Invacare 1034177 User Manual

PATIENT OPERA TED WHEEL LOCKS FOR 2000/4000 RECLINER WHEELCHAIRS KIT NUMBERS 1034177
Installation, Assembly and Operating Instructions
NOTE: Check ALL parts for shipping damage. In case of shipping damage, DO NOT use. Contact Carrier/ Dealer for further instruction.
SAFETY SUMMARY
The following recommendations are made for safe in­stallation and adjustment of the Patient Operated Wheel Locks:
GENERAL WARNINGS
DO NOT install or use this kit without first read­ing and understanding this instruction sheet. If you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel be­fore attempting to install this kit - otherwise, in­jury or damage may occur.
INSTALLATION WARNINGS
After making adjustments and BEFORE use, make sure all attaching hardware is securely tightened.
INSTALLATION CAUTIONS
When securing the strap and stud assembly to the pivot arm assembly, DO NOT overtighten the attaching hardware.
3. Remove the existing tube clamp from the wheel­chair frame.
4. Install the plug button into end of the NEW pivot arm assembly.
5. Insert the NEW strap and stud assembly through the pivot arm assembly and secure with locknut.
6. Install the NEW tube clamp over the frame and measure 4-3/16-inches from the back tube of the frame assembly to the center of the tube clamp.
7. Insert the NEW 2-inch mounting screw through the NEW tube clamp, bushing, strap/stud assem­bly and thin washer.
8. Install locknut onto 2-inch mounting screw. T ighten until snug fit is obtained.
9. Insert the pivot arm spacer into the end of the NEW pivot arm assembly .
10. Insert the existing axle mounting screw through the rear wheel, pivot arm spacer, NEW pivot arm assembly and axle lug.
1 1. Install locknut onto axle mounting screw. Securely
tighten.
12. Attach the wheel lock assembly to the NEW pivot arm assembly by using the 3/4-inch mounting screw and heavy washer.
REPLACING THE PATIENT OPERATED WHEEL LOCKS (FIGURE 1)
NOTE: Refer to INSTALLATION WARNINGS and CAUTIONS in the SAFETY SUMMARY in this instruc-
tion sheet. NOTE: If converting from attendant operated wheel
lock, refer to CONVERTING FROM ATTENDANT OPERA TED WHEEL LOCKS in this instruction sheet.
1. Remove the axle mounting screw and locknut that secure the rear portion of the existing patient op­erated wheel lock, pivot arm spacer and rear wheel to the wheelchair frame.
2. Remove the existing 2-inch mounting screw, bush­ing, thin washer, and locknut that secure the ex­isting strap and stud assembly to the wheelchair frame. Refer to DETAIL “A” in FIGURE 1.
13. Repeat STEPS 1-12 for opposite wheel lock.
14. Adjust the wheel locks. Refer to ADJUSTING THE P ATIENT OPERA TED WHEEL LOCKS in this in­struction sheet.
1
Back Tube
(STEP 6)
2-inch
Mounting Screw
(STEPS 2,7)
Plug Button
(STEP 4)
Locknut
(STEPS 1,11)
Axle Lug
(STEP 10)
Wheelchair
Frame
Tube Clamp
(STEPS 3,6,7)
Bushing
(STEPS 2,5,7)
Strap/Stud
Assembly
(STEPS 2,7)
Thin Washer
(STEPS 2,7)
DETAIL A
Locknut
(STEPS 2,8)
Pivot Arm Assembly
(STEPS 1,4,5,9,10,12)
Pivot Arm
Spacer
(STEPS 1,9,10)
Locknut
(STEP 5)
Rear Wheel
(STEPS 1,10)
Heavy
Washer
(STEP 12)
3/4-inch
Mounting
Screw
(STEP 12)
Wheel Lock
Assembly
(STEP 12)
Axle Mounting
Screw
(STEPS 1,10,11)
FIGURE 1 - REPLACING THE PATIENT OPERATED WHEEL LOCKS
2
ADJUSTING THE PATIENT OPERATED WHEEL LOCKS (FIGURE 2)
NOTE: Refer to INSTALLATION WARNINGS in the SAFETY SUMMARY in this instruction sheet.
NOTE: If wheels are pneumatic, before adjusting or replacing the wheel lock assemblies, ensure that the tires are inflated to the recommended psi on the side wall of tire. The recommended tire pressure is located on the side wall of the tire.
1. Loosen the bolt and locknut that secure the wheel lock assembly to the wheelchair frame.
2. Adjust the position of wheel lock until the mea­surement between the rear wheel and the wheel lock shoe is between 5/32 and 5/16-inches.
USING THE PATIENT OPERATED THE WHEEL LOCKS (FIGURE 3)
WARNING
DO NOT attempt to stop a moving wheelchair with the wheel locks. WHEEL LOCKS ARE NOT BRAKES.
1. Ensure the wheelchair is not moving before en­gaging the wheel locks.
2. Perform one (1) of the following: A. Push-to-Lock - To engage, push the wheel
lock handle forward.
B. Pull-to-Lock - To engage, pull the wheel lock
handle backward.
3. Disengage the wheel locks by reversing STEP 2.
3. Securely tighten the bolt and locknut.
4. Engage wheel lock and ensure the wheelchair holds its position.
5. Repeat the above procedures until the wheel lock holds the wheelchair.
6. Repeat STEPS 1-5 for the opposite wheel lock.
Bolt and Locknut
Wheel Lock Handle
Wheel
Lock
Rear
Wheel
5/32 and 5/16-inches
Wheel Lock
Shoe
FIGURE 2 - ADJUSTING THE PATIENT
OPERATED WHEEL LOCKS
Unlocked
Position
Unlocked
Position
PUSH TO LOCK
Wheel Lock
PULL TO LOCK
Wheel Lock
Locked
Position
Locked
Position
FIGURE 3 - USING THE PATIENT OPERATED