NB: The control unit must always be located outside the incubator!
– Connect C ELLRO LL t o the control u nit with the cable
provided.
If connection has not been made, "Err 02" will appear on the
dis play on starting the instrument. Check the connection and
cancel the error message by pressing "ENTER".
– In the event of a power failure, any unstored data of both out-
puts will be automatically stored and recalled once the power has
been restored. The "Err 03" in the display can be cancelled by
pressing "ENTER".
Entering the program for working with CELLROLL
☞ Switch on the instrument by pressing the ON/OFF key. The
green LED under "RECALL" will light.
☞ Press "SELECT" and select the function "SPEED".
In the function "SPEED":
Selecting the output:
☞ Select OUTPUT 1 or OUTPUT 2 via key "< >".
Setting the rolling speed:
CELLROLL is equipped with two different motors. The control unit
will always select the right one.
"CELLROLL" Motor unit (0.1-2.0 rpm):
☞ Set the speed within the range 0.1 - 2.0 rpm via "∨" and "∧".
Confirm by pressing "ENTER".
"CELLROLL" Motor unit fast (2.0-6.0 rpm):
☞ Set the speed within the range 2.0 - 6.0 rpm via "∨" and "∧".
Confirm by pressing "ENTER".
In the function " ":
Selecting the pendular angle:
☞ For continuous rolling (no directional change), set the
pendulum angle to "0°" via "∨". Confirm by pressing "ENTER".
☞ For rocking (with directional change), set the pendulum angle
within the range 180° - 1440° (= 0.5 - 4.0 bottle revolutions).
Confirm by pressing "ENTER".
NB: If "rocking" has been selected, please observe the restrictions
impos ed below under " " and "–".
In the function "":
Set the rolling duration:
The duration of continuous rolling or rocking can be set to:
- A limited time between 00.05 (h/min) - 99.59 (h/min), or
- An unlimited time as "Con".
☞ To select a specific rolling duration
, set hours and/or minutes
via the "< >" key and alter via "∨" and "∧". Confirm by pressing
"ENTER".
☞ To select continuous rolling or rocking operation, activate the
h/min function and press the "∨" key until the display shows
less than 0.05 and "Con" appears. Confirm by pressing
"ENTER".
11
3.3 Abridged operating instructions - CELLROLL