C |
RE |
INVACA |
FACTORY REPAIR CENTERS
(OHIO) |
|
|
CORPORATE |
HEADQUARTERS |
|
899 CLEVELAND STREET |
||
ELYRIA, OHIO |
44036 |
|
(ELORIDA) |
|
|
INVACARE |
RESPIRATORY PRODUCTS |
|
699 AERO |
LANE |
|
_ SANFORD, |
FLORIDA 32771 |
MOBILITE BY INVACARE 2101 E. LAKE MARY BLVD. SANFORD, FLORIDA 32773
(TEXAS)
INVACARE FORT WORTH, TX DIST. CENTER 1825 WESTPARK DRIVE
GRAND PRAIRIE, TX 75050
(CALIFORNIA)
INVACARE CALIFORNIA DIST. CENTER 14030 E. 183RD STREET
LA PALMA, CALIFORNIA 90623
INVACARE TECHNICAL SERVICES “HOTLINE” 1-800-832-4707
CHOOSE |
ONE OF THE FOLLOWING OPTIONS |
|
“1” |
FOR: |
ELECTRONICS |
“2” |
FOR: |
RESPIRATORY PRODUCTS |
“3” |
FOR: |
HOMECARE BEDS, RECLINERS, LIFT-OUT CHAIRS |
|
|
AND THERAPEUTIC SUPPORT SURFACES |
“4” |
FOR: |
ANYTHING ELSE |
INVACARE’ |
TECHNICAL TRAINING 1-800-333-6900 X6697 ELYRIA, OHIO |
— A |
|
S
H
I
P p
I
N
G
/
H
A
N ‘D L
р
N
G
SHIPPING AND HANDLING
|
SHIPPING |
AND |
HANDLING |
INSTRUCTIONS |
|
|
|
||
This |
Service Manual describes, in detail, |
the Invacare 3, 5 |
and 6 |
Oxygen Concentrators. |
It has |
been |
|||
carefully written to explain concentrator |
operation, service and |
preventative |
maintenance. Before |
||||||
you |
install and operate the concentrator, |
please refer to |
OWNER'S MANUAL. |
|
|
|
|||
The |
Invacare concentrator should always be kept in the upright position to prevent |
cabinet damage |
while |
||||||
being |
transported. The shipping container has been designed to assure . maximum |
protection |
of the |
con- |
|||||
centrator. |
|
|
|
|
|
|
|
|
|
If the |
concentrator is to be |
reshipped |
by common carrier, it should |
be packed in a |
new carton. |
Additional |
|||
cartons are available from |
Invacare. |
|
|
|
|
|
|
|
The air compressor suspension system has been engineered to withstand severe motion and orientation.
UNPACKING |
(FIGURE 1) |
1. Check for |
any obvious damage to the carton or its contents. İf damage is evident, notify the carrier, or |
Invacare. |
|
2. Remove ail loose packing from the carton.
3.Carefully remove all the components from the carton.
NOTE: After |
inspection, |
unless the Invacare |
Oxygen |
Concentrator is to be used immediately, re- |
tain carton |
and packing |
materials for use in |
storing |
until use of the concentrator is required. |
INSPECTION
1. Examine exterior of the Invacare Oxygen Concentrator for nicks, dents, scratches or other damages. inspect all components.
STORAGE
1. Store the repackaged Invacare-Oxygen Concentrator ina dry area. 2, DO NOT place other objects on top of the repackaged concentrator.
Relief |
for |
|
|
Relief |
for Line |
|
Flowmeter |
| |
. |
Cord |
to |
face |
|
to face |
|
Back |
of |
Unit. |
||
|
|
|
||||
Front |
of |
|
|
|
|
|
Unit. |
|
|
|
|
|
|
FIGURE 1 - UNPACKING
POWER CONSUMPTION FORMULA
If a customer inquires as to the amount of electricity needed to run a concenirator and / or their monthly operating costs, here’s the formula :
|
|
|
|
|
|
|
|
IRC301 |
|
|
|
|
|
|
|
|
Example |
Wattage Rating |
of Concentrator |
|
Watts |
350 |
||||
Hours per Day of Usage |
|
x |
Hours © |
12 |
||||
Number |
of |
Days |
in Month |
|
x |
Days |
30 |
|
Divide |
by |
1000 |
|
|
|
+ |
1000 |
|
~Equals |
: Kilowatt |
Hours |
(KWH) |
= |
KWH |
126 |
||
To Calculate Monthly Operating Cost: |
|
|
|
|||||
Obtain |
the |
KWH |
Rate |
from |
your |
|
|
|
electric |
bill |
and |
Multiply : |
|
x |
$/ KWH |
$.08 |
|
Equals: |
Monthly |
Operating |
Cost |
=$ |
|
$ 10.08 |
SEMINAR FEEDBACK FORM
[1 Home O2 Systems (PLEASE CHECK APPROPRIATE BOX)
Instructor:
Location:
Date:
In an effort to continually |
improve the |
quality |
of this |
seminar, we would like |
you |
to rate us |
and provide us with |
|||
your comments and suggestions regarding this course. |
|
|
|
|
|
|||||
Please check the |
box |
which |
most closely |
reflects |
your judgement |
of the |
following |
areas: |
||
Subject Area |
| |
Excellent |
|Above Avg.| |
Average |
| |
Below |
Avg. |
Poor |
1.Contents of Seminar
2.Instructor
3.Service Manuals
4.Format
5.Equipment and Materials
Additional Comments and Suggestions:
Optional Information
Name:
Company:
Phone:
TABLE OF CONTENTS
Section1
Features .....ceeerenmeecaerenasrrerentenneosncataaeraramiseareocanta carrera neon 1-1
τοι μμο |
μμ |
1-2 |
Installation and Sequence of Operation... 1-6 |
||
Sens 02 Illustration ......c.secceeerereeeeasareneescasenecarcorenaaenenasscoracoennerensenaeno 1-10 |
||
Pneumatic Diagram ΟΡ |
μμ 1-11 |
Section2
Preventive Maintenance Record |
.......sseccsseccessercerssennensssnnersssennnenaceucess |
2-1 |
|
4-Way Va e C7ea7n/9 ProOceOO |
aneaasanarzsasananz |
2-4 |
|
Compressor Rebuilding Instructions... |
2-8 |
||
7 o77gs 2679 |
6 |
rr rannsarneraonosa |
2-17 |
Section 3
7ou//e Зпоойпа RC |
307507 инииинииииииитенивниииние |
3-1 |
|
Trouble Shooting IRC |
3LX/5LX/6LX............... |
rennes |
3-9 |
Section4
Parts Replacement Procedures urnes 4-1 to 4-30
Section |
5 |
|
|
|
oo |
|
|
ezio vicricrinierii nia ieini crinicorinieniseozinisvinionenioni |
5-1 |
Mo/ex |
Co77ecfo7C 7 |
ri vzivivene |
5-2 |
|
Pilot Valve Poppet Replacement Kit Instructions cesser |
5-9 |
|||
Venture |
Homefill Literature |
0000050007000000 000000*400088 0 040 00000 00000 00000 …042 00000 8 0 04 |
4008 |
INVACARE |
TECHNICAL TRAINING 1-800-333-6900 X6697 ELYRIA, OHIO |
FEATURES
Power indicator
Oxygen Outlet
Power Switch
Circuit Breaker
Elapsed Time
Meter
Cabinet Filter
(2) ea.
FEATURES
amacapPmnan
FEATURES - 3LX, SLX
Power Indicator Sens02 Status |
Indicator Lights (5LXO2, |
5LXO2T, 6LXO2 and 6LXO2T ONLY) |
|
Oxygen Outlet С) |
|
Kl< |
|
д |
|
|
|
Circuit |
Breaker |
||
Power Switch |
|
|
|
|
1 |
|
|
|
|
Elapsed |
Time |
Cabinet Filter |
||
Meter |
||||
(2) |
ea. |
|||
|
|
FEATURES - 5LXO2, 5LXO2T, 6LXO2, 6LXO2T
1-1
MOBILAIRE Ill /V
SPECIFICATIONS OBILAIREModelI IRC301(Νορ,/ 501
MOBILAIRE lit / V #5:
SPECIFICATIONS, CONTINUED
Filters: |
Cabinet |
. |
|
|
|
|
|
Compressor Inlet |
|
|
|
||
|
In-Line |
|
|
|
|
|
|
Exhaust |
|
|
|
|
|
|
Bacteria |
|
|
|
|
|
Safety Systems: |
Current |
overload |
or |
line surge |
shutdown. |
|
|
High |
temperature |
compressor |
shutdown. |
||
|
High |
pressure alarm |
w/compressor shutdown. |
|||
|
Low pressure alarm w/compressor shutdown. |
|||||
|
Battery Free™ Power Loss Alarm |
|||||
|
Flow Sentry™ Flow Monitor |
|
||||
|
SensO2 |
Oxygen |
Monitoring System |
|
|
GRC30102 |
/ |
IRC50102) |
|||
Width: |
18-inches |
(45.7 |
cm) |
||||
Height: |
23.75-inches |
(60.3 |
cm) |
||||
Depth: |
14-inches |
(35.6 |
cm) |
||||
Weight: |
54 |
lbs. |
(24.5 |
kg) |
(IRC501/IRC50102) |
||
|
51 |
lbs. |
(23.2 |
kg) |
(IRC301 /IRC30102) |
||
Shipping Weight: |
64 |
lbs. |
(29.1 |
kg) |
(IRC501/IRC50102) |
||
|
61 |
Ibs. |
(27.7 |
kg) |
@RC301/IRC30102) |
Operating |
Temperature |
|
|
|
|
|
|
|
|
|
|
|
|
||
|
Exhaust: |
Less |
|
than |
Ambient |
+ 35°F |
(+16°C) |
|
|||||||
|
Oxygen Outlet: |
Less |
|
than |
Ambient |
+ |
4°F |
(+2°C) |
|
||||||
Cabinet: |
|
|
Impact resistant flame-retardant plastic |
cabinet. |
|||||||||||
Regulatory |
Listing: |
ETL |
|
certified |
complying |
with |
UL 544 |
and UL 1431 |
|||||||
Recommended End-User Guidelines for |
|
|
|
|
|
|
|
|
|
|
|
|
|||
Optimum |
|
Performance: |
|
|
|
|
|
|
|
|
|
|
|
|
|
‘Temperature: |
32° |
- |
95°F |
(0° |
- 35°C) |
- |
|
|
|
||||||
Electrical: |
|
No |
extension |
cords |
|
|
|
|
|
||||||
Placement: |
|
No |
closer than 3-inches |
from |
wall |
|
|||||||||
‘Tubing: |
|
|
50 ft. |
maximum |
(do |
NOT pinch) |
|
||||||||
Room |
Characteristics: — * |
Smoke and soot-free |
|
|
|
|
|
||||||||
|
|
|
|
No |
confined |
spaces |
(closet) |
|
|
|
|||||
Relative |
Humidity: |
20 to |
90% |
|
|
|
|
|
|
|
|
||||
Time |
of |
Operation: |
-Not |
less than 30-minutes |
at |
a |
time |
|
|||||||
|
|
|
|
Up to 24 hours per day |
|
|
|
|
|||||||
Flow |
Rate: |
|
From |
5 |
I/min. |
|
or 3 |
l/min. to 1/2 Y/min. For flows less |
|||||||
. |
|
|
|
than |
|
1/2 |
1/min., |
use |
Mobilaire |
Pediatric |
Flowmeter. |
> |
Call: Toll Free 1-800-333-6900 |
Innovation0Henk Care ~ |
1-2 |
|
MOBILAIRE Ill /V
Model IRC301 / 501
MOBILAIRE IlIll /V “a“Sensosenso,.
STANDARD SPECIFICATIONS
Electrical Requirements:
Rated Current Input:
Sound Level;
Operating Altitude:
Ambient ‘Temperature Range:
Oxygen Output:
.
Concentration Levels:
. s Alarm Threshoids:
(RC30102/ IRC50102)
Flow, adjustable in 0.5 liter increments:
Power Consumption:
Pressure Relief Mechanism operates at:
SPECIFICATIONS
SPECIFICATIONS
120 |
VAC, |
60 |
Hz |
|
|
|
|
|
|
|
|
|||||
5.0 А |
|
ERC501 |
/IRC50102) |
3.5 |
A(IRC301 / IRC30102) |
|||||||||||
52 |
dB |
|
Average |
([RC301 |
/ IRC30102) |
|
|
|
||||||||
53 |
dB |
|
Average |
(IRCS01 |
/ IRCS0102) |
|
|
|
||||||||
Up |
|
to |
|
6,000 |
ft. |
(1828 meters) above sea level without |
||||||||||
degradation |
of performance. From 6,000 ft. |
|
|
|||||||||||||
(1828 meters) to 13,129 ft. (4000 meters) below |
|
|
||||||||||||||
90% |
|
efficiency. |
|
|
|
|
|
|
|
|
||||||
32°F |
- |
95°F |
(0° |
- 35°C) |
|
|
|
|
|
|
|
|||||
1/2 |
to |
|
5 |
min., |
|
5 psi |
+ 0.5 psi |
|
|
. |
|
|
||||
(34.5 |
kPa + |
3.45 kPa) (IRC501 |
/ TRCSO102) |
|
|
|||||||||||
1/2 |
|
to |
|
3 |
/min,, |
5 psi + 0.5 psi |
|
|
|
|
|
|||||
(345 kPa + 3.45 kPa) (RC301 |
/ IRC30102) |
|
|
|||||||||||||
IRC501 |
/ IRC50102 |
|
|
IRC301 |
/ IRC30102 |
|
|
|||||||||
95% |
|
+ |
2% |
at |
1 |
Ymin. |
|
|
95% |
+ |
2% at 1-2 Yenin. |
|
||||
95% |
|
+ |
2% at |
2 |
l/min. |
|
|
92% |
+ |
3% |
at 3 /min. |
|
|
|||
95% |
|
+ |
2% |
at |
3 |
Ymin. |
|
|
|
|
|
|
|
|
||
94% |
|
+ |
3% |
at |
4 |
Ymin. |
|
|
|
|
|
|
|
|
||
91% |
|
= |
4% |
at |
5 |
/min. |
|
|
|
|
|
. |
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
Indicators |
|
|
||
|
|
|
|
|
|
|
|
|
Internal Switch |
|
Internal Switch |
|||||
O, |
Purity |
|
|
|
|
Set |
at |
73% |
|
|
Set |
at |
85% |
|||
Over |
85% |
(22%) |
|
|
Green |
© |
|
|
Green |
© |
||||||
73% |
(238) |
085% |
(22%) |
Yellow |
|
|
|
Red |
m |
|||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Continuous Audible Alarm |
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Sieve-GARD™ Compressor |
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Shut-Dowa |
||
Below |
73% |
(23%) |
|
|
Red |
m |
|
|
|
|
|
|||||
|
|
|
|
|
|
|
|
|
Continuous Audible |
Alarm: |
|
|
||||
|
|
|
|
|
|
|
|
|
Sieve-GARD™ Compressor |
|
|
|||||
|
|
|
|
|
|
|
|
|
Shut-Down |
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
- |
|
|
|
|
|
1/2 |
|
- 5 |
min. |
(ERCS01 / IRC50102) |
1/2 - 3 /min. |
ŒRC301 /IRC30102) |
||||||||||
450W |
|
(ARC501 |
/ IRC59102) |
350% |
(RC301 |
/ IRC30102) |
||||||||||
35 |
psi |
+ |
3.5 |
psi |
(241 |
kPa |
+ |
24.1 |
kPa) |
|
|
|
Call: Toll Free 1-800-333-6900
InnovationinHesich Care —
1-3
S P
E C 1
F
i C A T 1 O N S
SPECIFICATIONS
SPECIFICATIONS
STANDARD |
SPECIFICATIONS |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
Electrical |
Requirements: |
120 |
VAC |
+10% |
(132 VAC/108 VAC), 60 Hz |
|
|
|
|
|
|||||||||
Rated |
Current Input: |
3.6 |
A |
(3LX) |
|
3.4 A (5LX, 5LXO2, 5LXO2T, 6LX02, 6LXO2T) |
|||||||||||||
Sound |
Level: |
|
52 |
dB |
Average (3LX) |
|
|
|
|
|
|
|
|||||||
|
|
|
|
|
53 |
dB Average (SLX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T) |
|
||||||||||||
Operating |
Altitude: |
Up |
to |
6,000 ft. (1828 meters) above sea level without degradation of |
|||||||||||||||
|
|
|
|
|
performance. |
From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below |
|||||||||||||
|
|
|
|
|
90% |
efficiency. |
|
|
|
|
|
|
|
|
|
||||
Ambient |
Temperature Range: |
50°F |
- 95°F |
|
(10° |
|
- 35°C) |
|
|
|
|
|
|
||||||
*Oxygen |
Output |
Concentration |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
Levels: |
|
|
|
|
@LX, |
5LXO2, |
5LXO2T, |
6LXO2, |
GLXO2T) |
(3LX) |
|
|
|
|
|||||
|
|
|
|
|
95.6% |
to |
93% |
at |
1-3 Umin. |
|
95.6% |
to |
93% |
at |
1-2 L/min. |
||||
|
|
|
|
|
95.6% |
to |
91% |
at |
4 |
Limin. |
|
|
95.6% |
to |
87% |
at |
3 Limin. |
||
|
|
|
|
|
95.6% |
to |
87% |
at |
5 |
Limin. |
|
|
|
|
|
|
|
||
|
|
|
|
|
95.6% |
to |
87% |
at |
6 |
Umin.(6LXO2, |
6LX02T) |
|
|
|
|
|
|||
|
|
|
|
|
* Stated |
performance |
maximum achieved |
after 30 min. of run time. |
|||||||||||
Maximum |
|
Outlet |
Pressure: |
(LX, |
5LXO2, |
5LXOZT, |
6LXO2, |
6LXO2T) |
(SLX) |
|
|
|
|
||||||
|
|
|
|
|
5 psi |
+ 0.5 |
psi. |
|
|
|
|
5 psi+ |
0.5 psi. |
|
|||||
|
|
|
|
|
(34.5 |
kPa |
+ |
3.45 |
|
kPa.) |
|
|
(34.5 |
kPa |
+ |
3.45 |
kPa.) |
SensO2 Alarm Thresholds:
lf no internal selector switch, 73% operation ONLY.
Flow Range:
Power Consumption:
Pressure Relief Mechanism Operational at:
Indicators
02 Purity |
Internal |
Switch |
|
Internal |
Switch |
||||||||
|
|
|
|
|
*Set at 73% |
|
|
Set at |
85% |
||||
Over 85% |
(+ 2%) |
GREEN |
Indicator |
Light |
GREEN |
Indicator Light |
|||||||
73% (+ |
3%) |
to |
85% (42%) |
YELLOW |
Indicator |
Light |
RED |
indicator Light - |
|||||
|
|
|
|
|
|
|
|
|
|
Continuous. Audible |
|||
|
|
|
|
|
|
|
|
|
|
Alarm |
Sieve |
GARD™ |
|
|
|
|
|
|
|
|
|
|
|
Compressor |
Shutdown |
||
Below |
73% |
(+ |
3%) |
RED |
Indicator Light |
- |
|
|
N/A |
|
|||
|
|
|
|
Continuous |
Audible |
|
|
|
|
|
|||
|
|
|
|
Alarm |
Sieve |
GARD™ |
|
|
|
|
|||
|
|
|
|
Compressor |
Shutdown |
|
|
|
|
||||
|
|
|
|
* Factory Preset at 73% |
|
|
|
|
|||||
1/2 - 5 Umin. (LX, SLXO2, |
5LXOZT, |
6LXO2, |
6LXO2T) |
12 - 3 |
L/min. |
3LX) |
|||||||
For flow rates less than 1/2 |
L/min., use |
the |
Invacare |
Pediatric |
Flowmeter |
||||||||
PF16. |
|
|
|
|
|
|
|
|
|
|
|
|
|
460W (6LXO2, 6LXO2T) 350W/400W (LX)
400W (SLX, 5LX02, 5LX02T)
35 psi + 3.5 psi (241 kPa + 24.1 kPa)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SPECIFICATIONS |
|
|||
|
|
|
|
|
SPECIFICATIONS, |
CONTINUED |
|
|
|
|
5 |
||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Filters: |
|
|
|
Cabinet |
|
|
|
|
|
|
|
|
|
|
|
|
|
E |
|||
|
|
|
|
|
Compressor |
inlet |
|
|
|
|
|
|
|
|
|
|
|||||
|
|
|
|
|
Bacteria |
|
|
|
|
|
|
|
|
|
|
|
|
A |
|||
Safety |
Systems: |
|
Current overload or line surge shutdown. |
|
|
|
|
F |
|||||||||||||
|
|
|
|
|
High |
temperature |
compressor shutdown. |
|
|
|
|
| |
|||||||||
|
|
|
|
|
High pressure alarm w/compressor shutdown. |
|
|
|
© |
||||||||||||
|
|
|
|
|
Low pressure alarm w/compressor shutdown. |
|
|
|
|
A |
|||||||||||
|
|
|
|
|
Battery Free™ Power Loss Alarm. |
|
|
|
|
Es |
|||||||||||
|
|
|
|
|
SensO2 |
Oxygen |
System |
(5LX02, |
5LXO2T, 6LX02, |
6LXO02T |
ONLY). |
| |
|||||||||
Width: |
|
|
|
|
18-inches (45.7 cm) |
|
|
|
|
|
|
|
|
o |
|||||||
Height: |
|
|
|
23.75-inches (60.3 cm) |
|
|
|
|
|
|
|
N |
|||||||||
Depth: |
|
|
|
14-inches (35.6 cm) |
|
|
|
|
|
|
|
|
S |
||||||||
Weight: |
|
|
|
54 |
Ibs. |
(24.5 |
kg) |
(All |
SLX |
and 6LX Models) |
53 lbs. (23.9 kg.) (3LX) |
|
|||||||||
‘Shipping |
Weight: |
|
59 |
Ibs. |
(27 |
kg) (All |
SLX and |
6LX |
Models) 58 |
Ibs. |
(26.1 |
kg.) |
(SLX) |
|
|||||||
Operating |
Temperature |
|
|
|
|
|
|
. |
|
|
|
|
|
|
|
|
|
|
|
||
Exhaust: |
|
Less |
than |
Ambient |
+ |
359 |
F |
(+169 |
C) |
|
|
|
|
|
|||||||
„Oxygen |
Output: |
|. |
Less |
than |
Ambient |
+ |
49F |
(+29 |
C) |
|
|
|
|
|
|||||||
Cabinet: |
|
|
|
Impact Resistant flame-retardant plastic cabinet. |
|
|
|
|
|||||||||||||
Regulatory |
Listing: |
|
ETL certified complying with applicable portions of UL 1431 |
|
|
||||||||||||||||
|
RECOMMENDED END-USER GUIDELINES FOR OPTIMUM PERFORMANCE:. |
|
|
||||||||||||||||||
Temperature: |
|
50° |
- 95°F |
(10° |
- 35°C) |
|
|
|
|
|
|
|
|
||||||||
Electrical: |
|
|
No |
extension |
cords. |
|
|
|
|
|
|
|
|
|
|||||||
Placement: |
|
|
No closer than 3-inches from any wall. |
|
|
|
|
|
|||||||||||||
Tubing: |
|
|
|
Crush-Proof |
Tubing |
- 50 |
feet (15.2 m.) maximum |
(do |
NOT |
pinch). |
|
||||||||||
Room |
Characteristics: |
|
Smoke and soot-free |
|
|
|
|
|
|
|
|
|
|||||||||
|
|
|
|
|
No |
confined |
spaces |
(closet). |
|
|
|
||||||||||
Relative |
Humidity: |
|
20 |
to |
60% . |
|
|
|
|
|
|
|
|
|
|
|
|
||||
Time |
of |
Operation: |
|
Up |
to |
24 hours per |
day. |
|
|
|
|
|
|
|
|
1-5
INSTALLATION AND
SEQUENCE OF OPERATION
INSTALLATION AND SEQUENCE OF OPERATION
Model IRC30102 | IRC50102 — Serial Numbers 93D to Present
pMOBILAIRE Y Y Е Sa,
Model IRC301 — Serial Numbers 90E to Present
Model IRC501 — Serial Numbers 90F to Present
INSTALLATION
When |
your new |
Mobilaire concentrator attives, it should |
|
be checked |
for |
proper |
operating |
conditions. |
|
|||||||||
1. |
ifthe unit has been in below-freezing |
temperatures, allow |
it |
to |
warm |
up for |
30 minutes |
at room temperature |
||||||||||
|
before operating. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
2. |
The IRC301/501 will need to |
be turned on |
for |
a few |
seconds |
to |
charge the Battery Free™ Alarm Circuit. |
|||||||||||
|
Connect power cord to receptacle and |
tum |
on |
concentrator. |
Tum |
flow |
control knob |
counterclockwise |
and |
|||||||||
|
flow will begin immediately. |
Set flow |
of the Mobilaire |
HI |
to |
3 |
/min. |
and Mobilaire |
V |
to 5 l/min. Tum |
unit |
|||||||
|
off. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
3. |
With the power cord unplugged, press |
On/Off switch to the “on” position. An audible alarm will sound. This |
||||||||||||||||
|
checks the Battery Free“ Alarm Circuit. Tum |
switch |
|
“off.” |
|
|
|
|
|
|
|
|
||||||
4. |
Connect cord |
to receptacle and turn on |
concentrator. |
3 |
BEEPS |
on |
startup. Check the |
oxygen concentration |
||||||||||
|
per |
specifications after 30-40 |
minutes |
running |
time. |
|
|
|
|
|
|
|
|
|
|
|
SEQUENCE |
OF |
OPERATION |
|
|
|
|
|
|
|
|
|
|
|
||
Apply |
120 VAC |
via the power |
switch |
energizes |
the |
compressor motor, |
hour meter, |
cooling |
fan |
and the printed |
|||||
circuit |
(P.C.) |
board. |
|
|
|
|
|
|
|
|
|
|
|
|
|
Room |
air enters the compressor via the |
cabinet |
filter, |
air |
inlet |
filter, |
and |
sound |
muffler. The |
air |
is compressed by |
||||
the wobble pistons in the compressor to a pressure of 20 |
PSI |
(137.72 |
kPa), then |
put |
through |
an |
in-line filter. |
||||||||
*Units |
built prior |
to 90K will |
run at 22 |
PSI (151.5 kPa). |
|
|
|
|
|
|
|
|
As increased pressure creates increased temperature, a heat exchanger is utilized to lower the temperature by
20° F (-6.7° C) as the air enters the four-way solenoid valve. It is then channeled to a sieve bed containing the adsorption material. A restrictor downstream of the sieve bed causes pressure to build up inside the sieve bed which is necessary for the adsorption process. À small amount of relatively pure oxygen enters a storage tank with the balance entering the top of the second bed. The nitrogen removed is purged from the bed through the four-way
valve into room |
air. A |
filter is |
located at the |
exhaust-end of |
the valve |
to |
muffle |
the nitrogen |
exhaust |
as it enters |
|||||||||||
the |
air. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
The |
oxygen not |
being |
used |
to purge is channeled into |
a storage tank. The pressurized oxygen is regulated down |
||||||||||||||||
to 5 |
PSI |
(34.4 kPa), enters |
an |
accurate flow-measuring |
device, |
flows |
through |
abacterial filterand out to'thee pa |
|||||||||||||
‘tient. |
i |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
The |
electrical |
energization |
of |
the |
solenoid valve |
is accomplished |
every |
10 to 14 seconds by |
the pressure sensor | |
||||||||||||
and |
P.C. board |
|
electronics |
when |
the pressure |
set |
point |
of 20 |
PSIG |
+ |
3 |
(137.72 kPa+ 51.32 |
kPa) is |
reached. The |
|||||||
time.between |
cycles is |
dependent |
on'altitude |
and |
flow |
rate. |
A "spring |
in |
the valve |
allows |
it to |
return |
to itsoriginal |
||||||||
position when de-energized.* |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
On |
the |
IRC501 |
|
/ 50102, a |
2-way |
pressure equalization valve |
opens |
just |
prior |
to |
the shift |
of |
the4-way valve.This |
allows high purity oxygen to enter the same bed from the top. This additional pressure allows the bed to start its cycle at a higher pressure. The 2-way valve will close just after the shift of the 4-way valve.
INVACARE |
Cal Toll Free 1-800-333-6900 |
InnovationinHealth Cate = |
|
|
1-6 |
EN |
|
INSTALLATIONAND |
|
Model IRC30102 |
/50102 * |
SEOUENCE OF |
OPERATION |
|
INSTALLATION AND SEOUENCE |
OF OPERATION, CONTINUED |
|
If power is interrupted, the Battery Free™ alarm |
will sound a short “beep,” |
with a |
long pause, intermittently. |
All |
||
IRC301’s and IRC501’s are equipped with a diagnostic |
alarm system |
that |
signals |
if the system’s pressure or |
elec- |
|
trical system malfunctions. The‘Troubleshooting |
Guide |
explains the |
alarm |
system’s |
signals and reasons in detail |
|
for your convenience. |
|
|
|
|
|
|
SensO, OXGYEN SENSOR TECHNOLOGY CERAMIC ZIRCONIA SENSOR Technical Description
The oxygen being produced by the concentrator flows out of the product tank and into the flowmeter. In-line between these components is a tee fitting that directs a small flow of oxygen through a precision orifice to the oxygen sensor mounted on the printed circuit board.
As the oxygen enters the sensor housing, it passes through a screen and contacts the electrode mounted on the top of the zirconia disk.
Electric current flowing to the top electrode and through the zirconia disk to the bottom electrode heats the disk to over 300° C. Oxygen molecules on top of the disk pick-up extra electrons and become oxygen ions. These
oxygen ions |
are |
attracted to the electrode on the bottom |
of the |
zirconia |
disk. Because |
of the crystal |
structure |
of |
|||||
the |
zirconia, |
only |
oxygen ions can pass through. Once |
oxygen |
|
ions |
reach |
the |
bottom |
electrode, the |
extra electrons |
||
are |
released |
from |
the oxygen ions and oxygen molecules retum |
to |
the |
air. |
The |
number of electrons |
is directly |
re- |
lated to the oxygen concentration. The electrons travel to the PC board where they are “counted” and the oxygen concentration “reading” is calculated.
A micro-processor on the PC board contains software that interprets the signal being received from the sensor. It compares the signal to clinical acceptable limits that have been selected with the indicator switch mounted on the
PC board. Signals outside of the clinical acceptable limits |
generate |
responses |
in the form |
of |
lighted LED’s, |
au- |
|||||||||||||||
dible indicators, and/or system shut-down. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
Operating Sequence |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
Once |
the power |
switch has been |
turned to |
ON, |
the SensO2 |
circuit |
will |
wait |
5 |
minutes |
for |
the |
concentrator |
fo |
be- |
||||||
- gin to produce |
clinical |
acceptable |
oxygen |
before activating. No LED’s |
|
will |
illuminate |
for |
at |
least 5 minutes. |
|
||||||||||
After |
5 minutes, when oxygen purity exceeds 85% + 2% the green |
LED |
will |
illuminate. |
|
|
|
|
|
||||||||||||
“If the |
oxygen level |
is |
not above 85% |
+ 2% after |
‘the first 5 |
minutes, LED's |
will remain |
off. The system |
will |
wait |
|||||||||||
for a |
maximum |
of |
30 |
minutes from start-up to reach 85% |
+ 2% before activating an alarm. Environmental fac- |
||||||||||||||||
tors, |
such as low voltage, high altitude, or age of the machine will |
affect |
the time |
required |
to |
reach 85% |
+ |
2%. |
|||||||||||||
If the |
oxygen level |
is |
not above 85% |
+ 2% |
within the first |
30minutes, |
the |
oxygen |
level alarm sequence |
will |
acti- |
vate, depending upon the position of the indicator switch. All units are preset at the factory at 73% + 3%.
“ When oxygen concentration is above 85% + 2%, the sensor measures oxygen purity every 10 minutes. If a read-
ing falls below 85% + 2% and the alarm indicator is in the 73% mode, the oxygen sensor measures purity con- tinuously and a yellow LED will illuminate. If the oxygen purity falls below 73% + 3% the red LED / Alarm /
Shut-Down |
mode will activate. If |
the oxygen |
purity falls |
below 85%+ 2% with the indicator switch set at85%, |
the red LED |
/ Alarm / Shut-Down |
mode will |
áctivate. |
o |
Call: Tol! Free 1-800-333-6900
innovation Health Care =
1-7 |
- |
: |
INSTALLATION/SEQUENCE OF OPERATION
INSTALLATION / SEQUENCE OF OPERATION MODEL 3LX and MODELS 5LX,5LX02, 5LXO2T, 6LXO2, 6LXO2T
INSTALLATION / VERIFICATION OF BATTERY FREE™ POWER LOSS ALARM
When your |
new |
Invacare |
concentrator |
arrives, it should |
be checked for proper operating conditions. |
|
1. Ifthe unit |
has |
been in |
below-freezing |
temperatures, |
allow it to warm up to room temperature (approxi- |
|
mately |
30 |
minutes) |
before operating. |
|
2. The concentratorwill need to be tumed on for4to 5 seconds to charge the Battery Free! Power Loss Alam. Connect power cord to outlet and tum the concentrator on. Tum flow contro! knob counterclockwise and flow will
begin |
immediately. |
Set flow rate of the |
Invacare |
3 to |
3 |
L/min., |
Invacare |
5 to 5 |
L/min. |
and the |
Invacare 6 to 6 |
L/min. |
|
Tum |
unit |
off. |
|
|
|
|
|
|
|
|
|
|
|
3. Unplug |
the power |
cord and press |
On/Off |
switch |
to the |
ON position. An intermittent audible alarm will |
|||||||
sound. This confirms proper operation of |
the |
Battery |
Free™ |
Power |
Loss |
Alarm. |
Turn On/Off |
switch |
|||||
OFF. |
|
|
|
|
|
|
|
|
|
|
|
|
|
4. Connect cord to outlet and turn on. concentrator. Unit will sound 3 BEEPS on startup.
5.Check the oxygen concentration per specifications after 30-40 minutes running time.
SEQUENCE |
OF |
OPERATION |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||||
Applying |
120 |
VAC via the power switch. energizes the compressor motor, |
hour |
meter, |
cooling |
fan |
and |
the |
||||||||||||||||||||||||||||||||
printed |
|
circuit |
(P.C.) |
board. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
Room |
|
air |
enters |
the |
compressor |
via the cabinet filter, compressor |
inlet |
filter, |
and |
sound |
muffler. |
The |
air is |
|||||||||||||||||||||||||||
compressed |
by |
|
the |
wobble |
pistons |
in the |
compressor |
|
to |
a pressure |
of |
20 |
PSI |
(137.72 |
kPa) |
BEFORE |
||||||||||||||||||||||||
Serial |
|
No. |
"97K" |
or 21 PSI (144.79 kPa) AFTER |
Serial |
No. |
"97K". |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||||||||||||
As |
increased |
pressure |
creates |
increased |
temperature, |
a |
heat exchanger |
is |
utilized |
to |
lower the |
tempera- |
||||||||||||||||||||||||||||
ture |
by |
20° F |
(6.7° C) before |
the |
air enters the.4-way valve. |
itis |
then channeled |
|
to |
a sieve |
bed |
containing ' |
||||||||||||||||||||||||||||
the |
adsorption |
maierial. |
A |
restrictor downstream |
of |
the |
sieve bed |
causes pressure |
to |
build |
up |
inside |
the. |
|||||||||||||||||||||||||||
. sieve |
bed |
which |
is necessary |
for |
the |
adsorption |
process. |
A |
small |
amount |
of |
relatively |
pure |
oxygen |
enters |
|||||||||||||||||||||||||
the |
top |
|
of |
the |
second bed with the balance |
entering |
a |
storage tank. The nitrogen removed |
is |
exhausted |
||||||||||||||||||||||||||||||
back |
from |
the |
bed through the four-way valve into room |
|
air. A muftler is located |
|
at |
the |
exhaust |
end |
of |
the |
||||||||||||||||||||||||||||
valve |
to |
muffle |
the |
sound |
of |
the |
exhaust |
as |
it |
exits |
the |
concentrator. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||||||||
The |
oxygen |
not being used to exhaust is channeled |
into the storage tank. The |
pressurized |
oxygen |
is |
||||||||||||||||||||||||||||||||||
regulated |
down |
|
to 5 PSI (34.4 kPa), enters an accurate |
|
flow-measuring |
device, |
|
fiows |
through |
a bacterial |
||||||||||||||||||||||||||||||
filter, |
a |
|
check |
valve and out to the patient. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||||||||
The |
electrical |
activation of the 4-way valve |
is accomplished |
every |
8 |
to |
21 |
seconds by the |
pressure |
sensor |
||||||||||||||||||||||||||||||
and |
P.C. |
board |
|
electronics |
when |
the |
pressure |
set |
point |
of 20 |
+ |
3 |
PSI |
(137.72 |
+ |
51.32 |
kPa) |
BEFORE |
||||||||||||||||||||||
Serial |
|
No. |
“97K" |
or |
21 |
+ |
3 |
PSI |
(144.79 |
+ |
51.32 |
kPa) |
AFTER |
|
Serial |
No. |
"97K" |
is |
reached. |
The |
time |
|||||||||||||||||||
between |
cycles |
|
is dependent |
on |
altitude, flow |
rate and |
internal |
environmental |
factors. |
|
|
|
- |
|
|
|
|
|||||||||||||||||||||||
On |
the |
|
5LX/5LX02/5LXO2T/6LX02/6LXO2T, |
|
a P.E. or |
Pressure |
Equalization |
|
valve |
opens |
just |
prior to |
||||||||||||||||||||||||||||
the |
shift |
of the |
4-way valve. |
This |
allows |
highly |
concentrated |
oxygen |
to |
enter |
the |
just exhausted |
bed |
from |
||||||||||||||||||||||||||
the |
top. |
This |
additional pressure allows the bed to start |
|
its |
cycle |
|
at |
a higher |
pressure. |
The |
P.E. |
valve |
will |
||||||||||||||||||||||||||
close |
just |
after |
the |
shift |
of |
the |
4-way |
valve. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
e A
MA
S
E
Q
U
E
N
Cc
E
ο
F
ο)
P
E
R
A
T i
о
N
SS
95
E
Q
U
E
N
с
Ε
o
F
οP
E
R
Α
T
|
ο
N
INSTALLATION / SEQUENCE OF OPERATION
INSTALLATION / SEQUENCE OF OPERATION, CONTINUED
If main power is lost, the Battery Free” Power Loss Alarm will sound a short "beep", with a long pause, intermittently. All units are equipped with a diagnostic alarm system that signals if the pneumatic pressure
or electrical systems malfunction. The Troubleshooting Guide |
explains the |
alarm system signals and |
|
reasons |
in detail for your convenience. |
É |
|
SENSO2 |
OXGYEN SENSOR TECHNOLOGY CERAMIC |
ZIRCONIA |
SENSOR |
Technical Description
The oxygen being produced by the concentrator flows out of the product tank and into the flowmeter.
In-line between these |
components is |
a |
tee |
fitting |
that directs a small flow of oxygen through a precision |
orifice to the oxygen |
sensor mounted |
on |
a |
printed |
circuit board. |
As the oxygen enters the sensor, it passes through a screen and contacts the sensing disk.
Electric |
current |
flowing |
through |
a metal film resistor heats the |
disk |
in excess |
of |
300°C. |
Oxygen |
molecules |
||||||||||||||||||
contact |
the electrode |
of |
the |
disk and |
pick-up |
extra |
electrons |
to become oxygen ions. These oxygen ions |
||||||||||||||||||||
are attracted to the electrode |
on |
the |
bottom |
of |
the |
zirconia |
disk. |
Because |
of |
the |
crystal |
structure of |
the |
|||||||||||||||
zirconia, only oxygen ions can |
pass through. When the oxygen ions reach the bottom electrode, the extra |
|||||||||||||||||||||||||||
electrons |
are |
released from the |
oxygen |
ions |
and |
oxygen |
molecules |
retum |
to |
the |
air. |
The |
number |
of |
||||||||||||||
electrons |
is directly related to the |
oxygen |
concentration. |
The |
electrons travel to the PC board where they |
|||||||||||||||||||||||
are “counted” |
and the |
oxygen |
concentration |
"reading" |
is |
calculated. |
|
|
|
|
|
|
|
|
|
|
||||||||||||
A microprocessor on the P.C. |
board |
contains |
software |
that |
interprets |
the signal |
being |
received |
from |
the |
||||||||||||||||||
sensor. |
It |
compares |
the signal |
to |
clinically acceptable |
limits that |
have |
been |
selected with |
the |
indicator |
|||||||||||||||||
switch |
mounted on |
the |
|
PC |
board. |
Signals |
outside of the |
|
clinically |
acceptable |
limits generate |
responses |
in |
the form of lighted LEDs, audible indicators, and/or system shut-down.
Operating Sequence
Once |
the power |
switch has been turned ON, the SensO, circuit will |
wait 5 minutes for the concentrator to |
|||
begin |
producing |
clinically acceptable oxygen |
before |
activating. No |
LEDs |
will illuminate for at least 5 |
minutes. |
|
|
|
|
|
|
After |
5 minutes, |
if the oxygen purity exceeds |
85% + |
2% the GREEN |
LED |
will illuminate. |
If the oxygen level is not above 85% + 2% after the first 5 minutes, LEDs will remain off. The system will
wait for a |
maximum |
of |
30 |
minutes |
from |
start-up to |
reach |
85% |
+ |
2% before |
activating an alarm. Environ- |
|||||||
mental |
factors |
such as low voltage, high altitude, |
or |
age |
of |
the |
machine |
will |
affect the |
time required |
to |
|||||||
reach |
85% |
+2%. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
If the |
oxygen |
level |
is not |
above 85% + |
2% |
within |
the first 30 minutes, the oxygen concentration alarm |
|||||||||||
sequence |
will |
activate, |
depending |
upon |
the |
position |
of |
the |
indicator switch. |
All units |
are preset at |
the |
||||||
factory at |
73% |
+ 3%. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
When oxygen concentration is above 85% + 2%, the sensor measures oxygen purity every 10 minutes.
If a reading |
falls below 85% + 2% and the alarm indicator is |
in the |
73% mode or no |
switch is present, |
the |
||||||
oxygen sensor measures purity continuously and a yellow |
LED |
will illuminate. |
If |
the |
oxygen |
purity |
falls |
||||
below 73% |
+ |
3% the |
RED LED/Alarm/Shut-Down mode |
will activate. |
lf the |
selector |
switch |
is present |
|||
and it is set |
at |
85%, |
the RED LED/Alarm/Shut-Down mode will |
activate |
if the |
oxygen |
purity |
falls below |
85% + 2%.
SENSO, TECHNICAL ILLUSTRATION
T-PIECE
< FLOW METER “τρ <— GAS Ve PRODUCT
ORIFICE
SENSOR HOUSING
SCREEN
TOP ELECTRODE
ZIRCONIA DISK
BOTTOM
ELECTRODE
ELECTRONS TO PC BOARD
©
GAS (INCLUDES OXYGEN)
|
OXYGEN |
6 |
|
|
|
+ |
|
<— ELECTRONS |
|
OXYGEN |
PICKS UP ELECTRONS |
|
||
TO FORM OXYGEN |
IONS |
|
||
OXYGEN |
IONS |
PASS |
THRU |
|
|
Y |
|
|
|
OXYGEN IONS |
RELEASE |
|
||
|
ELECTRONS |
|
|
|
TO |
FORM |
OXYGEN |
|
|
© ELECTRONS |
, |
|
ELECTRONS |
EROM |
|
|
|
||
|
OXYGEN |
TRANSFORMER |
NEUMATIC DIAGRAM
P G
PNEUMATIC DIAGRAM
BACTERIA
FILTER
OXYGEN |
|
FLOWMETER |
TEST POINT #1 |
|
|
||
|
|
$ |
|
ENRICHED |
|
|
|
|
|
|
|
PRODUCT |
| |
À |
|
TEST POINT #2 |
= |
|
|
|
|
|
|
|
RE |
|
PRODUCT TANK |
|
FLOW |
[|] . |
SENSO! |
|
|
|
RESTRICTOR |
92 SENSOR |
|
|
|
‘OPTIONAL „7 |
|
' |
pc. |
BOARD |
|
|
OXYGEN |
|
|
ВО |
A |
|
SENSOR |
|
|
|
|
|
INPUT |
|
|
|
|
SIEVE
COLUMN
AIR INLET
SOLENOID VALVE
MOBILAIRE Ill / V
Model IRC301 / 501
MOBILAIREill / V “#* SersO,
Model IRC30102 / 50102
CHECK VALVES .
ge
CROSSASSEMBLY-OVER - -
SIEVE
LA COLUMN
501 ONLY
RE.
VALVE
j
|
GROSS |
PARTICLE |
|
|
|
os |
|
|
|
FILTER |
|
|
|
|
|
FILTER |
ASS'Y | |
|
|
|
RELIEF |
|
' |
|
|
VALVE |
|
||||
|
RESONATOR |
|
|
|
=== |
- FIUER |
|
|
> |
|
AIR |
. |
HEAT |
|
|
|
|
COMPRESSOR |
|
EXCHANGER |
Call: Toll Free 1-800-333-6900
dnnovetionhhHeatth Care ~
1-1 1
|
|
|
|
|
|
PNEUMATIC |
DIAGRAM |
|
|||
|
|
PNEUMATIC |
DIAGRAM |
|
|
|
|
|
P |
||
|
|
|
|
|
|
|
N |
||||
|
INVACARE 3 & 5 OXYGEN CONCENTRATORS |
|
|
|
|
||||||
|
|
|
|
|
E |
||||||
|
|
MODELS |
IRC3LX, |
IRC5LX |
IRC5LXO2 |
|
|
|
|
|
Y |
|
|
RESTRICTOR |
|
|
TEST |
POINT |
2 |
|
|
|
M |
|
|
|
|
|
|
|
A |
||||
|
|
ASSEMBLY |
|
|
(BETWEEN PRODUCT |
TANK AND |
|
|
|
||
|
|
|
|
|
|
|
T |
||||
BACTERI |
|
IRC3LX & |
CHECK |
VALVES |
CHECK |
VALVE) |
|
|
|
||
FILTER A |
IRC5LX ONLY |
|
|
|
|
|
|
|
|
I |
|
|
FLOWMETER |
|
|
|
|
|
|
|
|
o |
|
|
|
|
|
|
|
|
|
|
|
||
/ |
|
PRESSURE |
|
|
|
RESTRICTOR |
|
D |
|||
|
|
|
|
is EXTERNAL |
on |
I |
|||||
|
Τη |
REGULATOR |
|
|
|
||||||
TEST POINT 1 |
|
|
|
|
|
IRC3LX ONLY, |
A |
||||
|
| |
|
|
|
|
||||||
(OXYGEN CONCENTRATOR | |
|
PRODUCT |
TANK |
|
It is |
INTERNAL |
to |
G |
|||
OUTLET) |
|
|
|
|
|||||||
|
|
|
|
|
|
P.E. VALVE |
for |
R |
|||
|
|
|
|
|
|
|
|||||
|
FLOW |
|
|
|
|
|
|
IRCSLX |
& |
|
A |
|
|
|
|
|
|
|
IRCSLXO2 |
|
|||
RESTRICTOR |
|
|
|
|
|
|
|
M |
|||
|
|
|
|
|
|
|
|
|
.
OPTIONAL
OXYGEN SENSOR
-ANS
P.C.BOARD
PI
RESTRICTOR
=
CROSSOVER ASSEMBLY
AIR INLET
- y CET] —
CABINET
(Goss PARTICLE)
FILTER
SIEVE
BED
SOLENOID |
PILOT |
VALVE |
|
RESONATOR
AR
COMPRESSOR
| RELIEF VALVE
di IP
HEAT
EXCHANGER
SIEVE
sxsLxo | BED
SLXO2T/
6LXO2/ 6LXO2T ONLY!
РЕ. МАМЕ
CH
|
|— |
4-WAY VALVE |
L |
i |
|
πι EXHAUST MUFFLER
LEGEND
_ ELECTRICAL Fora << =conNECTIONS
a ~I
2
INVACARE RESPIRATORY ©
OXYGEN CONCENTRATOR OPERATION
Invacare produces many models of Oxygen Concentrators.
Some of them are:
-IRC 301 3 Liter Machine
-IRC 501 |
5 |
Liter |
Machine |
-IRC3LX |
3 |
Liter |
Machine |
-IRC5LX |
5 |
Liter |
Machine |
-IRC5LXO, |
5 |
Liter |
Machine |
with |
Sensor |
for O, |
output |
In the following |
block |
diagram |
we |
show a |
typical |
oxygen concentrator and its |
|
componenis. |
|
|
|
|
|
|
|
Item #1 The Gross Particle Filter
The gross particle filter (20 p.p.i.) is used to stop large particles of dirt (i.e. lint, animal fur, etc.) from entering the concentrator.
Item #2 Intake Filter (Filter Assembly)
This filter removes particles down to 1 micron in size. It is used primarily as a compressor filter.
Item |
# 3 |
Resonators |
|
|
|
|
|
|
|
|
|||
|
|
This |
device |
does |
nothing more than reduce noise from |
the |
intake |
||||||
|
|
of |
the |
compressor. |
(Muffler) |
|
|
|
|
|
|||
item |
#4 |
Air |
Compressor |
|
|
|
|
|
|
||||
|
|
The function of the compressor |
is to compress |
filtered air to an |
|||||||||
|
|
average |
maximum |
working |
pressure of approximately |
20 |
p.s.i. |
||||||
|
|
This |
compressed |
air is routed to |
the molecular |
sieve beds |
which |
||||||
|
|
require air under pressure to produce oxygen. |
Invacare |
uses |
|||||||||
|
|
Thomas |
brand |
oil-less piston |
compressors in |
most of the |
|
||||||
|
|
concentrator |
we |
manufacture. |
|
|
|
|
|||||
Item |
#4A |
Capacitor |
|
|
|
|
|
|
|
|
|||
|
|
The capacitor is an electrical storage device used to help start the |
|||||||||||
|
|
compressor |
motor. |
|
|
|
|
|
INVACARE |
TECHNICAL TRAINING 4-800-333-6900 X6697 ELYRIA, OHIO |
1-13