Invacare IRC5PO2V User Manual

Invacare®Perfecto Concentrator
IRC5PO2V
enHomeFill®SystemCompatible
UserManual
₂ ₂
™VOxygen
©2016Invacare®Corporation Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted. MakingLife’sExperiencesPossibleisatrademarkintheU.S.A. Pine-solandLestoilareregisteredtrademarksofTheCloroxCompany. DawnisaregisteredtrademarkofTheProctorandGambleCompany.
Contents
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
1.3IndicationsForUse.............................6
1.4Description...................................6
1.5Contraindications..............................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................8
2.3RadioFrequencyInterference......................11
3Components....................................12
3.1ComponentIdentification.........................12
3.2PneumaticDiagram.............................14
4Accessories.....................................15
4.1OptionalAccessories............................15
5Setup..........................................16
5.1Unpacking....................................16
5.2Inspection....................................16
5.3Storage......................................16
6Usage..........................................17
6.1Introduction..................................17
6.2SelectaLocation...............................17
6.3SettingUptheConcentrator......................18
6.3.1SettingUptheHumidifier......................18
6.3.2TurningtheConcentratorOn...................20
6.3.3Flowrate..................................20
6.3.4SensO
OxygenPurityIndicator.................21
2
6.3.5InitialStartupoftheConcentrator................21
6.3.6ExplanationofOxygenPurityIndicatorLights........22
7Maintenance....................................23
7.1ServiceLife...................................23
7.2CleaningtheCabinet............................23
7.3CleaningtheCabinetFilter........................24
7.4HumidifierCleaningandThermicDisinfection..........24
7.5CleaningandDisinfectionBetweenPatients............25
7.6PreventiveMaintenanceChecklist...................26
8AfterUse.......................................27
8.1RecyclingInformation............................27
8.2WearandTear................................27
9Troubleshooting.................................28
9.1Troubleshooting...............................28
10Technicaldata..................................30
10.1Specifications.................................30
11Warranty......................................33
11.1NorthAmericanLimitedWarranty.................33
Invacare®Perfecto™VOxygenConcentrator

1General

1.1Symbols

Signalwordsareusedinthismanualandapplytohazardsorunsafe practiceswhichcouldresultinpersonalinjuryorpropertydamage. Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor efficient,trouble-freeuse.
SymbolsinDocumentation
GeneralWarningSign
Thebackgroundcolorinsidethe triangleisyellowonproductlabels.
Readmanual
Thecolorofthesymbolbackground isblueonproductlabels.
NoSmoking
Thecolorofthecirclewithdiagonal barisredonproductlabels.
NoOpenFlame
Thecolorofthecirclewithdiagonal barisredonproductlabels.
ClassII,DoubleInsulated
Protectedagainstsolidforeignobjectsof12.5 mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
IndoorUseONLY
KeepDry
4
1193322-A-04
-20F MIN
150F MAX
TransportandStorageTemperature
15%
MIN
95% MAX
TransportandStorageHumidity
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
O
Indicators
2
SymbolO
O
2
PurityIndicatorLights(LED)
2
SYSTEM
GREENIndicatorLight
OKAY
O
over85%
2
O 73%to85%
between
2
YELLOWIndicatorLight
A.YELLOWSolid
B.YELLOWFlashingSensor
Failure.Callaqualifiedtechnician.
SYSTEM FAILURE
O
Below73%
2
REDIndicatorLight
RefertoTroubleshooting.
General
ElectricalHazard

1.2IntendedUse

TheintendedfunctionanduseoftheInvacare®Perfecto™V
Manufacturer
OxygenConcentratoristoprovidesupplementaloxygentopatients withrespiratorydisorders,byseparatingnitrogenfromroomair,by wayofamolecularsieve.Itisnotintendedtosustainorsupportlife.
Callsupplier
DANGER! RiskofInjuryorDeath
SymbolsonProduct
Unitrunning
Thisproductistobeusedasanoxygensupplementand isnotintendedtobelife-supportingorlife-sustaining. ONLYusethisproductifthepatientiscapableof spontaneousbreath,abletoinhaleandexhalewithout
Unitnotrunning
theuseofamachine.
–DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
1193322-A-045
Invacare®Perfecto™VOxygenConcentrator
DANGER! RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientsonflow ratesgreaterthan5L/minmaybeatanincreasedrisk forseriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
caresupplierBEFOREusingthisproductifyouare prescribedaflowrategreaterthan5L/min.
WARNING! RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand specificationshasnotbeentestedandmayleadto productdamage,lossofproductfunction,orpersonal injury.
–DONOTusethisproductinanywayotherthan
describedintheSpecificationsandIntendedUse sectionsofthismanual.
DANGER! RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen concentratorinthemarkettoday,thisoxygen concentratorcanfailtoproduceoxygenduetopower failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient toswitchtotheirbackupsourceofoxygen.Referto Troubleshootingformoredetail.

1.3IndicationsForUse

TheintendedfunctionanduseoftheInvacare®Perfecto™V OxygenConcentratoristoprovidesupplementaloxygentopatients withrespiratorydisorders,byseparatingnitrogenfromroomair,by wayofamolecularsieve.Itisnotintendedtosustainorsupportlife.

1.4Description

TheInvacarePerfecto respiratorydisorderswhorequiresupplementaloxygen.Thedevice isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87% to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse ofanasalcannula.
TheInvacarePerfecto pressureswingadsorptionmethodologytoproducetheoxygen gasoutput.Ambientairentersthedevice,isfilteredandthen compressed.Thiscompressedairisthendirectedtowardoneof twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe patient.
TheInvacarePerfecto patientinahomeenvironmentorinaninstitutionalenvironment. Deviceoperatesatanominal120VAC/60Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified technicalpersonnelonly.
concentratorisusedbypatientswith
2
concentratorusesamolecularsieveand
2
concentratoriscapableofoperationbythe
2

1.5Contraindications

Therearenoknowncontraindications.
61193322-A-04
Safety

2Safety

2.1LabelLocations

A
B
C
D
1193322-A-047
Serialnumberlabelislocatedontheresonator intakeassembly.
SpecificationLabelislocatedonthebackofthe concentratoratthebase.
Invacare®Perfecto™VOxygenConcentrator

2.2GeneralGuidelines

Inordertoensurethesafeinstallation,assemblyandoperationofthe concentratortheseinstructionsMUSTbefollowed.
DANGER! RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor damage.Thissectioncontainsimportantinformationfor thesafeoperationanduseofthisproduct.
–DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand understandingtheseinstructionsandanyadditional instructionalmaterialsuchasusermanuals,service manualsorinstructionsheetssuppliedwiththis productoroptionalequipment.
–Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional, dealerortechnicalpersonnelbeforeattemptingto usethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot workingcorrectly,contactatechnicianorInvacare forrepair.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER! RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy substancesandothercombustiblesareeasilyignited andburnwithgreatintensityinoxygenenrichedairand whenincontactwithoxygenunderpressure.Toavoid fire,death,injuryordamage:
DONOTSMOKEwhileusingthisdevice.DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
NOSMOKINGsignsshouldbeprominentlydisplayed.Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby anytypeoffriction.
–Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutofthe roominwhichthisconcentratorislocatedandaway fromwhereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings, chaircushions,clothing,andawayfromheatedorhot surfacesincludingspaceheaters,stoves,andsimilar electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor ontheorderofamedicalpractitionerlicensedbya governmentalagencywherehe/shepractices.
–ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
81193322-A-04
Safety
DANGER! RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor injurytopersons:
–DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable parts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the concentratormustbelocatedinanotherroomat least7ft(2.1m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
–DONOTusefrayedordamagedACpowercords.
WARNING! RiskOfInjuryOrDamage
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare accessories.Accessoriesdesignedbyother manufacturershavenotbeentestedbyInvacareand arenotrecommendedforusewithInvacareproducts.
–Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis device.Youshouldcontactyourlocalhomecare supplierforrecommendationsonwhichofthese deviceswillbebestforyou.Theyshouldalsogiveyou adviceontheproperusage,maintenance,andcleaning.
DANGER! RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway contamination:
–DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics, cleaningagents,orchemicalvapors.
WARNING! RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorimpairedindividuals.
–Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING! RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing entanglement:
–Keepchildrenandpetsawayfromnasalcannulaand
tubing.
–Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/orimpairedpersons.
WARNING! RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.Check
withyourequipmentsupplierorphysicianto determinehowoftenthecannulashouldbereplaced.
–DONOTsharecannulasbetweenpatients.
1193322-A-049
Invacare®Perfecto™VOxygenConcentrator
WARNING! RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto you.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings shouldbechanged.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorbypulling
onthecord.
–DONOTuseextensioncordswithACpowercords
provided.
–Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING! RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhenplugged
in.
–Makesureconcentratorisoffwhennotinuse.
WARNING! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned tominimizeroutinepreventivemaintenance.Toprevent injuryordamage:
–Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas factorytrainedpersonnelshouldperformpreventive maintenanceorperformanceadjustmentsonthe oxygenconcentrator,exceptfortasksdescribedin thismanual.
–UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION! RiskofDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto water,callequipmentsupplier/qualifiedtechnicianfor examinationandrepair.
–NEVERdroporinsertanyobjectorliquidintoany
opening.
–ForindooruseONLY.
101193322-A-04

2.3RadioFrequencyInterference

WARNING! RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrom interferencewithwirelessequipment:
–Keepconcentratoratleast9.8ft(3.0m)awayfrom
wirelesscommunicationequipmentsuchaswireless homenetworkdevices,mobilephones,cordless phonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits specifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovide areasonableprotectionagainstelectromagneticinterferenceina typicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels ofelectromagneticemissionspermittedbytheabovestandards. Todetermineiftheemissionsfromtheconcentratorarecausing theinterference,turntheconcentratorOff.Iftheinterference withtheotherdevice(s)stops,thentheconcentratoriscausingthe interference.Insuchrarecases,interferencemaybereducedor correctedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweenthe
equipment.
Connecttheequipmentintoanoutletonacircuitdifferentfrom
thattowhichtheotherdevice(s)isconnected.
Safety
1193322-A-04
11
Invacare®Perfecto™VOxygenConcentrator

3Components

3.1ComponentIdentification

A
B
C
D
E
F
OxygenOutlet
Flowmeter
ElapsedTimeMeter
OxygenPurityIndicator Lights/FaultandPowerIndicator Lights
CircuitBreaker
PowerSwitch
12
1193322-A-04
Components
RearView
G
H
I
PowerCord
HomeFillOutletFitting
CabinetFilter
TheHomeFilloutletfittingHistobeusedonlyfor fillingoxygencylinderswiththeHomeFillhomeoxygen compressor.Theoutletfittingdoesnotaffectconcentrator performance.RefertotheHomeFillusermanual,part number1100873,forconnectionandoperatinginstructions. Whennotinuse,theplugprovidedwiththeconcentrator shouldbeinsertedintotheoutletfitting.Formore informationabouttheHomeFill,contactyourInvacare dealer.
1193322-A-0413
Invacare®Perfecto™VOxygenConcentrator

3.2PneumaticDiagram

14
1193322-A-04

4Accessories

4.1OptionalAccessories

Therearemanydifferenttypesofhumidifiers,oxygentubing,cannulas andmasksthatcanbeusedwiththisdevice.Youshouldcontact yourlocalhomecaresupplierforrecommendationsonwhichof thesedeviceswillbebestforyou.Theyshouldalsogiveyouadvice ontheproperusage,maintenance,andcleaning.
Thesupplyaccessories(nasalcannula,mask,tubing,humidifier,etc.) usedtodeliveroxygentothepatientneedtoincludeameansto reducethepropagationoffireintheaccessoriesforthesafetyofthe patientandothers.Ifacommerciallyavailable,fire-activatedflow stopdeviceisusedintheaccessoriessetup,itshouldbeplacedas closetothepatientaspracticable.
Thefollowingoptionalaccessoriesandreplacementparts(listed below)arealsoavailable:
StandardAdultNasalCannula,7ft(2.1m)-M3120
StandardHumidifier-InvacarePart#1155719
PreciseRx™PediatricFlowmeterAccessory-IRCPF16
HomeFillHomeOxygenCompressor-IOH200
OxygenSupplyTubing7ft(2.1m)-MS4107
OxygenSupplyTubing21ft(6.3m)-MS4121
OxygenTubingConnector-MS4301
Accessories
1193322-A-0415
Invacare®Perfecto™VOxygenConcentrator

5Setup

5.1Unpacking

1.Checkforanyobviousdamagetothecartonoritscontents.If damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecarton.
Unlesstheoxygenconcentratoristobeused IMMEDIATELY,leaveconcentratorinitspackagingfor storageuntiluseisrequired.

5.2Inspection

1.Inspect/examineexterioroftheconcentratorfornicks,dents, cracks,scratchesorotherdamage.
2.Inspectallcomponents.

5.3Storage

1.Storetherepackagedoxygenconcentratorinadryarea.Refer toSpecificationsforstoragetemperatureparameters.
2.DONOTplaceobjectsontopofrepackagedconcentrator.
161193322-A-04
Usage

6Usage

6.1Introduction

Youroxygenconcentratorisintendedforindividualuseindoors. Itisanelectronicallyoperateddevicethatseparatesoxygenfrom roomair.Itprovideshighconcentrationofoxygendirectlytoyou throughanasalcannula.Clinicalstudieshavedocumentedthat oxygenconcentratorsaretherapeuticallyequivalenttoothertypesof oxygendeliverysystems.
Yoursupplierwillshowyouhowtouseyouroxygenconcentrator. He/Sheshouldbecontactedwithanyquestionsorproblems regardingyouroxygenconcentrator.Thisusermanualwilltellyou aboutyourconcentratorandwillserveasareferenceasyouuse yourconcentrator.

6.2SelectaLocation

WARNING! RiskofInjury
Toavoidinjuryduringtherapy:
–TheoxygenconcentratorMUSTbeplacedonalevel
surfaceforuse.
–DONOTrelocatetheoxygenconcentratorwhilein
use.
WARNING! RiskofInjuryorDamage
Toavoidinjuryordamagefromairbornepollutants and/orfumesandforoptimalperformance:
–Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandtheair exhaustsarenotobstructed.
–NEVERblocktheairopeningsoftheconcentrator
orplaceitonasoftsurface,suchasabedorcouch, wheretheairopeningmaybeblocked.
–Keeptheopeningsfreefromlint,hairandsimilar
foreignitems.
–Keepconcentratoratleast12in(30.5cm)awayfrom
walls,draperiesandfurniture.
–Avoiduseinpresenceofpollutants,smokeorfumes,
flammableanesthetics,cleaningagentsorchemical vapors.
–Placeconcentratorinawellventilatedareatoavoid
airbornepollutantsand/orfumes.
–DONOTuseinacloset.
Youmayselectaroominyourhousewhereusingyouroxygen concentratorwouldbemostconvenient.Yourconcentratorcanbe easilyrolledfromroomtoroomonitscasters.
Youroxygenconcentratorwillperformbestwhenoperatedunder theconditionsoutlinedinSpecifications.
Usageinenvironmentsotherthanthosedescribedmayresultinthe needforincreasedequipmentmaintenance.Theairintakeofthe unitshouldbelocatedinawellventilatedareatoavoidairborne pollutantsand/orfumes.
1193322-A-0417
Invacare®Perfecto™VOxygenConcentrator

6.3SettingUptheConcentrator

1.Pluginpowercordtoanelectricaloutlet.
2.ConnectHumidifier(ifsoprescribed).

6.3.1SettingUptheHumidifier

WARNING! RiskofInjuryorDamage
Toavoidburnsfromsteamorhotwater,inhalationof waterand/orwaterdamagetotheconcentrator:
–DONOTfillhumidifierbottlewithhotwater.Allow
boiledwatertocooltoroomtemperaturebefore filling.
DONOToverfillhumidifier.Replacethehumidifiercapandsecurelytighten.
Confirmthatthecapisnotcross-threadedonthe humidifierbottle.
–DONOTreversetheoxygeninputandoutput
connections.Waterfromthehumidifierbottlewill travelthroughthecannulabacktothepatientifinput andoutputconnectionsarereversed.
–Whenusingtubingconnectionslongerthan7ft
(2.1m)inlength,positionthehumidifierasclose tothepatientaspossibletoallowformaximum humidificationoutput.
WARNING! RiskofInjuryorDamage
Failuretoproperlyinstallthehumidifierbottleorother accessoriestotheconcentratorwillimpacttheflowof oxygen.
–Tocheckforproperoperationoftheoxygen
concentratorandaccessories,placetheendofthe nasalcannulaunderthesurfaceofahalf-fullcupof water,onlywhentheoxygenisflowing,andlook forbubblesindicatingproperoperation.Wipeany excesswateroffthenasalprongswhenfinishedbefore placingonthepatient.
–Iftherearenobubbles,checkallconnections
(includinghumidifierbottleandotheraccessories,if applicable)andrepeat.Contactyourdealerorservice supplierimmediatelyifbubblesstilldonotappear.
181193322-A-04
Usage
HumidifierBottle
HumidifierBottlewithoutCap
withCap
1.RemovecapAfrombottleB.
2.Fillhumidifierwithboiledtapwaterorbottledwatertothelevel indicatedbythemanufacturer.Boiltapwaterforapproximately tenminutesandcooltoroomtemperaturepriortouse.
CAUTION! RiskofDamage
–Confirmthehumidifiercapisnotcross-threaded
onthehumidifierbottle.
3.Insertaflatheadscrewdriverintheplategrooveonthetopedge ofthefilteraccessdoorCandgentlyprythefilteraccessdoor off.
4.PullupandremovethehumidifierbottleadapterD.
5.AttachthehumidifierbottleadapterDtothehumidifier bottleBbyturningthewingnutEonthehumidifierbottle counterclockwiseuntilitissecurelyfastened.
1193322-A-0419
Invacare®Perfecto™VOxygenConcentrator
6.Selectthelocationforthehumidifierbottle/adapterassembly:
WhenplacingthehumidifierbottleinthecompartmentF
ontheconcentrator,attachthehumidifierbottle/adapter tubeGtotheoxygenoutletconnectorHonthe concentrator.
Whenusingtubingconnectionslongerthan7ft(2.1m) inlength,positionthehumidifierasclosetothepatient aspossibletoallowformaximumhumidificationoutput.
7.Attachthepatient’snasalcannulasupplytubetothehumidifier bottleoutlet.
8.Afterassembly,ensurethatoxygenisflowingthroughthenasal cannula.

6.3.2TurningtheConcentratorOn

1.PresspowerswitchAtoOnposition.
Allthepanellightsandtheaudiblealarmwillcome onforonesecond,confirmingthattheindicatorsare functioningproperly.

6.3.3Flowrate

WARNING! RiskofInjury
Toavoidinjuryfromexcessoxygenoradeficitof oxygen:
–DONOTchangetheL/minsettingontheflowmeter
unlessachangehasbeenprescribedbyyourphysician ortherapist.
201193322-A-04
2.Iftheflowrateontheflowmetereverfallsbelow0.5L/minDfor morethanaboutoneminute,thePotentialObstructionAlert willbetriggered.Thisisarapidbeepingoftheaudiblealarm. Checkyourtubingoraccessoriesforblockedorkinkedtubing oradefectivehumidifierbottle.Afterratedflowisrestoredto between0.5L/minand0.75L/min,thePotentialObstruction Alertwillturnoff.
APotentialObstructionAlertindicatesacondition thatmaybeassociatedwithapartialorcomplete obstructionofoxygenoutput.
Theuseofsomeaccessoriessuchasthe PreciseRx™PediatricFlowmeterandtheHomeFill compressorwilldeactivatethePotentialObstruction Alert.
Usage
DonotsettheflowabovetheREDring.Anoxygenflow greaterthan5L/minwilldecreasetheoxygenconcentration.
1.TurntheflowrateknobAtothesettingprescribedbyyour physicianortherapist.
ToproperlyreadtheflowmeterB,locatethe prescribedflowratelineontheflowmeter.Next,turn theflowknobuntiltheballCrisestotheline.Now, centertheballontheL/minlineprescribed.
1193322-A-04
6.3.4SensO
OxygenPurityIndicator
2
Thisfeaturemonitorsthepurityofoxygengeneratedbytheoxygen concentrator.Ifpurityfallsbelowfactorypresetstandards,indicator lightsonthecontrolpanelwillilluminate.

6.3.5InitialStartupoftheConcentrator

Concentratormaybeusedduringtheinitialwarm-uptime (approximately30min.)whilewaitingfortheO reachmaximum.
Whentheunitisturnedon,theyellowlightwillcomeonindicating theunitisoperatingandisintheinitialwarm-upperiod.Whenthe greenlightcomeson,theunitisinnormaloperatingmode(SYSTEM OK/O
greaterthan85%).Onceinnormaloperatingmode,thelights
2
willbecontrolleddependingontheoxygenconcentrationvalues.
2
purityto
21
Invacare®Perfecto™VOxygenConcentrator

6.3.6ExplanationofOxygenPurityIndicatorLights

Symbol
O
Purity
SYSTEM
A
O
2
OKAY
O
2
85%
O
B
between 73%and
85%
Indicator
2
Light
(LED)
Noactionis
GREEN
over
light
needed.Normal foroperation.
Immediatelycall supplier.Youmay
2
YELLOW
light
(Solid)
continuetouse theconcentrator unlessinstructed otherwisebyyour supplier.Becertain thatbackupoxygen isnearby.
Symbol
O
Purity
Indicator
2
Light
(LED)
GREEN
O
AB
2
O
unknown
lightwith
2
YELLOW
light
flashing
Explanation
Immediately callsupplier. Oxygensensoris malfunctioning.You maycontinuetouse theconcentrator.
ContinuousAudible Alarm
Explanation
Sieve-GARD™ Compressor
SYSTEM
C
FAILURE
O
below
2
REDlight
(Solid)
73%
Shutdown
SYSTEMFAILURE. Totalunit shutdown. Immediatelyswitch tobackupoxygen supplyandcall supplier.
D
CallSupplier
22
1193322-A-04
Maintenance

7Maintenance

7.1ServiceLife

WARNING! RiskofInjuryorDamage
Theproducthasbeentestedfortheservicelifestatedin thismanual.Useoftheproductbeyondthistimeperiod maycauseinjuryorproductdamage.
–ONLYusetheproductfortheservicelifestatedin
thismanual.DONOTexceedtheservicelifeofthe product.
–Performallmaintenanceaccordingtothe
recommendedscheduleinthismanual.
Theexpectedservicelifeofthisproductisthreeyearsofoperation whenusedinaccordancewiththesafetyinstructions,maintenance intervalsandcorrectuse,statedinthismanual.Theeffectiveservice lifecanvaryaccordingtofrequencyandintensityofuse.Referto theMaintenancesection.

7.2CleaningtheCabinet

DANGER! RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned tominimizeroutinepreventivemaintenance.Toprevent injuryordamage:
–Onlyqualifiedpersonnelshouldperformpreventive
maintenanceontheoxygenconcentrator.
–DONOTremovecabinet.
DANGER! RiskofInjuryorDamage
Liquidwilldamagetheinternalcomponentsofthe concentrator.Toavoiddamageorinjuryfromelectrical shock:
–TurnOfftheconcentratorandunplugthepowercord
beforecleaning.
–DONOTallowanycleaningagenttodripinsidethe
airinletandoutletopenings.
–DONOTsprayorapplyanycleaningagentdirectly
tothecabinet.
–DONOThosedowntheproduct.
CAUTION! RiskofDamage
Harshchemicalagentscandamagetheconcentrator. Toavoiddamage:
–DONOTcleanthecabinetorfilterwithalcohol
andalcoholbasedproducts(isopropylalcohol), concentratedchlorine-basedproducts(ethylene chloride),andoil-basedproducts(Pine-Sol®,Lestoil®) oranyotherharshchemicalagents.Onlyusemild liquiddishdetergent(suchasDawn®).
Ataminimum,preventivemaintenanceMUSTbeperformed accordingtothemaintenancerecordguidelines.Inplaces withhighdustorsootlevels,maintenancemayneedto beperformedmoreoften.RefertothePreventative MaintenanceChecklist.
1193322-A-0423
Invacare®Perfecto™VOxygenConcentrator
Periodicallycleantheconcentrator’scabinetasfollows:
1.Useadampcloth,orsponge,withamilddetergentsuchas Dawndishwashingsoaptogentlycleantheexteriorcase.
2.Allowtheconcentratortoairdry,oruseadrytowel,before operatingtheconcentrator.

7.3CleaningtheCabinetFilter

CAUTION! RiskofDamage
Toavoiddamagetotheinternalcomponentsoftheunit:
–DONOToperatetheconcentratorwithoutthefilter
installedorwithadirtyfilter.
Thereisonecabinetfilterlocatedonthebackofthecabinet.
1.Removethefilterandcleanasneeded.
Environmentalconditionsthatmayrequiremore frequentinspectionandcleaningofthefilterinclude,but arenotlimitedto:highdust,airpollutants,etc.
2.Cleanthecabinetfilterwithavacuumcleanerorwashwith amildliquiddishdetergent(suchasDawn)andwater.Rinse thoroughly.
3.Thoroughlydrythefilterandinspectforfraying,crumbling,tears andholes.Replacefilterifanydamageisfound.
4.Reinstallthecabinetfilter.
7.4HumidifierCleaningandThermic
Disinfection
Cleananddisinfecttheoxygenhumidifierdailytoreduce limestonedepositsandeliminatepossiblebacterial contamination.Followtheinstructionsprovidedbythe manufacturer.Ifnoneareprovided,followthesesteps:
1.Washhumidifierinsoapywaterandrinsewithasolutionoften partswaterandonepartvinegar.
2.Rinsethoroughlywithhotwater.
3.Airdrythoroughly.
Tolimitbacterialgrowth,airdrythehumidifierthoroughly aftercleaningwhennotinuse.RefertoSetUpforuse.
24
1193322-A-04

7.5CleaningandDisinfectionBetweenPatients

WARNING! RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto concentrator:
–Onlyqualifiedpersonnelshouldperformcleaning
anddisinfectionoftheoxygenconcentratorand accessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossiblepathogen exchangebetweenpatientsduetocontaminationof componentsoraccessories.Preventivemaintenanceshould alsobeperformedatthistimeifnecessary.
1.Disposeofandreplaceallpatientsideaccessoriesnotsuitable formultiplepatientuse,includingbutnotlimitedto:
NasalCannulaandTubing
Mask
Humidifier
2.Performmaintenanceproceduresdescribedinthismanualand itemsonthePreventiveMaintenanceChecklist
3.Checkconcentratorforpossibleexternaldamageorsignsthatit mayrequireserviceorrepair.
4.Ensureconcentratorfunctionsproperlyandallalarmsarein workingorder.
5.Beforerepackaginganddistributiontonewpatient,ensure packagingcontentscontaintheconcentratorandusermanual.
Maintenance
1193322-A-0425
Invacare®Perfecto™VOxygenConcentrator

7.6PreventiveMaintenanceChecklist

ModelNo:
ONEACHINSPECTION
RecordDateofService
RecordElapsedHoursonHourMeter
CleanCabinetFilter(s)(RefertoCleaningtheCabinetFilter.)
CheckPrescribedL/min.Flowrate
DURINGPREVENTIVEMAINTENANCESCHEDULE,ORBETWEENPATIENTS
Every3yearsofcontinuoususe(Equivalentto26,280hours)
CheckOxygenConcentration(greenindicatorlight)
Clean/ReplaceCabinetFilter(s)(RefertoCleaningtheCabinetFilter.)
Check/ReplaceOutletHEPAFilter*
Check/ReplaceCompressorInletFilter*
CheckPowerLossAlarm*
*Tobeconductedbysupplierorqualifiedservicetechnician.Refertoservicemanual.
SerialNo:
261193322-A-04
AfterUse

8AfterUse

8.1RecyclingInformation

Thisproductmaycontainsubstancesthatcouldbeharmfultothe environmentifdisposedofinplaces(landfills)thatarenotappropriate accordingtolegislation.
Followlocalgoverningordinancesandrecyclingplansregarding disposaloftheconcentratororcomponentsnormallyusedin operation.Theconcentratordoesnotgeneratewasteorresiduein operation.
DONOTdisposeoftheconcentratorinthenormalwaste
stream.
AnyaccessoriesnotpartoftheconcentratorMUSTbehandled
inaccordancewiththeindividualproductmarkingfordisposal.

8.2WearandTear

Invacarereservestherighttoaskforanyitembackthathasan allegeddefectinworkmanship.SeeWarrantythatshippedwiththe productforspecificwarrantyinformation.
Refertothisusermanualforproperpreventativemaintenance scheduleanduseoftheproduct.
Thisisjustageneralguidelineanddoesnotincludeitemsdamaged duetoabuseandmisuse.
Normalwearandtearitemsandcomponentsforthisproductare listedbelow.
Alltypesoffilters
Alltypesoftubing
Sieveisaporousfilteringmaterialandisconsideredawear item.Somefactorsthatcouldaffectsievemateriallifeinclude humidity,temperature,particulates,aircontaminates,air intake,vibrationandotherenvironmentalconditions.
1193322-A-0427
Invacare®Perfecto™VOxygenConcentrator

9Troubleshooting

9.1Troubleshooting

SYMPTOMPROBABLECAUSESOLUTION
Alarm:MainPowerLoss:
Nolightsilluminated.
ShortBeeps,longpause
Concentratornotoperating, powerswitchOn.
Beep...Beep...
Alarm:
REDlightilluminated.
Continuous
Concentratornotoperating, powerswitchOn.
Beep...
1.Powercordnotpluggedin.
2.Nopoweratoutlet.
3.Trippedcircuitbreaker.
SystemFailure:
1.Unitoverheatingduetoblocked
2.Insufficientpoweratoutlet.
3.Internalrepairsrequired.
1.Insertplugintooutlet.
2.Inspecthousecircuitbreakersorfuses.Ifproblemrecurs, useadifferentoutlet.
3.Push/resetcircuitbreaker.Ifproblemrecurs,callservice supplier.
1.Dooneorbothofthefollowing:
airintake.
a.Removeandcleancabinetfilters. b.Moveoxygenconcentratoratleast12in(30.5cm)away
fromwalls,draperiesorfurniture.
2.DONOTuseextensioncords.Movetoanotherelectrical outletorcircuit.
3.Callservicesupplier.
281193322-A-04
Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:
GREENlightilluminated.
Rapid
Beep...Beep...
Beep...Beep
PotentialObstructionAlert
1.Possibleinternalobstructioninthe oxygenpath.Kinkedorblocked tubing,cannulaorhumidifier.
2.Flowmetersetat0.5L/minorless.
1.Inspectforkinksorblockages.Correct,cleanorreplace item.Oncecorrected,turnpowerOfffor60secondsand thenturnpowerbackOn.
2.Resetflowmetertoprescribedflowrate.Ifprescribed flowrateisbelow0.5L/min,usethePediatricFlowmeter IRCPF16.
TheuseofthePediatricFlowmeterwilldeactivatethe PotentialObstructionAlert.
Alarm:
YELLOWlightIlluminated.
Shortbeeps,longpause
Beep...........Beep.............
SystemFailure:
1.Lowoxygenpurity.
2.Kinkedorblockedtubing,cannula orhumidifier.
3.Flowmetersetat0.5L/minorless.
1.CleanorReplacefilters.
2.Inspectforkinksorblockages.Correct,cleanorreplace item.Oncecorrected,turnpowerOfffor60secondsand thenturnpowerbackOn.
3.Checkflowmeterissetto1.0L/minormore.Referto Specifications.
GREENlightwithYELLOWLight
1.Internalrepairsrequired.1.Callservicesupplier.
Flashing.
1193322-A-0429
Invacare®Perfecto™VOxygenConcentrator

10Technicaldata

10.1Specifications

ElectricalRequirements:
RatedCurrentInput:3A
SoundPressureLevel:< 53dBAweightedfor3L/minflow
SoundPowerLevel:< 53dBAweightedfor5L/minFlow
Altitude:Upto8,000ft(2438m)abovesealevelwithoutdegradationof
OxygenOutputConcentrationLevels:87%to95.6%at0.5to5L/min
MaximumOutletPressure:5psi±0.5psi(34.5kPa±3.45kPa)
LowFlowAlarm:0.5to5L/min(maximum).Forflowrateslessthan1L/min,we
PotentialObstructionAlert:0L/minto0.5L/min
301193322-A-04
120VAC+10,–15%(132VAC/102VAC),60Hz
<53dBAweightedfor5L/minFlow
concentrationlevels.
WhilefillingaHomeFillSystem:Upto6,000ft(1828m)above sealevelwithoutdegradationofconcentrationlevels.
AtmosphericPressureRange:101.3kPa–75.0kPa
Concentrationlevelsachievedafterinitialwarm-upperiod (approximately30minutes)
recommendtheuseoftheInvacarePediatricFlowmeterAccessory (IRCPF16).
Theconcentratordetectsaconditionthatmayindicateapotential obstructionoftheoutputoxygen.Rapidaudiblebeepingalert(this alertisdeactivatedwhenaccessoriesareconnected)maybeassociated withflowsettingof0.5L/minorless.
Technicaldata
PowerConsumption:325WTypical
PressureReliefMechanismOperationalat:30psito45psi(207kPato310kPa)
Changeinmaximumrecommendedflowwhenbackpressureof7kPais
0.7L/min(BackPressureof7kPaapplied)
applied:
Filters:
Cabinet,OutletHEPAandCompressorInlet
SafetySystem:Currentoverloadorlinesurgeshutdown.
Hightemperaturecompressorshutdown.
HighPressureAlarmwithcompressorshutdown.
LowPressureAlarmwithcompressorshutdown.
BatteryFreePowerLossAlarm.
SensO
OxygenSystem.
2
PossibleObstructionAlert.
Width:15in±3/8in(38.1cm±1cm)
Height:
Depth:
23in±3/8in(58.4cm±1cm)
12in±3/8in(30.5cm±1cm)
Weight:40lbs±2lbs(18.1kg±1kg)
ShippingWeight:45lbs±2lbs(20.4kg±1kg)
OperatingAmbientTemperatureandHumidity:50°Fto95°F(10°Cto35°C)atupto60%relativehumidity
Cabinet:ImpactResistantflame-retardantplasticcabinetthatconformstoUL
94V–0
StandardsandRegulatoryListing:ETLcertifiedcomplyingwithAAMIES60601-1,IEC60601-1,IEC
60601-1-2,IEC60601-1-6,IEC60601-1-11,CSAC22.2#60601-1,ISO 80601-2-69
1193322-A-0431
Invacare®Perfecto™VOxygenConcentrator
Electrical:Noextensioncords.
Placement:
Nocloserthan12in(30.5cm)fromanywall,furniture,draperiesor similarsurfacestoassuresufficientairflow.
Avoiddeeppilecarpetsandheaters,radiatorsorhotairregisters.
Floorlocationonly.
Noconfinedspaces(Example:Noclosets).
Tubing:
7ft(2.1m)cannulawithamaximum50ft(15m)ofCrush-ProofTubing
(DONOTpinch)
TimeofOperation:Upto24hoursperday
RecommendedStorageandShippingTemperature:-20°Fto150°F(-29°Cto65°C)atupto95%relativehumidity
Measurementuncertaintyisincludedinthedevicespecification.AllconditionsatATPD.
321193322-A-04
Warranty

11Warranty

11.1NorthAmericanLimitedWarranty

THEWARRANTYBELOWHASBEENDRAFTEDTOCOMPLYWITH FEDERALLAWAPPLICABLETOPRODUCTSMANUFACTURED AFTERJULY4,1975.
Thiswarrantyisextendedonlytotheoriginalpurchaserwho purchasesthisproductwhennewandunusedfromInvacare Corporationoradealer.Thiswarrantyisnotextendedtoany otherpersonorentityandisnottransferableorassignabletoany subsequentpurchaserorowner.Coverageunderthiswarrantywill enduponanysuchsubsequentsaleorothertransferoftitletoany otherperson.Thiswarrantygivesyouspecificlegalrightsandyou mayalsohaveotherlegalrightswhichmayvaryfromstatetostate.
InvacareCorporationwarrantsitsPerfecto whenpurchasednewandunusedtobefreefromdefectsinmaterials andworkmanshipforaperiodofthreeyearsfromdateofpurchase fromInvacareoradealer,withacopyoftheseller’sinvoicerequired forcoverageunderthiswarranty.
DONOTOPENORATTEMPTTOSERVICETHE4-WAYVALVE BODY.
Thiswillvoidanyandallwarrantyapplicabletothevalvebody.If withinsuchwarrantyperiodsanysuchproductshallbeprovento InvacareCorporation'ssatisfactiontobedefective,suchproduct shallberepairedorreplaced,atInvacareCorporation'soption. Thiswarrantyonlyappliestothelaborforrepairsperformedby theInvacareServiceDepartmentorInvacareAuthorizedService Centers.Itdoesnotapplytothelaborperformedbythepurchaser oruser.Thiswarrantydoesnotincludenormalwearandtearor shippingcharges.InvacareCorporation'ssoleobligationandyour
OxygenConcentrators
2
exclusiveremedyunderthiswarrantyshallbelimitedtosuchrepair orreplacement.
Routinemaintenanceitems,suchasfilters,areexcludedfrom thiswarranty.Forwarrantyservice,pleasecontactInvacare Corporation'sServiceDepartmentatthetollfreenumberonthe backcover.Uponreceivingnoticeofanallegeddefectinaproduct, InvacareCorporationwillissueaserializedreturnauthorization. Itshallthenbetheresponsibilityofthepurchasertoreturnthe entireunitorremove,atpurchaser'scost,thedefectivecomponent part(s)identified,packthecomponentpart(s)inamannertoavoid shippingdamageandtoshipthecomponentpart(s)toeither InvacareCorporation'splantorservicecenterasspecifiedby InvacareCorporationinadvance.Defectivecomponentpart(s) mustbereturnedforwarrantyinspectionusingtheserialnumber asidentificationwithinthirtydaysofreturnauthorizationdate.Do notreturnproductstoourfactorywithoutpriorconsent.C.O.D. shipmentswillberefused;pleaseprepayshippingcharges.
LIMITATIONSANDEXCLUSIONS:THEFOREGOING WARRANTYSHALLNOTAPPLYTOPRODUCTSSUBJECTED TONEGLIGENCE,ACCIDENT,IMPROPEROPERATION, MAINTENANCEORSTORAGE,SOOTORSMOKE-FILLED ENVIRONMENTS,OROTHERTHANNORMALAPPLICATION, USEORSERVICE,ORTOPRODUCTSMODIFIEDWITHOUT INVACARECORPORATION'SEXPRESSWRITTENCONSENT (INCLUDING,BUTNOTLIMITEDTO,MODIFICATION THROUGHTHEUSEOFUNAUTHORIZEDPARTSOR ATTACHMENTS)ORTOPRODUCTSDAMAGEDBYREASON OFREPAIRSMADETOANYCOMPONENTWITHOUTTHE SPECIFICCONSENTOFINVACARECORPORATIONOR TOPRODUCTSDAMAGEDBYCIRCUMSTANCESBEYOND INVACARECORPORATION'SCONTROL.
1193322-A-0433
Invacare®Perfecto™VOxygenConcentrator
THEFOREGOINGEXPRESSWARRANTYISEXCLUSIVEANDIN LIEUOFANYOTHERWARRANTIESWHATSOEVER,WHETHER EXPRESSORIMPLIED,INCLUDINGTHEIMPLIEDWARRANTIES OFMERCHANTABILITYANDFITNESSFORAPARTICULAR PURPOSE,ANDTHESOLEREMEDYFORVIOLATIONSOFANY WARRANTYWHATSOEVER,SHALLBELIMITEDTOREPAIROR REPLACEMENTOFTHEDEFECTIVEPRODUCTPURSUANTTO THETERMSCONTAINEDHEREIN.THEAPPLICATIONOFANY IMPLIEDWARRANTYWHATSOEVERSHALLNOTEXTEND BEYONDTHEDURATIONOFTHEEXPRESSWARRANTY PROVIDEDHEREIN.INVACARESHALLNOTBELIABLEFORANY CONSEQUENTIALORINCIDENTALDAMAGESWHATSOEVER.
SOMESTATESDONOTALLOWTHEEXCLUSIONOR LIMITATIONOFINCIDENTALORCONSEQUENTIALDAMAGE, ORLIMITATIONOFHOWLONGANIMPLIEDWARRANTY LASTS,SOTHEABOVEEXCLUSIONANDLIMITATION MAYNOTAPPLYTOYOU.THISWARRANTYSHALLBE EXTENDEDTOCOMPLYWITHSTATE/PROVINCIALLAWS ANDREQUIREMENTS.
341193322-A-04
Notes
InvacareCorporation
USA
OneInvacareWay Elyria,OhioUSA 44035–2125 Tel:440–329–6000 Tel:800–832–4707 TechnicalServices Tel:440-329-6593 Tel:800-832-4707 www.invacare.com
1193322-A-042016-06-17
*1193322A-04*
Manufacturer
InvacareCorporation 2101E.LakeMaryBlvd. Sanford,FL32773 Tel:407–321–5630
Manufacturer
InvacareRehabilitation Equipment(Suzhou)Co.,Ltd. No.5WeixiRoad,SIP, Suzhou,Jiangsu,PRC215121 Tel:86–512–62586180 Fax:86–512–62586167
MakingLife’sExperiencesPossible™