Reliant Bath Lift (RBL250)
Pendant/Cable/Current Limit Device Replacement Kit
Assembly, Installation and Operating Instructions
Kit No. 1085144
NOTE: Check all parts for shipping damage. In case of damage, DO NOT use. Contact your Dealer for further instruction.
SAFETY SUMMARY
The following recommendations are made for the
safe replacement of the PENDANT/CABLE/CUR-
RENT LIMIT DEVICE:
GENERAL WARNINGS
DO NOT install or use this equipment without
first reading and understanding this instruction sheet. If you are unable to understand
the Warnings, Cautions or Instructions in this
Instruction Sheet and in the Owner's Operating and Maintenance Manual (Part No.
1081003) provided with the Bath Lift, contact
a healthcare professional, dealer or technical personnel before attempting to install this
kit, otherwise, injury or damage may occur.
This kit MUST be installed by an Invacare
Dealer or technical personnel.
INSTALLATION WARNINGS
After ANY adjustments, repair or service and
BEFORE use, make sure all attaching hardware is tightened securely - otherwise injury
or damage may occur.
Use proper lifting techniques to avoid injury. Keep back straight and bend knees
when lifting components of the bath lift.
REPLACING THE PENDANT/CABLE/
CURRENT LIMIT DEVICE
1. Secure the frame to the floor with the suction cups or by
standing on the frame base during removal of the seat.
2. Press the UP button on the pendant to raise the
seat all the way up (DETAIL "A").
3. Grasp the handgrip on the battery pack and pull
upwards to remove it from the motor shroud as
shown in DETAIL "B".
4. Press and hold the spring loaded chair release
button to release the seat (DETAIL "C").
5. While holding the button with one hand, grasp the
seat bottom firmly with your other hand.
6. Slide the seat up until the slide block disengages from
the groove in the frame as shown in FIGURE 1.
7. Grasp the handgrip on the battery pack and push
downwards to install onto the motor shroud.
8. Press the DOWN button on the pendant to lower
the bath lift to the lowest position (DETAIL "A").
9. If necessary, remove the bath lift frame from the
bathtub. Refer to
THE BATHTUB in this instruction sheet.
DETAIL "A"
STEPS 1 AND 8
Down Button
Pendant
REMOVING THE FRAME FROM
DETAIL "B"
STEPS 2 AND 7
PULL UP TO REMOVE
é
PUSH DOWN TO
INSTALL
ê
Handgrip
NOTE: Refer to the INSTALLATION WARNINGS in the
SAFETY SUMMARY of this instruction sheet.
Removing the Seat (FIGURE 1)
WARNING
To prevent pinching injury between seat
back and motor shroud when removing/
installing the seat, DO NOT grasp seat anywhere other than at the recess at the top
(DETAIL "C"), the opening at the bottom of
the seat or the lip at the top of the seat.
To prevent frame from tipping when removing seat, either suction the frame to the floor
or place one foot on the frame base when
removing/installing the seat.
CAUTION
The bath lift seat MUST be removed from the
top of the lifting cycle, otherwise, damage
to the unit may occur.
DO NOT grasp seat by side flaps, otherwise
damage to the bath lift or the bathtub may
occur.
Up
Button
DETAIL "C"
STEP 4
Spring Loaded Chair Release Button
LIFT THE SEAT
Seat
Recess
é
Slide
Block
Seat Bottom
FIGURE 1 - REMOVING THE SEAT
1
Groove
Frame
Motor
Shroud
Battery Pack
STEPS 5-6
Removing the Frame from the Bathtub
(FIGURE 2)
WARNING
ALWAYS lower the bath lift frame to the
lowest position BEFORE performing this
procedure. Otherwise, the bath lift frame
may tip during installation resulting in injury and/or damage.
CAUTION
ALL water MUST be drained from the bathtub BEFORE removal of the bath lift frame
to prevent damage to the bath lift frame
or suction cups.
ALWAYS release the suction cups from bath-
tub bottom BEFORE removing frame to prevent damage to the frame or bathtub.
1. Remove the seat, if not already done. Refer to
MOVING THE SEAT in this instruction sheet.
NOTE: Ensure bath lift is in lowest position before proceeding.
2. Grasp the handgrip on the battery pack and pull
upwards to remove it from the motor shroud.
3. Tilt the frame forward towards the drain or gently
pull up on the suction cup loops to release the suction cups from the bathtub.
4. Remove the frame from the bathtub.
5. Remove the motor shroud. Refer to
THE MOTOR SHROUD in this instruction sheet.
REMOVING
TILT FRAME
Frame
FORWARDS
RE-
WARNING
ALWAYS remove the bath lift from the bathtub BEFORE performing this procedure.
Otherwise - injury or damage may occur.
1. Remove the bath lift frame from the bathtub, if not
already done. Refer to
FROM THE BATHTUB in this instruction sheet.
2. Remove the mounting screw, two (2) self-locking
screws and three (3) washers securing the motor
shroud to the bath lift frame. Refer to FIGURE 3.
Washers
(STEP 2)
REMOVING THE FRAME
Motor Shroud
(STEP 3)
Mounting Screw
(STEP 2)
Washer
(STEP 2)
Self-Locking
Screws
(STEP 2)
FIGURE 3 - REMOVING THE MOTOR
SHROUD
CAUTION
The motor shroud is still connected to the
motor by the spade connectors and pendant cable. DO NOT attempt to forcibly
separate the motor shroud from the motor. Otherwise - damage to the motor
shroud and/or motor may occur.
3. Carefully lift the motor shroud off of the motor.
Suction Cup
Suction
Cup
FIGURE 2 - REMOVING THE FRAME FROM
Loop
Bathtub
THE BATHTUB
Removing the Motor Shroud (FIGURES 3
AND 4)
WARNING
ALWAYS lower the bath lift frame to the
lowest position BEFORE performing this
procedure. Otherwise, the bath lift frame
may tip during installation resulting in injury and/or damage.
4. Disconnect the RED and BLACK spade connector wires from the motor as shown in FIGURE 4.
5. Remove the two (2) mounting screws securing
the mounting bracket to the motor as shown in
FIGURE 4.
NOTE: Save the EXISTING two (2) mounting screws
for reuse when securing the NEW spade connector
wire and pendant cable to the motor.
6. Remove the mounting bracket and release the
spade connector wire and pendant cable as shown
in FIGURE 4.
7. Disconnect the battery connector from the motor
shroud. Refer to
CONNECTOR in this instruction sheet.
2
REMOVING THE BATTERY
Spade Connector Wires
BLACK Spade
Connector
RED
Spade
Connector
Nut
Connection to
Motor Shroud
Pendant
Cable
STEPS 1 - 2
YELLOW (+) Wire
Battery
Connector
Mounting
Screws
(Nuts NOT
SHOWN)
Mounting
Bracket
Mounting Screws
FIGURE 4 - REMOVING THE MOTOR SHROUD
- RELEASING THE CABLES
Removing the Battery Connector (FIGURE 5)
1. Remove the two (2) mounting screws and nuts
securing the battery connector to the motor shroud
as shown in FIGURE 5.
2. Carefully pull the battery connector away from the
motor shroud.
3. Loosen, but do not remove the two (2) flat screws
on the battery connector to release the YELLOW
(+) and BLUE (-) wires.
4. Pull the YELLOW (+) and BLUE (-) wires through
the motor shroud.
5. Remove the current limit device. Refer to
ING CURRENT LIMIT DEVICE in this instruction
sheet.
REMOV-
BLUE (-)
Wire
Flared End
YELLOW (+) Wire
DETAIL "A"
STEP 3
LOOSEN
Flat Screw
(DO NOT
REMOVE)
BLUE (-)
Wire
Insulated
Motor
Shroud
FIGURE 5 - REMOVING THE BATTERY
Battery
Connector
CONNECTOR
Wire
Bare
Wire
Removing Current Limit Device (FIGURE 6)
1. Remove the two (2) mounting screws and nuts securing the current limit device to the motor shroud.
NOTE: EXISTING mounting screws and nuts will be
used to secure NEW current limit device to motor
shroud.
2. Remove the pendant/cable/current limit device assembly.
3. Install NEW pendant/cable/current limit device assembly. Refer to
LIMIT DEVICE in this instruction sheet.
3
INSTALLING NEW CURRENT