Invacare®Perfecto
Concentrator
IRC5PO2V
enHomeFill®SystemCompatible
UserManual
₂
₂ ₂
™VOxygen
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.

©2016Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithout
priorwrittenpermissionfromInvacare.Trademarksareidentifiedby™and®.Alltrademarksare
ownedbyorlicensedtoInvacareCorporationoritssubsidiariesunlessotherwisenoted.
MakingLife’sExperiencesPossibleisatrademarkintheU.S.A.
Pine-solandLestoilareregisteredtrademarksofTheCloroxCompany.
DawnisaregisteredtrademarkofTheProctorandGambleCompany.

Contents
1General........................................4
1.1Symbols.....................................4
1.2IntendedUse..................................5
1.3IndicationsForUse.............................6
1.4Description...................................6
1.5Contraindications..............................6
2Safety..........................................7
2.1LabelLocations................................7
2.2GeneralGuidelines.............................8
2.3RadioFrequencyInterference......................11
3Components....................................12
3.1ComponentIdentification.........................12
3.2PneumaticDiagram.............................14
4Accessories.....................................15
4.1OptionalAccessories............................15
5Setup..........................................16
5.1Unpacking....................................16
5.2Inspection....................................16
5.3Storage......................................16
6Usage..........................................17
6.1Introduction..................................17
6.2SelectaLocation...............................17
6.3SettingUptheConcentrator......................18
6.3.1SettingUptheHumidifier......................18
6.3.2TurningtheConcentratorOn...................20
6.3.3Flowrate..................................20
6.3.4SensO
OxygenPurityIndicator.................21
2
6.3.5InitialStartupoftheConcentrator................21
6.3.6ExplanationofOxygenPurityIndicatorLights........22
7Maintenance....................................23
7.1ServiceLife...................................23
7.2CleaningtheCabinet............................23
7.3CleaningtheCabinetFilter........................24
7.4HumidifierCleaningandThermicDisinfection..........24
7.5CleaningandDisinfectionBetweenPatients............25
7.6PreventiveMaintenanceChecklist...................26
8AfterUse.......................................27
8.1RecyclingInformation............................27
8.2WearandTear................................27
9Troubleshooting.................................28
9.1Troubleshooting...............................28
10Technicaldata..................................30
10.1Specifications.................................30
11Warranty......................................33
11.1NorthAmericanLimitedWarranty.................33

Invacare®Perfecto₂™VOxygenConcentrator
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
–Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
–Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
–Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
SymbolsinDocumentation
GeneralWarningSign
Thebackgroundcolorinsidethe
triangleisyellowonproductlabels.
Readmanual
Thecolorofthesymbolbackground
isblueonproductlabels.
NoSmoking
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
NoOpenFlame
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
ClassII,DoubleInsulated
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
IndoorUseONLY
KeepDry
4
1193322-A-04

TransportandStorageTemperature
TransportandStorageHumidity
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
O
Indicators
2
SymbolO
O
2
PurityIndicatorLights(LED)
2
SYSTEM
GREENIndicatorLight
OKAY
O
over85%
2
O
73%to85%
between
2
YELLOWIndicatorLight
A.YELLOWSolid
B.YELLOWFlashingSensor
Failure.Callaqualifiedtechnician.
SYSTEM
FAILURE
O
Below73%
2
REDIndicatorLight
RefertoTroubleshooting.
General
ElectricalHazard
1.2IntendedUse
TheintendedfunctionanduseoftheInvacare®Perfecto₂™V
Manufacturer
OxygenConcentratoristoprovidesupplementaloxygentopatients
withrespiratorydisorders,byseparatingnitrogenfromroomair,by
wayofamolecularsieve.Itisnotintendedtosustainorsupportlife.
Callsupplier
DANGER!
RiskofInjuryorDeath
SymbolsonProduct
Unitrunning
Thisproductistobeusedasanoxygensupplementand
isnotintendedtobelife-supportingorlife-sustaining.
ONLYusethisproductifthepatientiscapableof
spontaneousbreath,abletoinhaleandexhalewithout
Unitnotrunning
theuseofamachine.
–DONOTuseinparallelorserieswithotheroxygen
concentratorsoroxygentherapydevices.
1193322-A-045

Invacare®Perfecto₂™VOxygenConcentrator
DANGER!
RiskofInjuryorDeath
Dependingontheirmedicalcondition,patientsonflow
ratesgreaterthan5L/minmaybeatanincreasedrisk
forseriousinjuryordeathintheeventoffailure.
–ALWAYSdiscussthisincreasedriskwithyourhealth
caresupplierBEFOREusingthisproductifyouare
prescribedaflowrategreaterthan5L/min.
WARNING!
RiskofInjuryorDamage
Useofthisproductoutsideoftheintendeduseand
specificationshasnotbeentestedandmayleadto
productdamage,lossofproductfunction,orpersonal
injury.
–DONOTusethisproductinanywayotherthan
describedintheSpecificationsandIntendedUse
sectionsofthismanual.
DANGER!
RiskofInjuryorDeath
WhileInvacarestrivestoproducethebestoxygen
concentratorinthemarkettoday,thisoxygen
concentratorcanfailtoproduceoxygenduetopower
failureordevicemalfunction.
–ALWAYShaveabackupsourceofoxygenreadily
available.
–Intheeventtheconcentratorfailstoproduceoxygen,
theconcentratorwillbrieflyalarmsignalingthepatient
toswitchtotheirbackupsourceofoxygen.Referto
Troubleshootingformoredetail.
1.3IndicationsForUse
TheintendedfunctionanduseoftheInvacare®Perfecto₂™V
OxygenConcentratoristoprovidesupplementaloxygentopatients
withrespiratorydisorders,byseparatingnitrogenfromroomair,by
wayofamolecularsieve.Itisnotintendedtosustainorsupportlife.
1.4Description
TheInvacarePerfecto
respiratorydisorderswhorequiresupplementaloxygen.Thedevice
isnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87%
to95.6%.Theoxygenisdeliveredtothepatientthroughtheuse
ofanasalcannula.
TheInvacarePerfecto
pressureswingadsorptionmethodologytoproducetheoxygen
gasoutput.Ambientairentersthedevice,isfilteredandthen
compressed.Thiscompressedairisthendirectedtowardoneof
twonitrogenadsorbingsievebeds.
Concentratedoxygenexitstheoppositeendoftheactivesievebed
andisdirectedintoanoxygenreservoirwhereitisdeliveredtothe
patient.
TheInvacarePerfecto
patientinahomeenvironmentorinaninstitutionalenvironment.
Deviceoperatesatanominal120VAC/60Hertzsupply.
Serviceinformationwillbeavailableuponrequesttoqualified
technicalpersonnelonly.
concentratorisusedbypatientswith
2
concentratorusesamolecularsieveand
2
concentratoriscapableofoperationbythe
2
1.5Contraindications
Therearenoknowncontraindications.
61193322-A-04

Safety
2Safety
2.1LabelLocations
A
B
C
D
1193322-A-047
Serialnumberlabelislocatedontheresonator
intakeassembly.
SpecificationLabelislocatedonthebackofthe
concentratoratthebase.

Invacare®Perfecto₂™VOxygenConcentrator
2.2GeneralGuidelines
Inordertoensurethesafeinstallation,assemblyandoperationofthe
concentratortheseinstructionsMUSTbefollowed.
DANGER!
RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor
damage.Thissectioncontainsimportantinformationfor
thesafeoperationanduseofthisproduct.
–DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasusermanuals,service
manualsorinstructionsheetssuppliedwiththis
productoroptionalequipment.
–Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional,
dealerortechnicalpersonnelbeforeattemptingto
usethisequipment.
–CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot
workingcorrectly,contactatechnicianorInvacare
forrepair.
–THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy
substancesandothercombustiblesareeasilyignited
andburnwithgreatintensityinoxygenenrichedairand
whenincontactwithoxygenunderpressure.Toavoid
fire,death,injuryordamage:
–DONOTSMOKEwhileusingthisdevice.
–DONOTusenearOPENFLAMEorIGNITION
SOURCES.
–DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
–NOSMOKINGsignsshouldbeprominentlydisplayed.
–Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby
anytypeoffriction.
–Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutofthe
roominwhichthisconcentratorislocatedandaway
fromwhereoxygenisbeingdelivered.
–Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings,
chaircushions,clothing,andawayfromheatedorhot
surfacesincludingspaceheaters,stoves,andsimilar
electricalappliances.
CAUTION!
Federal(statutory)lawrestrictsthisdevicesaletoor
ontheorderofamedicalpractitionerlicensedbya
governmentalagencywherehe/shepractices.
–ONLYalicensedmedicalpractitionermayorderthe
purchaseoruseofthisdevice.
81193322-A-04

Safety
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor
injurytopersons:
–DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceable
parts.
–Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroomat
least7ft(2.1m)fromthebath.
–DONOTcomeincontactwiththeconcentrator
whilewet.
–DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
–DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
–DONOTusefrayedordamagedACpowercords.
WARNING!
RiskOfInjuryOrDamage
–Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
–Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis
device.Youshouldcontactyourlocalhomecare
supplierforrecommendationsonwhichofthese
deviceswillbebestforyou.Theyshouldalsogiveyou
adviceontheproperusage,maintenance,andcleaning.
DANGER!
RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway
contamination:
–DONOTusetheconcentratorinthepresenceof
pollutants,smoke,fumes,flammableanesthetics,
cleaningagents,orchemicalvapors.
WARNING!
RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
–Closelysupervisewhenthisconcentratorisusedby
ornearchildrenorimpairedindividuals.
–Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
–Keepchildrenandpetsawayfromnasalcannulaand
tubing.
–Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/orimpairedpersons.
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
–Replacethenasalcannulaonaregularbasis.Check
withyourequipmentsupplierorphysicianto
determinehowoftenthecannulashouldbereplaced.
–DONOTsharecannulasbetweenpatients.
1193322-A-049

Invacare®Perfecto₂™VOxygenConcentrator
WARNING!
RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto
you.Topreventoxygendeprivation:
–Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings
shouldbechanged.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
–DONOTmoveorrelocateconcentratorbypulling
onthecord.
–DONOTuseextensioncordswithACpowercords
provided.
–Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
–NEVERleaveconcentratorunattendedwhenplugged
in.
–Makesureconcentratorisoffwhennotinuse.
WARNING!
RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned
tominimizeroutinepreventivemaintenance.Toprevent
injuryordamage:
–Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas
factorytrainedpersonnelshouldperformpreventive
maintenanceorperformanceadjustmentsonthe
oxygenconcentrator,exceptfortasksdescribedin
thismanual.
–UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION!
RiskofDamage
Topreventdamagefromliquidingress:
–Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto
water,callequipmentsupplier/qualifiedtechnicianfor
examinationandrepair.
–NEVERdroporinsertanyobjectorliquidintoany
opening.
–ForindooruseONLY.
101193322-A-04

2.3RadioFrequencyInterference
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrom
interferencewithwirelessequipment:
–Keepconcentratoratleast9.8ft(3.0m)awayfrom
wirelesscommunicationequipmentsuchaswireless
homenetworkdevices,mobilephones,cordless
phonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimits
specifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovide
areasonableprotectionagainstelectromagneticinterferenceina
typicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevels
ofelectromagneticemissionspermittedbytheabovestandards.
Todetermineiftheemissionsfromtheconcentratorarecausing
theinterference,turntheconcentratorOff.Iftheinterference
withtheotherdevice(s)stops,thentheconcentratoriscausingthe
interference.Insuchrarecases,interferencemaybereducedor
correctedbyoneofthefollowingmeasures:
•Reposition,relocate,orincreasetheseparationbetweenthe
equipment.
•Connecttheequipmentintoanoutletonacircuitdifferentfrom
thattowhichtheotherdevice(s)isconnected.
Safety
1193322-A-04
11

Invacare®Perfecto₂™VOxygenConcentrator
3Components
3.1ComponentIdentification
A
B
C
D
E
F
OxygenOutlet
Flowmeter
ElapsedTimeMeter
OxygenPurityIndicator
Lights/FaultandPowerIndicator
Lights
CircuitBreaker
PowerSwitch
12
1193322-A-04

Components
RearView
G
H
I
PowerCord
HomeFillOutletFitting
CabinetFilter
TheHomeFilloutletfittingHistobeusedonlyfor
fillingoxygencylinderswiththeHomeFillhomeoxygen
compressor.Theoutletfittingdoesnotaffectconcentrator
performance.RefertotheHomeFillusermanual,part
number1100873,forconnectionandoperatinginstructions.
Whennotinuse,theplugprovidedwiththeconcentrator
shouldbeinsertedintotheoutletfitting.Formore
informationabouttheHomeFill,contactyourInvacare
dealer.
1193322-A-0413

Invacare®Perfecto₂™VOxygenConcentrator
3.2PneumaticDiagram
14
1193322-A-04

4Accessories
4.1OptionalAccessories
Therearemanydifferenttypesofhumidifiers,oxygentubing,cannulas
andmasksthatcanbeusedwiththisdevice.Youshouldcontact
yourlocalhomecaresupplierforrecommendationsonwhichof
thesedeviceswillbebestforyou.Theyshouldalsogiveyouadvice
ontheproperusage,maintenance,andcleaning.
Thesupplyaccessories(nasalcannula,mask,tubing,humidifier,etc.)
usedtodeliveroxygentothepatientneedtoincludeameansto
reducethepropagationoffireintheaccessoriesforthesafetyofthe
patientandothers.Ifacommerciallyavailable,fire-activatedflow
stopdeviceisusedintheaccessoriessetup,itshouldbeplacedas
closetothepatientaspracticable.
Thefollowingoptionalaccessoriesandreplacementparts(listed
below)arealsoavailable:
•StandardAdultNasalCannula,7ft(2.1m)-M3120
•StandardHumidifier-InvacarePart#1155719
•PreciseRx™PediatricFlowmeterAccessory-IRCPF16
•HomeFillHomeOxygenCompressor-IOH200
•OxygenSupplyTubing7ft(2.1m)-MS4107
•OxygenSupplyTubing21ft(6.3m)-MS4121
•OxygenTubingConnector-MS4301
Accessories
1193322-A-0415

Invacare®Perfecto₂™VOxygenConcentrator
5Setup
5.1Unpacking
1.Checkforanyobviousdamagetothecartonoritscontents.If
damageisevident,notifythecarrier,oryourlocaldealer.
2.Removeallloosepackingfromthecarton.
3.Carefullyremoveallthecomponentsfromthecarton.
Unlesstheoxygenconcentratoristobeused
IMMEDIATELY,leaveconcentratorinitspackagingfor
storageuntiluseisrequired.
5.2Inspection
1.Inspect/examineexterioroftheconcentratorfornicks,dents,
cracks,scratchesorotherdamage.
2.Inspectallcomponents.
5.3Storage
1.Storetherepackagedoxygenconcentratorinadryarea.Refer
toSpecificationsforstoragetemperatureparameters.
2.DONOTplaceobjectsontopofrepackagedconcentrator.
161193322-A-04

Usage
6Usage
6.1Introduction
Youroxygenconcentratorisintendedforindividualuseindoors.
Itisanelectronicallyoperateddevicethatseparatesoxygenfrom
roomair.Itprovideshighconcentrationofoxygendirectlytoyou
throughanasalcannula.Clinicalstudieshavedocumentedthat
oxygenconcentratorsaretherapeuticallyequivalenttoothertypesof
oxygendeliverysystems.
Yoursupplierwillshowyouhowtouseyouroxygenconcentrator.
He/Sheshouldbecontactedwithanyquestionsorproblems
regardingyouroxygenconcentrator.Thisusermanualwilltellyou
aboutyourconcentratorandwillserveasareferenceasyouuse
yourconcentrator.
6.2SelectaLocation
WARNING!
RiskofInjury
Toavoidinjuryduringtherapy:
–TheoxygenconcentratorMUSTbeplacedonalevel
surfaceforuse.
–DONOTrelocatetheoxygenconcentratorwhilein
use.
WARNING!
RiskofInjuryorDamage
Toavoidinjuryordamagefromairbornepollutants
and/orfumesandforoptimalperformance:
–Locateandpositiontheconcentratorinawell
ventilatedspacesothattheairintakeandtheair
exhaustsarenotobstructed.
–NEVERblocktheairopeningsoftheconcentrator
orplaceitonasoftsurface,suchasabedorcouch,
wheretheairopeningmaybeblocked.
–Keeptheopeningsfreefromlint,hairandsimilar
foreignitems.
–Keepconcentratoratleast12in(30.5cm)awayfrom
walls,draperiesandfurniture.
–Avoiduseinpresenceofpollutants,smokeorfumes,
flammableanesthetics,cleaningagentsorchemical
vapors.
–Placeconcentratorinawellventilatedareatoavoid
airbornepollutantsand/orfumes.
–DONOTuseinacloset.
Youmayselectaroominyourhousewhereusingyouroxygen
concentratorwouldbemostconvenient.Yourconcentratorcanbe
easilyrolledfromroomtoroomonitscasters.
Youroxygenconcentratorwillperformbestwhenoperatedunder
theconditionsoutlinedinSpecifications.
Usageinenvironmentsotherthanthosedescribedmayresultinthe
needforincreasedequipmentmaintenance.Theairintakeofthe
unitshouldbelocatedinawellventilatedareatoavoidairborne
pollutantsand/orfumes.
1193322-A-0417

Invacare®Perfecto₂™VOxygenConcentrator
6.3SettingUptheConcentrator
1.Pluginpowercordtoanelectricaloutlet.
2.ConnectHumidifier(ifsoprescribed).
6.3.1SettingUptheHumidifier
WARNING!
RiskofInjuryorDamage
Toavoidburnsfromsteamorhotwater,inhalationof
waterand/orwaterdamagetotheconcentrator:
–DONOTfillhumidifierbottlewithhotwater.Allow
boiledwatertocooltoroomtemperaturebefore
filling.
–DONOToverfillhumidifier.
–Replacethehumidifiercapandsecurelytighten.
Confirmthatthecapisnotcross-threadedonthe
humidifierbottle.
–DONOTreversetheoxygeninputandoutput
connections.Waterfromthehumidifierbottlewill
travelthroughthecannulabacktothepatientifinput
andoutputconnectionsarereversed.
–Whenusingtubingconnectionslongerthan7ft
(2.1m)inlength,positionthehumidifierasclose
tothepatientaspossibletoallowformaximum
humidificationoutput.
WARNING!
RiskofInjuryorDamage
Failuretoproperlyinstallthehumidifierbottleorother
accessoriestotheconcentratorwillimpacttheflowof
oxygen.
–Tocheckforproperoperationoftheoxygen
concentratorandaccessories,placetheendofthe
nasalcannulaunderthesurfaceofahalf-fullcupof
water,onlywhentheoxygenisflowing,andlook
forbubblesindicatingproperoperation.Wipeany
excesswateroffthenasalprongswhenfinishedbefore
placingonthepatient.
–Iftherearenobubbles,checkallconnections
(includinghumidifierbottleandotheraccessories,if
applicable)andrepeat.Contactyourdealerorservice
supplierimmediatelyifbubblesstilldonotappear.
181193322-A-04

Usage
HumidifierBottle
HumidifierBottlewithoutCap
withCap
1.RemovecapAfrombottleB.
2.Fillhumidifierwithboiledtapwaterorbottledwatertothelevel
indicatedbythemanufacturer.Boiltapwaterforapproximately
tenminutesandcooltoroomtemperaturepriortouse.
CAUTION!
RiskofDamage
–Confirmthehumidifiercapisnotcross-threaded
onthehumidifierbottle.
3.Insertaflatheadscrewdriverintheplategrooveonthetopedge
ofthefilteraccessdoorCandgentlyprythefilteraccessdoor
off.
4.PullupandremovethehumidifierbottleadapterD.
5.AttachthehumidifierbottleadapterDtothehumidifier
bottleBbyturningthewingnutEonthehumidifierbottle
counterclockwiseuntilitissecurelyfastened.
1193322-A-0419

Invacare®Perfecto₂™VOxygenConcentrator
6.Selectthelocationforthehumidifierbottle/adapterassembly:
•WhenplacingthehumidifierbottleinthecompartmentF
ontheconcentrator,attachthehumidifierbottle/adapter
tubeGtotheoxygenoutletconnectorHonthe
concentrator.
Whenusingtubingconnectionslongerthan7ft(2.1m)
inlength,positionthehumidifierasclosetothepatient
aspossibletoallowformaximumhumidificationoutput.
7.Attachthepatient’snasalcannulasupplytubetothehumidifier
bottleoutlet.
8.Afterassembly,ensurethatoxygenisflowingthroughthenasal
cannula.
6.3.2TurningtheConcentratorOn
1.PresspowerswitchAtoOnposition.
Allthepanellightsandtheaudiblealarmwillcome
onforonesecond,confirmingthattheindicatorsare
functioningproperly.
6.3.3Flowrate
WARNING!
RiskofInjury
Toavoidinjuryfromexcessoxygenoradeficitof
oxygen:
–DONOTchangetheL/minsettingontheflowmeter
unlessachangehasbeenprescribedbyyourphysician
ortherapist.
201193322-A-04

2.Iftheflowrateontheflowmetereverfallsbelow0.5L/minDfor
morethanaboutoneminute,thePotentialObstructionAlert
willbetriggered.Thisisarapidbeepingoftheaudiblealarm.
Checkyourtubingoraccessoriesforblockedorkinkedtubing
oradefectivehumidifierbottle.Afterratedflowisrestoredto
between0.5L/minand0.75L/min,thePotentialObstruction
Alertwillturnoff.
APotentialObstructionAlertindicatesacondition
thatmaybeassociatedwithapartialorcomplete
obstructionofoxygenoutput.
Theuseofsomeaccessoriessuchasthe
PreciseRx™PediatricFlowmeterandtheHomeFill
compressorwilldeactivatethePotentialObstruction
Alert.
Usage
DonotsettheflowabovetheREDring.Anoxygenflow
greaterthan5L/minwilldecreasetheoxygenconcentration.
1.TurntheflowrateknobAtothesettingprescribedbyyour
physicianortherapist.
ToproperlyreadtheflowmeterB,locatethe
prescribedflowratelineontheflowmeter.Next,turn
theflowknobuntiltheballCrisestotheline.Now,
centertheballontheL/minlineprescribed.
1193322-A-04
6.3.4SensO
OxygenPurityIndicator
2
Thisfeaturemonitorsthepurityofoxygengeneratedbytheoxygen
concentrator.Ifpurityfallsbelowfactorypresetstandards,indicator
lightsonthecontrolpanelwillilluminate.
6.3.5InitialStartupoftheConcentrator
Concentratormaybeusedduringtheinitialwarm-uptime
(approximately30min.)whilewaitingfortheO
reachmaximum.
Whentheunitisturnedon,theyellowlightwillcomeonindicating
theunitisoperatingandisintheinitialwarm-upperiod.Whenthe
greenlightcomeson,theunitisinnormaloperatingmode(SYSTEM
OK/O
greaterthan85%).Onceinnormaloperatingmode,thelights
2
willbecontrolleddependingontheoxygenconcentrationvalues.
2
purityto
21

Invacare®Perfecto₂™VOxygenConcentrator
6.3.6ExplanationofOxygenPurityIndicatorLights
Symbol
O
Purity
SYSTEM
A
O
2
OKAY
O
2
85%
O
B
between
73%and
85%
Indicator
2
Light
(LED)
Noactionis
GREEN
over
light
needed.Normal
foroperation.
Immediatelycall
supplier.Youmay
2
YELLOW
light
(Solid)
continuetouse
theconcentrator
unlessinstructed
otherwisebyyour
supplier.Becertain
thatbackupoxygen
isnearby.
Symbol
O
Purity
Indicator
2
Light
(LED)
GREEN
O
AB
2
O
unknown
lightwith
2
YELLOW
light
flashing
Explanation
Immediately
callsupplier.
Oxygensensoris
malfunctioning.You
maycontinuetouse
theconcentrator.
ContinuousAudible
Alarm
Explanation
Sieve-GARD™
Compressor
SYSTEM
C
FAILURE
O
below
2
REDlight
(Solid)
73%
Shutdown
SYSTEMFAILURE.
Totalunit
shutdown.
Immediatelyswitch
tobackupoxygen
supplyandcall
supplier.
D
——
CallSupplier
22
1193322-A-04

Maintenance
7Maintenance
7.1ServiceLife
WARNING!
RiskofInjuryorDamage
Theproducthasbeentestedfortheservicelifestatedin
thismanual.Useoftheproductbeyondthistimeperiod
maycauseinjuryorproductdamage.
–ONLYusetheproductfortheservicelifestatedin
thismanual.DONOTexceedtheservicelifeofthe
product.
–Performallmaintenanceaccordingtothe
recommendedscheduleinthismanual.
Theexpectedservicelifeofthisproductisthreeyearsofoperation
whenusedinaccordancewiththesafetyinstructions,maintenance
intervalsandcorrectuse,statedinthismanual.Theeffectiveservice
lifecanvaryaccordingtofrequencyandintensityofuse.Referto
theMaintenancesection.
7.2CleaningtheCabinet
DANGER!
RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned
tominimizeroutinepreventivemaintenance.Toprevent
injuryordamage:
–Onlyqualifiedpersonnelshouldperformpreventive
maintenanceontheoxygenconcentrator.
–DONOTremovecabinet.
DANGER!
RiskofInjuryorDamage
Liquidwilldamagetheinternalcomponentsofthe
concentrator.Toavoiddamageorinjuryfromelectrical
shock:
–TurnOfftheconcentratorandunplugthepowercord
beforecleaning.
–DONOTallowanycleaningagenttodripinsidethe
airinletandoutletopenings.
–DONOTsprayorapplyanycleaningagentdirectly
tothecabinet.
–DONOThosedowntheproduct.
CAUTION!
RiskofDamage
Harshchemicalagentscandamagetheconcentrator.
Toavoiddamage:
–DONOTcleanthecabinetorfilterwithalcohol
andalcoholbasedproducts(isopropylalcohol),
concentratedchlorine-basedproducts(ethylene
chloride),andoil-basedproducts(Pine-Sol®,Lestoil®)
oranyotherharshchemicalagents.Onlyusemild
liquiddishdetergent(suchasDawn®).
Ataminimum,preventivemaintenanceMUSTbeperformed
accordingtothemaintenancerecordguidelines.Inplaces
withhighdustorsootlevels,maintenancemayneedto
beperformedmoreoften.RefertothePreventative
MaintenanceChecklist.
1193322-A-0423

Invacare®Perfecto₂™VOxygenConcentrator
Periodicallycleantheconcentrator’scabinetasfollows:
1.Useadampcloth,orsponge,withamilddetergentsuchas
Dawndishwashingsoaptogentlycleantheexteriorcase.
2.Allowtheconcentratortoairdry,oruseadrytowel,before
operatingtheconcentrator.
7.3CleaningtheCabinetFilter
CAUTION!
RiskofDamage
Toavoiddamagetotheinternalcomponentsoftheunit:
–DONOToperatetheconcentratorwithoutthefilter
installedorwithadirtyfilter.
Thereisonecabinetfilterlocatedonthebackofthecabinet.
1.Removethefilterandcleanasneeded.
Environmentalconditionsthatmayrequiremore
frequentinspectionandcleaningofthefilterinclude,but
arenotlimitedto:highdust,airpollutants,etc.
2.Cleanthecabinetfilterwithavacuumcleanerorwashwith
amildliquiddishdetergent(suchasDawn)andwater.Rinse
thoroughly.
3.Thoroughlydrythefilterandinspectforfraying,crumbling,tears
andholes.Replacefilterifanydamageisfound.
4.Reinstallthecabinetfilter.
7.4HumidifierCleaningandThermic
Disinfection
Cleananddisinfecttheoxygenhumidifierdailytoreduce
limestonedepositsandeliminatepossiblebacterial
contamination.Followtheinstructionsprovidedbythe
manufacturer.Ifnoneareprovided,followthesesteps:
1.Washhumidifierinsoapywaterandrinsewithasolutionoften
partswaterandonepartvinegar.
2.Rinsethoroughlywithhotwater.
3.Airdrythoroughly.
Tolimitbacterialgrowth,airdrythehumidifierthoroughly
aftercleaningwhennotinuse.RefertoSetUpforuse.
24
1193322-A-04

7.5CleaningandDisinfectionBetweenPatients
WARNING!
RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto
concentrator:
–Onlyqualifiedpersonnelshouldperformcleaning
anddisinfectionoftheoxygenconcentratorand
accessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossiblepathogen
exchangebetweenpatientsduetocontaminationof
componentsoraccessories.Preventivemaintenanceshould
alsobeperformedatthistimeifnecessary.
1.Disposeofandreplaceallpatientsideaccessoriesnotsuitable
formultiplepatientuse,includingbutnotlimitedto:
•NasalCannulaandTubing
•Mask
•Humidifier
2.Performmaintenanceproceduresdescribedinthismanualand
itemsonthePreventiveMaintenanceChecklist
3.Checkconcentratorforpossibleexternaldamageorsignsthatit
mayrequireserviceorrepair.
4.Ensureconcentratorfunctionsproperlyandallalarmsarein
workingorder.
5.Beforerepackaginganddistributiontonewpatient,ensure
packagingcontentscontaintheconcentratorandusermanual.
Maintenance
1193322-A-0425

Invacare®Perfecto₂™VOxygenConcentrator
7.6PreventiveMaintenanceChecklist
ModelNo:
ONEACHINSPECTION
RecordDateofService
RecordElapsedHoursonHourMeter
CleanCabinetFilter(s)(RefertoCleaningtheCabinetFilter.)
CheckPrescribedL/min.Flowrate
DURINGPREVENTIVEMAINTENANCESCHEDULE,ORBETWEENPATIENTS
Every3yearsofcontinuoususe(Equivalentto26,280hours)
CheckOxygenConcentration(greenindicatorlight)
Clean/ReplaceCabinetFilter(s)(RefertoCleaningtheCabinetFilter.)
Check/ReplaceOutletHEPAFilter*
Check/ReplaceCompressorInletFilter*
CheckPowerLossAlarm*
*Tobeconductedbysupplierorqualifiedservicetechnician.Refertoservicemanual.
SerialNo:
261193322-A-04

AfterUse
8AfterUse
8.1RecyclingInformation
Thisproductmaycontainsubstancesthatcouldbeharmfultothe
environmentifdisposedofinplaces(landfills)thatarenotappropriate
accordingtolegislation.
Followlocalgoverningordinancesandrecyclingplansregarding
disposaloftheconcentratororcomponentsnormallyusedin
operation.Theconcentratordoesnotgeneratewasteorresiduein
operation.
•DONOTdisposeoftheconcentratorinthenormalwaste
stream.
•AnyaccessoriesnotpartoftheconcentratorMUSTbehandled
inaccordancewiththeindividualproductmarkingfordisposal.
8.2WearandTear
Invacarereservestherighttoaskforanyitembackthathasan
allegeddefectinworkmanship.SeeWarrantythatshippedwiththe
productforspecificwarrantyinformation.
Refertothisusermanualforproperpreventativemaintenance
scheduleanduseoftheproduct.
Thisisjustageneralguidelineanddoesnotincludeitemsdamaged
duetoabuseandmisuse.
Normalwearandtearitemsandcomponentsforthisproductare
listedbelow.
•Alltypesoffilters
•Alltypesoftubing
Sieveisaporousfilteringmaterialandisconsideredawear
item.Somefactorsthatcouldaffectsievemateriallifeinclude
humidity,temperature,particulates,aircontaminates,air
intake,vibrationandotherenvironmentalconditions.
1193322-A-0427

Invacare®Perfecto₂™VOxygenConcentrator
9Troubleshooting
9.1Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:MainPowerLoss:
Nolightsilluminated.
ShortBeeps,longpause
Concentratornotoperating,
powerswitchOn.
Beep...Beep...
Alarm:
REDlightilluminated.
Continuous
Concentratornotoperating,
powerswitchOn.
Beep...
1.Powercordnotpluggedin.
2.Nopoweratoutlet.
3.Trippedcircuitbreaker.
SystemFailure:
1.Unitoverheatingduetoblocked
2.Insufficientpoweratoutlet.
3.Internalrepairsrequired.
1.Insertplugintooutlet.
2.Inspecthousecircuitbreakersorfuses.Ifproblemrecurs,
useadifferentoutlet.
3.Push/resetcircuitbreaker.Ifproblemrecurs,callservice
supplier.
1.Dooneorbothofthefollowing:
airintake.
a.Removeandcleancabinetfilters.
b.Moveoxygenconcentratoratleast12in(30.5cm)away
fromwalls,draperiesorfurniture.
2.DONOTuseextensioncords.Movetoanotherelectrical
outletorcircuit.
3.Callservicesupplier.
281193322-A-04

Troubleshooting
SYMPTOMPROBABLECAUSESOLUTION
Alarm:
GREENlightilluminated.
Rapid
Beep...Beep...
Beep...Beep
PotentialObstructionAlert
1.Possibleinternalobstructioninthe
oxygenpath.Kinkedorblocked
tubing,cannulaorhumidifier.
2.Flowmetersetat0.5L/minorless.
1.Inspectforkinksorblockages.Correct,cleanorreplace
item.Oncecorrected,turnpowerOfffor60secondsand
thenturnpowerbackOn.
2.Resetflowmetertoprescribedflowrate.Ifprescribed
flowrateisbelow0.5L/min,usethePediatricFlowmeter
IRCPF16.
TheuseofthePediatricFlowmeterwilldeactivatethe
PotentialObstructionAlert.
Alarm:
YELLOWlightIlluminated.
Shortbeeps,longpause
Beep...........Beep.............
SystemFailure:
1.Lowoxygenpurity.
2.Kinkedorblockedtubing,cannula
orhumidifier.
3.Flowmetersetat0.5L/minorless.
1.CleanorReplacefilters.
2.Inspectforkinksorblockages.Correct,cleanorreplace
item.Oncecorrected,turnpowerOfffor60secondsand
thenturnpowerbackOn.
3.Checkflowmeterissetto1.0L/minormore.Referto
Specifications.
GREENlightwithYELLOWLight
1.Internalrepairsrequired.1.Callservicesupplier.
Flashing.
1193322-A-0429

Invacare®Perfecto₂™VOxygenConcentrator
10Technicaldata
10.1Specifications
ElectricalRequirements:
RatedCurrentInput:3A
SoundPressureLevel:< 53dBAweightedfor3L/minflow
SoundPowerLevel:< 53dBAweightedfor5L/minFlow
Altitude:Upto8,000ft(2438m)abovesealevelwithoutdegradationof
OxygenOutputConcentrationLevels:87%to95.6%at0.5to5L/min
MaximumOutletPressure:5psi±0.5psi(34.5kPa±3.45kPa)
LowFlowAlarm:0.5to5L/min(maximum).Forflowrateslessthan1L/min,we
PotentialObstructionAlert:0L/minto0.5L/min
301193322-A-04
120VAC+10,–15%(132VAC/102VAC),60Hz
<53dBAweightedfor5L/minFlow
concentrationlevels.
WhilefillingaHomeFillSystem:Upto6,000ft(1828m)above
sealevelwithoutdegradationofconcentrationlevels.
AtmosphericPressureRange:101.3kPa–75.0kPa
Concentrationlevelsachievedafterinitialwarm-upperiod
(approximately30minutes)
recommendtheuseoftheInvacarePediatricFlowmeterAccessory
(IRCPF16).
Theconcentratordetectsaconditionthatmayindicateapotential
obstructionoftheoutputoxygen.Rapidaudiblebeepingalert(this
alertisdeactivatedwhenaccessoriesareconnected)maybeassociated
withflowsettingof0.5L/minorless.

Technicaldata
PowerConsumption:325WTypical
PressureReliefMechanismOperationalat:30psito45psi(207kPato310kPa)
Changeinmaximumrecommendedflowwhenbackpressureof7kPais
0.7L/min(BackPressureof7kPaapplied)
applied:
Filters:
Cabinet,OutletHEPAandCompressorInlet
SafetySystem:Currentoverloadorlinesurgeshutdown.
Hightemperaturecompressorshutdown.
HighPressureAlarmwithcompressorshutdown.
LowPressureAlarmwithcompressorshutdown.
BatteryFreePowerLossAlarm.
SensO
OxygenSystem.
2
PossibleObstructionAlert.
Width:15in±3/8in(38.1cm±1cm)
Height:
Depth:
23in±3/8in(58.4cm±1cm)
12in±3/8in(30.5cm±1cm)
Weight:40lbs±2lbs(18.1kg±1kg)
ShippingWeight:45lbs±2lbs(20.4kg±1kg)
OperatingAmbientTemperatureandHumidity:50°Fto95°F(10°Cto35°C)atupto60%relativehumidity
Cabinet:ImpactResistantflame-retardantplasticcabinetthatconformstoUL
94V–0
StandardsandRegulatoryListing:ETLcertifiedcomplyingwithAAMIES60601-1,IEC60601-1,IEC
60601-1-2,IEC60601-1-6,IEC60601-1-11,CSAC22.2#60601-1,ISO
80601-2-69
1193322-A-0431

Invacare®Perfecto₂™VOxygenConcentrator
Electrical:Noextensioncords.
Placement:
Nocloserthan12in(30.5cm)fromanywall,furniture,draperiesor
similarsurfacestoassuresufficientairflow.
Avoiddeeppilecarpetsandheaters,radiatorsorhotairregisters.
Floorlocationonly.
Noconfinedspaces(Example:Noclosets).
Tubing:
7ft(2.1m)cannulawithamaximum50ft(15m)ofCrush-ProofTubing
(DONOTpinch)
TimeofOperation:Upto24hoursperday
RecommendedStorageandShippingTemperature:-20°Fto150°F(-29°Cto65°C)atupto95%relativehumidity
Measurementuncertaintyisincludedinthedevicespecification.AllconditionsatATPD.
321193322-A-04

Warranty
11Warranty
11.1NorthAmericanLimitedWarranty
THEWARRANTYBELOWHASBEENDRAFTEDTOCOMPLYWITH
FEDERALLAWAPPLICABLETOPRODUCTSMANUFACTURED
AFTERJULY4,1975.
Thiswarrantyisextendedonlytotheoriginalpurchaserwho
purchasesthisproductwhennewandunusedfromInvacare
Corporationoradealer.Thiswarrantyisnotextendedtoany
otherpersonorentityandisnottransferableorassignabletoany
subsequentpurchaserorowner.Coverageunderthiswarrantywill
enduponanysuchsubsequentsaleorothertransferoftitletoany
otherperson.Thiswarrantygivesyouspecificlegalrightsandyou
mayalsohaveotherlegalrightswhichmayvaryfromstatetostate.
InvacareCorporationwarrantsitsPerfecto
whenpurchasednewandunusedtobefreefromdefectsinmaterials
andworkmanshipforaperiodofthreeyearsfromdateofpurchase
fromInvacareoradealer,withacopyoftheseller’sinvoicerequired
forcoverageunderthiswarranty.
DONOTOPENORATTEMPTTOSERVICETHE4-WAYVALVE
BODY.
Thiswillvoidanyandallwarrantyapplicabletothevalvebody.If
withinsuchwarrantyperiodsanysuchproductshallbeprovento
InvacareCorporation'ssatisfactiontobedefective,suchproduct
shallberepairedorreplaced,atInvacareCorporation'soption.
Thiswarrantyonlyappliestothelaborforrepairsperformedby
theInvacareServiceDepartmentorInvacareAuthorizedService
Centers.Itdoesnotapplytothelaborperformedbythepurchaser
oruser.Thiswarrantydoesnotincludenormalwearandtearor
shippingcharges.InvacareCorporation'ssoleobligationandyour
OxygenConcentrators
2
exclusiveremedyunderthiswarrantyshallbelimitedtosuchrepair
orreplacement.
Routinemaintenanceitems,suchasfilters,areexcludedfrom
thiswarranty.Forwarrantyservice,pleasecontactInvacare
Corporation'sServiceDepartmentatthetollfreenumberonthe
backcover.Uponreceivingnoticeofanallegeddefectinaproduct,
InvacareCorporationwillissueaserializedreturnauthorization.
Itshallthenbetheresponsibilityofthepurchasertoreturnthe
entireunitorremove,atpurchaser'scost,thedefectivecomponent
part(s)identified,packthecomponentpart(s)inamannertoavoid
shippingdamageandtoshipthecomponentpart(s)toeither
InvacareCorporation'splantorservicecenterasspecifiedby
InvacareCorporationinadvance.Defectivecomponentpart(s)
mustbereturnedforwarrantyinspectionusingtheserialnumber
asidentificationwithinthirtydaysofreturnauthorizationdate.Do
notreturnproductstoourfactorywithoutpriorconsent.C.O.D.
shipmentswillberefused;pleaseprepayshippingcharges.
LIMITATIONSANDEXCLUSIONS:THEFOREGOING
WARRANTYSHALLNOTAPPLYTOPRODUCTSSUBJECTED
TONEGLIGENCE,ACCIDENT,IMPROPEROPERATION,
MAINTENANCEORSTORAGE,SOOTORSMOKE-FILLED
ENVIRONMENTS,OROTHERTHANNORMALAPPLICATION,
USEORSERVICE,ORTOPRODUCTSMODIFIEDWITHOUT
INVACARECORPORATION'SEXPRESSWRITTENCONSENT
(INCLUDING,BUTNOTLIMITEDTO,MODIFICATION
THROUGHTHEUSEOFUNAUTHORIZEDPARTSOR
ATTACHMENTS)ORTOPRODUCTSDAMAGEDBYREASON
OFREPAIRSMADETOANYCOMPONENTWITHOUTTHE
SPECIFICCONSENTOFINVACARECORPORATIONOR
TOPRODUCTSDAMAGEDBYCIRCUMSTANCESBEYOND
INVACARECORPORATION'SCONTROL.
1193322-A-0433

Invacare®Perfecto₂™VOxygenConcentrator
THEFOREGOINGEXPRESSWARRANTYISEXCLUSIVEANDIN
LIEUOFANYOTHERWARRANTIESWHATSOEVER,WHETHER
EXPRESSORIMPLIED,INCLUDINGTHEIMPLIEDWARRANTIES
OFMERCHANTABILITYANDFITNESSFORAPARTICULAR
PURPOSE,ANDTHESOLEREMEDYFORVIOLATIONSOFANY
WARRANTYWHATSOEVER,SHALLBELIMITEDTOREPAIROR
REPLACEMENTOFTHEDEFECTIVEPRODUCTPURSUANTTO
THETERMSCONTAINEDHEREIN.THEAPPLICATIONOFANY
IMPLIEDWARRANTYWHATSOEVERSHALLNOTEXTEND
BEYONDTHEDURATIONOFTHEEXPRESSWARRANTY
PROVIDEDHEREIN.INVACARESHALLNOTBELIABLEFORANY
CONSEQUENTIALORINCIDENTALDAMAGESWHATSOEVER.
SOMESTATESDONOTALLOWTHEEXCLUSIONOR
LIMITATIONOFINCIDENTALORCONSEQUENTIALDAMAGE,
ORLIMITATIONOFHOWLONGANIMPLIEDWARRANTY
LASTS,SOTHEABOVEEXCLUSIONANDLIMITATION
MAYNOTAPPLYTOYOU.THISWARRANTYSHALLBE
EXTENDEDTOCOMPLYWITHSTATE/PROVINCIALLAWS
ANDREQUIREMENTS.
341193322-A-04

Notes

InvacareCorporation
USA
OneInvacareWay
Elyria,OhioUSA
44035–2125
Tel:440–329–6000
Tel:800–832–4707
TechnicalServices
Tel:440-329-6593
Tel:800-832-4707
www.invacare.com
1193322-A-042016-06-17
*1193322A-04*
Manufacturer
InvacareCorporation
2101E.LakeMaryBlvd.
Sanford,FL32773
Tel:407–321–5630
Manufacturer
InvacareRehabilitation
Equipment(Suzhou)Co.,Ltd.
No.5WeixiRoad,SIP,
Suzhou,Jiangsu,PRC215121
Tel:86–512–62586180
Fax:86–512–62586167
MakingLife’sExperiencesPossible™