.
WARNING! RiskofInjuryorDamage
Improperuseofthisproductmaycauseinjury ordamage.
–Usethepushbarsonthemastatalltimesto
pushorpullthelift.
–Avoidusingtheliftonanincline.Invacare
recommendsthattheproductonlybeusedon aatsurface.
–Duringtransfer,withthepatientsuspendedin
aslingattachedtothelift,donotrollliftover unevensurfacesthatcouldcausethelifttotip over.
–Duringtransfer,withthepatientsuspendedin
aslingkeepthelegsatthemaximumopen positionforoptimumstability .Ifitbecomes necessarytomovethroughanarrowpassage, closethelegsonlyaslongasrequiredand returntomaximumopenpositionassoonas possible.
WARNING! RiskofInjury
Injurytothepatientand/orassistantmayoccur
–Alwaysbeawareofthepositionofthelifting
armandthepatient.
–Alwaysbeawareofthepositionofthefoot
plate,especiallythepatient'spositiononthe footplate.
–Beforepositioningthelegsoftheliftaround
thepatient,makesurethatthepatient'sfeet areoutofthewayofthefootplate.
–Alwaysensurethatthefootplateand/or
patient'sfeetaredrybeforeperforminga transfer .
WARNING! RiskofEntrapmentorStrangulation
Thehandcontrolcordcancauseinjuryif improperlypositionedandsecured.
–Alwaysbeawareofthelocationofthehand
controlcordinrelationtothepatientand caregivers.
–Donotallowthehandcontrolcordtobecome
entangledaroundthepatientandcaregivers.
–Thehandcontrolmustbesecuredproperly.
ALWAYSstorethehandcontrolproperlywhen notinuse.
WARNING! RiskofEntrapmentorStrangulation
Itemsinthepatient’ssurroundingscancause entrapmentstrangulationduringlifting.Toavoid entrapmentorstrangulation:
–Beforelifting,checkthatthepatientis
completelyfreeofhis/hersurroundings.
WARNING! RiskofEntrapment
Thereisariskofentrapmentbetweenthehooks andthesling.
–Usecautionwhenlifting. –Neverputhandsorngersonornearthe
hookswhenlifting.
–Ensurethepatient’shandsandngersareaway
fromthehooksbeforelifting.
WARNING! RiskofInjuryorDamage
Damagetopartsofthelift(handcontrol,castors, etc.)causedbyimpactwiththeoor,wallsor otherstationaryobjectsmaycausedamageto theproductandleadtoinjury.
–Donotallowpartsofthelifttoimpactthe
oor,wallsorotherstationaryobjects.
–Alwaysstorethehandcontrolproperlywhen
notinuse.
181654371-A
WARNING! Riskofinjury
Injurytothepatientand/orassistantmayoccur:
–Ifliftingarmand/orlegsupportisnotadjusted
correctlytothepatientheight.
–Iftheliftingarmadjustmentisnotlocked,
damagedorwithwornparts.
IMPORTANT!
Alltransferproceduresdescribedinthefollowing canbeperformedbyone(1)assistant.However, Invacarerecommendsthattheproceduresare performedbytwo(2)assistantswhenever possible.
Onlyliftthepatientashighasrequired.Alower centerofgravityprovidesbetterstability,makesthe patientfeelmoresecureandthelifteasiertomove.

6.2Attachingtheslingtothelift

A
C
B
A
C
B
D
E
WARNING! RiskofInjury
Usageofwrongordamagedslingscancausethe patienttofallorcauseinjurytoassistants.
–UseanInvacareapprovedslingthatis
recommendedbytheindividual’sdoctor ,nurse ormedicalassistantforthecomfortandsafety oftheindividualbeinglifted.
–Invacareslingsandpatientliftaccessoriesare
specicallydesignedtobeusedinconjunction withInvacarepatientlifts.
–Aftereachlaundering(inaccordancewith
instructionsonthesling),inspectsling(s)for wear,tear,andloosestitching.
–Bleached,torn,cut,frayed,ordamagedslings
areunsafeandcouldresultininjury.Discard immediately.
–DONOTalterslings.
PatientT ransfer
1.PlacetheloopsAoftheslingBoverthehooksCon theliftingarm.
WARNING! RiskofInjury
Improperlyattachedoradjustedslingscancause thepatienttofallorcauseinjurytoassistants.
–Besuretochecktheslingattachmentseach
timetheslingisremovedandreplaced,to ensurethatitisproperlyattachedbeforethe patientisremovedfromastationaryobject (bed,chairorcommode).
–DONOTuseanykindofplasticback
incontinencepadorseatingcushionbetween patientandslingmaterialthatmaycausethe patienttoslideoutoftheslingduringtransfer.
–Positionthepatientintheslingasdirectedby
theinstructionsprovidedwiththesling.
–Adjustmentsforpatientsafetyandcomfort
shouldbemadebeforemovingthepatient.
Theloopsoftheslingmaybeequippedwithcolorcoded strapsprovidingdifferentlength,toplacethepatientin variouspositions.Matchthecorrespondingcolorsoneach sideoftheslingforanevenliftofthepatient.
Invacare®StandAssistsling
WARNING! RiskofInjury
–UsingaStandAssistslingthepatientmust
beabletosupportthemajorityoftheirown weight.
–MakesurethebottomedgeoftheStandAssist
slingisplacedonthelowerbackofthepatient andthepatient`sarmsareoutsideofthesling.
–Thebeltmustbesnugtopreventthepatient
fromslidingoutofthesling.
Invacare®TransferStandAssistsling
WARNING! RiskofInjury
–MakesurethebottomedgeoftheTransfer
StandAssistslingisplacedatthebaseofthe patient`sspineandthepatient`sarmsare outsideofthesling.
–Donotraisethepatienttoafullstanding
positionusingtheTransferStandAssistsling.
1.PlacetheupperloopsAoftheslingBoverthehooks Contheliftingarm.
2.PlacethelowerloopsDoftheslingBoverthehooks Eontheboom.
1654371-A19
Invacare®ISA™

6.3Transferringapatient

Liftingfromaseatingsurface
1.Beforeproceeding,reviewtheinformationandobserve allwarningsindicatedin2Safety,page5 Transfer ,page18.
2.Ifapplicable,ensurethebrakesofthestationaryobject (e.g.bed,wheelchairetc.)areapplied.
3.Positiontheslingaroundthepatient.Refertoyoursling usermanual.
4.Adjusttheliftingarmtothepatientheight,ifrequired. See5Usage,page13.
5.Unlocktherearcastorsofthelift.See5Usage,page13.
6.Openthelegsofthelifttomaximum.See5Usage, page13.
7.Usethepushbartomovetheliftintoposition.
Fortransferringapatientfromabedthelegsof theliftareplacedunderthebed
Fortransferringapatientfromawheelchair , commode,chairetc.thelegsoftheliftareplaced oneachsideofthestationaryobject.
8.Instructthepatienttoplacetheirfeetontothefoot plateandtheirlowerlegsagainstthelegsupport.Assist thepatientifnecessaryandensureproperplacementof thefeetonthefootplate
and6Patient
9.Ensuretheupperedgeofthelegsupportcushionis placedabout2ngersbelowthekneecaps(patella) ofthepatient.Adjusttheheightofthelegsupportif necessary.See5Usage,page13.
10.Ifapplicable,adjustandapplythelegbandaroundthe patientslowerlegs.See5Usage,page13.
11.Lowerthepatientliftforeasyattachmentofthesling. See5Usage,page13.
12.Locktherearcastorsofthelift.See5Usage,page13
13.Attachthesling.See6.2Attachingtheslingtothelift,
page19
14.Unlocktherearcastorsofthelift.See5Usage,page13.
15.Instructthepatienttoholdontothehandgripsonboth sidesoftheliftingarm.
16.Instructthepatienttoleanbackintothesling.
17.PresstheUParrowbuttononthehandcontroltoraise thepatientabovethestationaryobject.See5Usage,
page13
18.Beforemovingthepatient,checkagaintomakesure thattheslingisproperlyconnectedtothehooksofthe liftingarm.Ifanyattachmentsarenotproperlyinplace, lowerthepatientbackontothestationaryobjectand correcttheproblem.
19.Usingthepushbar,movetheliftawayfromthe stationaryobject.
Loweringtoaseatingsurface
1.Beforeproceeding,reviewtheinformationandobserve allwarningsindicatedin2Safety,page5and6Patient Transfer ,page18.
2.Ensurethepatientisraisedhighenoughtoclearthe surfacetobetransferredto.Raisethepatientand/or lowerthestationaryobjectifapplicable.
3.Ifapplicable,ensurethebrakesofthestationaryobject (e.g.bed,wheelchairetc.)areapplied.
4.Positionthepatientasfarovertheseatingsurfaceas possible.
Fortransferringapatienttoabedthelegsofthe liftareplacedunderthebed
Fortransferringapatienttoawheelchair,commode, chairetc.thelegsoftheliftareplacedoneach sideofthestationaryobject.
5.PresstheDOWNarrowbuttonandlowerthepatient ontotheseatingsurface.See5Usage,page13
6.Locktherearcastorsofthelift.See5Usage,page13
7.Unhooktheslingfromallattachmentpointsonthe standassistlift.
8.Ifapplicable,removethelegbandfromaroundthe patientslowerlegs.See5Usage,page13
9.Instructthepatienttolifttheirfeetoffofthefootplate. Assistthepatientifnecessary.
10.Unlocktherearcastorsofthelift.See5Usage,page13
11.Movetheliftawayfromthestationaryobject.
12.Removetheslingfromaroundthepatient.
ThepatientcanremainintheInvacare®Stand Assistslingwhileusingthecommode.Inthis casetheslingmustnotbeplacedaroundthe patientagainbeforeliftingfromthecommode.
201654371-A
TransportationandStorage

7TransportationandStorage

7.1Generalinformation

Duringtransportation,orwhenthepatientliftisnottobe usedforsometime,theemergencystopbuttonshouldbe pushedin.
Adisassembledliftcanbelocatedinthepackagingbox,for transportationandstorage.
Thepatientliftmustbestoredatnormalroomtemperature. Ifitisstoredinadamp,coldorwetenvironmentthenthe motorandothermountingpartsmaybepronetocorrosion. See11.4Environmentalconditions,page28
.

7.2Disassemblingthelift

1.Removethelegspreaderleverifattached.
2.Lowertheboomandclosethelegscompletely.
3.Activatetheemergencystopbuttonandapplycastor brakes.
4.Adjusttheliftingarmtominimumlength.See5.5 Adjustingtheliftingarmtopatientbodyheight,page13.
5.Reversetheproceduredescribedin4.6Installingthe legsupport,page11.
6.Reversetheproceduredescribedin4.3Installingthe masttothebase,page10.
7.Reversetheproceduredescribedin4.4Installingthe footplate,page11.
1654371-A
21
Invacare®ISA™

8Maintenance

8.1Generalmaintenanceinformation

CAUTION! Riskofinjuryordamage
–Donotperformanymaintenanceorservice
procedureswhiletheproductisinuse.
Followthemaintenanceproceduresdescribedinthismanual tokeepyourproductincontinuousservice.
Furthermaintenanceandinspectionprocedures,tobe performedbyaqualiedtechnician,aredescribedinthe servicemanualofthisproduct.Servicemanualscanbe obtainedfromInvacare.

8.2Dailyinspections

WARNING! Riskofinjuryordamage
Damagedorwornpartscanaffectthesafetyof thelift.
–Thepatientliftshouldbecheckedeachtime
itisused.
–Donotusetheliftifdamageisfoundoryou
questionthesafetyofanypartofthelift. ContactyourInvacareproviderimmediately andensuretheliftisnotuseduntilrepairsare made.
CAUTION! Riskofelectricshockandproductdamage
–Switchoffthedeviceanddisconnectfrom
mains,ifapplicable.
–Whencleaningelectroniccomponentsconsider
theirprotectionclassregardingwateringress.
–Makesurethatnowatersplashestotheplug
orthewalloutlet.
–Donottouchthepowersocketwithwethands.
IMPORTANT!
Wronguidsormethodscanharmordamage theproduct.
–Allcleaningagentsanddisinfectantsusedmust
beeffective,compatiblewithoneanotherand mustprotectthematerialstheyareusedto clean.
–Neverusecorrosiveuids(alkalines,acidetc.)
orabrasivecleaningagents.Werecommend anordinaryhouseholdcleaningagentsuchas dishwashingliquid,ifnotspeciedotherwisein thecleaninginstructions.
–Neveruseasolvent(cellulosethinner ,acetone
etc.)thatchangesthestructureoftheplastic ordissolvestheattachedlabels.
–Alwaysmakesurethattheproductiscompletely
driedbeforetakingintouseagain.
Forcleaninganddisinfectioninclinicalorlong-term careenvironments,followyourin-houseprocedures.
DailyInspectionChecklist
qVisuallyinspectthepatientlift.Checkallpartsfor
externaldamageorwear.
qCheckallhardware,attachmentpointsandstressed
parts,suchasslings,hooksandanypivotpointsforsigns ofwear,cracking,fraying,deformationordeterioration.
qVerifythatthehandcontrolisfunctional(liftingand
legmovements).
qChargethebatteryeverydaytheliftisused. qChecktheemergencystopfunction. qChecktheservicelight(CBJCarecontrolunitonly).
Theservicelightisindicatedbyawrench symbolonthecontrolunitandifapplicable onthehandcontrol.
–Iftheservicelightisashingyellowtheliftrequires
service.DonotusetheliftandcontactyourInvacare providerforservice.
–Iftheservicelightisnotashingtheliftisreadyfor
use.

8.3CleaningandDisinfection

8.3.1Generalsafetyinformation
CAUTION! Riskofcontamination
–T akeprecautionsforyourselfanduse
appropriateprotectiveequipment.
8.3.2CleaningIntervals
IMPORTANT!
Regularcleaninganddisinfectionenhances smoothoperation,increasestheservicelifeand preventscontamination. Cleananddisinfecttheproduct
–regularlywhileitisinuse, –beforeandafteranyserviceprocedure, –whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser .
8.3.3Cleaninginstructions
IMPORTANT!
–Theproductdoesnottoleratecleaningin
automaticwashingsystems,withhigh-pressure cleaningequipmentorsteam.
Cleaningthelift
Method:Wipeoffwithawetclothorsoftbrush.
Max.temperature:40°C
Solvent/chemicals:Ordinaryhouseholdcleaningagentand water .
Drying:Wipedrywithasoftcloth.
Cleaningthesling
Refertothewashinginstructionsontheslingandtothe slingmanualforcleaningdetails.
22
1654371-A
Maintenance
8.3.4Disinfectioninstructions
Informationonrecommendeddisinfectants andmethodscanbefoundonhttps://vah­online.de/en/for-users.
Indomesticcare
Method:Followtheapplicationnotesfortheused disinfectantandwipedisinfectallaccessiblesurfaces.
Disinfectant:Ordinaryhouseholddisinfectant.
Drying:Allowtheproducttoair-dry.
Inclinicalandlongtermcare
Followyourin-housedisinfectionproceduresandonlyuse thedisinfectantsandmethodsspeciedtherein.

8.4Serviceinterval

WARNING! Riskofinjuryordamage
Servicemustbeperformedonlybyaqualied technician.
–ContactyourInvacareproviderforservice.
Servicemustbeperformedatleastevery12monthsunless otherwisestatedinlocalrequirements.
8.4.1LOLERStatement
TheUKHealthandSafetyExecutive’sLiftingOperationsand LiftingEquipmentRegulations1998,requireanyequipment thatisusedintheworkplacetoliftaloadbesubjectto safetyinspectiononasixmonthlybasis.Pleaserefertothe HSEwebsiteforguidancewww.hse.gov.uk.
Thepersonresponsiblefortheequipmentmustensure adherencetoLOLERregulations.
1654371-A23
Invacare®ISA™

9AfterUse

9.1Disposal

WARNING! EnvironmentalHazard
Devicecontainsbatteries. Thisproductmaycontainsubstancesthatcould beharmfultotheenvironmentifdisposedof inplaces(landlls)thatarenotappropriate accordingtolegislation.
–DONOTdisposeofbatteriesinnormal
householdwaste.
–BatteriesMUSTbetakentoaproperdisposal
site.Thereturnisrequiredbylawandfreeof
charge. –Doonlydisposedischargedbatteries. –Coverterminalsoflithiumbatteriespriorto
disposal. –Forinformationonthebatterytypeseebattery
labelorchapter11TechnicalData,page27
Beenvironmentallyresponsibleandrecyclethisproduct throughyourrecyclingfacilityatitsendoflife.
Disassembletheproductanditscomponents,sothedifferent materialscanbeseparatedandrecycledindividually.
Thedisposalandrecyclingofusedproductsandpackaging mustcomplywiththelawsandregulationsforwaste handlingineachcountry.Contactyourlocalwaste managementcompanyforinformation.

9.2Reconditioning

Thisproductissuitableforreuse.T oreconditiontheproduct foranewuser ,carryoutthefollowingactions:
Cleaninganddisinfection
Inspectionaccordingtoserviceplan
Fordetailedinformationaboutinspection,cleaningand disinfection,consulttheusermanualandservicemanual forthisproduct.
.
24
1654371-A

10Troubleshooting

10.1Identifyingfaultsandpossiblesolutions

SymptomsFaultsSolution
Patientliftfeelsloose
Castors/brakesnoisyorstiff
NoisyordrysoundfrompivotsPivotsarewornordamagedContactyourInvacareprovider
Electricactuatorfailstoraiseorlower theboomorlegsfailtoopenorclose whenbuttonispressed
UnusualnoisefromactuatorActuatorisdamagedContactyourInvacareprovider
Boomwillnotlowerwhenactivating themanualemergencyrelease
Thecontrolunitemitsabeepingsound duringlifting,andthemotorstops
BatterydoesnotchargePowercordisnotproperlyconnectedto
LiftingarmcannotbeadjustedTheredleverontheboomisnot
Legsupportcannotbeinstalledor adjusted
Mast/BasejointlooseRetightentheMast/Baseconnection.
See4.3Installingthemasttothebase,
page10
ManuallegspreadermechanismlooseContactyourInvacareprovider
FluffordebrisinbearingsCleancastorsfromuffanddebris
Castors/brakesaredamaged
Handcontroloractuatorconnector loose
BatterylowChargethebattery.See5.9Charging
EmergencystopbuttonispressedinRotatetheemergencystopbutton
Batterynotconnectedproperlyto controlunit
TheconnectingterminalsaredamagedContactyourInvacareprovider
PowercordconnectedtopoweroutletDisconnectpowercordfrompower
Boomactuatororlegactuatoris damaged
Max.loadisexceeded
Boomrequiresaminimumweightload tolowerfromtheuppermostposition
Max.loadisexceeded
thecontrolunit/andorpoweroutlet
PowercordisdamagedContactyourInvacareprovider
pushedandheldupwards
Liftingarmisalreadyatthemaximum orminimumposition
TheadjustmentmechanismisdamagedContactyourInvacareprovider
Boomisinwrongposition
Thereleasehandlenotpushedandheld upwards
TheadjustmentmechanismisdamagedContactyourInvacareprovider
Legbandiscloggingthebackoftheleg support
ContactyourInvacareprovider
Connecthandcontroloractuator connector.Ensureconnectorsareseated properlyandfullyconnected
thebattery,page16
clockwiseuntilitpopsout
Reconnectthebatterytothecontrol unit.See5.9.3Optionalbatterycharger,
page17
outlet
ContactyourInvacareprovider
Reducetheload(andtheliftwill functionnormally)
Pulldownslightlyontheboom
Reducetheload(andtheliftwill functionnormally)
Ensureconnectorsareseatedproperly andfullyconnected.See5.9Charging
thebattery,page16
See5.5Adjustingtheliftingarmto patientbodyheight,page13
See5.5Adjustingtheliftingarmto patientbodyheight,page13
Movetheboomtohorizontalposition. See5.3Raising/LoweringanElectricLift,
page13
See4.6Installingthelegsupport,page 11
Clearthebackofthelegsupport
Troubleshooting
1654371-A25
Invacare®ISA™
SymptomsFaultsSolution
MastcannotbeassembledonthebaseBoompointinginthewrongdirectionBringtheboomincorrectposition.See
4.3Installingthemasttothebase,page 10
LiftdoesnotrollThecastorsarelockedUnlockthecastorsSee5.2
Locking/Unlockingtherearcastors,page 13
ContactyourInvacareprovideriftheabovedoesnotsolveyourproblems.
261654371-A

11TechnicalData

11.1Maximumsafeworkingload

TechnicalData
Max.SafeWorkingLoad(patient+sling)

11.2Dimensionsandweights

ISA™
COMPACT
STANDARD
ISA™
ISA™ PLUS
140kg160kg180kg200kg
ISA™
XPLUS
*Forwarddirection
Dimensions[mm]
ISA™
COMPACT
STANDARD
ISA™
ISA™ PLUS
Front/rearcastordiameter75/75100/100100/100100/100100/125
Max.reachat600mm(a)*
Max.reachfrombase(b)*
Baselength(c)
Totallengthwithoutlever(n)
Totallengthwithlever(l)*
Reachfrombasewithlegsspreadto700 mm(d)
Max.height(f)*
Liftingrange(g)*
Min.heightatSlingHook-up(h
Max.heightatSlingHook-up(h
)*
min
)*
max
Totalwidth(legsopen),externalmeasure(i)
Totalwidth(legsopen)internalmeasure(j)
110-270110-270110-270110-270110-270
340-500340-500340-500340-500340-500
900925108012201280
925925
n/an/a
980-1000980-100011451290
420420225225225
1635-17851650-18001650-18001650-18001650-1800
505-675505-675505-675505-675505-675
945-970960-985960-985960-985960-985
1475-16201490-16351490-16351490-16351490-1635
865885111011701170
755
73596010201020
ISA™
XPLUS
1280
n/a
1654371-A
27
Invacare®ISA™
Dimensions[mm]
Totalwidth(legsclosed),externalmeasure (r)
Min.internalwidth(s)
Internalwidthatmaximumreach(k)*
Turningdiameter
Heighttoupperedgeoflegs(m)
Min.freeheight(p)
*Rangedependingonfullyshortenedtofullyextendedpositionofliftingarm
WeightsMainParts[kg]
Totalweight
Weight,mast,incl.battery,
Weightmanuallegsection
Weightelectricallegsection
Legsupport
Footplate(high,lowandinclined)
515535635640640
380380495495495
640-725640-725775-870775-870775-870
10601060126014001430
100115115115115
ISA™
COMPACT
2035353535
ISA™
COMPACT
43484952
152021
n/a
ISA™
STANDARD
ISA™
STANDARD
1920
202123
5
4
ISA™ PLUS
ISA™ PLUS
ISA™
XPLUS
ISA™
XPLUS
n/a

11.3Electricalsystem

ISA™
COMPACT
Voltageoutput Voltagesupply
280mA(for
Maximumcurrentinput
Protectionclass(entiredevice)IPX4* InsulationclassClassIIequipment
TypeBAppliedPart
Soundlevel Workingability Intermittent(periodicmotoroperation) Batteryspecications Batterytype Manualemergencylowering
Electricemergencylowering/lifting
CBJHome)/
400mA(for
CBJ2)
Appliedpartcomplyingwiththespeciedrequirementsforprotection
againstelectricalshockaccordingtoIEC60601-1.
82fulllifts(withbatterybetween100%–50%offullcapacity)
Yes/No(forCBJ2
andCBJHome)
ISA™
STANDARD
24VDC
100–240VAC,50/60Hz
280mA(for
CBJHome)/
400mA(for
CBJ1andCBJ2)
45–50dB(A)
10%,max.2min/18min
2x12V/2.9Ah
Leadacid(servicefree,sealed)
Yes(atbottomofactuator)
Yes/Yes
(forCBJ1) Yes/No(for CBJ2andCBJ
Home)
CBJ1,CBJ2and
(forCBJ1and
Yes/No(for CBJ2andCBJ
280mA(for
CBJHome)/
400mA(for
CBJCare)
Yes/Yes
CBJCare)
ISA™ PLUS
Home)
ISA™
XPLUS
400mA(forCBJ1
andCBJcare)
Yes/Yes(for CBJ1andCBJ
care)
*IPx4-Thesystemisprotectedagainstwatersplashedfromanydirection.

11.4Environmentalconditions

Storageand
transportation
Temperature
281654371-A
-10°Cto+50°C+5°Cto+40°C
Operation
Relativehumidity20%to90%at30°C,notcondensing Atmospheric
pressure
800hPato1060hPa
TechnicalData
Allowtheproducttoreachoperationtemperature beforeusage:
Warmingupfromminimumstoragetemperature mighttakemorethan30minutes.
Coolingdownfrommaximumstorage temperaturemighttakemorethan5minutes.

11.5Materials

Component
Base,legs,mastandboomSteel,powder-coated Boomconnector,boltsand
nuts Handgrips
Slinghooks LegsupportcushionSelfskinPU Footplatecover
Actuatorhousing,hand control,mastprotector, castorsandotherplastic parts
Legband
Material
Steel,zinc-plated
TPE TPU
PVC,PU
Materialaccordingto marking(PA,PP ,PE)
PA,PES,PVC,PU
Allcomponentsoftheproductareeithercorrosionresistant orcorrosionprotected.

11.6Operatingforcesofcontrols

ControlOperatingforce
Buttonsoncontrolunit Buttonsonhandcontrol
FootpedalonISA™COMPACTwith 100mmcastors*
LegspreaderleveronISA™ COMPACT
LegspreaderleveronISA™ STANDARD
LegspreaderleveronISA™PLUS
*Forallothermodelsandcastorsizes,aleverforthe manuallegspreaderismandatory.
4N 4N
270N
64N
70N
85N
1654371-A29
Invacare®ISA™

12Electromagneticcompatibility(EMC)

12.1GeneralEMCinformation

MedicalElectricalEquipmentneedstobeinstalledandusedaccordingtotheEMCinformationinthismanual.
ThisproducthasbeentestedandfoundtocomplywithEMClimitsspeciedbyIEC/EN60601-1-2forClassBequipment.
PortableandmobileRFcommunicationsequipmentcanaffecttheoperationofthisproduct.
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabove standard.Todetermineiftheemissionfromthisproductiscausingtheinterference,runandstoprunningthisproduct.If theinterferencewiththeotherdeviceoperationstops,thenthisproductiscausingtheinterference.Insuchrarecases, interferencemaybereducedorcorrectedbythefollowing:
Reposition,relocate,orincreasetheseparationbetweenthedevices.

12.2Electromagneticemission

Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis productshouldassurethatitisusedinsuchanenvironment.
Emissionstest
RFemissions CISPR11
RFemissions CISPR11
Harmonic emissions IEC61000-3-2
Voltageuctuations /ickeremissions IEC61000-3-3
ComplianceElectromagneticenvironment-guidance
ThisproductusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
GroupI
ClassB
ClassA
Complies
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby electronicequipment.
Thisproductissuitableforuseinallestablishmentsincludingdomestic establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.

12.3ElectromagneticImmunity

Guidanceandmanufacturer´sdeclaration
Thisproductisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthis productshouldassurethatitisusedinsuchanenvironment.
Immunitytest
Electrostatic discharge(ESD)
IEC61000-4-2
Electrostatic transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Test/Compliancelevel
±8kVcontact
±2kV,±4kV,±8kV,± 15kVair
±2kVforpowersupply lines;100kHzrepetition frequency
±1kVforinput/output lines;100kHzrepetition frequency
±1kVlinetoline
±2kVlinetoearth
Electromagneticenvironment–guidance
Floorsshouldbewood,concreteorceramictile.Ifoorsarecoveredwith syntheticmaterial,therelativehumidityshouldbeatleast30%.
Mainspowerqualityshouldbethatofatypicalcommercialorhospital environment.
Mainspowerqualityshouldbethatofatypicalcommercialorhospital environment.
301654371-A
Loading...