IKA C 200 User Manual

CALORIMETER SYSTEM
C 200
Recommended for learning and teaching purpo­ses as well as industrial laboratory use
Space saving, low cost combustion calorimeter to determine calorifi c values of liquid and solid samples
Calorifi c value measurement in accordance with DIN 51900, ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711
Four different test procedures isoperibol, dyna­mic, manual and time-controlled allow ideal ad­aptation to various tasks
Clear, self-explanatory and easy to operate dis­play
Includes external power pack, complies with all global voltages from 100 - 240 V AC, 50/60 Hz; calorimeter is powered with a low operating vol­tage 24 V DC
Highly operator maintenance friendly
Manual operating mode (learn mode): ignition and end of measurement are activated by user; changes are displayed in one minute intervals; all calculations need to be performed manually
Isoperibol, dynamic and time-controlled modes: automatic ignition and calculation of calorifi c va­lue; result is displayed
With Calwin C 5040 software: monitoring and vi­sualisation of measuring procedure (data export and further processing is possible in Excel and / or Word etc.)
The C 200 system includes the following components:
Basic device (measuring cell) C 200 incl. power pack and ignition adaptor
Decompostion vessel C 5010
Oxygen station C 248
SYSTEM PROPERTIES
CALORIMETER SYSTEM
C 200
With the addition of available accessories the system can be adapted to individual laboratory tasks. The system offers the following features:
Clear and self-explanatory menu
Range of measurement: max. 40.000 J (equals a temperature increase of approx. 4 K in the inner vessel)
TECHNICAL DATA
Four different possible test procedures:
Measuring mode Measuring time
isoperibol approx. 17 min
dynamic approx. 8 min
manual (isoperibol) approx. 17 min
time-controlled 14 min
ACCESSORIES
C 5010 IKA®-decomposition vessel, standard
C 5010.4 Attachment for combustible crucible
C 5010.5 Crucible holder, big
C 5010.6 Venting button
C 5030 Venting station
C 5040 CalWin
C 21 Pelleting press
C 29 Pressure gauge, oxygen
C 200.1 Measuring cup 2.000 ml
®
, calorimeter software
External power supply unit:
• Rated voltage/frequency
• Input power max.
Calorimeter:
• Rated voltage
• Input power max.
Fuse 1x 2.5 AT
On-time continuous operation
Protection class according to DIN EN 60529
Protection class III
Ambient temperature 20°C ... 25°C (constant)
Ambient humidity 80 %
Dimensions (WxDxH) 400 x 400 x 400 mm
Weight 21 kg
Range of measurement 40.000 J
Measuring mode isoperibol
Interfaces 1 x serial (RS 232)
100 - 240 V AC / 50/60 Hz 120 W
24V DC 5A 100 W
IP 21
dynamic manual time-controlled
1 x parallel (Centronics)
CONSUMABLES
C 710.4 Cotton thread, cut to length (x500)
C 5010.3 Ignition wire, spare (x5)
C 4 Quartz dish
C 5 Set of VA combustion crucibles (x25)
C 6 Quartz crucible, big
C 710.2 Set of VA combustion crucibles, big (x25)
C 9 Gelatine capsules (x100)
C 10 Acetobutyrate capsules (x100)
C 12 Combustion bags, 40 x 35 mm (x100)
C 12A Combustion bags, 70 x 40 mm (x100)
C 43 Benzoic acid (NBS 39i, 30 g)
C 43A Benzoic acid (100 g)
C 723 Benzoic acid, blister package (x50)
C 14 Combustible crucible (x100)
C 15 Paraffin strips (x600)
IKA®-WERKE GMBH & CO.KG
Janke & Kunkel-Str. 10, D-79219 Staufen Tel: +49 7633 831-0 Fax: +49 7633 831-98 E-Mail: sales@ika.de www.ika.net
We recommend the IKA® analytical mill A 11 basic for sample homogenizing.
Subject to technical change. Informa-
tion not binding for delivery. Our General Terms
and Conditions apply.
Loading...