Federal (USA) law restricts this device to sale by or on the order of a
physician.
Publication Name
Integra Camino ICP Monitor User’s Manual
Integra Part Number 60903733 Rev. B
Effective Date September 2012
Tradema rk Acknowledgements
Integra, the Integra logo, and Camino are registered trademarks of Integra
LifeSciences Corporation or its subsidiaries in the United States and/or
other countries. Delta-Cal is a trademark of Utah Medical Products, Inc.
Sani-Cloth is a registered trademark of Professional Disposables
International, Inc. Aptimax, Sealsure, and Sterrad are registered trademarks
of Johnson and Johnson. Tyvec is a registered trademark of E.I. duPont.
Kimguard and One-Step are registered trademarks of Kimberly-Clark
Worldwide, Inc.
Packaging and Label Symbols................................................................................................................ v
Software Symbols .................................................................................................................................. vi
List of Abbreviations .............................................................................................................................. vii
Chapter 1: System Overview
Indications for Use / Intended Use ..........................................................................................................1
Intended Patient Population ........................................................................... 1
Description of the Integra Camino ICP Monitor............................................ 2
Reviewing the User’s Manual ......................................................................... 2
List of Warnings for Using the Monitor ......................................................... 3
Parts of the Monitor......................................................................................... 6
Indications for Use / Intended Use
The Integra® Camino® ICP Monitor is indicated for use by qualified neurosurgeons
or neurointensivists for measurement of intracranial pressure and temperature.
Contraindications
The Integra Camino ICP Monitor and its accessories are contraindicated for use in
a Magnetic Resonance (MR) environment.
Intended User
The Integra Camino ICP Monitor is intended to be used by the following qualified
medical and biomedical professionals:
•A qualified neurosurgeon should perform the placement and handling of the
physician, or physician’s assistant) should perform the operation of the monitor.
Chapter 9 provides instructions for testing and maintaining the monitor. The
procedures in this chapter are intended to be performed by the hospital’s
biomedical engineering staff.
Intended Patient Population
Patients undergoing treatment with this monitor under the cranial applications are
expected to have had a TBI, undergone a major neurosurgical procedure, or some
other traumatic, ischemic or hemorrhagic incident requiring controlled
monitoring of ICP and brain temperature.
Chapter 1 • System Overview • 1
Description of the Integra Camino ICP Monitor
The Integra Camino ICP Monitor is a compact, portable device that provides tools
for continuously determining and monitoring intracranial pressure (ICP) and
intracranial temperature (ICT) directly in the brain, depending on which catheters
are connected to the system. This monitor supports the following catheters:
•Series of Integra® Camino® Fiber Optic Catheters (110-4 series) for measuring both
ICP and temperature.
•Integra® Camino® Flex Catheter for measuring ICP values.
All Integra catheters measure their respective values at the tip of the catheter. This
design eliminates the need for a fluid-filled system to communicate pressure (and
carry pressure waves) to an external transducer.
Key Functions of Monitor
During clinical use, the Integra Camino ICP Monitor provides several key functions
to facilitate the process for monitoring and analyzing patient data:
•Touch screen interface for evaluating patient ICP/ICT data and setting patient
parameters (see
•Physiological alarm that activates if the patient’s Mean ICP value exceeds a
user-specified limit for more than 5 seconds (see
page 35)
page 41)
•Rechargeable lithium ion battery that supplies power to monitor during patient
transport (see
•Storage of patient’s ICP trend data for up to 5 days (see page 39)
•Outputs for transferring patient data to a patient bedside monitor (see page 30)
•Outputs for extracting patient data to remote media types via USB drive or digital
streaming (see
For instructions on using the Integra catheters, see the directions for use supplied
with each respective catheter.
Reviewing the User’s Manual
Integra recommends that all physicians, nurses, and technicians who will be using,
operating, and maintaining the Integra Camino ICP Monitor review this user’s
manual prior to using the system. If there are additional questions after reading
this manual, contact Integra.
page 21)
page 57)
2 • Chapter 1 • System Overview
List of Warnings for Using the Monitor
Failure to observe one or more of the following warnings could compromise
patient safety or result in measurement errors.
Warnings
• Use of the Integra Camino ICP Monitor is restricted to one patient at a time.
• The Integra Camino ICP Monitor and its accessories are contraindicated for use in a
Magnetic Resonance (MR) environment.
• Always verify the high ICP alarm limit is set appropriately for each patient prior to
treatment.
• Selecting the
limit indefinitely. Use caution if this feature is selected. To re-enable this alarm, press
the
Alarm On and Accept buttons.
• No modification of the Integra Camino ICP Monitor is allowed.
• The Integra Camino ICP Monitor is a sensitive electronic device. When using the
monitor, always handle with care. If damage is suspected, contact Integra.
• Read the user’s manual from the patient bedside monitor’s manufacturer before
connecting the Integra Camino ICP Monitor to a patient bedside monitor.
• To prevent injury to the patient, user, or other persons, or damage to the monitor,
always verify that the monitor is clamped securely to the equipment pole.
• To reduce the risk of electric shock, do not disassemble the Integra Camino ICP
Monitor. Refer all servicing to qualified service personnel at Integra.
• To prevent electrical shock, only use the AC power adapter supplied by Integra
(REF # MONPWR). Using a different AC power adapter may not provide protection
against electric shock.
• Danger - Possible explosion hazard if used in the presence of flammable anaesthetics.
• Only use Integra supplied accessories on the Integra Camino ICP Monitor. This applies
in particular to catheters, catheter cables, battery, AC power adapter, and
USB-to-R232 adapter cable.
Alarm Off feature on the Alarm panel will disable the high ICP alarm
• If the Integra Camino ICP Monitor loses power and shuts down while it is connected to
a patient bedside monitor, do not use the ICP values on the patient bedside monitor
for patient measurements; the ICP values on the patient bedside monitor will be
invalid.
• Connect the monitor to an AC power supply immediately if the low battery alarm is
activated.
Chapter 1 • System Overview • 3
Warnings
• When using the battery:
-Do not heat above 80 °C.
-Do not open battery.
-Do not dispose of in fire.
-Do not short circuit as battery may ignite, explode, leak, or get hot causing
personal injury.
-Replace battery with same part number only (REF # BAT1001).
-Use of another battery may present a risk of fire or explosion.
• To prevent injury to the patient, user, or other persons, make sure that the battery
cover is closed securely during monitor use.
• Do not connect an Integra Camino Fiber Optic Catheter and an Integra Camino Flex
Catheter simultaneously to the monitor. The Integra Camino ICP Monitor is designed
to report ICP measurements with either the Flex Catheter or Fiber Optic Catheter
connected to the monitor, but not both together.
• To prevent possible patient injury resulting from incorrect ICP measurements, always
perform the steps listed on page 23 before implanting a new Integra Camino Fiber
Optic Catheter into the patient. In particular, always make sure to adjust the new Fiber
Optic Catheter’s ICP value to zero before implantation while the catheter is in the air.
Never attempt to re-zero a catheter while the catheter is inside the patient.
• Once the Integra Camino Fiber Optic Catheter has been zeroed to the Integra Camino
ICP Monitor, do not replace the Fiber Optic Catheter Cable being used for patient
measurement. Replacing the Fiber Optic Catheter Cable with another cable after the
catheter has already been zeroed may result in inaccurate patient measurements.
• Once the Integra Camino Fiber Optic Catheter has been zeroed to the Integra Camino
ICP Monitor, do not transfer this zeroed catheter to any other monitor. Transferring a
zeroed catheter to a different monitor may result in inaccurate ICP measurements.
• To prevent possible patient injury resulting from incorrect ICP measurements, always
perform each of the steps listed on page 26 before implanting a new Integra Camino
Flex Catheter into the patient. In particular, always leave the Flex Catheter in the air
until the monitor completes the autozero (i.e. initialization) process successfully.
• Once the Integra Camino Flex Catheter has been initialized (autozeroed) by the Integra
Camino ICP Monitor, do not replace the Integra Camino Flex Extension Cable being
used for patient measurement. Replacing the Flex Extension Cable with another cable
after the Flex Catheter has already been initialized may result in inaccurate ICP
measurements.
• Once the Integra Camino Flex Catheter has been initialized (i.e. autozeroed) by the
Integra Camino ICP Monitor, do not transfer this initialized catheter to any other
monitor. Transferring an initialized catheter to a different monitor may result in
inaccurate ICP measurements.
4 • Chapter 1 • System Overview
Warnings
• The Integra Camino ICP Monitor will only store the Mean ICP data from the most
recent 5 days. All stored trend data older than 5 days will be lost. If monitoring is
continued for more than 5 days, placement of a new catheter under sterile conditions
is recommended. Note that replacing a catheter will reset the trend data. Please
extract any data that you wish to retain prior to replacing the catheter.
• Do not autoclave or immerse the Integra Camino ICP Monitor in liquid as damage may
occur. If the monitor is exposed to liquids, turn off the unit, remove the AC power
adapter, dry the unit thoroughly, and send to biomed staff for evaluation before
reapplying power.
• Only use the cleaning agents listed in Chapter 7 for cleaning and disinfecting the
Integra Camino ICP Monitor system. Using solvents or cleaning agents not listed in
Chapter 7 may damage the plastic exterior of the Integra Camino ICP Monitor.
Chapter 1 • System Overview • 5
Parts of the Monitor
2
3
4
1
About the Front Panel
The Integra Camino ICP Monitor contains hardware, software, and electrical
components that support specific Integra catheters for monitoring the patient’s
ICP and temperature. The following section provides information on the different
parts of the monitor.
The front panel contains:
NumberItemDescription
1HandleHandle used for carrying the monitor.
2Power StatusGreen LED button that indicates the monitor is
being powered by the AC power adapter. Note
that this button does not illuminate if the monitor
3Power ButtonTurns the monitor on and off. This button is
4Touch ScreenProvides software tools for viewing data and
is being powered by the battery.
illuminated when the power is on.
controlling parameters for monitoring the
patient’s ICP and ICT levels.
6 • Chapter 1 • System Overview
About the Rear Panel
1
2
5
6
4
3
The rear panel contains:
NumberItemDescription
1USB PortConnection port for extracting trend data via
USB transfer or digital streaming.
2AC Power Adapter PortConnection port for the AC power cord.
3PMIO PortConnection port for PMIO cable. This cable
is used to connect the Integra Camino ICP
Monitor to a patient bedside monitor.
4Pole ClampClamping system for securing monitor to an
5Air VentGrated opening that allows air being
6Cable StrapRubber strap used to secure AC power
equipment pole.
circulated by the internal cooling fan to leave
the monitor.
adapter during transport.
Chapter 1 • System Overview • 7
About the Right Panel
1
2
3
The right panel contains:
NumberItemDescription
1Temperature PortConnection port for the temperature connector on
the Fiber Optic Catheter Cable.
2Pressure PortConnection port for the ICP connector on the
3Pressure PortConnection port for the ICP connector on the
Fiber Optic Catheter Cable.
8 • Chapter 1 • System Overview
About the Bottom Panel
1
2
The bottom panel contains:
About the Left Panel
NumberItemDescription
1Battery Door CoverRemovable cover for accessing/replacing the
14.4V lithium ion battery.
2Air VentGrated opening that allows air being circulated
by the internal cooling fan to leave the monitor.
The left panel does not contain any usable connector ports or buttons.
Chapter 1 • System Overview • 9
This page is intentionally left blank.
10 • Chapter 1 • System Overview
CHAPTER 2 SETTING UP SYSTEMFORTHE FIRST TIME
Procedures for Initial Setup
Step 1: Unpack the System (REF # CAM02)
Remove the contents from the Integra® Camino® ICP Monitor shipping box and
verify the following items are included.
Warning
Only use Integra supplied accessories on the Integra Camino ICP
Monitor. This applies in particular to catheters, catheter cables, battery,
AC power adapter, and USB-to-R232 adapter cable.
ContentDescription
• Integra Camino ICP Monitor
•Quantity = 1
• AC power adapter (18 V DC , 1.67 A, 30 W)
•REF # MONPWR
•Quantity = 1
• 14.4 V lithium ion battery
•REF # BAT1001
•Quantity = 1
Chapter 2 • Setting Up System for the First Time • 11
ContentDescription
• Integra® PMIO patient bedside monitor cable
• REF # PMIOMPM
•Quantity = 1
• Integra® Camino® Fiber Optic Catheter Cable
• REF # CAMCABL
•Quantity = 1
• Integra® Camino® Flex Extension Cable
•User manual
• REF # UM-CAM02
(where
XX refers to the country’s language code)
•Quantity = 1
• REF # FLEXEXT
•Quantity = 1
• USB-to-RS232 adapter cable
• REF # EXPORTCAB
•Quantity = 1
XX
After unpacking the contents, inspect the shipment for any signs of damage or
loss. If any damages are discovered, notify the carrier, the supplier, and retain all
shipping cartons for examination.
12 • Chapter 2 • Setting Up System for the First Time
Step 2: Install the Battery
Perform the following steps with a Phillips screwdriver and the Integra-supplied
14.4 V lithium ion battery.
1. Make sure the monitor is unplugged and turned off.
2. Turn the monitor upside down so the handle is facing downward.
3. Remove the 2 screws to take off the battery cover.
4. Verify the Integra logo on the battery is facing up and insert the battery:
A. Align the battery’s connectors on the front of the battery to the connector
slots on the monitor.
B. Slide the battery’s connector into the monitor’s connector slots until the
battery is fully inserted into place.
5. Secure the battery cover to the monitor by re-inserting the two small screws.
Chapter 2 • Setting Up System for the First Time • 13
Step 3: Prepare the AC Power Adapter with Region-Specific Plug
Backside of AC power adapter
USUKEuropeAustralia
Region-Specific Adapter Plugs
Prongs on AC power adapter
Slots on
adapter plug
1. Remove the AC power adapter from the package and attach the region-specific
adapter plug to the backside of the AC power adapter.
A. Align the two slots on the adapter plug over the two prongs on the AC
power adapter.
B. Push the adapter plug down over the two prongs until the entire base of
the plug sits flush against the AC power adapter.
14 • Chapter 2 • Setting Up System for the First Time
C. Gently slide the adapter plug forward until it snaps into place.
Press the release button to
remove adapter plug
Connector end of
AC power adapter
2. To remove the adapter plug, press the release button on the AC power adapter
and pull the plug away from the adapter.
Step 4: Plug the Monitor into AC Power
Perform the following steps with the Integra-supplied AC power adapter:
1. Place the monitor on a flat surface.
2. Connect the AC power adapter to the Integra Camino ICP Monitor:
A. On the back of the monitor, attach the connector end of the AC power
adapter into the port labeled
Input: 18 V .
B. Insert the plug end of the AC power adapter into a grounded AC wall
outlet.
3. Turn on the monitor. On the front of the monitor, press the power button; the
Integra Camino ICP Monitor will display the Integra logo before initiating the
monitor setup process.
Chapter 2 • Setting Up System for the First Time • 15
4. After the initial setup process completes, the Integra Camino ICP Monitor
Settings tab
Set Language
button
sounds a one-second startup tone and displays the
Main panel on the touch
screen.
Caution
The purpose of the startup tone verifies that the audio alarms are
functioning correctly. If this tone does not sound during the startup
process, contact Integra for service.
Step 5: Set the Time and Date
1. On the touch screen, press the Settings tab and select Set Time and Date.
2. On the displayed panel, press the desired field (hour, minutes, date, month, or
year) and use the arrows to specify the appropriate setting. Note that you may
adjust each of these settings prior to accepting them in the following step.
3. Press Accept; the Integra Camino ICP Monitor will display the selected time/date
on the touch screen.
Step 6: Specify the Language
1. On the touch screen, press the
Settings tab (fourth tab from the left)
2. Press Set Language.
16 • Chapter 2 • Setting Up System for the First Time
3. In the displayed Language: menu,
use the arrows to select the desired
language.
4. Press Accept; the Integra Camino ICP Monitor will display all of the text that
appears on the touch screen in the selected language.
Step 7: Use AC Power to Charge the Battery to Full Capacity
1. Turn off the monitor. On the front of the monitor, press the power button.
2. Keep the Integra Camino ICP Monitor on AC power with the monitor turned off
for 5 hours. This will re-charge the battery to full capacity.
3. After 5 hours, turn on the monitor by pressing the power button.
4. On the touch screen, view the battery power symbol on the status
bar to verify the symbol displays four green bars; this indicates the
battery has full charge.
Chapter 2 • Setting Up System for the First Time • 17
This page is intentionally left blank.
18 • Chapter 2 • Setting Up System for the First Time
CHAPTER 3 SETTING UP SYSTEMFOR CLINICAL USE
Setting Up System for Clinical Use.............................................................. 19
About the Integra Catheters ......................................................................... 23
Connecting the Integra® Camino® Fiber Optic Catheters (110-4 Series) 24
Connecting the Integra® Camino® Flex Catheters .................................... 27
Connecting to a Patient Bedside Monitor (if applicable) ........................... 30
Storing the System........................................................................................ 33
Setting Up System for Clinical Use
The following section contains instructions for positioning the monitor, powering
the monitor, and connecting catheters to the monitor prior to clinical use.
Positioning the Monitor
The Integra® Camino® ICP Monitor is intended to be positioned on a hard flat
surface or securely clamped to an equipment pole or bed support next to the
patient. The distance between the patient and the monitor is restricted by the
length of the Integra catheter cable. It is not permitted to extend the Integra
catheter cable with non-Integra extension cables.
Caution
To prevent the monitor from overheating:
•Do not block the air vents on the rear and bottom of the unit
•Do not place the monitor on a soft or uneven surface, which may result in
blockage of the monitor’s air vents. For example, do not place the monitor
on the bed during patient transport.
Attaching to Equipment Pole (if applicable)
The Integra Camino ICP Monitor includes a clamp for attaching the monitor to an
equipment pole. To attach:
1. On the rear of the monitor, fit the pole clamp around the equipment pole.
Chapter 3 • Setting Up System for Clinical Use • 19
2. Tighten the knob on the pole clamp to secure the monitor to the equipment
Pole clamp attached
Pole Clamp
to equipment pole
Connector end of
AC power adapter
AC power adapter
pole.
Note that the pole clamp supports equipment poles between 0.5 and 1.2 inches
(12.7 and 30.5 mm) in diameter.
Warning
To prevent injury to the patient, user, or other persons, or damage to the
monitor, always verify that the monitor is clamped securely to the
equipment pole.
Caution
To prevent liquid from dripping inside the monitor and damaging the
internal components, do not mount the monitor underneath an I.V. bag or
tube feed. If liquid does drip onto the monitor, dry the monitor
immediately.
Powering the System On and Off
Turning On the System
Warning
To prevent electrical shock, only use the AC power adapter supplied by
Integra (REF # MONPWR). Using a different AC power adapter may not
provide protection against electric shock.
1. Plug the monitor to an AC power outlet:
A. On the back of the monitor, attach the connector end of the AC power
adapter into the port labeled
Input: 18 V .
20 • Chapter 3 • Setting Up System for Clinical Use
B. Insert the plug end of the AC power adapter into a grounded AC wall outlet.
Power Button
2. Turn on the monitor:
A. On the front of the monitor, press the power button.
B. Once the button illuminates, the Integra logo will appear
on the touch screen for a few seconds before initiating the
setup process.
3. After the setup process completes, the monitor will sound a one-second
startup tone and display the
Caution
Main panel:
The purpose of the startup tone verifies that the audio alarms are
functioning correctly. If this tone does not sound during the startup
process, contact Integra for service.
Warning
To prevent injury to the patient, user, or other persons, or damage to the
monitor, make sure to position the cables so that they are free from all foot
traffic.
Turning Off the System
On the front of the monitor, press the power button. If the monitor ever freezes
and does not turn off, press and hold the power button for several seconds to
perform a forced shutdown of the system.
Using the Battery for Power
The Integra Camino ICP Monitor includes a rechargeable 14.4 V lithium ion battery
that supplies power to the monitor for at least 1.5 hours when the battery is fully
charged; the battery is only intended for use during patient transport.
•To charge the battery to full capacity, turn off the monitor and plug it into an AC
•To ensure the battery maintains charge during patient use, always plug the
If the battery exhibits problems powering the monitor for 1.5 hours, perform the
"Battery Charge" test on page 86 to ensure the battery is functioning properly.
Battery Power Indicator
On the touch screen, a battery power indicator appears in the status
bar that displays the amount of battery charge available (see page 36).
outlet for at least 5 hours prior to use.
monitor to an AC outlet whenever possible.
•If the battery charge level drops to 15 minutes or less, the monitor will activate
visual warnings.
•If the battery charge level drops to 5 minutes or less, the monitor will sound an
alarm.
Chapter 3 • Setting Up System for Clinical Use • 21
For more information on responding to low battery alarms, see page 52. For
specific information on testing/replacing the battery, see page 85.
Warning
When using the battery:
• Do not heat above 80 °C.
• Do not open battery.
• Do not dispose of in fire.
• Do not short circuit as battery may ignite, explode, leak, or get hot
causing personal injury.
• Replace battery with same part number only (REF # BAT1001).
• Use of another battery may present a risk of fire or explosion.
Warning
To prevent injury to the patient, user, or other persons, make sure that the
battery cover is closed securely during monitor use.
Caution
To ensure appropriate battery power with use of the Integra Camino ICP
Monitor, only use batteries supplied by Integra (REF # BAT1001).To o r de r
replacement batteries, contact Integra.
Storing the Battery
Notice
The Integra Camino ICP Monitor has been designed to operate during
use while a battery is installed in the unit (even if the battery is not being
used for power). As a result, always use the monitor with a battery
installed.
If the Integra Camino ICP Monitor will not be used for several months or longer,
remove the battery from the monitor prior to storing the monitor. Ensure the
battery is stored in a cool, dry, and well-ventilated area.
Notice
If the Integra Camino ICP Monitor is not used for a long duration of time,
the battery will lose charge. Always make sure the battery is charged to
full capacity prior to use. The Integra Camino ICP Monitor will charge the
battery while the monitor is plugged into AC power.
22 • Chapter 3 • Setting Up System for Clinical Use
About the Integra Catheters
The Integra Camino ICP Monitor supports the following Integra catheters:
CatheterMeasurementsDescription
Integra Camino Fiber Optic
(110-4 Series)
Integra Camino FlexICP
ICP/Temperature
For complete instructions on using catheters, see the directions for use supplied
by Integra with each respective catheter.
Warning
Do not connect a Fiber Optic Catheter and a Flex Catheter simultaneously
to the monitor. The Integra Camino ICP Monitor is designed to report ICP
measurements with either the Flex Catheter or Fiber Optic Catheter
connected to the monitor, but not both together.
Fiber optic pressure/temperature catheters and
pressure transducer-tipped catheters.
A silicon strain gauge pressure transducer-tipped 9/10
FR catheter.
Chapter 3 • Setting Up System for Clinical Use • 23
Connecting the Integra® Camino® Fiber Optic Catheters (110-4 Series)
Pressure
Fiber Optic Catheter Cable
Temperature
connector from
connector from
catheter cable
catheter cable
Connecting the Fiber Optic Catheters (110-4 series) to the Integra Camino ICP
Monitor requires the Integra Camino Fiber Optic Catheter Cable
(REF # CAMCABL)
Warning
To prevent possible patient injury resulting from incorrect ICP
measurements, always perform each of the following steps before
implanting a new Fiber Optic Catheter into the patient. In particular,
always make sure to adjust the new Fiber Optic Catheter’s ICP value to
zero before implantation while the catheter is in the air. Never attempt to
change the zero adjustment while the catheter is inside the patient.
Notice
Before implanting the catheter into the patient, ensure that the
temperature is at a reasonable value, such as room temperature.
To c on ne ct:
1. On the monitor’s right side, connect the Fiber Optic Catheter Cable to the ICP
and temperature ports:
:
A. Attach the cable’s ICP connector (P) into the port labeled ICP (CAMCABL).
To a tt ac h:
• Align the red dot on the cable connector with the red triangle on the
• Verify the cable is completely connected to the monitor before
B. If temperature will be monitored for the patient, attach the cable’s
temperature connector (
attach:
• Align the red dot on the cable connector with the red triangle on the
• Verify the cable is completely connected to the monitor before
24 • Chapter 3 • Setting Up System for Clinical Use
monitor’s port and push firmly.
proceeding.
T) into the port labeled ICT (CAMCABL). To
monitor’s port and push firmly.
proceeding.
2. When operating in the sterile field, select the desired Integra® Camino®
ICP transducer from
Temperature thermistor from
catheter
ICP connector from
catheter cable
Temperature connector from
catheter cable
catheter
Zero adjustment tool
Transducer from catheter
Pressure Monitoring Kit (model 110-4 series). For instructions on using the kit,
see the directions for use supplied with each kit.
3. Remove the catheter from its sterile tray and firmly attach its ICP and
temperature connections to the Fiber Optic Catheter Cable:
A. Attach the ICP transducer from the catheter to the ICP connector on the
Fiber Optic Catheter Cable.
B. If applicable, attach the temperature thermistor from the catheter to the
temperature connector on the Fiber Optic Catheter Cable.
4. Note that when the temperature catheter is first connected and in open air
(room temperature), the Integra Camino ICP Monitor will:
• Sound an alarm; and
•Display "Temperature out of accuracy range" on the
status bar.
To silence the alarm temporarily (3 minutes) while the
catheter is still in air, press the yellow alarm button.
5. Verify that the ICP value reads 0 mmHg on the monitor. To
verify, press the
value does not read
A. Press the Scale button on the Main tab to select a waveform range of
-10 to 20 mmHg. Note that pressing the Scale button each time advances
to the next range.
B. Remove the black zero adjustment tool from the catheter kit.
C. Use the tool to adjust the screw located on the bottom of the catheter’s
transducer to increase/decrease the ICP value.
Main tab on the touch screen and view the ICP value. If the ICP
0 mmHg:
D. Set the ICP value to 0 mmHg.
Chapter 3 • Setting Up System for Clinical Use • 25
6. Prior to implanting the catheter into the patient, disconnect the Fiber Optic
Catheter from the Fiber Optic Catheter Cable.
7. After implanting the catheter into the patient, reconnect the catheter to the
Fiber Optic Catheter Cable.
Warning
Once the Fiber Optic Catheter has been zeroed to the Integra Camino
ICP Monitor, do not replace the Fiber Optic Catheter Cable being used
for patient measurement. Replacing the Fiber Optic Catheter Cable with
a different cable after the ICP catheter has already been zeroed may
result in inaccurate patient measurements.
Warning
Once the Fiber Optic Catheter has been zeroed to the Integra Camino
ICP Monitor, do not transfer this zeroed catheter to any other monitor.
Transferring a zeroed catheter to a different monitor may result in
inaccurate ICP measurements.
Notice
If the patient only requires an ICP catheter, the temperature value on the
touch screen will appear blank (with two dashed lines).
26 • Chapter 3 • Setting Up System for Clinical Use
Connecting the Integra® Camino® Flex Catheters
Flex Extension Cable
Black Barrel
Connect to
Monitor
Connecting the Integra Camino Flex Catheter to the Integra Camino ICP Monitor
requires the Integra Camino Flex Extension Cable (REF # FLEXEXT).
Warning
To prevent possible patient injury resulting from incorrect ICP
measurements, always perform each of the following steps before
implanting a new Integra Camino Flex Catheter into the patient. In
particular, always leave the Flex Catheter in the air until the monitor
successfully completes the initialization (i.e. autozero) process.
To c on ne ct:
1. On the monitor’s right side, attach the end of the Flex Extension Cable closest
the black barrel to the port labeled
ICP (FLEXEXT).
2. Remove the Flex Catheter from its packaging. For complete instructions on
using this catheter, see the directions for use supplied in the Flex Catheter
packaging.
Chapter 3 • Setting Up System for Clinical Use • 27
3. When operating in the sterile field, attach the male end of the Flex Catheter
Female end
Male end of
Flex Catheter
Flex Extension Cable
ICP value on
Main panel
connector to the female end of the Flex Extension Cable connector (note the
alignment arrows on each connector):
4. When attached correctly, the Integra Camino ICP Monitor will display
"
Catheter is initializing" in the status bar.
5. Verify the following message appears to ensure that the initialization (i.e.
autozero) process has successfully completed:
28 • Chapter 3 • Setting Up System for Clinical Use
If the monitor reports a "Catheter failure" message in the status bar, replace
with a new catheter.
6. Press Accept; the ICP value, which may be non-zero, will appear on the Main
panel.
Notice
Note the ICP value displayed may be non-zero; this value reflects the
hydration offset of the catheter while it is still in air. This value will
self-adjust automatically when the catheter is implanted into the patient.
7. Prior to implanting the catheter into the patient, disconnect the Flex Catheter
from the Flex Extension Cable. This will prevent the monitor from measuring
ICP while the catheter is traveling, which may trigger alarms caused by the
sudden changes in pressure (see the following section for details).
8. After implanting the catheter into the patient, reconnect the Flex Catheter to
the Flex Extension Cable.
Warning
Once the Flex Catheter has been initialized (i.e. autozeroed) by the
Integra Camino ICP Monitor, do not transfer this initialized catheter to
any other monitor. Transferring an initialized catheter to a different
monitor may result in inaccurate ICP measurements.
Warning
Once the Flex Catheter has been initialized (i.e. autozeroed) by the
Integra Camino ICP Monitor, do not replace the Flex Extension Cable
being used for patient measurement. Replacing the Flex Extension
Cable with a different cable after the Flex Catheter has already been
initialized may result in inaccurate ICP measurements.
Transient Pressure Changes During Implantation or Removal of
Catheter May Trigger Alarm
If the Flex Catheter is connected to the monitor during the implantation or
removal of a catheter, a sudden change in ICP measurements may occur while the
catheter is traveling. Depending on the level of pressure change, the ICP
measurements may temporarily exceed the ICP accuracy range for the monitor
(between -10 mmHg and 125 mmHg) and trigger one of the following alarms:
•"ICP out of accuracy range"; or
• "ICP catheter failure"
If either of these alarms is triggered during implantation or removal of the
catheter, disconnect and re-connect the Flex Catheter to the Flex Extension Cable.
Chapter 3 • Setting Up System for Clinical Use • 29
Connecting to a Patient Bedside Monitor (if applicable)
The Integra Camino ICP Monitor provides outputs for connecting to a patient
bedside monitor. This connection requires two Integra cables:
CablesREF #Description
PMIO PMIOMPM
Monitor Adapter CableICPXX and ICTXX
To determine which monitor adapter cables are required for your particular patient
bedside monitor, contact Integra (see page 89).
Warning
Read the user’s manual from the patient bedside monitor’s manufacturer
before connecting the Integra Camino ICP Monitor to a patient bedside
monitor.
Procedures for Synchronizing the Two Monitors
Perform the following steps to verify that both the Integra Camino ICP Monitor
and the patient bedside monitor report the same ICP values, ±1 mmHg or 1 % of
monitor reading (whichever is greater).1 Note that these steps should be
performed in each of the following situations:
•When first connecting the Integra Camino ICP Monitor to a patient bedside
monitor.
Main cable for connecting Integra
monitor to patient bedside monitor.
Adapter cables that fit between the
PMIO cable and the vendor-specific
patient bedside monitor.
•If the Integra Camino ICP Monitor becomes disconnected from a patient bedside
monitor (e.g. due to patient transport to CT or OR), and then needs to be
re-connected.
•If, during use, the reported ICP values between the Integra Camino ICP Monitor
and the patient bedside monitor differ by more than ± 1 mmHg or 1 % of monitor
reading (whichever is greater).
Step 1: Connect the Two Monitors
Notice
To cancel the synchronization process during any of the following steps,
press
Done.
Connect the Integra Camino ICP Monitor to the patient bedside monitor:
A. On the rear panel of the Integra monitor, attach the connector end of the
PMIO cable to the port labeled
1.If the ICP value is above 100 mmHg, the difference between both monitors can
differ by 2 mmHg.
30 • Chapter 3 • Setting Up System for Clinical Use
PMIO.
• Align the red dot on the PMIO cable connector with the red triangle on
PMIO cable
AC power adapter cable
the monitor’s port and push firmly.
• Verify the cable is completely connected to the monitor before
proceeding.
B. On the other end of the PMIO cable, attach the monitor adapter cable for
pressure (and temperature if applicable).
C. Attach the PMIO cable and adapter cable to the temperature/pressure
modules on the patient bedside monitor.
Notice
The Integra Camino ICP Monitor does not measure Cerebral Perfusion
Pressure (CPP) ranges. As a result, the CPP connector on the PMIO cable
will not be used.
Step 2: Press the "Synchronize to Monitor" Button
On the Integra Camino ICP Monitor’s touch screen, press
Synchronize to Monitor button; the Integra monitor
the
will transmit an ICP value of
0 mmHg for up to one minute.
A countdown will appear to display the amount time left to
zero the patient bedside monitor.
Step 3: Zero the Patient Bedside Monitor to the Integra Monitor
On the patient bedside monitor, follow the manufacturer’s instructions to zero the
monitor.
Step 4: Check Additional Values to Confirm Synchronization
Notice
The following synchronization checks at 25 mmHg, 50 mmHg, and
100 mmHg are not required, but are recommended by Integra.
On the Integra ICP Camino’s touch screen, press the Check Additional Values
button to transmit additional ICP values to the patient bedside monitor for scaling
verification:
•25 mmHg
•50 mmHg
•100 mmHg
Chapter 3 • Setting Up System for Clinical Use • 31
Note that pressing Check Additional Values each time advances to the next ICP
value. If you are not checking additional values, proceed to the final step,
"Complete the Synchronization Process" on page 32.
A. On the Integra Camino ICP Monitor’s touch screen, press the Check
Additional Values
button to transmit an ICP value at 25 mmHg for up to
one minute to the patient bedside monitor.
A countdown will appear to display the amount time left to confirm that the
value is also displayed on the patient bedside monitor.
B. Once the ICP value on the patient bedside monitor stabilizes, verify that the
patient bedside monitor reads the same value as the Integra Camino ICP
Monitor:
25 mmHg, ± 1 mmHg
C. Repeat steps A to B in this section for each additional value that you want to
check.
Step 5: Complete the Synchronization Process
After completing the synchronization process, press Done. Once the ICP value on
the patient bedside monitor stabilizes, verify that the patient bedside monitor
reads the same ICP value as the Integra Camino ICP Monitor (±1 mmH or 1 % of
monitor reading, whichever is greater).
During patient treatment, Integra recommends comparing the ICP values between
the two monitors while the clinical staff is:
•Recording ICP
•Tre at in g IC P
•Per the hospital’s unit policy.
Warning
If the Integra Camino ICP Monitor loses power and shuts down while it is
connected to a patient bedside monitor, do not use the ICP values on the
patient bedside monitor for patient measurements; the ICP values on the
patient bedside monitor will be invalid.
32 • Chapter 3 • Setting Up System for Clinical Use
Notice
Cable strap
AC power adapter cord
After completing the synchronization process, always use the Integra
Camino ICP Monitor’s measurements over the patient bedside monitor’s
measurements. If drift occurs between the two monitors, use the Integra
Camino ICP Monitor’s ICP values for the patient’s measurements and
repeat the procedures for synchronizing the two monitors (see
page 30).
About Pressure and Temperature Measurements on the Patient Bedside Monitor
The Integra Camino ICP Monitor is designed to measure tissue pressure between
the range of -10 mmHg to 125 mmHg, and tissue temperature between the range of
°C and 42 °C. If the pressure or temperature value is outside these specific ranges
30
on the Integra Camino ICP Monitor while it is connected to the patient bedside
monitor, the Integra monitor will activate a low priority alarm for either an ICP or
temperature accuracy range alarm condition (see
alarm condition, the Integra monitor will also transmit either a pressure value of
mmHg or a temperature value of 15 °C to the patient bedside monitor to indicate
-15
that the Integra monitor is unable to make accurate measurements.
page 54). Depending on the type of
Storing the System
If the Integra Camino ICP Monitor will not be used for several months or longer,
remove the battery from the monitor prior to storing the monitor. During storage,
keep the monitor in a dry location that meets the following environmental
conditions:
•Temperature: -20 °C to 50 °C
•Humidity: 25% to 80% RH, non-condensing
Also ensure the storage location is safe from any liquids that may drip inside the
monitor and damage its internal components.
Using the Cable Strap to Wrap the AC Power Adapter
The Integra Camino ICP Monitor also includes an adjustable strap for securing the
AC power adapter cord during patient transport or storage. To use this cable strap,
wrap the rubber strap around the bundled adapter cord and secure it by inserting
the round plastic fastener through the desired hole.
Chapter 3 • Setting Up System for Clinical Use • 33
This page is intentionally left blank.
34 • Chapter 3 • Setting Up System for Clinical Use
CHAPTER 4 MONITORINGTHE PATIENT’S ICP AND
T
EMPERATURE
About the Touch Screen ............................................................................... 35
About the Alarms........................................................................................... 37
Monitoring the Patient’s ICP and Temperature........................................... 37
Setting the High ICP Alarm Limit ................................................................. 41
Customizing the User Settings..................................................................... 44
About the Touch Screen
The Integra® Camino® ICP Monitor includes a touch screen for evaluating and
controlling parameters for monitoring the patient’s ICP and temperature. The
touch screen provides the following tabs for accessing and activating different
parameter:
TabDescription
MainProvides tools for evaluating the patient’s Mean ICP (mmHg) and
temperature (°C). For details, see page 37.
TrendProvides tools for monitoring for up to a 5-day history of the patient’s
intracranial pressure values. For details, see page 39.
AlarmProvides tools for specifying a high ICP alarm limit the monitor will
tolerate before sounding an alarm. For details, see page 37.
SettingsProvides tools for specifying languages, date and time. This panel also
contains settings for specifying the animated waveform graph that
appears on the
Chapter 4 • Monitoring the Patient’s ICP and Temperature • 35
Main panel. For details, see page 44.
About the Synchronize to Monitor Button
Status bar
If you are planning to display patient data from the Integra
Camino ICP Monitor to a patient bedside monitor, press the
Synchronize to Monitor button to initiate the process of
synchronizing the two monitors with each other. This process ensures that both
monitors are displaying identical ICP values for the patient. Note that this feature
provides synchronization checks at 0 mmHg, 25 mmHg, 50 mmHg, and 100
mmHg. Pressing the
the next measurement. For instructions on synchronizing the monitors, see page
30.
Reviewing the Status Bar
When using the Integra Camino ICP Monitor, the status bar on the bottom of the
touch screen will display messages to indicate the current states of alarms,
catheters, and AC/battery power.
Check Additional Values button each time will advance to
Verifying Status of Battery and AC Power
The Integra Camino ICP Monitor may be powered by either AC power or battery
charge. To determine the status of power for the monitor, view the battery and
plug symbols that appear on the bottom right of the touch screen
SymbolStatus
Indicates AC power is being used.
Indicates AC power is not being used/available.
Indicates battery power is being used to power the monitor. Each green
bar represents the level of remaining battery charge (see following
section).
Indicates AC power is being used to charge the battery.
Indicates that either a battery is not connected or a battery is not
functioning properly.
For instructions on powering the monitor, see page 20.
:
36 • Chapter 4 • Monitoring the Patient’s ICP and Temperature
Verifying Amount of Battery Charge Available
1
2
3
4
5
The Integra Camino ICP Monitor displays the following symbols on the status bar
to indicate battery charge levels
SymbolColorAvailable Charge
Green75% to 100%
Green50% to 75%
Green25% to 50%
GreenLess than 25%
Flashing YellowLess than 5 minutes
If the level of battery charge falls to 15 minutes or less, the monitor will display an
error message on the status bar. If the level of battery charge falls to 5 minutes or
less, the monitor will activate a technical alarm. For more information, see page
52.
About the Alarms
The Integra Camino ICP Monitor activates one physiological alarm for exceeding
the high ICP alarm limit and several technical alarms for indicating
equipment-related problems. For instructions on setting the high ICP alarm limit,
see page 41. For detailed information on responding to each of these alarm types,
see Chapter 5.
Monitoring the Patient’s ICP and Temperature
Press the Main tab to view the patient’s Mean ICP and temperature values as
numerical data. The ICP may also be viewed as an animated waveform.
Chapter 4 • Monitoring the Patient’s ICP and Temperature • 37
The information on this screen includes:
ParameterDescription
1Displays a graph showing the ICP in mmHg versus time as an
animated waveform (solid or line). To specify the waveform type,
press the
Settings tab. For instructions, see page 45.
2Displays the Mean ICP in mmHg as numerical values.
3Displays the high ICP alarm limit for the patient. If the patient’s Mean
ICP exceeds this limit for more than 5 seconds, an alarm will sound.
To specify this limit, see page 42.
4Displays tissue temperature in degrees Celsius (°C).
5Specifies the scales of pressure ranges for the ICP waveform.
About the Numerical Display of Mean ICP Values
The numerical ICP value that is displayed on the touch screen is the patient’s Mean
ICP value, which is a time-averaged representation of the continuously measured
ICP. This displayed value is rounded to the nearest 1 mmHg.
Scaling the ICP Waveform Ranges
If the ICP waveforms on the Main panel require resizing for clarity,
press the
Scale button to specify different waveform ranges. The
ranges include:
Ranges (mmHg)Increments (mmHg)
0 to 305
0 to 5010
0 to 10010
0 to 15025
-10 to 205
Pressing the
Scale button each time advances to the next range. Note that the
purpose of the -10 to 20 mmHg range is to provide appropriate scaling when
zeroing the Fiber Optic Catheter (see page 24 instructions on zeroing the Fiber
Optic Catheter).
38 • Chapter 4 • Monitoring the Patient’s ICP and Temperature
Monitoring Trend Data
1
2
3
4
5
8
6
7
Press the Trend tab to view a graph of recorded Mean ICP trend data in mmHg at
one-minute intervals. The Integra Camino ICP Monitor stores up to five days of
trend data that may be viewed in specific ranges of time. This data may also be
extracted from the monitor to other media types by either USB transfer or by
digital streaming. For instructions on extracting data, see page 57.
The information on this panel includes:
ItemsDescription
1
Extract Data
Copies the patient’s trend data to a USB drive that has been connected
to the USB port on the rear of the monitor.
2
Data Streaming Enabled
Indicates that the feature for streaming patient trend data digitally to
another media type has been activated. This feature requires the use of
the USB-to-RS232 adapter cable (see Chapter 6).
3
ICP Scale
Specifies the range of the ICP values displayed on the Trend graph.
4
Time Scale
Specifies the range of time displayed on the Trend graph.
5Displays the patient’s ICP trend data.
6Displays Mean ICP in mmHg as a numerical value.
7Displays the high ICP alarm limit. If the patient’s Mean ICP value
exceeds this limit for more than 5 seconds, an alarm will sound. To
specify this limit, see page 42.
8Displays tissue temperature in degrees Celcius (°C).
Chapter 4 • Monitoring the Patient’s ICP and Temperature • 39
Scaling Time Ranges for Trend Data
Press the Time Scale button to specify different time ranges to view
the trend data. The time ranges include:
• 3 hours
•12 hours
•24 hours
•48 hours
• 120 hours
Pressing the Time Scale button each time advances to the next time range.
Scaling the ICP Ranges for Trend Data
If the ICP waveforms on the Trend panel require resizing for optimal
viewing, press the
ICP Scale button to specify different waveform
ranges. The ranges include:
Ranges (mmHg)Increments (mmHg)
0 to 305
0 to 5010
0 to 10010
0 to 15025
-10 to 205
Pressing the
Conditions that Reset Trend Data
Depending on the catheter type, some conditions may result in the loss of trend
data during the recording process. For details, see page 59.
ICP Scale button each time advances to the next waveform range.
40 • Chapter 4 • Monitoring the Patient’s ICP and Temperature
Setting the High ICP Alarm Limit
6
5
2
1
4
3
Press the Alarm tab to specify settings for controlling the physiological alarm.
This panel includes arrows for specifying the high ICP alarm limit value, turning
the alarm on and off, and restoring the high ICP alarm limit from a customized
setting to the factory default setting (20 mmHg).
The information on this panel includes:
ItemDescription
Alarm On
1
Activates the physiological alarm settings. If this option is selected, the
monitor will sound an alarm if the patient’s displayed Mean ICP exceeds the
specified high ICP alarm limit for more than 5 seconds. By default, this
option is selected and set to 20 mmHg.
Alarm Off
2
Deactivates the physiological alarm settings. If this option is selected, the
monitor will disable the high ICP alarm and prevent a high ICP alarm from
sounding. In this mode, the monitor will display
ICP Alarm Off in the
status bar. Note that technical alarms cannot be disabled with this option.
High ICP Alarm Limit
3
Specifies the maximum ICP value that the monitor will tolerate before
activating the alarm. To specify this limit, press the arrows to a value
between -10 mmHg and +125 mmHg in 1 mmHg increments.
Restore Defaults
4
Restores customized settings for the high ICP alarm limit to the factory
default settings (20 mmHg). Pressing this button also activates the
button.
On
ICP
5
Displays the Mean ICP in mmHg as numerical values.
Alarm
6
Temperature
Displays tissue temperature in degrees Celsius (°C).
Chapter 4 • Monitoring the Patient’s ICP and Temperature • 41
About the High ICP Alarm Limit Calculation
high ICP
alarm limit
high ICP
alarm limit
The Integra Camino ICP Monitor bases the high alarm limit on the patient’s Mean
ICP value, which is represented by the numerical ICP value that appears on the
touch screen (see page 38). The high ICP alarm is triggered if the patient’s
displayed Mean ICP value continuously remains above the high ICP alarm limit for
a period of 5 seconds.
Specifying the High ICP Alarm Limit
Warning
Always verify the high ICP alarm limit is set appropriately for each patient
prior to treatment.
1. On the touch screen, press the Alarm tab.
2. Press the Alarm On button; this will activate the alarm settings.
3. In the High ICP Alarm Limit field, use the arrows to specify the high ICP alarm
limit value for the patient. Note that the range limits are between -10 to +125
mmHg in 1 mmHg increments.
4. Press Accept. If the patient’s Mean ICP value exceeds this specified limit for more
than 5 seconds, the monitor will activate the physiological alarm. Note that
pressing
this alarm, see
Touch Screen Labels Indicating High ICP Alarm Limit
Cancel will revert all changes back to the previous values. To respond to
page 51.
Note that the alarm limit specified in step 3 appears in the Main and Trend screens
as the following indicator labels:
•A horizontal line in the graph corresponding to the specified high ICP alarm limit
•A numerical value next to the alarm limit symbol
42 • Chapter 4 • Monitoring the Patient’s ICP and Temperature
Disabling the High ICP Alarm
To turn off the high ICP alarm and prevent the Integra Camino
ICP Monitor from sounding a physiological alarm, press
Off
and then Accept If this option is selected, the monitor will
display
ICP Alarm Off in the status bar. The monitor will also remove the
corresponding high ICP alarm limit indicator labels from the
screens. To turn on the high ICP alarm, press
Warning
Selecting the Alarm Off feature will disable the physiological alarm
indefinitely. Use caution if this feature is selected.
Silencing the High ICP Alarm Temporarily
To silence the physiological alarm temporarily, press the active
alarm symbol on the status bar. This will silence the alarm for
either 3 minutes or until the patient’s Mean ICP value falls within
the specified limit. In this silenced state, the status bar will display
and change the alarm symbol from active to inactive (see page 48). If the patient’s
Mean ICP value does not fall to within the specified alarm limit, the alarm tone will
resume.
Alarm
Main and Trend
Alarm On and then Accept.
Audio Paused
Restoring Default High ICP Alarm Limit Values
After specifying the high ICP alarm limit value on the Alarm tab,
the Integra Camino ICP Monitor saves this current value into
memory . If the monitor is turned off for any duration of time
and then turned back on, the last saved high ICP alarm limit value will be restored.
To restore the high ICP alarm limit from a customized setting to the factory default
settings (20 mmHg), press
Restore Defaults button will also activate the Alarm On mode automatically.
Restore Defaults and then Accept. Note that pressing
Chapter 4 • Monitoring the Patient’s ICP and Temperature • 43
Customizing the User Settings
1
2
3
4
Press the Settings tab to control the appearance of specific graphical and textual
information that appears on the touch screen. This panel also includes a button
for determining system information regarding software and firmware.
The information on this panel includes:
NumberDescription
Set Waveform
1
Specifies the type of ICP waveform (line or fill) that is displayed on the
Main panel.
Set Time and Date
2
Specifies the time and date that is displayed on the monitor. The
purpose of the time/date display is for the
Extract Data feature (see
page 57).
3
4
Set Language
Specifies the language that is displayed on the software interface.
System Information
Displays information regarding the system’s software and firmware.
Service Mode
Provides diagnostic information for errors. This mode appears on the
System Information panel. This mode is password protected and is only
accessible by Integra staff.
44 • Chapter 4 • Monitoring the Patient’s ICP and Temperature
Specifying Waveform Type
Waveform set to Fill mode.
The Integra Camino ICP Monitor provides options for displaying the ICP waveform
as either a line or filled graph.
1. On the the touch screen, press the Settings tab.
2. Press Set Waveform.
3. Specify the waveform type by selecting either Line or Fill.
4. Press Accept.
The selected waveform type will appear on the Main panel.
Specifying Time and Date
The Integra Camino ICP Monitor provides tools for setting the current time and
date that appears on the touch screen.
To set this information:
1. On the touch screen, press the Settings tab.
2. Press the Set Time and Date button.
3. On the displayed panel, press the desired field (hour, minutes, date, month, or
year) and use the arrows to specify the appropriate setting. Note that you may
adjust each of these settings prior to accepting them in the following step.
4. Press Accept; the Integra Camino ICP Monitor will display the selected time/date
on the touch screen.
Chapter 4 • Monitoring the Patient’s ICP and Temperature • 45
Specifying Languages
The Integra Camino ICP Monitor provides options for displaying the onscreen text
indifferent languages:
The list of supported languages include:
•English •Spanish•Danish • Chinese (Simplified)
•French•Dutch•Polish• Portuguese (Brazilian)
•Italian• Japanese•Korean
•German
• Russian
To change languages from any screen:
1. On the touch screen, press the Settings tab.
2. On the left side of the panel, press Set Language.
3. In the displayed Language: menu, use the arrows to select the desired language.
4. Press Accept; the Integra Camino ICP Monitor will display all onscreen text to the
selected language.
Determining System Information
If you experience technical problems with your Integra Camino ICP
Monitor, you may need to provide Integra with information regarding
the system’s software and firmware versions. To determine this
information, press the information symbol.
About Extract Log and Service Mode Buttons
The Extract Log button only appears on the System
Information
panel if a USB drive is connected to the Integra
Camino ICP Monitor. Note that when attaching a USB drive, this
button may take several seconds to appear. Pressing the
Extract Log button will
copy system log information from the monitor to the USB drive. During a
troubleshooting session, the Integra technical staff may request this log file.
The
Service Mode button provides diagnostic information for
Integra staff to troubleshooting errors. This button is only
intended for Integra staff.
46 • Chapter 4 • Monitoring the Patient’s ICP and Temperature
CHAPTER 5 RESPONDINGTO PHYSIOLOGICALAND
• Low battery level (15 minutes or less)
• Multiple catheters are connected
• Catheter initializing
• No catheter is connected
• Catheter initialization failure
• Cannot write to log file
T
ECHNICAL ALARMS
About the Two Alarm Types ......................................................................... 47
Understanding the Alarm Symbols.............................................................. 48
About the Technical Messages .................................................................... 47
How the Monitor Prioritizes the Alarms....................................................... 48
Responding to the Physiological Alarm (ICP above alarm limit).............. 51
Responding to Technical Alarms................................................................. 51
About the Two Alarm Types
The Integra® Camino® ICP Monitor activates two types of audio/visual alarms to
indicate problems that require immediate attention:
Alarm TypeDescription
PhysiologicalAlerts the nurse/physician that the patient’s Mean ICP value being monitored has exceeded the
specified limit for more than 5 seconds.
TechnicalAlerts the nurse/physician that there is a problem with the monitor’s battery, electrical
components, software version, or catheter connections. These technical alarms include:
• Irreversible System Failures
• ICP catheter failure
• Temperature sensor failure
• Low battery level (5 minutes or less)
• Monitor overheating
Warning
Do not block the alarm speaker with any materials that might muffle the
alarm sound.
• Cooling fan failure
• Mean ICP is out of accuracy range
• Temperature range is out of accuracy range
• Battery failure
About the Technical Messages
In addition to the technical alarms, the Integra Camino ICP Monitor also displays
technical messages on the status bar to indicate problems with:
For information on responding to these messages, see page 67.
Chapter 5 • Responding to Physiological and Technical Alarms • 47
Understanding the Alarm Symbols
On the status bar, the Integra Camino ICP Monitor displays different variations of
the alarm bell symbol to indicate three different alarm states: Active, Audio
Paused, and Inactive:
SymbolStateDescription
(yellow borders)
Audio PausedIndicates an active physiological or technical alarm that
(crossed-out yellow borders)
InactiveIndicates no physiological or technical alarms are active.
ActiveIndicates an active physiological or technical alarm. In
this state, the status bar displays the description of the
error that activated the alarm.
has been paused temporarily by pressing the active
alarm symbol. In this state, the status bar displays
Audio Paused. After 3 minutes of silence, the Audio
Paused
message disappears and the alarm will
re-activate automatically.
(greyed out)
How the Monitor Prioritizes the Alarms
The Integra Camino ICP Monitor assigns each error condition that may activate a
physiological or technical alarm type with a range of clinical priorities between
medium and low. These priorities are determined by two factors:
•The potential harm that may occur to the patient
•How quickly the onset of potential harm may occur if the user fails to respond
to the cause of the alarm.
Clinical Priority
MediumReversible injuryPrompt
LowReversible injuryDelayed
Potential Harm If
Alarm Is Ignored
Timeframe for
Potential Harm to Occur
48 • Chapter 5 • Responding to Physiological and Technical Alarms
Audio and Visual Indicators for Medium and Low Priority Alarms
Depending on the clinical priority of the error condition that is causing the alarm,
the Integra Camino ICP Monitor activates audio and visual indicators to alert the
user of the condition
of three pulses, with a pulse
separation of 180 milliseconds and a
burst separation of 3 seconds.
• The volume of each alarm pulse
measures a sound pressure of 70 dB.
Low• Sounds an alarm containing a single
pulse, with a burst separation of
16seconds .
• The volume of the single alarm pulse
measures a sound pressure of 69 dB.
For specific information on which audio/visual indicators occur for each medium
priority and low priority alarm type, see
above alarm limit) on page 51, and Responding to Technical Alarms on page 51.
• A message will appear on the status bar that
identifies the error
• If a system failure occurs, a standalone message
window will appear that identifies an error code and
description of the failure
• The borders surrounding specific symbols (e.g,
alarm, battery) on the touch screen will flash yellow
• For some error conditions, the patient values will
flash yellow
• A message will appear on the status bar that
identifies the error.
• The patient values will appear yellow (not flashing)
• For battery-related errors, the battery symbol
indicates the battery has no remaining charge.
Responding to the Physiological Alarm (ICP
Priorities of Physiological and Technical Alarms
The Integra Camino ICP Monitor includes one medium priority physiological alarm
that gets activated if a patient’s Mean ICP value exceeds the user specified limit for
more than 5 seconds; there are no other physiological alarms. If two or more error
conditions occur simultaneously that can activate a physiological alarm and a
technical alarm, the physiological alarm will always take precedence (unless the
technical alarm is caused by an irreversible system failure, which results in the
monitor being unusable; see page 51 for details). The remaining technical alarms
will only become announced after the physiological alarm is resolved.
How the Monitor Prioritizes Between Two Technical Alarms
If two or more error conditions of the same clinical priority occur simultaneously
that can activate different technical alarms, the Integra Camino ICP Monitor
prioritizes which alarm it activates based on how the alarm condition affects the
safety of the patient or stability of the equipment. Note that a technical alarm
with a higher priority will always be activated over a technical alarm with a lower
priority; a lower priority alarm will not be announced until the higher priority
alarm is resolved.
Chapter 5 • Responding to Physiological and Technical Alarms • 49
List of Priorities for Each Alarm
Notice
If any of the software, firmware, or electrical components inside the
monitor suffers a serious malfunction during a physiological alarm, the
Integra Camino ICP Monitor will activate an irreversible technical alarm
that will override the physiological alarm and indicate a system failure. If
this irreversible technical alarm is activated, the monitor will no longer
report patient values and the user will need to contact Integra for service.
For more information on this error condition, see page 51.
The following table lists the priorities of each alarm condition:
Alarm PriorityType of Error ConditionType of Clinical Priority
1Irreversible System Failures (see page 51)Medium
2High ICP alarm (see page 51)Medium
3ICP catheter failure (see page 52)Medium
4Temperature sensor failure (see page 52)Medium
5Low battery level (5 minutes or less) (see page 52)Medium
6Monitor overheating (see page 53)Medium
7Cooling fan failure (see page 53)Low
8Mean ICP is out of accuracy range (see page 54)Low
9Temperature range is out of accuracy range (see page 54)Low
10Battery failure (see page 55)Low
11Low Battery level (15 minutes or less)N/A
12Multiple catheters are connectedN/A
13Catheter initializingN/A
14No catheter is connectedN/A
15Catheter initialization failureN/A
16Cannot write to log fileN/A
50 • Chapter 5 • Responding to Physiological and Technical Alarms
Responding to the Physiological Alarm (ICP above alarm limit)
If the patient’s Mean ICP value exceeds the specified limit for more than 5 seconds,
the Integra Camino ICP Monitor will activate a medium priority physiological
alarm that will:
•Sound an alarm containing a burst of three pulses
•Display the following error message on the status bar: "ICP above alarm limit"
•Flash the alarm symbol on status bar
•Flash the Mean ICP value in yellow on the Main, Trend, and Alarm screens
To silence the alarm temporarily, press the flashing alarm symbol on the touch
screen. This will silence the alarm for either 3 minutes or until the patient’s Mean
ICP value falls within the specified alarm limit for at least one measurement. After
3 minutes, the alarm will reactivate. To turn off the alarm completely, press the
Alarm tab and select Alarm Off and Accept. For instructions on specifying the
high ICP alarm limit, see page 41.
Warning
Selecting the Alarm Off feature will disable the high ICP alarm limit
indefinitely. Use caution if this feature is selected.
Responding to Technical Alarms
Use the following guidelines to respond to each technical alarm.
Responding to Irreversible System Failure Alarms
If any of the software, firmware, or electrical components inside the monitor
suffers a serious malfunction during use, the Integra Camino ICP Monitor will stop
reporting measurement values on the touch screen and activate a medium priority
technical alarm. During this alarm mode, the monitor will:
•Sound an alarm containing a burst of three pulses
•Display a standalone message window that specifies an error code and a
description of the applicable system failure, which include the following:
•Sensor board failure
•Power board failure
•General software programming error
•Incompatible firmware error
For a complete list of error codes, see page 73. Please note that each of these
alarm conditions are irreversible. If any of these alarms occur, tend to the patient’s
need, note the specific error code, and contact Integra for service.
Chapter 5 • Responding to Physiological and Technical Alarms • 51
Responding to ICP Catheter Failure Alarm
If the ICP catheter fails during initialization or use, the monitor will activate a medium
priority alarm that will:
•Sound an alarm containing a burst of three pulses
•Display the following error message on status bar: "Catheter failure"
•Remove the ICP and temperature values from the screen
To resolve this problem, disconnect the catheter. To resume monitoring ICP, the
entire catheter must be replaced.
Silencing the Catheter Failure Alarm Temporarily
Press the flashing alarm symbol on the touch screen. This will silence the alarm for
3 minutes. If the catheter is not disconnected within 3 minutes,
reactivate.
Responding to Temperature Sensor Failure Alarm
If the temperature catheter fails during use, the monitor will activate a medium
priority alarm that will:
•Sound an alarm containing a burst of three pulses
•Display the following error message on status bar: "Temperature sensor
failure"
the alarm will
•Remove the ICP and temperature values from the screen
To resolve this problem, disconnect the temperature cable. To resume monitoring
temperature, the entire catheter must be replaced.
Silencing the Catheter Failure Alarm Temporarily
Press the flashing alarm symbol on the touch screen. This will silence the alarm for
3 minutes. If the catheter is not disconnected within 3 minutes,
reactivate.
Responding to Low Battery Alarm
The Integra Camino ICP Monitor includes a replaceable 14.4 V lithium ion battery
that supplies power to the monitor for at least 1.5 hours. The battery is intended to
power the monitor during patient transport only; use AC power whenever
possible.
15 Minutes or Less of Remaining Power
If the battery power level drops to 15 minutes or less of remaining charge, the
Integra Camino ICP Monitor will display the following message on status bar:
Battery is low" If this occurs, plug in the monitor to an AC power outlet. This
"
warning message will not disappear until the monitor is connected to AC power.
the alarm will
52 • Chapter 5 • Responding to Physiological and Technical Alarms
5 Minutes or Less of Remaining Power
If the battery power level drops to 5 minutes or less of remaining charge, the
Integra Camino ICP Monitor will activate a medium priority alarm that will:
•Sound an alarm containing a burst of three pulses
•Display the following message on the status bar: "Battery is low"
•Flash the battery symbol in yellow on status bar
If this occurs, plug in the monitor to an AC power outlet immediately. This alarm
will stop once the monitor is connected to AC power
Warning
Connect the monitor to an AC power supply immediately if the low
battery alarm is activated.
Silencing the Low Battery Alarm Temporarily
Press the flashing alarm symbol on the touch screen. This will silence the alarm for
3 minutes while you plug the monitor into an AC power outlet to charge the
battery. If the battery does not receive charge within 3 minutes, the alarm will
reactivate.
The Integra Camino ICP Monitor battery may be replaced by the hospital’s
biomedical engineer. For instructions, see page 86.
Responding to Monitor Overheating Alarm
If the temperature inside the Integra Camino ICP Monitor exceeds 80 °C, the
monitor will activate a medium priority alarm that will:
•Sound an alarm containing a burst of three pulses
•Display the following error message on status bar: "Monitor overheating,
check vent"
To resolve this problem, check the vents on the rear and bottom panels of monitor
to ensure nothing is blocking them. If something is blocking them, remove it
immediately. Once the temperature inside the monitor falls below 80
will stop and the message will disappear.
Silencing the Monitor Overheating Alarm Temporarily
Press the flashing alarm symbol on the touch screen. This will silence the alarm for
3 minutes while you clean the vent from any obstructions. If the temperature
inside the monitor does not fall below 80
reactivate.
Responding to Cooling Fan Failure Alarm
The Integra Camino ICP Monitor contains an internal fan that circulates air to cool
the internal electronic components. If the monitor detects the cooling fan has
stopped running,
the monitor will activate a low priority technical alarm that will:
°C, this alarm
°C within 3 minutes, the alarm will
•Sound an alarm containing a burst of one pulse every 16 seconds
•Display the following error message on status bar: "Cooling fan failure"
Chapter 5 • Responding to Physiological and Technical Alarms • 53
If this problem occurs, turn off the monitor to prevent possible overheating. After
turning off the monitor, contact Integra for service.
Silencing the Cooling Fan Alarm Temporarily
Press the yellow alarm symbol on the touch screen. This will silence the alarm for 3
minutes while you turn off the monitor’s power to prevent possible overheating. If
the monitor remains on for 3 minutes while there is still a cooling fan error
condition, the alarm will reactivate.
Responding to Accuracy Range Alarm
The Integra Camino ICP Monitor is designed to measure ICP and tissue
temperature between specific clinical ranges:
MeasurementsRange
ICPBetween -10 to +125 mmHg that includes the following levels of
accuracy depending on catheter type:
Fiber Optic Catheter:
• -10 to 50 mmHg (±2 mmHg)
• 51 to 125 mmHg (±6%)
Flex Catheter:
• -10 to 10 mmHg (±2 mmHg)
• 11 to 33 mmHg (±3 mmHg)
• 34 to 125 mmHg: ± (6% + 1 mmHg)
Te m p e r a t u r e
Between 30 °C to 42 °C (±0.3 °C)
If the measured ICP or temperature falls outside these ranges, the monitor will
activate a low priority alarm that will:
•Sound an alarm containing a burst of one single pulse every 16 seconds
•Display one of the following messages on the status bar:
"ICP is out of accuracy range" or "Temperature is out of accuracy range"
•Display the Mean ICP or temperature value in yellow on the Main, Trend, and
Alarm screens
To resolve this problem, the Mean ICP or temperature must fall within the rated
accuracy ranges listed in the above table.
Silencing the ICP and Temperature Accuracy Range Alarm Temporarily
Press the yellow alarm symbol on the touch screen. This will silence the alarm for 3
minutes while the Mean ICP or temperature values fall back within the rated
accuracy ranges displayed in the above table. If the Mean ICP or temperature
values remain outside the rated accuracy ranges for the 3 minutes, the alarm will
reactivate.
54 • Chapter 5 • Responding to Physiological and Technical Alarms
Responding to Battery Failure Alarm
If the battery control system inside the Integra ICP Camino Monitor fails during
the use of the monitor, the monitor will no longer be able to report the amount of
remaining charge that battery has. If this problem occurs, the monitor will activate
a low priority technical alarm that will:
•Sound an alarm containing a burst of one single pulse every 16 seconds
•Battery power symbol on the touch screen will appear crossed-out.
•Displays the following message on the status bar:
"Battery Failure"
If this battery failure alarm activates, do not run the monitor on battery power. If
necessary, plug the monitor into an AC outlet to maintain power. After patient
treatment, try the following:
•Turn off the monitor and then turn it back on.
•Replace the battery.
If this alarm condition persists, contact Integra for service.
Silencing the Battery Failure Alarm Temporarily
Press the yellow alarm symbol on the touch screen. This will silence the alarm for 3
minutes. If the monitor remains on for 3 minutes while there is still a reported
battery failure, the alarm will reactivate.
Chapter 5 • Responding to Physiological and Technical Alarms • 55
This page is intentionally left blank.
56 • Chapter 5 • Responding to Physiological and Technical Alarms
CHAPTER 6 EXTRACTING TREND DATAFOR REMOTE
2011-04-20-13:38, 18.3
Mean ICP (mmHg)
Year
Month
Day
Time
E
VALUATION
About Data Extraction................................................................................... 57
Extracting Data to USB Drive ............. .......................................................... 57
Extract Data via Digital Streaming ............................................................... 59
Conditions That Reset Trend Data During Recording ............................... 59
About Data Extraction
The Integra® Camino® ICP Monitor provides two options for extracting Mean ICP
trend data:
•External USB Drive
•Digital streaming via the USB-to-RS232 adapter cable
Both media types are connected to the monitor via a USB connection that is
located on the rear panel.
Notice
The USB connector port on the Integra Camino ICP Monitor is only
intended for connecting a USB drive and USB-to-RS232 adapter cable for
extracting ICP trend data. The USB port is not intended to be connected
to a network connection (e.g. USB bluetooth adapter, USB ethernet
adapter, etc).
Extracting Data to USB Drive
Use the Extract Data feature to export up to 5 days worth of Mean ICP trend data
from the monitor to an external USB drive. When using this feature, the Integra
Camino ICP Monitor copies the data to the USB drive as a comma separated value
(
.csv) file. This .csv file records the date, time, and Mean ICP values in one-minute
intervals and displays the information as ASCII text. The following text provides a
sample
To extract the data to a USB drive:
.csv trend data entry:
Chapter 6 • Extracting Trend Data for Remote Evaluation • 57
1. On the rear panel, attach the USB drive to the USB connector.
USB drive
AC power adapter
PMIO cable
2. On the touch screen, press the Trend tab and select Extract Data.
3. Verify the following message appears:
4. Press Accept, remove the USB drive, and view the .csv file in the desired output
device (e.g. laptop).
Notice
To ensure the complete transfer of trend data to the USB drive, do not
remove the USB drive until the message specified in step
confirming the extraction.
Note
In the .csv file, the timestamp for the data extraction is based on the
monitor’s date/time settings that are specified on the
countries where daylight savings are observed, the time must be
adjusted manually. For instructions, see
How the Monitor Stores Trend Data for Up to 5 Days
The Integra Camino ICP Monitor will only store the Mean ICP data from the most
recent 5 days. All stored trend data older than 5 days will be lost.
Warning
If monitoring is continued for more than 5 days, placement of a new
catheter under sterile conditions is recommended. Note that replacing a
catheter will reset the trend data. Please extract any data that you wish
to retain prior to replacing the catheter.
3 appears
Settings tab. In
page 45.
58 • Chapter 6 • Extracting Trend Data for Remote Evaluation
Extract Data via Digital Streaming
USB end of
USB-to-RS232 adapter cable
adapter cable
Use the streaming data feature to export Mean ICP trend data digitally from the
Integra Camino ICP Monitor to a remote device via a USB-to-RS232 adapter cable.
Note that the Integra supplied USB-to-RS232 adapter cable is required to use this
feature. When streaming data, the monitor reports the mean ICP value in mmHg of
the current ICP trend data in one-minute intervals. The USB-to-RS232 adapter
cable delivers the information as 9600 baud, 8 data bits, 1 stop bit, even parity
configurations.
To connect the USB-to-RS232 adapter cable to a computer, you need a female to
female, null modem RS232 cable. To extract the data via digital streaming:
1. On the rear panel, attach the USB end of the USB-to-RS232 adapter cable to the
USB connector.
2. On the touch screen, press Trend and verify the following message appears on
screen: "
3. Connect the other end of the USB-to-RS232 adapter cable to the external
data streaming enabled"
monitoring device.
Conditions That Reset Trend Data During Recording
Depending on the catheter type being used to record the patient’s Mean ICP trend
data, the following conditions will result in trend data being reset for that patient:
CatheterConditions
Integra® Camino® Fiber Optic
(110-4 Series)
• Monitor is turned off and then turn back on.
• Battery discharges completely, resulting in the monitoring
being turned off and then turned back on.
• The current ICP catheter is disconnected from the monitor and
then re-connected.
• The current temperature catheter is disconnected from the
monitor and then re-connected.
• The current catheter is replaced by a different catheter.
® Camino® Flex
Integra
•The current catheter is replaced by a different catheter.
Chapter 6 • Extracting Trend Data for Remote Evaluation • 59
Warning
Replacing a catheter will reset the trend data. Please extract any data
that you wish to retain prior to replacing the catheter.
60 • Chapter 6 • Extracting Trend Data for Remote Evaluation
CHAPTER 7 CLEANINGAND STERILIZINGTHE SYSTEM
Cleaning the System and Components....................................................... 61
Note that each of these components should be cleaned immediately after
contamination.
Warning
Only use the cleaning agents listed in this section for cleaning and
disinfecting the Integra Camino ICP Monitor system. Using solvents or
cleaning agents not listed in the cleaning guidelines may damage the
plastic exterior of the Integra Camino ICP Monitor.
Chapter 7 • Cleaning and Sterilizing the System • 61
Warning
Do not autoclave or immerse the Integra Camino ICP Monitor in liquid as
damage may occur. If the monitor is exposed to liquids, turn off the unit,
remove the AC power adapter, dry the unit thoroughly, and send to
biomed staff for evaluation before reapplying power.
Caution
The Fiber Optic Catheter Cable and Flex Extension Cable should not be
cleaned using automated cleaning methods.
ItemGuideline
Preparation for Cleaning
Limits on Reprocessing
Recommended Manual
Cleaning Method
Ensure that all the connections between the cables, catheter, cable strap, AC
power adapter, USB-to-RS232 serial adapter, and monitor have been removed
before cleaning the Integra Camino ICP Monitor.
• The Fiber Optic Catheter Cable may be sterilized thirty (30) times and cleaned/
disinfected one hundred (100) times. Do not use the cable after reaching these
reprocessing limits. For tracking purposes, use the reprocessing record sheet
(doc # 92903867) supplied with your Fiber Optic Catheter Cable.
• The Flex Extension Cable may be sterilized thirty (30) times. Do not use the cable
after reaching this sterilization limit. For tracking purposes, use the reprocessing
record sheet (doc # MAN0000005) supplied with your FLEX Extension Cable.
• There are no limits on reprocessing the Integra Camino ICP Monitor and cable
strap.
Using either 70% IPA with a lint free wipe or a Super Sani-Cloth® (or equivalent
solution), thoroughly wipe all surfaces at least three (3) times and then inspect
the surfaces for visible residues. If residues remain, use a new Super Sani-Cloth
(or equivalent solution) or lint free wipe soaked with 70% IPA and continue
wiping the surfaces until they are visibly free of residues. If the Flex Extension
Cable or the Fiber Optic Catheter Cable cannot be thoroughly cleaned, discard
and replace the cables.
Recommended
Disinfection Method
Using either 70% IPA with a lint free wipe OR Super Sani-Cloth (or equivalent
solution), thoroughly wipe all surfaces once. Ensure that the surfaces remain
visibly wet for a minimum of two (2) minutes and use additional wipe(s), if
needed, to ensure continuous two (2) minute wet contact time. Let the
Camino ICP Monitor and any of its system components
Inspection and Functional
Te s t i n g
Containment and
Transportation
62 • Chapter 7 • Cleaning and Sterilizing the System
After each reprocessing event, visually inspect the Integra Camino ICP
Monitor and any of its system components for any wear and tear.
It is recommended that the Integra Camino ICP Monitor and any of its system
components be cleaned and disinfected as soon as is reasonably practical
after use.
Integra
air dry prior to reuse.
Sterilizing the Integra® Camino® Fiber Optic Catheter Cable
The Fiber Optic Cable (REF # CAMCABL) is provided non-sterile.
Notice
The Fiber Optic Catheter Cable must be sterilized whenever it is to be
used in the sterile field. Prior to sterilization, clean this cable using the
guidelines listed on page 61.
Limits on Sterilization
The Fiber Optic Catheter Cable may be sterilized thirty (30) times and cleaned/
disinfected one hundred (100) times. Do not use the cable after reaching these
reprocessing limits. For tracking purposes, use the reprocessing record sheet (doc
# 92903867) supplied with your cable.
Sterilization Parameters
The Integra Camino Fiber Optic Catheter Cable supports EtO and Sterrad®
sterilization methods.
Notice
Prior to EtO sterilization, the cable should be individually packaged in a
self-sealing sterilization pouch with dimensions 10" x 15" (or equivalent
pouches designed for EtO gas sterilization).
Notice
Prior to Sterrad sterilization:
•The cable should be placed in a 12" x 15" Tyvek® Self-Seal Pouch (or
equivalent pouches designed for Sterrad sterilization);
- or -
•The cable should be placed in a 23" x 11" x 4" Sterrad® compatible
Aptimax® tray (or an equivalent instrument tray designed for Sterrad
sterilization) and wrapped in Kimguard® One-Step® Kc400 Sterilization
Wrap, then secured with Sterrad® compatible Sealsure® chemical
indicator tape (or equivalent sterilization wrap and indicator tape
designed for Sterrad sterilization).
Chapter 7 • Cleaning and Sterilizing the System • 63
Following Sterilization
EtO Sterilization Cycle Parameters
for Fiber Optic Cable
ParameterSpecification
Temperature55 ±3 ºC
Exposure Time 60 minutes
Humidity50-80% RH
Aeration at 55ºC> 8 hours
Concentration725 ±25 mg/L
Sterrad® 100S Cycle Parameters
for Fiber Optic Cable
StageCycle Duration
Pre-ConditioningPre-Plasma10 minutes
Injection I6 minutes
Phase 1 - Exposure 1Diffusion I2 minutes
Plasma I2 minutes
Injection II2 minutes
Phase 2 - Exposure 2Diffusion II2 minutes
Plasma II2 minutes
Injection H2O2 Volume
(59% nominal concentration by weight)
1800
μl ±50 μl
Allow cables to cool down to ambient temperature before starting the monitoring
process. Before each use, visually inspect the sterilized cables for any possible
defects. Do not use defective cables.
64 • Chapter 7 • Cleaning and Sterilizing the System
Sterilizing the Integra® Camino® Flex Extension Cable
The Flex Extension Cable (REF # FLEXEXT) is provided non-sterile.
Caution
A sterile Flex Extension Cable is required during implantation of the
catheter. Prior to sterilization, clean this cable using the guidelines listed
on page 61.
Limits on Sterilization
The Flex Extension Cable may be sterilized thirty (30) times. Do not use this cable
after reaching the sterilization limit. For tracking purposes, use the reprocessing
record sheet (doc # MAN0000005) supplied with your cable.
Sterilization Parameters
The Flex Extension Cable supports steam sterilization only. Prior to sterilization,
the cables should be packaged individually in a self-sealing sterilization pouch
with dimensions 7.5" x 13" or larger.
The Option 2 and Option 3 sterilization cycles are not considered by the United
States Food and Drug Administration (US FDA) to be standard sterilization cycles.
Users should only use sterilizers and accessories (such as sterilization wraps,
sterilization pouches, chemical indicators, biological indicators, and sterilization
containers) that have been cleared by the US FDA for the selected sterilization
cycle specifications (time and temperature).
Following Sterilization
Allow cables to cool down to ambient temperature before starting the monitoring
process. Before each use, visually inspect the sterilized cables for any possible
defects. Do not use defective cables.
Containment and Transportation of Sterilized Cables
It is recommended that the FLEX Extension Cable be sterilized as soon as possible
after use.
Storage of Sterilized Cables
Sterilized cables should be stored at room temperature in sterile pouches.
Chapter 7 • Cleaning and Sterilizing the System • 65
About Single-Use Only Catheters
Integra catheters are intended for single use only. Reuse of the device
can result in contamination and/or disease transmission. This product
should not be resterilized. Resterilization may affect the performance
characteristics and the safety of the device.
Disposal of the Monitor System and Components
Unless stipulated otherwise in the sales contract, the end-of-life
Integra Camino ICP Monitor, cables and AC power adapter must be
returned to Integra for disposal or disposed according to the hospital
recommended procedure. The devices must be clean and free from
blood or other biological contaminants when returned for disposal.
Please contact Integra for further details.
66 • Chapter 7 • Cleaning and Sterilizing the System
CHAPTER 8 TROUBLESHOOTINGTHE SYSTEM
About the Troubleshooting Process............................................................ 67
Responding to System Status Messages.................................................... 67
Responding to Problems During Use.......................................................... 69
Responding to System Failure Messages................................................... 73
About the Troubleshooting Process
This chapter provides guidelines for responding to technical errors, system
failures, and system messages that appear on the touch screen. For instructions
on responding to alarms, see Chapter 5.
Responding to System Status Messages
The following table lists each error message alphabetically that appears on the
touch screen:
System MessageCauseRecommendation
Battery failure
Battery is low
Cannot write to log file
Catheter failure
Battery’s control system has failed.Plug Integra® Camino® ICP Monitor into AC
power outlet (see page 20).
After patient treatment, turn off Integra
Camino ICP Monitor and then turn back on.
Replace the battery (see page 86).
If problem persists, contact Integra.
Battery has 15 minutes or less of
charge.
Battery has 5 minutes or less of
charge (accompanied by alarm
tone).
Cannot write data to log file that is
used by Integra for servicing.
Sudden pressure change that may
occur during implantation or
removal of Flex Catheter.
Plug Integra Camino ICP Monitor into AC
power outlet (see page 20).
Plug Integra Camino ICP Monitor into AC
power outlet (see page 20).
Not resolvable. Contact Integra.
Disconnect and re-connect Flex Catheter
after completing the implantation or
removal process (see page 29).
Chapter 8 • Troubleshooting the System • 67
System MessageCauseRecommendation
Catheter breaks during use.Disconnect catheter and replace with new
one (see directions for use supplied with
catheter).
Catheter fails during initialization.Disconnect catheter and replace with new
one (see directions for use supplied with
catheter).
Catheter is initializing
Cooling fan failure
Data export failed. Could
not detect or write to a
USB drive.
ICP above alarm limit
ICP is out of accuracy
range
Integra Camino ICP Monitor is in
process of initializing new catheter
connection.
Cooling fan inside Integra Camino
ICP Monitor has stopped running.
Problem with connection or
compatibility of USB drive.
Patient’s Mean ICP value has
exceeded user-specified high ICP
alarm limit for more than 5 seconds.
Mean ICP value has exceeded the
range that the Integra Camino ICP
Monitor can measure accurately
(between -10 mmHg to
+125 mmHg).
Wait until Integra Camino ICP Monitor
completes catheter initialization process.
Turn off Integra Camino ICP Monitor to
prevent overheating and contact Integra (see
page 53).
See troubleshooting details on page 72: No
transfer of trend data via USB connection.
Respond to patient’s need immediately.
Once the patient’s Mean ICP value falls
within the alarm limit, the alarm will turn off
(see page 51).
If message appears during implantation or
removal of the Flex Catheter, disconnect and
re-connect the Flex Catheter to the Flex
Extension Cable (see page 29).
Monitor overheating,
check vent
More than one catheter
connected.
Temperature inside Integra Camino
ICP Monitor has exceeded 80 °C.
Two ICP catheters types (Fiber
Optic and Flex) are connected
simultaneously.
No catheter is connected
No ICP catheter is connected to
Integra Camino ICP Monitor.
68 • Chapter 8 • Troubleshooting the System
If message appears during patient
treatment, do not rely on ICP measurements
until the Mean ICP value falls within the
accuracy range.
Check air vent on rear panel of Integra
Camino ICP Monitor. If anything is blocking
vent, remove it (see page 53).
Remove the ICP catheter that will not be
used during patient treatment.
Connect ICP catheter (see page 24 for
connecting fiber optic catheters; see page 27
for connecting Integra Camino Flex
Catheters).
System MessageCauseRecommendation
Temperature is out of
accuracy range
Temperature value has fallen under
range that the Integra Camino ICP
Monitor can measure accurately
Do not rely on temperature measurements
until the temperature value falls within the
accuracy range (see page 54).
(between 30 °C to 42 °C).
Temperature sensor
failure
Temperature catheter broken.Disconnect temperature catheter and
replace with new one (see directions for use
supplied with catheter).
Responding to Problems During Use
The following table provides guidelines for responding to technical problems that
may occur with the Integra Camino ICP Monitor during use.
ProblemCauseRecommendation
Integra Camino ICP
Monitor does not operate
with AC power.
AC power adapter not connectedAttach AC power adapter to Integra Camino
ICP Monitor (see page 20).
Faulty AC power adapter cordAttach AC adapter cord to Integra Camino ICP
Monitor.
• Perform AC power test to ensure AC power
works (see page 85).
Integra Camino ICP
Monitor does not operate
on battery power.
• Contact Integra to order new AC power
adapter cord.
Battery not installed.• Attach AC power adapter to Integra Camino
ICP Monitor (see page 13).
• On touch screen, check battery indicator on
status bar to ensure the battery is installed
(see page 36).
• Install battery (see page 86).
Battery not charged• Attach AC power adapter to Integra Camino
ICP Monitor (see page 13).
• On touch screen, check battery indicator on
status bar to ensure the battery is installed
(see page 36).
Faulty battery• Attach AC adapter cord to Integra Camino
ICP Monitor.
• Perform battery charge test to ensure
battery holds charge properly (see page 86).
• Contact Integra to order new battery
Chapter 8 • Troubleshooting the System • 69
ProblemCauseRecommendation
Touch screen freezesSystem hardware error• Press and hold the power button for several
seconds to perform a forced shutdown of
the system.
• Press the power button to turn on the
system.
Integra Camino ICP
Monitor does not power on
or off when power button
is pressed.
The Mean
ICP value on the
touch screen displays two
dashes (
- -).
Firmware error• Disconnect AC adapter cord from Integra
Camino ICP Monitor and remove battery.
•Re-insert battery.
• Reconnect AC adapter cord to Integra
Camino ICP Monitor.
• Press the power button.
Poor connection between Fiber
Optic Catheter Cable and Integra
Make sure the cable is properly connected to
the Integra Camino ICP Monitor.
Camino ICP Monitor.
Poor connection between Fiber
Optic Catheter Cable or Flex
Extension Cable and the catheter.
Replace the catheter (see directions for use
supplied with catheter).
Faulty catheterReplace the catheter (see directions for use
supplied with catheter).
Damaged Fiber Optic Catheter
Cable
Replace Fiber Optic Catheter Cable (see page
24) .
Damaged Integra Camino Flex
Catheter Cable.
The temperature value on
Faulty catheterReplace the catheter (see directions for use
the touch screen displays
two dashes (
- -).
ICP only catheter in useReplace the catheter with ICP/temperature
Damaged Fiber Optic Catheter
Cable
ICP waveform appears too
ICP waveform not scaled correctly.On touch screen, press
small or too large.
70 • Chapter 8 • Troubleshooting the System
Replace Integra Camino Flex Catheter Cable
(see page 27).
supplied with catheter).
catheter (see directions for use supplied with
catheter).
Replace Fiber Optic Catheter Cable (see page
24) .
Main tab and use the
Scale button to adjust scaling factor (see
page 38).
ProblemCauseRecommendation
ICP waveform appears as a
solid wave instead of a line
wave.
ICP waveform appears as a
line wave instead of solid
wave.
Time/date appears
incorrectly.
Incorrect language is
displayed.
Discrepancy of
ICP
measurements between
the Integra Camino ICP
Monitor and the patient
bedside monitor.
Waveform Style is set to
Waveform Style is set to
mode.
Fill mode.On touch screen, press Settings tab and
Set Waveform->Line (see page 45).
select
Line
On touch screen, press Settings tab and
Set Waveform->Fill (see page 45).
select
Time and/or date is set incorrectly.On touch screen, press
Set Time and Date to specify current
select
time and date (see page 45).
Choice of language is set
incorrectly.
On touch screen, press
Set Language to specify language (see
select
page 46).
Patient bedside monitor is not
synchronized to the Integra
Camino ICP Monitor correctly.
Incorrect patient bedside monitor
adapter cable being used.
On the Integra Camino ICP Monitor, press
Synchronize to Monitor to re-calibrate the
two monitors (see page 30).
Verify correct patient bedside monitor cable is
being used (see page 30).
Settings tab and
Settings tab and
Discrepancy of
temperature
measurements between
the Integra Camino ICP
Monitor and the patient
bedside monitor.
PMIO cable or patient bedside
monitor adapter cable is faulty.
PMIO cable or patient bedside
monitor adapter cable is loose.
Patient bedside monitor has
malfunctioned.
Incorrect patient bedside monitor
adapter cable being used.
PMIO cable or patient bedside
monitor adapter cable is faulty.
PMIO cable or patient bedside
monitor adapter cable is loose.
Replace PMIO cable or patient bedside
monitor adapter cable (see page 30).
Check cable connections between Integra
Camino ICP Monitor and patient bedside
monitor (see page 30).
Refer to manufacturer’s troubleshooting
guide for the patient bedside monitor.
Verify correct patient bedside monitor cable is
being used (see page 30).
Replace PMIO cable or patient bedside
monitor adapter cable (see page 30).
Check cable connections between Integra
Camino ICP Monitor and patient bedside
monitor (see page 30).
Chapter 8 • Troubleshooting the System • 71
ProblemCauseRecommendation
No transfer of trend data
via digital streaming with
USB-to-RS232 adapter
cable.
No transfer of trend data
via USB connection.
Patient bedside monitor has
malfunctioned.
Loose USB-to-RS232 adapter
cable.
Refer to manufacturer’s troubleshooting
guide for the patient bedside monitor.
• Reconnect USB-to-RS232 adapter cable
(see page 59).
• On touch screen, press
Data Streaming Enabled message appears
Trend tab and verify
(see page 59).
Faulty USB-to-RS232 adapter
cable.
• Replace USB-to-RS232 adapter cable (see
page 59).
• On touch screen, press
Data Streaming Enabled message appears
Trend tab and verify
(see page 59).
External monitoring device is not
configured correctly for digital
See digital streaming requirements on page
59.
streaming.
No USB drive is connected.• Connect USB drive to rear panel (see page
57).
• On touch screen, press
Extract Data button. Verify confirmation
Trend tab and select
message appears that indicates successful
USB transfer (see page 57).
USB drive not seated properly in
monitor.
• Check connection of USB drive on rear
panel (see page 57).
• On touch screen, press
Extract Data button. Verify confirmation
Trend tab and select
message appears that indicates successful
USB transfer (see page 57).
USB drive has been removed
before completion of data
transfer.
• Re-connect USB drive to rear panel (see
page 57).
• On touch screen, press
Extract Data button. Verify confirmation
Trend tab and select
message appears that indicates successful
USB transfer (see page 57).
USB drive is corrupted.• Connect new USB drive to Integra Camino
ICP Monitor (see page 57).
• On touch screen, press
Extract Data button. Verify confirmation
Trend tab and select
message appears that indicates successful
USB transfer (see page 57).
72 • Chapter 8 • Troubleshooting the System
ProblemCauseRecommendation
Brand of USB drive is not
recognized by Integra Camino ICP
Monitor.
Insert different brand of USB drive into rear
panel of Integra Camino ICP Monitor (see
page 57).
Responding to System Failure Messages
If a system failure occurs, the Integra Camino ICP Monitor will display a message
window that identifies an error code and description of the problem. The
following table lists each error code associated with a system failure:
Testing High ICP Alarm Limit ....................................................................... 84
Testing AC Power and Battery Charge........................................................ 85
Inserting A New Battery ................................................................................ 86
Testing Synchronizing to Patient Bedside Monitor.................................... 87
About These Procedures
Notice
The procedures in this chapter are intended for the hospital’s biomedical
engineering staff.
Perform the following operational checks to verify that the Integra® Camino® ICP
Monitor is operating safely and effectively. Some checks may be helpful to
diagnose problems that may occur with the system. None of these checks require
disassembling the monitor.
Notice
The operational checks may be performed in any order.
Use Fiber Optic Catheters for Each Test
Each of the following tests requires the use of the:
• Pressure simulator, between 0 and 125 mmHg (e.g.
Delta-Cal™ 650-950)
• Tubing and manifold assembly
Notice
Before performing this test, verify that the ICP Monitor is turned off and
not connected to any cables.
1. Plug the Integra Camino ICP Monitor to an AC wall outlet.
2. On the right side of the monitor, attach the Fiber Optic Catheter Cable and a
functional Fiber Optic Catheter to the
3. Turn on the monitor. On the touch screen, press the Main tab and verify the
ICP (CAMCABL) port.
ICP value reads 0. If not, use the black zero adjustment tool from the Fiber
Optic Catheter kit to set the ICP value to 0 mmHg.
4. Insert the Fiber Optic Catheter into the manifold.
5. Connect the other end of the manifold to the pressure simulator.
6. If necessary, adjust the pressure simulator to 0 mmHg.
7. Verify that the ICP values on both the pressure simulator and the Integra
monitor’s touch screen read 0 mmHg.
78 • Chapter 9 • Testing and Preventive Maintenance
8. Adjust the pressure simulator control knob to the following pressures and
Testing Pressure Output
Note the following test guidelines:
verify the Integra ICP monitor displays the ICP value within the stated limits. It
will also be necessary to take into account the settings accuracy of the pressure
simulator:
Applied Pressure (mmHg)ICP Monitor (mmHg)
0-2 to 2
2018 to 22
4038 to 42
6056 to 64
10094 to 106
125117 to 133
GuidelinesRequirements
Maintenance Interval
Equipment
Once every three months is suggested.
• Digital Multimeter (DMM), 1 mV (resolution)
• PMIO cable (part #PMIO-MPM)
• DC bench power supply (10 V DC)
Notice
Before performing this test, verify that the Integra Camino ICP Monitor is
turned off and not connected to any cables.
1. Plug the Integra Camino ICP Monitor to an AC wall outlet.
2. On the rear of the monitor, connect a PMIO cable to the PMIO connector, but
not to an external patient
3. Using the DC bench power supply and the Digital Multimeter (DMM), do the
bedside monitor.
following:
A. Set DC power supply to 10 ± 0.1 volts and connect “+” and “-” leads to pins “1”
4” of the six pin cannon connector, respectively.
and“
B. Connect the “+” and “-” leads of the DMM to pins “2” and “3” of the six pin
cannon connector, respectively.
Chapter 9 • Testing and Preventive Maintenance • 79
4. Turn on the monitor. On the touch screen, press Synchronize to Monitor to
5. Turn off all test equipment, power down monitor, and then disconnect
Testing Temperature Input
Note the following test guidelines:
GuidelinesRequirements
test the following ICP values.
Simulated Pressure (mmHg)Simulated Voltage (mVDC)
00.000 ± .25
251.250 ± .25
502.500 ± .25
1005.000 ± .25
Note that pressing Check Additional Values each time will advance to the
next fixed ICP value:
Chapter 9 • Testing and Preventive Maintenance • 81
Notice
Before performing this test, verify that the Integra Camino ICP Monitor is
turned off and not connected to any cables.
1. Plug the Integra Camino ICP Monitor to an AC wall outlet.
2. On the right side of the monitor, connect a Fiber Optic Catheter Cable to the
ICP and Temperature connections of the Camino.
3. Using a functional Fiber Optic Catheter, firmly connect the pressure transducer
connector to the Fiber Optic Catheter Cable.
4. On the rear of the monitor, connect a PMIO cable to the PMIO connector, but
not to an external patient
5. Connect the Digital Multimeter (DMM) leads to a ¼” jack connector:
bedside monitor.
•Set red to tip
•Set black to barrel.
•Set DMM to ohms.
6. Fill a water bath or beaker with warm water at approximately 32 °C and do the
following:
A. Place the Fiber Optic Catheter in the water.
B. If necessary, place thermometer in water.
C. On the touch screen, press the Main tab and verify the Temperature value
matches the thermometer reading to ± .3 of a degree C.
7. Using the table that appears after step 10, verify that the measured resistance
is between the low and high limits for the displayed temperature value.
8. Fill a water bath or beaker with warm water at approximately 38 °C and do the
following:
A. Place the Fiber Optic Catheter in the water.
B. On the touch screen, press the Main tab and verify the Temperature value
match the thermometer reading of ± .3 of a degree C.
9. Using the table that appears after step 10, verify that the measured resistance
is between the low and high limits for the displayed temperature value.
10. Disconnect DMM, and turn off monitor.
82 • Chapter 9 • Testing and Preventive Maintenance
Output Temperature Range: 37 °C to 39 °C
Temperature (°C) Low High
3713371367
37.113351361
37.213301359
37.313251354
37.413201349
37.513101344
37.613051334
37.713001329
37.812951324
37.912901319
3812851314
38.112801309
38.212751304
38.312701299
38.412661294
38.512611290
38.612561285
38.712511280
38.812421275
38.912381266
3912331262
Output Temperature Range: 31 °C to 33 °C
Temperature (°C)Low High
3117271751
31.117181751
31.217091742
31.316921742
31.416921733
31. 516841716
31.616761708
31.716761700
31.816681700
31.9166 01692
3216521684
32.116351684
32.216351676
32.316271659
32.416271651
32 .516201651
32 .616121644
32.716041636
32.815971628
32.915821628
3315821621
Temperature vs. Resistance Table
Chapter 9 • Testing and Preventive Maintenance • 83
Testing High ICP Alarm Limit
Note the following test guidelines:
GuidelinesRequirements
Maintenance Interval
Equipment
Once every three months is suggested.
• Integra Camino Fiber Optic Catheter Cable
• Integra Camino Fiber Optic Catheter (either part #
110-4BT or 110-4HMT)
• Pressure source device (either graduated drainage bag;
pressure gauge, between 0 and 125 mmHg (minimum);
or electronic pressure simulator, 0 to 125 mmHg
(minimum)
Notice
Before performing this test, verify that the Integra Camino ICP Monitor is
turned off and not connected to any cables.
1. Plug the Integra Camino ICP Monitor to an AC wall outlet.
2. On the right side of the monitor, connect a Fiber Optic Catheter Cable to the
ICP and Temperature connectors.
3. Using a functional Fiber Optic Catheter, firmly connect the pressure transducer
connector to the Fiber Optic Catheter Cable. Do not connect the temperature
transducer connector if you are using a 110-4BT or 110-4HMT model catheter.
4. Use one of the following pressure sources to provide pressure to the catheter
by using any of the methods described in the
ICP Input Tests section:
•Graduated Drainage Bag (page 76)
•Pressure Simulator (page 79)
5. Turn on the monitor. On the touch screen, press the Alarm tab and set the High
ICP Alarm Limit
6. Set the pressure source to 10 mmHg.
7. On the touch screen, press the Main tab and verify the ICP value displays
to 4 mmHg.
approximately 10 ± 2 mmHg. After about 5 seconds, verify the monitor:
•Sounds an alarm containing a burst of three pulses
•Displays the following error message on status bar: "ICP above alarm limit"
•Flashes the alarm symbol on status bar
•Flashes the current ICP value in yellow on the Main, Trend, and Alarm screens
8. Silence the alarm temporarily. On the touch screen, press the flashing alarm
symbol and verify the monitor:
•Silences the alarm
84 • Chapter 9 • Testing and Preventive Maintenance
•Changes the active alarm symbol to the silenced alarm symbol
•Displays Audio Paused on the status bar.
9. Wait approximately 3 minutes and verify the monitor re-activates the alarms
specified in step
10.Decrease simulated pressure to 0 mmHg, and verify the alarms turn off.
11. On the touch screen, press the Alarm tab and select Alarm Off.
Testing AC Power and Battery Charge
AC Power
Integra recommends performing this test once every three months:
1. Plug the Integra Camino ICP Monitor to an AC power outlet.
2. Press the power button to turn on the monitor.
3. Verify the following:
•On the front of the monitor, check that the LED AC power indicator that
appears above the power button is illuminated.
•On the touch screen, verify the plug icon appears in the status bar.
7.
Low Battery Alarm
4. On the rear of the monitor, disconnect the power cord and verify
the following:
•On the front of the monitor, check that the green LED AC power indicator
turns off.
•On the touch screen, verify the plug icon appears crossed-out in
the status bar
.
If you experience AC power problems, contact Integra.
Integra recommends performing this test once every three months:
1. Run the Integra Camino ICP Monitor on battery power until the following
message appears: "
Battery is low", indicating that the battery has 15 minutes
or less of remaining charge before the monitor shuts off. Note a fully charged
battery will take over an hour before falling to the charge level that displays the
Battery is low message.
2. Keep running the monitor on battery power for another 10 minutes. After 10
minutes, verify that the monitor will:
•Sound an alarm containing a burst of three pulses
•Display the following error message on status bar: "Battery is low"
•Flash the battery symbol in yellow on the status bar
These audible and visual alarms indicate that the battery has approximately 5
minutes or less of remaining charge before the monitor shuts off. If problems with
low battery alarm persist, contact Integra.
Chapter 9 • Testing and Preventive Maintenance • 85
Battery Charge
1. Run the Integra Camino ICP Monitor on battery power until the remaining
battery charge is depleted. Note the when the battery power falls to
approximately 5 minutes or less of remaining charge, the monitor will activate
an audible alarm. After the alarm sounds for approximately 5 minutes, the
monitor will automatically shut down.
2. Plug the monitor into an AC adapter outlet to start re-charging the battery.
3. Press the power button to turn on the monitor. On the touch
screen, verify that the battery power symbol that appears on the
status bar displays the electrical charge symbol.
4. Press the power button again to turn off the monitor. Continue charging the
battery to full capacity; The time for re-charging the battery while the monitor
is turned off will take less than 5 hours.
Notice
If the battery takes more than 5 hours to charge to full capacity, try
replacing the battery with a new one. To order a battery, contact Integra.
5. Once the battery reaches full charge, the battery power symbol will
display four green bars.
6. Unplug the monitor and run it on battery power only and verify that the battery
Inserting A New Battery
The Integra Camino ICP Monitor requires the Integra-supplied 14.4 V lithium ion
battery (REF # BAT1001)to provide battery power to the monitor during transport.
To order a replacement battery, contact Integra.
To i ns ert th e ba tt er y :
1. Verify that the Integra Camino ICP Monitor is turned off and the AC power
2. Turn the monitor upside down so the handle is facing downward.
3. Remove the 2 screws and take off the battery cover.
4. If necessary, remove the old battery by gently pulling the battery from the
5. Insert the new battery:
charge lasts a minimum of 1 hour and 30 minutes. If the battery charge does
not last a minimum of 1 hour and 30 minutes, contact Integra to order a new
battery.
adapter is unplugged.
monitor’s connector slots. Make sure to dispose of the old battery according to
local regulations.
A. Verify the Integra logo on the battery label is facing up.
B. Align the battery’s connectors on the front of the battery to the connector
slots on the monitor.
86 • Chapter 9 • Testing and Preventive Maintenance
C. Slide the battery’s connector into the monitor’s connector slots until the
battery is fully inserted into place.
Notice
The battery for the Integra Camino ICP Monitor can only be inserted in
one orientation.
6. Secure the battery cover to the monitor by re-inserting the two small screws.
Warning
To prevent injury to the patient, user, or other persons, make sure to close
the battery cover securely. Note that the battery cover should remain
closed at all times during the monitor use.
Notice
If the monitor is not used, the battery will lose charge. Always make sure
the battery is charged to full capacity prior to use. The Integra Camino
ICP Monitor will charge the battery while the monitor is plugged into AC
power.
Determining Software Version
On the touch screen, press the Settings tab and select System Information.
Testing Synchronizing to Patient Bedside Monitor
Perform the procedures for synchronizing the Integra Camino ICP Monitor to a
patient bedside monitor on page 30.
Chapter 9 • Testing and Preventive Maintenance • 87
This page is intentionally left blank.
88 • Chapter 9 • Testing and Preventive Maintenance
CHAPTER 10 CONTACTING INTEGRAFOR TECHNICAL
CALIBRATION
VERIFIED.
NEXT SCHEDULED
MAINTENANCE DUE
S
UPPORTAND ANNUAL MAINTENANCE
About Technical Support
If the Integra® Camino® ICP Monitor fails to perform as specified, and the cause
cannot be determined, do not use or attempt to repair it. Instead, contact Integra
for technical service:
For service and repairs outside the United States, contact your local authorized
Integra representative.
Warning
To reduce the risk of electric shock, do not disassemble the Integra
Camino ICP Monitor. Refer all servicing to qualified service personnel at
Integra.
Warning
No modification of the Integra Camino ICP Monitor is allowed.
About Annual Maintenance
The Integra Camino ICP Monitor requires annual maintenance to be performed by
Integra at their service center to ensure proper function and calibration of the
monitor. During this process, Integra will also evaluate the battery’s usage,
number of re-charge cycles, and calibration status and replace it with a new one if
necessary.
•To determine the next due date for annual maintenance, see the label
that appears on the rear panel.
•To schedule the Integra Camino ICP Monitor for annual maintenance,
contact Integra.
CALIBRATION
VERIFIED.
NEXT SCHEDULED
MAINTENANCE DUE
72903759 Rev 1
Chapter 10 • Contacting Integra for Technical Support and Annual Maintenance • 89
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.