Invacare PLATINUM 5 User Manual

0 (0)

OPERATOR'S MANUAL - ENGLISH MANUAL DE INSTRUCCIONES - SPANISH MANUEL D’UTILISATION - FRENCH BENUTZERHANDBUCH - GERMAN

NANUAL DO OPERADOR - BRAZILIAN PORTUGUESE

PLATINUM 5

OXYGEN CONCENTRATORS STANDARD,

WITH/SENSO2®AND HOMEFILL™ II

CONCENTRADORES DE OXÍGENO STANDARD,

CON/SENSO2®Y HOMEFILL™ II

CONCENTRATEURS D’OXYGENE STANDARD,

AVEC / SENSO2 ® ET HOMEFILL™ II

SAUERSTOFFKONZENTRATOREN STANDARD,

MIT/SENSO2®UND FÜR HOMEFILL™ II

CONCENTRA DO RES DE OXIGENIO STANDARD,

COH/SENSO2®E HOMEFILL™ II

DEALER: THIS MANUAL MUST BE GIVEN TO THE END USER.

USER: BEFORE USING THIS CONCENTRATOR, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.

DISTRIBUIDOR: ENTREGUE ESTE MANUAL AL USUARIO FINAL. USUARIO: ANTES DE OPERAR EL CONCENTRADOR, LEA TOTALMENTE ESTE MANUAL Y CONSÉRVELO COMO REFERENCIA PARA EL FUTURO.

DISTRIBUTEUR: CE MANUEL DOIT ÊTRE REMIS À L’UTILISATEUR. UTILISATEUR: AVANT D’UTILISER CE CONCENTRATEUR, BIEN LIRE CE MANUEL ET LE CONSERVER POUR S’Y RÉFÉRER ULTÉRIEUREMENT.

FÜRHÄNDLER:DIESESHANDBUCHMUSSDEMENDBENUTZERÜBERGEBEN WERDEN.

FÜRBENUTZER:LESENSIEDIESESHANDBUCHVORBENUTZUNGDIESES KONZENTRATORSUNDBEWAHRENSIEESZUMNACHSCHLAGENFÜRDIE ZUKUNFT AUF.

REPRESENTANTE: ESTE MANUAL DEVE SER ENTREGUE AO USUÁRIO.

USUÁRIO: ANTES DE USAR ESSE CONCENTRADOR, LEIA ESTE MANUAL E GUARDE-O PARA REFERÊNCIAS FUTURAS.

SPECIAL NOTES

Invacare 5 with/SensO2®

Invacare 5

Model IRC5LXO2AW

Model IRC5LXAW

Invacare 5 con/SensO2®

Invacare 5

Modelo IRC5LXO2AW

Modelo IRC5LXAW

Invacare 5 avec/SensO²®

Invacare 5

Modèle IRC5LXO2AW

Modèle IRC5LXAW

Invacare 5 mit/SensO2®

Invacare 5

Modell IRC5LXO2AW

Modell IRC5LXAW

Invacare 5 con/SensO2®

Invacare 5

Modelo IRC5LXO2AW

Modelo IRC5LXAW

SPECIAL NOTES

WARNING/CAUTION notices used in this manual apply to hazards or unsafe practices which could result in personal injury or property damage.

NOTICE

THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.

WARNING

DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL.

IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE SERIOUS INJURY OR PROPERTY DAMAGE MAY RESULT.

SAVE THESE INSTRUCTIONS

DANGER

Users MUST NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this product is located. NO SMOKING signs should be prominently displayed. Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air. Failure to observe this warning can result in severe fire, property damage and cause physical injury or DEATH.

2

TABLE OF CONTENTS

CONCENTRATOR

 

TABLE OF CONTENTS

 

SPECIAL NOTES .....................................................................

2

SAFETY SUMMARY ................................................................

5

FEATURES ................................................................................

8

SPECIFICATIONS ...................................................................

9

HANDLING .............................................................................

12

Unpacking ........................................................................................................

12

Inspection ........................................................................................................

12

Storage ..............................................................................................................

12

INTRODUCTION ..................................................................

13

OPERATING INSTRUCTIONS ...........................................

13

Select a Location ............................................................................................

13

Recommended End-User Guidelines

 

for Optimum Performance ....................................................................

13

Plug in Power Cord ......................................................................................

14

Connect Humidifier (if so prescribed) ...................................................

14

Power Switch ..................................................................................................

16

Flowrate ............................................................................................................

16

SensO2®Oxygen Purity Indicator

 

(if the unit features the O2 Sensor) .....................................................

17

Initial Startup of the Concentrator ..........................................................

17

Explanation of Oxygen Purity Indicator Lights .....................................

18

Indicators (if the unit does not feature the O2 Sensor).....................

18

CONCENTRATOR MAINTENANCE ................................

19

Cleaning the Cabinet Filters .......................................................................

19

Cabinet .............................................................................................................

20

Humidifier ........................................................................................................

20

TROUBLESHOOTING GUIDE ............................................

20

PREVENTIVE MAINTENANCE RECORD ........................

22

WARRANTY ...........................................................................

23

3

TABLE OF CONTENTS

 

SPANISH OWNER’S MANUAL

 

(MANUAL DEL USARIO) .................................................

24

FRENCH OWNER’S MANUAL

 

(MANUEL D’UTILISATION) ...........................................

49

GERMAN OWNER’S MANUAL

 

(BENUTZERHANDBUCH) ...............................................

73

BRAZILIAN PORTUGUESE OWNER’S MANUAL

 

(NANUAL DO OPERADOR) ...........................................

97

4

SAFETY SUMMARY

SAFETY SUMMARY

CAUTION

Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel.

READ THE FOLLOWING INFORMATION BEFORE OPERATING THIS PRODUCT.

WARNING

The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.

A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless expressly recommended by Invacare.

For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.

If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call Qualified Service Personnel for examination and repair.

Keep the cord away from HEATED or HOT surfaces.

Do not move or relocate concentrator by pulling on the cord.

NEVER drop or insert any object into any opening.

NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air opening may be blocked. Keep the openings free from lint, hair and the like.

5

SAFETY SUMMARY

WARNING (CONTINUED)

Fill humidifier with water to the level shown by the manufacturer. DO NOT overfill.

Keep unit at least 7.5 cm (3 in) away from walls, draperies, furniture, and the like.

Invacare recommends that Crush-Proof oxygen tubing (supplied by Invacare) be used with this product and NOT exceed 15.2 m (50 ft) in length.

MAINTENANCE

The Invacare Oxygen Concentrator was specifically designed to minimize routine preventive maintenance at intervals of once per year. Only professionals of the healthcare field or persons fully conversant with this process such as authorized or factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator.

RADIO FREQUENCY INTERFERENCE

Most electronic equipment is influenced by Radio Frequency Interference (RFI). CAUTION should be exercised with regard to the use of portable communications equipment in the area around such equipment.

TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS.

Avoid using while bathing. If continuous usage is required by the physician’s prescription:

The concentrator must be located in another room at least 2.1 m (7 ft) from the bath.

DO NOT come in contact with the concentrator while wet.

DO NOT place or store product where it can drop into water or other liquid.

DO NOT reach for product that has fallen into water. UNPLUG IMMEDIATELY.

A product should NEVER be left unattended when plugged in.

6

SAFETY SUMMARY

WARNING (CONTINUED)

This device is to be used only in accordance with the prescription of a physician and this Owners Manual. If at any time the patient or attendant conclude that the patient is receiving an insufficient amount of oxygen, the supplier and/or physician should be contacted immediately. No adjustments should be made to the flowrate unless prescribed by a physician.

Close supervision is necessary when this product is used near children or physically-challenged individuals.

Use this product for only intended use as described in this manual.

DO NOT use parts, accessories or adapters other than those authorized by Invacare.

DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.

Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impair the performance.

In certain circumstances oxygen therapy can be hazardous. Invacare recommends that you seek medical advice before using this machine.

This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining.

Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.

OXYGEN CONCENTRATION

WARNING

Take care not to set the flow above RED ring. An oxygen flow greater than 5 L/min will decrease the oxygen concentration.

7

FEATURES

 

FEATURES

 

FRONT VIEW

 

Oxygen Purity Indicator Lights

Oxygen Outlet

(SensO2 model only)

 

(Contact Invacare for avail-

Flowmeter

ability in your country)/ Fault

 

and Power Indicator Lights

 

(LX model only)

 

Power Switch

 

Circuit Breaker

 

Elapsed Time Meter

 

Cabinet Filter

 

(one of two shown)

REAR VIEW

ACCESSORIES (NOT SHOWN):

 

Venture® HomeFill II home oxygen

 

compressor - IOH200AW

 

Contact local Invacare representative

 

for available accessory information.

Power

Outlet

Cord

Fitting

NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the Venture HomeFill II home oxygen compressor.

The outlet fitting does not affect concentrator performance. Refer to the HomeFill II owner's manual, for connection and operating instructions. When not in use, the plug provided with the concentrator should be inserted into the outlet fitting. For more information about the HomeFill II, contact your Invacare dealer.

8

SPECIFICATIONS

SPECIFICATIONS

Alternating Current

Type BF Equipment

Running

NOT Running

Attention-Consider Accompanying Documents

DO NOT Smoke

 

 

Class II

 

Electrical Requirements:

230 VAC ± 10% (253 VAC/207 VAC), 50 Hz

Rated Current Input:

2.0 A

Sound Level:

44 dB Average

Altitude:

Up to 1828 m(6,000 ft) above sea level without

 

 

 

degradation of concentration levels. NOT

 

 

 

RECOMMENDED FOR USE ABOVE

 

 

 

1828 M (6000 FT).

*Oxygen Output

ALL 5LXA/5LXO2A MODELS

Concentration Levels:

 

 

 

93% minimum at 1 to 3 L/min

 

 

 

91% minimum at 4 L/min

 

 

 

87% minimum at 5 L/min (maximum

 

 

 

recommended flow)

 

 

 

* Stated concentration levels

 

 

 

achieved after initial warm-up period

 

 

 

(approximately 30 minutes).

Maximum Outlet Pressure:

34.5 ± 3.45 kPa

 

 

 

(5 ± 0.5 psi)

5LXO2 INDICATORS:

Label

Status

Indicator Lights (LED)

Symbol

 

 

 

 

 

 

SYSTEM FAILURE

RED Continuous Audible Alarm Sieve-GARDTM

 

Below 73%

Compressor Shutdown

 

73% to 85%

YELLOW

O2

SYSTEM OKAY

GREEN

Over 85%

 

 

 

 

 

 

 

 

 

9

SPECIFICATIONS

SPECIFICATIONS (CONTINUED)

5LX INDICATORS:

Label

Status

Indicator Lights (LED)

Symbol

 

 

 

 

 

I/O

SYSTEM OKAY

GREEN

 

SYSTEM FAILURE

RED

 

 

 

Flow Range:

Low Flow Alarm:

Average Power

Consumption:

Pressure Relief Mechanism Operational at:

Change in maximum recommended flow when back pressure

of 7kPa is applied: Filters:

Safety System:

Width:

Height:

Depth:

Weight:

Shipping Weight:

1/2 - 5 L/min.

Flowrates less than 1 L/min., are not recommended.

0 L/min. to 0.5 L/min., Rapid Audible Alarm Beeping (No Accessories Connected)

395W

241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)

0.7 L/min.

Cabinet (2), Outlet HEPA and Compressor Inlet.

Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm.

SensO2 Oxygen System (SensO2 Model) Low Flow Alarm.

46.7 cm ± 1 cm (18 3/8” ± 3/8”)

67.0 cm ± 1 cm (26 3/8” ± 3/8”)

36.5 cm ± 1 cm (14 3/8” ± 3/8”) 23.2 kg ± 0.5 kg (51 lbs. ± 1 lb) 26.3 kg ± 0.5 kg (58 lbs. ± 1 lb)

10

SPECIFICATIONS

SPECIFICATIONS (CONTINUED)

Operating Ambient:

Temperature:

Exhaust:

Oxygen Output:

Cabinet:

Regulatory Listing:

Electrical:

Placement:

Tubing:

Relative Humidity:

Time of Operation:

Recommended

Storage and Shipping

Temperature:

10°C - 35°C (50°F - 95°F) at 20-60% relative humidity

Less than Ambient + 19°C (+ 35°F). Less than Ambient + 2°C (+ 4°F).

Impact resistant flame-retardant plastic cabinet that conforms to UL 94-V0.

5LXA/5LXO2A Models

ETL certified complying with: EN 55011

EN 61000-3-2

EN 61000-3-3

IEC 60601-1, AI, A2 IEC 60601-1-2

5LXO2A Model only

ISO 8359 MDD

No extension cords.

No closer than 7.5 cm (3 in) from any wall, furniture, draperies, or similar surfaces.

Crush-Proof Tubing - 2.1 m (7 ft) cannula with a maximum 15.2 m (50 ft) extension

(DO NOT PINCH). 20 to 60%

Up to 24 hours per day.

-29°C to 65°C (-20°F to 150°F), at 15-95% relative humidity

11

HANDLING

HANDLING

The concentrator should always be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator. If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare.

UNPACKING

1.Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer.

2.Remove all loose packing from the carton.

3.Carefully remove all the components from the carton.

FIGURE 1 - PACKAGING

NOTE: Unless the oxygen concentrator is to be used immediately, retain

containers and packing materials for storage until use of the concentrator is required.

INSPECTION

1.Examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all components.

STORAGE

1.Store the repackaged oxygen concentrator in a dry area.

2.DO NOT place other objects on top of the repackaged concentrator.

12

INTRODUCTION/ OPERATING INSTRUCTIONS

INTRODUCTION

Your oxygen concentrator is intended for individual use in the home. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems.

Your provider will show you how to use your oxygen concentrator. He/She should be contacted with any questions or problems regarding your oxygen concentrator. This owner’s manual will tell you about your concentrator and will serve as a reference as you use your concentrator.

OPERATING INSTRUCTIONS

SELECT A LOCATION

You may select a room in your house where using your oxygen concentrator would be most convenient. Your concentrator can be

easily rolled from room to room on its casters.

Your oxygen concentrator will perform best when operated under the following conditions. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated

area to avoid airborne pollutants and/or fumes.

RECOMMENDED END-USER GUIDELINES FOR OPTIMUM PERFORMANCE:

Temperature:

10°C - 35°C (50°F - 95°F)

Electrical:

No extension cords.

Placement:

No closer than 7.5 cm (3 in) from the wall,

 

furniture, draperies, or similar surfaces.

Tubing and Cannula:

2.1 m (7 ft) cannula with a maximum 15.2 m

 

(50 ft) of Crush-Proof Tubing (DO NOT

 

PINCH).

Environment:

Smoke and soot-free. No confined spaces

 

(Example: NO closets).

Relative Humidity:

20 to 60%

Time of Operation:

Up to 24 hours per day.

Flow Rate

From 1/2 L/min. to 5 L/min.

 

Flowrates less than 1 L/min., are not

 

recommended.

Minimum OperatingTime:

30 Minutes

13

OPERATING INSTRUCTIONS

NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away from walls, draperies or furniture to assure sufficient air flow. Avoid deep pile carpets and heaters, radiators or hot air registers.

PLUG IN POWER CORD

1. Plug in power cord to an electrical outlet.

CONNECT HUMIDIFIER (IF SO PRESCRIBED)

1.Fill humidifier with distilled water to the level indicated by the manufacturer. (Recommended humidifiers: Invacare 3260, 003-40, 006-40, or M5120).

WARNING

DO NOT overfill humidifier.

Replace the humidifier cap and securely tighten.

DO NOT reverse the oxygen input and output connections. Water from the humidifier bottle WILL travel through the cannula back to the patient.

Humidifier Bottle

FIGURE 2 - FILLING THE HUMIDIFIER

2. Place the humidifier in the humidifier compartment.

Humidifier Bottle

FIGURE 3 - HUMIDIFIER COMPARTMENT

14

OPERATING INSTRUCTIONS

3.Remove the Filter Access Door located on the side of the concentrator. Insert a flathead screwdriver in the plate groove on the top edge of the Filter Access Door and gently pry the Filter Access Door open (FIGURE 4, DETAIL “A”).

4.The humidifier adapter is next to the inlet filter. Pull up and remove the humidifier bottle adapter (FIGURE 4, DETAIL “B”).

5.Attach it to the humidifier by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened (FIGURE 4, DETAIL “C”).

DETAIL “A”

Plate Groove (Insert Flat

Head Screwdriver Here)

Filter Access Door

DETAIL “B”

Humidifier Bottle

Adapter

DETAIL “C”

Filter

 

Access

Humidifier Bottle

Door

Adapter

FIGURE 4 - ATTACHING THE HUMIDIFIER

BOTTLE ADAPTER

15

OPERATING INSTRUCTIONS

6.Attach oxygen tubing from the humidifier bottle to oxygen outlet connector on the oxygen concentrator.

Oxygen Outlet

Oxygen Connector

Tubing

Humidifier Outlet

FIGURE 5 - ATTACHING THE OXYGEN TUBING

7.Attach the cannula/patient supply tubing to the humidifier bottle outlet.

8.After assembly, ensure that oxygen is flowing through the cannula.

9.Replace the filter access door on the side of concentrator by snapping it back into the plate groove.

WARNING

For optimum performance, Invacare recommends that the concentrator be operated a minimum of 30 minutes during any use. Shorter periods of operation may adversely affect the life of the product.

POWER SWITCH

1.Press power switch to the ON position. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functioning properly. After one second, the only green SYSTEM OK/POWER light will stay on.

On/Off (I/O)

Power Switch

FIGURE 6 - POWER SWITCH

FLOWRATE

1.Turn the flowrate knob to the setting prescribed by your physician or therapist.

16

OPERATING INSTRUCTIONS

WARNING

DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.

NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed (FIGURE 7).

2.If the flowrate on the flowmeter ever falls below 0.5 L/min. for more than about one minute, the LOW FLOW ALARM will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than

0.5 L/min., the Low Flow audible alarm will go off.

Flow Knob

Ball

FIGURE 7 - ADJUSTING THE FLOWRATE-

PLATINUM 5 WITH SENSO2 SHOWN

NOTE: The use of some accessories such as the Homefill II compressor will deactivate the Low Flow Alarm.

SENSO2 OXYGEN PURITY INDICATOR

(IF YOUR UNIT FEATURES THE O2 SENSOR FEATURE)

This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.

INITIAL STARTUP OF THE CONCENTRATOR

NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.

When the unit is turned on, the green light will come on (SYSTEM OK / O2 greater than 85%). After 5 minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values. The explanation of the indicator light functions are as follows.

17

OPERATING INSTRUCTIONS

EXPLANATION OF OXYGEN PURITY INDICATOR LIGHTS (FIGURE 8):

GREEN LIGHT ( O2 ) - Normal Operation

YELLOW LIGHT ( ) - Call Supplier IMMEDIATELY. You may continue to use the concentrator unless instructed otherwise by your supplier. Be certain that BACKUP OXYGEN is nearby.

RED LIGHT ( )- Total unit shutdown. Switch IMMEDIATELY to BACKUP OXYGEN supply. Call Supplier IMMEDIATELY.

GREEN LIGHT - with YELLOW LIGHT FLASHING - Call Supplier IMMEDIATELY. Oxygen sensor malfunctioning; you may continue to use the concentrator.

OXYGEN PURITY

Auto Shut - Down

RED

Use Backup

Call Supplier

 

Below Normal

YELLOW

Normal O2

GREEN

FIGURE 8 - O2 DISPLAY CONTROL PANEL

INDICATORSIF YOUR UNIT DOES NOT FEATURE THE O2 SENSOR

LX MODELS INDICATOR LIGHT EXPLANATION (FIGURE 9):

RED LIGHT ( )- Total Unit Shut-Down. Switch IMMEDIATELY to a BACK-UP OXYGEN supply. Call Supplier IMMEDIATELY.

GREEN LIGHT (I/O)- On/Off. System okay.

Auto Shut - Down

RED

Use Backup

 

Call Supplier

Normal I/O

GREEN

FIGURE 9 - DISPLAY CONTROL PANEL

18

MAINTENANCE

CONCENTRATOR MAINTENANCE

NOTE: Normal maintenance is to be performed by the user or caregiver. The following outlines normal maintenance procedures.

NOTE: The Invacare Concentrators are specifically designed to minimize routine preventive maintenance at intervals of once per year. In places with high dust or soot levels, maintenance may need to be performed more often. Preventive maintenance must be performed at a minimum of one (1) year in service to assure years of additional reliability. Only authorized or factory-trained personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator.

A. CLEANING THE CABINET FILTERS

NOTE: There are two (2) cabinet filters. One (1) is located on each side of the cabinet (FIGURE 1).

1.Remove each filter and

CLEAN AT LEAST ONCE A WEEK depending on environmental conditions (FIGURE 2).

NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc.

Cabinet Filter (located on both sides of concentrator)

FIGURE 1 - CABINET FILTER LOCATION

2.Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly (FIGURE 2).

3.Dry the filters thoroughly before reinstallation.

CAUTION

DO NOT operate the concentrator without the filters installed.

FIGURE 2 - CLEANING THE CABINET FILTERS

19

MAINTENANCE/TROUBLESHOOTING

B. CABINET

WARNING

UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet.

1.Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.

C. HUMIDIFIER

NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these steps:

1.Clean the humidifier every day.

2.Wash it in soapy water and rinse it with a solution of ten (10) parts water and one (1) part vinegar.

3.Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier.

WARNING

DO NOT overfill Humidifier Bottle.

CONCENTRATOR

TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

 

 

Alarm:

Main Power Loss:

1.

Insert plug into outlet.

SHORT BEEPS,

1. Power cord NOT

2.

Inspect house circuit

LONG PAUSE

 

plugged In.

 

 

breakers or fuses. If

 

 

 

 

Concentrator

2.

No power at outlet.

 

problem recurs, use a

NOT operating,

3.

Tripped circuit breaker.

 

different outlet.

power switch ON.

3.

Call Service Provider.

Beep......Beep......

 

 

 

 

 

 

 

 

 

 

Alarm:

System Failure:

 

 

CONTINUOUS

1.

Unit overheating due

1a. Remove and clean

Concentrator

 

to blocked air intake.

 

cabinet filters.

 

 

1b.

Move oxygen concentrator at

NOT operating,

 

 

PowerSwitchON.

 

 

 

least 7.5 cm (3 in) away from

 

 

 

walls, draperies or furniture.

Beeeeeeeep....

 

 

 

 

 

 

 

 

2.

Insufficient power at

2.

Do not use extension

 

 

outlet.

 

cords. Move to another

 

 

 

 

electrical outlet or circuit.

 

3.

Internal repairs required.

3.

CallServiceProvider.

 

 

 

 

 

 

 

 

 

 

20

TROUBLESHOOTING

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

YELLOW or RED

1. Low oxygen purity.*

1.

Clean or replace filters.

LIGHT

2. Kinked or blocked

2.

Inspect for kinks or

Illuminated.

tubing, cannula or

 

blockages. Correct,

 

 

Alarm:

humidifier.*

 

clean or replace item.

 

 

 

Once corrected, turn

CONTINUOUS

 

 

 

power OFF for 60

On red light only.

 

 

 

seconds and then turn

 

 

 

 

power back ON.

 

3. Flowmeter set at

3.

Check flowmeter is set

 

1/2 L/min. or less.*

 

to 1.0 L/min. or more.

 

 

 

 

Refer to Recommended

 

 

 

 

End-User Guidelines in

 

 

 

 

the Operating Instructions

 

 

 

 

of this manual.

 

4. Unit overheating

4a.

Remove and clean

 

due to blocked air

 

cabinet filters.

* Only applies to

intake.

4b.

Move concentrator at

SensO2 models.

 

 

 

least 7.5 cm (3 in) from

 

 

 

 

walls, draperies, and

 

 

 

 

furniture.

 

5. Insufficient power at

5.

Do not use extension

 

outlet.

 

cords. Move to another

 

 

 

 

electrical outlet or circuit.

 

6. Internal repairs required.

6.

Call Service

 

 

 

 

Provider.

 

 

 

 

 

GREEN LIGHT

1.

Internal repairs required.

1.

Call Service Provider.

with YELLOW

Light Flashing.

 

 

 

 

 

 

 

 

Alarm:

LOW FLOW ALARM

1a.

Inspect for kinks or

RAPID

1a.

Kinked or blocked

 

blockages. Correct,

Beep....Beep...

 

tubing, cannula or

 

clean or replace item.

 

humidifier.

 

Once corrected, turn

Beep...Beep

 

 

 

 

 

power OFF for 60

 

 

 

 

 

 

 

 

seconds and then turn

 

 

 

 

power back ON.

 

1b. Flowmeter set at

1b.

Reset flowmeter to

 

 

0.5 L/min. or less.

 

prescribed flowrate.

 

 

 

 

Flowrates less than

 

 

 

 

1L/min. are not

 

 

 

 

recommended.

 

 

 

 

 

21

Invacare PLATINUM 5 User Manual

22

PREVENTIVE MAINTENANCE RECORD

INVACARE PLATINUM OXYGEN CONCENTRATOR

ON EACH INSPECTION

Record Date of Service

Record Elapsed Hours On Hour Meter

Clean Cabinet Filter(s)

Check Prescribed L/min. Flow Rate

EVERY 90 DAYS

Check Oxygen Concentration

(UNITS WITHOUT SensO2)

ANNUALLY OR DURING PREVENTIVE MAINTENANCE SCHEDULE

Clean/Replace Cabinet Filter(s)

Check Outlet HEPA Filter*

Check Compressor Inlet Filter*

Check Oxygen Concentration

(SensO2 MODELS)

Check Power Loss Alarm

AS REQUIRED

Clean Heat Exchanger

Replace Exhaust Muffler

Rebuild Top End of Compressor

*Refer to Preventive Maintenance section of Service Manual.

NO SERIAL IRC.NO MODEL

RECORD MAINTENANCEPREVENTIVE

.

 

 

 

WARRANTY

LIMITED WARRANTY

This warranty is extended only to the original purchaser of our products and does not affect statutory rights.

Invacare warrants the Oxygen Concentrators IRC5LXAW and IRC5LXO2AW to be free from defects in materials and workmanship for a period of five (5) years from date of purchase. The 4-way valve body carries a lifetime warranty from date of purchase. DO NOT open or attempt to service valve body. This will void any and all warranty applicable to the valve body. Routine maintenance items, such as filters, etc..., are excluded from cover under this warranty. If within such warranty period the product shall be proven to Invacare’s reasonable satisfaction to be defective, it shall be repaired or replaced, at Invacare’s option; Invacare’s sole obligation, and your exclusive remedy under this warranty being limited to such repair or replacement.

For warranty service, please contact Invacare Service Department during normal working hours. Upon receiving notice of an alleged defect in a product, Invacare will issue a serialised return authorisation. The product shall then be returned to either Invacare’s plant or service center as specified by Invacare. Defective component part(s) must be returned for warranty inspection using the serial number as identification within thirty (30) days of return authorisation date. DO NOT return products to our factory without prior consent.

LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY DOES NOT COVER NORMAL WEAR AND TEAR AND SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, COMMERCIAL OR INSTITUTIONAL USE, PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORISED PARTS OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE.

23

MANUAL DE INSTRUCCIONES

PLATINUM 5

CONCENTRADORES DE OXÍGENO STANDARD, CON/SENSO2®Y HOMEFILL™ II

INFORMACIÓN SOBRE EL

CONCENTRADOR

Invacare 5 con/SensO2

Invacare 5

Modelo IRC5LXO2

Modelo IRC5LX

DISTRIBUIDOR: ENTREGUE ESTE MANUAL

AL USUARIO FINAL.

USUARIO: ANTES DE UTILIZAR EL CONCENTRADOR,

LEALASINSTRUCCIONESDEESTEMANUALY

CONSÉRVELO COMO REFERENCIA PARA EL FUTURO.

24

NOTAS ESPECIALES

NOTAS ESPECIALES

Las ADVERTENCIAS y PRECAUCIONES indicadas en este manual se refieren a manipulaciones delicadas o peligrosas que pudieran causar lesiones personales o dañar el producto.

AVISO

LA INFORMACIÓN CONTENIDA EN ESTE DOCUMENTO ESTÁ SUJETA A CAMBIOS SIN PREVIO AVISO.

ADVERTENCIA

NO PONGA EN FUNCIONAMIENTO ESTE EQUIPO SIN ANTES LEER Y COMPRENDER ESTE MANUAL. SI NO ENTIENDE LAS ADVERTENCIAS, PRECAUCIONES E INSTRUCCIONES, PÓNGASE EN CONTACTO CON LOS SERVICIOS TÉCNICOS DE INVACARE ANTES DE USAR EL EQUIPO, DE LO CONTRARIO PODRÍA SUFRIR LESIONES O CAUSAR DAÑOS.

CONSERVE ESTAS INSTRUCCIONES

PELIGRO

Los usuarios NO DEBEN FUMAR mientras utilicen este dispositivo. Mantenga toda cerilla o cigarrillos que estén encendidos u otras fuentes inflamables fuera de la sala en la cual esté el producto. Coloque avisos de NO FUMAR donde puedan ser vistos por todos. Las telas y demás materiales que normalmente no son inflamables se encienden y queman fácilmente con una gran intensidad en el aire enriquecido con oxígeno. Si no se acata esta advertencia se pueden producir incendios, daños materiales graves y producir lesiones físicas graves o LETALES.

25

CONTENIDO

 

CONTENIDO

 

NOTAS ESPECIALES ...........................................................

25

MEDIDAS DE SEGURIDAD .................................................

28

CARACTERÍSTICAS .............................................................

31

ESPECIFICACIONES ............................................................

32

MANIPULACIÓN ...................................................................

35

Desembalaje.....................................................................................................

35

Inspección ........................................................................................................

35

Almacenaje .......................................................................................................

35

INTRODUCCIÓN ..................................................................

36

INSTRUCCIONES DE FUNCIONAMIENTO ...................

36

Escoja el lugar de uso ...................................................................................

36

Pautas para un rendimiento óptimo

 

recomendadas al usuario final ...............................................................

36

Enchufe el cable eléctrico ...........................................................................

37

Conecte el humidificador (si ha sido prescrito)...................................

37

Interruptor de encendido...........................................................................

39

Caudal del flujo...............................................................................................

39

Alarma por bajo flujo ....................................................................................

40

Indicador SensO2®de pureza del oxígeno (si la

 

unidad incluye la función detectora de O2) ......................................

40

Puesta en marcha del concentrador........................................................

40

Explicación de las luces indicadoras ........................................................

41

Indicadores (si la unidad no incluye el detector de O2).....................

41

MANTENIMIENTO DEL CONCENTRADOR ..................

42

Limpieza de los filtros de entrada .............................................................

42

Armario ............................................................................................................

43

Humidificador .................................................................................................

43

26

CONTENIDO

CONTENIDO (CONTINUACIÓN)

 

GUÍA PARA LA SOLUCIÓN DE

 

PROBLEMAS DEL CONCENTRADOR ..........................

43

REGISTRO DEL MANTENIMIENTO

 

PREVENTIVO .....................................................................

46

GARANTÍA DEL CONCENTRADOR ................................

47

FRENCH OWNER’S MANUAL

 

(MANUEL D’UTILISATION) ...........................................

49

GERMAN OWNER’S MANUAL

 

(BENUTZERHANDBUCH - DEUTSCH) ........................

73

BRAZILIAN PORTUGUESE OWNER’S MANUAL

 

(NANUAL DO OPERADOR) ...........................................

97

27

MEDIDAS DE SEGURIDAD

MEDIDAS DE SEGURIDAD

PRECAUCIÓN

“Precaución: Este aparato sólo debe usarse bajo la prescripción médica o de personal autorizado”.

Peligro de descarga eléctrica. NO desmonte este aparato. Sólo personal calificado debe dar servicio al equipo.

LEA LA SIGUIENTE INFORMACIÓN ANTES DE PONER EN FUNCIONAMIENTO ESTE EQUIPO.

ADVERTENCIA

La oxigenoterapia debe realizarse con sumo cuidado para reducir el riesgo de incendio. Cualquier material combustible o no en el aire es fácilmente inflamable en concentraciones altas de oxígeno. Por razones de seguridad, es preciso mantener todas las fuentes inflamables lejos del producto, y preferiblemente fuera de la habitación donde sea utilizado. Coloque señales de NO FUMAR en lugares visibles.

Puede producirse una inflamación espontánea y violenta si aceite, grasas u otras substancias entran en contacto con el oxígeno bajo presión. ALEJE estas substancias del concentrador de oxígeno, de sus tubos, conexiones y de todo otro equipo afín. NO use ningún tipo de lubricante a menos que éste sea recomendado por Invacare.

Para obtener un rendimiento óptimo, Invacare recomienda mantener los concentradores encendidos y funcionando por un mínimo de 30 minutos cada vez. Periodos de funcionamiento inferiores pueden reducir la vida útil máxima del producto.

Póngase en contacte con personal de servicio calificado para que repare el equipo en caso de que éste tenga el cable o enchufe dañados, si no funciona adecuadamente, si se ha caído, dañado o caído al agua.

Mantenga el cable lejos de superficies TEMPERADAS o CALIENTES.

No tire del cable eléctrico para mover ni trasladar el concentrador.

NUNCA deje caer ni introduzca ningún objeto en ninguna abertura.

28

MEDIDAS DE SEGURIDAD

ADVERTENCIA (CONTINUACIÓN)

NUNCA obstruya las aberturas de aire del producto ni ubique el equipo sobre una superficie blanda, como una cama o sofá donde la abertura de aire quede obstruida. Elimine todas las pelusas, cabello y demás impurezas. Llene con agua el humidificador hasta el nivel indicado por el fabricante.

NO REBASE EL NIVEL.

Mantenga el equipo a 7.5 centímetros (3 in) de las paredes, cortinas, muebles, etc.

Invacare recomienda usar el equipo con tubos antienredos (proporcionados por Invacare) que no sobrepasen los 15.2 metros (50 pies) de largo.

MANTENIMIENTO

El concentrador de oxígeno Invacare ha sido especialmente diseñado para minimizar el mantenimiento preventivo rutinario una vez al año. Sólo personal médico o personas con conocimientos técnicos sobre el equipo, tales como el personal autorizado o capacitado de la fábrica, pueden llevar a cabo el mantenimiento preventivo o realizar ajustes de rendimiento en el concentrador de oxígeno.

INTERFERENCIA DE RADIOFRECUENCIA

La mayoría de los equipos eléctricos se ven afectados por interferencia de radiofrecuencia. TENGA CUIDADO al usar equipos de comunicación portátiles cerca o alrededor de dichos equipos.

PARA REDUCIR EL RIESGO DE QUEMADURAS, ELECTROCUCIÓN, INCENDIO O LESIONES PERSONALES.

SIEMPRE desenchufe el producto INMEDIATAMENTE después de usarlo. No lo use mientras se bañe. Si el médico ha recomendado el uso continuo del equipo:

Coloque el concentrador en otra habitación, a por lo menos 2.1 metros (7 pies) de la bañera.

NO toque el concentrador si usted está mojado.

NO coloque ni guarde el producto en un lugar donde pueda caer al agua u otros líquidos.

NO recoja el producto si éste ha caído al agua. DESENCHÚFELO DE INMEDIATO.

SIEMPRE preste atención al equipo cuando esté enchufado.

29

MEDIDAS DE SEGURIDAD

ADVERTENCIA (CONTINUACIÓN)

Este dispositivo debe usarse tal como lo indica la receta médica y el manual del usuario. Si en cualquier momento el paciente o el asistente considera que el paciente está recibiendo una cantidad insuficiente de oxígeno, se debe comunicar de inmediato al proveedor y/o al médico. No haga ningún ajuste al caudal del flujo a menos que lo indique el médico.

Se debe supervisar continuamente el dispositivo cuando se use cerca de niños o personas discapacitadas.

Use este producto para lo cual fue diseñado, tal como se describe en este manual.

NO use partes, accesorios ni adaptadores distintos de los autorizados por Invacare.

NO conecte el concentrador en paralelo ni en serie con otros concentradores de oxígeno ni ningún otro dispositivo para la oxigenoterapia.

El uso de algunos accesorios de administración de oxígeno o ciertos humidificadores no especificados para ser usados con el concentrador pueden disminuir el rendimiento del mismo.

La terapia con oxígeno puede ser peligrosa en ciertas circunstancias. Invacare recomienda consultar al médico antes de usar esta máquina.

Este equipo está diseñado para usarse sólo como suplemento de oxígeno y no como un respirador artificial para pacientes terminales.

Evite crear chispas cerca del equipo médico de oxígeno. Ello incluye chispas de la electricidad estática causada por cualquier tipo de fricción.

CONCENTRACIÓN DE OXÍGENO

ADVERTENCIA

Procure no fijar el flujo por encima de la marca ROJA. Un flujo de oxígeno superior a 5 L/min. disminuirá la concentración de oxígeno.

30